CALIBRATION PROCEDURES STATE OF THE ART
|
|
- Cecily West
- 5 years ago
- Views:
Transcription
1 CALIBRATION PROCEDURES STATE OF THE ART H. Ossenbrink European Commission, DG Joint Research Centre, Institute for Environment and Sustainability, Renewable Energies Unit, Ispra, Italy ABSTRACT Solar Cell Calibration plays an important role for determining the efficiency of photovoltaic devices under defined conditions. Laboratories involved in PV characterisation developed a number of calibration methods and procedures, which progress towards decreasing uncertainty, and worldwide acceptance. This paper gives an overview about the methods used, their specific features and applicability. The overview is limited to reference devices, and does not discuss standard performance measurements on PV devices. 1. INTRODUCTION The most visible figure-of-merit of photovoltaic solar cells is their efficiency. Since first solar cells flew in space, measurement of efficiency and power delivered under defined conditions was a key to improve solar cells, to understand parameters influencing their performance and to forecast their economic value. Calibration is today understood as the process, to determine the short-circuit current of photovoltaic devices under reference conditions with respect to device temperature, irradiance level and spectral distribution. As it would be economically not feasible, to calibrate all solar cells or modules in production with the required precision and uncertainly, use is made of reference cells, which, once thoroughly calibrated, serve as comparison in a simplified measurement procedure of short circuit current of the device under test. Calibration of reference devices in turn can make use of sophisticated methods, which developed in particular since the use of solar cells in terrestrial applications. A key element in calibration is the reference to standard test conditions (STC), which define a solar cell junction temperature of 25 C, an irradiance exposed to of 1000 W/m 2, and a tabular solar spectral distribution of Air Mass 1.5, which corresponds to a sun elevation of 37 at sea level. Whilst temperature and irradiance can be established either in the laboratory or under natural conditions, or at least corrected for, the reference solar spectral distribution is neither to simulate with sufficient match, nor can it be found routinely under natural sunlight. Almost all calibration methods have their own, characteristic approach to take this problem into account. In contrast to PV solar cells, thermopile detectors, which measure heat generated by the incident light, are spectrally almost unselective, representing practically a black body, and calibration of such devices can be performed with less sophisticated methods. A set of thermopile detectors are also the primary reference for the World Radiometric Reference, established by the World Meteorological Organization, and serve as a traceability standard to determine irradiance in the magnitude of solar irradiance. 2. RELEVANCE In today s terrestrial PV markets, the sales price of PV modules is almost entirely related to the maximum power it can deliver at STC. Competition between manufacturer is largely based on the offered unit price per Watt. The maximum power of the modules leaving the factory is established within the quality control systems in place, and refers either to calibrated reference modules, or to calibrated reference cells placed together with the device under test in the measurement plane of a solar simulator. Assuming the world-wide sales of PV modules for 2003 being 500 MW, a global calibration uncertainty of ±3% for the production module would be the equivalent of over- or underestimating the delivered modules by ±15MW. Assuming average sales price of 4 EUR per Watt, this uncertainty translates to ±60 Mio EUR sales volume. Moreover, as profit margins in the whole commercialisation chain are most probably less than 10%, calibration uncertainty can be the decisive factor on profitability of a PV enterprise. 3. REQUIREMENTS 3.1. Uncertainty The basic requirement, which all calibration methods developed so far strive to minimise, is overall calibration uncertainty. The major components of uncertainty are: 1. Bias error. This is a shift of calibration values caused by factors not corrected for, or by poor experimental design. Typical bias errors are for instance drift or ageing of electronic components or misalignment to the light source. 2. Precision error: This includes all uncertainties of subsystems, such as Voltmeters and Analogue to Digital converters, or temperature controls. 3. Random error: basically this error is can caused by the random nature of atmosphere or noise in the signal conditioning. 4. Repeatability: Almost any experiment, when repeated shows slight variations, which are beyond the pure random error. It can be caused by slight changes in the experimental set-up, by the measurement sequence or procedure, or even by human factors. Repeatability can never be better than the random error.
2 Even though uncertainty can be broken down in the above factors which can be estimated or measured, a control of uncertainty can only established by repeated intercomparison campaigns (Round-Robins) involving as many methods and operators as possible [5], [6], [7] Costs There are not to many calibration laboratories in the world, which routinely are engaged in solar cell calibration. It is in all their interest to contain costs for equipment purchase and set-up as low as possible Duration The time required to execute a complete calibration including data analysis and verification is not only the major factor for customer satisfaction, but the associated labour costs may well limit calibration throughput to very few cells per year, worsening the economic balance of a calibration laboratory quite considerably Availability Ideally, a calibration task shall be executed immediately after receipt of the test devices. Outdoor methods, however, rely on actual weather conditions, which often are variable. Also, outdoor measurements should not lead to another bias error caused by climatic conditions, which are difficult to correct for, such as humidity or excessive atmospheric aerosol content Universal Applicability Solar reference cells may incorporate different photovoltaic technologies, which might interfere with specific features of the calibration method. Typical examples are slow response to fast light-pulses or chopped light, or impossibility to control the junction temperature in case of thin-film reference devices. 