FluoroPRO. DICOM conformance Statement

Size: px
Start display at page:

Download "FluoroPRO. DICOM conformance Statement"

Transcription

1 FluoroPRO DICOM conformance Statement Version 1.1

2 Intended Audience It is assumed that any readers of this document are familiar with the DICOM standard Credits and Trademarks CADRAN TM is a trade mark of Cambridge Computed Imaging Ltd. Windows is a registered trademark of Microsoft Corporation SQL Server TM is trademark of Microsoft Corporation FluoroPRO is trademark of Virtual Imaging, Inc. Revision History Revision Comments Date 1.1 Initial version February 28

3 CONFORMANCE STATEMENT OVERVIEW The Cadran Image Storage Managment system implements: provides an overview of the network services supported by FluoroPRO. Network Services SOP Classes User of Service (SCU) Provider of Service (SCP) Transfer X-Ray RadioFluoroscopic Image Storage Secondary Capture Image Storage Yes Yes Query / Retrieve Images Option No Workflow Management Modality Worklist Yes No Storage Commitment Push Model Modality Performed Procedure Step Print Management Basic Grayscale Print Management No No Yes Option Option No No No Media Services Media Storage Application Profile Write Files (FSC or FSU) General Purpose CD-R Yes No General Purpose DVD-RAM Yes No Read Files (FSR)

4 Table of Contents 1. Introduction Purpose: Scope: References: Equipment/Materials: Responsibilities: Definitions: Instructions: Implementation Model Application Data Flow Diagrams Functional definition of AEs STORE_SCP AE: STORE_SCU AE: QR CLIENT AE: PRINT_SCU AE: WORK_SCU AE: MEDIA STORAGE AE: Sequencing of Real World Activities File Meta Information Options AE Specifications STORE_SCP AE Specification Association establishment policies General Number of Associations Asynchronous nature Implementation Identifying Information Association initiation policy Association acceptance policy Real World Activity Receive Echo Associated Real-World Activity Reply to Echo Request.17 Version 1.1 Page 4 of 4 12 th Feb 28

5 Proposed Presentation Contexts SOP Specific Conformance Real World Activity Receive Images from a Remote AE Associated Real-World Activity Accepted Presentation Contexts SOP Specific Conformance Real World Activity Respond to Storage Commitment requests from a Remote AE STORE_SCU AE Specification Association establishment policies General Number of Associations Asynchronous nature Implementation Identifying Information Association initiation policy Association acceptance policy Real World Activity Send Images to Remote AE Associated Real-World Activity Manual Operation System Operation Proposed Presentation Contexts SOP Specific Conformance QR_Client AE Specification Association establishment policies General Number of Associations Asynchronous nature Implementation Identifying Information Association initiation policy Association Acceptance Policy Real World Activity Query a Remote AE Associated Real-World Activity Manual Operation...25 Version 1.1 Page 5 of 5 12 th Feb 28

6 Proposed Presentation Contexts SOP Specific Conformance Real World Activity Move Images from a Remote AE Associated Real-World Activity Manual Operation Proposed Presentation Contexts PRINT_SCU AE Specification Association establishment policies General Number of Associations Asynchronous nature Implementation Identifying Information Association initiation policy Association Acceptance Policy Real World Activity Printing of DICOM images on a Remote AE Associated Real-World Activity Proposed Presentation Contexts SOP Specific Conformance WORK_SCU AE Specification Association establishment policies General Number of Associations Asynchronous nature Implementation Identifying Information Association initiation policy Association Acceptance Policy Real World Activity Query a Remote AE Associated Real-World Activity Manual Operation Proposed Presentation Contexts SOP Specific Conformance Network Communication Profiles Supported Communication Stacks...34 Version 1.1 Page 6 of 6 12 th Feb 28

7 8.2 OSI Stack TCP/IP Stack Physical media support Extensions/Specializations/Privatizations Configuration Media Interchange Implementation Model Application Data Flow Functional Definition of AEs Functional Definition of Offline-Media Application Entity Sequencing of Real World Activities File Meta Information Options AE Specifications Offline-Media Application Entity Specification Real-World Activities Activity Export to CD-R Meda Storage Application Profiles Options AE Title / Presentation Address Mapping Support of Extended Character Sets IOD Contents IOD CONTENTS Created SOP instances X-Ray Radiofluoroscopic Image IOD Secondary Capture IOD Common Modules X-Ray Radiofluoroscopic Image Modules Secondary Capture Image Modules Attribute mapping Coerced/Modified Fields DATA DICTIONARY OF PRIVATE ATTRIBUTES Standard Extended / Specialized / Private SOP Classes X-Ray Radiofluoroscopic Image Storage SOP Class...47 Version 1.1 Page 7 of 7 12 th Feb 28

8 14.4 PRIVATE TRANSFER SYNTAXES...47 Version 1.1 Page 8 of 8 12 th Feb 28

9 1. Introduction 1.1 Purpose: Define the DICOM Conformance statement associated with the Acquisition system 1.2 Scope: This document describes the DICOM Conformance statement in accordance with the document DICOM PS 3.2 Conformance. 1.3 References: DICOM PS 3.2 Conformance DICOM PS 3.3 Information Object Definitions DICOM PS 3.4 Service Class Specifications DICOM PS 3.5 Data Structures and Encoding DICOM PS 3.6 Data Dictionary DICOM PS 3.7 Message Exchange DICOM PS 3.8 Network Communication Support for Message Exchange DICOM PS 3.1 Media Storage and File Format for Data Interchange DICOM PS 3.11 Media Storage Application Profiles DICOM PS 3.12 Media Format and Physical Media for Media Interchange 1.4 Equipment/Materials: N/A 1.5 Responsibilities: N/A 1.6 Definitions: AE Application Entity FSC - File Set Creator FSR - File Set Reader FSU - File Set Updater RWA Real Worls Activity SCU - Service Class User SCP - Service Class Provider SOP Service Object Pair UID Unique Identifier ISM Image Storage Manager 1.7 Instructions: The rest of this document is written in the format specified for DICOM Conformance statements in the DICOM PS 3.2 Conformance standard document. Version 1.1 Page 9 of 9 12 th Feb 28

10 2. Implementation Model implements the necessary DICOM services to: Download work lists from an information system Save images across the network or to a suitable removable medium Print images Optionally, licences can be purchased to provide: Query / retrieve of remote images Receipt of remote images Implementation uses the Cadran libraries, provided by Cambridge Computed Imaging Ltd, which in turn uses the Cedara MergeCOM-3 Toolkit. 3. Application Data Flow Diagrams Figure 1 Store SCP Data flow Local Remote Image Stored in Image File Stored in File System systen Remote Remote System Image System Storage Request List of SOPs created Remote System Sends Storage Commit Request Status of SOPs on local system interrogated. Response sent back to Remote AE INFIMED_ STORE_SCP SCP AE Remote System Receives Storage Commit Notification Respond to Echo request Remote AE Sends Echo Request DICOM standard Version 1.1 Page 1 of 1 12 th Feb 28

11 Figure 2 Store SCU data flow Local System Request Image Storage INFIMED_ STORE_SCU SCU AE Image Stored Remotely Figure 3 QR SCU Client data flow Image Stored in File System Remote System Image Storage Request QR CLIENT AE Local Query / Retrieve Request Remote System Queried Version 1.1 Page 11 of th Feb 28

12 Figure 4 Work SCU Data Flow Local Remote Local Query / Retrieve Request WORK SCU AE Remote System Queried DICOM standard Figure 5 Print SCU Data Flow Local Remote Format Local Print System from Sundial Request Viewer Image Storage Remote Image Basic StoredPrint Remotely SCP Maintain Database Local Queue of Images Queried to Print INFIMED_ PRINT SCU SCU AE Accept Respond Printer to Echo and Print Request Job Updates Remote AE Sends Remote Print Job SCP Echo Request DICOM standard Version 1.1 Page 12 of th Feb 28

13 Figure 6 Media Storage AE Data flow Display Directory Create CD Media Storage AE CD-R/ DVD-R/ DVD-RAM Storage Medium Images Stored in File System Version 1.1 Page 13 of th Feb 28

14 4. Functional definition of AEs FluoroPRO DICOM functionality described above is spread over the following applications: STORE_SCP (a C-STORE SCP), STORE_SCU (ac-store SCU), a C- STORE SCU in response to C-MOVE requests from a remote Query retrieve SCU), QR_CLIENT (a Query Retrieve C-MOVE and C-FIND SCU), WORK_SCU (a Modality Worklist C-FIND SCU), PRINT_SCU (a DICOM Print SCU), and MEDIA STORAGE. 4.1 STORE_SCP AE: STORE_SCP can operate as a command line application or as a Windows NT Service, and provides C-STORE SCP functionality to FluoroPRO. Generally STORE_SCP is configured to begin automatically on system startup and run continuously until shutdown. As a SCP for C-STORE, FluoroPRO listens for storage proposals from remote AEs. Following acceptance by FluoroPRO of an association request from a remote AE, the remote AE transmits the DICOM Information Objects to FluoroPRO. FluoroPRO stores the received Information Object and updates its database. A response is sent to the remote AE indicating the result status of the storage operation. STORE_SCP also functions as a SCP for C-ECHO, and as such listens for communication verification requests from remote AEs, and replies with the C-ECHO response. 4.2 STORE_SCU AE: can operate as a command line application or as a Windows NT Service, and provides C-STORE SCU functionality to FluoroPRO. As a SCU for C-STORE, a user interface command causes FluoroPRO to request an association with a remote system. Following acceptance of the association by the remote system, FluoroPRO transmits the DICOM Information Objects. FluoroPRO returns the result status of the storage request to the user interface accordingly. 4.3 QR CLIENT AE: QR_CLIENT is a windows application that provides C-FIND SCU and C-MOVE SCU functionality to FluoroPRO. As an SCU for C-FIND, a user interface command causes FluoroPRO to request an association with a remote system. Following acceptance of the association by the remote system, FluoroPRO transmits the DICOM Query requests. The remote system returns the matches as DICOM Query responses to FluoroPRO, which displays them in the user interface. As a SCU for C-MOVE, a user interface command causes FluoroPRO to request an association with a remote system. Following acceptance of the association by the remote system, FluoroPRO transmits the DICOM Retrieve requests to the remote system. If a C-MOVE is requested, FluoroPRO will wait for a final response from the remote system returns which it displays the in the user interface. Version 1.1 Page 14 of th Feb 28

