Perturbation Method for Magnetic Field Calculations of Nonconductive Objects

Size: px
Start display at page:

Download "Perturbation Method for Magnetic Field Calculations of Nonconductive Objects"

Transcription

1 Perturbation Method for Magnetic Field Calculations of Nonconductive Objects Mark Jenkinson,* James L. Wilson, and Peter Jezzard Magnetic Resonance in Medicine 52: (2004) Inhomogeneous magnetic fields produce artifacts in MR images including signal dropout and spatial distortion. A novel perturbative method for calculating the magnetic field to first order (error is second order) within and around nonconducting objects is presented. The perturbation parameter is the susceptibility difference between the object and its surroundings (for example, 10 ppm in the case of brain tissue and air). This method is advantageous as it is sufficiently accurate for most purposes, can be implemented as a simple convolution with a voxel-based object model, and is linear. Furthermore, the method is simple to use and can quickly calculate the field for any orientation of an object using a set of precalculated basis images. Magn Reson Med 52: , Wiley- Liss, Inc. Key words: susceptibility; distortion; simulation; field calculation Theoretical calculation of the magnetic B field, given a distribution of tissue, allows modeling of various phenomena, such as MRI signal dropout, geometric distortion, interaction of B field and motion effects, manipulation of the B field using active and weakly magnetic passive shims, and respiration effects. Existing methods for calculating the B field use full finite element calculations (1,2) or approximate solutions to Maxwell s equations given either surface models of matter interfaces (3 5), voxelbased elements (6), spherical elements (7,8), or Fourier representations (9). By using a perturbation approach to solving Maxwell s equations (10), a linear first-order solution can be found which is fast and appropriate for most MR imaging applications. In addition, the perturbation method allows the magnitude of the errors to be calculated, and hence the accuracy and appropriateness of the method to be estimated for various applications. THEORY Assuming the object is nonconductive (so J 0), the relevant Maxwell s equations (ƒ H 0 and ƒ B 0) can be reduced to a single equation by using the magnetic scalar potential (11): H ƒ. This gives: 0 ƒ 1 ƒ 0, [1] Oxford Centre for Functional Magnetic Resonance Imaging of the Brain (FMRIB), University of Oxford, Department of Clinical Neurology, John Radcliffe Hospital, Headington, Oxford OX3 9DU, UK. rant sponsors: UK MRC, laxosmithkline, UK EPSRC. *Correspondence To: M. Jenkinson, Oxford Centre for Functional Magnetic Resonance Imaging of the Brain (FMRIB), University of Oxford, Department of Clinical Neurology, John Radcliffe Hospital, Headington, Oxford OX3 9DU, UK. mark@fmrib.ox.ac.uk Received 3 November 2003; revised 17 March 2004; accepted 3 April 2004 DOI /mrm Published online in Wiley InterScience ( Wiley-Liss, Inc. 471 where is the tissue dependent susceptibility. Let the susceptibility,, be expanded as: 0 1, [2] where 0 is the susceptibility of air ( ) and is a constant, equal to the difference in the susceptibility of the tissue under consideration and air (e.g., in the case of brain tissue and air). Note that 1 is a scaled version of the susceptibility difference and can take on continuous values when using a complex tissue model, but will take values of 0 or 1 for a simple two-tissue model. Similarly, expand in a series: [3] This perturbation expansion in can be substituted back into Eq. 1 to give: ƒ [4] 1 0 ƒ 2 1 ƒ 1 ƒ 0 0, [5] for the zeroth and first-order terms in. The first-order equation is a 3D Poisson equation: with solution: ƒ ƒ 1 1 ƒ 0 [6] 0 1 x x x f x dx, [7] where (x) (4 r) 1 is the reen s function, r x x 2 y 2 z 2 and f 1 ƒ 1 1 ƒ 0. 0 This convolution can also be more concisely written as 1 f. The z-component of the B field is given by: B z H z z z z O z 2. [8]

2 472 Jenkinson et al. The zeroth-order term is B z (0) 0 (1 0 ) 0 / z, and the first-order term is: Using the fact that: B z B z z. [9] f f x x x f holds for any and f, together with Eqs. 4, 5, 7 and 9, gives: 2 B 1 z 1 B 1 z y z 1 B y 2 Lorentz Correction x z 1 B x 2 z 1 B z. [10] The solution derived above is valid for a continuous medium but not for a medium composed of discrete particles. For MR calculations, however, it is the field that is applied externally to the discrete nuclei that is of interest. This field can be calculated from the continuous media solution using the Lorentz Correction (11,12) (or local field representation). The corrected field is given by B LC B 2 3 0M, where M H is the magnetization of the material. Therefore, the Lorentz Corrected field can be expressed in terms of the scalar potential, (the continuous media solution) as B LC 0 (1 /3)ƒ. Comparing this with the uncorrected (continuous media) solution, B 0 (1 )ƒ, shows that the Lorentz Corrected solution can be found simply by replacing all instances of with /3 in equations involving only and terms. Consequently, the zeroth and first-order corrected fields can be written as: B LC ƒ 0 [11] B 1 LC ƒ ƒ 1 [12] which, together with Eq. 10 gives: 1 B LC,z B LC,z y z 1 B LC,y 2 x z 1 B LC,x 2 z 2 1 B LC,z. [13] For the remainder of this article only the Lorentz Corrected fields will be used and the LC subscript will be dropped. Single Voxel Solution Equation 13 allows the first-order B z (1) field to be calculated if the zeroth-order field B (0) and the susceptibility distribution 1 are known. The zeroth-order field represents the field which would be present if there were no object in the scanner. For example, with a constant field in the z direction, B x (0) B y (0) 0 and B z (0) B 0, so that Eq. 13 simplifies considerably. The susceptibility distribution represents the object in the scanner and needs to be specified at each point in space. For complicated functions, though, the required convolutions are difficult, if not impossible, to do analytically. For most purposes, however, it is sufficient to approximate the object using small rectangular volume elements (voxels). The advantage of this is that the convolution can be done analytically for a single voxel. Consider a single voxel of dimensions (a, b, c) and, without loss of generality, let it be centered at the origin (0, 0, 0) with a susceptibility of 1 1 within the voxel and 1 0 outside the voxel. iven this, the required convolutions in Eq. 13 can be written as: 2 v z 1 B v 1 x x B v x x 2 v z x dx x a/2 x a/2 y b/2 dx y b/2 dy z c/2 z c/2dz 2 v z x B v, [14] where v stands for either x, y, or z and it is assumed that B v (0) is not spatially varying. The last integral can be easily calculated from the indefinite integral, which we will denote here as F (x). That is: F x 2 v z x B v dx [15] giving the single voxel solution as: H v,z x 2 v z 1 B v F x a 2, y b 2 2, z c F x a 2, y b 2, z c 2 F x a 2, y b 2, z c 2 F x a 2, y b 2, z c 2 F x a 2, y b 2, z c 2 F x a 2, y b 2, z c 2 F x a 2, y b 2, z c 2 F x a 2, y b 2, z c 2, where the following specific solutions can be substituted: For a constant (normalized) field along the z-axis: B (0) (x) (0, 0, 1):

