Exam 8N080 - Introduction MRI

Size: px
Start display at page:

Download "Exam 8N080 - Introduction MRI"

Transcription

1 Exam 8N080 - Introduction MRI Friday January 23 rd 2015, h For this exam you may use an ordinary calculator (not a graphical one). In total there are 6 assignments and a total of 65 points can be earned. You are allowed to take this exam with you afterwards. If you don t quite understand a certain assignment, don t hesitate to ask for a translation. Good luck! 1) An MRI operator wants to image an organ with a T 2 of 30 ms using a T 2-weighted spin-echo sequence. The tissue surrounding the organ has a T 2 of 50 ms. The proton density of the organ ρρ oooooooooo = 0.8, the proton density of the surrounding tissue ρρ tttttttttttt = 0.5. (total of 11 points) a. What is the approximate signal formula for a spin-echo sequence that is T 2-weighted (i.e. TR>>T 1)? (1 point) b. Sketch the signal amplitude of both the organ and the tissue as a function of the echo time TE and in the same graph, i.e. sketch SS tttttttttttt (TE) and SS oooooooooo (TE). (2 points) c. The contrast between the organ and the surrounding tissue is defined as the difference in signal intensity. Derive a formula for the contrast CC(TE). (2 points) d. Show that in this case we should choose TE as short as possible for best contrast. (3 points) e. Also explain why in MRI we can t take TE=0. (1 point) f. Now assume ρρ oooooooooo = ρρ tttttttttttt. What TE should we choose in this case for best contrast? (2 points) a) SS = ρρ ee TTTT TT2 1 ee TTTT TT1, and if we use TR>>T 1 we get SS ρρ ee TTTT TT2 b) See the figure below; intersecting curves since the organ has a higher ρρ but also a faster decay. 1

2 c) CC = ρρ oooooooooo ee TTTT/TT 2,oooooooooo ρρ tttttttttttt ee TTTT/TT 2,tttttttttttt. Notice that the signal difference can be both positive and negative; see the plot in b). d) If we calculate dddd dddddd = 0, we will find a good TE value (local maximum) to the right of the intersection of the curves. We get dddd dddddd = ρρ oooooooooo ee TTTT gggggggg TT2,oooooooooo + ρρ tttttttttttt TT 2,oooooooooo ee TTTT gggggggg TT 2,tttttttttttt 1 1 TT 2,oooooooooo TT 2,tttttttttttt = 0 which yields TTTT gggggggg = ln TT 2,ttttttttttttρρ oooooooonn TT2,oooooooooo ρρ / = tttttttttttt TT 2,tttttttttttt 73.6 mmmm. The contrast (absolute signal difference) at this value is SS tttttttttttt (TTTT) SS oooooooooo (TTTT) = Notice that this is smaller than the signal difference at TE=0, which is =0.3! So indeed it s a good TE value, but not the optimal one! We d better choose the shortest possible TE. (Note that if for technical reasons or so TE cannot be made shorter than around 25 ms, we d better aim for the calculated value of 73.6 ms, but no points are subtracted if you didn t calculate this critical value of 25 ms). e) We cannot take TE=0 because the echo can only occur is we apply gradients or refocussing RF-pulses after the excitation pulse. A certain amount of time is needed to apply these gradients/pulses. f) The equation that was derived in d) can now be used, ρρ oooooooooo and ρρ tttttttttttt now cross out since they are now equal. So we now get TTTT oooooooooooooo = ln TT 2,tttttttttttt / TT 2,oooooooooo 1 1 TT 2,oooooooooo TT 2,tttttttttttt =38.3 ms 2) For a certain scan we define a field of view (FOV) of 30x15 cm and we want to create an image of 256x128 pixels, see the figure below. The centre of our coordinate system (xx, yy) coincides with the middle of the FOV. The xx-direction is the read-out direction, the corresponding gradient strength is GG xx =7 mt/m. The gyromagnetic ratio γγ= MHz/T. (total of 11 points) a. When recording the echo s, the signal that comes from the receive coil is typically demixed before it is sampled. This demixing means that the so called carrier frequency ωω 0 is subtracted. Explain that this means that the signal actually becomes independent of the main magnetic field strength BB 0. (2 points) 2

3 b. In the remainder of this assignment we assume that the signal has already been demixed. What is the maximum signal frequency ff mmmmmm during read-out? (2 points) c. What sample frequency ff ss do we then minimally need? (1 point) d. What is the acquisition time of a single echo, TT eeeehoo? Hint: how many samples do we acquire per echo? (1 point) e. In this case we sample kk-space (kk xx, kk yy ) by taking samples in time (related to step size ddkk xx ) and by taking different values for GG yy (related to kk yy ). It is given that the step size dddd yy (related to dddd yy ) must be chosen such that the maximum phase step is π radians. Derive the value of dddd yy. The duration of the phase encoding gradient TT pppp =0.8 ms. (2 points) f. It is given that the repetition time TR of this sequence is 500 ms, and that we acquire 1 echo per excitation. What is the total acquisition time TT aaaaaa? (1 point) g. Assume we rotate the FOV 90 o around the origin. Should we now increase or decrease ff ss? And what about dddd yy? Explain your answers. (2 points) a) The precession frequency in the presence of gradients is generally given by ωω = γγ BB 0 + GG xx xx + GG yy yy + GG zz zz = ωω 0 + γγ GG xx xx + GG yy yy + GG zz zz. If we subtract ωω 0, the term that is dependent on BB 0 drops from the equation. (We can interpret this at looking at the spins from within the rotating frame of reference rather than the static laboratory frame). b) We can use ff = γγ BB. During read-out, GG 2ππ xx is on, and the local field then is highest at xx = FFFFFF xx /2. So we get ff mmmmmm = γγ GG 2ππ xx c) ff ss = 2ff mmmmmm =89.4 khz (Nyquist) FFFFFF xx 2 = 44.7 khz. d) TT eeeehoo = 256 tt ss = 256 = 2.9 ms, with tt ff ss the sample time. ss e) The largest phase steps are made at the borders of the FOV in phase encode direction, so in this case yy-direction. As an equation we then get ππ = γγ dddd FFFFFF yy yy which can be 2 1 rewritten to dddd yy = = 0.2 mt/m. γγffffff yy TT pppp f) TT aaaaaa = = 64 s g) ff ss can be decreased since the FOV gets narrower in read-out direction, so the maximum frequency during read out goes down as well. And from the equation in e) we can see that a larger FFFFFF yy entails a smaller phase encoding step size dddd yy. 3) Given is the schematic of a multi-echo sequence, see the figure below. The top row represents the RF pulses, the following rows indicate the switching of the 3 gradient systems, and the bottom row indicates the data acquisition (read-out; sampling indicated by vertical bars). (total of 10 points) 3

