1 Electron Paramagnetic Resonance PH425 System outline Michael Kabatek & Jessica Lovewell Version 1.1 This outline is intended to introduce you to the PH425 EPR system. The system consists of a field controlled lock-in detection scheme which you will be using to study EPR of a solid black powered sample of the free radical diphenyl picryl hydrazil (DPPH). Understanding the system will be essential in order to perform EPR experiments in this laboratory. Understanding this system will allow one to perform a complete power calibrated measurement of EPR. It will allow for measurement of not only the resonance conditions of black powered DPPH but also the energy and power absorption associated with the resonance. The system schematic is shown in figure 1. Figure 1. The field controlled lock-in detection scheme used to measure EPR in solid black powder sampled of DPPH. The system consists of a microwave source, electromagnet, helical cavity, lock-in amplifier, and computer with interface cards.
2 Measurement techniques are covered in an additional theory handout (ELECTRON PARAMAGNETIC RESOSANCE), be sure you read and understand the experiment description before continuing. The components which comprise the system are the microwave source, electromagnet, frequency counter, lock-in amplifier, diode detector, helical cavity, 60Hz reference, and computer with interface cards. Measurements are preformed by sweeping the applied magnetic field, and recording the output voltage from the lock-in. This can be automated in a number of ways using the computer and labview program. The microwave source will allow you to set the power and frequency of microwaves which are pumped into the sample. The microwave source is not computer controlled so you will have to manually adjust the frequency and power of microwaves pumped into the sample. The microwave source is shown below: The microwave source is picked off, sent through a 20 db attenuator, and input into a digital frequency counter so that one can accurately read the pumping frequency. The microwave source should be connected into the helical cavity which contains the DPPH sample. The output lead of the helical cavity will then go through a diode detector and into the lock-in amplifier. Note that the 60Hz modulation is also connected to both the helical cavity and lock-in amplifier and neither are computer controlled.
3 The lock-in amplifier is shown below (left) and the 60Hz reference/field modulator is shown below (right): The lock-in amplifier output is connected to one of the lab jack data acquisition unit (DAQ) inputs (AI0 - analog input 0), which is connected to the computer via USB. This will allow you to read the voltage output from the lock-in on the computer. The DAQ also has a second function which is to control the current supplied to electromagnet which will control the applied magnetic field. One of the DAQ s outputs (AO0 - analog output 0), is connected to a high current electronic measurements inc. (EMI) power supply. The positive lead of AO0 is connected to V AMP IN (3) on the TBIO1 bus on the back of the EMI power supply. The GND lead of AO0 is connected to the GND (8) on the TBI bus on the back of the EMI power supply The voltage applied by the DAQ to the back of the EMI power supply will control the current supplied to the magnet.
4 Because the DAQ cannot sink the current required to drive the magnet the current supplied to the magnet is recorded using a Keithly A175 digital multimeter (DMM) connect to the computer via a GPIB card in the computer. (shown below): A conversion will have to be done by the experimentalist in order to convert the current read by the DMM to magnetic field. Data relating the magnetic field to the current driving the magnet is located on the desktop of the computer. A labview program is provided to automate sweeping the magnetic field, recording the current supplied to the magnet, and recording the voltage output from the lock-in. Using the labview program one can change the starting current, current sweep resolution, time delay between measurements, and the filename to which the data is output as well as the number of data points to be taken. The program is located on the desktop (EPR1.vi). The front panel of the program is shown above, and the block diagram of the program can be found by click Window from the file menu and then show block diagram
5 It would be wise to become familiar with using and programming in labview. Feel free to modify and improve the program to your liking but be sure you leave the original program intact. One idea would be to add a graph to the program so that the program makes a graph of lock-in output versus current supplied to the magnet in real time We will give a brief overview of the functionality of the program, by looking at the block diagram. The program consists of a for loop in which the number of data points can be set on the front panel. Inside the for loop is the main function of the program which executes in a sequence called a stacked sequence frame. The first frame of the sequence is shown below. The first frame tells the lab jack to output a voltage on AO0 depending on the values of Magnet offset and Magnet Sweep inc. set by the user on the front panel.
6 The next frame in the sequence is a simple delay. This delay is a delay set by the user in order to allow the system to reach a steady state before recording data. The delay should be set in accordance with the time constant on the lock-in amplifier. The next frame in the sequence records the current supplied to the magnet from the Keithly multimeter via a GPIB card installed in the computer.