3 Perturbative B 0 Calculation 473 F x 2 x z dxdydz 1 4 sinh 1 y x 2 z 2. [17] For a constant (normalized) field along the y-axis: B (0) (x) (0, 1, 0): F(x) is the same as for the x-axis case except all x and y terms are swapped. Combining Voxels Due to the linearity of Eq. 13 the single voxel solutions can be added together to give the total field: B 1 z x 1 x H x x, [18] x where x are the source points (locations of the voxel centers) and x is the field point where the field is evaluated. This equation takes the form of a discrete convolution of the voxel-based susceptibility map, 1, with the single voxel solution, H, if both field points and source points lie on the same rectangular grid. Therefore, the calculation can be efficiently implemented by using the 3D Fast Fourier Transform (FFT) as: B z 1 1 Z 1 H, [19] where (... ) is the FFT and Z(... ) is a zero-padding function, used to ensure that there is no periodic wrap-around in the convolution. Note that in previous sections the equations have been general enough so that any field point could be chosen. However, in order to take advantage of the efficiency of the discrete convolution the generality of the field point locations must be restricted to a grid (with spacing equal to the source voxel size and with the points located at the source voxel centers). radient of the Perturbed Field FI. 1. Spherical object model (a), analytical field (b) and perturbation field calculation (c) for the case of a sphere of radius R 0 8 voxels. The displayed scale of the field maps is 0.4 ppm. F x 2 z 2 dxdydz 1 4 atan xy. [16] zr For a constant (normalized) field along the x-axis: B (0) (x) (1, 0, 0): The exact analytical gradient of the perturbed field B z (1) can be calculated by simply replacing the convolution kernel, H, with its gradient: where q stands for x, y or z. 1 B z q 1 x H q x x x [20] Changing Orientation The field produced by the object in another orientation can be calculated by rotating the applied B (0) field. Note that this calculation is in the reference frame of the object, not the scanner, and so calculating the scanner defined z component of the perturbed field requires projection of the full B (1) field vector onto the scanner frame z-axis unit vector. Hence the desire to express Eqs. 10 and 13 in a general

4 474 Jenkinson et al. form containing B x (0) and B y (0) as well as B z (0). Furthermore, since B (1) is a linear function of B (0), a rotation can be reconstructed using a basis set of B (1) images resulting from B (0) fields in the x, y and z directions: 1 B x 1 1,0,0 B 1 x 0,1,0 B 1 x B 0 01 R B 1 z 1 y 1,0,0 B 1 y 0,1,0 B 1 y B 1 z 1,0,0 B 1 z 0,1,0 B 1 z 0,0,1 0 1 R 0, [21] where R isa3 3 matrix (13) that represents the rotation from the scanner coordinates, x sc, to the object coordinates, x ob, so that x ob Rx sc, and B p (1) (qˆ) is the field calculated in the object coordinate system with the p direction (x, y, or z) specifying the component of the field resulting from an applied field B (0) qˆ, where qˆ is a unit axis vector either xˆ (1, 0, 0), ŷ (0, 1, 0) or ẑ (0, 0, 1). For example, p y and q x gives B y (1) (xˆ), which represents the y component of the field due to a unit applied field along the x direction (all in the object coordinate system). Note that R. [001] T represents the scanner unit z-axis vector (the direction of the applied field) as represented in the object coordinate system. This vector specifies the direction of the applied B 0 field in the object coordinate system. The matrix of perturb fields, [B p (1) (qˆ)], represents a set of nine basis images, which can be precalculated and then combined as specified above to give the desired field at any orientation. This does not involve further approximation; it is precisely the same perturbed field that would be calculated for the object in the new orientation. In addition, although the perturbed field is linear in the basis images, it is not linear in the rotation angles, since the elements of R are nonlinear functions of these angles. A similar calculation can be done for the gradients of the field, ƒb (1), although this requires 27 basis images. VALIDATION AND RESULTS Analytical Sphere The analytical magnetic field (including Lorentz Correction) produced by a spherical object of radius R 0 and susceptibility i inside a medium of susceptibility e is given analytically by (12) as: B B 0ẑ i R e e r B 0 3 cos rˆ ẑ r R 0 B 0 ẑ r R 0, [22] where the angles and are defined by rˆ (cos sin, sin sin, cos ) and ẑ (0, 0, 1). This expression gives all three components of the field: B (B x,b y,b z ). Figure 1 shows, qualitatively, the z component of the field distribution (less B 0 ) for a sphere, comparing the analytical solution given in Eq. 22 with the solution calculated with the perturbation method for the case of Figure 2 shows plots of this field taken along the z-axis through the centers of these spheres for a range of different radii, R 0 8, 16, 32 voxels, where in each case FI. 2. Plots of the B z distribution along the z-axis for spherical objects of radius R 0 8, 16, 32 voxels. Results are shown for the analytical calculation (theory), the perturbation calculation, and the difference between the two.