4. METHODS AND STANDARDS In the following, the variety of calibration methods will be briefly described, with reference to published standards Thermopile Detectors Thermopile detectors serve to determine solar irradiance regardless of spectral distribution. They have in common that solar irradiance is converted into heat in a black absorber. The temperature rise in relation to a reference temperature is proportional to the incoming irradiation. Temperatures are measured by a number of thermocouples connected in series ( Pile ) in order to increase the thermovoltage measured to practical levels of some millivolts. Thermopile detectors are different by their field of view: Pyranometers have a hemispheric field of view, allowing measurement of solar irradiance form the whole sky, whilst pyrheliometers apply collimation to restrict the field of view to typically 5. The apparent angle of the sun s disc is in comparison Pyrheliometers consequently need to be mounted a tracking platform ensuring alignment with the sun s position. A classification of these instruments is published as ISO Pyrheliometers ISO describes the calibration method for pyrheliometers, to be performed under natural sunlight. Both the test and reference pyrheliometer a mounted on a sun-tracking platform, and the values read. The standard pre-scribes minimum irradiance levels as well as a data analysis procedure. This measurement method and its standard has a very high for the whole chain of calibration, as the reference pyrheliometer can also be one of the primary or secondary references in use world-wide within the World Radiometric Reference. Such primary references are Practical Absolute Cavity Radiometers (PACRAD), which absorb incident light on a ideal blackbody, realised as an precision aperture in front of a cavity, which houses a usually cone-shaped, highly absorptive surface. A second, identical cavity within the same case is shielded from sunlight and electrically heated to track the temperature of the first cavity, which is exposed to sunlight. The electrical energy required for heating of the reference cavity divided by the aperture area of the first cavity establishes the irradiance value. Uncertainty of the method depends to a large extent on the time-constant of the instruments involved and the corresponding reading intervals where a compromise has to be found between precision and atmospheric fluctuations. Figure 1 one shows measured uncertainties of the Joint Research Centre s Cavity Radiometers, established in a number of intercomparisons with the World Radiometric Reference. WRR Factor Initial Calibration World Radiometric Reference (WRR) Factors PMO PMO Initial Calibration Re-Calibration IPC-IX (2000) Year Figure 1. Uncertainty and calibration factors relative to the World Radiometric Reference in a 10-yr period of the two JRC Reference Cavity Radiometers. Uncertainties ar less ±0.2% Pyranometers Calibration of Pyranometers with their hemispheric field of view is referenced to pyrheliometers and requires a set-up, which accounts for the diffuse irradiance absorbed. All test methods apply for this purpose a shading device aligned such that no direct irradiance from the sun can fall onto the pyranometers thermopile surface. The shading device is constructed such that its geometry
3 (diameter and distance to the pyranometer) matches the receiving geometry of the pyrheliometer. After reading the diffuse irradiation value of the pyranometer under shaded condition, the shade is removed and a reading of the total irradiance is taken. The difference between total and diffuse irradiance corresponds to the direct irradiance as determined by the pyrheliometer and established the calibration value of the pyranometer. Calibration uncertainty is much influenced by the design of the shading device, and the linearity of the pyranometer under test. ISO describes the calibration methods, taking into account to set-ups: one where the pyranometer under test is mounted co-planar with the pyrheliometer, and a second where the pyranometer is mounted fixed horizontally, and the shadowing device follows the sun. The first method is mechanically simpler, but may need corrections for ground albedo effects. There are a number of methods describing transfer of calibration between reference and field pyranometers, such as in ISO an ASTM Silicon Solar Cells Different from thermopiles, solar cells are spectrally sensitive. Photons with wavelengths greater than 1050 nm are not absorbed at all in silicon cells, and for wavelength below this band-gap limit the quantum efficiency varies. Consequently, even if solar irradiance is constant as measured by a thermopile detector, which integrates over all wavelengths, the short circuit current of a solar cell may vary with the relative spectral distribution. The spectral distribution varies with atmospheric conditions, and even if the standard sun elevation of 37 defining Airmass 1.5 can be found at least twice a day, the variability of the diffuse and direct component of sunlight may bias the measurement results. However, simulated sunlight is even more difficult to match with the standard Air Mass 1.5 spectral distribution, and requires in any case thorough correction of spectral mismatch Simulated Sunlight This method requires a spectroradiometer to determine the spectral distribution of the light source during the calibration. The short-circuit current of the reference solar cell is measured under the simulator s light together with the spectral irradiance. Together with the spectral responsivity data of the reference cell measured along the method described under 4.2.3, a spectral mismatch calculation can be performed to correct for the differences between the simulator and the tabulated standard spectral spectral distribution. The absolute irradiance during calibration can be derived either from the absolute irradiance calibration for each wavelength of the spectroradiometer, or from an intermediate transfer thermopile, limited by filters to the wavelength range of the spectroradiometer. The advantage of the method is that it can be applied for any solar cell technology, in particular when a steadystate simulator is used. Major uncertainty source is the temperature control of the solar cell, and the calibration uncertainty of the spectroradiometer. For a full description of the method, refer to ISO/DIS Natural Sunlight Direct Sunlight Method. This method follows in principle the calibration of pyrheliometers, described in above, with the test pyrheliometer replaced by the reference solar cell. The solar cell is mounted with an aperture close matching the geometry of the reference pyrheliometer. In addition, spectral irradiance of the direct sunlight is measured, in order to perform a spectral mismatch correction, which requires the spectral responsivity of the solar cell. If meteorological conditions are good and repeatable, the mismatch factor can be small (<2%) and most of all, a constant for different measurement campaigns. In this case, the spectroradiometer can be omitted, or replaced by the measurement of selective wavelengths, which allow a model to compute the actual spectral irradiance. One of the principal disadvantages of this method is on one hand the limited field of view of the reference cell which might not correspond to the later use, the other being the fact that the spectral irradiance of the direct sun light is different from the tabulated standard IEC Furthermore, practical considerations limit this method to reference solar cells smaller than about 25 cm 2, as the collimator system required for larger cells becomes rather unhandy. In taking readings one needs to consider the settling time constants for the reference pyrheliometer are at least three orders of magnitude slower than of a solar cell Global Sunlight Method. This method is an adoption of the above described pyranometer calibration method The silicon reference cell is mounted on a sun-tracking device together with a pyranometer and pyrheliometer. The pyranometer is shaded such to measure only the diffuse component of sunlight, the direct one is given by the pyrheliometer readings. The method avoids the principal disadvantage of the direct sunlight method, and might benefit from the closer match of the actual spectral distribution with the standard reference spectrum. If atmospherical conditions are carefully selected, one can achieve high repeatability even without the measurement of the spectral irradiance. Fig.2. Set-Up for calibration comparison of the global and direct calibration method. Mounted are: Collimator for reference cell, uncollimated reference cell, pyranometer as control, shadowed pyranometer for diffuse irradiance, two pyrheliometers and two cavity radiometers (from left).
