MKII - PORTABLE CORROSION METER.

Size: px
Start display at page:

Download "MKII - PORTABLE CORROSION METER."

Transcription

1 0 STOMER: TITLE: MKII - PORTABLE CORROSION METER. NUAL CUMENT NO. : C Tot. no. of pages: 125 OJECT DOC. NO. : For distribution MH LK MB For use including SandLog Terminal LK RJ RGB Functions For use including ML Terminal Function RGB RJ RGB FOR ACCEPTANCE RGB RJ RGB FOR INTERNAL CHECK RGB RJ REV DATE REVISION BY CHK APP

2 MultiCorr MKII - User Manual Page: 2 WARRANTY CorrOcean warrants that the MultiCorr MKII Corrosion Instrument is free from manufacturing defects and will operate in accordance with the procedures detailed in this Users Manual. If during the first year, from the date of purchase, the MultiCorr MKII becomes defective due solely to poor workmanship or design CorrOcean will, at it's sole discretion, repair or replace the defective unit. Damage to the MultiCorr MKII due to misuse, physical damage or other abuse shall not be covered under any warranty repairs. The repair or replacement of batteries is not considered part of the one year warranty and is specifically excluded from any repair or replacement during the unit warranty period. Limitations of Liability Except for the express warranty statement above, CorrOcean grants no warranties, either expressed or implied, with regard to this product, including all implied warranties of merchantability and fitness for a particular purpose, and the stated express warranty is in lieu of all liabilities or obligations of CorrOcean for damages arising out of or in connection with the delivery, use or performance of the MultiCorr MKII product. CorrOcean assumes no responsibility for any economic or other losses incurred by the Purchaser, User or that any Third Party might experience from the use of the MultiCorr MKII, poor interpretation of readings, analysis of data obtained or from conclusions made based on use of the MultiCorr MKII. In no event shall CorrOcean be liable for lost profits or consequential damages even if CorrOcean is informed of the possibility of such damage. In no event will CorrOcean be liable for any claim or demand of a third party. Copyright 1995 by CorrOcean as, Trondheim, Norway IBM is a registered Trade Mark of the IBM Corporation

3 MultiCorr MKII - User Manual Page: 3 CONTENTS 1 INTRODUCTION Capability and Functions Probe Compatibility Data Transmission To PC Manual Use and Information 7 2 REFERENCE DOCUMENTS 8 3 HANDLING AND STORAGE General Packing Initial Inspection Storage Charging NiCd Batteries Connecting to a Personal Computer Connecting to a Probe 11 4 MULTICORR MKII INSTRUMENT General Entry Keys Data Screen Probe Identification System - Probes Measured by MultiCorr MKII Probe Identification System - Probes measured by MultiLog Probe Identification System - Probes Measured by SandLog 15 5 MENU DESCRIPTION The Main Menu Operator Mode Advanced Mode Terminal mode Menu Paths 18 6 PROBE IDENTIFICATION AND METER SETUP Selecting Probe Type and Assigning a Probe Number Installing and Editing Probe Tag Identification Reading Setup 22 7 DEFINING PROBE CHARACTERISTICS AND PARAMETERS General Entering Electrical Resistance (ER) Probe Parameters Entering Linear Polarization Resistance (LPR) Probe Parameters Entering Galvanic Probe Parameters Entering Analog Probe Parameters Entering Polarization Parameters 31 8 PROBE MEASUREMENTS Taking A Reading In The Operator Mode Taking A Reading In The Advanced Mode 34 9 PROBE MEASUREMENT READINGS AND DISPLAY

4 MultiCorr MKII - User Manual Page: 4 INFORMATION General Measurement Error Messages Selecting The Probe to Be Monitored Raw Data and Final Monitoring Electrical Resistance (ER) Probes Monitoring Linear Polarization Resistance (LPR) Probes Monitoring Galvanic Probes Monitoring Analog Probes, Reference Electrodes and CP Systems Polarization Readings DISPLAYING MEASUREMENT DATA Selecting A Probe Displaying Raw Data Displaying Readings Displaying ER Probe Corrosion Rate MultiCorr MKII SYSTEM FEATURES Memory and Status Setting Date and Time System Reset and Default Settings Clearing The Memory Changing The Pass Code TRANSMITTING READINGS TO A COMPUTER Data Reception Software Transmission Cable Connections Specifying Data to be Transmitted Setting the Transmission Speed Starting Data Transmission Data Transmission Troubleshooting AUTOMATIC READINGS General Selecting Reading Intervals Selecting A Probe For Automatic Readings Starting Automatic Readings Stopping Automatic Readings Automatic Reading Technical Information TERMINAL MODE Global settings for MultiCorr MKII 56

5 MultiCorr MKII - User Manual Page: 5 15 TERMINAL MODE - MULTILOG MultiLog Terminal Main Menu Logger Manager Transfer Data TERMINAL MODE - SANDLOG SandLog Terminal Main Menu Logger Manager Transfer Data 87 APPENDIX A: MultiCorr MKII Specifications 95 APPENDIX B: ER Probe Technical Information 97 APPENDIX C: LPR Probe Technical Information 105 APPENDIX D: Galvanic Probe Technical Information 111 APPENDIX E: Analog Probe Technical Information 113 APPENDIX F: Potentiostatic and Potentiodynamic Procedures 114 APPENDIX G: MultiCorr MKII Error Messages 115 APPENDIX H: Service and Technical Information Sources 117 APPENDIX I: Hazardous Area Certificate 118 APPENDIX J: MultiCorr MKII Menu Pattern 123

6 MultiCorr MKII - User Manual Page: 6 1 INTRODUCTION The MultiCorr MKII is a software based, computerized, hand held corrosion meter and data acquisition unit suitable for the interrogation of different types of corrosion probes. It is housed in a rugged rubber coated housing that is splash resistant and intrinsically safe. Each MultiCorr MKII can be individually configured to the user's specific needs at the time of purchase. Additionally, the unit can have its software upgraded at any time to add Functions which were not included at the time the unit was purchased. Each MultiCorr MKII has the capacity to store up to 3,000 sets of readings. Readings may be stored in memory along with a unique probe number, an alphanumerical tag designation, along with an automatically generated reading date and time. Written records are not required, but can be easily maintained. 1.1 Capability and Functions The MultiCorr MKII is capable of monitoring a wide variety of corrosion probes, so there is no need to purchase or use different corrosion monitoring instruments when measuring different types of corrosion probes. Shown below is a summary of the Functions available in the MultiCorr MKII Instrument. Function Capability ER Probe Read and Monitor Electrical Resistance Probes LPR Probe Read and Monitor Linear Polarization Probes Galvanic Probe Read and Monitor Galvanic Probes Analog Probe Read and Monitor Reference Electrodes, Monitored Anodes, Temperature and Pressure Sensors Polarization Conduct several types of polarization measurements. Potentiostatic, potentiodynamic and cyclic polarization may be performed. Automatic Conduct Automatic Readings at Uniform Time Intervals Terminal Function for MultiLog Setup of MultiLog Stations, data transfer from MultiLog to the MultiCorr MKII, data transfer from the MultiCorr MKII to PC, etc. Terminal Function for SandLog Setup of Sandlog Stations, data transfer from SandLog to the MultiCorr MKII, data transfer from the MultiCorr MKII to PC, etc. The user can include all of the above Functions in a single instrument or chose any possible combination of Functions at the time of purchase. For instance, you may configure an instrument to read only ER Probes and take automatic readings or have an instrument that can read ER, LPR and Galvanic Probes, take Automatic readings and perform Polarization measurements. The MultiCorr MKII can grow with your operations since any Function which was not included when the unit was delivered may be added at a later date.

