Applying Hounsfield unit density calibration in SkyScan CT-analyser

Similar documents
Enhanced material contrast by dual-energy microct imaging

How does the ROI affect the thresholding?

Adaptive thresholding. in CTAn

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units

How to make color-coded 3D models for structure thickness in CTVox

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

Method note. Page 1 of 7

Fundamentals of CT imaging

Micro-CT Methodology Hasan Alsaid, PhD

Manual for Bruker-microCT CT-Analyser v

DUE to beam polychromacity in CT and the energy dependence

microct Systems for Life Science Innovation with Integrity High Resolution Microtomography for in vivo and ex vivo Applications Microtomography

The user's guide. CT-Analyser. Contents. (version )

LAB DEMONSTRATION COMPUTED TOMOGRAPHY USING DESKCAT Lab Manual: 0

Using X-Ray Micro Tomography to assess the quality of Packaging Foil

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Micro-CT analysis of mouse long bones (proximal tibia, distal femur)

Image and dataset registration in Dataviewer - Example of expanded clay

FINDING THE TRUE EDGE IN CTA

Quality control phantoms and protocol for a tomography system

Ch. 4 Physical Principles of CT

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

Physical bases of X-ray diagnostics

Comparison of Scatter Correction Methods for CBCT. Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis

Corso di laurea in Fisica A.A Fisica Medica 4 TC

ML reconstruction for CT

Computed tomography (Item No.: P )

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Computed tomography of simple objects. Related topics. Principle. Equipment TEP Beam hardening, artefacts, and algorithms

BRUKER MICROCT. NRecon User Manual. Bruker microct July, 2016

CT vs. VolumeScope: image quality and dose comparison

Coordinate Measuring Machines with Computed Tomography

The user's guide. CT-Volume (version 2.0) Contents

A high definition Mueller polarimetric endoscope for tissue characterisation

Experiment 6. Snell s Law. Use Snell s Law to determine the index of refraction of Lucite.

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Modifications for P551 Fall 2014

Surface and volume rendering of large datasets

CALIBRATIONS FOR ANALYZING INDUSTRIAL SAMPLES ON MEDICAL CT SCANNERS

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system

Tomographic Reconstruction

A closer look at CT scanning

BME I5000: Biomedical Imaging

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Digital Image Processing

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

PROVIDE A UNIFORM DOSE IN THE SMALL ANIMAL.

UNCOMPROMISING QUALITY

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

Oxygen XRay Step by Step

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

AnalySIS Tutorial part 2

Moscow-Bavarian Joint Advanced Student School 2006 / Medical Imaging Principles of Computerized Tomographic Imaging and Cone-Beam Reconstruction

Visualizing Dynamics with the Multi Frame FLIM Script

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings)

Standard Operating Procedure of Triboindenter (Hysitron TI 950)

[PDR03] RECOMMENDED CT-SCAN PROTOCOLS

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

Synoptics Limited reserves the right to make changes without notice both to this publication and to the product that it describes.

Note: FLIM-FRET is a robust method to determine the FRET efficiency of a suited donor acceptor pair.

A new Concept for High-Speed atline and inlinect for up to 100% Mass Production Process Control

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Reflection and Refraction of Light

FACULTY AND STAFF COMPUTER FOOTHILL-DE ANZA. Office Graphics

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel

RSNA, /rg

Methods for data preprocessing

Image Acquisition Systems

Simplified statistical image reconstruction algorithm for polyenergetic X-ray CT. y i Poisson I i (E)e R } where, b i

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2)

Biomedical Image Analysis. Mathematical Morphology

Alignment and Other Challenges in Reconstructing Cryotomograms with IMOD

Basics of treatment planning II

Digital Volume Correlation for Materials Characterization

Gamepad Controls. Figure 1: A diagram of an Xbox controller. Figure 2: A screenshot of the BodyViz Controller Panel. BodyViz 3 User Manual 1

S. Guru Prasad, Ph.D., DABR

Effect of CT number stopping power conversion uncertainties on dose distributions

Tools Menu (Frequently Used Features)

Computer-Tomography I: Principles, History, Technology

J. Ling Chinese Academy of Science Institute for Ceramics Shanghai, China

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

A new Concept for High-Speed atline and inlinect for up to 100% Mass Production Process Control

Manual for Microfilament Analyzer

Loma Linda University Medical Center Dept. of Radiation Medicine

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

Application Note AN-1703 Strain Filter Selection

MEDICAL IMAGING 2nd Part Computed Tomography

Computed tomography - outline

Contents. Introduction... iii. CHAPTER 1 - Getting Started... 5

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Cornell Spectrum Imager (CSI) Open Source Spectrum Analysis with ImageJ Tutorial

Refraction Corrected Transmission Ultrasound Computed Tomography for Application in Breast Imaging

