Image Optimisation. -How to Get the Most out of your Echo Machine. Katarina Winkler, Biomedicinsk analytiker, Barnhjärtcentrum i Lund
|
|
- Scott Richardson
- 5 years ago
- Views:
Transcription
1 Image Optimisation -How to Get the Most out of your Echo Machine analytiker, Barnhjärtcentrum i Lund
2
3 Image Optimisation Helpful tips to improve the 2D-Image and Doppler registrations. Common adjustments for the most routinely used applications on the ultrasound machine. Frequently made beginner mistakes. I will not go deep into physics
4 Image Optimisation What is good quality image?
5 Applications on the Ultrasound Machine 2D-Imaging M-mode Spectral Doppler Color Doppler Tissue Doppler 3D-Imaging
6 Presets Create presets. Do not use factory presets. All machines at the lab MUST have the same presets. Enables uniformed exams.
7 2D Optimisation GAIN iscan ZOOM Depth Focus
8 Higher frequency 2D Optimisation Selection of Transducer Better resolution - Less penetration Lower frequency Better penetration - Less resolution Broad band transducers have a range of frequencies.
9 2D Optimisation Adjust the sector size to increase Framerate Depth Set the depth to a minimum according to what need to be imaged. Sector width Set the width as narrow as possible. Lateral resolution
10 2D Optimisation Sector size Excessive depth Depth: 18 cm Wide sector angle Fame Rate: 45 Hz Adjusted sector size Depth:12 cm Decreased sector angle Frame Rate: 76 Hz
11 Overall Gain 2D Optimization Gain settings Adjusts the Brightness in the Image Equal in the whole image Influences Receiving Signals
12 2D Optimization Gain settings TGC Time-Gain Compensation Adjusts Gain at a Specific Depth Compensate for the Attenuation of Ultrasound
13 2D Optimisation Gain settings
14 2D Optimisation Gain settings iscan/auto Optimisation The machine analyses the tissue and automatically adjusts the gain settings Good starting point
15 2D Optimisation Gain settings Over gained settings Correct gain settings
16 2D Optimisation Focus Works as a photo camera lens Lateral Resolution Multiple focus zones Decreases Frame Rate!
17 2D Optimisation
18 2D Optimisation Zoom Small region of interest High Frame rate High resolution
19 2D Optimisation Frame Rate 87 Hz Frame Rate 120 Hz
20 2D Optimisation 2 nd Harmonic Imaging Reduces Image Artefacts by Filtering out Harmonic signals. Send at a lower frequency and listen for a higher frequency. Higher overtone less noise but also less energy.
21 2D Optimisation Fundamental 2nd Harmonic
22 2D Optimisation Create a 2D image with highest detail Use the highest frequency transducer without compromising penetration Use minimal depth setting Use small sector width Optimise gain settings Use zoom function for small area of interest Use the focal zone
23 M-mode Plots movements of structures along the cursor and are displayed along a time axis. Track fast movements Very High Resolution
24 M-mode Wrong positioning LVID: 60 mm Accurate positioning LVID: 49 mm
25 Ultrasound Doppler F = 2F 0V C cos θ Angle between the blood flow and the ultrasound beam is very important Stay as parallel to flow as possible Never exceed 20 degrees
26 Ultrasound Doppler
27 Ultrasound Doppler Spectral Doppler The scale is set to high Correct setting
28 Ultrasound Doppler Spectral Doppler Slow sweep speed 50 mm/s Fast sweep speed 150 mm/s
29 Ultrasound Doppler Spectral Doppler Over gained settings Correct gain settings
30 Ultrasound Doppler Spectral Doppler Spectral Doppler signals for accurate measurements. Doppler angle <20º Adjust the Scale Adjust Sweep speed Optimise Gain settings
31 Ultrasound Doppler Color doppler 2D-gain set too high Correct settings
32 Ultrasound Doppler Color doppler Color Doppler Gain
33 Ultrasound Doppler Color doppler Color Box Size
34 Ultrasound Doppler Color doppler Color Doppler Scale: 66 cm/s Color Doppler Scale: 53 cm/s
35 Ultrasound Doppler Color doppler Color Doppler Scale: 67 cm/s Color Doppler Scale: 123 cm/s
36 Ultrasound Doppler Color doppler Well defined Color Doppler jets with high Frame Rates Decrease 2D gain Adjust Color Doppler Gain Minimize Color Box Adjust Color Doppler Scale
37 What is good a quality image? 2D image with highest detail information Spectral doppler for accurate measurements Color doppler with high frame rates and well defined jets
