Lucy Phantom MR Grid Evaluation

Similar documents
Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

S. Guru Prasad, Ph.D., DABR

Use of MRI in Radiotherapy: Technical Consideration

Computational Algorithm for Estimating the Gradient Isocenter of an MRI Scanner

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM

Page 1 of 9. Protocol: adult_other_adni3basichumanprotocol25x_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION

COBRE Scan Information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology

Dosimetric Analysis Report

MRI Physics II: Gradients, Imaging

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

Functional MRI. Jerry Allison, Ph. D. Medical College of Georgia

Fmri Spatial Processing

Role of Parallel Imaging in High Field Functional MRI

Comparison of Two Phantoms for End-to-End SRS/SBRT Testing. Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center

8/3/2016. MR-guided RT: Commissioning and Quality Control. Topics for MR-guided RT system commissioning & QC

XI Signal-to-Noise (SNR)

A Virtual MR Scanner for Education

MapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays

SIEMENS MAGNETOM Avanto syngo MR B15

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

Evaluation of AutoQA Lite TM Image Quality Measurement Software

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

Novel methods and tools for MR Commissioning and Quality Control

Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction

ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T

Following on from the two previous chapters, which considered the model of the

Protocol. Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070

A Study of Motion Tracking Accuracy of Robotic Radiosurgery Using a Novel CCD Camera Based End-to-end Test System

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing

3-D Compounding of B-Scan Ultrasound Images

SIEMENS MAGNETOM Verio syngo MR B17

Breast MRI Accreditation Program Clinical Image Quality Guide

MR Advance Techniques. Vascular Imaging. Class III

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

SIEMENS MAGNETOM Skyra syngo MR D13

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging

Medical Image Processing: Image Reconstruction and 3D Renderings

MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays

Basic fmri Design and Analysis. Preprocessing

An approach for measuring the spatial orientations of a computed-tomography simulation system

Optimisation of Toshiba Aquilion ONE Volume Imaging

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc.

GafChromic Protocol Multi-Channel Film Dosimetry + Gamma Map Analysis

VALIDATION OF DIR. Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology

Patient Set-ups and Tumor Localizations

PET-CT in Radiation Treatment Planning

Micro-CT Methodology Hasan Alsaid, PhD

Geometric distortion evaluation using a multi-orientated water-phantom at 0.2 T MRI

A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

Sources of Distortion in Functional MRI Data

Virtual Phantoms for IGRT QA

Integrating spatially resolved 3D MALDI imaging mass spectrometry with in vivo MRI

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

SIEMENS MAGNETOM Avanto syngo MR B15

New Technology Allows Multiple Image Contrasts in a Single Scan

Imaging protocols for navigated procedures

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health

Functional MRI in Clinical Research and Practice Preprocessing

M R I Physics Course

A complete distortion correction for MR images: I. Gradient warp correction

Lecture 6 Geometric Transformations and Image Registration. Lin ZHANG, PhD School of Software Engineering Tongji University Spring 2013

Version 5.6. Quick Start Guide

UNIVERSITY OF SOUTHAMPTON

8/3/2017. Contour Assessment for Quality Assurance and Data Mining. Objective. Outline. Tom Purdie, PhD, MCCPM

Investigation of a 3D system distortion correction method for MR images

A NEW PHANTOM AND GRADIENT ISOCENTER ESTIMATION FOR MRI DISTORTION CORRECTION. A Thesis. Presented to the. Faculty of. California State University,

Supplementary Information

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE

Help Guide. mm Copyright Mirada Medical Ltd, Mirada Medical RTx 1

Toward Magnetic Resonance Only Treatment Planning: Distortion Mitigation And Image- Guided Radiation Therapy Validation

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images

MR-Guided Mixed Reality for Breast Conserving Surgical Planning

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

A Study of Medical Image Analysis System

TG 132: Use of Image Registration and Fusion in RT

Applications Guide for Interleaved

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

MRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison

Super-resolution Reconstruction of Fetal Brain MRI

DICOM Correction Item

Fast Imaging Trajectories: Non-Cartesian Sampling (1)