5 Perturbative B 0 Calculation 475 Table 1 Quantitative Error Measurements of B z for Spherical Objects R 0 Mean Median P90 P95 P99 P99.9 Max Quantitative error measurements of the absolute difference in B z (in units of ppm) between the analytical result and the perturbation calculations for a spherical object of radius R 0 voxels, with The maximum field value (less B 0 ) for the analytical solution is 6.33 ppm. P90 represents the 90th percentile, etc. the voxel size is mm. This illustrates the relationship between spatial extent and size of the error. To investigate the dependence of the error on voxel size more quantitatively, the absolute difference between the analytical and perturbation-based calculated fields were generated for a larger range of radii: R 0 4, 8, 16, 32, 64 voxels. In each case, several measurements of the distribution of this absolute difference (error) were made mean, median, maximum and 90th, 95th, 99th and 99.9th percentiles which are shown in Table 1. Note that the measurements include all field points in the 3D volume, not just along an axis, as in Fig. 2. In addition, to test the ability of the method to calculate other components of the field the same absolute difference calculations for the x component of the field, B x, were generated. The results from this test are shown in Table 2. Note that the x-component of the field was calculated using a form of Eq. 13 derived for B x rather than B z (where the first term becomes zero and the last term interchanges x and z). The tables indicate that the errors are very small overall and that as the resolution of the object improves (i.e., more voxels or, equivalently, increased radius) the averaged errors (mean, median, percentiles) all decrease, such that for R 0 32 voxels, 95% of the voxels, or field points, have an error less than 1% of the maximum field, and 99% of the voxels have an error less than 10% of the maximum field. However, some larger errors still persist at any resolution, indicated by the maximum error values. These errors occur in a very small number of voxels which are typically located near the surface of the object, where the susceptibility changes sharply. These results are very similar to those shown in Refs. 1, 2, 5, 6 despite the range of different object models (e.g., boundary element methods vs. voxels) and approximations used. However, superior performance is shown by Ref. 3, where the object is modeled with piecewise planar patches and a boundary-element style calculation is used. In addition, the method presented in Refs. 7, 8, which modeled the object as a collection of spheres, shows very accurate performance, although a direct comparison is difficult from the available results. However, these methods are more numerically intensive and require significantly greater amounts of computational time. In Vivo Human Head An experimentally acquired field map of an in vivo human head was used to validate the method in practice. The MR field map sequence used a symmetric-asymmetric spinecho pair (14,15) (2.5 ms asymmetry time; voxels of size mm). The 3D theoretical field map was calculated using a 3D object susceptibility map ( mm), that was created by segmenting a whole-head CT image into bone, tissue, and air and then assigning a value of 1.0 (for 1 ) to the voxels classified as bone or tissue and a value of 0.0 to those classified as air. This susceptibility map was then registered to the magnitude component of the MR image corresponding to the acquired field map in order for the theoretically calculated field map to be registered to the experimentally acquired map. A full 3D theoretical field map was calculated where the source and field points were taken as the voxel centers from the input susceptibility map (1 mm spacing). Figure 3 shows 2D slices from the 3D CT image used to define the object susceptibility map, plus slices from both the 3D experimentally acquired field map and the 3D field map calculated using the voxel-based perturbation method described above (execution time was 9 min on a 1.8 Hz Athlon, 2 B memory running Linux). Note that both field maps have been masked so that only brain tissue is in- Table 2 Quantitative error measurements of B x for spherical objects R 0 Mean Median P90 P95 P99 P99.9 Max Quantitative error measurements of the absolute difference in B x (in units of ppm) between the analytical result and the perturbation calculations for a spherical object of radius R 0 voxels, with The maximum field value for the analytical solution is 4.75 ppm. P90 represents the 90th percentile, etc.

6 476 Jenkinson et al. FI. 3. Corresponding axial 2D slices from 3D images of an in vivo human head, showing the CT image used to derive the susceptibility map (a), the experimentally acquired field map (b), and the field map from the voxel-based perturbation calculation (c). The displayed scale of the field maps is 0.4 ppm. cluded (although the simulation included all tissues present, with brain bone 1.0) and have had the first and second-order spherical harmonics removed in order to factor out the effect of the shims on the field maps. Qualitatively it can be seen that the match is good. Quantitatively (see Table 3) the mean absolute difference between the field maps is ppm, while the typical range of the field values (used for the display range in Fig. 3) is 0.4 ppm. Furthermore, 90% of the voxels were in error by less than 0.1 ppm. The calculated error can be compared with the neglected second-order terms in the perturbation expansion. These second-order terms have an approximate magnitude of ppm, which is two orders of magnitude less than the observed errors. Therefore, it is likely that the observed errors are due to inaccuracies in the voxel-based modeling, as investigated in the previous section. DISCUSSION In this article we present a perturbation method for calculating the magnetic B field for an object with varying spatial susceptibility. A fast, first-order calculation is presented for voxel-based objects, using the analytical voxel solution. The accuracy of this method was tested using the analytical solution for a sphere as well as by empirical comparison with a human head dataset. These results indicate that highly localized errors of less than 1 ppm are achieved generally, which is very similar to other calculation methods, and sufficient for most MR imaging purposes. This implementation is available as a free download Table 3 Quantitative error measurements on in vivo human brain Mean Median P90 P95 P99 P99.9 Max Quantitative error measurements of the absolute difference in B z (in units of ppm) between the experimentally measured field map and the perturbation calculations for the in vivo human brain after removal of first- and second-order spherical harmonics. P90 represents the 90th percentile, etc. from mark/b0calc and will be distributed with future versions of FSL ( fsl). There are several main contributions of this work. The first is the use of a principled, perturbation method for arriving at the field approximation. This is useful in that it allows the magnitude of the error terms (second-order and higher) to be estimated, which then permits the relevant applicability of the method to be assessed. For instance, the method cannot be used for metallic objects where J 0 or objects where the susceptibility difference,, is large, but it can be used for some substances with slightly higher susceptibility than biological tissues, such as graphite (15). It is also possible, although potentially analytically intractable, to extend the approximation to higher orders to increase the accuracy. In addition, the formulation of the perturbation equations is separate from the object model specification and could be used with other object models, such as boundary element methods. Another significant contribution of this work is the ability to calculate more than just the z component of the field. In particular, the x and y components can be calculated just as easily (although separately) as well as the gradients of the fields (evaluated at the voxel centers), and formulations are provided for all these cases. More interestingly, it is possible to calculate the field for different object orientations by linearly combining precalculated basis images. This allows the field to be determined, without further approximation, at any orientation and in a very efficient manner. Such calculations will allow the interaction between susceptibility fields and motion artifacts to be explored more easily, a current research interest of the authors. In the field calculations used here there are two main sources of approximation beyond the requirement for zero conductivity: 1) neglecting all perturbation terms beyond first-order, and 2) representing the object by a voxel-based model. The first approximation limits the range of objects for which this method could be applied, as discussed above. The second approximation is potentially more limiting, as the use of a voxel-based model for the object will cause errors that are not as easily estimated as the pertur-