38 Thank you!
KNOBOLOGY. Jamie Ternes, RVT William Beaumont Hospital Royal Oak, Michigan. From the Beginning
KNOBOLOGY Jamie Ternes, RVT William Beaumont Hospital Royal Oak, Michigan From the Beginning Power On/Off Patient information Presets Probe selection Scan 1 Modalities /Gray scale/2d Pulse Wave Doppler
More informationCertificate in Clinician Performed Ultrasound (CCPU)
Certificate in Clinician Performed Ultrasound (CCPU) Syllabus Physics Tutorial Physics Tutorial Purpose: Training: Assessments: This unit is designed to cover the theoretical and practical curriculum for
More informationImaging Guide Pancreatic Imaging
Imaging Guide Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Imaging System Ver 1.0 Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Imaging System Course Objectives: This guide
More information연구용유방초음파질관리 원광대학병원김혜원
연구용유방초음파질관리 원광대학병원김혜원 Why US QC? Quality control (QC) testing of ultrasound scanners is important to verify the proper and consistent operation of these devices. main goal ; quality improvement Guidelines
More informationQuick reference guide. MyLab50
Quick reference guide MyLab50 Quick reference guide MyLab50 Introduction This Quick Guide describes only the basic operating procedures for the diagnostic ultrasound MyLab50 Family, named in the following
More informationQuick Guide. EPIQ 7 Ultrasound System. English
Quick Guide EPIQ 7 Ultrasound System English Contents For easy reference, each section is color-coded. The color of each section title correlates with the color band on each page in that section. Controls
More informationACUSON P300 Ultrasound System Control Panel Diagram
www.siemens.com ACUSON P300 Ultrasound System Control Panel Diagram 1 Power On / Off 1 2 TGC Controls 3 Left Mouse Click 4 Action 4 7 5 Right Mouse Click 6 Trackball 2 3 6 5 8 7 Pointer 8 Freeze 16 Color
More informationLOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company.
TM LOGIQ V2 Ultrasound Part of LOGIQ Vision Series Imagination at work The brilliance of color. The simplicity of GE. Now you can add the advanced capabilities of color Doppler to patient care with the
More informationNuclear Associates and
Nuclear Associates 84-317 and 84-317-7000 Multipurpose Tissue/Cyst Ultrasound Phantoms Users Manual February 2005 Manual No. 84-317-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in
More informationGE Healthcare. Agile Ultrasound. The Next Revolution in Ultrasound Imaging
Agile Ultrasound The Next Revolution in Ultrasound Imaging Abstract Diagnostic use of ultrasound has greatly expanded over the past couple of decades because it offers many advantages as an imaging modality.
More informationSimplicity, quality, and flexibility
Ultrasound Lumify Simplicity, quality, and flexibility Philips Lumify ultrasound system specifications Contents 1 Introduction 3 1.1 Applications 3 2 System overview 4 2.1 System architecture 4 2.2 Imaging
More informationOPTICAL COHERENCE TOMOGRAPHY:SIGNAL PROCESSING AND ALGORITHM
OPTICAL COHERENCE TOMOGRAPHY:SIGNAL PROCESSING AND ALGORITHM OCT Medical imaging modality with 1-10 µ m resolutions and 1-2 mm penetration depths High-resolution, sub-surface non-invasive or minimally
More informationSONOLINE G40 ultrasound system
SONOLINE G40 ultrasound system Quick Scan Card: General/OB For system software version 2.0 Press the Power Switch on the upper left side of the keyboard. A blue light will indicate the power is on. Getting
More informationUNIVERSITY OF SOUTHAMPTON
UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions
More informationWorkhorse ADCP Multi- Directional Wave Gauge Primer
Acoustic Doppler Solutions Workhorse ADCP Multi- Directional Wave Gauge Primer Brandon Strong October, 2000 Principles of ADCP Wave Measurement The basic principle behind wave the measurement, is that
More informationDose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them
Dose Distributions George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Purpose To familiarize the resident with dose distributions and the factors that affect them
More informationPREMIER DIAGNOSTIC ULTRASOUND. Diagnostic ultrasound. that delivers unparalleled. image resolution.
