Noise weighting with an exponent for transmission CT

Size: px
Start display at page:

Download "Noise weighting with an exponent for transmission CT"

Transcription

1 doi: / /2/4/ RECEIVED 13 January 2016 REVISED 4 June 2016 ACCEPTED FOR PUBLICATION 21 June 2016 PUBLISHED 27 July 2016 PAPER Noise weighting with an exponent for transmission CT Gengsheng L Zeng 1,2 and Wenli Wang 3 1 Department of Engineering, Weber State University, Ogden, UT 84408, USA 2 Department of Radiology, University of Utah, Salt Lake City, UT 84108, USA 3 Toshiba Medical Research Institute USA, Inc. Vernon Hills, IL 60061, USA larryzeng@weber.edu Keywords: computed tomography, iterative reconstruction, filtered backprojection algorithm, noise weighted image reconstruction Abstract It is widely believed that the correct weighting function is the reciprocal of the noise variance of the associated measurement. Many researchers are making great efforts to find the accurate variance for the measurements for imaging systems so that they can hopefully achieve an optimal reconstruction. An optimal solution in the context of this paper is referred to as the image that reaches optimum according to a criterion or criteria among a group of candidates, regardless how the images in the group are obtained. This optimal solution is not a theoretical concept, but is simply the best of the bunch. The goal of the paper is to investigate how the weighting function affects the image noise when the image contrast is pre-specified in an iterative algorithm for x-ray CT. This paper makes some interesting observations: there is no universal optimal weighting function. The noise weighting function can introduce artifacts. The optimal noise weighting varies with the object to be reconstructed and targeted image contrast in an iterative image reconstruction algorithm and in a filtered backprojection algorithm that incorporates the projection noise. It is suggested that an exponent be used in the weighting function so that the artifacts caused by the weighting function can be reduced. 1. Introduction One of the advantages of using iterative algorithms to reconstruct a tomographic image is the ability to model and suppress the measurement noise (Kuhl and Edwards 1963, Shepp and Vardi 1982, Langer and Carson 1984, Fessler 1994). Recently we have shown that the filtered backprojection (FBP) can be extended to model and suppress the measurement noise too (Zeng and Zamyatin 2013, Zeng 2014). In all these algorithms, the noise-control weighting function is normally set up as the reciprocal of the noise variance associated with the measurement (Geman and McClure 1987). This weighting function assignment is supported by the general maximum likelihood theory, and its fundamental principle is explained intuitively below. When the measurements are over-determined, there are more redundant measurements than the number of the unknowns. Figure 1 shows a simple example with two unknowns x 1 and x 2 as well as three measurements. In the x 1 x 2 solution space, the three measurements are represented as three lines. If the measurements are consistent (i.e., noiseless), the three lines should meet at a single point in the x 1 x 2 solution space and this point is the unique solution. Due to noise, the three measurements are not consistent and the solution does not exist. A maximum likelihood method selects an approximate solution that is closer to less noisy measurements and farther away from noisier measurements. If the number of unknowns is the same (or almost the same) as the number of the measurements, a unique solution can always be obtained regardless whether the measurements are noisy or not. Figure 2 shows a simple case of two unknowns (x 1 and x 2 ) and two noisy measurements. The unique solution in this case is most likely not the true solution because the measurements are noisy and deviate from the true values. This wrong, noisy, and unique solution can be obtained by using an iterative algorithm using a large enough iteration number so that the algorithm converges. This wrong, noisy, and unique solution can also be obtained by an analytical algorithm such as the 2016 IOP Publishing Ltd

2 Figure 1. An over-determined system with twp unknowns and three measurements. (a) The measurements are noiseless and the system is consistent. A unique solution exists. (b) The measurements are noisy and the system is inconsistent. An approximate solution can be determined by the relative noise variance of the measurements. Figure 2. An well-determined system with two unknowns and two noisy measurements. (a) Even though the measurements are noisy, the system is consistent and a unique noisy solution exists. (b) If an iterative algorithm is stopped before convergence, many pseudo solutions can be obtained. These pseudo solutions are influenced by the weighting function, and may be closer to the true solution than the unique noisy solution. FBP algorithm which is popular in image reconstruction community (Radon 1917, Bracewell 1956, Shepp and Logan 1974). In terms of image reconstruction, a wrong solution here is a solution to the system of linear equations, which are corrupted with noise. If the system matrix is invertible, the solution exists and is unique. Due the noise, this solution is no longer the same as the original image. If an iterative algorithm is stopped before convergence, a sequence of pseudo solutions is obtained. A pseudo solution is any intermediate solution at any iteration in an iterative algorithm, regardless whether the algorithm has converged or not. These pseudo solutions are influenced by the weighting function, and may be closer to the true solution than the unique noisy solution (see figure 2(b)). It is important to notice that the weighting function is effective before the algorithm converges. In other words, the weighting function determines the path of the algorithm towards the converged noisy solution. Due to the involvement of the weighting function, some pseudo solutions (i.e., the intermediate solutions) may be closer to the true solution than the ultimate noisy solution. We argue that the situation described in figure 2 applies to most tomographic medical imaging systems such as x-ray computed tomography (CT), positron emission tomography (PET) and single photon emission CT (SPECT). This is the reason that when an iterative algorithm is used to reconstruct an image, the algorithm must stop early otherwise a very noisy and useless image will result. This paper considers the iterative algorithm that stops before convergence, and investigates how the weighting function influences the early solutions. The result of this paper will apply to the extend FBP algorithm that models and suppresses the noise. 2