7 Perturbative B 0 Calculation 477 bation approximation errors and appear to dominate, as demonstrated in the results on the human head. Quantitative investigation of the object modeling errors was conducted on a range of spherical objects and the results are shown in Fig. 2 and Tables 1 and 2. These indicate that the spatial extent of larger errors is limited to a few voxels, which lie near the surface of the sphere (Fig. 1). This is to be expected, as this is closest to the area where the voxel-based object model deviates from the real, continuous object. It also suggests that using higher-resolution images leads to both smaller and more spatially localized errors overall, although at the cost of extra computational effort (the number of computations for N voxels is proportional to N log N). For models of the human head with air-filled cavities, the significant errors are therefore only likely to occur within a few voxels of the air tissue boundaries. Alternative models, such as boundary element methods (1 5,7,8), are likely to be physically accurate in capturing the object shape, but have two main disadvantages. One is that boundary meshes are more difficult to instantiate from typical voxel-based images (although one possible solution to this problem is given in Ref. 3) and the second is that they require more computation for the field calculation as each element (triangle of the mesh) is normally unique and requires separate calculations. In contrast, voxel-based models (6,10) are easy to instantiate and very efficient to calculate (using Fast Fourier Transforms). Furthermore, the numerical results on the spherical object indicate that similar errors are obtained by many of the proposed methods. Finally, all of these object models have an advantage over finite Fourier representations (9) since they can ensure that the object has finite spatial extent, which is not possible with the Fourier method. ACKNOWLEDMENT We thank Dr. Bob Cox for suggesting that a perturbation calculation could be useful for magnetic field calculations in notes from a workshop presentation on motion in FMRI. REFERENCES 1. Li S, Dardzinski BJ, Collins CM, Yang QX, Smith MB. Three-dimensional mapping of the static magnetic field inside the human head. Magn Reson Med 1996;36: Collins CM, Yang B, Yang QX, Smith MB. Numerical calculations of the static magnetic field in three-dimensional multi-tissue models of the human head. Magn Reson Imag 2002;20: Hwang SN, Wehrli FW. The calculation of the susceptibility-induced magnetic field from 3D NMR images with applications to trabecular bone. J Magn Reson B 1995;109: Balac A, Caloz. Magnetic susceptibility artifacts in magnetic resonance imaging: calculation of the magnetic field disturbances. IEEE Trans Magn 1996;32: Munck JD, Bhagwandien R, Muller SH, Verster FC, Herk MB. The computation of MR image distortions caused by tissue susceptibility using the boundary element method. IEEE Trans Med Imag 1996;15: Yoder D, Changchien E, Paschal CB, Fitzpatrick JM. MRI simulator with static field inhomogeneity. In: SPIE Proc Med Imag. Image Proc San Diego, February 2002; vol Case TA, Durney CH, Ailion DC, Cutillo A, Morris AH. A mathematical model of diamagnetic line broadening in lung tissue and similar heterogeneous systems: calculations and measurements. J Magn Reson 1987;73: Christman RA, Ailion DC, Case TA, Durney CH, Cutillo A, Shioya S, oodrich KC, Morris AH. Comparison of calculated and experimental nmr spectral broadening for lung tissue. Magn Reson Med 1996;35: Marques JP, Bowtell R. Evaluation of a Fourier based method for calculating susceptibility induced magnetic field perturbations. In: Proc 11th Annual Meeting ISMRM. Toronto, p Jenkinson M, Wilson J. Jezzard P. Perturbation calculation of B0 field for non-conducting materials. In: Proc 10th Annual Meeting ISMRM, Honolulu, p Schwinger J, DeRaad L Jr, Milton KA, Tsai WY. Classical electrodynamics. New York: Perseus Books; Haacke EM, Brown RW, Thompson MR, Venkatesan R. Magnetic resonance imaging: physical principles and sequence design. New York: Wiley-Liss; Foley J, Dam A, Feiner S, Hughes J. Computer graphics: principles and practice in C. Boston: Addison Wesley; Jezzard P, Balaban R. Correction for geometric distortion in echo planar images from B0 field variations. Magn Reson Med 1995;34: Wilson JL, Jenkinson M, Jezzard P. Optimization of static field homogeneity in human brain using diamagnetic passive shims. Magn Reson Med 2002;48:

Sources of Distortion in Functional MRI Data

Sources of Distortion in Functional MRI Data Human Brain Mapping 8:80 85(1999) Sources of Distortion in Functional MRI Data Peter Jezzard* and Stuart Clare FMRIB Centre, Department of Clinical Neurology, University of Oxford, Oxford, UK Abstract:

More information

Motion Robust Magnetic Susceptibility and Field Inhomogeneity Estimation Using Regularized Image Restoration Techniques for fmri

Motion Robust Magnetic Susceptibility and Field Inhomogeneity Estimation Using Regularized Image Restoration Techniques for fmri Motion Robust Magnetic Susceptibility and Field Inhomogeneity Estimation Using Regularized Image Restoration Techniques for fmri Desmond Tec Beng Yeo 1,, Jeffrey A. Fessler 1,, and Bolye Kim 1 1 Department

More information

k-space Interpretation of the Rose Model: Noise Limitation on the Detectable Resolution in MRI

k-space Interpretation of the Rose Model: Noise Limitation on the Detectable Resolution in MRI k-space Interpretation of the Rose Model: Noise Limitation on the Detectable Resolution in MRI Richard Watts and Yi Wang* Magnetic Resonance in Medicine 48:550 554 (2002) Noise limitation on the detected

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

Motion Correction in fmri by Mapping Slice-to-Volume with Concurrent Field-Inhomogeneity Correction

Motion Correction in fmri by Mapping Slice-to-Volume with Concurrent Field-Inhomogeneity Correction Motion Correction in fmri by Mapping Slice-to-Volume with Concurrent Field-Inhomogeneity Correction Desmond T.B. Yeo 1,2, Jeffery A. Fessler 2, and Boklye Kim 1 1 Department of Radiology, University of

More information

CHAPTER 4 RAY COMPUTATION. 4.1 Normal Computation

CHAPTER 4 RAY COMPUTATION. 4.1 Normal Computation CHAPTER 4 RAY COMPUTATION Ray computation is the second stage of the ray tracing procedure and is composed of two steps. First, the normal to the current wavefront is computed. Then the intersection of

More information

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction

A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction Tina Memo No. 2007-003 Published in Proc. MIUA 2007 A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction P. A. Bromiley and N.A. Thacker Last updated 13 / 4 / 2007 Imaging Science and

More information

Functional MRI in Clinical Research and Practice Preprocessing

Functional MRI in Clinical Research and Practice Preprocessing Functional MRI in Clinical Research and Practice Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization

More information

SPM8 for Basic and Clinical Investigators. Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous,

More information

Basic fmri Design and Analysis. Preprocessing

Basic fmri Design and Analysis. Preprocessing Basic fmri Design and Analysis Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial filtering

More information

Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction

Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction Nan-kuei Chen 1,5 and Alice M. Wyrwicz 4 * Magnetic Resonance in Medicine 51:147 153 (004) Odd even echo inconsistencies result

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing Functional Connectivity Preprocessing Geometric distortion Head motion Geometric distortion Head motion EPI Data Are Acquired Serially EPI Data Are Acquired Serially descending 1 EPI Data Are Acquired

More information

Partial k-space Reconstruction

Partial k-space Reconstruction Chapter 2 Partial k-space Reconstruction 2.1 Motivation for Partial k- Space Reconstruction a) Magnitude b) Phase In theory, most MRI images depict the spin density as a function of position, and hence

More information

K-Space Trajectories and Spiral Scan

K-Space Trajectories and Spiral Scan K-Space and Spiral Scan Presented by: Novena Rangwala nrangw2@uic.edu 1 Outline K-space Gridding Reconstruction Features of Spiral Sampling Pulse Sequences Mathematical Basis of Spiral Scanning Variations