7 MultiCorr MKII - User Manual Page: Probe Compatibility The MultiCorr MKII is a versatile instrument which accepts corrosion probes of different designs and from different manufacturers. There are default settings within the MultiCorr MKII for each type of probe, and the user can easily modify these settings so that any probe can be read regardless of the manufacturer. Because of variations in the size and type of connectors it may be necessary to have different cables for different manufacturers. When ordering, please specify the manufacturer and probe type and CorrOcean will provide the required probe cables to suit the user's needs. 1.3 Data Transmission To PC The MultiCorr MKII can communicate directly with any IBM compatible Personal Computer (PC), which has been loaded with the CorrOcean Corrlist Software provided with the MultiCorr MKII. It can also communicate directly with other CorrOcean Software programs such as CorrTrend and CorrChart. CorrTrend is an advanced corrosion data management program and CorrChart can be utilized to automatically plot and present polarization curves. Data transfer to the PC can be performed using either direct data transfer cables or via a modem. When used as a terminal, the data are transferred into the CorrOcean software MultiCom or SandRater (for MultiLog and SandLog data respectively) 1.4 Manual Use and Information This manual includes information on the use of every feature of the MultiCorr MKII Instrument. The manual provides details on the use of functions which may not have been included in the MultiCorr MKII purchased by the user. Any MultiCorr MKII function may be added to your instrument after it has been purchased. Call the nearest CorrOcean office listed in the back of this manual for more information on how to upgrade your MultiCorr MKII. It is recommended that the MultiCorr MKII Instrument be used while reading this manual so that the various options may be tried as the user is reading the manual.

8 MultiCorr MKII - User Manual Page: 8 2 REFERENCE DOCUMENTS The MultiCorr MKII portable corrosion monitoring system and related software are also described in the following documents, available from CorrOcean. Descriptive Literature MultiCorr MKII Corrosion Meter Data Sheet 155 CorrTrend Software Data Sheet 154 MultiCorr MKII/CorrChart Software Data Sheet 156 MultiLog Data Sheet 158 The CorrOcean Sand Monitoring System Data Sheet 180 Operating Manuals Corrlist Software User Manual Document No.: CorrTag Software User Manual Document No.: CorrTrend Software Inst. and User Manual Document No.: (A): , (M): CorrChart Software User Manual Document No.: MultiCom - User Manual Document No.: MultiLog Stations - Installation/Operation/Maintenance Manual Document No.: SandRater - Sand Monitoring Program Operating Manual Document No.: Installation Procedure Sand Monitoring System Document No.: CorrOcean Sand Monitoring System User and Maintenance Manual Document No.:

9 MultiCorr MKII - User Manual Page: 9 3 HANDLING AND STORAGE 3.1 General The MultiCorr MKII is built into a rugged splash proof case and is designed for operation in an offshore, industrial and/or hazardous environment. Carefully selected materials have been used to meet these requirements. For example, the application of a silicone rubber cover on the bottom part provides thermal insulation, shock absorption and a friction mantle. The silicone rubber can withstand environmental temperatures up to 550 F (300 C). However, the user must remember this is an electronic instrument and if it is left on a hot surface and the internal temperature is allowed to exceed 160 F (70 C), the instrument may be permanently damaged. The specifications on the MultiCorr MKII are described in APPENDIX A "MultiCorr MKII Specifications". 3.2 Packing The instrument comes complete with a battery charger, PC data transmission cable, Corrlist software and a users manual. All of the components are packed in a plastic transportation case filled with shock absorbing foam. The plastic case is meant for shipment and storage only, and is not intended for use in a hazardous area. 3.3 Initial Inspection The instrument was carefully inspected both mechanically and electronically prior to shipment. It should be free of physical damage upon receipt. To confirm this, the instrument should be inspected for any damage that may have occurred during shipment. If any damage is observed please contact CorrOcean immediately. 3.4 Storage The instrument may be stored or shipped in environments within the following limits: Temperature: -5 F to 160 F (-20 C to 70 C) Humidity: Up to 95% The instrument must be protected from extreme temperatures which may cause condensation to form within the instrument case. It is recommended that the instrument is stored in a dry, cool place (not colder than -5 F (-20 C)) to minimize battery discharge. To prevent loss of stored data and keep the internal clock running, recharging should be carried out at least once every 3 months. This recharging should be more often if the unit is stored at temperatures beyond 85 F (30 C) or if the battery is old.

10 MultiCorr MKII - User Manual Page: Charging Charging time for empty batteries is approximately 12 hours. The instrument may also be operated during charging. Charging should only take place in nonhazardous areas. LOW BATT is displayed on the instrument's status line when the battery charge drops below a certain limit. The batteries should be recharged as soon as possible after this. If operation continues, the power will automatically be switched off when the voltage drops below the secure operation limit. The automatic switch off function is to insure that there is sufficient battery charge to save stored data. If the instrument has automatically switched itself off under these conditions, it will not be possible to make measurements until it has been recharged. NiCd batteries can occasionally develop a "memory" which limits the available user time. If this occurs deplete the battery through use and then recharge. This process of draining the battery and recharging may have to be repeated. 3.6 NiCd Batteries The MultiCorr MKII NiCd batteries have a design life of 3 to 5 years depending on the use and care. The battery package was made especially for the MultiCorr MKII instrument and are designed not only for a specific power usage but also for use in a hazardous environment. Changing the battery should only be conducted by personnel authorized by CorrOcean, to insure that the hazardous area requirements are maintained. Warning: NiCd batteries slowly loose their charge even when not in use. For this reason it is important to charge the batteries at least every 3 months, even if the MultiCorr MKII is not being utilized. If the NiCd batteries are not recharged when the LOW BATT Warning appears it is possible that stored data may be lost if the battery charge level drains below the minimum level necessary to maintain the MultiCorr MKII memory. The battery voltage can be seen by selecting STATUS in the system or acq. setup menues. Both the positive and the negative battery should be at more than 6.25 V when they are charged. The voltages are however depending on the temperature. 3.7 Connecting to a Personal Computer The battery charging cable also contains a cable branch for connecting it to a PC RS-232 port. The MultiCorr MKII may be connected to a PC for data transfer whether or not the battery charger is connected to an outlet. It is recommended that the charger be connected when data transmission is being accomplished, but this is not required. The standard connector on the PC Data Cable is a 25 pin, female connector. Some computer ports may require the use of a 9 pin female connector, in which case an adapter may be utilized. Connection to a PC should only take place in nonhazardous areas.

