Evaluation of a Combined Array-Planar Crystal for Gamma-ray Scintillation Imagers
|
|
- Angela Stewart
- 5 years ago
- Views:
Transcription
1 Evaluation of a Combined Array-Planar Crystal for Gamma-ray Scintillation Imagers R.Pani 1, R.Scafè 2, R.Pellegrini 1, A.Soluri 3, N.Burgio 2, C.Ciavola 2, M.N.Cinti 1, L.Indovina 1, G.De Vincentis 1, F.Garibaldi 4, F.Cusanno 4, G.M.Urciuoli 4 and E.Cisbani 4 1 Department of Experimental Medicine, University of Rome "La Sapienza", Rome, Italy 2 ENEA, C.R. Casaccia, Rome, Italy 3 Institute of Biomedical Technologies, CNR, Rome, Italy 4 Laboratory of Physics, ISS, Rome, Italy
2 R697 ( 8 inch ) Hamamatsu PSPMT Family ( crossed wire anode) R3292 ( 5 inch ) R2486 ( 3 inch ) R59 ( 1 inch )
3 SPEM Single Photon Emission Mammography A new scintigraphic technique based upon a small FOV dedicated gamma camera The main advantages of this detection system with respect to Anger camera are the compactness and the light weight SPEM detection system SPEM detection head
4 SPEM Single Photon Emission Mammography SPEM detection system allows to perform scintigrafic breast projection as Rx mammography Uncompressed Breast Compressed Breast Cranio Caudal projection
5 Spatial resolution in PSM From: J.A.Sorenson and M.E.Phelps Physics in Nuclesr Medicine Grune & Strattom ANGER CAMERA 1-15 cm Object collimator distance OCD (cm )
6 Spatial resolution in SPEM Moderate compression 3 cm SPEM
7 PSM (Prone Scinti Mammography) by using commercially available Gamma Camera Proper positioning in PSM Prone scintimammography is the only useful view allowed for a patient
8 SPEM Single Photon Emission Mammography 99m Tc Sestamibi scintimammography: Right Breast Carcinoma 13 mm sized a b a: SPEM Camera scan clearly shows the lesion as an inhomogeneous area of increase uptake of tracer (arrow) b: The same breast imaged by commercially available Anger Camera showing the lesion as a small hot spot (arrow)
9 99m Tc Sestamibi Scintimammography a b Right Breast Carcinoma 7 mm sized c d RCC uncompressed RCC compressed a and b SPEM Camera imaging with breast uncompressed (a) and mildly compressed: into the red circle the lesion is shown as an area of increased uptake of tracer c: The same breast imaged by commercially available Anger Camera: no pathological uptake is shown d: The corresponding X-Ray mammography- the lesion is surrounded by a red circle
10 2 x 2 PSPMT Hamamatsu R 59-C8 With flanges 2 x 2 PSPMT Hamamatsu R 76-C8 25 mm Flangeless 3 mm dead zone 2 mm
11 Diagram of Multi-PSPMT camera (All dimensions in mm). External dimension 27.7 ±.5 25 ±.5 22 photocathode 18 anode Flanges (2 mm) Overall dead zone 5.7 mm Scintillation light distribution
12 Read-out electronics and acquisition system Camera base PSDD PC Σx, Πx, Σy, Πy Quad ADC FAST 774 FAST Intf box PC FAST MP card
13 CsI(Na) planar crystal 55 x 55 x 3 mm 3 2 mm quartz CsI(Tl) crystal array 3 x 3 x 3 mm 3 pixel size 3 mm quartz 2 x 2 PSPMT Hamamtsu R59-C8 Flood filed irradiation 14 kev 13
14 1 mm quartz PSPMT R59-C8 Hamamatsu CsI(Tl) crystal array 3 x 3 x 3 mm 3 Flood filed irradiation 14 kev 3 mm quartz 5 mm quartz
15 Multi PSPMT Spatial Resolution CsI(Tl) crystal array 3x3x3 mm 3 / 3 mm quartz 1 cm Counts FWHM 2.5 mm resolution [mm FWHM] position [pixel] CsI(Na) planar crystal / 2 mm quartz position [mm]
16 6 counts per channel (a.u.) x 2 PSPMT R59-C8 FWHM=21% flood spot CsI(Na) planar crystal 14 kev irradiation Channels Counts per channel CsI(Tl) crystal array 3 x 3 x 3 mm 3 / 3 mm quartz FWHM=29% Counts per channel Channels Channels Channels 16
17 SCHEMA CAPC
18 COMBINED ARRAY- PLANAR CRYSTAL CsI(Na) CAPC 3 x 3 mm 2 /.25 mm dead zone
19 2 x 2 PSPMT Hamamtsu R76-C8 Individual PSPMT Energy Response from Flood Field Irradiation at 14 kev Counts Channels
20 R59-C8 PSPMT Crystal array / 3 mm quartz CAPC Flood Field Irradiation R59-C8 R76-C8 Counts CsI(Tl) CsI(Na) CsI(Na) Channe Channels Counts 45 Channels Counts Channels
21 Counts FWHM=25% R76 without flanges Multi PSPMT- CsI(Na) CAPC Energy Resolution Counts R59 with flanges Channels FWHM=22% Counts FWHM=22% Canali Channels kev spot irradiation Channels
