NANOSPEC 4150 STANDARD OPERATING PROCEDURES

Size: px
Start display at page:

Download "NANOSPEC 4150 STANDARD OPERATING PROCEDURES"

Transcription

1 NANOSPEC 4150 STANDARD OPERATING PROCEDURES Version: 1.0 JAN 2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center

2 TABLE OF CONTENTS 1.0 INTRODUCTION HARDWARE OPERATING PROCEDURES LOADING A WAFER FOR TESTING SELECTING A RECIPE USING F3 AUTO MODE RUNNING A RECIPE USING F8 OR F9 MANUAL MODES DATABASE 13 2

3 1.0 Introduction The Model 4150 NanoSpec is an instrument designed to measure the thickness of films deposited on various substrates used in semiconductor and MEMs fabrication. The Model 4150 provides accurate and repeatable measurement s for films ranging from 15nm - 75,000nm in thickness. Two layer films may also be measured on the 4150, but the lower layer thickness must be known. 2.0 Hardware Visible light source Motorized X, Y & Z axis stage System accepts 75mm to 200mm wafer sizes Auto focusing Range of apertures, 5X, 10X, 15X & 50x Custom software to handle measurements and database. 3

4 3.0 Operating Procedures 3.1 Loading a Wafer for Testing Log into the main system computer by entering NANO as the user ID. Then tab down to the Password box and enter The measurement software should now open You will notice two separate wafer loading platforms on the system (Fig. 1). Load the wafer to be measured onto the left side platform. Be careful to get the wafer centered on the platform. There are stops to help center full sized wafers on the stage platform. Fig The right side platform is for a bare silicon reference wafer. A reference is only necessary if it has been longer than 24hrs since the previous reference was performed by the system. Load your bare clean silicon reference wafer onto the right hand platform to run the reference scan. NOTE -This SOP will describe the process of running both the reference scan and measurement scans. If it has been < 24hrs since the prior reference scan was performed then you do not need the reference scan (answer NO when prompted to run it) and you may just skip the following steps which refer to it. 4

5 3.2 Selecting a Recipe using F3 Auto Mode At the top of the computer screen you will see a number of software tabs numbered F1 F10 (Fig. 2). These tabs correspond to your Keyboard F keys. The tabs you ll use to select a recipe will depend on the type of measurement you are going to perform. Tab F3 Auto will run a measurement recipe based on a multi-point preprogrammed map of the wafer. The total measurement points are configurable by staff from 1 to 49 points. The 5 and 9 point tests are the most common on a full sized wafer. Tab F8 User will allow a custom single point test to be run by the user. The recipes in tab F9 STD are single point tests that were set up by the vendor. Pressing the particular F-button will open a second window to allow viewing of the recipe types available for that particular tab. Fig Now press F3 Auto and the system will ask if you want to save the measurements in the database. Answer YES to store the measurements for future access After this another prompt box will open asking for you to enter Parameters. Enter a name or ID that you would like to store the data under. 5

6 3.2.4 Next a recipe selection window will open (Fig. 3). The recipes listed display the recipe #, the # of measurement points, the film type and the substrate. Select the appropriate recipe # from this window and press ENTER. Fig Now the system will prompt as to whether it s been > 24 hours and if you need to run a reference scan. Answer accordingly. If you answer NO then skip ahead to step If you selected YES then the system will tell you to Prepare Substrate for Reference This is your cue to put the clean bare silicon wafer onto the right hand stage if you have not already done so, then press enter. The system will then prompt as to whether or not you would like for the stage to move to the reference position automatically, answer YES. The stage will now move to the center of the right hand platform. Another window will open notifying you that the joystick is active so that you may manually move to a different reference position if required. Once the wafer is properly positioned you are ready to take the reference scan press the OK button and the scan will commence. 6

7 3.2.6 If you selected a two layer measurement recipe in step then a window will now open with a prompt to enter the UNDERLAYER thickness. This is your first layer on the wafer. You must know what this thickness is as the system cannot measure and identify both layers. This step will be skipped if a one layer measurement recipe was selected prior Next the system will ask whether the Cassette/Wafer is ready (Fig. 4). This is referring to the wafer to be measured. If you have not already done so sit your wafer or sample centered onto the left hand platform then press OK. Fig. 4 7

8 3.2.8 The stage will now move to the first measurement point which is programmed into the system. The Joystick Active prompt will come up and you may move the stage manually using the joystick to the correct location if needed. You may also manually focus on the test area if needed by rotating the large dial under the left hand side of the stage (Fig. 5). Once you are sure the measurement point is in view of the camera then select OK. Fig A final box will alert you to Place Wafer, Close UV Shutter, Focus, Etc. Nothing should be needed as this should have all been completed by now. Do not worry about the UV shutter we are not using the UV lamp on this system so the UV shutter will always be closed. Press OK to begin the measurement The system will now begin measuring the first test point. This takes approx. 5 seconds and then the system moves automatically to the next test point The Joystick Active prompt will come on again once the next test point is under the objective. You may move the stage to a better test position if needed. The press OK to begin the next measurement. 8

9 Repeat and continue this process until all of the measurement data points have been collected. Then a final prompt will open asking if you would like to Continue for Another Wafer Y/N. Answer No to this and the screen will now display your individual test point measurements and the Max, Min, Mean, Standard Dev, and Uniformity (Fig. 6) for the test points you just measured. In addition, you may view the Fit curve for the measured data by pressing the F2 Graph button on the screen. This will pull up the fit graph in the lower right hand corner of the screen. Pressing the F2 button again will remove it from the screen. Fig Once you are through you may Press the Alt F6 buttons to once again activate the joystick and manually move the stage to a position which is easy to remove your test sample and the reflectance substrate When you are finished with the System then Press the F5 Logoff button. 9

