Running Multiple Experiments on the Same Sample Automatically. (multizg, spooler/queue, iconnmr)

Size: px
Start display at page:

Download "Running Multiple Experiments on the Same Sample Automatically. (multizg, spooler/queue, iconnmr)"

Transcription

1 Running Multiple Experiments on the Same Sample Automatically (multizg, spooler/queue, iconnmr)

2 Multiple Experiment Setup If you have one sample that requires full characterization (e.g. 1 H, 13 C, COSY, HMBC etc.) and you do not want to wait until the previous experiment finishes to start the next one. There are ways to setup you experiments and have them run automatically. For systems without an auto sampler, this can only be done on one sample at a time. There are various way in which this can be achieved, here touches upon three common methods: Using the multizg command Setting up a queue of experiments Through iconnmr (only available for UTL300 with an autosampler)

3 Multizg Multizg runs experiments that have been setup sequentially within the same experiment folder automatically. Setup 1. Create a new experiment folder, insert sample and do the preacquisition processes as usual (e.g. lock, tune, shim, adjust receiver gain) 2. Make new experiment files for each type of experiment that you want to acquire. Keep the same folder but change EXPNO. 3. Go back to the first experiment of the list and execute the multizg command. A popup window will ask How many experiment do you want to perform so enter the number of experiments that you created. Advantages Only one command needed Disadvantages Once the multizg is running it is difficult to stop the entire process Additional experiments cannot be appended to the list

4 Spooler / Queue A better alternative is to use the spooler command and setup a study queue. Setup 1. Create a new experiment folder, insert sample and do the preacquisition processes as usual (e.g. lock, tune, shim, adjust receiver gain) 2. Enter the command spooler in the command line which will bring up the spooler window. This lists all the commands that are in queue. 3. To add a command to the queue list use the command qu <command> (e.g. qu zg). You can input any type of Topspin command, acquisition and processing. 4. If multiple experiments are desired, read in the next experiment number then queue the commands for that experiment. 5. Alternatively, for multiple EXPNOs use the command qumulti. This will bring up a New Job window with all experiments listed in the current data set. Enter the command to be executed and check off all EXPNOs for it to be executed on.

5 Spooler Window When using the lock command, ensure you specify which solvent to lock on. If not, you will be prompted to select the solvent and the queue will not continue until then. When doing multiple nuclei on one channel ensure that you retune the probe before the next experiment is acquired as seen above. To delete a command, right click on command line and select Delete in options menu.

6 Advantages over multizg Spooler / Queue Displays all experiments to be run in a easy to use GUI Able to stop acquisitions easily in the event of improper parameter setup Typing stop in the command line will auto suspend all spooler commands Delete any or all items in the spooler by right clicking and select delete Able to queue any Topspin command (i.e. atma, rga, ef, apk, etc.) Displays exact end time for each experiment Able to manipulate experimental parameters for acquisitions already in the spooler (i.e. number of scans, number of 2D/3D increments, etc.) and end time will be automatically recalculated Not necessary to run experiments that are in sequential EXPNOs or even under the same experiment name

7 iconnmr IconNMR should only be used on the UTL300 spectrometer equipped with an autosampler. Steps: Open Topspin, if it is not already open Type iconnmr on the command line Choose Automation Pick userid and provide password with prompt

8 Populate 6 fields: Disk: N:\ (auto populates) File Name Experiment number (auto populates) Solvent (from drop down) Experiment (from drop down): N PROTON.f or N PROTON Title (if needed) Highlight experiment to copy and click copy button and change number to the number of samples you have to run. Then edit sample name and/or text Highlight experiments to run and click submit button

9 Status light will be yellow after submitting Once all samples are submitted click the green start icon A new window will pop up, click start again Status light will stay yellow until sample is progress, at which time it turns green

10 current function highlighted. Also new pop up window allows you to stop the automation run Lower portion of the screen summarizes experiment and provides any errors in the remarks.