15 4.4 PRINT_SCU AE: PRINT_SCU can operate as a command line application or as a Windows NT Service, and provides DICOM Print SCU functionality to FluoroPRO. Images are preformatted by the CADRAN Viewer application, and a request to print them submitted. PRINT_SCU interrogates this print request list and sends the requested pages to the remote Basic Print SCP. PRINT_SCU waits for Print Job status reports from the remote Print SCP, and updates local the records of Print Job Status accordingly. 4.5 WORK_SCU AE: WORK_SCU is a windows application that provides Modality Worklist C-FIND SCU to CADRAN applications. As an SCU for C-FIND, a user interface command causes a request for an association with a remote system. Following acceptance of the association by the remote system, FluoroPRO transmits the DICOM Query requests. The remote system returns the matches as DICOM Query responses to FluoroPRO, which displays them in the user interface. 4.6 MEDIA STORAGE AE: The Media Storage AE can perform the following functions: It can initialize a piece of media, writing a new DICOM File-set onto the media. It can display a directory listing of the File-set on a piece of media. It can copy SOP instances from the media onto local storage. 5. Sequencing of Real World Activities NA 6. File Meta Information Options Implementation Class UID = Implementation Version = CADRAN_21 The Implementation Class UID is part of the File Meta Information written into every file and therefore necessary for any device that acts as an FSC. Version 1.1 Page 15 of th Feb 28

16 7. AE Specifications 7.1 STORE_SCP AE Specification The STORE_SCP AE provides Standard Conformance to the DICOM V3. SOP Classes listed in Table 1 as an SCP. Note that the STORE_SCP AE is highly configurable, and as such not all classes in the list may be accepted when using the default configuration, i.e. configuration changes may be required (see section 1). Table 1 SOP Class UID SOP Class Verification SOP Class Secondary Capture Image Storage Multi-Frame Greyscale Byte Secondary Capture Image Storage Multi-Frame Greyscale Word Secondary Capture Image Storage Multi-Frame True Colour Secondary Capture Image Storage Standard Xray Angio Standard Xray RF Association establishment policies General The AE Title of the STORE_SCP AE is a configurable parameter. The default title is FP_ST_SCP. The STORE_SCP AE waits for an association as an SCP of the Storage Services. The maximum PDU size is configurable from a minimum of 4,96 bytes Number of Associations The STORE_SCP AE will handled up to 5 storage SCP associations simultaneously Asynchronous nature The STORE_SCP AE does not support asynchronous operations Implementation Identifying Information The Implementation Class Unique Identifier (UID) for the STORE_SCP AE is: The Implementation version name for the STORE_SCP AE is: CADRAN_21 Version 1.1 Page 16 of th Feb 28

17 Association initiation policy The AE Title of the STORE_SCP AE is a configurable parameter. The default title is FP_ST_SCP. STORE_SCP AE will attempt to establish an association to a remote Storage Commit SCU to facilitate a N-EVENT-REPORT reply to a N-ACTION Storage Commit request. The maximum PDU size is configurable from a minimum of 4,96 bytes Association acceptance policy The STORE_SCP AE accepts associations for the store and verification services Real World Activity Receive Echo Associated Real-World Activity Reply to Echo Request The associated activity is a C-ECHO request Proposed Presentation Contexts The STORE_SCP AE will propose Presentation Contexts as shown in Table 2. Table 2 SCP Presentation Context Table Abstract Syntax Transfer Syntax UID List UID List Verification Service Class Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian Role SCP Extended Negotiation SOP Specific Conformance The STORE_SCP AE provides standard conformance to the DICOM Verification Service Class as a SCP Real World Activity Receive Images from a Remote AE Associated Real-World Activity The associated activity is an incoming C-STORE request. Following acceptance of the association the STORE_SCP AE receives the images on the open association. The internal database is updated on transfer of each image. An appropriate response is returned. Version 1.1 Page 17 of th Feb 28

18 Accepted Presentation Contexts STORE_SCP AE will accept presentation contexts as shown in Table 3. Note that configuration changes (as outlined in section 1) may be required for STORE_SCP to accept a paticular Abstract Syntax/Transfer Syntax pair. Note that when STORE_SCP receives an uncompressed image of implicit little endian or explicit big endian, it is synchronously transformed to explicit little endian before storage. Uncompressed X- ray angio images (Abstract syntax UID ) are asynchronously transformed to transfer syntax JPEG lossless Hier14 (UID ) before storage. Images of all other transfer syntaxes are stored in the transfer syntax in which they are received. Table 3 Presentation Context Table Abstract Syntax Transfer Syntax UID List UID List SC Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian RLE JPEG Baseline JPEG Lossless Hier Role SCP Extended Negotiation Multi-Frame Greyscale Byte Secondary Capture Image Storage Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian RLE SCP JPEG Baseline JPEG Lossless Hier Multi-Frame Greyscale Word Secondary Capture Image Storage Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian RLE SCP JPEG Baseline JPEG Lossless Hier Version 1.1 Page 18 of th Feb 28

19 Multi-Frame True Colour Secondary Capture Image Storage Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian RLE SCP JPEG Baseline JPEG Lossless Hier Xray Angio Explicit VR Little Endian Implicit VR Little Endian SCP Explicit VR Big Endian RLE JPEG Baseline JPEG Lossless Hier Xray RF Explicit VR Little Endian Implicit VR Little Endian SCP Explicit VR Big Endian RLE JPEG Baseline JPEG Lossless Hier SOP Specific Conformance The STORE_SCP AE conforms to the Storage SOP Class as an SCP at Level 2 (Full) as defined in PS3.4 (B.4.1). After storage, STORE_SCP AE responds as shown in Table 4. Table 4 Service Status Further Meaning Comments Status Codes Related Fields Refused Out of resources A FIFO policy can be configured to delete old data. A7xx (,92) Error Identifier does not match SOP Class Incorrectly formed request A9xx (,91) (,92) Version 1.1 Page 19 of th Feb 28

20 Error Cannot Incomprehensible Cxxx (,91) understand data (,92) Success Real World Activity Respond to Storage Commitment requests from a Remote AE. Please note that this functionality is only present in versions 2.1 and later of FluoroPRO. 7.2 STORE_SCU AE Specification The STORE_SCU AE provides Standard Conformance to the DICOM V3. SOP Classes listed in Table 5 as a SCU. Note that the STORE_SCU AE is highly configurable, and as such, not all classes in the list may be accepted when using the default configuration, i.e. configuration changes may be required (see section 1). Table 5 SOP Class UID SOP Class Secondary Capture Image Storage Multi-Frame Greyscale Byte Secondary Capture Image Storage Multi-Frame Greyscale Word Secondary Capture Image Storage Multi-Frame True Colour Secondary Capture Image Storage Standard Xray Angio Standard Xray RF Association establishment policies General The AE Title of the STORE_SCU AE is a configurable parameter. The default title is FP_ST_SCU. STORE_SCU AE will attempt to establish an association when it is invoked from the user interface or through a system message to facilitate a C-MOVE request. The maximum PDU size is configurable from a minimum of 4,96 bytes Number of Associations The STORE_SCU AE will only initiate on assocaition at a time Asynchronous nature The STORE_SCU AE does not support asynchronous operations. Version 1.1 Page 2 of 2 12 th Feb 28

21 Implementation Identifying Information The Implementation Class Unique Identifier (UID) for the STORE_SCU AE is: The Implementation version name for the STORE_SCU AE is: CADRAN_ Association initiation policy The STORE_SCU AE attempts to initiate an association for the following real world activity: Send images from STORE_SCU AE to a Remote AE Association acceptance policy The STORE_SCU AE does not accept associations Real World Activity Send Images to Remote AE Associated Real-World Activity The associated real-world activity is a C-STORE Request initiated by the user when images are selected for transfer to a remote AE Manual Operation The user selects the remote system and image sets to transfer in the user interface. Committing the transfer causes STORE_SCU AE to initiate an association with the remote AE for a C-STORE request System Operation In certain situations, STORE_SCU AE will initiate an asociation within a user request for a remote system for a C-STORE request Proposed Presentation Contexts STORE_SCU AE will accept presentation contexts as shown in Table 6. Note that FluoroPRO is not capable of translating to or from a particular encapsulated transfer syntax. If an image is received by the STORE_SCP in a given encapsulated transfer syntax, it can only be sent by the STORE_SCU in that same transfer syntax. STORE_SCU can freely translate between explicit little endian, implicit little endian and explicit big endian transfer syntaxes. Table 6 Store SCU Presentation Context Table Version 1.1 Page 21 of th Feb 28

22 Abstract Syntax Transfer Syntax UID List UID List SC Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian RLE JPEG Baseline JPEG Lossless Hier Role SCU Extended Negotiation Multi-Frame Greyscale Byte Secondary Capture Image Storage Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian RLE SCU JPEG Baseline JPEG Lossless Hier Multi-Frame Greyscale Word Secondary Capture Image Storage Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian RLE SCU JPEG Baseline JPEG Lossless Hier Multi-Frame True Colour Secondary Capture Image Storage Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian RLE SCU JPEG Baseline JPEG Lossless Hier Version 1.1 Page 22 of th Feb 28

23 Xray Angio Explicit VR Little Endian Implicit VR Little Endian SCU Explicit VR Big Endian RLE JPEG Baseline JPEG Lossless Hier Xray RF Explicit VR Little Endian Implicit VR Little Endian SCU Explicit VR Big Endian RLE JPEG Baseline JPEG Lossless Hier Table 7 shows those transfer syntaxes for which there is limited supported by STORE_SCU. FluoroPRO is able to receive images in these transfer syntax, but will not be able to translate them to another encapsulated syntax. Any images received in these transfer syntaxes will be stored in the transfer syntax in which they are received, and can only be transmitted by the STORE SCU in this same syntax. Table 7 Possibly Supported Transfer syntax list UID JPEG Extended JPEG Extended JPEG Spec Non-Hier JPEG Spec Non-Hier JPEG Full Prog Non-Hier JPEG Full Prog Non-Hier JPEG Lossless Non-Hier JPEG Lossless Non-Hier JPEG Extended Hier JPEG Extended Hier JPEG Spec Hier JPEG Spec Hier JPEG Full Prog Hier JPEG Full Prog Hier JPEG Lossless Hier Version 1.1 Page 23 of th Feb 28

