1 Microprocessor Benchtop Conductivity/ Resistivity Meter SC-2300 Operation Manual 1

2 Content Table 1. Specifications 1 2. Configuration 2.1 Display Socket Keypad 3 3. Operation 3.1 Measurement Measurement function and file bit shift Temperature function and compensation Auto Read 5 4. Settings Block diagram of settings Parameter settings Entry of set-up Settings of Settings of data deletion Settings of baud rate (BAUD) Settings of system time Deletion of all stored data System reset 10 Block diagram of calibration and temperature compensation settings Settings of calibration parameter Electrode coefficient and calibration parameter Standard solution calibration parameter Temperature compensation coefficient 13 Block diagram of operations in Auto/manual data storage Auto storage Manual storage Data full Data readout and output Auto transmission Manual readout and transmission Calibration Block diagram of Calibration Standard solution preparation Calibration Entry of calibration 21 2

3 5.3.1 Electrode coefficient calibration Standard solution single-point calibration Standard solution five-point calibration Error message Instruction set for RS Alternative fittings Failure recovery Maintenance 24 3

4 1. Specifications Model SC-2300 Function µs/ms/mω/salt/tds/temp Cond us/cm ~ ms/cm in 5 measuring ranges for manual or auto switch Res. 0.00~ MΩ- cm Range Salt 0.0~70.0ppt TDS 0~2000ppm TEMP C Cond. 0.01µS/cm Res MΩ-cm Resolution Salt 0.1 ppt TDS 1 ppm TEMP 0.1 C Cond. 0.5%(±1Digit) Res. 1%(±1Digit) Accuracy Salt 0.5%(±1Digit) TDS 0.5%(±1Digit) TEMP ±0.2 C(±1Digit) Temperature PT-1000 or NTC 30K auto recognition Compensation Auto / Manual selectable Temperature Non-linear / 0.00~20.00% Linear Coefficient Ref. Temperature T ref 25 Manual calibration Single point known standard solution or electrode constant calibration Standard solution calibration Single or up to 5-point known standard solution calibration; default five standard solution 10/84/1413 us/cm & 12.88/100 ms/cm adjustable RS232 interface YES Output Data logging 450 sets (measuring value, temp., date, time & ID) Ambient temperature 0~50 C Storage Temperature -20~+70 C Display 3/4 LCD / backlit Power 4 x AA / AC ADAPTOR Battery Around hrs according to measuring range and whether the backlit is used Protection Type IP43 Dimension (L x W x H) 220 x 190 x 70mm Weight 0.8kg 1

5 2. Configuration 2.1: Display: Set-up Measurement Read Measurement unit Auto Auto temperature compensation Calibration Time reminder Data storage Data transferring speed Low battery reminder Storage ID Data reading /Transferring 2.2 Socket: Data Storage ID POWER : DC 6V Adaptor socket RS-232 : RS-232 output socket REC. : no function at the moment T/P : External temperature probe socket (prohibited if using electrode which has combined with temperature sensor). EC/DO : Resistivity or Conductivity electrode waterproof socket 2

6 2.3 Keypad: : Power switch Push it to power on and push it again to power off : Cond./Res./Salt/TDS shift key; pushing it at any time to back to : Enter calibration : Auto read Push one time for startup and push again for cancellation. : Upward or leftward key : Downward or rightward key : (ENTER) confirmation, execution, and confirmation for RS-232 manual output : (Recall) Data recall key and value selection key : (Store) Data storage key + : Enter system parameter setup, see ch : Delete all stored data, see ch : System reset, seech : Shift among five manual ranges and the auto range recognition, see ch : Enter calibration parameter settings, see ch : Enter auto continual storage settings, see ch : Enter auto continuation data transmission settings, see ch.4.6 3

7 3. Operation 3.1 Measurement: Push key to power on the instrument. It will automatically show last and be ready for new. 3.2 Measurement function shifting: 1. PushMODE to shift among Cond., Salt, and TDS in Cond. Mode. This function is unavailable in Res.. 2. PushENT+MODE to shift among five manual ranges(20.00µs /200.0µS /2000µS /20.00 ms /200.0 ms) and auto range shifting. When entering into auto range shifting, symbol will disappear after 3 seconds. Push Pushto shift to Salt Push Push to adjust the salt value. Note: Factor range is 0.4~1.5 Pushto shift TDS Push Push to adjust Total Dissolved Solids. Note: Factor range is 0.4~1.5 Push Push Push Push Remark: Before measuring salt and TDS, please finish the calibration according to the conductivity calibration steps, and then set the factor value according to the standard solution of salt and TDS. 4

8 3.3 Temperature and compensation: 1. Auto Temp. Compensation: Auto recognition of temperature probe system. (NTC30K or PT1000) Push or to adjust the temperature within in ±5. Push keys simultaneouslyto back to the original value. 2. Manual Temp. Compensation: Enter MTC 25 automatically if no temperature probe. Push or to adjust the temperature. Push simultaneously to back to the default value. 3. Temperature compensation: The temperature of the to-be-tested solution may not be 25. For standardization, the conductivity value of the solution which is not at the temperature of 25 will be calculated to that of solution at the temperature of According to the temperature coefficient required by the, three kinds of temperature compensation can be applied: Linear(Lin), Non-linear(nLin), and non compensation (0.00). See the setting method in ch Users can set by themselves as needed. 5. Temp. Coefficient (TC for abbreviation), the conductivity value of the solution will increase as the temperature rises. TC may differ from different solution. See the formula as follows: C 25 Conductivity at 25 Formula 1: Ct = C Ct 25 { 1+β( T 25 ) } Conductivity at T T Temp. of the tested solution Formula 2: β = ( CtC β Temp. Coefficient 25 ) / { C 25 ( T25 )} 6. How to measure the TC of the solution: If the user wants to get a more accurate value, the above formula can be applied to measure the TC of the solution and set an appropriate TC on the machine. Take 0.01N KCl as an instance, set the TC as non compensation (0.00%) in the machine. Control the same solution respectively at 25 and 20. You could acquire at 25(C 25 ) for 1413µS; and the at 20(C) for 1278µS According to formula 2, β1.91%. 3.4 Auto Read: In the, push AUTO to start and push ENT to measure. Then, push ENTto make the next. PushAUTOor MODE at any time will allow backing to the (See the following charts). The symbol twinkles for about 10 seconds to get a stable value. PushAUTO The symbol stop twinkling and a value will appear after the is stable. PushENT 1. PushENTfor the next 2. Push AUTOorMODE Back to the 5

9 4. Settings Block diagram of settings PushMODE+, Release + MODE to enter Settings of Power off Data overwriting settings Baud rate settings System time settings PushSTO+, Push+, Release + STO Release + to deleteallstored data for system reset (Data Clean) (RS 232) Pushor to select conductivity or specific resistivity Push or to select auto overwriting or cancel overwriting Pushor to select other transmission speed: 1200, 2400, 4800 or 9600 Push or to modify the year Push or to modify the month Push or to modify the day : ENTER key Push or to modify the hour Push or to modify the minute 6 6

