Dynamic Imaging With Multiple Resolutions Along Phase-Encode and Slice-Select Dimensions

Size: px
Start display at page:

Download "Dynamic Imaging With Multiple Resolutions Along Phase-Encode and Slice-Select Dimensions"

Transcription

1 Magnetic Resonance in Medicine 45: (2001) Dynamic Imaging With Multiple Resolutions Along Phase-Encode and Slice-Select Dimensions Lawrence P. Panych, 1 * Lei Zhao, 1 Ferenc A. Jolesz, 1 and Robert V. Mulkern 2 An implementation is reported of an imaging method to obtain MUltiple Resolutions along Phase-encode and Slice-select dimensions (MURPS), which enables dynamic imaging of focal changes using a graded, multiresolution approach. MURPS allows one to trade spatial resolution in part of the volume for improved temporal resolution in dynamic imaging applications. A unique method of Hadamard slice encoding is used, enabling the varying of the phase encode and slice resolution while maintaining a constant effective TR throughout the entire 3-D volume. MURPS was implemented using a gradient-recalled echo sequence, and its utility was demonstrated for MR temperature monitoring. In this preliminary work, it has been shown that changes throughout a large volume can be effectively monitored in times that would normally only permit dynamic imaging in one or a very few slices. Magn Reson Med 45: , Wiley-Liss, Inc. Key words: dynamic multiresolution imaging; temperature monitoring; Hadamard encoding; spatially-selective RF excitation; non-fourier encoding; adaptive imaging Dynamic MRI methods for monitoring spatially localized changes usually involve reducing volume coverage to improve temporal resolution. This is a good tradeoff when changes are clearly limited to a small, well-defined volume of interest (VOI). However, in some applications, such as monitoring the progress of heating in focused ultrasound or interstitial laser surgery, this tradeoff is less than ideal because changes are not restricted to a well-defined volume. To achieve a reasonable temporal resolution, it then becomes necessary to monitor a relatively small VOI comprising very few slices where the dominant change is experienced. Changes may occur in other slices throughout the volume but, in the interests of temporal resolution, they may not be monitored. Here we propose a graded, multiresolution image acquisition as an alternative to the all or nothing approach. In a standard acquisition, the parameters affecting spatial resolution slice thickness and number of phase encodes are constant for all slice acquisitions. For example, if it is determined that 3-mm slices and 256 phase encodes are necessary to spatially resolve changes in one slice, the same parameters must be used for all slices, thereby limiting the volume that can be imaged. If, however, slice thickness and phase-encode resolution are varied within an acquisition, we reason that it will be possible to effectively monitor changes throughout a much larger volume without sacrificing temporal resolution. This depends, of course, on whether or not the requisite spatial resolution to resolve changes is also variable throughout the volume. For applications such as temperature mapping this is indeed the case. Sharp and rapid temperature changes occur close to the source of energy deposition; however, in regions distant from the source, the changes are slower and more spatially diffuse. In previous work, we and others have investigated multiresolution approaches based on non-fourier encoding with wavelets (1 9). In wavelet encoding, RF pulses were designed to excite spatial profiles shaped like wavelet functions. The special features of the wavelet transform make it particularly suitable for reconstructing images with spatially variable resolution, and this led us to propose adaptive imaging strategies based on multiresolution wavelet encoding (2). Our previous work with wavelets involved variation of resolution in a single RF select direction only. Other groups (10 14) have reported imaging methods that also involve expansions in terms of non-fourier basis functions with spatially variable resolution. In these methods, variable resolution is achieved through a nonuniform sampling of k-space in phase-encoding directions combined with a special reconstruction method that projects the selected Fourier basis set onto the non-fourier basis functions. We propose a method to obtain MUltiple Resolutions along Phase-encode and Slice-select dimensions (MURPS). In the MURPS approach, resolution parameters are constant within any single slice; however, both slice thickness and phase-encode resolution are varied for different slices. By reducing the number of phase encodes and increasing the slice thickness for some slices, we are able to directly trade reduced spatial resolution in some parts of the volume for an overall increase in either temporal resolution or volume coverage. Currently MURPS is being developed for MR temperature monitoring, but it also has potential for use with other applications, such as fmri and other dynamic contrast studies. 1 Department of Radiology, Brigham and Women s Hospital, Harvard Medical School, Boston, Massachusetts. 2 Department of Radiology, Children s Hospital, Harvard Medical School, Boston, Massachusetts. Presented in part at the 8th Annual Meeting of ISMRM, Denver, Grant sponsor: Whitaker Foundation; Grant number: TF ; Grant sponsor: NIH; Grant numbers: R29-CA70314; P01-CA67165; R01-NS *Correspondence to: Lawrence P. Panych, Ph.D., Department of Radiology, Brigham and Women s Hospital, 75 Francis Street, Boston, MA Received 2 October 2000; revised 3 January 2001; accepted 8 February Wiley-Liss, Inc. 940 METHODS Multiresolution Encoding Method Here we describe a MURPS implementation for acquiring seven slices, each with variable slice thickness and variable phase-encode resolution but with constant effective TR. The drawing at the top of Fig. 1 demonstrates the MURPS acquisition schematically, with each block in the drawing representing a view of a voxel in the slice-phase plane. The seven slices are arranged horizontally. For sim-

2 Dynamic Multiresolution Imaging 941 excited four times in the 12 TR periods. However, because reconstruction of the B slices requires two separate Hadamard encodes (B1 B2, B1 B2), only two phase encodes can be collected in the same time period in which slice A is phase encoded four times. The volume covered by the C slices is also excited four times in the 12 TR periods. Because reconstruction of the C slices requires four separate Hadamard encodes (C1 C2 C3 C4, C1 C2 C3 C4, C1 C2 C3 C4, and C1 C2 C3 C4), only one phase encode can be collected in the same time period in which slice A is phase encoded four times and the B slices are phase encoded twice. Note that with this scheme there is a relationship between the number of phase encodes and the Hadamard encoding factor. The reduction in the number of phase encodes is inversely proportional to the Hadamard encoding factor so that the product of the number of phase encodes and the Hadamard encoding factor is constant for all slices. For example, when the number of phase encodes acquired for a slice is one fourth of the full resolution case, then four slices are Hadamard encoded. Likewise if the number of phase encodes is reduced by half, the number of Hadamard-encoded slices is 2. This relationship is necessary in order to maintain a constant effective TR for all slices. There is no such restriction on the slice thickness. FIG. 1. The upper drawing is a schematic depiction of voxel sizes in the slice and phase-encoding directions for a seven-slice MURPS acquisition method. The lower drawing is of excitation profiles for the seven MURPS slices. The table lists resolution parameters for the seven-slice MURPS scheme used in brain imaging experiments. plicity, only a small number of voxels are shown in the phase-encoding direction (i.e., the vertical direction in the drawing). In the frequency-encoding direction (throughplane in Fig. 1) the resolution is constant for all slices. Ideal excitation profiles for the seven slices are shown below the drawing in Fig. 1. The table at the bottom of Fig. 1 lists parameters that were used in some of our experiments. In this example, a relatively thin (5-mm) center slice (labeled A) was acquired at full in-plane resolution. In an application such as the monitoring of heating by focused ultrasound, this slice should be placed so as to cover the location of maximum energy deposition. For two neighboring slices (B1 and B2), slice thicknesses were increased (to 10 mm) and phaseencode resolution was reduced by a factor of 2. For the outer four slices (C1, C2, C3, and C4), slice thicknesses were again varied and phase-encode resolution was reduced by another factor of 2. In the MURPS implementation described here, the B and C slices are separately Hadamard encoded (15 18) to enable a reduced number of phase encodes in the outer slices while still maintaining a fixed TR for all slices. Figure 2 shows the encoding scheme for the first 12 TR periods. After every three TR periods, the entire imaging volume (A, B, and C slices) is excited once; thus the effective TR throughout the volume is 3 TR. At the end of the 12 TR periods, as shown in Fig. 2, slice A has been excited four times and four phase encodes of the slice have been collected. The volume covered by the B slices has also been FIG. 2. Hadamard encoding profiles used for acquisition of data to reconstruct seven MURPS slices as shown in Fig. 1. From top to bottom, a different profile is used in succession on each of 12 successive TR periods. This sequence of profiles is used repeatedly for 3*N TR periods, where N is the number of phase encodes collected for slice A. The order of phase encoding is shown for the first 12 TR periods.