24 JPEG Lossless Hier SOP Specific Conformance The DICOM images sent by STORE_SCU AE conform to the standard DICOM IOD definitions (see PS3.3). 7.3 QR_Client AE Specification The QR_CLIENT AE provides Standard Conformance to the DICOM V3. SOP Classes listed in Table 8 as a SCU. Table 8 SOP Class UID SOP Class Study Root Query/Retrieve Information Model Find Study Root Query/Retrieve Information Model Move Association establishment policies General The AE Title of the QR_CLIENT AE is a configurable parameter. The default title is FP_QR_SCU. QR_CLIENT AE will attempt to establish an association when it is invoked from the user interface. The maximum PDU size is configurable from a minimum of 4,96 bytes Number of Associations The QR_CLIENT AE will only initiate one association at a time Asynchronous nature The QR_CLIENT AE does not support asynchronous operations Implementation Identifying Information The Implementation Class Unique Identifier (UID) for the QR_CLIENT AE is: The Implementation version name for the QR_CLIENT AE is: CADRAN_ Association initiation policy The QR_CLIENT AE attempts to initiate an association for the following real world activity: Version 1.1 Page 24 of th Feb 28

25 The Query of objects on a Remote AE Association Acceptance Policy The QR_CLIENT AE does not accept associations Real World Activity Query a Remote AE Associated Real-World Activity Querying is associated with the C-FIND request Manual Operation Querying is a manual operation performed from the user interface. The user may select the remote AE, the tpe of query and the desired keys from the available options. On initiation of the query from the user interface, an association is attempted with the remote AE. Once the association is established, QR_CLIENT will perform as many C- FIND requests as are necessary to complete the query. Ther results from the query, successful or otherwise are displayed in the user interface Proposed Presentation Contexts QR_CLIENT AE will propose presentation contexts as shown in Table 9. Table 9 Query SCU Presentation Context Table Abstract Syntax Transfer Syntax UID List UID List Study Root Query/Retrieve Information Model FIND Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian Role SCU Extended Negotiation SOP Specific Conformance QR_CLIENT AE can request hierarchical queries only. Relational queries are not implemented. The keys available for display and searching are shwon in Table 1, along with the level of query required to retrieve them. A search may be initiated on those DICOM tags shown. Table 1 Description DICOM Tag Study/Series/Ima ge Comments Patient's ID (1,1) Study Search Version 1.1 Page 25 of th Feb 28

26 Patient's Patient's Birth Date (1,2) Study Search (1,3) Study Not displayed Patient's Sex (1,4) Study Not displayed Study Instance UID Modalities in Study (2,D) Study Unique key (8,61) Study Search Study Date (8,2) Study Search Study Time (8,3) Study Not displayed Accession Number Referring Physician's name Study Description (8,5) Study Not displayed (8,9) Study Not displayed (8,13) Study Study ID (2,1) Study Not displayed Number of study related Series Number of study related Images Series Instance UID (2,126) Study (2,128) Study (2,E) Series Unique Key Modality (8,6) Series Series Description Contrast/Bolu s agent Body Part Examined Series Number (8,13E) Series (18,1) Series (18,15) Series (2,11) Series Version 1.1 Page 26 of th Feb 28

27 Number of Series related Images SOP InstanceUID Image Number Image Position (Patient) Image Orientation (Patient) (2,129) Series (8,18) Image Unique Key (2,13) Image (2,32) Image Not displayed (2,37) Not displayed Real World Activity Move Images from a Remote AE Associated Real-World Activity Moving images from the remote AE after a query operation is associated with the C- MOVE request Manual Operation Moving is a manual operation which follows a query operation. The user may select the items to move from the results generated by the previous query. The user can also select the destination AE of the move operation. On initiation of the move from the user interface, an association is attempted with the remote AE. Following acceptance of the association, QR_CLIENT will perform as many C-MOVE requests as are necessary to complete the move. The results from the move, successful or otherwise are displayed in the user interface Proposed Presentation Contexts QR_CLIENT AE will propose presentation contexts as shown in Table 11. Table 11 Move SCU Presentation Context Table Abstract Syntax Transfer Syntax UID List UID List Study Root Query/Retrieve Information Model MOVE Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian Role SCU Extended Negotiation Version 1.1 Page 27 of th Feb 28

28 7.4 PRINT_SCU AE Specification The PRINT_SCU AE provides Standard Conformance to the DICOM V3. SOP Classes listed in Table 12 as a SCU. Table 12 SOP Class UID Print Job SOP Class Association establishment policies Basic Grayscale Print Management (META) Basic Film Session Basic Film Box Basic Grayscale Image Box Printer General The AE Title of the PRINT_SCU AE is a configurable parameter. The default title is PRINT_SCU. PRINT_SCU AE will attempt to establish an association when it is invoked from the user interface. The maximum PDU size is configurable from a minimum of 4,96 bytes Number of Associations The PRINT_SCU AE will only initiate one association at a time Asynchronous nature The QR_CLIENT AE does not support asynchronous operations Implementation Identifying Information The Implementation Class Unique Identifier (UID) for the PRINT_SCU AE is: The Implementation version name for the QR_CLIENT AE is: CADRAN_ Association initiation policy The PRINT_SCU AE attempts to initiate an association for the following real world activity: The Printing of DICOM images on a Remote AE. The PRINT_SCU AE initiates an association for the appropriate Print Services Class that corresponds to the set of images requested to be printed. The association is Version 1.1 Page 28 of th Feb 28

29 closed when all images have been printed and all print jobs have completed. The association is aborted should a time-out occur Association Acceptance Policy PRINT_SCU does not accept associations Real World Activity Printing of DICOM images on a Remote AE Associated Real-World Activity Once the Print Image association has been established, PRINT_SCU, sends a Basic Film Session, N_CREATE message to the Basic Print SCP. This is followed by an Basic Film Box, N_CREATE message. The PRINT_SCU then sends a Basic Grayscale Image Box, N_SET message. Finally, an N_ACTION message is sent to instruct the Basic Print SCP to print either at the Basic Film Session (if collation is supported by the remote SCP) or at the Basic Film Box level (if collation is not supported). While the association with the remote SCP is open, PRINT_SCU will listen for and process N_EVENT_REPORT messages from the remote SCP. If the remote SCP supports the Print Job SOP Class, then the SCU will periodically poll the remote SCP for Print Job Status reports until all submitted print jobs are finished Proposed Presentation Contexts PRINT_SCU AE will propose presentation contexts as shown in Table 13. Table 13 Presentation Context Table Abstract Syntax Transfer Syntax UID List UID List Basic Grayscale Print Management (META) Print Job Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian Role SCU. SCU Extended Negotiation SOP Specific Conformance Attribute values for SOP classes proposed by PRINT_SCU are specified in Table 14. Table 14 SOP Class Command Attribute name Valid Range Default Version 1.1 Page 29 of th Feb 28

30 Basic Film Session Basic Film Box Basic Greyscale image box N_CREATE N_ACTION N_CREATE N_ACTION N_SET Number of Copies Medium Type 1 1 PAPER, CLEAR FILM, BLUE FILM Film Destination N/A Film Session Label Memory Allocation Referenced Print Job Sequence Image Display format Film Orientation N/A N/A Customisable in software PORTRAIT, LANDSCAPE Film Size ID N/A Magnification Type N/A STANDARD\1,1 PORTRAIT Max Density N/A Configuration Information N/A Smoothing Type N/A Border Density N/A Empty Image Density N/A Min Density N/A Trim N/A Referenced Print Job Sequence Image Position 1 1 Samples per pixel 1 1 Photometric Interpretation Rows MONOCHRO ME2 Customisable in software MONOCHROME 2 Version 1.1 Page 3 of 3 12 th Feb 28

31 Printer Print Job N_GET/ N_EVENT_ REPORT N_EVENT_ REPORT Columns Pixel Aspect Ratio Bits Allocated Bits Stored High Bit Pixel Representation Customisable in software 1/1 1/1 Customisable in software Customisable in software Customisable in software Polarity N/A Magnification Type N/A Smoothing Type N/A Requested Image Size N/A Printer Status * Printer Status Info * Printer * Manufacturer * Manufacturer Model Number * Software Version * Execution Status * Execution Status Info * Print Priority * Creation Date * Creation Time * Printer * Originator * Version 1.1 Page 31 of th Feb 28

32 7.5 WORK_SCU AE Specification The CADRAN_WORK_SCP AE provides Standard Conformance to the DICOM V3. SOP Classes listed in Table 15 as a SCU. Table 15 SOP Class UID SOP Class Modality Worklist Find Find Association establishment policies General The AE Title of the WORK_SCU AE is a configurable parameter. The default title is WORK_SCU. WORK_SCU AE will attempt to establish an association when it is invoked from the user interface. The maximum PDU size is configurable from a minimum of 4,96 bytes Number of Associations WORK_SCU will only initiate one association at a time Asynchronous nature The WORK_SCU AE does not support asynchronous operations Implementation Identifying Information The Implementation Class Unique Identifier (UID) for the WORK_SCU AE is: The Implementation version name for the WORK_SCU AE is: CADRAN_ Association initiation policy The WORK_SCU AE attempts to initiate an association for the following real world activity: The Query of objects on a Remote AE Association Acceptance Policy WORK_SCU does not accept associations. Version 1.1 Page 32 of th Feb 28

33 7.5.2 Real World Activity Query a Remote AE Associated Real-World Activity Querying is associated with the C-FIND request Manual Operation Querying is a manual operation performed from the user interface. The user may select the remote AE, the type of query and the desired keys from the available options. On initiation of the query from the user interface, an association is attempted with the remote AE. Once the association is established, WORK_SCU will perform as many C-FIND requests as are necessary to complete the query. Ther results from the query, successful or otherwise are displayed in the user interface Proposed Presentation Contexts WORK_SCU AE will propose presentation contexts as shown in Table 16. Table 16 Presentation Context Table Abstract Syntax Transfer Syntax UID List UID List Modality Worklist Find Explicit VR Little Endian Implicit VR Little Endian Explicit VR Big Endian Role SCU Extended Negotiation SOP Specific Conformance The keys available for display and searching by WORK_SCU AE are shown in Table 17. Table 17 Description DICOM Tag Required? Search? Study Instance UID (2,D) Y N Requested Procedure ID (4,11) Y N Requested Procedure Description (32,16) C N Referring Physician s name (8,9) N N Accession Number (8,5) N N Patients (1,1) Y Y Version 1.1 Page 33 of th Feb 28

