Quality control phantoms and protocol for a tomography system

Similar documents
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

X-ray CT analyses for paleontological subjects in the ICP system

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

Recognition and Measurement of Small Defects in ICT Testing

Pipe Inspection with Digital Radiography

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

Image Acquisition Systems

Introduction to Biomedical Imaging

Enhanced material contrast by dual-energy microct imaging

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Digital Image Processing

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

DUE to beam polychromacity in CT and the energy dependence

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Coordinate Measuring Machines with Computed Tomography

S. Guru Prasad, Ph.D., DABR

Translational Computed Tomography: A New Data Acquisition Scheme

Thickness Measurement of Metal Plate Using CT Projection Images and Nominal Shape

CT Systems and their standards

Investigation on reconstruction methods applied to 3D terahertz computed Tomography

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications

Simulation of Internal Backscatter Effects on MTF and SNR of Pixelated Photon-counting Detectors

CIVA Computed Tomography Modeling

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

ISO ISO ISO OHSAS ISO

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Applying Hounsfield unit density calibration in SkyScan CT-analyser

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package.

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.

A study on image quality provided by a kilovoltage cone-beam computed tomography

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

James R Halama, PhD Loyola University Medical Center

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system

Central Slice Theorem

3D X-ray Microscopy Characterization. Metal Additively Manufactured Parts. Application Note

INTERNATIONAL STANDARD

Development of an automated bond verification system for advanced electronic packages

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Traceable 3D X-Ray Measurements ZEISS METROTOM

Methods for Quantitative Characterization of Large-Scale High Energy Computed Tomography Systems

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

X-rays see all X-ray computed microtomography (µct) a novel technique available at CERN for material and metrological inspection

High-resolution X-ray CT Inspection of Honeycomb Composites Using Planar Computed Tomography Technology

3D Computed Tomography (CT) Its Application to Aerospace Industry

Comparison of Scatter Correction Methods for CBCT. Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis

Ch. 4 Physical Principles of CT

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

X-ray Industrial Computed Laminography (ICL) Simulation Study of Planar Objects: Optimization of Laminographic Angle

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM

Frequency-based method to optimize the number of projections for industrial computed tomography

ML reconstruction for CT

Advanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies

Characterization of microshells experimented on Laser Megajoule using X-Ray tomography

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT

Case Studies in the Use of Computed Tomography for Non-Destructive Testing, Inspection and Measurement

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units

Radon Transform and Filtered Backprojection

Suitability of a new alignment correction method for industrial CT

How does the ROI affect the thresholding?

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

연구용유방초음파질관리 원광대학병원김혜원

Introduction to Positron Emission Tomography

MEDICAL IMAGE ANALYSIS

Dr. Javier Santillan, San Carlos, CA

Simulation of Radiographic Testing for POD Assessment

Spatial Resolution Properties in Penalized-Likelihood Reconstruction of Blurred Tomographic Data

Computed tomography (Item No.: P )

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

CT vs. VolumeScope: image quality and dose comparison

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Simulation Based POD Estimation for Radiographic Testing of Turbine Blades

Advanced Reconstruction Techniques Applied to an On-Site CT System

Comparison of Reconstruction Methods for Computed Tomography with Industrial Robots using Automatic Object Position Recognition

CERTIFICATION OF PERSONNEL ENGAGED IN COMPUTED RADIOGRAPHIC TESTING (CRT) AND/OR COMPUTED RADIOGRAPHIC INTERPRETATION (CRI)

GE s Revolution CT MATLAB III: CT. Kathleen Chen March 20, 2018

Evaluation of Computed Tomography Techniques for Material Identification in Low Level and Intermediate Level Waste

Reconstruction Methods for Coplanar Translational Laminography Applications

Calibration of bi-planar radiography with minimal phantoms

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

INTERNATIONAL STANDARD

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

INDUSTRIAL SYSTEM DEVELOPMENT FOR VOLUMETRIC INTEGRITY

SUPPLEMENTARY INFORMATION

Medical Image Reconstruction Term II 2012 Topic 6: Tomography

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Transcription:

Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing (NDT) in industrial applications need a specific quality assurance protocol. Specific phantoms have been chosen to establish the test protocol of our own-developed tomograph system in the technological center Aimen. The principal image quality parameters to monitorize are the spatial resolution, noise and contrast sensibility. The system noise can be also considered a uniformity measurement, from the mean value of different regions of interest s noise values. Contrast sensibility is defined from the signal to noise ratio, which must be calculated for different acquisition parameters. The system allows varying the object-to-source distance, so the phantoms are used to characterize the magnification (along with the pixel size) in the tomography system. The study for characterizing these parameters is carried out for different acquisition parameters in the CT system. The different materials of the phantoms (steel, aluminium and PMMA) allow the characterization of image quality values for different industrial materials, as those inspected in a NDT department. This analysis is repeated over a time period to finally fix a specific quality assurance protocol and its periodicity. Keywords: tomography, quality control, spatial resolution, noise, beam hardening 1 Introduction Industrial tomography systems are becoming usual in NDT applications, within different areas of interest, and usually oriented to specific production. The Technological Centre AIMEN is oriented to welding technology and different research areas. Different pieces (in terms of size and density) must be analyzed: welds, aeronautical, biological probes, etc. Therefore, a customized CT/DR system was developed for inspecting different pieces by adjusting every acquisition (time, magnification) and energy parameter. Both the hardware and the software (reconstruction by the filtered backprojection algorithm, FBP) along with a specific visualization tool, were developed in collaboration with the University of Santiago de Compostela. As every imaging system, our own-developed tomograph requires a quality assurance protocol to provide the reliability our inspection services and researchs demand. The principal imaging parameters to monitorize are the spatial resolution, uniformity, noise and contrast sensibility [1-2]. Artifacts must also be characterized, as those of beam hardening in high attenuation pieces. On other side, since the system magnification is variable (the rotary stage is fixed to a horizontal linear stage to vary the object to detector distance) the scale must be calibrated, thus giving the pixel size in length units. Recently in 2012, Spanish standards have been published following European ones related with tomography in non-destructive testing, EN 16016-1:4 [3-6], related with the corresponding ISO standards. The aim of this work is to establish simple tests to calculate basic parameters to monitorize the behaviour and time evolution of the CT system and to analyze the correction needs for maintaining the initial tomographic image quality. 375

2 Methods and Materials The analysis of image parameters and establishment of a quality control protocol is based on imaging phantoms of different materials: steel (Fe), aluminium (Al) and PMMA (for low-density pieces). In this way, the image quality parameters are set for a range of different attenuation values, and they are studied for different values of energy, magnification (pixel size) and projections (see Table 1). Energy (kv) 50, 100, 150, 180, 225 Number of projecctions 360,, 1200 Pixel size, mm (magnification).16 (5),.2 (4),.4 (2),.5 (1.6),.6 (1.33) Table 1: Different acquisition parameters for our CT quality control. 2.1 Phantoms Uniform blocks are used for calculating the uniformity, spatial resolution, contrast sensitivity and image to noise ratio. Steel and aluminium blocks are cilinders of 25 mm, 45 mm diameter respectively. PMMA phantom is a rectangular block of 35.5 mm x 45 mm section size. The energy values for the study vary depending on the material: 180, 225 kev for Fe, 100, 150, 180, 225 kev for Al and 50, 100, 150, 180, 225 kev for PMMA. 2.2 Image parameters The system image quality parameters to study are the uniformity, noise, wire detectability, contrast resolution, beam hardening correction, spatial resolution, magnification scale and alignment (of detector center-rotation axis). These values are analyzed completely once a year, for different energies, number of projections for image acquisition and magnification (scale), following Table 1. The parameters and the corresponding calculations are explained below. The images for measurements are the reconstructed ones for the different phantoms, with grey levels in units of cm -1 (linear attenuation coefficient dimensional units). No further calibration (for example, to Hounsfield units) is used since it will go beyond the scope, industrial non destructive testing, of our CT. 2.2.1 Magnification An object of known dimensions is radiographed in consecutive magnification positions. The inverse of the magnification is linear dependent with the horizontal position of the reference object. Performing the linear fit (by least squares method) the parameters for calculating the pixel size (in mm or other length units) are obtained, thus leading to a known dimensioning accuracy. 2.2.2 Alignment The detector center to rotation axis and x-ray focus alignment is decisive for a correct tomographic reconstruction and very important for all the magnification values. A first rough alignment is performed aided with a laser source, and later checked by imaging of a thin wire. The sinogram (tomographic data before reconstruction) is a sine curve and it must be centered in the central channel of the detector (640 in this case). Moreover, the reconstructed image must be a point if every element is correctly aligned. 2.2.3 Detectability A set of wires from industrial IQI (image quality indicators) are imaged to control the detectability for Fe and Al. Wire diameters are 0.1, 0.125, 0.16, 0.2, 0.25, 0.32 and 0.4 mm. It should be remembered that a feature smaller than a single pixel can affect it to such an extent that it will appear with a visible contrast relative to adjacent pixels. In this way, the individual detectability will be greater than the spatial resolution from the f 50 parameter, and accounts for the smallest feature resolved. 376