3 942 Panych et al. We are free to set slices to any convenient thickness. As shown in the table of Fig. 1, it is not even necessary to maintain constant slice thickness within a group of Hadamard-encoded slices (e.g., the C group). In MURPS, the signal-to-noise ratio (SNR) varies between slices according to slice thickness, number of phase encodes, and number of Hadamard encodes. The SNR in any slice with respect to slice A is equal to (S/ S A ) (HP A )/P, where S is the slice thickness, P is the number of phase encodes, and H is the number of Hadamard encodes. S A is the thickness of the slice A, and P A is the number of phase encodes for the slice. Since there is no Hadamard encoding of slice A, H A 1, and is not included in the equation. As discussed above, the product of the number of phase encodes and the Hadamard encoding factor is constant for all slices, so that P P A /H; thus, the relative SNR in any slice with respect to slice A is simply equal to H (S/S A ). A 2D spoiled gradient-echo (SPGR) sequence was modified for the MURPS implementation as follows. First, the sequence was altered so that a set of arbitrary RF pulses (as defined in external files) could be preloaded and then different RF pulses could be used for each excitation. Twelve pulses designed to excite the profiles shown in Fig. 2 were computed on a workstation and preloaded into the imager. Each of the 12 RF pulses (of 5.12-ms duration and 30 flip angle) was used in succession, with the sequence repeating after every 12 excitations. The imaging pulse sequence was further modified to execute a phase-encoding order, as denoted in Fig. 2, for the first 12 excitations. On each set of excitations, the phase-encoding order was simply equal to the order from the previous set of excitations plus {4,2,1,4,2,1,4,2,1,4,2,1}. The phase-encode ordering for each TR period was programmed directly in the pulse sequence, although it is also possible to read the order from an external file. Brain Imaging Experiments The seven-slice multiresolution MURPS acquisition scheme was implemented on our 1.5T Signa LX system (GE Medical Systems, Milwaukee, WI) and used to acquire a set of slices from a human volunteer after obtaining informed consent according to institutional review protocol. Seven slices of varying thickness and in-plane resolution were acquired in an interleaved fashion using the modified SPGR sequence (TR/TE, 30/8.6 ms). Slice A was acquired at full in-plane resolution ( acquisition matrix, 240-mm FOV) and at a thickness of 5 mm. The B slices were acquired at half the phase-encode resolution ( acquisition matrix) and the C slices were acquired at one fourth of the phase-encode resolution ( acquisition matrix). The B and C slices were two or three times the thickness of the center slice. As shown in the table of Fig. 1, with this scheme voxel volume increases in graded steps from 5 mm 3 in the center slice to 60 mm 3 in the outer slices. The total time for encoding the seven multiresolution slices is 3*N*TR, where N is the number of phase encodes collected for slice A (i.e., 256 in the brain imaging experiments). The total volume covered is 7.5 cm 24 cm 24 cm. To cover the same volume with 5-mm slices at the highest in-plane resolution would require 15*N*TR. Thus, the temporal resolution is increased fivefold over what it would be at full spatial resolution. Or, for the same temporal resolution, there is a fivefold increase in volume coverage with MURPS over a standard acquisition at constant full spatial resolution. A straightforward tradeoff is made, decreasing spatial resolution in some parts of the volume in order to increase either temporal resolution or volume coverage. Temperature Mapping Experiments MURPS was implemented on a Signa 1.5T LX system for dynamic temperature monitoring using the proton resonant shift method (19 21). A glass tube was inserted into the center of an orange and placed in the MRI system. A catheter with attached syringe was inserted into the glass tube so that hot water could be injected into the tube at the start of imaging. A doped water phantom was also placed within the imaged volume to monitor changes due to slow drift of the static magnetic field during the experiment. The method of acquisition was as described in the previous sections with resolution parameters as follows: Numbers of phase encodes in the A, B, and C slices, respectively, were 120, 60, and 30, and slice thicknesses were 80% of those used in the brain imaging experiments (i.e., 4 mm in the center slice, A, and 12 mm in the edge slices, C1 and C4). A dynamic series of images was acquired at the rate of one set of seven multiresolution slices every 12.6 s over a period of approximately 1 h. Each set of multiresolution slices was reconstructed and phase-difference images were computed by performing a complex division of the image set at each time point with the final image set in the dynamic series, then taking the phase of the result. For each voxel, a time series of 252 phase-difference values was obtained. These were then converted directly to temperature change given a reported shift in the proton resonance of approximately ppm per C of change (21). Prior to conversion from phase to temperature values, the phase was unwrapped using a library function in Matlab (The Mathworks, Natick, MA) that removes abrupt wraparound phase changes. The average phase was also measured in a VOI in the unheated doped-water phantom to separate phase changes due to heating from the background drift in phase due to static magnetic field changes. RESULTS Brain Imaging Experiments The upper part of Fig. 3 shows seven reconstructed axial brain slice images using MURPS. The variable resolution is evident across the volume with highest resolution (and lower SNR) in slice A and lowest resolution in slices C1 and C4. Note the uniform contrast behavior throughout the volume. The total volume covered in 23 s is 75 mm 240 mm 240 mm. At full resolution (5-mm 3 voxels) 115 s would be required to cover the same volume. The table at the bottom of Fig. 3 gives the theoretical and experimental SNR of the MURPS slices with respect to slice A. There is a relatively close agreement between the theoretical SNR values and those determined experimentally, thus validat-

