Correction Factor Kit v. 1.1 (14 March 2008)

Size: px
Start display at page:

Download "Correction Factor Kit v. 1.1 (14 March 2008)"

Transcription

1

2 Correction Factor Kit v. 1.1 (14 March 2008)

3 F-3026 Correction Factor Kit Operation Manual v (All HORIBA Jobin Yvon companies were formerly known as Jobin Yvon) USA: HORIBA Jobin Yvon Inc., 3880 Park Avenue, Edison, NJ , Toll-Free: jobinyvon Tel: , Fax: , France: HORIBA Jobin Yvon S.A.S., 16-18, rue du Canal, Longjumeau Cedex, Tel: +33 (0) , Fax: +33 (0) , Japan: HORIBA Ltd., JY Optical Sales Dept, Higashi-Kanda, Daiji Building, Higashi-Kanda Chiyoda-ku, Tokyo , Tel: +81 (0) , Germany: +49 (0) Italy: UK: +44 (0) China: +86 (0) i

4 Copyright 2008 by HORIBA Jobin Yvon. All rights reserved. No part of this work may be reproduced, stored, in a retrieval system, or transmitted in any form by any means, including electronic or mechanical, photocopying and recording, without prior written permission from HORIBA Jobin Yvon. Requests for permission should be requested in writing. Origin is a registered trademark of OriginLab Corporation. Spectralon is a registered trademark of Labsphere, Inc. Windows is a registered trademark of the Microsoft Corporation. Information in this manual is subject to change without notice, and does not represent a commitment on the part of the vendor. March 2008 Part Number ii

5 Table of Contents 0: Introduction About the Correction Factor Kit Chapter overview Disclaimer Safety summary Additional risks of tungsten-halogen lamps : Unpacking & Installation Electrical requirements Unpacking : System Description Introduction Components : Operation Set up the accessory Run the scan Analyze the data Remove the accessory : Maintenance : Troubleshooting Introduction Basic troubleshooting steps Further assistance : Technical Specifications Introduction Specifications : Correction-Factor-Kit Record of Use Form : Index iii

6 iv

7 0: Introduction About the Correction Factor Kit Introduction The F-3026 Correction Factor Kit is an NIST-traceable accessory for the FluoroMax and Fluorolog spectrofluorometers for calibrating the relative spectral response of the spectrofluorometer, and includes a Spectralon integrating sphere as a sampling optic. Each Correction Factor Kit includes a regulated 12 VDC power supply, a calibration certificate, and a CD with data files for use with FluorEssence software. The Correction Factor Kit provides 50 hours of operation before recalibration is necessary. Note: Keep this and the other reference manuals near the system. 0-1

8 Chapter overview 1: Unpacking & Installation How to unpack and install the accessory. 2: System Description Components of the Correction Factor Kit. 3: Operation How to set up and use the Correction Factor Kit. 4: Maintenance How to keep the Correction Factor Kit running properly for years to come. 5: Troubleshooting Potential sources of problems, their most probable causes, and possible solutions. Introduction 6: Technical Specifications Instrument specifications and computer requirements. 7: Correction-Factor-Kit Record of Use Form 8: Index A log to keep track of time that the accessory is used before recalibration is required. 0-2

9 Disclaimer By setting up or starting to use any HORIBA Jobin Yvon product, you are accepting the following terms: Introduction You are responsible for understanding the information contained in this document. You should not rely on this information as absolute or all-encompassing; there may be local issues (in your environment) not addressed in this document that you may need to address, and there may be issues or procedures discussed that may not apply to your situation. If you do not follow the instructions or procedures contained in this document, you are responsible for yourself and your actions and all resulting consequences. If you rely on the information contained in this document, you are responsible for: Adhering to safety procedures Following all precautions Referring to additional safety documentation, such as Material Safety Data Sheets (MSDS), when advised As a condition of purchase, you agree to use safe operating procedures in the use of all products supplied by HORIBA Jobin Yvon, including those specified in the MSDS provided with any chemicals and all warning and cautionary notices, and to use all safety devices and guards when operating equipment. You agree to indemnify and hold HORIBA Jobin Yvon harmless from any liability or obligation arising from your use or misuse of any such products, including, without limitation, to persons injured directly or indirectly in connection with your use or operation of the products. The foregoing indemnification shall in no event be deemed to have expanded HORIBA Jobin Yvon s liability for the products. HORIBA Jobin Yvon products are not intended for any general cosmetic, drug, food, or household application, but may be used for analytical measurements or research in these fields. A condition of HORIBA Jobin Yvon s acceptance of a purchase order is that only qualified individuals, trained and familiar with procedures suitable for the products ordered, will handle them. Training and maintenance procedures may be purchased from HORIBA Jobin Yvon at an additional cost. HORIBA Jobin Yvon cannot be held responsible for actions your employer or contractor may take without proper training. Due to HORIBA Jobin Yvon s efforts to continuously improve our products, all specifications, dimensions, internal workings, and operating procedures are subject to change without notice. All specifications and measurements are approximate, based on a standard configuration; results may vary with the application and environment. Any software manufactured by HORIBA Jobin Yvon is also under constant development and subject to change without notice. Any warranties and remedies with respect to our products are limited to those provided in writing as to a particular product. In no event shall HORIBA Jobin Yvon be held li- 0-3

10 Introduction able for any special, incidental, indirect or consequential damages of any kind, or any damages whatsoever resulting from loss of use, loss of data, or loss of profits, arising out of or in connection with our products or the use or possession thereof. HORIBA Jobin Yvon is also in no event liable for damages on any theory of liability arising out of, or in connection with, the use or performance of our hardware or software, regardless of whether you have been advised of the possibility of damage. 0-4

11 Safety summary Introduction The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture and intended use of instrument. HORIBA Jobin Yvon assumes no liability for the customer s failure to comply with these requirements. Certain symbols are used throughout the text for special conditions when operating the instruments: Warning: A hazardous condition exists, or danger exists that could damage the equipment. HORIBA Jobin Yvon is not responsible for damage arising out of improper use of the equipment. Caution: Intense ultraviolet, visible, or infrared light! Wear light-protective goggles, full-face shield, skin-protection clothing, and light-blocking gloves. Do not stare into light. Warning: Hot! This symbol warns the user that hot equipment may be present, and could create a risk of fire or burns. Note: General information is given concerning operation of the equipment. Read this manual before using or servicing the instrument. Wear appropriate safety goggles to protect the eyes. 0-5

12 Introduction Additional risks of tungsten-halogen lamps Among the dangers associated with tungsten-halogen lamps are: Burns caused by contact with a hot bulb or its casing. Fire ignited by hot bulb. Interaction of other nearby chemicals with intense ultraviolet, visible, or infrared radiation. Damage caused to apparatus placed close to the lamp. Explosion or mechanical failure of the lamp. Visible radiation Any very bright visible light source will cause a human aversion response: we either blink or turn our head away. Although we may see a retinal afterimage (which can last for several minutes), the aversion response time (about 0.25 seconds) normally protects our vision. This aversion response should be trusted and obeyed. NEVER STARE AT ANY BRIGHT LIGHT-SOURCE FOR AN EXTENDED PERIOD. Overriding the aversion response by forcing yourself to look at a bright light-source may result in permanent injury to the retina. This type of injury can occur during a single prolonged exposure. Excessive exposure to visible light can result in skin and eye damage. Visible light sources that are not bright enough to cause retinal burns are not necessarily safe to view for an extended period. In fact, any sufficiently bright visible light source viewed for an extended period will eventually cause degradation of both night and color vision. Appropriate protective filters are needed for any light source that causes viewing discomfort when viewed for an extended period of time. For these reasons, prolonged viewing of bright light sources should be limited by the use of appropriate filters. The blue-light wavelengths ( nm) present a unique hazard to the retina by causing photochemical effects similar to those found in UV-radiation exposure. Infrared radiation Warning: Tungsten-halogen lamps can be dangerous. Please read the following precautions. Infrared (or heat) radiation is defined as having a wavelength between 780 nm and 1 mm. Specific biological effectiveness bands have been defined by the CIE (Commission International de l Eclairage or International Commission on Illumination) as follows: IR-A (near IR) ( nm) IR-B (mid IR) ( nm) IR-C (far IR) (3000 nm 1 mm) 0-6