10 3.3 Running a Recipe using F8 or F9 Manual Modes These two modes are primarily for single point tests or users which are using partial wafers. See step first then proceed to the next step Press F8 for all User configured recipes or F9 for all Standard recipes that were loaded by the vendor. The system will Ask if you would like to store the measurements select Yes After this another prompt box will open asking for you to enter Parameters. Enter a name or ID that you would like to store the data under. It would be best to use your name and a date code. Once you enter this another Parameters box for the wafer will open. Enter a wafer ID if you like Next a Program Selection window will open (Fig. 7). The recipes listed display the recipe #, the # of measurement points, the film type and the substrate. Select the appropriate recipe # from this window and press ENTER. Fig. 7 10

11 3.3.5 You will now be prompted to identify which lens you are using (Fig. 8). Look at the lens turret on the microscope and on the side of the objective in use is printed the magnification of the lens. Typically 5X or 10X lenses are used. Make sure the Selected Lens matches and press Enter. Fig If you selected a two layer measurement recipe in step then a window will now open with a prompt to enter the UNDERLAYER thickness. This is your first layer on the wafer. You must know what this thickness is as the system cannot measure and identify both layers. It will only measure the top layer. This step will be skipped if a one layer measurement recipe was selected prior Now the system will prompt as to whether it s been > 24 hours and if you need to run a reference scan. Answer accordingly. If you answer NO then skip ahead to step If you selected YES then the system will tell you to Prepare Substrate for Reference This is your cue to put the clean bare silicon wafer onto the right hand stage if you have not already done so, then press enter. The system will then prompt as to whether or not you would like for the stage to move to the reference position automatically, answer YES. The stage will now move to the center of the right hand platform. Another window will open notifying you that the joystick is active so that you may manually move to a different reference position if required. Once the wafer is properly positioned you are ready to take the reference scan press the OK button and the reference scan will commence. 11

12 3.3.8 Once you have finished the prior step you will need to Press Alt F6 simultaneously to activate the joystick. Manually move the right stage which contains your sample under the measuring objective. Once the sample is positioned under the objective for measurement press OK Now you may adjust the Focus if necessary of the sample image by manually rotating the stage adjustment (See Fig. 5) You are now ready to measure, Press F10 to perform a single point measurement A final box will alert you to Place Wafer, Close UV Shutter, Focus, Etc. Nothing should be needed as this should have all been completed by now. Do not worry about the UV shutter we are not using the UV lamp on this system so the UV shutter will always be closed. Press OK to begin the measurement. The measurement will take approx. 5 seconds to complete If you wish to perform another single point measurement proceed to Step and move the stage to a new test point using the joystick. Continue as needed using steps 3.3.8, & until all test points have been measured then proceed to the next step Once you are finished with the testing press button F6 Done to display the calculated data for the measured test points (See Fig. 6). A parameters prompt will come on the screen for you to enter or modify the Wafer ID. Enter information if needed then press OK You may now press button F2 to display the Fit Curve graph for your measurement data in the lower right hand corner. Pressing F2 again will turn the graph off If you are finished measuring samples with the system then press the F5 Logoff button. 12

13 4.0 Database To access the stored user database files press the F1 button from the main software screen (See. Fig. 2) This will open a new window (Fig. 9)from which you can pick out a stored data file that you would like to view. Files are listed by date and the ID information which was entered by the user in the Parameters box in steps or Fig Highlight the desired file then click on the Show Measurement Data button on top of the screen. 13

14 4.1.4 A new window will open displaying the data file. Scroll to the right in this window and you will see the thickness measurements. Highlight the column of measurements you wish to view (Fig. 10). Then click on the MAP button at the bottom right of the data file window. Fig A new window will open displaying the map of the wafer and the measured points (Fig. 11). Along with the wafer map will be the other statistics for the measured data. Fig Once you are finished viewing the Wafer Map click on File from the header and then Exit. You will now be at the main Data selection window. To exit this window, once again click on File from the header and select Exit to return to the main system measurement screen. 14

ABM's High Resolution Mask Aligner Features:

ABM's High Resolution Mask Aligner Features: ABM's High Resolution Mask Aligner is a very versatile instrument with interchangeable light sources which allow Near-UV (405-365 nm) as well as Mid- and Deep-UV (254 nm, 220 nm) exposures in proximity

More information

R I T Title: Nanometrics Nanospec AFT #1 Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 03/28/2011

R I T Title: Nanometrics Nanospec AFT #1 Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 03/28/2011 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Nanometrics Nanospec AFT. All users are expected to have read and understood this

More information

NANOSPEC 6100 SERIES OPERATOR USERS MANUAL. Automated Film Thickness Measurement System P/N P3

NANOSPEC 6100 SERIES OPERATOR USERS MANUAL. Automated Film Thickness Measurement System P/N P3 NANOSPEC 6100 SERIES OPERATOR USERS MANUAL Automated Film Thickness Measurement System P/N 7500-1092 P3 Nanometrics Incorporated 1550 Buckeye Drive, Milpitas, California 95035-7418 Telephone 408 435 9600