Bruker NMR Topspin Manual

Bruker NMR Topspin Manual Bruker NMR Topspin Manual Topspin is an icon driven software package. This means that instead of reading words and typing commands in the software most of the things will be carried out by clicking icons

More information

NMR Data Acquisition and Processing Procedure

NMR Data Acquisition and Processing Procedure NMR Data Acquisition and Processing Procedure Dr. Jianfeng Zhu (Research Officer) Please DO NOT remove from NMR lab! Things to Avoid when Using NMR Lab Here's a list of stuff that I have seen that you

More information

Quick Guide to Topspin

Quick Guide to Topspin Quick Guide to Topspin 1. Log into the workstation. Right mouse click on the desktop anywhere and select new terminal. Type topspin [enter] (if this fails, create the launcher as described next). Alternatively,

More information

Vanderbilt Small Molecule NMR Center

Vanderbilt Small Molecule NMR Center Vanderbilt Small Molecule NMR Center Instructions for Using TOPSPIN version 2.0 DRX 400: Setting up 1H, 13C (1H decoupled), and 31P NMR Experiments Getting started: Sample Preparation: Make sure your sample

More information

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER DISCLAIMER Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 This document is intended to be a brief, bare-bones user s

More information

DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015

DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015 DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015 The Bruker AVANCE III HD 400 has a broadband (BBO) probe with a 1 H Channel and X-Channel tunable from (30-300 MHz). Both channels

More information

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin.

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin. TopSpin QuickStart Preliminaries Sign Up Sign up for instrument time at calendar.yahoo.com/hmcnmr The password is nmr. During the summer, 1100-1200 and 1600-1700 are reserved for walk-on use (15 min maximum

More information

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation The GVSU chemistry department owns 2 NMR spectrometers. A JEOL Eclipse 300 MHz and a Varian Inova 400 MHz. Due to their strong, constant magnetic

More information

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer Dr. Zhenming Du Email: zdu@gsu.edu Phone: 3-5538 Revised 04/02/2015 LC=Left click; DC=double

More information

DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015

DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015 DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015 The Bruker AVANCE III HD 401 has a broadband (BBO) probe with a 1 H Channel and X-Channel tunable from (30-300 MHz). Both channels

More information

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500 VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500 August 9, 2016 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they

More information

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751)

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751) Dissimilar This manual is intended to be only a very brief introduction for using the Bruker 400 MHz NMR spectrometer (PS751) at the California State University LA NMR Facility. For complete information

More information

IconNMR Operating Procedures

IconNMR Operating Procedures IconNMR Operating Procedures Revised: 7-1-2015 This guide provides basic instructions for acquiring NMR spectra using the IconNMR automation program. 1. Insert your sample(s) in the SampleXpress sample

More information

Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12

Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12 Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12 DOSY Diffusion Ordered SpectroscopY. NMR diffusion experiments, like DOSY, are used to determine the diffusion coefficients of solute

More information

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 January 10, 2019 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they are shorter; you

More information

1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR. Spectrometer without the Sample Changer and IconNMR

1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR. Spectrometer without the Sample Changer and IconNMR 1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR Spectrometer without the Sample Changer and IconNMR Scheduling NMR Time: Any use of the 700 MHz NMR spectrometer requires prior

More information

Bruker Topspin NMR Training Manual

Bruker Topspin NMR Training Manual Bruker Topspin NMR Training Manual Rev 180521 Perry Pellechia University of South Carolina NMR Facility Subject Page Number Overview of Operation... 1 Gauging Samples in Spinners... 2 Starting and logging

More information

NMR Spectrometer Crib-Sheet

NMR Spectrometer Crib-Sheet NMR Spectrometer Crib-Sheet Sample Preparation: 1. Dissolve your sample in a deuterated solvent. For 1 H NMR use ~ 1 mm concentrations. For 13 C NMR, use ~ 10 mm. 2. Use clean dry tubes to avoid contaminating

More information

Bruker Avance 400 MHz Instructions

Bruker Avance 400 MHz Instructions Bruker Avance 400 MHz Instructions General Policy: 1) All users must pass the training course before they will obtain their own access to the spectrometer. 2) A key to Bruker Avance 400 Room is accessible

More information

Acquiring Data in VnmrJ 4.2A

Acquiring Data in VnmrJ 4.2A Acquiring Data in VnmrJ 4.2A Linux Primer Initial Steps Changing Password Disabling Screen Lock Reservations Reservation Terminals Rules and Proper Etiquette System Identification Working with VnmrJ 4.2A

