1 Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Krista Chytyk MSc student Supervisor: Dr. Boyd McCurdy

2 Introduction The objective of cancer radiotherapy is to destroy the malignancy by maximizing the dose to the tumour while minimizing the dose to the normal tissues The accuracy of this delivery directly affects patient outcome because small variations in dose may lead to large discrepancies in tissue response dose errors of greater than 3% can have a considerable effect on patient outcome * * A. Brahme, Dosimetric precision requirements in radiation therapy, Acta. Radiologica. Oncology 23 Fasc. 5, (1984).

3 Geometric verification of treatments reduces the uncertainty in the delivery location carried out with radiographic film, but has been largely replaced by electronic portal imaging devices (EPIDs) a-si EPIDs have shown potential as a radiation dosimeter * *

4 Performing dosimetric verification while the treatment is administered allows for appropriate adjustments to be made for the remainder of the treatment or treatments Measured and predicted portal images can be compared to determine whether the plan is delivering correct dose distributions

5 There had been no investigation into the accuracy of a-si portal dose image prediction using the parallel kernel assumption applying parallel dose kernels in a divergent clinical geometry where radiation fluence is not truly parallel Parallel (convolution) Tilted (superposition) Incident diverging fluence Parallel kernels Tilted kernels

6 Methods and Materials Monte Carlo Simulations Dose kernels for the EPID were obtained at a number of incident angles for a range of monoenergetic photon energies using EGSnrc 0 to 14 every 2 0.1, 2, 6 and 18 MeV user code DOSXYZnrc Incident tilted tilted photon beams Incident perpendicular photon beams back -x-axis EPID case +x-axis Phosphor Proximal half Distal half

7 Simulated detector design specifications provided by Varian Medical Systems (Palo Alto, CA) uniform epoxy front cover followed by a 22 mm air gap and the imaging cassette 1 mm thick copper buildup plate, a Lanex Fast-B phosphor (gadolinium oxysulphide) screen and an a-si photodiode array 3 cm slab of uniform water buildup was added directly upstream of the detector front cover simulated area of the detector for dose kernel generation was approximately 40 x 40 cm 2

8 Dose Kernel Analysis The effect of angle on the symmetry of the dose kernel about the incident pencil beam was investigated kernels were divided into the negative-x half (proximal half) and the positive-x half (distal half), as in figure

9 Tilted vs. Parallel Kernels Clinical impact was investigated by applying the parallel and tilted kernels to a step function representing incident fluence The convolution (parallel kernels) and superposition (tilted kernels) methods of dose calculation were employed for two field sizes and two SDDs 27.2 x 20.4 and 30 x 30 cm 2 at isocenter 105 and 140 cm D = ## $ detector ( x', y') " k( x! x', y! y') dx' dy'

10 For the superposition method, the appropriate tilted kernel for an angle of incidence for a voxel was interpolated from the available kernels This kernel was then rotated azimuthally and translated to the position of the voxel of interest, then regridded to the EPID image pixel coordinates The results of the convolution versus full superposition approach were evaluated by the χ- comparison test * with Δd = mm (the detector resolution) and ΔD = 1.0% % = D # D c 2 ( r) $ D + # d 2 " r ( r)! D r 2 * A Bakai, M Alber and F Nusslin, A revision of the gamma-evaluation concept for the comparison of dose distributions, Phys. Med. Biol. 48, (2003).

