1.0 Introduction. 1.1 General

Size: px
Start display at page:

Download "1.0 Introduction. 1.1 General"

Transcription

1

2 1.0 Introduction The Linac Coherent Light Source (LCLS) project at the Stanford Linear Accelerator Center (SLAC) is intended to create an x-ray free electron laser source of unprecedented brightness and short-time pulse duration. At a point after the electron beam has been diverted from the path of the photon beam, a series of stoppers must be remotely inserted into the photon beam path to allow entry into the Front End Enclosure and down-beam areas. The stoppers must be capable of stopping both the photon and electron beams, in case the electron-diverting magnets fail. 1.1 General The photon/electron stoppers are part of the Personnel Protection System (PPS). The primary control of the stoppers will be through the PPS. Status of the stopper position and Burn-Through-Monitor (BTM) integrity may be provided to other controls monitoring systems, such as the Beam Containment System (BCS). 1.2 Scope The parameters required to design and fabricate the photon/electron stoppers must be established to meet requirements of the PPS and BCS systems. The design parameters described herein only apply to the LCLS photon/electron stoppers located in the electron beam dump area; ST1and ST Beam Line Device Two Photon/Electron Stoppers are to be located in the LCLS E-Beam Dump enclosure. One stopper is located between the main dump line electromagnet string and the safety dump line permanent magnet bend string. The second stopper is located between the safety magnet permanent magnet bend string and the Wall 1 Collimator. The primary function of the Photon/Electron stoppers is to stop photons emitted from the LCLS FEL Undulator during access to the FEE, while electrons continue to be delivered to the main dump. The stoppers must stop the spontaneous radiated photons, FEL photons, and Bremsstrahlung γ rays. 2 of 9 that this is the correct version prior to use.

3 Photon damage simulations and testing of materials has been performed by LLNL. The results indicate that 10mm thick Boron Carbide (B 4 C) on the front face of the photon/electron stopper will stop the FEL photons without experiencing damage itself, and will reduce the total transmitted pulse energy to non-damaging levels for any downstream component. The secondary function of the photon/electron stoppers is to stop electrons in case the electromagnet main dump bend magnets fail while the FEE is in access. (In the case of ST2, both the electromagnet main dump bend magnets and the permanent magnet safety dump bend strings must fail for ST2 to see electrons.) Safety system function information may be found in PRD LCLS Personnel Protection System Requirements. Many stoppers have been built at SLAC, and are still in use in every beam line. For ease of maintenance and reliability, the vertical actuator design originally built for the PEPII injection beamlines will be used. The stopper block however, will be re-designed for the specific needs in the photon line in the E-Dump enclosure. The actuator is shown in Figure 1. Figure 1. Stopper actuator mechanism 3 of 9 that this is the correct version prior to use.

4 The stopper block will be made in multiple segments and with two BTM's to meet the requirement to detect the unlikely event of a burn-through caused by either the photon beam or electron beam. The first absorber segment will be B 4 C, 10mm thick, to stop the FEL pulse. A thick titanium-alloy wafer will spoil the electron beam to decrease the density of energy deposition in the following copper segment. A tungsten segment will follow the copper segment to further attenuate radiation escaping the downstream face of the copper segment. (see Figure 1). The stopper block will be water-cooled. A photon BTM will be placed behind the B 4 C and titanium to detect failure of the photon stopper element. An electron BTM will be placed downstream of the copper segment, upstream of the tungsten segment. The maximum energy deposition is in the bulk of the copper. The electron beam must burn through the copper to cause failure of the electron BTM. The electron BTM must fail prior to appreciable melting of the tungsten element. Figure 1. Stopper Block Material Stack 4 of 9 that this is the correct version prior to use.

5 The BTM's will be pressurized with ultrahigh vacuum (UHV)-compatible, filtered, dry air. Upon failure, pressure loss in the BTM will be detected using an electronic pressure sensor. The stopper block will be water-cooled. The device will only see the electron beam in accidental exposure or low-power conditions for testing performed by RP; however, the stopper will be designed to accept full beam power. 3.0 Device Control The stopper is actuated via a solenoid valve and must have a pair each of in and out limit switches. It will be equipped with a panel-mount cable connector. Specific cable connector types must be negotiated between mechanical and controls engineering and recorded in the cable plant database. The control hardware should be the same as for other LCLS in/out air-actuated devices, with the addition of BTM-failure detection. Indications of in/out and BTM status must be available to the PPS. The pressurized air electron and photon BTM's will be instrumented with an electronic pressure gauge. The signal from the pressure gauge must be made available to the PPS System. In addition to pressure monitoring, BCS Protection Ion Chambers (PIC's) will be placed at each stopper, as required in the SLAC Radiation Safety Systems Technical Basis Document. The PIC's will shut-off the electron beam to protect the stoppers from damage and thereby protect personnel (a Machine Protection System [MPS]-like function performed by the BCS for PPS stoppers). 4.0 Design Parameters The basic design parameters of the stopper are shown in Tables 1 and 2. Electron beam parameters are taken from PRD Electron Beam Power Absorbing Device Performance Requirements, PRD Insertable Beam Dumps in the Linac and LTU, and PRD Electron Dump-Line Requirements. The maximum single-pulse temperature rise was calculated for ST1 by the Radiation Physics group using conservative input parameters of a φ20 µm, 2 kw beam (120 Hz). Scaling the RP data to φ40 µm and 5kW yields the values indicated in Table 1. Scaling was 5 of 9 that this is the correct version prior to use.