4 Spectral Responsivity Calibrations Absolute Responsivity. Spectral responsivity calibrations are based on the fact, that the calibration value of a solar cell can be computed for any reference solar spectral distribution, if the absolute responsivity for each wavelength interval is known. All methods for spectral responsivity are based on short-circuit current measured of solar cells exposed to near monochromatic light of known wavelength. This can be achieved by applying band-pass filters of a monochromator between the light source and the solar cell. However, as irradiance of filtered light is low, it needs to be insured that the solar cell is linear in its short circuit, as the number of photons converted per unit of time and surface area is up to three orders of magnitude smaller than at exposure to natural sunlight. The application of additional white bias light up to 1000 W/m 2 can resolve this problem, but requires then in addition temperature control of the cell to be calibrated. The absolute, monochromatic irradiance can be determined either when both the light source level and the absorptivity of the monochromator or filters are known, or by switching the cell under test with a reference detector calibrated for selected wavelengths. This method is described as Differential Spectral Responsivity Method in ISO/DIS Relative Responsivity. If the uncertainty of the absolute responsivity from the previous described methods is insufficient, the measurements can be verified by one or more of the previous natural sunlight or solar simulator methods, as in the spectral mismatch calculation only relative responsivity is required. 5. TRACEBILITY ISSUES According to international agreements, all calibration value of commercial significance must refer to SI-Units. In particular, calibration laboratories accredited according to ISO need to maintain an unbroken chain of traceability. IEC refers to the World Radiometric Reference as the primary standard, as the SI radiometric scale is demonstrated only for irradiance levels far below natural sunlight. Comparison [4] have shown meanwhile the equivalence of both radiometric references. An international standard is in preparation, IEC PWI, which shall define procedures for establishing the traceability of the calibration of photovoltaic reference devices. It will define requirements for calibration procedures used in maintenance of the traceability chain, such as uncertainty, and list some of methods in use. 6. UNCERTAINTY LIMITS All calibration procedures need to establish and regularly confirm their uncertainty limits either through verification measurements or Round-Robin intercomparisons.. References [1], [2] and [3] are examples of such uncertainty estimates. From intercomparisons held in recent years, uncertainties between ±1.0% and ±2.0% are realistic and worldwide achievable. A current Round-Robin focuses on the establishment of a World Photovoltaic Scale (WPVS), which would consist in a set of standardised reference cells compared repeatingly, which shall serve the calibration laboratories as a primary laboratory reference [8]. 7. CONCLUSIONS Calibration methods in use today have evolved to better uncertainties, which are consistently near ±1.5%. The different methods, when carefully applied, have shown in various intercomparisons that the calibration factors achieved are in good agreements. The establishment of the World Photovoltaic Scale, together with a new international standard on traceability of calibrations to SI units will underpin fair and transparent commercialisation of photovoltaic technology on all the world markets. 8. ACKNOWLEDGENTS The author wishes to thank W. Zaaiman of JRC, Italy for much of the material referenced in this publications, and K. Emery of NREL, USA for many fruitful discussions. 9. REFERENCES 9.1. Selected Standards Publications Solar Spectra and Light Simulation IEC Measurement principles for terrestrial photovoltaic (PV) devices with reference spectral irradiance data. IEC Solar Simulator requirements. ISO Reference solar spectral irradiance at ground at different receiving conditions Part1: Direct normal and hemispherical solar irradiance for Air Mass 1.5., ASTM 891 Tables for Terrestrial Direct Normal Solar Spectral Irradiance for Air Mass 1.5 ASTM 892 Tables for Terrestrial Solar Spectral Irradiance for Air Mass 1.5 for a 37 Tilted Surface. ASTM 927 Specification for Solar Simulation for Terrestrial Photovoltaic Testing Calibrations (Thermopiles) ISO 9059 Calibration of field pyrheliometers by comparison to a reference pyrheliometer, 1990 ISO 9060 Specification and classification of instruments for measuring hemispherical solar and direct radiation, ISO 9846 Solar Energy - Calibration of a pyranometer using a pyrheliometer. ISO 9847 Solar Energy - Calibration of field pyranometers by comparison to a reference pyranometer. ASTM 816 Method for Calibration of Secondary Reference Pyrheliometers and Pyrheliometers for Field Use
5 ASTM 824 ASTM 913 ASTM 941 ASTM 1144 Method for Transfer of Calibration from Reference to Field Pyranometers Method for Calibration of Reference Pyranometers with Axis Vertical by the Shading Method Test Method for Calibration of Reference Pyranometers with Axis Tilted by the Shading Method Test Method for Calibration of Non- Concentrator Terrestrial Photovoltaic Primary Reference Cells under Direct Irradiance Calibrations (Solar Cells)) IEC Requirements for reference solar cells. IEC Procedures for establishing the traceability of the calibration of photovoltaic reference devices. (PWI: Proposed Work Item) IEC Requirements for reference solar modules. IEC Computation of spectral mismatch error introduced in the testing of a photovoltaic device. IEC Guidance for the measurement of spectral response of a photovoltaic (PV) device. ISO/DIS15387 Space systems Single-junction space solar cells Measurement and calibration procedures ASTM 1144 Test Method for Calibration of Non- Concentrator Terrestrial Photovoltaic Primary Reference Cells under Direct Irradiance ASTM 1039 Method for Calibration and Characterisation of Non-Concentrator Terrestrial Photovoltaic Reference Cells under Global Irradiation ASTM 1040 Specification for Physical Characteristics of Non-Concentrator Terrestrial Photovoltaic Reference Cells ASTM 1125 Test Method for Calibration of Primary Non-Concentrator Terrestrial Photovoltaic Reference Cells using a Tabular Spectrum ASTM 1362 Test Method for Calibration of Non- Concentrator Photovoltaic Secondary Reference Cells Electrical Performance Measurements IEC Procedure for temperature and irradiance corrections to measured I-V characteristics of crystalline silicon photovoltaic IEC IEC ASTM 973 ASTM Literature Measurement of photovoltaic currentvoltage characteristics. Determination of the equivalent cell temperature (ECT) of photovoltaic (PV) devices by the open-circuit voltage ASTM 1021 Methods for Measuring the Spectral Response of Photovoltaic Cells Test Method for Determination of the Spectral Mismatch Parameter between a Photovoltaic Device and a Photovoltaic Reference Cell Test Method for Determination the Linearity of a Photovoltaic Device with respect to the Test Parameter [1] K. Whitfield and C.R. Osterwald, Procedure for the determination of Photovoltaic Module Outdoor Electrical Performance, Prog. Photovolt: Res. Appl.2001; 9: [2] J.S. Henzing and W.H. Knap, Uncertainty in pyranometer and pyrheliometer measurements at KNMI in De Bilt, TR-235, ISBN: , [3] W. Zaaiman and A. Realini, Photovoltaic Device Calibration Intercomparison Test Report,TNI.98.18, European Commission, Joint Research Centre [4] J. Romero, N.P.Fox and C. Froehlich: Improved comparison of the World Radiometric Reference and the SI radiometric scale. Metrologia, 1995/96, 32, [5] The results of the 1984/1985 Round Robin calibration of reference solar cells for the Summit Working Group on Technology, Growth and Employment, EUR, European Commission, Joint Research Centre 0613, [6] The results of the PEP 87 Round Robin Calibration of Reference Solar Cells and Modules, PTB Bericht, PTB-Opt-31, ISBN , [7] The results of the PEP 93 Intercomparison of Reference Cell Calibrations and Newer Technology Performance Measurements, NREL/TP , March [8] The results of the First World Photovoltaic Scale Recalibration, NREL/TP , March 2000.