22 2 x 2 PSPMT Hamamatsu 76-C8 CAPC CsI(Na) Spot irradiation Counts Position (pixels)
23 Counts Counts CsI(Na) CAPC Spatial Resolution 14 kev spot irradiation 5 mm FWHM 2.7 mm FWHM mm Position(Pixels) FWHM mm R76-C8 R59-C8 5 mm Position(Pixels)
Fits you like no other
Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room
More informationFits you like no other
Fits you like no other Philips BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase
More informationCherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar
Doctoral Thesis Time-of-Flight Time-of-Flight Positron Positron Emission Emission Tomography Tomography Using Using Cherenkov Cherenkov Radiation Radiation Rok Dolenec Supervisor: Prof. Dr. Samo Korpar
More informationValidation of the Small Animal Biospace Gamma Imager Model Using GATE Monte Carlo Simulations on the Grid
Validation of the Small Animal Biospace Gamma Imager Model Using GATE Monte Carlo Simulations on the Grid Joe Aoun, Vincent Breton, L. Desbat, B. Bzeznik, M. Leabadand, J. Dimastromatteo To cite this version:
More informationIntroduction to Emission Tomography
Introduction to Emission Tomography Gamma Camera Planar Imaging Robert Miyaoka, PhD University of Washington Department of Radiology rmiyaoka@u.washington.edu Gamma Camera: - collimator - detector (crystal
More informationValidation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition
Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.
More informationCorso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET
Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 5 SPECT, PET Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes
More informationIntroduction to Positron Emission Tomography
Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator
More informationImproving spatial resolution in SPECT with the combination of PSPMT based detector and iterative reconstruction algorithms
Computerized Medical Imaging and Graphics 27 (2003) 307 313 www.elsevier.com/locate/compmedimag Improving spatial resolution in SPECT with the combination of PSPMT based detector and iterative reconstruction
More informationCharacterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide
2005 IEEE Nuclear Science Symposium Conference Record M04-8 Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide J. S. Karp, Senior Member, IEEE, A. Kuhn, Member, IEEE, A. E. Perkins,
More informationProject GAMMA: Adaptive Methods for Medical Applications
Project GAMMA: Adaptive Methods for Medical Applications V. Solovov, A. Morozov, V. Chepel, V. Domingos, R.Martins, L. Pereira, F. Neves, A. Lindote, F. Fraga, I.Lopes, F. Alves LIP-Coimbra and Coimbra
More informationSymbia E and S System Specifications Answers for life.
www.siemens.com/mi Symbia E and S System Specifications Answers for life. Symbia E: Work with confidence. What does it mean to work with confidence? It is clarity not only image clarity but the clarity
More informationSPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.
SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce
More informationConflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program
James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives 1. Be familiar with recommendations
More informationInterpolating Silicon Photomultipliers
Interpolating Silicon Photomultipliers Peter Fischer, Heidelberg University, Germany (Presenter) Claudio Piemonte, FBK, Italy We present the novel Interpolating Silicon PhotoMultiplier (ISiPM) topology
More informationMonte-Carlo simulation for scatter correction compensation studies in SPECT imaging using GATE software package
Nuclear Instruments and Methods in Physics Research A 569 (2006) 404 408 www.elsevier.com/locate/nima Monte-Carlo simulation for scatter correction compensation studies in SPECT imaging using GATE software
More informationPositron. MillenniumVG. Emission Tomography Imaging with the. GE Medical Systems
Positron Emission Tomography Imaging with the MillenniumVG GE Medical Systems Table of Contents Introduction... 3 PET Imaging With Gamma Cameras PET Overview... 4 Coincidence Detection on Gamma Cameras...