RT_Image v0.2β User s Guide

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing

Thank-You Members of TG147 TG 147: QA for nonradiographic

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

RADIOMICS: potential role in the clinics and challenges

Transcription:

Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is designed to assess spatial distortion in MR images. Spatial distortions appear as geometric warping of scanned objects in MR images and may be caused by either the scanner (for example, magnetic field in-homogeneities, susceptibility effects, gradient field nonlinearities, signal to noise ratio, SNR, etc.) or the patient (chemical shift, blood-flow). Spatial distortions introduce contouring errors and therefore misrepresentation of target volume and organs at risk (OARs) resulting in erroneous treatment plans. It is important to evaluate the magnitude of spatial distortions in MR image scans. A B C D Figure 1. Lucy phantom and accessories used in MR distortion correction grid evaluation and testing. (A) Lucy 3D QA phantom; (B) MR grid insert; (C) CT marker cylinders; and (D) MRI marker cylinders. 1 P a g e

II. Methods and Materials: The QA phantom and accessories required to perform distortion correction testing are: Lucy 3D QA phantom, MR distortion grid and CT/MR marker cylinders (figure 1). Figure 2. MR distortion grid insert in planar (left) and 3D (right) views. Lucy phantom (Figure 1A) is a spherical (140mm diameter) Lucite phantom designed to perform Endto-End QA (or process QA) for stereotactic radiosurgery treatments. In addition, the phantom can also be used to evaluate performance of individual components of the SRS program: imaging, planning and treatment. Several inserts are available with the Lucy phantom for this purpose. The MR distortion grid (Figure 1B) consists of 60 x 60 x 30 mm array of orthogonal wires (in 10 x 10 mm grid spacing) embedded in an acrylic box (85 x 85 x 45 mm outer dimensions). The acrylic box is filled with MR contrast enhancing manganese chloride solution (0.1mM MnCl 2 ). The 2mm thick wires are composed of glass reinforced polyamide. As per design specifications, the grid is accurate to within 0.1 mm in the XY directions and 0.2 mm in the Z direction. A set of four marker cylinders for CT and MRI (Figure 1C, 1D) are provided with the Lucy phantom to evaluate positional and fusion accuracy. Each CT marker cylinder contains five 2mm diameter aluminum spheres spaced 5mm center to center. The MR marker cylinders are identical to CT markers except that the aluminum spheres are replaced by mineral oil. 2 P a g e

A. CT scan: First, the phantom was first CT scanned to obtain grid baseline dimensions that could be used as template against which to compare MR images. This also ensured that the phantom had been constructed to the relevant specifications. The MR grid was placed in the center of the Lucy phantom with CT markers in place. The phantom was positioned on the CT scanner couch and aligned with lasers (or physical axes of the scanner). CT scans of the grid were acquired on a multi-slice CT scanner (Phillips Brilliance Big Bore 16-slice). Scan parameters were: axial (non-helix), slice width = 0.75mm, 120 kv, 455 mas, image matrix 512x512, FOV = 350 mm. CT images were imported in a treatment planning system (TPS) and grid separation measured with the measuring tool. Grid spacing was measured along the long dimensions and diagonal in both central and peripheral parts of the phantom. Grid spacing was verified to be within +/- 0.1 mm. In addition, CT marker spacing was also checked. Figure 3. CT scan of grid insert imported into the treatment planning system. Shown are (clockwise from top left): 3D, axial, coronal and sagittal views. 3 P a g e

Figure 4. Axial CT scan of grid insert. Grid dimensions were measured along the long dimensions (60 mm) and diagonal (shown as dashed lines) in the central and peripheral sections of the grid. B. MR scans: With the MR marker cylinders in place, the phantom was aligned with the MR scanner lasers (scanner axes) and images acquired as per our SRS scan protocol (see table 2) used for all patient studies. T1 and T2 MR Image sets were obtained in axial and coronal planes on a 1.5 T MR scanner using 1mm and 2mm slice separation. All images were imported into our SRS treatment planning system. For each image modality, appropriate window/level setting was selected to achieve the best image contrast. The grid wires appear as dark pixels and the fiducial markers as bright spots on CT and MR image sets. Center of each marker was identified and the measurement tool available in the TPS was used to check grid separation in the central and peripheral parts of the grid image (axial plane). Measurements were performed in central and peripheral axial planes. MR T1 coronal image set was reconstructed in to axial images prior to measurement. In addition, the separation between fiducial markers in the marker cylinders was measured. Each measurement was repeated three times to minimize random errors. To obtain distortion data, a one-to-one correspondence was established between grid point locations in MR and CT data sets. 4 P a g e