7 The final frame in the sequence records the voltage output from the lock-in amplifier into the AI0 input of the lab jack and writes the data into a spreadsheet file specified by the user. For each data point recorded the program averages over 100 voltage values from the lock-in and outputs a mean and standard deviation of each point. The standard deviation read from the multimeter is close to negligible thus one can use the instrumental uncertainty associated with this specific multimeter. These four sequences are run each iteration of the outer for loop, which is set by the user on the front panel. The program will output a file containing three columns [current, lock-in voltage, lock-in standard deviation]. You will need to configure the program s options based on your lock-in settings; you will want to configure the program to optimally record data based on the time constant of the lock-in. Before running the program you should ensure all the required equipment is turned on, and you know at what current/magnetic field value the resonance occurs. You will need to set the magnet offset and Magnet Sweep Inc setting to a reasonable value such that when you run the program it will sweep the magnetic field through the resonance of the sample. You will need to adjust all the parameters to your liking in order to get adequate data.
Teaching Computer-Based Data Acquisition and Analysis Research Skills Using LabVIEW
Teaching Computer-Based Data Acquisition and Analysis Research Skills Using LabVIEW OUTLINE: John Essick Reed College Reed Advanced Laboratory LabVIEW-Based Instruction Required Cost and Time for Instruction
More informationThe data acquisition components, and their relationship to each other, are shown below.
Data acquisition system Part (1) Digital inputs output and counter You can think of a data acquisition system as a collection of software and hardware that connects you to the physical world. A typical
More informationHands-on Lab 2: LabVIEW NI-DAQ Basics 2
Hands-on Lab 2: LabVIEW NI-DAQ Basics 2 Recall that the final objective is position regulation using computer-controlled state feedback. Computer control requires both software, like LabVIEW and hardware,
More informationIntroduction to Labview and Temperature Measurement
Introduction to Labview and Temperature Measurement Objective This lab is intended to familiarize you with the LABVIEW software and the data acquisition board used in this class and with temperature measurements
More informationAgilent 3630A Triple DC Power Supply. Agilent 34401A Digital Multimeter (DMM)
Agilent E3630A Triple DC Power Supply and Agilent 34401A Digital Multimeter (DMM) Agilent 3630A Triple DC Power Supply The DC power supply used in this lab is the Agilent E3630A Triple DC power supply.
More informationEE 100 DMM and DC Power Supply LabVIEW Experiment Guide
1. Objectives EE 100 DMM and DC Power Supply LabVIEW Experiment Guide The purpose of this experiment is to show you how to interface your DMM and DC Power Supply to LabVIEW. I. World of LabVIEW LabVIEW
More informationPHYS 5061 Lab 1: Introduction to LabVIEW
PHYS 5061 Lab 1: Introduction to LabVIEW In this lab, you will work through chapter 1 and 2 of Essick s book to become familiar with using LabVIEW to build simple programs, called VI s in LabVIEW-speak,
More informationPHYS 5061 Lab 6 Programmable Instruments
Introduction PHYS 5061 Lab 6 Programmable Instruments This lab introduces the computer control of bench lab instruments for data acquisition and uses a programmable digital multimeter as part of a measurement
More informationLABORATORY USE OF A MICROCOMPUTER (Last Revision: August 27, 2000)
LABORATORY USE OF A MICROCOMPUTER (Last Revision: August 27, 2000) ABSTRACT: A program, LabVIEW, is used to enable a microcomputer to acquire and manipulate data and to generate output voltages. TEXT REFERENCE:
More informationSection 1 Establishing an Instrument Connection
Manual for Sweep VI Fall 2011 DO NOT FORGET TO SAVE YOUR DATA TO A NEW LOCATION, OTHER THAN THE TEMP FOLDER ON YOUR LAB STATION COMPUTER! FAILURE TO DO SO WILL RESULT IN LOST DATA WHEN YOU LOG OUT! 1.1.