34 Patients ID (1,2) Y Y Patients Birth Date (1,3) N N Patients Sex (1,4) N N Modality (8,6) Y Y Scheduled Station AE (4,1) Y Y Scheduled Procedure Start Date Scheduled Procedure Start Time Scheduled Performing Physicians Scheduled Procedure Step ID Scheduled Procedure Step Description 8. Network Communication Profiles (4,2) Y Y (4,3) Y Y (4,6) N Y (4,9) Y N (4,7) C N 8.1 Supported Communication Stacks FluoroPRO provides DICOM V3. TCP/IP Network Communication Support as defined in Part 8 of the DICOM Standard (PS 3.8 Network Communication Support for Message Exchange). 8.2 OSI Stack No OSI Stack communications are provided. 8.3 TCP/IP Stack FluoroPRO inherits its TCP/IP stack from the operating system on which it executes. These are currently one of the following: Microsoft Windows XP 8.4 Physical media support FluoroPRO is indifferent to the physical medium over which TCP/IP executes; it inherits this from the operating system on which it executes. 9. Extensions/Specializations/Privatizations Not applicable. Version 1.1 Page 34 of th Feb 28

35 1. Configuration The FluoroPRO products obtains their configuration information from the following files: merge.ini - Identifies the other three configuration files and also contains message logging parameters. The location of this file is specified by specific register settings, which may be set as described in the operator s manual. mergecom.pro - Defines run-time parameters for FluoroPRO product applications, such as the network timeouts and PDU size, etc. Most parameters in this file should never be changed, as doing so could disrupt the DICOM interface conformance. mergecom.app Defines applications on other network nodes (such as supported image types, transfer syntaxes, application entities, etc.) to which connections are possible. mergecom.srv - Service and sequence definitions. This file does not need to be altered, as long as standard DICOM image types are transferred. Most parameters in this file should never be changed, as doing so could disrupt the DICOM interface conformance. Database network nodes can also be defined within the database to which the application is connected. 11. Media Interchange 11.1 Implementation Model Application Data Flow Export to CD-R Offline-Media Application Entity CD-R Storage Medium Figure 7. Application data flow fiagram for media storage Version 1.1 Page 35 of th Feb 28

36 Functional Definition of AEs Functional Definition of Offline-Media Application Entity Selection of the Export to Media button will open the Exporter Application. One or more studies may be selected for export to the selected media Sequencing of Real World Activities At least one study must be selected before media creation can commence. Blank media should be placed in the writer before commencing though a reminder will be displayed if it is not present File Meta Information Options The implementation information written to the File Meta Header in each file is: Table 18 Implementation class UID AE Specifications Offline-Media Application Entity Specification The Offline-Media Application Entity provides standard conformance to the Media Storage Service Class. The Application Profiles and roles are listed below: Table 19 Application Profiles Supported Real World Activity Role STD-GEN-CD Export to CD-R FSC SRD-GEN-DVD-RAM Export to DVD-RAM FSC Real-World Activities Selecting the Write to Disk command button will initiate the process of copying all the studies listed in the right hand window pane to CD/DVD. The program will inform you how many disks will be required, and will prompt you to continue or cancel at several points during the processing Activity Export to CD-R Meda Storage Application Profiles The Offline-Media Application Entity support the STD-GEN-CD and STD-GEN-DVD- RAM Application Profile Options Version 1.1 Page 36 of th Feb 28

37 The Offline-Media Application Entity supports the SOP Classes and Transfer Syntaxes listed in the Table below: Table 2 Information Object Definition Media Storage Directory Storage X-Ray Radio Fluoroscopic Image Storage SOP Class UID Transfer Syntax Explicit VR Little Endian Explicit VR Little Endian Transfer Syntax UID AE Title / Presentation Address Mapping Presentation address mapping for remote SCP s is configured in the mergecom.app file. This is where the Host, Port Number, and Application Title map an Application Entity (AE) Title to a Presentation Address in TCP/IP for the provider to which you wish to connect is stored. The Presentation Address of local SCP and SCU s is defined as shown in the FluoroPRO products operation manuals. Note: The host name maps to an IP address as specified by your host table. Generally port 14 should be used for standard connectivity; since this is the well-defined port for a DICOM server. 13. Support of Extended Character Sets Not supported. 14. IOD Contents 14.1 IOD CONTENTS Created SOP instances Table 21 specifies the attributes of an X-Ray Radiofluoroscopic Image transmitted by the FluoroPRO storage application. The following tables use a number of abbreviations. The abbreviations used in the Presence of column are: VNAP ANAP ALWAYS EMPTY Value Not Always Present (attribute sent zero length if no value is present) Attribute Not Always Present Always Present Attribute is sent without a value Version 1.1 Page 37 of th Feb 28

38 The abbreviations used in the column: MWL USER AUTO MPPS CONFIG the attribute value source Modality Worklist the attribute value source is from User input the attribute value is generated automatically the attribute value is the same as that use for Modality Performed Procedure Step the attribute value source is a configurable parameter NOTE: All dates and times are encoded in the local configured calendar and time. Date, Time and Time zone are configured using the Service/Installation Tool X-Ray Radiofluoroscopic Image IOD Table 21. IOD of created RF SOP instances IE Module Reference Presence of Module Patient Patient Table 23 ALWAYS Study General Study Table 24 ALWAYS Patient Study Table 25 ALWAYS Series General Series Table 26 ALWAYS Equipment General Equipment Table 27 ALWAYS Image General Image Table 29 ALWAYS Image Pixel Table 3 ALWAYS Cine Table 31 Only if Multi-frame Multi-Frame Table 32 Only if Multi-frame Frame Pointers Only if Multi-frame X-Ray Image Table 33 ALWAYS X-Ray Acquisition Table 34 ALWAYS Modality LUT Table 35 Only if Pixel Intensity Relationship (28,14) is LOG VOI LUT Table 36 ALWAYS SOP Common Table 37 ALWAYS Private Application ALWAYS Secondary Capture IOD Table 22. IOD of Created Secondary Capture IODs IE Module Reference Presence of Module Patient Patient Table 23 ALWAYS Version 1.1 Page 38 of th Feb 28

39 Study General Study Table 24 ALWAYS Patient Study Table 25 ALWAYS Series General Series Table 26 ALWAYS Equipment General Equipment Table 27 ALWAYS SC Equipment ALWAYS Image General Image Table 29 ALWAYS Image Pixel Table 3 ALWAYS Cine Table 31 Only if Multi-frame Multi-Frame Table 32 Only if Multi-frame Frame Pointers Only if Multi-frame SC Image Module SC Multi-frame Image Module Only if Multi-frame SOP Common Table 37 ALWAYS Common Modules Table 23. Patient module of created SOP instances Attribute Tag VR Value Patient s Patient ID (1,1) PN (1,2) LO Patient s Birth Date (1,3) DA Patient s Sex (1,4) CS Patient Comments (1,4) LT From Modality Worklist or user input. Values supplied via Modality Worklist will be entered as received. Values supplied via user input will contain all 5 components (some possibly empty). Maximum 64 characters. From Modality Worklist or user input. Maximum 64 characters. From Modality Worklist or user input From Modality Worklist or user input From User Input. Maximum 124 characters. Presence of Value VNAP VNAP VNAP VNAP ANAP MWL/ USER MWL/ USER MWL/ USER MWL/ USER USER Table 24. General study module of created SOP instances Attribute Tag VR Value Study Instance UID (2,D) UI From Modality Worklist or generated by device Presence of Value ALWAYS MWL/ AUTO Version 1.1 Page 39 of th Feb 28

40 Study Date (8,2) DA <yyyymmdd> ALWAYS AUTO Study Time (8,3) TM <hhmmss> ALWAYS AUTO Referring Physician s (8,9) PN Study ID (2,1) SH From Modality Worklist or User Input Requested Procedure ID from Worklist or User Input VNAP VNAP MWL/ USER MWL/ USER Accession Number (8,5) SH From Modality Worklist VNAP MWL Study Description (8,13) LO Comment text box in study list. Maximum 124 characters. VNAP USER Table 25. Patient study module of created SOP instances Attribute Tag VR Value Patient s Age (1,11) AS Calculated from DoB input on base of actual Date Presence of Value VNAP AUTO Patient s Size (1,12) DS Patient height (m) from user input VNAP USER Patient s Weight (1,13) DS From user input VNAP USER Table 26. General series module of created SOP instances Attribute Tag VR Value Presenc e of Value Modality (8,6) CS RF ALWAYS AUTO Series Instance UID (2,E ) UI Generated by device ALWAYS AUTO Series Number (2,11) IS Generated by device ALWAYS AUTO Series Date (8,21) DA <yyyymmdd> ALWAYS AUTO Series Time (8,31) TM <hhmmss> ALWAYS AUTO Performing Physician s Series Description (8,15) PN (8,13E ) LO Physician field in Study list. Maximum 64 characters. From User or Modality Worklist. Organ from Study list. Maximum 512 characters. VNAP ANAP USER/ MWL USER Table 27. General equipment module of created SOP instances Attribute Tag VR Value Presence of Value Manufacturer (8,7) LO MANUFACTURER ALWAYS AUTO Institution (8,8) LO From Configuration VNAP CONFIG Version 1.1 Page 4 of 4 12 th Feb 28