3 This paper introduces a parameter in the weighting function; this parameter should be 1 according to theoretical derivation. Our motivation of introducing this parameter is as follows. The theoretically derived weighting function assumes the convergence of the algorithm. In practice, the iterative algorithm is terminated before reaching the convergence. Therefore, this small perturbation of early stopping may require a small deviation of the parameter from its theoretical value in order to obtain a better image. Here better is task dependent. The task in our paper is to find a solution with the smallest noise standard deviation for a given image contrast. As expected, this deviation from its theoretical value 1 is small in numerical studies. 2. Methods An iterative algorithm is used to minimize or maximize an objective function. The objective function usually consists of two parts: the data fidelity term and the regularization term. One popular way to do regularization is to control the number of iterations (i.e., the stopping rules). Our stopping rule is that the iteration terminates when the image contrast reaches a pre-specified value. When the iteration stops, the reconstructed image depends on the noise weighting factors. The state-of-the-art iterative image reconstruction algorithms use a fixed weighting factor that is the reciprocal of the image variance. This paper points out that if a control parameter is introduced to the weighting factor, the resultant reconstruction can be further improved in terms of the reduction of noise The gradient descent iterative algorithm The gradient descent iterative algorithm considered in this paper has the following form (Elbakri and Fessler 2002) X å ( + ) ( ) i k 1 = Xi k j - a å () ( å jn n k - j) Ajiwj A X P n A w A j ji j å n jn, () 1 ( ) where X k i is the ith image pixel at the kth iteration, P j is the jth line-integral (ray-sum) measurement value, A ji is the contribution of the ith image pixel to the jth measurement, w j is the weighting factor for the jth measurement, and α is a constant to prevent the algorithm from divergence. The purpose of the denominator åjaw ji j ånajn is to normalize the step size so that the step size is independent from the system matrix A and the weighting function w j. Thus the value of α is always 1 in (Elbakri and Fessler 2002). However, this scaled step size does not always work and the algorithm may diverge for many situations. We set α to 0.1 in this paper. The summation over the index n is the projector and the summation over the index j is the backprojector. Figure 3. A computer generated torso phantom is used for transmission CT data generation. The linear attenuation coefficients are labeled in the unit of per mm. Two regions of interest (ROIs) are defined for image evaluation and for the image contrast calculation The noise-weighted FBP algorithm A noise-weighted FBP algorithm was recently developed to model and suppress noise (Zeng and Zamyatin 2013, Zeng 2014). This algorithm emulates the gradient descent algorithm and contains a control index k, which is similar to the iteration number in an iterative algorithm. This noise-weighted FBP algorithm is almost the same as the conventional FBP algorithm, except for the ramp filter. In a conventional FBP algorithm, the ramp filter is ω,whereω is the frequency. In the noiseweighted FBP algorithm, the ramp filter is modified by a window function and is expressed as a k w j H ( w) = w, withw ¹ 0 w and H () 0 = 0, () 2 where α is a positive constant to prevent the algorithm from divergence. In this paper, α is set to 0.1. The implementation of (2) is in the Fourier domain of the sinogram. The weighting factors w j in (2) for all projection bins are quantized into 11 discrete values, and each of these 11 quantized weighting factors produces a filtered sinogram. A combined sinogram from these filtered sinograms is formed pointby-point according to the variance of the original sinogram. The details of the implementation can be found in (Zeng and Zamyatin 2013) Data generation and noise model The computer simulations in this paper are based on a scaled-down x-ray CT fan-beam imaging geometry with a curved detector. The image array was , the pixel size was 1.52 mm 1.52 mm, the number of views was 400 over 360, the number of detection channels was 400, and the focal length was 240 mm. The x-ray source flux had I 0 = 10 4 counts, which corresponds to a low-dose imaging setup. The phantom shown in figure 3 is 355 mm 187 mm 3

4 Figure 4. Iterative reconstructions with various parameter γ and the normalized standard deviation in ROI 2 for Case (A): the default phantom as shown in figure 3 is used, and I 0 = The target contrast is ellipse with water background (μ = 0.02 mm 1 ), three high contrast regions (μ = mm 1 ) with diameter 48 mm, two low-contrast regions (μ = mm) with diameter 36 mm, surrounded by outer layers of fat (μ = mm 1 ) and skin (μ = mm 1 ), in which ROI 1 (high contrast object) and ROI 2 (water background) are used to evaluate the image quality. The projection data were generated in the prelog format with the Poisson noise model and Gaussian electronic noise. No beam-hardening effects are simulated. The pre-log data were then converted into the post-log data for image reconstruction. If a pre-log data is less than one, it is changed to one before taking logarithm to avoid negative post-log sinogram values. The iterative algorithm was implemented according to (1), and the iterative algorithm stops when a prespecified image contrast is reached. This value was set up as 0.53 and 0.57 (where the true contrast is 0.6). The reconstructed images are compared with the normalized standard deviation value in ROI 2. The normalized standard deviation value is the standard deviation value divided by the mean value. The phantom is shown in figure 3. A popular approach to assigning the weighting factor is to let w j be the reciprocal of the noise variance of the ray-sum measurement. This approach is justified by using the likelihood function as the objective function for an optimization problem (Aitken 1935). The philosophy is that we should trust the less noisy measurements more than noisier measurements. 4

5 Biomed. Phys. Eng. Express 2 (2016) Figure 5. Iterative reconstructions with various parameter γ and the normalized standard deviation in ROI 2 for Case (B): the phantom is changed to an obese version, and I0 = 104. The target contrast is In x-ray CT imaging, the noise in measured transmission data can be approximately described by a Poisson distribution, i.e., var(i) I, where I denotes an x-ray intensity transmission measurement (Hsieh 1998). If the additive electronic noise σ2 is also considered for the detection system, the total variance of the pre-log transmission measurement is then var (I) I + σ2. Here s 2 = 6.32 was chosen in our simulated low-count x-ray CT data generation. After log conversion the noise variance is described by var(i)/i2 (I + σ2)/ I2. If the x-ray source flux I0 is stable and consistent, measurement intensity I can be written as I = I0 exp (-p ) according to Beer s law, where p is the ray-sum or the total attenuation along the ray. The variance of post-log measurement p can thus be expressed as 5 var ( p )» I0 exp ( - p ) + s 2 I + s2 =. I2 I02 exp ( - 2p ) (4) According to our previous experiments, the image noise was not reduced by including the electronic noise variance σ2 in the weighting factors because the value σ2 is the same for every measurement. Some other researchers (Sauer and Bouman 1993) also choose not to include the electronic noise variance in the weighting function. We suggest discarding the components that are common to all measurements from the weighting function. Even though s 2 = 6.32 was used in data generation, we assume s 2 = 0 during image reconstruction. The conventional weighing factor is inversely proportional to the noise variance. Thus, the conventional weighing factor can be

