STANDARD OPERATING PROCEDURE AA SERIES AA-160 PRECISION BALANCE

Size: px
Start display at page:

Download "STANDARD OPERATING PROCEDURE AA SERIES AA-160 PRECISION BALANCE"

Transcription

1 Table of Contents Page 1. Scope 2 2. Objective 2 3. Definitions 2 4. Responsibility 3 5. Procedure 3 6. Precautions References Appendices Revision History 12 Original Authorised Copy for: Reading Copy Written by: Reviewed by: Approved by: Signature/Date: Signature/Date: Signature/Date: Page 1 of 14

2 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the AA Series AA-160 Precision Balance in the laboratories of the Pharmacy Department, University of Malta. 2. Objective To describe the procedure for the installation, configuration, operation, calibration, maintenance and troubleshooting of the AA Series AA-160 Precision Balance. 3. Definitions 3.1. Adjustable Feet: Circular screws situated at the bottom of the balance that are manipulated to level the balance out Auto CAL Key: To initiate automatic calibration using internal weights Calibration: A procedure to check a measuring instrument against an accurate standard to determine any deviations and correct for errors Calibration Variance: The corrections made to the calibration values to compensate for differences between the actual weight being used and the desired value Configuration Parameters Set-Up Menu: Allows the user to adapt the balance to different weighing situations (see /A1) Message Display: Part of the dual LCD display that uses text to provide additional information Number Display: Part of the dual LCD display that continuously shows the weighing results On/Off Key: To turn the display ON or OFF or to enter the desired option in the Set-Up menu Print Key: To transmit sample weight to printer or computer. Is also used to scroll through the different options in the Set-Up Print menu. Page 2 of 14

3 3.10. Select Key: To scroll through the configuration parameters of a particular option in the Configuration Parameters Set-Up menu Set-Up Key: To access the Configuration Parameters Set-Up menu and then to cycle through its options. Is also used to access the Set-Up Print menu Tare: Allows the balance to display the weight of the sample only and not the combined weight of the sample and its container Tare Key: To set the Numeric Display to zero. Is also used to exit Set- Up menu and return the balance to normal weighing mode U symbol: An unstable indicator that is automatically displayed whenever the balance is not stable and then disappears when the balance becomes stable Weighing Pan: The area on top of which the material to be weighed is placed. 4. Responsibility 4.1. The members of the Department of Pharmacy (staff and students) are responsible for following this SOP The designated Laboratory Officer or Laboratory Assistant is responsible for ensuring that this SOP is followed. 5. Procedure 5.1. Installation Place the balance on a stable surface, away from any vibrations Turn the adjustable feet at the bottom of the balance clockwise or anticlockwise until the air bubble in the spirit level is in the centre of the marked circle Insert power supply connector in the plug on the rear of the balance and plug the power supply into the mains supply Press the On/Off key to switch on the balance Wait until the start-up message and then are displayed before starting to operate the balance. Page 3 of 14

4 5.2. Configuration Press the Set-Up key briefly but repeatedly until the desired option is displayed (see /A1 Configuration Parameters Set-Up Table) Press the On/Off key to enter into the desired option Press the SELECT key briefly but repeatedly until the desired configuration parameter is displayed Press the TARE key to save changes and return the balance to normal weighing mode Operation Place the material that needs to be weighed on the weighing pan Close the draft shield Wait until the [U] symbol disappears Read off the value from the display. If using the tare function: Place an empty weighing container on the weighing pan Close the draft shield Wait until the [U] symbol disappears Press the Tare key Open draft shield and place the material to be weighed in the weighing container Close draft shield Wait until the [U] symbol disappears Read off the value from the display Calibration Auto Calibration Press the On/Off Key to switch on the balance Wait until the [U] symbol disappears from the message display Press the TARE key Press the AUTO CAL key. Page 4 of 14

5 Wait until the message display changes from [CAL] to [CAL-100-*] to [TARING] and finally to [GRAMS] to indicate that the auto-calibration process is complete External Calibration 5.5. Maintenance Press the On/Off Key to switch on the balance Wait until the [U] symbol disappears from the message display Press the TARE key Place the standard calibration weight on the weighing pan Wait for the [U] symbol to disappear and the calibration weight value together with [GRAM] to be displayed Press the AUTO CAL key Wait until [XXX.0000] and [CAL-XXX-] are displayed Remove the standard weight Perform the calibration procedure once again if it is to be repeated with a different standard weight See /A2 Calibration Variance Table if calibration variance needs to be modified Disconnect the balance from its power supply before attempting to clean and leave it in its normal operating position Use a clean fine brush to remove any surrounding powder Use a moistened cloth with a mild cleaning agent to clean and remove any spilt material Ensure that the air bubble in the spirit level on the front end of the scale is in the centre of the marked circle. Page 5 of 14

6 5.6. Troubleshooting Fault Possible Cause/s Possible Solution/s - Display previously turned - Press the On/Off Key off by the On/Off key Both displays are blank - Power cord not connected properly - Power cut or an improper voltage supply present - Attach the power cord accordingly - Check if a proper power supply is present - Temporary fault - Disconnect power cord, wait for at least 5 seconds and reconnect [OVER] is displayed Exceeded weighing capacity Reduce the weight - Pan is not properly - Install pan properly installed [UNDER] is displayed - Pan is obstructed - Move balance to a better [GRAM U] continues to be displayed [NO CAL] is displayed during calibration process - Air currents present around balance - Sample not stationary Balance capacity is exceeded location - Move balance to a better location or close all draft shields - Ensure sample is made stationary Use a 50g, 100g or 150g standard calibration weight to complete calibration Page 6 of 14

7 5.7. Flow Charts Installation Start Place balance on a stable surface, away from any vibrations Turn adjustable feet until air bubble in spirit level is in the centre of the marked circle Insert power supply connector in the plug on rear of balance and plug power supply into mains supply Press On/Off key to switch on balance Wait until start-up message and are displayed before use End Page 7 of 14