13 Introduction The skin and eyes absorb infrared radiation (IR) as heat. Workers normally notice excessive exposure through heat sensation and pain. Infrared radiation in the IR-A that enters the human eye will reach (and can be focused upon) the sensitive cells of the retina. For high irradiance sources in the IR-A, the retina is the part of the eye that is at risk. For sources in the IR-B and IR-C, both the skin and the cornea may be at risk from flash burns. In addition, the heat deposited in the cornea may be conducted to the lens of the eye. This heating of the lens is believed to be the cause of so called glassblowers cataracts because the heat transfer may cause clouding of the lens. Retinal IR Hazards (780 to 1400 nm): possible retinal lesions from acute high irradiance exposures to small dimension sources. Lens IR Hazards (1400 to 1900 nm): possible cataract induction from chronic lower irradiance exposures. Corneal IR Hazards (1900 nm to 1 mm): possible flashburns from acute high irradiance exposures. Who is likely to be injured? The user and anyone exposed to the radiation or tungstenhalogen lamp shards as a result of faulty procedures. Injuries may be slight to severe. 0-7

14 Introduction 0-8

15 1: Unpacking & Installation Electrical requirements Unpacking & Installation In addition to those required for the instrument, at least one extra outlet must be available for the Correction Factor Kit. Voltage Factory preset at 110 V ± 5% at 60 Hz or 220 V ± 5% at 50 Hz Three-conductor power cord connected to the system frame (earth) ground. This ground provides a return path for fault current from equipment malfunction or external faults. Warning: For all instruments, ground continuity is required for safe operation. Any discontinuity in the ground line can make the instrument unsafe for use. Do not operate this system from an ungrounded source. HORIBA Jobin Yvon is not liable for damage from line surges and voltage fluctuations. A surge protector is strongly recommended for minor power fluctuations. For more severe voltage fluctuations, a generator or an uninterruptible power supply is suggested. Improper line voltages can damage the equipment severely. 1-1

16 Unpacking When the Correction Factor Kit arrives The accessory is delivered disassembled. The shipment should include: Power cords Power supply Source/integrating sphere module Special gap-bed Sample-compartment cover Manual CD with irradiance data Thumbscrews Calibration datasheet Unpacking & Installation Note: The source/integrating sphere module may arrive premounted atop the special gapbed. General checklist for the Correction Factor Kit Quantity Item Part Number 1 Source/integrating sphere module VDC power supply 1 12 VDC power cord 1 Mains power cord 1 Special gap-bed 1 Extended lid CD with irradiance data 1 Calibration datasheet 1 F-3026 Correction Factor Kit Operation Manual Brass thumbscrew Examine the shipping boxes carefully. If damage is evident, do not continue with the installation. Notify HORIBA Jobin Yvon and the shipper at once. Note: Many public carriers do not recognize claims for concealed damage reported later than 15 days after delivery. For a shipping damage claim, inspection by the carrier agent is required. Therefore, the original packing material should be kept as evidence. While HORIBA Jobin Yvon is not responsible for damage occurring during transit, the company will extend every effort to aid and advise. To avoid damage, unpack the equipment as close as possible to the selected location of the instrument. HORIBA Jobin Yvon suggests that you unpack the equipment in order to set the major components on the table, while also checking that all parts have arrived undamaged. Package contents The Correction Factor Kit is contained in packing cartons. If any other special accessories were included in the order, they accompany the accessory, unless specified by 1-2

17 Unpacking & Installation HORIBA Jobin Yvon. See the Performance Test Report or the sales order for a list of all items shipped or back-ordered. Unpacking instructions 1 Open the shipping carton. 2 Remove the foam-injected top piece and any other shipping restraints in the carton. 3 With at least one helper, carefully lift the components from the carton and Warning: Watch your fingers! put them on the table on which the system will rest. 4 Unpack all other boxes containing power supplies, detectors, cables, and manu- Note: Do not dispose of shipping cartons until after installation. als. Store all these items together in a safe location. 5 Inspect for previously hidden damage. Notify the carrier and HORIBA Jobin Yvon if any damage is found. 6 Check the packing list and sales order to verify that all components are present. 1-3

18 Unpacking & Installation 1-4

19 2: System Description System Description Introduction The Correction Factor Kit consists of the following major components: Module with light-source and integrating sphere mounted on a special gap-bed Special sample-compartment cover Power supply for module Irradiance data Components Module with light-source and integrating sphere The light-source (tungsten-halogen lamp) and integrating sphere are combined in a single module, mounted on a special gap-bed. This special gap-bed replaces the normal gap-bed that was included with your instrument. The 12 VDC power cord for the light source is threaded through the front of the sample compartment in a light-tight manner. Every 50 hours the accessory ought to be recalibrated. Special sample-compartment cover Because the light-source and integrating sphere module is taller than the standard sample compartment, a special sample-compartment cover is included to make the interior of the sample compartment light-tight. Power supply for module An external regulated 12 VDC power supply supplies electricity to the light source. This power supply is connected to your standard AC wall outlet (mains). Irradiance data For relative calibration of the module, irradiance data from the near-uv through IR are included on a CD with the Kit. These data can be loaded into FluorEssence to provide relative irradiance data from the module. A separate datasheet is also included. 2-1

20 System Description 2-2

21 3: Operation Set up the accessory. 1 Attach the correctionfactor light-source to its gap-bed with the brass thumbscrews. Use this orientation. Be sure the power cable doesn t block optical apertures. Operation Note: The source/integrating sphere module may arrive premounted atop the special gap-bed. 2 Plug the power cable into the front of the gap-bed. 3 Remove the standard sample-compartment gap-bed, and insert the correction-factor kit gapbed into the sample compartment. Fix the gap-bed firmly to the sample compartment with screws through the front panel. 3-1

22 Run the scan. 1 Start FluorEssence and initialize the instrument. Turn the xenon lamp OFF, and the lamp for the correction-factor kit ON. ON Operation Warning: Wear appropriate eye-protection against UV, visible, and IR when the tungsten-halogen lamp is on and the sample-compartment is uncovered. Note: We recommend keeping a log of the amount of time the source is on. Calibration is necessary after every 50 h of use. 2 Put the raised lid on the sample compartment. Make sure it fits securely, and there is no light leak into the sample compartment. Caution: Do not operate the accessory without the lid completely covering the sample compartment. 3 Collect an emission scan using the following scan parameters: Range: nm; increment: 5 nm; integration time: 1 s 3-2

23 Analyze the data. 1 Import the standard irradiance file for the calibrated lamp in the correction-factor kit: a In the File menu, click Import, and choose SPC Data. Operation The Open window appears. b Browse for and choose the irradiance file on the CD-ROM (included with this kit). Note: Choose the correct irradiance file associated with the desired detector. c Click the Open button. The irradiance file opens. 2 Identify the datalabel names for the two graphs. Right-click on the curve; the data name appears at the bottom. 3-3