10 4.1 Parameter settings: Entry of set-up : In the power-off status, push MODEfirst and then push to power on the machine; then please release first and then MODEto enter the parameter setting. Use or to select Settings of : Push ENTto enter Push or to select MΩ-cmµS/cm PushENTto confirm it Enter Settings of data deletion Settings of data deletion: Enter Settings of data deletion PushENTto enter Push or to select auto overcast or cancel auto overcast PushENTto confirm it Enter Settings of baud rate 7

11 4.1.4 Settings of baud rate (BAUD): Enter Settings of RS-232 baud rate PushENTto enter Push or Push or to select to select Push to select or PushENTto confirm it Enter Settings of system 8

12 4.1.5 Settings of system time: Enter Settings of system time PushENTto enter Push or to modify the year, then pushentto confirm it Push or to modify the month, then pushentto confirm it Push or to modify the date, then pushentto ensure Push or to modify the hour, then pushentto confirm it Push or to modify the minute, then push ENTto confirm it Back to Setting of mod, or push MODEto back to 9 7

13 4.2 Deletion of all stored data: In the power-on status, pushstofirst andthen ; after that release first and then releasestoto enter the data deletion. (Data Clean) PushENTto confirm to delete all stored data or push MODEto back to the Back to the 4.3 System reset: In the power-off status, pushcalfirst and then ; after that, release then releasecalto enter the system reset page. first and (Initial Setting) PushENTto confirm the system reset; or pushmode to back to the Back to the Factory defaults: Data deletion settings: Auto (BAUD) settings: 4800 Data transmission settings: read the stored data on the display (Sto disp) 10 8

14 Block diagram of calibration and temperature compensation settings Settings in conductivity Settings in specific resistivity Power on Calibration and temperature compensation settings. Push ENT+CAL Power on PushENT+CAL Electrode coefficient calibration settings Standard solution calibration settings Measurement temperature coefficient settings Measurement temperature settings Push to adjust compensation coefficient Push to adjust compensation coefficient Push to adjust compensation coefficient Measurement Measurement ENTER key 10 11

15 4.4 Settings of calibration parameter: In the power-on status, pushentfirst and thenpushcalto enter set-up of calibration parameter. Use to select Electrode coefficient and calibration parameter: Electrode coefficient and calibration parameter PushENTto enter the settings of calibration temperature compensation coefficient = PushENTto confirm Push to select or Push or to select PushENTto confirm it Push or to set the temperature PushENTto confirm it Back to 12

16 4.4.2 Standard solution calibration parameter: (Only suitable when the unit is conductivity) PushENTto confirm it Push or to select PushENTto confirm it Back to Temperature compensation coefficient: PushENTto confirm it Push to select or Push or to select PushENTto confirm it Push or to set the coefficient value PushENTto confirm it Back to 13

17 Block diagram of operations in Power on Auto store Push ENT+STO Storage Manual store pushsto Data store full Transmission Auto transmission PushENT+RCL Manual read and transmission PushRCLto shift circularly among 4 s The stored data is transmitted to the display The stored data is transmitted through RS232 The calibrated data is transmitted to the display The stored data is transmitted through RS232 Appear the original ID Push or to set the ID number Appear the original ID Push or to set the ID number Back to the In case of data full unable to store, will twinkle for 5 seconds Back to the If twinkles, Push or to set the minute Push or to set the second Appear recently stored data Push or to modify the inquiry site After transmission, back to the automatically CELL direct calibration Std simple point calibration Std five point calibration After transmission, back to the Push or to set the minute Push or to set the second If twinkles, it means that the data is stored Continual and data it means that the data is stored Continual transmission PushENT+RCL Cancel continual storage Read the ID, and push or to inquire other data through ID Read the stored year Read the stored date Back to the Back to the PushENT+STOto close Cancel continual storage Read the stored time Back to the Back to the ENTER key 14 14

18 4.5 Auto/manual data storage: Auto storage: In the, pushent first and then pushstoto enter the auto continual storage set-up Manual storage: In the, pushsto to enter manual store page. Appear the original ID. Push ENTto confirm it PushENTto confirm it Push or to set ID number; PushENTto confirm it and enter settings of automatic storage time. Push or to set the minute, and push for confirmation and settings of second Pus or to set ID number; Push ENTto confirm the storage Back to the Data full: When the data full storage, the display will twinkle as a reminder, and five seconds later, it will back to the original. Push or to set the seconds, and push for confirmation. (The minimum value is 5 seconds) If twinkles, it means that the data is stored According to step 4.2, delete all the data. If under auto overwriting, there is no reminder of data full. PushENTSTOto cancel the auto store Back to 15

19 4.6 Data readout and output: Auto transmission: In the, pushent first and then pushrclto enter auto transmission Manual readout and transmission: In the, pushrclto shift among 4 s circularly, and pushent to enter. PushENTto enter, and the recently stored data will appear Push or to set the minute, and pushentfor confirmation and enter the second setting. Push or to modify the inquiry site. Push Enter to confirm it Push or to set the seconds, and pushentfor confirmation. (The minimum value is 5 seconds) If twinkles, it means that Read the ID and push to inquire other data according to the ID. Push enter to read the data value and data storage year. the data is stored PushENTRCLto cancel the auto store Back to PushENTto read the data storage date PushENTto read the stored time Push ENTto read the temperature value of the data PushENT Back to 16

20 B. Transmit the stored data through RS-232 Transmit all stored data through RS-232 C. Read the calibrated data on the display C.1 Read the calibrated data on the display PushENTto transmit After the transmission, back to PushENT, and it will show the standard solution value and electrode coefficient PushENT, and it will show the calibrated temperature compensation coefficient value PushENT, and it will show the calibration record year PushENT,and it will show the calibration record date PushENTand it will show and the calibration record time PushENT Back to 17

21 C.2 Read simple-point calibration data of standard solution (when the calibration parameter is set as Std 1P calibration). PushENTto enter, and the data and electrode coefficient of standard solution appear PushENTand the calibration record year appears PushENT, and the calibration record date appears PushENTand the calibration record time appears PushENT Back to 18

22 C.3 Read five-point calibration data of standard solution (when the calibration parameter is set as Std 5P calibration). PushENTand the calibration record date appears PushENTand the calibration record time PushENT, show the calibration value of Buffer 1. If there is no data, it will show NULL. PushENT PushENT, show the calibration value of Buffer 2. If there is no data, it will show NULL. Back to PushENT, show the calibration value of Buffer 3. If there is no data, it will show NULL. D. Transmit the stored data through RS-232 Transmit all stored data through RS-232 PushENT, show the calibration value of Buffer 4. If there is no data, it will show NULL. PushENTto transmit PushENT, show the calibration value of Buffer 5. If there is no data, it will show NULL. After the transmission, back to PushENTand the calibration record year appears 19