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa

Transcription:

1 Bruker-microCT Method note Applying Hounsfield unit density calibration in SkyScan CT-analyser Hounsfield units (HU) are a standard unit of x-ray CT density, in which air and water are ascribed values of 0 and 1000 respectively. It has been found over several decades of CT imaging to be a useful general CT density calibration owing to the approximate linearity of the calibration curve of materials of different x-ray density, such that calibration with water and air data points gives a useful and highly reproducible reference curve even for materials (such as biological calcified tissue) with CT density significantly higher than that of water. It is conventional to displace the HU value downwards by 1000 units, so that the values for water and air are 0 and -1000 respectively. SkyScan CT-analyser allows you to either adopt this convention or not, as you choose. You may have noticed that HU calibration is possible in SkyScan scanners at the point of image reconstruction. (This involves HU calibration factors in the scanner control software.) The HU calibration done by CT-analyser is a second, alternative option, for post-hoc calibration of a dataset after reconstruction, by reference to a scanned phantom (tube of water). It is probably an easier, more flexible and more accurate method, and how to do it is set out in this method note. Bruker microct Page 1 of 9 September 2005

2 Bruker-microCT Method Note: HU calibration BEFORE YOU SCAN THE OBJECT The ability to accurately measure density from a micro-ct scan of an object is not something that can be taken for granted. Certain artefacts of micro-ct imaging can compromise density measurement. An exact correlation between apparent grey level in the reconstructed image and object density cannot be assumed. Two key points should be considered: 1. Beam hardening. This relates to the wide x-ray energy distribution from a micro-ct x-ray source, and selective removal of low energy photons as the x-rays pass through your object. Uncorrected, this phenomenon gives the surface layer an exaggerated high density. If you scan an object to measure density, you should minimise beam hardening by (a) applying an aluminium filter during the scan, and (b) applying the correct beam hardening correction during reconstruction. SkyScan NRecon software includes a profile line tool useful for assessing beam hardening and optimising its correction. Also, keep the applied voltage to the minimum needed to give adequate transmission through your object. 2. Sample hydration. If the sample is wet, such as a piece of biological tissue, it is very important to maintain a level of hydration close to natural during the scan, and prevent drying. Drying in particular will cause problems, both by causing inaccuracy in density measurement and by causing movement artefacts in the scan due to shrinkage. So wet biological samples should be maintained moist (in a physiological 0.9% saline if possible) and enclosed to prevent drying. Drying can be prevented either by wrapping in parafilm or cling wrap plastic (PVC-free if possible), or by scanning the object inside a narrow plastic tube of water or saline. 3. Streak artefacts, 360 scanning. Scans over 360 will also give a slightly more uniform image density than over 180 (removing asymmetric beam hardening). In some 180 degree scan reconstructions, a high density object surrounded by lower density medium can cause dark streak artefacts around the dense object. These can be minimised by scanning over 360 rather than 180. This will improve the accuracy of density measurement. Note however that for 360 scans it is important to check that the postalignment correction has the right value. 4. Apply sufficient smoothing. If an image is too noisy then the density peaks on the attenuation histogram are widely spread out, compromising the accuracy of density measurement. Smoothing should be applied in NRecon to remove excess noise from the image. For density measurement you do not really lose any image data by doing this, you can in fact improve the image quality. Bruker microct Page 2 of 9 September 2005

3 Bruker-microCT Method Note: HU calibration STEPS TO APPLY HU CALIBRATION The steps to calibrating density in a sample scanned by a SkyScan micro- CT instrument are as follows: 1. Scan your sample (noting points 1-3 above for scanning for CT density). Reconstruct the dataset, taking care to minimise beam hardening, and find and apply the optimal beam hardening correction. When setting contrast limits in reconstruction, the lower limit should be zero (this simplifies density calibration). The upper limit should be above the upper end of the sample absorption histogram, so that minimal saturation occurs in the image (i.e. minimise the number of pixels reaching the maximum density 255, black grey level ). Note that you can double-click on the density histogram in NRecon to change the y axis to a logarithmic scale. 2. Make your micro-ct HU phantom. This is a plastic tube filled with water, with a diameter about the same as the sample object(s) you are scanning. The plastic should be low density polypropylene, acrylic, polystyrene or polycarbonate are acceptable (PVC is not). 3. Scan the water phantom, and reconstruct it. All the parameters of the scan (tube settings, rotation step, etc.) and all the reconstruction settings including contrast limits, and beam hardening correction, must be identical between the scanned sample and the scanned water phantom. Thus the reconstructed Bruker microct Page 3 of 9 September 2005

