Doc. No: D Rev. 01 Last revision: TEST1 ROLLER 20 ROLLER 10/10PN ROLLER 20PN INTERFACING MANUAL

Size: px
Start display at page:

Download "Doc. No: D Rev. 01 Last revision: TEST1 ROLLER 20 ROLLER 10/10PN ROLLER 20PN INTERFACING MANUAL"

Transcription

1 Doc. No: D Rev. 01 Last revision: TEST1 ROLLER 20 ROLLER 10/10PN ROLLER 20PN INTERFACING MANUAL

2 REVISIONS Revision n 00 08/07/ /07/200 8 Date Author Software TEST1 Software ROLLER2 0 Software ROLLER10/20P N Comments AP 4.05x 5.02 Creation AP 601F 6.01C R P ROLLER10/20PN INTEGRATION INTERFACING LATEX PROTOCOL CONTACT ADDRESS MANUFACTURER SIRE Analytical Systems S.r.l. Via Biella 121/ UDINE ITALY tel / fax / WORLDWIDE EXCLUSIVE DISTRIBUTOR ALIFAX s.p.a. Alifax S.p.A. via Petrarca, 2/ Polverara (PD) - Italy Tel Fax

3 INDEX 1. INTRODUCTION ADD PARTS to INSTALL USEFUL TOOLS FOR DEBUGGING SYMBOLS TEST1 RS232 PORTS features: SERIAL CABLE PIN CONNECTIONS TO CONNECT TEST1 TO PC-LAB: TEST1 protocol options RS232 test function to check the communication of TEST QUERY PROTOCOL project for TEST Protocol description: ROLLER20 (old model) RS232 PORTS features: SERIAL CABLE PIN CONNECTIONS TO CONNECT ROLLER20 TO PC-LAB: ROLLER 20 protocol options RS232 test function to check the communication of ROLLER20 (OLD MODEL) QUERY PROTOCOL project for ROLLER Protocol description: ROLLER 10-10PN - 20PN RS232 PORTS features: SERIAL CABLE PIN CONNECTIONS TO CONNECT ROLLER10PN/20PN TO PC-LAB: ROLLER10PN/20PN PROTOCOL OPTIONS: RS232 test to check the communication of ROLLER10PN/20PN QUERY PROTOCOL project for ROLLER 10-10PN / ROLLER20PN Protocol explanation (BCI Configuration): TEST1, ROLLER10PN-ROLLER20PN and ROLLER20(old model) Multiple connections INTERFACING LATEX PROTOCOL LATEX PROTOCOL INTERFACE EXPLICATION Up to Sw. version 6.01C LATEX PROTOCOL INTERFACE EXPLICATION From Sw. version 6.01D (Test 1 and new Roller 10-10PN / 20PN) Note: The paragraphs written with the italic blue characters (as on this note) indicate a variation or an addition to the previous version of the manual.

4 1. INTRODUCTION This manual is focused to give detailed information about serial ports connections available on TEST1 ROLLER20 (old model) ROLLER10/10PN/20PN instruments and communication protocol features. It should help Software House Companies during the modification or the installation of their program in the Host Computer. COMMON PARTS 2. ADD PARTS to INSTALL A standard cable to connect two Personal Computers via serial port. This cable must have: D-type 25 pin female connector at one side and a D-type 9 pin female connector at the opposite side for TEST1 and ROLLER20 (OLD MODEL). D-type 9 pin female connector at one side and a D-type 9 pin female connector at the opposite side for ROLLER10PN/20PN. 3. USEFUL TOOLS FOR DEBUGGING A PC computer with a communication program installed (like HyperTerminal of Windows) to test the instrument communications. HyperTerminal allows the technician to check the communication messages byte by byte. To run it follow the path: start, all programs, accessories, communication, HyperTerminal; run it and choose new connection. Type a preferred name e.g. TEST1, press OK and then EXIT. Open the curtain menu pressing File on the left upper side of HyperTerminal window and Property; a new window will be opened. Press Connection and choose the serial port where the serial cable has been connected: e.g.com1. Press Configuration and choose the speed which should be 9600Bps. Press OK twice. 4. SYMBOLS The below symbols will be displayed on HyperTerminal if the program is configured with Terminal character. = STX symbol = ETX symbol = ACK symbol = NAK symbol CHECKSUM = XOR calculation of part of R received message to check the correctness of it. 1

5 5. TEST1 RS232 PORTS features: Communication ports parameters are fixed and cannot be modified. Baud rate 9600 Parity N Data bit number 8 Stop bit 1 HyperTerminal program must have the same parameters to a correct data exchanging. PORT2 connection. This instrument communicates through Serial Port 2. This port uses a 25 pin male (DB25) serial connector. The used pins are: no. 2 for TXDATA, no. 3 for RXDATA and 7 for ground. The remaining pins are connected together as nullmodem configuration (4-5-8 and 6-20). Serial Port 1 cannot be used for communication. It is connected internally to the INTERNAL BAR CODE READER (IBCR if installed) or it can be connected to an EXTERNAL BAR CODE READER (EBCR) The maximum number of ASCII characters for each patient ID is 15 (fifteen). PORT1 connection 6. SERIAL CABLE PIN CONNECTIONS TO CONNECT TEST1 TO PC-LAB: DB9 F DB25 F 2

6 7. TEST1 protocol options The right choice of the protocol enables the Query functionality. The table below specify different configurations: PROTOCOL Query DESCRIPTION R message structure SYMBOLS NOTES DAT8 No Patient ID codes, STX 1 byte Checksum note having length of R 1 byte characters of 8 digits at Not used 2 byte maximum, could Identification no 8 byte be read by an Not used 6 byte external bar code ESR result 4 byte reader (EBCR) or Checksum 1 byte they could be introduced typing ETX 1 byte them manually. The output message structure will be followed by the sedimentation curve date (like STD) if KINET option in the COMMUNICA- TION menu is enabled. TOTAL = 24 byte Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. PROTOCOL Query DESCRIPTION R message structure SYMBOLS NOTES DAT15 Yes It works by an STX 1 byte internal (IBCR) R 1 byte or external bar code reader Workstation number 2 byte (EBCR). It Identification no 15 byte accepts the Rack no 2 byte maximum Position 2 byte patient identity Cycle/Bayer (Note) 2 byte code with 15 ESR result 4 byte characters of Checksum 1 byte length. This ETX 1 byte protocol will be activated automatically if an IBCR is connected. TOTAL = 31 byte Checksum note Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. Bayer note: This field is managed by Bayer parameter on tech/communication Menu: If Bayer is 0, the field deals with the number of Cycle otherwise if Bayer is 1, the field will contain '01' (hex 30 and 31) to maintain the Bayer protocol compatibility. DAT8 either DAT15 option can be choice accessing in the COMMUNICATION MENU passing through TECH MENU. Press key 5 from MAIN MENU and keys 1 and subsequently 9 as password and then key 3. If the internal bar code reader is installed in a TEST1 instrument, DAT15 will be forced automatically at powering on. In the TEST1 COMMUNICATION MENU, RS232 Protocol option can be recalled pressing key 1- DAT08 2- DAT15 This message will be displayed Option (1) Option (2) Enables the patient identification number length to 8 bytes. Enables the patient identification number length to15 bytes. 3

7 8. RS232 test function to check the communication of TEST1 It simulates one transmission of 10 TEST1 results to host computer through RS232 serial port (SERIAL PORT 2). If DAT8 is chosen, a message with 8 ID digits will be sent to the Host. Selecting DAT15, the ID digit number to send to the Host is going to be 15. DAT8 test transmission and its explanation: Field name Symbol Hex value Length Description STX 02 1 char The first field is the same to all ten strings R 52 1 char The second field is the same to all ten strings Workstation The third field is the same to all ten strings 2 char Patient ID The furth field is different from the others strings 8 char Rack no. 2 char The fifth field is the same to all ten strings Position 2 char The sixth field is different from the others Bayer (Note2) 2 char The seventh field is the same to all ten strings Result 4 char The eighth field is different from the others Checksum (Note1) 1 char The nighth field is different from the others. ETX 03 1 char The tenth field is the same to all ten strings. Note : what above explained refers to the TEST Strings instrument send to host computer, this string is displayed above this table. Note1: Checksum note Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. Total length 24 char Note2: Bayer field remains activated to, '01' (hex 30 and 31) because it is only a test. DAT15 test transmission and its explanation: Field name Symbol Hex value Length Description STX ( 02 1 char The first field is the same to all ten strings R 52 1 char The second field is the same to all ten strings Workstation The third field is the same to all ten strings 2 char Patient ID The furth field is different from the others strings 15 char Rack no. 2 char The fifth field is the same to all ten strings Position 2 char The sixth field is different from the others Bayer (Note2) 2 char The seventh field is the same to all ten strings Result 4 char The eighth field is different from the others Checksum The nighth field is different from the others. 1 char (Note1) ETX 03 1 char The tenth field is the same to all ten strings. Total length 31 char Note : what above explained refers to the TEST Strings instrument send to host computer, this string is displayed above this table. Note1: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. Note2: Bayer field remains activated to, '01' (hex 30 and 31) 4

8 In the COMMUNICATION MENU, RS232 function can be recalled pressing. COMMUNICATION RS232 TEST TX This message will be displayed 9. QUERY PROTOCOL project for TEST1 This document describes the serial protocol (Query Protocol) used by TEST1 with the Internal Bar Code Reader (IBCR). It queries HOST Computer through Serial Port 2 (RS232C). This protocol is the same of QUERY TEST1 program; a Windows software designed to be loaded in a Personal Computer (LAB PC) installed in a laboratory and connected to HOST through a network. QUERY protocol uses two files to exchange data for the communication with HOST by using the LAB PC like a bridge between TEST1 and HOST Computer. Further information of QUERY TEST1 program software can be taken from the QUERY TEST1 installation and using manual. TEST1 Settings for a correct communication COMMUNICATION MENU in the TECH MENU, allows the configuration of some parameters for the correct communication protocol. The accessing in it is allowed by typing key 5 from MAIN MENU, keys 1 and subsequently 9 for the password requesting and key 3. COMMUNICATION 1. RS232 PROTOCOL 2. RS232 TEST 3. KINET. SETUP 4. SETUP [ * ] 5. ENABLE BAYER 6. MEASURE MISSING ID 7. TIME OUT UART 8. TEST1 NUMBER 9. CLEAR 5