Removing Systematic Errors from Rotating Shadowband Pyranometer Data
Removing Systematic Errors from Rotating Shadowband Pyranometer Data Frank Vignola Solar Radiation Monitoring Laboratory Department of Physics University of Oregon Eugene, OR 9743-274 fev@uoregon.edu ABSTRACT
More informationRecommended Practices for PV Module Outdoor Characterization and Power Rating
Recommended Practices for PV Module Outdoor Characterization and Power Rating Gabi Friesen Scuola Universitaria Professionale della Svizzera Italiana Lugano / Canobbio, Switzerland Christian Reise Fraunhofer
More informationModule Shading Guide. Field Shading Guide and Avoiding Device Damage
Module Shading Guide The purpose of this document is to provide an understanding of the shading impact on First Solar Series 3 and Series 4 PV modules (hereafter referred to as FS Series PV Modules ) as
More informationCalibration Procedures for LI-COR Spectroradiometers, Radiation Sensors & Lamps
Application Note #109 Calibration Procedures for LI-COR Spectroradiometers, Radiation Sensors & Lamps Introduction This report summarizes LI-COR factory calibration procedures for all quantum sensors,
More informationCHARACTERIZATION OF BIFACIAL SILICON SOLAR CELLS AND MODULES: A NEW STEP
CHARACTERIZATION OF BIFACIAL SILICON SOLAR CELLS AND MODULES: A NEW STEP C. Duran, P. Hering, T. Buck, K. Peter International Solar Energy Research Center, ISC Konstanz ev., Rudolf Diesel Str 15, D-78467,
More informationPERFORMANCE COMPARISON OF FIXED, SINGLE, AND DUAL AXIS TRACKING SYSTEMS FOR SMALL PHOTOVOLTAIC SYSTEMS WITH MEASURED DIRECT BEAM FRACTION
PERFORMANCE COMPARISON OF FIXED, SINGLE, AND DUAL AXIS TRACKING SYSTEMS FOR SMALL PHOTOVOLTAIC SYSTEMS WITH MEASURED DIRECT BEAM FRACTION Paula E. Logan Department of Technology and Environmental Design
More informationBEAM TILTED CORRELATIONS. Frank Vignola Department of Physics University of Oregon Eugene, OR
BEAM TILTED CORRELATIONS Frank Vignola Department of Physics University of Oregon Eugene, OR 9743-1274 fev@uoregon.edu ABSTRACT A model is described and evaluated that calculates beam irradiance from tilted
More informationMACHINE VISION AS A METHOD FOR CHARACTERIZING SOLAR TRACKER PERFORMANCE
MACHINE VISION AS A METHOD FOR CHARACTERIZING SOLAR TRACKER PERFORMANCE M. Davis, J. Lawler, J. Coyle, A. Reich, T. Williams GreenMountain Engineering, LLC ABSTRACT This paper describes an approach to
More informationSRI LANKA STANDARD CODE OF PRACTICE FOR GRID CONNECTED PHOTOVOLTAIC POWER SYSTEMS SLS 1522:2016
SRI LANKA STANDARD CODE OF PRACTICE FOR GRID CONNECTED PHOTOVOLTAIC POWER SYSTEMS SLS 1522:2016 BY H S W Karunaratne Assistant Director - Engineering Sri Lanka Standards Institution. 1 DEFINITIONS Ac module:
More informationMeasurement and Standards: The Role of NIST
Measurement and Standards: The Role of NIST Dr. Kent Rochford, Acting Director Electrical and Electronics Engineering Laboratory (EEEL) National Institute of Standards and Technology Department of Commerce
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. CFV SOLAR TEST LABORATORY, INC Ste. A University Blvd. SE Albuquerque, NM 87106
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 CFV SOLAR TEST LABORATORY, INC. 5600 Ste. A University Blvd. SE Albuquerque, NM 87106 Jim Crimmins (Authorized Representative) Phone: 505 998 0102 Email: Jim.Crimmins@cfvsolar.com
More informationUser Interface: Simple Model of Atmospheric Radiative Transfer of Sunshine
User Interface: Simple Model of Atmospheric Radiative Transfer of Sunshine April 2003 WINDOWS Version Introduction This document provides instructions for installing and running the User Interface for
More informationTHE DESIGN AND TESTING OF A LOW-MATERIAL-COST PARABOLIC-TROUGH PV CONCENTRATOR
THE DESIGN AND TESTING OF A LOW-MATERIAL-COST PARABOLIC-TROUGH PV CONCENTRATOR Clive K.Weatherby Dept. of Cybernetics, University of Reading, Whiteknights, PO Box 225, Reading, Berkshire, RG6 6AY, UK.
More informationMeterless Laser Power/Energy Measurement Simplifies Embedding
White Paper Meterless Laser Power/Energy Measurement Simplifies Embedding Traditional laser power and energy measurement instruments typically comprise a sensor head connected to separate meter electronics.
