Introduction. Quality Assurance for Image- Guided Radiation Therapy. Justification for IGRT. Image-Guided Radiation Therapy

Similar documents
Automated Quality Assurance for Image-Guided Radiation Therapy

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

8/3/2016. Image Guidance Technologies. Introduction. Outline

Optical Guidance. Sanford L. Meeks. July 22, 2010

Thank-You Members of TG147 TG 147: QA for nonradiographic

Quick Reference Datasheet For All RIT113 Packages

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

CBCT: Past, Present and Future. Disclosures. Computed Tomography. ATec Education Course Feb 28-Mar3, Douglas Moseley PhD, DABR

Image Guidance and Beam Level Imaging in Digital Linacs

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

The team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.

Use of TrueBeam developer mode for imaging QA

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Financial disclosure. Onboard imaging modality for IGRT

Evaluation report. X-ray tomographic image guided radiotherapy systems CEP10071

1. Learn to incorporate QA for surface imaging

Digital phantoms for the evaluation of a software used for an automatic analysis of the Winston-Lutz test in image guided radiation therapy

Feasibility study of performing IGRT system daily QA using a commercial QA device

Evaluation of the Elekta Synergy concept for patient positioning in image guided radiotherapy

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package.

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field

Digital Tomosynthesis for Target Localization

Initial Clinical Experience with 3D Surface Image Guidance

Patient Set-ups and Tumor Localizations

URGENT IMPORTANT FIELD SAFETY NOTIFICATION

Measurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images

Protocol. Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070

Image Acquisition Systems

Version 5.6. Quick Start Guide

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC

Mech. Engineering, Comp. Science, and Rad. Oncology Departments. Schools of Engineering and Medicine, Bio-X Program, Stanford University

ICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588

A study on image quality provided by a kilovoltage cone-beam computed tomography

Real Time Tumor Motion Tracking with CyberKnife

Ch. 4 Physical Principles of CT

MR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner

DAILY LINAC QA BEAM QA

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Tumor motion during liver SBRT

Clinical assessment and characterization of a dual-tube kilovoltage X-ray localization system in the radiotherapy treatment room

ISO ISO ISO OHSAS ISO

8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques

Using a research real-time control interface to go beyond dynamic MLC tracking

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Quality control phantoms and protocol for a tomography system

Introduction to Biomedical Imaging

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

8/3/2016. Outline. The EPID Strikes Back: Future EPID Technology and Applications. Active Matrix Flat-Panel Imagers (AMFPIs)

TG-148 overview. Introduction. System Overview. System Overview. QA for Helical Tomotherapy: Report of the AAPM Task Group 148. Conflict of Interest:

Using the ACR CT accreditation phantom for routine image quality assurance on both CT and CBCT imaging systems in a radiotherapy environment

Micro-CT Methodology Hasan Alsaid, PhD

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques

Basics of treatment planning II

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

ph fax Deming Way Middleton WI USA

State-of-the-Art IGRT

Philips SPECT/CT Systems

Facility Questionnaire PART I (General Information for 3DCRT and IMRT)

S. Guru Prasad, Ph.D., DABR

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Basic Radiation Oncology Physics

MapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF

Evaluation of AutoQA Lite TM Image Quality Measurement Software

Virtual Phantoms for IGRT QA

Clinical implementation of photon beam flatness measurements to verify beam quality

MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

DUAL-ENERGY CT IN PROTON THERAPY

Brilliance CT Big Bore.

DEVELOPING AND IMPLEMENTING A HIGH PRECISION SETUP SYSTEM

Table of Contents. About AQUA

Raising the Bar in IMRT QA

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

Overview of Proposed TG-132 Recommendations

Quality assurance of a helical tomotherapy machine

Determination of rotations in three dimensions using two-dimensional portal image registration

BME I5000: Biomedical Imaging

Advanced Radiotherapy

Digital Image Processing

Lag correction model and ghosting analysis for an indirect-conversion flat-panel imager

Radiation Oncology treatment facility design. Simulators, CT scanners, HDR Brachytherapy

UNCOMPROMISING QUALITY

Transcription:

Introduction Quality Assurance for Image- Guided Radiation Therapy Jean-Pierre Bissonnette, Ph.D., MCCPM Princess Margaret Hospital, Toronto, Canada IGRT What is it? Rationale Equipment Quality Assurance Acceptance Commissioning Quality Control Geometric integrity Image quality Patient-specific QA Image-Guided Radiation Therapy Frequent imaging during a course of treatment as used to direct radiation therapy. It is distinct from the use of imaging to enhance target and organ delineation in the planning of radiation therapy. Justification for IGRT Accuracy: verify target location (QA) Precision: tailor PTV margins (patient-specific) Adaptation to on-treatment changes Correct & moderate setup errors Assess anatomical changes Re-planning ( naïve or explicit) + + + 1