PREMIER DIAGNOSTIC ULTRASOUND Diagnostic ultrasound that delivers unparalleled image resolution. High Resolution Goes Ultra THE NEW GENERATION EYE CUBED TM M A K E S I T P O S S I B L E T O EVALUATE OCULAR
More informationBeginner s Guide to Vevo 2100
PN 12090 Beginner s Guide to Vevo 2100 December 2008 This Beginner s Guide to Vevo 2100 provides a quick overview of the Vevo 2100 imaging modes and provides key tips to allow users to start a study, acquire
More informationTopics in Linguistic Theory: Laboratory Phonology Spring 2007
MIT OpenCourseWare http://ocw.mit.edu 24.910 Topics in Linguistic Theory: Laboratory Phonology Spring 2007 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.
More informationGE Medical Systems Training In Partnership LOGIQ 400. Quick Start Guide A Training in Partnership Program
GE Medical Systems Training In Partnership LOGIQ 400 Quick Start Guide 2237879 100 A Training in Partnership Program TiPTM Regulatory Requirement This product complies with regulatory requirements of
More informationbelow. AOTF Module / Optics R O B U ST. R E L I A B L E. A C C U R A T E AOTF Video Adapter (Series VA310) KEY FEATURES POTENTIAL APPLICATIONS
KEY FEATURES VIS-NIR-SWIR Range 400-1700 nm Low-Power Consumption High-Switching Speed: 16,000 λ/sec Real Time Image Display Solid State / No Moving Parts High-Spectral Resolution Immune to vibration No
More informationTELEMED. Echo Wave II Software LB-2. Ultrasound Consolle. User Manual
SmartUs - ClarUs - LS128 - LS64 - Echo Blaster 64 - Echo Blaster 128 Echo Wave II Software LB-2 Ultrasound Consolle User Manual T E L E M E D Ultrasound Medical Systems Telemed Medical Systems WEB: Telemed
More informationstep into the future Ultrasound Diagnostic Imaging System
step into the future Ultrasound Diagnostic Imaging System Moving into a new era for ultrasound diagnostics. With a proven track record in medical imaging, FUJIFILM has partnered with SonoSite, an innovator
More informationKULLEGG MARIA REGINA BOYS SECONDARY MOSTA HALF-YEARLY EXAMINATIONS 2012/2013. SUBJECT: PHYSICS Form 4 TIME: 1 HR 30 MIN
KULLEGG MARIA REGINA BOYS SECONDARY MOSTA HALF-YEARLY EXAMINATIONS 2012/2013 SUBJECT: PHYSICS Form 4 TIME: 1 HR 30 MIN NAME : CLASS : INDEX NO : Track 3 Answer ALL questions in the spaces provided on the
More informationAcoustic Camera Nor848A
Product Data Acoustic Camera Nor848A Three array sizes with up to 384 microphones available Real time virtual microphone Digital microphones, no extra acquisition unit needed Intuitive software Plug and
More informationRobot-Based Solutions for NDT Inspections: Integration of Laser Ultrasonics and Air Coupled Ultrasounds for Aeronautical Components
19 th World Conference on Non-Destructive Testing 2016 Robot-Based Solutions for NDT Inspections: Integration of Laser Ultrasonics and Air Coupled Ultrasounds for Aeronautical Components Esmeralda CUEVAS
More informationSemi-Interactive Console Layout and Instructions
Semi-Interactive Console Layout and Instructions A video of the SIC accompanies these instructions, which you may refer to for further understanding. Click here to open the video link. NOTE: The images
More informationADOBE 9A Adobe After Effects 7.0 Professional ACE.