9 The configuration for a correct Query protocol functioning is: RS232 PROTOCOL= It has to be set to DAT15 (Working with an Internal Bar Code Reader it is forced automatically). To access inside and select DAT15 press key 1 in the COMMUNICATION menu. KINET. SETUP= It must be kept to 0. Press key 3 in the COMMUNICATION menu to access inside and select 0. This parameter, when activated, sends a kinetic curve data for factory checking after having transmitted R message. SETUP [ * ]= ENABLE BAYER MEASURE MISSING ID= TIME OUT UART= Press key 4 in the COMMUNICATION menu to access inside the option and 1 to enable the asterisk character transmission to the host. It works when a analysed sample gives a low value of hematocrit (around 20% and less). The position of it in the message is explained in the Warnings (next page). Press key 5 in the COMMUNICATION menu to access Bayer option and 1 to enable it to guarantee the Bayer protocol specifications (field Cycle/Bayer = 01). If Bayer protocol is not requested this option must be kept to 0 (disable). Press key 6 in the COMMUNICATION menu to access to the option and key 1 to allow the analysis of samples without any identification number or not readable by IBCR. Keeping it to 0, these samples won t be analysed. Press key 7 in the COMMUNICATION menu to access this option to set the waiting time (from 2 to 99 seconds) if the increasing of the waiting time to receive T message and the confirmation byte (ACK) for having received R message is requested. The maximum time to receiving T message and the confirmation byte (ACK) for having received R message remain set to 2 seconds keeping it to OFF. If within this time no message comes in, the instrument will go ahead. This option works if ACK option, available pressing 9 in the COMMUNICATION menu is enabled. Verify of the parameters setting in the flag list printing: Pressing 6 in the TECH MENU, the flag list printing is going to be obtained like in the following example: TEST1 UK ver. 6.01F Jun xx:xx:xx. TEST1 SN. xxx # AVAILABILITY xxxx # GLOBAL COUNT. xxxx BoosterY Model Fact MIXING CYCLES 180 MIXING SPEED 1 Smart warning 2000 Instrument number N 1 RS232 DAT15 FLOW CONT. ON OFFSET LEV. FLOW 556 T FS fact Miss. ID set. 1 Change ID 0 ACK set 0 Print setup 1 Magnif Set 1 Calib. setup 1 Bayer. setup 1 [ * ] setup 0 Kinet.setup 0 Mixer speed 1 Offset car. 0 Offset nee. 0 Washing time 20 Vs. 0 TimeOut 20 6

10 10. Protocol description: 1. The internal Scanner reads the bar code (patient ID) from the label stuck on the tube, the instrument creates Q message and sends it to HOST computer through RS232 serial port (or to PC LAB PC if it is connected like a bridge). <STX>Q HOST computer verifies if the patient identity code is present in the working list and, accordingly, it replies with "T" message. TEST1 waits to receiving it within two seconds and then it goes ahead. If ACK option is off, TEST1 is going to remain waiting up to 2 seconds for receiving it, otherwise if it is on, TEST1 is going to wait for all the set time (TIMEOUT from 2 to 99 seconds). TIMEOUT UART time and ACK are selectable in the COMMUNICATION MENU. T message is going to have the result field filled with 0001, if the sample requires the analysis. The result field will be filled with 0000 if the sample doesn t require the analysis. <STX>T <STX>T e<ETX> 0000d<ETX> 3. After measuring (24 seconds about), TEST1 sends R message to host replacing the result field value (0001) with the ESR read value which should be between 0002 and 0120 or negative. Errors messages are reported like: -001 for SM (Sample Missing) error, -002 for NR (Not Readable) error and -004 for NF (No flow) error. <STX>R <STX>R f <ETX> l <ETX> 4. Host computer should send ACK confirmation byte ( symbol displayed on the HyperTerminal program) to TEST1 for having accepted R message. TEST1 will ignore it, if ACK is off otherwise it will wait for all the set time (TIMEOUT from 2 to 99 seconds) if ACK is OFF. ACK = Hexadecimal code 06 Warnings: 1 If a patient identity (ID) read by the TEST1 Scanner doesn t find its own correspondent code in the Host computer working list, T message will be replaced by <NAK> message ( symbol displayed on the HyperTerminal program). The sample will be however analysed and R message sent to host. (NAK Message) Hex code 21 2 If a particular sample has a very low hematocrit (around 20% and less) and * option in the communication menu is enabled, the result field of R message, sent to the host, will contains an asterisk set in the first position on the left and its own ESR result (i.e. <STX>R *015[ <ETX> If the analysis of a particular sample generates an error and * option in the communication menu is enabled, ESR result field will contain 0 in the first left position, - (minus) in the second and the error code (i.e. <STX>R \ <ETX> 3 If the internal Scanner cannot read the patient ID and MISS ID option in the Communication setup is 0, Q message won t be created. The sample without its own patient identity won t be processed. 7

11 4 If the internal Scanner cannot read the patient ID and MISS ID option in the Communication setup is 1, Q message won t be created. The sample without its own patient identity will be, however, processed and R message sent to host with an identity code generated by the instrument itself. The new patient ID, in this case, will be composed by 10 characters: cycle number (from 01 to 99) (2 characters), instrument s/n (4 characters), rack number (from 01 to 04-2 characters) sample position in the rack (from 01 to 15-2 characters). Example: Instrument s/n Rack number Cycle number Sample position in the rack A complete R string for the transmission could be: [Workstation number] [Instrument s/n] [Rack number] [ESR] [Checksum] <STX>R f <ETX> [Cycle number] [Sample position] [Cycle number (Note 2)] 8

12 11. ROLLER20 (old model) RS232 PORTS features: Communication ports parameters are fixed and cannot be modified. Baud rate 9600 Parity N Data bit number 8 Stop bit 1 HyperTerminal program must have the same parameters to a correct data exchanging. PORT2 connection. This instrument communicates to host through Serial Port 2.This port uses a 25 pin male (DB25) serial connector and the used pins are: no. 2 for TXDATA, no. 3 for RXDATA and 7 for ground. The remaining pins are connected together as null-modem configuration (4-5-8 and 6-20). PORT1 connection. Serial Port 1 has been reserved for the connection of an external SCANNER The maximum number of ASCII characters for each patient ID is 15 (fifteen). 12. SERIAL CABLE PIN CONNECTIONS TO CONNECT ROLLER20 TO PC-LAB: DB9 F DB25 F

13 13. ROLLER 20 protocol options The right choice of the protocol enables the Query functionality. The table below specify the different configurations: PROTOCOL Query DESCRIPTION R message structure SYMBOLS NOTES BCI Yes It works by an STX 1 byte internal (IBCR) or R 1 byte external bar code Workstation number reader (EBCR). It 2 byte accepts up to 15 Identification n 15 byte characters of maximum patient Rack number 2 byte identity code. The Position 2 byte instrument enables Cycle/Bayer 2 byte Query protocol which ESR result 4 byte enquires the host if a Checksum 1 byte specific sample ETX 1 byte needs to be TOTAL = 31 byte analyzed or not. STD, DAT8 DAT15 or BCI protocol can be selected by software accessing in the SETUP menu. Checksum note: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. Bayer note: This field is managed by Bayer parameter on Tech Menu. If Bayer is 0, the field deals with the number of Cycle otherwise if Bayer is 1, the field will contain '01' (hex 30 and 31) to maintain the Bayer protocol compatibility. In the SETUP MENU list, RS232 Protocol can be recalled by pressing or until the following message is displayed: SETUP RS232 PROTOCOL Confirm the function by pressing The following menu will be displayed: 1 - STD 2 - DAT DAT15 4 BCI _ 10

14 14. RS232 test function to check the communication of ROLLER20 (OLD MODEL) It simulates one test transmission of 10 test results to host computer through RS232 serial port (SERIAL PORT 2). If DAT8 is chosen, a message with 8 ID digits will be sent to the Host. Selecting DAT15 or BCI, the ID digit number to send to the Host is going to be 15. STD test transmission and its examination: No test is available in STD test DAT8 test transmission and its examination: Field name Symbol Hex value Length Description STX 02 1 char The first field is the same to all ten strings R 52 1 char The second field is the same to all ten strings Workstation The third field is the same to all ten strings 2 char Patient ID The furth field is different from the others strings 8 char Rack no. 2 char The fifth field is the same to all ten strings Position 2 char The sixth field is different from the others Bayer (Note 2) 2 char The seventh field is the same to all ten strings Result 4 char The eighth field is different from the others Checksum (Note 1) 1 char The nighth field is different from the others. ETX 03 1 char The tenth field is the same to all ten strings. Total length 24 char Note : what above explained refers to the TEST Strings instrument send to host computer, this string is displayed above this table. Checksum note: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. Bayer note: Bayer field remains activated to, '01' (hex 30 and 31) because it is only a test. DAT15 test transmission and its examination: Field name Symbol Hex value Length Description STX 02 1 char The first field is the same to all ten strings R 52 1 char The second field is the same to all ten strings Workstation The third field is the same to all ten strings 2 char Patient ID The furth field is different from the others strings 15 char Rack no. 2 char The fifth field is the same to all ten strings Position 2 char The sixth field is different from the others Bayer (Note 2) 2 char The seventh field is the same to all ten strings Result 4 char The eighth field is different from the others Checksum (Note 1) 1 char The nighth field is different from the others. ETX 03 1 char The tenth field is the same to all ten strings. Total length 31 char Note : what above explained refers to the TEST Strings instrument send to host computer, this string is displayed above this table. Checksum note: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. Bayer note: Bayer field remains activated to, '01' (hex 30 and 31) because it is only a test. 11