More information

Following on from the two previous chapters, which considered the model of the

Following on from the two previous chapters, which considered the model of the Chapter 5 Simulator validation Following on from the two previous chapters, which considered the model of the simulation process and how this model was implemented in software, this chapter is concerned

More information

Three-dimensional numerical simulations of susceptibility-induced magnetic field inhomogeneities in the human head

Three-dimensional numerical simulations of susceptibility-induced magnetic field inhomogeneities in the human head Magnetic Resonance Imaging 20 (2002) 759 770 Three-dimensional numerical simulations of susceptibility-induced magnetic field inhomogeneities in the human head Trong-Kha Truong a, Bradley D. Clymer b,

More information

Partial k-space Recconstruction

Partial k-space Recconstruction Partial k-space Recconstruction John Pauly September 29, 2005 1 Motivation for Partial k-space Reconstruction a) Magnitude b) Phase In theory, most MRI images depict the spin density as a function of position,

More information

Fmri Spatial Processing

Fmri Spatial Processing Educational Course: Fmri Spatial Processing Ray Razlighi Jun. 8, 2014 Spatial Processing Spatial Re-alignment Geometric distortion correction Spatial Normalization Smoothing Why, When, How, Which Why is

More information

Biomagnetic inverse problems:

Biomagnetic inverse problems: Biomagnetic inverse problems: Magnetic resonance electrical property tomography (MREPT) and magnetoencephalography (MEG) 2018 Aug. 16 The University of Tokyo Takaaki Nara 1 Contents Measurement What is

More information

2. Creating Field Maps Using the Field Map GUI (Version 2.0) in SPM5

2. Creating Field Maps Using the Field Map GUI (Version 2.0) in SPM5 1. Introduction This manual describes how to use the Field Map Toolbox Version 2.0 for creating unwrapped field maps that can be used to do geometric distortion correction of EPI images in SPM5. 1. 1.

More information

3D Surface Reconstruction of the Brain based on Level Set Method

3D Surface Reconstruction of the Brain based on Level Set Method 3D Surface Reconstruction of the Brain based on Level Set Method Shijun Tang, Bill P. Buckles, and Kamesh Namuduri Department of Computer Science & Engineering Department of Electrical Engineering University

More information

MRI Physics II: Gradients, Imaging

MRI Physics II: Gradients, Imaging MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes

More information

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005 Field Maps John Pauly October 5, 25 The acquisition and reconstruction of frequency, or field, maps is important for both the acquisition of MRI data, and for its reconstruction. Many of the imaging methods

More information

The SIMRI project A versatile and interactive MRI simulator *

The SIMRI project A versatile and interactive MRI simulator * COST B21 Meeting, Lodz, 6-9 Oct. 2005 The SIMRI project A versatile and interactive MRI simulator * H. Benoit-Cattin 1, G. Collewet 2, B. Belaroussi 1, H. Saint-Jalmes 3, C. Odet 1 1 CREATIS, UMR CNRS

More information

A Transmission Line Matrix Model for Shielding Effects in Stents

A Transmission Line Matrix Model for Shielding Effects in Stents A Transmission Line Matrix Model for hielding Effects in tents Razvan Ciocan (1), Nathan Ida (2) (1) Clemson University Physics and Astronomy Department Clemson University, C 29634-0978 ciocan@clemon.edu

More information

Role of Parallel Imaging in High Field Functional MRI

Role of Parallel Imaging in High Field Functional MRI Role of Parallel Imaging in High Field Functional MRI Douglas C. Noll & Bradley P. Sutton Department of Biomedical Engineering, University of Michigan Supported by NIH Grant DA15410 & The Whitaker Foundation

More information

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 3, JUNE 2003 405 Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging Edward J. Soares, Kevin W. Germino,

More information

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford Diffusion MRI Acquisition Karla Miller FMRIB Centre, University of Oxford karla@fmrib.ox.ac.uk Diffusion Imaging How is diffusion weighting achieved? How is the image acquired? What are the limitations,

More information

Optimizing Flip Angle Selection in Breast MRI for Accurate Extraction and Visualization of T1 Tissue Relaxation Time

Optimizing Flip Angle Selection in Breast MRI for Accurate Extraction and Visualization of T1 Tissue Relaxation Time Optimizing Flip Angle Selection in Breast MRI for Accurate Extraction and Visualization of T1 Tissue Relaxation Time GEORGIOS KETSETZIS AND MICHAEL BRADY Medical Vision Laboratory Department of Engineering

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

Susceptibility Distortion Correction for Echo Planar Images with Non-uniform B-Spline Grid Sampling: A Diffusion Tensor Image Study

Susceptibility Distortion Correction for Echo Planar Images with Non-uniform B-Spline Grid Sampling: A Diffusion Tensor Image Study Susceptibility Distortion Correction for Echo Planar Images with Non-uniform B-Spline Grid Sampling: A Diffusion Tensor Image Study M.O. Irfanoglu 1,L.Walker 3,S.Sammet 2, C. Pierpaoli 3,andR.Machiraju

More information

Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques

Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques Sea Chen Department of Biomedical Engineering Advisors: Dr. Charles A. Bouman and Dr. Mark J. Lowe S. Chen Final Exam October

More information

Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm

Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm L.P. Panych 1,3, B. Madore 1,3, W.S. Hoge 1,3, R.V. Mulkern 2,3 1 Brigham and

More information

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Sensors & Transducers 2013 by IFSA http://www.sensorsportal.com Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Shinji Ohyama, Masato Mukouyama Graduate School

More information

IMAGE reconstruction in conventional magnetic resonance

IMAGE reconstruction in conventional magnetic resonance IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 24, NO. 3, MARCH 2005 325 Conjugate Phase MRI Reconstruction With Spatially Variant Sample Density Correction Douglas C. Noll*, Member, IEEE, Jeffrey A. Fessler,

More information

A MORPHOLOGY-BASED FILTER STRUCTURE FOR EDGE-ENHANCING SMOOTHING

A MORPHOLOGY-BASED FILTER STRUCTURE FOR EDGE-ENHANCING SMOOTHING Proceedings of the 1994 IEEE International Conference on Image Processing (ICIP-94), pp. 530-534. (Austin, Texas, 13-16 November 1994.) A MORPHOLOGY-BASED FILTER STRUCTURE FOR EDGE-ENHANCING SMOOTHING

More information

Constrained Reconstruction of Sparse Cardiac MR DTI Data

Constrained Reconstruction of Sparse Cardiac MR DTI Data Constrained Reconstruction of Sparse Cardiac MR DTI Data Ganesh Adluru 1,3, Edward Hsu, and Edward V.R. DiBella,3 1 Electrical and Computer Engineering department, 50 S. Central Campus Dr., MEB, University

More information

ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T

ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T Revision date: 12/13/2010 Overview Susceptibility Weighted Imaging (SWI) is a relatively new data acquisition and processing

More information

Basic principles of MR image analysis. Basic principles of MR image analysis. Basic principles of MR image analysis

Basic principles of MR image analysis. Basic principles of MR image analysis. Basic principles of MR image analysis Basic principles of MR image analysis Basic principles of MR image analysis Julien Milles Leiden University Medical Center Terminology of fmri Brain extraction Registration Linear registration Non-linear

More information

Supplementary Data. in residuals voxel time-series exhibiting high variance, for example, large sinuses.

