1 Signet 5700 ph/ Monitor Instructions English * * Rev. K /06 English/ Contents 1. Power Connections. Compatible Electrode/Preamp Wiring. 4 0 ma Current Output Connections 4. Relay Connections 5. Relay Operation CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. 6. Menu Functions 7. Parts and Accessories 8. Specifi cations 9. Quick Reference Menu Parameters 10. Troubleshooting 11. Maintenance Remove terminal blocks for easy wiring Freq. IN Freq. IN Sen. Pwr. Iso. Gnd Gnd Gnd PLS AUX output output Std. Sensor Open Collector Sensor LISTED UL 77CJ 14 V 10 W 1 Total reset Patent No. D76,8 Flow 1 Freq. IN Freq. IN Sen. Pwr. Iso. Gnd Gnd Gnd PLS AUX output output Std. Sensor Open Collector Sensor LISTED UL 77CJ 14 V 10 W 1 Total reset Patent No. D76,8 Flow Power Connections. Compatible Electrode/Preamp Wiring CAUTION! Never connect 110 VAC or 0 VAC to rear power terminals. High voltage AC will damage instrument and void warranty. 14 V 10 W 5700 Terminals External power supply 1 4 VDC OR 1 4 VAC mv Input Brown wire Silver (earth ground) Brown Iso. Gnd Blue wire Blue T Green wire Green T White wire White Red Black V Red Wire V Black wire 40 ma ph/ Preamp: Signet 70 TwistLock Electrodes 5700 Terminals DryLoc Electrodes Preamp: Signet 7601 Signet 760 = Double Insulated = DC or AC power Technical Notes: To reduce the possibility of noise interference, isolate AC power lines from signal lines. Maximum 40 ma loop impedance (sec. ) is affected by the supply voltage. Use six conductor shielded cable for cable extensions up to 10 m (400 ft) max. (#550064) Cable shield must be maintained through cable splice.
2 . 4 0 ma Current Output Connections Technical Notes: **1/8A fuse recommended (customer supplied) * 40 ma output is internally powered (nonisolated), maximum loop impedance 50 Ω with a 1 V instrument supply voltage, 950 Ω with a 4 V instrument supply voltage. To isolate output and prevent ground loop problems: 1. Use monitor device with isolated inputs, or. Use separate DC supply for 5700 and monitor device, or. Power 5700 with1 4 VAC step down transformer 4. Relay Connections Two internal relay contact sets (COM, NO, and NC) may be used for external device control. Front panel LED annunciators indicate the activation status of each relay. Each relay can control up to two devices simultaneously, as shown. Relay operation modes include Low alarm, High alarm, and Proportional Pulse (sec. 5). Common device connections include: Pulse mode metering pump control Pulse mode solenoid valve control Low or High mode warning lamps Low or High mode bells or sirens Low or High mode external heavyduty relay Wiring Example Right Device A IS powered when relay is deenergized (front panel LED off ). Power is discontinued when the relay setpoint is reached (front panel LED on ). Device B IS NOT powered when the relay is deenergized. Power is applied after the relay setpoint is reached Terminals Fuse** 1/8A 40 ma Relay ph/ Relay 1 NO C NC NO C NC External AC/DC power Relay Deenergized NO C NC Device A Device B Relay Energized Technical Notes: Maximum relay contact ratings: 5 0 VDC, 5 15 VAC, or 50 VAC An external heavyduty relay must be used for devices with surge currents or operating currents that exceed the above specifi cations. * 5700 Terminals 4 0 ma NO C NC 5. Relay Operation A. LOW alarm mode The relay is energized when the drops below the setpoint, and is deenergized when the rises above the setpoint plus hysteresis (sec. 6.D, 6.E). Low setpoint= B. HIGH alarm mode The relay is energized when the rises above the setpoint and is deengergized when the falls below the setpoint plus hysteresis (sec. 6.D, 6.E). High setpoint= Adjustable hysteresis= Adjustable hysteresis= Relay energized= Relay deenergized= Relay energized= Relay deenergized= C. Proportional PULSE mode The proportional pulse relay configuration is primarily designed for metering pump control. The operator is prompted to enter a minimum and maximum setpoint and maximum pulse rate for the assigned relay (sec. 6.F, 6.G). Relay pulse width is fixed at 10 ms. Refer to the operation examples below. Metering pump chemical addition (dry contact activation type required) Example 1 (right): As the process rises above the minimum pulse setpoint (500 mv) the relay begins pulsing; triggering the metering pump for chemical addition. As the continues to rise, pulsing accelerates proportionally until the maximum programmed pulse rate of 00 pulses/minute and setpoint (500 mv) are reached, forcing the process back down to intended levels (e.g. 500 mv) mv 0 pulses Relay 1 Rate: 00 Pulses/min > Relay 1 or Relay 00 pulses 500 mv 500 mv 1000 mv Example (right): As the process falls below the minimum pulse setpoint (500 mv) the relay begins pulsing; triggering the metering pump for chemical addition. As the continues to decrease, pulsing accelerates proportionally until the maximum programmed pulse rate of 00 pulses/minute and setpoint (500 mv) are reached, forcing the process back up to intended levels (e.g. 500 mv). 00 pulses Relay 1 or Relay 0 pulses 1000 mv 500 mv 500 mv 1000 mv Signet 5700 ph/ Monitor Instructions