4 a. Indicate for each of the 3 gradients whether it is used as slice selection gradient, phase encoding gradient, or frequency encoding gradient. Explain your answers. (3 points) b. Is this sequence T 2 or T 2* weighted? Explain your answer. (1 point) c. Gradient lobes A and B are of the same magnitude, but opposing sign. Explain why gradient lobe B is twice as long. (1 point) d. Copy the following schematic to your answering sheet and draw the kk-space trajectory of this sequence. Indicate time instances 1-9 clearly in your sketch. Make sure to use the fact that the positive GG yy gradient lobes have the same magnitude as the negative lobe, but only 1/3 the width (time duration). (3 points) e. In which direction do you expect most relaxation-induced artefacts in the image, phase encoding direction or frequency encoding direction? Explain your answer. (2 points) a) GG xx is the slice selection gradient since it is on during the application of the RF pulse; GG yy is the phase encoding gradient since its on before each read out; GG zz is the frequency 4

5 encoding gradient since it is on during data-acquisition (read-out), as can be seen from the timing of the sampling on the lower time line. b) This is a gradient echo sequence without refocussing RF pulses, so it s a T 2-weighted sequence c) So that the dephasing caused by gradient lobe A is compensated exactly halfway gradient lobe B, since phase is proportional to gradient amplitude and duration (φφ = γγγγγγ). This means that the echo that occurs during the application of B has maximal amplitude in the middle of the read-out d) See the sketch below. Note that the last echo is at height kk yy = 0 because the cumulative effect of the yy-gradient is zero for this third echo. e) In phase encoding direction, since most relaxation occurs between echo s (time interval TR) rather than within echo s (time interval shorter than TE). 4) Given is an MRI sequence with a repetition time TR of 100 ms. We are imaging a tissue with a T 1 of 150 ms and we use a flip angle α=30 o. The equilibrium magnetization value MM 0 =1. See the schematic below. Each time point (i) is directly prior or after an α-pulse, and the duration of these pulses may be neglected (they are very short). (total of 8 points) a. Calculate MM zz directly after the first α-pulse; time point (1) in the schematic. (1 point) b. This MM zz now has until the next α-pulse to recover towards MM 0. Draw this recovery process MM zz (tt) and derive the corresponding formula. (3 points) c. Use this formula to find the value of MM zz directly prior to the second α-pulse, i.e. time point (2). (1 point) d. The second α-pulse is now applied. Again, MM zz has a period TR to recover towards MM 0. Find the value of MM zz directly prior to the third α-pulse, i.e. time point (3). (2 points) 5

6 e. Now find the amplitude of the transversal magnetization MM xx,yy at time point (4). (1 point) a) MM zz (tt 1 ) = cos(αα) MM b) See the figure below. This is T 1 relaxation starting at cos(αα) MM 0 and recovering towards MM 0. The formula is MM zz (tt) = MM 0 + (cos(αα) MM 0 MM 0 )ee tt/tt 1, with tt the time after the α-pulse. c) Just plug in tt = TTTT in the equation: MM zz (TTTT)=0.93 d) We have a similar recovery process as sketched in b), only now we start at 0.93 MM 0. Deriving a similar equation and again plugging in tt = TTTT gives MM zz (tt 3 ) = e) MM TT (tt 4 ) = sin(αα) ) The following questions are about the kk-space. (total of 13 points) a. Briefly explain what the kk-space is. Use the terms raw data, image, spatial frequency and Fourier transform in your answer. (3 points) b. Link the following spatial frequency images A-E to their kk-space positions 1-5. (2 points) c. Often MR images are based of full kk-space coverage. For a certain application, fast imaging is needed, and only part of the kk-space (in this case 50%) is acquired, see the figure below. Which partial kk-space coverage A-E would you prefer? Motivate your answer, and for each other option give a reason why you didn t choose it. It is given 6

7 that the horizontal direction is the frequency encoding direction, and the vertical direction is the phase encoding direction. (4 points) d. In the above figure, one can find 2 pairs of partial kk-spaces which have the same acquisition time. Which 2 pairs are these? Explain your answer. (2 points) e. Given below is an actual kk-space. Explain why most signal intensity is found in the middle. (2 points) kk-space: bright (lots of signal) near the origin a) The kk-space represents the raw data of an MRI scanner. It is a way to store the data (samples) from the individual echo s in a 2D matrix. The kk-space is linked to the MR image via a Fourier transform. It represents the spatial frequencies which are present in the image; the brighter a certain position in kk-space, the stronger the corresponding spatial frequency. b) The further away from the center of kk-space, the higher the spatial frequency A: large number of periods in both xx and yy 5 B: Moderate number of periods in xx 3 C: One period in xx, two periods in yy 2 7