8 Configuration Start Press Set-Up key briefly but repeatedly until desired option is displayed (see /A1) Press On/Off key to enter desired option Press SELECT key briefly but repeatedly until desired configuration parameter is displayed Press TARE key to save changes and return balance to normal weighing mode End Page 8 of 14

9 Operation Start Using tare function No Place material on weighing pan Yes Place empty weighing container on weighing pan Close draft shield Wait until [U] symbol disappears Press Tare Key Open draft shield and place material to be weighed in weighing container Close draft shield Wait until [U] symbol disappears Read off value from display End Page 9 of 14

10 Calibration Start Press On/Off key to switch on balance Wait until [U] symbol disappears Press TARE Key Yes Performing external calibration Place standard weight on weighing pan No Perform auto calibration by pressing the AUTO CAL key Wait until message display changes from [CAL] to [CAL-100-*-] to [TARING] and finally to [GRAMS] Wait for [U] to disappear and calibration weight value together with [GRAM] to be displayed Press AUTO CAL key Wait until [XXX.000] and [CAL-XXX-] are displayed Remove standard weight Repeat calibration with different standard weight No End Yes Page 10 of 14

11 Maintenance Start Balance needs cleaning No Yes Disconnect balance from power supply Leave balance in its normal operating position and do not tilt sideways Any surrounding powder No Yes Remove using a clean fine brush Any spilt material No Yes Remove using a moistened cloth with a mild cleaning agent Yes Air bubble in spirit level is in centre of marked circle No Manipulate adjustable feet until this is achieved End Page 11 of 14

12 6. Precautions 6.1. Do not install balance near extremes of heat and cold Protect balance from dust and air draughts Keep weighing pan clean and dry to prevent inaccuracies Do not move balance during weighing Ensure that the maximum tolerated weight of 160g is not exceeded during weighing procedures and also when using the tare function Do not leave any items on the balance when this is not in use Only use the balance if the spirit level is correct Adjust the air bubble in the spirit level each time the balance is moved to a different location Avoid static charge buildup of the material to be weighed and of the weighing container When transferring balance to a significantly warmer environment, allow it to acclimatise for approximately 2 hours at room temperature after it has been disconnected from the mains Do not unplug balance as this is designed to be continuously plugged in Ensure that an audible beep is heard whenever one of the keys is pressed if the beeper option is activated. If this beep is not heard, press the key again until the beep is heard. 7. References Denver Instrument Company. AA Operating Instructions. Series Electronic Analytical Balances 8. Appendices /A1 Configuration Parameters Set-Up Table /A2 Calibration Variance Table 9. Revision History Version Number Amendments/ Reasons for change 01 Initial Release Page 12 of 14

13 /A1 Configuration Parameters Set-Up Table Option Function CURRENT DEFAULT FILTER AUTO ZERO BEEPER To display the current selected configuration parameters To reset the balance to the default factory configuration parameters To choose the desired filter settings from [FT FAST] in ideal conditions, [FT NORMAL] in standard laboratory conditions and [FT SLOW] in areas with drafts and vibrations To choose to automatically compensate for zero drift [A-Z ON] or to turn off this correction feature [A-Z OFF] To choose between short beeps [BP SHORT], long beeps [BP LONG] or else to turn off the beeper function [BP OFF] Page 13 of 14

14 /A2 Calibration Variance Table Procedure Message Display Press On/Off key to switch on balance Press AUTO CAL key Press SET UP key repeatedly until calibration value to be modified is displayed for e.g. for 100g CAL CAL-100- Press On/Off key V Press SELECT key repeatedly until desired + or sign is displayed V or V 00.0 Press On/Off key Press SELECT key repeatedly to change first digit (between 0 and 1) Press On/Off key when desired digit is displayed Press SELECT key repeatedly to change next digit (between 0 and 9) Press On/Off key when desired digit is displayed V +/- x0.0 V +/- 1x.0 Repeat the last 4 steps for subsequent digits Page 14 of 14

STANDARD OPERATING PROCEDURE ACB SERIES ACB-600H PRECISION BALANCE

STANDARD OPERATING PROCEDURE ACB SERIES ACB-600H PRECISION BALANCE Ref.. Table of Contents Page 1. Scope 2 2. Objective 2 3. Definitions 2 4. Responsibility 3 5. Procedure 3 6. Precautions 14 7. References 14 8. Appendices 14 9. Revision History 14 Original Authorised

More information

PX Series Balances. Quick Start Guide. Please download the user manual from

PX Series Balances. Quick Start Guide. Please download the user manual from Scan QR Code with your WebCam for downloading the user instruction manual. PX Series Balances Quick Start Guide Please download the user manual from www.ohaus.com. EN-1 1. INSTALLATION 1.1 Select the

More information

Operating instruction Platform scale

Operating instruction Platform scale KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating instruction Platform scale KERN ECB Version

More information

Denver Instrument Company. M-Series. Analytical Balances. Operation Manual Rev. F

Denver Instrument Company. M-Series. Analytical Balances. Operation Manual Rev. F Denver Instrument Company M-Series Analytical Balances Operation Manual 700714.1 Rev. F Table Of Contents Disclaimer Notice.................................... 4 Introduction.......................................

More information

AL-Series INSTRUCTION MANUAL AL-311 AL-201S. American Weigh Scales, Inc. Analytical Balances American Weigh Scales, Inc. pcs% kgct.