More informationNuclear Instruments and Methods in Physics Research A
Nuclear Instruments and Methods in Physics Research A 621 (2010) 590 594 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima
More informationSymbia T Series System Specifications Answers for life.
www.siemens.com/mi System Specifications Answers for life. 2 Diagnostic Excellence Today s healthcare institutions have increasingly diverse patient needs, greater budget limitations and the need for higher
More informationConstructing System Matrices for SPECT Simulations and Reconstructions
Constructing System Matrices for SPECT Simulations and Reconstructions Nirantha Balagopal April 28th, 2017 M.S. Report The University of Arizona College of Optical Sciences 1 Acknowledgement I would like
More informationImplementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system
Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation
More informationMedical Imaging BMEN Spring 2016
Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,
More informationA region segmentation based algorithm for building crystal position lookup table in scintillation detector
A region segmentation based algorithm for building crystal position lookup table in scintillation detector WANG Hai-Peng ( 王海鹏 ) 1,2,3 YUN Ming-Kai ( 贠明凯 ) 1,2 LIU Shuang-Quan ( 刘双全 ) 1,2 FAN Xin ( 樊馨
More informationReview of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA
Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.
More informationValidation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation camera dedicated to small animal imaging
Validation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation camera dedicated to small animal imaging D Lazaro 1, I Buvat 2, G Loudos 3,4, D Strul 5, G Santin 6, N Giokaris
More informationSmall Angle Gamma Ray Scatter: What Is The Impact On Image Quality
ISPUB.COM The Internet Journal of Medical Technology Volume 4 Number 2 Small Angle Gamma Ray Scatter: What Is The Impact On Image Quality G CUrrie, J Wheat Citation G CUrrie, J Wheat.. The Internet Journal
More informationPositron Emission Tomography
Physics 656 Seminar on Physical Fundamentals of Medical Imaging Positron Emission Tomography Ahmed Qamesh Outline What is PET? PET mechanism Radionuclide and its synthesis Detection concept and Development
More informationDeliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors
MS116 Characterization of light production, propagation and collection for both organic and inorganic scintillators D10.2 R&D on new and existing scintillation materials: Report on the light production,
More informationJames R Halama, PhD Loyola University Medical Center
James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives Be familiar with the tests recommended
More informationPhoton counting spectral CT versus conventional CT: comparative evaluation for breast imaging application
Physics in Medicine & Biology Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application To cite this article: Polad M Shikhaliev and Shannon G Fritz 2011
More informationVALIDATION OF GATE MONTE CARLO SIMULATIONS FOR INDIUM 111 IMAGING
VALIDATION OF GATE MONTE CARLO SIMULATIONS FOR INDIUM IMAGING Karine Assié, Isabelle Gardin, Pierre Véra, and Irène Buvat Abstract-- Accurate quantification of indium ( In) Single Photon Emission Computed
More informationSPECT: Physics Principles and Equipment Design
SPECT: Physics Principles and Equipment Design Eric C. Frey, Ph.D., Professor Division of Medical Imaging Physics Russell H. Morgan Department of Radiology and Radiological Science Disclosures Johns Hopkins
More informationCHAPTER 11 NUCLEAR MEDICINE IMAGING DEVICES
CHAPTER 11 M.A. LODGE, E.C. FREY Russell H. Morgan Department of Radiology and Radiological Sciences, Johns Hopkins University, Baltimore, Maryland, United States of America 11.1. INTRODUCTION Imaging
More information8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial
8//0 AAPM0 Scientific Symposium: Emerging and New Generation PET: Instrumentation, Technology, Characteristics and Clinical Practice Aug Wednesday 0:4am :pm Solid State Digital Photon Counting PET/CT Instrumentation
More informationBiomedical Imaging. Computed Tomography. Patrícia Figueiredo IST
Biomedical Imaging Computed Tomography Patrícia Figueiredo IST 2013-2014 Overview Basic principles X ray attenuation projection Slice selection and line projections Projection reconstruction Instrumentation
More informationIntroduction to Biomedical Imaging
Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray
More informationBME I5000: Biomedical Imaging
1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.
