4D Auto MVQ (Mitral Valve Quantification)

Size: px
Start display at page:

Download "4D Auto MVQ (Mitral Valve Quantification)"

Transcription

1 4D Auto MVQ (Mitral Valve Quantification) Federico Veronesi, Glenn Reidar Lie, Stein Inge Rabben GE Healthcare Introduction As the number of mitral valve repairs is on the rise, so is the need for mitral valve quantification. Due to the complex anatomy and physiology of the mitral valve apparatus, 4D visualization and quantification is becoming increasingly important. The use of 4D imaging prior to, during and after these repairs is mandatory for certain procedures. 4D quantification tools for the mitral valve have existed for some time and are available from several vendors. These tools provide more objective and accurate measurements as compared to 2D quantification tools, where access to desired views may be more difficult. The 4D Mitral Valve Quantification tools are usually automated to achieve quick access to the measurements of interest. GE Healthcare has over the years introduced a series of quantification tools for 2D (e.g. 2D Strain, AFI and 2D Auto EF) as well as for 4D (4D Auto LVQ, 4D LV Mass, 4D Strain, 4D Auto AVQ). GE Healthcare has also licensed tools from TomTec. These tools, 4D RV Function and Mitral Valve Assessment, were implemented as plug-ins into our 4D ultrasound systems (Vivid E9, Vivid S70 with 4D and Vivid E95) as well as the EchoPAC work station. We have now implemented GE Healthcare s 4D Auto MVQ tool as a replacement for TomTec s Mitral Valve Assessment. * This tool is seamlessly integrated into the Vivid systems workflow providing quick access to 4D raw data analysis. gehealthcare.com

2 Method 4D Auto MVQ is a quantification tool for rapid semi-automated detection of morphology of the mitral valve, including the annulus and the leaflets. Segmentation of the MV during systole is performed using a multistep approach. First, from the user-identified landmarks the MV annulus is identified by means of template matching along an a-priori 3D saddle-shaped line. Then, the MV leaflets are represented by a surface which during segmentation is deformed using image data together with the tracked annulus. The segmentation is represented as a state estimation problem and solved with an extended Kalman filter. From the detected surface, the MV coaptation line and possibly MV orifice is detected considering shape and thickness of anterior and posterior MV leaflets. 4D Auto MVQ workflow description The fully integrated tool requires a 4D data set containing the mitral valve acquired with a TTE or a TEE probe, either in a full volume or zoomed acquisition. Care must be taken during acquisition to ensure that the complete valve, with the ring/annulus is included, and that the volume rate exceeds 12 volumes per second for adequate temporal assessment if dynamic systolic performance is of interest. Speed of access to stored 4D images (from pressing the 4D Auto MVQ button on the touch panel until the 4D data is available for analysis) is twice as fast compared to its predecessor. The 4D rendering quality of the valve is similar to that of the original scanner acquisition throughout the entire quantification procedure. The workflow is divided into stages (see Figure 1) accessible both through the touch panel as well as on the main screen. When entering the tool, the first step is to adjust 2D gain, 4D gain and zoom if needed. These controls are available throughout the remaining stages to always ensure optimal image quality. Then one can set/adjust the End Systolic (ES) and End Diastolic (ED) time events if needed. This is followed by an Alignment stage (see Figure 2) to ensure a standardized and expected view is obtained where the vertical axis crosses through the center of the MV, while the horizontal axis is parallel to the mitral valve. Six Landmark points (see Figure 3) are then placed (two mitral annulus points, two anterior/posterior points, the coaptation point plus the aorta point). Using the above described method, the tool then generates a model for the leaflets and the annulus which can easily be edited by the operator through an intuitive and flexible user interface. The editing offers undo/redo capabilities, and one can at any time go back and forth between the various stages as one observes the results and the potential need for editing. Both systolic and full cycle loops can be played, though the tool only displays the model during systole. Figure 2. Alignment stage Figure 1. Stages Figure 3. Landmark stage

3 The tool offers three layouts at the Review stage (Biplane, Scallops and Commissures) depending on assessment needs (See Figure 4). In the Review stage, the 3D model can be checked in all slices. To visualize the model in different slices, the user must rotate and translate the reference slices in 2D view (dotted lines) and observe the segmentation in the other interactive views. The 3D model is adjusted by clicking and dragging contours on 2D views when needed. Five layouts are offered at the Results stage, see Figure 5 (Biplane, Scallops, Compact, Dynamic and Measurements). The Results layouts show all the Annulus and Leaflets parameters including three that were not supported in the predecessor tool (Inter-trigonal Distance, Anterior Leaflet Angle and Billowing Height). Figure 4. Review Biplane The Laser Lines introduced on GE systems years ago can be offered as an overlay on to the model to ease orientation both in the Review and Results stages. The ten most commonly requested and used measurements are transferred into the Worksheet for reporting when selecting the Approve & Exit button. Now, the tool also stores the complete setup and the results together with the exact screen view. When later recalling this particular image/loop and entering the 4D Auto MVQ tool, all prior alignments, landmarks, models and results are recalled for ease of continued analysis and review. This also serves as quality control. Figure 5. Results Compact Additionally, for documentation purposes at any time during use of the tool, image screen shots can be stored to the clipboard and be part of the study.