ADOBE 9A0-060 Adobe After Effects 7.0 Professional ACE http://killexams.com/exam-detail/9a0-060 QUESTION: 137 You have a composition with five layers. You want to animate a Lens Flare effect independently
More informationOptimized Design of 3D Laser Triangulation Systems
The Scan Principle of 3D Laser Triangulation Triangulation Geometry Example of Setup Z Y X Target as seen from the Camera Sensor Image of Laser Line The Scan Principle of 3D Laser Triangulation Detektion
More informationKULLEGG MARIA REGINA BOYS SECONDARY MOSTA HALF-YEARLY EXAMINATIONS 2012/2013. SUBJECT: PHYSICS Form 4 TIME: 1 HR 30 MIN NAME :
KULLEGG MARIA REGINA BOYS SECONDARY MOSTA HALF-YEARLY EXAMINATIONS 2012/2013 SUBJECT: PHYSICS Form 4 TIME: 1 HR 30 MIN NAME : CLASS : INDEX NO : Track 2 Answer ALL questions in the spaces provided on the
More informationUGEO G60 Performance in style
CT-UGEO-G60-CWW-MCI-0731-EN Performance in style 2012 Samsung Electronics All Rights Reserved. Samsung Electronics Reserves The Right To Modify The Design, Packaging, Specifications And Features Shown
More informationEcho Blaster 64/128 and LogicScan 64/128 Series Ultrasound Systems Echo Wave II Software User Manual
Echo Blaster 64/128 and LogicScan 64/128 Series Ultrasound Systems Echo Wave II Software User Manual Internet page: http://www.telemed.lt Information E-mail: info@telemed.lt Ultrasound Medical Systems
More informationAN acoustic array consists of a number of elements,
APPLICATION NOTE 1 Acoustic camera and beampattern Jørgen Grythe, Norsonic AS, Oslo, Norway Abstract The wavenumber-frequency response of an array describes the response to an arbitrary plane wave both
More informationA THREE-DIMENSIONAL PHASED ARRAY ULTRASONIC TESTING TECHNIQUE
A THREE-DIMENSIONAL PHASED ARRAY ULTRASONIC TESTING TECHNIQUE So KITAZAWA, Naoyuki KONO, Atsushi BABA and Yuji ADACHI HITACHI, Ltd., Japan Mitsuru ODAKURA HITACHI-GE Nuclear Energy, Ltd., Japan Introduction
More informationCoding and Modulation in Cameras
Mitsubishi Electric Research Laboratories Raskar 2007 Coding and Modulation in Cameras Ramesh Raskar with Ashok Veeraraghavan, Amit Agrawal, Jack Tumblin, Ankit Mohan Mitsubishi Electric Research Labs
More informationMaternal monitoring, where and when. it matters
Mobile Obstetrics Monitoring (MOM) Maternal monitoring, where and when it matters Providing expectant mothers with enhanced care to address maternal mortality is a top priority for many communities. Philips
More informationBuxton & District U3A Digital Photography Beginners Group Lesson 6: Understanding Exposure. 19 November 2013
U3A Group Lesson 6: Understanding Exposure 19 November 2013 Programme Buxton & District 19 September Exploring your camera 1 October You ve taken some pictures now what? (Viewing pictures; filing on your
More informationSDI Speed Dome Camera OSD Menu
SDI Speed Dome Camera OSD Menu Indoor/ Outdoor Ver 1.2 00P9SH720ZXSEA2 Table of Contents 1. OSD Menu Tree...4 2. Configuration Menu...7 2.1 VIDEO TYPE...8 2.2 DEFAULT CAMERA...8 2.3 BACKLIGHT...8 2.4 FOCUS...8
More informationVevo770 Ultrasound. Quick Operation Guide AIPT
Vevo770 Ultrasound Quick Operation Guide Vevo770 Probe head (707B/704) Preparation The acoustic membrane is very fragile and sensitive, it must be handled with care. The probe head should be filled with
More informationNEW! PORTABLE DIAGNOSTIC ULTRASOUND PLATFORM. Considered to be the gold standard in ophthalmic diagnostic ultrasound,
NEW! PORTABLE DIAGNOSTIC ULTRASOUND PLATFORM Considered to be the gold standard in ophthalmic diagnostic ultrasound, Eye Cubed features advanced movie mode using the fastest sampling Eye One rate available,
More informationQuick Start Guide for Shooting Video with the Panasonic GH4
Quick Start Guide for Shooting Video with the Panasonic GH4 Two options for using this camera: Option 1: The simplest method is to use the C1 Preset that has been entered into the camera. Option 2: Put
More informationIASI spectral calibration monitoring on MetOp-A and MetOp-B
IASI spectral calibration monitoring on MetOp-A and MetOp-B E. Jacquette (1), B. Tournier (2), E. Péquignot (1), J. Donnadille (2), D. Jouglet (1), V. Lonjou (1), J. Chinaud (1), C. Baque (3), L. Buffet
More informationTo Measure a Constant Velocity. Enter.