More information

Filmetrics F40 UV Thin Film Measurement System Standard Operating Procedure

Filmetrics F40 UV Thin Film Measurement System Standard Operating Procedure Filmetrics F40 UV Thin Film Measurement System Standard Operating Procedure NYU Tandon School of Engineering Nanofabrication Facility Contents 1 Startup 2 Baseline and reference check 2.1 Baseline setting

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: Film-Sense Coral Name: film-sense Revision Number: 1 Model: FS-1 Revisionist: T. Whipple Location: Bay 1 Date: 5/16/2017 1 Description The Film Sense FS-1 Multi-Wavelength Ellipsometer

More information

` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE

` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE ` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE Version: 1.0 April-2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research Center (NRC) TABLE OF CONTENTS 1. Introduction...2 1.1

More information

FILMETRICS F20 STANDARD OPERATION PROCEDURE

FILMETRICS F20 STANDARD OPERATION PROCEDURE Arizona State University NanoFab FILMETRICS F20 STANDARD OPERATION PROCEDURE Rev C Table of Contents Contents Table of Contents...1 1. Purpose / Scope...2 2. Reference Documents...2 3. Equipment / Supplies

More information

LEXT 3D Measuring LASER Microscope

LEXT 3D Measuring LASER Microscope LEXT 3D Measuring LASER Microscope Warning: This instrument may only be operated by those who have been trained by AAF staff and have read and signed the AAF laboratory policies. A) STARTUP 1. Computer

More information

for R&D and Production Advanced Technology OAI is a Silicon Valley-based manufacturer of advanced precision equipment

for R&D and Production Advanced Technology OAI is a Silicon Valley-based manufacturer of advanced precision equipment Advanced Technology for R&D and Production OAI is a Silicon Valley-based manufacturer of advanced precision equipment for the MEMS, Semiconductor, Nanotechnology, Microfluidics, MEMS Micro TAS and Flat

More information

SUSS SB6E BONDER SOP. October 2013

SUSS SB6E BONDER SOP. October 2013 SUSS SB6E BONDER SOP October 2013 The SUSS SB6 bonder is capable of silicon fusion, anodic, and eutectic bonding. There are two separate heads; one for anodic bonding, and another for fusion bonding. Definitions

More information

NRF Ellipsometer SOP Revision /19/15 Page 1 of 14. Ellipsometer SOP

NRF Ellipsometer SOP Revision /19/15 Page 1 of 14. Ellipsometer SOP Page 1 of 14 Ellipsometer SOP The J. A. Woollam M88 is a spectroscopic ellipsometer used to measure film thickness and optical constants of transparent/semi-transparent thin films. It uses a Xenon arc

More information

Perkin-Elmer Lambda 3B Spectrophotometer

Perkin-Elmer Lambda 3B Spectrophotometer Written by Mark Csele Written by Jeff Brown Operation Turn on the Lambda 3B and the Computer Remove all samples from the sample chamber. Turn on the lamp for the range being investigated using the switches

More information

Operating Procedure for Horiba Raman Microscope

Operating Procedure for Horiba Raman Microscope Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch

More information

Equipment Standard Operating Procedure Kimberly Appel

Equipment Standard Operating Procedure Kimberly Appel Date Created: June 26, 2004 Date Modified: The Flexus 2320 Equipment Standard Operating Procedure Kimberly Appel 1. Purpose 1.1. The Flexus Thin Film Stress Measuring Apparatus (TFSMA) measures the changes

More information

Dual-Angle Imaging Camera for Determination of Dielectric Thin Film Thickness Product Requirements Document Professor. Jennifer Kruschwitz

Dual-Angle Imaging Camera for Determination of Dielectric Thin Film Thickness Product Requirements Document Professor. Jennifer Kruschwitz Dual-Angle Imaging Camera for Determination of Dielectric Thin Film Thickness Product Requirements Document Professor. Jennifer Kruschwitz Customer has reviewed and approved the final product review document.

More information

AUTOFOCUS SENSORS & MICROSCOPY AUTOMATION IR LASER SCANNING CONFOCAL MICROSCOPE IRLC DEEP SEE. Now See Deeper than ever before

AUTOFOCUS SENSORS & MICROSCOPY AUTOMATION IR LASER SCANNING CONFOCAL MICROSCOPE IRLC DEEP SEE. Now See Deeper than ever before AUTOFOCUS SENSORS & MICROSCOPY AUTOMATION IR LASER SCANNING CONFOCAL MICROSCOPE IRLC DEEP SEE Now See Deeper than ever before Review and inspection of non visible subsurface defects Non visible and subsurface

More information

Welcome to the Opus Bank Mobile Banking App User Guide

Welcome to the Opus Bank Mobile Banking App User Guide Welcome to the Opus Bank Mobile Banking App User Guide 1.855.678.7226 servicecenter@opusbank.com www.opusbank.com Table of Contents Enroll in Mobile Banking & Install App... 3 Login to Personal Online

More information

Zeiss AxioImager.Z2 Fluorescence Protocol

Zeiss AxioImager.Z2 Fluorescence Protocol Zeiss AxioImager.Z2 Fluorescence Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

Flexus 2320 SOP Revision 1 01/25/16 Page 1 of 7. Flexus 2320 SOP

Flexus 2320 SOP Revision 1 01/25/16 Page 1 of 7. Flexus 2320 SOP Page 1 of 7 Flexus 2320 SOP The Flexus 2320 determines wafer curvature by measuring the angle of deflection of a laser beam off the surface 6, 4, 3 and 2 diameter substrates. Film stress is determined