More information

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy PINMRF Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy INCLUDING: AV-III-400-HD w/ 5mm BBFO SmartProbe 369 WTHR AV-III-500-HD w/ 5mm BBFO

More information

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS April 16, 2018 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they are shorter; you

More information

Processing data with Bruker TopSpin

Processing data with Bruker TopSpin Processing data with Bruker TopSpin This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and

More information

On Artemis and Cronus, make sure you are adjusting the db scale and not the watts

On Artemis and Cronus, make sure you are adjusting the db scale and not the watts Solvent Suppression - Presaturation (NOTE: There are many types of solvent suppression this is presaturation and all exchangeable protons will be attenuated-instructions for other types follow) 1. Turn

More information

DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400

DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400 1 DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400 400: The Bruker DPX 400 has a broadband probe with 1 H Channel and X-Channel tunable from (30-300 MHz). Both Channels may be tuned researchers.

More information

CITRIX NAVIGATION & ACCESSING myhr

CITRIX NAVIGATION & ACCESSING myhr INTRODUCTION This guide details how to log into Citrix and navigate to the myhr Home page. If you have any difficulty throughout this process please contact ICT (extension 43000). After 20 minutes of inactivity,

More information

IconNMR (>TopSpin 3.0)

IconNMR (>TopSpin 3.0) IconNMR (>TopSpin 3.0) Dr. Sven Augner Bruker BioSpin, Rheinstetten, Germany Innovation with Integrity How to start an IconNMR set 2 IconNMR Select the respective holder 2x IconNMR Select the disk Base

More information

UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X

UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X TABLE OF CONTENTS UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X VERSION 3 February, 2016 Disclaimer... 2 Conventions... 2 Routine 1D NMR Experiments Procedure for acquiring 1D Proton spectrum...

More information

EQNMR Facility Rules of Use

EQNMR Facility Rules of Use New User Package for East Quad 600 NMR Date: 01/25/08 Welcome to the East Quad NMR facility. This document contains information for new users using the 600MHz NMR at our facility. In this package, you

More information

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry NMR Spectroscopy with VnmrJ University of Toronto, Department of Chemistry Walk-up interface 1 Logging in 1 Starting VnmrJ 1 Inserting sample into the magnet or sample changer 1 Enter sample information

More information

PINMRF. Bruker Avance / AV NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy

PINMRF. Bruker Avance / AV NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy PINMRF Bruker Avance / AV NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy INCLUDING: Avance DRX500-1 w/ 5mm TBI Probe - 367 WTHR Avance DRX500-2 w/ 5mm BBO or TXI

More information

KeePass Password Safe: Password Manager

KeePass Password Safe: Password Manager KeePass Password Safe: Password Manager The recommended from Information Services at the University of Regina password manager is KeePass Password Safe. This documents is designed to assist with the creation

More information

See XwinNMR Acquisition Manual, Sections 1.4, 1.5 (see list of acquistion parameter descriptions starting , pg. A-39), 1.6

See XwinNMR Acquisition Manual, Sections 1.4, 1.5 (see list of acquistion parameter descriptions starting , pg. A-39), 1.6 Bruker Avance360 X-Nucleus Experiments Page 13 III. X-Nucleus Acquisitions Example Session See XwinNMR Acquisition Manual, Sections 1.4, 1.5 (see list of acquistion parameter descriptions starting 1.5.2.4,

More information

VnmrJ Walkup. Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No , Rev. A 0808

VnmrJ Walkup. Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No , Rev. A 0808 VnmrJ Walkup Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No. 01-999377-00, Rev. A 0808 VnmrJ Walkup Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software

More information

Welcome to the NMR Center

Welcome to the NMR Center 1 Welcome to the NMR Center The NMR center in chemistry consists of six Bruker NMR instruments Field/Name Location Comments Operating Software 300nb (narrow bore) 119 Havemeyer walk-on, routine 1 H Topspin

More information

KJM Proton T1 Spectra on the AVI-600 and AVII-600. Version 1.0

KJM Proton T1 Spectra on the AVI-600 and AVII-600. Version 1.0 KJM 9250 Proton T1 Spectra on the AVI-600 and AVII-600 Version 1.0 Professor Emeritus Alistair Lawrence Wilkins, University of Waikato, New Zealand. February 2018 Proton T1 Spectra on the AVI-600 and AVII-600

