Uniformity scan of the 6cm tubes. Jingbo Wang Argonne National Laboratory, Lemont, IL

Size: px
Start display at page:

Download "Uniformity scan of the 6cm tubes. Jingbo Wang Argonne National Laboratory, Lemont, IL"

Transcription

1 Uniformity scan of the 6cm tubes Jingbo Wang Argonne National Laboratory, Lemont, IL

2 Update of the test system Realized the computer control for the multi-channel HV supply, the 2D translation stage and the keysight oscilloscope. Integrated all the codes into a single python script. Two testing options enabled: 1) Usual data acquisition and offline analysis. 2) Immediate measurements using the scope functions 2

3 2D scan All the strips are chained one by one. Signals are readout from two ends. 2D translation stage (Newport ESP300 motor) has a 5 cm scanning range in both x and y directions. Step size: 2 mm in both directions Scanning area: 5 cm 5 cm Left Grid spacer Right 3

4 Gain uniformity: Tube#44 In single-pe mode, the signal of the direct incident photon is contaminated by the scattered photons. No matter where the laser spot is, the tube is always able to detect the scattered photons, which makes the gain look more uniform. For example, even the laser spot is on the spacer, the tube gives a good pulse due to the scattered photon somewhere else. In this case, the spacer is indistinguishable in the 2D gain map. In multi-pe mode, the pulses produced by the main laser beam is much higher than that produced by the scattered photons, which reveals the spacers in the 2D map. Left Random scattered photon Main laser beam Right 4

5 Uniformity: Tube#44 Left figure is the absolute gain map measured in single PE mode, but the result is contaminated by the scattered photons. Right figure is a map of the number of the collected electrons, measured in multi-pe mode. It shows the overall uniformity of the tube. It is a combination of the QE and gain uniformities. Multi-PE 5

6 Uniformity: Tube#45 and Tube#47 Tube#47 was measured to have a decent QE (%14) and a low gain (2e6). Even in multi-pe mode, the spacers are barely distinguishable in the 2D map. Increasing the light level should help to reveal the spacers. 2e6 gain and 100 ps time resolution of Tube#47 were measured with the laser spot on the upper-left quadrant. The same measurements should be done for other quadrants. Multi-PE Multi-PE 6

7 Is there any way to measure the absolute gain? Option 1: multi-pe measurment Assume the number of collected electrons N e responses linearly to the number of input photons as: where G is gain. If N photon and QE are known, gain G can be calculated. N photon can be calibrated by a device (silicon photomultiplier for instance) with a known QE. QE can be measured separately in the optical table. Option 2: single-pe measurement In single PE mode, the total number of collected electrons N e consists of two parts: First, block the laser beam just in front of the tube and measure the number of collected electrons (G*N scatterred ) produced by the scattered photons. Then, unblock the laser and measure the total number of collected electrons (N e ). Then the gain G can be calculated. Option 3: single-pe measurement using a fiber 7

8 Conclusion The bule laser test facility has been updated from a manual system to a automatically computer control system. Uniformity scan is available in multi-pe mode. This can be a routine test for every new tube. It is difficult to measure the uniformity of the absolute gain due to the contamination from the scattered photons. Overall uniformity can be scanned in multi-photoelectron mode. There are two possible options for measuring the absolute gain uniformity with the current setup. A better solution is using a fiber. Next Try to measure the absolute gain uniformity. Redo the performance tests for all three working tubes. Write a documentation for the laser system. 8

Development of planar microchannel plate photomultiplier at Argonne National Laboratory

Development of planar microchannel plate photomultiplier at Argonne National Laboratory Development of planar microchannel plate photomultiplier at Argonne National Laboratory Junqi Xie on behalf of the detector R&D group Argonne National Laboratory, Argonne, IL Email: jxie@anl.gov The Technology

More information

Lecture 05. First Example: A Real Lidar

Lecture 05. First Example: A Real Lidar Lecture 05. First Example: A Real Lidar Brief review of lidar basics K Doppler lidar system architecture K lidar signal estimate from lidar equation Comparison of estimate to reality Summary Review of

More information

MCP Photodetector Development at Argonne. Lei Xia ANL - HEP

MCP Photodetector Development at Argonne. Lei Xia ANL - HEP MCP Photodetector Development at Argonne Lei Xia ANL - HEP Introduction large area photodetector development Development of ALD coated large area MCP Development of ceramic/glass packaging for large area

More information

Laser Diagnostic for High-Energy, Laser Fusion Drivers

Laser Diagnostic for High-Energy, Laser Fusion Drivers UCRL-JC-119065 PREPRINT Laser Diagnostic for High-Energy, Laser Fusion Drivers D S. C. Burkhart Wm. Behrendt Ian Smith This paper was prepared for submittal to the Conference on Lasers and Electro-Optics

More information

Optical Diffraction and Interference using Single Photon Counting

Optical Diffraction and Interference using Single Photon Counting Optical Diffraction and Interference using Single Photon Counting Optical Diffraction and Interference using Single Photon Counting In this experiment the wave and quantum properties of light can be studied