15 BCI test transmission and its examination: Field name Symbol Hex value Length Description STX 02 1 char The first field is the same to all ten strings R 52 1 char The second field is the same to all ten strings Workstation The third field is the same to all ten strings 2 char Patient ID The furth field is different from the others strings 15 char Rack no. 2 char The fifth field is the same to all ten strings Position 2 char The sixth field is different from the others Bayer (Note 2) 2 char The seventh field is the same to all ten strings Result 4 char The eighth field is different from the others Checksum (Note 1) 1 char The nighth field is different from the others. ETX 03 1 char The tenth field is the same to all ten strings. Total length 31 char Note : what above explained refers to the TEST Strings instrument send to host computer, this string is displayed above this table. Checksum note: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. Bayer note: Bayer field remains activated to, '01' (hex 30 and 31) because it is only a test. In the TECH MENU, RS232 function can be recalled pressing or for many times until the following message is displayed. COMMUNICATION RS232 TEST TX Press to run the test 12

16 15. QUERY PROTOCOL project for ROLLER20. This paragraph describes the serial protocol (Query Protocol) used by ROLLER20 with an External Bar Code Reader (EBCR) connected to Serial Port 2 (RS232C). This protocol is the same used by QUERY TEST1 program; a Windows software designed to be loaded in a Personal Computer (LAB PC) installed in a laboratory and connected to HOST through a network. QUERY protocol uses two files to exchange data for the communication with HOST by using the LAB PC like a bridge between ROLLER20 and HOST Computer. Further information of QUERY TEST1 program software can be taken from the QUERY TEST1 installation and using manual. Settings for a Query protocol correct communication on ROLLER20 The configuration for a correct Query protocol functioning is: RS232 PROTOCOL= It must be set to BCI (And an External Bar Code Reader must be installed). This function is available in the SETUP MENU. It can be recalled by pressing or until the following message is displayed: SETUP RS232 PROTOCOL Confirm the function by pressing The following menu will be displayed: 1 - STD 2 - DAT DAT15 4 BCI _ Make your choice and press CLEAR twice to go back MAIN MENU Chose 4 for a Query correct communication: 13

17 KINET. SETUP= This function is available in the SETUP MENU. It can be recalled by pressing or until the following message is displayed: SETUP KINET. SETUP Confirm the function by pressing The following message will be displayed: KINET. SETUP= 0 1 -Y 0-N _ It must be set to 0 SETUP [ * ]= This function is available in the TECH MENU. It can be recalled by pressing or until the following message is displayed: TEC. CAL MENU 19 [ * ] SETUP Confirm the function by pressing The following message will be displayed: 1- [ * ] [ Y ] 0- [ * ] [ N ] The asterisk function activated allows the instrument to send the asterisk parameter to the Host computer through the serial port. It will be displayed near the ESR value if the sample analyzed has a haematocrit value low (< 20%). IDENTITY ESR *045 Example of ESR result (045) with the * parameter enabled and sent to the host computer with the patient identification together. IDENTITY ESR Example of a sample analysed with a haematocrit value over 20% and * parameter enabled. The ESR result (5) will be sent to the host computer using this form. Note: Having disabled the asterisk function, the instrument cannot send the asterisk character to the host if the analysed sample had a haematocrit value lower then 20% but it will be printed out next the ESR result however. 14

18 ENABLE BAYER Bayer communication protocol detects the Bayer field value (the previous field of the ESR result) in the R message. If Bayer function in the COMMUNICATION menu is enabled, the field will contain '01' (hex 30 and 31) to maintain the Bayer protocol compatibility otherwise if it is disabled, the field will report the number of Cycle. This function is available in the TECH MENU. It can be recalled by pressing or until the following message is displayed: TEC. CAL MENU 7 ENABLE BAYER Confirm the function by pressing The following message will be displayed: 1- ENABLE [Y] 0- DISABLE [N] Query setup= This function is available in the TECH MENU. It can be recalled by pressing or until the following message is displayed: TEC. CAL MENU 17 QUERY SETUP Confirm the function by pressing The following message will be displayed: 1- Query [ Y ] 0- Query [ N ] If Query is enabled (default), the instrument will be able to interrogate the Host computer for the analysis requesting of each specific sample. ACK= This function is available in the TECH MENU. It can be recalled by pressing or until the following message is displayed: TEC.CAL MENU 18 ACK SETUP Confirm the function by pressing The following message will be displayed: 1 ACK [Y] 0 ACK [N] This function allows to extend the waiting time for receiving T message (Query answer from the HOST computer) and ACK (acknowledge message confirmation from the HOST computer for having received R message). To modify the extension time (in seconds), it is available the TIMEOUT function explained in the Protocol description (Paragraph 15). 15

19 Parameters verification: In the SETUP menu you will find FLAG LIST function. Press or until the following message is displayed: FLAGS SETUP LIST Confirm the function by pressing Below is an example ROLLER 20 UK ver. 6.01C ROLLER20 SN. xxx # AVAILABILITY xxxx # GLOBAL COUNT. xxxx GAIN MIXING CYCLES 180 MIXING SPEED 1 RS232 BCI FLOW CONT. ON OFFSET LEV. FLOW 800 OT OFFSET LEV. 700 T ROLLER n. 1 ACK setup 0 Query setup 1 Print setup 1 Calib. setup 1 Bayer. setup 1 [ * ] setup 0 Kinet.setup 0 Offset nee. 0 Vs 0 TimeOut 20 Warning SMART

20 16. Protocol description: 1. The patient bar code ID stuck on each sample tube can be read by using of an External Bar Code Reader connected at <STX>Q the ROLLER Serial Port1. The instrument will create and send Q message to the HOST computer through RS232 serial port (or to LAB PC if used like a bridge) if Query Setup option is 1. (Note 1) 2. HOST computer will verify if the patient identity has been inserted in the working list and then it replies with "T" message. If ACK option (in the TECH MENU list) is 0, ROLLER 20 remains waiting up to 2 seconds for receiving it. otherwise if ACK option is 1 it will wait for all the set time (TIMEOUT from 2 to 99 seconds). TIMEOUT UART time option is located in the TECH MENU list. (Note 2) As soon as the instrument receives T message, it will be checked. The wheel will rotate to one position to allow the operator to insert the sample tube, if T message has the result field <STX>T compiled with 0001 (that means the sample requires the analysis). The wheel remains stopped, if the result field will be <STX>T (the analysis of that particular sample is not requested). 0001e<ETX> 0000d<ETX> The procedure above explained will be repeated for the first to the 18 th sample tubes before starting the analysis. 3. After measuring (20 seconds about), ROLLER 20 will compile R messages replacing the result field value (0001) with an ESR value between values of 0002 and 0120 or a negative one. For reading errors, the instrument will report the error codes like: -001 for SM error, -002 for NR error and -004 for NF error. <STX>R <STX>R F <ETX> \ <ETX> 4. Host computer should send an ACK confirmation byte ( symbol on hyper terminal program) to ROLLER 20 for having accepted R message. The instrument will ignore it, if ACK option is 0. It will wait for all the set time (TIMEOUT from 2 to 99 seconds) if ACK option is 1. ACK Hex 06 Warnings: 1. If the patient identity (ID) read by the Scanner doesn t find its own correspondent in the Host computer working list, T message will be replaced by <NAK> message Hexadecimal code 21 ( symbol on hyper terminal program), the wheel will however rotate to the next position to allow the operator to insert the tube. The sample will be analysed and R message sent to host. 2. If a particular sample has a very low haematocrit (less then 20%) and * option in the TECH MENU is enabled, the result field of R message sent to the host will be composed by an asterisk in the first position on the left and the rest its own ESR result (i.e. *015 ). If the analysis of a particular sample generates an error, ESR result field will have 0 in the first left position - (minus) for the second position and the rest of the field the error code (i.e ). 3. If no label has been stuck on the tube or the external Scanner cannot read the patient ID, Q message won t be created. The wheel will rotate to accept the sample by pressing ENTER key 17

21 The sample will be however processed and R message sent to host with an identity code generated by itself. The new patient ID in this case will be composed by 10 characters: cycle number (2 characters from 01 to 99), instrument s/n (4 characters), wheel number ( 2 characters) sample position in the wheel (from 01 to 18-2 characters). Example: Instrument s/n Wheel number Cycle number Sample position in the wheel A complete R string for the transmission will be: [Workstation number] [Instrument s/n] [Wheel number] [ESR] [Checksum] <STX>R I <ETX> [Cycle number] [Sample position] [Cycle number (Note 3)] Note 1: If Query Setup option is 0, no Query protocol will be created; the instrument is going to accept and analyse all samples. Note 2: Follow the instruction looking at the Chapter 13-8 of the technical manual for TIMEOUT UART activation. Note 3: This field is related to Bayer parameter on SETUP. If Bayer is 0, the field contains the number of Cycle. If Bayer is 1, the field contains '01' (hex 30 and 31) to keep compatibility with Bayer protocol. Note 4: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. 18

22 17. ROLLER 10-10PN - 20PN RS232 PORTS features: PORT2 connection. The instrument communicates through Serial Port 1.This port uses a 9 pin male (DB9) serial connector. The used pins are: n 2 for TXDATA, n 3 for RXDATA and 7 for ground. The remaining pins are connected together as nullmodem configuration. This cable must have a D-type 9 pin female connector from one side and a D-type 9 pin female connector at the opposite side. PORT1 connection Serial Port 1 has been reserved for the connection of an external SCANNER The maximum number of ASCII characters for each patient ID is 15 (fifteen). 18. SERIAL CABLE PIN CONNECTIONS TO CONNECT ROLLER10PN/20PN TO PC-LAB: DB9 F DB9 F 19