41 Station (8,11) SH From Configuration ALWAYS CONFIG Manufacturer s Model Device Serial Number Software Version (8,19) LO From Implementation ALWAYS AUTO (18,1) LO From Configuration ALWAYS CONFIG (18,12) LO From Configuration ALWAYS CONFIG Table 28. Private application module of created SOP instances Attribute Tag VR Value Presence of Value Private Creator (9,xx) LO Cadranism Private Group ALWAYS AUTO Annotation String (9,xx1) LO Zero or greater length. XML rendition of annotation information VNAP AUTO X-Ray Radiofluoroscopic Image Modules Table 29. General image module of created RF SOP instances Attribute Instance Number Patient Orientation Acquisition Date Tag VR Value Presence of Value (2,13) IS Generated by device ALWAYS AUTO (2,2) CS Zero length EMPTY AUTO (8,22) DA <yyyymmdd> ALWAYS AUTO Content Date (8,23) DA <yyyymmdd> ALWAYS AUTO Acquisition Time (8,32) TM <hhmmss> ALWAYS AUTO Content Time (8,33) TM <hhmmss> ALWAYS AUTO Acquisition Number Image Comments (2,12) IS Generated by device (currently blank) ALWAYS AUTO (2,4) LT From user input. Maximum 124 characters. VNAP USER Table 3. Image pixel module of created RF SOP instances Attribute Tag VR Value Presence of Value Pixel Data (7FE,1) OW The Pixel Data itself does not contain any burned-in annotation. ALWAYS AUTO Version 1.1 Page 41 of th Feb 28

42 Table 31. Cine module of created RF sop Attribute Tag VR Value Presence of Value Frame Time (18,163) DS Only if multi-frame. ANAP AUTO Table 32. Multi-frame module of created RF SOP instances Attribute Number of Frames Tag VR Value Presence of Value (28,8) IS Only if multi-frame ANAP AUTO Table 33. X-ray image module of created RF SOP instances Attribute Frame Increment Pointer Image Type Pixel Intensity Relationship Samples per Pixel Photometric Interpretation Tag VR Value Presence of Value (28,9) AT <18,163> only if multi-frame ANAP AUTO (8,8) CS ORIGINAL\PRIMARY\SINGLE PLANE\DAS (acquired images) ORIGINAL\DERIVED\SINGLE PLANE (post-processed images) ALWAYS AUTO (28,14) CS LIN or LOG ALWAYS AUTO (28,2) US 1 ALWAYS AUTO (28,4) CS MONOCHROME2 ALWAYS AUTO Rows (28,1) US 124 ALWAYS AUTO Columns (28,11) US 124 ALWAYS AUTO Bits Allocated (28,1) US 16 ALWAYS AUTO Bits Stored (28,11) US 12 ALWAYS AUTO High Bit (28,12) US 11 ALWAYS AUTO Pixel Representation (28,13) US H ALWAYS AUTO Table 34. X-ray acquisition module of created RF SOP instances Attribute Tag VR Value Presence of Value KVP (18,6) DS From Acquisition parameters ALWAYS AUTO Version 1.1 Page 42 of th Feb 28

DICOM 3.0 Conformance Statement for PlatinumOne Systems

DICOM 3.0 Conformance Statement for PlatinumOne Systems DICOM 3.0 Conformance Statement for PlatinumOne Systems 0.0 Revision History: Revision ECN Date Description A 3032 09/12/2002 New Release 1.0 Purpose: Define the DICOM Conformance statement associated

More information

SCORE 3D Workstation DICOM CONFORMANCE STATEMENT

SCORE 3D Workstation DICOM CONFORMANCE STATEMENT Document No.: S517-E105 Revision No.: A DIGITAL ANGIOGRAPHY SYSTEM DAR-9500f SCORE 3D Workstation DICOM CONFORMANCE STATEMENT Revision History Rev. Date Content of Change First 2013.09 Newly issued. A

More information

Punctual Dicom Workstation

Punctual Dicom Workstation DICOM Conformance Statement Punctual Dicom Workstation Company Name Product Name Product Version 2.0 Document number 2008-1001 Date June 1, 2009 Punctual Software Inc. Punctual Dicom Workstation TABLE

More information

Raffaello Conformance Statement for DICOM V3.0

Raffaello Conformance Statement for DICOM V3.0 Raffaello Conformance Statement for DICOM V3.0 Copyright 2003-2006 by I.M.S. s.r.l. DOCUMENT VERSIONS Version Date Author Changes 1.00 04-Oct-03 IMS s.r.l. First Version 1.01 27-Feb-04 IMS s.r.l. Updated

More information

AVIA (Dx MM) DICOM 3.0 Conformance Statement

AVIA (Dx MM) DICOM 3.0 Conformance Statement AVIA (Dx MM) DICOM 3.0 Conformance Statement AVIA (Dx MM) is a product of Medasys Digital Systems. Table of Contents 1INTRODUCTION... 4 1.1SCOPE AND AUDIENCE... 4 1.2REFERENCES... 4 1.3ACRONYMS AND ABBREVIATIONS...

More information

Copyright FUJIFILM Corporation, Japan. August, th Edition 897N100760F

Copyright FUJIFILM Corporation, Japan. August, th Edition 897N100760F DICOM Conformance Statement ***** FDR-1000AWS CR-IR363AWS for DICOM Storage DICOM Storage Commitment DICOM MWM DICOM MPPS DICOM Print DICOM Query / Retrieve DICOM Media Storage (Standard) August, 2010

More information

DAR-8000 Series DICOM Conformance Statement

DAR-8000 Series DICOM Conformance Statement Series DICOM Conformance Statement for 03.01.xx No. : S517-1070 Rev. : D [NO TEXT] 2 0.0 Revision History: Revision Date Description First Edition 2005/08/03 New Release A 2005/11/09 Added Worklist B 2006/03/02

More information

1 Introduction Scope and audience References Acronyms and abbreviations 3. 2 Implementation model... 4

1 Introduction Scope and audience References Acronyms and abbreviations 3. 2 Implementation model... 4 NAME FONCTION VISA AUTHOR Alain Battistini Senior Engineer CONFORM TO ORIGINAL REVIEW Jean-François Hoh Senior Engineer CONFORM TO ORIGINAL APPROVAL Pascal Gilbert-Jeantet CONFORM TO ORIGINAL Revision

More information

Version 7 DICOM Conformance Statement. Document Version 3.02, June 2009

Version 7 DICOM Conformance Statement. Document Version 3.02, June 2009 Version 7 DICOM Conformance Statement Document Version 3.02, June 2009 1 Conformance Statement Overview The application described in this Conformance Statement VEPRO EMR Manager PACS is a collection of

More information

Aloka ProSound DICOM

Aloka ProSound DICOM 11/18/09 Page 1 of 17 Aloka ProSound DICOM Aloka ProSound C3 and C3cv Ultrasound Imaging System ProSound Versions 1.0 and 1.1 DICOM 11/18/09 Page 2 of 17 Table of Contents 1 Introduction...4 1.1 Scope...4

More information

Version 9 DICOM Conformance Statement. Version 3.05, September 2015

Version 9 DICOM Conformance Statement. Version 3.05, September 2015 Version 9 DICOM Conformance Statement Version 3.05, September 2015 1 Conformance Statement Overview The application described in this conformance statement, VEPRO EMR Manager PACS, is a collection of processes

More information

Technical Publication. DICOM Conformance Statement. iplan 3.0. Document Revision 1. November 17, Copyright BrainLAB AG

Technical Publication. DICOM Conformance Statement. iplan 3.0. Document Revision 1. November 17, Copyright BrainLAB AG Technical Publication DICOM Conformance Statement Document Revision 1 November 17, 2009 2009 Copyright BrainLAB AG 1 Conformance Statement Overview This is a conformance statement for the BrainLAB software

More information

DICOM CONFORMANCE STATEMENT

DICOM CONFORMANCE STATEMENT Document No.: S517-1087 Revision: B DAR-7500 DICOM CONFORMANCE STATEMENT MEDICAL SYSTEMS DIVISION Revision History History Date Revision Description 1 st Edition July 2007 - A July.2008 Entirely recreated.

More information

DBSWIN DICOM Conformance Statement. DBSWIN DD-DICOM Interface. DICOM Conformance Statement V2.1

DBSWIN DICOM Conformance Statement. DBSWIN DD-DICOM Interface. DICOM Conformance Statement V2.1 DBSWIN DD-DICOM Interface DICOM Conformance Statement V2.1 DÜRR DENTAL Page 1 / 23 Conformance Statement Overview The DBSWIN software controls the digital Dürr Dental imaging products. The DD-DICOM Interface

More information

DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL SSA-640A V5.0

DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL SSA-640A V5.0 DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM TM MODEL SSA-640A V5.0 TOSHIBA MEDICAL SYSTEMS CORPORATION 2015 ALL RIGHTS RESERVED Trademarks Viamo is a trademark of Toshiba Medical Systems

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement SCENARIA Tokyo, Japan E1E-HC0002-01 Hitachi, Ltd. 2016. All rights reserved. History Revision Description Date Rev. 1.0 Initial. 2016/04/01 ii E1E-HC0002 Contents History...