More information

Spinning Disk Protocol

Spinning Disk Protocol Spinning Disk Protocol 1) System Startup F Please note our sign-up policy You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time

More information

TECHNICS MACRO-RIE 8800 STANDARD OPERATING PROCEDURES

TECHNICS MACRO-RIE 8800 STANDARD OPERATING PROCEDURES TECHNICS MACRO-RIE 8800 STANDARD OPERATING PROCEDURES Version: 1.0 Sept 2017 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1.0 DESCRIPTION.... 3 2.0 SAFETY....... 3

More information

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis Start Up: Upon entering the laser room turn on the wall mounted Laser Power Button by pulling it away from the wall. Turn on Shutter controllers (toggle switch on back of unit). There should be a U in

More information

5.2.2 Pins on the alignment detector are fragile. Use caution when inserting and removing the detector. Report any broken pins to staff.

5.2.2 Pins on the alignment detector are fragile. Use caution when inserting and removing the detector. Report any broken pins to staff. Title: Woollam Vase Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Woollam VASE. All users are expected to have read and understood

More information

Sample study by 3D optical profiler Contour Elite K for KTH university.

Sample study by 3D optical profiler Contour Elite K for KTH university. Sample study by 3D optical profiler Contour Elite K for KTH university Samuel.lesko@bruker.com Objectives Objectives Main goals for the visit consist of evaluating 3D optical profiler: Confirm capability

More information

NXQ8000 Series Mask Aligner

NXQ8000 Series Mask Aligner NXQ8000 Series Mask Aligner The NXQ8000 Production Mask Aligner and Front to Back Overlay Inspection System integrates the latest in Robotic Automation with state of the art next generation alignment stage

More information

GDN Gladson Delivery Network

GDN Gladson Delivery Network GDN Gladson Delivery Network GENERAL DESIGN The Web site is set up like a shopping cart. First, you will add items to your cart. Second, you will submit your cart for processing. And lastly, once your

More information

ksa MOS Ultra-Scan Performance Test Data

ksa MOS Ultra-Scan Performance Test Data ksa MOS Ultra-Scan Performance Test Data Introduction: ksa MOS Ultra Scan 200mm Patterned Silicon Wafers The ksa MOS Ultra Scan is a flexible, highresolution scanning curvature and tilt-measurement system.

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual J.A. Woolam M-2000V Spectroscopic Ellipsometer Version 1.0 Page 1 of 13 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Features 2.3 Sample

More information

.txt - Exporting and Importing. Table of Contents

.txt - Exporting and Importing. Table of Contents .txt - Exporting and Importing Table of Contents Export... 2 Using Add Skip... 3 Delimiter... 3 Other Options... 4 Saving Templates of Options Chosen... 4 Editing Information in the lower Grid... 5 Import...

More information

Production Assistance for Cellular Therapies (PACT) PACT Application System User s Guide

Production Assistance for Cellular Therapies (PACT) PACT Application System User s Guide Production Assistance for Cellular Therapies (PACT) PACT Application System User s Guide Version 1.0 February 9, 2017 Version 1.0 TABLE OF CONTENTS 1.0 Getting Started... 1 1.1 Access to the Internet...

More information

Biology Shared Instrumentation Facility Olympus IX70 Inverted Microscope and Simple PCI Imaging System

Biology Shared Instrumentation Facility Olympus IX70 Inverted Microscope and Simple PCI Imaging System Biology Shared Instrumentation Facility Olympus IX70 Inverted Microscope and Simple PCI Imaging System Use Protocols: A. Reservation and Starting Up B. Sample Observation (No Image Capture) Transmitted

More information

Fig. 3. Fig. 4. Fig. 5

Fig. 3. Fig. 4. Fig. 5 From lunchboxorders.com click on the green ORDER LUNCH button at the top right-hand corner of the page as shown below (Fig. 1) to proceed to the ordering site, lunchboxorders.net Fig. 1 Once you ve reached

More information

LECO LM247AT Microhardness Tester & Amh43 Software

LECO LM247AT Microhardness Tester & Amh43 Software Instrument Operation Instructions Department of Materials Science & Engineering Equipment: Location: Access: Contact: LECO LM247AT Microhardness Tester and Control Software 3355 Hoover Hall Open for undergraduate

More information

SUSS MJB4. Manual Aligner For Research, Development and Operator Assisted Production October, 2009

SUSS MJB4. Manual Aligner For Research, Development and Operator Assisted Production October, 2009 SUSS MJB4 Manual Aligner For Research, Development and Operator Assisted Production October, 2009 Overview Product Portfolio Aligner MA/BA 8 MA200Compact LithoFab200 MJB4 MA300Plus MA/BA 6 MA150e LithoPack300

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: University of Minnesota Nano Fabrication Center Coral Name: hs-scope Revision Number: 1.5 Model: HS200A Revisionist: M. Fisher Location: Bay 1 Date: 9/12/2013 1 Description The Hyphenated

More information

Using Report Builder in Total Grant Solution (TGS)

Using Report Builder in Total Grant Solution (TGS) Using Report Builder in Total Grant Solution (TGS) How to set up a new report (Note: The steps on pages 1 & 2 only need to be done the first time you create a new report; skip to the second step on page

