1 DISCLAIMER Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 This document is intended to be a brief, bare-bones user s guide for NMR data collection using Bruker s IconNMR Automation software for the 400 MHz Avance-IIIHD Bruker NMR spectrometer located in Chem93 managed by the UC Davis NMR Facility. For detailed help with both routine and advanced NMR experiments, please consult Bruker s User Guides, which can be found within the Resources tab on our website, nmr.ucdavis.edu. CONVENSIONS Some user control will be done in Bruker s Topspin 3.5 interface. For Topspin interaction, keyboard input is shown as boldface type in this manual. Note that in TS the enter key must be used after the command is typed; this is assumed through-out this manual and enter key strokes are not given explicitly. Commands in TS are typed in on the TS command line near the bottom of the TS window; again this is assumed and will not generally be stated explicitly herein. LMB, MMB, and RMB are used to indicate actions of the left, middle, and right mouse button respectively. On a PC the mouse wheel acts as the MMB. Click, and double click refer to pressing the LMB. GENERAL PROCEDURE You will find that the general procedure for acquiring NMR data on all NMR spectrometers is essentially the same. IconNMR handles the majority of the tasks listed below. The general procedure is as follows: 1 Sample Preparation 2 Login and Startup 3 Setup Initial Parameters 4 Check/Edit Acquisition parameters 5 Insert Sample 6 Lock onto Solvent 7 Tune the Probe 8 Shim 9 Check Receiver Gain 10 Acquire 11 Process Data 12 Remove Sample 13 Logout, Sign Logbook

2 Experiment: Parameter Set: Instrument: Routine Proton NMR PROTON Chem Bruker 400 using IconNMR Sample Preparation 1. Dissolve your sample in an appropriate deuterated NMR solvent. Make sure there is no un-dissolved material. If there is, you will need to either centrifuge or filter your sample to remove crystals/debris. TIP: use a small amount of KimWipe in a Pasteur Pipette to quickly filter out any undissolved material 2. Transfer about ul of solvent into a clean NMR tube. We recommend high-quality NMR tubes - rated 600 MHz or higher, but economy tubes will be OK for routine work at lower fields (400 MHz and below). Your sample height should be about 4-5 cm. Login and Startup 1. Log into your iconnmr user account. If a previous user is currently logged in, you will need to select Change User, then log into your account. NOTE: If your IconNMR account has not yet been created, please reach out to Bennett Addison: jbaddison@ucdavis.edu

3 Set Initial Parameters 1. Identify an empty holder in the BACS120 autosampler. We strongly recommend you use the holders in the inner ring: holders Place your sample into an available blue spinner, set depth using the depth gauge, and drop your sample off in the appropriate holder. Do NOT drop off your sample while the carousel or sample-arm is moving. Use inner ring: Inner ring is the lower numbers, outer ring higher numbers 3. Identify the matching holder in IconNMR menu and double-click to add an experiment. 4. Enter your details: Data Directory, Experiment Name, Experiment Number, Solvent, Experiment/Parameter Set. Experiment Number. The default is increments of 10 (10, 20, 30 ) for each sample/holder, or increments of 1 (10, 11, 12 ) if you are collecting multiple experiments on the same sample/holder. Experiment: Select your experiment from the dropdown list. If there is an experiment you would like to run but it is not on the list, contact Bennett. Directory: Your directory should be based on your PI. It is set automatically when you log in. Experiment Name. Users often include initials, lab notebook page, date, and any additional info if desired. Note do not use spaces, special characters, etc. Solvent: Select your solvent from the drop-down menu

4 Check/Edit Acquisition Parameters 1. If desired, hit the equals button to edit standard Proton acquisition parameters including number of scans, spectral width, acquisition time, offset frequency, etc. Note: The default acquisition parameters are good, but feel free to change them as desired. If you would like to change any other parameters not accessible in this list, you can select Parameters Edit All, which brings you to Topspin. You may need to do this for some 2D experiments. 2. Edit the title if desired. You can add any sample/experiment description in this text box. You may find it useful later for your bookkeeping. 3. Edit the Start Time if desired. The clock is set to 24 hr time. Note, if you set a start time between the hours of 8pm 8am (night-time queue), please make a reservation on the online scheduler. 11 pm Start Time 4. If you would like to also collect another experiment on this sample (e,g, C13CPD), you can highlight the current experiment and then hit the Add button. Another experiment will appear below with experiment number 11 and an empty experiment type. Select the appropriate experiment and edit acquisition parameters if necessary. Details are provided in later pages. Note: This is red only because a parameter (ns) has been changed