AL-Series INSTRUCTION MANUAL AL-311 AL-201S. American Weigh Scales, Inc. Analytical Balances American Weigh Scales, Inc. pcs% kgct. American Weigh Scales, Inc. AL-Series Analytical Balances INSTRUCTION MANUAL Max = 310g d=1mg e=10d STANDBY pcs% kgct ON/OFF MODE CAL UNIT PRINT AL-311 AL-201S Max = 210g d=0.1mg e=10d kgct STANDBY pcs%

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Sartorius Laboratory. L 610, L2200S, L2200P Electronic Precision Balances Installation and Operating Instructions 1

More information

1.0 Description. 2.0 Unpacking. 3.0 Installation

1.0 Description. 2.0 Unpacking. 3.0 Installation ES-H, ES-HA Series Precision Balance Thank you for purchasing the Model ES-H and ES-HA precision balance. Please read all operating instructions carefully before using and note the following items to ensure

More information

Precision Balance W3200 Series Operating Manual

Precision Balance W3200 Series Operating Manual Precision Balance W3200 Series Operating Manual Version 9-19-16 1 2 CONTENTS 1. INTRODUCTION. 4 2. INSTALLATION... 4 2-1. Unpacking.... 5 2-2. Specifications.... 5 2-3. Environmental Requirements... 5

More information

STANDARD OPERATING PROCEDURE TENSOVAL COMFORT

STANDARD OPERATING PROCEDURE TENSOVAL COMFORT Page 1 of 13 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the Tensoval Comfort in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy Department, University

More information

EHC-WF. User/Technical Manual. Contents subject to change without notice

EHC-WF. User/Technical Manual. Contents subject to change without notice EHC-WF User/Technical Manual Contents subject to change without notice Version 1.0 7/2017 TABLE OF CONTENTS 1. INTRODUCTION... 1 General and Safety Information... 1 Specifications... 1 2. Unpacking and

More information

Oxford E Series. Instruction Manual

Oxford E Series. Instruction Manual Oxford E Series Instruction Manual E Series User Manual Contents. Introduction................................................................... Unpacking...................................................................

More information

Adam Equipment ABL. Software revision D

Adam Equipment ABL. Software revision D Adam Equipment ABL Software revision 8.01.170127-2285D Adam Equipment Company 2018 Easy Reference: Model name of the scale: Serial number of the unit: Software revision number (Displayed when power is

More information

Sartorius Research. R 160 P Electronic Semi-Microbalance

Sartorius Research. R 160 P Electronic Semi-Microbalance Sartorius Research. R 160 P Electronic Semi-Microbalance Installation and Operating Instructions 1 Power receptacle, fuse, 8 Level indicator voltage selector 9 ON/OFF key 2 Menu access switch 10 Print

More information

CONTENTS. PRECAUTIONS Introduction Installation... 7

CONTENTS. PRECAUTIONS Introduction Installation... 7 2 CONTENTS PRECAUTIONS... 4 1. Introduction... 6 2. Installation... 7 2.1 Unpacking... 7 2.2 Illustration of the installation for plastic shied... 7 2.3 Selecting the location... 8 2.4 Leveling the scale...

More information

IDS-801 Series Compact Bench Counting Scale. User Manual

IDS-801 Series Compact Bench Counting Scale. User Manual IDS-801 Series Compact Bench Counting Scale User Manual 1. INTRODUCTION... 3 1.1. Product Description... 3 1.2. General Features... 3 1.3. Safety Precautions... 3 2. INSTALLATION... 4 2.1. Unpacking...

More information

Table of Contents. Laboratory Balances. Instruction Manual - TORBAL AGC Series

Table of Contents. Laboratory Balances. Instruction Manual - TORBAL AGC Series Laboratory Balances Table of Contents Cautionary Notes and Precautions...3 1. Specification... 4 2. Keys and Display Indicators... 6 3. Commands and abbreviations...7 4. Parts Description...8 5. Unpacking

More information

Sartorius handy. H 160, H160 P. Electronic Precision Balance Installation and Operating Instruction

Sartorius handy. H 160, H160 P. Electronic Precision Balance Installation and Operating Instruction Sartorius handy. H 160, H160 P. Electronic Precision Balance Installation and Operating Instruction 2 1 Draft shield lid 2 Glass cylinder of the round draft shield 3 Weighing pan 4 Tare compensating weight,

More information

KW serial. User s guide. Bench Scales UGKW-E0302

KW serial. User s guide. Bench Scales UGKW-E0302 KW serial Bench Scales User s guide UGKW-E0302 KW serial bench scale user s manual Table of Contents SECTION 1 INTRODUCTION 1 SECTION 2 SPECIFICATIONS.. 2 SECTION 3 INSTALLATION.3 SECTION 4 KEY DESCRIPTIONS

More information

YJ Carat and Gold Scales

YJ Carat and Gold Scales 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com YJ Carat and Gold Scales Instruction Manual EN-2 YJ Series YJ Scale Mark Standard

More information

MWP SERIES. High Precision

MWP SERIES. High Precision MWP SERIES High Precision 1. INTRODUCTION------------------------------------------------1 2. INSTALLATION---------------------------------------------------2 2.1 Unpacking---------------------------------------------------------------

More information

STANDARD OPERATING PROCEDURE ACCU-CHEK GO

STANDARD OPERATING PROCEDURE ACCU-CHEK GO Page 1 of 21 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the ACCU-CHEK Go in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy Department, University

More information

Kilotech. KRS SWP Series. Counter-top weighing scale. NTEP (USA) and Measurement Canada approved.

Kilotech. KRS SWP Series. Counter-top weighing scale. NTEP (USA) and Measurement Canada approved. Kilotech KRS SWP Series Counter-top weighing scale NTEP (USA) and Measurement Canada approved www.kilotech.com KILOTECH KRS SWP series Weighing Scale NOTE: Le manuel en Français est disponible au www.kilotech.com.

More information

BBA-Series Precision Balances

BBA-Series Precision Balances Schematic BBA-Series Precision Balances BBA-600/BBA-1200 Service Manual 11 Schematic 1 10 Troubleshooting You have purchased a quality precision weighing instrument that requires handling with care. Trouble

More information

Userʼs Guide. 1200g High Precision Scale. 1200g High Precision Scale. Legal for Trade. Includes RS-232 interface port. UG Version 1.

