Monte-Carlo-Based Scatter Correction for Quantitative SPECT Reconstruction

Similar documents
Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Advanced Scatter Correction for Quantitative Cardiac SPECT

Concepts, Applications, and Requirements for Quantitative SPECT/CT. Conflict of Interest Disclosure

Slab-by-Slab Blurring Model for Geometric Point Response Correction and Attenuation Correction Using Iterative Reconstruction Algorithms

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

IN single photo emission computed tomography (SPECT)

Iterative SPECT reconstruction with 3D detector response

3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction

Research Article Optimisation of Simultaneous Tl-201/Tc-99m Dual Isotope Reconstruction with Monte-Carlo-Based Scatter Correction

SINGLE-PHOTON emission computed tomography

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques

GPU implementation for rapid iterative image reconstruction algorithm

Improved activity estimation with MC-JOSEM versus TEW-JOSEM in 111 In SPECT

Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Constructing System Matrices for SPECT Simulations and Reconstructions

Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy.

Modeling and Incorporation of System Response Functions in 3D Whole Body PET

Introduction to Positron Emission Tomography

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

S rect distortions in single photon emission computed

SPECT images are degraded by photon attenuation, by

ATTENUATION CORRECTION IN SPECT DURING IMAGE RECONSTRUCTION USING INVERSE MONTE CARLO METHOD A SIMULATION STUDY *

A Weighted Least Squares PET Image Reconstruction Method Using Iterative Coordinate Descent Algorithms

Reconstruction from Projections

USING cone-beam geometry with pinhole collimation,

SPECT (single photon emission computed tomography)

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

Q.Clear. Steve Ross, Ph.D.

Planar and SPECT Monte Carlo acceleration using a variance reduction technique in I 131 imaging

SPECT reconstruction

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA

The Emory Reconstruction Toolbox Version 1.0

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

Simulations in emission tomography using GATE

Quantitative I-131 SPECT Reconstruction using CT Side Information from Hybrid Imaging

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam

Organ Dose estimates for Radio-Isotope Therapy treatment planning purposes

Fits you like no other

A dedicated tool for PET scanner simulations using FLUKA

Fits you like no other

Spatiotemporal processing of gated cardiac SPECT images using deformable mesh modeling

A publicly accessible Monte Carlo database for validation purposes in emission tomography

Scatter correction improvement based on the Convolution Subtraction Technique in SPECT imaging

Preliminary study. Aim

Continuation Format Page

Development and Evaluation of Image Reconstruction Algorithms for a Novel Desktop SPECT System

A Fast GPU-Based Approach to Branchless Distance-Driven Projection and Back-Projection in Cone Beam CT

DUAL energy X-ray radiography [1] can be used to separate

Single and Multipinhole Collimator Design Evaluation Method for Small Animal SPECT

Resolution and Noise Properties of MAP Reconstruction for Fully 3-D PET

Monte-Carlo simulation for scatter correction compensation studies in SPECT imaging using GATE software package

Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation

Accelerated Monte Carlo simulation for scatter correction in SPECT. Hugo de Jong

In SPECT clinical practice, filtered backprojection is still

SPECT: Physics Principles and Equipment Design

Image-based Monte Carlo calculations for dosimetry

Symbia E and S System Specifications Answers for life.

Introduc)on to PET Image Reconstruc)on. Tomographic Imaging. Projec)on Imaging. Types of imaging systems

CORRECTION OF THE RESPIRATORY MOTION OF THE HEART FROM 4D MYOCARDIAL IMAGES

James R Halama, PhD Loyola University Medical Center

Xi = where r is the residual error. In order to find the optimal basis vectors {vj}, the residual norms must be minimized, that is

Improvement of contrast using reconstruction of 3D Image by PET /CT combination system

Few-view single photon emission computed tomography (SPECT) reconstruction based on a blurred piecewise constant object model

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

A new approach to scatter correction in SPECT images based on Klein_Nishina equation

*Health Physics Department, **Nuclear Medicine Department, Azienda Ospedaliera Maggiore della Carità, Novara, Italy

Image reconstruction for PET/CT scanners: past achievements and future challenges

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

VALIDATION OF GATE MONTE CARLO SIMULATIONS FOR INDIUM 111 IMAGING

NIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2010 February 9.