4 Bruker-microCT Method Note: HU calibration water images will look quite pale if your object is more dense that water this is OK. The one exception is that post-alignment correction is unique to each scan, and differences between scans in post-alignment do not compromise comparative measurements of density (or morphometry). This parameter is automatically calculated by NRecon and should be individually optimised for each scan. 4. Open the reconstructed dataset of the water tube in CTAn. Select a circular region of interest (ROI) of water as shown (right). Make this into a volume of interest over a number of layers of water. It is best to use image levels with water only, not levels containing a denser object (e.g. bone) surrounded by water, since the dense object can slightly change the density of neighbouring water. 5. Having selected the water volume of interest (VOI), go to the next, binary page of CTAn. Above the density histogram are two tabs, from image and from dataset. Click on from dataset to integrate the density histogram over all crossections. 6. At the right end of the buttons under Histogram is the button toggle VOI view. Please make sure that this button is pressed. This will make the density measurement correspond to the region/volume of interest. (If it is not pressed, then the output density values will be taken from the whole dataset, without reference to the VOI, resulting in a calibration error.) Note that clicking on the log button above the histogram display, can make the histogram easier to see. 7. Open the HU calibration window by pressing on the button above the density histogram indicated by the red arrow in the image below. The HU calibration box has three columns of paired boxes, for corresponding values of grey level ( index ) and Bruker microct Page 4 of 9 September 2005

5 Bruker-microCT Method Note: HU calibration The density page in CTAn for applying HU calibration. With the VOI of water selected, and the average grey level of water displayed as Mean (total), the HU calibration button can be pressed (red arrow). HU. As indicated in the images below, enter values 0 for grey level and -1000 for HU in the first column. (Remember during NRecon reconstruction you set the contrast minimum to zero, so that zero greyscale corresponds to zero attenuation.) In the second column, enter the measured mean grey value for the water VOI here 104.83, and the corresponding HU value for water, which is 0. Bruker microct Page 5 of 9 September 2005

6 Bruker-microCT Method Note: HU calibration Click on OK to implement the calibration the window will close. Please note that on reopening the HU calibration box, the numbers will have changed as shown in the image (left). However the calibration you entered is unchanged, but the displayed numbers have just been normalised to show the maximum density range, corresponding to the extreme grey level values of 0 and 255. If you make a note of the HU number shown above 1433 which corresponds to the 255 grey level, this number will allow you to implement the same HU calibration on other datasets such as your scanned sample or animal datasets. This number, here 1433, is called the HU calibration number. 8. Apply the HU calibration from the water tube, to your measured sample. Open the sample scan dataset in CTAn, go to the binary page, and click on which will open the density range calibration window, and enter the values exactly as in the second such window shown above (i.e. 255 grey = 1433 HU). Remember to set also the HU value of -1000 to correspond with the grey scale value of 0. 9. Now the sample scan dataset is calibrated to HU. To measure HU within a volume of interest (VOI), select a VOI from your sample the example below shows a mouse lung. Then go to the density page of CTAn and click on the from dataset tab to show the density histogram. Also make sure the toggle VOI view button is pressed. Bruker microct Page 6 of 9 September 2005

7 Bruker-microCT Method Note: HU calibration Then read the density value in calibrated HU for the manually selected lung ROI (black shape around the lung in the image above right) from the summary data. The value as shown below is -529.43 HU. Note that since we want to measure HU in the binarised white lung pixels only, then the HU value is read from the mean under the Selection heading, not the mean (total) which would output HU from the whole black area. Bruker microct Page 7 of 9 September 2005

8 Bruker-microCT Method Note: HU calibration 10. Saving your calibration. To apply the HU calibration that you have obtained from the water phantom to a sample dataset, you should record the HU number corresponding to the maximum grey level 255. In the example given above, 255 grey = 1433 HU. For example you can save it in a text file. 11. Saving a calibrated dataset. You can embed a HU calibration into a dataset in the following way. First apply the HU calibration from the button, as described above. Then go to the ROI page in CTAn, and select a volume of interest. This can either be a part of the dataset or the whole dataset (in the case that you choose no ROI). Then save a new dataset from the ROI using the fourth of the ROI function buttons (see the window below). This new saved dataset will have the HU calibration that you just applied embedded in it (written into the text header of each bmp file). So when you re-open this new dataset you will not need to apply the HU calibration the correct HU calibration will already be set. 12. Measuring HU from custom processing. Once a HU calibration has been made for the open dataset in CTAn, density as HU can also be output from the custom processing. Here for example one can import the greyscale lung image into a binary mask of the lungs (see image below). Bruker microct Page 8 of 9 September 2005

9 Bruker-microCT Method Note: HU calibration With the ROI set as the lung binary mask as shown in the above image, density is measured and output in custom processing using the histogram plugin: The HU density data is output in the report file with the following format: Bruker microct Page 9 of 9 September 2005