11 Results Dose Kernel Comparison 1E-15 Dose/Fluence (Gy*cm 2 /N) 1E-16 1E-17 1E-18 1E-19 1E-20 1E-21 0 kernel 14 kernel (distal) 14 kernel (proximal) 1E-22 1E Position (cm)

12 Superposition vs. Convolution Fraction of Voxels Fraction of Voxels MeV 2 MeV !-Comparison Value 6 MeV 18 MeV (a) (b)

13 Discussion and Conclusions Increasing asymmetry in the χ-comparison distributions with increasing incident photon beam energy Positive values of the χ-comparison indicate the parallel kernel assumption underestimates the tilted kernel superposition results slightly greater tendency for the parallel kernel assumption to underestimate the superposition results with increasing incident photon energy This study validates the practice of applying parallel dose kernels in a divergent clinical geometry to determine the dose deposited in an a-si EPID a full superposition is not required for dosimetric accuracy of better than 1.5% for extreme cases and 1.0% for clinically relevant scenarios

14 Questions?

Proton dose calculation algorithms and configuration data

Proton dose calculation algorithms and configuration data Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam

More information

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization

More information

Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code

Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code American Journal of Biomedical Engineering 216, 6(4): 124-131 DOI: 1.5923/j.ajbe.21664.3 Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code Ankit Kajaria 1,*, Neeraj Sharma

More information

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS Computational Medical Physics Working Group Workshop II, Sep 30 Oct 3, 2007 University of Florida (UF), Gainesville, Florida USA on CD-ROM, American Nuclear Society, LaGrange Park, IL (2007) ELECTRON DOSE

More information

Michael Speiser, Ph.D.

Michael Speiser, Ph.D. IMPROVED CT-BASED VOXEL PHANTOM GENERATION FOR MCNP MONTE CARLO Michael Speiser, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX September 1 st, 2012 CMPWG Workshop Medical

More information

2D DOSE MEASUREMENT USING A FLAT PANEL EPID

2D DOSE MEASUREMENT USING A FLAT PANEL EPID 2D DOSE MEASUREMENT USING A FLAT PANEL EPID by Seng Boh Lim B.A.Sc. (Hons.), University of British Columbia, 1994 M.A.Sc., University of British Colombia, 1996 M.Eng, University of British Columbia, 2002

More information

Basic Radiation Oncology Physics

Basic Radiation Oncology Physics Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides

More information

Comparison of absorbed dose distribution 10 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm

Comparison of absorbed dose distribution 10 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm Journal of Physics: Conference Series PAPER OPEN ACCESS Comparison of absorbed dose distribution 1 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm To cite

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate

More information

Assesing multileaf collimator effect on the build-up region using Monte Carlo method

Assesing multileaf collimator effect on the build-up region using Monte Carlo method Pol J Med Phys Eng. 2008;14(3):163-182. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0014-0 website: http://www.pjmpe.waw.pl M. Zarza Moreno 1, 2, N. Teixeira 3, 4, A. P. Jesus 1, 2, G. Mora 1 Assesing multileaf

More information

Implementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43

Implementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43 International Journal of Innovation and Applied Studies ISSN 2028-9324 Vol. 6 No. 3 July 2014, pp. 635-641 2014 Innovative Space of Scientific Research Journals http://www.ijias.issr-journals.org/ Implementation

More information

CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS

CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS J. Heese, J. Wulff, A. Winnebeck, A. Huggins, M. Schillo VARIAN PARTICLE THERAPY JUERGEN HEESE New gantry developments Viewpoint from user and vendor perspective

More information

A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study

A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study Indrin J. Chetty, a) Mihaela Rosu, Neelam Tyagi, Lon H. Marsh, Daniel

More information

Effect of Statistical Fluctuation in Monte Carlo Based Photon Beam Dose Calculation on Gamma Index Evaluation

Effect of Statistical Fluctuation in Monte Carlo Based Photon Beam Dose Calculation on Gamma Index Evaluation Effect of Statistical Fluctuation in Monte Carlo Based Photon Beam Dose Calculation on Gamma Index Evaluation Yan Jiang Graves 1,2, Xun Jia 1, Steve B. Jiang 1 1 Center for Advanced Radiotherapy Technologies