4 Dynamic Multiresolution Imaging 943 FIG. 3. At top are seven brain slices acquired using the MURPS method. See Fig. 1 for a description of resolution parameters. The seven slices cover a volume of cm. The vertical is the frequency-encode direction, and the horizontal is the phase-encode direction. Below the MURPS set are images of virtual slices created from a reference set of 15 slices of 5-mm thickness and phase-encode resolution of 256, which cover the same total volume as the seven slices above. The virtual slices were created by averaging appropriate 5-mm slices to simulate the thicker slices, and by filtering the images in the phase-encode direction. The table below the images gives the theoretical and experimental SNR of each MURPS slice with respect to slice A. Relative SNR is equal to H (S/S A ) where S is the slice thickness, H is the Hadamard encoding factor, and S A is the thickness of slice A. ing our expression for the variation in SNR between MURPS-acquired slices. A reference set of 15 slices of 5-mm thickness covering the same volume as above was also acquired. Phaseencode resolution was 256 for this set. The slices were acquired sequentially with a TR of 90 ms so as to match the T 1 weighting of the multiresolution set (where the effective TR was ms). From this reference set of slices, virtual slices were created by averaging appropriate slices to simulate the thicker slices in the MURPS set, and then filtering the images in the phase-encode direction. As expected, the virtual set is very similar to the MURPS result except for a few understandable differences. Because the reference set was acquired sequentially, in-flowing spins were not saturated and blood vessels in the virtual image set appear bright. In MURPS, the entire volume is excited every three shots, minimizing the in-flow effect from vessels. Other minor contrast differences are likely due to slightly different flip angle settings in the two acquisitions. There is also a small but noticeable dark artifact in the frontal region of slice C4 in the MURPS result (Fig. 3, see arrow). We attribute this to susceptibility-induced losses in the slice direction due to the relatively large slice thickness. Note that it does not appear in the corresponding virtual slice because the image was created by averaging thinner slices. The left side of Fig. 4 shows mid-brain slices for axial, sagittal, and coronal orientations after an interpolated data set was created from the seven original MURPS slices. The interpolated slices were obtained using an interpolation function in Matlab. For comparison, we also created a comparable interpolated data set from the 15 reference slice images, and mid-brain slices for the three orientations are shown on the right of Fig. 4. The coronal slice on the left of Fig. 4 shows variable resolution in both the phase and slice directions, causing it to appear slightly more isotropically smoothed than the sagittal slice, which has variable resolution only in the slice direction. The dark streaks in the sagittal view on the left are due to thin horizontal features that are not broadened because the vertical is the frequency-encoded direction. However, the features are very poorly resolved in the slice direction, giving the streaked appearance. The central axial slice is, as expected, virtually identical with respect to resolution in the two sets. It should be noted that the interpolated images in Fig. 4 do not provide any information that does not also exist in Fig. 3. These interpolated data sets are presented only to demonstrate the tradeoffs that are made when using MURPS to increase temporal resolution or volume coverage. Temperature Mapping Experiments The left side of Fig. 5 shows three-plane views of an interpolated data set computed from seven magnitude multiresolution MURPS images of the phantom (orange with tube of hot water). Voxel sizes varied from 2 2

5 944 Panych et al. FIG. 4. Left: Mid-brain slices for axial, sagittal, and coronal orientations after creation of an interpolated data set from the MURPS slices shown at the top of Fig. 3. Right: Mid-brain slices for axial, sagittal, and coronal orientations after creation of an interpolated data set using 15 5-mm slices covering the same volume as the seven MURPS slices. 4 mm in the center slice to mm in the edge slices. Temperature-difference maps in three-plane views were computed and are also shown in Fig. 5 (right side). The temperature maps are obtained from phase-difference images using the linear relationship between the phase change (after unwrapping) and temperature change. The maps in Fig. 5 show the temperature difference at the beginning of the cooling period with respect to the baseline temperature at the end of the cooling period. As expected, the largest temperature change occurs close to the hot water tube. The images clearly demonstrate the ability of the multiresolution acquisition method to depict temperature change throughout a large, 3D volume. Figure 6 shows a segmented gray-scale representation of the temperature changes ocurring over a 30-min period along one column through the phantom in each of the FIG. 5. Left image set: Three-plane views of interpolated data set obtained from magnitude images of a phantom (orange with hot water tube inserted in center) acquired using the MURPS method. Right image set: Three-plane views of interpolated temperature difference maps for the period just after injection of hot water into the tube. The maps show the temperature change in C with respect to baseline temperature. White represents the hottest temperature, as shown on the scale at the far right.

6 Dynamic Multiresolution Imaging 945 FIG. 6. Evolution of heat over a 30-min period along one plane (8 cm wide) cutting through the phantom in all seven MURPS slices. The location of the plane is shown in the set of slice reference images on the right side of the figure. The maps are segmented into four temperature regions that are represented in the figure by successively lighter shades of gray (see color legend at bottom). White represents the hottest temperature region. seven MURPS slices. The maps are segmented into four regions of temperature change: [T 2 ], [2 T 4 ], [4 T 10 ], and [T 10 ], which are represented in the figure by successively lighter shades of gray, with white representing the highest temperature region. The position of the column in each slice is shown on the right in Fig. 6. The columns are displayed side by side for each time point so that the horizontal direction represents time and the vertical direction represents space. Figure 6 demonstrates temperature change with time in each of these slices along the column indicated in the right-hand panel. The figure demonstrates that, as expected, the time lag to reach peak temperature is a function of the slice distance from the hot water tube (which is located in slice A), and that temperature changes are more spatially and temporally diffuse in regions further away from the heat source. It should be noted that the data shown in Fig. 6 are only a small fraction of that available from the complete MURPS data set. DISCUSSION In this work we present MURPS, a multiresolution imaging scheme for dynamic imaging with variable slice thickness and phase-encode resolution. The MURPS method allows one to trade spatial resolution in part of the volume for improved temporal resolution or increased volume coverage in dynamic imaging applications. MURPS was implemented using a modified SPGR sequence, and its utility was demonstrated for MR temperature monitoring. In the preliminary work presented here, we have shown that temperature changes throughout a large volume can be effectively monitored in times that would normally only permit monitoring in one or a very few slices. The potential for the use of MURPS with other applications in which we have applied a multiresolution imaging approach, such as fmri (22), and in dynamic contrast studies (4), is currently under investigation.