4 Validation The 4D Auto MVQ tool has been compared to the MV Assessment tool by TomTec on a set of 20 4D datasets including both normal subjects and patients affected by mitral valve disease. Results of the comparison are reported in table 1 as average ± standard deviation. For further information regarding the validation, contact your local representative. Limitations: The measurements obtained from the two products compared in the validation span a clinically significant range. However, the number of subjects analyzed is limited and further studies on specific populations are needed to identify the clinical benefit of using advanced 3D tools to characterize the MV. Both 4D Auto MVQ (and MV Assessment TomTec) use an automated segmentation algorithm to initiate a 3D model from where the tool s measurements are calculated. There is an inherent risk that these automated segmentation algorithms may give inaccurate 3D models. That means that the tool results should not be approved without review and, if necessary, editing of the 3D model. If the tools are used without review and editing, the expected accuracy will be negatively affected. Table 1. Comparison between 4D Auto MV and MVA Assessment tool by TomTec Annulus Perimeter Measurement unit Values MVQ Values TomTec TomTec - MVQ Difference cm 13.7 ± ± ± 0.58 * Annulus Area 3D cm ± ± ± 1.37 * AL-PM diameter cm 4.44 ± ± ± 0.25 * A-P diameter cm 3.76 ± ± ± 0.34 * Annulus Height mm 7.05 ± ± ± 1.58 * Mitral Annular Excursion Leaflet Angle Mitral-Aortic Angle Anterior Leaflet Area Leaflet Area mm 6.58 ± ± ± 2.1 * ± ± ± 6.85 * ± ± ± 8.59 * cm ± ± ± 1.0 * cm ± ± ± 1.0 * Tenting Height cm 0.62 ± ± ± 0.26 * Anterior Leaflet Length Leaflet Length *Not significantly different (Two-tails paired Student s t-test with P>0.05) cm 2.48 ± ± ± 0.23 * cm 1.77 ± ± ± 0.24 *

5 Imagination at work * Available on the Vivid E95 and Vivid S70N 4D Productivity Elevated release General Electric Company All rights reserved. GE Healthcare reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE Healthcare representative for the most current information. Please visit GE, the GE Monogram, imagination at work, Vivid and EchoPAC are trademarks of General Electric Company or one of its subsidiaries. GE Healthcare, a division of General Electric Company. TomTec is a trademark of TomTec Imaging Systems GmbH. Third party trademarks are the property of their respective owners. GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC, a General Electric Company, doing business as GE Healthcare. JB48903XX

GE Healthcare. Vivid 7 Dimension 08 Cardiovascular ultrasound system

GE Healthcare. Vivid 7 Dimension 08 Cardiovascular ultrasound system GE Healthcare Vivid 7 Dimension 08 Cardiovascular ultrasound system ltra Definition. Technology. Performance. Start with a system that s proven its worth in LV quantification and 4D imaging. Then add even

More information

GE Healthcare. Vivid E9 XDclear. with

GE Healthcare. Vivid E9 XDclear. with GE Healthcare Vivid E9 XDclear with capabilities for every day Whether in TTE or TEE, capture the entire heart in a single beat. Image a valve, or the entire ventricle with excellent image quality. And

More information

GE Healthcare. Vivid E9 XDclear. with. healthymagination. GE imagination at work

GE Healthcare. Vivid E9 XDclear. with. healthymagination. GE imagination at work GE Healthcare with Vivid E9 XDclear healthymagination GE imagination at work Extraordinary capabilities Whether in TTE or TEE, capture the entire heart in a single beat. Image a valve, or the entire ventricle

More information

EchoPAC Software Only v201

EchoPAC Software Only v201 EchoPAC Software Only v201 Product Description The EchoPAC Application is a comprehensive clinical software package for viewing, analyzing, and reporting of multi-dimensional echo, vascular and abdominal

More information

What you should know about the new release

What you should know about the new release T.A8.4401-01 TomTec-Arena 1.1 What you should know about the new release Congratulations to the purchase of your new TomTec Software TomTec-Arena and thank you for choosing TomTec Imaging Systems GmbH

More information

csound 2.0 Vivid E95 gehealthcare.com Background

csound 2.0 Vivid E95 gehealthcare.com Background csound 2.0 Vivid E95 Background Through the Healthymagination initiative, GE Healthcare continuously invests in innovations that help lower the cost, increase access, and improve the quality of healthcare.

More information

A powerful, software-based beamformer image reconstruction platform

A powerful, software-based beamformer image reconstruction platform csound A powerful, software-based beamformer image reconstruction platform Figure 1: Vivid S70 and E90/95; first GE systems built upon the csound platform Background GE s cardiovascular ultrasound imaging

More information

Ultrasound. Q-Station software. Streamlined workflow solutions. Philips Q-Station ultrasound workspace software

Ultrasound. Q-Station software. Streamlined workflow solutions. Philips Q-Station ultrasound workspace software Ultrasound Q-Station software Streamlined workflow solutions Philips Q-Station ultrasound workspace software Managing your off-cart workf low Everyone is being asked to do more with fewer resources it

More information

Accurate Regression-based 4D Mitral Valve Segmentation from 2D MRI Slices

Accurate Regression-based 4D Mitral Valve Segmentation from 2D MRI Slices Accurate Regression-based 4D Mitral Valve Segmentation from 2D MRI Slices Dime Vitanovski 2,3, Alexey Tsymbal 2, Razvan Ioan Ionasec 1, Andreas Greiser 4, Edgar Mueller 4, Xiaoguang Lu 1, Gareth Funka-Lea

More information

LOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company.

LOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company. TM LOGIQ V2 Ultrasound Part of LOGIQ Vision Series Imagination at work The brilliance of color. The simplicity of GE. Now you can add the advanced capabilities of color Doppler to patient care with the

More information

3D Echocardiography Development Timeline

3D Echocardiography Development Timeline 3D Acquisition Strategies and Display Lissa Sugeng, MD, MPH Associate Professor of Medicine Director of the Yale Echo Lab and Echo Core Lab (YCRG) And Peter Flueckiger, MD Advanced Imaging Cardiology Fellow

More information

Modeling Mitral Valve Leaflets from Three-Dimensional Ultrasound

Modeling Mitral Valve Leaflets from Three-Dimensional Ultrasound Modeling Mitral Valve Leaflets from Three-Dimensional Ultrasound Robert J. Schneider 1, William C. Burke 1, Gerald R. Marx 3, Pedro J. del Nido 2, and Robert D. Howe 1 1 Harvard School of Engineering and

More information

Using TeraRecon intuition Viewer with STAT Quick Reference Guide

Using TeraRecon intuition Viewer with STAT Quick Reference Guide Using TeraRecon intuition Viewer with STAT Quick Reference Guide 1. Launching intuition 2. 3D Assessment of Coronary Arteries for Stenosis 3. Time Volume Analysis (TVA) for Determination of Left Ventricular

More information

This is your imaging dashboard. There is a bevy of valuable information to read from here. Let s break it up.

This is your imaging dashboard. There is a bevy of valuable information to read from here. Let s break it up. The Volume Strikes Back Press the (4D) button. This is your imaging dashboard. There is a bevy of valuable information to read from here. Let s break it up. In the Left GUI (Graphic User Interface), you

More information

Volumetric Analysis of the Heart from Tagged-MRI. Introduction & Background

Volumetric Analysis of the Heart from Tagged-MRI. Introduction & Background Volumetric Analysis of the Heart from Tagged-MRI Dimitris Metaxas Center for Computational Biomedicine, Imaging and Modeling (CBIM) Rutgers University, New Brunswick, NJ Collaboration with Dr. Leon Axel,

More information

Technical Principles of Transthoracic Three-Dimensional Echocardiography

Technical Principles of Transthoracic Three-Dimensional Echocardiography Technical Principles of Transthoracic Three-Dimensional Echocardiography 2 Stein Inge Rabben 2.1 Introduction The last years we have experienced a rapid development in three-dimensional echocardiography

More information

icatvision Quick Reference

icatvision Quick Reference icatvision Quick Reference Navigating the i-cat Interface This guide shows how to: View reconstructed images Use main features and tools to optimize an image. REMINDER Images are displayed as if you are

More information

Mitral Annulus Segmentation From 3D Ultrasound Using Graph Cuts

Mitral Annulus Segmentation From 3D Ultrasound Using Graph Cuts Mitral Annulus Segmentation From 3D Ultrasound Using Graph Cuts The MIT Faculty has made this article openly available. Please share how this access benefits you. Your story matters. Citation As Published

More information

PASS Sample Size Software

PASS Sample Size Software Chapter 941 Introduction In PASS, it is easy to study power and sample size calculations for a range of possible parameter values. When at least 2 input parameters vary, you can create stunning 3D power

More information

3D Surface Plots with Groups

3D Surface Plots with Groups Chapter 942 3D Surface Plots with Groups Introduction In PASS, it is easy to study power and sample size calculations for a range of possible parameter values. When at least 3 input parameters vary, you

More information

Chapter 9 Conclusions

Chapter 9 Conclusions Chapter 9 Conclusions This dissertation has described a new method for using local medial properties of shape to identify and measure anatomical structures. A bottom up approach based on image properties

More information

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc. Unique Features of the GE Senoclaire Tomosynthesis System Tyler Fisher, M.S., DABR Therapy Physics, Inc. Conflict of Interest Disclosure I have no conflicts to disclose. Learning Objectives Overview of

More information

QAngio XA 7.3. Quick Start Manual. July 31, v6.0

QAngio XA 7.3. Quick Start Manual. July 31, v6.0 QAngio XA 7.3 Quick Start Manual July 31, 2018 9.04.250.73.6 v6.0 Medis medical imaging systems bv Schuttersveld 9, 2316 XG Leiden, the Netherlands http://www.medis.nl Medis medical imaging systems bv

More information

Analysis of CMR images within an integrated healthcare framework for remote monitoring

Analysis of CMR images within an integrated healthcare framework for remote monitoring Analysis of CMR images within an integrated healthcare framework for remote monitoring Abstract. We present a software for analyzing Cardiac Magnetic Resonance (CMR) images. This tool has been developed

More information

Contour LS-K Optical Surface Profiler

Contour LS-K Optical Surface Profiler Contour LS-K Optical Surface Profiler LightSpeed Focus Variation Provides High-Speed Metrology without Compromise Innovation with Integrity Optical & Stylus Metrology Deeper Understanding More Quickly

More information

Presenting System 1* v18.1

Presenting System 1* v18.1 Presenting System 1* v18.1 Jackie Tappan Senior Staff Technical Product Manager Bently Nevada* Software Development Team Bently Nevada is pleased to present System 1 Version 18.1. The added capabilities

