Effect of Scattering on the Image. Reducing Compton Scatter with a Grid
|
|
- Lenard Jeffery Nichols
- 5 years ago
- Views:
Transcription
1 Effect of Scattering on the Image Increasing Compton scattering degrades image. Webb 21 Reducing Compton Scatter with a Grid Grids Parallel (focused at infinity) Linear Focused (see figure) Moving grids avoid grid lines. Grid increases exposure by grid conversion factor (GCF) (typically 3-8) Scattering worse for high kv photons and thick body parts. Linear Focused Grid 22 1
2 Air Gaps and Scanning Slits Air Gaps Increasing space between patient and the detector reduces scatter, because scattered photons diverge at a given angle. Must trade off with increased unsharpness as will be seen. Scanning Slits Mechanical slits that move together in front and in back of the patient. Only photons traveling a straight path make it through. Requires longer exposure, but not greater radiation to patient. 23 Film-Screen Detectors Intensifying screens consist of a phosphor and a mirror on both sides of the photographic film to convert x-rays into visible photons. Luminescence (photons generated by something besides heat, which is incandescence) 2 types of Luminescence involved here, both caused by incoming high energy photons (x-rays). Fluorescence - good (fast re-emission of lower energy photon) Phosphorescence - bad (much slower re-emission due to forbidden energy state transitions) Conversion Efficiency # light photons per x-ray photon (typically 10 3 ) 24 2
3 From Film To Detectors Film is basically almost completely gone UPMC is now filmless Detectors have great advantages Ability to adjust gain and offset afterwards Permits computational techniques such as Dual-Energy, which eliminates ribs from the chest X-ray by using two scans with different KeV. Linear combination (subtraction) to eliminate bones. Much easier to store, retrieve, and transmit the images. Dual Energy X-ray 25 X-ray Image Intensifier (Fluoroscopy) Fluoroscopy - like night vision goggles, but with extra input phosphor to convert x-rays to light photons, Light photons hit photocathode, generating electrons 26 Electrons are accelerated and focused at anode on output phosphor. 3
4 X-ray Image Intensifier (Fluoroscopy) Curved input phosphor/photocathode produces photoelectrons, which are focused by electrostatic lenses (dynodes) onto a fluorescent screen (output phosphor). Huge amplification ~200,000 light photons per x-ray photon. Magnification can be changed dynamically by voltage on dynodes. Webb 27 Image Formation - Linear Attenuation Recall Intensity = Energy x Flux As we have seen, attenuation of polyenergetic x-ray source (Bremsstrahlung) by a non-homogeneous structure yields where r(x,y) is the length of the path, S 0 (E) is spectrum of the incident x-rays, s is the distance from the x-ray source along the path and I(x,y) is the intensity of x-rays remaining. 28 4
5 Geometric Effects - Beam Divergence Diverging beam of x-rays yields inverse square law; r varies with (x,y). Uneven intensity between I O at detector origin and I r elsewhere on a flat detector, even before object attenuation. 29 Flat detector is not orthogonal to the beam except at the detector origin causing intensity on the detector I d to be lower than that at origin by Combines with beam divergence on previous slide to yield Geometric Effects - Obliquity 30 5
6 Anode Heel Effect stronger intensity beam in the cathode direction and no variation orthogonal to the cathode-anode direction Less attenuation from anode itself in this direction. Can be compensated for by a filter or used to compensate for variations in patient thickness 31 Path length through patient Consider homogeneous flat slab of thickness L Ignoring anode heel, and combining with obliquity and beam divergence: 32 6
7 Effect of shading artifact The preceding geometric effects create a gradual change in intensity with position that can be ignored by the radiologist, who can just concentrate locally. However this low frequency shading artifact can be undesirable for automated image analysis, and the models just presented may be used to cancel the effect, insofar as the models are accurate. 33 Depth-dependent object magnification Example of stick object object magnification M(z) 34 7
8 Imaging Equation with Geometric Effects When object is located at detector face, intensity at detector is When Things closer to x-ray origin appear larger 35 Blur due to extended source (geometric unsharpness) Non-zero spot size of the x-ray emission source, needed to prevent it from melting. 36 8