5 Submit your Experiment 1. When you are satisfied with all experiments and are ready to collect your data, highlight the appropriate experiment or holder directory, and hit the Submit icon. AUTOMATION PROCESSES: After hitting Submit, the autosampler will take over. The automation software will handle the following processes: Insert Sample, Tune the Probe, Lock onto Solvent, Shim, Check Receiver Gain, Acquire Data, Process Data, Remove Sample, Backup Data to Google Drive. Current Process If Flashing: in Progress If Green: Completed If Clear: Skipped or Not Completed

6 Data Processing 1. To look at your spectra, simply double-click on your completed experiment in the Preceding Experiments list. If data acquisition and processing is complete, you will see a checkmark under Acq and Proc by your experiment. When you double click, you will be sent to Topspin for spectral viewing. Identify your completed experiment and double-click. The list is ordered chronologically. Spectral Processing Commands and Tips: 1. Navigate to the Process tab, and hit Proc. Spectrum button. This will perform Fourier transform (efp), autophase your data (apk), and perform an automatic baseline correction (abs). 2. Click and drag to zoom in 3. Double-click or hit the Full View button to zoom out. 4. Use middle mouse scroll wheel to zoom up and down 5. Click and hold the Hand icon to shift the baseline up or down or side to side 6. Return to IconNMR when done processing / viewing spectrum

7 Logout Procedure: 1. When the autosampler has returned your sample to the holder, you can remove your sample and spinner. Caution: do NOT remove your sample while the carousel is moving, and do NOT reach across the center where the sensor is located. This is especially problematic when there is a sample currently in the magnet. Tripping the sensor with your hand may result in shutdown of the autosampler. If this happens, be sure to call Bennett or an NMR TA. 2. When you are done and no longer using your chosen holder, return to IconNMR, select your completed holder, and hit Delete. The holder should become available. 3. Sign the log book, including Date, Name/PI, Nucleus, Drop-Off Time, Pickup Time, and check if you have deleted your completed run in IconNMR.

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751)

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751) Dissimilar This manual is intended to be only a very brief introduction for using the Bruker 400 MHz NMR spectrometer (PS751) at the California State University LA NMR Facility. For complete information

More information

Bruker Avance 400 MHz Instructions

Bruker Avance 400 MHz Instructions Bruker Avance 400 MHz Instructions General Policy: 1) All users must pass the training course before they will obtain their own access to the spectrometer. 2) A key to Bruker Avance 400 Room is accessible

More information

UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X

UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X TABLE OF CONTENTS UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X VERSION 3 February, 2016 Disclaimer... 2 Conventions... 2 Routine 1D NMR Experiments Procedure for acquiring 1D Proton spectrum...

More information

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500 VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500 August 9, 2016 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they

More information

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation The GVSU chemistry department owns 2 NMR spectrometers. A JEOL Eclipse 300 MHz and a Varian Inova 400 MHz. Due to their strong, constant magnetic

More information

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 January 10, 2019 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they are shorter; you

More information

DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015

DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015 DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015 The Bruker AVANCE III HD 401 has a broadband (BBO) probe with a 1 H Channel and X-Channel tunable from (30-300 MHz). Both channels

More information

General Information for Automated Inova 500 MHz NMR (Indy)

General Information for Automated Inova 500 MHz NMR (Indy) General Information for Automated Inova 500 MHz NMR (Indy) Primary Contacts: NMR Facility Manager: David VanderVelde, x3004, davidv@caltech.edu GLA: Pamela Tadross, x6131, 646-220-5789 (cell), 204 Church,

More information

DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015

DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015 DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015 The Bruker AVANCE III HD 400 has a broadband (BBO) probe with a 1 H Channel and X-Channel tunable from (30-300 MHz). Both channels

More information

Agilent VnmrJ 3.2. Quick Start Guide

Agilent VnmrJ 3.2. Quick Start Guide Agilent VnmrJ 3.2 Quick Start Guide VnmrJ 3.2 Interface 3 Overview 4 Prepare Sample 5 Load Sample 6 Enter Sample Information 7 Build Study Queue 8 Run Study Queue 8 Process Data 9 Plot Data 10 This Quick