Supplementary Data. in residuals voxel time-series exhibiting high variance, for example, large sinuses. Supplementary Data Supplementary Materials and Methods Step-by-step description of principal component-orthogonalization technique Below is a step-by-step description of the principal component (PC)-orthogonalization

More information

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging 1 CS 9 Final Project Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging Feiyu Chen Department of Electrical Engineering ABSTRACT Subject motion is a significant

More information

A Coupled 3D/2D Axisymmetric Method for Simulating Magnetic Metal Forming Processes in LS-DYNA

A Coupled 3D/2D Axisymmetric Method for Simulating Magnetic Metal Forming Processes in LS-DYNA A Coupled 3D/2D Axisymmetric Method for Simulating Magnetic Metal Forming Processes in LS-DYNA P. L Eplattenier *, I. Çaldichoury Livermore Software Technology Corporation, Livermore, CA, USA * Corresponding

More information

Computational Algorithm for Estimating the Gradient Isocenter of an MRI Scanner

Computational Algorithm for Estimating the Gradient Isocenter of an MRI Scanner Computational Algorithm for Estimating the Gradient Isocenter of an MRI Scanner R Cai 1, KE Schubert 1, R Schulte 2 ritchiecai@r2labsorg, keith@r2labsorg, rschulte@dominionllumcedu 1 School of Computer

More information

Application of Tatian s Method to Slanted-Edge MTF Measurement

Application of Tatian s Method to Slanted-Edge MTF Measurement Application of s Method to Slanted-Edge MTF Measurement Peter D. Burns Eastman Kodak Company, Rochester, NY USA 465-95 ABSTRACT The 33 method for the measurement of the spatial frequency response () of

More information

A TEMPORAL FREQUENCY DESCRIPTION OF THE SPATIAL CORRELATION BETWEEN VOXELS IN FMRI DUE TO SPATIAL PROCESSING. Mary C. Kociuba

A TEMPORAL FREQUENCY DESCRIPTION OF THE SPATIAL CORRELATION BETWEEN VOXELS IN FMRI DUE TO SPATIAL PROCESSING. Mary C. Kociuba A TEMPORAL FREQUENCY DESCRIPTION OF THE SPATIAL CORRELATION BETWEEN VOXELS IN FMRI DUE TO SPATIAL PROCESSING by Mary C. Kociuba A Thesis Submitted to the Faculty of the Graduate School, Marquette University,

More information

Automatic segmentation of the cortical grey and white matter in MRI using a Region Growing approach based on anatomical knowledge

Automatic segmentation of the cortical grey and white matter in MRI using a Region Growing approach based on anatomical knowledge Automatic segmentation of the cortical grey and white matter in MRI using a Region Growing approach based on anatomical knowledge Christian Wasserthal 1, Karin Engel 1, Karsten Rink 1 und André Brechmann

More information

design as a constrained maximization problem. In principle, CODE seeks to maximize the b-value, defined as, where

design as a constrained maximization problem. In principle, CODE seeks to maximize the b-value, defined as, where Optimal design of motion-compensated diffusion gradient waveforms Óscar Peña-Nogales 1, Rodrigo de Luis-Garcia 1, Santiago Aja-Fernández 1,Yuxin Zhang 2,3, James H. Holmes 2,Diego Hernando 2,3 1 Laboratorio

More information

3-D Gradient Shimming

3-D Gradient Shimming 3-D Gradient Shimming 3-D Gradient Shimming on imaging and micro-imaging systems Technical Overview Introduction Advantage statement The GE3DSHIM method is a 3-D gradient shimming procedure developed for

More information

1 (5 max) 2 (10 max) 3 (20 max) 4 (30 max) 5 (10 max) 6 (15 extra max) total (75 max + 15 extra)

1 (5 max) 2 (10 max) 3 (20 max) 4 (30 max) 5 (10 max) 6 (15 extra max) total (75 max + 15 extra) Mierm Exam CS223b Stanford CS223b Computer Vision, Winter 2004 Feb. 18, 2004 Full Name: Email: This exam has 7 pages. Make sure your exam is not missing any sheets, and write your name on every page. The

More information

Learning Algorithms for Medical Image Analysis. Matteo Santoro slipguru

Learning Algorithms for Medical Image Analysis. Matteo Santoro slipguru Learning Algorithms for Medical Image Analysis Matteo Santoro slipguru santoro@disi.unige.it June 8, 2010 Outline 1. learning-based strategies for quantitative image analysis 2. automatic annotation of

More information

Improved Shim Method Based on the Minimization of the Maximum Off-Resonance Frequency for Balanced Steady- State Free Precession (bssfp)

Improved Shim Method Based on the Minimization of the Maximum Off-Resonance Frequency for Balanced Steady- State Free Precession (bssfp) Improved Shim Method Based on the Minimization of the Maximum Off-Resonance Frequency for Balanced Steady- State Free Precession (bssfp) Jongho Lee, 1,2 * Michael Lustig, 2 Dong-hyun Kim, 3 and John M.

More information

LATEST TRENDS on APPLIED MATHEMATICS, SIMULATION, MODELLING

LATEST TRENDS on APPLIED MATHEMATICS, SIMULATION, MODELLING 3D surface reconstruction of objects by using stereoscopic viewing Baki Koyuncu, Kurtuluş Küllü bkoyuncu@ankara.edu.tr kkullu@eng.ankara.edu.tr Computer Engineering Department, Ankara University, Ankara,

More information

Three-dimensional analytical field calculation of pyramidal-frustum shaped permanent magnets Janssen, J.L.G.; Paulides, J.J.H.; Lomonova, E.A.