3 ENTER 6. Menu Functions To access either CALIBRATE or OPTIONS menus, press and hold the ENTER key as illustrated below: Press & hold ENTER key for access: ENTER Menus: VIEW menu (sec. 6.1): The VIEW menu is displayed during standard operation. The operator can navigate freely through the menu by pressing either UP or DOWN arrow keys. CALIBRATE Menu (sec. 6.): The CALIBRATE menu contains all critical display setup and output parameters. A simple security code feature prevents unauthorized tampering. The operator is required to enter a simple access code for menu access. The same code also unlocks OPTIONS menus. VIEW (sec. 6.1) CALIBRATE (sec. 6.) OPTIONS (sec. 6.4) seconds 5 seconds OPTIONS Menu (sec. 6.4): The OPTIONS menu contains setup and display features that are seldom accessed for minor display or output adjustments. Reversible Dial Your 5700 includes a reversible dial face for ph use. See enclosed ph manual for operation details. A B 0 14 ph ph 0 ALARM 1 ALARM C ALARM 1 ALARM D 6.1 VIEW Menu G 1 Choose: Change: Save: Menu Displays A G: (Factory default displays shown in menu column 1) A. mv/raw mv A. Shows calibrated solution mv and Raw (uncalibrated) input mv from electrode/preamplifi e r B. C. Loop output B. Loop output display: shows the loop current output level. C. range display: shows the programmed min and max meter range (sec. 6.A) D. E. Range Relay 1 Relay D. Relay 1 display: this display shows the programmed operation mode and setpoint for relay 1 (sec. 6.DE) E. Relay display: this display shows the programmed operation mode and setpoint for relay (sec. 6.DE) F. G. Last calibration Buffer calibration A Press and release to access EASY CAL Procedure See EASY CAL Procedure (Section 6.) F. Last calibration: this display shows a user defi ned setup date for maintenance records. This feature is not an internal timer or calender. G. EASY CAL: press the right arrow key to access the EASY CAL buffer calibration procedure (sec. 6.). Signet 5700 ph/ Monitor Instructions
4 6. EASY CAL Procedure Requirements This procedure simplifi es system calibration using standard ph 4.0 and 7.0 buffers saturated with Quinhydrone (customer supplied). If standard ph 4.0 and 7.0 buffers and Quinhydrone are not available, the system can be calibrated using the CALIBRATE menu standard and slope settings (sec. 6.B, 6.C). Quinhydrone (customer supplied) is the oxidizer measured by the electrode and is essential for EASY CAL calibration. To guarantee saturation with ph 4.0 and 7.0 buffers, mix 1/8 g (1/4 ml) of Quinhydrone powder for each 50 ml of ph buffer solution (sec. 7). Press keys in sequence to continue, will appear during code entry. 1 To Calibrate: Response: To Accept: Place electrode tip in first ph buffer saturated with Quinhydrone; Allow for stabilization Press ENTER to accept first buffer calibration. Displayed Data mv: Shows the calibrated solution oxidation reduction potentioal (). Input mv: Shows the actual Raw (uncalibrated) preamplifi er electrode input ph mv ph mv 1 0 seconds* For best results, gently stir the submerged electrode for approximately 5 seconds during the stabilization period (step ). Place electrode tip in second (different) ph buffer saturated with Quinhydrone; Allow for stabilization Press ENTER to accept second buffer calibration. ph mv ph mv 0 seconds* Display returns to VIEW Menu in s To exit EASY CAL at any time without saving changes: quick press Technical Note: The differences between input mv and mv is a good indication of the electrode's condition. Differences in excess of 50 mv may indicate a need to service the electrode. Refer to electrode manual for maintenance recommendations. Theoretical mv values 5 C mv Signet 5700 ph/ Monitor Instructions
5 6. CALIBRATE Menu Requirements System calibration is possible with two known solutions within 1000 mv 000 mv. The EASY CAL procedure (sec. 6.) is recommended when standard ph buffers and Quinhydrone are available. If EASY CAL is performed, manual calibration steps BC below are not required and should be skipped. Press keys in sequence will appear during code entry. to continue, H A. B. C. 1 Choose: Range Electrode standard Meter min max Allow for stabilization Change: Save: Menu Settings A H: (Factory default displays shown in menu column 1) A. Sets Min Max meter dial range, 1000 mv 000 mv (factory installed dial, 1000 mv 1000 mv). Contact factory for custom dial confi gurations. Does not effect 4 to 0 ma output B. Sets electrode standard to any value from 1000 mv to 000 mv (standard value must be 10 mv from slope value, step C) D. Electrode slope Relay Allow for stabilization Relay Low or High Mode Selected Mode Setpoint ENTER C. Sets electrode slope to any value from 1000 mv to 000 mv (slope value must be 10 mv from standard value, step B) Menu items DG repeat for relay setup. E. Relay Hysteresis "SAVING" briefly displays D. Sets relay operation mode Low or High, and (mv) setpoint, 1000 mv 000 mv (sec. 5AB) F. Relay Pulse Mode Selected, step D E. Sets relay hysteresis, mv. Set to zero to disable feature (sec. 5AB) G. Relay Relay Mode Pulse rate Range F. Sets relay minimum and maximum pulse setpoint, 1000 mv 000 mv (sec. 5C) G. Sets relay pulse rate, pulses/minute. H. To return to VIEW: Last calibration A quick press Optional To restore original value: quick press H. Sets user defi ned setup date for maintenance records. This feature is not an internal timer or calender Quick Reference Calibration Procedures: Point Quick Calibration (recommended): 1. Set solution standard (step B). Set solution slope (step C) 1Point Calibration (optional): 1. Set solution standard (step B) Signet 5700 ph/ Monitor Instructions 5