More information

gpmc: GPU-Based Monte Carlo Dose Calculation for Proton Radiotherapy Xun Jia 8/7/2013

gpmc: GPU-Based Monte Carlo Dose Calculation for Proton Radiotherapy Xun Jia 8/7/2013 gpmc: GPU-Based Monte Carlo Dose Calculation for Proton Radiotherapy Xun Jia xunjia@ucsd.edu 8/7/2013 gpmc project Proton therapy dose calculation Pencil beam method Monte Carlo method gpmc project Started

More information

Intensity-modulated radiation therapy dose maps: the matchline effect

Intensity-modulated radiation therapy dose maps: the matchline effect University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2006 Intensity-modulated radiation therapy dose maps: the matchline

More information

MCNP4C3-BASED SIMULATION OF A MEDICAL LINEAR ACCELERATOR

MCNP4C3-BASED SIMULATION OF A MEDICAL LINEAR ACCELERATOR Computational Medical Physics Working Group Workshop II, Sep 3 Oct 3, 7 University of Florida (UF), Gainesville, Florida USA on CD-ROM, American Nuclear Society, LaGrange Park, IL (7) MCNP4C3-BASED SIMULATION

More information

Improvements in Monte Carlo simulation of large electron fields

Improvements in Monte Carlo simulation of large electron fields SLAC-PUB-12908 October 2007 Improvements in Monte Carlo simulation of large electron fields Bruce A Faddegon 1, Joseph Perl 2 and Makoto Asai 2 1 University of California San Francisco Comprehensive Cancer

More information

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them Dose Distributions George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Purpose To familiarize the resident with dose distributions and the factors that affect them

More information

Verification of dynamic and segmental IMRT delivery by dynamic log file analysis

Verification of dynamic and segmental IMRT delivery by dynamic log file analysis JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 Verification of dynamic and segmental IMRT delivery by dynamic log file analysis Dale W. Litzenberg,* Jean M. Moran, and Benedick

More information

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. 2-D D Dose-CT Mapping in Geant4 Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. Table of Contents Background & Purpose Materials Methods

More information

On compensator design for photon beam intensity-modulated conformal therapy

On compensator design for photon beam intensity-modulated conformal therapy On compensator design for photon beam intensity-modulated conformal therapy Steve B. Jiang a) and Komanduri M. Ayyangar Department of Radiation Therapy, Medical College of Ohio, 3000 Arlington Avenue,

More information

Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy

Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy Journal of Physics: Conference Series Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy To cite this article: Lukas A Hirschi et al 2007 J. Phys.: Conf. Ser. 74 021008 View the article online

More information

Three-dimensional IMRT verification with a flat-panel EPID

Three-dimensional IMRT verification with a flat-panel EPID Three-dimensional IMRT verification with a flat-panel EPID S. Steciw Department of Medical Physics, Cross Cancer Institute, 11560 University Avenue, Edmonton, Alberta T6G IZ2, Canada B. Warkentin Department

More information

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization

More information

8/3/2016. Outline. The EPID Strikes Back: Future EPID Technology and Applications. Active Matrix Flat-Panel Imagers (AMFPIs)

8/3/2016. Outline. The EPID Strikes Back: Future EPID Technology and Applications. Active Matrix Flat-Panel Imagers (AMFPIs) 8//6 The EPID Strikes Back: Future EPID Technology and Applications Larry E. Antonuk Department of Radiation Oncology University of Michigan, Ann Arbor Acknowledgements: Youcef El-Mohri, Qihua Zhao (U.

More information

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions LINAC and MLC QA for IMRT Ping Xia, Ph.D., Department of Radiation Oncology Disclosure Received research support from Siemens Medical Solutions University of California San Francisco Therapy Series (SAM)

More information

Chapter 9 Field Shaping: Scanning Beam

Chapter 9 Field Shaping: Scanning Beam Chapter 9 Field Shaping: Scanning Beam X. Ronald Zhu, Ph.D. Department of Radiation Physics M. D. Anderson Cancer Center Houston, TX June 14-18, 2015 AAPM - Summer School 2015, Colorado Spring Acknowledgement

More information

Photon beam dose distributions in 2D

Photon beam dose distributions in 2D Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation

More information

Calculating percent depth dose with the electron pencilbeam redefinition algorithm

Calculating percent depth dose with the electron pencilbeam redefinition algorithm JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 2, SPRING 2007 Calculating percent depth dose with the electron pencilbeam redefinition algorithm Michael J. Price, 1,2,a Kenneth R. Hogstrom,

More information

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta Use of Monte Carlo modelling in radiotherapy linac design David Roberts, PhD Senior Physicist Elekta Contents Overview of Elekta What we do Where we use Monte Carlo Codes and resources Example : Agility

More information

Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation

Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation Danial Seifi Makrani 1, Hadi Hasanzadeh 2*, Tayyeb Allahverdi Pourfallah 3, Arash Ghasemi 4, Majid

More information

Qalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004

Qalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004 Qalitätssicherung an Multileafkollimatoren Dr. Lutz Müller Würzburg jan 2004 IMRT Verification - present Target Volume Constraints Inverse Planning Algorithm Fluence Map Leaf Sequencer Leaf & Gantry sequence

More information

Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy

Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 3, summer 2011 Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy

More information

Validation of GEANT4 Monte Carlo Simulation Code for 6 MV Varian Linac Photon Beam

Validation of GEANT4 Monte Carlo Simulation Code for 6 MV Varian Linac Photon Beam Validation of GEANT4 Monte Carlo Code for 6 MV Varian Linac Photon Beam E. Salama ab*, A.S. Ali c, N. Emad d and A. Radi a a Physics Department, Faculty of Science, Ain Shams University, Cairo, Egypt;

More information

Dynalog data tool for IMRT plan verification

Dynalog data tool for IMRT plan verification Dynalog data tool for IMRT plan verification Poster No.: R-0051 Congress: 2014 CSM Type: Scientific Exhibit Authors: V. Sashin; FOOTSCRAY/AU Keywords: Computer applications, Radiation physics, Experimental,

More information

Megan A. Wood, M.S. Under the direction of Larry DeWerd, Ph.D. University of Wisconsin Medical Radiation Research Center (UWMRRC)

Megan A. Wood, M.S. Under the direction of Larry DeWerd, Ph.D. University of Wisconsin Medical Radiation Research Center (UWMRRC) Megan A. Wood, M.S. Under the direction of Larry DeWerd, Ph.D. University of Wisconsin Medical Radiation Research Center (UWMRRC) NCCAAPM Spring Meeting May 3, 2013 Introduction FFF background Project

More information

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations A. Al-Basheer, G. Sjoden, M. Ghita Computational Medical Physics Team Nuclear & Radiological Engineering University

More information

FAST, precise. qa software

FAST, precise. qa software qa software FAST, precise Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, rotational or brachytherapy plans no film, no phantoms, no linac

More information

Monte Carlo evaluation of the dosimetric uncertainty in matched 6 MV Elekta and Varian linear accelerators

Monte Carlo evaluation of the dosimetric uncertainty in matched 6 MV Elekta and Varian linear accelerators Monte Carlo evaluation of the dosimetric uncertainty in matched 6 MV Elekta and Varian linear accelerators Dr Jessica E Lye PhD, BSc Hons. A thesis submitted in partial fulfilment of the requirements for

More information

GPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC

GPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC GPU applications in Cancer Radiation Therapy at UCSD Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC Conventional Radiotherapy SIMULATION: Construciton, Dij Days PLANNING:

More information

Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC)

Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC) Dosimetry Check for convolution/superposition collapsed cone (CC), Page 1 of 11 Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC) File: CCReleaseNotes.doc Date:

More information

On the clinically relevant detector resolution and error detection capability of COMPASS 3D plan verification

On the clinically relevant detector resolution and error detection capability of COMPASS 3D plan verification On the clinically relevant detector resolution and error detection capability of COMPASS 3D plan verification white paper Nick Narloch Application Specialist, IBA Dosimetry Introduction What is COMPASS?