More informationNI mydaq HANS-PETTER HALVORSEN, Telemark University College Department of Electrical Engineering, Information Technology and Cybernetics
Telemark University College Department of Electrical Engineering, Information Technology and Cybernetics NI mydaq HANS-PETTER HALVORSEN, 2012.01.20 Faculty of Technology, Postboks 203, Kjølnes ring 56,
More informationUNIVERSITY OF CALIFORNIA BERKELEY Department of Electrical Engineering and Computer Sciences. B. E. BOSER Oscilloscopes EECS 100
UNIVERSITY OF CALIFORNIA BERKELEY Department of Electrical Engineering and Computer Sciences B. E. BOSER Oscilloscopes EECS 100 Multimeters are convenient to check supply voltages and currents, bias levels,
More informationNI Technical Symposium ni.com
NI Technical Symposium 2016 1 New Hardware Technologies for DAQ Applications Nicholas Haripersad Field Sales Engineer Session Objectives Introduce the three major NI DAQ platforms Present future trends
More informationDetermination of Drag Coefficient
DEPARTMENT OF MECHANICAL, INDUSTRIAL AND MANUFACTURING ENGINEERING MIMU 505 - MEASUREMENT AND ANALYSIS Determination of Drag Coefficient You will need to bring a zip disk or USB storage device to the lab
More informationENGR 1000, Introduction to Engineering Design
ENGR 1000, Introduction to Engineering Design Unit 2: Data Acquisition and Control Technology Lesson 2.1: Programming Line Outputs for the NI USB-6008 in LabVIEW Hardware: 12 VDC power supply Several lengths
More informationData Acquisition Card or DMM: Which is Right for Your Application? By John Tucker and Joel Roop Keithley Instruments, Inc.
Data Acquisition Card or DMM: Which is Right for Your Application? By John Tucker and Joel Roop Keithley Instruments, Inc. When automating voltage measurements, is it better to go with a programmable Digital
More informationStrain and Force Measurement
NORTHEASTERN UNIVERSITY DEPARTMENT OF MECHANICAL, INDUSTRIAL AND MANUFACTURING ENGINEERING MIMU 0-MEASUREMENT AND ANALYSIS Strain and Force Measurement OBJECTIVES The primary objective of this experiment
More informationLabVIEW Core 1. What You Need To Get Started. File Locations. The course installer places the course files in the following location: ni.
LabVIEW Core 1 What You Need To Get Started LabVIEW Core 1 Course Manual (online) LabVIEW Core 1 Exercise Manual (online) LabVIEW Core 1 Course CD (preloaded on S Share) Multifunction DAQ device File Locations
More informationStep-by-Step Data Acquisition Part II Exercise 2: Generating an Analog Output Waveform
Step-by-Step Data Acquisition Part II Exercise 2: Generating an Analog Output Waveform In this exercise, you will use the DAQ Assistant to build a LabVIEW VI that generates and outputs an analog waveform.
More informationNI R Series. Contents. Software CALIBRATION PROCEDURE
CALIBRATION PROCEDURE NI R Series NI PCIe-784xR, NI PXI-784xR, NI PXIe-784xR, NI USB-784xR, NI PCIe-785xR, NI PXI-785xR, NI PXIe-785xR, and NI USB-785xR This document contains the verification and adjustment
More informationLabview Lab 2. Vern Lindberg. April 16, 2012
Labview Lab 2 Vern Lindberg April 16, 2012 1 Temperature Measurement Thermistors are sensitive semiconductor devices that can measure temperature over a restricted temperature range. The thermistors we
More informationHDice Review. Peter Bonneau Detector Support Group
Peter Bonneau Detector Support Group 1 Detector Support Group 2 1. Search for semi-flexible NMR cables with low loss or controlled temperature variation. Status: Completed. Conducted extensive research
More informationNI PCIe-784xR, NI PXI-784xR, NI PXIe-784xR, NI USB-784xR, NI PCIe-785xR, NI PXI-785xR, NI PXIe-785xR, NI USB-785xR, and NI PXIe-786xR
CALIBRATION PROCEDURE NI R Series NI PCIe-784xR, NI PXI-784xR, NI PXIe-784xR, NI USB-784xR, NI PCIe-785xR, NI PXI-785xR, NI PXIe-785xR, NI USB-785xR, and NI PXIe-786xR This document contains the verification
More informationAIO LN-USB. Features. N Series for USB Multifunction DAQ Unit (8ch AI, 2ch AO, 16ch DIO) AIO LN-USB 1. Ver.1.01
N Series for USB Multifunction DAQ Unit (8ch AI, 2ch AO, 16ch DIO) AIO-120802LN-USB This product is a USB2.0-compliant analog I/O unit that extends the analog I/O function of USB port of PCs. Compact design