Userʼs Guide. 1200g High Precision Scale. 1200g High Precision Scale. Legal for Trade. Includes RS-232 interface port. UG Version 1. Userʼs Guide 1200g High Precision Scale Legal for Trade 1200g High Precision Scale Includes RS-232 interface port UG Version 1.2_12-30-13 !!!! CALIBRATION WARNING!!!! Calibration AND inspection of calibration

More information

American Weigh Scales H-110. User Manual. H-110 (110lb x 0.05lb) Copyright 2010 American Weigh Scales, Inc. All rights reserved. Rev. 4.

American Weigh Scales H-110. User Manual. H-110 (110lb x 0.05lb) Copyright 2010 American Weigh Scales, Inc. All rights reserved. Rev. 4. American Weigh Scales H-110 User Manual H-110 (110lb x 0.05lb) Copyright 2010 American Weigh Scales, Inc. All rights reserved. Rev. 4.0 H-110 Thank you for purchasing the H-110 digital hanging scale. Please

More information

Receiver REC 300 Digital. Operating instructions

Receiver REC 300 Digital. Operating instructions Receiver REC 300 Digital GB Operating instructions Operating instructions GB The STABILA REC 300 Digital is a simple-to-use receiver for the rapid capture of rotating lasers. The REC 300 Digital receiver

More information

DWP-PC SERIES PRICE COMPUTING SCALE User s Guide

DWP-PC SERIES PRICE COMPUTING SCALE User s Guide DWP-PC SERIES PRICE COMPUTING SCALE User s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT Specifications subject to change without prior notice CONTENTS 1. INSTALLATION

More information

OHAUS SCOUT BALANCE SCALE

OHAUS SCOUT BALANCE SCALE H-5848, H-5849, H-5850 π H-5851, H-5852, H-5853 H-5854, H-5855 1-800-295-5510 uline.com OHAUS SCOUT BALANCE SCALE CONTROL PANEL OVERVIEW OF CONTROLS CONTROL FUNCTIONS BUTTON Zero Yes Print Units No Function

More information

Weighing Indicator M701. Handrail scales. REV: M1, April,05,2014 (v )

Weighing Indicator M701. Handrail scales. REV: M1, April,05,2014 (v ) Weighing Indicator M701 Handrail scales REV: M1, April,05,2014 (v3.09-3.014) CONTENTS 1. PRECAUTIONS..2 2. INTRODUCTION 3 3. SPECIFICATIONS...4 3.1Specifications...4 3.2 Load Cell Specification......5

More information

STANDARD OPERATING PROCEDURE TONOPORT V

STANDARD OPERATING PROCEDURE TONOPORT V Page 1 of 24 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the Tonoport V in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy Department, University

More information

This document hosted by: Ranger 7000 Scales Instruction Manual

This document hosted by:  Ranger 7000 Scales Instruction Manual This document hosted by: www.oldwillknottscales.com Ranger 7000 Scales Instruction Manual RANGER 7000 SCALES EN-1 TABLE OF CONTENTS 1 INTRODUCTION...4 1.1 Description... 4 1.2 Features... 4 1.3 Definition

More information

INTRODUCTION INSTALLATION

INTRODUCTION INSTALLATION INTRODUCTION SNUG II precision balance series contains five models ---- SNUG II -150,300, 600, 1500 and 3000 with weighing capacities of 150g, 300g, 600g, 1500g and 3000g respectively. The display resolution

More information

EC100. Counting Scale. Operations Manual (V1611)

EC100. Counting Scale. Operations Manual (V1611) EC100 Counting Scale Operations Manual (V1611) Anyload Transducer Co. Ltd Website: www.anyload.com Email: info@anyload.com Fax: +1 866 612 9088 North America Toll Free: 1-855-ANYLOAD (269 5623) TABLE OF

More information

1. INTRODUCTION This manual contains installation, operation and maintenance instructions. Please read the manual completely before using the balance.

1. INTRODUCTION This manual contains installation, operation and maintenance instructions. Please read the manual completely before using the balance. 1. INTRODUCTION This manual contains installation, operation and maintenance instructions. Please read the manual completely before using the balance. 1.1 Safety Precautions Please follow these safety

More information

American Weigh Scales, Inc. AMW-BENCH. D i g i t a l T a b l e t o p S c a l e. User Manual

American Weigh Scales, Inc. AMW-BENCH. D i g i t a l T a b l e t o p S c a l e. User Manual g American Weigh Scales, Inc. AMW-BENCH D i g i t a l T a b l e t o p S c a l e Hold Mode Tare ON/OFF AMW-500i AMW-1000 AMW-2000 User Manual AMW-BENCH Manual Thank you for purchasing the American Weigh

More information

Acaia Lunar Operation Manual

Acaia Lunar Operation Manual Acaia Lunar Operation Manual Using Your Acaia Coffee Scale Identification The instructions in this operation manual apply to the Acaia Lunar. You have purchased a quality precision weighing scale that

More information

Sartorius XQ 6100, XQ 6100 B.