6 Figure 6. Iterative reconstructions with various parameter γ and the normalized standard deviation in ROI 2 for Case (C): the phantom is changed to a slightly thinner version, and I 0 = The target contrast is assigned as w = exp (-p). ( 5) This paper introduces a new parameter γ to the weighting function: w = exp (-gp). ( 6) In our implementation of the weighting function (6) the post-log data p is first smoothed by a five-point running average low-pass filter in the detector channel direction, in order to reduce the noise propagation from the weighting function to the reconstruction. However, the post-log projections used in (1) are not pre-filtered. The image quality is evaluated by the normalized standard deviation in ROI 2. The normalized standard deviation is defined as the ratio of the standard deviation and the mean value. Performing systematic comparison studies with scanned data is difficult because the ground true is unknown. However, we did apply our method of using an artificial parameter γ in real patient studies, see (Zeng and Zamyatin 2013), where γ = 0.3 was used. 3. Results Resolution is closely related to contrast. Usually higher resolution leads to larger noise standard deviation. If the resolution is fixed, a better reconstruction method can produce less noisy images. This is the whole point of the paper. By fixing the image contrast, we vary a 6

7 Figure 7. Iterative reconstructions with various parameter γ and the normalized standard deviation in ROI 2 for Case (D): the phantom is changed to a very thin version, and I 0 = The target contrast is control parameter in the weighting function and observe how the image noise and artifacts change accordingly Iterative reconstructions The effects of the exponent γ are illustrated with the following variations. A. The default phantom as shown in figure 3 is used, and I 0 = The target contrast is B. The phantom is changed to an obese version. C. The phantom is changed to a slightly thinner version. D. The phantom is changed to a very thin version. E. Same as (A), but the target contrast is changed to F. Same as (A), but I 0 is increased to The computer simulation results are summarized in figures 4 9. In all these figures, one can make the following observations. When γ is small, there are severe noise induced streaking artifacts. As the value of γ increases, the streaking artifacts are gradually suppressed. After passing the optimal value of γ, a larger value of γ causes severe low-frequency shadowing artifacts. Sometimes these low-frequency shadowing artifacts may be mistaken as the beam-hardening artifacts. The cause of the shadowing artifacts is the extremely small values of the weighting factors w j, due to the large p values. The shadowing artifacts are caused by 7

8 Figure 8. Iterative reconstructions with various parameter γ and the normalized standard deviation in ROI 2 for Case (E): the phantom is as in figure 3, and I 0 = The target contrast is changed to the improper weighting factors, and are not caused by discarding negative sinogram values. The shadowing artifacts can appear with ideally generated, i.e., noiseless (i.e., I 0 ), line-integral sinogram when improper weighting factors are used during image reconstruction. When the weighting factors w j are too small, some important tomographic information is neglected, resulting in limited data artifacts (similar to metal artifacts). The optimal parameter γ depends on the object shape and image contrast (maybe less on dose I 0 ). The main idea of using a new exponent parameter γ is to reduce those large p values to some extend so that the over-suppressed tomographic information can be available for image reconstruction FBP reconstructions Image reconstruction results using the noise-weighted FBP algorithm with a modified ramp filter (2) are listed in figure 10. The FBP results have the same trend as that in the iterative reconstruction results. For a small parameter γ, we see streaking artifacts. The streaking artifacts are suppressed with a larger γ. However, when the parameter γ is too large, the low-frequency shadowing artifacts appear. An optimal parameter γ should be used. Thus only three representative images are shown. Similar to the iterative algorithm s iteration number, the parameter k is selected when the pre-specified image contrast is reached. We must point out that the parameter k in the FBP algorithm and the number of iteration k in the iterative algorithm, in general, are not the same; they are closely 8

9 Figure 9. Iterative reconstructions with various parameter γ and the normalized standard deviation in ROI 2 for Case (F): the phantom is in figure 3, but I 0 is increased to The target contrast is Figure 10. FBP reconstructions with various parameter γ and the normalized standard deviation in ROI 2 for Case (C): the phantom is the slightly thinner version as in figure 6, and I 0 = The target contrast is The values of parameter k are 9100, , and , respectively. 9

10 related. Similarly, the γ values in both algorithms are not thesame,buthavethesametrend. All images are displayed with the same linear gray scale. 4. Conclusions This paper uses an exact known noise model to investigate the effects of the weighting function. This paper suggests that a weighting function that is a power function of the reciprocal of the noise variance, w = 1/ variance g, should be used. When γ = 0, the weighting function is a constant without any variation. A larger γ gives a larger variation of the weighting function. When γ = 1, the weighting function is the so-called correct weighting which is widely used among researchers. Both an iterative gradient descent algorithm and an analytic noise-weighted FBP algorithm are used for the investigation. In order for the weighting function to be effective, the iteration number k in the iterative algorithm or the noise control parameter k in the FBP algorithm must be small enough so that the algorithm is not converged yet. Our computer simulations show that the optimal weighting scheme depends on the object and the pre-specified image contrast. Therefore, there is no universal optimal weighting function. The so-called correct weighting function is sub-optimal. For a given object, the optimal weighting function is image contrast dependent, which in turn will be determined by the noise. The power function weighting function suggested in this paper is only one of many options that can help to achieve a desired image quality. References Aitken A C 1935 On least squares and linear combinations of observations Proc. R. Soc. Edinburgh Bracewell R N 1956 Strip integration in radio astronomy Aus. J. Phys Elbakri A and Fessler J A 2002 Statistical image reconstruction for polyenergetic x-ray computed tomography IEEE Trans. Med. Imaging Fessler J A 1994 Penalized weighted least-squares image reconstruction for positron emission tomography IEEE Trans. Med. Imaging Geman S and McClure D E 1987 Statistical methods for tomographic image reconstruction Bull. Int. Stat. Inst. LII Hsieh J 1998 Adaptive streak artifact reduction in computed tomography resulting from excessive x-ray photon noise Med. Phys Kuhl D E and Edwards R Q 1963 Image separation radioisotope scanning Radiology Langer K and Carson R 1984 EM reconstruction algorithms for emission and transmission tomography J. Comput. Assist. Tomogr Radon J 1917 Über die Bestimmung von Funktionen durch ihre integralwerte längs gewisser Mannigfaltigkeiten Ber. Verh. Sächs. Akad. Wiss. Leipzig, Math. Nat. K Sauer K and Bouman C 1993 A local update strategy for iterative reconstruction projections IEEE Trans. Signal Process Shepp L A and Logan B F 1974 The Fourier reconstruction of a head section IEEE Trans. Nucl. Sci. NS Shepp L A and Vardi Y 1982 Maximum likelihood reconstruction for emission tomography IEEE Trans. Med. Imaging Zeng G L and Zamyatin A 2013 A filtered backprojection algorithm with ray-by-ray noise weighting Med. Phys Zeng G L 2014 Model-based filtered backprojection algorithm: a tutorial Biomed. Eng. Lett