Rationale against IGRT IGRT Technologies Increased complexity Find new sources of error Patient dose Redefining workload (more?) Therapy, Physics, Oncology Time Resources/Infrastructure PRIMATOM Ultrasound TomoTherapy Hi-Art kv Radiographic Elekta Synergy Portal Imaging Varian OBI Artiste kv CT MV CT kv and MV Cone-beam CT IGRT Technologies EPID: Image Acquisition Modes PRIMATOM kv CT Ultrasound TomoTherapy Hi-Art MV CT kv Radiographic Elekta Synergy Portal Imaging Varian OBI kv and MV Cone-beam CT Artiste Localization Image Pre-Port Verification Exposure Portal During Treatment Double Exposure Movie-Loop Multiple Images During a Single Treatment Measurement Tools Med Phys. 28 (5) 712-737 2

IGRT Technologies Ultrasound kv Radiographic Portal Imaging kv Radiographs & Fluoroscopy Reference high contrast anatomy, or implanted markers. More explicit information than MV portal imaging. Lower dose than MV portal imaging. Real-time monitoring with fluoroscopy. Commercial examples: BrainLab, Accuray Varian OBI Elekta XVI PRIMATOM TomoTherapy Hi-Art Elekta Synergy Varian OBI Artiste kv CT MV CT kv and MV Cone-beam CT Real-time Tumortracking System for Gated Radiotherapy Highly Integrated System (4 x- ray tubes, 4 Image Intensifiers) Temporal Resolution: 30 fps Spatial Targeting Precision: 1.5 mm @ 40 mm/s Shirato H et al., Hokkaido University School of Medicine, Sapporo, Japan. Range of motion w.r.t. Tx port (4 patients with Ca Lung): With real-time gating: 2.5-5.3 mm Without real-time gating: 9.6-38.4mm Shirato H et al., Hokkaido University School of Medicine, Sapporo, Japan. 3

Targeting System X-ray sources IG Technologies Synchrony camera Linear accelerator Manipulator Robotic Delivery System Ultrasound kv Radiographic Portal Imaging Image detectors PRIMATOM TomoTherapy Hi-Art Elekta Synergy Varian OBI Artiste kv CT MV CT kv and MV Cone-beam CT Optically Guided 3D Ultrasound Optically Guided 3D Ultrasound CCD-Camera 2D-Ultrasound Probe with IRELD tracking device NOMOS: BATCAM system Varian: SonArray Resonant Medical: Restitu Restore Prostate Table mounted Passive tracking device Ultrasound Images are displayed for the operator in real time on the screen as they are acquired. 4

US for Image-Guided RT Axial Images IGRT Technologies Ultrasound kv Radiographic Portal Imaging Uncorrected Kuban DA, Dong L, et al Semin Radiat Oncol 15(3):180-191 (2005). Van den Heuvel, et al Med Phys 30(11):2878-2887 (2003) Artignan et al, IJROBP 59(2):595-601 (2004) Ramos Poli et al, IJROBP 67(5) 1430-1437 (2007) Shifts to be applied via 3D couch translation PRIMATOM kv CT TomoTherapy Hi-Art MV CT Elekta Synergy Varian OBI kv and MV Cone-beam CT Artiste Soft-tissue Imaging for Guidance IGRT Technologies Reference to internal soft-tissue anatomy. Stronger correlation between imaged contrasts and target anatomy. Computed Tomography kv and MV Review technologies, and applications PRIMATOM Ultrasound TomoTherapy Hi-Art kv Radiographic Elekta Synergy Portal Imaging Varian OBI Artiste kv CT MV CT kv and MV Cone-beam CT 5

Tomotherapy - MVCT Conventional CT detector General Electric Xe filled cavities >700 detector elements readout cycle 300Hz Utilizes treatment beam Lower X-ray energy 3.5MV instead of 6MV Linac detuned Quantitative Imaging with MVCT Avoids artifacts and photon starvation for highly attenuating and high-z materials Facilitates contouring, planning, and dose reconstruction GE Lightspeed PET/CT Tomotherapy MVCT IGRT Technologies X-Ray Volume Imaging Platforms Ultrasound kv Radiographic Portal Imaging Elekta Synergy Varian OBI Artiste PRIMATOM TomoTherapy Hi-Art Elekta Synergy Varian OBI Artiste kv CT MV CT kv and MV Cone-beam CT 6