23 19. ROLLER10PN/20PN PROTOCOL OPTIONS: The choice of the protocol enables the Query functionality. Patient ID codes, having length of characters of 8 digits at maximum, could be read by an external bar code reader (EBCR) or they could be introduced typing them manually. The tables below specify different configurations: (NO QUERY) DAT8 transmission and its examination: Field name Symbol Hex value Length Description STX 02 1 char Start transmission R 52 1 char R states for result message Workstation Is the number identifying the instrument if more than 2 char one is connected to Host computer Patient ID 8 char Patient identification Wheel no. 2 char Indicate the progressive number of analysis made Position 2 char Indicate the sample position inside the wheel Bayer (Note 2) 2 char See note 2 Result 4 char ESR result Checksum (Note 1) 1 char See note 1 ETX 03 1 char End transmission Total length 24 char Checksum note: See note at the end of this page Note2: Bayer field is not used, in this case show s the progressive analysis cycle (NO QUERY) DAT15 transmission and its examination: Field name Symbol Hex value Length Description STX 02 1 char Start transmission R 52 1 char R states for result message Workstation Is the number identifying the instrument if more than 2 char one is connected to Host computer Patient ID 15 char Patient identification Rack no. 2 char Indicate the progressive number of analysis made Position 2 char Indicate the sample position inside the wheel Bayer (Note 2) 2 char See note 2 Result 4 char ESR result Checksum (Note 1) 1 char See note 1 ETX 03 1 char End transmission Total length 31 char Note1: See note at the end of this page Note2:see previous table (QUERY) BCI transmission and its examination: Field name Symbol Hex value Length Description STX 02 1 char Start transmission R 52 1 char R states for result message Workstation Is the number identifying the instrument if more than 2 char one is connected to Host computer Patient ID 15 char Patient identification Rack no. 2 char Indicate the progressive number of analysis made Position 2 char Indicate the sample position inside the wheel Bayer (Note 2) 2 char See note 2 Result 4 char ESR result Checksum (Note 1) 1 char See note 1 ETX 03 1 char End transmission Total length 31 char Note1: See note at the end of this page Note2:see previous table Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. 20

24 NO transmission: PROTOCOL Query DESCRIPTION R message structure SYMBOLS NOTES NO Host NO No transmission In the Communication menu is possible to choose the appropriate protocol and options: To activate the below option the password 1990 is compulsory. From Main Screen, the path to enable it is Setup, Log In Out, LOGIN A keyboard to type the password is going to be displayed. Type the password:1990 then press the left arrow to confirm or Clear to exit without changing anything, in both cases, the instrument is going to display again the previous image, then press Back. 21

25 20. RS232 test to check the communication of ROLLER10PN/20PN To activate the below option the password 1990 is compulsory. From Main Screen, the path to enable it is Setup, Log In Out, LOGIN A keyboard to type the password is going to be displayed. Type the password:1990 then press the left arrow to confirm or Clear to exit without changing anything, in both cases, the instrument is going to display again the previous image, then press Back. Pressing Comm menu and TEST it is possible to send a communication test to host computer through RS232 serial port. If DAT8 is chosen, the test message having 8 ID digits will be sent to the Host like: <0x2>R01TESTR10A U<0x3> <0x2>R01TESTR10B S<0x3> <0x2>R01TESTR10C U<0x3> <0x2>R01TESTR10D ^<0x3> <0x2>R01TESTR10E V<0x3> <0x2>R01TESTR10F T<0x3> <0x2>R01TESTR10G N<0x3> <0x2>R01TESTR10H N<0x3> <0x2>R01TESTR10I H<0x3> <0x2>R01TESTR10J @<0x3> Selecting DAT15 or BCI, the ID digit number to send to the Host is going to be 15 like: <0x2>R01TESTR10A <0x2>R01TESTR10B <0x2>R01TESTR10C <0x2>R01TESTR10D <0x2>R01TESTR10E <0x2>R01TESTR10F <0x2>R01TESTR10G <0x2>R01TESTR10H <0x2>R01TESTR10I <0x2>R01TESTR10J t<0x3> r<0x3> t<0x3> <0x7f><0x3> w<0x3> u<0x3> o<0x3> o<0x3> i<0x3> a<0x3> If NO HOST is configured, the instrument will not communicate with Host Computer 22

26 21. QUERY PROTOCOL project for ROLLER 10-10PN / ROLLER20PN. This paragraph describes the serial protocol (Query Protocol) used by ROLLER10PN /ROLLER20PN with an External Bar Code Reader (EBCR) connected to Serial Port 1 (25 PIN). This protocol is the same used by QUERY TEST1 program; a Windows software designed to be loaded in a Personal Computer (LAB PC) installed in a laboratory and connected to HOST through a network. QUERY protocol uses two files to exchange data for the communication with HOST by using LAB PC like a bridge between ROLLER10PN /ROLLER20PN and HOST Computer. Further information of QUERY TEST1 program software can be taken from the QUERY TEST1 installation and using manual. Fields to check looking at the flag list. 23

27 22. Protocol explanation (BCI Configuration): 1. The patient bar code ID stuck on each sample tube can be read by using an External Bar Code Reader connected to Roller 10 Serial Port1. The instrument will create and send Q <STX>Q a<ETX> message to HOST computer through RS232 serial port (or to LAB PC if it is used like a bridge) if Query Setup option is 1. (Note 3) 2. HOST computer verifies if the patient identity has been inserted in the working list and then it replies with "T" message. Roller 10 remains waiting for receiving it on time (within 2 seconds), if ACK option is disabled. It will wait for all the set time if ACK option is enabled. (Follow the instruction looking on page 22 of this manual.) 3. As soon as the instrument receives T message, it is checked. 4. The wheel is going to rotate to one position to allow the operator to insert the sample tube, if T message has the result field compiled with 0001 (that means the sample requires the <STX>T <STX>T e<ETX> 0000d<ETX> analysis). The wheel will remain stop, if the result field is 0000 (the analysis of that particular sample is not requested). Before starting the analysis, the explained above, from 1 to 2, procedure will be repeated for the further 10 th (Roller 10) or the 20 th (Roller 20) sample tubes. 5. After measuring (20 seconds about), Roller 10 compile R messages replacing the result field value (0001) with an ESR value that, read on the paper could be between values of 0002 and 0120 or a negative one. For reading errors, the instrument will report the error <STX>R <STX>R f <ETX> l <ETX> codes like: for SM error, -002 for NR error and -004 for NF error. 6. Host computer should send an ACK confirmation byte ( symbol looking at the hyper terminal program) to Roller 10 for having accepted R message. The Hexadecimal code 0x06 instrument will ignore it, if ACK option is disabled. It will wait for all the set time (TIMEOUT from 2 to 60 seconds) if ACK option is enabled. Protocol explanation (DAT 8 DAT 15 Configuration): In this case the instrument doesn t query the Host Computer, it just perform the analisis and send the result to host computer as explaned at points 3 and 4. Warnings: 1. If the patient identity (ID) read by the Scanner doesn t find its own correspondent in the Host computer working list, T message will be replaced by <NAK> message ( symbol looking at the hyper terminal program). If MISS ID option is enabled, the wheel Hexadecimal code 0x21 will rotate to the next position to allow the operator to insert the tube. The sample, then will be analysed and R message sent to host. 2. If no label has been stuck on the tube or the external Scanner cannot read the patient ID, Q message won t be created. The wheel is going to rotate to accept the sample tube by pressing Next button on touchscreen display. 3. The sample is going to be processed and R message sent to host with an identity code generated by itself. The new patient ID in this case will be composed by 10 characters: 4. Cycle number (2 characters from 01 to 99), instrument s/n (4 characters), wheel number ( 2 characters) sample position in the wheel (from 01 to 10-2 characters). 24

28 Example: Instrument s/n Wheel number Cycle number Sample position in the wheel A complete R string for the transmission will be: [Workstation number] [Instrument s/n] [Wheel number] [ESR] [Checksum] <STX>R f <ETX> [Cycle number] [Sample position] [Cycle number (Note 5)] Note 1: Note 2: Note 3: Note 4: Note 5: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. <STX> character ( symbol looking at the hyper terminal program) in the serial transmission is represented by 0x02 hexadecimal bytes and <ETX> ( symbol looking at the hyper terminal program) by 0x03. If Query Setup option is 0, no Query protocol will be created; the instrument will accept and analyse all samples. Follow the instruction looking at the of the technical manual for TIMEOUT UART activation. This field is related to Bayer parameter on SETUP. If Bayer is 0, the field contains the number of Cycle. If Bayer is 1, the field contains '01' (hex 30 and 31) to keep compatibility with Bayer protocol. 25

29 23. TEST1, ROLLER10PN-ROLLER20PN and ROLLER20(old model) Multiple connections Query protocol is able to manage multiple connections by using the same PC-LAB with multiple serial ports. In that case every instrument connected to the PC-LAB has to have an identification number different for each and the PC-LAB the same number of serial ports as the instruments. Our instruments have an option (TEST1 number in the COMMUNICATION menu (for TEST1) and ROLLER number in the TECH MENU (for ROLLER20) to assign a different identification number to each instrument, R message should be different for each like in the following example: [Workstation number] <STX> R [Workstation number] <STX> R [Workstation number] <STX> R [Checksum] I <ETX> [Checksum] z <ETX> [Checksum] ~ <ETX> 26