11 MultiCorr MKII - User Manual Page: Connecting to a Probe The MultiCorr MKII is connected to the probe using a flexible retractable probe cable. The probe cable and connector assembly will be different for the different probe types, and may also be different for probes from different manufacturers. Probe cables can be supplied by CorrOcean according to user's specifications. For most commercial corrosion probes, probe cables can be supplied on short notice. Please contact CorrOcean for more information. The probe cable has an 11 pin male Fisher connector on the instrument side. There is a white orientation dot on the Fisher connector that should be visible on the top of the connector when inserting the cable into the MultiCorr MKII. Always insert and remove the cable by grasping the connector housing and never pull directly on the cable itself. The internal wiring of probe cables may be different even when the connector appears to fit the probe. Make sure that both the probe and cable are compatible for both probe type (e.g. ER, LPR) and manufacturer. NOTE: The instrument housing is electrically conductive. When measuring on LPR and galvanic probes, the instrument must be held in the hand, or otherwise electrically isolated from ground.

12 MultiCorr MKII - User Manual Page: 12 4 MULTICORR MKII INSTRUMENT 4.1 General The user interface of MultiCorr MKII is implemented with an efficient and easy to learn on screen menu system. In this manual the definition of a menu is a list of functions which can be selected with a single numerical key entry. Throughout this manual the keyboard entries are enclosed in brackets [ ] to distinguish them from other numerical listings. The definition of a window is a display screen with specified contents. Probe information may also be edited or new information may be added utilizing CorrOcean's CorrTag Software which is described later in this manual. Since the MultiCorr MKII can store hundreds of different probes in its memory the operator will learn how to select a specific probe so that readings are calculated and stored for that specific probe. In this manual the probe which is ready to accept and store readings at any given time is referred to as the Active Probe. 4.2 Entry Keys The key board on the MultiCorr MKII is generally self descriptive but it is helpful to understand the use of these keys as they are used the MultiCorr MKII. Key Use Description [On/Off] On or Off Use to turn the MultiCorr MKII screen on or off. The MultiCorr MKII has an internal battery saver which will turn off the instrument if no key is pressed within 2 minutes of the last key entry. [Del] Delete This key is used to delete the last erroneous character entered. You may also use the left arrow key to make a deletion. [Exit] Exit The [Exit] key is used to leave a menu or a window. Terminating data entry with the [Exit] key leaves the previous value unchanged. [Ent] Enter When entering parameters such as probe numbers or the surface area of an LPR Probe electrode, the input must be terminated with the [Ent] key. The input value will not be accepted until the entered value is within legal limits. When an illegal entry is made, the display cursor will return to the first position in the input field and a new value may be entered.

13 MultiCorr MKII - User Manual Page: 13 Arrows Up/Down Once probe parameters have been entered into the unit the operator can display the next probe in sequence by pressing the down arrow or returning to a prior probe with the up arrow. When the menu is set to display data the operator can move through the reading summary using the Up and Down arrows. These arrows are also utilized when selecting the alphanumeric tag numbers. This use is described more fully later in the manual. Arrows Left/Right The left arrow may be used to delete an erroneous character entry. These arrows are also utilized when selecting the alphanumeric tag numbers. This use is described more fully later in the manual. 4.3 Data Screen The MultiCorr MKII uses a Liquid Crystal Display (LCD) screen. As a battery saving feature the screen will automatically go blank if no key has been pressed during any two (2) minute interval. The screen will also go blank during the periods between automatic readings. The LCD display utilizes a standard pattern for the display of information. Line Location First Line selection. Description Provides the name of the menu. This corresponds to the previous Second Line This is a warning information line. When the battery is low a "LOW BATT" warning will appear. When the instrument is set to automatically take readings an "AUTO-ON" Warning will appear. Additionally, if the MultiCorr MKII memory is full a "MEMORY FULL" Warning is displayed. Lines 3-8 Menu selection alternatives are listed when a menu is displayed. When the unit is in an operational mode, the last two lines (7-8) display the Active Probe in memory. Both the probe number and tag number are displayed. Line 3-8 is also used to display any electrical measuring errors which may be detected by the MultiCorr MKII. The meaning of these error messages is provided in APPENDIX G - MultiCorr MKII Error messages.

14 MultiCorr MKII - User Manual Page: Probe Identification System - Probes Measured by MultiCorr MKII The MultiCorr MKII uses a standard method to identify each probe so that the user can readily identify the probe which is being interrogated. It is important for the user to understand this identification system prior to reading the manual. The following identification labels have a specific meaning that is described below. Probe No Each probe is assigned a unique number that includes both the type of Probe (e.g. ER, LPR, Galvanic) and a number from 0 to 99. For instance ER 23, LPR 99 or Glv 0. All ER Probe Numbers will start with "ER" and all LPR Probe Numbers will start with "LPR". The MultiCorr MKII will distinguish between probes of the same type by the number following the probe type designation. A total number of 200 different probes may be defined. Probe Numbers must be assigned before the MultiCorr MKII will take a reading. Tag No The Tag Number allows the user to assign a name to each probe that can assist in identifying the probe. The Tag Number may be up to 10 alphanumeric characters in length. For instance, the user could tag probe ER 12 with the additional identification of "Riser A67". When this probe is selected the MultiCorr MKII screen would display the following as the Active Probe. Probe: ER 12 Tag No: Riser A67 Tag Numbers are not required to be entered before the MultiCorr MKII takes a reading. However, it is recommended that Tag Numbers are assigned. 4.5 Probe Identification System - Probes measured by MultiLog The probes measured by MultiLog are identified by tag no. etc. as defined in the MultiCom software (it is referred to document for description of the probe identification system). However, to reduce complexity and data amount to be transferred, tag numbers etc. are not given in the probe identification in the MultiLog stations and Terminal. Identification of the probes are given by MultiLog station no, probe type and number. Probe number starts at 1 for each probe type for each logger. When MultiLog data are stored in the MultiCorr MKII memory, "tag number" will be automatically generated. This number is used only for display of probe data and is not transmitted when data are sent to PC. However, these tag numbers will be part of the total capacity of 200 tag numbers, and are cleared by reset of the MultiCorr MKII only (please see section 11.3 for reset instructions).