More informationFRAUNHOFER INSTITUTE FOR SOLAR ENERGY SYSTEMS ISE
FRAUNHOFER INSTITUTE FOR SOLAR ENERGY SYSTEMS ISE Mirror qualification for concentrating solar collectors Anna Heimsath Head of Team Concentrating Collectors Fraunhofer Institute for Solar Energy Systems
More informationIRRADIANCE DISTRIBUTION OF IMAGE SURFACE IN MICROLENS ARRAY SOLAR CONCENTRATOR
IRRADIANCE DISTRIBUTION OF IMAGE SURFACE IN MICROLENS ARRAY SOLAR CONCENTRATOR Ali H. Al-Hamdani 1, Hayfa G. Rashid 2 and Alaa B. Hasan 3 1 Department of Laser and Optoelectronics Engineering, University
More informationPhilpot & Philipson: Remote Sensing Fundamentals Interactions 3.1 W.D. Philpot, Cornell University, Fall 12
Philpot & Philipson: Remote Sensing Fundamentals Interactions 3.1 W.D. Philpot, Cornell University, Fall 1 3. EM INTERACTIONS WITH MATERIALS In order for an object to be sensed, the object must reflect,
More informationA SIMPLE METHOD OF DETERMINING THE INFLUENCE OF THE OVERHANG ON WINDOW SOLAR GAINS
Eleventh International IBPSA Conference Glasgow, Scotland July 27-30, 2009 A SIMPLE METHOD OF DETERMINING THE INFLUENCE OF THE OVERHANG ON WINDOW SOLAR GAINS Dominik Włodarczyk 1, Henryk Nowak 1 1 Division
More informationInstruction Manual for Modelling of Trap Detectors. Helsinki University of Technology. Metrology Research Institute. Instruction Manual for
Page 1 (10) Helsinki University of Technology Metrology Research Institute Instruction Manual for Modelling Version: 2.2 Date of Issue: December 21, 2005 Page 2 (10) Table of Contents 1. INTRODUCTION 3
More informationEvaluation of Satellite Ocean Color Data Using SIMBADA Radiometers
Evaluation of Satellite Ocean Color Data Using SIMBADA Radiometers Robert Frouin Scripps Institution of Oceanography, la Jolla, California OCR-VC Workshop, 21 October 2010, Ispra, Italy The SIMBADA Project
More informationSHAILRON Technology SOLAR MONITORING SYSTEM
SHAILRON Technology SOLAR MONITORING SYSTEM In the following sections the individual system components are further described. EKO s solar monitoring system is configured with high accurate sensors and
More informationInteractive comment on Quantification and mitigation of the impact of scene inhomogeneity on Sentinel-4 UVN UV-VIS retrievals by S. Noël et al.
Atmos. Meas. Tech. Discuss., www.atmos-meas-tech-discuss.net/5/c741/2012/ Author(s) 2012. This work is distributed under the Creative Commons Attribute 3.0 License. Atmospheric Measurement Techniques Discussions
More informationInteractive comment on Quantification and mitigation of the impact of scene inhomogeneity on Sentinel-4 UVN UV-VIS retrievals by S. Noël et al.
Atmos. Meas. Tech. Discuss., 5, C741 C750, 2012 www.atmos-meas-tech-discuss.net/5/c741/2012/ Author(s) 2012. This work is distributed under the Creative Commons Attribute 3.0 License. Atmospheric Measurement
More informationDesign of Hexagonal Micro Lenses Array Solar Concentrator
ISSN: 235-328 Design of Hexagonal Micro Lenses Array Solar Concentrator Alaa Bader Hassan, Sabah Ali Hussein Department of Physics, College of Education Ibn Al-Haitham for Pure Sciences, University of
More informationSPN1 Measurement Best Practices Solar Irradiance Measurements with Delta T Devices SPN1 Sunshine Pyranometer
Solar Irradiance Measurements with Delta T Devices SPN1 Sunshine Pyranometer by John Wood SPN1 Best Practice version 3 Contents 1. Introduction... 3 2. Principle of operation... 3 3. Measurement site and
More informationWhy Modules bought anonymously on the Market should be tested in Addition to IEC Certification?
Why Modules bought anonymously on the Market should be tested in Addition to IEC Certification? 5th SMET Advanced PV Technology Conference München, 27-29.05.2009 Dipl.-Ing. Willi Vaaßen TÜV Immissionsschutz
More informationInvestigation of an Alternative Method for Measuring the Spectral Irradiance of a High-Powered Focused Light Source
Investigation of an Alternative Method for Measuring the Spectral Irradiance of a High-Powered Focused Light Source Speaker / Author: I. Kruger * Co-author: R. Sieberhagen ** National Metrology Institute
More informationMission Solar Energy Product Models
Cover Page RETC PAN File Test Report December 16, 2016 Mission Solar Energy Product Models MSE265SO5T, MSE270SO5T, MSE275SO5T Issued by: Issued to: Trevor Charles Udwin Product Manager Renewable Energy
More informationInstruction Manual of Modelling of Trap Detectors Using Matlab
Aalto University School of Electrical Engineering Metrology Research Institute Mikko Puranen Tuomas Poikonen Timo Dönsberg Petri Kärhä Instruction Manual of Modelling of Trap Detectors Using Matlab Version
More informationPV Module Installation Instructions
Name 1601A Page 1 / 8 Name 1601A Page 2 / 8 1 Introduction Thanks for purchasing PV modules of HT SOLAR ENERGY JOINT STOCK COMPANY This manual refers to PV modules manufactured and sold by HT Turkey. This
More informationPOLYHEDRAL SPECULAR REFLECTOR
32 C h a p t e r 3 POLYHEDRAL SPECULAR REFLECTOR The goal of the Full Spectrum Photovoltaics Project was to design and prototype a 50% module efficiency photovoltaic system. Of the three designs we initially
More informationGEOG 4110/5100 Advanced Remote Sensing Lecture 2
GEOG 4110/5100 Advanced Remote Sensing Lecture 2 Data Quality Radiometric Distortion Radiometric Error Correction Relevant reading: Richards, sections 2.1 2.8; 2.10.1 2.10.3 Data Quality/Resolution Spatial
More informationField data requirements for the validation of PV module performance models
Field data requirements for the validation of PV module performance models Dipl phys. Gabi Friesen 4th PV Performance Modelling and Monitoring Workshop OUTLINE Introduction Test issues and requirements
More informationAnalysis Ready Data For Land (CARD4L-ST)
Analysis Ready Data For Land Product Family Specification Surface Temperature (CARD4L-ST) Document status For Adoption as: Product Family Specification, Surface Temperature This Specification should next
More informationBird Solar Model Source Creator
Bird Solar Model Source Creator INTRODUCTION This knowledge base article describes a script that generates a FRED source that models the properties of solar light incident on a tilted or solar-tracking
More informationMission Solar Energy Product Models
Cover Page RETC PAN File Test Report October 27, 2016 Mission Solar Energy Product Models MSE280SQ5T, MSE285SQ5T, MSE290SQ5T, MSE295SQ5T Issued by: Issued to: Trevor Charles Udwin Product Manager Renewable
More informationTwo slit interference - Prelab questions
Two slit interference - Prelab questions 1. Show that the intensity distribution given in equation 3 leads to bright and dark fringes at y = mλd/a and y = (m + 1/2) λd/a respectively, where m is an integer.