More information

Progress of the Thomson Scattering Experiment on HSX

Progress of the Thomson Scattering Experiment on HSX Progress of the Thomson Scattering Experiment on HSX K. Zhai, F.S.B. Anderson, D.T. Anderson HSX Plasma Laboratory, UW-Madison Bill Mason PSL, UW-Madison, The Thomson scattering system being constructed

More information

Spherical Crystal X-ray Imaging for MTW, OMEGA, and OMEGA EP

Spherical Crystal X-ray Imaging for MTW, OMEGA, and OMEGA EP Spherical Crystal X-ray Imaging for MTW, OMEGA, and OMEGA EP C.STOECKL, G. FISKEL, R. K. JUNGQUIST, P. M. NILSON, AND W. THEOBALD University of Rochester, Laboratory for Laser Energetics Spherical Crystal

More information

High Voltage system for the Double Chooz experiment

High Voltage system for the Double Chooz experiment High Voltage system for the Double Chooz experiment Fumitaka Sato Junpei Maeda Takayuki Sumiyoshi Kento Tsukagoshi (Tokyo Metropolitan University) for the Double Chooz collaboration 1 7m Double Chooz detector

More information

Section 2 PROGRESS IN LASER FUSION

Section 2 PROGRESS IN LASER FUSION Section 2 PROGRESS IN LASER FUSION 2.A Optical Fiducials for X-Ray Streak Cameras X-ray streak cameras are the primary instruments for studying the transient nature of laser-produced plasmas. However,

More information

Introduction. Experiment A: Snell s Law. Physics 1CL REFLECTION AND REFRACTION OF LIGHT Summer Session II 2010

Introduction. Experiment A: Snell s Law. Physics 1CL REFLECTION AND REFRACTION OF LIGHT Summer Session II 2010 Introduction This laboratory is a quantitative investigation of the reflection and refraction of light off optical interfaces. An optical interface is a boundary between two transparent media of different

More information

PMT Candidates for SSD Challenge for PMTs in air-shower detection highest linear dynamic range

PMT Candidates for SSD Challenge for PMTs in air-shower detection highest linear dynamic range PMT Candidates for SSD Challenge for PMTs in air-shower detection highest linear dynamic range Karl-Heinz Becker, Karl-Heinz Kampert, Alex Kääpä, Christian Pauly, Sven Querchfeld, Julian Rautenberg OBSERVATORY

More information

Locating the neutrino interaction vertex with the help of electronic detectors in the OPERA experiment

Locating the neutrino interaction vertex with the help of electronic detectors in the OPERA experiment Locating the neutrino interaction vertex with the help of electronic detectors in the OPERA experiment S.Dmitrievsky Joint Institute for Nuclear Research, Dubna, Russia LNGS seminar, 2015/04/08 Outline

More information

Ultrafast web inspection with hybrid dispersion laser scanner

Ultrafast web inspection with hybrid dispersion laser scanner Ultrafast web inspection with hybrid dispersion laser scanner Hongwei Chen, 1,2,3 Chao Wang, 1,4 Akio Yazaki, 1,5 Chanju Kim, 1 Keisuke Goda, 1,6,7,8, * and Bahram Jalali 1,6,7 1 Department of Electrical

More information

Synchronization and pump-probe experiments

Synchronization and pump-probe experiments Synchronization and pump-probe experiments by Stefan Düsterer Outline o Pump-probe probe infrastructure VUV + VUV VUV + optical o o Temporal overlap and jitter As example: the pump-probe probe chamber

More information

Production and Quality Assurance of Detector Modules for the LHCb Silicon Tracker

Production and Quality Assurance of Detector Modules for the LHCb Silicon Tracker Production and Quality Assurance of Detector Modules for the LHCb Silicon Tracker Olaf Steinkamp for Dmytro Volyanskyy Physik-Institut der Universität Zürich 10th ICATPP Conference on Astroparticle, Particle,

More information

Engineered Diffusers Intensity vs Irradiance

Engineered Diffusers Intensity vs Irradiance Engineered Diffusers Intensity vs Irradiance Engineered Diffusers are specified by their divergence angle and intensity profile. The divergence angle usually is given as the width of the intensity distribution

More information

CyFlow ML. Healthcare Immunology Pathology Microbiology Cell Biology. Multilaser 16 Parameter Desktop FCM System

CyFlow ML. Healthcare Immunology Pathology Microbiology Cell Biology. Multilaser 16 Parameter Desktop FCM System CyFlow ML Healthcare Immunology Pathology Microbiology Cell Biology Multilaser 16 Parameter Desktop FCM System 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements

More information

Expected Performances of the scintillator counters Time Of Flight system of the AMS-02 experiment

Expected Performances of the scintillator counters Time Of Flight system of the AMS-02 experiment Expected Performances of the scintillator counters Time Of Flight system of the AMS-02 experiment Cristina Sbarra for the AMS-TOF Bologna group (sbarra@bo.infn.it) INFN-Bologna (Italy) Cristina Sbarra

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

New approach to jitter reduction of an x-ray streak camera in accumulation mode

New approach to jitter reduction of an x-ray streak camera in accumulation mode New approach to jitter reduction of an x-ray streak camera in accumulation mode J. Liu *a, A. G. MacPhee b, C. Liu a, B. Shan c, Z. Chang c and J. Wang a a Advanced Photon Source, Argonne National Laboratory,