More informationEstablishing a Validated Automated System for Attenuation Measurement at DC-18GHz Frequencies
International Journal of Information and Computation Technology. ISSN 0974-2239 Volume 3, Number 6 (2013), pp. 521-526 International Research Publications House http://www. irphouse.com /ijict.htm Establishing
More informationni.com/training Quizzes LabVIEW Core 1 ni.com/training Courses Skills learned: LabVIEW environment Certifications Skills tested: LabVIEW environment
LabVIEW Core 1 What You Need To Get Started LabVIEW Core 1 Course Manual LabVIEW Core 1 Exercise Manual LabVIEW Core 1 Course CD Multifunction DAQ device GPIB interface DAQ Signal Accessory, wires, and
More informationCHAPTER IV Computerized Data-acquisition Systems
CHAPTER IV Computerized Data-acquisition Systems The signal outputting a sensor is usually an analog signal. Post-processing methods involve, however, complex mathematical formulations. The data recorded
More informationBE/EE189 Design and Construction of Biodevices Lecture 5. BE/EE189 Design and Construction of Biodevices - Caltech
BE/EE189 Design and Construction of Biodevices Lecture 5 LabVIEW Programming Data acquisition DAQ system Signals and signal conditioning Nyquist frequency NI ELVIS II NI-DAQmx and DAQ assistant LabVIEW
More informationDept. of Electrical, Computer and Biomedical Engineering. Instrumentation for the data acquisition laboratory
Dept. of Electrical, Computer and Biomedical Engineering Instrumentation for the data acquisition laboratory Purpose of the lab activity Design and make simple systems for data acquisition from detectors,
More informationPHY 351/651 LABORATORY 1 Introduction to LabVIEW
PHY 351/651 LABORATORY 1 Introduction to LabVIEW Introduction Generally speaking, modern data acquisition systems include four basic stages 1 : o o A sensor (or transducer) circuit that transforms a physical
More informationVME Data Acquisition System, ADC Read
VME Data Acquisition System, ADC Read Abstract: UTA-HEP/LC 0023 Shashwat Udit University of Texas at Arlington August 25, 2008 This document presents the design concept and the functionality of the newly
More informationTABLE OF CONTENTS. AEROFLEX EUROPTEST WPN9000 Getting Started V 2.2 Page 1 / 37
TABLE OF CONTENTS WPN9000 Main Display...2 1 WPN9000 Interface...3 Top Bar Display...4 Left Bar Menu...5 Memory:...5 Kphi Measurement...6 Measurement...6 Cablings system and PN9000 Phase Noise Test Set...7
More informationScannerCal Accuracy Verification System for the FlowScanner
ScannerCal Product Bulletin ScannerCal Accuracy Verification System for the FlowScanner www.fisher.com Product Bulletin ScannerCal ScannerCal Accuracy Verification System Verify the calibration and accuracy
More information8000 Series Digital Multimeter THE DMM DESIGNED FOR METROLOGISTS
8000 Series Digital Multimeter THE DMM DESIGNED FOR METROLOGISTS Key Features Function Range Notes DC Voltage 0 1000V With accuracies of up to 4ppm, and exceptional low noise performance AC Voltage 0 1000V
More informationINTRODUCTION TO LABVIEW
INTRODUCTION TO LABVIEW 2nd Year Microprocessors Laboratory 2012-2013 INTRODUCTION For the first afternoon in the lab you will learn to program using LabVIEW. This handout is designed to give you an introduction
More informationLab 4: Introduction to ELVIS II+ Introduction to ELVIS II+
Page 1 of 12 Laboratory Goals Introduction to ELVIS Lab 4: Introduction to ELVIS Familiarize students with the National Instruments hardware ELVIS Identify the capabilities of ELVIS Make use of the built
More informationM Series USB-6221/6225/6229/6251/6255/6259 OEM Devices
USER GUIDE NI USB-x/x OEM M Series USB-///// OEM Devices This document provides dimensions, pinouts, and information about the connectors, switch, LEDs, and chassis ground of the National Instruments USB-
More informationMeterless Laser Power/Energy Measurement Simplifies Embedding
White Paper Meterless Laser Power/Energy Measurement Simplifies Embedding Traditional laser power and energy measurement instruments typically comprise a sensor head connected to separate meter electronics.