Three-dimensional analytical field calculation of pyramidal-frustum shaped permanent magnets Janssen, J.L.G.; Paulides, J.J.H.; Lomonova, E.A. Three-dimensional analytical field calculation of pyramidal-frustum shaped permanent magnets Janssen, J.L.G.; Paulides, J.J.H.; Lomonova, E.A. Published in: IEEE Transactions on Magnetics DOI: 10.1109/TMAG.2009.2021861

More information

Invariant Recognition of Hand-Drawn Pictograms Using HMMs with a Rotating Feature Extraction

Invariant Recognition of Hand-Drawn Pictograms Using HMMs with a Rotating Feature Extraction Invariant Recognition of Hand-Drawn Pictograms Using HMMs with a Rotating Feature Extraction Stefan Müller, Gerhard Rigoll, Andreas Kosmala and Denis Mazurenok Department of Computer Science, Faculty of

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

A Diagonal Split-cell Model for the High-order Symplectic FDTD Scheme

A Diagonal Split-cell Model for the High-order Symplectic FDTD Scheme PIERS ONLINE, VOL. 2, NO. 6, 2006 715 A Diagonal Split-cell Model for the High-order Symplectic FDTD Scheme Wei Sha, Xianliang Wu, and Mingsheng Chen Key Laboratory of Intelligent Computing & Signal Processing

More information

arxiv: v1 [cs.cv] 6 Jun 2017

arxiv: v1 [cs.cv] 6 Jun 2017 Volume Calculation of CT lung Lesions based on Halton Low-discrepancy Sequences Liansheng Wang a, Shusheng Li a, and Shuo Li b a Department of Computer Science, Xiamen University, Xiamen, China b Dept.

More information

Modeling and Analysis of the Electric Field and Potential Distribution in a Wire-Cylinder Air Gap

Modeling and Analysis of the Electric Field and Potential Distribution in a Wire-Cylinder Air Gap Modeling and Analysis of the Electric Field and Potential Distribution in a Wire-Cylinder Air Gap KONSTANTINOS N. KIOUSIS, ANTONIOS X. MORONIS Technological Educational Institute (TEI) of Athens Energy

More information

This Time. fmri Data analysis

This Time. fmri Data analysis This Time Reslice example Spatial Normalization Noise in fmri Methods for estimating and correcting for physiologic noise SPM Example Spatial Normalization: Remind ourselves what a typical functional image

More information

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy Chenyang Xu 1, Siemens Corporate Research, Inc., Princeton, NJ, USA Xiaolei Huang,

More information

Automatic Gradient Preemphasis Adjustment: A 15-Minute Journey to Improved Diffusion-Weighted Echo-Planar Imaging

Automatic Gradient Preemphasis Adjustment: A 15-Minute Journey to Improved Diffusion-Weighted Echo-Planar Imaging Automatic Gradient Preemphasis Adjustment: A 15-Minute Journey to Improved Diffusion-Weighted Echo-Planar Imaging Vincent J. Schmithorst* and Bernard J. Dardzinski Magnetic Resonance in Medicine 47:208

More information

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Doug Dean Final Project Presentation ENGN 2500: Medical Image Analysis May 16, 2011 Outline Review of the

More information

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Advanced Image Reconstruction Methods for Photoacoustic Tomography Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic

More information

Solving a Two Dimensional Unsteady-State. Flow Problem by Meshless Method

Solving a Two Dimensional Unsteady-State. Flow Problem by Meshless Method Applied Mathematical Sciences, Vol. 7, 203, no. 49, 242-2428 HIKARI Ltd, www.m-hikari.com Solving a Two Dimensional Unsteady-State Flow Problem by Meshless Method A. Koomsubsiri * and D. Sukawat Department

More information

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig

More information

We have proposed a new method of image reconstruction in EIT (Electrical Impedance

We have proposed a new method of image reconstruction in EIT (Electrical Impedance Impedance tomography using internal current density distribution measured by nuclear magnetic resonance Eung Je Woo, Soo Yeol Lee, Chi Woong Mun Kon Kuk University, Department of Biomedical Engineering

More information

Research Collection. Localisation of Acoustic Emission in Reinforced Concrete using Heterogeneous Velocity Models. Conference Paper.

Research Collection. Localisation of Acoustic Emission in Reinforced Concrete using Heterogeneous Velocity Models. Conference Paper. Research Collection Conference Paper Localisation of Acoustic Emission in Reinforced Concrete using Heterogeneous Velocity Models Author(s): Gollob, Stephan; Vogel, Thomas Publication Date: 2014 Permanent

More information

A Novel Two-step Method for CT Reconstruction

A Novel Two-step Method for CT Reconstruction A Novel Two-step Method for CT Reconstruction Michael Felsberg Computer Vision Laboratory, Dept. EE, Linköping University, Sweden mfe@isy.liu.se Abstract. In this paper we address the parallel beam 2D

More information

CMP 2 Grade Mathematics Curriculum Guides

CMP 2 Grade Mathematics Curriculum Guides CMP 2 Grade 7 2007 2008 Mathematics Curriculum Guides Wisconsin Mathematics Standard MPS Learning Target Wisconsin Assessment Descriptors for Mathematics Curriculum Throughout The Year A. Mathematical

More information

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions EE225E/BIOE265 Spring 2013 Principles of MRI Miki Lustig Assignment 8 Solutions 1. Nishimura 7.1 P)jc%z 5 ivcr3. 1. I Due Wednesday April 10th, 2013 (a Scrhon5 R2iwd b 0 ZOms Xn,s r cx > qs 4-4 8ni6 4

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

Skåne University Hospital Lund, Lund, Sweden 2 Deparment of Numerical Analysis, Centre for Mathematical Sciences, Lund University, Lund, Sweden

Skåne University Hospital Lund, Lund, Sweden 2 Deparment of Numerical Analysis, Centre for Mathematical Sciences, Lund University, Lund, Sweden Volume Tracking: A New Method for Visualization of Intracardiac Blood Flow from Three-Dimensional, Time-Resolved, Three-Component Magnetic Resonance Velocity Mapping Appendix: Theory and Numerical Implementation

More information

Learning to Identify Fuzzy Regions in Magnetic Resonance Images

Learning to Identify Fuzzy Regions in Magnetic Resonance Images Learning to Identify Fuzzy Regions in Magnetic Resonance Images Sarah E. Crane and Lawrence O. Hall Department of Computer Science and Engineering, ENB 118 University of South Florida 4202 E. Fowler Ave.