Cone-Beam CT: From Slice to Cone Cone-Beam CT: From Slice to Cone fan beam x-ray source Linear Array Detector Many Rotations fan beam x-ray source Linear Array Detector Many Rotations Conventional CT Conventional CT cone beam x-ray source Single Rotation cone beam x-ray source Single Rotation Cone-Beam CT asi Flat-panel Detector Cone-Beam CT asi Flat-panel Detector Megavoltage CBCT Uses treatment beam (6MV). Imaging/Tx share isocentre. Very low dose-rate (0.005 MU/deg) beam-pulse triggered image acquisition a-si Panel EPID optimized for MV Typical acquisition: Half rotation (200 degrees, ~ 45s) ~ 2min reconstruction (~256 3, 0.5mm) (27 cm) 3 FOV Typical dose: 2 to 9 cgy Immune from electron density artifacts CT MVCBCT (9MU) Courtesy of J. Pouliot Cone-Beam Computed Tomography Features soft-tissue contrast patient imaged in the treatment position 3-D isotropic spatial resolution geometrically precise calibrated to linac treatment iso-centre Limitations NOT fast acquisition 0.5-2 minutes NOT diagnostic quality Truncation artifacts Image lag/ghosting No scatter rejection 7

Current Paradigm in External Beam Radiation Therapy QA ❶Acceptance testing Meets specifications in tender ❷Clinical Commissioning Prepare for clinical work ❸Periodic QC Testing Ensure stable, reproducible performance ❹Patient-specific QA Current Paradigm in External Beam Radiation Therapy QA ❶Acceptance testing Meets specifications in tender ❷Clinical Commissioning Prepare for clinical work ❸Periodic QC Testing Ensure stable, reproducible performance ❹Patient-specific QA Acceptance Testing Verify that equipment meets specifications agreed upon in the purchase Verify that all safety systems are operational Follow manufacturer s customer acceptance procedure Test any other clauses in the purchasing contract Establish baseline values Current Paradigm in External Beam Radiation Therapy QA ❶Acceptance testing Meets specifications in tender ❷Clinical Commissioning Prepare for clinical work ❸Periodic QC Testing Ensure stable, reproducible performance ❹Patient-specific QA 8

Clinical Commissioning Prepare equipment and staff for clinical work Training Safety Comprehensive of baseline values for QA No guidance - yet! Clinical factors to consider Accuracy of process Staff workload Patient tolerance Dose Resources (time, staffing) Applicability Clinical context Clinical Commissioning: Accuracy Implementation is greatly facilitated when performed in parallel with existing image guidance Portal imaging with implanted markers Ultrasound Head-to-head comparison CBCT vs US US vs portal imaging CBCT vs in-room CT During Commissioning Dry Runs Chose phantom that allows for independent verification of accuracy Treat phantom exactly like a live patient Planning scan (test orientation info!) Treatment plan (place isocentre!) R&V system Remote setup correction automated couch Have therapists perform setup and treatment Image or localization review Identify and solve problems before they re clinical problems (workarounds) Current Paradigm in External Beam Radiation Therapy QA ❶Acceptance testing Meets specifications in tender ❷Clinical Commissioning Prepare for clinical work ❸Periodic QC Testing Ensure stable, reproducible performance What can go wrong? ❹Patient-specific QA 9

IGRT systems QC Safety Safety Geometric accuracy System stability Image quality System infrastructure Dose Test all interlocks Door kv source arm Flat panel arm Terminate key Visual inspection No loose covers Hanging wires Test all collision detection devices Test all relevant radiation monitors Geometric accuracy: coincidence with MV isocentre Coincidence with MV isocentre Image reconstruction isocentre Point of interest Linac mechanical isocentre Variations of the Winston-Lutz test used for brain stereotactic QA Lutz, Winston, & Maleki, IJROBP 14, pp. 373-81 (1988) y Linac radiation isocentre z x Calibrated isocentre 10