24 3 From the File menu, choose New... The New window appears. Operation 4 Choose Graph, then the OK button. A new plot appears on the FluorEssence main window. 5 In the Analysis menu, choose Simple Math... The Simple Math Tool window opens. 3-4

25 Operation a Choose the data corresponding to the irradiance file. b Choose divide as the math function c For Operand, select the data corresponding to the emission spectrum obtained using the correction-factor kit. d Select the Add New radio button, and click the Apply button. 6 The new graph now displays the correction factor. Normalize the graph as follows: a Right-click on the correction-factor curve and click Go to Data. b Highlight the B(Y) column, right click and select Normalize. c Note down the Min. number. d Open a new graph from the file menu. e Again, choose Simple Math... from the Analysis menu and choose the data name corresponding to the correction-factor data file. f Select the divide function. g For Operand, select constant and enter the Min. number noted previously. h Select the Add New radio button and click the Apply button. 3-5

26 7 Export the normalized correction-factor graph. In the File menu, click Export, and select SPC Data. Operation 8 Apply the correction factor into your system configuration by following the instructions in your instrument manual. Remove the accessory. 1 Remove the sample-compartment cover. Warning: Always wear appropriate eye-protection against bright UV, visible, and IR light when the tungsten-halogen lamp is on and the sample-compartment is uncovered. 2 Turn OFF the accessory. OFF Warning: The accessory may be HOT. Let the accessory cool before removal! 3-6

27 Operation Note: We recommend keeping a log of the amount of time the source is on. Calibration is necessary after every 50 h of use. 3 Remove the accessory s sample-compartment gap-bed, and return the standard gap-bed to the sample compartment. 3-7

28 Operation 3-8

29 4: Maintenance Maintenance Every 50 hours, the Correction Factor Kit must be recalibrated. Call the Spex Fluorescence Service Department at (732) , or fax us at (732) , for instructions on how to ship the Kit to us for recalibration. 4-1

30 Maintenance 4-2

31 5: Troubleshooting Introduction Troubleshooting Because your spectrofluorometer has many user settings and adjustable parameters, many apparent system failures are simply problems with the experimental setup, sample geometry, or choice of sample. If the symptoms indicate that a problem may be a system failure, contact the Spex Fluorescence Service Department, or your local distributor. Refer to the spectrofluorometer s Operation Manual and the FluorEssence User s Guide for help with troubleshooting problems in steady-state operation. This chapter is intended to isolate and solve problems regarding operation of the Correction Factor Kit with the system only. 5-1

32 Basic troubleshooting steps Troubleshooting When an apparent problem is first observed, follow the basic steps and suggestions outlined below to troubleshoot the initial problem. In many cases, shutting down the system, carefully restarting, and redoing an experiment helps to remove an unseen or outside issue that was causing the apparent problem. If the suggestions that follow do not help to remove the problem, note the steps that led up to the problem, and proceed to the next section. 1 Reboot. If this is the first time the problem has occurred, try restarting the system. Turn off all components and accessories. After a cool-down period, turn the system on and see if the failure persists. Many apparent software and hardware problems are actually conflicts with other software or peripheral devices currently active on the system. 2 Check the connections. With the power to all components and accessories turned off, make sure all accessories and cables are attached and configured properly. Verify that line cords provide the proper input AC voltage, and that all appropriate accessories and power supplies are powered ON during use of the system. 3 Verify system performance. Verify that the system is properly calibrated in steady-state mode. Print out each spectrum, and note the peak intensities and wavelength positions. 4 Duplicate the error. Try to duplicate the problem, and write down the steps required to do so. Our service engineers will attempt the same with a test system. Depending on the nature of the problem, a service visit may not be required. 5-2

33 Further assistance... Troubleshooting Read all software and accessory manuals before contacting the Spex Fluorescence Service Department. Often the manuals show the problem s cause and a method of solution. Technical support is available for both hardware and software troubleshooting. Before contacting the service department, however, complete the following steps. 1 If this is the first time the problem has occurred, try turning off the system and accessories. After a cool-down period, turn everything back on. 2 Make sure all accessories are properly configured, and turned on as needed. 3 Following the instructions in Operation, run a xenon-lamp scan and a water-raman scan to make sure the system is properly calibrated. Print the spectrum for each and note the peak intensities. 4 Check this chapter to see if the problem is discussed. 5 Visit our web site at See if the question is addressed in the Systems or FAQs sections of the site. 6 Try to duplicate the problem and write down the steps required to do so. The service engineers will try to do the same with a test system. Depending on the problem, a service visit may not be required. 7 If an error message appears in FluorEssence, write down the exact error displayed. 8 Determine FluorEssence s version number. a Choose the Help menu. b Choose About FluorEssence

34 The About FluorEssence window opens. Near the bottom are the FluorEssence and Origin version numbers. Troubleshooting c Click the OK button to close the window. 9 Write down the software s version numbers, along with the purchase dates, model numbers, system configuration, and serial numbers of the instrument and its accessories. 10 Call the Spex Fluorescence Service Department at (732) Be prepared to describe the malfunction and the attempts, if any, to correct it. Have serial and version numbers of all software and equipment handy, along with all relevant spectra (sample, polarization ratio, xenon-lamp scan, emission calibration, etc.). If the problem persists or is unlisted, call the Spex Fluorescence Service Department at (732) , or fax us at (732)

35 6: Technical Specifications Introduction Each F-3026 Correction Factor Kit consists of: An excitation source and integrating sphere in one module A special gap-bad for the sample compartment A special cover for the sample compartment 12 VDC output power supply A CD with irradiance file Specifications Excitation source Tungsten-halogen lamp Time between calibrations 50 h Calibrated range nm Calibration method As per NIST Handbook 150-2E Output data Absolute values in µw/nm Integrating sphere size 1.5 (3.8 cm) diameter Integrating sphere material Spectralon Electrical power input V AC, Hz, 0.7 A Technical Specifications 6-1

36 Technical Specifications 6-2

37 Correction-Factor-Kit Record of Use Form 7: Correction-Factor-Kit Record of Use Form Because the spectral output of a tungsten-halogen bulb can vary slightly over time, we recommend recalibration of the correction factor kit every 50 hours. Use the record-ofuse form on the next page to keep a log of usage for the correction factor kit. Contact the Spex Fluorescence Service Department for instructions on recalibration, at (732) , or fax us at (732) , when the 50 hours of usage is reached. 7-1

38 Correction-Factor-Kit Record of Use Form Correction-Factor-Kit Record of Use Page of In Service Date Operator Current Date Time On Time Off Total Time (Hours/Min.) Total Hours / 7-2

39 8: Index Key to the entries: Times New Roman font...subject or keyword Arial font...command, menu choice, or data-entry field Arial Condensed Bold font...dialog box Courier New font...file name or extension A About FluorEssence window About FluorEssence Add New radio button Analysis menu Apply button B B(Y) column C Index calibration certificate D damage danger divide E emission scan Export F File menu , 3-6 FluorEssence 0-1, 2-1, 3-2, 3-4, 5-1, Fluorolog FluoroMax G gap-bed...1-2, 2-1, 3-1, 3-7 Go to Data Graph H hazardous condition Help menu hot equipment I Import infrared radiation integrating sphere...0-1, 1-2, 2-1, 6-1 intense light irradiance data...1-2,

40 irradiance file...3-3, 3-5, 6-1 S Index M Material Safety Data Sheets Min , 7-2 MSDS N New window New NIST , 6-1 normalization Normalize O OK button , 5-4 Open button Open window Operand P power supply , 1-2, 2-1, 6-1 R read this manual notice recalibration...0-1, 4-1, 7-1 record-of-use form safety goggles notice safety precautions sample-compartment cover , 2-1, 3-6 Service Department...4-1, 5-1, 5-3 4, 7-1 Simple Math Tool window Simple Math SPC Data...3-3, 3-6 Spectralon...0-1, 6-1 symbols T thumbscrews...1-2, 3-1 troubleshooting tungsten-halogen lamp , 3-2, 6-1, 7-1 tungsten-halogen lamp risks V visible radiation W water-raman scan X xenon lamp xenon-lamp scan

41 HORIBA JOBIN YVON INC Park Ave., Edison, New Jersey , U.S.A. [Design Concept] The HORIBA Group application images are collaged in the overall design. Beginning from a nano size element, the scale of the story develops all the way to the Earth with a gentle flow of the water.