6 6.4 OPTIONS Menu Press keys in sequence will appear during code entry. to continue, E A. 1 Choose: Change: Save: Menu Settings A E: (Factory default displays shown in menu column 1) Contrast Low High A. Selects display contrast: 5 levels B. Loop output 4 ma setpoint B. Sets 4 ma setpoint: mv. 4 ma and 0 ma setpoints are reversible C. D. Loop output 0 ma setpoint ENTER "SAVING" briefly displays C. Sets 0 ma setpoint: mv. 0 ma and 4 ma setpoints are reversible D. Adjusts 4 ma output:.0 to 5.0 ma (overrides factory 4.00 ma calibration) E. Loop output 4 ma adjust E. Adjusts 0 ma output: 19 to 1 ma (overrides factory 0.00 ma calibration) To return to VIEW: Loop output A quick press 0 ma adjust To restore original value: quick press 7. Parts and Accessories Splashproof rear cover # (code ) 5 x 5 inch adapter plate for Signet retrofi t # (code ) Optional surface mount bracket # (code ) Power supply, 10 VAC 4 VAC, # ph buffer kit, ph 4.0, 7.0, 10.0, # (code ) Front snapon bezel, # (code ) 5700 ph/ instruction sheet, # (code ) 6 Signet 5700 ph/ Monitor Instructions
7 8. Specifications General Compatible sensors: Signet 70 ph/ Preamplifi er/ Sensor System Accuracy: ±0.% of full scale Input range: : 1000 to 000 mv, optically isolated (standard 1000 to 1000 mv dial included) Enclosure: Rating: NEMA 4X/IP65 front Dimensions: 1/4 DIN, 96 x 96 x 88 mm (.8 x.8 x.5 in.) Case: ABS plastic Keypad: Sealed 4key silicone rubber Weight: Approximately 500 g (18 oz.) Display: Type: Microprocessor controlled aircore meter movement and backlit Alphanumeric x 16 LCD Update rate: <1s Contrast: User selected Relay annunciators: LEDs Electrical Power requirements: 1 to 4 VDC or 1 to 4 VAC, unregulated, 5060 Hz, 10 W max. Relay contacts ( sets): Mechanical SPDT contacts Max. voltage rating: 5 0 VDC, 5 15 VAC, or 50 VAC, (power factor = 1.0) Hysteresis: User adjustable Current output: 4 to 0 ma, nonisolated, internally powered, fully adjustable and reversible Update rate: <1s Max loop impedance: 50 Ω with a 1 V instrument supply voltage, 950 Ω with a 4 V instrument supply voltage Accuracy: ±0.1% of max. range Noise immunity: Noise emissions: Safety: EN5008 EN55011 EN Environmental Operating temp.: 10 to 55 C (14 to 11 F), 50 C (1 F) with optional rear cover Storage temp.: 15 to 80 C (5 to 176 F) Relative humidity: 0 to 95%, noncondensing Altitude: 4000 m max. Pollution degree: Agency Approvals CE, UL listed Manufactured under ISO 9001 Dimensions 96 mm (.8 in.) Panel Gasket Optional Splashproof Rear Cover # RELAY 1 RELAY Mounting Panel Mounting Clamp ( ea.) mm (.8 in.) 91 mm (.6 in.) 88 mm (.5 in.) ENTER Side View Signet Monitor Front View 76 mm ( in.) 96 mm (.8 in.) 88 mm (.5 in.) Signet 5700 ph/ Monitor Instructions (continued) 7
8 Dimensions 88 mm (.5 in.) UL LISTED 77CJ Signet Model # V 10 W mv Input Brown wire Iso. Gnd Blue wire T Green wire T White wire V Red Wire Relay V Black wire NO C NC NO C NC 40 ma Relay 1 Patent No. D 76,8 ph/ 88 mm (.5 in.) 9 x 9 mm (.6 x.6 in.) Rear View Panel cutout 9. Quick Reference Menu Parameters 9.1 VIEW Menu Setup Parameters (sec. 6.1) Menu Parameters Display Description Range Factory Default A. mv Raw sensor mv 1000 mv 000 mv 1000 mv 000 mv n/a n/a B. Current loop output 1 ma n/a C. Min Max meter and dial range 1000 mv 000 mv 1000 mv 1000 mv D. Relay 1 mode Relay 1 setpoint Low, High, or Pulse, mv Low 500 mv E. Relay mode Relay setpoint Low, High, or Pulse, mv High 500 mv F. Last calibration date G. EASY CAL procedure Two ph buffers saturated with Quinhydrone required: ph 4= 87 mv, ph 7= 64 mv, (sec. 6.) n/a 8 Signet 5700 ph/ Monitor Instructions
9 9. CALIBRATE Menu Setup Parameters (sec. 6.) Menu Parameters Display Description Range Factory Default A. Min max meter/dial range mv mv B. Manual electrode standard calibration mv n/a C. D. E. Manual electrode slope calibration Relay 1 mode Relay 1 setpoint Relay 1 hysteresis mv Low or High mv mv n/a Low 500 mv 10 mv Relay mode and setpoint displays repeat for relay setup F. Relay 1 mode Relay 1 range Pulse mv mv G. Relay 1 pulse rate pulses/minute 10 pulses/minute H. Last calibration date OPTIONS Menu Setup Parameters (sec. 6.4) Menu Parameters Display Description Range Factory Default A. Display contrast 0 to 5 B. 4 ma setpoint mv 1000 mv C. 0 ma setpoint mv 1000 mv D. 4 ma adjust.0 to 5.0 ma 4.00 ma E. 0 ma adjust 19 to 1 ma 0.00 ma Signet 5700 ph/ Monitor Instructions 9
10 10. Troubleshooting Display Problem Solution ph 5.0 C (Example ) ph electrode installed in preamplifi er Remove ph electrode and replace with recommended electrode (sec. ).. CHECK SENSOR?. Out Of Range Use Manual Cal Electrode not installed in preamplifi er mv input from preamplifi er out of range ( 1000 mv) Missing or wrong electrode identifi cation resistance applied to rear T and T terminals. ph buffers other than ph 4 and 7 saturated with Quinhydrone used during EASY CAL procedure; ph 4= 87 mv, ph 7= 64 mv 1) Verify preamplifi er/electrode input connections and electrode installation (sec. ). ) Disconnect Green and White preamplifi er wires from rear T and T terminals. Measure electrode identifi cation resistance across disconnected wires for 9.0 kω 10.8 kω. If measured resistance is out of range: Wrong