Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Similar documents
Automatic Ascending Aorta Detection in CTA Datasets

Optimization of scanner parameters for dual energy micro-ct

Cardiac Dual Energy CT: Technique

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.

Ch. 4 Physical Principles of CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION

Enhanced material contrast by dual-energy microct imaging

Spectral analysis of non-stationary CT noise

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Introduction to Biomedical Imaging

Scatter Correction Methods in Dimensional CT

DICOM. Supplement 188 Multi-Energy CT Imaging. DICOM Working Group 21 Computed Tomography

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

DUE to beam polychromacity in CT and the energy dependence

BME I5000: Biomedical Imaging

Image Acquisition Systems

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

A Radiometry Tolerant Method for Direct 3D/2D Registration of Computed Tomography Data to X-ray Images

CT vs. VolumeScope: image quality and dose comparison

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel

Fast iterative beam hardening correction based on frequency splitting in computed tomography

A New Approach for the Enhancement. of Dual-energy Computed Tomography Images. Kyung Kook Park

Dual-Energy CT Based Monochromatic Imaging

The Near Future in Cardiac CT Image Reconstruction

Empirical dual energy calibration EDEC for cone-beam computed tomography

Automatic Segmentation of the Aortic Dissection Membrane from 3D CTA Images

ML reconstruction for CT

Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications 1

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Advanced Multi Material Decomposition of Dual Energy in Computed Tomography Image

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology.

Photon counting spectroscopic CT with dynamic beam attenuator

Motion Compensation from Short-Scan Data in Cardiac CT

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application

The VesselGlyph: Focus & Context Visualization in CT-Angiography

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Acknowledgments and financial disclosure

Quantitative image-based spectral reconstruction for computed tomography

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology

Some reference material

Proton dose calculation algorithms and configuration data

Computer-Tomography I: Principles, History, Technology

MEDICAL IMAGING 2nd Part Computed Tomography

Fundamentals of CT imaging

Dual-Energy CT: General Principles

CT issues in PET / CT scanning. ImPACT technology update no. 4

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

DUAL-ENERGY CT IN PROTON THERAPY

Computed Tomography. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Sep 26th/Oct 3rd, 2016

Micro-CT Methodology Hasan Alsaid, PhD

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

Quantitative Attenuation Correction for PET/CT Using Iterative Reconstruction of Low-Dose Dual-Energy CT

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

Midterm Review. Yao Wang Polytechnic University, Brooklyn, NY 11201

arxiv: v1 [physics.med-ph] 18 Apr 2016

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

Digital phantoms for the evaluation of a software used for an automatic analysis of the Winston-Lutz test in image guided radiation therapy

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

arxiv: v2 [cond-mat.mtrl-sci] 5 Jan 2010

Deep Scatter Estimation (DSE): Accurate Real-Time Scatter Estimation for X-Ray CT using a Deep Convolutional Neural Network

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners

Respiratory Motion Compensation for C-arm CT Liver Imaging

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

RADIOMICS: potential role in the clinics and challenges

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study

Application of MCNP Code in Shielding Design for Radioactive Sources

Michal E. Kulon, MD 1,2 1. Peter Komlosi, MD, PhD 3 3. Radiology Universe Institute, 2

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.

Digital Image Processing

Tomographic Reconstruction

CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona

Computer-Tomography II: Image reconstruction and applications

RADIOLOGY AND DIAGNOSTIC IMAGING

INTERNATIONAL STANDARD

3D Guide Wire Navigation from Single Plane Fluoroscopic Images in Abdominal Catheterizations

Spectral CT reconstruction with an explicit photon-counting detector model: a one-step approach

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

Power Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis

A closer look at CT scanning

UNIVERSITY OF SOUTHAMPTON

High dynamic range magnetic resonance flow imaging in the abdomen

Separate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising

Physical bases of X-ray diagnostics

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Transcription:

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision Laboratory, ETH Zurich, Switzerland 2 Institute of Diagnostic Radiology, University Hospital Zurich, Switzerland 3 Medical Image Analysis Center, University of Basel, Switzerland saur@vision.ee.ethz.ch Abstract. A drawback of the commonly used single source computed tomography systems (CT) is that different materials might show very similar attenuation at any selected radiation energy. However, the assessment of atherosclerosis requires good differentiation between vessel lumen, calcium, adipose, and surrounding tissue. Dual energy CT (DECT) simultaneously measures attenuations at two energies and therefore can improve the differentiation to some extent. A tissue cancelation and enhancement algorithm for dual energy data was already proposed in 1981 and evaluated on experimental settings with a stationary X-ray source. For this study, we adapted this algorithm for DECT and propose its usage as a pre-processing step for the assessment of atherosclerosis. On clinical DECT patient data and with fixed parameters we could show a simultaneous contrast enhancement between 8% and 67% among all targeted tissues. 1 Introduction Each material has a characteristic X-ray attenuation depending on its atomic number, density and the incident energy of the X-ray beam. Single source computed tomography systems (CT) characterize tissues with only one value such that different materials might show very similar attenuation. For the assessment of atherosclerosis, however, a precise differentiation between different tissues is needed: Calcified hard plaques are detected by differentiating calcium from the vessel lumen (blood/iodine), whereas for non-calcified soft plaques, vessel lumen, adipose and surrounding tissue must be distinguished. A dual energy CT system (DECT) has two X-ray sources and therefore allows to simultaneously measure the X-ray attenuation at low and high energy levels such that two attenuation values are present for each volume element (voxel). This additional dimension can be used for better material discrimination. Dual energy X-ray acquisitions have been discussed and evaluated on basic phantoms since the early 1980s [1, 2, 3] until the first commercial DECT system was introduced 2005 [4]. On this DECT system, Eusemann et al. [5] evaluated different

62 Saur et al. blending techniques to combine the low and high energy images into one improved single image whereas Johnson et al. [6] used unpublished algorithms to estimate the iodine content of each voxel from the low and high energy images. In 1981, Lehmann et al. [1] proposed a tissue cancelation algorithm by decomposing the low and high energy images into two basis material images and combining them afterwards into a single projection image. Depending on a single parameter, the algorithm cancels out the contrast between two materials to enhance a third one, or it enhances the contrast between materials - both demonstrated on experimental settings with a stationary X-ray source. We adapted this algorithm for clinical DECT and propose its usage as a pre-processing step for the assessment of atherosclerosis. We applied the method on clinical dualenergy (DE) data and measured the contrast enhancement and signal to noise ratio between several tissues of interest. With one fixed parameter setting we could simultaneously enhance the contrast between all targeted tissues. 2 Methods The principle of the tissue cancelation algorithm by Lehmann et al. [1] is summarized in Fig. 1 and described below in detail. First, the original low and high energy images are decomposed into two basis material images to map the density distribution of two arbitrarily chosen basis materials. Then, the two bases are combined in a cancelation step into a single projection image. The two attenuation coefficients µ(low) and µ(high) obtained by DECT for each voxel are decomposed into a basis pair ρ A and ρ B of two chosen basis materials A and B by ρ A = 1 α [µ(high)µ B(low) µ(low)µ B (high)] (1) ρ B = 1 α [µ(low)µ A(high) µ(high)µ A (low)] (2) CANCELATION ALGORITHM DUAL ENERGY CT SCAN HIGH ENERGY LOW ENERGY µ (high) µ (low) BASIS DECOMPOSITION ρ A ρ B BASIS A BASIS B CANCELATION STEP φ C PROJECTION / CANCELATION X-RAY SOURCE TISSUE specifications chemical composition XOP XCOM SIMULATION emission spectra absorption spectra COMPUTATION OF μ i (U) LOOK UP TABLE µ A (low) µ A (high) µ B (low) µ B (high) Fig. 1. Workflow of the tissue cancelation algorithm proposed by Lehmann et al. [1]