15 MultiCorr MKII - User Manual Page: Probe Identification System - Probes Measured by SandLog The probes measured by SandLog are identified by the SandLog address and probe number. A SandLog is measuring one or two sand probes. A sand probe may have 2, 3 or 4 measuring elements, and each element is displayed individually. SandLog address and probe number must be defined before measurements are initiated, please see section for logger setup. Also see figure 32 for data display in SandLog. The corresponding sand probe identification is found when data is transferred to the MultiCorr MKII. Section describes display of the stored data, an example is shown in figure 37. For complete sand monitoring system setup, it is referred to "SandRater Sand Monitoring Program Operating Manual".

16 MultiCorr MKII - User Manual Page: 16 5 MENU DESCRIPTION 5.1 The Main Menu The MultiCorr MKII Main Menu is displayed when the meter is turned on. This menu allows the user to select two operating modes which are defined as the "Operator Mode" and the "Advanced Mode". Operator Mode This is a limited use mode designed to allow only field measurements and storage. This limited operational mode permits the user to quickly take readings without the need to utilize the expanded menu selection. Advanced Mode This menu mode allows the corrosion specialist to set probe parameters, take automatic readings, set engineering units, transmit data to a PC and perform a wide variety of specialized operations. Some areas of this menu are protected by a pass code to insure that inadvertent changes are not made to the system. Field measurements may also be taken using the mode. Terminal Mode This menu mode allows setup of CorrOcean MultiLog and/or SandLog stations, check of data and data retrieval from MultiLog/SandLog and data transfer to relevant PC software. Some areas of this menu are protected by a pass code to insure that inadvertent changes are not made to the system. Shown in Figure 1 is the Main Menu of the MultiCorr MKII. To enter either the Operator Mode or the Advanced Mode the user simply presses the corresponding selection number located in front of the desired selection. The MultiCorr MKII will immediately move to the first menu level for that operating mode. Figure 1 shows the menu with five (5) types of probe functions and terminal function installed. Only the functions which are actually installed in the user's MultiCorr MKII will be displayed. *** MultiCorr MKII *** Probes: ER - LPR Glv - Ana Pol 1: Operator Mode 2: Advanced Mode 3: Terminal Mode Figure 1

17 MultiCorr MKII - User Manual Page: Operator Mode There are only 2 possible selections in this mode. The user can either decide to take a measurement on the Active Probe or change the Active Probe prior to making a measurement. The details on how to install probe information, take measurements and store the results are provided later in this manual. Shown in Figure 2 is the Operator Mode Main Menu Window. Details on making measurements are provided in Section 8.0. * MEASURE * 1: Measure 2: Select Probe Probe = ER 12 Tag No: Riser A67 Figure Advanced Mode The Advanced Mode allows the user to make measurements in the same manner as the Operator Mode and also allows the use of the advanced capabilities in the MultiCorr MKII. The most sensitive functions in the Advanced Mode are protected with a pass code to prevent data loss or corruption caused by unauthorized operators. Shown in Figure 3 is the Advanced Mode Main Menu Window. There are six (6) general capabilities that can be accessed through the use of the Advanced Mode. The use of these capabilities are described in more detail later in this manual. For now it is only important to understand the general capabilities of the MultiCorr MKII and how these are arranged in the menu structure. 1: Acq. Setup 2: Measure 3: Show Data 4: System 5: Transmit 6: Automatic Figure 3 * MAIN MENU * 1. Acquisition Setup This menu route allows the user to enter the various parameters for each type of probe, enter the tag identification and establish data reading variables such as the type of engineering units. 2. Measure The user can take field measurements exactly the same way as it is accomplished in the Operator Mode. 3. Show Data The user can view all the stored readings for any probe including raw data, calculated results and corrosion rates. 4. System Setup This menu route allows the user to set the date and time, clear the memory of readings, clear the unit of probe parameters, change the pass code and also check the remaining memory. 5. Transmit This feature allows the stored data to be transferred to a computer.

18 MultiCorr MKII - User Manual Page: Automatic THIS IS AN OPTIONAL FUNCTION THAT MAY NOT APPEAR ON ALL UNITS. It allows the user to automatically take and store readings at regular intervals. This function may be added to any MultiCorr MKII. 5.4 Terminal mode The choice of terminal mode allows user to use the MultiCorr MKII as a terminal for MultiLog and/or SandLog stations (depending on which functions installed). If both terminal functions are installed the choice of terminal mode will bring you into a submenu where you can choose between SandLog and MultiLog terminal. When used as a Sand Terminal you cannot operate MultiLog stations and vice versa. If only one terminal function is installed the above mentioned submenu is skipped, and you are brought directly into the installed terminal menu. 5.5 Menu Paths As menu selections are made, the MultiCorr MKII will move to the next appropriate menu. This means that if the MultiCorr MKII needs additional information (For instance, probe parameters) it will automatically move to the appropriate menu. When the unit has the necessary information to proceed with an entry it will skip unnecessary menus. This feature saves the user time and minimizes confusion. There are some general rules which will assist your understanding of this manual and which are applicable to all menus. 1. Throughout this manual the keyboard entries are enclosed in brackets [ ] to distinguish them from other numerical information listings. 2. The Enter Key [Ent] is used to confirm your entry and to move to the next screen. 3. The Exit Key [Exit] is used to leave the entries unchanged and moves you forward in the menu path.

19 MultiCorr MKII - User Manual Page: 19 6 PROBE IDENTIFICATION AND METER SETUP There are general procedures that apply to all probes and these procedures are discussed in this section. Additionally, the user will learn how to check the operational status of the MultiCorr MKII and select the type of engineering units which are to be used for the probe readings. The intent of this section of the manual is only to familiarize the user with general procedures for identifying and labeling all types of probes. Section 7 takes the operator on a step by step procedure for entering the parameters for each specific type of probe. Before taking any measurements the user must assign a unique number to each probe. This number will define the type of probe (For instance ER or LPR) and allow the MultiCorr MKII to know where the readings will be stored. At the same time that probe number is assigned you may also assign a location name, called a Tag Number, to assist you in recognizing the probe you are measuring. Note that if a Tag Number is not assigned, it is possible to enter the Operator Mode and change the probe parameters without entering the Pass Code. To protect the probe parameter setup, a Tag Number must be assigned for the probe. All of the steps necessary to complete each operation are given along with the keys that need to be pressed and the results that will be obtained. 6.1 Selecting Probe Type and Assigning a Probe Number This procedure allows the user to assign a Probe Number without also assigning an identifying Tag Number. 1 [On/Off] Turn on MultiCorr MKII 2 [2] Go to Advanced Mode 3 [1] Go to Acquisition Set Up 4 [1] To Edit Probe 5 [Pass Code number] Enter your personal passcode 6 [Ent] Enter pass code and move to next menu 7 [1] To select a Probe Type 8 [1] through [5] Select the type of probe which will be assigned a number NOTE: Since every MultiCorr MKII is configured for each user, the meter you are using may not have 5 options from which to select. For your general information the five (5) possible key selections and the respective probe functions are as follows: [1] ER [4] Analog [2] LPR [5] Polarization types [3] Galvanic In addition, the two teminal modes, MultiLog and SandLog, are available functions.