7 946 Panych et al. The MURPS implementation described here uses Hadamard slice encoding, enabling the varying of the phaseencode resolution while maintaining constant effective TR. This is achieved when the product of the number of phase encodes and the Hadamard encoding factor is constant for all slices. Sequential, non-hadamard MURPS acquisitions are also possible (i.e., all phase encodes for one slice are acquired before acquisition of another slice begins). Therefore, constant effective TR can be achieved without Hadamard encoding, with the advantage that the temporal resolution of each slice acquisition can be varied along with spatial resolution. A fully sequential acquisition, however, is only time efficient for short TR applications, and SNR would suffer in regions of medium and long T 1. It is also possible to design a non-hadamard MURPS implementation that combines both sequential and interleaved acquisitions. For example, if N/4 phase encodes of slices [A,B1,C1] are followed by N/4 phase encodes of [A,B1,C2], followed in turn by N/4 phase encodes of [A,B2,C3] and, then, N/4 phase encodes of [A,B2,C4], we would achieve the same effective TR as Hadamard MURPS along with a variable number of phase encodes in the A, B, and C slices. The absence of Hadamard encoding in this scheme has the advantage of avoiding potential peak power problems and the need to design special RF pulses. These advantages, however, would have to be weighed against several disadvantages. SNR per unit time in Hadamard MURPS would be 100% and 40% higher, respectively, in the C and B slices compared to the non-hadamard implementation. Furthermore, in the non-hadamard MURPS acquisition, several extra dummy excitations would be needed to allow for signal equilibrium to be reached in the B and C slices. In addition, the temporal sampling window is not constant for all sets of slices, as it is for Hadamard MURPS. Hadamard encoding may introduce some constraints in terms of peak RF power. This was not a problem in our work, however, due to the fact that a gradient-echo sequence with relatively short TR was used, so flip angles in the range of 30 were sufficient. For spin-echo imaging, peak power constraints are more severe and the implications need to be investigated. Because of the flip angles used, we did not expect distortions in the Hadamard profiles due to nonlinear effects. When higher flip angles are used, however, special RF design methods must be employed with Hadamard MURPS to ensure that the desired profiles are excited (23). SNR varies considerably between MURPS slices due to the large variation in voxel sizes. In the implementation reported here, SNR in the central, high-resolution slice is the same as would be obtained in a normal high-resolution acquisition for an equivalent TR. In the MURPS acquisition, however, the SNR increases as one moves away from the central slice, and it is a full 12 times greater in the outer slices than in the central slice. Generally, because changes tend to become much smaller as one moves away from the focal VOI, an important advantage of MURPS is that it provides increasing SNR, which enables more accurate measures of the changes where they tend to be small. It should be noted that MURPS is not pulse-sequence dependent and, in principle, it is possible to implement it for any kind of sequence, including fast imaging approaches such as echo-planar imaging. Also, although we implemented MURPS using seven multiresolution slices, any number of variations is possible. It is not necessary, for example, to limit oneself to a single high-resolution slice located at the center of the imaged volume. Slices can be arranged in any fashion with one, two, or more highresolution slices mixed with other slices of varying thickness. Furthermore, it is not even necessary that slices be contiguous. Because RF pulses are designed off-line and read by the pulse sequence prior to scanning, it is easy to provide a flexible choice of slice thickness and location. We are currently investigating dynamically adaptive approaches in which such a choice is made on-line during a dynamic scan session (22). Finally, MURPS is compatible with other dynamic imaging methods that involve reconstructing information from a reduced number of phase encodes, thereby raising the possibility of hybrid approaches. For example, combining MURPS with keyhole approaches (24) involves a minimum of modifications simply replacing the appropriate data from a high-resolution static baseline dataset with the multiresolution dynamic data obtained in MURPS. Extended keyhole-type methods, such as time resolved imaging of contrast kinetics (TRICKS) (25,26), which segment k-space into regions with different update rates can also be combined with MURPS by redesigning the phase-encode acquisition scheme within individual slices. Constrained reconstruction methods, such as reduced-encoding imaging by generalized-series reconstruction (RIGR) (27 29) and related methods (30), use loworder phase encodes to reconstruct higher-resolution dynamic data based on assumptions about the dynamics, and can also be combined with MURPS to reconstruct data in the slices in which we have reduced phase encoding. More complex methods that involve nonuniform k-space sampling, such as locally focused MRI (11 14), are also compatible because MURPS does not require the use of any particular set of reduced phase encodes. Combining MURPS with locally focused MRI would enable variable resolution imaging within individual slices as well as between slices, further extending the multiresolution capability of both methods. ACKNOWLEDGMENTS The authors acknowledge Dr. Seung-Schik Yoo for his assistance with some of the experimental work, and Dr. Kullervo Hynynen and Nathan McDannold for the useful discussions on applications of the method for temperature mapping. REFERENCES 1. Panych LP, Jakab PD, Jolesz FA. An implementation of wavelet encoded MRI. J Magn Reson Imaging 1993;3: Panych LP, Jolesz FA. A dynamically adaptive imaging algorithm for wavelet encoded MRI. Magn Reson Med 1994;32: Panych LP, Zientara GP, Saiviroonporn P, Yoo S-S, Jolesz FA. Digital wavelet-encoded MRI: a new wavelet-encoding methodology. J Magn Reson Imaging 1998;8: Shimizu K, Panych LP, Mulkern RV, Yoo S-S, Schwartz RB, Kikinis R, Jolesz FA. Partial wavelet encoding: a new approach for accelerating temporal resolution in contrast-enhanced MR imaging. J Magn Reson Imaging 1999;9:

8 Dynamic Multiresolution Imaging Healy DM, Weaver JB. Two applications of wavelet transforms in magnetic resonance imaging. IEEE Trans Inform Theory 1992;38: Weaver JB, Xu Y, Healy DM, Driscoll JR. Wavelet-encoded MR imaging. Magn Reson Med 1992;24: Gelman N, Wood ML. Wavelet encoding for three-dimensional gradient-echo MR imaging. Magn Reson Med 1996;36: Gelman N, Wood ML. Wavelet encoding for improved SNR and retrospective slice thickness adjustment. Magn Reson Med 1998;39: Peters RD, Wood ML. Multilevel wavelet-transform encoding in MRI. J Magn Reson Imaging 1996;6: Cao Y, Levin DN. MR imaging with spatially variable resolution. J Magn Reson Imaging 1992;2: Cao Y, Levin DN, Yao L. Locally focused MRI. Magn Reson Med 1995;34: Yao L, Cao Y, Levin DN. 2D locally focused MRI: applications to dynamic and spectroscopic imaging. Magn Reson Med 1996;36: Kuperman VY, Nagle SK, Levin DN. Locally focused MRI of interventions. J Magn Reson Imaging 1998;8: Kuperman VY, Nagle SK, Alley MT, Glover GH, Levin DN. Locally focused contrast-enhanced carotid MRA. J Magn Reson Imaging 1999; 9: Oh C, Park HW, Cho ZH. Line-integral projection reconstruction (LPR) with slice encoding techniques: multislice regional imaging in NMR tomography. IEEE Trans Med Imaging 1984;MI-3: Bolinger L, Leigh JS. Hadamard spectroscopic imaging (HSI) for multivolume localization. J Magn Reson 1988;80: Goelman G, Harihara Subramanian V, Leigh JS. Transverse Hadamard spectroscopic imaging techniques. J Magn Reson 1990;89: Gonen O, Arias-Mendoza F, Goelman B. 3D localized in vivo 1 H spectroscopy of human brain using a hybrid of 1D-Hadamard with 2Dchemical shift imaging. Magn Reson Med 1997;37: de Poorter J, de Wagter C, de Deene Y, Thomsen C, Stahlberg F, Achten E. Noninvasive MRI thermometry with the proton resonance frequency (PRF) method: in vivo results in human muscle. Magn Reson Med 1995;33: Chung AH, Hynynen K, Colucci V, Oshio K, Cline HE, Jolesz FA. Optimization of spoiled gradient-echo phase imaging for in vivo localization of a focused ultrasound beam. Magn Reson Med 1996;36: Lewa CJ, Certaines JD. Body temperature mapping by magnetic resonance imaging. Spect Lett 1994;27: Yoo S-S, Guttmann CRG, Zhao L, Guttmann CRG, Panych LP. Real-time adaptive functional MRI. Neuroimage 1999;10: Cunningham CH, Wood ML. Method for improved multiband excitation profiles using the Shinnar-Le Roux transform. Magn Reson Med 1999;42: van Vaals JJ, Brummer ME, Dixon MT, Tuithof HH, Engels H, Nelson RC, Gerety BM, Chezmar JL, den Boer JA. Keyhole method for accelerating imaging of contrast agent uptake. J Magn Reson Imaging 1993; 3: Korosec FR, Frayne R, Grist TM, Mistretta CA. Time-resolved contrastenhanced 3D MR angiography. Magn Reson Med 1996;36: Mistretta CA, Grist TM, Koresec FR, Frayne R, Peters DC, Mazaheri Y, Carrol TJ. 3D time-resolved contrast-enhanced MR DSA: advantages and tradeoffs. Magn Reson Med 1998;40: Liang Z-P, Lauterbur PC. An efficient method for dynamic magnetic resonance imaging. IEEE Trans Med Imaging 1994;13: Webb AG, Liang ZP, Magin RL, Lauterbur PC. Applications of reducedencoding MR imaging with generalized-series reconstruction (RIGR). J Magn Reson Imaging 1993;3: Chandra S, Liang ZP, Webb A, Lee H, Morris HD, Lauterbur PC. Application of reduced-encoding imaging with generalized-series reconstruction (RIGR) in dynamic MR imaging. J Magn Reson Imaging 1996; 6: Oesterle C, Hennig J. Improvement of spatial resolution of keyhole effect images. Magn Reson Med 1998;39:

Dynamic Imaging with Multiple Resolutions Along Phase-Encode and Slice-Select Dimensions

Dynamic Imaging with Multiple Resolutions Along Phase-Encode and Slice-Select Dimensions Dynamic Imaging with Multiple Resolutions Along Phase-Encode and Slice-Select Dimensions Lawrence P. Panych, Lei Zhao, Ferenc A. Jolesz and Robert V. Mulkern Departments of Radiology, Brigham and Women

More information

Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm

Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm L.P. Panych 1,3, B. Madore 1,3, W.S. Hoge 1,3, R.V. Mulkern 2,3 1 Brigham and

More information

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous,

More information

Constrained Reconstruction of Sparse Cardiac MR DTI Data

Constrained Reconstruction of Sparse Cardiac MR DTI Data Constrained Reconstruction of Sparse Cardiac MR DTI Data Ganesh Adluru 1,3, Edward Hsu, and Edward V.R. DiBella,3 1 Electrical and Computer Engineering department, 50 S. Central Campus Dr., MEB, University

More information

Spiral keyhole imaging for MR fingerprinting

Spiral keyhole imaging for MR fingerprinting Spiral keyhole imaging for MR fingerprinting Guido Buonincontri 1, Laura Biagi 1,2, Pedro A Gómez 3,4, Rolf F Schulte 4, Michela Tosetti 1,2 1 IMAGO7 Research Center, Pisa, Italy 2 IRCCS Stella Maris,

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

K-Space Trajectories and Spiral Scan

K-Space Trajectories and Spiral Scan K-Space and Spiral Scan Presented by: Novena Rangwala nrangw2@uic.edu 1 Outline K-space Gridding Reconstruction Features of Spiral Sampling Pulse Sequences Mathematical Basis of Spiral Scanning Variations

More information

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street MRI is located in the sub basement of CC wing. From Queen or Victoria, follow the baby blue arrows and ride the CC south

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

k-space Interpretation of the Rose Model: Noise Limitation on the Detectable Resolution in MRI

k-space Interpretation of the Rose Model: Noise Limitation on the Detectable Resolution in MRI k-space Interpretation of the Rose Model: Noise Limitation on the Detectable Resolution in MRI Richard Watts and Yi Wang* Magnetic Resonance in Medicine 48:550 554 (2002) Noise limitation on the detected

More information

Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA)

Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA) Magnetic Resonance in Medicine 53:981 985 (2005) Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA) Felix A. Breuer, 1 * Peter Kellman, 2 Mark A. Griswold, 1 and Peter M. Jakob 1 Current

More information

Role of Parallel Imaging in High Field Functional MRI

Role of Parallel Imaging in High Field Functional MRI Role of Parallel Imaging in High Field Functional MRI Douglas C. Noll & Bradley P. Sutton Department of Biomedical Engineering, University of Michigan Supported by NIH Grant DA15410 & The Whitaker Foundation

More information

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution.

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution. MRES 7005 - Fast Imaging Techniques Module 5: Dynamic Imaging and Phase Sharing (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review Improving Temporal Resolution True-FISP (I) True-FISP (II) Keyhole

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005 Field Maps John Pauly October 5, 25 The acquisition and reconstruction of frequency, or field, maps is important for both the acquisition of MRI data, and for its reconstruction. Many of the imaging methods

More information

MRI Physics II: Gradients, Imaging

MRI Physics II: Gradients, Imaging MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction

A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction Tina Memo No. 2007-003 Published in Proc. MIUA 2007 A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction P. A. Bromiley and N.A. Thacker Last updated 13 / 4 / 2007 Imaging Science and

More information

Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction

Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction Nan-kuei Chen 1,5 and Alice M. Wyrwicz 4 * Magnetic Resonance in Medicine 51:147 153 (004) Odd even echo inconsistencies result

More information

High Spatial Resolution EPI Using an Odd Number of Interleaves

High Spatial Resolution EPI Using an Odd Number of Interleaves Magnetic Resonance in Medicine 41:1199 1205 (1999) High Spatial Resolution EPI Using an Odd Number of Interleaves Michael H. Buonocore* and David C. Zhu Ghost artifacts in echoplanar imaging (EPI) arise

More information

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE MRES 7005 - Fast Imaging Techniques Module 4 K-Space Symmetry Review K-Space Review K-Space Symmetry Partial or Fractional Echo Half or Partial Fourier HASTE Conditions for successful reconstruction Interpolation

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Dynamic Contrast enhanced MRA

Dynamic Contrast enhanced MRA Dynamic Contrast enhanced MRA Speaker: Yung-Chieh Chang Date : 106.07.22 Department of Radiology, Taichung Veterans General Hospital, Taichung, Taiwan 1 Outline Basic and advanced principles of Diffusion

More information

Super-resolution Reconstruction of Fetal Brain MRI

Super-resolution Reconstruction of Fetal Brain MRI Super-resolution Reconstruction of Fetal Brain MRI Ali Gholipour and Simon K. Warfield Computational Radiology Laboratory Children s Hospital Boston, Harvard Medical School Worshop on Image Analysis for

More information

Use of Multicoil Arrays for Separation of Signal from Multiple Slices Simultaneously Excited

Use of Multicoil Arrays for Separation of Signal from Multiple Slices Simultaneously Excited JOURNAL OF MAGNETIC RESONANCE IMAGING 13:313 317 (2001) Technical Note Use of Multicoil Arrays for Separation of Signal from Multiple Slices Simultaneously Excited David J. Larkman, PhD, 1 * Joseph V.