8 D: One period in both xx and yy 1 E: One period in yy 4 c) For speed up, we need to reduce the number of phase encodes. This means we want partial coverage in yy-direction, and this applies to kk-spaces A, C and E. However, in E the phase encoding steps are too large, which would cause aliasing (fold-in). The difference between A and C is that in A we only loose high frequency content (details) in yydirection, if this is undesirable than C is the better option, although its acquisition time is slightly higher. B and D don t offer a reduction in number of phase encodes and therefore don t give a shorter acquisition time. Furthermore, D again gives aliasing (this time because the sample frequency is too low). d) As said, the number of lines in yy-direction corresponds to the number of echo s, which is a measure for the acquisition time. The number of lines in yy-direction is the same for A and E (half the number of echo s as in the full kk-space), and for B and D (partial readout of all echo s and undersampling of all echo s, respectively). e) Let s say yy again is the phase encoding direction and xx again is the read-out direction. Phase encoding introduces a phase difference in yy-direction, which corresponds to signal loss (dephasing). This dephasing is lowest for zero phase encoding (centre of kk-space in yy-direction), so that s were most signal is found along the kk yy axis. Furthermore, the echo intensity is highest at the middle of read-out (refocussing in xx-direction), which is the middle of kk-space in xx-direction. So indeed the center of k-space (low kk xx and low kk yy ) is brightest. 6) Traditional MR sequences (such as spin-echo) fill kk-space along rectangular pathways, using rectangular phase encoding and frequency encoding gradients. An alternative is provided below; frequency encoding and phase encoding here are a cosine and a sine of linearly increasing amplitude, respectively. Samples are acquired continuously, from the moment of the excitation pulse onwards. (total of 12 points) a. Sketch the kk-space trajectory of this sequence, with GG ffffffff and GG pphaaaaaa corresponding to kk xx and kk yy, respectively. (4 points) b. The 4 corners of kk-space are not filled by this sequence. Discuss what effect this has on the resulting image. (2 points) 8

9 c. Explain that for traditional image reconstruction, the data needs to be interpolated (this is called regridding ). (3 points) d. Is this sequence T 2 or T 2* weighted? (1 point) e. Explain what consequences these smooth gradient shapes have on hardware demands and scanner noise. (2 points) a. (4 pt) See the figure below b. (2 pt) The further away from the center of kk-space, the higher the corresponding spatial frequency. So if we don t fill the corners, we miss the highest spatial frequencies. However, notice that this also means that the resolution (level of detail) is highest in diagonal directions for rectangular kk-spaces. So effectively, the resolution is more homogenous for this trajectory (i.e. the same in all radial directions) c. (3 pt) The samples are not located on a grid, so we cannot directly apply a fast Fourier transform (FFT) for image reconstruction; this requires a 2D matrix as input (i.e. a rectangular grid of data). Hence the term regridding. d. (1 pt) T 2* weighted, since there are no refocussing pulses (only a single 90 o excitation pulse). e. (2 pt) It s quite difficult to create perfectly rectangular gradient shapes; due to the inductance of the gradient coils, some time is needed for the coil currents to start up/switch off, so the upslope/downslope is not perfectly vertical. This problem is much less in smoother gradient shapes in which coil current varies much slower. Furthermore, rapid switching of the gradients causes rapid flux change, which in turn induce eddy currents in the scanner plating. This causes mechanical vibrations and thus noise. This is less so for smooth gradient shapes (slower flux changes, less eddy currents, less noise). 9

10 THE END 10

Midterm Review

Midterm Review Midterm Review - 2017 EE369B Concepts Noise Simulations with Bloch Matrices, EPG Gradient Echo Imaging 1 About the Midterm Monday Oct 30, 2017. CCSR 4107 Up to end of C2 1. Write your name legibly on this

More information

MRI Physics II: Gradients, Imaging

MRI Physics II: Gradients, Imaging MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes

More information

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE MRES 7005 - Fast Imaging Techniques Module 4 K-Space Symmetry Review K-Space Review K-Space Symmetry Partial or Fractional Echo Half or Partial Fourier HASTE Conditions for successful reconstruction Interpolation

More information

Imaging Notes, Part IV

Imaging Notes, Part IV BME 483 MRI Notes 34 page 1 Imaging Notes, Part IV Slice Selective Excitation The most common approach for dealing with the 3 rd (z) dimension is to use slice selective excitation. This is done by applying

More information

XI Signal-to-Noise (SNR)

XI Signal-to-Noise (SNR) XI Signal-to-Noise (SNR) Lecture notes by Assaf Tal n(t) t. Noise. Characterizing Noise Noise is a random signal that gets added to all of our measurements. In D it looks like this: while in D

More information

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford Diffusion MRI Acquisition Karla Miller FMRIB Centre, University of Oxford karla@fmrib.ox.ac.uk Diffusion Imaging How is diffusion weighting achieved? How is the image acquired? What are the limitations,

More information

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions EE225E/BIOE265 Spring 2013 Principles of MRI Miki Lustig Assignment 8 Solutions 1. Nishimura 7.1 P)jc%z 5 ivcr3. 1. I Due Wednesday April 10th, 2013 (a Scrhon5 R2iwd b 0 ZOms Xn,s r cx > qs 4-4 8ni6 4