Revisit of the Ramp Filter

Revisit of the Ramp Filter IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 62, NO. 1, FEBRUARY 2015 131 Revisit of the Ramp Filter Gengsheng L. Zeng, Fellow, IEEE Abstract An important part of the filtered backprojection (FBP) algorithm

More information

Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm

Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm 548 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 5, MAY 2000 Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm Gengsheng L. Zeng*, Member, IEEE, and Grant T. Gullberg,

More information

Reconstruction of Tomographic Images From Limited Projections Using TVcim-p Algorithm

Reconstruction of Tomographic Images From Limited Projections Using TVcim-p Algorithm Reconstruction of Tomographic Images From Limited Projections Using TVcim-p Algorithm ABDESSALEM BENAMMAR, AICHA ALLAG and REDOUANE DRAI Research Center in Industrial Technologies (CRTI), P.O.Box 64, Cheraga

More information

Total Variation and Tomographic Imaging from Projections

Total Variation and Tomographic Imaging from Projections Total Variation and Tomographic Imaging from Projections Per Christian Hansen & Jakob Heide Jørgensen Technical University of Denmark Joint work with Dr. Emil Sidky & Prof. Xiaochuan Pan Univ. of Chicago

More information

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004 Spiral ASSR Std p = 1.0 Spiral EPBP Std p = 1.0 Kachelrieß et al., Med. Phys. 31(6): 1623-1641, 2004 256 slices (0/300) Advantages of Cone-Beam Spiral CT Image quality nearly independent of pitch Increase

More information

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam Home Search Collections Journals About Contact us My IOPscience Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam projections This content has been downloaded from IOPscience.

More information

Gengsheng Lawrence Zeng. Medical Image Reconstruction. A Conceptual Tutorial

Gengsheng Lawrence Zeng. Medical Image Reconstruction. A Conceptual Tutorial Gengsheng Lawrence Zeng Medical Image Reconstruction A Conceptual Tutorial Gengsheng Lawrence Zeng Medical Image Reconstruction A Conceptual Tutorial With 163 Figures Author Prof. Dr. Gengsheng Lawrence

More information

Reconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization

Reconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization 1 Reconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization T. Humphries and A. Faridani Abstract Recent work in CT image reconstruction has seen increasing interest

More information

Medical Image Reconstruction Term II 2012 Topic 6: Tomography

Medical Image Reconstruction Term II 2012 Topic 6: Tomography Medical Image Reconstruction Term II 2012 Topic 6: Tomography Professor Yasser Mostafa Kadah Tomography The Greek word tomos means a section, a slice, or a cut. Tomography is the process of imaging a cross

More information

AN ELLIPTICAL ORBIT BACKPROJECTION FILTERING ALGORITHM FOR SPECT

AN ELLIPTICAL ORBIT BACKPROJECTION FILTERING ALGORITHM FOR SPECT 1102 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 40, NO. 4, AUGUST 1993 AN ELLIPTICAL ORBIT BACKPROJECTION FILTERING ALGORITHM FOR SPECT Grant. T. Gullberg and Gengsheng L. Zeng, Department of Radiology,

More information

A Weighted Least Squares PET Image Reconstruction Method Using Iterative Coordinate Descent Algorithms

A Weighted Least Squares PET Image Reconstruction Method Using Iterative Coordinate Descent Algorithms A Weighted Least Squares PET Image Reconstruction Method Using Iterative Coordinate Descent Algorithms Hongqing Zhu, Huazhong Shu, Jian Zhou and Limin Luo Department of Biological Science and Medical Engineering,

More information

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University of Michigan Paul E. Kinahan Radiology Department,

More information

Cover Page. The handle holds various files of this Leiden University dissertation

Cover Page. The handle   holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/887/48289 holds various files of this Leiden University dissertation Author: Plantagie, L. Title: Algebraic filters for filtered backprojection Issue Date: 207-04-3

More information

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and

More information

THE FAN-BEAM scan for rapid data acquisition has

THE FAN-BEAM scan for rapid data acquisition has 190 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 26, NO. 2, FEBRUARY 2007 Hilbert Transform Based FBP Algorithm for Fan-Beam CT Full Partial Scans Jiangsheng You*, Member, IEEE, Gengsheng L. Zeng, Senior

More information

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Sensors & Transducers 2013 by IFSA http://www.sensorsportal.com Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Shinji Ohyama, Masato Mukouyama Graduate School

More information

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig

More information

CENTER FOR ADVANCED COMPUTATION

CENTER FOR ADVANCED COMPUTATION CENTER FOR ADVANCED COMPUTATION MODIFIED EXPECTATION MAXIMIZATION ALGORITHM FOR TRANSMISSION TOMOGRAPHY PATIWAT KAMONPET VENKATARAMA KRISHNAN Technical Report CAC T94-01 January 28, 1994 Parts of this

More information

Radon Transform and Filtered Backprojection

Radon Transform and Filtered Backprojection Radon Transform and Filtered Backprojection Jørgen Arendt Jensen October 13, 2016 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering Center for Fast Ultrasound Imaging Department

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

Algebraic Iterative Methods for Computed Tomography

Algebraic Iterative Methods for Computed Tomography Algebraic Iterative Methods for Computed Tomography Per Christian Hansen DTU Compute Department of Applied Mathematics and Computer Science Technical University of Denmark Per Christian Hansen Algebraic

More information

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy T. Rohlfing a, D. Zerfowski b, J. Beier a, P. Wust a, N. Hosten a, R. Felix a a Department of

More information

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION 2009 IEEE Nuclear Science Symposium Conference Record M03-7 FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION Wonseok Huh, Jeffrey A. Fessler, Adam M. Alessio, and Paul E. Kinahan Department

More information

Tomographic Image Reconstruction in Noisy and Limited Data Settings.