42

Addendum to Fluorolog Tau-3 Lifetime System Operation Manual

Addendum to Fluorolog Tau-3 Lifetime System Operation Manual Addendum to Fluorolog Tau-3 Lifetime System Operation Manual Pockels-cell power supplies http://www.jobinyvon.com (All HORIBA Jobin Yvon companies were formerly known as Jobin Yvon) USA: HORIBA Jobin Yvon

More information

User s Guide for software version 2.1

User s Guide for software version 2.1 FluorEssence User s Guide v. 2.0 (6 Dec 2006) User s Guide for software version 2.1 http://www.jobinyvon.com (All HORIBA Jobin Yvon companies were formerly known as Jobin Yvon) 1 Copyright 2006 by HORIBA

More information

STANDARD OPERATING PROCEDURE: HORIBA SPECTROFLUOROMETER_STEADY STATE

STANDARD OPERATING PROCEDURE: HORIBA SPECTROFLUOROMETER_STEADY STATE STANDARD OPERATING PROCEDURE: HORIBA SPECTROFLUOROMETER_STEADY STATE Purpose of this Instrument: Essential tool for characterizing the relationship between absorbed and emitted photos at specified wavelengths.

More information

220F and 1700F Fiber Adapters

220F and 1700F Fiber Adapters 220Fand1700F Fi beradapt er s Oper at i onmanual Par tnumberj80112r ev.a 220F and 1700F Fiber Adapters J80112 rev. A (11 Mar 2014) 220F and 1700F Fiber Adapters Operating Manual www.horiba.com i 220F

More information

Benchmark Plus Microplate Reader User Manual. This Instrument for Laboratory Use Only Rev C

Benchmark Plus Microplate Reader User Manual. This Instrument for Laboratory Use Only Rev C Benchmark Plus Microplate Reader User Manual This Instrument for Laboratory Use Only 4100172 Rev C 4100172 Rev C Benchmark Plus System with Incubator Catalog Number 170-6930 Benchmark Plus Reader with

More information

SDrive-500 Shutter Control Unit

SDrive-500 Shutter Control Unit SDrive-500 Shutter Control Unit Operation Manual Rev. A http://www.horiba.com/scientific i Copyright 2012 by HORIBA Instruments Incorporated. All rights reserved. No part of this work may be reproduced,

More information

EOS-6000 Series Optical A/B Switch User Manual DC Version

EOS-6000 Series Optical A/B Switch User Manual DC Version EOS-6000 Series Optical A/B Switch User Manual DC Version For more information on this and other products: Contact Sales at EMCORE 626-293-3400, or visit www.emcore.com. Table of Contents Table of Contents...2

More information

SATA II HDD Canister KISS DA 435 Quick Reference Guide

SATA II HDD Canister KISS DA 435 Quick Reference Guide SATA II HDD Canister KISS DA 435 Quick Reference Guide If it s embedded, it s Kontron 1. Table of Contents SATA II HDD Canister KISS DA 435 1. Table of Contents 1. Table of Contents... 1 2. Important Information...

More information

ImagingSpectrometer ihr320 ihr550. OperationManual PartnumberJ81092rev.G

ImagingSpectrometer ihr320 ihr550. OperationManual PartnumberJ81092rev.G ihrfulyautomated ImagingSpectrometer ihr320 ihr550 OperationManual PartnumberJ81092rev.G ihr Fully Automated Imaging Spectrometer Operation Manual http://www.horiba.com Rev. G i Copyright 2012, 2013 by

More information

Automated Tuner System Power Distribution Hub

Automated Tuner System Power Distribution Hub User Guide Automated Tuner System Power Distribution Hub Model MT1020B MT1020-340 (Rev B) 12/11 User Guide Automated Tuner System Power Distribution Hub Model MT1020B 2900 Inland Empire Boulevard Ontario,

More information

3-4 SAS/SATA II HDD Canister Entry version USER S MANUAL XC-34D1-SA10-0-R. Document number: MAN A

3-4 SAS/SATA II HDD Canister Entry version USER S MANUAL XC-34D1-SA10-0-R. Document number: MAN A 3-4 SAS/SATA II HDD Canister Entry version XC-34D1-SA10-0-R USER S MANUAL Document number: MAN-00077-A ii Preface Important Information Warranty Our product is warranted against defects in materials and

More information

Polarizers v. 3.0 (12 Apr 2007)

Polarizers v. 3.0 (12 Apr 2007) Polarizers v. 3.0 (12 Apr 2007) Polarizers v. 3.0 (12 Apr 2007) Polarizers for Spex Spectrofluorometers and FluorEssence Software Operation Manual Rev. 3.0 i Polarizers v. 3.0 (12 Apr 2007) Copyright HORIBA

More information

Contents. HP E1586A Rack Mount Terminal Panel User s Manual

Contents. HP E1586A Rack Mount Terminal Panel User s Manual Contents HP E1586A Rack Mount Terminal Panel User s Manual Description... 5 Connecting to VXIbus Instruments... 5 Interconnect Cables... 5 Terminal Block Connections... 6 Using the Terminal Panel for Reference

More information

20W Tungsten-Halogen Light Source. Operation Manual

20W Tungsten-Halogen Light Source. Operation Manual 20W Tungsten-Halogen Light Source - ASB-W-020 - Operation Manual Important Safety Notices 1. Never look directly into the light beam, including through the cooling fan while light is on, as this can cause

More information

HX1 Bluetooth Stereo Headset

HX1 Bluetooth Stereo Headset HX1 Bluetooth Stereo Headset Ref. nr.: 130.115 INSTRUCTION MANUAL V1.1 ENGLISH Congratulations to the purchase of this Max product. Please read this manual thoroughly prior to using the product in order

More information

ENGLISH Flexstrip Set RGB

ENGLISH Flexstrip Set RGB MANUAL ENGLISH Flexstrip Set RGB V1 Highlite International B.V. Vestastraat 2 6468 EX Kerkrade the Netherlands Table of contents Warning... 2 Safety Instructions... 2 Operating Determinations... 4 Connection

More information

User Manual. pe-100 fiber. DOC-014 Iss 3 1

User Manual. pe-100 fiber. DOC-014 Iss 3 1 User Manual pe-100 fiber DOC-014 Iss 3 1 Table of Contents 1. Introduction... 3 2. Safety Precautions... 4 3. pe-100 fiber Version... 6 4. Combined pe-100 fiber Version... 7 5. Operation Manual Control...

More information

User Guide. Control Box. RoscoLED TM.