9 Blur due to extended source (geometric unsharpness) The physical extent of blurring caused by an extend source depends on the size of the source and the location of the object. Larger source, or greater distance from patient to detector, means more blur. penumbra (Latin for partial shadow) Webb 37 Principle of Source Magnification For a given z, convolution of the source shape with the object shape (with magnification effects also included). source magnification m(z) (negative because image is inverted) as opposed to object magnification M(z) disk function Ignoring geometric effects, including 1/r 2, so that all rays are passing through hole where k is constant for a given z 38 9
10 Solving for k as a function of z (this is more detail than you really need to know ) Recall that geometric effects are being ignored (including 1/r 2 ) integration yields DC component To find k take the Fourier transform, using the scale law As object approaches detector, integrated intensity independent of z dilution from source magnification so that no blurring occurs 39 Superposition of pin holes in one z plane Convolving the magnified (by M) object s transmittivity by the magnified (by m) and scaled (by geometric factors) source function. geometric factors source magnification object magnification see k in previous slide For objects close to the detector, convolution in 3D over entire object the object will have unity magnification, and will not be blurred by the source focal spot regardless of the focal spot size (convolution with impulse function)
11 PA / AP chest X-ray PA (posterior-anterior) X-rays hit the back first. AP (anterior-posterior) heart is blurred and magnified. from Lucy Squire 41 Mammography Very low X-ray energy Lower radiation dose Better soft tissue (PE effect) Less Compton scattering Looking for micro-calcifications. Very small focal spot. Minimize geometric unsharpness by Large focal-spot-to-film distance Small film-to-patient distance May use magnifying glass to see micro-calcifications. sprojects.mmi.mcgill.ca 42 11
12 Film-Screen Blurring Isotropic shower of light photons form spot on the film. Film-screen impulse response h(x,y) to x-ray photon impulse. Cascades with previous response derived for a single plane transmittivity. 43 Film-Screen Blurring typical MTF Thin phosphor, less blurring so higher resolution, but lower efficiency (% of x-ray photons stopped) thus higher noise. Thus a thicker phosphor may actually be better
13 Digital detectors have replaced film Large, high resolution digital cameras (like film) Still need phosphor to convert x-ray to visible light Permits dynamic adjustment of contrast after the fact Enables digital storage and distribution Radiographs were last hold-out in medical imaging Kodak has filed for bankruptcy Enables use of Pulsed X-Ray tubes (specialized scan) Very short pulses (12 nanosecond) High intensity but short duration, so reasonable dose. Permits movies like fluoroscopy, but much higher shutter speed Need digital detector to store sequence of snapshots Used in dynamic studies of motion in spine, joints, etc
14 Chapter 6 Computed Tomography Originally Computed Axial Tomography ( CAT scan) 1 Before computed tomography Motion tomogram produced by moving source and film in opposite directions to focus on a particular plane different planes temporomandibular joint exostosis in auditory canal CT has almost completely replaced this. Though CT has poorer in-plane resolution, it eliminated overlapping structures 2 1
15 Computed Tomography Invented mid-1960s Through 1970s and 1980s profound change in diagnostic imaging. Reduced exploratory surgery. CT numbers in Hounsfield units for quantitative analysis. Same fundamental physics. We will discuss geometries of generations of CT scanners. Mathematics of reconstruction Same can be used in SPECT, PET, MRI. gantry 3 Subdural Hematoma Note the midline shift and distorted ventricles
16 Lung Tumor contrast (iodine) in vasculature CT always seen from the patients feet, tumor in left lung. 5 Liver Tumors Liver is on patient s right. 6 3
17 Kidney and Intestines Oral and intravenous contrast were administered prior to the image. 7 Hepatic veins with contrast Reconstructed in the coronal plane. 8 4