electrode installed in preamplifi er (ph electrode = 5 C (77 F)) Faulty preamplifi er contacts, see preamplifi er manual for additional information Faulty electrode, see electrode manual for additonal information ) To verify instrument: Disconnect Green and White preamplifi er wires from rear T and T terminals, then install a 10 kω fi xed resistor across the terminals. Disconnect Brown and Blue preamplifi er wires from rear mv Input and Iso. Gnd terminals. Install shorting strap (jumper) across mv Input and Iso. Gnd terminals. Power instrument and verify 0 mv on the Input mv display (sec. 6.B). If 0 mv ±5 mv are not displayed, the instrument requires factory calibration. If Input mv are within ±5 mv, calibrate instrument Set Standard for 0 mv (sec. 6.B). If instrument displays correctly after calibration, instrument is ok; preamplifi er and/or electrode problem exists. Refer to preamplifi er and electrode manual for troubleshooting information. Use ph 4 and 7 buffers (ONLY) saturated with Quinhydrone for EASY CAL procedure (sec. 6.). The EASY CAL procedure simplifi es manual Standard and Slope calibration in the CALIBRATE Menu (sec. 6). The instrument can be calibrated using two buffers of known by this method (sec. 6.B, 6.C). If manual calibration is performed, the EASY CAL procedure is not required. 4.! Same Buffer Same ph buffer used for EASY CAL solution #1 and solution # Use different buffer saturated with Quinhydrone for EASY CAL procedure solution #1 and solution # (sec. 7). Do not calibrate both points with the same buffer! 5. : 000 mv Input: 000 mv (display stuck at 000 mv) Excessive mv input from preamplifi er Faulty preamplifi er, refer to preamplifi er manual for additional information To verify instrument input: Disconnect Brown and Blue preamplifi er wires and install shorting strap (jumper) across rear mv input and Iso Gnd terminals. Power instrument and verify 0 mv on display. If 0 mv is not displayed, calibrate Standard input for 0 mv (sec. 6.B) If instrument displays correctly after calibration, instrument is ok. If error condition persists, instrument requires factory service. 10 (continued) Signet 5700 ph/ Monitor Instructions
11 Display Problem Solution 6.! must be 000 or less (mv) input greater than 000 mv, or improperly entered during Standard or Slope calibration A) Use solution of known within mv range B) Enter mv value within 1000 mv 000 mv during standard or slope calibration (sec. 6.B, 6.C) 7. Slope Too Close To Standard 8. Value Must Be 1000 Or More 9. Value Must Be 000 Or Less 10. Value Must Be 00 Or Less 11. SETUP READ ERROR Press any Key slope calibration too close to standard calibration Relay, 4 ma, or 0 ma (mv) setpoint entered out of range Relay, 4 ma, or 0 ma (mv) setpoint entered out of range Relay pulse setting greater than 00 pulses per minute Power fault occurred while saving setup menu entry (CALIBRATE Menu sec. 6. only) Calibrate slope using an solution 10 mv from the standard solution (sec. 6.C). Enter setpoint within mv range (sec. 6., 6.4) Enter setpoint within mv range (sec. 6., 6.4) Enter relay pulse setting within 0 00 pulses per minute (sec. 6.F) Press any key to reload factory defaults then reprogram system setup parameters. 11. Maintenance Clean the instrument case and front panel with a soft cloth and a mild soap solution. Notes: Signet 5700 ph/ Monitor Instructions 11
12 George Fischer Signet, Inc. 401 Aerojet Avenue, El Monte, CA U.S.A. Tel. (66) Fax (66) For Worldwide Sales and Service, visit our website: Or call (in the U.S.): (800) Rev. K /06 English/ George Fischer Signet, Inc Printed in U.S.A. on recycled paper
SIGNET 5700 ph/orp Monitor - ORP Instructions
SIGNET 5700 ph/orp Monitor ORP Instructions ENGLISH 35700.0901 H (6/01) English CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections. Follow
More informationSignet 5700 ph/orp Monitor - ph Instructions
Signet 5700 ph/orp Monitor ph Instructions English *35700.0901* 35700.0901 Rev. K 3/06 English/pH CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output
More informationSignet 5900 Salinity Monitor
Signet 5900 Salinity Monitor *35900.0901* 35900.0901 Rev. E 3/06 English Contents 1. Power Connections 2. Compatible Sensor Wiring 3. 4 20 ma Current Output Connections 4. Relay Connections 5. Relay Operation
More information+GF+ SIGNET 5800CR Conductivity/Resistivity Monitor Instructions
GF SIGNET 5800CR Conductivity/Resistivity Monitor Instructions ENGLISH 35800CR.0901 D1/98 CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections.
More information+GF+ SIGNET ph/orp Transmitter Instructions
GF SIGNET 8750 ph/orp Transmitter Instructions ENGLISH 8750.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.
More informationSignet 5800CR Conductivity/Resistivity Monitor
Signet 5800CR Conductivity/Resistivity Monitor *35800CR.0901* 35800CR.0901 Rev. H 3/06 English Contents 1. Power Connections 2. Compatible Sensor Wiring 3. 4 20 ma Current Output Connections 4. Relay Connections
More informationContents. 1. Installation
ENGLISH 8750.090 C/0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Installation. Specifications. Electrical
More information+GF+ SIGNET 5900 Salinity Monitor Instructions
+GF+ SIGNET 5900 Salinity Monitor Instructions ENGLISH 35900.0901 (B8/99), English CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections.