30 24. INTERFACING LATEX PROTOCOL LATEX PROTOCOL INTERFACE EXPLICATION Up to Sw. version 6.01C (Test 1), 6.01B (Roller20 old model) The three R Latex messages are sent, to the serial port, after having analysed the three referent Latex of the Control kit and at the end of the washing, by withdrawing water from the fifth and sixth inserted tube, for rinsing the capillary. It should be understood R Latex messages are not available by using Calibration Latex kit. Messages R message for Latex Field name Hex Length value Description STX (Note 2) 02 1 char R 52 1 char Means the instrument is sending a Result Workstation 2 char Identifies the workstation if more instrument are connected, normally is 01 LATEXSTDL 9 char Is the identification of the latex tube Field 1 1 char Can be 1, 2 or 3 and identifies level 1, 2 and 3 of latex ESR Value 3 char This is the expected Latex ESR value Space char Two spaces Field 2 2 char Fixed to 01 Field 3 2 char The same as Field 1 Field 4 2 char The same as Field 2 ESR Value 4 char This is the detected Latex ESR value Checksum (Note 1) 1 char ETX 03 1 char Total length 31 char Note 1: Note 2: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum value calculated by the analyzer is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value. Software House must adopt the same procedure when calculating checksum of the strings received from the analyzer. <STX> character ( symbol looking at the hyper terminal program) in the serial transmission is represented by 0x02 hexadecimal bytes and <ETX> ( symbol looking at the hyper terminal program) by 0x03. here are some examples: <0x2>R01LATEXSTDL ^<0x3> <0x2>R01LATEXSTDL ><0x3> <0x2>R01LATEXSTDL ><0x3> 27

CONNECTION TO A HOST COMPUTER

CONNECTION TO A HOST COMPUTER CONNECTION TO A HOST COMPUTER Foreword: Hardware data The Communication between the CUBE30TOUCH and an external PC may be done in two possible ways: 1- Using a USB connection: Connect a standard A-B USB

More information

11 Serial Communications

11 Serial Communications 11.1 COMMUNICATIONS PACKAGES There are a number of communication packages that can be used with the Quantum III to facilitate setup, record parameter data, view internal activity on a soft-scope and permit

More information

EWBS Receiving Module Communication specifications

EWBS Receiving Module Communication specifications 2017/01/05 EWBS Receiving Module Communication specifications v1.00 Century Revision history Revision Date Changes 1.00 2017/01/05 Initial Release. 1 Serial communication specifications Asynchronous serial

More information

Operating instructions

Operating instructions Operating instructions SPM Bi-Directional Communication Protocol Your Uptime Is Our Top Priority Congratulations on your purchase of the SPM Bi-Directional Communications Protocol. It will provide you

More information

Lab Establishing a Console Connection to a Router or Switch Instructor Version

Lab Establishing a Console Connection to a Router or Switch Instructor Version Lab 5.2.7 Establishing a Console Connection to a Router or Switch Instructor Version Objective Create a console connection from a PC to a router and switch using the proper cable Configure HyperTerminal

More information

Configuration Manual PULSAR C CCD SCANNER. Table of Contents

Configuration Manual PULSAR C CCD SCANNER. Table of Contents Table of Contents PULSAR C CCD SCANNER Configuration Manual Metrologic Instruments GmbH Dornier Strasse 2 82178 Puchheim Germany Tel +49 89 890190 Fax +49 89 89019200 www.europe.metrologic.com Metrologic

More information

OPERATING MANUAL AND TECHNICAL REFERENCE

OPERATING MANUAL AND TECHNICAL REFERENCE MODEL WFG-D-130 HIGH SPEED DIGITAL 3 AXIS FLUXGATE MAGNETOMETER OPERATING MANUAL AND TECHNICAL REFERENCE December, 2012 Table of Contents I. Description of the System 1 II. System Specifications.. 2 III.

More information

USER INTERFACE RS232-C

USER INTERFACE RS232-C USER INTERFACE RS232-C Cat..: 28062 V 1.1 - March 02 Revision Table Revision Table Manual Version Date List of Modifications V 1.0 vember 95 Original Version V 1.1. March 97 Modified Pages: 11, 14, 24

More information

Desktop Reader and Wall Reader Adaptor USER MANUAL DOCUMENT REVISIONS

Desktop Reader and Wall Reader Adaptor USER MANUAL DOCUMENT REVISIONS Desktop Reader and Wall Reader Adaptor USER MANUAL Version: 2.31 Distribution: SALTO customers DOCUMENT REVISIONS Version Date Modifications 1.2 5/03/2004 1.3 11/03/2004 Note on Wall Reader Adaptor about

More information

Communication Protocol for Solder Feeder (SFR) Programmer s Guide

Communication Protocol for Solder Feeder (SFR) Programmer s Guide www.jbctools.com Communication Protocol for Solder Feeder (SFR) Programmer s Guide SFR Communication Protocol for Automation This communication protocol was developed to use JBC devices, such as the Automatic

More information

Microframe Corporation

Microframe Corporation A0262: RS232 Remote Display Operating Manual A0262-7010 *A0262-7010* A0262 INSTALLATION & SPECIFICATION GUIDE ITEM NO: A0262 REVISION DATE: 11/09 Microframe Corporation 604 S. 12th Street Broken Arrow,

More information

It is the installer's responsibility to follow all instructions in this manual and to follow correct electrical practice.

It is the installer's responsibility to follow all instructions in this manual and to follow correct electrical practice. MCD Modbus Module Instructions Important User Information INSTALLATION INSTRUCTIONS: MCD MODBUS MODULE Order Code: 175G9000 1. Important User Information Observe all necessary safety precautions when controlling

More information

1. Introduction. 2. Installation MODBUS INTERFACE

1. Introduction. 2. Installation MODBUS INTERFACE 5551.C 8473.C MODBUS INTERFACE PIM-MB-1 Modbus Interface 1. Introduction AuCom soft starters can be controlled and monitored across an RS485 serial communication network using the Modbus RTU and AP ASCII

More information

Card Dispensing/Collecting Machine

Card Dispensing/Collecting Machine A 1 OF 36 2010. 03. 02. Card Dispensing/Collecting Machine Series. 3,,. http://www.securetech-corp.com E-mail : sales@securetech-corp.com A 2 OF 36 2010. 03. 02. REVISION HISTORY No DATE DESCRIPTION REV

More information

MS Protocol Converter. User Manual. Firmware version 2.0 ISI. Instrumental Solutions, Inc.

MS Protocol Converter. User Manual. Firmware version 2.0 ISI. Instrumental Solutions, Inc. MS1-2150 Protocol Converter User Manual Firmware version 2.0 ISI Instrumental Solutions, Inc. WWW.ISIDEV.NET Introduction The MS1-2150 protocol converter can take readings from 1 or more ModBus registers

More information

Flex Series User Guide

Flex Series User Guide User Programmable Current 4..20mA Digital RS485 Dual & Single Axis Up to 360º 2016 Flex Series User Guide Sensor Installation, Wiring, Flexware App Instructions Page 1 of 33 Page 2 of 33 Table of Contents

More information

Revision 1.2. July 24, COM Protocol Manual. for MDC and ADC N 11th St - San Jose CA

Revision 1.2. July 24, COM Protocol Manual. for MDC and ADC N 11th St - San Jose CA Revision 1.2 July 24, 2017 COM Protocol Manual for MDC and ADC www.mountztorque.com - 1080 N 11th St - San Jose CA 95112-408.292.2214 1 1 Overview and Communication Specifications 1.1 Overview If the PC

More information

MSOC Communication Commonality with DPU/TPU/GPU 2000R Protective Relays

MSOC Communication Commonality with DPU/TPU/GPU 2000R Protective Relays ABB Application Note MSOC AN-64A-00 Substation Automation and Protection Division MSOC Communication Commonality with DPU/TPU/GPU 2000R Protective Relays Introduction There seems to be some confusion with

More information

THERMOSALD ISX RS485 MODBUS (V5) 3E S.r.l. - Via del Maccabreccia 37/a LIPPO DI CALDERARA (BOLOGNA) THERMOREGULATOR for PULSE WELDING

THERMOSALD ISX RS485 MODBUS (V5) 3E S.r.l. - Via del Maccabreccia 37/a LIPPO DI CALDERARA (BOLOGNA) THERMOREGULATOR for PULSE WELDING THERMOREGULATOR for PULSE WELDING THERMOSALD ISX ADDRESS OF UNITS WITH DIP SWITCHES ADDRESS OF UNITS FROM 1 TO 255 BROADCASTING ADDRESS = 0 CONNECTION OF 3E PANEL UP TO 255 UNITS CHANGE ALL VARIABLE FROM

More information

DirectNET Host. Communications Programs. In This Chapter...

DirectNET Host. Communications Programs. In This Chapter... Communications Programs In This Chapter.... Why do you need a communications program? Modes of Operation Protocol Components Controlling the Communications Initiating the Request Acknowledging the Request

More information

USB TO RS-232/RS-422/RS-485 ADAPTER

USB TO RS-232/RS-422/RS-485 ADAPTER USB TO RS-232/RS-422/RS-485 ADAPTER For Android User s Manual UTS-232AD / UTS-422AD / UTS-485AD Table of Contents Introduction...2 System Requirements...2 Features...2 Specifications...3 Install Application...4

More information

LomaLink User s Guide

LomaLink User s Guide LomaLink User s Guide The Serial Communications System for Loma Inspection Equipment Loma Part Number: 811153 Spectrum Inspection Systems Issue: K Southwood Date: July 2007 Farnborough Hampshire GU14 0NY

More information

PS232. RS-232 to PS/2 Keyboard Port Adapter Part # SA0009 (Version 4.0) Copyright 2003 L3 Systems, Inc. Redmond

PS232. RS-232 to PS/2 Keyboard Port Adapter Part # SA0009 (Version 4.0) Copyright 2003 L3 Systems, Inc. Redmond PS232 RS-232 to PS/2 Keyboard Port Adapter Part # SA0009 (Version 4.0) Copyright 2003 L3 Systems, Inc. Redmond Quick Reference Command Description Pg ~H Help Screen Displays short command reference 4 ~V

More information

VEEMUX Series. SM-4X4-DVI SM-8X8-DVI DVI Video Matrix Switch Installation and Operation Manual

VEEMUX Series. SM-4X4-DVI SM-8X8-DVI DVI Video Matrix Switch Installation and Operation Manual VEEMUX Series SM-4X4-DVI SM-8X8-DVI DVI Video Matrix Switch Installation and Operation Manual MAN114 Rev Date 10/7/2009 TRADEMARK VEEMUX is a registered trademark of Network Technologies Inc in the U.S.