User Guide. Control Box. RoscoLED TM. RoscoLED TM Control Box User Guide This guide applies to the following RoscoLED Control Box models: RoscoLED Control Box 300W/Static White (293 22250 0000) RoscoLED Control Box 400W/VariWhite (293 22260

More information

FastGene FAS-Nano. Catalogue Number GP06-LED

FastGene FAS-Nano. Catalogue Number GP06-LED FastGene FAS-Nano Catalogue Number GP06-LED Content 1. SAFETY INFORMATION 2 2. CONTACT DETAILS 2 3. WARRANTY 3 4. REGULATORY NOTICE 3 5. IMPORTANT NOTICE 3 6. SPECIFICATION 4 7. NOMENCLATURE 5 8. OPERATING

More information

D-Pack 6 Classic ORDERCODE 50315

D-Pack 6 Classic ORDERCODE 50315 D-Pack 6 Classic ORDERCODE 50315 Congratulations! You have bought a great, innovative product from Showtec. The Showtec D-Pack 6 Classic brings excitement to any venue. Whether you want simple plug-&-play

More information

ENGLISH Flexstrip Set IP44 RGB

ENGLISH Flexstrip Set IP44 RGB MANUAL ENGLISH Flexstrip Set IP44 RGB V1 Highlite International B.V. Vestastraat 2 6468 EX Kerkrade the Netherlands Table of contents Warning... 2 Safety Instructions... 2 Operating Determinations... 4

More information

Emerson Network Power provides customers with technical support. Users may contact the nearest Emerson local sales office or service center.

Emerson Network Power provides customers with technical support. Users may contact the nearest Emerson local sales office or service center. Liebert PSA iton User Manual Version: V2.8 Revision date: November 14, 2005 Emerson Network Power provides customers with technical support. Users may contact the nearest Emerson local sales office or

More information

OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide

OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide 2004-2009, AFL Telecommunications, all rights reserved. COM4-00-1001 Revision E, 2009-06-16 Specifications

More information

Xxxxx Xxxxxx OM-CP-QUADVOLT. OM-CP-OCTVOLT 8-Channel Low Level DC Voltage Data Logger. 4-Channel Low Level DC Voltage Data Logger INSTRUCTION SHEET

Xxxxx Xxxxxx OM-CP-QUADVOLT. OM-CP-OCTVOLT 8-Channel Low Level DC Voltage Data Logger. 4-Channel Low Level DC Voltage Data Logger INSTRUCTION SHEET OM-CP-QUADVOLT Xxxxx Xxxxxx 4-Channel Low Level DC Voltage Data Logger OM-CP-OCTVOLT 8-Channel Low Level DC Voltage Data Logger INSTRUCTION SHEET MQS5086/1216 Shop online at omega.com SM e-mail: info@omega.com

More information

QUICK START GUIDE ODT-HH-MAH300 Handheld

QUICK START GUIDE ODT-HH-MAH300 Handheld FACTORY AUTOMATION QUICK START GUIDE ODT-HH-MAH300 Handheld Laser Class 2M EN 60825-1 The General Terms of Delivery for Products and Services of the Electrical Industry, published by the Central Association

More information

Operation Instruction for CNI Model with PSU-III-LCD

Operation Instruction for CNI Model with PSU-III-LCD Operation Instruction for CNI Model with PSU-III-LCD Caution-Use of controls or adjustments or performance of procedures other than those specified in this manual may result in hazardous radiation exposure.

More information

User Manual. For more information, visit

User Manual. For more information, visit User Manual H1 For more information, visit www.humaxdigital.com/me 2 What s in the box? H1 Remote Control / Batteries Quick Start Guide AC Adaptor Quick start Guide Note: Accessories may vary according

More information

XBDM. 1015LV, 1020LV, 1030LV, 1020HV Models USER & INSTALLATION MANUAL BYPASS DISTRIBUTION MODULE

XBDM. 1015LV, 1020LV, 1030LV, 1020HV Models USER & INSTALLATION MANUAL BYPASS DISTRIBUTION MODULE XBDM 1015LV, 1020LV, 1030LV, 1020HV Models USER & INSTALLATION MANUAL www.xpcc.com 2013 Xtreme Power Conversion Corporation. All rights reserved. Table of Contents IMPORTANT SAFETY INSTRUCTIONS:... 4 INTRODUCTION...

More information

EQUIPMENT Operation Manual

EQUIPMENT Operation Manual EQUIPMENT Operation Manual 7700 Label # 988830 Loctite Handheld LED Light Sources Part Number 1427231 Model 7700 Light Source II Part Number 98595 Model 7703 Light Source TABLE OF CONTENTS 1. PLEASE OBSERVE

More information

CVU-200-KIT. 200 V Bias Tee Kit. Description. Parts list / October 2014 *P A* 1

CVU-200-KIT. 200 V Bias Tee Kit. Description. Parts list / October 2014 *P A* 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com CVU-200-KIT 200 V Bias Tee Kit Description The CVU-200-KIT Bias Tee Kit consists of three 2600-RBT-200

More information

CVU-3K-KIT. 3 kv Bias Tee Kit. Description. Parts list / October 2014 *P * 1

CVU-3K-KIT. 3 kv Bias Tee Kit. Description. Parts list / October 2014 *P * 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com CVU-3K-KIT 3 kv Bias Tee Kit Description The CVU-3K-KIT Bias Tee Kit consists of three bias tees for

More information

TVAC25100 TVAC25110 User manual

TVAC25100 TVAC25110 User manual TVAC25100 TVAC25110 User manual Version 11/2010 Original user manual. Keep for future use. 12 Introduction Dear Customer, Thank you for purchasing this product. This product meets the requirements of the

More information

TC100 Precision Thermocouple Calibrator

TC100 Precision Thermocouple Calibrator TC100 Precision Thermocouple Calibrator Table of Contents 1. Introduction.....................1 2. Accessories....................5 3. Set-Up Basic/Advanced...........5 4. Operating Procedure.............6

More information

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com Model 2460-KIT Screw Terminal Connector Kit Description The Model 2460-KIT Screw Terminal Connector Kit

More information

Contents Safety.. 1 General. 1 Electrical.. 1 Warning 1 Working Principle 2 Unpacking Instructions 2 Specifications 3 Installation.. 3 Introduction..

Contents Safety.. 1 General. 1 Electrical.. 1 Warning 1 Working Principle 2 Unpacking Instructions 2 Specifications 3 Installation.. 3 Introduction.. Contents Safety.. General. Electrical.. Warning Working Principle 2 Unpacking Instructions 2 Specifications 3 Installation.. 3 Introduction.. 4 Description of keys and operating method.. 5 Operating Procedures.

More information

UHP-T-LED User Manual

UHP-T-LED User Manual UHP-T-LED User Manual Ver 03 July 2014 UHP-T-LED System Description The UHP-T-LED is an Ultra High Power LED light source for fluorescence microscopy, Optogenetics, High power illumination and other applications.

More information

CX1 Outdoor Color-Changing Wireless Bluetooth Speaker Please fully charge the speaker prior to first use!

CX1 Outdoor Color-Changing Wireless Bluetooth Speaker Please fully charge the speaker prior to first use! CX1 Outdoor Color-Changing Wireless Bluetooth Speaker 130.109 Please fully charge the speaker prior to first use! 1 ENGLISH Congratulations to the purchase of this Max product. Please read this manual

More information

Obtaining Documentation and Submitting a Service Request, page xvii Safety Warnings, page xvii Safety Guidelines, page xx

Obtaining Documentation and Submitting a Service Request, page xvii Safety Warnings, page xvii Safety Guidelines, page xx Preface Obtaining Documentation and Submitting a Service Request, page xvii Safety s, page xvii Safety Guidelines, page xx Obtaining Documentation and Submitting a Service Request For information on obtaining

More information

8 Button RS232/IR. Control Panel. MuxLab Inc A / SE A

8 Button RS232/IR. Control Panel. MuxLab Inc A / SE A 8 Button RS232/IR Control Panel 500816 MuxLab Inc. 2016 94-000833-A / SE-000833-A SAFETY PRECAUTIONS To insure the best use from the product, please read all instructions carefully before using the device.