6 achieved by linearly scaling the deposited energy per radial bin by 5 kw/2 kw and also scaling the deposited energy by the ratio of the areas of the increased radial bin to the original bin size. The average temperature rise in the stopper block materials at 120 Hz is a maximum of 4 C in the titanium. The titanium material is an alloy; Ti-6AL-4V. The tungsten material is a machinable, sintered tungsten alloy suitable for UHV applications. Electron Beam Parameter Value Units Horizontal RMS Beam Size 40 µm Vertical RMS Beam Size 40 µm Longitudinal RMS Beam Size 20 µm Nominal electron energy 4.3 to 14 GeV Bunch charge range 0.2 to 1 nc Maximum beam rate 120 Hz Max. average beam power 5 kw Max. reasonable insertion/extraction time 5 sec Table 1: Electron Beam Parameters for the Photon/Electron Stoppers. Stopper Stack Materials Value Units Photon Absorber Material B 4 C - Photon Absorber Length 10 mm Electron Absorber Materials Ti, W & Cu - Total Electron Absorber Length (Radiation Lengths) 25 X o Maximum Average Temperature Rise in Ti 4 C Maximum Single Pulse Temperature Rise in B4C 114 C Maximum Single Pulse Temperature Rise in Ti 176 C Maximum Single Pulse Temperature Rise in Cu 60 C Table 2: Stopper Block Material Parameters. 5.0 Procurement/Fabrication/Assembly The basis of estimate indicates that the stopper mechanical parts will be machined by an outside vendor and the final assembly will be performed at SLAC. Alternatives to the basis are allowed to reduce schedule and/or cost. The stopper controls hardware is anticipated to be identical to stoppers installed in the upstream areas of LCLS, with the addition of BTM failure- 6 of 9 that this is the correct version prior to use.

7 detection hardware. 5.1 Materials All parts to be installed inside the vacuum enclosure must conform to the SLAC vacuum standard for machined components, including materials requirements including materials source data and machining lubricant restrictions. See the latest revision of SLAC FP for details. All test results shall be recorded in the stopper traveler. 5.2 Machined Components All vacuum enclosure parts must be leak checked prior to assembly. All invacuum parts must be fabricated in accordance to SLAC vacuum standard SLAC FP All machined parts must be inspected for conformance to drawing dimensions and tolerances. Acceptance of parts is to be recorded on the component traveler. 5.2 Purchased Components All purchased components, including controls hardware, must conform to SLAC safety policies and practices. Controls hardware must be acceptablyrated by a nationally-recognized testing laboratory or pass internal SLAC electrical safety inspections. 5.3 Vacuum Assembly Assembly of in-vacuum parts must be performed in a SLAC Clean Room. Leak testing of components and subassemblies shall be performed in such a manner as to identify leaks at as early a time as possible. 5.3 Alignment Fiducials Included in the design of the stopper vacuum chambers are provisions for tooling ball sockets. Each stopper vacuum chamber must have the tooling ball sockets fiducialized on a CMM with respect to the vacuum chamber beam axis. 5.4 Identification 7 of 9 that this is the correct version prior to use.

8 Each stopper assembly shall have affixed to the outer surface on both sides The identifier LCLS Photon/Electron Stopper ST-<X> where <X> is the unit number. Total weight of the stopper assembly in pounds. Each stopper shall also have a white, painted arrow on each side, which points in the direction shown in the drawing, with Beam Direction lettered on the shaft of the arrow (with at least 1 cm high characters) as illustrated in Figure 3. Beam Direction Figure 3: Beam direction arrow. 6.0 Tests and Measurements Tests and measurements must be performed to assure that the installed stopper system will perform as designed. Tests and measurements shall include part and assembly qualification, electrical safety inspection, actuation testing, limit switch tests, BTM pressure tests, BTM pressure signal verification, leak checks, assembly fiducialization and vacuum qualification bake-out. Tests shall be performed in the laboratory to qualify the device prior to installation. Additional tests shall be performed after installation to qualify the system prior to delivery of electron or photon beams to the installed region. All tests and measurements shall be recorded on a device traveler which corresponds to the unique device serial number. 7.0 Installation Installation of the Photon/Electron Stoppers and control system shall conform to the Davis Bacon Act where required. Precision alignment of the beam line device is required; to be performed by the SLAC Metrology group. 8 of 9 that this is the correct version prior to use.

9 Interconnection and testing of controls shall be performed following vacuum interconnection. Additional field testing shall be performed by SLAC Radiation Physics. 8.0 Applicable Standards and Codes Design, fabrication, testing and acceptance must be in conformance to the following codes and SLAC standards. 1 SLAC SLAC Guidelines for Operations SLAC-I SLAC ES&H Manual 3 SLAC Technical Specification Fabrication of UHV Components FP NEMA National Electrical Manufactures Association: Standard 250 Enclosures for Electrical Equipment (1000 Volts Max.), section SLAC SLAC Radiation Safety Systems Technical Basis Document 9 of 9 that this is the correct version prior to use.

Engineering Specifications for the X-Ray Focusing Lens System

Engineering Specifications for the X-Ray Focusing Lens System ENGINEERING SPECIFICATION DOCUMENT (ESD) Doc. No. SP-391-000-58 R0 LUSI SUB-SYSTEM Diagnostics Engineering Specifications for the X-Ray Focusing Lens System Prepared by: Marc Campell Signature Date Design

More information

Hutch Protection System (HPS) Stopper Acceptance

Hutch Protection System (HPS) Stopper Acceptance Printed copies of this document must be verified as being current prior to use by checking the effective date on the Systems Document Master List. Table of Contents 1.0 Introduction... 1 2.0 Applicability...

More information

Ray-Trace Study for the LCLS-II Safety Dump Lines

Ray-Trace Study for the LCLS-II Safety Dump Lines LCLS-TN-12-6 Ray-Trace Study for the LCLS-II Safety Dump Lines J. Welch October 2, 2012 Contents 1 Summary 1 2 Safety Dump Lines in LCLS-II 1 3 Fundmental Assumptions 2 4 Analysis 2 4.1 Effect of Earth

More information

BFW Traveler: Assembly and Alignment Procedure Serial Number: BFW # / VC # / WC # / Base #

BFW Traveler: Assembly and Alignment Procedure Serial Number: BFW # / VC # / WC # / Base # BFW Traveler: Assembly and Alignment Procedure Serial Number: BFW # / VC # / WC # / Base # NOTE - This document: 1. records the assembly and alignment steps for the LCLS Undulator Beam Finder Wire (BFW).