More information

NMR Spectrometer Crib-Sheet

NMR Spectrometer Crib-Sheet NMR Spectrometer Crib-Sheet Sample Preparation: 1. Dissolve your sample in a deuterated solvent. For 1 H NMR use ~ 1 mm concentrations. For 13 C NMR, use ~ 10 mm. 2. Use clean dry tubes to avoid contaminating

More information

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer Dr. Zhenming Du Email: zdu@gsu.edu Phone: 3-5538 Revised 04/02/2015 LC=Left click; DC=double

More information

Vanderbilt Small Molecule NMR Center

Vanderbilt Small Molecule NMR Center Vanderbilt Small Molecule NMR Center Instructions for Using TOPSPIN version 2.0 DRX 400: Setting up 1H, 13C (1H decoupled), and 31P NMR Experiments Getting started: Sample Preparation: Make sure your sample

More information

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS April 16, 2018 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they are shorter; you

More information

MNova Version Installation Instructions.

MNova Version Installation Instructions. MNova Version 12.0.0 Installation Instructions. Note: This Document contains information for a new NMR user. Carefully read and accomplish each section before moving to the next. If you can t finish a

More information

c. Saturday and Sunday: 10 minute time blocks and unlimited time.

c. Saturday and Sunday: 10 minute time blocks and unlimited time. Scheduling 400 MHz A with 100 slot sample changer: a. Monday to Friday from 8 am to 8 pm: 10 minute blocks, 30min max. b. Monday to Friday from 8 pm to 8 am: 10 minute blocks, unlimited time. c. Saturday

More information

North Carolina State University Department of Chemistry Varian NMR Training Manual

North Carolina State University Department of Chemistry Varian NMR Training Manual North Carolina State University Department of Chemistry Varian NMR Training Manual by J.B. Clark IV & Dr. S. Sankar 1 st Edition 05/15/2009 Section 1: Essential Operations for Basic 1D Spectra Preparing

More information

Acquiring Data in VnmrJ 4.2A

Acquiring Data in VnmrJ 4.2A Acquiring Data in VnmrJ 4.2A Linux Primer Initial Steps Changing Password Disabling Screen Lock Reservations Reservation Terminals Rules and Proper Etiquette System Identification Working with VnmrJ 4.2A

More information

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry NMR Spectroscopy with VnmrJ University of Toronto, Department of Chemistry Walk-up interface 1 Logging in 1 Starting VnmrJ 1 Inserting sample into the magnet or sample changer 1 Enter sample information

More information

NMR Data Acquisition and Processing Procedure

NMR Data Acquisition and Processing Procedure NMR Data Acquisition and Processing Procedure Dr. Jianfeng Zhu (Research Officer) Please DO NOT remove from NMR lab! Things to Avoid when Using NMR Lab Here's a list of stuff that I have seen that you

More information

Bruker Topspin NMR Training Manual

Bruker Topspin NMR Training Manual Bruker Topspin NMR Training Manual Rev 180521 Perry Pellechia University of South Carolina NMR Facility Subject Page Number Overview of Operation... 1 Gauging Samples in Spinners... 2 Starting and logging

More information

Bruker NMR Topspin Manual

Bruker NMR Topspin Manual Bruker NMR Topspin Manual Topspin is an icon driven software package. This means that instead of reading words and typing commands in the software most of the things will be carried out by clicking icons

More information

1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR. Spectrometer without the Sample Changer and IconNMR

1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR. Spectrometer without the Sample Changer and IconNMR 1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR Spectrometer without the Sample Changer and IconNMR Scheduling NMR Time: Any use of the 700 MHz NMR spectrometer requires prior

More information

Varian Solution NMR Automation Procedure

Varian Solution NMR Automation Procedure Menu System System Tool Bar Command Line Graphics Tool Bar Vertical Panels Protocols NMR Data Display NMR Graphics Area Study Queue Horizontal Panels Activity Monitoring Varian Solution NMR Automation

More information

DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400

DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400 1 DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400 400: The Bruker DPX 400 has a broadband probe with 1 H Channel and X-Channel tunable from (30-300 MHz). Both Channels may be tuned researchers.