More information

surface Image reconstruction: 2D Fourier Transform

surface Image reconstruction: 2D Fourier Transform 2/1/217 Chapter 2-3 K-space Intro to k-space sampling (chap 3) Frequenc encoding and Discrete sampling (chap 2) Point Spread Function K-space properties K-space sampling principles (chap 3) Basic Contrast

More information

Role of Parallel Imaging in High Field Functional MRI

Role of Parallel Imaging in High Field Functional MRI Role of Parallel Imaging in High Field Functional MRI Douglas C. Noll & Bradley P. Sutton Department of Biomedical Engineering, University of Michigan Supported by NIH Grant DA15410 & The Whitaker Foundation

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

Steen Moeller Center for Magnetic Resonance research University of Minnesota

Steen Moeller Center for Magnetic Resonance research University of Minnesota Steen Moeller Center for Magnetic Resonance research University of Minnesota moeller@cmrr.umn.edu Lot of material is from a talk by Douglas C. Noll Department of Biomedical Engineering Functional MRI Laboratory

More information

Applications Guide for Interleaved

Applications Guide for Interleaved Applications Guide for Interleaved rephase/dephase MRAV Authors: Yongquan Ye, Ph.D. Dongmei Wu, MS. Tested MAGNETOM Systems : 7TZ, TRIO a Tim System, Verio MR B15A (N4_VB15A_LATEST_20070519) MR B17A (N4_VB17A_LATEST_20090307_P8)

More information

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 5. Solutions

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 5. Solutions EE225E/BIOE265 Spring 211 Principles of MRI Miki Lustig Handout Assignment 5 Solutions 1. Matlab Exercise: 2DFT Pulse sequence design. In this assignment we will write functions to design a 2DFT pulse

More information

A Virtual MR Scanner for Education

A Virtual MR Scanner for Education A Virtual MR Scanner for Education Hackländer T, Schalla C, Trümper A, Mertens H, Hiltner J, Cramer BM Hospitals of the University Witten/Herdecke, Department of Radiology Wuppertal, Germany Purpose A

More information

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most

More information

MRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison

MRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison MRI Imaging Options Frank R. Korosec, Ph.D. Departments of Radiolog and Medical Phsics Universit of Wisconsin Madison f.korosec@hosp.wisc.edu As MR imaging becomes more developed, more imaging options

More information

Unit WorkBook 4 Level 5 ENG U36 Advanced Mechanical Principles 2018 UniCourse Ltd. All rights Reserved. Sample

Unit WorkBook 4 Level 5 ENG U36 Advanced Mechanical Principles 2018 UniCourse Ltd. All rights Reserved. Sample 018 UniCourse Ltd. All rights Reserved. Pearson BTEC Levels 5 Higher Nationals in Engineering (RQF) Unit 36: Advanced Mechanical Principles Unit Workbook 4 in a series of 4 for this unit Learning Outcome

More information

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health MRI image formation Indiana University School of Medicine and Indiana University Health Disclosure No conflict of interest for this presentation 2 Outlines Data acquisition Spatial (Slice/Slab) selection

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

UNIVERSITY OF SOUTHAMPTON

UNIVERSITY OF SOUTHAMPTON UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions

More information

New Technology Allows Multiple Image Contrasts in a Single Scan

New Technology Allows Multiple Image Contrasts in a Single Scan These images were acquired with an investigational device. PD T2 T2 FLAIR T1 MAP T1 FLAIR PSIR T1 New Technology Allows Multiple Image Contrasts in a Single Scan MR exams can be time consuming. A typical

More information

Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 2004

Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 2004 Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 24 1 Alec Chi-Wah Wong Department of Electrical Engineering University of Southern California 374 McClintock

More information

Single-sided Magnet Diffusion Measurements

Single-sided Magnet Diffusion Measurements Single-sided Magnet Diffusion Measurements Kuldeep S. Panesar April 2013 (kuldeep.panesar@gmail.com) Contents 1. Initial configuration... 2 2. Preparation for diffusion measurement... 5 3. Starting a diffusion

More information

Advanced Imaging Trajectories

Advanced Imaging Trajectories Advanced Imaging Trajectories Cartesian EPI Spiral Radial Projection 1 Radial and Projection Imaging Sample spokes Radial out : from k=0 to kmax Projection: from -kmax to kmax Trajectory design considerations

More information

Section Graphs of the Sine and Cosine Functions

Section Graphs of the Sine and Cosine Functions Section 5. - Graphs of the Sine and Cosine Functions In this section, we will graph the basic sine function and the basic cosine function and then graph other sine and cosine functions using transformations.

More information

Head motion in diffusion MRI

Head motion in diffusion MRI Head motion in diffusion MRI Anastasia Yendiki HMS/MGH/MIT Athinoula A. Martinos Center for Biomedical Imaging 11/06/13 Head motion in diffusion MRI 0/33 Diffusion contrast Basic principle of diffusion

More information

COBRE Scan Information

COBRE Scan Information COBRE Scan Information Below is more information on the directory structure for the COBRE imaging data. Also below are the imaging parameters for each series. Directory structure: var/www/html/dropbox/1139_anonymized/human:

More information

Getting Started with montaj GridKnit

Getting Started with montaj GridKnit Getting Started with montaj GridKnit Today s explorers use regional and local geophysical compilations of magnetic, radiometric and apparent resistivity data to perform comprehensive geologic interpretations

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Following on from the two previous chapters, which considered the model of the

Following on from the two previous chapters, which considered the model of the Chapter 5 Simulator validation Following on from the two previous chapters, which considered the model of the simulation process and how this model was implemented in software, this chapter is concerned