More information

Novel Magnetic Field Mapping Technology for Small and Closed Aperture Undulators

Novel Magnetic Field Mapping Technology for Small and Closed Aperture Undulators Novel Magnetic Field Mapping Technology for Small and Closed Aperture Undulators Erik Wallen and Hyun-Wook Kim 06.06.2017 Outline Introduction - Measurement systems at LBNL - Activities at LBNL - Need

More information

The Anfatec Level AFM a short description. Atomic Force Microscopy - approved devices for affordable prices

The Anfatec Level AFM a short description. Atomic Force Microscopy - approved devices for affordable prices The Anfatec Level AFM a short description Atomic Force Microscopy - approved devices for affordable prices Our system is complete for almost all typical applications. It provides all basic modes as: contact

More information

Confocal Raman Imaging with WITec Sensitivity - Resolution - Speed. Always - Provable - Routinely

Confocal Raman Imaging with WITec Sensitivity - Resolution - Speed. Always - Provable - Routinely Confocal Raman Imaging with WITec Sensitivity - Resolution - Speed Always - Provable - Routinely WITec GmbH, Ulm, Germany, info@witec.de, www.witec.de A modular microscope series An Example: FLIM optical

More information

User Operation Manual

User Operation Manual User Operation Manual Manual Version: 1.3 For Single Cell Bioanalyzer (SCB-402) Doc# 102-SCB-402-1.3 Table of Contents 1. General Information... 1 1.1 How to Use This Manual... 1 1.2 Operation Safety...

More information

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar Doctoral Thesis Time-of-Flight Time-of-Flight Positron Positron Emission Emission Tomography Tomography Using Using Cherenkov Cherenkov Radiation Radiation Rok Dolenec Supervisor: Prof. Dr. Samo Korpar

More information

Status of the TORCH time-of-flight detector

Status of the TORCH time-of-flight detector Status of the TORCH time-of-flight detector Neville Harnew University of Oxford (On behalf of the TORCH collaboration : the Universities of Bath, Bristol and Oxford, CERN, and Photek) August 7-9, 2017

More information

GEANT4 is used for simulating: RICH testbeam data, HCAL testbeam data. GAUSS Project: LHCb Simulation using GEANT4 with GAUDI.

GEANT4 is used for simulating: RICH testbeam data, HCAL testbeam data. GAUSS Project: LHCb Simulation using GEANT4 with GAUDI. Status of GEANT4 in LHCb S. Easo, RAL, 30-9-2002 The LHCbexperiment. GEANT4 is used for simulating: RICH testbeam data, HCAL testbeam data. GAUSS Project: LHCb Simulation using GEANT4 with GAUDI. Summary.

More information

8/3/2016. Outline. The EPID Strikes Back: Future EPID Technology and Applications. Active Matrix Flat-Panel Imagers (AMFPIs)

8/3/2016. Outline. The EPID Strikes Back: Future EPID Technology and Applications. Active Matrix Flat-Panel Imagers (AMFPIs) 8//6 The EPID Strikes Back: Future EPID Technology and Applications Larry E. Antonuk Department of Radiation Oncology University of Michigan, Ann Arbor Acknowledgements: Youcef El-Mohri, Qihua Zhao (U.

More information

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

Meterless Laser Power/Energy Measurement Simplifies Embedding

Meterless Laser Power/Energy Measurement Simplifies Embedding White Paper Meterless Laser Power/Energy Measurement Simplifies Embedding Traditional laser power and energy measurement instruments typically comprise a sensor head connected to separate meter electronics.

More information

Pixel Crosstalk and Correlation with Modulation Transfer Function of CMOS Image Sensor

Pixel Crosstalk and Correlation with Modulation Transfer Function of CMOS Image Sensor Pixel Crosstalk and Correlation with Modulation Transfer Function of CMOS Image Sensor M.Estribeau*, P.Magnan** SUPAERO Integrated Image Sensors Laboratory avenue Edouard Belin, 4 Toulouse, France ABSTRACT

More information

arxiv: v1 [physics.ins-det] 13 Jan 2015

arxiv: v1 [physics.ins-det] 13 Jan 2015 The Assembly of the Belle II TOP Counter Boqun Wang, On behalf of the Belle II PID Group Department of Physics, University of Cincinnati, Cincinnati, OH, USA University of Cincinnati preprint UCHEP-14-01

More information

TORCH: A large-area detector for precision time-of-flight measurements at LHCb

TORCH: A large-area detector for precision time-of-flight measurements at LHCb TORCH: A large-area detector for precision time-of-flight measurements at LHCb Neville Harnew University of Oxford ON BEHALF OF THE LHCb RICH/TORCH COLLABORATION Outline The LHCb upgrade TORCH concept

More information

Ray Optics. Lecture 23. Chapter 23. Physics II. Course website:

Ray Optics. Lecture 23. Chapter 23. Physics II. Course website: Lecture 23 Chapter 23 Physics II Ray Optics Course website: http://faculty.uml.edu/andriy_danylov/teaching/physicsii Let s finish talking about a diffraction grating Diffraction Grating Let s improve (more