To Measure a Constant Velocity Apparatus calculator, black lead, calculator based ranger (cbr, shown), Physics application this text, the use of the program becomes second nature. At the Vernier Software
More informationSmaart v8.3 Release Overview
Smaart v8.3 Release Overview Copyright 2018 Rational Acoustics Table of Contents 8.3 Release Notes... 3 Roland OCTA-CAPTURE Control Integration... 4 Gain Tracking...4 A note about Gain Tracking accuracy...
More informationSimrad EK80. Software Release Note Introduction
Simrad EK80 Software Release Note 1.12.2 Introduction This document describes the changes introduced with the new software version. Product: EK80 Software version: 1.12.2 This software controls all functionality
More informationDANNOVO HD USB 3.0 PTZ Video Conference Camera User Manual
DANNOVO HD USB 3.0 PTZ Video Conference Camera User Manual V 2.3(ENGLISH VERSION) Please read this Manual before set up Camera, and stick to its requirements strictly for Safety reason. Suggest you save
More informationRange Sensors (time of flight) (1)
Range Sensors (time of flight) (1) Large range distance measurement -> called range sensors Range information: key element for localization and environment modeling Ultrasonic sensors, infra-red sensors
More informationChapter 2: Wave Optics
Chapter : Wave Optics P-1. We can write a plane wave with the z axis taken in the direction of the wave vector k as u(,) r t Acos tkzarg( A) As c /, T 1/ and k / we can rewrite the plane wave as t z u(,)
More informationPhysics 210 Medical Physics Midterm Exam Fall 2012 October 12, 2012
Physics 210 Medical Physics Midterm Exam Fall 2012 October 12, 2012 Name Problem 1 /32 Problem 2 /32 Problem 3 /24 Total /88 I affirm that I have carried out my academic endeavors with full academic honesty.
More informationCalibration of Recorded Digital Counts to Radiance
Graduate Radiometry, 1051-719 Laboratory No. 2 Fall Quarter, 2006 Calibration of Recorded Digital Counts to Radiance INTRODUCTION The goal of this experiment is to calibrate your digital camera, using
More informationAP Physics Problems -- Waves and Light
AP Physics Problems -- Waves and Light 1. 1975-4 (Physical Optics) a. Light of a single wavelength is incident on a single slit of width w. (w is a few wavelengths.) Sketch a graph of the intensity as
More informationACTIVITYDETECTION 2.5
ACTIVITYDETECTION 2.5 Configuration Revision 1 2018 ACIC sa/nv. All rights reserved. Document history Revision Date Comment 1 23/11/18 First revision for version 2.5 Target public This document is written
More informationEnFocus Your Upgrade Path to High Performance Intrasurgical OCT
Your Upgrade Path to High Performance Intrasurgical OCT is FDA 510(k) Cleared > Ultra HD OCT EXTEND YOUR MICROSCOPE S POTENTIAL WITH INTRASURGICAL OCT Brilliant Images, Sub-Surface Knowledge is an intrasurgical
More informationUsing the Spectrum Management Tools
CHAPTER 4 Revised: June 28, 2007, Using the Trace Window CBT 3.3 interfaces with the Cisco CMTS to provide return path spectrum analysis, which is provided in the following tasks: Trace Window To monitor
More informationGE Healthcare. Venue 40. an innovation of healthymagination
GE Healthcare Venue 40 an innovation of healthymagination Product Description Venue 40 gives physicians precision and exceptional image quality in an intuitive and affordable ultrasound system. Venue is
More informationOperation of Digital Features of HD Lenses
Operation of Digital Features of HD Lenses The zoom can be operated in the following five operation modes. 1. Operation by Zoom Seesaw Control Lever 2. QuickZoom Operation 3. Auto Cruising Zoom Operation
More informationHome Lab 7 Refraction, Ray Tracing, and Snell s Law
Home Lab Week 7 Refraction, Ray Tracing, and Snell s Law Home Lab 7 Refraction, Ray Tracing, and Snell s Law Activity 7-1: Snell s Law Objective: Verify Snell s law Materials Included: Laser pointer Cylindrical
More informationAP* Optics Free Response Questions
AP* Optics Free Response Questions 1978 Q5 MIRRORS An object 6 centimeters high is placed 30 centimeters from a concave mirror of focal length 10 centimeters as shown above. (a) On the diagram above, locate
More information10 Working With Setup Tools
10 Working With Setup Tools The Setup window allows you to set defaults and control various functions of the Terason software. The Setup window tabs display sets of controls for defaults and other settings.