More informationElectrical Metrology Applications of LabVIEW Software
Journal of Software Engineering and Applications, 2013, 6, 113-120 http://dx.doi.org/10.4236/jsea.2013.63015 Published Online March 2013 (http://www.scirp.org/journal/jsea) 113 Hala M. Abdel Mageed, Ali
More information1. Learn about LabView software and its different components
SfwrEng 4aa3/4ga3 Lab 1 Lab Sessions: Week starting Sept. 21, 2009. Pre-lab reports Due: Week Starting Sept. 21, 2009 at the start of lab sessions. Lab-Reports Due: Week Starting Oct. 5, 2009 at the start
More informationc01.qxd p /18/01 11:03 AM Page 1 Fundamentals
c01.qxd p001-017 10/18/01 11:03 AM Page 1 Fundamentals c01.qxd p001-017 10/18/01 11:03 AM Page 2 OVERVIEW Welcome to the world of LabVIEW! This chapter gives you a basic explanation of LabVIEW and its
More informationNI PXI Contents CALIBRATION PROCEDURE
CALIBRATION PROCEDURE NI PXI-5422 Contents This document contains step-by-step instructions for writing a calibration procedure for the NI PXI-5422 arbitrary waveform generator. Conventions... 1 Introduction...
More informationD115 The Fast Optimal Servo Amplifier For Brush, Brushless, Voice Coil Servo Motors
D115 The Fast Optimal Servo Amplifier For Brush, Brushless, Voice Coil Servo Motors Ron Boe 5/15/2014 This user guide details the servo drives capabilities and physical interfaces. Users will be able to
More informationLab Exercise 2: Data Acquisition with NI LabVIEW
Objective When you have completed this exercise, you will be able: To use the DAQ Assistant to acquire analog data measurements with NI LabVIEW To use Write to Measurement express VI to log real time data
More informationPersonal Daqs USB Data Acquisition Modules Models /55 & /56
USB Data Acquisition Modules Models /55 & /56 Features Multi-function data acquisition modules attach to PCs via Universal Serial Bus (USB) Ultra low-power design requires no external power or batteries
More informationLabJack U12 Quickstart Guide
LabJack U12 Quickstart Guide Revision 1.03 9/19/2001 LabJack Corporation www.labjack.com support@labjack.com For the latest version of the quickstart guide, the user s guide, or software, go to www.labjack.com.
More informationECONseries Low Cost USB DAQ
ECONseries Low Cost USB Data Acquisition Modules ECONseries Low Cost USB DAQ The ECONseries is a flexible yet economical series of multifunction data acquisition modules. You choose the number of analog
More informationINNOVATION IN MAGNETICS DS-2201/8 DAS1. Data Acquisition System.
INNOVATION IN MAGNETICS DS-2201/8 DAS1 Data Acquisition System www.bartington.com DAS1 Data Acquisition System DAS1 Data Acquisition System The DAS1 is a multi-sensor data acquisition system designed to
More informationAdvanced NI-DAQmx Programming Techniques with LabVIEW
Advanced NI-DAQmx Programming Techniques with LabVIEW Agenda Understanding Your Hardware Data Acquisition Systems Data Acquisition Device Subsystems Advanced Programming with NI-DAQmx Understanding Your
More information3,600 W to 10,800 W 120 V to 800 V 22.5 A to 330 A
3,600 W to 10,800 W 120 V to 800 V 22.5 A to 330 A 2 Electronic Energy Recycling Load ERI Series Syntax Interface Overview RS-232 USB LAN GPIB O CAN Analog Analog isolated O Standard O Option /not available
More informationEinfach & Flexibel USB-Messtechnikprodukte von Agilent Technologies
Einfach & Flexibel USB-Messtechnikprodukte von Agilent Technologies Topics USB Modular Instruments Overview USB Modular Instruments Features Examples Typical Test Environment Source Power Supplies: E3600A
More informationData Acquisition Laboratory
Session 2559 Data Acquisition Laboratory Asad Yousuf Savannah State University Abstract The essential element to automate your system for data collection and analysis is termed as the data acquisition.
More informationFeatures. High-precision Analog input board (Low Profile size) for PCI Express AI-1616L-LPE AI-1616L-LPE 1. Ver.1.01
High-precision Analog input board (Low Profile size) for PCI Express AI-1616L-LPE *Specifications, colors and design of the products are subject to change without notice. This product is a multi-function,
More informationNI USB-622x/625x/628x OEM
USER GUIDE NI USB-x/x/x OEM M Series USB-/////// OEM Devices This document provides dimensions, pinouts, and information about the connectors, switch, LEDs, and chassis ground of the National Instruments
More informationUSB 1608G Series USB Multifunction Devices
USB Multifunction Devices Features 16-bit high-speed USB devices Acquisition rates ranging from 250 ks/s to 500 ks/s differential (DIFF) or 16 singleended (SE) analog inputs (softwareselectable) Up to
More informationKistler LabAmp. Electronics & Software. Charge Amplifier and Data Acquisition Unit for Dynamic Measurements. Type 5165A...