More information

: Imaging Systems Laboratory II. Laboratory 2: Snell s Law, Dispersion and the Prism March 19 & 21, n 1 n 2

: Imaging Systems Laboratory II. Laboratory 2: Snell s Law, Dispersion and the Prism March 19 & 21, n 1 n 2 05-3: Imaging Systems Laboratory II Laboratory : Snell s Law, Dispersion and the Prism March 9 &, 00 Abstract. This laboratory exercise will demonstrate two basic properties of the way light interacts

More information

A Further Study on Large Size LSO and LYSO Crystal Samples

A Further Study on Large Size LSO and LYSO Crystal Samples October 25, 2005 IEEE NSS, N12-6, Puerto-Rico 1 A Further Study on Large Size LSO and LYSO Crystal Samples Jianming Chen, Liyuan Zhang, Ren-yuan Zhu California Institute of Technology BGO, LSO & LYSO Samples

More information

A NEW TIMING CALIBRATION METHOD FOR SWITCHED CAPACITOR ARRAY CHIPS TO ACHIEVE SUB-PICOSECOND RESOLUTIONS

A NEW TIMING CALIBRATION METHOD FOR SWITCHED CAPACITOR ARRAY CHIPS TO ACHIEVE SUB-PICOSECOND RESOLUTIONS Stefan Ritt, Paul Scherrer Institute, Switzerland A NEW TIMING CALIBRATION METHOD FOR SWITCHED CAPACITOR ARRAY CHIPS TO ACHIEVE SUB-PICOSECOND RESOLUTIONS 13 March 2014 Workshop on Picosecond Photon Sensors,

More information

J. Ling Chinese Academy of Science Institute for Ceramics Shanghai, China

J. Ling Chinese Academy of Science Institute for Ceramics Shanghai, China DEVELOPMENT AND APPLICATION OF LOCAL 3-D CT RECONSTRUCTION SOFTWARE FOR IMAGING CRITICAL REGIONS IN LARGE CERAMIC TURBINE ROTORS E. A. Sivers, D. L. Holloway, and W. A. Ellingson Argonne National Laboratory

More information

REVISIONS LTR. ECN DESCRIPTION DATE APPROVED - NA Original Document 11/13/02

REVISIONS LTR. ECN DESCRIPTION DATE APPROVED - NA Original Document 11/13/02 REVISIONS LTR. ECN DESCRIPTION DATE APPROVED - NA Original Document 11/13/02 SHEET REVISION STATUS SHEET 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 REVISION - - - - - - - - - - CONTROLLED COPY DIST.

More information

Time-Resolved measurements by FEL spontaneous emission: A proposal for sub-picosecond pumps & probe structural and spectrometric investigations

Time-Resolved measurements by FEL spontaneous emission: A proposal for sub-picosecond pumps & probe structural and spectrometric investigations Time-Resolved measurements by FEL spontaneous emission: A proposal for sub-picosecond pumps & probe structural and spectrometric investigations V. Rossi Albertini, B. Paci & P. Perfetti Istituto di Struttura

More information

Nonlinear optics and two photon microscopy. Table of contents. Sam Whiteley and Seth Parker PHYS 173/BGGN 266 July 13, 2014

Nonlinear optics and two photon microscopy. Table of contents. Sam Whiteley and Seth Parker PHYS 173/BGGN 266 July 13, 2014 Nonlinear optics and two photon microscopy Sam Whiteley and Seth Parker PHYS 173/BGGN 266 July 13, 2014 Table of contents 1. Introduction 2. Optical setup 3. Initial images and troubleshooting 4. Determining

More information

Alibava: A startup guide

Alibava: A startup guide Table of Contents Alibava: A startup guide Alibava Systems Basic connections and initialization of the system...3 Laser setup connections...4 Radioactive setup connections...4 Probing the Beetle ouput

More information

Scan System Parallelization

Scan System Parallelization Scan System Parallelization Recent Scan System Additions Kay Kasemir, Oct. 2015 n Sample n Neutron Detector n Events Data Collection Nexus File Position, Temperature,.. Experiment Control IOCs Histogram

More information

BENCHMIKE 283 SERIES. Off-Line ID/OD/Wall measurement systems for pipe and tube samples

BENCHMIKE 283 SERIES. Off-Line ID/OD/Wall measurement systems for pipe and tube samples BENCHMIKE 283 SERIES Off-Line ID/OD/Wall measurement systems for pipe and tube samples Measure manufactured parts or cut samples fast and with the highest accuracy Perform reliable, non-contact measurements

More information

PF32 QUICK START GUIDE. v1.4.9

PF32 QUICK START GUIDE. v1.4.9 PF32 QUICK START GUIDE v1.4.9 WARRANTY AND DISCLAIMERS Use of this product and the associated software implies acceptance of the Photon Force terms of use. Bare electronic components are susceptible to

More information

Single Photon Counting Module

Single Photon Counting Module Description Laser Components COUNT series of s has been developed to offer a unique combination of high quantum efficiency, wide dynamic range and ease of use for photon counting applications. Combining

More information

Classical Scattering Aerosol Spectrometer

Classical Scattering Aerosol Spectrometer Page 1 of 8 SOP DRAFT - MEASUREMENT OF THE AMBIENT AIR PARTICLE SIZE DISTRIBUTION USING THE PMS CSAS-100-HV Identification code: CSAS-100-HV SOP Working SOP pages APPROVALS Issue Date: /. Local PI: / /

More information

ksa MOS Ultra-Scan Performance Test Data

ksa MOS Ultra-Scan Performance Test Data ksa MOS Ultra-Scan Performance Test Data Introduction: ksa MOS Ultra Scan 200mm Patterned Silicon Wafers The ksa MOS Ultra Scan is a flexible, highresolution scanning curvature and tilt-measurement system.