More information

Varian Solution NMR Automation Procedure

Varian Solution NMR Automation Procedure Menu System System Tool Bar Command Line Graphics Tool Bar Vertical Panels Protocols NMR Data Display NMR Graphics Area Study Queue Horizontal Panels Activity Monitoring Varian Solution NMR Automation

More information

2 Dimensional NMR User s Brief Guide

2 Dimensional NMR User s Brief Guide 2D NMR Handout 7/17/18 Weiguo Hu 2 Dimensional NMR User s Brief Guide (This handout presumes your thorough familiarity with the 1D handout commands and their use!) Options There are a vast number of 2D

More information

Bruker AVANCE III HD 400 MHz Spectrometer Operation

Bruker AVANCE III HD 400 MHz Spectrometer Operation Bruker AVANCE III HD 400 MHz Spectrometer Operation The spectrometer is located in Room 312 of Physical Sciences I. It is equipped with a SampleXpress sample changer for full automation. To utilize the

More information

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy Delta V5 operation Widetron Technologies Corp. Wu, Chung Ying Connect to spectrometer 1. Start Delta V5 (double click ). 2. Open Spectrometer Control (click ). 3. Select spectrometer on Available Instruments.

More information

Océ Posterizer Pro. POP into retail. User manual Quick Reference Manual

Océ Posterizer Pro. POP into retail. User manual Quick Reference Manual - Océ Posterizer Pro POP into retail o User manual Quick Reference Manual Copyright copyright-2010 Océ All rights reserved. No part of this work may be reproduced, copied, adapted, or transmitted in any

More information

TopSpin 1.3 patch level 10. DPX 200, DPX 300 and DRX 500. Windows XP

TopSpin 1.3 patch level 10. DPX 200, DPX 300 and DRX 500. Windows XP KJM 5280 1 H nmr introductory user manual I Version 1.5 TopSpin 1.3 patch level 10 DPX 200, DPX 300 and DRX 500 Windows XP F. Rise May 2018 1 Welcome to the wonderful and amazing world of Nuclear Magnetic

More information

CONTENTS 1D EXPERIMENTS D EXPERIMENTS APPENDIXES... 83

CONTENTS 1D EXPERIMENTS D EXPERIMENTS APPENDIXES... 83 Jari Koivisto, Ph.D. Aalto University School of Science and Technology Faculty of Chemistry and Materials Sciences Department of Chemistry Bruker Avance DPX400 User Manual Basic and Advanced 1D and 2D

More information

MNova Version Installation Instructions.

MNova Version Installation Instructions. MNova Version 12.0.0 Installation Instructions. Note: This Document contains information for a new NMR user. Carefully read and accomplish each section before moving to the next. If you can t finish a

More information

Installing TopSpin 4

Installing TopSpin 4 Installing TopSpin 4 Note: the screenshots in this guide are from the Windows installation. The Mac installation should be similar, but you must follow the prompts carefully. Start 1. Make an account with

More information

Nicolet is50 FTIR User s Booklet

Nicolet is50 FTIR User s Booklet Nicolet is50 FTIR User s Booklet I. Getting started 1) Start your session by launching the LSA Chemistry Recharge software. This software keeps track of the amount of time that you are using the instrument.

More information

ONE DIMENSIONAL NMR SPECTROSCOPY. DPX 200 and DPX 300

ONE DIMENSIONAL NMR SPECTROSCOPY. DPX 200 and DPX 300 1 1H, 13C, 15N, 19F and 31P ONE DIMENSIONAL NMR SPECTROSCOPY (Xwinnmr on PC, W2Kpro, version 3.5) on DPX 200 and DPX 300 Department of Chemistry University of Oslo February 2005 Version 1 (To be used from

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in the Thermo TraceFinder 3.0 analytical software. For detailed descriptions

More information

Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start

Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start Sample preparation for the Bruker OminiFlex MALDI-TOF Mass Spectrometer is of crucial importance. Please take care when preparing your

More information

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Guide for Basic 1D NMR Spectroscopy

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Guide for Basic 1D NMR Spectroscopy PINMRF Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Guide for Basic 1D NMR Spectroscopy INCLUDING: AV-III-400-HD w/ 5mm BBFO SmartProbe 369 WTHR AV-III-500-HD w/ 5mm BBFO Cryoprobe

More information

Bruker AVANCE-360 User s Guide for the UWChemMRF

Bruker AVANCE-360 User s Guide for the UWChemMRF Bruker AVANCE-360 User s Guide for the UWChemMRF by CG Fry last revised: 2004.08.01 Bruker Avance360 Introduction Page 2 UWChemMRF User Guide for XwinNMR I. Introduction... 3 a) Account information...