More information

Leica Cyclone 5.4 Technical Specifications

Leica Cyclone 5.4 Technical Specifications Leica Cyclone 5.4 Technical Specifications HDS Scanner control and operation Scan Scan Scan Register Model Survey Viewer Acquire and display image Acquire image at specified resolution (high, medium, low)

More information

A Semi-automatic Endocardial Border Detection Method for 4D Ultrasound Data

A Semi-automatic Endocardial Border Detection Method for 4D Ultrasound Data A Semi-automatic Endocardial Border Detection Method for 4D Ultrasound Data Marijn van Stralen 1, Johan G. Bosch 1, Marco M. Voormolen 2,3, Gerard van Burken 1, Boudewijn J. Krenning 3, Charles T. Lancée

More information

Development of 3D Model-based Morphometric Method for Assessment of Human Weight-bearing Joint. Taeho Kim

Development of 3D Model-based Morphometric Method for Assessment of Human Weight-bearing Joint. Taeho Kim Development of 3D Model-based Morphometric Method for Assessment of Human Weight-bearing Joint Taeho Kim Introduction Clinical measurement in the foot pathology requires accurate and robust measurement

More information

PREMIER DIAGNOSTIC ULTRASOUND. Diagnostic ultrasound. that delivers unparalleled. image resolution.

PREMIER DIAGNOSTIC ULTRASOUND. Diagnostic ultrasound. that delivers unparalleled. image resolution. PREMIER DIAGNOSTIC ULTRASOUND Diagnostic ultrasound that delivers unparalleled image resolution. High Resolution Goes Ultra THE NEW GENERATION EYE CUBED TM M A K E S I T P O S S I B L E T O EVALUATE OCULAR

More information

8/3/2017. Contour Assessment for Quality Assurance and Data Mining. Objective. Outline. Tom Purdie, PhD, MCCPM

8/3/2017. Contour Assessment for Quality Assurance and Data Mining. Objective. Outline. Tom Purdie, PhD, MCCPM Contour Assessment for Quality Assurance and Data Mining Tom Purdie, PhD, MCCPM Objective Understand the state-of-the-art in contour assessment for quality assurance including data mining-based techniques

More information

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package.

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package. So much is asked of medical physicists these days We Can Help Automated Radiation Therapy Phantom Analysis in Seconds QC for Therapy OBI s and CT simulators. Radia s Catphan/OBI module performs analysis

More information

Robust Live Tracking of Mitral Valve Annulus for Minimally-Invasive Intervention Guidance

Robust Live Tracking of Mitral Valve Annulus for Minimally-Invasive Intervention Guidance Robust Live Tracking of Mitral Valve Annulus for Minimally-Invasive Intervention Guidance Ingmar Voigt 1, Mihai Scutaru 2, Tommaso Mansi 3, Bogdan Georgescu 3, Noha El-Zehiry 3, Helene Houle 4, and Dorin

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

BUZZ - DIGITAL OR WHITEPAPER

BUZZ - DIGITAL OR WHITEPAPER Video Inputs for 2x SDI, 2x DVI, Svideo, Composite, VGA Connection of additional 3 Full HD displays full HD Streaming, Recording (2 channels) OR Portal DICOM Image Viewing Video routing Integration of

More information

PalmScan PRO 5 in 1 Biometer Your Complete Biometry System - Fully Configurable

PalmScan PRO 5 in 1 Biometer Your Complete Biometry System - Fully Configurable B-Scan A-Scan Pachymeter Keratometer UBM PalmScan PRO 5 in 1 Biometer Your Complete Biometry System - Fully Configurable A-Scan + IOL Calculator Keratometer Pachymeter B-Scan UBM www.micromedinc.com 818.222.3310

More information

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement Application Annex: HeartNavigator R2.0.5 Koninklijke Philips N.V. 2015 All rights are reserved. Document Number: ICAP-PF.0015391 Issued by: Philips Medical Systems Nederland

More information

Segment CT Instructions For Use - English

Segment CT Instructions For Use - English Segment CT Instructions For Use - English February 28, 2018 Software platform v2.2 R6239 MEDVISO AB http://www.medviso.com Griffelvägen 3 SE-224 67 Lund Sweden Tel: +46-76-183 6442 ii Contents 1 Terms

More information

Resona 7. New Waves in Ultrasound Innovation. Premium Ultrasound System

Resona 7. New Waves in Ultrasound Innovation. Premium Ultrasound System Resona 7 Premium Ultrasound System New Waves in Ultrasound Innovation With over 20 years of experience, Mindray hosts a wide range of ultrasound imaging solutions including cart-based and portable systems.

More information

Improving Access to Healthcare. CLINiC. Clinical Care Device. Capsa Cart & Wall Mount Solutions Featuring the CLINiC

Improving Access to Healthcare. CLINiC. Clinical Care Device. Capsa Cart & Wall Mount Solutions Featuring the CLINiC Improving Access to Healthcare CLINiC Clinical Care Device Capsa Cart & Wall Mount Solutions Featuring the CLINiC CLINiC Clinical Care Device Options Simple Design Supports Multiple Peripherals Mounting

More information

NUCLeUS: Your Partner Before, During & After Surgery

NUCLeUS: Your Partner Before, During & After Surgery NUCLeUS: Your Partner Before, During & After Surgery Today s operating rooms are continuously evolving as new medical equipment arrives on the scene. Any future oriented hospital should consider tomorrow

More information

Resona 7. Premium Ultrasound System. New Waves in Ultrasound Innovation

Resona 7. Premium Ultrasound System. New Waves in Ultrasound Innovation Resona 7 Premium Ultrasound System New Waves in Ultrasound Innovation New Waves in Ultrasound Innovation Since the company was founded, Mindray is continuously exploring new ways to improve diagnostic

More information

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals.