More information

Polarization Ray Trace, Radiometric Analysis, and Calibration

Polarization Ray Trace, Radiometric Analysis, and Calibration Polarization Ray Trace, Radiometric Analysis, and Calibration Jesper Schou Instrument Scientist Stanford University jschou@solar.stanford.edu 650-725-9826 HMI00905 Page 1 Polarization Ray Trace Purpose

More information

Dr. Larry J. Paxton Johns Hopkins University Applied Physics Laboratory Laurel, MD (301) (301) fax

Dr. Larry J. Paxton Johns Hopkins University Applied Physics Laboratory Laurel, MD (301) (301) fax Dr. Larry J. Paxton Johns Hopkins University Applied Physics Laboratory Laurel, MD 20723 (301) 953-6871 (301) 953-6670 fax Understand the instrument. Be able to convert measured counts/pixel on-orbit into

More information

Refraction and Polarization of Light

Refraction and Polarization of Light Chapter 9 Refraction and Polarization of Light Name: Lab Partner: Section: 9.1 Purpose The purpose of this experiment is to demonstrate several consequences of the fact that materials have di erent indexes

More information

Operation Manual. Autocollimator Test Wedge for quick testing of Electronic Autocollimators

Operation Manual. Autocollimator Test Wedge for quick testing of Electronic Autocollimators for quick testing of Electronic Autocollimators For Ident.-No.: 223 244 Version date 05.05.2004 1. Basics The is designed for quick testing of accuracy of electronic autocollimators. If a light beam passes

More information

MCP and Photocathode Testing and Systems Integration At the Advanced Photon Source B. Adams, K. Attenkofer, M. Chollet, Z. Insepov, J. McPhate, O.

MCP and Photocathode Testing and Systems Integration At the Advanced Photon Source B. Adams, K. Attenkofer, M. Chollet, Z. Insepov, J. McPhate, O. MCP and Photocathode Testing and Systems Integration At the Advanced Photon Source B. Adams, K. Attenkofer, M. Chollet, Z. Insepov, J. McPhate, O. Siegmund, D. Walters, M. Wetstein, Z. Yusof for the LAPPD

More information

Calibration Contents

Calibration Contents Calibration Contents Calibration: Table of Contents... 1 Calibration: Instruction Set 1... 2 Starting Calibration: Manual and Multiple Calibration: Instruction Set 2a... 8 Manual Calibration Calibration:

More information

Physics 210 Medical Physics Midterm Exam Fall 2012 October 12, 2012

Physics 210 Medical Physics Midterm Exam Fall 2012 October 12, 2012 Physics 210 Medical Physics Midterm Exam Fall 2012 October 12, 2012 Name Problem 1 /32 Problem 2 /32 Problem 3 /24 Total /88 I affirm that I have carried out my academic endeavors with full academic honesty.

More information

SERGEY STEPANOV. Argonne National Laboratory, Advanced Photon Source, Lemont, IL, USA. October 2017, ICALEPCS-2017, Barcelona, Spain

SERGEY STEPANOV. Argonne National Laboratory, Advanced Photon Source, Lemont, IL, USA. October 2017, ICALEPCS-2017, Barcelona, Spain BEAMLINE AND EXPERIMENT AUTOMATIONS FOR THE GENERAL MEDICAL SCIENCES AND CANCER INSTITUTES STRUCTURAL BIOLOGY FACILITY AT THE ADVANCED PHOTON SOURCE (GM/CA@APS) SERGEY STEPANOV Argonne National Laboratory,

More information

Refraction and Polarization of Light

Refraction and Polarization of Light Chapter 9 Refraction and Polarization of Light Name: Lab Partner: Section: 9.1 Purpose The purpose of this experiment is to demonstrate several consequences of the fact that materials have di erent indexes

More information

BENCHMIKE PRO NEW. The Industry s leading off-line diameter & ovality measurement system

BENCHMIKE PRO NEW. The Industry s leading off-line diameter & ovality measurement system BENCHMIKE PRO NEW The Industry s leading off-line diameter & ovality measurement system Industry s leading sample inspection system Measure manufactured cut samples fast and with the highest accuracy in

More information

VDL 002 VariDelay I. Operation Manual

VDL 002 VariDelay I. Operation Manual VDL 002 VariDelay I Miniature Manual Variable Fiber Optic Delay Line Operation Manual May 6, 2009 General Photonics Corp. Ph: (909) 590-5473 5228 Edison Ave. Fax: (909) 902-5536 Chino, CA 91710 USA www.generalphotonics.com

More information

Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015

Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015 Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI Floris Jansen, GE Healthcare July, 2015 PET/MR 101 : challenges Thermal Workflow & Apps RF interactions?!!