23 5. Calibration Block diagram of Calibration PushCAL Power on Conductivity Use keys to select known standard solution value Specific resistivity Back to Use keys to select suitable coefficient C=.0100 C=0.100 C=0.475 C=10.00 Coefficient valve shinesuse keys to adjust coefficient value Back to Use keys to select standard solution value, including: 10.00uS/cm, 84.0uS/cm, 1413uS/cm, 12.88mS/cm, ms/cm five default values The instrument is reading coefficient PushMODE Back to PushMODE Back to Use keys to select known standard solution value The instrument is reading coefficient Use keys to select known standard solution value The instrument is reading coefficient PushMODE The instrument is reading coefficient Use keys to select known standard solution value PushMODE The instrument is reading coefficient Back to Back to Back to Use keys to select suitable coefficient C=.0100 C=0.100 Coefficient valve shinesuse keys to adjust coefficient value to the known standard solution value PushMODE Use keys to select known standard solution value Back to The instrument is reading coefficient ENTER key 20

24 5.1 Standard solution preparation: Put reagent KCl in an oven of for 5 hours, and then put it in a dry container until it cools down to the ambient temperature. Dissolve g KCl powder into 1 kilolitre pure water to be 0.01N KCl standard solution (1413µS/cm). 5.2 Calibration : Measured functions Calibration method Preset temperature compensation method Set CELL CONSTANT Linear(Lin) 2.00 adjustable Std 1P Prepare 5 selectable kinds of According to standard solution default Conductivity Buffer Std 5P Prepare 5 selectable kinds of According to standard solution default Buffer Special resistivity Set CELL CONSTANT Non-Linear(nLin) 5.3 Entry of calibration : Temperature coefficient produces a great influence on of conductivity, so it is suggested to control the temperature of standard solution within 25±3 during the calibration Electrode coefficient calibration: 1. When the calibration parameter is set as electrode coefficient setting (see 4.4), pushcalto enter the electrode coefficient setting page, and then pushentto enter the next page. 2. Push or to select electrode coefficient default 10.00, 0.475, or (resistivity only includes and ), in order to select a suitable electrode coefficient value. After selecting the suitable default, pushentto enter the next page. 3. Then, the electrode coefficient value on display begins to twinkle. Push or to adjust the electrode coefficient value until equal to the buffer s standard. Then, pushentto back to the Standard solution single-point calibration: 1. When the calibration parameter is set as single-point calibration (see 4.4), pushcalto enter Standard solution single-point calibration, and then pushentto enter the calibration solution selection page. 2. The display will show the first preset buffer s standard value and temperature value. Push or to modify the standard value, including five default values 10.00µS, 84.0µS, 1413µS, 12.88mS and 100.0mS. After selecting the suitable default, pushentto enter the next page. 3. The symbol begins to twinkle, indicating that the instrument begins auto calibration. After the calibration, the display shows the calibrated electrode coefficient automatically. 4. After the calibration, the display shows the calibrated electrode coefficient automatically. PushENTto back to the. 21

25 5.3.3 Standard solution five-point calibration: 1. When the calibration parameter is set as five-point calibration (see 4.4), pushcalto enter Standard solution five-point calibration, and then pushentto enter the first CAL1 page. PushMODE at any time will allow you intermit the calibration and back to the. 2. Push or to select calibration point CAL1~5. Their default values are respectively 10.00µS, 84.0µS, 1413µS, 12.88mS and 100.0mS. You can select any calibration point to make single-point, two-point, three-point, four-point, and five-point calibration. Clean and dry the electrode, and then put it into the buffer solution. Push ENTto enter the calibration page. 3. The symbol begins to twinkle, indicating that the instrument begins auto calibration. After the calibration, the display shows the calibrated electrode coefficient automatically. 4. After the calibration, the display shows the calibrated electrode coefficient automatically. 5. PushENTto enter the second calibration (CAL2) and finish the first point calibration. PushMODEto store the calibration data and the system will exit calibration and back to. 6. Other CAL2~CAL5 calibration applies the same steps as CAL1. Feature: The design of this instrument is free of the calibration sequence, so after finishing any point calibration, you can pushentto enter the next point calibration, and pushmodeto store the calibration data. Then, exit the calibration and back to the to finish the calibration of any one or a few points Error messages: When the display shows error messages of failure calibration, please refer to chapter 8 Failure Recovery to remove the errors and restart the calibration. 22

26 6. Instruction set for RS232 The communication between the instrument and the Windows Hyper Terminal is set as follows: 1. Click Starton the lower left, and select All Programs Accessories Communication Hyper Terminal, and the display will show a window of Hyper Terminal. 2. Set icon and online name for Hyper Terminal before entering the ONLINE window. 3. Select communication terminal (for example, select connect COM1) in Use online (N):in ONLINE window, and then enter the window COM1 content. 4. In transmission bit per second (B):in COM1 content, select baud rate=9600 (It should be consistent with controller settings), data bit (D)=8, parity check(p)=none, stop bit (S)= 1, flow control (E) =none. 5. Click disconnet 6. Click filecontent, and enter the window XXX content. 7. Click setting valueascii settingto enter ASCII setting window. 8. In ASCII setting window, select newline at the end of each line (S),Respond to the input characters (E),Add LF at the end of each input line (A) and newline if exceeding the width of terminal (W), and then leave ASCII setting window by pushing confirm. 9. Push confirmto leave XXX content window. 10. Push callto link the communication. 7. Alternative fittings Calibration solution types and description Order number Conductivity electrode C Special resistivity electrode C Special resistivity electrode C Flow cell 8-TF µs KCl Calibration Solution in 16-oz(about 473ml) glass bottle µs KCl Calibration Solution in 500-ml leak-proof bottle ,413 µs KCl Calibration Solution in 500-ml leak-proof bottle ,880µS KCl Calibration Solution in 500-ml leak-proof bottle ,000S KCl Calibration Solution in 16-oz(about 473ml) glass bottle

27 8. Failure recovery Failure phenomenon Possible causes Disposition ERR1 The electrode coefficient has a too large deviation when using the standard solution to calibrate. Replace new standard solution for another calibration. For others, please refer to the disposition of deviation in the ERR2 Measurement deviation The is unstable when using the standard solution in calibration. The electrode is contaminated. The electrode coefficient (C) is wrongly set. The temperature coefficient (TC) is wrongly set. The electrode is damaged or broken-down. There are air bubbles around the electrode. Instrument fault The electrode is influenced by air bubbles. The electrode is polluted. Electrode fault. Instrument fault. The electrode plugs is not inserted into the required place. following. Please refer to the disposition of unstable in the following. Clean the electrode. Modify the electrode coefficient. Modify the temperature compensation. coefficient. Readjust the coefficient. Stir the sample with the electrode to remove bubbles. Send to the original factory for repair Unstable Stir the sample with the electrode to remove bubbles. Clean the electrode. Replace the electrode. Send to the original factory for repair. The conductivity value is Check the electrode plugs and reinsert it into zero or the special resistivity the required place. value is out of the range. The electrode is broken. Replace the electrode. Temperature deviation Not soak in the sample deeply enough. Deepen the soaked length to ensure the accuracy of temperature. Deviation of standard comparison. Use standard thermometer comparison to correct the difference in temperature. The temperature sensor is broken. Replace the electrode. Instrument fault. Send it back to the local distributor for repair. Temperature is wrongly The electrode plugs is not inserted into Check the electrode plugs and reinsert it into displayed. the required place. the required place. The temperature sensor is broken. Replace the electrode. Instrument fault Send it back to the local distributor for repair. 9. Maintenance 9.1 Maintenance of instrument: Please keep it in dry and ventilated environment. Use slightly wet soft cloth to wipe the instrument s surface. For pollution by grease, please remove the stain with waxed soft cloth. Solvent is strictly prohibited. 9.2 Battery replacement: a. When the display showslobat during the instrument works, please replace with alkali dry cells. b. Please power off the instrument before open the battery cover. c. Take out the old battery and put in 4 new AA batteries in correct directions. Reinstall the battery cover to finish the battery replacement. d. Caution! The battery replacement should be finished within 20 seconds; otherwise the date data will disappear and should be reset. 9.3 Maintenance of electrode: The substances on the measuring side of electrode will influence the value and cause deviation. Please clean and maintain it regularly. Use soft brush or cotton stick to clean the graphite surface of electrode. Pollution types Cleaning methods Pollution by grease Use mild degreasing agent to clean the electrode. Pollution by dirt Soak in 2~5% HCl for about 5 minutes. 24