20 MultiCorr MKII - User Manual Page: 20 9 [Ent] To move to Next Menu Screen 10 Enter Any No The user can assign any number from 0 through 99. The numbers do not have to be consecutive. It is often helpful to assign the numbers in the same sequence that the operator encounters the probes while taking field readings. HINT: If additional probes will be added to the reading circuit, skip some numbers so consecutive number spaces will be available at a future date. 11 [Ent] Enters the number you selected. From this point on the entries depend on the type of probe being utilized. Section 7 of this manual provides details on how to enter the parameters for each type of probe. The same steps as shown above will be repeated, but the background information will be excluded. 6.2 Installing and Editing Probe Tag Identification The MultiCorr MKII provides two (2) Methods for installing Tag Numbers and probe parameters. 1. Manual installation through the use of key entries on the MultiCorr MKII. 2. Automated downloading of probe files through the use of CorrTag Software Manual Probe and Tag Number Entry Using The MultiCorr MKII This procedure allows the user to assign a Probe Number and also assign an identifying Tag Number. Generally, this is a preferred manual procedure to just assigning a probe number as shown above in Section 6.1. However, to assign a Tag Number using the MultiCorr MKII the user must know the pass code assigned to that particular MultiCorr MKII. 1 [On/Off] Turn on MultiCorr MKII 2 [2] Go to Advanced Mode 3 [1] Go to Acquisition Set Up 4 [1] Edit probe 5 [Pass Code Number] Enter your personal passcode 6 [Ent] To Enter Pass Code and move to the next Menu 7 [1] To Select Probe 8 [1] through [5] Select the type of probe which will be assigned a number 9 [Ent] Enter the probe type you selected 10 Enter Any No The user can assign any number from 0 through 99. The numbers do not have to be consecutive. It is often helpful to assign the numbers in the same sequence that the operator encounters the probes while taking field readings. HINT: If additional probes will be added to the reading circuit, skip some numbers so consecutive number spaces will be available at a future date. 11 [Ent] Enters the number you selected. NOTE: As explained in Section 6.1 above, at this point the available functions in each MultiCorr MKII will vary. 12 [2] Edit Tag no.

21 MultiCorr MKII - User Manual Page: 21 Figure 4 shows the Edit Tag Number window through which an identifying Tag Number may be assigned. Tag Numbers may be up to 10 characters long. To keep the meter as simple as possible, the MultiCorr MKII does not have an alphanumeric keyboard. The alphabet is displayed on the bottom of the screen. Characters are entered by the use of the arrow keys and then selected using the Enter Key [Ent]. 1. Move the Cursor to the desired character using the Arrow Keys and select it by pressing the Enter Key [Ent]. 2. Numbers are selected directly by pressing the Number Keys on the MultiCorr MKII Keyboard. ---EDIT TAG NO-- PROBE = ER 12 { } ABCDEFGHIJKLMNO PQRSTUVWXYZ /( ) + &#$ * <accept> Figure 4 3. Note that you may select an open space in the Tag Number by placing the cursor under the open space (after Z on the screen) and pressing [Ent]. An open space counts as 1 of the maximum 10 allowable characters. 4. To delete an erroneous entry press the Delete Key [Del]. To accept the completed Tag Number move the cursor to [accept] and press the Enter Key [Ent]. From this point on the entries depend on the type of probe being utilized. Section 7 of this manual provides details on how to enter the parameters for each type of probe. The same steps as shown above will be repeated.

22 MultiCorr MKII - User Manual Page: Probe Information Installation Using CorrTag Software CorrOcean has developed a PC software package called CorrTag which allows the MultiCorr MKII user to enter all of the Probe Numbers, Tag Numbers and Probe Parameters in a computer file and then download this information directly into the MultiCorr MKII Meter. When using CorrTag the user can create files containing the probe information for different locations. This allows the user to simply clear all of the existing probe information from the meter and download just the file containing information needed for the geographic area where probe readings will be taken. Some illustrative examples of how CorrTag may be used are as follows: 1. If a plant has more than 200 probes of a given type, simply divide the plant into areas of less than 200 probes and create a separate file for each area. 2. If there are distinct geographical areas where probes are to be measured the user can create a separate file of the probe parameters for each area. Before visiting a location the user can clear the meter, download the probe information and quickly proceed with taking measurements without having to reenter probe information. This also removes the possibility that readings could be mistakenly entered into a probe that is not located in that area. Details on the use of CorrTag are provided in the CorrTag Data Sheets and user manual. The steps necessary to set up the MultiCorr MKII to receive CorrTag files are as follows: 1 [On/Off] Turn on MultiCorr MKII 2 [2] Go to Advanced Mode 3 [1] Go to Acquisition Set Up 4 [2] Load from PC 5 [Pass Code Number] Enter your personal passcode 6 [Ent] To Enter Pass Code and move to the next Menu The MultiCorr MKII is now set to receive any desired CorrTag file. The actual initiation on the data downloading is started from the PC software menu. If your MultiCorr MKII does not provide for a selection between "Manual" and "Using PC" then the internal electronics will need to be upgraded before you may use CorrTag Software. NOTE: If the MultiCorr MKII is not reset to eliminate all of the existing Probe Numbers and parameters the CorrTag files will override any existing Probe Numbers that are the same as in the downloaded file. Resetting the MultiCorr MKII is detailed in Section Reading Setup The MultiCorr MKII allows the user to define both the number of samples taken with each reading and the engineering units for the measurement of ER and LPR Probes.