More information

Fast Neutron Resonance Radiography in a Pulsed Neutron Beam. V. Dangendorf,

Fast Neutron Resonance Radiography in a Pulsed Neutron Beam. V. Dangendorf, Fast Neutron Resonance Radiography in a Pulsed Neutron Beam V. Dangendorf,, G. Laczko,, C. Kersten Physikalisch-Technische Bundesanstalt /Braunschweig,, Germany A. Breskin,, R. Chechik,, D. Vartsky Weizmann

More information

SensL s Second Generation of Red-Enhanced Silicon Photomultipliers. V 35 mm 23.0 Breakdown Voltage Range 4 ± 0.5 V

SensL s Second Generation of Red-Enhanced Silicon Photomultipliers. V 35 mm 23.0 Breakdown Voltage Range 4 ± 0.5 V SensL s Second Generation of Red-Enhanced Silicon Photomultipliers MicroRB sensors are the second release of Silicon Photomultipliers (SiPM) from SensL s R-Series range. The MicroRB sensors provide further

More information

Tutorial: Instantaneous Measurement of M 2 Beam Propagation Ratio in Real-Time

Tutorial: Instantaneous Measurement of M 2 Beam Propagation Ratio in Real-Time Tutorial: Instantaneous Measurement of M 2 Beam Propagation Ratio in Real-Time By Allen M. Cary, Jeffrey L. Guttman, Razvan Chirita, Derrick W. Peterman, Photon Inc A new instrument design allows the M

More information

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY BACKSCATTER DATA Young S. Ham and C.F. Poranski Materials Chemistry Branch Naval Research Laboratory Washington DC 20375 E.C. Greenawald Geo-Centers,

More information

BENCHMIKE 283 SERIES. BETA LaserMike. Off-Line ID/OD/Wall measurement systems for pipe and tube samples

BENCHMIKE 283 SERIES. BETA LaserMike. Off-Line ID/OD/Wall measurement systems for pipe and tube samples BENCHMIKE 283 SERIES Off-Line ID/OD/Wall measurement systems for pipe and tube samples Measure manufactured parts or cut samples fast and with the highest accuracy Perform reliable, non-contact measurements

More information

Single Photon Counting Module COUNT blue -Series

Single Photon Counting Module COUNT blue -Series Single Photon Counting Module COUNT blue -Series Description Laser Components COUNT blue series of Single Photon Counting Modules has been developed to offer a unique combination of high quantum efficiency,

More information

Industrial Fiber Beam Delivery for Ultrafast Lasers

Industrial Fiber Beam Delivery for Ultrafast Lasers Photonic Tools GmbH, 2017 Industrial Fiber Beam Delivery for Ultrafast Lasers System Technology and Industrial Application Bastian Kruschke 15.11.2017 Swissphotonics Workshop Specialty Optical Fibers Motivation

More information

LIFA KEY FEATURES APPLICATIONS. Fluorescence Lifetime Attachment LIFA 15001A02 16/03/2015

LIFA KEY FEATURES APPLICATIONS. Fluorescence Lifetime Attachment LIFA 15001A02 16/03/2015 LIFA Fluorescence Lifetime Attachment LIFA 151A2 16/3/215 The LIFA is a dedicated system for Fluorescence Lifetime Imaging Microscopy (FLIM). It allows the generation of lifetime images on any widefield

More information

Autofocus Systems in Machine Vision. Thomas Schäffler

Autofocus Systems in Machine Vision. Thomas Schäffler Autofocus Systems in Machine Vision Thomas Schäffler Agenda Definition: what is Autofocus (AF) Why do we need AF? Components of an AF system How to check for focus? How to adjust focus? Example system

More information

LECTURE 15. Dr. Teresa D. Golden University of North Texas Department of Chemistry

LECTURE 15. Dr. Teresa D. Golden University of North Texas Department of Chemistry LECTURE 15 Dr. Teresa D. Golden University of North Texas Department of Chemistry Typical steps for acquisition, treatment, and storage of diffraction data includes: 1. Sample preparation (covered earlier)

More information

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them Dose Distributions George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Purpose To familiarize the resident with dose distributions and the factors that affect them

More information

Two slit interference - Prelab questions

Two slit interference - Prelab questions Two slit interference - Prelab questions 1. Show that the intensity distribution given in equation 3 leads to bright and dark fringes at y = mλd/a and y = (m + 1/2) λd/a respectively, where m is an integer.

More information

Laser Contouring Method for Wall Thickness Metrology on OMEGA Shock Breakout Targets

Laser Contouring Method for Wall Thickness Metrology on OMEGA Shock Breakout Targets Laser Contouring Method for Wall Thickness Metrology on OMEGA Shock Breakout Targets R. J. Sebring 1, J. J. Bartos 1, R. D. Day 1, J. M. Edwards 1, A. Nobile 1, G. Rivera 1 and R. E, Olson 2. 1. Materials

More information

Fits you like no other

Fits you like no other Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room

More information

2003 TEC beam test : a quick overview

2003 TEC beam test : a quick overview 2003 TEC beam test : a quick overview Presentation: A petal in the beam line! Power supplies, cooling, interlock... The available measurements. Some plots taken from the online monitoring. Post X5 studies.