More information

Compressed Sensing And Joint Acquisition Techniques In Mri

Compressed Sensing And Joint Acquisition Techniques In Mri Wayne State University Wayne State University Theses 1-1-2013 Compressed Sensing And Joint Acquisition Techniques In Mri Rouhollah Hamtaei Wayne State University, Follow this and additional works at: http://digitalcommons.wayne.edu/oa_theses

More information

Sparse sampling in MRI: From basic theory to clinical application. R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology

Sparse sampling in MRI: From basic theory to clinical application. R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology Sparse sampling in MRI: From basic theory to clinical application R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology Objective Provide an intuitive overview of compressed sensing

More information

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 16 Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 3.1 MRI (magnetic resonance imaging) MRI is a technique of measuring physical structure within the human anatomy. Our proposed research focuses

More information

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution.

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution. MRES 7005 - Fast Imaging Techniques Module 5: Dynamic Imaging and Phase Sharing (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review Improving Temporal Resolution True-FISP (I) True-FISP (II) Keyhole

More information

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005 Field Maps John Pauly October 5, 25 The acquisition and reconstruction of frequency, or field, maps is important for both the acquisition of MRI data, and for its reconstruction. Many of the imaging methods

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

2D spatially selective excitation pulse design and the artifact evaluation

2D spatially selective excitation pulse design and the artifact evaluation EE 591 Project 2D spatially selective excitation pulse design and the artifact evaluation 12/08/2004 Zungho Zun Two-dimensional spatially selective excitation is used to excite a volume such as pencil

More information

Plane Wave Imaging Using Phased Array Arno Volker 1

Plane Wave Imaging Using Phased Array Arno Volker 1 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic More Info at Open Access Database www.ndt.net/?id=16409 Plane Wave Imaging Using Phased Array

More information

Sampling, Ordering, Interleaving

Sampling, Ordering, Interleaving Sampling, Ordering, Interleaving Sampling patterns and PSFs View ordering Modulation due to transients Temporal modulations Slice interleaving Sequential, Odd/even, bit-reversed Arbitrary Other considerations:

More information

Magnetic Resonance Angiography

Magnetic Resonance Angiography Magnetic Resonance Angiography Course: Advance MRI (BIOE 594) Instructors: Dr Xiaohong Joe Zhou Dr. Shadi Othman By, Nayan Pasad Phase Contrast Angiography By Moran 1982, Bryan et. Al. 1984 and Moran et.

More information

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous,

More information

Compressed Sensing for Rapid MR Imaging

Compressed Sensing for Rapid MR Imaging Compressed Sensing for Rapid Imaging Michael Lustig1, Juan Santos1, David Donoho2 and John Pauly1 1 Electrical Engineering Department, Stanford University 2 Statistics Department, Stanford University rapid

More information

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street MRI is located in the sub basement of CC wing. From Queen or Victoria, follow the baby blue arrows and ride the CC south

More information

FINAL EXAM REVIEW CH Determine the most precise name for each quadrilateral.

FINAL EXAM REVIEW CH Determine the most precise name for each quadrilateral. FINAL EXAM REVIEW CH 6 1. Determine the most precise name for each quadrilateral. 2. Find the measures of the numbered angle for the following parallelogram a. A(-5,7), B(2,6), C(-4,0), D(5,-3) 1 3 2 35

More information

The SIMRI project A versatile and interactive MRI simulator *

The SIMRI project A versatile and interactive MRI simulator * COST B21 Meeting, Lodz, 6-9 Oct. 2005 The SIMRI project A versatile and interactive MRI simulator * H. Benoit-Cattin 1, G. Collewet 2, B. Belaroussi 1, H. Saint-Jalmes 3, C. Odet 1 1 CREATIS, UMR CNRS

More information

Partial k-space Reconstruction

Partial k-space Reconstruction Chapter 2 Partial k-space Reconstruction 2.1 Motivation for Partial k- Space Reconstruction a) Magnitude b) Phase In theory, most MRI images depict the spin density as a function of position, and hence

More information

Fast Imaging Trajectories: Non-Cartesian Sampling (1)

Fast Imaging Trajectories: Non-Cartesian Sampling (1) Fast Imaging Trajectories: Non-Cartesian Sampling (1) M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.05.03 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business

More information

Partial k-space Recconstruction

Partial k-space Recconstruction Partial k-space Recconstruction John Pauly September 29, 2005 1 Motivation for Partial k-space Reconstruction a) Magnitude b) Phase In theory, most MRI images depict the spin density as a function of position,

More information

Section 1: Introduction to Geometry Points, Lines, and Planes

Section 1: Introduction to Geometry Points, Lines, and Planes Section 1: Introduction to Geometry Points, Lines, and Planes Topic 1: Basics of Geometry - Part 1... 3 Topic 2: Basics of Geometry Part 2... 5 Topic 3: Midpoint and Distance in the Coordinate Plane Part

More information

Assignment 2. Due Feb 3, 2012

Assignment 2. Due Feb 3, 2012 EE225E/BIOE265 Spring 2012 Principles of MRI Miki Lustig Assignment 2 Due Feb 3, 2012 1. Read Nishimura Ch. 3 2. Non-Uniform Sampling. A student has an assignment to monitor the level of Hetch-Hetchi reservoir

More information

Phase Difference Reconstruction. Outline

Phase Difference Reconstruction. Outline Advanced MRI Phase Difference Reconstruction Faik Can MERAL Outline Introduction Quantitative Description Arctangent operation ATAN2 Phased-Array Multiple Coil Data Correction of Predictable Phase Errors