Coincidence with MV isocentre Coincidence with MV isocentre Direct method Place object directly at radiation isocentre Calibrate IGRT device against that object + Burn beam isocentre directly into the image dataset + Highly accurate Takes 2 hours to perform Indirect method Place object at surrogate of radiation isocentre (i.e., lasers) Calibrate IGRT device against that object + Minutes to perform + Can calibrate daily Subject to laser imprecision and drift Direct method examples: Elekta Synergy MVCT Cyberknife Sharpe et al, Med. Phys. 33, 136-144, 2006 Morin et al, Med. Phys. 34, 2634, 2007 Calibration using MV Imaging Gantry Angle (degrees) Required Shift [mm] 1. MV Localization (0 o ) of BB; collimator at 0 and 90 o. 2. Repeat MV Localization of BB for gantry angles of 90 o, 180 o, and 270 o. 3. Analyze images and adjust BB to Treatment Isocentre (± 0.3 mm) 0 90 180 270 R 1.58 ± 0.89 A 0.19 ± 0.55 I 0.78 ± 0.54 +1mm θ gantry u -1mm -180 v θ gantry +180 θ gantry 4. Measure BB Location in kv radiographic coordinates (u,v) vs. θ gantry. 5. Analysis of Flex Map and Storage for Future Use. Reconstruction 6. Employment of Flex Map During Routine Clinical Imaging. 11

Long-term Stability: Flexmap Flexmap 1.25 R esidu al disp lacem ent (m m ) 1 0.75 0.5 0.25 0-0.25-0.5-0.75-1 -1.25-1.5 95% confidence interval = 0.25 mm -180-135 -90-45 0 45 90 135 180 Gantry angle (degrees) v u 12 calibrations over 28 months A plot of the apparent travel of a point as a function of gantry angle. Removes the effect of component flexes and torques prior to reconstructions. Ties the 3D image matrix to the radiation isocentre of the accelerator. Effect of absent Calibration Effect of Incorrect Calibration Blur Image translocation 12

Coincidence with MV isocentre: MVCT Coincidence with MV isocentre: MVCT Reconstruction Program Courtesy of O. Morin Coincidence with MV isocentre: MVCT Courtesy of O. Morin Geometric accuracy: CyberKnife Courtesy of O. Morin 13

Coincidence with MV isocentre Indirect method examples Varian OBI BATCAM, SonArray, Resonant In-room CT CTVision Isocenter accuracy 2D-2D Cube phantom Marker phantom Courtesy of S. Yoo Isocenter over gantry rotation Mechanical accuracy Tolerance Displacement < 2mm Preparation Phantom with a center marker 0, 90, 180, and 270 Tolerance Mechanical pointer Displacements ± 2 mm Courtesy of S. Yoo Courtesy of S. Yoo 14

3D ultrasound device calibration IRLED or Passive Markers 2D Ultrasound Probe with IRLED tracking array Accuracy of Optically Guided 3D Ultrasound Calibration wires Intersection of calibration wires with image plane A B DP Courtesy of W. Tomé Courtesy of W. Tomé Tomé et al., Med. Phys. 29(8), 1781-1788 (2002). Geometric calibration - BATCAM Geometric calibration - Restitu 15

Geometric accuracy: in room CT Daily Geometry QC 8 mm BB CCD camera Align phantom with lasers Acquire portal images (AP & Lat) & assess central axis Acquire CBCT Difference between predicted couch displacements (MV & kv) should be < 2 mm Computer Kuriyama et al., IJORBP 55, 428-435, 2003 Daily Geometry QC Warms up the tube Checks for sufficient disk space Tests remote-controlled couch correction Can be well-integrated in QC performed by therapists 1. Shift BB embedded in cube from isocentre. θ gantry 2. MV Localization of BB for gantry angles of 0 o and 90 o. Reconstruction 3. kv Localization with cone-beam CT 4. Compare kv and MV localizations; tolerance is ± 2 mm 5. Use automatic couch to place BB to isocentre; verify shift with imaging 16

2D2D match and couch shift Phantom Setup at Iso kv images (AP/Lat) 2D2D match Before matching Match DRR s graticule to off centered marker After matching Apply couch shift Remote shift couch Verify the position in room 17

Image quality Image quality Scale Spatial resolution (MTF) Noise Uniformity Signal Linearity (CT numbers) AAPM CT Performance Phantom CatPhan 500 phantom Scale Uniformity Geometric calibration to tie isocentre to centre of volumetric reconstruction Scale to relate all pixels to isocentre Bissonnette et al., Med Phys 35, pp. 1807-1815 (2008) Standard CT tests Cupping, capping Baseline nonuniformity index: CT CT max max CT + CT min min Bissonnette et al., Med Phys 35, pp. 1807-1815 (2008) 18