More information

PI-MTE3. Datasheet LARGE-FORMAT, IN-VACUUM CAMERAS. Direct detection of soft x-rays. Up to 4k x 4k sensor formats. Reads out as fast as 3 full fps

PI-MTE3. Datasheet LARGE-FORMAT, IN-VACUUM CAMERAS. Direct detection of soft x-rays. Up to 4k x 4k sensor formats. Reads out as fast as 3 full fps PI-MTE3 LARGE-FORMAT, IN-VACUUM CAMERAS Datasheet Direct detection of soft x-rays Up to 4k x 4k sensor formats Reads out as fast as 3 full fps PI-MTE3 DATASHEET Highlights Breakthrough In-Vacuum Performance

More information

Open access to photonic integration technologies

Open access to photonic integration technologies Open access to photonic integration technologies Academic and Industrial examples of photonic integrated circuits Katarzyna Ławniczuk k.lawniczuk@tue.nl What is photonic integration technology? multiple

More information

Charged Particle Tracking at Cornell: Gas Detectors and Event Reconstruction

Charged Particle Tracking at Cornell: Gas Detectors and Event Reconstruction Charged Particle Tracking at Cornell: Gas Detectors and Event Reconstruction Dan Peterson, Cornell University The Cornell group has constructed, operated and maintained the charged particle tracking detectors

More information

PH 481/581 Physical Optics Winter 2018

PH 481/581 Physical Optics Winter 2018 PH 481/581 Physical Optics Winter 2018 Laboratory #1 Week of January 15 Read: Section 5.2 (pp.151-175) of "Optics" by Hecht Do: 1. Experiment I.1: Thin Lenses 2. Experiment I.2: Alignment Project 3. Experiment

More information

Fits you like no other

Fits you like no other Fits you like no other Philips BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase

More information

CT Systems and their standards

CT Systems and their standards CT Systems and their standards Stephen Brown Engineering Measurement 11 th April 2012 Industrial X-ray computed tomography: The future of co-ordinate metrology? Burleigh Court, Loughborough University

More information

PRODUCT INFORMATION Model 1000XP Wide Range Particle Spectrometer (WPS )

PRODUCT INFORMATION Model 1000XP Wide Range Particle Spectrometer (WPS ) PRODUCT INFORMATION Model 1000XP Wide Range Particle Spectrometer (WPS ) With the lower size limit extended from 10nm to 5 nm, the WPS can now count and size aerosol particles automatically from 5nm to

More information

High Versatility High Throughput Functional Testing. Robert Polster, David Calhoun, Keren Bergman

High Versatility High Throughput Functional Testing. Robert Polster, David Calhoun, Keren Bergman High Versatility High Throughput Functional Testing Robert Polster, David Calhoun, Keren Bergman Challenges of High-Throughput Functional Testing for Scalable Manufacturing Numerous Functional Applications

More information

Using Photons Drift Time to Reconstruct Nuclear Processes and PET Event Topologies. Andrey Elagin University of Chicago

Using Photons Drift Time to Reconstruct Nuclear Processes and PET Event Topologies. Andrey Elagin University of Chicago Using Photons Drift Time to Reconstruct Nuclear Processes and PET Event Topologies Andrey Elagin University of Chicago IRRMA-X, Chicago, July 11, 2017 Outline Ultra-fast timing frontier Optical tracking

More information

Supported by. Alignment of the Thomson scattering diagnostic on NSTX. B.P. LeBlanc and A. Diallo Princeton Plasma Physics Laboratory

Supported by. Alignment of the Thomson scattering diagnostic on NSTX. B.P. LeBlanc and A. Diallo Princeton Plasma Physics Laboratory NSTX-U Supported by Coll of Wm & Mary Columbia U CompX General Atomics FIU INL Johns Hopkins U LANL LLNL Lodestar MIT Lehigh U Nova Photonics ORNL PPPL Princeton U Purdue U SNL Think Tank, Inc. UC Davis

More information

X-Ray fluorescence and Raman spectroscopy

X-Ray fluorescence and Raman spectroscopy X-Ray fluorescence and Raman spectroscopy Advanced physics laboratory (nd part) 4CFU Catalini Letizia, De Angelis Giulia Vittoria, Piselli Verdiana Abstract In this paper we report about two different

More information

NEW OPTICAL MEASUREMENT TECHNIQUE FOR SI WAFER SURFACE DEFECTS USING ANNULAR ILLUMINATION WITH CROSSED NICOLS

NEW OPTICAL MEASUREMENT TECHNIQUE FOR SI WAFER SURFACE DEFECTS USING ANNULAR ILLUMINATION WITH CROSSED NICOLS NEW OPTICAL MEASUREMENT TECHNIQUE FOR SI WAFER SURFACE DEFECTS USING ANNULAR ILLUMINATION WITH CROSSED NICOLS Satoru Takahashi 1, Takashi Miyoshi 1, Yasuhiro Takaya 1, and Takahiro Abe 2 1 Department of