On the other side, a set of duplex wires is used to establish the detectability of close features, giving another value for the blurring process of the CT system. The wire diameters and spacing are: 0.10, 0.13, 0.16, 0.20, 0.26, 0.32, 0.40, 0.50, 0.64, 0.80, 1.00, 1.26 and 1.16 mm. The IQI image shows that detectability alone is often non sufficient, since features must be discriminated (detected and resolved). 2.2.4 Uniformity The uniformity is calculated from 5 different regions of interest (roi) in the phantom image. The uniformity is the mean value (in units of cm -1, linear attenuation coefficient) for every roi divided by the mean value of the area covering the whole phantom image. The maximum of these values will be computed as the final uniformity value (dimensionless). Beam hardening corretion The uniformity value is also used to analyze the beam hardening contribution and its correction by software filtering. This filtering corrects the measured signal to the expected signal without beam hardening with a polinomial fit. The fit calculation is performed separately for Al and Fe pieces, by imaging a step-phantom with different widths. The measured signal is fitted to the expected theoretical one [7], after adjusting the x-ray energy and the attenuation coefficients (from the NIST database [8]). After the correction, the comparison between corrected and non corrected uniformity values, as well as the line profiles of the reconstructed images, gives the contribution of this filtering. 2.2.5 Noise Noise measurement is computed as the mean value of noise values of different background rois in the images. Noise individual values are computed as the standard deviation of the pixels values. 2.2.6 Contrast Contrast is determined from the signal to noise ratio (SNR) and the simple definition as the relative difference of a feature from the background, as a percentage. The signal to noise ratio (SNR) accounts for the contrast resolution, and it is calculated as the ratio of the mean value and noise (as the standard deviation) of a given region of interest covering the phantom image. The established contrast value is calculated as the ratio of the mean values of similar rois covering the block piece image and the background. 2.2.7 Spatial resolution The spatial resolution is defined as the f 50 parameter from the modulation transfer function (MTF), calculated from the edge spread function (ESF). This parameter is calculated from the procedure described in the standards [2]. First, different line profiles are acquired for the phantom image. These profiles are aligned and averaged (to reduce system and quantization noise) obtaining the ESF profile. The system line spread function (LSF) is obtained by the discrete derivative of the ESF and finally the system modulation transfer will be obtained by the discrete transformation of the LSF. The f 50 parameter is the spatial frequency, in units of line pairs per mm (lp/mm) for the 0.5 value of the normalized MTF (by convention, MTF is normalized to unity). 377

Figure 1: MTF calculation: example of an ESF profile (right), averaged ESF (left). Figure 2: MTF calculation: LSF (left) and finally normalized MTF (right) The PSF function accounts for the blurring function describing the image system, that is, the response of the system when imaging an ideal point object, while the MTF describes its differential ability to reproduce spatial frequencies. It is used to measure system performance and the effect on the effective contrast of small features. 3 Results For simplicity, only the global characteristics and image parameters of magnification 4 (pixel size 0.2 mm) and 225 kev (X-ray source maximum energy) are shown below, along with their analysis. 3.1 System magnification The linear fit for the magnification values and the horizontal position of the piece are: M -1 = 5.3 10-4 dfo + 3.8 10-17 dfo stands for object to source distance (in mm). This magnification scale is fixed on an initial calibration of the CT system and checked periodically (once a year) through measurement of a piece of known dimensions. 3.2 Alignment The alignment is checked for every magnification value, by imaging a thin wire. The sinogram must be centered in the central detector channel (640), as shown in Figure 3. 378

Figure 3: Sinogram of the wire (left) and corresponding reconstructed image (right). 3.3 Detectability Figure 4: Example of IQI reconstructed CT slice (left) and duplex IQI profile from reconstruction (right) Number of wires detected 1200 projections projections Duplex IQI 4 (up to 0.8 mm) 3 (up to 1 mm) Fe 7 (up to 0.1 mm) 7 (up to 0.1 mm) Al 5 (up to 0.16 mm) 5 (up to 0.16 mm) Table 2: Wire detectability of the system. The smallest feature detected is smaller than the pixel size for this magnification value (pixel size of 0.2 mm). 3.4 Uniformity Material 1200 proj proj 360 proj Steel 1.36 + 0.04 1.35 + 0.04 1.27 + 0.04 Steel - beam hardening corrected 1.30 + 0.02 1.29 + 0.02 1.25 + 0.02 Aluminium 1.072 + 0.003 1.074 + 0.003 1.069 + 0.002 Aluminium - beam hardening corrected 1.085 + 0.002 1.086 + 0.003 1.086 + 0.003 PMMA 1.012 + 0.001 1.011 + 0.001 1.012 + 0.001 Table 3: Uniformity results for Fe, Al, PMMA and for beam hardening Fe, Al corrections. 379