More information

DICOM Conformance Statement * /02* /02 MADE IN GERMANY

DICOM Conformance Statement * /02* /02 MADE IN GERMANY EN DICOM 9000-608-109/02 *9000-608-109/02* MADE IN GERMANY Overview Vet-Exam Intra controls the digital DÜRR MEDICAL imaging products. The Vet-Exam Intra DICOM Interface translates between DICOM applications

More information

DICOM. Conformance Statement. Panorama Release 6. Panorama 0.6T Panorama 0.23T Panorama 0.23T I/T Panorama 0.23T R/T

DICOM. Conformance Statement. Panorama Release 6. Panorama 0.6T Panorama 0.23T Panorama 0.23T I/T Panorama 0.23T R/T DICOM Conformance Statement Panorama Release 6 Panorama 0.6T Panorama 0.23T Panorama 0.23T I/T Panorama 0.23T R/T Koninklijke Philips Electronics N.V. 2005 All rights are reserved. This DICOM Conformance

More information

Villa Sistemi Medicali DRF 4343

Villa Sistemi Medicali DRF 4343 Digitec S.r.l. Via Caduti Lecchesi a Fossoli, 17 23900 LECCO Italy digitec@digiteclecco.com Villa Sistemi Medicali DRF 4343 Network and Media Interchange DICOM Conformance Statement August 13, 2009 Draft

More information

Whole Body X-ray CT System Supria. DICOM Conformance Statement

Whole Body X-ray CT System Supria. DICOM Conformance Statement Whole Body X-ray CT System Supria DICOM Conformance Statement Tokyo, Japan (SN-C633E) Copyright Hitachi Medical Corporation. 2013, 2014. All rights reserved. History Revision Description Date Rev. 1.0

More information

DICOM Conformance Statement. CR-VW 674 for DICOM Storage DICOM Print DICOM MWM DICOM MPPS DICOM Media Storage

DICOM Conformance Statement. CR-VW 674 for DICOM Storage DICOM Print DICOM MWM DICOM MPPS DICOM Media Storage DICOM Conformance Statement CR-VW 674 for DICOM Storage DICOM Print DICOM MWM DICOM MPPS DICOM Media Storage December, 2011 7th Edition Copyright FUJIFILM Corporation, Japan 1 CONTENTS 1... CONFORMANCE

More information

DICOM CONFORMANCE STATEMENT FOR VANTAGE-GALAN TM VERSION V5.0

DICOM CONFORMANCE STATEMENT FOR VANTAGE-GALAN TM VERSION V5.0 DICOM CONFORMANCE STATEMENT FOR VANTAGE-GALAN TM VERSION V5.0 CANON MEDICAL SYSTEMS CORPORATION 2018 ALL RIGHTS RESERVED Trademarks VANTAGE-GALAN is trademark of Canon Medical Systems Corporation. This

More information

Technical Publication. DICOM Conformance Statement. BrainSUITE NET 1.5. Document Revision 1. March 5, Copyright BrainLAB AG

Technical Publication. DICOM Conformance Statement. BrainSUITE NET 1.5. Document Revision 1. March 5, Copyright BrainLAB AG Technical Publication DICOM Conformance Statement Document Revision 1 March 5, 2009 2009 Copyright BrainLAB AG 1 Conformance Statement Overview This is a conformance statement for the BrainLAB software

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement DICOM Pro Plus as an SCU Array Corporation Index 0. INTRODUCTION...3 1. IMPLEMENTATION MODEL...4 1.1 APPLICATION DATA FLOW DIAGRAM...4 1.2 FUNCTIONAL DEFINITION OF APPLICATION

More information

S517-E118 Dec DICOM Conformance Statement

S517-E118 Dec DICOM Conformance Statement Dec. 2014 DICOM Conformance Statement for NO TEXT 2 Overview: This conformance statement details the compliance to DICOM 3.0 of Side Station RAD for RADspeed Pro system. Table below provides an overview

More information

OASIS V4.0 DICOM Conformance Statement Rev. 3

OASIS V4.0 DICOM Conformance Statement Rev. 3 0BMR Imaging System OASIS V4.0 DICOM Conformance Statement Rev. 3 Tokyo, Japan E1E-BM7403-3 Copyright Hitachi Medical Corporation. 2011. All rights reserved. Revision History Revision Date Change Description

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL HDR-08A

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL HDR-08A DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL HDR-08A TOSHIBA MEDICAL SYSTEMS CORPORATION 2008-2010 ALL RIGHTS RESERVED Trademarks This document may include trademarks or registered

More information

DICOM. Conformance Statement For VINNO Medical Ultrasound System Revision 2.0. Copyright 2012 by VINNO Technology (Suzhou) Co., Ltd.

DICOM. Conformance Statement For VINNO Medical Ultrasound System Revision 2.0. Copyright 2012 by VINNO Technology (Suzhou) Co., Ltd. DICOM Conformance Statement For VINNO Medical Ultrasound System Revision 2.0 Copyright 2012 by VINNO Technology (Suzhou) Co., Ltd. 1 / 37 Revision History Document version Date Author Description 1.0 2014-07-10

More information

Technical Publication. DICOM Conformance Statement. Patient Browser 2.1. Document Revision 2. April 13, Copyright BrainLAB AG

Technical Publication. DICOM Conformance Statement. Patient Browser 2.1. Document Revision 2. April 13, Copyright BrainLAB AG Technical Publication DICOM Conformance Statement 2.1 Document Revision 2 April 13, 2011 2011 Copyright BrainLAB AG 1 Conformance Statement Overview This is a conformance statement for the BrainLAB software.

More information

DICOM Conformance Statement

DICOM Conformance Statement For SonixTouch Q+, SonixMDP/SP/OP Q+ and SonixOne Diagnostic Ultrasound Systems (and Systems with Software Version 6.1.0 and Newer) Analogic Corporation 8 Centennial Drive Peabody, MA 01960 USA bkultrasound.com

More information

Fluoroscan InSight V5.0.6 DICOM Conformance Statement

Fluoroscan InSight V5.0.6 DICOM Conformance Statement Fluoroscan InSight V5.0.6 DICOM Conformance Statement MAN-02996 Rev 002 Page 1 of 30 1. INTRODUCTION... 4 1.1. Purpose of the Document... 4 1.2. References... 4 1.3. Definitions... 4 2. IMPLEMENTATION

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement ECLOS Tokyo, Japan (SN-C357E) Copyright Hitachi Medical Corporation. 2007, 2008. All rights reserved. History Revision Description Date Rev. 1.0 Initial. 2007/02/01 Rev 2.0

More information

Artwork consists of twenty three (23) 8½ x 11 inch pages.

Artwork consists of twenty three (23) 8½ x 11 inch pages. Artwork consists of twenty three (23) 8½ x 11 inch pages. REV AUTHORED BY DATE P KUPOVICH 9/17/04 REV DRAFTED BY DATE G CORRIDORI 9/17/04 PROPRIETARY: This document contains proprietary data of Hologic,

More information

X-Porte DICOM Conformance Statement

X-Porte DICOM Conformance Statement 21919 30th Dr. SE, Bothell, WA 98021-3904 USA Telephone 1.425.951.1200 Facsimile 1.425.951.1201 www.sonosite.com Document Number: Revision: Title: D10711 D X-Porte DICOM Conformance Statement CHANGE HISTORY:

More information

DICOM CONFORMANCE STATEMENT

DICOM CONFORMANCE STATEMENT MultiScan G-Arm System Versions DigiArc 100AU/100AE DICOM CONFORMANCE STATEMENT Whale Imaging 1 / 16 REVISION HISTORY REVISION DATE 1.0 August 5.2013 For information contact Headquarters Whale Imaging

More information

Whole Body X-ray CT System Supria. DICOM Conformance Statement

Whole Body X-ray CT System Supria. DICOM Conformance Statement Whole Body X-ray CT System Supria DICOM Conformance Statement Tokyo, Japan E1E-HC0001-02 Hitachi, Ltd. 2016. All rights reserved. ii E1E-HC0001 History Revision Description Date Rev. 1.0 Initial. 2016/4/1

More information

REGIUS CONSOLE CS-2. Revision History. Date Version Description 02/08/2004 Ver This Conformance Statement applies to DICOM3.0 (2000 version).

REGIUS CONSOLE CS-2. Revision History. Date Version Description 02/08/2004 Ver This Conformance Statement applies to DICOM3.0 (2000 version). Revision History Date Version Description 02/08/2004 Ver.1.00 This Conformance Statement applies to DICOM3.0 (2000 version). REGIUS CONSOLE CS-2 DICOM 3.0 Conformance Statement Ver.1.00 02/08/2004 1 Contents

More information

MediaWorkStation. MWS DICOM Push Interface. DICOM Conformance Statement. Version 2.0. English

MediaWorkStation. MWS DICOM Push Interface. DICOM Conformance Statement. Version 2.0. English MediaWorkStation MWS DICOM Push Interface Version 2.0 DICOM Conformance Statement English 1-2014 Abbreviations AE DICOM IE IOD MPPS MWL MWS PACS PDU SOP SCP SCU Application Entity Digital Imaging and Communication

More information

Sonovision DICOM Conformance Statement

Sonovision DICOM Conformance Statement Sonovision DICOM Conformance Statement Document Version: 1.0 Product Name: Sonovision Release: 2.0 Manufacturer: Peridot Technologies Date: February 15, 2011 Document Used By: Sales & Marketing, Customer

More information

HS40 Ultrasound System

HS40 Ultrasound System HS40 Ultrasound System DICOM Conformance Statement Revision 1.0 System Version 1.00 Created: Apr 17, 2017 Last Updated: Apr 17, 2017 HS40 DICOM Conformance Statement 0 COVER PAGE Company Name: SAMSUNG

More information

DICOM Conformance Statement for GALILEI. Software Version V6.0

DICOM Conformance Statement for GALILEI. Software Version V6.0 DICOM Conformance Statement for GALILEI Software Version V6.0 Version 1.0, December 6 th, 2011 Contents Revision History... 2 Purpose... 2 References... 2 Terms and Definitions... 2 Abbreviations... 3

More information

DxMM/DxWin DICOM 3.0 Conformance Statement. Document Reference (Référence du document) : 99/ Oct30/ABA/MM103/398B

DxMM/DxWin DICOM 3.0 Conformance Statement. Document Reference (Référence du document) : 99/ Oct30/ABA/MM103/398B DxMM/DxWin DICOM 3.0 Conformance Statement Document Reference (Référence du document) : 99/ Oct30/ABA/MM103/398B Table of Contents (Table des Matières) 1 Introduction 6 1.1 Scope and Audience 6 1.2 References

More information

DICOM CONFORMANCE STATEMENT FOR Diagnostic Ultrasound System

DICOM CONFORMANCE STATEMENT FOR Diagnostic Ultrasound System DICOM CONFORMANCE STATEMENT FOR Diagnostic Ultrasound System Intellectual Property 2011 SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. ALL RIGHTS RELEASE 1 Conformance Statement Overview This document

More information

Philips Medical Systems DICOM Conformance Statement. Inturis Suite R2.2

Philips Medical Systems DICOM Conformance Statement. Inturis Suite R2.2 Philips Medical Systems DICOM Conformance Statement Inturis Suite R2.2 Document Number XZR 080-010044 8 February 2002 Copyright Philips Medical Systems Nederland B.V. 2002 Page ii DICOM Conformance Statement

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL ADR-1000A /R2 (MIIXR0003EAE)

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL ADR-1000A /R2 (MIIXR0003EAE) DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL ADR-1000A /R2 (MIIXR0003EAE) 1998-2004 IMPORTANT! (1) No part of this manual may be copied or reprinted, in whole or in part, without