More informationUniversity Physics (Prof. David Flory) Chapt_37 Monday, August 06, 2007
Name: Date: 1. If we increase the wavelength of the light used to form a double-slit diffraction pattern: A) the width of the central diffraction peak increases and the number of bright fringes within
More informationMeasurement of pinna flare angle and its effect on individualized head-related transfer functions
PROCEEDINGS of the 22 nd International Congress on Acoustics Free-Field Virtual Psychoacoustics and Hearing Impairment: Paper ICA2016-53 Measurement of pinna flare angle and its effect on individualized
More informationInstallation and Configuration Guide for Queue Management camera HIKVISION TECHNICAL SUPPORT TEAM
Installation and Configuration Guide for Queue Management camera HIKVISION TECHNICAL SUPPORT TEAM 2018-5-4 1. Environment 1 DS-2CD7126G0/L-IZS Installation and Configuration Guide for Queue Management
More informationSizing and evaluation of planar defects based on Surface Diffracted Signal Loss technique by ultrasonic phased array
Sizing and evaluation of planar defects based on Surface Diffracted Signal Loss technique by ultrasonic phased array A. Golshani ekhlas¹, E. Ginzel², M. Sorouri³ ¹Pars Leading Inspection Co, Tehran, Iran,
More informationCIS 580, Machine Perception, Spring 2015 Homework 1 Due: :59AM
CIS 580, Machine Perception, Spring 2015 Homework 1 Due: 2015.02.09. 11:59AM Instructions. Submit your answers in PDF form to Canvas. This is an individual assignment. 1 Camera Model, Focal Length and
More information10 Working With Setup Tools
10 Working With Setup Tools Setup Tools Overview The Setup window allows you to set defaults and control various functions of the Terason software. The Setup window tabs display sets of controls for defaults
More informationDevelopment and validation of a short-lag spatial coherence theory for photoacoustic imaging
Development and validation of a short-lag spatial coherence theory for photoacoustic imaging Michelle T. Graham 1 and Muyinatu A. Lediju Bell 1,2 1 Department of Electrical and Computer Engineering, Johns
More information1 About the Terason Ultrasound System
1 About the Terason Ultrasound System Overview The Terason Ultrasound System is an easy-to-use, portable ultrasound system that produces high resolution images. The microminiaturized ultrasound system
More informationResona 7. New Waves in Ultrasound Innovation. Premium Ultrasound System
Resona 7 Premium Ultrasound System New Waves in Ultrasound Innovation With over 20 years of experience, Mindray hosts a wide range of ultrasound imaging solutions including cart-based and portable systems.