More information

Fast Imaging Trajectories: Non-Cartesian Sampling (1)

Fast Imaging Trajectories: Non-Cartesian Sampling (1) Fast Imaging Trajectories: Non-Cartesian Sampling (1) M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.05.03 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business

More information

A Doubly Adaptive Approach to Dynamic MRI Sequence Estimation

A Doubly Adaptive Approach to Dynamic MRI Sequence Estimation 1168 IEEE TRANSACTIONS ON IMAGE PROCESSING, VOL. 11, NO. 11, OCTOBER 2002 A Doubly Adaptive Approach to Dynamic MRI Sequence Estimation William Scott Hoge, Member, IEEE, Eric L. Miller, Member, IEEE, Hanoch

More information

3-D Gradient Shimming

3-D Gradient Shimming 3-D Gradient Shimming 3-D Gradient Shimming on imaging and micro-imaging systems Technical Overview Introduction Advantage statement The GE3DSHIM method is a 3-D gradient shimming procedure developed for

More information

MR Advance Techniques. Vascular Imaging. Class III

MR Advance Techniques. Vascular Imaging. Class III MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions EE225E/BIOE265 Spring 2013 Principles of MRI Miki Lustig Assignment 8 Solutions 1. Nishimura 7.1 P)jc%z 5 ivcr3. 1. I Due Wednesday April 10th, 2013 (a Scrhon5 R2iwd b 0 ZOms Xn,s r cx > qs 4-4 8ni6 4

More information

New Technology Allows Multiple Image Contrasts in a Single Scan

New Technology Allows Multiple Image Contrasts in a Single Scan These images were acquired with an investigational device. PD T2 T2 FLAIR T1 MAP T1 FLAIR PSIR T1 New Technology Allows Multiple Image Contrasts in a Single Scan MR exams can be time consuming. A typical

More information

High dynamic range magnetic resonance flow imaging in the abdomen

High dynamic range magnetic resonance flow imaging in the abdomen High dynamic range magnetic resonance flow imaging in the abdomen Christopher M. Sandino EE 367 Project Proposal 1 Motivation Time-resolved, volumetric phase-contrast magnetic resonance imaging (also known

More information

Sparse sampling in MRI: From basic theory to clinical application. R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology

Sparse sampling in MRI: From basic theory to clinical application. R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology Sparse sampling in MRI: From basic theory to clinical application R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology Objective Provide an intuitive overview of compressed sensing

More information

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Doug Dean Final Project Presentation ENGN 2500: Medical Image Analysis May 16, 2011 Outline Review of the

More information

Estimating Arterial Wall Shear Stress 1

Estimating Arterial Wall Shear Stress 1 DEPARTMENT OF STATISTICS University of Wisconsin 1210 West Dayton St. Madison, WI 53706 TECHNICAL REPORT NO. 1088 December 12, 2003 Estimating Arterial Wall Shear Stress 1 John D. Carew 2 Departments of

More information

Parallel Imaging. Marcin.

Parallel Imaging. Marcin. Parallel Imaging Marcin m.jankiewicz@gmail.com Parallel Imaging initial thoughts Over the last 15 years, great progress in the development of pmri methods has taken place, thereby producing a multitude

More information

Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging

Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging Nan-kuei Chen and Alice M. Wyrwicz* Magnetic Resonance in Medicine 41:1206 1213 (1999) A novel and effective technique is described

More information

TITLE: Regularized Reconstruction of Dynamic Contrast-Enhanced MR Images for Evaluation of Breast Lesions

TITLE: Regularized Reconstruction of Dynamic Contrast-Enhanced MR Images for Evaluation of Breast Lesions AD Award Number: W81XWH-08-1-0273 TITLE: Regularized Reconstruction of Dynamic Contrast-Enhanced MR Images for Evaluation of Breast Lesions PRINCIPAL INVESTIGATOR: Kimberly A. Khalsa CONTRACTING ORGANIZATION:

More information

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most

More information

ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T

ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T Revision date: 12/13/2010 Overview Susceptibility Weighted Imaging (SWI) is a relatively new data acquisition and processing

More information

Unaliasing by Fourier-Encoding the Overlaps Using the Temporal Dimension (UNFOLD), Applied to Cardiac Imaging and fmri

Unaliasing by Fourier-Encoding the Overlaps Using the Temporal Dimension (UNFOLD), Applied to Cardiac Imaging and fmri 1999 ISMRM YOUNG INVESTIGATORS MOORE AWARD PAPERS Magnetic Resonance in Medicine 42:813 828 (1999) Unaliasing by Fourier-Encoding the Overlaps Using the Temporal Dimension (UNFOLD), Applied to Cardiac

More information

Compressed Sensing for Rapid MR Imaging

Compressed Sensing for Rapid MR Imaging Compressed Sensing for Rapid Imaging Michael Lustig1, Juan Santos1, David Donoho2 and John Pauly1 1 Electrical Engineering Department, Stanford University 2 Statistics Department, Stanford University rapid

More information

Alternate Biplanar MR Navigation for Microwave Ablation of Liver Tumors

Alternate Biplanar MR Navigation for Microwave Ablation of Liver Tumors Magnetic Resonance in Medical Sciences, Vol. 4, No. 2, p. 89 94, 2005 TECHNICAL NOTE Alternate Biplanar MR Navigation for Microwave Ablation of Liver Tumors Koichiro SATO 1 *, Shigehiro MORIKAWA 2,ToshiroINUBUSHI

More information

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford Diffusion MRI Acquisition Karla Miller FMRIB Centre, University of Oxford karla@fmrib.ox.ac.uk Diffusion Imaging How is diffusion weighting achieved? How is the image acquired? What are the limitations,

More information

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

Fiber Selection from Diffusion Tensor Data based on Boolean Operators

Fiber Selection from Diffusion Tensor Data based on Boolean Operators Fiber Selection from Diffusion Tensor Data based on Boolean Operators D. Merhof 1, G. Greiner 2, M. Buchfelder 3, C. Nimsky 4 1 Visual Computing, University of Konstanz, Konstanz, Germany 2 Computer Graphics

More information

Modeling of MR-guided HIFU for Breast and Brain Therapy

Modeling of MR-guided HIFU for Breast and Brain Therapy Modeling of MR-guided HIFU for Breast and Brain Therapy Douglas A. Christensen, Allison Payne, Nick Todd, Scott Almquist, Alexis Farrer and Dennis L. Parker University of Utah Salt Lake City, Utah Overview

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Basic fmri Design and Analysis. Preprocessing

Basic fmri Design and Analysis. Preprocessing Basic fmri Design and Analysis Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial filtering