Contrast enhancement with dual energy CT 63 Table 1. For each possible material combination of lumen, tissue, adipose, and calcium, we determined the cancelation angle ϕ max that maximized the mean contrast enhancement factor (CEF) measured between the cancelation image and the 140 kv acquisition over all data sets. lumen tissue adipose calcium max CEF (ϕ max ) lumen - 1.58 (0.81) 2.82 (0.84) 1.30 (0.81) max CEF (ϕ max ) tissue 1.58 (0.81) - 1.16 (0.80) 1.38 (0.81) max CEF (ϕ max) adipose 2.82 (0.84) 1.16 (0.80) - 1.48 (0.81) max CEF (ϕ max ) calcium 1.30 (0.81) 1.38 (0.81) 1.48 (0.81) - with α = µ B (low)µ A (high) µ B (high)µ A (low). The attenuation coefficient µ i (U) of the basis material i at the tube voltage U is calculated by µ i (U) = 0 N U (E)µ i (E)dE 0 N U (E)dE where N U (E) is the photon energy distribution at tube voltage U and µ i (E) is the attenuation spectrum for material i. We used [7] to compute N U (E) of a Tungsten X-ray tube (5 mm Al filter) and for µ i (E) we used a photon cross sections database [8]. For this study, we considered all possible basis material pair combinations between iodine, calcium, water, polymethyl methacrylate, adipose and blood. The projection, or cancelation, image is computed by C = ρ A sin(ϕ) + ρ B cos(ϕ) with ϕ, the cancelation angle, as a parameter. For the assessment of atherosclerosis, a good contrast between vessel lumen, soft tissue, adipose, and calcium is desired. Therefore, we manually selected a region R i within a data set for each material i (i = 1..4). The mean intensity Īi and standard deviation σ i are computed within each R i for every data set. The cancelation angle ϕ is varied between π and +π with a step of 0.01. For each selected ϕ, the contrast enhancement factor (CEF) between material i and j is computed as CEF ij = Īc i Īc j Ī ci + Īc j / Ī hi Īh j Ī hi + Īh j, (3) where CEF represents the enhancement ratio of the contrast in the cancelation image (index c) and the original high energy image (index h). Hence, a CEF>1 should be achieved. Further, the signal to noise ratio SNR i = Īi/σ i is measured for each material i in the low and high energy, and cancelation image, respectively. The cancelation angle is only dependent on the materials to be enhanced [1] such that we expected different optimal cancelation angles for the various targeted tissues which, however, should be universally applicable for all patient images. We evaluated this by measuring the CEF and SNR in dependence on ϕ. Further, we determined the optimal angles maximizing the contrast between two targeted tissues.

