Primary Use. Operating Principle
|
|
- Emmeline McGee
- 5 years ago
- Views:
Transcription
1 Primary Use The Leica DVM6 is an optical microscope that has the ability observe samples at a high magnification at a high resolution. The microscope allows users to view their sample with up to a 2350x magnification while not losing out on resolution or quality. This instrument is capable of producing images up to 1 micron in accuracy. The accuracy and resolution make the microscope a useful product in many MSE projects. The microscope lets you manually adjust the tilt of the lense system as well as the rotation and movement of the sample stage. This allows the user to produce any angle of picture with ease. There are currently two objectives at varying working distance and magnification allowing the user to interchange the lense system to achieve different images. There are also two different stages: one solid fill and one with a controllable backlight. The system uses a computer software that allows for seamless interface with the microscope and easy manipulation of the image. The software comes equipped with a variety of tools that allows the user to find the perfect image of his/her sample. Operating Principle The Leica microscope uses a converging lense system to produce a magnified image. Using a digital camera, the sample being observed is viewed and analyzed on an attached computer/monitor. Light passes through a series of lenses which either spread out light rays making the sample appear larger or further magnify the image. The sample will be in focus when it is at a sufficient distance away from the objective lense, and can be altered by raising and lowering the stage. This microscope relies heavily on image processing. The capabilities expand beyond 2D analysis and can generate a 3D surface map of a sample. The software can automatically take images of the sample through a predetermined range of distance. Safety Procedures The Leica Microscope has a very low possibility of injuring the user, however there are several safety precautions to keep in mind so the device does not break. When coming close to the lense system or handling the objectives, gloves are needed to protect the delicate systems from oil and dust. This is also important when handling samples on and around the stage. Always be aware of the distance between the lense system and the stage in order to prevent collisions when focusing or tilting the system. When tilting the system, never forcefully tilt the system without pressing the black handle and the silver handle together on the back of the system. The tilting of the system should be done with care, place one hand on the handles and the other hand on the white part of the head so the system doesn't fall of break against the stage. Be aware the device produces beeps when it detects the head getting too close to the stage.
2 General Procedure 1. Steps for startup: a. Remove red cover over the microscope b. Turn on the microscope with the switch on the back and wait for the indicator light on the left side to turn green Figure 1a-b. The power switch and green light referred to in step 1b. c. Once the light is green, open the Leica LAS X Hardware Configurator Figure 2: Shortcut to Leica Hardware Configurator. d. Click the Hardware Setup button in the top left e. Make sure the Port and the Stage both match in saying DVM6 f. Click Test g. If it says it was successfully completed you can click Apply then OK and you can close out of the Hardware Configurator software i. If it says there was an error you might have a problem with the connection between the microscope and the computer ii. If you open the LAS X and an image is not being generated, then restart either the computer or the microscope until the connection is achieved
3 h. Next open the LAS X software and follow the steps in the popup menus to continue to the software 2. Adjusting the Magnification a. To control the magnification of the microscope, you can twist the big red wheel above the objective lense system b. Counterclockwise decreases the magnification, clockwise increases it c. You can also control the digital magnification with the scroll wheel in the software, however the digital magnification will not increase the resolution of the image Figure 3: The red magnification adjustment wheel highlighted in blue. 3. Focusing the Microscope a. After finding the magnification of your choice, you can focus the lenses by twisting the adjustment wheels towards the bottom of the microscope below the stage b. The bigger wheel is for rough focusing while the smaller wheel is for fine adjustments c. Secondary, if you left click and drag on the live image, a menu will pop up. Simply click the auto-focus and the highlighted area you have selected with adjust to focus
4 Figure 4: Focusing wheels on the microscope. 4. Movement of the Stage a. You can move the stage itself to move the location of your sample in either the X or the Y direction b. Also you can adjust the rotation of the stage by twisting it using the black knob on the bottom of the left side of the stage c. You can monitor the angle of rotation on the popup on the software 5. Movement of the Lense System a. To adjust the tilt of the lense system by slightly pressing the black and silver handle sticking up towards the back of the microscope, behind the lense system Note: Never try tilting the system without slightly pressing the two handles together b. You can monitor the angle of the tilt on the popup on the software Figure 5: Gripping Procedure for Tilting the Microscope.
5 6. Opening a Previous Image a. The Explorer Menu lets the user navigate through computer files and open previous images b. Open the Explorer Menu in the top left on the screen c. Open the Folder or location where the image is located and open the Multifocus Image to use the full capabilities of the 3D image Figure 6: The explorer menu (top-left) for step Finding the Correct Exposure and Settings a. In the Image Menu, the top left button turns on the auto exposure. This is used in most situation to create a good image quick and easy b. To go in further detail or create a perfect image, the exposure, the brightness, and the contrast can be manually adjusts Figure 7: The image menu for step 7.