Sartorius XQ 6100, XQ 6100 B. Sartorius XQ 6100, XQ 6100 B. Electronic Paint Mixing Scales Installation and Operating Instructions Sartorius XQ 6100, XQ 6100 B. Please read these installation and operating instructions carefully before

More information

Trade Approved Version NMI: S520

Trade Approved Version NMI: S520 @Weigh BW/BWS Series Trade Approved Version NMI: S520 Weighing Indicator User s guide UGBW-E1.09 Table of Contents SECTION 1 INTRODUCTION. 1 SECTION 2 SPECIFICATIONS... 2 SECTION 3 INSTALLATION.3 3. 1

More information

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD AD-8923-BCD Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD4002137 2010 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated

More information

acaia Coffee Scale Operation Manual

acaia Coffee Scale Operation Manual acaia Coffee Scale Operation Manual Using Your acaia Coffee Scale Identification The instructions in this operation manual apply to the acaia Coffee Scale. You have purchased a quality precision weighing

More information

uline.com OVERVIEW OF CONTROLS. k g. Tare Menu. Exit. Print. Units PRINT

uline.com OVERVIEW OF CONTROLS. k g. Tare Menu. Exit. Print. Units PRINT 3000SERIES 0 CO On/Zero Off Yes Print Units No PCS NET TARE Function Mode Back Tare Menu Exit k g π H-2587, H-4594 OHAUS DEFENDER 3000 DIGITAL SCALE 1-800-295-5510 uline.com OVERVIEW OF CONTROLS CONTROL

More information

Explorer Balances Instruction Manual

Explorer Balances Instruction Manual Explorer Balances Instruction Manual EN-1 TABLE OF CONTENTS 1. INTRODUCTION...5 1.1 Description... 5 1.2 Features... 5 1.3 Safety Precautions... 5 2. INSTALLATION...6 2.1 Unpacking... 6 2.2 Installing

More information

Instruction Manual for. Torbal AGCN Series Balances

Instruction Manual for. Torbal AGCN Series Balances Instruction Manual for Torbal AGCN Series Balances 101 Witmer Road, Suite 700 Horsham, PA 19044 Phone 1-800-521-0754 Fax 215-442-9202 www.labrepco.com sales@labrepco.com Table of Contents Chapter 1.Cautionary

More information

User Manual. Models: BL 303 BL 602 BL 2002 BL 3001 BL 6001 BL 30K1

User Manual. Models: BL 303 BL 602 BL 2002 BL 3001 BL 6001 BL 30K1 User Manual Models: BL 303 BL 602 BL 2002 BL 3001 BL 6001 BL 30K1 Index GLASS PROTECTION 3 ERRORE. IL SEGNALIBRO NON È DEFINITO. INSTALLATION 6 KEYBOARD 6 START 7 DATE AND TIME SETTINGS 7 USER CALIBRATION

More information

Compact Balances. EK-610j EK-6100j. in an Authorized A&D Dealer

Compact Balances. EK-610j EK-6100j.   in an Authorized A&D Dealer Compact Balances EK-610j EK-6100j 1WMPD4001387 This manual and Marks All safety messages are identified by the following, WARNING or CAUTION, of ANSI Z535.4 (American National Standard Institute: Product

More information

DWP-98 Series. Counting scale. User s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT

DWP-98 Series. Counting scale. User s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT c DWP-98 Series Counting scale User s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT Specifications subject to change without prior notice CONTENTS 1. INSTALLATION

More information

Ranger 7000 Scales Instruction Manual

Ranger 7000 Scales Instruction Manual Ranger 7000 Scales Instruction Manual RANGER 7000 SCALES EN-1 TABLE OF CONTENTS 1. INTRODUCTION...4 1.1 Description... 4 1.2 Features... 4 1.3 Definition of Signal Warnings and Symbols... 4 1.4 Safety

More information

Explorer Semi-Micro Balances Instruction Manual

Explorer Semi-Micro Balances Instruction Manual Explorer Semi-Micro Balances Instruction Manual EXPLORER SEMI-MICRO BALANCES EN-1 TABLE OF CONTENTS 1. INTRODUCTION...5 1.1 Description... 5 1.2 Features... 5 1.3 Safety Precautions... 6 2. INSTALLATION...7

More information

EDGE. Digital Kitchen Scale. 11lb x 0.1oz 5000g x 1g

EDGE. Digital Kitchen Scale. 11lb x 0.1oz 5000g x 1g EDGE Digital Kitchen Scale 11lb x 0.1oz 5000g x 1g EDGE Manual Thank you for purchasing the American Weigh EDGE Digital Kitchen Scale. Please read all operating instructions carefully before use. Scales

More information

Operating instructions METTLER TOLEDO BD balances

Operating instructions METTLER TOLEDO BD balances Operating instructions METTLER TOLEDO BD balances 0.00 g Startup procedure 1 Unpacking, standard equipment The standard equipment contains: BD balance, Weighing pan, permanently mounted with BD6000, AC

More information

AL-Series Analytical and Toploading Balances

AL-Series Analytical and Toploading Balances AL-Series Analytical and Toploading Balances 0.000 g I/O Print Menu 0.0000 g Print Menu I/O Operation Manual 902327.1 Rev. C Specifications Analytical Models Weigh Range Readability Linearity Pan Size

More information

CONTENTS INTRODUCTION 1 INSTALLATION 1 KEYS 3 MAXIMUM DISPLAY & DISPLAY RESOLUTION 4 BATTERY STATUS INDICATION 5 POWER ON SETUP 5

CONTENTS INTRODUCTION 1 INSTALLATION 1 KEYS 3 MAXIMUM DISPLAY & DISPLAY RESOLUTION 4 BATTERY STATUS INDICATION 5 POWER ON SETUP 5 CONTENTS INTRODUCTION 1 INSTALLATION 1 KEYS 3 MAXIMUM DISPLAY & DISPLAY RESOLUTION 4 BATTERY STATUS INDICATION 5 POWER ON SETUP 5 COUNT VALUE DISPLAY 6 WEIGHING UNIT SETUP 7 INITIAL UNIT SETUP 7 AUTOMATIC

More information

Platinum RTD Thermometer MODEL NO

Platinum RTD Thermometer MODEL NO Platinum RTD Thermometer MODEL NO. 93410-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com

More information

Standard Analytical Balance

Standard Analytical Balance Standard Analytical Balance Operating Manual written for WBA - 220 witeg Labortechnik GmbH Contents 1 General... 3 2 Safety Instructions... 3 2.1 Description of symbols... 3 2.2 Instructions... 3 2.3 Product