23 MultiCorr MKII - User Manual Page: Number of Samples per Reading Normally a user would make a probe reading and obtain a display of the result for that reading. However in some cases, especially when the system to be measured is not stable, it may be advantageous to have a single reading represent the average of multiple samples. The MultiCorr MKII allows the user to establish the number of samples that can be averaged to obtain a single displayed reading. The actual raw data samples that make up the single reading may also be viewed by the operator so that the actual measured stability may be reviewed. The procedure for displaying raw data is discussed in Section 10.0, Displaying Measurement Data. The default value set in the MultiCorr MKII is 1 sample for each reading. The user can set the MultiCorr MKII to take up to 50 samples each time a measurement is made. Regardless of the number of samples which is defined only a single average result will be displayed. Setting the number of samples to a number above 1 will enhance the stability of the readings, but it will also increase the time required to perform a measurement. Increasing the number of samples per reading will also increase battery consumption and the total number of readings which may be obtained on a single charge will be reduced. Unless there is a specific reason to suspect a stability problem using the default value of 1 sample per reading is probably adequate. The additional sample readings are only made on the actual reading element step so increasing the number of samples, to say 3, does not require three times as much time as a single sample size. The user may want to test out some multiple sample sizes to optimize the stability of the readings and the time it takes to obtain measurement results. To set the number of samples follow the steps shown below: 1 [On/Off] Turn on MultiCorr MKII 2 [2] Go to Advanced Mode 3 [1] Go to Acquisition Set Up Menu 4 [3] Go to Reading Set Up Menu 5 [Pass Code Number] Enter your personal passcode 6 [Ent] To enter pass code and move to next menu 7 [1] Select No of Samples 8 Enter [1] through [50] Defines number of samples per reading 9 [Ent] Enters sample number and returns to Reading Set Up Menu 10 [Exit] Returns to Acquisition Set Up Menu

24 MultiCorr MKII - User Manual Page: Selection of Engineering Units The MultiCorr MKII allows the user to set their preferred engineering units. There are two (2) types of units for each type of probe. To select the engineering units follow the steps shown below: 1 [On/Off] Turn on MultiCorr MKII 2 [2] Go to Advanced Mode 3 [1] Go to Acquisition Set Up Menu 4 [3] Go to Reading Set Up Menu 5 [Pass Code Number] Enter your personal passcode 6 [Ent] To enter pass code and move to next menu 7 [2] Go to Engineering Unit Menu 8 Enter [1] through [2] Defines the units for every Type of Probe based upon which entry [1] through [2] is pressed. See the Engineering Units selection Chart below. 9 [Ent] Confirms entry and returns user to Read Set Up Menu 10 [Exit] Returns to Acquisition Set Up Menu Once a selection has been made for one type of probe, it establishes the units for all the Probe Functions. For instance, if you select Key [2] (Mils) for an ER probe then the LPR probes will read in mpy and the ER probes in mils/mpy. Listed below is a summary of the units that will be established when making an engineering units selection. The user may also change the engineering units after the data has been transferred to a Corrlist file. Engineering Unit Selection Chart Selection Option Function [1] [2] ER Probes mm, um, mm/y, um/y, (autoscaling) mils, mpy LPR Probes mm/y, um/y (autoscaling) mpy Galvanic Probes A, ma, ua (autoscaling) A, ma, ma (autoscaling) Analog Probes V, mv, uv (autoscaling) V, mv, uv (autoscaling) Polarization probes mv, ma mv, ma

25 MultiCorr MKII - User Manual Page: 25 7 DEFINING PROBE CHARACTERISTICS AND PARAMETERS 7.1 General The concept for the operation of the MultiCorr MKII is that the meter shall be capable of handling different Probe Types and different Probe Manufacturers. In order to accomplish this versatility, the user enters the probe parameters for each individual Probe. The MultiCorr MKII provides an easy and logical selection menu approach to assist the user in entering probe parameters. Probe parameters can be entered from either the Operator Mode or the Advanced Mode which were discussed in Section 5 of this manual. However, identification Tag Numbers can only be installed from the Advanced Mode so all of the examples will use the Advanced Mode menu. The probe parameters should be installed by the system or corrosion engineer. Tag information is protected by a pass code in the meter to avoid tampering or accidental changes by a field technician. A pass code number is supplied with the MultiCorr MKII at the time of delivery but it may be changed by the operator using the Advanced Mode and the existing Pass Code. Once a Tag Number has been defined for a Probe, the probe parametes are protected by the Pass Code. In MultiCorr MKII, raw data, such as resistance for ER Probes and current and voltage for LPR and galvanic Probes, is stored. Thus, when changing probe parameters, the displayed measurement values will change for all measurements for that Probe stored in the instrument. However, the raw data will remain unchanged. Additional Technical Information on each type of probe is provided in the Appendices of this manual. The Appendices also provide details on the formulas used to calculate displayed values and how corrosion rates are calculated.

26 MultiCorr MKII - User Manual Page: Entering Electrical Resistance (ER) Probe Parameters To install ER Probe parameters and assign the Probe Tag Number follow the steps provided below: 1 [On/Off] Turn on MultiCorr MKII 2 [2] Go to Advanced Mode 3 [1] Go to Acquisition Set Up 4 [1] Go to Edit Probe 5 [Pass Code Number] Enter your personal passcode 6 [Ent] To enter Pass Code and move to the next Menu 7 [1] Go to Select Probe 8 [1] To select an ER probe 9 [Ent] To confirm the selection and move to next menu screen 10 Enter Any No The user can assign any number from 0 through b [Ent] To create if not created earlier 11 [1] through [4] See Explanation of Choices below [1] T10 Tubular ER Probe with a 10 mil element thickness [2] T20 Tubular ER Probe with a 20 mil element thickness [3] Wireloop Standard ER Wireloop. A standard ER wireloop is defined as a wireloop probe where the measuring element is twice as long as the Reference Element. If your wireloop is a nonstandard design refer to APPENDIX B "ER Probe Technical Information" [4] Special This can be used for specially designed ER Probes or tubular probes other than T10 and T20. For installing and editing probe tag identification, please see section 6.2.1, step 12 and following instructions. NOTE: If you selected [1] T10 or [2] T20 no additional information is required and the MultiCorr MKII will skip steps Enter the Multiplier If you want to accept the default value provided for the Multiplier proceed with Step 13. The Multiplier is identical to the element thickness. NOTE: See APPENDIX B "ER Probe Technical Information" for a detailed explanation of the Multiplier and a listing of common ER probes. Basically the multiplier is equal to: Element Thickness (Mils) times 25 OR Element Thickness (mm) times [Ent] This Accepts the Multiplier and moves the cursor to Conceal %

27 MultiCorr MKII - User Manual Page: Enter % Conceal Some ER Probes do not have 100% of the measuring element exposed to the corrosive environment. Most elements on ER Probes are fully exposed and have a conceal % of zero (0). If you want to accept the default value for Conceal % proceed with Step 15 without entering a new conceal %. 15 [Ent] Accepts the Conceal % you assigned and displays the last window in this menu path For installing and editing probe tag identification, please see section 6.2.1, step 12 and following instructions. 16 Press either [1] or [2] Press [1] if you want to set up additional probes Press [2] if you are finished. This will also return you to the Edit Probe Menu.