More informationAX-PET : A novel PET concept with G-APD readout
AX-PET : A novel PET concept with G-APD readout Matthieu Heller CERN - PH/DT Marie Curie network MC-PAD Matthieu.heller@cern.ch On behalf of the AX-PET collaboration https://twiki.cern.ch/twiki/bin/view/axialpet
More informationValidation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation camera dedicated to small-animal imaging
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 49 (24) 271 285 PHYSICS IN MEDICINE AND BIOLOGY PII: S31-9155(4)68367-9 Validation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation
More informationIn-vivo dose verification for particle therapy
In-vivo dose verification for particle therapy D.R. Schaart, NCS Lustrum, 5-Oct-2012 HollandPTC 1 Protons vs. photons Photons Protons Dennis R. Schaart Delft University of Technology 2 Protons: the promise
More informationProgram Modeling Concepts: Lesson-2: DATA FLOW GRAPH MODELS
Program Modeling Concepts: Lesson-2: DATA FLOW GRAPH MODELS 1 1. DFG 2 Data Flow graph (DFG)( A data-flow means that a program flow such that specifically the data only determines all the program execution
More informationDigital Scatter Removal in Mammography to enable Patient Dose Reduction
Digital Scatter Removal in Mammography to enable Patient Dose Reduction Mary Cocker Radiation Physics and Protection Oxford University Hospitals NHS Trust Chris Tromans, Mike Brady University of Oxford
More informationDesign and assessment of a novel SPECT system for desktop open-gantry imaging of small animals: A simulation study
Design and assessment of a novel SPECT system for desktop open-gantry imaging of small animals: A simulation study Navid Zeraatkar and Mohammad Hossein Farahani Research Center for Molecular and Cellular
More informationAn educational tool for demonstrating the TOF-PET technique
Nuclear Instruments and Methods in Physics Research A 471 (2001) 200 204 An educational tool for demonstrating the TOF-PET technique T.Bȧack a, *, J. Cederkȧall a, B. Cederwall a, A. Johnson a, A. Kerek
More informationMediso AnyScan S Single-Head and Dual-Head SPECT
Mediso AnyScan S Single-Head and Dual-Head SPECT ANYSCAN S (SINGLE-HEAD & DUAL-HEAD SPECT) Low Cost of Ownership Small Footprint Absolute Imaging Solutions tm is the exclusive U.S. source for Mediso Medical
More informationV/Q Imaging Protocols
Elizabeth ailey ppsc(mrs) M Department of Nuclear Medicine Royal North Shore Hospital, Sydney, ustralia email: ebailey@nsccahs.health.nsw.gov.au The V/Q scan has traditionally used planar imaging techniques.
More informationWe are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors
We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,500 108,000 1.7 M Open access books available International authors and editors Downloads Our
More informationMediso AnyScan. Single-Head and Dual-Head SPECT
Mediso AnyScan S Single-Head and Dual-Head SPECT ANYSCAN S (SINGLE-HEAD & DUAL-HEAD SPECT) Low Cost of Ownership Associated Imaging Services is the exclusive U.S. source in the Midwest for Mediso Medical
More informationFlat Panel Detectors. P-Series ISS. Dynamic Visualization SmartSwitch. Wireless. Portable. Technology
Flat Panel Detectors P-Series Portable Wireless ISS Technology DV Dynamic Visualization SmartSwitch Introducing P-Series Flat Panel Detectors DR solutions made especially for private practices like yours.