More information

Fmri Spatial Processing

Fmri Spatial Processing Educational Course: Fmri Spatial Processing Ray Razlighi Jun. 8, 2014 Spatial Processing Spatial Re-alignment Geometric distortion correction Spatial Normalization Smoothing Why, When, How, Which Why is

More information

Sampling, Ordering, Interleaving

Sampling, Ordering, Interleaving Sampling, Ordering, Interleaving Sampling patterns and PSFs View ordering Modulation due to transients Temporal modulations Slice interleaving Sequential, Odd/even, bit-reversed Arbitrary Other considerations:

More information

Accelerated MRI by SPEED with Generalized Sampling Schemes

Accelerated MRI by SPEED with Generalized Sampling Schemes Magnetic Resonance in Medicine 70:1674 1681 (201) Accelerated MRI by SPEED with Generalized Sampling Schemes Zhaoyang Jin 1 * and Qing-San Xiang 2 Purpose: To enhance the fast imaging technique of skipped

More information

Imaging Notes, Part IV

Imaging Notes, Part IV BME 483 MRI Notes 34 page 1 Imaging Notes, Part IV Slice Selective Excitation The most common approach for dealing with the 3 rd (z) dimension is to use slice selective excitation. This is done by applying

More information

TOF-MRA Using Multi-Oblique-Stack Acquisition (MOSA)

TOF-MRA Using Multi-Oblique-Stack Acquisition (MOSA) JOURNAL OF MAGNETIC RESONANCE IMAGING 26:432 436 (2007) Technical Note TOF-MRA Using Multi-Oblique-Stack Acquisition (MOSA) Ed X. Wu, PhD, 1,2 * Edward S. Hui, BEng, 1,2 and Jerry S. Cheung, BEng 1,2 Purpose:

More information

XI Signal-to-Noise (SNR)

XI Signal-to-Noise (SNR) XI Signal-to-Noise (SNR) Lecture notes by Assaf Tal n(t) t. Noise. Characterizing Noise Noise is a random signal that gets added to all of our measurements. In D it looks like this: while in D

More information

Motion artifact reduction technique for dual-contrast FSE imaging

Motion artifact reduction technique for dual-contrast FSE imaging Magnetic Resonance Imaging 20 (2002) 455 462 Motion artifact reduction technique for dual-contrast FSE imaging Eugene G. Kholmovski a, *, Alexei A. Samsonov a, Dennis L. Parker a,b a Department of Physics,

More information

Redundancy Encoding for Fast Dynamic MR Imaging using Structured Sparsity

Redundancy Encoding for Fast Dynamic MR Imaging using Structured Sparsity Redundancy Encoding for Fast Dynamic MR Imaging using Structured Sparsity Vimal Singh and Ahmed H. Tewfik Electrical and Computer Engineering Dept., The University of Texas at Austin, USA Abstract. For

More information

Functional MRI in Clinical Research and Practice Preprocessing

Functional MRI in Clinical Research and Practice Preprocessing Functional MRI in Clinical Research and Practice Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization

More information

Zigzag Sampling for Improved Parallel Imaging

Zigzag Sampling for Improved Parallel Imaging Magnetic Resonance in Medicine 60:474 478 (2008) Zigzag Sampling for Improved Parallel Imaging Felix A. Breuer, 1 * Hisamoto Moriguchi, 2 Nicole Seiberlich, 3 Martin Blaimer, 1 Peter M. Jakob, 1,3 Jeffrey

More information

Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing

Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing Peng Hu, Ph.D. Associate Professor Department of Radiological Sciences PengHu@mednet.ucla.edu 310-267-6838 MRI... MRI has low

More information

A Study of Medical Image Analysis System

A Study of Medical Image Analysis System Indian Journal of Science and Technology, Vol 8(25), DOI: 10.17485/ijst/2015/v8i25/80492, October 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 A Study of Medical Image Analysis System Kim Tae-Eun

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health MRI image formation Indiana University School of Medicine and Indiana University Health Disclosure No conflict of interest for this presentation 2 Outlines Data acquisition Spatial (Slice/Slab) selection

More information

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing Functional Connectivity Preprocessing Geometric distortion Head motion Geometric distortion Head motion EPI Data Are Acquired Serially EPI Data Are Acquired Serially descending 1 EPI Data Are Acquired

More information

Partially Parallel Imaging With Localized Sensitivities (PILS)

Partially Parallel Imaging With Localized Sensitivities (PILS) Partially Parallel Imaging With Localized Sensitivities (PILS) Magnetic Resonance in Medicine 44:602 609 (2000) Mark A. Griswold, 1 * Peter M. Jakob, 1 Mathias Nittka, 1 James W. Goldfarb, 2 and Axel Haase

More information

SPM8 for Basic and Clinical Investigators. Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

MRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD

MRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD MRI When to use What sequences Govind Chavhan, MD Assistant Professor and Staff Radiologist The Hospital For Sick Children, Toronto Planning Acquisition Post processing Interpretation Patient history and

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 BOLD and CBV functional maps showing EPI versus line-scanning FLASH fmri. A. Colored BOLD and CBV functional maps are shown in the highlighted window (green frame) of the raw EPI

More information

VD-AUTO-SMASH Imaging

VD-AUTO-SMASH Imaging Magnetic Resonance in Medicine 45:1066 1074 (2001) VD-AUTO-SMASH Imaging Robin M. Heidemann, Mark A. Griswold, Axel Haase, and Peter M. Jakob* Recently a self-calibrating SMASH technique, AUTO-SMASH, was

More information

M R I Physics Course

M R I Physics Course M R I Physics Course Multichannel Technology & Parallel Imaging Nathan Yanasak, Ph.D. Jerry Allison Ph.D. Tom Lavin, B.S. Department of Radiology Medical College of Georgia References: 1) The Physics of

More information

General and Efficient Super-Resolution Method for Multi-slice MRI

General and Efficient Super-Resolution Method for Multi-slice MRI General and Efficient Super-Resolution Method for Multi-slice MRI D.H.J. Poot 1,2,V.VanMeir 2, and J. Sijbers 2 1 BIGR, Erasmus Medical Center, Rotterdam 2 Visionlab, University of Antwerp, Antwerp Abstract.

More information

SIEMENS MAGNETOM Avanto syngo MR B15

SIEMENS MAGNETOM Avanto syngo MR B15 \\USER\INVESTIGATORS\Ravi\ADNI-phantom\QC Phantom-Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer

More information

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 16 Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 3.1 MRI (magnetic resonance imaging) MRI is a technique of measuring physical structure within the human anatomy. Our proposed research focuses

More information

NIH Public Access Author Manuscript J Magn Reson Imaging. Author manuscript; available in PMC 2009 October 1.