More information

STUDY ASSISTANT. Adding a New Study & Submitting to the Review Board. Version 10.03

STUDY ASSISTANT. Adding a New Study & Submitting to the Review Board. Version 10.03 STUDY ASSISTANT Adding a New Study & Submitting to the Review Board Version 10.03 Contents Introduction... 3 Add a Study... 3 Selecting an Application... 3 1.0 General Information... 3 2.0 Add Department(s)...

More information

1D and 2D Experiments Step-by-Step Tutorial

1D and 2D Experiments Step-by-Step Tutorial 1D and 2D Experiments Step-by-Step Tutorial Basic Experiments User Guide Version 004 The information in this manual may be altered without notice. BRUKER BIOSPIN accepts no responsibility for actions taken

More information

Running. Time out. Event wait. Schedule. Ready. Blocked. Event occurs

Running. Time out. Event wait. Schedule. Ready. Blocked. Event occurs Processes ffl Process: an abstraction of a running program. ffl All runnable software is organized into a number of sequential processes. ffl Each process has its own flow of control(i.e. program counter,

More information

Community Fund Grant Application Technical Assistance. United Way of the Dutchess-Orange Region

Community Fund Grant Application Technical Assistance. United Way of the Dutchess-Orange Region Community Fund Grant Application Technical Assistance United Way of the Dutchess-Orange Region Create an Application New & Returning Users Browsers Supported Include: IE9, IE10, IE11, Mozilla Firefox,

More information

Varian Solution NMR Procedure

Varian Solution NMR Procedure System Tool Bar Command Line Vertical Panels Protocols Menu System User Tool Bar NMR Data Display Graphics Tool Bar NMR Graphics Area Study Queue Horizontal Panels Hardware Bar Varian Solution NMR Procedure

More information

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised 4-21-2015 Relaxation times are divided into two types: longitudinal, which concerns change in magnetization along the

More information

CompuLEAD: Installing LeadLINK Plus!

CompuLEAD: Installing LeadLINK Plus! CompuLEAD: Installing LeadLINK Plus! LEADlink Plus! Sales Lead Management Software, by CompuSystems, is a new, innovative and creative way of capturing sales leads through a CompuLEAD data recorder into

More information

DOCUMENT IMAGING REFERENCE GUIDE

DOCUMENT IMAGING REFERENCE GUIDE January 25, 2017 DOCUMENT IMAGING REFERENCE GUIDE AppXtender Web Access version 7 Kent State University Division of Information Services AppXtender Web Access Help: For questions regarding AppXtender Web

More information

JAMS 7.X Getting Started Guide

JAMS 7.X Getting Started Guide Table of Contents JAMS Overview 2 Working with Servers 3-4 The JAMS Client Interface 5 JAMS Scheduler Overview 6 Defining Folders and Jobs 7-10 1 2018 MVP Systems Software, Inc. All Rights Reserved. JAMS

More information

e-bridge Color Profile Tool Quick Start Guide

e-bridge Color Profile Tool Quick Start Guide e-bridge Color Profile Tool Quick Start Guide 1 Contents 1. Installation... 3 1.1. Installing the e-bridge Color Profile Tool Software... 3 1.1. Removing the e-bridge Color Profile Tool... 4 1.2. Installing

More information

Agilent VnmrJ 3.2. Quick Start Guide

Agilent VnmrJ 3.2. Quick Start Guide Agilent VnmrJ 3.2 Quick Start Guide VnmrJ 3.2 Interface 3 Overview 4 Prepare Sample 5 Load Sample 6 Enter Sample Information 7 Build Study Queue 8 Run Study Queue 8 Process Data 9 Plot Data 10 This Quick