ACTIVITY quantification with SPECT has three main

Abstract I. INTRODUCTION

ANALYSIS OF CT AND PET/SPECT IMAGES FOR DOSIMETRY CALCULATION

Fast Timing and TOF in PET Medical Imaging

Automatic Lesion Detection for Measuring Response using Dynamic FDG-PET

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Rapid Emission Tomography Reconstruction

PET vs. SPECT: in the Context of Ongoing Developments (Review Article) ABSTRACT

Image Acquisition Systems

A Projection Access Scheme for Iterative Reconstruction Based on the Golden Section

Simultaneous measurement of noise and spatial resolution in PET phantom images

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide

Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging

Quantitative imaging for clinical dosimetry

Mathematical methods and simulations tools useful in medical radiation physics

A slice-by-slice blurring model and kernel evaluation using the Klein-Nishina formula for 3D

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 1, FEBRUARY

A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET

Super-resolution Reconstruction of Fetal Brain MRI

GPU Based Convolution/Superposition Dose Calculation

RADIOLOGY AND DIAGNOSTIC IMAGING

Transcription:

Monte-Carlo-Based Scatter Correction for Quantitative SPECT Reconstruction Realization and Evaluation Rolf Bippus 1, Andreas Goedicke 1, Henrik Botterweck 2 1 Philips Research Laboratories, Aachen 2 Fachhochschule Lübeck, Bildgebende Verfahren in der Medizintechnik rolf.bippus@philips.com Abstract. Quantitative SPECT as well as simultaneous acquisition of multiple isotopes with SPECT in the clinical field, although clinically interesting, are still limited by reconstruction artifacts and computing power. As a considerable step in this direction, we have implemented an efficient reconstructor with variance reduced Monte-Carlo-simulation in the forward and/or backward projection of an OS-EM iteration. Apart from a quantitatively accurate scatter estimation the integrated MC simulation allows us to include all effects that are relevant for the multiple isotope task. The reconstruction problem is explicitly formulated as a combined maximum-likelihood estimation for all unknowns in one step and implemented efficiently on the basis of the OS-EM algorithm using the Effective Scatter Source approach. The algorithm has been evaluated on simulated as well as measured phantom data. Two isotopes (Tc99m, Tl201) were tested on the voxelized NCAT phantom and evaluated quantitatively. In addition we performed phantom measurements to evaluate the method on a Philips CardioMD TM with a Vantage TM line source for attenuation correction. A clear advantage of the proposed approach is its robustness and generalizability. It is currently being evaluated in clinical applications like simultaneous dual isotope cardiac imaging. 1 Introduction Absolute quantification as well as simultaneous acquisition of multiple isotopes are receiving increasing interest in clinical applications (e.g. [1], [2]). However patient scatter, down-scatter plus collimator effects such as penetration and Pbfluorescence result in data contamination. Basically one can distiguish between energy and/or spatial distribution based scatter estimation and correction from reconstruction based methods [3]. The OS-EM (Ordered Subset Expectation Maximization) approach we follow with our implementation falls into the second category. Scatter is estimated based on the current activity estimate in the forward projection step. Using the approximative scheme of effective scatter source (ESS) Estimation, originally published by Frey et al. [4], scatter estimation is effectively separated from other effects in the algorithm. The original approach uses precomputed scatter kernels to

Quantitative SPECT Reconstruction 371 obtain the ESS via convolution. In contrast we apply variance reduced Monte- Carlo simulation to estimate the ESS. An attenuation map is used as a physical model of the patient. Using one of the reconstruction based scatter estimation methods one can differentiate sequential and simultaneous reconstruction as depicted in Fig. 1. We follow the scheme of simultaneous reconstruction(fig. 1(b)) where the cross contamination estimation is integrated into the forward projector and all estimates isotope activities are forward projected into all energy windows and updated in each sub-iteration of the OS-EM algorithm. 2 Methods The full update equations for the ordered subset EM algorithm for all isotope activities (typicall 1 or 2) f (isotope i in voxel ǫ) on all available data are given by equations (1) and (2). The forward projection step computes(1) the contributions of isotopes i to all projection pixels p in each energy-window e, based on the recent activity estimate, where H p,e is the normalized system operator. In the backprojection step (2), the correction factors are obtained from the ratio of the observed and the estimated projection values. ν (k) (p,e) = ν (k) i (p,e) = f (k) H p,e, Hp,e = 1 (1) i i ǫ p,e f (k) f (k+1) = f (k) 1 n p e Hp,e ν(p, e) ν (k) (p,e), n = p,e H p,e (2) 2.1 Approximations / Implementation Using brute force counting statistics with MC-simulation to obtain the ν (k) (p,e) in equation (1) is not feasible. Therefore we applied the principle of convolution based forced detection. Simulating the photon tracks, at each interaction of the photons with the patient the effective cross section is calculated for the photon being redirected towards the detector for a set of intermediary energy intervals. These are accumulated in 3D and are treated as secondary emission distributions, the ESS. These are then projected onto all camera positions and energy windows, taking (c) (d) Fig. 1. Sequential (a) vs. simultaneous (b) dual isotope reconstruction scheme.