More information

Three-Dimensional Laser Scanner. Field Evaluation Specifications

Three-Dimensional Laser Scanner. Field Evaluation Specifications Stanford University June 27, 2004 Stanford Linear Accelerator Center P.O. Box 20450 Stanford, California 94309, USA Three-Dimensional Laser Scanner Field Evaluation Specifications Metrology Department

More information

Beam Position Monitor For High Energy Particle Beams

Beam Position Monitor For High Energy Particle Beams Beam Position Monitor For High Energy Particle Beams Features 16 by 16 strip ionization chamber sensor High vacuum compatible Small insertion length (152.4 mm) Robust thin steel foil vacuum windows Very

More information

LUSI SUB-SYSTEM XCS Physics Requirements for the LUSI Large Offset Monochromator. Doc. No. SP R0

LUSI SUB-SYSTEM XCS Physics Requirements for the LUSI Large Offset Monochromator. Doc. No. SP R0 PHYSICS REQUIREMENT DOCUMENT (PRD) Doc. No. SP-391-00-16 R0 LUSI SUB-SYSTEM XCS Physics Requirements for the Aymeric Robert Author, LUSI Scientist Signature Date David Fritz LUSI Scientist Signature Date

More information

Novel Magnetic Field Mapping Technology for Small and Closed Aperture Undulators

Novel Magnetic Field Mapping Technology for Small and Closed Aperture Undulators Novel Magnetic Field Mapping Technology for Small and Closed Aperture Undulators Erik Wallen and Hyun-Wook Kim 06.06.2017 Outline Introduction - Measurement systems at LBNL - Activities at LBNL - Need

More information

12 Alignment. Technical Synopsis

12 Alignment. Technical Synopsis 12 Alignment Technical Synopsis This section describes the procedures and methods used to position the LCLS-II components with their required accuracy. Most of the alignment requirements for the injector,

More information

300W 1.5kW, QUBE Fiber Lasers

300W 1.5kW, QUBE Fiber Lasers Technical Specification 300W 1.5kW, QUBE Fiber Lasers CONTENTS 1.1 SCOPE 2 1.2 OPTICAL SPECIFICATION 2 1.3 BEAM DELIVERY FIBER SPECIFICATION 3 1.4 ALIGNMENT LASER 3 1.5 POWER DISTRIBUTION 3 1.6 WATER COOLING

More information

ph fax Deming Way Middleton WI USA

ph fax Deming Way Middleton WI USA www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 3120 Deming Way Middleton WI 53562-1461 USA 2007 Standard Imaging, Inc. 1239-21 (03.07) beam qa LINAC AND TOMOTHERAPY QA The NEW

More information

FC 1800 #1 Operation Manual University of Notre Dame

FC 1800 #1 Operation Manual University of Notre Dame \ FC 1800 #1 Operation Manual University of Notre Dame For Training Please contact: Dave Heemstra 8 June 2015 Page 1 of 12 General Precautions Contacts For problems, clarification of procedures, or general

More information

NeXLev Product Specification TB2144

NeXLev Product Specification TB2144 NeXLev Product Specification TB2144 Revision B Specification Revision Status Revision SCR No. Description Initial Date - S0161 New Release DM 4/5/2006 A S1357 Update copper material callout, DM 8/31/09

More information

Manual Control Unit GFC 32

Manual Control Unit GFC 32 Manual Control Unit 1400004_EN/05.2017 Index 1. Main features 3 2. Technical features 3 3. Installation guidelines 4 4. Preliminary checks 5 5. Electrical connections 5 6. Settings 6 7. Remote control

More information

Manual Control Unit GFC 16

Manual Control Unit GFC 16 Manual Control Unit 1400003_EN/04.2017 Index 1. Main features 3 2. Tekniske data 3 3. Installation guidelines 4 4. Preliminary checks 5 5. Electrical connections 5 6. Settings 5 7. Remote 7 8. Shut down

More information

LCLS Undulator Quadrupole Fiducialization Plan

LCLS Undulator Quadrupole Fiducialization Plan LCLS-TN-07-7 LCLS Undulator Quadrupole Fiducialization Plan Zachary Wolf, Michael Levashov, Eric Lundahl, Ed Reese, Catherine LeCocq, Robert Ruland Stanford Linear Accelerator Center August 14, 2007 Abstract

More information

Governing Standards. All wiring and components shall meet the applicable requirements of the NEC.

Governing Standards. All wiring and components shall meet the applicable requirements of the NEC. 1.1 SCOPE. This section covers the furnishing and installation of Video Surveillance Systems (VSS), including Video Management System (VMS) software, Cameras, and all associated equipment, devices, and

More information

Front Load Single Sensor

Front Load Single Sensor INFICON Front Load Single crystal sensors offer proven reliability and durability and have the best thermal stability of any sensor head on the market. The front load design allows for easy insertion of

More information

Manual Simatek Control Unit GFC 32

Manual Simatek Control Unit GFC 32 Manual Simatek Control Unit GFC 32 Original instructions 1400004_GB Ver. 2014.01.09 Index 1. Main features 3 2. Technical features 3 3. Installation guidelines 4 4. Preliminary checks 5 5. Electrical connections

More information

UW Purchase Specification Section Central Cooling Water Meter Page 1 of 6

UW Purchase Specification Section Central Cooling Water Meter Page 1 of 6 Page 1 of 6 PART 1 GENERAL 1.01 DESCRIPTION A. Purpose 1. This section covers central cooling water meters for use in the Owner's central cooling water systems. 1.02 QUALIFICATIONS A. Approved manufacturers

More information

APPLICATION SPECIFICATION TITLE PAGE REVISION

APPLICATION SPECIFICATION TITLE PAGE REVISION 1 of 10 E 1 OBJECTIVE...2 2 SCOPE...2 3 GENERAL...2 4 DRAWINGS AND APPLICABLE DOCUMENTS...2 5 APPLICATION REQUIREMENTS...2 5.1 PCB layout... 2 5.2 Keep-out zone... 2 5.3 Packaging... 2 5.4 Multiple connector

More information

Radiation Protection in Experimental Hutches at European XFEL. Safety and Radiation Protection Group

Radiation Protection in Experimental Hutches at European XFEL. Safety and Radiation Protection Group Radiation Protection in Experimental Hutches at European XFEL Safety and Radiation Protection Group Topics 2 Typical Hutch Layout at XFEL Shielding Calculations for Radiation Hutches Design of Optical

More information

Front Load Dual Sensor

Front Load Dual Sensor INFICON Front Load Dual crystal sensors offer proven reliability and durability and have the best thermal stability of any sensor head on the market. The dual sensor provides a backup crystal and is essential