More information

JAMS 7.X Getting Started Guide

JAMS 7.X Getting Started Guide Table of Contents JAMS Overview 2 Working with Servers 3-4 The JAMS Client Interface 5 Customizing the JAMS Client 6-8 JAMS Scheduler Overview 9 Defining Folders and Jobs 10-13 1 2017 MVP Systems Software,

More information

General Information for Automated Inova 500 MHz NMR (Indy)

General Information for Automated Inova 500 MHz NMR (Indy) General Information for Automated Inova 500 MHz NMR (Indy) Primary Contacts: NMR Facility Manager: David VanderVelde, x3004, davidv@caltech.edu GLA: Pamela Tadross, x6131, 646-220-5789 (cell), 204 Church,

More information

Vnmrj 3.1 New and improved features

Vnmrj 3.1 New and improved features Vnmrj 3.1 New and improved features Automation& Hardware Printing USB printers are supported. Using CUPS, a generic print dialog and JPG, PDF, & other formats have replaced the previous method of sending

More information

Basic Procedure 1D NMR Data Collection

Basic Procedure 1D NMR Data Collection Basic Procedure 1D NMR Data Collection Log into your UNIX account and start XWIN- NMR program At the prompt login:, type your (e.g. jdw, mmj, mjt) or click on your group icon (in some spectrometers

More information

How To Title: Scan and File into ICS Purpose: To scan and file information into the patient s medical record Audience: All Staff Date: June 21, 2013

How To Title: Scan and File into ICS Purpose: To scan and file information into the patient s medical record Audience: All Staff Date: June 21, 2013 There are two ways to import medical records into ICS: A. Scan paper document(s) utilizing a scanner B. Import documents from a file, folder or Right Fax Acquiring images through a scanner: 1. Check that

More information

MX-FR41U. Data Security Kit Operation Guide. Contents

MX-FR41U. Data Security Kit Operation Guide. Contents MX-FR41U Data Security Kit Operation Guide Contents Introduction...2 When the data security kit is installed...4 Security settings of the data security kit...4 Document control function...4 Document Filing...4

More information

Scan to Quick Setup Guide

Scan to  Quick Setup Guide Xerox WorkCentre M118i Scan to E-mail Quick Setup Guide 701P42574 This guide provides a quick reference for setting up the Scan to E-mail feature on the Xerox WorkCentre M118i. It includes procedures for:

More information

VnmrJ Walkup. Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No , Rev. 0604

VnmrJ Walkup. Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No , Rev. 0604 VnmrJ Walkup Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No. 01-999266-00, Rev. 0604 VnmrJ Walkup Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No. 01-999266-00, Rev. 0604 VnmrJ Walkup

More information

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP) 500 MHz Solution-state NMR Procedure (Bruker AVANCE Machines running TopSpin under WINDOWS XP) Jerry Hu, x7914, jghu@mrl.ucsb.edu Shamon Walker, x3248 shamonwalker@mrl.ucsb.edu Materials Research Laboratory

More information

An Introduction to Google Chrome

An Introduction to Google Chrome An Introduction to Google Chrome Box is an online file sharing and cloud content management service and it provides features such as: Getting Google Chrome 1. To get started with Google Chrome, you have

More information

TopSpin 3.5 Tips and Tricks. Nicole Kruse US Applications Pre PANIC Workshop February 20 th 2017

TopSpin 3.5 Tips and Tricks. Nicole Kruse US Applications Pre PANIC Workshop February 20 th 2017 TopSpin 3.5 Tips and Tricks Nicole Kruse US Applications Pre PANIC Workshop February 20 th 2017 Non-Uniform Sampling Incorporated into TopSpin 3.0 Higher Resolution NUS HSQC NS = 2 TD = 2048 NUS @ 10 %

More information

TOCSY 15. Goto. Introduction Pulse Sequence Diagram AVANCE User s Guide Bruker 167

TOCSY 15. Goto. Introduction Pulse Sequence Diagram AVANCE User s Guide Bruker 167 Chapter TOCSY 15 Introduction 15.1 Goto TOtal Correlation SpectroscopY provides a different mechanism of coherence transfer than COSY for 2D correlation spectroscopy in liquids. In TOCSY, cross peaks are

More information

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1)