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals. GE Healthcare CLINICAL GALLERY Discovery * MR750w 3.0T This brochure is intended for European healthcare professionals. NEURO PROPELLER delivers high resolution, motion insensitive imaging in all planes.

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION doi:10.1038/nature12009 Supplementary Figure 1. Experimental tilt series of 104 projections with a tilt range of ±72.6 and equal slope increments, acquired from a Pt nanoparticle using HAADF- STEM (energy:

More information

Fluorescence Tomography Source Reconstruction and Analysis

Fluorescence Tomography Source Reconstruction and Analysis TECHNICAL NOTE Pre-clinical in vivo imaging Fluorescence Tomography Source Reconstruction and Analysis Note: This Technical Note is part of a series for Fluorescence Imaging Tomography (FLIT). The user

More information

SHARP Pen Software. Version 2.2 OPERATION MANUAL

SHARP Pen Software. Version 2.2 OPERATION MANUAL SHARP Pen Software Version 2.2 OPRATION MANUAL Contents Introduction...3 Starting and xiting the Software...4 Pen Software modes and how to start the modes...4 xiting the software...5 Layout of Basic Screen...6

More information

syngo.mr Neuro 3D: Your All-In-One Post Processing, Visualization and Reporting Engine for BOLD Functional and Diffusion Tensor MR Imaging Datasets

syngo.mr Neuro 3D: Your All-In-One Post Processing, Visualization and Reporting Engine for BOLD Functional and Diffusion Tensor MR Imaging Datasets syngo.mr Neuro 3D: Your All-In-One Post Processing, Visualization and Reporting Engine for BOLD Functional and Diffusion Tensor MR Imaging Datasets Julien Gervais; Lisa Chuah Siemens Healthcare, Magnetic

More information

EPSRC Centre for Doctoral Training in Industrially Focused Mathematical Modelling

EPSRC Centre for Doctoral Training in Industrially Focused Mathematical Modelling EPSRC Centre for Doctoral Training in Industrially Focused Mathematical Modelling More Accurate Optical Measurements of the Cornea Raquel González Fariña Contents 1. Introduction... 2 Background... 2 2.

More information

PART I GravoStyle5-Laser Introduction

PART I GravoStyle5-Laser Introduction PART I GravoStyle5-Laser Introduction I. INTRO GravoStyle 5 Laser is designed is a component of GravoStyle5 for use with the Gravograph/New Hermes and other manufacturer Laser Engravers. Combined with

More information

ilook 25 Next Generation in Portable Ultrasound Delivering the Personal Imaging Tool

ilook 25 Next Generation in Portable Ultrasound Delivering the Personal Imaging Tool Delivering the Next Generation in Portable Ultrasound Designed to enhance the ease, efficiency, accuracy and reliability of vascular access User-Friendly System Controls Flexible System Design > Facilitate

More information

XD 3.6 Release Notes

XD 3.6 Release Notes XD 3.6 Release Notes 1 Introduction 1.1 Purpose The purpose of these Release Notes is to: 1.2 Scope Communicate any known limitations for Mirada s XD release. Provide information on certain product features.

More information

T-SCAN 3 3D DIGITIZING

T-SCAN 3 3D DIGITIZING T-SCAN 3 3D DIGITIZING 2 T-SCAN 3: THE HANDHELD LASER SCANNER Launching the innovative concept of an intuitive-to-use high-precision laser scanner a few years ago, Steinbichler Optotechnik, as the first

More information

Morphometric Analysis of Biomedical Images. Sara Rolfe 10/9/17

Morphometric Analysis of Biomedical Images. Sara Rolfe 10/9/17 Morphometric Analysis of Biomedical Images Sara Rolfe 10/9/17 Morphometric Analysis of Biomedical Images Object surface contours Image difference features Compact representation of feature differences

More information

Case TAVR080: Aortic Valve Step-by-Step Guide

Case TAVR080: Aortic Valve Step-by-Step Guide Case TAVR080: Aortic Valve Step-by-Step Guide 3D printing the aortic valve The 3D printed anatomical model is not for diagnostic use. M-06235C 208 Vital Images, Inc. www.vitalimages.com Case Airways -

More information

By your side. Anytime. Anywhere.

By your side. Anytime. Anywhere. By your side. Anytime. Anywhere. Portability without compromise If unrestricted access and diagnostic performance at the point of care are your priorities, Viamo c100 is your ideal solution. Viamo c100

More information

YOUR WAY TO PRECISE OUTCOMES

YOUR WAY TO PRECISE OUTCOMES Microscope systems for markerless IOL alignment during cataract surgery YOUR WAY TO PRECISE OUTCOMES 2D and 3D IOL guidance systems 2 YOUR WAY TO PRECISE OUTCOMES Drive minimum residual astigmatism for

More information

Object Identification in Ultrasound Scans

Object Identification in Ultrasound Scans Object Identification in Ultrasound Scans Wits University Dec 05, 2012 Roadmap Introduction to the problem Motivation Related Work Our approach Expected Results Introduction Nowadays, imaging devices like

More information

EnSite NavX Navigation and Visualization Technology EnSite Fusion Registration Module Procedure Guide