More information

POWER SUPPLY UNIT PSU-3

POWER SUPPLY UNIT PSU-3 INSTRUCTION MANUAL NEITS INSTRUMENTS CO., LTD. International Division 2013.09 Thank you for purchasing the NEITZ Power Supply Unit PSU-3. This unit has been designed for the use with NEITZ Binocular Indirect

More information

It s Under Control. Installation and Operation Guide CPB-1. Control Port Connecting Block V 1.1

It s Under Control. Installation and Operation Guide CPB-1. Control Port Connecting Block V 1.1 Installation and Operation Guide 70-210043-17 V 1.1 Copyright 2008 Remote Technologies Incorporated All rights reserved. 2 DECLARATIONS 117 612 914 DECLARATION OF CONFORMITY (DOC) The Declaration of Conformity

More information

Owner s Instruction Manual

Owner s Instruction Manual Owner s Instruction Manual Advanced Healthcare Telephone Model 5150 Contents IMPORTANT SAFETY INSTRUCTIONS...3 BOX CONTENTS...4 FEATURES...4 ON/OFF SWITCH...4 DIAL BUTTONS...4 RECEIVER VOLUME CONTROL...4

More information

SIRIUS CAPACITOR MODULE User Manual. Model number: A-1.7C-M-SD-G Version 1.0; Release Date: April 2018

SIRIUS CAPACITOR MODULE User Manual. Model number: A-1.7C-M-SD-G Version 1.0; Release Date: April 2018 SIRIUS CAPACITOR MODULE User Manual Model number: 3550-48-A-1.7C-M-SD-G Version 1.0; Release Date: April 2018 Introduction The Sirius Capacitor Module ( Sirius ) is supercapacitor-based storage that uses

More information

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only.

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only. Concorde 600 Power Supply Operation Manual Cat.no. R10-1001011 *This instrument is intended for laboratory use only http://www.recenttec.com E-mail : support@recenttec.com Version 1.1 Packing List x 1

More information

Vista SD Portable Remote Control Pendant

Vista SD Portable Remote Control Pendant S&C Vista SD Underground Distribution Switchgear Pad-Mounted and Vault-Mounted Styles Outdoor Distribution (17.5 kv and 29 kv) With Visi-Gap Load Interrupter Switches and Visi-Gap Fault Inetrrupters Vista

More information

EPS Power Supply

EPS Power Supply EPS - 600 Power Supply Installation and Operation Manual Version 1.0 *This instrument is intended for laboratory use only Index A. Important Notice ----------------------------------------------------------------

More information

User s Manual. PCI Bus Expansion Adapter for PCI Bus PC-Slot EAD(PCI)BE PCI Bus Expansion Adapter for Low Profile PCI PC-Slot EAD(LPCI)BE

User s Manual. PCI Bus Expansion Adapter for PCI Bus PC-Slot EAD(PCI)BE PCI Bus Expansion Adapter for Low Profile PCI PC-Slot EAD(LPCI)BE PC-HELPER PCI Bus Expansion Adapter for PCI Bus PC-Slot EAD(PCI)BE PCI Bus Expansion Adapter for Low Profile PCI PC-Slot EAD(LPCI)BE User s Manual CONTEC CO.,LTD. Check Your Package Thank you for purchasing

More information

Operating Instructions

Operating Instructions 9000 Operating Instructions Contents Introduction 1 Operating Instructions 2-5 Demonstrations 6-8 Storing/Handling/Cleaning 9 Safety Precautions 9-10 Specifications 10 FCC Compliance Statement 11-12 Limited

More information

Operating Instructions

Operating Instructions Bracken Hill South West Industrial Estate Peterlee Co Durham SR8 2SW ENGLAND Tel: +44(0)191 5863511 www.seaward.co.uk sales@seaward.co.uk service@seaward.co.uk Part Number 344A550 Revision 1 2006 Seaward

More information

Logitech Alert 700i/750i System Requirements & Support Guide

Logitech Alert 700i/750i System Requirements & Support Guide Logitech Alert 700i/750i System Requirements & Support Guide Contents System Requirements............................ 3 Product Information............................ 4 Contact Us..................................

More information

Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide

Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide Safety precautions Observe the following safety precautions before using this product and any associated instrumentation.

More information

This 4200-RM Rack Mount Kit is for installation in 4200-CAB series cabinets only.

This 4200-RM Rack Mount Kit is for installation in 4200-CAB series cabinets only. Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 (440) 248-0400 Fax: (440) 248-6168 www.keithley.com Model 4200-RM Rack Mount Kit Packing List Introduction NOTE This 4200-RM Rack Mount

More information

USER MANUAL. Uninterruptible Power Supply Line-interactive VCL Series UPS VA. GE Critical Power

USER MANUAL. Uninterruptible Power Supply Line-interactive VCL Series UPS VA. GE Critical Power Critical Power USER MANUAL Uninterruptible Power Supply Line-interactive VCL Series UPS 400 600 800 1000 1500 VA GE Consumer & Industrial SA General Electric Company CH 6595 Riazzino (Locarno) Switzerland

More information

Installation and Operation

Installation and Operation S&C TripSaver II Cutout-Mounted Recloser Outdoor Distribution (15 kv and 25 kv) Cordless Power Module Installation and Operation Table of Contents Section Page Section Page Introduction Qualified Persons....

More information

Installation manual plugs and connectors with screw connection (16/32 A)

Installation manual plugs and connectors with screw connection (16/32 A) EN Installation manual plugs and connectors with screw connection (16/32 60003213 Issue 04.2016 2016-04-01 Table of contents 1 About this manual 3 1.1 Structure of the warnings 3 1.2 Symbols used 4 1.3

More information

MX3 BT LED Party Tube

MX3 BT LED Party Tube MX3 BT LED Party Tube Ref. nr.: 130.121 INSTRUCTION MANUAL V1.1 ENGLISH Congratulations to the purchase of this Max product. Please read this manual thoroughly prior to using the product in order to benefit

More information

Manual No: Revision: D. Mag Retrofit Probe. Installation Guide

Manual No: Revision: D. Mag Retrofit Probe. Installation Guide Manual No: 577013-791 Revision: D Mag Retrofit Probe Installation Guide Notice Veeder-Root makes no warranty of any kind with regard to this publication, including, but not limited to, the implied warranties

More information

9040/9040UK. Users Manual. Phase Rotation Indicator

9040/9040UK. Users Manual. Phase Rotation Indicator 9040/9040UK Phase Rotation Indicator Users Manual PN 2438546 April 2005 2005 Fluke Corporation. All rights reserved. Printed in China All product names are trademarks of their respective companies. LIMITED

More information

Dear Customer, User Memo: Please visit us on facebook or twitter! Thank you for purchasing this product.