More information

Operate the Hydrolab SDI-12 / Modbus / RS232 TTY Communications Module (HL Series Sonde)

Operate the Hydrolab SDI-12 / Modbus / RS232 TTY Communications Module (HL Series Sonde) Operate the Hydrolab SDI-12 / Modbus / RS232 TTY Communications Module (HL Series Sonde) 04/2018, Edition 1 User Manual Overall Contents Part A Operate the Hydrolab SDI-12 Communications Module Part B

More information

API Hessen Network Protocol Revision C June 24, 1998

API Hessen Network Protocol Revision C June 24, 1998 Revision C June 24, 1998 TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TAPI) 9970 CARROLL CANYON ROAD SAN DIEGO, CA 92131-1106 USA TOLL-FREE: 800-324-5190 FAX: +1 858-657-9816 PHONE: +1 858-657-9800 E-MAIL:

More information

Xi2000-BT Series Configuration Guide

Xi2000-BT Series Configuration Guide U.S. Default Settings Sequence Reset Scanner Xi2000-BT Series Configuration Guide Auto-Sense Mode ON UPC-A Convert to EAN-13 OFF UPC-E Lead Zero ON Save Changes POS-X, Inc. 2130 Grant St. Bellingham, WA

More information

education and advancement of the marine electronics industry and to waypoint, present Lat/Long, and a binary status byte.

education and advancement of the marine electronics industry and to waypoint, present Lat/Long, and a binary status byte. Q: What is NMEA What is NMEA? The National Marine Electronics Association is dedicated to the education and advancement of the marine electronics industry and the market which it serves. It is a non-profit

More information

C30xx Digital communication

C30xx Digital communication C30xx Digital communication Table of contents 1 Introduction...3 2 Digital ports...3 3 Command protocol to send...3 4 Command protocol to receive...4 5 Command table...5 6 Command specifications + examples...6

More information

User s Manual. Xi3000 Scanner. Table of Contents

User s Manual. Xi3000 Scanner. Table of Contents Xi3000 Scanner User s Manual Table of Contents Restore Default Settings... 1 Exit Setup without Changes... 1 Configure Through RS232... 1 List Setting... 1 Buzzer Settings... 2 Reading Redundancy Setting...

More information

Specification. Current Consumption Range 8m * Rotational Angle +/- 50 degrees * Shear Angle +/- 40 degrees *

Specification. Current Consumption Range 8m * Rotational Angle +/- 50 degrees * Shear Angle +/- 40 degrees * HX11TR Ultrasonic Positioning Device The HX11TR can be set up to operate as a ultrasonic signal receiver, ultrasonic transmitter, ultrasonic caller and ultrasonic transponder. It is small size and economical.

More information

Instruction Manual. External Output Manual. Liquid Scintillation System LSC Model

Instruction Manual. External Output Manual. Liquid Scintillation System LSC Model Instruction Manual External Output Manual Liquid Scintillation System Model LSC-8000 Thank you for purchasing this product. The instruction manual of this product consists of the Operation Manual the External

More information

SIOX-RS232C Converter

SIOX-RS232C Converter L40 SIOX-RS232C Converter p 2 TELEFRANG AB TABLE OF CONTENTS SIOX-RS232C Converter General Description This module connects a CRT terminal or other RS232C serially communicating I/O device as a station

More information

Planar Simplicity Series

Planar Simplicity Series Planar Simplicity Series RS232 PROTOCOL Document 020-1285-00 1. INTRODUCTION 1.1 Purpose The purpose of this document is to explain in detail the commands and steps that can be used to control a Planar

More information

DiViS DVR CAP Series Rev Chance- i USA Corp.

DiViS DVR CAP Series Rev Chance- i USA Corp. DiViS DVR CAP Series Hardware Installation Guide Rev. 1.1 Digital Video Security System Digital Video Recorder www.divisdvr.com *All contents of this document may change without prior notice. Chance-i

More information

Leica LP C (Laser Printer for Cassettes) System Requirement & Specifications

Leica LP C (Laser Printer for Cassettes) System Requirement & Specifications Leica LP C (Laser Printer for Cassettes) System Requirement & Specifications LP C Req. & Spec. Page 1 of 9 Version 1.1 Copyright 2010 Leica Microsystems All rights reserved All company names, product names,

More information

Ultrabeam RCU-06 USB port description

Ultrabeam RCU-06 USB port description Abstract The Ultrabeam RCU-06 is an electronic controller used in a variety of Ultrabeam antennas. The main task of the controller is to tune the antenna by moving the (radiating) elements, based on the

More information

Connecting UniOP to Datalogic Barcode Readers

Connecting UniOP to Datalogic Barcode Readers Connecting UniOP to Datalogic Barcode Readers This Technical Note contains the information needed to connect UniOP to Datalogic Barcode Scanners. Contents 1. Introduction...1 2. Designer setup...1 2.1

More information

ecov-110 User s Manual (V0.9f_Eng)

ecov-110 User s Manual (V0.9f_Eng) ecov-110 User s Manual (V0.9f_Eng) 1. Product Introduction 2. Windows Utility 3. Web Console 4. Command Mode 5. Connection Test Appendix A. ecov-110 Firmware Upgrade Appendix B. ecov-110 Serial Advanced

More information

Application Note BDLxxxx RS232 SERIAL INTERFACE COMMUNICATION PROTOCOL (SICP V1.82)

Application Note BDLxxxx RS232 SERIAL INTERFACE COMMUNICATION PROTOCOL (SICP V1.82) Application Note BDLxxxx RS232 SERIAL INTERFACE COMMUNICATION PROTOCOL (SICP V1.82) Table of Contents 1. INTRODUCTION... 1 1.1 PURPOSE... 1 1.2 DEFINITIONS, ABBREVIATIONS AND ACRONYMS... 1 2. COMMAND PACKET

More information

REV: C JAN 2008 GIGA-TMS REGISTERED TO ISO

REV: C JAN 2008 GIGA-TMS REGISTERED TO ISO PCR340 Series Dual Frequency RFID Reader Manual Part Number: TM951119 REV: C JAN 2008 GIGA-TMS REGISTERED TO ISO 9001:2000 8F, No.31, Lane 169, Kang-Ning St., Hsi-Chih Taipei Hsien, 221 Taiwan TEL:(886)

More information

VIGOR PROGRAMMABLE LOGICAL CONTROLLERS VIGOR ELECTRIC CORP.

VIGOR PROGRAMMABLE LOGICAL CONTROLLERS VIGOR ELECTRIC CORP. VIGOR PROGRAMMABLE LOGICAL CONTROLLERS VIGOR ELECTRIC CORP. 232G RX TX VB-1COM VIGOR + 24V IN - SHORT FOR TR 485G D+ D- VB-1COM Serial Foreword This manual contains text, diagrams and explanations which

More information

PRE32 RS232 protocol v1.21 Tomas Andersson

PRE32 RS232 protocol v1.21 Tomas Andersson PRE32 RS232 protocol 2014-11-07 v1.21 Tomas Andersson Description This document describes the RS232 protocol used to control the PRE32 device. Command structure Commands are sent to the device using the

More information

Room Control Module (2000i-RCM) Interface Command Structure. for the

Room Control Module (2000i-RCM) Interface Command Structure. for the Room Control Module (2000i-RCM) Interface Command Structure for the 2000i-DVS Interactive Whiteboard 99-00544-00 Rev A0 FCC Warning This equipment has been tested and found to comply with the limits for

More information

FACTORY AUTOMATION. MANUAL OHV-F230-B17 PROFINET Gateway for OHV Handheld

FACTORY AUTOMATION. MANUAL OHV-F230-B17 PROFINET Gateway for OHV Handheld FACTORY AUTOMATION MANUAL OHV-F230-B17 PROFINET Gateway for OHV Handheld R With regard to the supply of products, the current issue of the following document is applicable: The General Terms of Delivery

More information

Menu & commands booklet FOR BAR CODE SCANNERS * *

Menu & commands booklet FOR BAR CODE SCANNERS * * Menu & commands booklet FOR BAR CODE SCANNERS *201002993520* Caution! This menu book may be revised or withdrawn at any time. Copyright 1993-99, Datalogic S.p.A. All Rights Reserved. No part of this publication

More information

RS485 INSTRUCTION MANUAL. DM4500 Series. Page 1 of 6

RS485 INSTRUCTION MANUAL. DM4500 Series. Page 1 of 6 RS485 INSTRUCTION MANUAL DM4500 Series Page 1 of 6 INDEX 1.1. RS485 OUTPUT...3 1.1.1. Introduction... 1.1.2. RS output menu diagram... Page 2 of 6 1.1 RS4 OUTPUT 1.1.1 Introduction The RS485 output option

More information

User Guide. Modbus Module. For Digistart soft starters. Part Number: Issue: 3.