More information

Wide Guy: Inverted Widefield Microscope

Wide Guy: Inverted Widefield Microscope Wide Guy: Inverted Widefield Microscope Kyle Marchuk Adam Fries Jordan Briscoe August 2017 Contents 1 Introduction 2 2 Initial Setup 3 2.1 Hardware Startup........................................... 3

More information

RHD Online Progress Reporting System in Nustar. RHD Reviewer User Guide

RHD Online Progress Reporting System in Nustar. RHD Reviewer User Guide RHD Online Progress Reporting System in Nustar RHD Reviewer User Guide TABLE OF CONTENTS TABLE OF CONTENTS... 2 QUICK GUIDE... 3 DETAILED GUIDE... 4 GETTING STARTED... 4 When to complete a review... 4

More information

B4 Nitride Furnace Instructions

B4 Nitride Furnace Instructions B4 Nitride Furnace Instructions The Nitride furnace is a LPCVD furnace capable of depositing standard stoichiometric (Si3N4) silicon nitride, as well as silicon rich/low stress silicon nitride. The Nitride

More information

Budget Exercise for Intermediate Excel

Budget Exercise for Intermediate Excel Budget Exercise for Intermediate Excel Follow the directions below to create a 12 month budget exercise. Read through each individual direction before performing it, like you are following recipe instructions.

More information

Standard Operating Procedure for the Horiba FluroMax-4

Standard Operating Procedure for the Horiba FluroMax-4 Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.

More information

Nishtha Bhatia Washington High School July 31 st, 2014

Nishtha Bhatia Washington High School July 31 st, 2014 Nishtha Bhatia Washington High School July 31 st, 2014 MY PROJECTS Programming Photolithography o Incorporate image slider to Nanolab website homepage o MNL s current i-line PR OiR 897-10i is discontinued

More information

Leica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b

Leica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b Leica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b User Guide Biomedicum Imaging Unit (BIU) University of Helsinki www.biu.helsinki.fi 11.11.2016 1 GENERAL...1

More information

menu of icons which you can select which item you want to use to startup the computer.

menu of icons which you can select which item you want to use to startup the computer. To begin, insert the media that came with your computer into your CD / DVD drive. While the media is in the computer, power the computer off and restart the system. Upon restart you will need to press

More information

Grant Lifecycle Manager Learning Lab Applicant Tutorial

Grant Lifecycle Manager Learning Lab Applicant Tutorial VIDEO 1: Site Access & Account Creation You will access the system via the log on page. The Foundation may have provided you with a link to this page, or it may be hyperlinked on their Foundation website.

More information

2. This will bring you to a new window asking what type of policy you are inserting. Personal Policies. Commercial Policies. All Other Policies

2. This will bring you to a new window asking what type of policy you are inserting. Personal Policies. Commercial Policies. All Other Policies How to Insert a New Client File To insert a new client file you can do this one of two ways. The first way is to manually insert the client file through Hawksoft. You would use this method if you are writing

More information

EPiSERVER Content Management System

EPiSERVER Content Management System Last Updated: 11/05/2014 Refreshable/Rotator Hero Slider Blocks EPiSERVER Content Management System A Refreshable Hero Slider is created and housed in the Global Components Folders within the department

More information

5 Setting Preferences 15 Preferences 15 Configure Chart Colors 16

5 Setting Preferences 15 Preferences 15 Configure Chart Colors 16 CRITERION Vantage 3 Acquire Training Manual Contents 1 Introduction 3 Collecting Data 3 2 Starting the Program 3 Logging In and Logging Out 3 Logging In 3 Logging in as an Administrator 3 Logging in as

More information

Price Checker Setup Guide for AIMsi

Price Checker Setup Guide for AIMsi Price Checker Setup Guide for AIMsi Table of Contents Configuring the Price Checker Shuttle... 2 Configuring the Wireless SG15+ Price Checker Shuttle... 5 PriceChecker Software Setup... 9 AIMsi Setup...

More information

FEI Helios NanoLab 600 TEM specimen prep recipe Nicholas G. Rudawski (352) (office) (805) (cell) Last updated: 07/16/18

FEI Helios NanoLab 600 TEM specimen prep recipe Nicholas G. Rudawski (352) (office) (805) (cell) Last updated: 07/16/18 FEI Helios NanoLab 600 TEM specimen prep recipe Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 07/16/18 This recipe is essentially a composite of several established

More information

Application Note #554 VXI Universal Surface Measurements for 3D Optical Microscopes

Application Note #554 VXI Universal Surface Measurements for 3D Optical Microscopes Surface detail of smooth AMOLED substrate Detail of LED wafer and interactive cursors Application Note #554 VXI Universal Surface Measurements for 3D Optical Microscopes MEMS inertial sensor Bruker has

More information

Fluorimeter User s Booklet

Fluorimeter User s Booklet Fluorimeter User s Booklet 1) Verify that the instrument power and the computer are both off. Then turn on the lamp. The red indicator light will glow when the lamp is on. Turn on lamp switch here. This

More information

Start-up system. Nikon Wide Field Microscope Ee1078. Complete manual

Start-up system. Nikon Wide Field Microscope Ee1078. Complete manual Start-up system Switch on the power socket located on the floor Switch devices on in the order of: o Mercury lamp control-unit switch on and ignite o Halogen lamp o Incubator control-box o Microscope stand

More information

Quick Start Guide - Contents. Opening Word Locating Big Lottery Fund Templates The Word 2013 Screen... 3

Quick Start Guide - Contents. Opening Word Locating Big Lottery Fund Templates The Word 2013 Screen... 3 Quick Start Guide - Contents Opening Word... 1 Locating Big Lottery Fund Templates... 2 The Word 2013 Screen... 3 Things You Might Be Looking For... 4 What s New On The Ribbon... 5 The Quick Access Toolbar...