28 25

Introduction: Features:

Introduction: Features: Index Features:... 3 Specifications:... 4 Accessories:... 5 Functions of Keyboard:... 7 Preparation:... 8 ... 8 ... 8 Calibration:... 8 ... 8 ... 9 ...

More information

C100 Handheld Cond./TDS/Sal. Meter. Instruction Manual

C100 Handheld Cond./TDS/Sal. Meter. Instruction Manual C100 Handheld Cond./TDS/Sal. Meter Instruction Manual Table of Contents 1. Introduction ---------------------------------------------------------------------------------------------------------------------------

More information

A150 Benchtop Conductivity Meter Instruction Manual

A150 Benchtop Conductivity Meter Instruction Manual A150 Benchtop Conductivity Meter Instruction Manual BANTE INSTRUMENTS CO., LTD A150 Benchtop Conductivity Meter 1 Introduction Thank you for selecting the A150 benchtop conductivity meter. This manual

More information

DDS-12DW Benchtop Conductivity Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

DDS-12DW Benchtop Conductivity Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD DDS-12DW Benchtop Conductivity Meter Instruction Manual BANTE INSTRUMENTS CO., LTD DDS-12DW Benchtop Conductivity Meter 1 Introduction Thank you for selecting the DDS-12DW benchtop conductivity meter.

More information

INSTRUCTION MANUAL CONDUCTIVITY-METER MODEL CC-01

INSTRUCTION MANUAL CONDUCTIVITY-METER MODEL CC-01 INSTRUCTION MANUAL CONDUCTIVITY-METER MODEL CC-01 27618366 / 74 EL-221, MIDC Electronic Zone, Mhape, Navi Mumbai-400701. Tel: CHAPTER 1 INTRODUCTION Contech CC-01 measures Conductivity, Total Dissolved

More information

Benchtop Multi-parameter Water Quality Meter

Benchtop Multi-parameter Water Quality Meter LMPWM-A Series Benchtop multi-parameter water quality meter LMPWM-A series is comes with different measurement modes for quality and testing of water on different parameters. It measures ph, Conductivity,

More information

Portable Conductivity Meter M1410. User Manual 1.002

Portable Conductivity Meter M1410. User Manual 1.002 Portable Conductivity Meter M1410 User Manual 1.002 General information This manual describes the operation of the Portable Conductivity Meter M1410. Please read this entire manual before operating this

More information

WK7200. Water Quality Test Kit USER S MANUAL

WK7200. Water Quality Test Kit USER S MANUAL WK7200 USER S MANUAL Water Quality Test Kit INTRODUCTION: Thank you for selecting model WK7200 microprocessor-based waterproof Water Quality Test Kit. It is used to measure a wide range of ph, ORP, Conductivity,

More information

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements www.adwainstruments.com Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting

More information

USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter

USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter www.adwainstruments.com 1 Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting operations. This instrument

More information

Instruction Manual. Laboratory EC Meter NeoMet Series. EC-400L (EC/TDS/ Salinity/Temp)

Instruction Manual. Laboratory EC Meter NeoMet Series. EC-400L (EC/TDS/ Salinity/Temp) Laboratory EC Meter NeoMet Series EC-400L (EC/TDS/ Salinity/Temp) istek, Inc. Room #1011, Hanshin IT?Tower 235 Guro-Dong, Guro-Gu, Seoul, Korea Phone : +82-2-2108-8400 Fax : +82-2-2108-8430 Website: www.istek.co.kr

More information

Bante902 Benchtop ph/conductivity Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

Bante902 Benchtop ph/conductivity Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD Bante902 Benchtop ph/conductivity Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante902 Benchtop ph/conductivity Meter 1 Introduction Thank you for selecting the Bante902 benchtop ph/conductivity

More information

Conductivity and TDS Meter Pen Style Water Quality Meter

Conductivity and TDS Meter Pen Style Water Quality Meter User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Conductivity and TDS Meter Pen Style Water Quality Meter Model EC150

More information

User's Guide. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500

User's Guide. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500 User's Guide ExStik EC500 ph/conductivity / TDS / Salinity / Temperature Meter EC500 Introduction Congratulations on your purchase of the ExStik EC500 ph/conductivity/total Dissolved Solids (TDS) / Salinity

More information

Datalogging Conductivity/ TDS Meter. Instruction Manual

Datalogging Conductivity/ TDS Meter. Instruction Manual Datalogging Conductivity/ TDS Meter 850039 Instruction Manual 1 TABLE OF CONTENTS 1. INTRODUCTION...3 2. PANEL DESCRIPTION...4 3. MEASURING PROCEDURE...5 3-A General Measurement...5 3-B Auto and Manual

More information

CX3100 Intelligent Conductivity Transmitter

CX3100 Intelligent Conductivity Transmitter CX3100 Intelligent Conductivity Transmitter Operation Manual CX3100 CONTENTS Precautions for Installation Brief Instruction 1 1. Specifications 4 2. Assembly and Installation 2.1 Transmitter Installation

More information

User's Guide. ExStik EC400. Conductivity / TDS / Salinity / Temperature Meter MODE HOLD

User's Guide. ExStik EC400. Conductivity / TDS / Salinity / Temperature Meter MODE HOLD User's Guide ExStik EC400 Conductivity / TDS / Salinity / Temperature Meter MODE HOLD Introduction Congratulations on your purchase of the ExStik EC400 Conductivity/Total Dissolved Solids (TDS) / Salinity

More information

Bante902P Portable ph/conductivity Meter Instruction Manual

Bante902P Portable ph/conductivity Meter Instruction Manual Bante902P Portable ph/conductivity Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante902P Portable ph/conductivity Meter 1 Introduction Thank you for selecting the Bante902P portable ph/conductivity

More information

Operation Manual. MODEL ph/ec80. Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter. ph/ec80

Operation Manual. MODEL ph/ec80. Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter. ph/ec80 Operation Manual MODEL ph/ec80 Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter ph/ec80 CONTENTS INITIAL INSPECTION AND ASSEMBLY...2 BEFORE YOUR FIRST USE......2 A. Replace the

More information

Operation Manual MODEL 3173/3173R. MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R

Operation Manual MODEL 3173/3173R. MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R Operation Manual MODEL 3173/3173R MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R INITIAL INSPECTION..... 2 GENERAL INTRODUCTION.. 2 USING THE METER. 3 THE DISPLAY....