More information

DICOM CONFORMANCE STATEMENT

DICOM CONFORMANCE STATEMENT Ultrasonix Medical Corporation DICOM CONFORMANCE STATEMENT For SonixTablet, SonixTouch, 2010 SonixMDP/SP/OP and Legacy SonixMDP Diagnostic Ultrasound Systems (Software Version 6.0.3 and above) Ultrasonix

More information

KARL STORZ OR1 FUSION V1.3.1 DICOM Conformance Statement

KARL STORZ OR1 FUSION V1.3.1 DICOM Conformance Statement KARL STORZ OR1 FUSION V1.3.1 DICOM Conformance Statement PRODUCT INFO OR1 OR1 87 1.2 02/2018/PI-E 1/21 1 Table of Content 1 Abbreviations...3 1.1 Definitions...3 1.2 References...3 2 CONFORMANCE STATEMENT

More information

SonoWin Picture Archiving System

SonoWin Picture Archiving System Page 1 of 13 SonoWin Picture Archiving System DICOM SSCU Module Version 4.1 DICOM Conformance Statement Page 2 of 13 1 DOCUMENT HISTORY...4 2 INTRODUCTION...4 2.1 SOURCES OF INFORMATION...4 2.2 ACRONYMS

More information

Printlink III-ID/IV. Revision History. Date Version Description 20/08/2004 Ver First edition

Printlink III-ID/IV. Revision History. Date Version Description 20/08/2004 Ver First edition Revision History Date Version Description 20/08/2004 Ver. 1.00 First edition NOTE: Descriptions in this document are subject to change without prior notice. Printlink III-ID/IV DICOM 3.0 Conformance Statement

More information

Technical Publications

Technical Publications Technical Publications Direction DOC0637451 Revision 1 Vivid i version 10 for DICOM Copyright 2009 By General Electric Co. g Do not duplicate GE Ultrasound THIS PAGE LEFT INTENTIONALLY BLANK TABLE OF CONTENTS

More information

DAR Revision Date: July,

DAR Revision Date: July, S517-E053B DAR-9000 Revision Date: July, 25. 2005 DICOM service classes in this manual are provided as an optional item of DAR-9000. For your suitable system configuration, please consult with SHIMADZU

More information

KARL STORZ AIDA V1.3.1 DICOM Conformance Statement PRODUCT INFO OR1 OR /2017/PI-E 1/32

KARL STORZ AIDA V1.3.1 DICOM Conformance Statement PRODUCT INFO OR1 OR /2017/PI-E 1/32 KARL STORZ AIDA V1.3.1 DICOM Conformance Statement PRODUCT INFO OR1 OR1 92 2.0 02/2017/PI-E 1/32 Change History Version Date Changes Reason Editor BB-02 07.02.2017 Whole document (table numbering, spelling)

More information

dicom PACS DX-R Version 3.0

dicom PACS DX-R Version 3.0 dicompacs DX-R Version 3.0 TABLE OF CONTENTS 1. Introduction... 3 1.1. Revision History... 3 1.2. Abbreviations and Acronyms... 4 2. Implementation Model... 5 2.1. Application Data Flow Diagram... 5 2.2.

More information

DICOM CONFORMANCE STATEMENT FOR DP10/DP20/DP30 SERIES DIGITAL ULTRASONIC DIAGNOSTIC IMAGING SYSTEM

DICOM CONFORMANCE STATEMENT FOR DP10/DP20/DP30 SERIES DIGITAL ULTRASONIC DIAGNOSTIC IMAGING SYSTEM DICOM CONFORMANCE STATEMENT FOR DP10/DP20/DP30 SERIES DIGITAL ULTRASONIC DIAGNOSTIC IMAGING SYSTEM Version V1.0 01-10-2013 Mindray 2013 All rights reserved SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO.,

More information

XPACS DICOM 3.0 Conformance Statement Document Version October 2004

XPACS DICOM 3.0 Conformance Statement Document Version October 2004 XPACS DICOM 3.0 Conformance Statement Document Version 1.2.0 28 October 2004 XPACS DICOM 3.0 CONFORMANCE STATEMENT Revision 1.3.0 2 8 October 2004 Page 1 of 25 Table of Contents 0 INTRODUCTION... 4 0.1

More information

nstream Version 3.1 DICOM Conformance Statement

nstream Version 3.1 DICOM Conformance Statement Image Stream Medical nstream Version 3.1 DICOM Conformance Statement Revision History: Revision Date Author Notes A 03/02/2006 TMT/EP Initial Control A 03/05/2006 TMT Minor cosmetic changes A 03/07/2006

More information

MyLab Ultrasound Scanner. DICOM Conformance Statement. Document Version 4.0. Date 19 OCT 2005

MyLab Ultrasound Scanner. DICOM Conformance Statement. Document Version 4.0. Date 19 OCT 2005 MyLab Ultrasound Scanner DICOM Conformance Statement Document Version 4.0 Date 19 OCT 2005 1 CONFORMANCE STATEMENT OVERVIEW MyLab is an Ultrasound scanner made by ESAOTE S.p.A.; its software is based upon

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement for (For Rev.01.01.00 or later) Apr. 2015 NO TEXT 2 Overview: This conformance statement details the compliance to DICOM 3.0 of Side Station i3 for SONIALVISION G4 system. Table

More information

Technical Publications

Technical Publications Technical Publications Direction FC250039 Revision 16 Vivid7 version 6 for DICOM Copyright 2007 By General Electric Co. g Do not duplicate GE Ultrasound THIS PAGE LEFT INTENTIONALLY BLANK TABLE OF CONTENTS

More information

MediPlus TM PACS&RIS Version 2.6

MediPlus TM PACS&RIS Version 2.6 Revision 2.0, Publication #7143-202 MediPlus TM PACS&RIS Version 2.6 DICOM 3.0 Conformance Statement June 2007 Revision 2.0, Publication #7143-202 MediPlus PACS&RIS V2.6 DICOM Conformance Statement Revision

More information

Technical Publications

Technical Publications DIRECTION 5177444-100 REV 02 Technical Publications Direction 5177444-100 Revision 02 LOGIQ e version 4.x.x for DICOM Copyright 2006 By General Electric Co. g Do not duplicate GE Ultrasound DIRECTION 5177444-100

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER. Aquilion ONE TM,Aquilion TM V4.63 OR LATER (TSX-301A,TSX-301B)

DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER. Aquilion ONE TM,Aquilion TM V4.63 OR LATER (TSX-301A,TSX-301B) No.MIICT0055EAA DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER Aquilion ONE TM,Aquilion TM V4.63 OR LATER (TSX-301A,TSX-301B) TOSHIBA MEDICAL SYSTEMS CORPORATION 2010 ALL RIGHTS RESERVED

More information

Technical Publications

Technical Publications Technical Publications Direction 2316173-100 Revision 8.00 LOGIQ 7 CONFORMANCE STATEMENT for DICOM Do not duplicate Copyright ª 2008 By GE Healthcare THIS PAGE LEFT INTENTIONALLY BLANK REVISION HISTORY

More information

1 CONFORMANCE STATEMENT OVERVIEW

1 CONFORMANCE STATEMENT OVERVIEW PMOD Technologies Ltd PMOD v3.9 Version 1.0 Date: 2017.11.02 1 CONFORMANCE STATEMENT OVERVIEW This document is a DICOM Conformance statement for PMOD application version 3.9xx, where xx denotes the build

More information

Technical Publications

Technical Publications GE HEALTHCARE DIRECTION 5178040-100REV 2.0 LOGIQ P5 2.X.X Technical Publications Direction 5178040-100 Revision 2.0 LOGIQ P5 version 2.x.x for DICOM Copyright ª 2007 By General Electric Co. g Do not duplicate

More information

Technical Publications

Technical Publications Technical Publications Direction 5141741-100 Revision 01 LOGIQ5 Expert version 4.x.x for DICOM Copyright 2005 By General Electric Co. g Do not duplicate GE Ultrasound THIS PAGE LEFT INTENTIONALLY BLANK

More information

Sep, th Edition 897N101668H

Sep, th Edition 897N101668H DICOM Conformance Statement Console Advance DR-ID 300CL/700CL 800CL/900CL for DICOM Storage DICOM Storage Commitment DICOM MWM DICOM MPPS DICOM Print DICOM Query / Retrieve DICOM Media Storage DICOM Dose

More information

Technical Publications

Technical Publications DOC0445444 REV 2 Technical Publications DOC0445444 Revision 2 CONFORMANCE STATEMENT for DICOM Do not duplicate Copyright 2008 By GE Healthcare DOC0445444 REV 2 THIS PAGE LEFT INTENTIONALLY BLANK DOC0445444

More information

QPC XSCAN32 DICOM 3.0 Conformance Statement Document Version October 2004

QPC XSCAN32 DICOM 3.0 Conformance Statement Document Version October 2004 QPC XSCAN32 DICOM 3.0 Conformance Statement Document Version 1.3.0 28 October 2004 QPC XSCAN32 DICOM 3.0 CONFORMANCE STATEMENT Revision 1.3.0 2 8 October 2004 Page 1 of 22 Table of Contents 0 INTRODUCTION...

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM. Infinix Celeve-i series / Infinix-i series Model DFP-8000 series V4.