More information

DEKTAK XT MECHANICAL PROFILOMETER

DEKTAK XT MECHANICAL PROFILOMETER EPFL STI CMi DEKTAK XT MECHANICAL PROFILOMETER The DektakXT stylus surface profiler is an advanced thin and thick film step height measurement tool with the following characteristics: - equipment compatible

More information

Olympus IX-70 Imaging Protocol

Olympus IX-70 Imaging Protocol Olympus IX-70 Imaging Protocol 1) System Startup Please note our sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused

More information

Operating Instructions

Operating Instructions Operating Instructions Measurement of Liquid Viscosity The viscosity is measured using the Brookfield Viscometer (see Figure 1a) (a) (b) Figure 1. (a) Top view of the Brookfield Viscometer. The viscosity

More information

Mobile App User Guide

Mobile App User Guide Download the Mobile App iphone and ipad To find our Freedom Credit Union Mobile App just scan the appropriate QR code to the right with your Apple device: iphone Download Or you can find it through the

More information

Hitachi U-2001 UV/Vis Spectrometer Updated November 14, 2017

Hitachi U-2001 UV/Vis Spectrometer Updated November 14, 2017 Hitachi U-2001 UV/Vis Spectrometer Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml If you have any problems with the

More information

Panalytical MRD X-Ray Diffraction SOP

Panalytical MRD X-Ray Diffraction SOP Panalytical MRD X-Ray Diffraction SOP Table of Contents 1.0 Safety 2.0 Training 3.0 Sample Preparation 4.0 Pre-Operation 5.0 Sample Height Adjustment 6.0 Running Programs 7.0 Sample Unloading 8.0 Post-Operation

More information

PC Agent Quick Start. Opening the Agent. Logging in to the Agent. Connected Backup. Version 8.6. Revision 0

PC Agent Quick Start. Opening the Agent. Logging in to the Agent. Connected Backup. Version 8.6. Revision 0 T E C H N I C A L N O T E Version 8.6 PC Agent Quick Start Revision 0 Use this document as a quick reference for common Connected Backup PC Agent tasks. If the Agent is not on your computer, contact your

More information

SIMDELL USER MANUAL Rent Arrears

SIMDELL USER MANUAL Rent Arrears FROM SIMDELL USER MANUAL Rent Arrears Software for The Providers of Social Housing Page 1 of 111 Release - August 2007 Revision History Date of next revision: June 2008 Revision date Previous revision

More information

Connection Guide (for D SLR Cameras) En

Connection Guide (for D SLR Cameras) En SB7J02(11)/ 6MB41011-02 Connection Guide (for D SLR Cameras) En Table of Contents Introduction...2 The Interface...2 Trouble Connecting?...2 More on SnapBridge...2 What You ll Need...3 What SnapBridge

More information

INSIGHT Installation Instructions

INSIGHT Installation Instructions This is the New Icon that will appear on your computer screen after the INSIGHT v8.1 software is installed. IMPORTANT: These instructions pertain to version 8.1.29. INSIGHT v8.1.29 requires Windows 8,

More information

Determination of Drag Coefficient

Determination of Drag Coefficient DEPARTMENT OF MECHANICAL, INDUSTRIAL AND MANUFACTURING ENGINEERING MIMU 505 - MEASUREMENT AND ANALYSIS Determination of Drag Coefficient You will need to bring a zip disk or USB storage device to the lab

More information

Olympus DeltaVision Microscope Start-Up and Shut-Down Instructions.

Olympus DeltaVision Microscope Start-Up and Shut-Down Instructions. DeltaVision Instructions: 1. Start-up (Olympus DV) 2. Basic operation (Olympus DV) 2.1 set-up 2.2 designing and running an experiment 3. Transferring files via FTP 4. Deconvolving files 5. File conversion

More information

AERIES INSTRUCTIONS for EXCESSIVE and SARB LETTERS

AERIES INSTRUCTIONS for EXCESSIVE and SARB LETTERS AERIES INSTRUCTIONS for EXCESSIVE and SARB LETTERS Updating Attendance/Enrollment Totals, Printing the Letter, Printing the Letter Log, Editing Letter Log Best Practices and Policies Contents Updating

More information

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Karl Suss MA150 Mask Aligner. All users are expected to have read and understood

More information

Windows 95/98 Infoway Connection Instructions 1/4/2002; rev. 1/9/2002

Windows 95/98 Infoway Connection Instructions 1/4/2002; rev. 1/9/2002 Windows 95/98 Infoway Connection Instructions 1/4/2002; rev. 1/9/2002 The following are the basic steps for setting up your computer for a PPP connection to the library s Infoway Internet service. A PPP

More information

PD Prism Finder. English Instructions

PD Prism Finder. English Instructions PD Prism Finder English Instructions Special Features 1. The finder produces an unreversed, laterally correct and bright image, has a built-in Silicon Diode metering system and an eyepiece that is raised

More information

NGForms. Acceptance Test Plan. v1.1. Drexel CS Senior Design Dan DeCamillo Hung Dinh Julian Lico Raymond Ross, Jr.