64 Saur et al. Table 2. We measured the mean (±std) contrast enhancement factor (CEF) for the cancelation image (ϕ = 0.79) against the high energy 140 kv acquisition for each possible material combination. Further, the mean (±std) signal to noise ratio (SNR) in the cancelation, the 80 kv, and the 140 kv images was measured. lumen tissue adipose calcium SNR cancelation 61.6 (±28.2) 39.6 (±17.8) 60.9 (±23.5) 11.3 (±1.6) SNR 140kV 51.4 (±11.7) 37.8 (±10.9) 56.7 (±15.5) 14.7 (±6.3) SNR 80kV 42.9 (±22.2) 27.3 (±14.2) 40.4 (±18.3) 9.5 (±0.7) CEF lumen - 1.27 (±0.06) 1.67 (±0.34) 1.16 (±0.03) CEF tissue 1.27 (±0.06) - 1.08 (±0.08) 1.19 (±0.03) CEF adipose 1.67 (±0.34) 1.08 (±0.08) - 1.23 (±0.04) CEF calcium 1.16 (±0.03) 1.19 (±0.03) 1.23 (±0.04) - 3 Results The presented method was evaluated on five abdominal DE scans showing both lipid and calcified plaques in the descending aorta. The data sets were acquired on a Siemens Definition DECT scanner (Siemens Medical Solutions, Forchheim, Germany). The study was approved by the local ethics committee. A clinical abdominal DE protocol was used with a tube voltage of 80 kv for the low and 140 kv for the high energy image, respectively. The in-plane resolution and slice spacing ranged between 0.68-0.87 mm with a slice thickness of 2 mm. In each patient, a region for lumen and tissue (both 7 7 9 voxels), for adipose (5 5 1 voxels), and for calcium (3 3 1 voxels) was selected by an experienced reader. Among all basis material combinations, the pair calciumadipose showed the best results in terms of magnitude and robustness for the intended contrast enhancement such that this basis pair was used for this study. Figure 3(a) exemplarily shows the CEF between adipose and calcium in dependence of ϕ for all data sets. Our initial expectation regarding patient independent characteristics could be justified. The mean CEF and mean SNR over all data sets were computed and for each possible tissue combination, we determined an optimal cancelation angle ϕ max that maximized CEF (Table 1). As all ϕ max were very similar, we decided to determine one single optimal angle ϕ that simultaneously enhances the contrast among all targeted tissues. Based on one data set, we chose the optimal angle ϕ = 0.79 that maximized the CEF among all tissue combinations but also showed a good SNR for each material (Fig. 3(b)). A higher CEF would have been possible with a slightly larger ϕ but that would have resulted in a worse SNR. The optimal angle ϕ was then fixed and applied on all data sets. The contrast between all targeted tissues could be simultaneously enhanced between 8% and 67% (Table 2). Except for calcium, the SNR in the cancelation image was also higher for the materials than the SNR measured in the 80 kv and 140 kv data sets, respectively.

Contrast enhancement with dual energy CT 65 Fig. 2. (a) Contrast enhancement factor (CEF) between adipose and calcium in dependence on the cancelation angle ϕ for all data sets. The selected ϕ (dashed) simultaneously enhanced the contrast among all targeted tissues in the cancelation image compared to the 140 kv image - exemplarily shown in (b) and (c). 2 1.5 CEF 1 0.5 0 2 0 2 (a) cancelation angle [radian] (b) cancelation image (c) 140 kv image 4 Discussion and conclusion This study evaluated for the first time the performance of a tissue cancelation algorithm on real DECT patient data for contrast enhancement. We propose to use this procedure as a pre-processing step for the assessment of atherosclerosis to improve the robustness of both the manual reading and intensity-based detection algorithms. We could show by measuring the contrast enhancement between targeted tissues that there is one fixed, patient-independent optimal angle for the assessment of atherosclerosis as it enhances the contrast between all targeted tissues. Future work will concentrate on incorporating this contrast enhancement into radiological workstations and into algorithms for automatic detection and quantification of atherosclerosis. Acknowledgement. This work has been supported by the CO-ME/NCCR research network of the Swiss National Science Foundation. References 1. Lehmann LA, Alvarez RE, Macovski A, et al. Generalized image combinations in dual KVP digital radiography. Med Phys. 1981;8(5):659 667. 2. Marziani M, Taibi A, Tuffanelli A, et al. Dual-energy tissue cancellation in mammography with quasi-monochromatic x-rays. Phys Med Biol. 2002;47(2):305. 3. Taibi A, Fabbri S, Baldelli P, et al. Dual-energy imaging in full-field digital mammography: A phantom study. Phys Med Biol. 2003;48(13):1945 1956. 4. Flohr T, McCollough C, Bruder H, et al. First performance evaluation of a dualsource CT (DSCT) system. Eur Radiol. 2006;16(2):256 268. 5. Eusemann C, Holmes III DR, Schmidt B, et al. Dual energy CT: How to best blend both energies in one fused image? Procs SPIE. 2008;6918(1):691803 8. 6. Johnson T, Krauß B, Sedlmair M, et al. Material differentiation by dual energy CT: Initial experience. Eur Radiol. 2007;17(6):1510 1517. 7. del Rio MS, Dejus RJ. XOP: Recent developments. Procs SPIE. 1998;3448(1):340 345. 8. Berger MJ, Hubbel JH. XCOM: Photon Cross Sections Database. National Institute of Standards and Technology; 1998.