EnSite NavX Navigation and Visualization Technology EnSite Fusion Registration Module Procedure Guide EnSite NavX Navigation and Visualization Technology EnSite Fusion Registration Module Procedure Guide The EnSite Fusion Module enables Dynamic Registration, allowing a DIF model (a surface rendering of

More information

Ultrasound To Go. MySono U5

Ultrasound To Go. MySono U5 Ultrasound To Go MySono U5 Ultrasound To Go With the introduction of the MySono U5, Samsung Medison brings you a fully featured ultrasound imaging system to go. Delivering exceptional image quality and

More information

The Supreme 3D/4D Ultrasound. The Supreme 3D/4D Ultrasound. CT-V20-ICM

The Supreme 3D/4D Ultrasound. The Supreme 3D/4D Ultrasound.   CT-V20-ICM The Supreme 3D/4D Ultrasound The Supreme 3D/4D Ultrasound www.medison.com info@medison.com CT-V20-ICM-06.20.2008 The Supreme 3D/4D Ultrasound Since launching the first commercially available 3D ultrasound

More information

Quantitative Three-Dimensional Imaging of the Posterior Segment with the Heidelberg Retina Tomograph

Quantitative Three-Dimensional Imaging of the Posterior Segment with the Heidelberg Retina Tomograph Quantitative Three-Dimensional Imaging of the Posterior Segment with the Heidelberg Retina Tomograph Heidelberg Engineering GmbH, Heidelberg, Germany Contents 1 Introduction... 1 2 Confocal laser scanning

More information

SHARP Pen Software. Version 2.3 OPERATION MANUAL

SHARP Pen Software. Version 2.3 OPERATION MANUAL SHARP Pen Software Version 2.3 OPRATION MANUAL Contents Introduction...3 Starting and xiting the Software...4 Pen Software modes and how to start the modes...4 xiting the software...5 Layout of Basic Screen...6

More information

Anatomic Growth Modelling of Cleft Palate Shape. S.K. Chua*, S.H. Ong*, K.W.C. Foong**

Anatomic Growth Modelling of Cleft Palate Shape. S.K. Chua*, S.H. Ong*, K.W.C. Foong** ABSTRACT Anatomic Growth Modelling of Cleft Palate Shape S.K. Chua*, S.H. Ong*, K.W.C. Foong** *Dept of Electrical and Computer Engineering, **Dept of Preventive Dentistry, National University of Singapore

More information

연구용유방초음파질관리 원광대학병원김혜원

연구용유방초음파질관리 원광대학병원김혜원 연구용유방초음파질관리 원광대학병원김혜원 Why US QC? Quality control (QC) testing of ultrasound scanners is important to verify the proper and consistent operation of these devices. main goal ; quality improvement Guidelines

More information

Dell Canvas Layout. Version 1.0 User s Guide

Dell Canvas Layout. Version 1.0 User s Guide Dell Canvas Layout Version 1.0 User s Guide Notes, cautions, and warnings NOTE: A NOTE indicates important information that helps you make better use of your product. CAUTION: A CAUTION indicates either

More information

CADMATIC is a registered trademark of CADMATIC Oy. All other trademarks are the property of their respective owners.

CADMATIC is a registered trademark of CADMATIC Oy. All other trademarks are the property of their respective owners. Diagram Outfitting & Plant Design Hull Information Management: eshare, ebrowser, ego Release highlights 2018T2 Version 2018T2 includes many additions that increase the productivity of work in 3D and the

More information

From Eye to Insight Hand-held, contact-free Optical Coherence Tomography for pediatrics ENVISU C2300 OCT Take OCT to your patient FDA 510(k) Cleared

From Eye to Insight Hand-held, contact-free Optical Coherence Tomography for pediatrics ENVISU C2300 OCT Take OCT to your patient FDA 510(k) Cleared From Eye to Insight Hand-held, contact-free Optical Coherence Tomography for pediatrics ENVISU C2300 OCT Take OCT to your patient FDA 510(k) Cleared ENVISU C-CLASS EASY ON YOUR PATIENT The origins of pathology

More information

QMass 8.1. Quick Start Manual. June 13, v2.0

QMass 8.1. Quick Start Manual. June 13, v2.0 QMass 8.1 Quick Start Manual June 13, 2017 9.10.250.81.3 v2.0 Medis medical imaging systems bv Schuttersveld 9, 2316 XG Leiden, the Netherlands http://www.medis.nl Medis medical imaging systems bv Medis

More information

A Comprehensive Method for Geometrically Correct 3-D Reconstruction of Coronary Arteries by Fusion of Intravascular Ultrasound and Biplane Angiography

A Comprehensive Method for Geometrically Correct 3-D Reconstruction of Coronary Arteries by Fusion of Intravascular Ultrasound and Biplane Angiography Computer-Aided Diagnosis in Medical Imaging, September 20-23, 1998, Chicago IL Elsevier ICS 1182 A Comprehensive Method for Geometrically Correct 3-D Reconstruction of Coronary Arteries by Fusion of Intravascular

More information

SAFI Sample Projects. Design of a Steel Structure. SAFI Quality Software Inc. 3393, chemin Sainte-Foy Ste-Foy, Quebec, G1X 1S7 Canada