More information

Bruker AVANCE III HD 400 MHz Spectrometer Operation

Bruker AVANCE III HD 400 MHz Spectrometer Operation Bruker AVANCE III HD 400 MHz Spectrometer Operation The spectrometer is located in Room 312 of Physical Sciences I. It is equipped with a SampleXpress sample changer for full automation. To utilize the

More information

Quick Guide to Topspin

Quick Guide to Topspin Quick Guide to Topspin 1. Log into the workstation. Right mouse click on the desktop anywhere and select new terminal. Type topspin [enter] (if this fails, create the launcher as described next). Alternatively,

More information

Running Multiple Experiments on the Same Sample Automatically. (multizg, spooler/queue, iconnmr)

Running Multiple Experiments on the Same Sample Automatically. (multizg, spooler/queue, iconnmr) Running Multiple Experiments on the Same Sample Automatically (multizg, spooler/queue, iconnmr) Multiple Experiment Setup If you have one sample that requires full characterization (e.g. 1 H, 13 C, COSY,

More information

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1)

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1) Menu System System Tool Bar Command Line Graphics Tool Bar Vertical Panels Protocols NMR Data Display NMR Graphics Area Study Queue Horizontal Panels Activity Monitoring Varian Solution NMR Automation

More information

Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12

Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12 Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12 DOSY Diffusion Ordered SpectroscopY. NMR diffusion experiments, like DOSY, are used to determine the diffusion coefficients of solute

More information

Processing data with Bruker TopSpin

Processing data with Bruker TopSpin Processing data with Bruker TopSpin This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and

More information

UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS

UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS TABLE OF CONTENTS Disclaimer... 2 Conventions... 2 Routine 1D NMR Experiments Procedure for acquiring 1D Proton spectrum...

More information

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin.

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin. TopSpin QuickStart Preliminaries Sign Up Sign up for instrument time at calendar.yahoo.com/hmcnmr The password is nmr. During the summer, 1100-1200 and 1600-1700 are reserved for walk-on use (15 min maximum

More information

Varian Solution NMR Procedure

Varian Solution NMR Procedure System Tool Bar Command Line Vertical Panels Protocols Menu System User Tool Bar NMR Data Display Graphics Tool Bar NMR Graphics Area Study Queue Horizontal Panels Hardware Bar Varian Solution NMR Procedure

More information

Basic Procedure 1D NMR Data Collection

Basic Procedure 1D NMR Data Collection Basic Procedure 1D NMR Data Collection Log into your UNIX account and start XWIN- NMR program At the prompt login:, type your (e.g. jdw, mmj, mjt) or click on your group icon (in some spectrometers

More information

NMR Users Guide Organic Chemistry Laboratory

NMR Users Guide Organic Chemistry Laboratory NMR Users Guide Organic Chemistry Laboratory Introduction The chemistry department is fortunate to have a high field (400 MHz) Nuclear Magnetic Resonance (NMR) spectrometer. You will be using this instrument

More information

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP) 500 MHz Solution-state NMR Procedure (Bruker AVANCE Machines running TopSpin under WINDOWS XP) Jerry Hu, x7914, jghu@mrl.ucsb.edu Shamon Walker, x3248 shamonwalker@mrl.ucsb.edu Materials Research Laboratory

More information

IconNMR Operating Procedures

IconNMR Operating Procedures IconNMR Operating Procedures Revised: 7-1-2015 This guide provides basic instructions for acquiring NMR spectra using the IconNMR automation program. 1. Insert your sample(s) in the SampleXpress sample

More information

VnmrJ Walkup. Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No , Rev. 0604

VnmrJ Walkup. Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No , Rev. 0604 VnmrJ Walkup Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No. 01-999266-00, Rev. 0604 VnmrJ Walkup Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No. 01-999266-00, Rev. 0604 VnmrJ Walkup

More information

Welcome to the NMR Center

Welcome to the NMR Center 1 Welcome to the NMR Center The NMR center in chemistry consists of six Bruker NMR instruments Field/Name Location Comments Operating Software 300nb (narrow bore) 119 Havemeyer walk-on, routine 1 H Topspin

More information

NMR Training VNMRJ 4.2A/VNMRS 400

NMR Training VNMRJ 4.2A/VNMRS 400 NMR Training VNMRJ 4.2A/VNMRS 400 The VNMRS/new 400 MHz NMR uses VNMRJ 4.2A. Although VNMRJ is different than VNMRalmost ALL typed commands that you know from VNMR should work. LOGIN: Your username is

More information

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy PINMRF Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy INCLUDING: AV-III-400-HD w/ 5mm BBFO SmartProbe 369 WTHR AV-III-500-HD w/ 5mm BBFO