More information10.4 Interference in Thin Films
0. Interference in Thin Films You have probably noticed the swirling colours of the spectrum that result when gasoline or oil is spilled on water. And you have also seen the colours of the spectrum shining
More informationGlare Spread Function (GSF) - 12 Source Angle
Normalized Pixel Value POWERED BY OPTEST SOFTWARE Stray Light Measurement on LensCheck Lens Measurement Systems 1 Glare Spread Function (GSF) - 12 Source Angle 0.1 0.01 0.001 0.0001 0.00001 0.000001 1
More informationA Comparative Performance Study of Diffuse Fraction Models Based on Data from Vienna, Austria
Proceedings of the 2 nd ICAUD International Conference in Architecture and Urban Design Epoka University, Tirana, Albania, 08-10 May 2014 Paper No. 200 A Comparative Performance Study of Diffuse Fraction
More informationModel of Photovoltaic Generation During Cloudy Days
Model of Photovoltaic Generation During Cloudy Days Dionysios C. Aliprantis, Chengrui Cai Iowa State University Sandia National Laboratory June 11, 2012 Outline Motivation Literature review on cloud pattern
More informationENHANCEMENT OF DIFFUSERS BRDF ACCURACY
ENHANCEMENT OF DIFFUSERS BRDF ACCURACY Grégory Bazalgette Courrèges-Lacoste (1), Hedser van Brug (1) and Gerard Otter (1) (1) TNO Science and Industry, Opto-Mechanical Instrumentation Space, P.O.Box 155,
More informationBifacial PV Performance Models: Comparison and Field Results
Bifacial PV Performance Models: Comparison and Field Results Chris Deline, B. Marion, S. MacAlpine (NREL) J. Stein (SNL), F. Toor (U. Iowa), S. Ayala (U. Ariz.) BiFiPV 2017 Workshop Konstanz Germany October
More informationinter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE
Copyright SFA - InterNoise 2000 1 inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering 27-30 August 2000, Nice, FRANCE I-INCE Classification: 3.1 FACTORS AFFECTING
More information1.3 Customized Solutions (OEM)
1.3 Customized Solutions (OEM) Ophir The World Leading Source for Custom Designed Laser Measurement Solutions 1.3.1 Introduction Many laser systems manufacturers need to have a measuring capability built
More informationTechniques to Reduce Measurement Errors in Challenging Environments
Techniques to Reduce Measurement Errors in Challenging Environments Jon Semancik VTI Instruments Corp. 2031 Main Street Irvine, CA 92614 949-955-1894 jsemancik@vtiinstruments.com INTRODUCTION Test engineers
More informationSL 200 instrument can process the energy intensity emitted by the solar radiation in a precise place of the earth.
Solarimeter SL 200 Table of contents 1 Introduction...4 2 General information...4 2.1 Measured units...4 2.2 Use...4 3 Operating principle...5 3.1 Keyboard presentation...5 3.2 Instrument offers 3 groups
More informationspecular diffuse reflection.
Lesson 8 Light and Optics The Nature of Light Properties of Light: Reflection Refraction Interference Diffraction Polarization Dispersion and Prisms Total Internal Reflection Huygens s Principle The Nature
More informationComprehensive characterisation and analysis of PV module performance under real operating conditions
PROGRESS IN PHOTOVOLTAICS: RESEARCH AND APPLICATIONS Prog. Photovolt: Res. Appl. 2017; 25:218 232 Published online 8 December 2016 in Wiley Online Library (wileyonlinelibrary.com)..2848 RESEARCH ARTICLE
More informationJune 19, 2013 PVMC SDO
Introduction to International QA Task Force John Wohlgemuth & Sarah Kurtz June 19, 2013 PVMC SDO NREL is a national laboratory of the U.S. Department of Energy, Office of Energy Efficiency and Renewable
More informationAnalysis Ready Data For Land
Analysis Ready Data For Land Product Family Specification Optical Surface Reflectance (CARD4L-OSR) Document status For Adoption as: Product Family Specification, Surface Reflectance, Working Draft (2017)
More informationOptics Vac Work MT 2008
Optics Vac Work MT 2008 1. Explain what is meant by the Fraunhofer condition for diffraction. [4] An aperture lies in the plane z = 0 and has amplitude transmission function T(y) independent of x. It is
More informationMu lt i s p e c t r a l
Viewing Angle Analyser Revolutionary system for full spectral and polarization measurement in the entire viewing angle EZContrastMS80 & EZContrastMS88 ADVANCED LIGHT ANALYSIS by Field iris Fourier plane
More informationSolar Panel Irradiation Exposure efficiency of solar panels with shadow
Solar Panel Irradiation Exposure efficiency of solar panels with shadow Frits F.M. de Mul MEDPHYS Software & Services 2012 www.medphys.nl email: info(at)medphys.nl Solar Panel Irradiation 1. Local Times,
More informationPhoto Emission Tech., Inc. 760 Calle Plano, Camarillo, CA USA Tel: (805) Fax: (805)
IQE1800 SYSTEM INTRODUCTION Quantum Efficiency (QE) measurements provide a primary characterization method for completed photovoltaic devices. QE system provides electronics and software designed for fully
More informationEvaluation of radiative power loading on WEST metallic in-vessel components
Evaluation of radiative power loading on WEST metallic in-vessel components M-H. Aumeunier 1, P. Moreau, J. Bucalossi, M. Firdaouss CEA/IRFM F-13108 Saint-Paul-Lez-Durance, France E-mail: marie-helene.aumeunier@cea.fr