6 8. Controlling the illumination of the sample a. To get a easy image of a basic sample, the default ring light should be used. The intensity can be controlled as well as which lights are illuminated by picking the correct preset under the stage drop down b. To get a better overhead light, the Coaxial light should be used. This works well on metal samples where grain boundaries need to be analyzed c. The Backlight can be put on the stage to give the sample better light from the bottom. This works well with thin films and transparent samples Figure 8: The illumination menu for step Creating a multifocus image using the Start Up method a. Click on the Start Up button b. Switch on the Calculate an Extended Depth of Field switch c. Set the maximum number of steps to 60. If necessary, the user can raise the number of steps to 100 d. Close the menu and get the lowest point of the image into focus e. Click the Z-Stack button on the bottom toolbar of the software f. Monitor the process and stop the process once the highest point of the image just goes out of focus to insure the full image has been captured
7 Figure 9: The start-up multifocus menu for step Creating a multifocus image using the Custom method a. Click on the Custom button b. If there is a previous preset in the menu, click the trash button in the menu c. To start, get the lowest point of the image in focus and click begin d. Then get the highest point of the image in focus and click end e. Close out of the menu and click the Z-Stack button on the bottom toolbar of the software to run the program 11. Using a multifocus image to create a 3D model a. After an image has been Z-Stacked, click the 3D model button on the right side toolbar to open the 3D model b. A creat map can be created with different intensities as well as adding axis scallings, scale bars, and a heat map legend c. Different overlays and measurements can be added in the measurement tab Different step heights, volumes, and profiles can be created as overlays d. In order to return to the 2D view, click the 2D button on the bottom toolbar
8 Figure 10a-b: The button used to access 3d view (left) and the 3d workspace (right). 12. Stitching together multiple images together to create a photo of the entire sample a. Click the Stage Menu button to begin. b. Click the Spiral button c. After closing the menu, click the Stage icon button on the bottom toolbar d. This will begin the process and it will finish automatically e. Now a high resolution image has been created of the whole sample Figure 11. The stage menu button for step Saving an image a. To save an image, click the capture button on the bottom toolbar b. In order to save the image with overlays, select the stamp button on the top of the right toolbar. If the capture button has a stamp on it, then the overlays will be saved
9 Figure 12. Location of the save image button for step Create overlays on an image a. The top right of the toolbar has a menu to add and control overlays b. To measure distance, click the distance to distance button and click on two points that are being measured c. A scaler bar can be added through the scale bar menu and can be customized d. Annotations can be added that can have text added to point out certain features of the image e. The overlay menu can control which overlays are shown and if they will be saved or not Figure 13. The overlay menu for step Changing the objective a. Put on nitrile gloves b. Loosen locking screw on the left side of the microscope head c. Grip the objective with one hand and slowly pull directly outward until fully clear of the microscope head d. Without setting the objective down, place it in its designated box (located in clear containers beneath the Mark-10 e. Carefully lift the new objective out of its box and align it to the tracks on the microscope head
10 f. Slowly push the objective in all the way and tighten the locking screw Figure 14: The locking screw highlighted in blue.
LEXT 3D Measuring LASER Microscope
LEXT 3D Measuring LASER Microscope Warning: This instrument may only be operated by those who have been trained by AAF staff and have read and signed the AAF laboratory policies. A) STARTUP 1. Computer
More informationOperating Procedure for Horiba Raman Microscope
Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch
More informationM40 Microscope User s Manual
M40 Microscope User s Manual for M40 and M40RT Microscope Components: Trinocular Port Eyepieces Beam Splitter Field Diaphragm Adjustment Aperture Diaphragm Adjustment Filter Slots Analyzer Polarizer Hex
More informationINDEX. Quick Start Guide. Intermediate level guide. Digital Microscope. Displaying a scale. Split-screen display 3
Digital Microscope Quick Start Guide INDEX Displaying a scale Split-screen display 5 6 Using the mouse and operating the motorized XY stage Changing the lens Observing from an angle Additional uses for
More informationLab 5 Microscopy. Introduction. Carrying the Microscope. Depth of Focus. Parts of the Compound Microscope. Magnification
Lab 5 Microscopy Introduction The microscope is an instrument that contains one or more lenses and is used to view objects that are too small be seen with the unaided eye. A magnifying glass is a simple
More informationIntroduction. Features and Highlights
Introduction The Nodal Ninja Ultimate M2 Giga (M2G) manual pano head is an ideal solution for capturing high resolution Gigapixel or mosaic type imagery. The M2G will accommodate telephoto lenses up to
More informationTable of Contents. Chapter 1. Safety Precautions Notice Chapter 2. Components and accessories Chapter 3. Camera components...