More information

EU Contribution to ITER Diagnostics. Glenn Counsell Project Manager Diagnostics Fusion for Energy

EU Contribution to ITER Diagnostics. Glenn Counsell Project Manager Diagnostics Fusion for Energy EU Contribution to ITER Diagnostics Glenn Counsell Project Manager Diagnostics Fusion for Energy 1 DIAGNOSTICS PROCUREMENTS EU SCOPE OF SUPPLY EU Industry EFLs EU is the largest supplier of Diagnostics

More information

COORDINATE MEASUREMENT MACHINES AS AN ALIGNMENT TOOL *

COORDINATE MEASUREMENT MACHINES AS AN ALIGNMENT TOOL * .. SLAC-PUB-5506 March 1991 (W COORDINATE MEASUREMENT MACHINES AS AN ALIGNMENT TOOL * Bernd T. Wand Stanford Linear Accelerator Center Stanford University Stanford,CA 94309 1. ABSTRACT In February of 1990

More information

1. INTRODUCTION 2. REFERENCE MATERIAL. 1 of 8

1. INTRODUCTION 2. REFERENCE MATERIAL. 1 of 8 Application Specification 67 Positions 0.5 Pitch M.2(NGFF) MINICARD 114-115006 19May14 Rev. B NOTE All numerical values are in metric units [with U.S. customary units in brackets]. Dimensions are in millimeters.

More information

SSPP-TEV-035 HITCHHIKER ADVANCED CARRIER EQUIPMENT RANDOM VIBRATION TEST PLAN BASIC RELEASE MARCH 1997 GODDARD SPACE FLIGHT CENTER GREENBELT, MARYLAND

SSPP-TEV-035 HITCHHIKER ADVANCED CARRIER EQUIPMENT RANDOM VIBRATION TEST PLAN BASIC RELEASE MARCH 1997 GODDARD SPACE FLIGHT CENTER GREENBELT, MARYLAND HITCHHIKER ADVANCED CARRIER EQUIPMENT RANDOM VIBRATION TEST PLAN SSPP-TEV-035 BASIC RELEASE MARCH 1997 GODDARD SPACE FLIGHT CENTER GREENBELT, MARYLAND PREPARED BY: 5050 POWDER MILL ROAD BELTSVILLE, MD

More information

3a. IEM Indoor Metal Clad Medium Voltage Switchgear 25KV a. Section INDOOR METAL CLAD MEDIUM VOLTAGE SWTICHGEAR (Std.

3a. IEM Indoor Metal Clad Medium Voltage Switchgear 25KV a. Section INDOOR METAL CLAD MEDIUM VOLTAGE SWTICHGEAR (Std. 3a. IEM Indoor Metal Clad Medium Voltage Switchgear 25KV 16346-1 3a. Section 16346 INDOOR METAL CLAD MEDIUM VOLTAGE SWTICHGEAR (Std. Relays) Part 1 General 1.1 CONDITIONS AND REQUIREMENTS: A. Refer to

More information

2. INITIAL ALIGNMENT 2.1 Tolerances The FFTB is composed of four straight sections as shown in fig. 1. FFTB composed of 4 Straight Sections

2. INITIAL ALIGNMENT 2.1 Tolerances The FFTB is composed of four straight sections as shown in fig. 1. FFTB composed of 4 Straight Sections SLAC-PUB-11424 A DYNAMIC ALIGNMENT SYSTEM FOR THE FINAL FOCUS TEST BEAM* R.E. Ruland, V.E. Bressler, G. Fischer, D. Plouffe Stanford Linear Accelerator Center Stanford University 0. ABSTRACT The Final

More information

TiTan Ion pumps (MINI - 75S) (100L LX) (150TV - 600TV)

TiTan Ion pumps (MINI - 75S) (100L LX) (150TV - 600TV) Titan ion pumps TiTan Ion pumps TiTan Ion pumps Ion pumps are used in a wide variety of high and ultra-high vacuum (UHV) environments. They can reach the lowest possible vacuum for an economical cost.

More information

XES PCDS to LUSI XCS Instrument ICD DRAFT

XES PCDS to LUSI XCS Instrument ICD DRAFT INTERFACE CONTROL DOCUMENT (ICD) Doc. No. SP-391-001-25 R0 LUSI SUB-SYSTEM XCS XES PCDS to LUSI XCS Instrument ICD DRAFT Perry Anthony Author Signature Date Gunther Haller PCDS Manager Signature Date Aymeric

More information

Engineering Specification Doc. for the Primary, Guard and Monochromatic Slit Systems

Engineering Specification Doc. for the Primary, Guard and Monochromatic Slit Systems ENGINEERING SPECIFICATION DOCUMENT (ESD) Doc. No. SP-391-000-54 R0 LUSI SUB-SYSTEM Diagnostics Engineering Specification Doc. for the Primary, Guard and Monochromatic Slit Systems Prepared by: Richard

More information

PRODUCT SPECIFCATION GS MINI-SAS HD External Cable to Board Connector System. 1 of 14 F. Roger Cheng TYPE NUMBER

PRODUCT SPECIFCATION GS MINI-SAS HD External Cable to Board Connector System. 1 of 14 F. Roger Cheng TYPE NUMBER 1 of 14 F TABLE OF CONTENTS 1.0 OBJECTIVE... 3 2.0 SCOPE... 3 3.0 APPLICABLE DOCUMENTS... 3 3.1 FCI SPECIFICATIONS... 3 3.2 OTHER STANDARDS AND SPECIFICATIONS... 3 3.3 FCI PRODUCT QUALIFICATION TEST REPORTS...