NGForms. Acceptance Test Plan. v1.1. Drexel CS Senior Design Dan DeCamillo Hung Dinh Julian Lico Raymond Ross, Jr. NGForms Acceptance Test Plan v1.1 Drexel CS Senior Design 2009-2010 Dan DeCamillo Hung Dinh Julian Lico Raymond Ross, Jr. Trung Tran 1 Contents 1 Introduction... 3 1.1 User Acceptance Definition... 3 1.2

More information

PONEMAH V5.30 WITH NOLDUS MEDIA RECORDER 4.0 QUICK START GUIDE

PONEMAH V5.30 WITH NOLDUS MEDIA RECORDER 4.0 QUICK START GUIDE TECHNICAL NOTE PONEMAH V5.30 WITH NOLDUS MEDIA RECORDER 4.0 QUICK START GUIDE DSI has partnered with Noldus Information Technology, the leading solution provider for human and animal behavioral research

More information

TMB LASERFICHE WEBLINK GUIDE... 1 INTRODUCTION... 2 CONNECTING TO THE WEBSITE... 2 WEBLINK LOGIN... 3 BROWSING... 3 LASERFICHE FILE FORMATS...

TMB LASERFICHE WEBLINK GUIDE... 1 INTRODUCTION... 2 CONNECTING TO THE WEBSITE... 2 WEBLINK LOGIN... 3 BROWSING... 3 LASERFICHE FILE FORMATS... TEXAS MEDICAL BOARD INTRODUCTION TO WEBLINK Page 1 of 20 TMB LASERFICHE WEBLINK GUIDE Table of Contents TMB LASERFICHE WEBLINK GUIDE... 1 INTRODUCTION... 2 CONNECTING TO THE WEBSITE... 2 WEBLINK LOGIN...

More information

Computer. Test Fixture SPA. Key board

Computer. Test Fixture SPA. Key board Semiconductor Parameter Analyzer (Operating instructions) A PDF copy of this document should be available on the Windows Desktop of the SPA computer Read this document fully before attempting to operate

More information

BIZPRAC 12 GUIDE RE-INSTALLATION

BIZPRAC 12 GUIDE RE-INSTALLATION BIZPRAC 12 GUIDE RE-INSTALLATION BEFORE YOU START 1. Please ensure that your computer/s meets the Minimum System Requirements: http://www.bizprac.com/wp-content/uploads/bizprac- Minimum-System-Requirements.pdf.

More information

External Funds Transfer Part I Adding an External Account

External Funds Transfer Part I Adding an External Account External Funds Transfer Part I Adding an External Account 1. Log in to your Online Banking profile. In the column on the left, scroll down to Services and select Other Services. 2. From the list, select

More information

REGION & ZONE INSTRUCTION GUIDE. This document contains instructions to establish and administer your Region/Zone/Club structure.

REGION & ZONE INSTRUCTION GUIDE. This document contains instructions to establish and administer your Region/Zone/Club structure. REGION & ZONE INSTRUCTION GUIDE 1 This document contains instructions to establish and administer your Region/Zone/Club structure. To establish the region/zone/club structure, follow these steps. 1st 2nd

More information

MOTOR Data Management System Users Guide

MOTOR Data Management System Users Guide MOTOR Data Management System Users Guide Version 1.0 November 24, 2003 Prepared by the Collaborative Studies Coordinating Center TABLE OF CONTENTS PAGE 1 MOTOR DMS... 4 2 OVERVIEW OF DATA COLLECTION...

More information

If you sync with your home machine, you may update there. The best solution is for you to update to ios5 with the computer you sync the device with.

If you sync with your home machine, you may update there. The best solution is for you to update to ios5 with the computer you sync the device with. ios5.1.0 UPDATE INSTRUCTIONS BELTON ISD These instructions are made for Belton ISD teachers and staff to update ipads and computers to ios5. If you sync with your home machine, you may update there. The

More information

TRAINING GUIDE. Advanced Crystal 1

TRAINING GUIDE. Advanced Crystal 1 TRAINING GUIDE Advanced Crystal 1 Using Crystal Reports with Lucity Advanced Examples 1 The fifth of a seven-part series, this workbook is designed for Crystal Reports users with some experience, who wish

More information

How to Create and Submit an Amendment in INSPIR II

How to Create and Submit an Amendment in INSPIR II How to Create and Submit an Amendment in INSPIR II Login into INSPIR and from your Home page, click on My Studies to list all the studies that you have access to Locate the study and click on the Open

More information

TRAINING GUIDE. Advanced Crystal 1

TRAINING GUIDE. Advanced Crystal 1 TRAINING GUIDE Advanced Crystal 1 Using Crystal Reports with Lucity Advanced Examples 1 The fifth of a seven-part series, this workbook is designed for Crystal Reports users with some experience, who wish

More information

Carestream Vita user quick guide. Software version 3.2 From April 2012

Carestream Vita user quick guide. Software version 3.2 From April 2012 Carestream Vita user quick guide Software version 3.2 From April 2012 1 Carestream Vita user quick guide Software version 3.2 from April 2012 1. To switch your Vita on Press the power button on the PC

More information

Operation Manual of J.A. Woollam Ellipsometer

Operation Manual of J.A. Woollam Ellipsometer Operation Manual of J.A. Woollam Ellipsometer 1) Press both the Lamp power and lamp Ignition buttons to turn them on (The buttons are located on the front panel of the bottom control box). 2) Put your

More information

Basics. Mouse The mouse normally looks like a little arrow, but it can change depending on what you are doing

Basics. Mouse The mouse normally looks like a little arrow, but it can change depending on what you are doing Robert DiChiara 1 Basics Mouse The mouse normally looks like a little arrow, but it can change depending on what you are doing Red X Button It is found on the top right of every window, and if you press

More information

TopPoint User's Guide

TopPoint User's Guide TopPoint User's Guide LT2152/ 2000 Gilson, Inc. All rights reserved August 2000 Table of Contents 1 Introduction 2 Operation 3 Troubleshooting Steps to Using TopPoint... 1-2 Accessing Online Help... 1-3...