More information

Seized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division

Seized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division SHIMADZU ULTRAVIOLET/VISIBLE SPECTROPHOTOMETER (UV/VIS) Instrument Shimadzu UV-2401 UV/VIS Spectrophotometer (Serial Number

More information

PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University

PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University 1. INTRODUCTION...1 1.1. About this Worksheet... 1 1.2. A Very Brief

More information

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample

More information

EQNMR Facility Rules of Use

EQNMR Facility Rules of Use New User Package for East Quad 600 NMR Date: 01/25/08 Welcome to the East Quad NMR facility. This document contains information for new users using the 600MHz NMR at our facility. In this package, you

More information

VnmrJ Walkup. Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No , Rev. A 0808

VnmrJ Walkup. Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No , Rev. A 0808 VnmrJ Walkup Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No. 01-999377-00, Rev. A 0808 VnmrJ Walkup Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software

More information

SmartSoft AES Operator s Guide

SmartSoft AES Operator s Guide SmartSoft AES Operator s Guide Part No. 702064 Rev. A Physical Electronics USA, PHI, SMART-Tool, SmartSoft, MultiPak and Watcher are trademarks of ULVAC-PHI, Inc. All other trademarks are the property

More information

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy Delta V5 operation Widetron Technologies Corp. Wu, Chung Ying Connect to spectrometer 1. Start Delta V5 (double click ). 2. Open Spectrometer Control (click ). 3. Select spectrometer on Available Instruments.

More information

ONE DIMENSIONAL NMR SPECTROSCOPY. DPX 200 and DPX 300

ONE DIMENSIONAL NMR SPECTROSCOPY. DPX 200 and DPX 300 1 1H, 13C, 15N, 19F and 31P ONE DIMENSIONAL NMR SPECTROSCOPY (Xwinnmr on PC, W2Kpro, version 3.5) on DPX 200 and DPX 300 Department of Chemistry University of Oslo February 2005 Version 1 (To be used from

More information

TopSpin 1.3 patch level 10. DPX 200, DPX 300 and DRX 500. Windows XP

TopSpin 1.3 patch level 10. DPX 200, DPX 300 and DRX 500. Windows XP KJM 5280 1 H nmr introductory user manual I Version 1.5 TopSpin 1.3 patch level 10 DPX 200, DPX 300 and DRX 500 Windows XP F. Rise May 2018 1 Welcome to the wonderful and amazing world of Nuclear Magnetic

More information

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP) 500 MHz Solution-state NMR Procedure (Bruker AVANCE Machines running TopSpin under WINDOWS XP) Jerry Hu, x7914, jghu@mrl.ucsb.edu Shamon Walker, x6079 shamonwalker@mrl.ucsb.edu Materials Research Laboratory

More information

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised 4-21-2015 Relaxation times are divided into two types: longitudinal, which concerns change in magnetization along the

More information

Troubleshooting Acquisition Related Problems

Troubleshooting Acquisition Related Problems Troubleshooting Acquisition Related Problems This handout contains pertinent information to help you when a problem arises. Please look through the topics below to identify and find resolutions to the

More information

Getting low-resolution GC-MS data on the JEOL GCMate Introduction

Getting low-resolution GC-MS data on the JEOL GCMate Introduction Getting low-resolution GC-MS data on the JEOL GCMate Introduction GC-MS is a technique that combines gas phase separation and mass-based detection/characterization of analytes. It s a very sensitive method

More information

CONTENTS 1D EXPERIMENTS D EXPERIMENTS APPENDIXES... 83

CONTENTS 1D EXPERIMENTS D EXPERIMENTS APPENDIXES... 83 Jari Koivisto, Ph.D. Aalto University School of Science and Technology Faculty of Chemistry and Materials Sciences Department of Chemistry Bruker Avance DPX400 User Manual Basic and Advanced 1D and 2D

More information

PINMRF. Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy

PINMRF. Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy PINMRF Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy INCLUDING: Inova-300-1 w/ 5mm 4-nucleus probe 365 WTHR Inova-300-2 w/ 5mm 4-nucleus probe 4100 BRWN Table of Contents

More information

KJM Proton T1 Spectra on the AVI-600 and AVII-600. Version 1.0

KJM Proton T1 Spectra on the AVI-600 and AVII-600. Version 1.0 KJM 9250 Proton T1 Spectra on the AVI-600 and AVII-600 Version 1.0 Professor Emeritus Alistair Lawrence Wilkins, University of Waikato, New Zealand. February 2018 Proton T1 Spectra on the AVI-600 and AVII-600