28 MultiCorr MKII - User Manual Page: Entering Linear Polarization Resistance (LPR) Probe Parameters To install LPR Probe parameters and assign the Probe Tag Number follow the steps provided below: 1 [On/Off] Turn on MultiCorr MKII 2 [2] Go to Advanced Mode 3 [1] Go to Acquisition Set Up 4 [1] Go to Edit Probe 5 [Pass Code Number] Enter your personal passcode 6 [Ent] To enter Pass Code and move to the next Menu 7 [1] Go to Select Probe 8 [2] To select LPR Probes 9 [Ent] To confirm selection and move to next screen 10 Enter Any No The user can assign any number from 0 through b [Ent] To create if not created earlier 11 [1] through [3] See Explanation of Choices below [1] Standard This is a 3 electrode LPR probe with electrode area 300 mm 2 and a B Constant of 24 mv. The MultiCorr MKII will skip directly to step 16 if you make this selection. [2] Special 3 el If your LPR Probe has another surface area/ B Constant and has 3 electrodes Press [2] [3] Special 2 el If your LPR Probe has another surface area or has 2 electrodes Press [3] 12 Enter B Constant If you want to accept the default value provided for the B Constant proceed with Step 12 NOTE: The B constant varies with each type of metal and environment. Values may be obtained from Tables or Tafel Curves. In the MultiCorr MKII a value of B=24mV has been used as the default value. This is a typical value for carbon steel in saline water. For additional information see APPENDIX C "LPR Probe Technical Information". 13 [Ent] Accepts the B Constant you assigned and moves the cursor to Area of Electrodes 14 Enter Electrode Area Enter the surface area of the LPR Working Electrode. For information about default values, please see section For additional information on the area of some commonly used probes see APPENDIX C "LPR Probe Technical Information". If you want to accept the default value for the Electrode area proceed with Step [Ent] Accepts the Electrode Area you assigned and displays the last window in this menu path.

29 MultiCorr MKII - User Manual Page: 29 For installing and editing probe tag identification, please see section 6.2.1, step 12 and following instructions. 16 Press either [1] or [2] Press [1] if you want to set up additional probes Press [2] if you are finished. This will also return you to the Acquisition Set Up Menu. 7.4 Entering Galvanic Probe Parameters To install Galvanic Probe parameters and assign the Probe Tag Number follow the steps provided below: [On/Off] Turn on MultiCorr MKII 2 [2] Go to Advanced Mode 3 [1] Go to Acquisition Set Up 4 [1] Go to Edit Probe 5 [Pass Code Number] Enter your personal passcode 6 [Ent] To enter Pass Code and move to the next Menu 7 [1] To select Probe 8 [3] To select Galvanic Probes 9 [Ent] To confirm the selection of Galvanic Probe and to move to next menu screen 10 Enter Any No The user can assign any number from 0 through 199 For installing and editing probe tag identification, please see section 6.2.1, step 12 and following instructions. 11 Press either [1] or Press [1] if you want to set up additional probes [exit] Press [exit] if you are finished. This will also return you to the Acquisition Set Up Menu. NOTE: Additional information on Galvanic Probes may be found in APPENDIX D "Galvanic Probe Technical Information".

MS3600E. High Resolution ER Data Logger Operator s Manual

MS3600E. High Resolution ER Data Logger Operator s Manual MS3600E High Resolution ER Data Logger Operator s Manual Metal Samples A Division of Alabama Specialty Products, Inc. 152 Metal Samples Rd., Munford, AL 36268 Phone: (256) 358-4202 Fax: (256) 358-4515

More information

700 Series 200 Amp Clamp Meters

700 Series 200 Amp Clamp Meters 700 Series 200 Amp Clamp Meters #61-700 #61-701 #61-702 1 2 3 6 5 7 4 8 1. Non-contact voltage (NCV) (#61-701 and #61-702) With the NCV tab on the tip of the clamp close to an AC voltage, press the NCV

More information

User's Guide. MiniTec TM Series Model MN25 MultiMeter

User's Guide. MiniTec TM Series Model MN25 MultiMeter User's Guide MiniTec TM Series Model MN25 MultiMeter Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a

More information

Operations Manual EcoSense ph100

Operations Manual EcoSense ph100 Operations Manual EcoSense ph100 Portable ph, mv and Temperature Instrument English Français Español Deutsch Italiano WARRANTY The EcoSense ph100 Instrument is warranted for one year from date of purchase

More information

PCM-7140 Pulsed Current Source Operation Manual

PCM-7140 Pulsed Current Source Operation Manual PCM-7140 Pulsed Current Source Operation Manual Directed Energy, Inc. 1609 Oakridge Dr., Suite 100, Fort Collins, CO 80525 (970) 493-1901 sales@ixyscolorado.com www.ixyscolorado.com Manual Document 7650-0031

More information

I/O SIGNAL CONDITIONER

I/O SIGNAL CONDITIONER Technical Data Sheet No. TD9809M Rev. F Date of Issue: December 9, 2009 OPERATING MANUAL I/O SIGNAL CONDITIONER CAUTION: THIS PRODUCT DOES NOT PROVIDE GALVANIC ISOLATION. DO NOT ATTEMPT USE OF THIS PRODUCT

More information

5.8.1 General System Description Example System description System Operation 30

5.8.1 General System Description Example System description System Operation 30 Page: 2 of 125 Table of contents: 1 INTRODUCTION 5 1.1 GUIDE TO THE USER MANUAL 5 1.2 ABBREVIATIONS AND DEFINITIONS 5 2 MAIN DATA 5 2.1 COPYRIGHT & LICENCE NOTICE 5 2.2 MULTITREND PLATFORM REQUIREMENTS

More information

REMOTE FRONT PANEL EMULATION

REMOTE FRONT PANEL EMULATION REMOTE FRONT PANEL EMULATION USER REFERENCE GUIDE 2500 Application Software Series Part Number: 3-9003-170 Revision 1.0 MARCH 1991 EMULATOR GUIDE DANIEL INDUSTRIES, INC. REMOTE FRONT PANEL EMULATION USER

More information

PMDX-105. I/O Option Riser Board User s Manual. Document Revision: 1.1 Date: 7 September 2004 PCB Revision: PCB-443A

PMDX-105. I/O Option Riser Board User s Manual. Document Revision: 1.1 Date: 7 September 2004 PCB Revision: PCB-443A PMDX-105 I/O Option Riser Board User s Manual Date: 7 September 2004 PMDX Web: http://www.pmdx.com 7432 Alban Station Blvd., A105 Phone: +1 (703) 912-4991 Springfield, VA 22150-2321 USA FAX: +1 (703) 912-5849

More information

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: This meter is

More information

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE

More information

CORD-XL Dual-Channel Electronic Chart Recorder User s Manual

CORD-XL Dual-Channel Electronic Chart Recorder User s Manual CORD-XL Dual-Channel Electronic Chart Recorder User s Manual Rohrback Cosasco Systems Inc. 11841 E. Smith Ave Santa Fe Springs, CA 90670 Tel: (562) 949-0123 Fax: (562) 949-3065 P/N 720701-Manual Rev E