More information

Instruction manual for T3DS calculator software. Analyzer for terahertz spectra and imaging data. Release 2.4

Instruction manual for T3DS calculator software. Analyzer for terahertz spectra and imaging data. Release 2.4 Instruction manual for T3DS calculator software Release 2.4 T3DS calculator v2.4 16/02/2018 www.batop.de1 Table of contents 0. Preliminary remarks...3 1. Analyzing material properties...4 1.1 Loading data...4

More information

Use of MRI in Radiotherapy: Technical Consideration

Use of MRI in Radiotherapy: Technical Consideration Use of MRI in Radiotherapy: Technical Consideration Yanle Hu, PhD Department of Radiation Oncology, Mayo Clinic Arizona 04/07/2018 2015 MFMER slide-1 Conflict of Interest: None 2015 MFMER slide-2 Objectives

More information

A Transmission Line Matrix Model for Shielding Effects in Stents

A Transmission Line Matrix Model for Shielding Effects in Stents A Transmission Line Matrix Model for hielding Effects in tents Razvan Ciocan (1), Nathan Ida (2) (1) Clemson University Physics and Astronomy Department Clemson University, C 29634-0978 ciocan@clemon.edu

More information

Visit MathNation.com or search "Math Nation" in your phone or tablet's app store to watch the videos that go along with this workbook!

Visit MathNation.com or search Math Nation in your phone or tablet's app store to watch the videos that go along with this workbook! Topic 1: Introduction to Angles - Part 1... 47 Topic 2: Introduction to Angles Part 2... 50 Topic 3: Angle Pairs Part 1... 53 Topic 4: Angle Pairs Part 2... 56 Topic 5: Special Types of Angle Pairs Formed

More information

Eureka Math. Grade 7, Module 6. Student File_A. Contains copy-ready classwork and homework

Eureka Math. Grade 7, Module 6. Student File_A. Contains copy-ready classwork and homework A Story of Ratios Eureka Math Grade 7, Module 6 Student File_A Contains copy-ready classwork and homework Published by the non-profit Great Minds. Copyright 2015 Great Minds. No part of this work may be

More information

PHASE-ENCODED, RAPID, MULTIPLE-ECHO (PERME) NUCLEAR MAGNETIC RESONANCE IMAGING

PHASE-ENCODED, RAPID, MULTIPLE-ECHO (PERME) NUCLEAR MAGNETIC RESONANCE IMAGING PHASE-ENCODED, RAPID, MULTIPLE-ECHO (PERME) NUCLEAR MAGNETIC RESONANCE IMAGING Mark Steven Lawton Master of Engineering Thesis Lawrence Berkeley Laboratory University of California Berkeley, California

More information

How to cite this article: emagres,2016,vol5: DOI / emrstm1454

How to cite this article: emagres,2016,vol5: DOI / emrstm1454 CSI and SENSE CSI Michael Schär 1,BernhardStrasser 2 & Ulrike Dydak 3,4 1 Johns Hopkins University, Baltimore, MD, USA 2 Medical University of Vienna, Vienna, Austria 3 Purdue University, West Lafayette,

More information

Image Sampling and Quantisation

Image Sampling and Quantisation Image Sampling and Quantisation Introduction to Signal and Image Processing Prof. Dr. Philippe Cattin MIAC, University of Basel 1 of 46 22.02.2016 09:17 Contents Contents 1 Motivation 2 Sampling Introduction

More information

Image Sampling & Quantisation

Image Sampling & Quantisation Image Sampling & Quantisation Biomedical Image Analysis Prof. Dr. Philippe Cattin MIAC, University of Basel Contents 1 Motivation 2 Sampling Introduction and Motivation Sampling Example Quantisation Example

More information

Excerpt from the Proceedings of the COMSOL Conference 2010 Paris

Excerpt from the Proceedings of the COMSOL Conference 2010 Paris Excerpt from the Proceedings of the COMSOL Conference 2010 Paris Simulation of Flaw Signals in a Magnetic Flux Leakage Inspection Procedure O. Nemitz * and T. Schmitte Salzgitter Mannesmann Forschung GmbH

More information

Fast methods for magnetic resonance angiography (MRA)

Fast methods for magnetic resonance angiography (MRA) Fast methods for magnetic resonance angiography (MRA) Bahareh Vafadar Department of Electrical and Computer Engineering A thesis presented for the degree of Doctor of Philosophy University of Canterbury,

More information

ksa 400 Growth Rate Analysis Routines

ksa 400 Growth Rate Analysis Routines k-space Associates, Inc., 2182 Bishop Circle East, Dexter, MI 48130 USA ksa 400 Growth Rate Analysis Routines Table of Contents ksa 400 Growth Rate Analysis Routines... 2 1. Introduction... 2 1.1. Scan

More information

K-Space Trajectories and Spiral Scan

K-Space Trajectories and Spiral Scan K-Space and Spiral Scan Presented by: Novena Rangwala nrangw2@uic.edu 1 Outline K-space Gridding Reconstruction Features of Spiral Sampling Pulse Sequences Mathematical Basis of Spiral Scanning Variations

More information

Constrained Reconstruction of Sparse Cardiac MR DTI Data

Constrained Reconstruction of Sparse Cardiac MR DTI Data Constrained Reconstruction of Sparse Cardiac MR DTI Data Ganesh Adluru 1,3, Edward Hsu, and Edward V.R. DiBella,3 1 Electrical and Computer Engineering department, 50 S. Central Campus Dr., MEB, University

More information

Similar Polygons Date: Per:

Similar Polygons Date: Per: Math 2 Unit 6 Worksheet 1 Name: Similar Polygons Date: Per: [1-2] List the pairs of congruent angles and the extended proportion that relates the corresponding sides for the similar polygons. 1. AA BB

More information

A more accurate account of the effect of k-space sampling and signal decay on the effective spatial resolution in functional MRI

A more accurate account of the effect of k-space sampling and signal decay on the effective spatial resolution in functional MRI A more accurate account of the effect of k-space sampling and signal decay on the effective spatial resolution in functional MRI Denis Chaimow 1 and Amir Shmuel 1,2 1 Centre for Magnetic Resonance Research

More information

Lesson 10. Homework Problem Set Sample Solutions. then Print True else Print False End if. False False True False False False

Lesson 10. Homework Problem Set Sample Solutions. then Print True else Print False End if. False False True False False False Homework Problem Set Sample Solutions 1. Perform the instructions in the following programming code as if you were a computer and your paper were the computer screen. Declare xx integer For all xx from

More information

FMRI Pre-Processing and Model- Based Statistics

FMRI Pre-Processing and Model- Based Statistics FMRI Pre-Processing and Model- Based Statistics Brief intro to FMRI experiments and analysis FMRI pre-stats image processing Simple Single-Subject Statistics Multi-Level FMRI Analysis Advanced FMRI Analysis

More information

Albertson AP Calculus AB AP CALCULUS AB SUMMER PACKET DUE DATE: The beginning of class on the last class day of the first week of school.

Albertson AP Calculus AB AP CALCULUS AB SUMMER PACKET DUE DATE: The beginning of class on the last class day of the first week of school. Albertson AP Calculus AB Name AP CALCULUS AB SUMMER PACKET 2017 DUE DATE: The beginning of class on the last class day of the first week of school. This assignment is to be done at you leisure during the

More information

Sampling, Ordering, Interleaving

Sampling, Ordering, Interleaving Sampling, Ordering, Interleaving Sampling patterns and PSFs View ordering Modulation due to transients Temporal modulations Timing: cine, gating, triggering Slice interleaving Sequential, Odd/even, bit-reversed

More information

TOPICS 2/5/2006 8:17 PM. 2D Acquisition 3D Acquisition

TOPICS 2/5/2006 8:17 PM. 2D Acquisition 3D Acquisition TOPICS 2/5/2006 8:17 PM 2D Acquisition 3D Acquisition 2D Acquisition Involves two main steps : Slice Selection Slice selection is accomplished by spatially saturating (single or multi slice imaging) or

More information

Supplementary Information

Supplementary Information Supplementary Information Magnetic resonance imaging reveals functional anatomy and biomechanics of a living dragon tree Linnea Hesse 1,2,*, Tom Masselter 1,2,3, Jochen Leupold 4, Nils Spengler 5, Thomas

More information

Lesson 24: Matrix Notation Encompasses New Transformations!

Lesson 24: Matrix Notation Encompasses New Transformations! Classwork Example 1 Determine the following: a. 1 0 0 1 3 b. 1 0 7 0 1 1 c. 1 0 3 5 0 1 1 d. 1 0 3 1 0 1 7 6 e. 9 1 0 1 3 1 0 1 f. 1 0 cc aa 0 1 bb dd xx yy 0 g. 1 zz ww 0 1 Date: 1/5/15 S.14 Example Can

More information

Computational Medical Imaging Analysis

Computational Medical Imaging Analysis Computational Medical Imaging Analysis Chapter 2: Image Acquisition Systems Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

Fmri Spatial Processing

Fmri Spatial Processing Educational Course: Fmri Spatial Processing Ray Razlighi Jun. 8, 2014 Spatial Processing Spatial Re-alignment Geometric distortion correction Spatial Normalization Smoothing Why, When, How, Which Why is

More information

Section 6: Triangles Part 1

Section 6: Triangles Part 1 Section 6: Triangles Part 1 Topic 1: Introduction to Triangles Part 1... 125 Topic 2: Introduction to Triangles Part 2... 127 Topic 3: rea and Perimeter in the Coordinate Plane Part 1... 130 Topic 4: rea

More information

PHASE-SENSITIVE AND DUAL-ANGLE RADIOFREQUENCY MAPPING IN. Steven P. Allen. A senior thesis submitted to the faculty of. Brigham Young University

PHASE-SENSITIVE AND DUAL-ANGLE RADIOFREQUENCY MAPPING IN. Steven P. Allen. A senior thesis submitted to the faculty of. Brigham Young University PHASE-SENSITIVE AND DUAL-ANGLE RADIOFREQUENCY MAPPING IN 23 NA MAGNETIC RESONANCE IMAGING by Steven P. Allen A senior thesis submitted to the faculty of Brigham Young University in partial fulfillment

More information

1.2 Round-off Errors and Computer Arithmetic

1.2 Round-off Errors and Computer Arithmetic 1.2 Round-off Errors and Computer Arithmetic 1 In a computer model, a memory storage unit word is used to store a number. A word has only a finite number of bits. These facts imply: 1. Only a small set

More information

Chapter 2: Understanding Data Distributions with Tables and Graphs

Chapter 2: Understanding Data Distributions with Tables and Graphs Test Bank Chapter 2: Understanding Data with Tables and Graphs Multiple Choice 1. Which of the following would best depict nominal level data? a. pie chart b. line graph c. histogram d. polygon Ans: A

More information

G Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine. Compressed Sensing

G Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine. Compressed Sensing G16.4428 Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine Compressed Sensing Ricardo Otazo, PhD ricardo.otazo@nyumc.org Compressed