More information

CX2000 Intelligent Conductivity Transmitter

CX2000 Intelligent Conductivity Transmitter CX2000 Intelligent Conductivity Transmitter Operation Manual Lin 10.00mS/cm MTC PM 12:00 2013/04/10 25.0 C ACT B.L. SETUP CAL MODE ENTER Conductivity Transmitter CX2000 S/N: 123456789 1 ISO 9001-2008 Cercate

More information

User Guide. ExStik II EC400. Conductivity / TDS / Salinity / Temperature Meter

User Guide. ExStik II EC400. Conductivity / TDS / Salinity / Temperature Meter User Guide ExStik II EC400 Conductivity / TDS / Salinity / Temperature Meter Introduction Congratulations on your purchase of the ExStik II EC400 Conductivity/Total Dissolved Solids (TDS) / Salinity /

More information

Oyster Meter ph-conductivity-tds-salinity-orp(mv)

Oyster Meter ph-conductivity-tds-salinity-orp(mv) User's Guide Oyster Meter ph-conductivity-tds-salinity-orp(mv) Model 341350A INTRODUCTION Congratulations on your purchase of the Extech's Oyster Series meter. This device measures ph, Conductivity, TDS,

More information

Operation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M

Operation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M Operation Manual Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS

More information

As a result, our corporate philosophy is focused on total customer satisfaction. The Headquarter of Adwa Instruments is located in Szeged, Hungary:

As a result, our corporate philosophy is focused on total customer satisfaction. The Headquarter of Adwa Instruments is located in Szeged, Hungary: Adwa Instruments Founded in 1998, Adwa Instruments is one of the leading manufacturers of instrumentation to measure, Conductivity, ORP, Dissolved Oxygen and erature for Research and Industry. Headquartered

More information

User Manual. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500. Additional User Manual Translations available at

User Manual. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500. Additional User Manual Translations available at User Manual ExStik EC500 ph/conductivity / TDS / Salinity / Temperature Meter EC500 Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the

More information

Microprocessor ph/orp Controller

Microprocessor ph/orp Controller PC3030A/PC-3050 Microprocessor ph/orp Controller Operating Manual CONTENTS 1 Specifications... 2 Assembly and installation... 2.1 Precautions for installation.. 2.2 Install the controller on the wall mount

More information

MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL

MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL 470 350/REV A/10-03 MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1-2 Installation 2 Displays

More information

Bante901 Benchtop ph/conductivity Meter Instruction Manual

Bante901 Benchtop ph/conductivity Meter Instruction Manual Bante901 Benchtop ph/conductivity Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante901 Benchtop ph/conductivity Meter 1 Introduction Thank you for selecting the Bante901 ph/conductivity meter.

More information

Large Display Salinity & Conductivity Pens &

Large Display Salinity & Conductivity Pens & Large Display Salinity & Conductivity Pens 850036 & 850037 Large Display Salinity & Conductivity Pens 850036 & 850037 Copyright 2013 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents

More information

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately: BOECO PORTABLE PH/ORP/TEMP METER MODEL PT-380 A Simultaneous displays ph or mv and temperature. A Convenient calibration with automatic buffer recognition of European and US buffer sets. Calibration data

More information

Electrochemistry Meters. Series SD (IP 67 waterproof) Electrochemistry Meters

Electrochemistry Meters. Series SD (IP 67 waterproof) Electrochemistry Meters Electrochemistry Meters Series SD (IP 67 waterproof) 82 The new AQUALYTIC SD series comprises a range of compact, easy-to-use, hand-held instruments for the accurate measurement of ph, ORP, Con, TDS or

More information

Handheld Water Quality Meters 100 Series

Handheld Water Quality Meters 100 Series ph ORP Ion Conductivity Resistivity TDS DO Salinity Handheld Water Quality Meters 100 Series METERS 2 YEAR W A R R A N T Y M A D E I N J A P A N www.horiba-laqua.com In the lab, in the field or anywhere

More information

The Economy Conductivity Benchtop Meter

The Economy Conductivity Benchtop Meter CON 500 The Economy Conductivity Benchtop Meter Conductivity meter SET MODE HOLD ENTER RANGE MEAS ON/OFF MI MR Model no: WD-35610-00 WD-35610-05 897 1997 00701-55 2 Table of contents 1. Introduction...4-5

More information

Professional Benchtop Conductivity meter BC3020

Professional Benchtop Conductivity meter BC3020 15,0,15,0,1,14,1,14,13,2,13,2,3,12,3,12,11,4,11,4,5,10,5,10,9,6,9,6,7,8,7,8 Professional Benchtop Conductivity meter BC3020 Operation Manual 0 INTRODUCTION Your purchase of this professional benchtop Conductivity

More information

Agilent 3200M Multi-Parameter Analyzer

Agilent 3200M Multi-Parameter Analyzer Agilent 3200M Multi-Parameter Analyzer User Guide Agilent Technologies Notices Agilent Technologies (Shanghai) Co., Ltd. 2012 No part of this manual may be reproduced in any form or by any means (including

More information

DO Meter Kit. Instruction Manual

DO Meter Kit. Instruction Manual DO Meter Kit 850048 Instruction Manual DO Meter Kit 850048 Copyright 2013 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted in any

More information

User's Guide. Online Meters. PH / ORP / TDS / Conductivity.

User's Guide. Online Meters. PH / ORP / TDS / Conductivity. User's Guide Online Meters PH / ORP / TDS / Conductivity www.rakiro.net Introduction Continuous monitoring of PH / ORP is commonly required in a wide variety of industries ranging from water, chemical,

More information

Bante 221 Portable ph/orp Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

Bante 221 Portable ph/orp Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD Bante 221 Portable ph/orp Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante 221 Portable ph/orp Meter 1 Introduction Thank you for selecting the Bante 221 portable ph/orp meter. This manual provides

More information

Instruction Manual CON 400 and CON 410 Waterproof Hand-held Conductivity/TDS Meter

Instruction Manual CON 400 and CON 410 Waterproof Hand-held Conductivity/TDS Meter Instruction Manual CON 400 and CON 410 Waterproof Hand-held Conductivity/TDS Meter Technology Made Easy... 68X248904 Rev 3 01/04 Preface This manual serves to explain the use of the Waterproof CON 400/410

More information

Integrated Instruction Manual WP 600 Series Meters

Integrated Instruction Manual WP 600 Series Meters Integrated Instruction Manual WP 600 Series Meters ph 600/ 610/ 620, COND 600/610, DO 600, PC 650, PD 650, CD 650, PCD 650 68X415307 Rev. 1 Oct 07 Part of Thermo Fisher Scientific Technology Made Easy...