Linearity of CT Numbers Linearity of CT numbers: 7 units (Synergy + OBI) 2000 1800 1600 Measured Hounsfield unit 1400 1200 1000 800 600 Unit 7 Unit 8 Unit 9 Unit 10 Unit 12 Unit 16 Unit 16 with annulus Unit 17 400 200 0 0 200 400 600 800 1000 1200 1400 1600 1800 2000 Theoretical Housfield unit Bissonnette et al., Med Phys 35, pp. 1807-1815 (2008) Bissonnette et al., Med Phys 35, pp. 1807-1815 (2008) Spatial Resolution (Synergy and OBI) MTF 1.2 1.0 0.8 0.6 0.4 Unit A Unit B Unit C Unit D Unit E Unit F Unit G Unit H Unit I Unit J Image quality phantom 20 cm diameter Four 2-cm sections: 1 solid water section for noise and uniformity 2 sections with inserts for contrast resolution Beads 4 sections 0.2 0.0 0 2 4 6 8 10 Spatial frequency (cm -1 ) Bissonnette et al., Med Phys 35, pp. 1807-1815 (2008) 1 section with bar groups for spatial resolution 12 beads for position accuracy Gayou & Miften, Med Phys 34, 3183-3192 (2007) Courtesy of M. Miften 19

(1) 1% (2) 3% (3) 5% (Brain) (4) 7% (Liver) (5) 9% (Inner bone) (6) 17% (Acrylic) (7) Air (8) 48% (Bone 50% mineral) 2 cm object with 1% contrast Image quality Gayou & Miften, Med Phys 34, 3183-3192 (2007) 1: 0.067 lp/mm 2: 0.1 lp/mm 3: 0.15 lp/mm 4: 0.2 lp/mm 5: 0.25 lp/mm 6: 0.3 lp/mm 7: 0.4 lp/mm 8: 0.5 lp/mm 9: 0.6 lp/mm 10: 0.8 lp/mm 11: 1.0 lp/mm Courtesy of M. Miften Resolution vs. Exposure kv-ct MV-CBCT 12 cgy Smallest visible bar group was 0.3 lp/mm for the 3 & 5 MU protocols 0.4 lp/mm for all other protocols. kv-ct was 0.6 lp/mm Courtesy of M. Miften Tomotherapy image quality: contrast-detail 1.25 mm objects resolved Image quality QA: 2D-2D Leeds phantom TOR 18FG (Leeds test objects Ltd, UK) Low contrast resolution with 1mm copper plate Spatial resolution Meeks et al., Med Phys 32, 2673-81, 2005 Courtesy of K. Langen Courtesy of S. Yoo 20

Image quality QA: 2D-2D Image Quality - 3D Ultrasound Fluoroscopic Radiographic Low contrast resolution Tolerance: > 11 12 disks Fluoro: 70 kvp, 32 ma, 6ms 11 13 discernable disks Radio: 75 kvp, 25 ma, 6ms 13 15 discernable disks Courtesy of S. Yoo Spatial resolution Tolerance: > 11 th group Fluoro: 50-80kVp, 80mA, 32ms 9 11 th group Radio: 50-80kVp, 80mA, 120ms 10 12 th group AAPM report #65, Med. Phys. 25, 1385-1406 (1998) Conclusions Conclusions Several QA program have been proposed for IGRT systems No formal guidance from AAPM task group reports - yet Elements common to all: Geometric accuracy and precision Image quality Recognize the value of IGRT systems as a measurement tool for new and existing processes. Faces new challenges. Daily QC of geometric accuracy is commonplace 21

References References CBCT Lehmann et al., JACPM 8, 21-36, 2007 Islam et al., Med Phys 33, 1573-1582, 2006 Saw et al., Med. Dosimetry 32, 80-85, 2007 Bissonnette et al., Med Phys 35, 1807-1815, 2008 Bissonnette et al., IJROBP 71, S57-S61, 2008 Yoo et al., Med Phys 33, 4431-4447, 2006 Jaffray et al., Modern Technology of Radiation Oncology Vol. 2, chapter 7, 2005 US Tomé et al. Med Phys 29, 1781-1788, 2002 Lachaine et al. Med Phys 32, 2154-2155, 2005 Bouchet et al., PMB 46, 559-77, 2001 MVCT & Tomotherapy Gayou & Miften, Med Phys 34, 499-506, 2007 Meeks et al., Med Phys 32, 2673-81, 2005 Langen et al., PMB 50, 4259-76, 2005 Morin et al., Med Phys 34, 2634, 2007 CT on rails Court et al., Med Phys 30, 1198-210, 2003 Kuriyama et al., IJORBP 55, 428-435, 2003 kv radiography Verellen et al., Radiother Oncol 67, 129-141, 2003 Murphy et al., Med Phys 23, 2043-9, 1996 CyberKnife Yan et al. Med Phys 30, 3052-60, 2003 AAPM reports: Diagnostic radiology #74 CT scanners #39 B-mode US #65 22