Tomographic Image Reconstruction in Noisy and Limited Data Settings. Tomographic Image Reconstruction in Noisy and Limited Data Settings. Syed Tabish Abbas International Institute of Information Technology, Hyderabad syed.abbas@research.iiit.ac.in July 1, 2016 Tabish (IIIT-H)

More information

THE general task in emission computed tomography (CT)

THE general task in emission computed tomography (CT) 228 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 17, NO. 2, APRIL 1998 List-Mode Likelihood: EM Algorithm and Image Quality Estimation Demonstrated on 2-D PET Lucas Parra* and Harrison H. Barrett, Member,

More information

Reconstruction from Projections

Reconstruction from Projections Reconstruction from Projections M.C. Villa Uriol Computational Imaging Lab email: cruz.villa@upf.edu web: http://www.cilab.upf.edu Based on SPECT reconstruction Martin Šámal Charles University Prague,

More information

Attenuation map reconstruction from TOF PET data

Attenuation map reconstruction from TOF PET data Attenuation map reconstruction from TOF PET data Qingsong Yang, Wenxiang Cong, Ge Wang* Department of Biomedical Engineering, Rensselaer Polytechnic Institute, Troy, NY 80, USA *Ge Wang (ge-wang@ieee.org)

More information

SPECT reconstruction

SPECT reconstruction Regional Training Workshop Advanced Image Processing of SPECT Studies Tygerberg Hospital, 19-23 April 2004 SPECT reconstruction Martin Šámal Charles University Prague, Czech Republic samal@cesnet.cz Tomography

More information

Splitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction

Splitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction Splitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction Hung Nien and Jeffrey A. Fessler Department of Electrical Engineering and Computer Science University of Michigan, Ann

More information

Simplified statistical image reconstruction algorithm for polyenergetic X-ray CT. y i Poisson I i (E)e R } where, b i

Simplified statistical image reconstruction algorithm for polyenergetic X-ray CT. y i Poisson I i (E)e R } where, b i implified statistical image reconstruction algorithm for polyenergetic X-ray C omesh rivastava, tudent member, IEEE, and Jeffrey A. Fessler, enior member, IEEE Abstract In X-ray computed tomography (C),

More information

Reconstruction in CT and relation to other imaging modalities

Reconstruction in CT and relation to other imaging modalities Reconstruction in CT and relation to other imaging modalities Jørgen Arendt Jensen November 1, 2017 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering Center for Fast Ultrasound

More information

Axial block coordinate descent (ABCD) algorithm for X-ray CT image reconstruction

Axial block coordinate descent (ABCD) algorithm for X-ray CT image reconstruction Axial block coordinate descent (ABCD) algorithm for X-ray CT image reconstruction Jeffrey A. Fessler and Donghwan Kim EECS Department University of Michigan Fully 3D Image Reconstruction Conference July

More information

A discrete convolution kernel for No-DC MRI

A discrete convolution kernel for No-DC MRI Inverse Problems Inverse Problems 3 (5) 856 (pp) doi:.88/66-56/3/8/856 A discrete convolution kernel for No-DC MRI Gengsheng L Zeng, and Ya Li 3 Department of Engineering, Weber State University, Ogden,

More information

PET image reconstruction algorithms based on maximum

PET image reconstruction algorithms based on maximum IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 18, NO. 4, APRIL 1999 293 A Theoretical Study of the Contrast Recovery and Variance of MAP Reconstructions from PET Data Jinyi Qi,* Member, IEEE, and Richard

More information

Adaptive algebraic reconstruction technique

Adaptive algebraic reconstruction technique Adaptive algebraic reconstruction technique Wenkai Lua) Department of Automation, Key State Lab of Intelligent Technology and System, Tsinghua University, Beijing 10084, People s Republic of China Fang-Fang

More information

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy.

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy. Modeling the Distance-Dependent Blurring in Transmission Imaging in the Ordered-Subset Transmission (OSTR) Algorithm by Using an Unmatched Projector/Backprojector Pair B. Feng, Member, IEEE, M. A. King,

More information

Non-Stationary CT Image Noise Spectrum Analysis

Non-Stationary CT Image Noise Spectrum Analysis Non-Stationary CT Image Noise Spectrum Analysis Michael Balda, Björn J. Heismann,, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen Siemens Healthcare, Erlangen michael.balda@informatik.uni-erlangen.de

More information

NIH Public Access Author Manuscript Int J Imaging Syst Technol. Author manuscript; available in PMC 2010 September 1.

NIH Public Access Author Manuscript Int J Imaging Syst Technol. Author manuscript; available in PMC 2010 September 1. NIH Public Access Author Manuscript Published in final edited form as: Int J Imaging Syst Technol. 2009 September 1; 19(3): 271 276. doi:10.1002/ima.20200. Attenuation map estimation with SPECT emission

More information

Central Slice Theorem

Central Slice Theorem Central Slice Theorem Incident X-rays y f(x,y) R x r x Detected p(, x ) The thick line is described by xcos +ysin =R Properties of Fourier Transform F [ f ( x a)] F [ f ( x)] e j 2 a Spatial Domain Spatial

More information

Superiorized polyenergetic reconstruction algorithm for reduction of metal artifacts in CT images

Superiorized polyenergetic reconstruction algorithm for reduction of metal artifacts in CT images 1 Superiorized polyenergetic reconstruction algorithm for reduction of metal artifacts in CT images T. Humphries 1 and A. Gibali 2 Abstract Artifacts caused by metal objects such as dental fillings, hip

More information

Adaptive Reconstruction Methods for Low-Dose Computed Tomography

Adaptive Reconstruction Methods for Low-Dose Computed Tomography Adaptive Reconstruction Methods for Low-Dose Computed Tomography Joseph Shtok Ph.D. supervisors: Prof. Michael Elad, Dr. Michael Zibulevsky. Technion IIT, Israel, 011 Ph.D. Talk, Apr. 01 Contents of this

More information

USING cone-beam geometry with pinhole collimation,

USING cone-beam geometry with pinhole collimation, IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 687 A Backprojection-Based Parameter Estimation Technique for Skew-Slit Collimation Jacob A. Piatt, Student Member, IEEE, and Gengsheng L.