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1) Menu System System Tool Bar Command Line Graphics Tool Bar Vertical Panels Protocols NMR Data Display NMR Graphics Area Study Queue Horizontal Panels Activity Monitoring Varian Solution NMR Automation

More information

Application Extender 16.3 Web Access

Application Extender 16.3 Web Access Application Extender 16.3 Web Access Pre-requisites for WebXtender client Firefox, Chrome or Edge browser For scanning, your friendly neighborhood desktop support person will need to install the specific

More information

The e-doc Manger. The e-doc Manager. Date and Time Stamps

The e-doc Manger. The e-doc Manager. Date and Time Stamps The e-doc Manger The e-doc Manager The e-doc Manager is the digital document folder for a loan file. All of a loan s files and documents are digitally stored in the e-doc Manager. To access a loan s e-doc

More information

NMR Training VNMRJ 4.2A/VNMRS 400

NMR Training VNMRJ 4.2A/VNMRS 400 NMR Training VNMRJ 4.2A/VNMRS 400 The VNMRS/new 400 MHz NMR uses VNMRJ 4.2A. Although VNMRJ is different than VNMRalmost ALL typed commands that you know from VNMR should work. LOGIN: Your username is

More information

Examplify: Downloading and Taking an Exam (Computer)

Examplify: Downloading and Taking an Exam (Computer) Examplify: Downloading and Taking an Exam (Computer) Once successfully logged into Examplify, there may or may not be exam(s) ready for download. If you do not see an exam you are scheduled to take, click

More information

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800 Walkup NMR Varian NMR Spectrometer Systems With VNMR 6.1C Software Pub. No. 01-999159-00, Rev. A0800 Walkup NMR Varian NMR Spectrometer Systems With VNMR 6.1C Software Pub. No. 01-999159-00, Rev. A0800

More information

vbound User Guide vbound User Guide Version Revised: 10/10/2017

vbound User Guide vbound User Guide Version Revised: 10/10/2017 vbound User Guide Version 4.1.1 Revised: 10/10/2017 Copyright 2014-2017 FFL Solutions Inc. Page 1 of 87 Table of Contents Using vbound...5 Starting vbound... 5 Bound Book List... 6 vbound Ribbon Menu...

More information

Introduction to Bruker Pulse Programs

Introduction to Bruker Pulse Programs Introduction to Bruker Pulse Programs Where to find the pulseprogram Computer: /opt/topspin/exp/stan/nmr/lists/pp/user or: cd $pp (changes direct to user folder) Topspin: edcpul, edpul Display: commandline:

More information

Kitsap County Online Permit Center How to Log In and Apply for a Right-of-Way Permit

Kitsap County Online Permit Center How to Log In and Apply for a Right-of-Way Permit Kitsap County Online Permit Center How to Log In and Apply for a Right-of-Way Permit Step 1: Go to https://co-kitsap-wa.smartgovcommunity.com/applicationpublic/applicationhome For new users select Sign

More information

FRAZER COMPUTING INTEGRATION

FRAZER COMPUTING INTEGRATION FRAZER COMPUTING INTEGRATION RouteOne / Frazer Computing Integration Quick Start Guide The Frazer system integration solution enables a two-way exchange of credit application information. Frazer integration

More information

c. Saturday and Sunday: 10 minute time blocks and unlimited time.

c. Saturday and Sunday: 10 minute time blocks and unlimited time. Scheduling 400 MHz A with 100 slot sample changer: a. Monday to Friday from 8 am to 8 pm: 10 minute blocks, 30min max. b. Monday to Friday from 8 pm to 8 am: 10 minute blocks, unlimited time. c. Saturday

More information

WORLDWIDE PANTS COLLECTION USER GUIDE! As of ! For best results, use Google Chrome as the recommended web browser.!