Electronics & Software Kistler LabAmp Charge Amplifier and Data Acquisition Unit for Dynamic Measurements This universal laboratory charge amplifier can be used wherever dynamic signals 1) of mechanical
More informationDesign of Virtual Multimeter Based on VIIS-EM Platform
Design of Virtual Multimeter Based on VIIS-EM Platform Zhengling Wu 1, Haiyan Wu 2 1 College of information and control engineering, Jilin Institute of Chemical Technology, Jilin, Jilin 132022, PR China;
More informationStudent Quick Reference Guide
Student Quick Reference Guide How to use this guide The Chart Student Quick Reference Guide is a resource for PowerLab systems in the classroom laboratory. The topics in this guide are arranged to help
More informationMeasurement & Automation Explorer (MAX) View and control your devices and software
1. Labview basics virtual instruments, data flow, palettes 2. Structures for, while, case,... editing techniques 3. Controls&Indicators arrays, clusters, charts, graphs 4. Additional lecture State machines,
More informationLabView Add-On Manual
Specialty Plants Specialty Experiment: PIV-plus-Feedforward Position Control Magnetic Levitation Plant LabView Add-On Manual Table of Contents 1. Objectives...1 2. Prerequisites...1 3. References...1 4.
More informationComputerized Measurement systems
Elektrisk Mätteknik, LTH Computerized Measurement systems 1. Introduction to Computerized measurements systems By Tomas Jansson, some text excerpts from NI application note AN007 regarding signal conditioning,
More informationNI-DAQmx Basic Course NITS John Shannon
NI-DAQmx Basic Course NITS 2005 John Shannon Agenda Intro to DAQMX Difference between traditional / DaqMX MAX and DAQ Assistance Analog Input / Output Property Nodes Buffered / Continuous Acquisition Analog
More informationSCXI Contents CALIBRATION PROCEDURE
CALIBRATION PROCEDURE SCXI -1122 Contents This document contains information and step-by-step instructions for calibrating the National Instruments SCXI-1122 module. Introduction... 2 What Is Calibration?...
More informationE/S/M/B Series Calibration Procedure for NI-DAQ mx
E/S/M/B Series Calibration Procedure for NI-DAQ mx Contents This document contains information about calibrating National Instruments E, S, M, and B Series data acquisition (DAQ) devices. This document
More informationUniversity of Illinois Department of Electrical and Computer Engineering ECE 451 Automated Microwave Measurements Laboratory
University of Illinois Department of Electrical and Computer Engineering ECE 451 Automated Microwave Measurements Laboratory Experiment No. 4 Introduction to the Windows XP Workstation using National Instruments
More informationAIO AH-PCI. Features. Packing List. 12-bits Analog I/O Board (High Gain) for PCI AIO AH-PCI 1. Ver.1.04
12-bits alog I/O Board (High Gain) for PCI AIO-121602AH-PCI This product is a PCI-bus compatible multifunction board equipped with analog input x 16ch, analog output x 2ch, digital input/output (unisolated
More informationCALIBRATION PROCEDURE NI PXIe Channel, Isolated Voltage/Current Analog Output Module
CALIBRATION PROCEDURE NI PXIe-4322 8-Channel, Isolated Voltage/Current Analog Output Module This document contains the verification and adjustment procedures for the National Instruments PXIe-4322 module.
More informationComputer Interfacing Using LabView
Computer Interfacing Using LabView Physics 258 Last revised September 25, 2005 by Ed Eyler Purpose: Note: To write a simple LabView program that digitizes data using an ADC on a data acquisition card,
More informationLabVIEW Tutorials. T.1 Temperature Measurement and Cold Junction Compensation (CJC) LUMS School of Science and Engineering
LabVIEW Tutorials T.1 Temperature Measurement and Cold Junction Compensation (CJC) LUMS School of Science and Engineering Umer Hassan & Muhammad Sabieh Anwar Introduction: In this tutorial we shall learn
More informationKistler LabAmp. Electronics & Software. Charge Amplifier and Data Acquisition Unit for Dynamic Measurements. Type 5165A...