More informationSignet Pressure Transmitter
Signet 850 Pressure English *850.090* 850.090 Rev. H /06 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More informationSIGNET ph/orp Transmitter
SIGNET 8750- ph/orp Transmitter ENGLISH 3-8750.090- D-4/03 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More information+GF+ SIGNET Conductivity/Resistivity Transmitter Instructions
GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions ENGLISH 8850.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to
More informationSignet Temperature Transmitter
Loop Output Output 7 Sensr IN (RED) Sensr V (BLACK) Signet 80 Temperature 80.090 English 80.090 Rev. J 0/ English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions
More informationSignet ph/orp Transmitter
Signet 87501 ph/orp *38750.0901* 38750.0901 Rev. H 06/10 English English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.
More informationSignet Flow Transmitter
Signet 8550 Flow *8550.090* English 8550.090 Rev. L 0/ English WARNING! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More informationSignet ph/orp Transmitter * *
Loop 7 6 5 Signet 8750- ph/orp *-8750.090-* CAUTION! Contents 0.0 ph 5.0C Signet ph/orp. Installation. Panel Installation gasket panel terminals mounting bracket 96 mm (.8 in.) 8050 8 mm (. in.) Optional
More informationUser s Guide PHCN ph/orp indicator ph Instructions. Shop online at. ENTER RELAY 1 RELAY 2
User s Guide 4 6 7 8 10 RELAY 1 RELAY 2 2 12 0 14 ENTER Shop online at www.omega.com email: info@omega.com PHCN5700 /ORP indicator Instructions OMEGAnet Online Service www.omega.com Internet email info@omega.com
More informationTable of Contents. 1 Introduction Introduction Front Panel Description Rear Panel Description 3
Chapter Page 1 Introduction 1 1.1 Introduction 1 1.2 Front Panel Description 2 1.3 Rear Panel Description 3 Table of Contents 2 Installation and Operation 4 2.1 Mounting Instructions 4 2.2 Power Connections
More informationSIGNET Level Transmitter
ENGLISH 850.090 A8/0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Description. Specifications. Panel
More informationFLS M9.06 PH/ORP MONITOR. safety instructions. packing List
FLS M9.06 PH/ORP MONITOR safety instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments
More informationFLS M9.06. ph/orp MONITOR
FLS M9.06 /ORP MONITOR Safety Instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments
More informationSignet 9900 Transmitter
Signet 9900 Transmitter Member of the SmartPro Family of Instruments Features Panel Mount Field Mount NEW! Modbus Module supports RS485 Serial Modbus Communications Multiple sensor types supported with
More informationFLOW MONITOR & TRANSMITTER
FLS M9.02 FLOW MONITOR & TRANSMITTER SAFETY INSTRUCTIONS General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other
More informationSignet 2750 DryLoc ph/orp Sensor Electronics
Signet 750 DryLoc ph/orp Sensor Electronics *750.090* 750.090 Rev. G /05 English WARNING! SAFETY INSTRUCTIONS. Depressurize and vent system prior to installation or removal.. Confi rm chemical compatibility
More information+GF+ SIGNET 9050CR Conductivity/Resistivity Controller
+GF+ SIGNET 9050CR Conductivity/Resistivity Controller Instruction Manual GEORGE FISCHER +GF+ Important Safety Information! CAUTION: Remove AC power to unit prior to wiring input and output connections.
More informationX Year mfg
4to20 ma Output S L Loop PWR Loop2 PWR N/C S L DATA 7 8 9 6,*(7 S L Output 8 S L SIGNET 8900 MultiParameter Controller Supplemental Manual *8900.090SUP* 8900.090SUP Rev. A 2/04 (English). Description This
More informationUDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS
UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS EN0I-6041 12/99 PRODUCT SPECIFICATION SHEET OVERVIEW The UDC 1000 and UDC 1500 are microprocessor-based 1/16 DIN and 1/8 DIN controllers
More informationMYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL
MYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM MYR QLC MANUAL 013114 2 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features WIRING
More informationMini Digital Multimeter
User Manual Mini Digital Multimeter Model MN15A Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the Extech MN15A MultiMeter. The MN15A offers
More informationAutoranging True RMS Multimeter User Manual
Autoranging True RMS Multimeter User Manual Please read this manual before switching the unit on. Important safety information inside. Contents Page 1. Safety Information... 4 2. Safety Symbols... 5 3.
More informationMini Digital Multimeter
User's Guide Mini Digital Multimeter Model MN15 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech MN15/MN16 Series
More informationHigh Voltage DC Meter
High Voltage DC Meter Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications
More informationMYRIAD LC1 LEVEL CONTROLLER INSTRUCTION MANUAL
MYRIAD LC1 LEVEL CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM MYR LC1 MANUAL 062114 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features Dimensions WIRING
More informationHIGH-VOLTAGE DC METER
HIGH-VOLTAGE DC METER Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications
More informationAnalyzer/Controller. ph/orp HART. Model 54e ph/orp ESSENTIAL INSTRUCTIONS WARNINGS RISK OF ELECTRICAL SHOCK
Instruction Sheet PN 51A-54epH/rev J October 2010 ph/orp HART Model 54e ph/orp Analyzer/Controller For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. ESSENTIAL
More information1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882
1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 This manual is intended for use in support of installation, commissioning and configuration of the 1/32-DIN Temperature
More information700 SERIES DPC DUPLEX PUMP CONTROLLER INSTRUCTION MANUAL
1 700 SERIES DPC DUPLEX PUMP CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM 700 DPC MANUAL 072114 2 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features WIRING
More informationOPERATING INSTRUCTION
OPERATING INSTRUCTION AUTORANGING MULTIMETER MAX Ω F C 10A MAX every 15 min. COM V SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation
More information5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL Fruitland Ave Los Angeles, CA SST7000 SST7100. Speed Switch / Transmitter
5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL 33309 3211 Fruitland Ave Los Angeles, CA 90058 SST7000 SST7100 Speed Switch / Transmitter Installation and Operation Manual Rev. C P/N145F-13112 PCO 00009270
More informationTempco Instruction Manual
Tempco Instruction Manual 1/16 DIN Solid State Temperature Controller Relay Output Solid State Output For Heating Model Numbers: TEC-901, TEC-902, TEC-905 Temperature controls in this series are designed
More informationAdapter Kit - QSI Electronics to TM Meter Installation Instructions
Adapter Kit - QSI Electronics to TM Meter Installation Instructions These installation instructions cover the adapter kit installation of the QSI1, QSI2 and QSI3 versions of communications electronics
More informationConductivity Transmitters
Conductivity Transmitters CDTX-90 Series Starts at $ 576 Electronics Only Display in µs, ms, kω, MΩ, ppm (TDS) Simulate Function Programmable Temperature Compensation Relay Options Dual Output Option Allows
More informationTABLE OF CONTENTS INTRODUCTION. 3. Analog Input Analog Output Digital Input Digital Output OPERATIONAL DESCRIPITON.. 7 PROGRAMMING AND INITIAL SETUP.