More information

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy Data ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy The ModuLeaf Mini Multileaf

More information

GPU-based fast gamma index calcuation

GPU-based fast gamma index calcuation 1 GPU-based fast gamma index calcuation 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Xuejun Gu, Xun Jia, and Steve B. Jiang Center for Advanced Radiotherapy Technologies

More information

87.55.K-, Gh, km, Ca. Agility, analytic model, Elekta, Monaco, Monte Carlo, virtual source

87.55.K-, Gh, km, Ca. Agility, analytic model, Elekta, Monaco, Monte Carlo, virtual source Received: 14 May 2018 Revised: 21 September 2018 Accepted: 27 September 2018 DOI: 10.1002/acm2.12485 RADIATION ONCOLOGY PHYSICS Monte Carlo and analytic modeling of an Elekta Infinity linac with Agility

More information

Verification measurements of an emc algorithm using a 2D ion chamber array

Verification measurements of an emc algorithm using a 2D ion chamber array JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Verification measurements of an emc algorithm using a 2D ion chamber array Mark D. Wanklyn, 1a Ghirmay Kidane, 2 and Liz Crees 2 Medical

More information

Evaluation of latent variances in Monte Carlo dose calculations with Varian TrueBeam photon phase-spaces used as a particle source

Evaluation of latent variances in Monte Carlo dose calculations with Varian TrueBeam photon phase-spaces used as a particle source Evaluation of latent variances in Monte Carlo dose calculations with Varian TrueBeam photon phase-spaces used as a particle source Eyad Alhakeem, 1, 2, Sergei Zavgorodni 2,1 1 Department of Physics and

More information

Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions

Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions

More information

Monaco VMAT. The Next Generation in IMRT/VMAT Planning. Paulo Mathias Customer Support TPS Application

Monaco VMAT. The Next Generation in IMRT/VMAT Planning. Paulo Mathias Customer Support TPS Application Monaco VMAT The Next Generation in IMRT/VMAT Planning Paulo Mathias Customer Support TPS Application 11.05.2011 Background What is Monaco? Advanced IMRT/VMAT treatment planning system from Elekta Software

More information

Indrin Chetty Henry Ford Hospital Detroit, MI. AAPM Annual Meeting Houston 7:30-8:25 Mon 08/07/28 1/30

Indrin Chetty Henry Ford Hospital Detroit, MI.   AAPM Annual Meeting Houston 7:30-8:25 Mon 08/07/28 1/30 Review of TG105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics,

More information

Image-based Monte Carlo calculations for dosimetry

Image-based Monte Carlo calculations for dosimetry Image-based Monte Carlo calculations for dosimetry Irène Buvat Imagerie et Modélisation en Neurobiologie et Cancérologie UMR 8165 CNRS Universités Paris 7 et Paris 11 Orsay, France buvat@imnc.in2p3.fr

More information

Monitor Unit (MU) Calculation

Monitor Unit (MU) Calculation Monitor Unit (MU) Calculation Timothy C. Zhu 1, Haibo Lin 1, and JiaJian Shen 2 1 University of Pennsylvania, Philadelphia, PA 2 Mayo Clinic, Phoenix, AZ Introduction Pencil-beam based dose/mu algorithms

More information

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Effects of the difference in tube voltage of the CT scanner on. dose calculation Effects of the difference in tube voltage of the CT scanner on dose calculation Dong Joo Rhee, Sung-woo Kim, Dong Hyeok Jeong Medical and Radiological Physics Laboratory, Dongnam Institute of Radiological

More information

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell

More information

Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study

Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study Received: 9 November 2016 Revised: 9 November 2016 Accepted: 28 November 2016 DOI: 10.1002/acm2.12039 RADIATION ONCOLOGY PHYSICS Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code

Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code reports of practical oncology and radiotherapy 1 5 (2 0 1 0) 64 68 available at www.sciencedirect.com journal homepage: http://www.rpor.eu/ Original article Measurement of depth-dose of linear accelerator