More informationPhilips SPECT/CT Systems
Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET
More informationMaster of Science Thesis. Activity quantification of planar gamma camera images
Master of Science Thesis Activity quantification of planar gamma camera images Marie Sydoff Supervisor: Sigrid Leide-Svegborn The work has been performed at Department of Radiation Physics Malmö University
More informationPreliminary Study of an Intra-operative PET Imaging Probe System
Preliminary Study of an Intra-operative PET Imaging Probe System by Sam Seoung Huh A dissertation submitted in partial fulfillment of the requirements for the degree of Doctor of Philosophy (Biomedical
More informationNuclear Medicine Imaging
Introduction to Medical Engineering (Medical Imaging) Suetens 5 Nuclear Medicine Imaging Ho Kyung Kim Pusan National University Introduction Use of radioactive isotopes for medical purposes since 1920
More informationSpherical Crystal X-ray Imaging for MTW, OMEGA, and OMEGA EP
Spherical Crystal X-ray Imaging for MTW, OMEGA, and OMEGA EP C.STOECKL, G. FISKEL, R. K. JUNGQUIST, P. M. NILSON, AND W. THEOBALD University of Rochester, Laboratory for Laser Energetics Spherical Crystal
More informationEducational Objectives SPECT/CT: Basics, Quality Assurance, and Clinical Applications
Educational Objectives SPECT/CT: Basics, Quality Assurance, and Clinical Applications 1. To understand the physics principles underlying SPECT/CT image acquisition and reconstruction 2. To understand the
More informationBEIJING HAMAMATSU PHOTON TECHNIQUES INC. Photon is our business BHP6602 Smart γ Camera Special NM Device for Details
BHP6602 Smart γ Camera Special NM Device for Details 1 / 6 Over View The BHP6602 is a smart γ camera just required less than eight square feet of floor space and provided DICOM compatibility. The redesigned
More informationA new PET prototype for proton therapy: comparison of data and Monte Carlo simulations
A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations V. Rosso, G.Battistoni, N. Belcari, N. Camarlinghi, A. Ferrari, S. Ferretti, A. Kraan, A. Mairani, N. Marino, J. E.
More informationFast Timing and TOF in PET Medical Imaging
Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part
More informationSiPMs for Čerenkov imaging
SiPMs for Čerenkov imaging Peter Križan University of Ljubljana and J. Stefan Institute Trends in Photon Detectors in Particle Physics and Calorimetry, Trieste, June 2-4, 2008 Contents Photon detectors
More informationSNIC Symposium, Stanford, California April The Hybrid Parallel Plates Gas Counter for Medical Imaging
The Hybrid Parallel Plates Gas Counter for Medical Imaging F. Anulli, G. Bencivenni, C. D Ambrosio, D. Domenici, G. Felici, F. Murtas Laboratori Nazionali di Frascati - INFN, Via E. Fermi 40, I-00044 Frascati,
More informationExtremely Fast Detector for 511 kev Gamma
Extremely Fast Detector for 511 kev Gamma V. Sharyy, D. Yvon, G. Tauzin, E.Delagnes, Ph. Abbon, J P. Bard, M. Kebbiri, M. Alokhina, C. Canot IRFU, CEA D. Breton, J. Maalmi LAL,IN2P3 Journée 2015 du Labex
More informationG3050 RadSearch: Radioactivity and Decommissioning Monitor
G3050 RadSearch: Radioactivity and Decommissioning Monitor 1. RadSearch overview RadSearch is a gamma radiation detection device based on medium-resolution gamma ray spectroscopy (MRGS). It is a real time,
More informationDetailed analysis of scatter contribution from different simulated geometries of X-ray detectors
Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors Elena Marimon 1,2*, Hammadi Nait-Charif 1, Asmar Khan 2, Philip A. Marsden 3, Oliver Diaz 4 1 Centre for
More informationAvailability Study of the Italian Electricity SCADA System in the Cloud
Availability Study of the Italian Electricity SCADA System in the Cloud Stefano Sebastio a, Antonio Scala b,a, and Gregorio D Agostino c,a a LIMS London Institute of Mathematical Sciences, London, UK b
More informationCorso di laurea in Fisica A.A Fisica Medica 4 TC
Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure
More information3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction
3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction Yuni K. Dewaraja, Member, IEEE, Michael Ljungberg, Member, IEEE, and Jeffrey A. Fessler, Member, IEEE Abstract- we
More informationGamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector
Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector K. Karafasoulis, K. Zachariadou, S. Seferlis, I. Papadakis, D. Loukas, C. Lambropoulos, C. Potiriadis Abstract Spectroscopic
More informationGeant4 in Brachytherapy
Geant4 in Brachytherapy 1. 2. 3. 4. 5. Brachytherapy: Brief Overview Clinical applications Basic research Ultrafast & biology applications Issues for the work group 1 Brachytherapy: Overview Brachy: Greek
More informationN23 CCD CAMERA FOR X-RAY IMAGE INTENSIFIER. CCD Camera Head. Control unit (CCU RACK) N23 SERIES
N23 CCD CAMERA FOR X-RAY IMAGE INTENSIFIER N23 Series CCD CAMERA The N23 CCD camera, manufactured using printed circuit board with SMD technology, offers better performance than NICAL s well know N20 camera
More informationTime-Resolved measurements by FEL spontaneous emission: A proposal for sub-picosecond pumps & probe structural and spectrometric investigations
Time-Resolved measurements by FEL spontaneous emission: A proposal for sub-picosecond pumps & probe structural and spectrometric investigations V. Rossi Albertini, B. Paci & P. Perfetti Istituto di Struttura
More informationA Further Study on Large Size LSO and LYSO Crystal Samples
October 25, 2005 IEEE NSS, N12-6, Puerto-Rico 1 A Further Study on Large Size LSO and LYSO Crystal Samples Jianming Chen, Liyuan Zhang, Ren-yuan Zhu California Institute of Technology BGO, LSO & LYSO Samples
More information3/19/2013. Outline. Objective. SPECT System Description. Calibration of an MR-Compatible, CZT Detector Based Stationary Small Animal SPECT system
Calibration of an MR-Compatible, CZT Detector Based Stationary Small Animal SPECT system Andrew Rittenbach 1, Jingyan Xu 1, AbdEl-Monem El-Sharkawy 1, William Edelstein 1, Kevin Parnham 2, James Hugg 2,
More informationGPU implementation for rapid iterative image reconstruction algorithm
GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial
More informationSemi-Quantitative Metrics in Positron Emission Tomography. Michael Adams. Department of Biomedical Engineering Duke University.
Semi-Quantitative Metrics in Positron Emission Tomography by Michael Adams Department of Biomedical Engineering Duke University Date: Approved: Timothy G. Turkington, Supervisor Adam P. Wax Terence Z.
More informationSimulations in emission tomography using GATE
Simulations in emission tomography using GATE Irène Buvat buvat@imed.jussieu.fr Laboratory of Functional Imaging, U678 INSERM, Paris, France Outline Emission tomography and need for simulations GATE short
More informationCOUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER
COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER By RAMI RIMON ABU-AITA A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY
More informationFPGA based charge fast histogramming for GEM detector
FPGA based charge fast histogramming for GEM detector Krzysztof T. Pozniak a, A. Byszuk a, M. Chernyshova b, R. Cieszewski a, T. Czarski b, W. Dominik c, K. Jakubowska b, G. Kasprowicz a, J. Rzadkiewicz
More informationStudy of a High-Resolution PET System Using a Silicon Detector Probe TESIS DOCTORAL UNIVERSITAT DE VALENCIA
thesis 2015/5/29 15:50 page 1 #1 thesis 2015/5/29 15:50 page 2 #2 thesis 2015/5/29 15:50 page 1 #3 Study of a High-Resolution PET System Using a Silicon Detector Probe Karol Brzeziński TESIS DOCTORAL UNIVERSITAT
More informationExpected Performances of the scintillator counters Time Of Flight system of the AMS-02 experiment
Expected Performances of the scintillator counters Time Of Flight system of the AMS-02 experiment Cristina Sbarra for the AMS-TOF Bologna group (sbarra@bo.infn.it) INFN-Bologna (Italy) Cristina Sbarra
More informationDesign and simulation of continuous scintillator with pixellated photodetector
University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 2001 Design and simulation of continuous scintillator with pixellated
More informationQuantitative capabilities of four state-of-the-art SPECT-CT cameras
Seret et al. EJNMMI Research 2012, 2:45 ORIGINAL RESEARCH Open Access Quantitative capabilities of four state-of-the-art SPECT-CT cameras Alain Seret 1,2*, Daniel Nguyen 1 and Claire Bernard 3 Abstract
More informationSPECT/CT Basics, Technology Updates, Quality Assurance, and Applications
SPECT/CT Basics, Technology Updates, Quality Assurance, and Applications S. Cheenu Kappadath, PhD Department of Imaging Physics University of Texas M D Anderson Cancer Center, Houston, Texas skappadath@mdanderson.org