SAFI Sample Projects. Design of a Steel Structure. SAFI Quality Software Inc. 3393, chemin Sainte-Foy Ste-Foy, Quebec, G1X 1S7 Canada SAFI Sample Projects Design of a Steel Structure SAFI Quality Software Inc. 3393, chemin Sainte-Foy Ste-Foy, Quebec, G1X 1S7 Canada Contact: Rachik Elmaraghy, P.Eng., M.A.Sc. Tel.: 1-418-654-9454 1-800-810-9454

More information

7D Cardiac Flow MRI: Techniques & Automation of Reconstruction

7D Cardiac Flow MRI: Techniques & Automation of Reconstruction Cleveland State University EngagedScholarship@CSU ETD Archive 2012 7D Cardiac Flow MRI: Techniques & Automation of Reconstruction Michael G. Ambrosia Cleveland State University Follow this and additional

More information

Creating a Basic Chart in Excel 2007

Creating a Basic Chart in Excel 2007 Creating a Basic Chart in Excel 2007 A chart is a pictorial representation of the data you enter in a worksheet. Often, a chart can be a more descriptive way of representing your data. As a result, those

More information

Roof Designer USER S GUIDE

Roof Designer USER S GUIDE Roof Designer USER S GUIDE Roof Designer-1 Roof Designer-2 Roof Designer The Roof Designer makes it easy to define and place custom roofs in your project. You can start the Roof Designer independently,

More information

Coke Drum Laser Profiling

Coke Drum Laser Profiling International Workshop on SMART MATERIALS, STRUCTURES NDT in Canada 2013Conference & NDT for the Energy Industry October 7-10, 2013 Calgary, Alberta, CANADA Coke Drum Laser Profiling Mike Bazzi 1, Gilbert

More information

Impact of 3D Laser Data Resolution and Accuracy on Pipeline Dents Strain Analysis

Impact of 3D Laser Data Resolution and Accuracy on Pipeline Dents Strain Analysis More Info at Open Access Database www.ndt.net/?id=15137 Impact of 3D Laser Data Resolution and Accuracy on Pipeline Dents Strain Analysis Jean-Simon Fraser, Pierre-Hugues Allard Creaform, 5825 rue St-Georges,

More information

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization

More information

Automatic Subthalamic Nucleus Targeting for Deep Brain Stimulation. A Validation Study

Automatic Subthalamic Nucleus Targeting for Deep Brain Stimulation. A Validation Study Automatic Subthalamic Nucleus Targeting for Deep Brain Stimulation. A Validation Study F. Javier Sánchez Castro a, Claudio Pollo a,b, Jean-Guy Villemure b, Jean-Philippe Thiran a a École Polytechnique

More information

Microsoft Office Excel

Microsoft Office Excel Microsoft Office 2007 - Excel Help Click on the Microsoft Office Excel Help button in the top right corner. Type the desired word in the search box and then press the Enter key. Choose the desired topic

More information

NEW! PORTABLE DIAGNOSTIC ULTRASOUND PLATFORM. Considered to be the gold standard in ophthalmic diagnostic ultrasound,

NEW! PORTABLE DIAGNOSTIC ULTRASOUND PLATFORM. Considered to be the gold standard in ophthalmic diagnostic ultrasound, NEW! PORTABLE DIAGNOSTIC ULTRASOUND PLATFORM Considered to be the gold standard in ophthalmic diagnostic ultrasound, Eye Cubed features advanced movie mode using the fastest sampling Eye One rate available,

More information

Kaluza C Analysis Software PROTECTS YOUR DATA AND YOUR REPUTATION.

Kaluza C Analysis Software PROTECTS YOUR DATA AND YOUR REPUTATION. Kaluza C Analysis Software PROTECTS YOUR DATA AND YOUR REPUTATION. KALUZA C ANALYSIS SOFTWARE Kaluza C Flow Cytometry Analysis Software is built upon our successful Kaluza Analysis research use platform.

More information

QIBA PET Amyloid BC March 11, Agenda

QIBA PET Amyloid BC March 11, Agenda QIBA PET Amyloid BC March 11, 2016 - Agenda 1. QIBA Round 6 Funding a. Deadlines b. What projects can be funded, what cannot c. Discussion of projects Mechanical phantom and DRO Paul & John? Any Profile

More information

What s new in Solutionix s scanners?

What s new in Solutionix s scanners? SOLUTIONIX 1 What s new in Solutionix s scanners? Solutionix provides fully automated 3D scanning solutions for both small and medium sized objects. With a powerful scan engine and affordable pricing,

More information

Mitral Annulus Segmentation from 3D Ultrasound Using Graph Cuts

Mitral Annulus Segmentation from 3D Ultrasound Using Graph Cuts IEEE TRANSACTIONS ON MEDICAL IMAGING 1 Annulus Segmentation from 3D Ultrasound Using Graph Cuts Robert J. Schneider, Douglas P. Perrin, Nikolay V. Vasilyev, Gerald R. Marx, Pedro J. del Nido, and Robert

More information

GE Healthcare. AppsLinq* remote courses catalogue

GE Healthcare. AppsLinq* remote courses catalogue GE Healthcare AppsLinq* remote courses catalogue AppsLinq * remote training for MR Magnetic Resonance AppsLinq* a GE Training in Partnership (TiP) program, revolutionizes applications training with live,

More information

Improve your image. ClearVue 350 ultrasound system

Improve your image. ClearVue 350 ultrasound system Improve your image ClearVue 350 ultrasound system Smart and intuitive The system looks sleek from the outside, but it s what s on the inside that you ll value most. ClearVue 350 introduces Philips proprietary