More information

Unscrew and remove the cap from the waste tank

Unscrew and remove the cap from the waste tank Stanford Cytoflex HTS User Guide 11232015 Starting up 1. Check the sheath fluid level every time you use the cytometer. This ensures that you do not run out of sheath fluid during an experiment. Replenish

More information

IconNMR (>TopSpin 3.0)

IconNMR (>TopSpin 3.0) IconNMR (>TopSpin 3.0) Dr. Sven Augner Bruker BioSpin, Rheinstetten, Germany Innovation with Integrity How to start an IconNMR set 2 IconNMR Select the respective holder 2x IconNMR Select the disk Base

More information

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800 Walkup NMR Varian NMR Spectrometer Systems With VNMR 6.1C Software Pub. No. 01-999159-00, Rev. A0800 Walkup NMR Varian NMR Spectrometer Systems With VNMR 6.1C Software Pub. No. 01-999159-00, Rev. A0800

More information

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted)

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted) Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted) Once FTIR data has been acquired for the gases evolved during your DSC/TGA run, you will process using the OPUS software package. Select

More information

IS2000. Administrative Operator s Guide

IS2000. Administrative Operator s Guide IS2000 Administrative Operator s Guide Table of Contents Logging Off... 7 Event Manager... 7 HARDWARE MANAGER... 8 Maneuvering the Hardware Tree... 8 Unlocking the Module... 8 Viewing the Hardware Tree...

More information

Premiere Pro Desktop Layout (NeaseTV 2015 Layout)

Premiere Pro Desktop Layout (NeaseTV 2015 Layout) Premiere Pro 2015 1. Contextually Sensitive Windows - Must be on the correct window in order to do some tasks 2. Contextually Sensitive Menus 3. 1 zillion ways to do something. No 2 people will do everything

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in the Thermo TraceFinder 3.0 analytical software. For detailed descriptions

More information

Windows 8.1 User Guide for ANU Staff

Windows 8.1 User Guide for ANU Staff Windows 8.1 User Guide for ANU Staff This guide has been created to assist with basic tasks and navigating Windows 8.1. Further tips for using Windows 8.1 can be found on the IT Services website, or by

More information

TOCSY 15. Goto. Introduction Pulse Sequence Diagram AVANCE User s Guide Bruker 167

TOCSY 15. Goto. Introduction Pulse Sequence Diagram AVANCE User s Guide Bruker 167 Chapter TOCSY 15 Introduction 15.1 Goto TOtal Correlation SpectroscopY provides a different mechanism of coherence transfer than COSY for 2D correlation spectroscopy in liquids. In TOCSY, cross peaks are

More information

Lab Assignment #1: Introduction to Creo ME 170

Lab Assignment #1: Introduction to Creo ME 170 Lab Assignment #1: Introduction to Creo ME 170 Instructor: Mike Philpott (email: mphilpot@illinois.edu) Date Due: One week from Start Day of Lab (turn in deadline 11pm night before next lab) Make sure

More information

Fitting NMR peaks for N,N DMA

Fitting NMR peaks for N,N DMA Fitting NMR peaks for N,N DMA Importing the FID file to your local system Any ftp program may be used to transfer the FID file from the NMR computer. The description below will take you through the process

More information

MestReC Cheat Sheet. by Monika Ivancic, July 1 st 2005

MestReC Cheat Sheet. by Monika Ivancic, July 1 st 2005 MestReC Cheat Sheet by Monika Ivancic, July 1 st 2005 This Cheat Sheet is to be used at UW-Madison as a quick guide to processing using the MestReC NMR software. You may find more help at the MestReC homepage

More information

Manual for Jasco-IR-spectrometer

Manual for Jasco-IR-spectrometer Manual for Jasco-IR-spectrometer Do not record spectra of hazardous volatile liquids! Do not record spectra of corrosive compounds! Do not record spectra of hard and sharp edged material (sand, broken

More information

Forms/Distribution Acrobat X Professional. Using the Forms Wizard

Forms/Distribution Acrobat X Professional. Using the Forms Wizard Forms/Distribution Acrobat X Professional Acrobat is becoming a standard tool for people and businesses to use in order to replicate forms and have them available electronically. If a form is converted