More information

3D Photography: Stereo

3D Photography: Stereo 3D Photography: Stereo Marc Pollefeys, Torsten Sattler Spring 2016 http://www.cvg.ethz.ch/teaching/3dvision/ 3D Modeling with Depth Sensors Today s class Obtaining depth maps / range images unstructured

More information

Suppression of crosstalk in SiPMs for the next generation gamma-ray observatory, CTA

Suppression of crosstalk in SiPMs for the next generation gamma-ray observatory, CTA Suppression of crosstalk in SiPMs for the next generation gamma-ray observatory, CTA 16aS41-3 Hiro Tajima, Akira Okumura, Yuki Nakamura, Anatolii Zenin, Nagoya University (for the CTA-Japan Consortium)

More information

Physics 1230: Light and Color

Physics 1230: Light and Color Physics 1230: Light and Color Chuck Rogers, Charles.Rogers@colorado.edu Matt Heinemann, Matthew.Heinemann@colorado.edu www.colorado.edu/physics/phys1230 HWK 1 is due later today at 5PM on D2L. Take the

More information

APV-25 based readout electronics for the SBS front GEM Tracker

APV-25 based readout electronics for the SBS front GEM Tracker APV-25 based readout electronics for the SBS front GEM Tracker Authors: Evaristo Cisbani, Paolo Musico Date: 26/June/2014 Version: 1.0 APV-25 based readout electronics for the SBS front GEM Tracker...

More information

1. INTRODUCTION 2. OPAL LIDAR OVERVIEW

1. INTRODUCTION 2. OPAL LIDAR OVERVIEW Characterization of the OPAL LiDAR under controlled obscurant conditions Xiaoying Cao, Philip Church, Justin Matheson Neptec Technology Corporation, 302 Legget Drive, Kanata, Ontario, Canada K2K 1Y5 ABSTRACT

More information

Lecture 16 (Total Internal Reflection and Dispersion) Physics Spring 2018 Douglas Fields

Lecture 16 (Total Internal Reflection and Dispersion) Physics Spring 2018 Douglas Fields Lecture 16 (Total Internal Reflection and Dispersion) Physics 262-01 Spring 2018 Douglas Fields Reading Quiz In order to have total internal reflection, the index of refraction of the material with the

More information

The roughness model in SHADOW

The roughness model in SHADOW The roughness model in SHADOW Introduction SHADOW contains a model for analysing the surface roughness. Surface roughness is defined as irregularities in the optical surface which produces dispersion (scattering)

More information

The RAMM / MIM Platform for Microscopy A Modular Alternative for Innovative Research APPLIED SCIENTIFIC INSTRUMENTATION

The RAMM / MIM Platform for Microscopy A Modular Alternative for Innovative Research APPLIED SCIENTIFIC INSTRUMENTATION The RAMM / MIM Platform for Microscopy A Modular Alternative for Innovative Research Why a Modular Microscope? RAMM BASIC The RAMM Frame Simple single-purpose microscope can be less expensive than a general

More information

Spectroscopic equipment. Multispectral Imaging

Spectroscopic equipment. Multispectral Imaging Spectroscopic equipment Multispectral Imaging Basic spectroscopic arrangement Source Sample Analyzer Detector Sun Lamps Lasers LEDs Synchrotron Plants Forests Tissue Cells Flames Chemical compounds etc.

More information

BENCHMIKE PRO NEW. The Industry s leading off-line ID/OD/Wall measurement system

BENCHMIKE PRO NEW. The Industry s leading off-line ID/OD/Wall measurement system BENCHMIKE PRO NEW The Industry s leading off-line ID/OD/Wall measurement system Industry s leading sample inspection system Measure manufactured cut samples fast and with the highest accuracy in the industry

More information

arxiv:physics/ v1 [physics.ins-det] 18 Dec 1998

arxiv:physics/ v1 [physics.ins-det] 18 Dec 1998 Studies of 1 µm-thick silicon strip detector with analog VLSI readout arxiv:physics/981234v1 [physics.ins-det] 18 Dec 1998 T. Hotta a,1, M. Fujiwara a, T. Kinashi b, Y. Kuno c, M. Kuss a,2, T. Matsumura

More information

Other Laser Surgery Laser Tonsillectomy Use CO 2 with mirror bouncing system Operation takes 15 minutes, no pain Cauterizes blood vessels & Lymphatic

Other Laser Surgery Laser Tonsillectomy Use CO 2 with mirror bouncing system Operation takes 15 minutes, no pain Cauterizes blood vessels & Lymphatic Other Laser Surgery Laser Tonsillectomy Use CO 2 with mirror bouncing system Operation takes 15 minutes, no pain Cauterizes blood vessels & Lymphatic vessels no blood in throat Patient eat & drink just

More information

Timing properties of MCP-PMT

Timing properties of MCP-PMT Photon Detector Workshop at Kobe, 27-29 June 27 Timing properties of MCP-PMT - Time resolution - Lifetime - Rate dependence K.Inami (Nagoya university, Japan) Introduction Photon device for TOP counter

More information