P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046 P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046 Table of Contents Chapter 1. Safety Precautions Notice...
More informationUse of the Binocular Microscope
Use of the Binocular Microscope Before you begin this learning module be sure that you have the following materials in front of you: A microscope A packet of lens paper and Kimwipes A glass slide of a
More informationPrism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis
Start Up: Upon entering the laser room turn on the wall mounted Laser Power Button by pulling it away from the wall. Turn on Shutter controllers (toggle switch on back of unit). There should be a U in
More informationUser Manual. Binocular Zoom Stereo Microscope with Boom Stand. MicroscopeNet.com
User Manual Binocular Stereo Microscope with Boom Stand Model W42C1 Series MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 1 1. Component Illustration... 2 2. Installation...3
More informationOlympus IX-70 Imaging Protocol
Olympus IX-70 Imaging Protocol 1) System Startup Please note our sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused
More informationAutodesk Fusion 360: Render. Overview
Overview Rendering is the process of generating an image by combining geometry, camera, texture, lighting and shading (also called materials) information using a computer program. Before an image can be
More informationUser manual for the 3-D macro lens model 2007
Ing. J. de Wijs. Populierstraat 44, 4131 AR Vianen, the Netherlands Tel/Fax. +31 (0)347-372242 e-mail: info@dewijs-3d.com Website: www.dewijs-3d.com Design and production of stereoscopic instruments. Bank:
More informationWritten By: Ben Eisenman
iphone 3GS Rear Panel Replacement Replace a broken rear case on your iphone 3GS. Written By: Ben Eisenman ifixit CC BY-NC-SA www.ifixit.com Page 1 of 22 INTRODUCTION The plastic rear half of the iphone.
More informationOlympus IX-70 Imaging Protocol
Olympus IX-70 Imaging Protocol 1) System Startup F Please note our sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for
More informationiphone 3G Headphone Jack Replacement Replace a broken audio port in an iphone 3G. Written By: irobot ifixit CC BY-NC-SA
iphone 3G Headphone Jack Replacement Replace a broken audio port in an iphone 3G. Written By: irobot ifixit CC BY-NC-SA www.ifixit.com Page 1 of 18 INTRODUCTION No audio? Replace the headphone jack! TOOLS:
More informationThe Microscope: A Tool of the Scientist
Name Class Date Microscope Lab The Microscope: A Tool of the Scientist Background Information (You may refer to page 61 of your textbook.) Problem One of the most important tools of a life scientist is
More informationSwift SM100 Series Stereo Microscope
Swift SM100 Series Stereo Microscope Use and Care Manual SWIFT OPTICAL Enduring Quality and Technical Excellence 2 SWIFT SM100 SERIES STEREO MICROSCOPE The Swift SM100 Series stereo microscope is a full
More informationLab 4 - The Microscope
Lab 4 - The Microscope Part A: The parts of the compound microscope. Obtain a microscope as indicated by your instructor. Always carry the microscope with one hand under the base and the other hand holding
More informationThe Paper Project Guide to Confocal Imaging at Home & in the Classroom
The Paper Project Guide to Confocal Imaging at Home & in the Classroom http://lifesciences.asu.edu/paperproject The Paper Project Guide to Confocal Imaging at Home & in the Classroom CJ Kazilek & Dennis
More informationVPC-64/ VPX-64 VIDEO POLE CAMERA OPERATION MANUAL
VPC-64/ VPX-64 VIDEO POLE CAMERA OPERATION MANUAL RESEARCH ELECTRONICS INTERNATIONAL 455 Security Drive Algood, TN 38506 U.S.A. +1 931-537-6032 http://www.reiusa.net/ COPYRIGHT RESEARCH ELECTRONICS INTERNATIONAL
More informationPhotogrammetry Guide to Creating Digital 3D Models
Photogrammetry Guide to Creating Digital 3D Models By Emily Hauf (SUNY Geneseo) and the Paleontological Research Institution Guide Version 1.0 Part I - Photography Create studio setup with black cloth
More informationVividia HM-250 HDMI/LCD/USB/TV Portable. Digital Measurement Microscope
Vividia HM-250 HDMI/LCD/USB/TV Portable Digital Measurement Microscope Table of Contents 1. Precaution. 3 2. Preparation..5 Package Contents.. 6 Part Names & Functions..6 3. Specifications.... 7 4. Use
More informationAVT Model Tripod.