Dear Customer, User Memo: Please visit us on facebook or twitter! Thank you for purchasing this product. Dear Customer, Thank you for purchasing this product. For optimum performance and safety, please read these instructions carefully. User Memo: Date of purchase: Dealer name: Dealer address: Dealer website:

More information

3M Constant Monitors WSMONITOR1, WSMONITOR 2, WSMONITOR3

3M Constant Monitors WSMONITOR1, WSMONITOR 2, WSMONITOR3 3M Constant Monitors WSMONITOR1, WSMONITOR 2, WSMONITOR3 User s Guide WSMONITOR1 WSMONITOR2 WSMONITOR3 Table of Contents Section Page SAFETY INFORMATION...3 Introduction...5 Installation...6 Operation...9

More information

User s Guide. OM-CP-PRHTEMP2000 Pressure, Humidity and Temperature Data Logger with LCD

User s Guide. OM-CP-PRHTEMP2000 Pressure, Humidity and Temperature Data Logger with LCD User s Guide OM-CP-PRHTEMP2000 Pressure, Humidity and Temperature Data Logger with LCD OM-CP-PRHTEMP2000 Product Overview The OM-CP-PRHTEMP2000 is a pressure, temperature and humidity data logger with

More information

3M Duplex Polishing Machine 6851-E with Universal Dual Holder Instructions

3M Duplex Polishing Machine 6851-E with Universal Dual Holder Instructions 3M Duplex Polishing Machine 6851-E with Universal Dual Holder Instructions Safety Information Read, understand and follow all safety information contained in these user instructions prior to use of the

More information

Omnitron Systems Technology, Inc. 1. iconverter. 19-Module Managed Power Chassis User s Manual

Omnitron Systems Technology, Inc. 1. iconverter. 19-Module Managed Power Chassis User s Manual Omnitron Systems Technology, Inc. 1 iconverter 19-Module Managed Power Chassis User s Manual 27 Mauchly, #201, Irvine, CA 92618 Phone: (949) 250-6510; Fax: (949) 250-6514 2 Omnitron Systems Technology,

More information

Wall. No opening (Example: LA-030-W)

Wall. No opening (Example: LA-030-W) Thank you very much for choosing an EIZO product. Please read this User s Manual carefully to familiarize yourself with safe and effective usage procedures. Please retain this manual for future reference.

More information

3M AEROSPACE & COMMERCIAL TRANSPORTATION DIVISION St. Paul, MN USA or or Made in USA

3M AEROSPACE & COMMERCIAL TRANSPORTATION DIVISION St. Paul, MN USA  or or Made in USA 3M AEROSPACE & COMMERCIAL TRANSPORTATION DIVISION St. Paul, MN 55144-1000 USA www.3m.com/aerospace or 1-800-364-3577 or 1-800-235-2376 Made in USA with US and Globally Sourced Materials 3M 2015. All rights

More information

EQUIPMENT Operation Manual

EQUIPMENT Operation Manual EQUIPMENT Operation Manual Loctite LED Light Sources Part Number 1427232 Model 7700-HD, LED Light Source II, 395 nm Peak Part Number 98665 Model 7703-HD, >400 nm Pure Visible Light Source 1 TABLE OF CONTENTS

More information

Powerware 3105 UPS User s manual

Powerware 3105 UPS User s manual Powerware 3105 UPS 2005 Eaton Corporation All Rights Reserved The contents of this manual are the copyright of the publisher and may not be reproduced (even extracts) unless permission granted. Every care

More information

Powerware 3105 UPS User s manual

Powerware 3105 UPS User s manual Powerware 3105 UPS 2005 Eaton Corporation All Rights Reserved The contents of this manual are the copyright of the publisher and may not be reproduced (even extracts) unless permission granted. Every care

More information

Xxxxx Xxxxxx Bridge / Strain Gauge Data Logger

Xxxxx Xxxxxx Bridge / Strain Gauge Data Logger -SERIES Xxxxx Xxxxxx Bridge / Strain Gauge Data Logger INSTRUCTION SHEET MQS5776/0117 Shop online at omega.com SM e-mail: info@omega.com For latest product manuals: www.omegamanual.info Product Notes The

More information

User Manual PUH4-H2. 4K HDMI Splitter 1x4. All Rights Reserved. Version: PUH4-H2_2016V1.2

User Manual PUH4-H2. 4K HDMI Splitter 1x4. All Rights Reserved. Version: PUH4-H2_2016V1.2 User Manual PUH4-H2 All Rights Reserved Version: PUH4-H2_2016V1.2 Preface Read this user manual carefully before using this product. Pictures shown in this manual is for reference only, different model

More information

AURA SSD FOR MAC PRO. Installation Guide

AURA SSD FOR MAC PRO. Installation Guide AURA SSD FOR MAC PRO Installation Guide CONTENTS Introduction... 1 1.1 System Requirements 1.2 Package Contents 1.3 About This Manual Installation... 2 2.1 Preparing the Mac Pro 2.2 Installing the Aura

More information

87421/22A Power Supply. Operating and Service Manual

87421/22A Power Supply. Operating and Service Manual 87421/22A Power Supply Operating and Service Manual Agilent Part Number: 87421-90001 Printed in USA April 2001 Supersedes: September 1998 Notice The information contained in this document is subject to

More information

Wa r n i n g-wa r n i n g-wa r n i n g

Wa r n i n g-wa r n i n g-wa r n i n g Installation Instructions of the Power Analyzer Wa r n i n g-wa r n i n g-wa r n i n g Read and understand this manual before connecting device. Death, fire or serious injury can occur from using equipment

More information

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide Model P4017 Single Channel USB Oscilloscope Quick Start Guide General Warranty BNC warrants that the product will be free from defects in materials and workmanship for 3 years from the date of purchase

More information

Reference Manual. HDMI FMC Card. Revision A 2018/02/ /02/ Macnica, Inc.

Reference Manual. HDMI FMC Card. Revision A 2018/02/ /02/ Macnica, Inc. Reference Manual HDMI FMC Card Revision A 2018/02/05 2018/02/05 2017 Macnica, Inc. http://www.m-pression.com 2018 Macnica, Inc. Disclaimer The information in this document (hereinafter called Information

More information

Agilent Technologies E5339A 38-Pin Low-Voltage Probe

Agilent Technologies E5339A 38-Pin Low-Voltage Probe Agilent Technologies E5339A 38-Pin Low-Voltage Probe Installation Note The 38-pin low-voltage probe provides a convenient way to connect two Agilent Technologies logic analyzer probe cables to a small

More information

IAQ-CALC TM Indoor Air Quality Meter

IAQ-CALC TM Indoor Air Quality Meter ENERGY AND COMFORT Indoor Air Quality IAQ-CALC TM Indoor Air Quality Meter Model 7535 Operation and Service Manual Copyright TSI Incorporated / May 2007 / All rights reserved. Address TSI Incorporated

More information

THECHARGEHUB.COM. User Manual. For Square & Round Models

THECHARGEHUB.COM. User Manual. For Square & Round Models THECHARGEHUB.COM User Manual For Square & Round Models User Manual THECHARGEHUB.COM 7-Port USB Universal Charging Station Table of Contents General Safety Information...2 Care and Maintenance...3 Introduction...4

More information

Start Here. All-in-One Printer. Print Copy Scan

Start Here. All-in-One Printer. Print Copy Scan Start Here All-in-One Printer Print Copy Scan Note: A USB 2.0 cable is required to connect your Kodak All-in-One printer to your Windows or Macintosh OS computer. Available in the Kodak online store or

More information

Keysight E5864A Removable Hard Drive for Series Logic Analyzers. Installation Guide

Keysight E5864A Removable Hard Drive for Series Logic Analyzers. Installation Guide Keysight E5864A Removable Hard Drive for 16850-Series Logic Analyzers Installation Guide Notices Keysight Technologies 2013-2014 No part of this manual may be reproduced in any form or by any means (including

More information

DVI Extender over Fiber with EMI Shielding

DVI Extender over Fiber with EMI Shielding DVI Extender over Fiber with EMI Shielding Model #: FO-DVI-1000M-EMI 2010 Avenview Inc. All rights reserved. The contents of this document are provided in connection with Avenview Inc. ( Avenview ) products.