More information

PRELIMINARY APPLICATION SPECIFICATION

PRELIMINARY APPLICATION SPECIFICATION SEARAY BOARD TO BOARD INTERCONNECT SYSTEM 4970 / 466 SMT Plug Connector (shown with kapton pad for pick and place) 4971 / 467 SMT Receptacle Connector (shown with kapton pad for pick and place) SEARAY

More information

Light. Form of Electromagnetic Energy Only part of Electromagnetic Spectrum that we can really see

Light. Form of Electromagnetic Energy Only part of Electromagnetic Spectrum that we can really see Light Form of Electromagnetic Energy Only part of Electromagnetic Spectrum that we can really see Facts About Light The speed of light, c, is constant in a vacuum. Light can be: REFLECTED ABSORBED REFRACTED

More information

METRIC MIL-C-83522/16B w/amendment 1 17 AUGUST 2004 SUPERSEDING MIL-C-83522/16B 11 May 1995 MILITARY SPECIFICATION SHEET

METRIC MIL-C-83522/16B w/amendment 1 17 AUGUST 2004 SUPERSEDING MIL-C-83522/16B 11 May 1995 MILITARY SPECIFICATION SHEET METRIC MIL-C-83522/16B 17 AUGUST 2004 SUPERSEDING MIL-C-83522/16B 11 May 1995 MILITARY SPECIFICATION SHEET CONNECTOR, FIBER OPTIC, SINGLE TERMINUS, PLUG, BAYONET COUPLING (ST STYLE), 2.5 MILLIMETERS DIAMETER

More information

DETAIL SPECIFICATION SHEET CONNECTOR, FIBER OPTIC, SINGLE TERMINUS, PLUG, BAYONET COUPLING (ST STYLE), 2.5 MILLIMETERS DIAMETER FERRULE, EPOXY

DETAIL SPECIFICATION SHEET CONNECTOR, FIBER OPTIC, SINGLE TERMINUS, PLUG, BAYONET COUPLING (ST STYLE), 2.5 MILLIMETERS DIAMETER FERRULE, EPOXY METRIC MIL-DTL-83522/16C 28 June 2013 SUPERSEDING MIL-C-83522/16B 11 May 1995 DETAIL SPECIFICATION SHEET CONNECTOR, FIBER OPTIC, SINGLE TERMINUS, PLUG, BAYONET COUPLING (ST STYLE), 2.5 MILLIMETERS DIAMETER

More information

Insertion Device Alignment for the Diamond Light Source

Insertion Device Alignment for the Diamond Light Source Insertion Device Alignment for the Diamond Light Source A. Mariani Diamond Light Source, Chilton, Didcot, Oxon. OX11 0DE, UK This paper covers the survey and alignment techniques selected for the build

More information

Automotive Electronics Council Component Technical Committee

Automotive Electronics Council Component Technical Committee ATTACHMENT 11 CHARGED DEVICE MODEL (CDM) ELECTROSTATIC DISCHARGE TEST Acknowledgment Any document involving a complex technology brings together experience and skills from many sources. The Automotive

More information

IT3D(M)-300S-BGA (37) Lead-Free THERMAL CYCLING TEST REPORT

IT3D(M)-300S-BGA (37) Lead-Free THERMAL CYCLING TEST REPORT TR066E-018 ITD(M)-00S-BGA (7) Lead-Free THERMAL CYCLING TEST REPORT - Post 6000 Cycles Report - APPROVED TY.ARAI Nov. 18, 009 CHECKED TM.MATSUO Nov. 17, 009 CHARGED TY.TAKADA Nov. 17, 009 HIROSE ELECTRIC

More information

Lyte-MV Digital Userguide Red & IR Versions

Lyte-MV Digital Userguide Red & IR Versions Product Overview Lyte-MV Digital Userguide Red & IR Versions Thanks for purchasing the Lyte-MV, a laser diode module emitting a uniform (non-gaussian) and welldefined line with a user-adjustable focus.

More information

IMPROVEMENT OF SCREEN MONITOR WITH SUPPRESSION OF COHERENT-OTR FOR SACLA

IMPROVEMENT OF SCREEN MONITOR WITH SUPPRESSION OF COHERENT-OTR FOR SACLA IMPROVEMENT OF SCREEN MONITOR WITH SUPPRESSION OF COHERENT-OTR FOR SACLA S. Matsubara 1, H. Maesaka 1,2, S. Inoue 3, Y. Otake 1,2 1: JASRI, 2: RIKEN SPring-8 Center, 3: SPring-8 Service Co. Ltd. Outline

More information

Front Load Single Sensor

Front Load Single Sensor INFICON Front Load Single crystal sensors offer proven reliability and durability and have the best thermal stability of any sensor head on the market. The front load design allows for easy insertion of

More information

Option G 1: Refraction

Option G 1: Refraction Name: Date: Option G 1: Refraction 1. The table below relates to the electromagnetic spectrum. Complete the table by stating the name of the region of the spectrum and the name of a possible source of

More information

PERFORMANCE SPECIFICATION

PERFORMANCE SPECIFICATION INCH-POUND MIL-PRF-6106/13F 17 November 2011 SUPERSEDING MIL-PRF-6106/13E 10 November 2000 PERFORMANCE SPECIFICATION RELAY, ELECTROMAGNETIC, 25 AMPERE, 3PST, NO, WITH 2 AMPERE 1PDT AUXILIARY CONTACTS,

More information

PDS: Rev :A STATUS:Released Printed: Jan 04, 2012

PDS: Rev :A STATUS:Released Printed: Jan 04, 2012 of 2 A.0 Objective This specification defines the performance, test, quality and reliability requirements of the USB3.0 Standard- A Receptacle. 2.0 Scope This specification is applicable to the termination

More information

SOLARIMMERSION IV Advanced Installation Manual v1.9

SOLARIMMERSION IV Advanced Installation Manual v1.9 SOLARIMMERSION IV Advanced Installation Manual v1.9 1 Contents 1. Overview 2. Technical Specifications 3. Installation Mounting Electrical Installation Clamp Installation Wiring Diagrams 4. Installation

More information

Type Approval Certificate

Type Approval Certificate Type Approval Certificate Type Approval Identification: SA 1491 AA3 Issuing Authority: NRCS, Legal Metrology, Type Approval Identification of Instrument: Utility Systems Model WMD in-line electronic pre-payment

More information

P-LED EALP03SXABA0. Lead (Pb) Free Product - RoHS Compliant

P-LED EALP03SXABA0. Lead (Pb) Free Product - RoHS Compliant Lead (Pb) Free Product - RoHS Compliant Feature Low profile. Uniform color. High flux output. Colorless clear resin. Brightness: 1125 to 2850 mlm at 30mA. Packaged in tubes for use with automatic insertion

More information

Install-It-Yourself Batch Oven Kits

Install-It-Yourself Batch Oven Kits Install-It-Yourself Batch Oven Kits Flexibility with standard assembled or installit-yourself ovens for preheating, drying and curing for applications up to 00 F (260 C). Durability provided from quality

More information

SCHOTT FLM 4 Fiber Lighting Modul. Operating Instructions

SCHOTT FLM 4 Fiber Lighting Modul. Operating Instructions SCHOTT FLM 4 Fiber Lighting Modul Operating Instructions Contents 1. Important information 3 2. Scope of Delivery 3 3. Intended use 3 4. Safety information 4 5. Operation 5 5.1. LED Optics Module 5 5.2.