More information

Multiparameter Waterproof Meters. Multiparameter measurement in compact, IP67 rated waterproof enclosures

Multiparameter Waterproof Meters. Multiparameter measurement in compact, IP67 rated waterproof enclosures Multiparameter Waterproof Meters Multiparameter measurement in compact, IP67 rated waterproof enclosures HI98195 Features HI98195 Multiparameter Waterproof Meter ph, ORP, EC, TDS, Resistivity, Salinity,

More information

24-hour Continuous Field Monitors

24-hour Continuous Field Monitors Instruction Manual 24-hour Continuous Field Monitors ph, ORP, and Conductivity Technology Made Easy... 68X254901 Rev. 5 08/03 Preface This manual serves to explain the use for all models of the Field Monitors

More information

HI 9828 Multiparameter Water Quality Meter...

HI 9828 Multiparameter Water Quality Meter... HI 9828 Multiparameter Water Quality Meter... Backlit LCD can display up to 12 Measurements at a time Feature Highlights Display up to 12 parameters Graphic LCD with backlight Meter features IP67 and probe

More information

HI 2300 Auto-ranging Microprocessor EC/TDS/NaCl/ºC Bench Meter

HI 2300 Auto-ranging Microprocessor EC/TDS/NaCl/ºC Bench Meter / LOCK C Instruction Manual HI 2300 Auto-ranging Microprocessor EC/TDS/NaCl/ºC Bench Meter T Functions Select Measurement RANGE Range ALT GLP ATC TC SETUP Start Calibration LOCK Lock Current Range Change

More information

for measuring ph values and redox potential, conductivity and dissolved oxygen Content

for measuring ph values and redox potential, conductivity and dissolved oxygen Content for measuring ph values and redox potential, conductivity and dissolved oxygen Content Portable ph meters with GLP functions handylab ph 11 and handylab ph 12 Page 32 Technical data handylab LF 11, handylab

More information

Handheld Water Quality Meters 100 Series

Handheld Water Quality Meters 100 Series ph ORP Ion Conductivity Resistivity TDS DO Salinity Handheld Water Quality Meters 100 Series METERS 2 YEAR W A R R A N T Y M A D E I N J A P A N www.horiba-laqua.com In the lab, in the field or anywhere

More information

Research Grade Meters

Research Grade Meters HI5521 HI5522 Research Grade Meters ph/orp/ise and EC/TDS/Resistivity/Salinity and Temperature bench Conductivity/TDS Measure up to Eight Parameters HI5521 and HI5522 are research grade benchtop instruments

More information

Instruction Manual COND 600/610 Waterproof Handheld Conductivity Meter

Instruction Manual COND 600/610 Waterproof Handheld Conductivity Meter Instruction Manual COND 600/610 Waterproof Handheld Conductivity Meter 68X415302 Rev.1 NOV 06 Technology Made Easy... Preface This manual serves to explain the use of the COND 600/610 Waterproof Handheld

More information

STARTER 3000 Bench ph Meter Instruction Manual

STARTER 3000 Bench ph Meter Instruction Manual STARTER 3000 Bench ph Meter Instruction Manual Table of Contents 1 Introduction... 2 1.1 Safety measures... 2 1.2 Display and controls... 3 2 Installation... 5 2.1 Package contents... 5 2.2 Installing

More information

Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter

Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter English Cond Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket

More information

Operations Manual EcoSense ph100a

Operations Manual EcoSense ph100a Operations Manual EcoSense ph100a Portable ph, mv and Temperature Instrument English CONTENTS GENERAL INTRODUCTION... 3 INITIAL INSPECTION... 3 THE INSTRUMENT... 3 BATTERY INSTALLATION... 3 Battery Disposal...

More information

TDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT. Pool Professional s Meter

TDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT. Pool Professional s Meter TDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT Pool Professional s Meter TRACER EC/TDS/SAL POCKETESTER TM CODE 1749-KIT TABLE OF CONTENTS Introduction... 4 Specifications... 4 Contents... 5 Parts &

More information

SD card real time data logger TDS ( Total dissolved solids ), Salt '20 us/200 us/2 ms/20 ms/200 ms CONDUCTIVITY METER.

SD card real time data logger TDS ( Total dissolved solids ), Salt '20 us/200 us/2 ms/20 ms/200 ms CONDUCTIVITY METER. SD card real time data logger TDS ( Total dissolved solids ), Salt '20 us/200 us/2 ms/20 ms/200 ms CONDUCTIVITY METER Model : CD-4318SD Your purchase of this CONDUCTIVITY METER with SD CARD DATA LOGGER

More information

Thermo Scientific Orion Star A214 ph/ise Benchtop Meter

Thermo Scientific Orion Star A214 ph/ise Benchtop Meter Orion Star A Meters Keep turning the pages for more great finds. Thermo Scientific Orion Star A214 / Benchtop Meter The Thermo Scientific Orion Star A214 / Benchtop Meter is the ideal choice for dedicated

More information

MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL CONTENTS

MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL CONTENTS MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL MANUEL D UTILISATION BETRIEBSHANDBUCH MANUALE OPERATIVO MANUAL DE OPERAÇÂO MANUAL DE FUNCIONAMIENTO 470 250/REV A/09-00 MODEL 470 PORTABLE CONDUCTIVITY/TDS

More information

USER GUIDE. Salinity Meter Pen Style Water Quality Meter. Model EC170

USER GUIDE. Salinity Meter Pen Style Water Quality Meter. Model EC170 USER GUIDE Salinity Meter Pen Style Water Quality Meter Model EC170 Introduction Congratulations on your purchase of the Extech Pen Style Water Quality instrument; the Model EC170 measures Salinity and

More information

Instruction Manual. CyberScan CON Bench Conductivity/ TDS/ Resistivity Meter. Technology Made Easy... 68X Rev 1 01/04

Instruction Manual. CyberScan CON Bench Conductivity/ TDS/ Resistivity Meter. Technology Made Easy... 68X Rev 1 01/04 Instruction Manual CyberScan CON 1500 Bench Conductivity/ TDS/ Resistivity Meter Technology Made Easy... 68X292335 Rev 1 01/04 PREFACE Thank you for selecting the Eutech Instruments CyberScan CON 1500

More information

Multiparameter Waterproof Meters. Multiparameter measurement in compact, IP67 rated waterproof enclosures

Multiparameter Waterproof Meters. Multiparameter measurement in compact, IP67 rated waterproof enclosures Multiparameter Waterproof Meters Multiparameter measurement in compact, IP67 rated waterproof enclosures HI98194 Features HI98194 Multiparameter Waterproof Meter ph, ORP, EC, TDS, Resistivity, Salinity,

More information

Industrial. Laboratory and Field. Accessories. Product Selection Guide. Analyzers 1 Controllers 2 Transmitters 3 Monitors 4

Industrial. Laboratory and Field. Accessories. Product Selection Guide. Analyzers 1 Controllers 2 Transmitters 3 Monitors 4 Product Selection Guide Industrial Analyzers 1 Controllers 2 Transmitters 3 Monitors 4 Laboratory and Field Benchtop Meters 5 Field Meters 5 Benchtop Controllers 7 Pocket Meters & Testers 9 Accessories

More information

Instruction Manual Mi 170 Bench Meter

Instruction Manual Mi 170 Bench Meter Instruction Manual Mi 170 Bench Meter GENERAL DESCRIPTION... 4 FUNCTIONAL DESCRIPTION... 2 SPECIFICATIONS... 5 OPERATIONAL GUIDE... 7 AUTORANGING... 9 EC/TDS CALIBRATION... 10 NaCl CALIBRATION... 11 GOOD

More information

HI 9828 Multiparameter Water Quality Meter...