NIH Public Access Author Manuscript J Magn Reson Imaging. Author manuscript; available in PMC 2009 October 1. NIH Public Access Author Manuscript Published in final edited form as: J Magn Reson Imaging. 2008 October ; 28(4): 1026 1032. doi:10.1002/jmri.21506. Evaluation of Referenceless Thermometry in MRI-Guided

More information

MRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison

MRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison MRI Imaging Options Frank R. Korosec, Ph.D. Departments of Radiolog and Medical Phsics Universit of Wisconsin Madison f.korosec@hosp.wisc.edu As MR imaging becomes more developed, more imaging options

More information

COBRE Scan Information

COBRE Scan Information COBRE Scan Information Below is more information on the directory structure for the COBRE imaging data. Also below are the imaging parameters for each series. Directory structure: var/www/html/dropbox/1139_anonymized/human:

More information

3-D Compounding of B-Scan Ultrasound Images

3-D Compounding of B-Scan Ultrasound Images 3-D Compounding of B-Scan Ultrasound Images Jochen F. Krücker, Charles R. Meyer, Theresa A. Tuthill, Gerald L. LeCarpentier, J. Brian Fowlkes, Paul L. Carson University of Michigan, Dept. of Radiology,

More information

A Transmission Line Matrix Model for Shielding Effects in Stents

A Transmission Line Matrix Model for Shielding Effects in Stents A Transmission Line Matrix Model for hielding Effects in tents Razvan Ciocan (1), Nathan Ida (2) (1) Clemson University Physics and Astronomy Department Clemson University, C 29634-0978 ciocan@clemon.edu

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Exam 8N080 - Introduction MRI

Exam 8N080 - Introduction MRI Exam 8N080 - Introduction MRI Friday January 23 rd 2015, 13.30-16.30h For this exam you may use an ordinary calculator (not a graphical one). In total there are 6 assignments and a total of 65 points can

More information

G Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine. Compressed Sensing

G Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine. Compressed Sensing G16.4428 Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine Compressed Sensing Ricardo Otazo, PhD ricardo.otazo@nyumc.org Compressed

More information

Automatic Gradient Preemphasis Adjustment: A 15-Minute Journey to Improved Diffusion-Weighted Echo-Planar Imaging

Automatic Gradient Preemphasis Adjustment: A 15-Minute Journey to Improved Diffusion-Weighted Echo-Planar Imaging Automatic Gradient Preemphasis Adjustment: A 15-Minute Journey to Improved Diffusion-Weighted Echo-Planar Imaging Vincent J. Schmithorst* and Bernard J. Dardzinski Magnetic Resonance in Medicine 47:208

More information

Development of fast imaging techniques in MRI From the principle to the recent development

Development of fast imaging techniques in MRI From the principle to the recent development 980-8575 2-1 2012 10 13 Development of fast imaging techniques in MRI From the principle to the recent development Yoshio MACHIDA and Issei MORI Health Sciences, Tohoku University Graduate School of Medicine

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Outline High Spectral and Spatial Resolution MR Imaging (HiSS) What it is How to do it Ways to use it HiSS for Radiation

More information

UNIVERSITY OF SOUTHAMPTON

UNIVERSITY OF SOUTHAMPTON UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions

More information

surface Image reconstruction: 2D Fourier Transform

surface Image reconstruction: 2D Fourier Transform 2/1/217 Chapter 2-3 K-space Intro to k-space sampling (chap 3) Frequenc encoding and Discrete sampling (chap 2) Point Spread Function K-space properties K-space sampling principles (chap 3) Basic Contrast

More information

Magnetic Resonance Angiography

Magnetic Resonance Angiography Magnetic Resonance Angiography Course: Advance MRI (BIOE 594) Instructors: Dr Xiaohong Joe Zhou Dr. Shadi Othman By, Nayan Pasad Phase Contrast Angiography By Moran 1982, Bryan et. Al. 1984 and Moran et.

More information

Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques

Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques Sea Chen Department of Biomedical Engineering Advisors: Dr. Charles A. Bouman and Dr. Mark J. Lowe S. Chen Final Exam October

More information

Calculating the Distance Map for Binary Sampled Data

Calculating the Distance Map for Binary Sampled Data MITSUBISHI ELECTRIC RESEARCH LABORATORIES http://www.merl.com Calculating the Distance Map for Binary Sampled Data Sarah F. Frisken Gibson TR99-6 December 999 Abstract High quality rendering and physics-based

More information

Magnetic Resonance Elastography (MRE) of Liver Disease

Magnetic Resonance Elastography (MRE) of Liver Disease Magnetic Resonance Elastography (MRE) of Liver Disease Authored by: Jennifer Dolan Fox, PhD VirtualScopics Inc. jennifer_fox@virtualscopics.com 1-585-249-6231 1. Overview of MRE Imaging MRE is a magnetic

More information

Partial k-space Reconstruction

Partial k-space Reconstruction Chapter 2 Partial k-space Reconstruction 2.1 Motivation for Partial k- Space Reconstruction a) Magnitude b) Phase In theory, most MRI images depict the spin density as a function of position, and hence

More information

A MORPHOLOGY-BASED FILTER STRUCTURE FOR EDGE-ENHANCING SMOOTHING

A MORPHOLOGY-BASED FILTER STRUCTURE FOR EDGE-ENHANCING SMOOTHING Proceedings of the 1994 IEEE International Conference on Image Processing (ICIP-94), pp. 530-534. (Austin, Texas, 13-16 November 1994.) A MORPHOLOGY-BASED FILTER STRUCTURE FOR EDGE-ENHANCING SMOOTHING

More information

Sampling, Ordering, Interleaving

Sampling, Ordering, Interleaving Sampling, Ordering, Interleaving Sampling patterns and PSFs View ordering Modulation due to transients Temporal modulations Timing: cine, gating, triggering Slice interleaving Sequential, Odd/even, bit-reversed

More information

MR Imaging* Implementation of Wavelet-encoded. Original Research

MR Imaging* Implementation of Wavelet-encoded. Original Research - Original Research mplementation of Wavelet-encoded MR maging* Lawrence P. Panych, MEng Peter D. Jakab, MEng Ferenc A. Jolesz, MD Reconstructions of images from wavelet-encoded data are shown. The method

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Supplementary methods

Supplementary methods Supplementary methods This section provides additional technical details on the sample, the applied imaging and analysis steps and methods. Structural imaging Trained radiographers placed all participants

More information

Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 2004

Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 2004 Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 24 1 Alec Chi-Wah Wong Department of Electrical Engineering University of Southern California 374 McClintock

More information

Knowledge-Based Segmentation of Brain MRI Scans Using the Insight Toolkit

Knowledge-Based Segmentation of Brain MRI Scans Using the Insight Toolkit Knowledge-Based Segmentation of Brain MRI Scans Using the Insight Toolkit John Melonakos 1, Ramsey Al-Hakim 1, James Fallon 2 and Allen Tannenbaum 1 1 Georgia Institute of Technology, Atlanta GA 30332,

More information

Sources of Distortion in Functional MRI Data

Sources of Distortion in Functional MRI Data Human Brain Mapping 8:80 85(1999) Sources of Distortion in Functional MRI Data Peter Jezzard* and Stuart Clare FMRIB Centre, Department of Clinical Neurology, University of Oxford, Oxford, UK Abstract:

More information