More information

Modulation factors calculated with an EPID-derived MLC fluence model to streamline IMRT/VMAT second checks

Modulation factors calculated with an EPID-derived MLC fluence model to streamline IMRT/VMAT second checks JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 6, 2013 Modulation factors calculated with an EPID-derived MLC fluence model to streamline IMRT/VMAT second checks Stephen Steciw, 1,2a Satyapal

More information

Medical Dosimetry 37 (2012) Medical Dosimetry. journal homepage:

Medical Dosimetry 37 (2012) Medical Dosimetry. journal homepage: Medical Dosimetry 37 (2012) 53-60 Medical Dosimetry journal homepage: www.meddos.org Characterization of responses of 2d array seven29 detector and its combined use with octavius phantom for the patient-specific

More information

Dose Calculations: Where and How to Calculate Dose. Allen Holder Trinity University.

Dose Calculations: Where and How to Calculate Dose. Allen Holder Trinity University. Dose Calculations: Where and How to Calculate Dose Trinity University www.trinity.edu/aholder R. Acosta, W. Brick, A. Hanna, D. Lara, G. McQuilen, D. Nevin, P. Uhlig and B. Slater Dose Calculations - Why

More information

A DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

A DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL A DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE MASTER OF

More information

Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code

Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code Iran. J. Radiat. Res., 2006; 4 (1): 7-14 Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code. Mesbahi * Department of Medical Physics, Medical School,

More information

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT Volumetric Modulated Arc Therapy - Clinical Implementation Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA Acknowledgement David M. Shepard, Ph.D. Muhammad K. N. Afghan, Ph.D. Fan Chen, Ph.D.

More information

Raising the Bar in IMRT QA

Raising the Bar in IMRT QA MapCHECK 2TM Raising the Bar in IMRT QA The leader in quick and precise measurement of modulated radiotherapy beams Benefits Proven solution for film-less rotational delivery and IMRT QA - More than 1500

More information

Clinical implementation of photon beam flatness measurements to verify beam quality

Clinical implementation of photon beam flatness measurements to verify beam quality JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Clinical implementation of photon beam flatness measurements to verify beam quality Simon Goodall, a Nicholas Harding, Jake Simpson,

More information

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters Monte Carlo Simulation for Neptun 1 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters M.T. Bahreyni Toossi a, M. Momen Nezhad b, S.M. Hashemi a a Medical Physics Research Center,

More information

Photon Dose Algorithms and Physics Data Modeling in modern RTP

Photon Dose Algorithms and Physics Data Modeling in modern RTP Photon Dose Algorithms and Physics Data Modeling in modern RTP Niko Papanikolaou, PhD Professor and Director of Medical Physics University of Texas Health Science Center of San Antonio Cancer Therapy &

More information

Significance of time-dependent geometries for Monte Carlo simulations in radiation therapy. Harald Paganetti

Significance of time-dependent geometries for Monte Carlo simulations in radiation therapy. Harald Paganetti Significance of time-dependent geometries for Monte Carlo simulations in radiation therapy Harald Paganetti Modeling time dependent geometrical setups Key to 4D Monte Carlo: Geometry changes during the

More information

Advanced Radiotherapy

Advanced Radiotherapy Advanced Radiotherapy IBA Concept for IMRT Verification Dr. Lutz Müller 1 IMRT Verification BIS BIS ADAS Fluence Model Delivered fluence Target Volume Constraints Inverse Backprojection Fluence Map Leaf

More information

Transitioning from pencil beam to Monte Carlo for electron dose calculations

Transitioning from pencil beam to Monte Carlo for electron dose calculations Transitioning from pencil beam to Monte Carlo for electron dose calculations Jessie Huang-Vredevoogd (jyhuang4@wisc.edu) University of Wisconsin NCC AAPM October 12, 2019 1 Topics to cover Background RayStation