User Guide. Modbus Module. For Digistart soft starters. Part Number: Issue: 3. User Guide Modbus Module For Digistart soft starters Part Number: 477-9-3 Issue: 3 General Information The manufacturer accepts no liability for any consequences resulting from inappropriate, negligent

More information

CR3000 Copy Reader. Installation Guide

CR3000 Copy Reader. Installation Guide CR3000 Copy Reader Installation Guide CONTENTS CONTENTS 1 CR3000 Installation Guide 1 What you get: 2 Installation Overview 3 Install the Reader 4 Mount the Reader 4 Connect the Reader Cables 5 Connect

More information

INSTRUCTION MANUAL SERIES

INSTRUCTION MANUAL SERIES INSTRUCTION MANUAL SERIES 1300 ( single board ) Appendix E complement for interactive software Rue du Puits-Godet 16 CH-2005 NEUCHATEL SA Phone:++41 (0)32 729 93 60 Fax: ++41 (0)32 724 10 23 Mail : info@irtsa.com

More information

INTERFACE SPECIFICATION

INTERFACE SPECIFICATION INTERFACE SPECIFICATION CELL-DYN 1800 SYSTEM LIST NO. 07H79-01 REVISION A ABBOTT DIAGNOSTICS DIVISION Abbott Laboratories Abbott Park, IL 60064, USA INTERFACE SPECIFICATION CELL-DYN 1800 SYSTEM Abbott

More information

Conitel ASYNC Adapter

Conitel ASYNC Adapter Conitel ASYNC Adapter TABLE OF CONTENTS SECTION 1 - DESCRIPTION...2 SECTION 2 - SPECIFICATIONS... SECTION - INSTALLATION...6 SECTION - CONTROLS AND INDICATORS...9 SECTION - NETWORK MANAGEMENT PORT...11

More information

T1K MODBUS Base Controller Specifications

T1K MODBUS Base Controller Specifications Base Controller 1 2 In This Chapter.... Base Controller Setting the DIP Switches Setting the Rotary Address Switches Port Pin out and Wiring RJ12 Serial Port Pin out and Wiring 2 2 Base Controller General

More information

Serial CFW-11. Communication Manual. Phone: Fax: Web:

Serial CFW-11. Communication Manual. Phone: Fax: Web: Motors Energy Automation Coatings Serial CFW-11 Communication Manual Language: English RS232 / RS485 Serial Communication Manual Series: CFW-11 Language: English Document Number: 0899.5741 / 03 Publication

More information

Hawki Advanced HyperTerminal Installation Instructions

Hawki Advanced HyperTerminal Installation Instructions Purpose Equipment Required This document provides a Novar Controls approved procedure for connecting a Hawki Advanced to a personal computer (PC) configured as an interactive ( dumb ) terminal through

More information

RM024 DVK USER GUIDE VERSION 1.2

RM024 DVK USER GUIDE VERSION 1.2 USER GUIDE VERSION 1.2 Americas: +1-800-492-2320 Asia: +852-2923-0610 REVISION HISTORY Version Revision Date Change Description Approved By 1.0 20 Dec 2012 Initial Release Chris Downey 1.1 15 Apr 2014

More information

DCA - 10 / DCA TTS Systems

DCA - 10 / DCA TTS Systems DCA - 10 / DCA - 20 Two Channell Load Cell Amplifie er Refere ence Manua al TTS Systems 14 Highpoint Business Village, Henwood,, Ashford, Kent,, TN24 8DH Telephone +44 (0)1233 624422 Telefax 0870 705 9678

More information

KRAMER ELECTRONICS LTD. USER MANUAL MODEL: VS-401USB 4x1 USB Switcher. P/N: Rev 1

KRAMER ELECTRONICS LTD. USER MANUAL MODEL: VS-401USB 4x1 USB Switcher. P/N: Rev 1 KRAMER ELECTRONICS LTD. USER MANUAL MODEL: VS-401USB 4x1 USB Switcher P/N: 2900-300029 Rev 1 Contents 1 Introduction 1 2 Getting Started 2 2.1 Achieving the Best Performance 2 2.2 Recycling Kramer Products

More information

THT-500-A/R (C5) 1. System Configuration COMMUNICATION INSTRUCTION MANUAL. No. THT5CE

THT-500-A/R (C5) 1. System Configuration COMMUNICATION INSTRUCTION MANUAL. No. THT5CE COMMUNICATION INSTRUCTION MANUAL THT-500-A/R (C5) No. THT5CE2 2018.04 This manual contains instructions for communication functions of the THT-500-A/R. 1. System Configuration 1.1 When Using USB Communication

More information

Addmaster Corporation

Addmaster Corporation IJ-1000 Ink-Jet Validation Printer Specification Addmaster Corporation Address: 225 East Huntington Drive Monrovia, CA 91016 Web: www.addmaster.com Phone: (626) 358-2395 FAX: (626) 358-2784 Document: ij1w.doc

More information

ECOV-110 User s Manual

ECOV-110 User s Manual ECOV-110 User s Manual 1. Product Introduction 2. Windows Utility 3. Web Console 4. Command Mode 5. Connection Test Appendix A. ECOV-110 Firmware upgrade ECOV-110 1. ECOV-110 Product Introduction: 1.1.

More information

WIZ125SR User Manual

WIZ125SR User Manual WIZ125SR User Manual ( Version 1.0 ) 2010 WIZnet Co., Ltd. All Rights Reserved. For more information, visit our website at http://www.wiznet.co.kr WIZ125SR User Manual (WIZnet Co., Ltd.) - 1 - Document

More information

Fluke Metrology Software

Fluke Metrology Software Fluke Metrology Software Version 7 MET/CAL 5500/CAL Getting Started Guide P/N 1275404 July 1999 Rev. 2, 9/04 1996-2004Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

ADDMASTER. Addmaster Corporation. IJ-3080 Journal/Validation Printer. Specification. IJ-3080 Specification

ADDMASTER. Addmaster Corporation. IJ-3080 Journal/Validation Printer. Specification. IJ-3080 Specification IJ-3080 Journal/Validation Printer Specification Provides the electrical, mechanical, and interface specifications of the IJ-3080 Journal/Validation Printer. Cover Models: IJ-3080 The Addmaster Model IJ-3080

More information

IB-3 PC Keyboard Wedge User s Manual. P/N Rev B

IB-3 PC Keyboard Wedge User s Manual. P/N Rev B IB-3 PC Keyboard Wedge User s Manual P/N 83-210040 Rev B Copyright 2006 by Microscan Systems, Inc., 1201 S.W. 7th Street, Renton, WA, U.S.A. 98057 (425) 226-5700 FAX: (425) 226-8682 ISO 9001:2000 Certification

More information

Communication Protocol Manual JOFRA CTC, ITC, MTC, ETC and Compact Copyright 2008 AMETEK Denmark A/S

Communication Protocol Manual JOFRA CTC, ITC, MTC, ETC and Compact Copyright 2008 AMETEK Denmark A/S Communication Protocol Manual JOFRA CTC, ITC, MTC, ETC and Compact Copyright 2008 AMETEK Denmark A/S Contents 1 Introduction...5 2 Protocol...5 2.1 Variables...5 2.2 Telegram structure...6 2.3 Packing

More information

EWAVE Inc Gracefield Ln. Dallas, Texas (972)

EWAVE Inc Gracefield Ln. Dallas, Texas (972) EWAVE Inc. 7419 Gracefield Ln. Dallas, Texas 75248 (972) 248-2931 www.electrowave.com STAMPER User s Manual Version 1.0 Ewave Radio Modems covered in this manual: STAMPER INTRODUCTION... 3 1.1 FEATURES

More information

Hematology Analyzer. Serial Communication Protocol. 1.7 release

Hematology Analyzer. Serial Communication Protocol. 1.7 release Abacus Arcus Hematology Analyzer Serial Communication Protocol 1.7 release DIATRON Messtechnik Ges.m.b.H. A-1141 Wien, Ameisgasse 49-51/2. AUSTRIA Tel.: (431) 914-85-00, 911-38-48 Fax: 914-85-07-15 Web:

More information

Serial Dongle RS232 Plug&Scan

Serial Dongle RS232 Plug&Scan Programming guide Serial Dongle RS232 Plug&Scan Version 2.56 Important notice: Every effort is made to ensure the accuracy of our product information; however, we accept no responsibility for errors or

More information

9146-R Smart Connector Connectors for the TE 9146-R

9146-R Smart Connector Connectors for the TE 9146-R Instruments 9146-R Smart Connector Connectors for the TE 9146-R Enables plug-and-play type operation of the 9146-R with a variety of input types. All transducer s information is contained within the plug

More information

Updating Reader Firmware

Updating Reader Firmware SWH-xxxx Readers Updating Reader Firmware Version C0 Document Part Number UM-096 July 2012 OVERVIEW This document describes how to download new firmware to Software House SWH-4xxx readers. The procedure

More information

Bluetooth V4.1 BLE RS-232 Serial Adapter Model: BLE-232B Bluetooth V4.2 BLE RS-232 Serial Adapter Model: BLE-232C

Bluetooth V4.1 BLE RS-232 Serial Adapter Model: BLE-232B Bluetooth V4.2 BLE RS-232 Serial Adapter Model: BLE-232C Bluetooth V4.1 BLE RS-232 Serial Adapter Model: BLE-232B Bluetooth V4.2 BLE RS-232 Serial Adapter Model: BLE-232C Applications: MSR card reader reader, RFID reader, IC card reader, Specifications: 1. Baud

More information

Supplement to Instruction Bulletin Micro-DCI 53ML5100A 53ML5100 MANUAL LOADING STATION REV. 1 FIRMWARE. PN24698 Rev. 1

Supplement to Instruction Bulletin Micro-DCI 53ML5100A 53ML5100 MANUAL LOADING STATION REV. 1 FIRMWARE. PN24698 Rev. 1 Supplement to Instruction Bulletin Micro-DCI 53ML5100A 53ML5100 MANUAL LOADING STATION REV. 1 FIRMWARE PN24698 Rev. 1 The Company MicroMod Automation, Inc. MicroMod Automation is dedicated to improving

More information

PIM Presco Interface Module

PIM Presco Interface Module INTRODUCTION The Interface Module (PIM) can be used to convert commonly used data formats to Nidac, or format data. PIM Interface Module Series 2 Revision B FEATURES Converts from,, Nidac, Clock & Data

More information

Overview. 08/06/2012 Page 2/47

Overview. 08/06/2012 Page 2/47 The information contained in this document has been developed solely for the purpose of providing general guidance to Cognex customers who need to configure communications between an DataMan reader and

More information

Mounting Dimensions / Viewing 2 Mounting Options 3. Wiring Configuration 4. Quick Set up Procedure 5. Changing Intensity 6.