More information

coding of various parts showing different features, the possibility of rotation or of hiding covering parts of the object's surface to gain an insight

coding of various parts showing different features, the possibility of rotation or of hiding covering parts of the object's surface to gain an insight Three-Dimensional Object Reconstruction from Layered Spatial Data Michael Dangl and Robert Sablatnig Vienna University of Technology, Institute of Computer Aided Automation, Pattern Recognition and Image

More information

User Manual 2.1.1(draft)

User Manual 2.1.1(draft) User Manual 2.1.1(draft) Rev. March 31, 2017 Introduction BallisticsARC has 3 modes; Chart, Map, and Comp. Each mode produces solutions independently, however data entered through the main menu will apply

More information

STIC AmSud Project. Graph cut based segmentation of cardiac ventricles in MRI: a shape-prior based approach

STIC AmSud Project. Graph cut based segmentation of cardiac ventricles in MRI: a shape-prior based approach STIC AmSud Project Graph cut based segmentation of cardiac ventricles in MRI: a shape-prior based approach Caroline Petitjean A joint work with Damien Grosgeorge, Pr Su Ruan, Pr JN Dacher, MD October 22,

More information

A Framework for Real-time Left Ventricular tracking in 3D+T Echocardiography, Using Nonlinear Deformable Contours and Kalman Filter Based Tracking

A Framework for Real-time Left Ventricular tracking in 3D+T Echocardiography, Using Nonlinear Deformable Contours and Kalman Filter Based Tracking 1 A Framework for Real-time Left Ventricular tracking in 3D+T Echocardiography, Using Nonlinear Deformable Contours and Kalman Filter Based Tracking Fredrik Orderud Norwegian University of Science and

More information

Focus on your needs. Ultrasound system HS60 SAMSUNG MEDISON CO., LTD. CT-HS60 V1.0-OB-FT EN

Focus on your needs. Ultrasound system HS60 SAMSUNG MEDISON CO., LTD. CT-HS60 V1.0-OB-FT EN Samsung Medison, an affiliate of Samsung Electronics, is a global medical company founded in 1985. With a mission to bring health and well-being to people's lives, the company manufactures diagnostic ultrasound

More information

ezimagex2 User s Guide Version 1.0

ezimagex2 User s Guide Version 1.0 ezimagex2 User s Guide Version 1.0 Copyright and Trademark Information The products described in this document are copyrighted works of AVEN, Inc. 2015 AVEN, Inc. 4595 Platt Rd Ann Arbor, MI 48108 All

More information

Using Pinnacle 16 Deformable Image registration in a re-treat scenario

Using Pinnacle 16 Deformable Image registration in a re-treat scenario Introduction Using Pinnacle 16 Deformable Image registration in a re-treat scenario This short Hands On exercise will introduce how the Deformable Image Registration (DIR) tools in Pinnacle can be used

More information

DAILY LINAC QA BEAM QA

DAILY LINAC QA BEAM QA BEAM QA DAILY LINAC QA The QA BeamChecker Plus allows for fast, reliable, and uncomplicated daily QA of Varian, Elekta, Siemens, and Accuray Treatment Machines. The QA BeamChecker Plus is specifically

More information

David Wagner, Kaan Divringi, Can Ozcan Ozen Engineering

David Wagner, Kaan Divringi, Can Ozcan Ozen Engineering Internal Forces of the Femur: An Automated Procedure for Applying Boundary Conditions Obtained From Inverse Dynamic Analysis to Finite Element Simulations David Wagner, Kaan Divringi, Can Ozcan Ozen Engineering

More information

M2M GEKKO. State-of-the-art phased-array flaw detector with TFM

M2M GEKKO. State-of-the-art phased-array flaw detector with TFM M2M GEKKO State-of-the-art phased-array flaw detector with TFM USER FRIENDLY FLAW DETECTOR Gekko is the only compact phased-array ultrasound testing (PAUT) instrumentation offering intuitive PA features

More information

Internal Organ Modeling and Human Activity Analysis

Internal Organ Modeling and Human Activity Analysis Internal Organ Modeling and Human Activity Analysis Dimitris Metaxas dnm@cs.rutgers.edu CBIM Center Division of Computer and Information Sciences Rutgers University 3D Motion Reconstruction and Analysis

More information

Section 7.2 Volume: The Disk Method

Section 7.2 Volume: The Disk Method Section 7. Volume: The Disk Method White Board Challenge Find the volume of the following cylinder: No Calculator 6 ft 1 ft V 3 1 108 339.9 ft 3 White Board Challenge Calculate the volume V of the solid

More information

MobileScan3D. MobileScan3D PlugIn for Geomagic. By Steintek GmbH

MobileScan3D. MobileScan3D PlugIn for Geomagic. By Steintek GmbH MobileScan3D PlugIn for Geomagic MobileScan3D V. 1.0 - Version February, 2006 For Geomagic Qualify 8 SR1 and subsequent releases For Geomagic Studio 8 and subsequent releases By Steintek GmbH Win3DS is

More information

User Guide. mk Config

User Guide. mk Config User Guide mk Config mk Config Register 1.1. CD-Start 4 1.2. Installation 5 1.3. Start 6 1.4. Layout of user interface and functions 7 1.4.1. Overview 7 1.4.2. Part buttons 8 1.4.3. Menus 9 1.4.3.1. Export

More information