More information

OP-W Series User Manual

OP-W Series User Manual User Manual Warnings Avoid using the balance in any harsh environment. Only trained professionals are allowed to do repair work. Turn power off before connecting/disconnecting the power source. Avoid overloading

More information

Instruction Manual for. Torbal AGZN Series Balances

Instruction Manual for. Torbal AGZN Series Balances Instruction Manual for Torbal AGZN Series Balances 101 Witmer Road, Suite 700 Horsham, PA 19044 Phone 1-800-521-0754 Fax 215-442-9202 www.labrepco.com sales@labrepco.com Table of Contents Chapter 1.Cautionary

More information

AELP PALLET BEAMS SCALES

AELP PALLET BEAMS SCALES Adam Equipment AELP PALLET BEAMS SCALES (P.N. 4327, Rev. A7, Mar 2016) Adam Equipment Company 2016 Adam Equipment Company 2016 CONTENTS 1.0 INTRODUCTION... 2 2.0 TECHNICAL SPECIFICATIONS... 3 3.0 UNPACKING

More information

Totalcomp TLI Indicator. Operation and Maintenance Manual Manual

Totalcomp TLI Indicator. Operation and Maintenance Manual Manual Totalcomp TLI Indicator Operation and Maintenance Manual Manual V1.08-0.06E DT: 04/29/2013 - - CONTENTS Precautions... 2 1. SPECIFICATIONS... 3 2. INTRODUCTION. 4 3. INSTALLATION... 5 Unpacking 5 Parts

More information

High-Precision Tuning Fork Electronic Balance. SJ-CE Series. Operation Manual IMPORTANT. To ensure safe and proper use of the balance, please read

High-Precision Tuning Fork Electronic Balance. SJ-CE Series. Operation Manual IMPORTANT. To ensure safe and proper use of the balance, please read 210015M01 High-Precision Tuning Fork Electronic Balance SJ-CE Series Operation Manual IMPORTANT To ensure safe and proper use of the balance, please read this manual carefully. After reading this manual,

More information

DL Series NEWDOLBO OWNER S MANUAL. Simple Weighing Scale 9002-DL

DL Series NEWDOLBO OWNER S MANUAL. Simple Weighing Scale 9002-DL Simple Weighing Scale DL Series NEWDOLBO OWNER S MANUAL 9002-DL1-0033-0 2005. 9 CONTENTS PRECAUTIONS 4 PREFACE 6 NAMES AND FUNCTIONS 6 INSTALLATIONS 8 OPERATIONS 9 1. General Weighing 9 2. Weighing with

More information

TWP-W. Weighing Indicator Service Manual

TWP-W. Weighing Indicator Service Manual TWP-W Weighing Indicator Service Manual CONTENTS 1. SPECIFICATIONS....3 2. INSTALATION AND PRECAUTIONS..4 Load cell connection......4 Power operation..5 Battery operation.5 3. NAME AND FUNCTIONS........6

More information

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE

More information

Quick Guide. bpro Retail scale

Quick Guide. bpro Retail scale Quick Guide bpro Retail scale 2 METTLER TOLEDO Quick Guide bpro Contents 1 Your new retail scale...4 1.1 Safety instructions...4 1.2 Further documents...4 1.3 Maintenance...4 1.4 Intended use...4 1.5 Device

More information

1. Introduction 2. Specifications 3. Symbol Definition and Button Location 4. Operation Instructions 5. Temperature Measurement 6.

1. Introduction 2. Specifications 3. Symbol Definition and Button Location 4. Operation Instructions 5. Temperature Measurement 6. CONTENTS TITLE PAGE 1. Introduction.. 1 2. Specifications... 1 3. Symbol Definition and Button Location... 2 4. Operation Instructions.... 3 4.1 Power-Up...3 4.2 Connection of the Temperature Probe...

More information

Nimbus Analytical Balances

Nimbus Analytical Balances LABORATRY SCALES AND BALANCES Nimbus Analytical Balances Capacity: 80g 250g Readability: 0.0001g RS-232 Bi-directional interface to communicate with printers or computers GLP compliant printouts for better

More information

This document hosted by: High-Precision Tuning Fork Electronic Balance. SJ-E Series. Operation Manual IMPORTANT

This document hosted by:   High-Precision Tuning Fork Electronic Balance. SJ-E Series. Operation Manual IMPORTANT This document hosted by: www.oldwillknottscales.com 210002M01 High-Precision Tuning Fork Electronic Balance SJ-E Series Operation Manual IMPORTANT To ensure safe and proper use of the balance, please read

More information

ESA SERIES ELECTRONIC BALANCE

ESA SERIES ELECTRONIC BALANCE ESA SERIES ELECTRONIC BALANCE ESA_AWT35-500064_AD.book Declarations of compliance United States This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant

More information

H-1671 REFERENCE. The scale is powered through the included power adapter. The scale can be powered by six AAA batteries (not included).

H-1671 REFERENCE. The scale is powered through the included power adapter. The scale can be powered by six AAA batteries (not included). H-1671 POSTAL SCALE Technical Specifications Capacity: 20.0 pounds (9 kg) Resolution: 0.1 oz. (2 g) Units: lb./oz. (kg/g) Auto-off: After two minutes of inactivity Operating Temperature: 40-104 F (4.5-40

More information

acaia lunar Operation Manual

acaia lunar Operation Manual a c a i a acaia lunar Operation Manual Using Your acaia Coffee Scale Identification The instructions in this operation manual apply to the acaia lunar. You have purchased a quality precision weighing scale

More information

Analytical Balance BM-20 BM-22 BM-252 BM-200 BM-300 BM-500

Analytical Balance BM-20 BM-22 BM-252 BM-200 BM-300 BM-500 Analytical Balance BM-20 BM-22 BM-252 BM-200 BM-300 BM-500 1WMPD4002195 2010 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated

More information

VWR BALANCES. A-Series Balances for High Precision with Advanced Features. B2-Series Balances are Rugged, Affordable, and Reliable