III. Results: MR image scan results are shown in Figures 5 and 6. Window level settings for optimum image contrast are shown in Table 1. Overall MR Axial T1 images with 1mm slice separation provide the best image quality. In general, T1 images are superior compared to other MR images and 1mm slice separation is better than 2mm. When grid dimensions were compared to the baseline values, MR Axial T1 image (1mm) set provided the best agreement with the smallest variance (standard deviation). Marker spacing in T1 Axial image set also agreed better with the baseline CT data. As shown in table 1, a comparison of peripheral vs. central grid spacing can provide an estimate of distortion in MR images. In general, MR image grid measurements were slightly less than those measured in CT image set. A slight bowing-in was noted in the central part of all MR images as evidenced by the MR central grid separation measurements being slightly smaller than those in the peripheral regions. Our results indicate a distortion correction of 0.3% 0.6% over the grid dimensions for MR Axial T1 images acquired at 1mm slice separation. In contrast, MR T2 images at 2mm slice separation resulted in relatively larger distortion correction, with other sequences showing intermediate results. Similarly, caution should be exercised when Coronal T1 images that have been reconstructed into axial plane are used in drawing targets. These reconstructed images tend to show greater variation caused by spatial distortion than their native axial images counterparts. IV. Summary: An estimate of MR distortion corrections is necessary to help improve target delineation in SRS planning. These results should also be helpful in estimating dose calculation errors when MRI images are solely used for treatment planning. The MR distortion grid when used with the Lucy phantom can provide distortion correction data for MR image sets. Furthermore, low distortion MR sequences may be investigated to minimize potential errors in radiotherapy planning. 5 P a g e

Figure 5. MRI scan (Axial T1, 1mm spacing) of distortion grid. Shown are clockwise from top left: 3D, axial, coronal and sagittal views. 6 P a g e

T1 Axial (1mm) T1 Coronal (1mm) T1 Axial (2mm) T2 Axial (2mm) Figure 6. MR scans of grid insert. Table 1. Window/Level settings, grid-dimensions and marker spacing for scan types used in this study. Scan Type Window/Level Grid Dimension (mm) Long Diagonal Peripheral Central Peripheral Central Markers Spacing CT -252-167 60 60 56.57 56.57 20 MRI Ax T1 (1mm) MRI Cor T1 (1mm) MRI Ax T1 (2mm) MRI Ax T2 (2mm) 0-425 0-700 0-270 0-383 Ave 59.80 59.62 56.55 56.20 19.97 S.D. 0.17 0.10 0.07 0.00 0.06 Ave 59.82 59.85 56.45 56.50 19.87 S.D. 0.11 0.14 0.21 0.00 0.15 Ave 59.80 59.75 56.35 56.30 20.10 S.D. 0.15 0.08 0.07 0.00 0.17 Ave 59.70 59.67 56.40 56.30 19.87 S.D. 0.15 0.12 0.14 0 0.15 7 P a g e

Table 2. MR Scan parameters used in this study. Scan Type Parameters Localizer Sagittal T1 Axial T2 Axial T1 Coronal T1 Contrast Pre Pre Pre Post Post TE 9 102 9 9 TR 600 2884 600 600 FOV 28 28 28 28 22.4 Slice thickness 5 3 2 1 or 2 1 or 2 Slice Separation 0 0 0 0 0 Image Parameters Plane 3 plane Sagittal Axial Axial Coronal Mode 2D 2D 2D 2D 2D Pulse Sequence Gradient Echo Spin Echo FSE-XL Spin Echo Spin Echo Image Option Seq, Fast none FC Fast none none Frequency 256 256 256 256 Phase 256 256 256 256 NEX 1 1 1 1 Phase FOV 1 0.75 0.75 1 Shim Auto Auto Auto Auto 8 P a g e