Electronics & Software Kistler LabAmp Charge Amplifier and Data Acquisition Unit for Dynamic Measurements This universal laboratory charge amplifier can be used wherever dynamic signals of mechanical quantities
More informationWhere to Start with NI ELVIS
Where to Start with NI ELVIS Conventions The National Instruments Educational Laboratory Virtual Instrumentation Suite (NI ELVIS) is a LabVIEW-based design and prototyping environment for university science
More informationESERIES CALIBRATION FIXTURE INSTALLATION GUIDE
ESERIES CALIBRATION FIXTURE INSTALLATION GUIDE E Series Calibration Fixture for E and S Series Devices The E Series calibration fixture connects the calibration equipment to the E or S Series device you
More informationITZ FLEX ULTRASTAB USER GUIDE /2 Page 1 of 32
ITZ FLEX ULTRASTAB USER GUIDE 140528/2 Page 1 of 32 TABLE OF CONTENTS TABLE OF CONTENTS... 2 LIST OF FIGURES... 3 LIST OF TABLES... 3 DOCUMENT INFORMATION... 3 INTRODUCTION... 4 RECEIVING AND UNPACKING...
More informationBSL Hardware Guide for MP35 and MP30-1 -
BSL Hardware Guide for MP35 and MP30-1 - MP35/30 ACQUISITION UNIT The MP35/30 data acquisition unit is the heart of the Biopac Student Lab PRO System. The MP35/30 has an internal microprocessor to control
More informationPARAMETER MIN TYP MAX UNITS COMMENTS
Phase Noise Measurements to -175dBc/Hz Power supply measurements to 1nV FFT Analysis from DC to 100kHz 0dB / 32dB / 64dB Gain Steps 100dB Dynamic Range Opto-Isolated RoHS PRODUCT SUMMARY The Holzworth
More informationComputerized Measurement Systems (EEMN10) 2016
Computerized Measurement Systems (EEMN10) 2016 CHRISTIAN ANTFOLK & JOSEFIN STARKHAMMAR Course information 2016 Course administrators: Christian Antfolk (christian.antfolk@bme.lth.se) Josefin Starkhammar
More informationComputerized Measurement Systems (EEMN10) 2014
Computerized Measurement Systems (EEMN10) 2014 CHRISTIAN ANTFOLK & JOSEFIN STARKHAMMAR Course information 2014 Course administrators: Christian Antfolk (christian.antfolk@bme.lth.se) Josefin Starkhammar
More informationComputerized Measurement Systems (EEMN10) 2015
Computerized Measurement Systems (EEMN10) 2015 CHRISTIAN ANTFOLK & JOSEFIN STARKHAMMAR Course information 2015 Course administrators: Christian Antfolk (christian.antfolk@bme.lth.se) Josefin Starkhammar
More informationDATENBLATT ADA16-32/2(PCI)F. HABEN SIE FRAGEN ODER WÜNSCHEN SIE EIN INDIVIDUELLES ANGEBOT? Unser Team berät Sie gerne persönlich.
DATENBLATT ADA16-32/2(PCI)F HABEN SIE FRAGEN ODER WÜNSCHEN SIE EIN INDIVIDUELLES ANGEBOT? Unser Team berät Sie gerne persönlich. TELEFON + 49 (0) 81 41/36 97-0 TELEFAX + 49 (0) 81 41/36 97-30 E-MAIL info@plug-in.de
More informationLab #4 The Data Acquisition (DAQ) Card. Read Before Coming to Lab: The Data Acquisition (DAQ) Card (from Appendix A)
The Data Acquisition (DAQ) Card Read Before Coming to Lab: The Data Acquisition (DAQ) Card (from Appendix A) Cards can be inserted into slots in computer mother boards to perform a variety of functions
More informationUniversity College of Southeast Norway. LM-900 Level Tank. Hans-Petter Halvorsen,
University College of Southeast Norway LM-900 Level Tank Hans-Petter Halvorsen, 2016.10.26 http://home.hit.no/~hansha Table of Contents Table of Contents... ii 1 Introduction... 1 2 System Description...