More information

Statistical iterative reconstruction using fast optimization transfer algorithm with successively increasing factor in Digital Breast Tomosynthesis

Statistical iterative reconstruction using fast optimization transfer algorithm with successively increasing factor in Digital Breast Tomosynthesis Statistical iterative reconstruction using fast optimization transfer algorithm with successively increasing factor in Digital Breast omosynthesis Shiyu Xu a and Zhenxi Zhang b and Ying Chen a,* a Department

More information

An FDK-like cone-beam SPECT reconstruction algorithm for non-uniform attenuated

An FDK-like cone-beam SPECT reconstruction algorithm for non-uniform attenuated Home Search Collections Journals About Contact us My IOPscience An FK-like cone-beam SPECT reconstruction algorithm for non-uniform attenuated projections acquired using a circular trajectory This content

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

NONLINEAR BACK PROJECTION FOR TOMOGRAPHIC IMAGE RECONSTRUCTION

NONLINEAR BACK PROJECTION FOR TOMOGRAPHIC IMAGE RECONSTRUCTION NONLINEAR BACK PROJECTION FOR TOMOGRAPHIC IMAGE RECONSTRUCTION Ken Sauef and Charles A. Bournant *Department of Electrical Engineering, University of Notre Dame Notre Dame, IN 46556, (219) 631-6999 tschoo1

More information

F3-A5: Toward Model-Based Reconstruction in Scanned Baggage Security Applications

F3-A5: Toward Model-Based Reconstruction in Scanned Baggage Security Applications F3-A5: Toward Model-Based Reconstruction in Scanned Baggage Security Applications Abstract While traditional direct reconstruction algorithms such as filtered back projection depend on the analytic inversion

More information

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study B. De Man, J. Nuyts, P. Dupont, G. Marchal and P. Suetens Medical Image Computing, ESAT-PSI, K.U.Leuven, B-3000 Leuven, Belgium Department

More information

SINGLE-PHOTON emission computed tomography

SINGLE-PHOTON emission computed tomography 1458 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 4, AUGUST 2012 SPECT Imaging With Resolution Recovery Andrei V. Bronnikov Abstract Single-photon emission computed tomography (SPECT) is a method

More information

Projection Space Maximum A Posterior Method for Low Photon Counts PET Image Reconstruction

Projection Space Maximum A Posterior Method for Low Photon Counts PET Image Reconstruction Proection Space Maximum A Posterior Method for Low Photon Counts PET Image Reconstruction Liu Zhen Computer Department / Zhe Jiang Wanli University / Ningbo ABSTRACT In this paper, we proposed a new MAP

More information

my working notes on HYPR for Math 597

my working notes on HYPR for Math 597 my working notes on HYPR for Math 597 Nasser M. Abbasi California State University, Fullerton. Summer 2008 page compiled on July 1, 2015 at 10:08pm 1 Notations and definitions 1. MLEM Maximum-Likelihood

More information

SPECT (single photon emission computed tomography)

SPECT (single photon emission computed tomography) 2628 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 5, OCTOBER 2006 Detector Blurring and Detector Sensitivity Compensation for a Spinning Slat Collimator Gengsheng L. Zeng, Senior Member, IEEE Abstract

More information

Iterative SPECT reconstruction with 3D detector response

Iterative SPECT reconstruction with 3D detector response Iterative SPECT reconstruction with 3D detector response Jeffrey A. Fessler and Anastasia Yendiki COMMUNICATIONS & SIGNAL PROCESSING LABORATORY Department of Electrical Engineering and Computer Science

More information

STATISTICAL image reconstruction methods for X-ray

STATISTICAL image reconstruction methods for X-ray IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 28, NO. 5, MAY 2009 645 Quadratic Regularization Design for 2-D CT Hugo R. Shi*, Student Member, IEEE, and Jeffrey A. Fessler, Fellow, IEEE Abstract Statistical

More information

A filtered backprojection MAP algorithm with nonuniform sampling and noise modeling

A filtered backprojection MAP algorithm with nonuniform sampling and noise modeling A filtered backprojection MAP algorithm with nonuniform sampling and noise modeling Gengsheng L. Zeng Citation: Medical Physics 39, 2170 (2012); doi: 10.1118/1.3697736 View online: http://dx.doi.org/10.1118/1.3697736

More information

Index. aliasing artifacts and noise in CT images, 200 measurement of projection data, nondiffracting

Index. aliasing artifacts and noise in CT images, 200 measurement of projection data, nondiffracting Index Algebraic equations solution by Kaczmarz method, 278 Algebraic reconstruction techniques, 283-84 sequential, 289, 293 simultaneous, 285-92 Algebraic techniques reconstruction algorithms, 275-96 Algorithms

More information

Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy

Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy Physics Department, Faculty of Applied Science,Umm Al-Qura

More information

Optimal threshold selection for tomogram segmentation by reprojection of the reconstructed image

Optimal threshold selection for tomogram segmentation by reprojection of the reconstructed image Optimal threshold selection for tomogram segmentation by reprojection of the reconstructed image K.J. Batenburg 1 and J. Sijbers 1 University of Antwerp, Vision Lab, Universiteitsplein 1, B-2610 Wilrijk,

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

Accelerating CT Iterative Reconstruction Using ADMM and Nesterov s Methods

Accelerating CT Iterative Reconstruction Using ADMM and Nesterov s Methods Accelerating CT Iterative Reconstruction Using ADMM and Nesterov s Methods Jingyuan Chen Meng Wu Yuan Yao June 4, 2014 1 Introduction 1.1 CT Reconstruction X-ray Computed Tomography (CT) is a conventional

More information

Resolution and Noise Properties of MAP Reconstruction for Fully 3-D PET

Resolution and Noise Properties of MAP Reconstruction for Fully 3-D PET IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 5, MAY 2000 493 Resolution and Noise Properties of MAP Reconstruction for Fully 3-D PET Jinyi Qi*, Member, IEEE, and Richard M. Leahy, Member, IEEE Abstract

More information

S rect distortions in single photon emission computed

S rect distortions in single photon emission computed IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 41, NO. 6, DECEMBER 1994 2807 A Rotating and Warping Projector/Backprojector for Fan-Beam and Cone-Beam Iterative Algorithm G. L. Zeng, Y.-L. Hsieh, and G. T.