DIVERSIFIED HEAT TRANSFER SERIES 700 STEAM GENERATOR CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE AT SIGMACONTROLS.COM SERIES 700 DHT STEAM GENERATOR MANUAL 042514 2 TABLE OF CONTENTS INTRODUCTION.
More informationUser Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at
User Manual 400Amp AC Clamp Meter + NCV Model MA430 Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of this Extech MA430 Clamp Meter. This
More information4801P ph/orp controller
4801P ph/orp controller Introduction: Thank you for choosing this ph/orp controller. This 1/16 DIN controller offers dual backlight LCD display, high accuracy, LED indicators, two control relays, and automatic
More informationUser's Guide. Extech AM A AC Analog Clamp Meter
User's Guide Extech AM300 300A AC Analog Clamp Meter Introduction Congratulations on your purchase of the Extech AM300 Analog Clamp Meter. This device measure AC Voltage and Current, DC Voltage, and Resistance.
More informationUser Guide True RMS Multimeter Extech EX205T
User Guide Extech EX205T True RMS Digital Multimeter Extech EX210T True RMS Digital Multimeter IR True RMS Multimeter Extech EX205T Introduction Thank you for selecting the Extech EX205T True RMS Auto-ranging
More informationS-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter
S-14 Compact Digital Multimeter S-14 Compact Digital Multimeter SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation at this meter
More informationOperating Manual. Dynisco /8 DIN Indicator Concise Product Manual
Dynisco 1490 1/8 DIN Indicator Concise Product Manual 59476-9 Operating Manual -1- CAUTION: Installation should be only performed by technically competent personnel. Local Regulations regarding electrical
More informationOPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS
OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS GB Contents Page 1 Introduction 4 2 Safety 5 3 Use 6 4 Checking on Delivered Equipment 6 5 General Specification and Dimensions 7 6 Positioning 10 7 Operating
More information700 SERIES TPC TRIPLEX PUMP CONTROLLER INSTRUCTION MANUAL
1 700 SERIES TPC TRIPLEX PUMP CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM 700SERIESTPCMANUAL071414 2 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features
More informationMODEL: M5DY SEN TRONIC AG CURRENT LOOP SUPPLY. [2] POWER INPUT AC Power M: V AC (Operational voltage range V, Hz)
Super-mini Terminal Block Signal Conditioners M5-UNIT CURRENT LOOP SUPPLY Functions & Features Powering a 0 ma DC current loop Applicable to smart transmitters Isolation between the input and output Fast
More informationMini Digital Multimeter
User's Guide Mini Digital Multimeter Model MN15 Introduction Congratulations on your purchase of the Extech MN15 MultiMeter. The MN15 offers AC/DC Voltage, AC/DC Current, Resistance, Diode, and Continuity
More informationOperating instructions. Speed monitor D / / 2014
Operating instructions Speed monitor D200 80005257 / 00 05 / 2014 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5 2.2 Target group...5
More informationOwner's Manual. True RMS Multimeter. Model No Safety Operation Maintenance Español
Owner's Manual True RMS Multimeter Model No. 82023 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance Español
More informationINSTALLATION DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER. Before installation:
DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER The DKM-409 is a precision instrument designed for displaying various AC parameters in 3-phase distribution panels. Thanks to its isolated
More informationSignet BC Batch Controller System
Signet 9900-1BC Batch Controller System *3-9900-1BC.090* 3-9900-1BC.090-1 Rev. A1 03/13 English Product Manual Look for the Quick Start icon to quickly set up your new 9900-1BC. Safety WARNING! Panel Mount
More informationDigital Proportional Valve Amplifier
Digital Proportional Valve Amplifier The Amplifier is a compact, low profile DIN mounted controller for use with proportional solenoid valves. The provides current to a valve coil in proportion to an input
More informationRXTH DUAL ROOM SENSOR / SWITCH
DUAL ROOM SENSOR / SWITCH FOR TEMPERATURE AND RELATIVE HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE
More informationUser's Guide. MiniTec TM Series Model MN25 MultiMeter
User's Guide MiniTec TM Series Model MN25 MultiMeter Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a
More informationSignet Conductivity/Resistivity Electrodes
Signet 2819-2823 Conductivity/Resistivity Electrodes 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - ¾ in. NPT Polypro - Tri-clamp 1-1½ in., 2 - Opt. ½ in. NPT 316 SS 316 SS or
More informationJUMO ecotrans ph 03 Microprocessor transmitter/ switching device for ph/redox voltage and temperature
Page 1/7 JUMO ecotrans 03 Microprocessor transmitter/ switching device for /Redox voltage and temperature with a 2-line LCD for mounting on a 35 mm DIN rail Brief description Depending on the configuration,
More informationModel: Pro93 TRUE RMS LEAKAGE CURRENT TESTER
Model: Pro93 TRUE RMS LEAKAGE CURRENT TESTER CONTENTS TITLE PAGE I. Safety Information.......1 Environmental Conditions... 1 Explanation of Symbols..... 1 II. Specification..... 2 General Specification...