More information

Synchronization and pump-probe experiments

Synchronization and pump-probe experiments Synchronization and pump-probe experiments by Stefan Düsterer Outline o Pump-probe probe infrastructure VUV + VUV VUV + optical o o Temporal overlap and jitter As example: the pump-probe probe chamber

More information

PLG110 Linear Stage PLG mm Travel PLG110 s. Frameless Motor With Ballscrew Linear Motor

PLG110 Linear Stage PLG mm Travel PLG110 s. Frameless Motor With Ballscrew Linear Motor 100mm Travel s Frameless Motor With Ballscrew Linear Motor Features Travel to 600mm Loads up to 25kg Precision Ballscrew Drive or Linear Motor Drive Integrated Frameless Servo Motor or NEMA23 Motor Mount

More information

Jun11 Rev A

Jun11 Rev A Product Specification MULTIGIG* Extreme 108-2072-2 28Jun11 Rev A 1. SCOPE 1.1. Content This document is specific to features, performance and testing of a metal shell and hardware kit which can be installed

More information

Model VD 23 Vacuum drying chambers for non-flammable solvents

Model VD 23 Vacuum drying chambers for non-flammable solvents Model VD 23 Vacuum drying chambers for non-flammable solvents A BINDER vacuum drying oven VD series gently dries materials with homogeneous temperature distribution. The patented expansion shelf technology

More information

Model VDL 115 Vacuum drying chambers for flammable solvents

Model VDL 115 Vacuum drying chambers for flammable solvents Model VDL 115 Vacuum drying chambers for flammable solvents A BINDER safety vacuum drying oven of the VDL series ensures maximum safety when drying organic solvents standard with TÜV/GS. The inner chamber

More information

Standard Operating Procedure for the Kaye Validator AVS System

Standard Operating Procedure for the Kaye Validator AVS System Document #: Revision: A Kaye Validator AVS Standard Operating Version 1.0 This is a version controlled document that is released with Amphenol internal approvals. The Approvals section is provided for

More information

VDL-004. Miniature Manual Variable Optical Delay Line. User Guide

VDL-004. Miniature Manual Variable Optical Delay Line. User Guide VDL-004 Miniature Manual Variable Optical Delay Line User Guide Version: 1.1 Date: July 24, 2018 General Photonics Corporation is located in Chino, California. For more information visit the company's

More information

Elastimold Molded Vacuum Recloser - MVR

Elastimold Molded Vacuum Recloser - MVR Subject: Elastimold Recloser Technical Specification Date: 4/24/12 File: MVR-RS-0412 Page: 1 of 6 1.0 SCOPE Elastimold Molded Vacuum Recloser - MVR 1.1 This specification covers electrical and mechanical

More information

Operating Instructions for Rotating Vane Flow Meter. Model: DRH-...

Operating Instructions for Rotating Vane Flow Meter. Model: DRH-... Operating Instructions for Rotating Vane Flow Meter Model: DRH-... 1. Contents 1. Contents... 2 2. Note... 3 3. Regulation Use... 3 4. Operating principles... 3 5. Instrument inspection... 4 6. Mechanical

More information

SPDS SERIES. Ultra-High Purity Diaphragm Valves (SPDS) Pneumatic Diaphragm Valves (SPDSA) Dual Containment Diapgragm Valves (DCS)

SPDS SERIES. Ultra-High Purity Diaphragm Valves (SPDS) Pneumatic Diaphragm Valves (SPDSA) Dual Containment Diapgragm Valves (DCS) SPDS SERIES Ultra-High Purity Diaphragm Valves (SPDS) Pneumatic Diaphragm Valves (SPDSA) Dual Containment Diapgragm Valves (DCS) SPDS SERIES This SPDS (diaphragm design) valve series is intended for bulk

More information

Document Title Issue Date. TBE/KBE-DOC TBE/KBE Document List TBE/KBE-REV TBE/KBE Revision List

Document Title Issue Date. TBE/KBE-DOC TBE/KBE Document List TBE/KBE-REV TBE/KBE Revision List Sid 1 (6) TBE/KBE DOCUMENT LIST Document Issue Date Supersedes TBE/KBE-DOC 12.0 2013-08-20 Issue 11.0 TBE/KBE-DOC TBE/KBE Document List 12.0 13-08-20 TBE/KBE-REV TBE/KBE Revision List 12.0 13-08-20 TBE/KBE

More information

Aug04 Rev D EC

Aug04 Rev D EC Product Specification 108-1163 18Aug04 Rev D EC 0990-1127-04 Modular Plugs, Thru-Hole and Surface Mount Jacks, Data and Telephone, PCB Mounted 1. SCOPE 1.1. Content This specification covers the performance,

More information

LITHIUM LENS FOR POSITRONS AND ANTIPROTONS IN COMPARISON 1

LITHIUM LENS FOR POSITRONS AND ANTIPROTONS IN COMPARISON 1 October 30, 2009 CBN 09-8 LITHIUM LENS FOR POSITRONS AND ANTIPROTONS IN COMPARISON 1 A.Mikhailichenko, Cornell University, LEPP, Ithaca, NY 14853 Abstract. In this report we compare parameters for Lithium

More information

SECTION PANELBOARDS

SECTION PANELBOARDS SECTION 16450 PART 1 - GENERAL 1.01 DESCRIPTION A. Section includes requirements for panelboards providing fault interrupting capability and overcurrent protective devices. 1.02 REFERENCE STANDARDS A.