HI 9828 Multiparameter Water Quality Meter... HI 9828 Multiparameter Water Quality Meter... Backlit LCD can display up to 12 Measurements at a time Feature Highlights Display up to 12 parameters Graphic LCD with backlight Meter features IP67 and probe

More information

INSTRUCTION MANUAL. Multiparameter Bench Meter. Mi 180. ph/mv/ec/tds/nacl/temperature.

INSTRUCTION MANUAL. Multiparameter Bench Meter. Mi 180. ph/mv/ec/tds/nacl/temperature. www.milwaukeetesters.com INSTRUCTION MANUAL Multiparameter Bench Meter Mi 180 ph/mv/ec/tds/nacl/temperature MARTINI instruments www.milwaukeetesters.com 1 Instruction Manual Mi 180 Bench Meter FUNCTIONAL

More information

703 Laboratory Conductivity Meter

703 Laboratory Conductivity Meter 703 Laboratory Conductivity Meter Sensoface Sensoface monitors the sensor and measuring equipment and provides information on sensor selection and handling. It reports clock memory loss and requests regular

More information

SP-2300/SP-2500 Microprocessor ph/orp/ion Benchtop Meter. Operation Manual

SP-2300/SP-2500 Microprocessor ph/orp/ion Benchtop Meter. Operation Manual SP-2300/SP-2500 Microprocessor ph/orp/ion Benchtop Meter Operation Manual 1 CONTENTS 1. Specifications...1 2. Panel Description...2 2.1 Front panel...2 2.2 Sockets...2 2.3 Display...3 2.4 Keypad...4 3.

More information

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810 Operation Manual Hand-held ph/mv/temperature Meter ph6810 CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS FUNCTIONS..... 4

More information

Instruction Manual HI Autoranging & Logging Portable Waterproof Microprocessor EC/TDS/NaCl/ C Meter

Instruction Manual HI Autoranging & Logging Portable Waterproof Microprocessor EC/TDS/NaCl/ C Meter Instruction Manual HI 98360 Autoranging & Logging Portable Waterproof Microprocessor EC/TDS/NaCl/ C Meter www.hannainst.com Dear Customer, Thank you for choosing a Hanna Instruments product. Please read

More information

Hanna Instruments HI9819X Multiparameter series

Hanna Instruments HI9819X Multiparameter series Hanna Instruments HI9819X series www.hannainst.com HI98194 Waterproof Meter ph / mv, ORP, EC, TDS, Resistivity, Salinity, Seawater σ, Dissolved Oxygen, Atmospheric Pressure and Waterproof IP67 rated waterproof,

More information

User Manual. Portamess 913 Cond and Portamess 913 X Cond

User Manual. Portamess 913 Cond and Portamess 913 X Cond User Manual Portamess 913 Cond and Portamess 913 X Cond Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender).

More information

Operating Instructions. SenseLine ph meter F410

Operating Instructions. SenseLine ph meter F410 Operating Instructions SenseLine ph meter F410 Contents Contents 1. Introduction 3 2. Safety measures 3 3. Installation 5 3.1 Installing the batteries 5 3.2 Fitting the wrist strap 5 4. Operating the F410

More information

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS OPERATION MANUAL JENCO MODEL 6307 MICROCOMPUTER BASED ph/conductivity/ Salinity/Temperature BENCH METER JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS INITIAL INSPECTION Carefully unpack

More information

ProLab 1. ProLab 1000: ph for the professional. ph Meter ProLab 1000: Superior technique in a classy metal body. Precise measurements

ProLab 1. ProLab 1000: ph for the professional. ph Meter ProLab 1000: Superior technique in a classy metal body. Precise measurements ProLab 1000: ph for the professional ProLab 1000: Superior technique in a classy metal body Precise measurements With a measuring range of -2.000 up to +20.000 ph and an accuracy of 0.003 ph the ProLab

More information

INSTRUCTION MANUAL ORP-METER MODEL COR-01

INSTRUCTION MANUAL ORP-METER MODEL COR-01 INSTRUCTION MANUAL ORP-METER MODEL COR-01 EL-221, MIDC Electronic Zone, Mhape, Navi Mumbai-400701. Tel: 022-61944000 CHAPTER 1 INTRODUCTION CONTECH ORP meter measures ph and mv, can used for Oxidation

More information

Professional Waterproof Meters

Professional Waterproof Meters HI9819X Series Professional Waterproof Meters New Quick Connect Probes (HI98190, HI98192 and HI98193 only) Benchtop Performance in a Portable Meter The HI9819X series of meters combine all of the features

More information

Instruction Manual. CyberScan DO Dissolved Oxygen Meter. Technology Made Easy... 68X Rev. 1 01/04

Instruction Manual. CyberScan DO Dissolved Oxygen Meter. Technology Made Easy... 68X Rev. 1 01/04 Instruction Manual CyberScan DO 1500 Dissolved Oxygen Meter Technology Made Easy... 68X292336 Rev. 1 01/04 PREFACE Thank you for selecting the Eutech Instruments CyberScan DO 1500 bench meter. The instruction

More information

315 ph/ion. Installation. Specifications. Installing the Batteries a595. Input Connectors

315 ph/ion. Installation. Specifications. Installing the Batteries a595. Input Connectors 315 /Ion Installation Input Connectors socket (automatic temperature compensation) socket data socket reference socket If you ordered an electrode with the 315, connect it as follows: Corning Incorporated

More information

HI HI 9034 HI HI 9130 HI HI 9635 Portable Waterproof Multi-Range Conductivity/TDS/ Fertilizer Meters

HI HI 9034 HI HI 9130 HI HI 9635 Portable Waterproof Multi-Range Conductivity/TDS/ Fertilizer Meters CONDUCTIVITY CFM BUF Instruction Manual HI 9033 - HI 9034 HI 9039 - HI 9130 HI 933100 - HI 9635 Portable Waterproof Multi-Range Conductivity/TDS/ Fertilizer Meters ins rum n s ms HI 933100 MICROPROCESSOR

More information

1. Introduction 2. Specifications 3. Symbol Definition and Button Location 4. Operation Instructions 5. Temperature Measurement 6.

1. Introduction 2. Specifications 3. Symbol Definition and Button Location 4. Operation Instructions 5. Temperature Measurement 6. CONTENTS TITLE PAGE 1. Introduction.. 1 2. Specifications... 1 3. Symbol Definition and Button Location... 2 4. Operation Instructions.... 3 4.1 Power-Up...3 4.2 Connection of the Temperature Probe...