AVT Model Tripod www.ravelliphoto.com Product Overview: The Ravelli AVT professional tripod is a high performance, fluid drag tripod that provides smooth continuous drag control and operates on both pan
More informationUser Manual. Trinocular Stereo Microscope
User Manual Trinocular Stereo Microscope Model V434 XV434 Series MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 2 1. Component Illustration... 3 2. Installation 4 3. Operation....6
More informationUP! Quick Start Guide
Personal Portable 3D Printer UP! www.pp3dp.com 1. Assemble Printer Open the box, take out the printer and accessories. Assemble the parts contained in the box as following procedure: 1. Unscrew the M4
More informationDV2. Alignment Procedure. Install DV2 on Microscope NOTE: PLEASE READ THE ENTIRE PROCEDURE BEFORE YOU BEGIN ALIGNMENT OF THE DV2. Alignment Procedure
H I G H - P E R F O R M A N C E E M C C D & C C D C A M E R A S F O R L I F E S C I E N C E S DV2 This document provides a straightforward, step-by-step outline of the alignment procedure for the Photometrics
More informationMi50 Inverted Microscope User s Manual. Fein.
Mi50 Inverted Microscope User s Manual Fein info@feinoptic.com Mi50 Microscope Components Siedentopf Trinocular Head Stage Plate Mechanical Stage Locking Screw Hex Wrench Storage Beam Splitter Dust Protector
More informationRue Gurnigel 48 CH-2501 Bienne Tél.: +41 (0) Fax: +41 (0)
OPTISCHE MESSTECHNIK OPTICAL MEASURING SYSTEMS SYSTEMES DE MESURE OPTIQUE marcel - aubert - sa Rue Gurnigel 48 CH-2501 Bienne Tél.: +41 (0)32 365 51 31 Fax: +41 (0)32 365 76 20 E-mail: info@marcel-aubert-sa.ch
More informationPrecautions The XRD uses x-ray radiation. The instrument is designed to prevent radiation exposure.
1 Operating Procedure for X-Ray Diffraction Purpose and Scope This document describes the procedures and policies for using the MSE Bruker D8 Focus XRD The scope of this document is to establish user procedures.
More informationRichter Optica. Instructions for: S6-BL, S6-TS, S6-SPS, S6-ILST Models
Richter Optica info@richter-optica.com Instructions for: S6-BL, S6-TS, S6-SPS, S6-ILST Models Trinocular port for c-mount adapter Zoom Knob Focusing Holder Clamp Diopter adjustable eyepieces Magnification
More informationG500 SERIES MICROSCOPE OPERATION MANUAL For use with MFR# G502, G502T, G504, G504T, G505, G505T.
G500 SERIES MICROSCOPE OPERATION MANUAL For use with MFR# G502, G502T, G504, G504T, G505, G505T. United Products & Instruments, Inc. 182 Ridge Road, Suite E Dayton, NJ 08810 USA T. 732 274 1155; F. 732
More informationCarestream Vita user quick guide. Software version 3.2 From April 2012
Carestream Vita user quick guide Software version 3.2 From April 2012 1 Carestream Vita user quick guide Software version 3.2 from April 2012 1. To switch your Vita on Press the power button on the PC
More informationSP mm F/2.8 Di VC USD (Model A009)
SP 70-200mm F/2.8 Di VC USD (Model A009) Thank you for purchasing the Tamron lens as the latest addition to your photographic equipment. Before using your new lens, please read the contents of this Owner
More informationBiology 105 Laboratory 1: Microscope
Biology 105 Laboratory 1: Microscope A. The Compound Microscope The microscope is a delicate, precision instrument. It must be treated gently, as the slightest bump or jar may damage the alignment of its
More informationBIOLOGICAL MICROSCOPE
BIOLOGICAL MICROSCOPE Instruction Manual LBM-102 B "Please read this manual carefully before using the instrument" Labnics Equipment Thank You for Buying a LABNICS BIOLOGICAL MICROSCOPE! For your personal
More informationOMM200 and OMM200T. Metallurgical Microscope
OMM200 and OMM200T Metallurgical Microscope Instruction Manual (Please read the manual carefully before using the microscope) Contents Safety ----------------------------------------------------------------------------
More informationContents escope Software Guide Execute Software Calibration and Measurement escope Software Interface Instructions Main Interface
Contents escope Software Guide Execute Software...3 Capture Photo...3 Capture Video...3 Time-lapsed Picture / Video...4 Picture / Video Save as...4 Calibration and Measurement Calibration...5 Measurement...7
More informationWelcome 1. Precaution