More information

Design Standard for Printed Electronics on Flexible Substrates

Design Standard for Printed Electronics on Flexible Substrates Design Standard for Printed Electronics on Flexible Substrates Developed by the D-61 Printed Electronics Design Subcommittee of the D-60 Printed Electronics Committee of IPC Users of this publication are

More information

Lyte-MV Analogue Userguide Red & IR Versions

Lyte-MV Analogue Userguide Red & IR Versions Product Overview Lyte-MV Analogue Userguide Red & IR Versions Thanks for purchasing the Lyte-MV, a laser diode module emitting a uniform (non-gaussian) and welldefined line with a user-adjustable focus.

More information

Product Specification

Product Specification Product Specification PR041-01 Revision A0 (Date 4-3-2017) Stewart Connector 11118 Susquehanna Trail South Glen Rock, Pa. 17327-9199 Ph: (717) 235-7512 Fax: (717) 235-4675 Product Specification PR0410-01

More information

Power Meter PowerMonitor 500

Power Meter PowerMonitor 500 ROCKWELL AUTOMATION PROCUREMENT SPECIFICATION PROCUREMENT SPECIFICATION PowerMonitor 500 NOTICE: The specification guidelines in this document are intended to aid in the specification of products. Specific

More information

Advanced Instrument Development, Inc Curtiss St. Downers Grove, IL U.S.A. Phone: (630) Fax: (630)

Advanced Instrument Development, Inc Curtiss St. Downers Grove, IL U.S.A. Phone: (630) Fax: (630) Three Output Ionization Chamber Calibration Procedure for Pre-Amplifier board 61171D The following adjustments apply to the calibration of a 61171D pre-amplifier board for a stationary three-output ion

More information

SECTION THERMAL UTILITY METERING INTERFACE

SECTION THERMAL UTILITY METERING INTERFACE SECTION 25 33 13 THERMAL UTILITY METERING INTERFACE GENERAL 1.1 SUMMARY A. Section includes metering and monitoring for interface points for utility chilled water, hot water, and steam fed to each building,

More information

INSTALLATION GUIDE Ultrasonic Sensors Series UFP

INSTALLATION GUIDE Ultrasonic Sensors Series UFP INSTALLATION GUIDE Ultrasonic Sensors Series UFP For further information please see the data sheet at www.waycon.biz/products/ultrasonic-sensors/ FIRST STEPS WayCon Positionsmesstechnik GmbH would like

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

Inter-University Accelerator Centre New Delhi

Inter-University Accelerator Centre New Delhi Inter-University Accelerator Centre New Delhi EOI No: IUAC/FEL/PC/May2017 Date: 26 May 2017 Notice inviting Expression of Interest for procurement of Photocathode deposition system to be used for producing

More information

Explosion-Protected Arc-Fault Protection ExAFCI Arc-Fault Circuit Interrupter

Explosion-Protected Arc-Fault Protection ExAFCI Arc-Fault Circuit Interrupter Installation, Operation & Maintenance Sheet Explosion-Protected Arc-Fault Protection ExAFCI Arc-Fault Circuit Interrupter > Contents 1 Contents 1 Contents...2 2 General Information...2 2.1 Manufacturer...2

More information

These devices require adequately trained and experienced operating personnel.

These devices require adequately trained and experienced operating personnel. MMa, 2017 PI Ceramic GmbH PL112 PL140 Rectangular PICMA Bender Piezo Actuators Figure 1: Different rectangular PICMA bender actuator models INFORMATION Before soldering, installing and operating the actuator,

More information

PROVIDE A UNIFORM DOSE IN THE SMALL ANIMAL.

PROVIDE A UNIFORM DOSE IN THE SMALL ANIMAL. Considerations in the Use of the RS 2000 X ray Irradiator for Biological Research (Primarily Small Animal, tissue, and cells) and the fallacy of the High KV spectrum. The performance goal for a small animal

More information

INSTALLATION DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER. Before installation:

INSTALLATION DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER. Before installation: DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER The DKM-409 is a precision instrument designed for displaying various AC parameters in 3-phase distribution panels. Thanks to its isolated

More information

TABLE OF CONTENTS 3.0 QUALITY ASSURANCE PROGRAM TEST EQUIPMENT AND INSTRUMENTATION 5 ATTACHMENTS A PHOTOGRAPHS A-1 B VIBRATION TEST DATA B-1

TABLE OF CONTENTS 3.0 QUALITY ASSURANCE PROGRAM TEST EQUIPMENT AND INSTRUMENTATION 5 ATTACHMENTS A PHOTOGRAPHS A-1 B VIBRATION TEST DATA B-1 Page No. 2 TABLE OF CONTENTS Page No. 1.0 INTRODUCTION 3 1.1 Scope 3 1.2 References 3 1.3 Test Specimen Description 3 1.4 Summary 3 2.0 TEST PROCEDURES AND RESULTS 4 2.1 Vibration Resonance Search Test

More information

Sputtering Sensor ADVANTAGES

Sputtering Sensor ADVANTAGES The INFICON is specifically designed for use in any sputtering process. The sensor body and cooling tubes are gold plated beryllium copper for maximum cooling efficiency in the sputtering environment.

More information

Model ISOXSHLD ISOCS Shield Systems

Model ISOXSHLD ISOCS Shield Systems Features Portable Germanium detector shield system for in situ assays Includes 25 mm and 50 mm thick lead shields Reduces interfering 1000 kev radiation by a factor of 7.5 (25 mm) and 60 (50 mm) Modular

More information

Meterless Laser Power/Energy Measurement Simplifies Embedding

Meterless Laser Power/Energy Measurement Simplifies Embedding White Paper Meterless Laser Power/Energy Measurement Simplifies Embedding Traditional laser power and energy measurement instruments typically comprise a sensor head connected to separate meter electronics.