More informationPerformance optimization of a PET insert for simultaneous. breast PET/MR imaging
Performance optimization of a PET insert for simultaneous breast PET/MR imaging PERFORMANCE OPTIMIZATION OF A PET INSERT FOR SIMULTANEOUS BREAST PET/MR IMAGING BY YICHENG LIANG, B.Sc. a thesis submitted
More informationMedical Imaging Modalities
Image Science Introduction Medical Imaging Modalities Ho Kyung Kim hokyung@pusan.ac.kr Pusan National University Projection Radiography Routine diagnostic radiography Chest x rays, fluoroscopy, mammography,
More informationA Method for Producing Simulated Mammograms: Observer Study
A Method for Producing Simulated Mammograms: Observer Study Payam Seifi M.Sc. Michael R. Chinander Ph.D. Robert M. Nishikawa Ph.D., FAAPM Carl J. Vyborny Translational Laboratory for Breast Imaging Research
More information218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997
218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997 Compton Scatter and X-ray Crosstalk and the Use of Very Thin Intercrystal Septa in High-Resolution PET Detectors Craig S. Levin, Member,
More informationEmission Computed Tomography Notes
Noll (24) ECT Notes: Page 1 Emission Computed Tomography Notes Introduction Emission computed tomography (ECT) is the CT applied to nuclear medicine. There are two varieties of ECT: 1. SPECT single-photon
More informationOptimization of light collection from crystal scintillators
Optimization of light collection from crystal scintillators V.M.Mokina1,2, F.А.Danevich1, V.V.Kobychev1, R.V.Kobychev1,3, H.Kraus4, V.Mikhailik4,5 1 Institute for Nuclear Research, Kyiv, Ukraine 2 3 National
More informationIEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 181 3-D Monte Carlo-Based Scatter Compensation in Quantitative I-131 SPECT Reconstruction Yuni K. Dewaraja, Member, IEEE, Michael Ljungberg,
More informationShadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS
Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic
More informationin PET Medical Imaging
Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part
More informationarxiv:physics/ v2 [physics.med-ph] 23 Nov 2005
arxiv:physics/0511065v2 [physics.med-ph] 23 Nov 2005 Experimental study of a liquid Xenon PET prototype module M.-L. Gallin-Martel 1, P. Martin, F. Mayet, J. Ballon, G. Barbier, C. Barnoux, J. Berger,
More informationAn Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste
An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste S.C. Kane 1, S. Croft 1, P. McClay 1, W.F. Mueller
More informationTHE COMBINED USE OF 3D DOSE ASSESSMENT AND HUMAN MOTION SIMULATION IN ALARA D&D PROBLEMS
THE COMBINED USE OF 3D DOSE ASSESSMENT AND HUMAN MOTION SIMULATION IN ALARA D&D PROBLEMS F. Vermeersch, SCK CEN, Boeretang 200, 2400 Mol Belgium Abstract Operations in a D&D environment are in most cases
More informationAdvantages of combining gamma scanning techniques and 3D dose simulation in dose optimisation problems.
Advantages of combining gamma scanning techniques and 3D dose simulation in dose optimisation problems. F.Vermeersch SCK CEN, Boeretang 200, 2400 Mol, Belgium Abstract In this paper we present a method
More informationO-ARM IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE
IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE SYSTEM FEATURES CATEGORY PHYSICAL DIMENSIONS IMAGING MODALITY PERFORMANCE SPECIFICATION Length Width Height Weight Gantry Opening Bore Diameter Single Plane
More informationFuture Prospects of Time-of-Flight PET: Scintillator, Detector and Depth-of- interaction (DOI) Technologies. TOF scanners. Clinical benefit of TOF PET
Future Prospects of Time-of-Flight PET: Scintillator, Detector and Depth-of- interaction (DOI) Technologies Suleman Surti Physics & Instrumentation Group Department of Radiology University of Pennsylvania
More informationMathematical methods and simulations tools useful in medical radiation physics
Mathematical methods and simulations tools useful in medical radiation physics Michael Ljungberg, professor Department of Medical Radiation Physics Lund University SE-221 85 Lund, Sweden Major topic 1:
More informationCh. 4 Physical Principles of CT
Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between
More informationSimulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images
Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),
More information