WORLDWIDE PANTS COLLECTION USER GUIDE! As of ! For best results, use Google Chrome as the recommended web browser.! WORLDWIDE PANTS COLLECTION USER GUIDE As of 3-19-15 For best results, use Google Chrome as the recommended web browser. NEW USER REGISTRATION 1. First time users will need to create an account. To create

More information

NC User Conference Tips and Tricks for SAS FM June 16, 2009

NC User Conference Tips and Tricks for SAS FM June 16, 2009 NC User Conference Tips and Tricks for SAS FM June 16, 2009 Reporting 1. CDA Formula (Formatted Statement): To switch out the cell reference for the hardcoded member or dim name, you can go into the formula

More information

My Practice Profile Attestation QUICK REFERENCE

My Practice Profile Attestation QUICK REFERENCE My Practice Profile Attestation QUICK REFERENCE My Practice Profile allows you to view, update and attest (update and attest capability is only available for the Optum ID administrator) group and provider

More information

DocAve 4.1 SharePoint Disaster Recovery Platform Recovery (SPDR PR) User Guide

DocAve 4.1 SharePoint Disaster Recovery Platform Recovery (SPDR PR) User Guide August 2007 DocAve 4.1 SharePoint Disaster Recovery Platform Recovery (SPDR PR) User Guide Additional user guides available at http://www.avepoint.com/support 1 Copyright 2001-2007 AvePoint, Inc. All rights

More information

TraceFinder Acquisition Quick Reference Guide

TraceFinder Acquisition Quick Reference Guide TraceFinder Acquisition Quick Reference Guide This quick reference guide describes the Thermo TraceFinder acquisition tasks assigned to the Technician role. Contents Acquisition Mode Selecting a Batch

More information

Data Acquisition with CP-2002/2003 Micro-GC Control

Data Acquisition with CP-2002/2003 Micro-GC Control Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598 Star Chromatography Workstation Version 6 Data Acquisition with CP-2002/2003 Micro-GC Control Operation Manual Varian, Inc. 2002

More information

Release Notes Spinsight Version 4.1

Release Notes Spinsight Version 4.1 Release Notes Spinsight Version 4.1 Spinsight 4.1 can be installed over any previous version of Spinsight. It is not necessary that you have 3.5.2 before installing 4.1. Details of new software features

More information

VNMRJ 2.1B Quick overview and BioPack features

VNMRJ 2.1B Quick overview and BioPack features VNMRJ 2.1B Quick overview and BioPack features VNMRJ 2.1B interface is a much more intuitive user interface than the VNMR6.1C. The Java interface is a layer on top of the conventional VNMR so all commands

More information

OnBase Quick Reference Guide Indexing Documents Unity Client

OnBase Quick Reference Guide Indexing Documents Unity Client OnBase Quick Reference Guide Indexing Documents Unity Client Indexing is the process of adding keyword values to documents in order to search for content (retrieve), route items through workflow, report

More information

Basic Navigation in Xtender

Basic Navigation in Xtender Basic Navigation in Xtender https://scan.iit.edu/appxtender/ Xtender is used by Illniois Institute of Technology to scan in paper documents for easy accessibility. Documents scanned within Xtender are

More information

SNS Locator Quick Start Tutorial

SNS Locator Quick Start Tutorial SNS Locator Quick Start Tutorial After starting the program, follow these five steps in order to quickly get started tracking vehicles: 1) Select the local coordinate system - Do this on the Settings dialog

More information

Authorized Send User s Guide Version 4.0

Authorized Send User s Guide Version 4.0 Canon Authorized Send User s Guide Version 4.0 08011-40-UD1-003 This page is intentionally left blank. 2 Authorized Send User s Guide Contents Preface...5 How to Use This Manual... 5 Symbols Used in This

More information

North Carolina State University Department of Chemistry Varian NMR Training Manual

North Carolina State University Department of Chemistry Varian NMR Training Manual North Carolina State University Department of Chemistry Varian NMR Training Manual by J.B. Clark IV & Dr. S. Sankar 1 st Edition 05/15/2009 Section 1: Essential Operations for Basic 1D Spectra Preparing

More information

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP) 500 MHz Solution-state NMR Procedure (Bruker AVANCE Machines running TopSpin under WINDOWS XP) Jerry Hu, x7914, jghu@mrl.ucsb.edu Shamon Walker, x6079 shamonwalker@mrl.ucsb.edu Materials Research Laboratory

More information

Agilent MassHunter Easy Access Software

Agilent MassHunter Easy Access Software Agilent MassHunter Easy Access Software Quick Start Guide Where to Find More Information 3 Installation 4 Sharing Configurations with Multiple Instruments 10 Installed Files 11 Using Easy Access Software

More information