More informationDM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER
DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: This meter is
More informationSHARK MULTI-PARAMETER CONTROLLER & ANALYZER USER S MANUAL
SHARK MULTI-PARAMETER CONTROLLER & ANALYZER USER S MANUAL Rev 3 AquaMetrix Inc. 1245 Maple Hill Ct., Unit 7 Newmarket, ON Canada, L3Y 9E8 Tel: (800) 742-1413 (905) 954-0841 Fax: (905) 954-0415 www.aquametrix.com
More informationLCD LCDS LCDRI CONTROLLER OPERATING MANUAL
This manual contains important safety information about installation and use of this equipment. Ignoring this information could result in injuries or damages. It is strictly forbidden to use this equipment
More informationINDICATOR FOR PROCESS CONTROL. Love Controls. LCI108J-0x INSTRUCTION MANUAL. EDITION: January Valid for models with software version P-1
Love Controls INDICATOR FOR PROCESS CONTROL LCI108-0x LCI108J-0x INSTRUCTION MANUAL EDITION: January 2001 Valid for models with software version P-1 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This
More informationUser's Guide. Digital Multimeter. Model MN42
User's Guide Digital Multimeter Model MN42 Introduction Congratulations on your purchase of the Extech MN42 MultiMeter. The MN42 offers AC/DC Voltage, DC Current, and Resistance testing. Proper use and
More informationEA401 SERIES PULSE INPUT FLOW RATEMETER/TOTALIZER INSTALLATION & OPERATING INSTRUCTIONS West Stetson Avenue Hemet, CA USA
EA401 SERIES PULSE INPUT FLOW RATEMETER/TOTALIZER INSTALLATION & OPERATING INSTRUCTIONS 3255 West Stetson Avenue Hemet, CA 92545-7799 USA Lit# 24508-67 99773 Rev. 1.1 02/25/0 Tel: 951-652-6811 Fax: 951-652-3078
More information407B 1/16 DIN Multi-Mode Timer
< Marsh Bellofram Group of Companies < Automatic Timing & Controls < Timer < 407B Series Downloadable Files: 407B Series 407B 1/16 DIN Multi-Mode Timer Selectable On-Delay/Off-Delay/Interval Timing Modes
More informationINSTALLATION MANUAL. LC 200 Electronic Overload Guard. Software versione PW0501 R 0.3
INSTALLATION MANUAL LC 200 Electronic Overload Guard Software versione PW0501 R 0.3 CONTENTS MAIN FEATURES LC 200 TECHNICAL FEATURES Page 2 SYMBOLS Page 3 WARNINGS Page 3 IDENTIFICATION DATA PLATE Page
More informationCDD Carbon Dioxide Transmitter
Introduction The OSA CO2 transmitter uses Infrared Technology to monitor CO2 levels within a range of 0 2000 ppm and outputs a linear 4-20 ma or 0-5/0-10 Vdc signal. The enclosure is designed to operate
More informationRTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions
Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA STANDARDS OPERATIONAL DIAGRAMS WIRING AND CONNECTIONS
More informationInstruction Manual. M Pump Motor Controller. For file reference, please record the following data:
Instruction Manual M Pump Motor Controller For file reference, please record the following data: Model No: Serial No: Installation Date: Installation Location: When ordering replacement parts for your
More informationSL2000E. Status Output. Electronic Code Lock. Features. Input. Exit Button Input. Introduction. Door Contact Input. Functional Description.
SL2000E Electronic Code Lock Features Door Relay output Status transistor output Aux transistor output Door Contact input Exit Button input INSTALLER code for programming MASTER code for arming/disarming
More informationNKQB24-1, NKQX24-1 On/Off, Electronic Fail-Safe, Direct Coupled, 24V
NKQB24-1, NKQX24-1 On/Off, Electronic Fail-Safe, Direct Coupled, 24V orque min. 54 in-lb for control damper surfaces up to 12 sq ft. Application For on/off, fail-safe control of dampers in HVAC systems.
More informationLDCL DIGITAL CONTROLLER OPERATING MANUAL
This manual contains important safety informations about installation and use of this equipment. Ignoring this informations could result in injuries or damages. It is strictly forbidden to use this equipment
More informationOPERATION MANUAL MODEL TT-LP
TURTLE TOUGH OPERATION MANUAL MODEL TT-LP 2-Wire Isolated ph/orp Transmitter turtletough.com.au CONTENT INITIAL INSPECTION 3 INTRODUCTION 3 ASSEMBLY 4 PREPARATION 5 Connection Diagram of Turtle Tough Sensors
More informationOperating instructions. Switching amplifier DN0210 DN / / 2015
Operating instructions Switching amplifier DN0210 DN0220 UK 80011079 / 00 01 / 2015 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5
More informationR5050. Model. Instruction Manual. TRMS AC/DC Clamp Meter. reedinstruments. www. com
Model R5050 TRMS AC/DC Clamp Meter Instruction Manual reedinstruments com Table of Contents Safety... 3 Features... 4 Specifications...4-6 Instrument Description...7-8 Measurement Procedures...9-12 Battery
More informationRТTH DUAL ROOM SWITCH
DUAL ROOM SWITCH FOR TEMPERATURE AND RELATIVE HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA
More informationAutoRanging Digital MultiMeter
Owner's Manual AutoRanging Digital MultiMeter Model No. 82175 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance
More informationUser's Guide. 800 Amp Clamp Meters. EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter
User's Guide 800 Amp Clamp Meters EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter Introduction Congratulations on your purchase of the EX710, EX720, or EX730 Clamp DMM.