Table of Contents EN Precaution....2 Preparation.. 4 Standard accessories....4 Parts Names & Functions...5 Computer System requirements.... 6 Technical Specifications 7 Install the software.. 7 Start Microscope.8
More informationR40POL Microscope User s Manual
R40POL Microscope User s Manual for R40POL and R40POL-RT Polarizing Microscopes Microscope Components: Trinocular Port Eyepieces Beam Splitter Analyzer Daylight Balancing Filter 12v 50w Halogen Reflected
More informationQuick Set-up Guide for the Biomedx Configured Olympus CX43
Quick Set-up Guide for the Biomedx Configured Olympus CX43 Page 1 of 15 Your microscope is supplied with an Olympus manual, please refer to that for more complete information on microscope operation. This
More informationimpact INSTRUCTIONS LiteTrek 4.0 Monolight lighting equipment and accessories
impact lighting equipment and accessories LiteTrek 4.0 Monolight INSTRUCTIONS Congratulations on your purchase of the Impact LiteTrek 4.0 Portable Monolight. Enjoy the many benefits of a portable strobe
More informationInstalling imac Intel 27" EMC 2390 Dual HDD or
Installing imac Intel 27" EMC 2390 Dual HDD or SSD Drive Installing a secondary HDD or SSD in the mid 2010 27" imac EMC 2390. Written By: Brett Hartt ifixit CC BY-NC-SA www.ifixit.com Page 1 of 23 INTRODUCTION
More informationD-SCOPE The indispensable analysis tool for all diamonds
www.hrdantwerp.be D-SCOPE The indispensable analysis tool for all diamonds The indispensable analysis tool for all diamonds HRD Antwerp has a long experience in research, examination and grading of diamonds.
More informationezimagex2 User s Guide Version 1.0
ezimagex2 User s Guide Version 1.0 Copyright and Trademark Information The products described in this document are copyrighted works of AVEN, Inc. 2015 AVEN, Inc. 4595 Platt Rd Ann Arbor, MI 48108 All
More informationZeiss AxioImager.Z2 Fluorescence Protocol
Zeiss AxioImager.Z2 Fluorescence Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge
More informationibook G4 14" 933 MHz-1.33 GHz Hard Drive
ibook G4 14" 933 MHz-1.33 GHz Hard Drive Replacement Written By: irobot ifixit CC BY-NC-SA www.ifixit.com Page 1 of 25 INTRODUCTION You can install hard drives up to 9.5mm thick. TOOLS: Coin (1) Phillips
More informationR I T. Title: Rudolf Ellipsometer Auto-EL IV. Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date:12/07/17 1 SCOPE
Title: Rudolf Ellipsometer Auto-EL IV Fabrication Laboratory Revision: C Rev Date:12/07/17 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the
More informationDesign and production of stereoscopic instruments. User manual for the 3-D macro lens model General. Exposure
Ing. J. de Wijs. Populierstraat 44, 4131 AR Vianen, the Netherlands Tel/Fax. +31 (0)347-372242 e-mail: info@dewijs-3d.com Website: www.dewijs-3d.com Design and production of stereoscopic instruments. Bank:
More informationComprehensive Scope 2. Instruction Manual T-19011C - Monocular T-19021C - Dual Viewing T-19031C - Binocular T-19041C - Trinocular
Comprehensive Scope 2 Instruction Manual T-19011C - Monocular T-19021C - Dual Viewing T-19031C - Binocular T-19041C - Trinocular Application Thank you for purchasing the new Ken-A-Vision Comprehensive
More informationOPTRON mobile. User Guide. magnified vision. Vers. 2.1AL Magnified Vision, Inc. 2013
OPTRON mobile User Guide Vers. 2.1AL Magnified Vision, Inc. 2013 magnified vision 2 Before operating this device, please read this user guide thoroughly and retain it for future reference. For questions,
More informationCanon EOS 6D Front and Top Body Panel Replacement
Canon EOS 6D Front and Top Body Panel Replacement Replace the front and top body panels of the EOS 6D. Written By: Justin Thomas ifixit CC BY-NC-SA www.ifixit.com Page 1 of 14 INTRODUCTION Use this guide
More informationUser Manual. Trinocular Stereo Microscope
User Manual Trinocular Stereo Microscope Model V436 XV436 Series MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 1 1. Component Illustration... 2 2. Installation 3 3. Operation....4
More informationFor documentation, both instruments are available with a trinocular head which permits mounting of a video or digital camera system.
PriorLux 1 1. Introduction The PriorLab and PriorLux upright compound microscopes are high quality instruments equipped with high resolution, chromatically corrected optics for excellent image quality.