18 First Generation (1G) CT scanner Single source and detector move in unison along a linear path. Then whole apparatus rotated around patient in the gantry. Arbitrary number of rays (paths). Arbitrary number of angular projections. Scattering goes mostly undetected and no collimators required No longer manufactured. 9 Second Generation (2G) CT scanner Additional detectors in an array. Still move linearly in unison, but can make larger rotation, reducing time (because multiple angles captured with each scan). Now scattering a problem and collimators required. This reduces efficiency and increases noise for a given dose. Payoff in reducing time is worth it. 10 5
19 Third Generation (3G) CT scanner More detectors in array, so fan beam covers entire patient. Linear motion no longer needed. Much faster. Collimators required to reduce scattering. Detectors smaller so less efficient. So for same SNR, higher dose required, but savings in time still win. 11 Fourth Generation (4G) CT scanner Complete ring of stationary detectors. Only the source rotates. Large detectors, since they lie on a large ring, so high detection efficiency. But collimators (between detectors) cannot be used, so noise is a problem. (Collimators still used axially to restrict beam to detectors overall) Overall image quality comparable to 3G 12 6
20 Fifth Generation (5G) CT scanner Electron Beam CT (EBCT) ultra-fast or cine CT Only ~150 in the world; 1 in Oakland Electronically steers beam across ring of stationary detectors No moving parts, 50ms per image Can image beating heart without gating to electrocardiogram US patent Sixth Generation (6G) CT scanner Helical (spiral) CT Rapid volumetric data acquisition (full torso in 30 seconds) Table moves smoothly as source (and in some scanners detectors) rotate continuously using slip rings. Fast enough for single breath-hold, reducing motion artifact. Becoming quite common. 14 7
21 Seventh Generation (7G) CT scanner Multislice CT Multiple rows of detectors, can detect simultaneously. Faster, higher resolution, many tricks can be played with combining detectors dynamically trading off resolution against signal-to-noise. Beautiful high-resolution cardiac studies gated to electrocardiogram. 15 Seventh Generation (7G) CT scanner Multislice CT
22 Cone-Beam CT Recent addition, a portable machine that spins around patient. Not as high quality an image, but used in the OR to guide intervention. Reconstructs 3D data sets from multiple projections on a flat sensor array X-ray source Most CT scanners have a single x-ray source (except EBCT) Expensive (need to be replaced frequently) Need to be constantly calibrated CT requires x-ray source to be approximated as monoenergetic. More filtering employed than with projection radiography Hardening the beam (removing lower energy x-rays) before it hits the patient (using aluminum and copper filter combinations). 18 9
23 CT Detectors All use photo-electric effect to produce photoelectrons, detected electronically. (a) Solid-state detectors, thicker crystals for higher efficiency. (b) Xenon gas tubes, less efficient, but highly directional (3G). (c) Multiple solid-state detector array, most common today. 19 Image Formation - Line Integrals Mathematically intractable for CT image reconstruction So we assume a single effective energy The basic measurement of a CT scanner is a line integral of the linear attenuation coefficient at the effective energy of the scanner (notice that I 0 is calibrated away)
24 Calibration and CT numbers Reference Intensity I 0 must be measured for each detector. However, Effective Energy still varies with each x-ray tube. To compensate, we normalize to The CT number in Hounsfield units (HU) is defined as Air HU Water 0 HU Bone ~1000 HU Metal > 3000 HU Contrast > 3000 HU because μ Air 0 Godfrey Hounsfield Reproducible across scanners +/- 2 HU 21 Parallel-Ray Reconstruction Project along a line Parametric form of projection Alternate form of projection using sifting 22 11
25 Radon Transform To characterize CT measurement, Johann Radon 1917 Attenuation coefficient of underlying unknown object. CT measurements along a given line at a given angle. 23 Example - Radon transform of unit disk due to symmetry, only need consider 24 12
MEDICAL IMAGING 2nd Part Computed Tomography
MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional
More informationIntroduction to Biomedical Imaging
Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray
More informationImage Acquisition Systems
Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS
More informationMEDICAL IMAGING 2nd Part Computed Tomography
MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional
More informationBME I5000: Biomedical Imaging
1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.