VWR BALANCES. A-Series Balances for High Precision with Advanced Features. B2-Series Balances are Rugged, Affordable, and Reliable VWR BALANCES A-Series Balances for High Precision with Advanced Features B2-Series Balances are Rugged, Affordable, and Reliable E-Series Balances for Entry Level Applications P-Series Balances for Portability

More information

Item ref: UK User Manual. MW-7 4-in-1 LED Mini Moving Head

Item ref: UK User Manual. MW-7 4-in-1 LED Mini Moving Head MW-7 4-in-1 LED Mini Moving Head Item ref: 150.449UK User Manual Please read through this manual thoroughly before use, any damage caused by misuse of the product will not be covered by the warranty. This

More information

Analytical Balance GH-120 GH-200 GH-300 GH-202 GH-252 1WMPD C

Analytical Balance GH-120 GH-200 GH-300 GH-202 GH-252 1WMPD C ふ Analytical Balance GH-120 GH-200 GH-300 GH-202 GH-252 1WMPD4000891C Contents Basic Operation 1. Introduction... 3 1.1. About This Manual... 3 1.2. Features... 4 1.3. Compliance... 4 2. Unpaking the Balance...

More information

MC Piece Counting Scale

MC Piece Counting Scale MC Piece Counting Scale User Manual Contents 1. Precautions 3 2. Installation.. 4 3. Name and functions. 6 Key board... 6 Display 8 4. Operation... 9 Power ON/OFF.9 Zero. 9 Tare. 9 5. Parts Counting...10

More information

DLA. DMX512 Analyzer. DLA Users Manual SV2_00 B.lwp copyright ELM Video Technology, Inc.

DLA. DMX512 Analyzer. DLA Users Manual SV2_00 B.lwp copyright ELM Video Technology, Inc. DLA DMX512 Analyzer DLA DLA-HH 1 Table Of Contents IMPORTANT SAFEGUARDS... 2 DLA OVERVIEW... 3 CONNECTION... 3 OPERATION... 3 HARDWARE SETUP... 4 DLA-HH (PORTABLE) LAYOUT... 4 CHASSIS LAYOUT... 4 DLA MENU

More information

TWP series. Weighing Indicator Service Manual

TWP series. Weighing Indicator Service Manual TWP series Weighing Indicator Service Manual CONTENTS 1. SPECIFICATIONS....3 2. INSTALATION AND PRECAUTIONS..4 Load cell connection......4 Power operation..5 Battery operation.5 3. NAME AND FUNCTIONS........6

More information

GraviTek Titration Scale Instruction Manual

GraviTek Titration Scale Instruction Manual GraviTek Titration Scale Instruction Manual Compliance Compliance to the following standards is indicated by the corresponding mark on the product This product conforms to the EMC directive 2004/108/EC

More information

QB Series. Operation Manual. Precision Balance. To be the best by every measure

QB Series. Operation Manual. Precision Balance. To be the best by every measure QB Series Precision Balance Operation Manual To be the best by every measure 42194 Contents About This Manual... 1 1.0 Introduction... 1 1.1 Unpacking and Inspection.........................................................

More information

OCEAN WATERPROOF WEIGHING SCALE USER MANUAL

OCEAN WATERPROOF WEIGHING SCALE USER MANUAL OCEAN WATERPROOF WEIGHING SCALE USER MANUAL Excell Precision Limited 2006. All rights reserved Worldwide. The information contained herein is the property of Excell Precision Limited and is supplied without

More information

WUNDER SA.BI.SRL EN USER MANUAL. MULTIFUNCTIONS ELECTRONIC SCALE MOD. W100 LED. User Manual

WUNDER SA.BI.SRL EN USER MANUAL.   MULTIFUNCTIONS ELECTRONIC SCALE MOD. W100 LED. User Manual www.wunder.it MULTIFUNCTIONS ELECTRONIC SCALE MOD. W100 LED User Manual Read this manual carefully before using the instrument W100 SCALE W100 INDICATOR INDICE 2. BEFORE USING PAG. 2 3. SPECIFICATIONS

More information

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION UWE-1707 SERIES CONTENTS ELECTRONIC WEIGHING INDICATOR OPERATION MANUAL 1. SPECIFICATIONS PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE SCALE 2. INSTALLATION 3. ROUTINE OPERATION

More information

User Manual - KVM Rear Kit Version

User Manual - KVM Rear Kit Version dedicated KVM switch and rackmount screen technology User Manual - KVM Rear Kit Version for LCD Console Drawer PS/2 DB-15 KVM - 801 back Designed and manufactured by Austin Hughes 751 Legal Information

More information

Operating Instructions. Precision and Analytical Balances ML-T

Operating Instructions. Precision and Analytical Balances ML-T Operating Instructions ML-T Table of Contents 1 Introduction 5 1.1 Conventions and symbols used in these operating instructions 5 2 Safety Information 6 2.1 Definition of signal warnings and symbols 6

More information

Adam Equipment LBK SERIES. (P.N. 9596, Revision D, July 2014)

Adam Equipment LBK SERIES. (P.N. 9596, Revision D, July 2014) Adam Equipment LBK SERIES (P.N. 9596, Revision D, July 2014) Easy Reference: Model name of the scale: Serial number of the unit: Software revision number (Displayed when power is first turned on): Date

More information

High Accuracy Gloss Meter Operation Manual. This manual is only suitable for our single angle and multi-angles gloss meters.