More informationIntroduction to LabVIEW and NI Hardware Platform
Introduction to LabVIEW and NI Hardware Platform Corrie Botha Platform-Based Approach 2 With LabVIEW, You Can Program the Way You Think 3 With LabVIEW, You Can Program the Way You Think The graphical,
More informationTemperature Measurements using the KPCI-3108 with the EXP-1800
Temperature Measurements using the KPCI-3108 with the EXP-1800 Introduction by Andy Toth Keithley Instruments Inc. Many applications require the use of data acquisition hardware to take high channel count
More informationSolar Module Quantum Efficiency Characterization System (LabVIEW Sheds Light on Solar Cell Quality Improvements)
Solar Module Quantum Efficiency Characterization System (LabVIEW Sheds Light on Solar Cell Quality Improvements) by Richard M. Brueggman President & CEO and Gregory C. Cala, Ph.D. Chief Technology Officer
More informationUSB 1608G Series USB Multifunction Devices
USB Multifunction Devices Features 16-bit high-speed USB devices Acquisition rates ranging from 250 ks/s to 500 ks/s differential (DIFF) or 16 singleended (SE) analog inputs (softwareselectable) Up to
More informationA Practical Exercise for the ADC
Richard Haberkamp Nikolas Roosevelt A Practical Exercise for the ADC Janusz Zalewski, Ph. D CDA 4170, Data Acquisition & Control Spring 2009 Table of Contents 1. Equipment Connectivity 3 1.1 Required Hardware
More informationHomework Assignment 9 LabVIEW tutorial
Homework Assignment 9 LabVIEW tutorial Due date: Wednesday, December 8 (midnight) For this homework assignment, you will complete a tutorial on the LabVIEW data acquistion software. This can be done on
More informationAO-1604LX-USB. Features. High-Resolution Analog Output Unit for USB AO-1604LX-USB 1. Ver.1.01
High-Resolution Analog Output Unit for USB AO-1604LX-USB This product is a USB2.0-compliant analog output unit that extends the analog output function of USB port of PCs. This product features high-precision
More informationLab Exercise 4: Analog & Digital I/O Signals
Objective To interface with analog & digital inputs & outputs. When you have completed this exercise, you will be able to: Configure and read analog input data Configure and write analog output data Configure
More informationOVERVIEW SYSTEM REQUIREMENTS
USER GUIDE OVERVIEW The AI-1 USB Audio Interface adds studio-quality Input and Output capabilities to your PC or Mac, turning your recording software into a full recording setup. With its high quality
More information48-Bit Isolated Digital I/O
48-Bit Isolated Digital I/O NI 6527 24 optically isolated digital inputs (0-28 VDC) 24 isolated, solid-state relay digital outputs (0-60 VDC, 0-30 V rms ) Switch up to 120 ma Digital filtering on inputs
More informationProCal & ProCal-Track Calibration Software Features In Depth
Calibration Software ProCal & ProCal Track Calibration Software Full versions of are available to evaluate from the Transmille CD. ProCal-Track runs for 30 days after installation ProCal runs only the
More informationAN-1055 APPLICATION NOTE
AN-155 APPLICATION NOTE One Technology Way P.O. Box 916 Norwood, MA 262-916, U.S.A. Tel: 781.329.47 Fax: 781.461.3113 www.analog.com EMC Protection of the AD7746 by Holger Grothe and Mary McCarthy INTRODUCTION
More informationMain Components of a Static Var Compensator (SVC)
Exercise 1 Main Components of a Static Var Compensator (SVC) EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the simplified diagram of an SVC. You will also be familiar
More informationMeasurement of display transfer characteristic (gamma, )
Measurement of display transfer characteristic (gamma, ) A. Roberts (BBC) EBU Sub group G4 (Video origination equipment) has recently completed a new EBU publication setting out recommended procedures
More informationAME 20213: Fundamentals of Measurements and Data Analysis. LEX-3: Wind Tunnel Pressure Measurement Uncertainty 9 October 2006
AME 20213: Fundamentals of Measurements and Data Analysis LEX-3: Wind Tunnel Pressure Measurement Uncertainty 9 October 2006 TA: John Schmitz Office: B028 Hessert Phone: 1-2317 E-mail : jschmitz@nd.edu
More informationInstructions for using Go!Link Data Acquisition with the Pharmacia UV-1
Instructions for using Go!Link Data Acquisition with the Pharmacia UV-1 Introduction The Go!Link interface, the Instrumentation Amplifier, and the Logger Lite software permit data acquisition from just
More informationThis Presentation Will
Investigating Basic Circuits Pre-Activity Discussion Digital Electronics 2014 Project Lead The Way, Inc. This Presentation Will Introduce you to basic circuits and their symbols. Introduce you to components
More informationIsolated, Field Configurable Analog Input 6B11 / 6B11HV FEATURES APPLICATIONS PRODUCT OVERVIEW
Isolated, Field Configurable Analog Input 6B11 / 6B11HV FEATURES Single-channel isolated signal-conditioning modules. Accepts outputs from Thermocouple, millivolt, volt and current signals. Complete microcomputer-based
More informationWith a digital input we can read two states, a high or low. A switch can be open or closed.
Page 1 of 6 PRODUCT INFORMATION B&B ELECTRONICS Data Acquisition Basics Data Acquisition hardware devices provide an interface between electrical signals a computer can read or can output to control things
More information