More information

Operating Instructions

Operating Instructions PRIMETEST 100 PRIMETEST 100 Bracken Hill South West Industrial Estate Peterlee Co Durham SR8 2SW ENGLAND Tel: +44(0)191 5863511 www.seaward.co.uk sales@seaward.co.uk service@seaward.co.uk Part Number

More information

IRC 1100 Kalibrierstrahler / Calibration Source

IRC 1100 Kalibrierstrahler / Calibration Source IRC 1100 Kalibrierstrahler / Calibration Source Betriebsanleitung User Manual IMPAC - Spezialist für berührungslose Temperaturmessung IMPAC - Specialist in non-contact thermometry Content 1 General...10

More information

MANUAL. ENGLISH Showtec PSA-161 Ordercode: Highlite International B.V. Vestastraat EX Kerkrade the Netherlands

MANUAL. ENGLISH Showtec PSA-161 Ordercode: Highlite International B.V. Vestastraat EX Kerkrade the Netherlands MANUAL ENGLISH Showtec PSA-161 V1 Highlite International B.V. Vestastraat 2 6468 EX Kerkrade the Netherlands Table of contents Warning... 2 Safety Instructions... 2 Operating Determinations... 4 Connection

More information

CrystalView DVI Multi INSTALLATION AND OPERATIONS MANUAL Stancliff Road Phone: (281)

CrystalView DVI Multi INSTALLATION AND OPERATIONS MANUAL Stancliff Road Phone: (281) CrystalView DVI Multi INSTALLATION AND OPERATIONS MANUAL 10707 Stancliff Road Phone: (281) 933-7673 Houston, Texas 77099 WWW.ROSE.COM LIMITED WARRANTY Rose Electronics warrants the CrystalView Multi to

More information

SmartBoom PRO PHS-SB100-4F, PHS-SB100-5M, PHS-SB100-U, PHS-SB200-4F, PHS-SB200-5M, PHS-SB200-U. Operating Manual

SmartBoom PRO PHS-SB100-4F, PHS-SB100-5M, PHS-SB100-U, PHS-SB200-4F, PHS-SB200-5M, PHS-SB200-U. Operating Manual SmartBoom Operating Manual PRO PHS-SB100-4F, PHS-SB100-5M, PHS-SB100-U, PHS-SB200-4F, PHS-SB200-5M, PHS-SB200-U Thank You We at Pliant Technologies, LLC want to thank you for purchasing the SmartBoom PRO

More information

* * Agilent Power Distribution Unit (PDU) Installation Guide

* * Agilent Power Distribution Unit (PDU) Installation Guide Agilent Power Distribution Unit (PDU) Installation Guide For use with Agilent PDU kits and PDU installation kits for Agilent instrument racks June 2008 Edition 7 E0608 *5000-0039* 5000-0039 Notice The

More information

Service Instructions. Conductor 420

Service Instructions. Conductor 420 Service Instructions Conductor 420 Listed, File No. E195547 Input: 480 VAC, 50/60 HZ. (Operating range 380-480 VAC) 20 AMP MAX OUTPUT W/25 AMP FUSE Output: 0-90% of Input Voltage Table of Contents General

More information

PR3400 Series 1.4Mp USB Cameras Hardware Guide

PR3400 Series 1.4Mp USB Cameras Hardware Guide PR3400 Series 1.4Mp USB Cameras Hardware Guide Manufactured by: SPOT Imaging Solutions, a division of Diagnostic Instruments, Inc. 6540 Burroughs Ave. Sterling Heights, MI 48314-2133 USA Toll-Free: 866-604-SPOT

More information

HDEXT50M USER MANUAL Extend HD Signals over CAT 5/5e/6 up to 164ft.(50m) All Rights Reserved Version: HDEXT50M_2017V1.2

HDEXT50M USER MANUAL Extend HD Signals over CAT 5/5e/6 up to 164ft.(50m) All Rights Reserved Version: HDEXT50M_2017V1.2 USER MANUAL Extend HD Signals over CAT 5/5e/6 up to 164ft.(50m) All Rights Reserved Version: _2017V1.2 Preface Read this user manual carefully before using this product. Pictures displayed in this manual

More information

Installation and Operation Manual Document Number

Installation and Operation Manual Document Number Mini Deuterium Halogen Light Source DT-Mini-2-GS Installation and Operation Manual Document Number 000-10000-110-02-1209 Offices: Ocean Optics, Inc. World Headquarters 830 Douglas Ave., Dunedin, FL, USA

More information

PC-HELPER. Expansion Adapter for Express Card Slot EAD-CE-EC. User s Manual CONTEC CO.,LTD.

PC-HELPER. Expansion Adapter for Express Card Slot EAD-CE-EC. User s Manual CONTEC CO.,LTD. PC-HELPER Expansion Adapter for Express Card Slot EAD-CE-EC User s Manual CONTEC CO.,LTD. Check Your Package Thank you for purchasing the CONTEC product. The product consists of the items listed below.

More information

EC Series. Safety Guide

EC Series. Safety Guide EC Series Safety Guide R5906018/02 01/07/2017 Barco Inc, Image Processing 3078 Prospect Park Drive, Rancho Cordova, CA, 95670, USA Phone: +1 916 859-2500 Fax: +1 916 859-2515 Support: www.barco.com/en/support

More information

Manual Version: V1.00. Video Decoder User Manual

Manual Version: V1.00. Video Decoder User Manual Manual Version: V1.00 Video Decoder User Manual Thank you for purchasing our product. If there are any questions, or requests, please do not hesitate to contact the dealer. Copyright Copyright 2016 Zhejiang

More information

1X2 HDMI Splitter with 3D Support

1X2 HDMI Splitter with 3D Support AV Connectivity, Distribution And Beyond... VIDEO WALLS VIDEO PROCESSORS VIDEO MATRIX SWITCHES EXTENDERS SPLITTERS WIRELESS CABLES & ACCESSORIES 1X2 HDMI Splitter with 3D Support Model #: SPLIT-HDM3D-2

More information

PDS8u POWER DISTRIBUTION SYSTEM USER'S MANUAL

PDS8u POWER DISTRIBUTION SYSTEM USER'S MANUAL PDS8u POWER DISTRIBUTION SYSTEM USER'S MANUAL 1 IMPORTANT SAFETY INSTRUCTION READ FIRST This symbol, whenever it appears, alerts you to the presence of uninsulated dangerous voltage inside the enclosure-voltage

More information

DP-222Q Color Video Door Phone Manual

DP-222Q Color Video Door Phone Manual DP-222Q Color Video Door Phone Manual * has 6 LEDs for nighttime operation Remotely and securely talk to visitors and unlock doors, gates, etc. from the Easily connect an secondary * Simple 2-wire connection

More information

Continuous Wave Single Frequency Lasers. Concerto Series. 442nm 25, 50, 75 & 100mW

Continuous Wave Single Frequency Lasers. Concerto Series. 442nm 25, 50, 75 & 100mW User Manual Continuous Wave Single Frequency Lasers Concerto Series 442nm 25, 50, 75 & 100mW Copyright 2012, KLASTECH GmbH. All rights reserved. Information presented herein is believed to be accurate

More information

DATA DOUBLER. Apple Mac mini (2009)

DATA DOUBLER. Apple Mac mini (2009) DATA DOUBLER Apple Mac mini (2009) TABLE OF CONTENTS Introduction... 1 1.1 Hardware Compatibility 1.1.1 Host Computer Compatibility 1.1.2 Drive Compatibility 1.2 Package Contents 1.3 About This Manual

More information

Led Illuminator and Light Stimulator. Model # LSD-1. Ver

Led Illuminator and Light Stimulator. Model # LSD-1. Ver Led Illuminator and Light Stimulator Model # LSD-1 Ver. 2.0.0 5.16 ALA Scientific Instruments Inc. 60 Marine Street, Farmingdale, NY 11735 Voice: +1 631.393.6401 fax: +1 631.393.6407 www.alascience.com

More information