More informationThe Microscope. Lab Exercise. Contents. Light Microscope. Objectives. Scanning Electron Microscope. Introduction. Transmission Electron Microscope
Lab Exercise The Microscope Contents Objectives 1 Introduction 1 Activity.1 Care of Microscopes 2 Activity.2 Parts of the Microscope 3 Activity.3 Viewing Images 4 Activity.4 Magnification of Images 4 Resutls
More informationWork with Shapes. Concepts CHAPTER. Concepts, page 3-1 Procedures, page 3-5
3 CHAPTER Revised: November 15, 2011 Concepts, page 3-1, page 3-5 Concepts The Shapes Tool is Versatile, page 3-2 Guidelines for Shapes, page 3-2 Visual Density Transparent, Translucent, or Opaque?, page
More informationFurther Information can be found at
Below is a step by step guide to assembling the Hurricane-Rig. Remember that this is a precision optical instrument. Excessive force can bend critical parts. If treated well it should give many years of
More informationVISIOFFICE 2 - USER MANUAL
VISIOFFICE 2 - USER MANUAL Summary A. Introduction...3 1. Main screen...3 a. Main Screen...3 b. Main Screen Menus...4 2. Client menu...4 a. New Client..4 b. Existing Client 4 3. Quick Jump...5 B. Creating
More informationOPERATING INSTRUCTIONS:
List OPERATING INSTRUCTIONS: Preparation For Installation, Product Identification Charging The Batteries ing And Adjusting The Lift Mounting Head s Brackets Mounting Clamps And Quick-Locking Arms Installation
More informationLynx Instruction Manual
Lynx Instruction Manual 1.0 INTRODUCTION 2.0 HARDWARE - 2.1 Slider 3.0 USER INTERFACE - 3.1 Main Menu - 3.2 Fire Test Shot - 3.3 Backlight - 3.4 Bluetooth - 3.5 Reset - 3.6 Motor Sleep - 3.7 Torque 4.0
More informationCustomizing FlipCharts Promethean Module 2 (ActivInspire)
Customizing FlipCharts Promethean Module 2 (ActivInspire) Section 1: Browsers The browsers (located on the left side of the flipchart) are menus for various functions. To view the browsers, click Main
More informationWelcome 1. Precaution
Table of Contents EN Precaution....2 Preparation.. 4 Standard accessories....4 Parts Names & Functions...5 Computer System requirements.... 6 Technical Specifications 7 Install the software.. 7 Start Microscope.8
More informationUniversity of Minnesota Nano Fabrication Center Standard Operating Procedure
Equipment Name: University of Minnesota Nano Fabrication Center Coral Name: hs-scope Revision Number: 1.5 Model: HS200A Revisionist: M. Fisher Location: Bay 1 Date: 9/12/2013 1 Description The Hyphenated
More informationProduct Overview. Features
APCF1 Model Tripod Product Overview The Ravelli APCF1 is a Professional Quality Carbon Fiber Tripod providing a solid base for high-end photographic equipment. This model is a mix of carbon fiber and magnesium
More informationSTEREO MICROSCOPE MODEL K400 / K500 / K700 INSTRUCTION MANUAL
STEREO MICROSCOPE MODEL K400 / K500 / K700 INSTRUCTION MANUAL Contents 1. Introduction...2 2. Nomenclature and Function....3 3. Assembly.....5 4. Operating the Microscope... 6 4.1 Interpupillary Distance
More informationRichter Optica. Instructions for Model: MDS2 Middle School Microscope
Richter Optica info@richter-optica.com Instructions for Model: MDS2 Middle School Microscope Eyepiece 360 Rotatable Head Revolving Nosepiece Carrying Handle Objectives Rack Stop Stage Stage Clips Disc
More informationMacro View All in One Scope Description. 2. Features. 3. Package Contents
Macro View All in One Scope 26700-104-00 1. Description Portable system ideal for inline use This self-contained video inspection system is a compact unit, that simplifies quality checks at any stage of
More informationClipArt and Image Files
ClipArt and Image Files Chapter 4 Adding pictures and graphics to our document not only breaks the monotony of text it can help convey the message quickly. Objectives In this section you will learn how
More informationWritten By: Allie Menard
Written By: Allie Menard ifixit CC BY-NC-SA www.ifixit.com Page 1 of 20 TOOLS: Paper Clip (1) Phillips #00 Screwdriver (1) Phillips #1 Screwdriver (1) Spudger (1) Tweezers (1) ifixit CC BY-NC-SA www.ifixit.com
More informationChapter 12 Notes: Optics