More information

Secure Transfer Site (STS) User Manual

Secure Transfer Site (STS) User Manual Secure Transfer Site (STS) User Manual (Revised 3/1/12) Table of Contents Basic System Display Information... 3 Command Buttons with Text... 3 Data Entry Boxes Required / Enabled... 3 Connecting to the

More information

HP Color LaserJet CM3530 MFP Series Fax

HP Color LaserJet CM3530 MFP Series Fax Fax my scan Send a fax 1 Place the original face-up in the document feeder or face-down on the glass. 2 Touch Fax. If prompted, type your user name and 3 Touch Fax Recipients. Enter a fax number. Touch

More information

Olympus IX-70 Imaging Protocol

Olympus IX-70 Imaging Protocol Olympus IX-70 Imaging Protocol 1) System Startup F Please note our sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for

More information

FTM 131 CANON EOS C100 WALKTHROUGH, PART I: 10 Steps to Record a Moving Image

FTM 131 CANON EOS C100 WALKTHROUGH, PART I: 10 Steps to Record a Moving Image FTM 131 CANON EOS C100 WALKTHROUGH, PART I: 10 Steps to Record a Moving Image Step 1: TRIPOD - Perform the following steps on your tripod, before you place the camera on it: - Extend the legs to set the

More information

Multi-User Importing Instructions from an Existing ACT! Database

Multi-User Importing Instructions from an Existing ACT! Database Multi-User Importing Instructions from an Existing ACT! Database Thank you for purchasing an Act4Advisors Network License. These instructions are designed to help you import data from an existing ACT!

More information

Plasma Quest RPCVD: Operating Procedure (6/24/02)

Plasma Quest RPCVD: Operating Procedure (6/24/02) Plasma Quest RPCVD: Operating Procedure (6/24/02) Considerations: *There are four basic steps that MUST exist in a recipe: Gas Stabilization, Plasma Tuning, Deposition/Process and End Step. Additional

More information

Calendar Basics Outlook 2016 for Windows

Calendar Basics Outlook 2016 for Windows Calendar Basics Outlook 2016 for Windows Texas Christian University Information Technology Table of Contents Navigate the Calendar... 1 Create and Edit New Calendar Items... 2 New Appointment... 2 Add

More information

xr Fiber Platform User Guide

xr Fiber Platform User Guide FOR USE WITH THE XR FIBER APPLICATION xr Fiber Platform User Guide The most current version of this User Guide can be found at: www.xrapid-environment.com/resources/ Important: This Guide describes how

More information

Spectroscopic Ellipsometer --- J. A. Woollam alpha-se

Spectroscopic Ellipsometer --- J. A. Woollam alpha-se Spectroscopic Ellipsometer --- J. A. Woollam alpha-se Introduction Figure 1: J. A. Woollam alpha-se spectroscopic ellipsometer An ellipsometer measures the change in polarization as light reflects or transmits

More information

web po user guide Supplier

web po user guide Supplier web po user guide Supplier web po user guide table of contents supplier section 1 before you begin section 2 getting started and the basics section 3 Web PO Supplier Administration section 4 Viewing Purchase

More information

PANASONIC TELEPHONE SYSTEM ADMINISTRATION MANUAL PANASONIC TVM VOIC SYSTEM ADMINSTRATION MANUAL

PANASONIC TELEPHONE SYSTEM ADMINISTRATION MANUAL PANASONIC TVM VOIC SYSTEM ADMINSTRATION MANUAL PANASONIC TELEPHONE SYSTEM ADMINISTRATION MANUAL & PANASONIC TVM VOICEMAIL SYSTEM ADMINSTRATION MANUAL PANASONIC KX-NCP ENTER PROGRAMMING MODE Double click on Phone System Maintenance Icon on your desktop

More information

Integrated ACH. Business Concentration Consumer Pre-authorized Debit

Integrated ACH. Business Concentration Consumer Pre-authorized Debit Integrated ACH Our new Integrated ACH Module will allow you to complete a host of electronic funds transfer activities via an easy to learn process outlined in the next several pages. Payment Options Transactions

More information

Module 2. Teenage Pregnancy Support Service 6.8

Module 2. Teenage Pregnancy Support Service 6.8 Module 2 Teenage Pregnancy Support Service 6.8 Contents TPSS Process Flowchart Explanation of Flowchart Colour Codes... 3 TPSS Process Flowchart...4-5 Teenage Pregnancy Advisor Recording Instructions Record

More information

MCS-TOUCHSCREEN Auto Disk Clean up

MCS-TOUCHSCREEN Auto Disk Clean up APP093 - MCS-Touchscreen Error 16 Auto Diskcleanup Procedure Micro Control Systems APPLICATION NOTE APP-093 MCS-TOUCHSCREEN Auto Disk Clean up Revision History Date Author Revision Description 07/31/2014

More information