More information

Non-Cartesian Parallel Magnetic Resonance Imaging

Non-Cartesian Parallel Magnetic Resonance Imaging Non-Cartesian Parallel Magnetic Resonance Imaging Dissertation zur Erlangung des naturwissenschaftlichen Doktorgrades der Bayerischen Julius-Maximilians-Universität Würzburg vorgelegt von Robin Heidemann

More information

Unit 4 Graphs of Trigonometric Functions - Classwork

Unit 4 Graphs of Trigonometric Functions - Classwork Unit Graphs of Trigonometric Functions - Classwork For each of the angles below, calculate the values of sin x and cos x ( decimal places) on the chart and graph the points on the graph below. x 0 o 30

More information

Image Enhancement. Digital Image Processing, Pratt Chapter 10 (pages ) Part 1: pixel-based operations

Image Enhancement. Digital Image Processing, Pratt Chapter 10 (pages ) Part 1: pixel-based operations Image Enhancement Digital Image Processing, Pratt Chapter 10 (pages 243-261) Part 1: pixel-based operations Image Processing Algorithms Spatial domain Operations are performed in the image domain Image

More information

ELECTRONOTES APPLICATION NOTE NO. 372

ELECTRONOTES APPLICATION NOTE NO. 372 ELECTRONOTES APPLICATION NOTE NO. 372 116 Hanshaw Rd Ithaca, NY 1485 January 28 INTERPOLATION ERROR IN SINEWAVE TABLES INTRODUCTION: The basic notion of interpolation is fundamental to many things we do

More information

Texture Mapping. Michael Kazhdan ( /467) HB Ch. 14.8,14.9 FvDFH Ch. 16.3, , 16.6

Texture Mapping. Michael Kazhdan ( /467) HB Ch. 14.8,14.9 FvDFH Ch. 16.3, , 16.6 Texture Mapping Michael Kazhdan (61.457/467) HB Ch. 14.8,14.9 FvDFH Ch. 16.3, 16.4.5, 16.6 Textures We know how to go from this to this J. Birn Textures But what about this to this? J. Birn Textures How

More information

LAB 2: DATA FILTERING AND NOISE REDUCTION

LAB 2: DATA FILTERING AND NOISE REDUCTION NAME: LAB TIME: LAB 2: DATA FILTERING AND NOISE REDUCTION In this exercise, you will use Microsoft Excel to generate several synthetic data sets based on a simplified model of daily high temperatures in

More information

A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data

A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data Alexey Samsonov, Julia Velikina Departments of Radiology and Medical

More information

Name: Tutor s

Name: Tutor s Name: Tutor s Email: Bring a couple, just in case! Necessary Equipment: Black Pen Pencil Rubber Pencil Sharpener Scientific Calculator Ruler Protractor (Pair of) Compasses 018 AQA Exam Dates Paper 1 4

More information

MR Advance Techniques. Vascular Imaging. Class III

MR Advance Techniques. Vascular Imaging. Class III MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Eureka Math. Geometry, Module 4. Student File_B. Contains Exit Ticket, and Assessment Materials

Eureka Math. Geometry, Module 4. Student File_B. Contains Exit Ticket, and Assessment Materials A Story of Functions Eureka Math Geometry, Module 4 Student File_B Contains Exit Ticket, and Assessment Materials Published by the non-profit Great Minds. Copyright 2015 Great Minds. No part of this work

More information

Lesson 1. Unit 2 Practice Problems. Problem 2. Problem 1. Solution 1, 4, 5. Solution. Problem 3

Lesson 1. Unit 2 Practice Problems. Problem 2. Problem 1. Solution 1, 4, 5. Solution. Problem 3 Unit 2 Practice Problems Lesson 1 Problem 1 Rectangle measures 12 cm by 3 cm. Rectangle is a scaled copy of Rectangle. Select all of the measurement pairs that could be the dimensions of Rectangle. 1.

More information

Massachusetts Institute of Technology. Department of Computer Science and Electrical Engineering /6.866 Machine Vision Quiz I

Massachusetts Institute of Technology. Department of Computer Science and Electrical Engineering /6.866 Machine Vision Quiz I Massachusetts Institute of Technology Department of Computer Science and Electrical Engineering 6.801/6.866 Machine Vision Quiz I Handed out: 2004 Oct. 21st Due on: 2003 Oct. 28th Problem 1: Uniform reflecting

More information

arxiv: v2 [physics.med-ph] 22 Jul 2014

arxiv: v2 [physics.med-ph] 22 Jul 2014 Multichannel Compressive Sensing MRI Using Noiselet Encoding arxiv:1407.5536v2 [physics.med-ph] 22 Jul 2014 Kamlesh Pawar 1,2,3, Gary Egan 4, and Jingxin Zhang 1,5,* 1 Department of Electrical and Computer

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

Dynamic Image and Fieldmap Joint Estimation Methods for MRI Using Single-Shot Trajectories

Dynamic Image and Fieldmap Joint Estimation Methods for MRI Using Single-Shot Trajectories Dynamic Image and Fieldmap Joint Estimation Methods for MRI Using Single-Shot Trajectories by Antonis Matakos A dissertation submitted in partial fulfillment of the requirements for the degree of Doctor

More information

2.1 Signal Production. RF_Coil. Scanner. Phantom. Image. Image Production

2.1 Signal Production. RF_Coil. Scanner. Phantom. Image. Image Production An Extensible MRI Simulator for Post-Processing Evaluation Remi K.-S. Kwan?, Alan C. Evans, and G. Bruce Pike McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University, Montreal,

More information