Chapter 12 Notes: Optics How can the paths traveled by light rays be rearranged in order to form images? In this chapter we will consider just one form of electromagnetic wave: visible light. We will be
More informationHow to add video effects
How to add video effects You can use effects to add a creative flair to your movie or to fix exposure or color problems, edit sound, or manipulate images. Adobe Premiere Elements comes with preset effects
More informationRichter Optica. Instructions for Model F-1 Compound Microscope
Richter Optica 6122 Innovation Way Carlsbad, California 92009 (800) 942-0528 US TOLL FREE (760) 438-0528 INTERNATIONAL Instructions for Model F-1 Compound Microscope Ocular Lens (eyepiece) Eyepiece Tube
More informationThis course will introduce the user the Windows 7 Operating System by covering the desktop, start button, task bar and the Computer section
This course will introduce the user the Windows 7 Operating System by covering the desktop, start button, task bar and the Computer section previously known as My Computer. The user will be introduced
More informationTechnical Data Sheet. PEAK Measuring Microscope 2008 Series
PEAK Measuring Microscope 2008 Series Pocket Stand Microscope 25x, 50x, 75x, 100x magnification 0.05-0.005 mm measuring scale Measure width, thickness, height & depth Peak 2008 Stand Microscopes are small,
More informationWide Guy: Inverted Widefield Microscope
Wide Guy: Inverted Widefield Microscope Kyle Marchuk Adam Fries Jordan Briscoe August 2017 Contents 1 Introduction 2 2 Initial Setup 3 2.1 Hardware Startup........................................... 3
More informationM20 Metallurgical Microscope User s Manual. Fein.
M20 Metallurgical Microscope User s Manual info@feinoptic.com M20 Microscope Components Lamp Housing C-Mount Adapter Eyepieces Analyzer Objective Lens Coarse Focus Mechanical Stage Stage X-Y Control Fine
More informationMacBook Core 2 Duo Clutch Cover
MacBook Core 2 Duo Clutch Cover Replacement Replace the clutch cover on your MacBook Core 2 Duo. Written By: Ben Eisenman ifixit CC BY-NC-SA www.ifixit.com Page 1 of 29 INTRODUCTION Replace the curved
More informationJEOL CarryScope SEM Revision /07/17 Page 1 of 7. JEOL CarryScope SEM
Page 1 of 7 JEOL CarryScope SEM The JEOL CarryScope is a compact and portable SEM that utilizes a standard tungsten filament. It is to be used for inspecting and measuring samples processed in the NRF
More informationXnView 1.9. a ZOOMERS guide. Introduction...2 Browser Mode... 5 Image View Mode...15 Printing Image Editing...28 Configuration...
XnView 1.9 a ZOOMERS guide Introduction...2 Browser Mode... 5 Image View Mode...15 Printing... 22 Image Editing...28 Configuration... 36 Written by Chorlton Workshop for hsbp Introduction This is a guide
More informationMotic Images Plus 3.0 ML Software. Windows OS User Manual
Motic Images Plus 3.0 ML Software Windows OS User Manual Motic Images Plus 3.0 ML Software Windows OS User Manual CONTENTS (Linked) Introduction 05 Menus and tools 05 File 06 New 06 Open 07 Save 07 Save
More informationTable of Contents. Revu ipad. v3.6. Navigation. Document Manager. File Access. Markups. Signature Tool. Field Verification Measurements
Table of Contents Navigation Document Manager File Access Markups Signature Tool Field Verification Measurements Editing Properties Tool Sets & the Tool Chest Markups List Forms Studio Sessions Studio
More informationABM's High Resolution Mask Aligner Features:
ABM's High Resolution Mask Aligner is a very versatile instrument with interchangeable light sources which allow Near-UV (405-365 nm) as well as Mid- and Deep-UV (254 nm, 220 nm) exposures in proximity
More informationIntroduction to Photography
Topic 5 - Basic DSLR Settings Learning Outcomes In this topic, you will learn about the functions of the buttons on the top, the side and the back of your camera. This will give you the platform to begin
More informationQuick Start Guide. Basic set-up for your Axis360 system
Quick Start Guide Basic set-up for your Axis360 system Table of Contents 1 Setting up Slide slider assembly attach belt to cart attach ballhead to cart connect motor to controller attach slider to tripod(s)
More informationOlympus DeltaVision Microscope Start-Up and Shut-Down Instructions.