More informationTomographic Reconstruction
Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)
More informationCorso di laurea in Fisica A.A Fisica Medica 4 TC
Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure
More informationCh. 4 Physical Principles of CT
Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between
More informationRADIOLOGY AND DIAGNOSTIC IMAGING
Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality
More informationBiomedical Imaging. Computed Tomography. Patrícia Figueiredo IST
Biomedical Imaging Computed Tomography Patrícia Figueiredo IST 2013-2014 Overview Basic principles X ray attenuation projection Slice selection and line projections Projection reconstruction Instrumentation
More informationSome reference material
Some reference material Physics reference book on medical imaging: A good one is The Essential Physics of Medical Imaging, 3 rd Ed. by Bushberg et al. ($170! new). However, there are several similar books
More informationShadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS
Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic
More informationMedical Imaging BMEN Spring 2016
Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,
More informationComputed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA
Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction
More informationMEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah
MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography
More informationProjection Radiography
Medical Physics Projection Radiography Ho Kyung Kim hokyung@pusan.ac.kr Pusan National University Projection radiography Most commonly used method of medical imaging utilizing x rays, called conventional
More informationComputer-Tomography II: Image reconstruction and applications
Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de
More informationA closer look at CT scanning
Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic
More informationRadiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada
Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging
More informationJoint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013
2455-5 Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage 29 April - 3 May, 2013 Lecture NoteBasic principles of X-ray Computed Tomography Diego Dreossi Elettra, Trieste
More informationCardiac Dual Energy CT: Technique
RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de
More informationPhysical bases of X-ray diagnostics
Physical bases of X-ray diagnostics Dr. István Voszka Possibilities of X-ray production (X-ray is produced, when charged particles of high velocity are stopped) X-ray tube: Relatively low accelerating
More informationDigital Image Processing
Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture
More informationMidterm Review. Yao Wang Polytechnic University, Brooklyn, NY 11201
Midterm Review Yao Wang Polytechnic University, Brooklyn, NY 11201 Based on J. L. Prince and J. M. Links, Medical maging Signals and Systems, and lecture notes by Prince. Figures are from the textbook.
More informationSpiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA
Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization
More informationCombining Analytical and Monte Carlo Modelling for Industrial Radiology
19 th World Conference on Non-Destructive Testing 2016 Combining Analytical and Monte Carlo Modelling for Industrial Radiology Carsten BELLON, Gerd-Rüdiger JAENISCH, Andreas DERESCH BAM Bundesanstalt für
More informationProton dose calculation algorithms and configuration data
Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam
More informationComputed tomography - outline
Computed tomography - outline Computed Tomography Systems Jørgen Arendt Jensen and Mikael Jensen (DTU Nutech) October 6, 216 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering
More informationFundamentals of CT imaging
SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform
More informationMoscow-Bavarian Joint Advanced Student School 2006 / Medical Imaging Principles of Computerized Tomographic Imaging and Cone-Beam Reconstruction
Line Integrals Line integrals represent the integral of some parameter of the object along the line (e.g. attenuation of x-rays) Object: f(x,y) Line: x cosθ + y sinθ = t Line integral / Radon transform:
More information3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH
3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk
More informationIndex. aliasing artifacts and noise in CT images, 200 measurement of projection data, nondiffracting
Index Algebraic equations solution by Kaczmarz method, 278 Algebraic reconstruction techniques, 283-84 sequential, 289, 293 simultaneous, 285-92 Algebraic techniques reconstruction algorithms, 275-96 Algorithms
More informationComputed Tomography. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Sep 26th/Oct 3rd, 2016
Computed Tomography Principles of Medical Imaging Prof. Dr. Philippe Cattin MIAC, University of Basel Contents Abstract 1 Computed Tomography Basics Introduction Computed Tomography Hounsfield's CT Prototype
More informationMulti-slice CT Image Reconstruction Jiang Hsieh, Ph.D.
Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced
More informationAP Physics: Curved Mirrors and Lenses
The Ray Model of Light Light often travels in straight lines. We represent light using rays, which are straight lines emanating from an object. This is an idealization, but is very useful for geometric
More informationProjection Radiography
Medical Physics Prince & Links 5 Projection Radiography Ho Kyung Kim hokyung@pusan.ac.kr Pusan National University Projection radiography Most commonly used method of medical imaging utilizing x rays,
More informationDUAL energy X-ray radiography [1] can be used to separate
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy
More informationMedical Images Analysis and Processing
Medical Images Analysis and Processing - 25642 Emad Course Introduction Course Information: Type: Graduated Credits: 3 Prerequisites: Digital Image Processing Course Introduction Reference(s): Insight
More informationDiagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology
Diagnostic imaging techniques Krasznai Zoltán University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology 1. Computer tomography (CT) 2. Gamma camera 3. Single Photon
More informationComputational Medical Imaging Analysis
Computational Medical Imaging Analysis Chapter 2: Image Acquisition Systems Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky
More information8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition
Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare
More informationMedical Image Processing: Image Reconstruction and 3D Renderings
Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,
More informationMicro-CT Methodology Hasan Alsaid, PhD
Micro-CT Methodology Hasan Alsaid, PhD Preclinical & Translational Imaging LAS, PTS, GlaxoSmithKline 20 April 2015 Provide basic understanding of technical aspects of the micro-ct Statement: All procedures
More informationCT Basics Principles of Spiral CT Dose. Always Thinking Ahead.
1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central
More informationThe branch of physics which studies light
Mr.V The branch of physics which studies light Geometric model XVI century by W Snell Wave Model XIX century by JC Maxwell Photon Model XX century by Planck, Einstein Models of Light Basic Concept Laws
More informationCT Systems and their standards
CT Systems and their standards Stephen Brown Engineering Measurement 11 th April 2012 Industrial X-ray computed tomography: The future of co-ordinate metrology? Burleigh Court, Loughborough University
More informationCoE4TN4 Image Processing. Chapter 5 Image Restoration and Reconstruction
CoE4TN4 Image Processing Chapter 5 Image Restoration and Reconstruction Image Restoration Similar to image enhancement, the ultimate goal of restoration techniques is to improve an image Restoration: a
More informationHIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS
23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,
More informationAutomated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4
More informationCLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0
Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed
More informationCT: Physics Principles & Equipment Design
CT: Physics Principles & Equipment Design James Kofler, Ph.D Radiology Mayo Clinic Rochester, MN June 27, 2012 Disclosures Nothing to disclose Learning Objectives Understand fundamental concepts of - CT
More informationC a t p h a n / T h e P h a n t o m L a b o r a t o r y
C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice
More informationFinancial disclosure. Onboard imaging modality for IGRT
Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty
More informationDEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE
Rajesh et al. : Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh V Acharya, Umesh Kumar, Gursharan
More informationCentral Slice Theorem
Central Slice Theorem Incident X-rays y f(x,y) R x r x Detected p(, x ) The thick line is described by xcos +ysin =R Properties of Fourier Transform F [ f ( x a)] F [ f ( x)] e j 2 a Spatial Domain Spatial
More informationBasic Radiation Oncology Physics
Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides
More informationFits you like no other
Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room
More informationComputer-Tomography I: Principles, History, Technology
Computer-Tomography I: Principles, History, Technology Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de History
More informationSimulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images
Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),
More informationEmission Computed Tomography Notes
Noll (24) ECT Notes: Page 1 Emission Computed Tomography Notes Introduction Emission computed tomography (ECT) is the CT applied to nuclear medicine. There are two varieties of ECT: 1. SPECT single-photon