High Accuracy Gloss Meter Operation Manual. This manual is only suitable for our single angle and multi-angles gloss meters. Content Introduction... 1 Cautions... 1 1. External structure description... 2 2. Switch on/off... 3 2.1 Startup... 3 2.2 Shutdown... 4 3. Basic operation... 4 3.1 Menu Operation... 4 3.2 Enter the main

More information

SELECTING THE LOCATION PREPARATION FOR USE. Pan Calibration Mass (only for SJX/E models with 64g capacity) Scoop (only for models with 64g capacity)

SELECTING THE LOCATION PREPARATION FOR USE. Pan Calibration Mass (only for SJX/E models with 64g capacity) Scoop (only for models with 64g capacity) UNPACKING Carefully remove your Scout balance and each of its components from the package. Store the packaging material to ensure safe storage and future transport. Included Components: Balance Quick Start

More information

1. Introduction Page 2. Specifications Page Installation Page Key Functions Page Operations Page 6. Check Weighing Page 6

1. Introduction Page 2. Specifications Page Installation Page Key Functions Page Operations Page 6. Check Weighing Page 6 Contents: 1. Introduction Page 2 Specifications Page 2 2. Installation Page 3 3. Key Functions Page 4 4. Operations Page 6 Check Weighing Page 6 Accumulation Page 7 Parts Counting Page 7 Print (RS232C)

More information

Application Guide for the Secura, Quintix and Practum Series of Models

Application Guide for the Secura, Quintix and Practum Series of Models Application Guide for the Secura, Quintix and Practum Series of Models Getting Started Quickly Density Checkweighing Counting Statistics Weighing Components Mixing Weighing in Percent Peak Hold Unstable

More information

English. User Manual Precision and Analytical Balances ML-T

English. User Manual Precision and Analytical Balances ML-T English User Manual ML-T 1 Safety Information 1.1 Definition of signal warnings and symbols Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring

More information

High-Precision, Advanced Tuning Fork Balance. Operation Manual IMPORTANT

High-Precision, Advanced Tuning Fork Balance. Operation Manual IMPORTANT 360003M01 High-Precision, Advanced Tuning Fork Balance CE Series Operation Manual IMPORTANT To ensure safe and proper use of the balance, please read this manual carefully. After reading this manual, store

More information

Acaia Pearl Operation Manual

Acaia Pearl Operation Manual Acaia Pearl Operation Manual Using Your Pearl FW 2.0.008 Identification The instructions in this operation manual apply to Acaia Coffee Scale. You have purchased a quality precision weighing scale that

More information

Barnant Company. Humidity Data Logger MODEL NO YR MIN HOLD MO MAX DAY STORE LOG RECALL PRINT C/ F PRINT SETUP.

Barnant Company. Humidity Data Logger MODEL NO YR MIN HOLD MO MAX DAY STORE LOG RECALL PRINT C/ F PRINT SETUP. Barnant Company Humidity Data Logger MODEL NO. 600-2050 %RH CAL F DP YR MIN HOLD MO MAX DAY STORE LOG RECALL PRINT STORE C/ F HOLD RECALL LOG MAX CLEAR PRINT MIN ON /OFF SETUP CAL Barnant Company 28W092

More information

Counting Scale. HC-3Ki. HC-15Ki HC-6Ki WM+PD

Counting Scale. HC-3Ki. HC-15Ki HC-6Ki WM+PD Counting Scale HC-30Ki HC-15Ki HC-6Ki HC-3Ki WM+PD41164 This manual and Marks All safety messages are identified by the following, WARNING or CAUTION, of ANSI Z535.4 (American National Standard Institute:

More information

PANTONE Formula Scale 3

PANTONE Formula Scale 3 PANTONE Formula Scale 3 Operating Manual Models: PFS3-5-22, PFS3-5-12 PANTONE is the property of Pantone, Inc. Copyright Cherlyn Electronics Limited, 2006 Portions Pantone, Inc., 2000 Introduction The

More information

Service Manual. Operators Manual. Ultegra Health Scale by Fairbanks Scales, Inc All rights reserved Revision 7 04/17

Service Manual. Operators Manual. Ultegra Health Scale by Fairbanks Scales, Inc All rights reserved Revision 7 04/17 Service Manual Operators Manual Ultegra Health Scale 2009-2017 by Fairbanks Scales, Inc. 50735 All rights reserved Revision 7 04/17 Amendment Record Ultegra Health Scale 50735 Manufactured by Fairbanks

More information

PCE Americas Inc. PCE Instruments UK Ltd. 711 Commerce Way Units 12/13, Southpoint Business Park Suite 8 Ensign Way Jupiter, FL 33458 Hampshire / Southampton USA United Kingdom, SO31 4RF Phone: 561-320-9162

More information

VWR BALANCES. A-Series Balances for High Precision with Advanced Features. B2-Series Balances are Rugged, Affordable, and Reliable

VWR BALANCES. A-Series Balances for High Precision with Advanced Features. B2-Series Balances are Rugged, Affordable, and Reliable VWR BALANCES A-Series Balances for High Precision with Advanced Features B2-Series Balances are Rugged, Affordable, and Reliable B2T-Series Balances Feature a 5 Color Touch Screen and Proximity Sensor

More information

Using the Cashmaster you will notice some changes in the way you work:

Using the Cashmaster you will notice some changes in the way you work: User Manual CONTENTS Introduction. 1 Box items and guidelines... 2 Maintenance... 2 Charging.. 2 Warning... 2 Unpacking... 3 Keys and display... 4 Cashmaster Sigma 170... 4 The keypad..... 5 The screen.......

More information

Using the Cashmaster you will notice some changes in the way you work:

Using the Cashmaster you will notice some changes in the way you work: User Manual CONTENTS Introduction. 1 Box items and guidelines... 2 Maintenance... 2 Charging.. 2 Warning... 2 Unpacking... 3 Keys and display... 4 Cashmaster Sigma 170... 4 The keypad..... 5 The screen.......

More information

6165 MONEY COUNTING SCALE

6165 MONEY COUNTING SCALE MANUAL English Nederlands Deutsch Français Español Italiano Português Polski Česky Slovenčina Magyar България 6165 MONEY COUNTING SCALE TABLE OF CONTENTS 01. INTRODUCTION page 3 1.1 Product 1.2 Display

More information