More information

Specification. Radiamatic Detectors and Accessories

Specification. Radiamatic Detectors and Accessories 4/92 Page 1 of 8 Radiamatic Detectors and Accessories Specification Function The Radiamatic temperature detector generates an emf proportional to the temperature of the object sighted. It is used with

More information

Product Specifications and Usage Information P4350 & P4355 MX2 Laser Micrometer and Gauge Blocks

Product Specifications and Usage Information P4350 & P4355 MX2 Laser Micrometer and Gauge Blocks DESCRIPTION The P4350 MX2 Laser Micrometer is a component of the DoseControl Dosimetry System used to measure the thickness of Harwell PMMA (Perspex) Dosimeters. The MX2 is designed to be mounted, connected,

More information

Assembly. 1 Compact version with valve plug 2 Short tube version with cable 3 Compact version with M12 plug 4 Short tube version with M12 plug

Assembly. 1 Compact version with valve plug 2 Short tube version with cable 3 Compact version with M12 plug 4 Short tube version with M12 plug Vibration Limit Switch Features Limit switch for liquids External function test with test magnet Onsite function check possible thanks to LED indication Large selection of process connections for hassle-free

More information

3mm Advanced Super Flux LEDs T4C-4PRB

3mm Advanced Super Flux LEDs T4C-4PRB 3mm Advanced Super Flux LEDs Lead (Pb) Free Product - RoHS Compliant Feature RoHS compliant. Piranha package. Colorless clear resin. Low thermal resistance Packaged in tubes for Automatic Insertion Equipment.

More information

PI31xx Product Family Reflow Soldering Guidelines

PI31xx Product Family Reflow Soldering Guidelines APPLICATION NOTE AN:302 PI31xx Product Family Reflow Soldering Guidelines Chris Swartz Principal Applications Engineer Contents Page Introduction 1 Storage 1 Solder Paste 1 Leaded Applications 1 PCB Considerations

More information

LIMO Line Laser System - PEX High Power Diode Laser System for Industrial Applications

LIMO Line Laser System - PEX High Power Diode Laser System for Industrial Applications General Description: Flat Panel Display The laser system is engineered to generate a homogeneous line of 80 mm x 0.2 mm. Componets: - 9x Laser modules - 9x Fibres - Beam Shaping Module - High power laser

More information

Manual Control Unit GFCD 16

Manual Control Unit GFCD 16 Manual Control Unit 1400002_EN/04.2017 Index 1. Main features 3 2. Technical features 3 3. Installation guidelines 4 4. Preliminary checks 5 5. Electrical connections 5 6. Filter taps 5 7. Settings 6 8.

More information

Manual# Installation Manual. 200E Series. DCU 210E/208E Engine Panel RP 210E/220E Remote Panel

Manual# Installation Manual. 200E Series. DCU 210E/208E Engine Panel RP 210E/220E Remote Panel Manual# 1006495 Installation Manual 200E Series DCU 210E/208E Engine Panel RP 210E/220E Remote Panel Installation Manual for the Marine Pro 200E Series ~~~ DCU 210E/208E Diesel Engine Control Unit RP 210E/220E

More information

View in Several Directions

View in Several Directions Multi-Sensor Machine Allows Two Measurements in One Setup View in Several Directions Measuring injection molded parts from the top and from the side in one setup has been impossible for medical device

More information

ECET 211 Electric Machines & Controls Lecture 6 Contactors and Motor Starters (1 of 2) Lecture 6 Contactors and Motor Starters

ECET 211 Electric Machines & Controls Lecture 6 Contactors and Motor Starters (1 of 2) Lecture 6 Contactors and Motor Starters ECET 211 Electric Machines & Controls Lecture 6 Contactors and Motor Starters (1 of 2) Text Book: Chapter 6, Electric Motors and Control Systems, by Frank D. Petruzella, published by McGraw Hill, 2015

More information

HIGH POWER CONTACT ELECTRICAL SPECIFICATIONS

HIGH POWER CONTACT ELECTRICAL SPECIFICATIONS HIGH POWER CONTACT ELECTRICAL SPECIFICATIONS 11.12.438 8.83.348 5.72.225 8.83.348 11.12.438 12.70.500 Receiver Contact Part # 610 149 101 38.48 1.515 36.20 1.425 12.70.500 ITA Contact Part # 610 150 101

More information

arxiv: v1 [physics.ins-det] 13 Jan 2015

arxiv: v1 [physics.ins-det] 13 Jan 2015 The Assembly of the Belle II TOP Counter Boqun Wang, On behalf of the Belle II PID Group Department of Physics, University of Cincinnati, Cincinnati, OH, USA University of Cincinnati preprint UCHEP-14-01

More information

Ultra High Vacuum Pumps

Ultra High Vacuum Pumps Ultra Hig Ultra High Vacuum Pumps TiTan TM Ion Pumps BOOSTIVAC TM Titanium Sublimation Pumping (TSP) Non-evaporable Getter (NEG) DIGITEL TM Ion Pump Controllers 310.00.02 Excerpt from the Leybold Full

More information

LASERMET LS-20 INSTRUCTION MANUAL

LASERMET LS-20 INSTRUCTION MANUAL LASERMET LS-20 INSTRUCTION MANUAL LASER SAFETY SHUTTER 00629-53-000 Issue 2 14 December 2012 Lasermet LS-20 Instruction Manual Contents DECLARATION OF CONFORMITY... 3 Concept... 4 Control Options... 4

More information

Short instructions. Olympus FV1000

Short instructions. Olympus FV1000 Olympus FV1000-1- Short instructions Short instructions. Olympus FV1000 Version 140416 This short instruction will guide you through the turn on and off sequence of the microscope system. For a detailed

More information

Physics Requirements for the LUSI X-ray Focusing Lens System

Physics Requirements for the LUSI X-ray Focusing Lens System PHYSICS REQUIREMENT DOCUMENT (PRD) Doc. No. SP-391-000-11 R1 LUSI SUB-SYSTEM CXI, XCS, XPP Physics Requirements for the Aymeric Robert LUSI Scientist, Author Signature Date David Fritz Sébastien Boutet

More information

Quick Start Installation and User Manual

Quick Start Installation and User Manual 1 Quick Start Installation and User Manual Contents 1. Overview 2. Technical Specifications 3. Installation Mounting Electrical Installation Clamp Installation Wiring Diagrams 4. Installation Settings

More information

High Performance Infrared Thermometer with High D:S Ratio, Adjustable Emissivity, and Built-in Laser Sighting

High Performance Infrared Thermometer with High D:S Ratio, Adjustable Emissivity, and Built-in Laser Sighting High Performance Infrared Thermometer with High D:S Ratio, Adjustable Emissivity, and Built-in Laser Sighting DHS235XEL WD1048 Rev A 04/10/10 Table of Contents 1. Product Introductions... 1 1-1 Features...

More information