More information

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Advanced Image Reconstruction Methods for Photoacoustic Tomography Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic

More information

Reconstruction in CT and relation to other imaging modalities

Reconstruction in CT and relation to other imaging modalities Reconstruction in CT and relation to other imaging modalities Jørgen Arendt Jensen November 16, 2015 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering Center for Fast Ultrasound

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Edge-Preserving Denoising for Segmentation in CT-Images

Edge-Preserving Denoising for Segmentation in CT-Images Edge-Preserving Denoising for Segmentation in CT-Images Eva Eibenberger, Anja Borsdorf, Andreas Wimmer, Joachim Hornegger Lehrstuhl für Mustererkennung, Friedrich-Alexander-Universität Erlangen-Nürnberg

More information

Introduc)on to PET Image Reconstruc)on. Tomographic Imaging. Projec)on Imaging. Types of imaging systems

Introduc)on to PET Image Reconstruc)on. Tomographic Imaging. Projec)on Imaging. Types of imaging systems Introduc)on to PET Image Reconstruc)on Adam Alessio http://faculty.washington.edu/aalessio/ Nuclear Medicine Lectures Imaging Research Laboratory Division of Nuclear Medicine University of Washington Fall

More information

Q.Clear. Steve Ross, Ph.D.

Q.Clear. Steve Ross, Ph.D. Steve Ross, Ph.D. Accurate quantitation (SUV - Standardized Uptake Value) is becoming more important as clinicians seek to utilize PET imaging for more than just diagnosing and staging disease, but also

More information

GE s Revolution CT MATLAB III: CT. Kathleen Chen March 20, 2018

GE s Revolution CT MATLAB III: CT. Kathleen Chen March 20, 2018 GE s Revolution CT MATLAB III: CT Kathleen Chen chens18@rpi.edu March 20, 2018 https://www.zmescience.com/medicine/inside-human-body-real-time-gifs-demo-power-ct-scan/ Reminders Make sure you have MATLAB

More information

Feldkamp-type image reconstruction from equiangular data

Feldkamp-type image reconstruction from equiangular data Journal of X-Ray Science and Technology 9 (2001) 113 120 113 IOS Press Feldkamp-type image reconstruction from equiangular data Ben Wang a, Hong Liu b, Shiying Zhao c and Ge Wang d a Department of Elec.

More information

A numerical simulator in VC++ on PC for iterative image reconstruction

A numerical simulator in VC++ on PC for iterative image reconstruction Journal of X-Ray Science and Technology 11 (2003) 61 70 61 IOS Press A numerical simulator in VC++ on PC for iterative image reconstruction Xiang Li a,, Ming Jiang c and Ge Wang a,b a Department of Biomedical

More information

A novel coupled transmission-reflection tomography and the V-line Radon transform

A novel coupled transmission-reflection tomography and the V-line Radon transform A novel coupled transmission-reflection tomography and the V-line Radon transform Rémi Regnier, M.K. Nguyen To cite this version: Rémi Regnier, M.K. Nguyen. A novel coupled transmission-reflection tomography

More information

STATISTICAL positron emission tomography (PET) image

STATISTICAL positron emission tomography (PET) image IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 23, NO. 9, SEPTEMBER 2004 1057 Accurate Estimation of the Fisher Information Matrix for the PET Image Reconstruction Problem Quanzheng Li, Student Member, IEEE,

More information

Validation of New Gibbs Priors for Bayesian Tomographic Reconstruction Using Numerical Studies and Physically Acquired Data 1

Validation of New Gibbs Priors for Bayesian Tomographic Reconstruction Using Numerical Studies and Physically Acquired Data 1 Validation of New Gibbs Priors for Bayesian Tomographic Reconstruction Using Numerical Studies and Physically Acquired Data 1 S. J. Lee # Member, IEEE, Y. Choi x Member, IEEE, and G. Gindi { Member, IEEE

More information

Simultaneous Estimation of Attenuation and Activity Images Using Optimization Transfer

Simultaneous Estimation of Attenuation and Activity Images Using Optimization Transfer 1 Sultaneous Estation of Attenuation and Activity Images Using Optization Transfer Matthew Jacobson and Jeffrey A. Fessler Department of Electrical Engineering and Computer Science University of Michigan,

More information

Expectation Maximization and Total Variation Based Model for Computed Tomography Reconstruction from Undersampled Data

Expectation Maximization and Total Variation Based Model for Computed Tomography Reconstruction from Undersampled Data Expectation Maximization and Total Variation Based Model for Computed Tomography Reconstruction from Undersampled Data Ming Yan and Luminita A. Vese Department of Mathematics, University of California,

More information

A Fast GPU-Based Approach to Branchless Distance-Driven Projection and Back-Projection in Cone Beam CT

A Fast GPU-Based Approach to Branchless Distance-Driven Projection and Back-Projection in Cone Beam CT A Fast GPU-Based Approach to Branchless Distance-Driven Projection and Back-Projection in Cone Beam CT Daniel Schlifske ab and Henry Medeiros a a Marquette University, 1250 W Wisconsin Ave, Milwaukee,

More information

A Curvelet based Sinogram Correction Method for Metal Artifact Reduction

A Curvelet based Sinogram Correction Method for Metal Artifact Reduction A based Sinogram Correction Method for Metal Artifact Reduction Kiwan Jeon 1 and Hyoung Suk Park 1 More info about this article: http://www.ndt.net/?id=3715 1 National Institute for Mathematical Sciences,

More information

Regularization in Tomography

Regularization in Tomography Dealing with Ambiguity and Noise Per Christian Hansen Technical University of Denmark About Me Professor of Scientific Computing at DTU Interests: inverse problems, tomography, regularization algorithms,

More information

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Benedikt Lorch 1, Martin Berger 1,2, Joachim Hornegger 1,2, Andreas Maier 1,2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg

More information

AN ANALYSIS OF ITERATIVE ALGORITHMS FOR IMAGE RECONSTRUCTION FROM SATELLITE EARTH REMOTE SENSING DATA Matthew H Willis Brigham Young University, MERS Laboratory 459 CB, Provo, UT 8462 8-378-4884, FAX:

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

Cover Page. The handle holds various files of this Leiden University dissertation.