More information

Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems. Bruce Curran University of Michigan Medical Center Ann Arbor, MI

Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems. Bruce Curran University of Michigan Medical Center Ann Arbor, MI Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems Bruce Curran University of Michigan Medical Center Ann Arbor, MI Disclosure Some of the work discussed in this talk was supported

More information

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it

More information

Comparison Study of Monte Carlo Simulations and Measurements of Relative Output Factors of 6 MV Photon Beam

Comparison Study of Monte Carlo Simulations and Measurements of Relative Output Factors of 6 MV Photon Beam Comparison Study of Monte Carlo Simulations and Measurements of Relative Output Factors of 6 MV Photon Beam Nakorn Phaisangittisakul 1*, Wanchaloem Saprangsi 2, Chirapha Tannanonta 2 and Sivalee Suriyapee

More information

Machine and Physics Data Guide

Machine and Physics Data Guide WWW..COM Machine and Physics Data Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 May / 2008 2008 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 www.standardimaging.com

More information

A fast and accurate GPU-based proton transport Monte Carlo simulation for validating proton therapy treatment plans

A fast and accurate GPU-based proton transport Monte Carlo simulation for validating proton therapy treatment plans A fast and accurate GPU-based proton transport Monte Carlo simulation for validating proton therapy treatment plans H. Wan Chan Tseung 1 J. Ma C. Beltran PTCOG 2014 13 June, Shanghai 1 wanchantseung.hok@mayo.edu

More information

Comprehensive evaluation and clinical implementation of commercially available Monte Carlo dose calculation algorithm

Comprehensive evaluation and clinical implementation of commercially available Monte Carlo dose calculation algorithm JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Comprehensive evaluation and clinical implementation of commercially available Monte Carlo dose calculation algorithm Aizhen Zhang,

More information

The MSKCC Approach to IMRT. Outline

The MSKCC Approach to IMRT. Outline The MSKCC Approach to IMRT Spiridon V. Spirou, PhD Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY Outline Optimization Field splitting Delivery Independent verification

More information

A SYSTEM OF DOSIMETRIC CALCULATIONS

A SYSTEM OF DOSIMETRIC CALCULATIONS A SYSTEM OF DOSIMETRIC CALCULATIONS INTRODUCTION Dose calculation based on PDD and TAR have Limitations The dependence of PDD on SSD Not suitable for isocentric techniques TAR and SAR does not depend on

More information

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac

More information

A secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans

A secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2013 A secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans Adam

More information

A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations

A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations V. Rosso, G.Battistoni, N. Belcari, N. Camarlinghi, A. Ferrari, S. Ferretti, A. Kraan, A. Mairani, N. Marino, J. E.

More information

AN ABSTRACT OF THE DISSERTATION OF

AN ABSTRACT OF THE DISSERTATION OF AN ABSTRACT OF THE DISSERTATION OF Alexander J Egan for the degree of Doctor of Philosophy in Medical Physics presented on March 31, 2014. Development of an accurate Monte Carlo treatment plan calculation

More information

Helical Tomotherapy Qualitative dose Delivery Verification

Helical Tomotherapy Qualitative dose Delivery Verification University of Tennessee, Knoxville Trace: Tennessee Research and Creative Exchange University of Tennessee Honors Thesis Projects University of Tennessee Honors Program 5-2003 Helical Tomotherapy Qualitative

More information

Investigation of photon beam output factors for conformal radiation therapy Monte Carlo simulations and measurements

Investigation of photon beam output factors for conformal radiation therapy Monte Carlo simulations and measurements INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N133 N143 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)32395-9 NOTE Investigation of photon beam output factors for conformal radiation therapy

More information

A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction

A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 4, FALL 2005 A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction Wendel Dean

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Measurement and Monte Carlo simulation for energy- and intensity-modulated electron radiotherapy delivered by a computer-controlled