More information

ON!Track User Manual. Mobile Release Manual Version 1.1

ON!Track User Manual. Mobile Release Manual Version 1.1 ON!Track User Manual Mobile Release 2.3.1 Manual Version 1.1 Table of Content What is ON!Track?... 2 How to download ON!Track from the App Store?... 3 How to navigate in ON!Track Mobile App?... 8 How to

More information

DCIF NMR Training Guide Varian Mercury 300 MHz Last Update

DCIF NMR Training Guide Varian Mercury 300 MHz Last Update DCIF NMR Training Guide Varian Mercury 300 MHz Last Update 10-12-07 1 This document is intended as an introductory guide to the Varian VNMR NMR processing / acquisition software. It is by no means intended

More information

TopSpin 3.5 Tips and Tricks. Nicole Kruse US Applications Pre PANIC Workshop February 20 th 2017

TopSpin 3.5 Tips and Tricks. Nicole Kruse US Applications Pre PANIC Workshop February 20 th 2017 TopSpin 3.5 Tips and Tricks Nicole Kruse US Applications Pre PANIC Workshop February 20 th 2017 Non-Uniform Sampling Incorporated into TopSpin 3.0 Higher Resolution NUS HSQC NS = 2 TD = 2048 NUS @ 10 %

More information

IMSERC NMR MANUAL 02: Basic Processing of Varian 1D NMR Data

IMSERC NMR MANUAL 02: Basic Processing of Varian 1D NMR Data IMSERC NMR MANUAL 02: Basic Processing of Varian 1D NMR Data Last updated: July 15, 2011 by Josh Kurutz This manual describes how to process NMR data on the offline processing computer in the IMSERC lab

More information

Ortelia Space Builder User Manual

Ortelia Space Builder User Manual Ortelia Space Builder User Manual http://ortelia.com 1 Table of Contents Introducing Ortelia Space Builder...2 System Requirements...3 1. Operating system:... 3 2. Hardware:... 3 Minimum Graphics card

More information

PANalytical Epsilon 5 XRF spectral acquisition, processing and reporting

PANalytical Epsilon 5 XRF spectral acquisition, processing and reporting QSM Approval: PANalytical Epsilon 5 XRF spectral acquisition, processing and reporting 1. Introduction and Scope 1.1. This Standard Operating Procedure provides instruction for the operation, acquisition

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in Thermo TraceFinder analytical software. For detailed descriptions

More information

PRINTING GROWER FIELD MAPS OFF THE WEB

PRINTING GROWER FIELD MAPS OFF THE WEB PRINTING GROWER FIELD MAPS OFF THE WEB 3-29-2018 A. Google Earth: Very easy to use; easy to print map (either directly or via extraction to Word); easy to scale up or down; locator makes finding field

More information

XPS1 Automated Multi-Sample Run Procedure

XPS1 Automated Multi-Sample Run Procedure XPS1 Automated Multi-Sample Run Procedure Follow the XPS Operating Procedure to load samples into the SAC chamber. Once the samples are in the SAC chamber, the following procedure can be used to automate

More information

Determination of Drag Coefficient

Determination of Drag Coefficient DEPARTMENT OF MECHANICAL, INDUSTRIAL AND MANUFACTURING ENGINEERING MIMU 505 - MEASUREMENT AND ANALYSIS Determination of Drag Coefficient You will need to bring a zip disk or USB storage device to the lab

More information

BUCKVIEW Advanced. User Guide

BUCKVIEW Advanced. User Guide BUCKVIEW Advanced User Guide Inside This Manual I. Inside This Manual... 2 II. Viewing and Managing Your Images... 3 Manage Image Folders...5 Manage Sites...6 Manage Locations...10 Erase Memory Card...14

More information

Agilent 6400 Series Triple Quadrupole LC/MS System

Agilent 6400 Series Triple Quadrupole LC/MS System Agilent 6400 Series Triple Quadrupole LC/MS System Quick Start Guide Where to find information 4 Getting Started 6 Step 1. Start the Data Acquisition software 7 Step 2. Prepare the LC modules 13 Step 3.

More information

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer Operating instructions - Cary 100 Bio UV-visible Spectrophotometer This document describes how to power up the spectrophotometer, set its measurement parameters, insert one or more of samples for analysis

More information

vbound User Guide vbound User Guide Version Revised: 10/10/2017

vbound User Guide vbound User Guide Version Revised: 10/10/2017 vbound User Guide Version 4.1.1 Revised: 10/10/2017 Copyright 2014-2017 FFL Solutions Inc. Page 1 of 87 Table of Contents Using vbound...5 Starting vbound... 5 Bound Book List... 6 vbound Ribbon Menu...