Cover Page. The handle   holds various files of this Leiden University dissertation. Cover Page The handle http://hdl.handle.net/1887/36132 holds various files of this Leiden University dissertation. Author: Bleichrodt, Folkert Title: Improving robustness of tomographic reconstruction

More information

Tomographic Algorithm for Industrial Plasmas

Tomographic Algorithm for Industrial Plasmas Tomographic Algorithm for Industrial Plasmas More info about this article: http://www.ndt.net/?id=22342 1 Sudhir K. Chaudhary, 1 Kavita Rathore, 2 Sudeep Bhattacharjee, 1 Prabhat Munshi 1 Nuclear Engineering

More information

Column-Action Methods in Image Reconstruction

Column-Action Methods in Image Reconstruction Column-Action Methods in Image Reconstruction Per Christian Hansen joint work with Tommy Elfving Touraj Nikazad Overview of Talk Part 1: the classical row-action method = ART The advantage of algebraic

More information

Iterative CT Reconstruction Using Curvelet-Based Regularization

Iterative CT Reconstruction Using Curvelet-Based Regularization Iterative CT Reconstruction Using Curvelet-Based Regularization Haibo Wu 1,2, Andreas Maier 1, Joachim Hornegger 1,2 1 Pattern Recognition Lab (LME), Department of Computer Science, 2 Graduate School in

More information

MAP Tomographic Reconstruction with a Spatially Adaptive Hierarchical Image Model

MAP Tomographic Reconstruction with a Spatially Adaptive Hierarchical Image Model 7 5th European Signal Processing Conference EUSIPCO) MAP Tomographic Reconstruction with a Spatially Adaptive Hierarchical Image Model Christophoros Nikou Department of Computer Science and Engineering

More information

Algebraic Iterative Methods for Computed Tomography

Algebraic Iterative Methods for Computed Tomography Algebraic Iterative Methods for Computed Tomography Per Christian Hansen DTU Compute Department of Applied Mathematics and Computer Science Technical University of Denmark Per Christian Hansen Algebraic

More information

ML reconstruction for CT

ML reconstruction for CT ML reconstruction for CT derivation of MLTR rigid motion correction resolution modeling polychromatic ML model dual energy ML model Bruno De Man, Katrien Van Slambrouck, Maarten Depypere, Frederik Maes,

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques IMAGING A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques Lai K. Leong, Randall L. Kruger, and Michael K. O Connor Section of Nuclear Medicine, Department

More information

ATTENUATION CORRECTION IN SPECT DURING IMAGE RECONSTRUCTION USING INVERSE MONTE CARLO METHOD A SIMULATION STUDY *

ATTENUATION CORRECTION IN SPECT DURING IMAGE RECONSTRUCTION USING INVERSE MONTE CARLO METHOD A SIMULATION STUDY * Romanian Reports in Physics, Vol. 66, No. 1, P. 200 211, 2014 ATTENUATION CORRECTION IN SPECT DURING IMAGE RECONSTRUCTION USING INVERSE MONTE CARLO METHOD A SIMULATION STUDY * S. AHMADI 1, H. RAJABI 2,

More information

Consistency in Tomographic Reconstruction by Iterative Methods

Consistency in Tomographic Reconstruction by Iterative Methods Consistency in Tomographic Reconstruction by Iterative Methods by M. Reha Civanlar and H.J. Trussell Center for Communications and Signal Processing Department of Electrical and Computer Engineering North

More information

Bias-Variance Tradeos Analysis Using Uniform CR Bound. Mohammad Usman, Alfred O. Hero, Jerey A. Fessler and W. L. Rogers. University of Michigan

Bias-Variance Tradeos Analysis Using Uniform CR Bound. Mohammad Usman, Alfred O. Hero, Jerey A. Fessler and W. L. Rogers. University of Michigan Bias-Variance Tradeos Analysis Using Uniform CR Bound Mohammad Usman, Alfred O. Hero, Jerey A. Fessler and W. L. Rogers University of Michigan ABSTRACT We quantify fundamental bias-variance tradeos for

More information

Improving Reconstructed Image Quality in a Limited-Angle Positron Emission

Improving Reconstructed Image Quality in a Limited-Angle Positron Emission Improving Reconstructed Image Quality in a Limited-Angle Positron Emission Tomography System David Fan-Chung Hsu Department of Electrical Engineering, Stanford University 350 Serra Mall, Stanford CA 94305

More information

IMAGE RECONSTRUCTION WITH BACK FILTERED PROJECTION ALGORITHMS, USED IN CULTURAL HERITAGE INVESTIGATIONS

IMAGE RECONSTRUCTION WITH BACK FILTERED PROJECTION ALGORITHMS, USED IN CULTURAL HERITAGE INVESTIGATIONS Romanian Reports in Physics 69, 405 (2017) IMAGE RECONSTRUCTION WITH BACK FILTERED PROJECTION ALGORITHMS, USED IN CULTURAL HERITAGE INVESTIGATIONS C. ARMEANU National Institute of Research and Development

More information

Regularization in Tomography

Regularization in Tomography Regularization in Tomography Dealing with Ambiguity and Noise Per Christian Hansen Technical University of Denmark About Me Professor of Scientific Computing at DTU Interests: inverse problems, tomography,

More information

M. Usman, A.O. Hero and J.A. Fessler. University of Michigan. parameter =[ 1 ; :::; n ] T given an observation of a vector

M. Usman, A.O. Hero and J.A. Fessler. University of Michigan. parameter =[ 1 ; :::; n ] T given an observation of a vector Uniform CR Bound: Implementation Issues And Applications M. Usman, A.O. Hero and J.A. Fessler University of Michigan ABSTRACT We apply a uniform Cramer-Rao (CR) bound [] to study the bias-variance trade-os

More information

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology.

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology. Photon Counting Detectors and Their Applications in Medical Imaging Principles, Pitfalls and Progress in Photon-Counting-Detector Technology Taly Gilat Schmidt, PhD Associate Professor Department of Biomedical

More information