Uniformity values are dimensionless. For this energy (225 kev), decreasing values (reflecting a better uniformity) can be found for Fe (see also Figure 5), while for Al the differences are negligible. This comes from the fact that the beam hardening effect at this energy is not important for the Al width. Figure 5: Fe phantom profiles with and without beam hardening correction 3.5 Noise Noise values were computed from the reconstructed images of the steel phantom (background values). From Table 4 it can be seen that the differences between different projections are negligible. 1200 proj proj 360 proj Mean of noise values (cm -1 ) 0.0076 + 0.0008 0.0080 + 0.0008 0.0080 + 0.005 3.6 Contrast Table 4: Noise values from reconstructed images Signal to noise ratio, SNR = mean(roi)/std(roi) (std: standard deviation) Material 1200 proj proj 360 proj Mean value Steel 46.5 46.4 43.1 45.3 + 1.5 Aluminium 366.6 397.5 284.3 382.8 + 10.8 PMMA 968.8 781.6 447.0 732 + 190 Table 5: SNR measured values The SNR values increase for diminishing densities. These values shows the beam hardening effect, which causes a cupping reconstructed image, is considerably more important in the Fe block than in the Al and PMMA ones. Contrast value, C = mean(piece)/mean(background) 380

Figure 6: Scheme of the background area for contrast measurement. Material 1200 proj proj 360 proj Mean value Steel 118.2 115.5 117.7 117.1 + 1.1 Aluminium 49.0 50.9 47.3 49.0 + 1.2 PMMA 62.1 63.8 1.9 62.6 + 0.8 Table 6: Contrast measured values. The significant difference is between the Fe results and the Al, PMMA ones. This accounts clearly for the higher attenuation on the Fe block. 3.7 Spatial resolution f 50 (lp/mm) 1200 proy proy 360 proy Mean value Steel 0.225 0.226 0.222 0.224 + 0.001 Aluminium 0.114 0.139 0.108 0.120 + 0.012 PMMA 0.186 0.189 0.177 0.184 + 0.004 Table 7: Spatial resolution as the spatial frequency (in line pairs per mm) for the 50% of the normalized MTF The f 50 values are low, as expected from the detectability of the duplex IQI. 0.8 mm and 1 mm (see Table 2) are equivalent to 0.25 and 0.31 lp/mm, making these results to be compatible. 4 Discussion The different values and characteristics of the system give a quantitative performance of the CT imaging. Moreover, the similarity between the parameter s measurements for different number of projections shows that for routine testing the whole set of measurements can be minimized. To follow this trend, and to establish a simpler quality assurance control, the acquisition parameters are restricted to relevant values in relation with the phantom s materials (see Table 8). We have established the system image quality parameters through the obtained values as well as the testing periodicity (Table 2). These values will be analyzed once a year to determine the image derivation and to readjust the specific quality assurance protocol. 381

Parameters Material Projections Periodicity Energy (kev) Uniformity Noise SNR, contrast Wire detetability Fe, Al, PMMA Fe, Al, PMMA Fe, Al, PMMA Fe, Al, IQI duplex, 1200 Monthly 225 Fe, Al / 100 PMMA 225, 100 225, 100 225 Beam hardening (uniformity) Spatial resolution Fe, Al / Fe, Al step phantoms Fe, Al, PMMA, radiography 6 months 225 225 Magnification scale Alignment (rotation ax-det.center) Fe block Fe wire Radiography Yearly 225 225 Table 8: Measurements, acquisition parameters and periodicity for the CT QA protocol 5 Conclusions The development of the protocol for the quality control of the AIMEN computed tomography system was established from the characterization and measurement of different parameters. Following the international standards regarding industrial tomography, several key parameters were chosen and simply established from uniform phantoms of Fe, Al and PMMA. In this way, the complete set of measurements was finally simplified in parameters and requirements to get a useful guide (see Table 8) for a reliable image quality control. This document also includes the geometrical calibration (magnification scale and alignment of the rotation axis and detector center), what is needed because of the variable magnification of our system. In addition, software filtering for beam hardening correction in Fe and Al pieces is checked with uniformity measurements. From the first results presented and with a quality assurance guide created, it is intended to generate an extense database of these measurements to study the derivation in time of the CT image quality. Acknowledgements The author would like to thank the help of her radiography colleagues at the NDT department. References [1] E 1570 00. Standard Practice for Computed Tomographic (CT) Examination. [2] E 1441 00. Standard Guide for Computed Tomography (CT) Imaging. [3] UNE-EN 16016-1. NDT Radiation methods. Computed Tomography. Part 1: Terminology. [4] UNE-EN 16016-2. NDT Radiation methods. Computed Tomography. Part 2: Principle, equipment and samples. [5] UNE-EN 16016-3. NDT Radiation methods. Computed Tomography. Part 3: Operation and interpretation. [6] UNE-EN 16016-4. NDT Radiation methods. Computed Tomography. Part 4: Qualification. [7] J. Hsieh. Computed tomography: principles, design, artifacts and recent advances, SPIE Press 2003. [8] http://physics.nist.gov/physrefdata/contents.html 382