Mounting Dimensions / Viewing 2 Mounting Options 3. Wiring Configuration 4. Quick Set up Procedure 5. Changing Intensity 6. Section Mounting Dimensions / Viewing 2 Mounting Options 3 Section 2 Wiring Configuration 4 Section 3 Quick Set up Procedure 5 Section 4 Changing Intensity 6 Section 5 Option Summary 7 Section 6 Option

More information

LT900SERIES. 1xN Multi-Channel Switch Operation Manual. For RS-232 Control with 16-pin Connector

LT900SERIES. 1xN Multi-Channel Switch Operation Manual. For RS-232 Control with 16-pin Connector LT900SERIES 1xN Multi-Channel Switch Operation Manual For RS-232 Control with 16-pin Connector Table of Contents General Information.. 1 General Specifications...... 2 Interface Connectors..... 3 RS232

More information

Doug Fleenor Design, Inc. RS-232 to DMX512 Interface, 2 Generation. March 8, 2010 (Software V1.2)

Doug Fleenor Design, Inc. RS-232 to DMX512 Interface, 2 Generation. March 8, 2010 (Software V1.2) Doug Fleenor Design, Inc. nd RS-232 to DMX512 Interface, 2 Generation March 8, 2010 (Software V1.2) The second generation RS-232 to DMX512 interface has numerous features beyond the original device. The

More information

USER MANUAL FOR GS100/GS1003G

USER MANUAL FOR GS100/GS1003G USER MANUAL FOR GS100/GS1003G 1 Table of Contents 1. INTRODUCTION... 3 2. FEATURES... 3 3. OPERATION... 3 4.CONNECTION DETAILS... 4 5.CONFIGURATION... 5 5.1 Hyper Terminal Setting... 5 5.2 GS100 Configuration...

More information

4.2 Programming with MotionWiz

4.2 Programming with MotionWiz 4. Programming with MotionWiz The slvd serial kit is supplied to enable communication between a PC and the drive. The kit includes an RS-4/RS- converter, relative 0V~ power supply and serial connection

More information

I/O Interface DNA300

I/O Interface DNA300 DNA300 OPERATIONS MANUAL 1 Ring Communications, Inc. Crisis Alert System I/O Interface DNA300 2 DNA300 OPERATIONS MANUAL 2006/06/13 DNA300 OPERATIONS MANUAL 3 Contents INTRODUCTION... 5 INSTALLATION...

More information

User Manual for VE GSM Modem

User Manual for VE GSM Modem User Manual for VE GSM Modem INTRODUCTION GSM/GPRS Smart Modem is a multi-functional, ready to use, rugged unit that can be embedded or plugged into any application. The Smart Modem can be controlled and

More information

BIS L , BIS Z-EL-...

BIS L , BIS Z-EL-... BIS L-400-..., BIS Z-EL-... Technical Description, User's Guide English www.balluff.com 1 2 3 4 5 6 7 8 9 Notes to the user 4 1.1 About this manual 4 1.2 Structure of the manual 4 1.3 Typographical conventions

More information

DSP860 Operation Manual

DSP860 Operation Manual DSP860 Operation Manual 1. Information 2. Introduction 3. Installation 4. Pin Assignment 5. DIP Switch Setting 6. Character Tables 7. Software Control (Command List) 8. Specifications 9. Instruction of

More information

DSP840 Operation Manual Version 3.0

DSP840 Operation Manual Version 3.0 DSP840 Operation Manual Version 3.0 1. Information 2. Introduction 3. Installation 4. Pin Assignment 5. DIP Switch Setting 6. Character Tables 7. Software Control (Command List) 8. Specifications 9. Instruction

More information

MODEL DSU Kbps Rack Mount CSU/DSU OPERATOR S MANUAL. 280 I-80 West Exit 1 PO Box 1330 Verdi NV 89439

MODEL DSU Kbps Rack Mount CSU/DSU OPERATOR S MANUAL. 280 I-80 West Exit 1 PO Box 1330 Verdi NV 89439 MODEL DSU-56 56 Kbps Rack Mount CSU/DSU OPERATOR S MANUAL 280 I-80 West Exit 1 PO Box 1330 Verdi NV 89439 TEL: 775-345-8000 FAX: 775-345-8010 E-MAIL: SUPPORT @S.NET DSU-56 Rack Mount TABLE OF CONTENTS

More information

NetBiter Serial Server User Manual

NetBiter Serial Server User Manual User Manual IntelliCom Innovation AB Linjegatan 3D SE-302 50 Halmstad SWEDEN Phone +46 35 18 21 70 Fax +46 35 17 29 09 email info@intellicom.se web www.intellicom.se Revision List Revision Date Author

More information

SCADA Point-to-Point Multiplexer

SCADA Point-to-Point Multiplexer SCADA Point-to-Point Multiplexer TABLE OF CONTENTS SECTION 1 - DESCRIPTION...2 SECTION 2 - SPECIFICATIONS... SECTION - INSTALLATION... SECTION - CONTROLS AND INDICATORS...9 SECTION - NETWORK MANAGEMENT

More information

RS232 User Guide. Planar Simplicity Series 4K Displays SL4364K SL5564K SL6564K SL7564K SL8664K. Simplicity Series 4K Displays User Guide A

RS232 User Guide. Planar Simplicity Series 4K Displays SL4364K SL5564K SL6564K SL7564K SL8664K. Simplicity Series 4K Displays User Guide A RS232 User Guide Planar Simplicity Series 4K Displays SL4364K SL5564K SL6564K SL7564K SL8664K 020-1344-00A Page 1 Copyright March 2018 by Leyard Optoelectronics Co., Ltd. and Planar Systems, Inc. All rights

More information

DFS/Cutter. Debugging Communication Problems

DFS/Cutter. Debugging Communication Problems DFS/Cutter Technical Support Bulletin Debugging Communication Problems Abstract: This TSB is written to help users debug communication problems where the cutter appears to ignore everything the computer

More information

KB232. PS/2 Keyboard RS-232 Adapter Part # SA0008 (Version 3.0) Copyright 2003 L3 Systems, Inc. Redmond

KB232. PS/2 Keyboard RS-232 Adapter Part # SA0008 (Version 3.0) Copyright 2003 L3 Systems, Inc. Redmond KB232 PS/2 Keyboard RS-232 Adapter Part # SA0008 (Version 3.0) Copyright 2003 L3 Systems, Inc. Redmond Command C Displays Configuration String CW= D Lkk=aa,bb P E H V T Quick Reference Notes Field

More information

REL 512 Connectivity With A Harris Westronics RTU Using DNP 3.0

REL 512 Connectivity With A Harris Westronics RTU Using DNP 3.0 ABB Application Note Substation Automation and Protection Division REL 512 AN-45A-99 REL 512 Connectivity With A Harris Westronics RTU Using DNP 3.0 ABSTRACT: DNP 3.0 is a popular communication protocol

More information

NL6000 Wireless Modem NL6000 User Manual Rev 1.1

NL6000 Wireless Modem NL6000 User Manual Rev 1.1 NL6000 Wireless Modem NL6000 User Manual Rev 1.1 WWW.RFNEULINK.COM 6. Procedures This section includes step-by-step procedures for configuring all the parameters that control the operation of the NL6000.

More information

Instruction manual CIU Prime type 880 (CIU858 emulation)

Instruction manual CIU Prime type 880 (CIU858 emulation) Instruction manual CIU Prime type 88 (CIU858 emulation) Version.2 November 2 Part no.: 446529 Enraf B.V. P.O. Box 82 26 AV Delft Netherlands Tel.: +3 5 26986, Fax: +3 5 269574 Email: Info@enraf.nl http://www.enraf.com

More information

Single-Axis Counter BDD 611

Single-Axis Counter BDD 611 Single-Axis Counter BDD 611 Technical Description, User's Guide english No. 611-150 D/E. Edition 0705; Subject to modification. Replaces edition 0701. www.balluff.com Balluff GmbH Schurwaldstrasse 9 73765

More information

CLOCKAUDIO. MR88 Automatic Microphone Mixer RS232 Programming Version 4.2

CLOCKAUDIO. MR88 Automatic Microphone Mixer RS232 Programming Version 4.2 CLOCKAUDIO MR88 Automatic Microphone Mixer RS232 Programming Version 4.2 Clockaudio Limited, 9 Stratfield Park Elettra Avenue, WATERLOOVILLE Hampshire. UK Tel : +44 (0)2392 251193 Fax : +44 (0)2392 251201

More information

RS485/RS232 Communications Interface

RS485/RS232 Communications Interface EUROTHERM DRIVES RS485/RS232 Communications Interface Technical Manual HA466357U001 Issue 3 Compatible with Version 4.x Software Copyright Eurotherm Drives Limited 2002 All rights strictly reserved. No

More information

Bluetooth/RS-232 Enhanced Data Communication

Bluetooth/RS-232 Enhanced Data Communication Bluetooth/RS-232 Enhanced Data Communication 9650-0307-01 Rev. E The issue date or revision level for this operation guide is shown on the front cover. ZOLL, and M Series are registered trademarks, and

More information

COMMUNICATION WITH OTHER SERIAL DEVICES

COMMUNICATION WITH OTHER SERIAL DEVICES Application Bulletin: #169 Date: November 2009 BLACKBELT COMMUNICATIONS INTERFACE The setup and operation of the Sentinel BLACKBELT communication options are explained in Error! Reference source not found.,

More information

Unique Micro Design First Choice of Professional Systems Intergrators. Model 361 Terminal Keyboard Wedge User Manual

Unique Micro Design First Choice of Professional Systems Intergrators. Model 361 Terminal Keyboard Wedge User Manual Model 361 User Manual Unique Micro Design First Choice of Professional Systems Intergrators Model 361 Terminal Keyboard Wedge User Manual Document Reference : DOC-M361-UM UMD Part Number : 6-0361-993-5

More information