More informationBifacial PV cell with reflector for stand-alone mast for sensor powering purposes
Downloaded from orbit.dtu.dk on: Mar 19, 219 Bifacial PV cell with reflector for stand-alone mast for sensor powering purposes Jakobsen, Michael Linde; Thorsteinsson, Sune; Poulsen, Peter Behrensdorff;
More informationWeather station AP 257/51
Gamma instabus Weather station GPS AP 257/61 Weather station AP 257/51 Technical product information 5WG1 257-3AB61 5WG1 257-3AB51 Weather station with sensors to survey temperature, brightness, rain and
More informationCalibration Techniques for NASA s Remote Sensing Ocean Color Sensors
Calibration Techniques for NASA s Remote Sensing Ocean Color Sensors Gerhard Meister, Gene Eplee, Bryan Franz, Sean Bailey, Chuck McClain NASA Code 614.2 Ocean Biology Processing Group October 21st, 2010
More informationIMPROVING THE ACCURACY OF LOW COST MEASUREMENT OF DIRECT NORMAL SOLAR IRRADIANCE
IMPROVING THE ACCURACY OF LOW COST MEASUREMENT OF DIRECT NORMAL SOLAR IRRADIANCE Jim Augustyn and Taylor Geer Augustyn + Company P.O. Box 7075 Berkeley CA 94707 Ed Kern and Ruel Little Schott Applied Power
More informationApplication basics/handbook
Application basics/handbook Inclination sensor / Tiltmeter GKAS2000 series with Modbus output a Dear Customer, thank you for purchasing this product. These Instruments are a results of a cooperation between
More informationCalibration of the depth-measuring device of Rockwell hardness testing machines
Calibration of the depth-measuring device of Rockwell hardness testing machines F. Löffler, A. Sawla, P. Strobel* Physikalisch-Technische Bundesanstalt, Braunschweig * Heinrich Bareiss Prüfgerätebau GmbH,
More informationDr. Larry J. Paxton Johns Hopkins University Applied Physics Laboratory Laurel, MD (301) (301) fax
Dr. Larry J. Paxton Johns Hopkins University Applied Physics Laboratory Laurel, MD 20723 (301) 953-6871 (301) 953-6670 fax Understand the instrument. Be able to convert measured counts/pixel on-orbit into
More informationImproving Traceability of Fluorescence Calibrations to Practical Colorimetric Applications
Improving Traceability of Fluorescence Calibrations to Practical Colorimetric Applications 9 th Biannual Joint US/CIE and CNC/CIE Technical Day 7 November 2013 Joanne Zwinkels, William Neil and Mario Noël
More informationBroadband and Wide Angle Antireflection Coatings for Solar Cell Applications Dr. Mohammed A. Hussein, Dr. Ali H. Al-Hamdani, Nibras S.
Broadband and Wide Angle Antireflection Coatings for Solar Cell Applications Dr. Mohammed A. Hussein University of Technology Dr. Ali H. Al-Hamdani Energy and Renewable Energy Technology Center/ University
More informationD&S Technical Note 09-2 D&S A Proposed Correction to Reflectance Measurements of Profiled Surfaces. Introduction
Devices & Services Company 10290 Monroe Drive, Suite 202 - Dallas, Texas 75229 USA - Tel. 214-902-8337 - Fax 214-902-8303 Web: www.devicesandservices.com Email: sales@devicesandservices.com D&S Technical
More information2.626 Fundamentals of Photovoltaics
MIT OpenCourseWare http://ocw.mit.edu 2.626 Fundamentals of Photovoltaics Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms. Efficiency Limits
More informationQuokka version 2: selective surface doping, luminescence. modeling and data fitting
Quokka version 2: selective surface doping, luminescence modeling and data fitting Andreas Fell 1 *, Keith R. McIntosh 2, Malcolm Abbott 2, Daniel Walter 1 1 Australian National University, Canberra, Australia
More informationPowered by nature. Certification and testing services for the photovoltaic sector. TÜV SÜD Product Service GmbH
Powered by nature Certification and testing services for the photovoltaic sector TÜV SÜD Product Service GmbH TÜV SÜD your trusted photovoltaic partner Solar energy is increasingly emerging as one of the
More informationMobile: web:
IMPROVING AND UNDERSTANDING kwh/kwp SIMULATIONS Steve Ransome 1 and Juergen Sutterlueti 2 1 Steve Ransome Consulting Limited (SRCL) Mobile: +44 7515 565010 mailto:steve@steveransome.com web: www.steveransome.com
More informationApplication Note - How to Simulate a SolarEdge System in PVsyst
Version 8, November 2017 Application Note - How to Simulate a SolarEdge System in PVsyst Version History Version 8 for PVsyst version 6.64 Version 7 for PVsyst version 5.53 PVsyst supports the design and
More informationVicarious Radiometric Calibration of MOMS at La Crau Test Site and Intercalibration with SPOT
Vicarious Radiometric Calibration of MOMS at La Crau Test Site and Intercalibration with SPOT M. Schroeder, R. Müller, P. Reinartz German Aerospace Center, DLR Institute of Optoelectronics, Optical Remote
More informationNRG INSTRUCTIONS Using the SPN1 Pyranometer on Symphonie Loggers
INTRODUCTION This application note provides information for using the Delta T SPN1 pyranometer with NRG Symphonie loggers. The Delta-T SPN1 Thermopile based pyranometer provides signals for Global Horizontal
More informationFluxGage. FluxGage. LED Luminaire Measurement System User Manual
FluxGage FluxGage LED Luminaire Measurement System User Manual 1 Acronyms... 3 2 Introduction... 4 2.1 Operation principle... 4 3 Specifications... 5 4 Mechanical and Electrical Installation... 9 4.1 Unpacking...