More informationFAST AND ACCURATE TRANSIENT ULTRASOUND PROPAGATION AND B-MODE IMAGING SIMULATION METHODS. Yi Zhu
FAST AND ACCURATE TRANSIENT ULTRASOUND PROPAGATION AND B-MODE IMAGING SIMULATION METHODS By Yi Zhu A THESIS Submitted to Michigan State University in partial fulfillment of the requirements for the degree
More informationProduct information. Hi-Tech Electronics Pte Ltd
Product information Introduction TEMA Motion is the world leading software for advanced motion analysis. Starting with digital image sequences the operator uses TEMA Motion to track objects in images,
More informationUSB PTZ Camera Controller User Manual
USB PTZ Camera Controller User Manual Version: 03 Issue Date: JUN 12, 2017 Copyright Information Copyrights Lumens Digital Optics Inc., All rights reserved. Lumens is a registered trademark of Lumens Digital
More informationSingle Slit Diffraction *
OpenStax-CNX module: m42515 1 Single Slit Diffraction * OpenStax This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 3.0 Discuss the single slit diraction
More informationResona 7. Premium Ultrasound System. New Waves in Ultrasound Innovation
Resona 7 Premium Ultrasound System New Waves in Ultrasound Innovation New Waves in Ultrasound Innovation Since the company was founded, Mindray is continuously exploring new ways to improve diagnostic
More informationLenses. Digital Cameras. Lenses. Lenses. Light focused by the lens. Lenses. Aperture Shutter Speed Holding a Camera Steady Shooting Modes ISO
Lenses Digital Cameras Light focused by the lens Lenses Aperture Shutter Speed Holding a Camera Steady Shooting Modes ISO Lenses Lenses Positive lens Converging lens Focuses image What is difference in
More informationValidation of aspects of BeamTool
Vol.19 No.05 (May 2014) - The e-journal of Nondestructive Testing - ISSN 1435-4934 www.ndt.net/?id=15673 Validation of aspects of BeamTool E. GINZEL 1, M. MATHESON 2, P. CYR 2, B. BROWN 2 1 Materials Research
More informationHow to Adjust the Back Focus for TCM-5311
How to Adjust the Back Focus for TCM-5311 Category Type Camera Model Firmware Version Configuration Note TCM-5311 N/A Publish Date 2009/09/03 Last Review 2009/09/03 Knowledge Type Function Type Video Quality
More informationMindray. Hand-carried Diagnostic Ultrasound System. Expanding the Envelope of Performance and Flexibility
M7 Mindray M7 Hand-carried Diagnostic Ultrasound System Expanding the Envelope of Performance and Flexibility M7 Diagnostic Ultrasound System Equipped for Quality Radiology Cardiology Outpatient Emergency
More informationAn Intuitive Explanation of Fourier Theory
An Intuitive Explanation of Fourier Theory Steven Lehar slehar@cns.bu.edu Fourier theory is pretty complicated mathematically. But there are some beautifully simple holistic concepts behind Fourier theory
More informationCONTENTS. Quick Start Guide V1.0
Quick Start Guide CONTENTS 1 Introduction... 2 2 What s in the box?... 3 3 Using your buzzard multispectral sensor... 4 3.1 Overview... 4 3.2 Connecting the power / remote trigger cable... 5 3.3 Attaching
More informationThe Most User-Friendly 3D scanner
The Most User-Friendly 3D scanner The Solutionix C500 is optimized for scanning small- to medium-sized objects. With dual 5.0MP cameras, the C500 provides excellent data quality at a high resolution. In
More informationCAUTION: NEVER LOOK DIRECTLY INTO THE LASER BEAM.
LABORATORY 12 PHYSICAL OPTICS I: INTERFERENCE AND DIFFRACTION Objectives To be able to explain demonstrate understanding of the dependence of a double slit interference pattern on slit width, slit separation
More informationSupplied Accessories.