More information

Agilent G6854 MassHunter Personal Pesticide Database

Agilent G6854 MassHunter Personal Pesticide Database Agilent G6854 MassHunter Personal Pesticide Database Quick Start Guide What is MassHunter Personal Pesticide Database? 2 Installation 3 Main Window 4 Getting Started 11 Database operations 12 Searching

More information

Navigating and Logging Out of Meditech

Navigating and Logging Out of Meditech Navigating and Logging Out of Meditech There are some basic tips when using Meditech: There are times when Meditech requires a command and it will not recognize the command if it is lower case and it expects

More information

CHEMISTRY DEPARTMENT Category Work Instruction

CHEMISTRY DEPARTMENT Category Work Instruction No. of pages: 1 of 10 1. PURPOSE The purpose of this procedure established is to provide clear instructions for steps or orders on how to run sample analysis using JEOL NMR ECX 400MHz SYSTEM JOEL equipment.

More information

Contributing Lists in Activity Center

Contributing Lists in Activity Center Contributing Lists in Activity Center Overview This tour details the use of Lists in Activity Center, including: 1. Selecting Contributing Lists for Student Input 2. Configuration Options for Contributing

More information

Welcome To Autotrak Monitor +

Welcome To Autotrak Monitor + Welcome To Autotrak Monitor + Dear Valued Customer welcome to the Monitor+. Here you will find an easy to follow break down of the Monitor+ website and examples to help you along the way. Please do not

More information

EMC Documentum TaskSpace

EMC Documentum TaskSpace EMC Documentum TaskSpace Version 6.7 User Guide EMC Corporation Corporate Headquarters: Hopkinton, MA 01748 9103 1 508 435 1000 www.emc.com EMC believes the information in this publication is accurate

More information

On Artemis and Cronus, make sure you are adjusting the db scale and not the watts

On Artemis and Cronus, make sure you are adjusting the db scale and not the watts Solvent Suppression - Presaturation (NOTE: There are many types of solvent suppression this is presaturation and all exchangeable protons will be attenuated-instructions for other types follow) 1. Turn

More information

MALDI Software Users Guide (simplified version)

MALDI Software Users Guide (simplified version) MALDI Software Users Guide (simplified version) Central Analytical Lab Ekeley E266 The Acquisition program is called Voyager Control Panel. Once you shoot your sample and determine it looks good, you transfer

More information

Welcome To Autotrak Alert + Help Menu

Welcome To Autotrak Alert + Help Menu Welcome To Autotrak Alert + Help Menu Dear Valued Customer welcome to Alert+. In this document you will find an easy to follow break down of the Alert+ website and examples to help you along the way. Please

More information

Polycom VSX 700 Notes

Polycom VSX 700 Notes Polycom VSX 700 Notes Speaker / subwoofer Polycom VSX 7000 Microphone Remote Using the Remote Placing a call There are 3 options for placing a call- 1. Enter the IP number of the site you wish to call

More information

Exporting distribution lists from Thunderbird to Outlook

Exporting distribution lists from Thunderbird to Outlook Exporting distribution lists from Thunderbird to Outlook PLEASE NOTE: Do not export the lists under Distribution Lists in your Thunderbird Address Book as these will no longer be maintained on the new

More information

Windows 8.1 User Guide for ANU Staff

Windows 8.1 User Guide for ANU Staff Windows 8.1 User Guide for ANU Staff This guide has been created to assist with basic tasks and navigating Windows 8.1. Further tips for using Windows 8.1 can be found on the IT Services website, or by

More information

Introduction And Overview ANSYS, Inc. All rights reserved. 1 ANSYS, Inc. Proprietary

Introduction And Overview ANSYS, Inc. All rights reserved. 1 ANSYS, Inc. Proprietary Introduction And Overview 2006 ANSYS, Inc. All rights reserved. 1 ANSYS, Inc. Proprietary The ANSYS Workbench represents more than a general purpose engineering tool. It provides a highly integrated engineering

More information

Process Eye Professional. Recall

Process Eye Professional. Recall Process Eye Professional Recall Process Eye Professional Recall User Manual SP104010.101 August 2005 As part of our continuous product improvement policy, we are always pleased to receive your comments

More information