More informationPowered by nature. Certification and testing services for the photovoltaic sector. TÜV SÜD Product Service GmbH
Powered by nature Certification and testing services for the photovoltaic sector TÜV SÜD Product Service GmbH TÜV SÜD your trusted photovoltaic partner Solar energy is increasingly emerging as one of the
More informationTextbook Reference: Physics (Wilson, Buffa, Lou): Chapter 24
AP Physics-B Physical Optics Introduction: We have seen that the reflection and refraction of light can be understood in terms of both rays and wave fronts of light. Light rays are quite compatible with
More informationPolarization Ray Trace, Radiometric Analysis, and Calibration
Polarization Ray Trace, Radiometric Analysis, and Calibration Jesper Schou Instrument Scientist Stanford University jschou@solar.stanford.edu 650-725-9826 HMI00905 Page 1 Polarization Ray Trace Purpose
More informationTracePro Stray Light Simulation
TracePro Stray Light Simulation What Is Stray Light? A more descriptive term for stray light is unwanted light. In an optical imaging system, stray light is caused by light from a bright source shining
More informationILLUMINATION HOMOGENEITY OF BIFACIAL SYSTEMS OUTDOOR MEASUREMENTS WITH SYSTEMATICALLY VARIED INSTALLATION CONDITIONS
ILLUMINATION HOMOGENEITY OF BIFACIAL SYSTEMS OUTDOOR MEASUREMENTS WITH SYSTEMATICALLY VARIED INSTALLATION CONDITIONS Thomas Baumann*, Markus Klenk, Nico Keller, Hartmut Nussbaumer, Franz Baumgartner ZHAW,
More informationLecture 24. Lidar Simulation
Lecture 24. Lidar Simulation q Introduction q Lidar Modeling via Lidar Simulation & Error Analysis q Functions of Lidar Simulation and Error Analysis q How to Build up Lidar Simulation? q Range-resolved
More informationAdvanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies
ECNDT 2006 - Th.3.4.2 Advanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies M. SIMON, I. TISEANU, C. SAUERWEIN Hans WÄLISCHMILLER, Meersburg, Germany M. SINDEL, M. BRODMANN,
More informationDevice for utilisation of solar energy. Technical field
1 Device for utilisation of solar energy Technical field The invention relates to a device for utilisation of solar energy, especially for its conversion into heat or electric energy, comprising an absorber
More informationUnderstanding MFC Metrology & Calibration
Understanding MFC Metrology & Calibration Factors Impacting MFC Performance in Biopharmaceutical Process Systems 1 Understanding MFC Metrology & Calibration Mass flow controllers (MFCs) precisely deliver
More informationCPV STANDARDS: UPDATE OF THE CURRENT TESTING STANDARDS AND FUTURE DEVELOPMENTS
CPV STANDARDS: UPDATE OF THE CURRENT TESTING STANDARDS AND FUTURE DEVELOPMENTS Paul F. Robusto, PhD Intertek Testing and Certification, Inc. 25791 Commercenter Drive, Lakeforest, CA 92630 paul.robusto@intertek.com
More informationSingle Photon Counting Module
Description Laser Components COUNT series of s has been developed to offer a unique combination of high quantum efficiency, wide dynamic range and ease of use for photon counting applications. Combining
More informationMONTE CARLO SIMULATION OF SOLAR RADIATION THROUGH A WATER-FILLED PRISMATIC LOUVER
Proceedings of the 16th International Heat Transfer Conference, IHTC-16 August 10-15, 2018, Beijing, China IHTC16-21877 MONTE CARLO SIMULATION OF SOLAR RADIATION THROUGH A WATER-FILLED PRISMATIC LOUVER
More informationSOLAR CELL SURFACE INSPECTION USING 3D PROFILOMETRY
SOLAR CELL SURFACE INSPECTION USING 3D PROFILOMETRY Prepared by Benjamin Mell 6 Morgan, Ste16, Irvine CA 92618 P: 949.461.9292 F: 949.461.9232 nanovea.com Today's standard for tomorrow's materials. 21
More informationT-Solar Overview. * Patent-pending
T-Solar T-Solar Overview The T-Solar system combines our best photovoltaic measurement technology into a system designed specifically for measuring textured samples. Based on the established M-2000 rotating
More informationMonitoring and Controlling Energy Efficiency of Photovoltaic Installations
Monitoring and Controlling Energy Efficiency of Photovoltaic Installations Features System Current and voltage supervision of solar panels Control of switching states Sources Shunt and inductive sensors
More informationinter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE
Copyright SFA - InterNoise 2000 1 inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering 27-30 August 2000, Nice, FRANCE I-INCE Classification: 2.4 NORD 2000. STATE-OF-THE-ART
More informationRetroSign GRX-7 1. MEASUREMENT PARAMETERS:
Specification for handheld Retroreflectometer for determining the coefficient of retroreflected luminance (R A ) of road traffic signs, high visibility clothing, license plates and reflective tapes 1.
More informationLecture 1 Image Formation.
Lecture 1 Image Formation peimt@bit.edu.cn 1 Part 3 Color 2 Color v The light coming out of sources or reflected from surfaces has more or less energy at different wavelengths v The visual system responds
More informationWeather station GPS AP 257/61 Weather station AP 257/51
Gamma instabus Weather station GPS AP 257/61 Weather station AP 257/51 Technical product information 5WG1 257-3AB61 5WG1 257-3AB51 Weather station with sensors to survey temperature, brightness, rain and
More informationSingle-Axis Tracked Bifacial System Results
5 th bifipv Workshop, Colorado 2018 Single-Axis Tracked Bifacial System Results Silvana Ayala Pelaez 1, Chris Deline 2, Peter Greenberg 3, Josh Stein 4, Raymond K. Kostuk 1 silvanaa@email.arizona.edu Outline
More informationA New Model for Optical Crosstalk in SinglePhoton Avalanche Diodes Arrays
A New Model for Optical Crosstalk in SinglePhoton Avalanche Diodes Arrays I. Rech, A. Ingargiola, R. Spinelli, S. Marangoni, I. Labanca, M. Ghioni, S. Cova Dipartimento di Elettronica ed Informazione Politecnico
More informationSpectraTec X. Multi- λ Laser Light Source BLUE SKY RESEARCH WAVELENGTHS
Multi- λ Laser Light Source BLUE SKY RESEARCH Up to four individual laser wavelengths One single mode PM fiber output Ultra-stable power output Compact : 165mm x 125mm x 34mm Integrated control and drive
More informationGODDARD SPACE FLIGHT CENTER. Future of cal/val. K. Thome NASA/GSFC
GODDARD SPACE FLIGHT CENTER Future of cal/val K. Thome NASA/GSFC Key issues for cal/val Importance of cal/val continues to increase as models improve and budget pressures go up Better cal/val approaches
More informationRadiance, Irradiance and Reflectance
CEE 6100 Remote Sensing Fundamentals 1 Radiance, Irradiance and Reflectance When making field optical measurements we are generally interested in reflectance, a relative measurement. At a minimum, measurements
More informationLecture 7 Notes: 07 / 11. Reflection and refraction
Lecture 7 Notes: 07 / 11 Reflection and refraction When an electromagnetic wave, such as light, encounters the surface of a medium, some of it is reflected off the surface, while some crosses the boundary
More informationBridging the Gap in Optical Network Testing by Bill Heselden
Bridging the Gap in Optical Network Testing by Bill Heselden TABLE OF CONTENTS: 1.0 Meeting Demand 02 2.0 Installation Testing 03 3.0 Commissioning the MON 05 The evolution of service is driving a desire
More information