Precautions. Safety Tips. Please read this manual carefully before using the camera. Avoid damage from stress, violent vibration or liquid intrusion during transportation, storage or installation. Take
More informationCinCam CCD - Technical Data -
- Technical Data - SENSOR DATA CCD-1201 CCD-1301 CCD-2301 CCD-2302 Format: 1/2 1/3 2/3 2/3 Active area: 6.5mm x 4.8mm 4.8mm x 3.6mm 9.0mm x 6.7mm 8.5mm x 7.1mm Number of pixel: 1388 x 1038 (1.4MPixel)
More informationFundamentals of CT imaging
SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform
More informationIntroduction to Raman spectroscopy measurement data processing using Igor Pro
Introduction to Raman spectroscopy measurement data processing using Igor Pro This introduction is intended to minimally guide beginners to processing Raman spectroscopy measurement data, which includes
More informationBLK-HD33IR HD-SDI Digital Video IR LED Dome Camera User Manual
BLK-HD33IR HD-SDI Digital Video IR LED Dome Camera User Manual Please read this manual before using your camera, and always follow the instructions for safety and proper use. Save this manual for future
More information1.1 The HeNe and Fourier Lab CCD Camera
Chapter 1 CCD Camera Operation 1.1 The HeNe and Fourier Lab CCD Camera For several experiments in this course you will use the CCD cameras to capture images or movies. Make sure to copy all files to your
More informationApplying Hounsfield unit density calibration in SkyScan CT-analyser
1 Bruker-microCT Method note Applying Hounsfield unit density calibration in SkyScan CT-analyser Hounsfield units (HU) are a standard unit of x-ray CT density, in which air and water are ascribed values
More informationLaserDec CL - Product Description -
CO 2 LASER BEAM PROFILER LaserDec CL - Product Description - The high performance LaserDec system is based on industry s unique imaging technique. It is designed for monitoring high-power CO 2 -lasers
More information10/16/14. Ultrasound Physics & Instrumentation 5 th Edition. License Agreement. Chapter Outline. All Copyright Laws Apply.
Ultrasound Physics & Instrumentation 5 th Edition Companion Presentation Frank R. Miele Pegasus Lectures, Inc. License Agreement This presentation is the sole property of Pegasus Lectures, Inc. No part
More information1D internal multiple prediction in a multidimensional world: errors and recommendations
1D internal multiple prediction 1D internal multiple prediction in a multidimensional world: errors and recommendations Pan Pan and Kris Innanen ABSTRACT Internal multiples are more difficult to estimate
More informationWe ll go over a few simple tips for digital photographers.
Jim West We ll go over a few simple tips for digital photographers. We ll spend a fair amount of time learning the basics of photography and how to use your camera beyond the basic full automatic mode.
More informationJohn R. Mandeville Senior Consultant NDICS, Norwich, CT Jesse A. Skramstad President - NDT Solutions Inc., New Richmond, WI
Enhanced Defect Detection on Aircraft Structures Automatic Flaw Classification Software (AFCS) John R. Mandeville Senior Consultant NDICS, Norwich, CT Jesse A. Skramstad President - NDT Solutions Inc.,
More informationA Level. A Level Physics. WAVES: Combining Waves (Answers) Edexcel. Name: Total Marks: /30
Visit http://www.mathsmadeeasy.co.uk/ for more fantastic resources. Edexcel A Level A Level Physics WAVES: Combining Waves (Answers) Name: Total Marks: /30 Maths Made Easy Complete Tuition Ltd 2017 1.
More informationReflectance & Lighting
Reflectance & Lighting Computer Vision I CSE5A Lecture 6 Last lecture in a nutshell Need for lenses (blur from pinhole) Thin lens equation Distortion and aberrations Vignetting CS5A, Winter 007 Computer
More informationVEVO 1100 Sole Source Specifications
VEVO 1100 Sole Source Specifications Table of Contents 1.SYSTEM... 2 1-1 Licensing for the User... 2 1-2 Safety... 2 1-3 Portability... 2 2.TRANSDUCERS... 2 2-1 MicroScan transducers... 2 3. IMAGING STATION...
More informationComputed tomography of simple objects. Related topics. Principle. Equipment TEP Beam hardening, artefacts, and algorithms
Related topics Beam hardening, artefacts, and algorithms Principle The CT principle is demonstrated with the aid of simple objects. In the case of very simple targets, only a few images need to be taken
More informationSupplied Accessories.
Precautions. Safety Tips. Please read this manual carefully before using the camera. Avoid damage from stress, violent vibration or liquid intrusion during transportation, storage or installation. Take
More information5.2. The Sine Function and the Cosine Function. Investigate A
5.2 The Sine Function and the Cosine Function What do an oceanographer, a stock analyst, an audio engineer, and a musician playing electronic instruments have in common? They all deal with periodic patterns.
More informationSystemVue 2011 Fundamentals (version ) LAB EXERCISE 1. SystemVue Basics
SystemVue 2011 Fundamentals (version 1-2012) LAB EXERCISE 1 SystemVue Basics This lab exercise introduces the basic use model of SystemVue, including using the built-in templates, running analyses and
More information