More information

Whole Body MRI Intensity Standardization

Whole Body MRI Intensity Standardization Whole Body MRI Intensity Standardization Florian Jäger 1, László Nyúl 1, Bernd Frericks 2, Frank Wacker 2 and Joachim Hornegger 1 1 Institute of Pattern Recognition, University of Erlangen, {jaeger,nyul,hornegger}@informatik.uni-erlangen.de

More information

Semi-Automatic Detection of Cervical Vertebrae in X-ray Images Using Generalized Hough Transform

Semi-Automatic Detection of Cervical Vertebrae in X-ray Images Using Generalized Hough Transform Semi-Automatic Detection of Cervical Vertebrae in X-ray Images Using Generalized Hough Transform Mohamed Amine LARHMAM, Saïd MAHMOUDI and Mohammed BENJELLOUN Faculty of Engineering, University of Mons,

More information

XI Conference "Medical Informatics & Technologies" VALIDITY OF MRI BRAIN PERFUSION IMAGING METHOD

XI Conference Medical Informatics & Technologies VALIDITY OF MRI BRAIN PERFUSION IMAGING METHOD XI Conference "Medical Informatics & Technologies" - 2006 medical imaging, MRI, brain perfusion Bartosz KARCZEWSKI 1, Jacek RUMIŃSKI 1 VALIDITY OF MRI BRAIN PERFUSION IMAGING METHOD Brain perfusion imaging

More information

Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging

Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging Nan-kuei Chen and Alice M. Wyrwicz* Magnetic Resonance in Medicine 41:1206 1213 (1999) A novel and effective technique is described

More information

Mathematics for Computer Graphics - Ray Tracing III

Mathematics for Computer Graphics - Ray Tracing III Mathematics for Computer Graphics - Ray Tracing III Dr. Philippe B. Laval Kennesaw State University November 12, 2003 Abstract This document is a continuation of the previous documents on ray tracing.

More information

CORRECTION OF IMAGE DISTORTION IN ECHO PLANAR IMAGE SERIES USING PHASE AND INTENSITY. Ning Xu. Dissertation. Submitted to the Faculty of the

CORRECTION OF IMAGE DISTORTION IN ECHO PLANAR IMAGE SERIES USING PHASE AND INTENSITY. Ning Xu. Dissertation. Submitted to the Faculty of the CORRECTION OF IMAGE DISTORTION IN ECHO PLANAR IMAGE SERIES USING PHASE AND INTENSITY By Ning Xu Dissertation Submitted to the Faculty of the Graduate School of Vanderbilt University in partial fulfillment

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

Optimal Sampling Geometries for TV-Norm Reconstruction of fmri Data

Optimal Sampling Geometries for TV-Norm Reconstruction of fmri Data Optimal Sampling Geometries for TV-Norm Reconstruction of fmri Data Oliver M. Jeromin, Student Member, IEEE, Vince D. Calhoun, Senior Member, IEEE, and Marios S. Pattichis, Senior Member, IEEE Abstract

More information

Institute of Cardiovascular Science, UCL Centre for Cardiovascular Imaging, London, United Kingdom, 2

Institute of Cardiovascular Science, UCL Centre for Cardiovascular Imaging, London, United Kingdom, 2 Grzegorz Tomasz Kowalik 1, Jennifer Anne Steeden 1, Bejal Pandya 1, David Atkinson 2, Andrew Taylor 1, and Vivek Muthurangu 1 1 Institute of Cardiovascular Science, UCL Centre for Cardiovascular Imaging,

More information

Detection of Unique Point Landmarks in HARDI Images of the Human Brain

Detection of Unique Point Landmarks in HARDI Images of the Human Brain Detection of Unique Point Landmarks in HARDI Images of the Human Brain Henrik Skibbe and Marco Reisert Department of Radiology, University Medical Center Freiburg, Germany {henrik.skibbe, marco.reisert}@uniklinik-freiburg.de

More information

Scientific Visualization Example exam questions with commented answers

Scientific Visualization Example exam questions with commented answers Scientific Visualization Example exam questions with commented answers The theoretical part of this course is evaluated by means of a multiple- choice exam. The questions cover the material mentioned during

More information

Volume Visualization

Volume Visualization Volume Visualization Part 1 (out of 3) Overview: Volume Visualization Introduction to volume visualization On volume data Surface vs. volume rendering Overview: Techniques Simple methods Slicing, cuberille

More information

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Outline High Spectral and Spatial Resolution MR Imaging (HiSS) What it is How to do it Ways to use it HiSS for Radiation

More information

Volume Illumination and Segmentation

Volume Illumination and Segmentation Volume Illumination and Segmentation Computer Animation and Visualisation Lecture 13 Institute for Perception, Action & Behaviour School of Informatics Overview Volume illumination Segmentation Volume

More information

XI Signal-to-Noise (SNR)

XI Signal-to-Noise (SNR) XI Signal-to-Noise (SNR) Lecture notes by Assaf Tal n(t) t. Noise. Characterizing Noise Noise is a random signal that gets added to all of our measurements. In D it looks like this: while in D

More information

Adaptive osculatory rational interpolation for image processing

Adaptive osculatory rational interpolation for image processing Journal of Computational and Applied Mathematics 195 (2006) 46 53 www.elsevier.com/locate/cam Adaptive osculatory rational interpolation for image processing Min Hu a, Jieqing Tan b, a College of Computer

More information

Consistency in Tomographic Reconstruction by Iterative Methods

Consistency in Tomographic Reconstruction by Iterative Methods Consistency in Tomographic Reconstruction by Iterative Methods by M. Reha Civanlar and H.J. Trussell Center for Communications and Signal Processing Department of Electrical and Computer Engineering North

More information

2.1 Signal Production. RF_Coil. Scanner. Phantom. Image. Image Production

2.1 Signal Production. RF_Coil. Scanner. Phantom. Image. Image Production An Extensible MRI Simulator for Post-Processing Evaluation Remi K.-S. Kwan?, Alan C. Evans, and G. Bruce Pike McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University, Montreal,

More information

Calculating the Distance Map for Binary Sampled Data

Calculating the Distance Map for Binary Sampled Data MITSUBISHI ELECTRIC RESEARCH LABORATORIES http://www.merl.com Calculating the Distance Map for Binary Sampled Data Sarah F. Frisken Gibson TR99-6 December 999 Abstract High quality rendering and physics-based

More information

Projection Technique for Vortex-Free Image Registration

Projection Technique for Vortex-Free Image Registration Projection Technique for Vortex-Free Image Registration Patrick Scheibe 1, Ulf-Dietrich Braumann 1,2, Jens-Peer Kuska 2 1 Translational Center for Regenerative Medicine (TRM) 2 Interdisciplinary Center

More information

3D Vascular Segmentation using MRA Statistics and Velocity Field Information in PC-MRA

3D Vascular Segmentation using MRA Statistics and Velocity Field Information in PC-MRA 3D Vascular Segmentation using MRA Statistics and Velocity Field Information in PC-MRA Albert C. S. Chung 1, J. Alison Noble 1, Paul Summers 2 and Michael Brady 1 1 Department of Engineering Science, Oxford

More information

Zigzag Sampling for Improved Parallel Imaging

Zigzag Sampling for Improved Parallel Imaging Magnetic Resonance in Medicine 60:474 478 (2008) Zigzag Sampling for Improved Parallel Imaging Felix A. Breuer, 1 * Hisamoto Moriguchi, 2 Nicole Seiberlich, 3 Martin Blaimer, 1 Peter M. Jakob, 1,3 Jeffrey

More information