More informationAN X-ray system with a large area detector, such as is used
IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 25, NO. 12, DECEMBER 2006 1573 Scatter Correction Method for X-Ray CT Using Primary Modulation: Theory and Preliminary Results Lei Zhu*, Member, IEEE, N. Robert
More informationImage Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD
Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image
More informationChapter 26 Geometrical Optics
Chapter 26 Geometrical Optics 26.1 The Reflection of Light 26.2 Forming Images With a Plane Mirror 26.3 Spherical Mirrors 26.4 Ray Tracing and the Mirror Equation 26.5 The Refraction of Light 26.6 Ray
More informationA prototype table-top inverse-geometry volumetric CT system
A prototype table-top inverse-geometry volumetric CT system Taly Gilat Schmidt a Department of Radiology, Stanford University, Stanford, California 94305 Josh Star-Lack NexRay, Inc., Los Gatos, California
More informationPrinciples of Computerized Tomographic Imaging
Principles of Computerized Tomographic Imaging Parallel CT, Fanbeam CT, Helical CT and Multislice CT Marjolein van der Glas August 29, 2000 Abstract The total attenuation suffered by one beam of x-rays
More informationREMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO
REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO Abstract C.S. Levin, Y-C Tai, E.J. Hoffman, M. Dahlbom, T.H. Farquhar UCLA School of Medicine Division
More informationCOMPARATIVE STUDIES OF DIFFERENT SYSTEM MODELS FOR ITERATIVE CT IMAGE RECONSTRUCTION
COMPARATIVE STUDIES OF DIFFERENT SYSTEM MODELS FOR ITERATIVE CT IMAGE RECONSTRUCTION BY CHUANG MIAO A Thesis Submitted to the Graduate Faculty of WAKE FOREST UNIVERSITY GRADUATE SCHOOL OF ARTS AND SCIENCES
More informationFits you like no other
Fits you like no other Philips BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase
More informationValidation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition
Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.
More informationA PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY
A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY BACKSCATTER DATA Young S. Ham and C.F. Poranski Materials Chemistry Branch Naval Research Laboratory Washington DC 20375 E.C. Greenawald Geo-Centers,
More informationScatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University
Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid by Yingxuan Chen Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang Fang Yin, Chair
More informationComputed tomography (Item No.: P )
Computed tomography (Item No.: P2550100) Curricular Relevance Area of Expertise: Biology Education Level: University Topic: Modern Imaging Methods Subtopic: X-ray Imaging Experiment: Computed tomography
More informationRange Sensors (time of flight) (1)
Range Sensors (time of flight) (1) Large range distance measurement -> called range sensors Range information: key element for localization and environment modeling Ultrasonic sensors, infra-red sensors
More informationNon-axially-symmetric Lens with extended depth of focus for Machine Vision applications
Non-axially-symmetric Lens with extended depth of focus for Machine Vision applications Category: Sensors & Measuring Techniques Reference: TDI0040 Broker Company Name: D Appolonia Broker Name: Tanya Scalia
More informationLecture 5. Prepared by:- KAMARUL AMIN BIN ABU BAKAR
Lecture 5 Prepared by:- KAMARUL AMIN BIN ABDULLAH @ ABU BAKAR LESSON OBJECTIVES At the end of the session, the students should be able to: Briefly explain the purpose, construction and principles of fluoroscopy.
More informationPhoton beam dose distributions in 2D
Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation
More informationWelcome to: Physics I. I m Dr Alex Pettitt, and I ll be your guide!
Welcome to: Physics I I m Dr Alex Pettitt, and I ll be your guide! Physics I: x Mirrors and lenses Lecture 13: 6-11-2018 Last lecture: Reflection & Refraction Reflection: Light ray hits surface Ray moves
More informationINTRODUCTION REFLECTION AND REFRACTION AT BOUNDARIES. Introduction. Reflection and refraction at boundaries. Reflection at a single surface
Chapter 8 GEOMETRICAL OPTICS Introduction Reflection and refraction at boundaries. Reflection at a single surface Refraction at a single boundary Dispersion Summary INTRODUCTION It has been shown that
More informationThe Near Future in Cardiac CT Image Reconstruction
SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de
More informationFirst CT Scanner. How it Works. Contemporary CT. Before and After CT. Computer Tomography: How It Works. Medical Imaging and Pattern Recognition
Computer Tomography: How t Works Medical maging and Pattern Recognition Lecture 7 Computed Tomography Oleh Tretiak Only one plane is illuminated. Source-subject motion provides added information. 2 How
More informationBasics of treatment planning II
Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!
More informationTomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS
Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate
More informationML reconstruction for CT
ML reconstruction for CT derivation of MLTR rigid motion correction resolution modeling polychromatic ML model dual energy ML model Bruno De Man, Katrien Van Slambrouck, Maarten Depypere, Frederik Maes,
More informationIntroduction to Positron Emission Tomography
Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator
More informationmicroct Systems for Life Science Innovation with Integrity High Resolution Microtomography for in vivo and ex vivo Applications Microtomography
microct Systems for Life Science High Resolution Microtomography for in vivo and ex vivo Applications Innovation with Integrity Microtomography SkyScan 1278 High Throughput Large Field of View in vivo
More information10/5/09 1. d = 2. Range Sensors (time of flight) (2) Ultrasonic Sensor (time of flight, sound) (1) Ultrasonic Sensor (time of flight, sound) (2) 4.1.
Range Sensors (time of flight) (1) Range Sensors (time of flight) (2) arge range distance measurement -> called range sensors Range information: key element for localization and environment modeling Ultrasonic
More informationBrilliance CT Big Bore.
1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.
More informationPhoton counting spectral CT versus conventional CT: comparative evaluation for breast imaging application
Physics in Medicine & Biology Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application To cite this article: Polad M Shikhaliev and Shannon G Fritz 2011
More informationPhilips SPECT/CT Systems
Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET
More informationDevelopments in Dimensional Metrology in X-ray Computed Tomography at NPL
Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Wenjuan Sun and Stephen Brown 10 th May 2016 1 Possible factors influencing XCT measurements Components Influencing variables Possible
More informationPhysical Optics. You can observe a lot just by watching. Yogi Berra ( )
Physical Optics You can observe a lot just by watching. Yogi Berra (1925-2015) OBJECTIVES To observe some interference and diffraction phenomena with visible light. THEORY In a previous experiment you
More informationApplication of 450 kv Computed Tomography to Engine Blocks with Steel Liners
Application of 450 kv Computed Tomography to Engine Blocks with Steel Liners Charles R. Smith, Kevin Holt BIR, Inc. Uwe Bischoff, Bernd Georgi, Ferdinand Hansen, Frank Jeltsch Volkswagen Commercial Vehicles
More informationNew Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC
New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it
More informationPhys 1020, Day 18: Questions? Cameras, Blmfld Reminders: Next Up: digital cameras finish Optics Note Final Project proposals next week!
Lights. Action. Phys 1020, Day 18: Questions? Cameras, Blmfld 15.1 Reminders: Next Up: digital cameras finish Optics Note Final Project proposals next week! 1 What have we learned in this section: 1) Lasers
More informationMEDICAL IMAGE ANALYSIS
SECOND EDITION MEDICAL IMAGE ANALYSIS ATAM P. DHAWAN g, A B IEEE Engineering in Medicine and Biology Society, Sponsor IEEE Press Series in Biomedical Engineering Metin Akay, Series Editor +IEEE IEEE PRESS
More informationScatter Correction Methods in Dimensional CT
Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University
More informationComputational Medical Imaging Analysis
Computational Medical Imaging Analysis Chapter 1: Introduction to Imaging Science Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky
More informationSystem Optimization and Patient Translational Motion Correction for Reduction of Artifacts in a Fan-Beam CT Scanner
Wright State University CORE Scholar Browse all Theses and Dissertations Theses and Dissertations 2012 System Optimization and Patient Translational Motion Correction for Reduction of Artifacts in a Fan-Beam
More informationFigure 1: Derivation of Bragg s Law
What is Bragg s Law and why is it Important? Bragg s law refers to a simple equation derived by English physicists Sir W. H. Bragg and his son Sir W. L. Bragg in 1913. This equation explains why the faces
More informationCT vs. VolumeScope: image quality and dose comparison
CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,
More information