More information

TM 2. User s Guide. Shop online at omega.com For latest product manuals: CDH-SD11

TM 2. User s Guide. Shop online at omega.com   For latest product manuals:   CDH-SD11 TM 2 User s Guide Shop online at omega.com e-mail: info@omega.com For latest product manuals: www.omegamanual.info CDH-SD11 Conductivity, TDS, and Salinity Meter with Real Time SD Card Data Logger omega.com

More information

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet English ph / ISE / COND / RDO / DO Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the

More information

Microprocessor SUNTEX EC-410/430. Conductivity Monitor

Microprocessor SUNTEX EC-410/430. Conductivity Monitor Microprocessor SUNTEX EC-410/430 Conductivity Monitor EC-410 μs/cm ms/cm MΩ cm 4-20mA Output ISOLATED 4-20mA OUTPUT Accurate measurement and signal transmission.. EC-430 Microprocessor Conductivity Monitor

More information

WalkLAB HC Professional Conductivity-TDS-Salinity Meter. Operations Manual 15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8

WalkLAB HC Professional Conductivity-TDS-Salinity Meter. Operations Manual 15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8 15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8 15,0,15,0,1,14,1,14,13,2,13,2,3,12,3,12,11,4,11,4,5,10,5,10,9,6,9,6,7,8,7,8 WalkLAB Professional Conductivity-TDS-Salinity Meter HC 9021 Operations Manual 0 INTRODUCTION

More information

Oakton Testers/Handheld Meters

Oakton Testers/Handheld Meters Table Of Contents Description Page EcoTestrTM ph/conductivity/salt Pen Testers 2 ph/orp/conductivity/tds/salinity Pen Testers 3 ph/orp/conductivity/tds/salinity Handheld Meters 4 Parts / Accessories 5-7

More information

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or F) Automatic

More information

EC-4310 Intelligent Conductivity Transmitter. Operation Manual

EC-4310 Intelligent Conductivity Transmitter. Operation Manual EC-4310 Intelligent Conductivity Transmitter Operation Manual CONTENTS Precautions for installation Brief Instruction 1 1. Specifications 4 2. Assembly and installation 2.1 Transmitter installation 5 2.2

More information

Waterproof Construction. Notepad Function. Automatic Datalogging. Automatic Calibration. Good Laboratory Practices. Serial Interface Port

Waterproof Construction. Notepad Function. Automatic Datalogging. Automatic Calibration. Good Laboratory Practices. Serial Interface Port Waterproof Construction At TPS we know that working in wet areas is sometimes unavoidable. The 90 series is designed for rugged field applications in wet environments. The enclosure, keypad and connectors

More information

Ideal portable ph. Automatic temperature compensation. Slimline waterproof casing

Ideal portable ph. Automatic temperature compensation. Slimline waterproof casing Portable // Laboratory part no: HI 991300 Measuring, conductivity, total dissolved solids and temperature in one slim lightweight casing and with one single probe - the HI 991300 is an ideal portable tester.

More information

SINAR TECHNOLOGY LARGE SEED MOISTURE SAMPLER. USER MANUAL

SINAR TECHNOLOGY LARGE SEED MOISTURE SAMPLER. USER MANUAL 1 SINAR TECHNOLOGY LARGE SEED MOISTURE SAMPLER. USER MANUAL 2 Read all instructions prior to operating your Sinar meter. Whenever using this unit, remember to follow the appropriate operating procedures.

More information

How to build a Conductivity Measurement System

How to build a Conductivity Measurement System How to build a Measurement System F B E C D A A Immersion system in open basin B Flow-through system C 1/2 pipe mounted system D 3/4 pipe mounted system E Side-entry system in vessel F Top-entry system

More information

Instruction Manual Cond 6+ Conductivity/Temp TDS 6+ TDS/Temp Salt 6+ Salinity/Temp

Instruction Manual Cond 6+ Conductivity/Temp TDS 6+ TDS/Temp Salt 6+ Salinity/Temp Instruction Manual Cond 6+ Conductivity/Temp TDS 6+ TDS/Temp Salt 6+ Salinity/Temp 68X243635 Rev 0 02/09 Preface This manual serves to explain the use of the Conductivity, TDS, and Salinity handheld meters.

More information

Conductivity meter Manual

Conductivity meter Manual Announcements Conductivity meter Manual 1.When using, please comply with the instruction of the operation procedures and the announcements. 2.If find the instrument working abnormally or damaged, please

More information

Tracer ph/tds/salt PockeTester TM Code 1766

Tracer ph/tds/salt PockeTester TM Code 1766 Tracer ph/tds/salt PockeTester TM Code 1766 Do Not allow ph sensor to dry out. Store Upright Before first use, tap meter against palm to move electrolyte to tip. See pages 10 and 19. ph/conductivity/tds/salinity

More information

Digital Thermometer MEM. PC-Link READ

Digital Thermometer MEM. PC-Link READ Digital Thermometer MEM PC-Link READ CONTENTS Digital Thermometer 1. Safety Information...1 2. Product Overview...3 3. Recognize Meter..... 4 3.1 Component...4 3.2 Display...5 3.3 Keys Description...6

More information

The Model 703 offers a wide range of practical features to meet the numerous requirements of everyday measuring tasks.

The Model 703 offers a wide range of practical features to meet the numerous requirements of everyday measuring tasks. Laboratory Meters The requirements for lab measurements become stricter every day. Quality assurance and measurement documentation in accordance with GLP are a must in many areas. 703 Laboratory Conductivity

More information

+GF+ SIGNET 5800CR Conductivity/Resistivity Monitor Instructions

+GF+ SIGNET 5800CR Conductivity/Resistivity Monitor Instructions GF SIGNET 5800CR Conductivity/Resistivity Monitor Instructions ENGLISH 35800CR.0901 D1/98 CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections.

More information

130 Salt Contamination Meter

130 Salt Contamination Meter 130 Salt Contamination Meter Conforms to SSPC Guide 15 Page 1 of 6 The Elcometer 130 Salt Contamination Meter quickly and accurately measures the level of soluble salts on surfaces. Features Fast reading

More information

Instruction Manual HI 9033 HI Waterproof Multi-Range EC and TDS Meters for Field Applications.

Instruction Manual HI 9033 HI Waterproof Multi-Range EC and TDS Meters for Field Applications. Instruction Manual HI 9033 HI 9034 Waterproof Multi-Range EC and TDS Meters for Field Applications www.hannainst.com PRELIMINARY EXAMINATION Dear Customer, Thank you for choosing a HANNA instruments product.

More information

Manual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER

Manual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER Manual No. 9835 Revision 0-401 MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER Please read this instruction manual carefully before using the instrument. This manual will provide you with all

More information

About Suntex. Online Measurement and Control. Portable and Laboratory Instruments. ISO 9001: 2000 Certificated since 2001

About Suntex. Online Measurement and Control. Portable and Laboratory Instruments. ISO 9001: 2000 Certificated since 2001 About Suntex Suntex Instruments Company Ltd., established in 1982, has gained prestige and trust from our customers for dedication of providing professional, high-quality water analysis products and first-rate

More information

CLEAN PH200/ PH500 ph/ mv/ ORP Ion Versatile Water Analysis Meter www.cleaninst.com E-mail: CS@cleaninst.com CLEAN INSTRUMENTS CO. LTD 2F No.140, Zhongxiao St., Zhonghe Dist., New Taipei City,Taiwan ph/

More information