More informationT M. IRDH375 Series. Digital Ground Fault Monitor / Ground Detector For Ungrounded (Floating) AC/DC Systems. Technical Bulletin NAE / 09.
T M 1 IRDH375 Series Digital Ground Fault Monitor / Ground Detector For Ungrounded (Floating) AC/DC Systems Technical Bulletin NAE101051 / 09.013 IRDH375 Series Digital ground fault monitor / ground detector
More informationPR602LCD v1.0 Installer Manual Rev. A
PR602LCD v1.0 Installer Manual Rev. A Introduction This document contents minimum information required for electrical connections and installation of the PR602LCD access controller in Roger Access Control
More informationM A C 3 Wind Speed Alarm & Controller
M A C 3 Wind Speed Alarm & Controller Installation Instructions Thank you for purchasing the MAC3 wind speed alarm and controller. This manual is designed to lead you through a step-by-step process to
More informationABB MEASUREMENT & ANALYTICS DATA SHEET. TB84PH ph /ORP / pion transmitter
ABB MEASUREMENT & ANALYTICS DATA SHEET TB84PH ph /ORP / pion transmitter 2 TB 8 4PH P H /ORP / P I ON TR AN S MIT TE R DS/ TB8 4PH-E N RE V. D Measurement made easy Hazardous area rated transmitter with
More informationNKQB24-1, NKQX24-1 On/Off, Fail-Safe, Operation, Direct Coupled, 24V
NKQB24-1, NKQX24-1 On/Off, Fail-Safe, Operation, Direct Coupled, 24V orque min. 54 in-lb for control damper surfaces up to 12 sq ft. Application For proportional modulation of dampers in HVAC systems.
More informationColor Mark Photoelectric Sensor E3S-VS
Sensing Supply voltage Output 12 to 80 ma, with 1.5 to 4 ma constant current source; 1.2 cm, 5 cm 100 ma, Color Mark Photoelectric Sensor Small Color Mark Sensor With Built-In DC Amplifier Fast, 1 ms maximum
More informationI-7520: RS-232 to RS-485. Input: RS-232 protocol Output:RS-485/RS-422 Speed: Self Tuner 'inside, auto switching baud rate, 300~ BPS
I-7520 RS-232 to RS-485 Converter I-7510 RS-485 Repeater RS-232 RS-485 I-7000 SERIES Common Specifications Isolation voltage: 3000VDC Speed: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115000 Dual watchdog
More informationMRC 7000 Proven Recording Reliability that Became the Industry Standard!
Data Sheet Recorder//Profiler DESCRIPTION The MRC7000 Recorder/ is a microprocessor based circular chart instrument capable of measuring, displaying, recording and controlling up to two process variables
More informationInstruction Manual. Liquitron DP5000 Series ph Controller. For file reference, please record the following data:
Instruction Manual Liquitron DP5000 Series ph Controller For file reference, please record the following data: Model No: Serial No: Installation Date: Installation Location: When ordering replacement parts
More informationUser's Guide. 1/16 th DIN ph Controller. Model 48PH2. Introduction
User's Guide 1/16 th DIN ph Controller Model 48PH2 Introduction Thank you for selecting the Extech Model 48PH2 controller. This 1/16 th DIN-size controller offers dual display, high resolution measurements,
More informationAdapter Kit - QSI Electronics to G2 Meter Installation Instructions
Adapter Kit - QSI Electronics to G2 Meter Installation Instructions These installation instructions cover the adapter kit installation of the QSI1, QSI2 and QSI3 versions of communications electronics
More informationPGR-3200 MANUAL INSULATION MONITOR
Tel: +1-800-832-3873 E-mail: techline@littelfuse.com www.littelfuse.com PGR-3200 MANUAL INSULATION MONITOR REVISION 3-E-040918 Copyright 2018 by Littelfuse, Inc. All rights reserved. Document Number: PM-1025-EN
More information7511 / / / 711 / / 731 / Preset 27
22 Timers Timers Electronic 23 7511 / 3410 23 6320 / 720-6300 24 ElectRo-MeCH 25 710 / 711 / 720 25 722 / 731 / 732 26 Preset 27 7932 27 Electronic 23 7511 Series 8 digit self powered LCD electronic timer
More informationASTAT XB/XBm Remote Operator
ASTAT XB/XBm Remote Operator User Manual 1 Introduction 1.1 Important User Information Observe all necessary safety precautions when controlling the soft starter remotely. Alert personnel that machinery
More informationTX ph/mv Transmitter/Controller
TX3000 - ph/mv Transmitter/Controller PRODUCT INSTRUCTION MANUAL TX3000 ph/mv Transmitter ESSENTIAL INSTRUCTIONS READ THIS BEFE USING YOUR TX3000 ph/mv TRANSMITTER! Thank you for choosing the TX3000 ph/mv
More informationSignet 2552 Metal Magmeter Flow Sensors
Signet 2552 Metal Magmeter Flow s Features Test certificate included Award winning hot-tap magnetic flow sensor up to DN2550 (102 in.) Patented Magmeter technology* Operating range 0.05 to 10 m/s (0.15
More informationCDD4 Series Room CO2 Transmitter Installation Instructions
CDD4 Series Room CO2 Transmitter Installation Instructions Introduction The CO2 transmitter uses Infrared Technology to monitor CO2 levels and outputs a linear 4-20 ma or 0-5/0-10 Vdc signal. Options include
More informationOWNER S MANUAL 9908-TE. HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER
OWNER S MANUAL 9908-TE HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER IMPORTANT! Read and understand this manual before using the instrument. Failure to understand and comply with
More information