More information

Suggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems

Suggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 3, Summer 2010 Suggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems Sanaz Hariri,

More information

REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT

REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT Anand P Santhanam Assistant Professor, Department of Radiation Oncology OUTLINE Adaptive radiotherapy for head and

More information

Evaluation of fluence-based dose delivery incorporating the spatial variation of dosimetric leaf gap (DLG)

Evaluation of fluence-based dose delivery incorporating the spatial variation of dosimetric leaf gap (DLG) JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Evaluation of fluence-based dose delivery incorporating the spatial variation of dosimetric leaf gap (DLG) Lalith K. Kumaraswamy,

More information

Automated Quality Assurance for Image-Guided Radiation Therapy

Automated Quality Assurance for Image-Guided Radiation Therapy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Automated Quality Assurance for Image-Guided Radiation Therapy Eduard Schreibmann, a Eric Elder, Tim Fox Department of Radiation

More information

A graphical user interface for an electron monitor unit calculator using a sector-integration algorithm and exponential curve-fitting method

A graphical user interface for an electron monitor unit calculator using a sector-integration algorithm and exponential curve-fitting method JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 A graphical user interface for an electron monitor unit calculator using a sector-integration algorithm and exponential curve-fitting

More information

Fast dose algorithm for generation of dose coverage probability for robustness analysis of fractionated radiotherapy

Fast dose algorithm for generation of dose coverage probability for robustness analysis of fractionated radiotherapy Fast dose algorithm for generation of dose coverage probability for robustness analysis of fractionated radiotherapy David Tilly 1,2), Anders Ahnesjö 1) 1) Medical radiation physics, department of immunology,

More information

A software tool for the quantitative evaluation of 3D dose calculation algorithms

A software tool for the quantitative evaluation of 3D dose calculation algorithms A software tool for the quantitative evaluation of 3D dose calculation algorithms William B. Harms, Sr., Daniel A. Low, John W. Wong, a) and James A. Purdy Washington University School of Medicine, Mallinckrodt

More information

Output factor comparison of Monte Carlo and measurement for Varian TrueBeam 6 MV and 10 MV flattening filter-free stereotactic radiosurgery system

Output factor comparison of Monte Carlo and measurement for Varian TrueBeam 6 MV and 10 MV flattening filter-free stereotactic radiosurgery system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 3, 2016 Output factor comparison of Monte Carlo and measurement for Varian TrueBeam 6 MV and 10 MV flattening filter-free stereotactic radiosurgery

More information

I Introduction 2. IV Relative dose in electron and photon beams 26 IV.A Dose and kerma per unit incident fluence... 27

I Introduction 2. IV Relative dose in electron and photon beams 26 IV.A Dose and kerma per unit incident fluence... 27 Notes on the structure of radiotherapy depth-dose distributions David W O Rogers Carleton Laboratory for Radiotherapy Physics Physics Department, Carleton University, Ottawa, Canada drogers at physics.carleton.ca

More information

State-of-the-Art IGRT

State-of-the-Art IGRT in partnership with State-of-the-Art IGRT Exploring the Potential of High-Precision Dose Delivery and Real-Time Knowledge of the Target Volume Location Antje-Christin Knopf IOP Medical Physics Group Scientific

More information

Monte Carlo modeling of a Novalis TX Varian 6 MV with HD-120 multileaf collimator

Monte Carlo modeling of a Novalis TX Varian 6 MV with HD-120 multileaf collimator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Monte Carlo modeling of a Novalis TX Varian 6 MV with HD-120 multileaf collimator Luis Alberto Vazquez-Quino, 1 Brian Massingill,

More information

Accuracy of treatment planning calculations for conformal radiotherapy van 't Veld, Aart Adeodatus

Accuracy of treatment planning calculations for conformal radiotherapy van 't Veld, Aart Adeodatus University of Groningen Accuracy of treatment planning calculations for conformal radiotherapy van 't Veld, Aart Adeodatus IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's

More information