DeltaVision Instructions: 1. Start-up (Olympus DV) 2. Basic operation (Olympus DV) 2.1 set-up 2.2 designing and running an experiment 3. Transferring files via FTP 4. Deconvolving files 5. File conversion
More informationTS-E50mm f/2.8l MACRO TS-E90mm f/2.8l MACRO TS-E135mm f/4l MACRO
TS-E50mm f/2.8l MACRO TS-E90mm f/2.8l MACRO TS-E135mm f/4l MACRO ENG Instructions Thank you for purchasing a Canon product. The Canon TS-E50mm f/2.8l MACRO, Canon TS-E90mm f/2.8l MACRO, and Canon TS-E135mm
More informationDigital Microscopes Zoomy 2.0 Digital Microscope What software/devices will I need to use a Zoomy? Connecting Zoomy 2.0
Digital Microscopes Zoomy 2.0 Digital Microscope Zoomy 2.0 is a handheld digital microscope that children can use to examine objects. Zoomy plugs into a computer using a built in USB cable. The computer
More informationMacBook Core 2 Duo Hard Drive Connector
MacBook Core 2 Duo Hard Drive Connector Replacement Written By: irobot ifixit CC BY-NC-SA www.ifixit.com Page 1 of 13 INTRODUCTION How to remove the hard drive connector. TOOLS: Coin (1) Phillips #0 Screwdriver
More informationContents. 1. Getting Started Preparations Preparing 3D Data Creating monofab Player AM Data Preparing the ARM-10 15
Printing Guide Contents 1. Getting Started 1 2. Preparations 2 3. Preparing 3D Data 3 4. Creating monofab Player AM Data 4 5. Preparing the ARM-10 15 6. Printing 19 1. Getting Started What is layered projection
More informationVXvue User Manual (For Human Use)
VXvue User Manual (For Human Use) Page 2 of 90 Revision History Version Date Description 1.0 2012-03-20 Initial Release Page 3 of 90 Contents Safety and Regulatory... 8 Safety Notice... 8 1. Introduction...
More informationMac mini Mid 2010 IR Sensor Replacement
Replace a broken IR sensor/sleep LED. Written By: Andrew Bookholt ifixit CC BY-NC-SA www.ifixit.com Page 1 of 18 INTRODUCTION Use this guide to replace your mini's IR sensor/sleep LED assembly. Replacing
More informationI N S T R U C T I O N S F O R. For Applied Photography in Science and Industry.
ILFORD I N S T R U C T I O N S F O R CAMERAS For Applied Photography in Science and Industry. ILFORD K.I. MONOBAR CAMERA 2 The llford Monobar Camera Type U represents a major advance in the design of 35
More informationHigh definition digital microscope. visioneng.us/lynxevo
visioneng.us/lynxevo High definition digital microscope Exceptional high resolution 1080p/60fps image quality Intuitive image capture and documentation Stand alone, wireless or PC connectivity High quality
More informationAdobe Premiere Pro CC 2015 Tutorial
Adobe Premiere Pro CC 2015 Tutorial Film/Lit--Yee GETTING STARTED Adobe Premiere Pro CC is a video layout software that can be used to create videos as well as manipulate video and audio files. Whether
More informationChapter 1 General Information
Chapter 1 General Information System Description Chapter 1 General Information 1.0 System Description The Affirm attaches to the Selenia Dimensions. A biopsy device attaches to the Affirm. X- and Y-axes
More informationPH 481/581 Physical Optics Winter 2018
PH 481/581 Physical Optics Winter 2018 Laboratory #1 Week of January 15 Read: Section 5.2 (pp.151-175) of "Optics" by Hecht Do: 1. Experiment I.1: Thin Lenses 2. Experiment I.2: Alignment Project 3. Experiment
More informationMacBook Pro 17" Models A1151 A1212 A1229 and A1261 LCD Replacement
MacBook Pro 17" Models A1151 A1212 A1229 and A1261 LCD Replacement Escrito por: Andrew Bookholt ifixit CC BY-NC-SA es.ifixit.com Página 1 de 21 INTRODUCCIÓN Use this guide to replace just the LCD rather
More informationkena Instruction Manual T-1050
kena Instruction Manual T-1050 2 Table of Contents Introduction / Diagram 4 Package Contents 5 Operational Requirements 5 Optional Accessories to Order 5 Specifications 6 Quick Start Guide 7 1. Install
More informationFast and Reliable Inspection of Printed Circuit Boards with Digital Microscopy for Quality Control, Failure Analysis, and Research and Development
From Eye to Insight MICROELECTRONICS TECHNICAL REPORT PART 2 Fast and Reliable Inspection of Printed Circuit Boards with Digital Microscopy for Quality Control, Failure Analysis, and Research and Development
More informationibook G4 12" 800 MHz-1.2 GHz RJ-11 Board
ibook G4 12" 800 MHz-1.2 GHz RJ-11 Board Replacement Written By: irobot ifixit CC BY-NC-SA www.ifixit.com Page 1 of 24 INTRODUCTION The standard telephone jack port for connecting to the 56k internal modem.
More information