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM. Infinix Celeve-i series / Infinix-i series Model DFP-8000 series V4. No. MIIXR0010EAL DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM Infini Celeve-i series / Infini-i series Model DFP-8000 series V4.40 TOSHIBA MEDICAL SYSTEMS CORPORATION 2011 ALL RIGHTS

More information

DICOM Conformance Statement VISIQ 1.0

DICOM Conformance Statement VISIQ 1.0 DICOM Conformance Statement VISIQ 1.0 Koninklijke Philips N.V. 2016 All rights are reserved. Document Number: 000343000000056 Rev A DICOM Conformance Statement: VISIQ 1.0 Page 1 of 26 Issued by: Philips

More information

Technical Publications

Technical Publications Technical Publications Direction DOC0913087 Revision 1.0 VOLUSON Voluson S6/S8 11.x.x Copyright 2011 by General Electric Co. Do not duplicate GE Medical Systems THIS PAGE LEFT INTENTIONALLY BLANK REVISION

More information

Technical Publications

Technical Publications GE HEALTHCARE DIRECTION 5184517-100REV 3.0 LOGIQ A5 3.X.X Technical Publications Direction 5184517-100 Revision 3.0 LOGIQ A5 version 3.x.x for DICOM Copyright 2009 By General Electric Co. Do not duplicate

More information

GEMINI DICOM Conformance Statement 16 Power PET/CT Imaging System Configuration

GEMINI DICOM Conformance Statement 16 Power PET/CT Imaging System Configuration GEMINI DICOM Conformance Statement 16 Power PET/CT Imaging System Configuration Positron Emission Tomography (PET) Technical Information Software Version 9.1 1 DICOM CONFORMANCE STATEMENT OVERVIEW 1.1

More information

Technical Publications

Technical Publications DIRECTION 2325843-100REV 6.01 LOGIQ 9 Technical Publications Direction 2325843-100 Revision 6.01 LOGIQ 9 CONFORMANCE STATEMENT for DICOM Do not duplicate Copyright 2005 By GE Healthcare DIRECTION 2325843-100REV

More information

Lumify 1.8.x DICOM Conformance Statement

Lumify 1.8.x DICOM Conformance Statement Lumify 1.8. DICOM Conformance Statement Lumify 1.8. 000683000000025 Rev A 2018-04-04 Koninklijke Philips Electronics N.V. 2018 All rights are reserved. Lumify 1.8. DICOM Conformance Statement 1 0.1 Revision

More information

Technical Publications

Technical Publications DIRECTION 2390126-100REV2 Technical Publications Direction 2390126-100 Revision 2 LOGIQ3 /Expert/Pro version 4.x.x for DICOM Copyright 2005 By General Electric Co. g Do not duplicate GE Ultrasound DIRECTION

More information

DAR-9400f / DAR-9500f DICOM Conformance Statement Revision J

DAR-9400f / DAR-9500f DICOM Conformance Statement Revision J S517-E057J Apr. 2016 DAR-9400f / DAR-9500f DICOM Conformance Statement Revision J Revision Date: April, 2016 Revision History Date Rev. Comment 2005.10.4 First First Revision 2006.11.10 A Correct the

More information

DICOM Conformance Statement. Forum

DICOM Conformance Statement. Forum Document Number: CG-LEH-MS-065 Revision: 2.0 Date: 2009-07-23 Author: Tobias Kurzke DICOM Conformance Statement Forum Version 2.0 Carl Zeiss Surgical GmbH Carl-Zeiss-Str. 22 73447 Oberkochen Germany www.meditec.zeiss.com

More information

RamSoft. PowerServer. DICOM Conformance Statement

RamSoft. PowerServer. DICOM Conformance Statement RamSoft PowerServer DICOM Conformance Statement PowerServer PACS DICOM Conformance Statement PowerServer PACS DICOM Conformance Statement... 2 1 Introduction... 4 2 Implementation Model... 4 2.1 Application

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER. Aquilion TM 32/64 (TSX-101A/D, TSX-101A/E)

DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER. Aquilion TM 32/64 (TSX-101A/D, TSX-101A/E) No. MIICT0041EA DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER Aquilion TM 32/64 (TSX-101A/D, TSX-101A/E) TOSHIBA MEDICAL SYSTEMS CORPORATION 2005 ALL RIGHTS RESERVED Trademarks Aquilion

More information

CONFORMANCE STATEMENT

CONFORMANCE STATEMENT Philips Medical Systems CONFORMANCE STATEMENT BV Family XA Document Number XPR 080-020061 July 15, 2002 Copyright Philips Medical Systems Nederland B.V. 2002 All rights reserved DICOM Conformance Statement

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement Application Annex: AneurysmFlow R1.1 On Interventional Workspot R1.5 Koninklijke Philips N.V. 2018 All rights are reserved. Issued by: Philips Medical Systems Nederland BV,

More information

Uscan. DICOM Conformance Statement

Uscan. DICOM Conformance Statement DICOM Conformance Statement - Uscan Uscan DICOM Conformance Statement Software Version: 4.1.1 Date: 3 August 2018 Signostics, Inc., a subsidiary of EchoNous, Inc. 8310 154th Ave NE, Suite 200 Redmond,

More information

DICOM. Conformance Statement. HD3 Release Rev A. Koninklijke Philips Electronics N.V All rights are reserved.

DICOM. Conformance Statement. HD3 Release Rev A. Koninklijke Philips Electronics N.V All rights are reserved. DICOM Conformance Statement HD3 Release 2.0 989605349120033 Rev A Koninklijke Philips Electronics N.V. 2006 All rights are reserved. 0.1 REVISION HISTORY Document Version Date of Issue Author Description

More information

Technical Publications

Technical Publications DIRECTION 2325843-100 REV 9.02 LOGIQ 9 Technical Publications Direction 2325843-100 Revision 9.02 LOGIQ 9 CONFORMANCE STATEMENT for DICOM Do not duplicate Copyright ª 2009 By GE Healthcare DIRECTION 2325843-100

More information

Technical Publications

Technical Publications Technical Publications Direction EP250409 Revision 21 EchoPAC PC version 5 for DICOM (From version 5.2.2) Copyright 2008 By General Electric Co. Do not duplicate GE Ultrasound THIS PAGE LEFT INTENTIONALLY

More information

Philips Medical Systems. Intera Document Number April 2003

Philips Medical Systems. Intera Document Number April 2003 Philips Medical Systems DICOM CONFORMANCE STATEMENT Intera 10.1 Document Number 4522 131 88821 10 April 2003 Copyright Philips Medical Systems Nederland B.V. 2003 All rights reserved DICOM Conformance

More information

PRO DATA Srl. MicroPrint. DICOM PRINT SCU Conformance Statement (MicroPrint-prnsrv-print-scu.sxw)

PRO DATA Srl. MicroPrint. DICOM PRINT SCU Conformance Statement (MicroPrint-prnsrv-print-scu.sxw) PRO DATA Srl MicroPrint DICOM PRINT SCU Conformance Statement (MicroPrint-prnsrv-print-scu.sxw) Microprint DCMIMGS - DICOM PRINT SCU Conformance Statement Pag 1 / 16 TABLE OF CONTENTS 1 INTRODUCTION...

More information

DICOM Conformance Statement FORUM

DICOM Conformance Statement FORUM Document Number: CG-LEH-MS-203 Revision: 1.0 Date: 2010-04-21 DICOM Conformance Statement FORUM Version 2.5 Carl Zeiss Surgical GmbH Carl-Zeiss-Str. 22 73447 Oberkochen Germany www.meditec.zeiss.com Document:

More information

DICOM Conformance Statement August 23, 2010 Version 1.0 ImageWorks Internal document number:

DICOM Conformance Statement August 23, 2010 Version 1.0 ImageWorks Internal document number: DICOM Conformance Statement August 23, 2010 Version 1.0 ImageWorks Internal document number: 0000062023 ImageWorks 250 Clearbrook Road, Suite 240 Elmsford, NY 10523 (914) 592-6100 1 Revision History Version

More information

Technical Publications

Technical Publications LOGIQ C SERIES 1.X.X Technical Publications Direction 5270029-100 Revision 1 LOGIQ C SERIES* version 1.X.X for DICOM Note: In the file, LOGIQ C SERIES means LOGIQ C2 & LOGIQ C3 & LOGIQ C5 & LOGIQ C2 PRO

More information

Cardiac Workstation Last Updated January 10,

Cardiac Workstation Last Updated January 10, . Camtronics Ltd. P.O. Box 950 900 Walnut Ridge Drive Hartland, WI 53029 Phone: 262-367-0700 Cardiac Workstation Last Updated January 10, 2001......... DICOM Conformance Statement AUTHOR: DATE: APPROVED

More information

dicom PACS 5 Version 5.2

dicom PACS 5 Version 5.2 dicompacs 5 Version 5.2 TABLE OF CONTENTS 1. Introduction... 3 1.1. Revision History... 3 1.2. Abbreviations and Acronyms... 4 2. Implementation Model... 5 2.1. Application Data Flow Diagram... 5 2.2.

More information

Hologic Physician s Viewer 5.0 DICOM Conformance Statement

Hologic Physician s Viewer 5.0 DICOM Conformance Statement Hologic Physician s Viewer 5.0 DICOM Conformance Statement MAN-00239 REV 001 Page 1 of 18 1 INTRODUCTION... 4 1.1 Purpose of the Document...4 1.2 References...4 1.3 Definitions...4 2 IMPLEMENTATION MODEL...

More information

Technical Publications

Technical Publications Technical Publications LOGIQ E9 CONFORMANCE STATEMENT for DICOM Do not duplicate Copyright 2011 By GE Healthcare REVISION HISTORY Revision Date Description Author 1 Draft May 27 th, 2008 Dominick Anggara

More information

Philips Medical Systems. Xcelera R1.1L1. Document Number XPR July 2003

Philips Medical Systems. Xcelera R1.1L1. Document Number XPR July 2003 Philips Medical Systems DICOM CONFORMANCE STATEMENT Xcelera R1.1L1 Document Number XPR 080-030073 11 July 2003 Copyright Philips Medical Systems Nederland B.V. 2003 All rights reserved 3 DICOM Conformance

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL DRAD-3000A, DRAD-3000E

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL DRAD-3000A, DRAD-3000E DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL DRAD-3000A, DRAD-3000E TOSHIBA MEDICAL SYSTEMS CORPORATION 2009 ALL RIGHTS RESERVED Trademarks This document may include trademarks

More information

DICOM Conformance Statement Product Name HCP DICOM Net Version

DICOM Conformance Statement Product Name HCP DICOM Net Version Title DICOM Conformance Statement Product Name HCP DICOM Net Version 5.00.00 Copyright - 2000-2013 SoftLink International Pvt. Ltd. SoftLink International Pvt. Ltd. Page 1 of 37 2, Anand Park, Aundh, Pune

More information

DICOM Conformance Statement RadWorks 2.1 Product Line

DICOM Conformance Statement RadWorks 2.1 Product Line RadWorks 2.1 Product Line Author : Frank Wessels Revision Date : August 1997 Version : 2.1a Document Number : 1996-1997 Applicare Medical Imaging B.V. Name Signature Date Consulting Radiologist Frits Barneveld-

More information