372 Bippus, Goedicke & Botterweck into account attenuation and collimator response functions (CRF). The CRFs, simulated beforehand and stored, are taking into account all major physical degradation effects. The principle is illustrated in Fig. 2. According to the Dual-Matrix aproach, the backprojector is approximated by the reversed PSFs and the attenuation (B ǫ p,e) as well as the relative weights ω i (p,e) of the isotope i contributing to pixel (p,e) in the forward projection, thus ignoring the spatial distribution of the scatter. f (k+1) = f (k) 1 n p e B ǫ p,eω (k) i (p,e) ν(p,e) ν (k) (p,e) (3) 2.2 Activity Estimation Quantitative results are obtained by defining volumes of interest (VOI) and calculating the mean activity within the VOI. We evaluated quantitative activities after 30 to 50 full iterations of the OS-EM algorithm. Corresponding images are not suited for visual inspection or VOI delineation due to high noise. Thus images for visual inspection or delineation use 4 full iterations of the OS-EM algorithm. For the simulated data we defined the VOIs directly from true activity images. 3 Results 3.1 NCAT Phantom Results The voxelized NCAT phantom was simulated such that the activities represent a patient dose of 25mCi Tc 99m and 3.5mCi of Tl 201. In the Tl 201 distribution a heart lesion was included, which was not present in the Tc 99m image. The lesion activity was chosen 15% of the normal myocardial Tl 201 activity. Fig. 2. Principle of down-scatter estimation in the forward projection step.

Quantitative SPECT Reconstruction 373 For quantitative comparison of true and reconstructed activity, we measured absolute activity in the 17 segments of a standard 17 segment polar plot representation. Heart re-orientation and segmentation were performed on the true activity image and kept constant for the reconstructed image. As before the values represent absolute activity estimates averaged over each segment of the heart. 3.2 Phantom Measurement Results Phantom measurements were performed using a Data Spectrum Cooperation Anthropomorphic Torso Phantom TM with a fillable heart insert. The phantom was filled according to an estimated biodistribution resulting from 90MBq Tl 201 and 650MBq Tc 99m administered to an adult. The heart contained 2 lesions of 5.6ml (lesion 1) and 12ml (lesion 2), each filled with Tc 99m only, which gives no Tl 201 and approx. 65% Tc 99m (due to the walls of the lesions) in the lesions. Resulting images and polar plots are shown in Fig. 4. 4 Conclusion Quantitative MC reconstruction for SPECT and multiple isotopes shows to be feasible. The more exact modeling of scatter within the object and collimator results in reduced image artifacts. The computational effort is reasonable, especially compared to other state of the art scatter corrections like ESSE. An advantage of the proposed approach is its robustness and generalizability: Fast adaptation to new isotope combinations is possible. Next steps will include the evaluation of its potential in clinical application of dual isotope cardiac imaging. (a) truth (b) 30 iterations Fig.3. NCAT, Tc 99m -Tl 201 SDI, Tl 201 mean activity in kbq/ml in each of 17 segments of standard polar plot.

374 Bippus, Goedicke & Botterweck Fig.4. Phantom measurements on the CardioMD. Cardiac SDI filled with Tl 201 and Tc 99m and two lesions filled with Tc 99m only. (a) (b) References 1. Berman D, Kang X, Tamarappoo B, et al. stress thallium-201/rest technetium- 99m sequential dual isotope high-speed myocardial perfusion imaging. JACC Cardiovasc Imaging. 2009;2(3):273 82. 2. Sgouros G. Dosimetry of internal emitters. J Nucl Med. 2005;46(1):273 82. 3. King J M A ad Glick, Pretorius PH, Wells RG, et al. Attenuation, Scatter and Spatial Resolution Compensation in SPECT. In: Wernick M, Aarsvold JS, editors. Emission Tomography. The Fundamentals of PET and SPECT. Acedemic Press; 2004. 4. Frey EC, Tsui BMW. A new method for modelling the spatially variant, objectdependent scatter response function in SPECT. In: Proc IEEE Nucl Sci Symp; 1996. p. 1082 6.