More information

PixController, Inc. Wireless Magnetic Switch Sensor For Doors, Windows, and Gates

PixController, Inc. Wireless Magnetic Switch Sensor For Doors, Windows, and Gates PixController, Inc. Wireless Magnetic Switch Sensor For Doors, Windows, and Gates Model: SEN-420 User s Manual Version 1.00 WARRANTY REGISTRATION PixController, Inc. warrants products sold by it and guarantees

More information

PixController, Inc. Wireless Switch Sensor For Normally Open (NO) and Normally Closed (NC) Sensors

PixController, Inc. Wireless Switch Sensor For Normally Open (NO) and Normally Closed (NC) Sensors PixController, Inc. Wireless Switch Sensor For Normally Open (NO) and Normally Closed (NC) Sensors Model: SEN-410 User s Manual Version 1.00 WARRANTY REGISTRATION PixController, Inc. warrants products

More information

INSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs

INSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs INSTRUCTION MANUAL Model 421504 Dual Input RTD Thermometer Measures two temperatures simultaneously Dual RTD probe inputs Clock and Elapsed Timer functions Special functions include Data Hold, MIN/MAX/AVG,

More information

Roxar CorrLog High Accuracy Corrosion Monitoring

Roxar CorrLog High Accuracy Corrosion Monitoring FA-T200-C Product Data Sheet 01.12.2016 High Accuracy Corrosion Monitoring Fast and reliable corrosion monitoring is a key for efficient corrosion management and control. In-line corrosion monitoring provides

More information

INSTRUCTION MANUAL. Model True RMS AC/DC 30A Mini Clamp-on Meter. Introduction. True RMS AC Current and Voltage

INSTRUCTION MANUAL. Model True RMS AC/DC 30A Mini Clamp-on Meter. Introduction. True RMS AC Current and Voltage INSTRUCTION MANUAL Model 380942 True RMS AC/DC 30A Mini Clamp-on Meter True RMS AC Current and Voltage Measure low current with high resolution to 0.1mA AC and 1mA DC Auto Power Off One touch DCA zero

More information

Secured Series: Hub Plus Kit Single Door Controller Package Installation Manual

Secured Series: Hub Plus Kit Single Door Controller Package Installation Manual Secured Series: Hub Plus Kit Single Door Controller Package Installation Manual This package is designed to simplify the connections to our Secured Series Hub Plus Controller. This will translate into

More information

User Manual Landscape Irrigation Controller

User Manual Landscape Irrigation Controller User Manual Landscape Irrigation Controller Drip Irrigation Micro Sprinklers Sprinklers Table of Contents Page 1 Introduction 2 2 About BERMAD BIC 1 Controller 2 3 Components Identification 2 4 LCD Display

More information

DVC-350C CALIBRATORS, INC HANDHELD BATTERY-POWERED VOLTAGE CALIBRATOR FEATURES INTRODUCTION GENERAL DESCRIPTION

DVC-350C CALIBRATORS, INC HANDHELD BATTERY-POWERED VOLTAGE CALIBRATOR FEATURES INTRODUCTION GENERAL DESCRIPTION CALIBRATORS, INC HANDHELD BATTERY-POWERED VOLTAGE CALIBRATOR FEATURES HAND-HELD, LIGHT WEIGHT 11 OUNCES (342 GRAMS) LABORATORY ACCURACY OF +/- 0.015% OF FSR AT 25c, TRACEABLE TO N.I.S.T. 4 ½ DIGIT LCD

More information

User Guide. ExStik TM Model RE300 Waterproof ORP Meter. Patent Pending RE300 ORP

User Guide. ExStik TM Model RE300 Waterproof ORP Meter. Patent Pending RE300 ORP 0 User Guide ExStik TM Model RE300 Waterproof ORP Meter Patent Pending 500 MV 1000 TM RE300 ORP ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE button 4. CAL

More information

STATIC METERS Models 211 & 212. SURVEYORSTAT Models 211XL & 212XL

STATIC METERS Models 211 & 212. SURVEYORSTAT Models 211XL & 212XL STATIC METERS Models 211 & 212 SURVEYORSTAT Models 211XL & 212XL Operating Manual 5/09 1.0 GENERAL DESCRIPTION The Model 211 and 212 Static Meters, shown in Figures 1.0-1a and b, are accurate, compact

More information

User's Manual. Model Heavy Duty Datalogging Module with Windows Software

User's Manual. Model Heavy Duty Datalogging Module with Windows Software User's Manual Model 380340 Heavy Duty Datalogging Module with Windows Software Stores data for later recall and analysis Can be used with any Extech Heavy Duty meter Selectable recording interval Battery

More information

AquaController Apex Probe Module 2

AquaController Apex Probe Module 2 AquaController Apex Probe Module 2 Setup Guide Table of Contents PROBE MODULE 2 INTRODUCTION... 1 FEATURES... 1 PHYSICAL INSTALLATION... 1 Initial Connections... 1 Startup... 2 VERIFY THE INSTALLATION...

More information

PIM-Mini Pulsed Current Source Operation Manual

PIM-Mini Pulsed Current Source Operation Manual PIM-Mini Pulsed Current Source Operation Manual Directed Energy, Inc. 1609 Oakridge Dr., Suite 100, Fort Collins, CO 80525 (970) 493-1901 sales@ixyscolorado.com www.ixyscolorado.com Manual Document 7650-0007

More information

User's Guide. Phase Sequence and Motor Rotation Tester Model

User's Guide. Phase Sequence and Motor Rotation Tester Model User's Guide Phase Sequence and Motor Rotation Tester Model 480403 Introduction Congratulations on your purchase of the Extech Model 408403 Motor and Phase Rotation Indicator. This handheld instrument

More information

USER MANUAL MODEL 2017P MODEL 2017P60. RS-232 to 20mA and. RS-232 to 60mA Current Loop Converters

USER MANUAL MODEL 2017P MODEL 2017P60. RS-232 to 20mA and. RS-232 to 60mA Current Loop Converters USER MANUAL MODEL 2017P RS-232 to 20mA and MODEL 2017P60 RS-232 to 60mA Current Loop Converters 07M2017P-E Doc# 073051UE Revised 5/7/96 SALES OFFICE (301) 975-1000 TECHNICAL SUPPORT (301) 975-1007 1.0

More information

USER MANUAL. MODEL 2014 Interface Powered RS-530 to V.35 Converter. SALES OFFICE (301) TECHNICAL SUPPORT (301)

USER MANUAL. MODEL 2014 Interface Powered RS-530 to V.35 Converter. SALES OFFICE (301) TECHNICAL SUPPORT (301) USER MANUAL MODEL 2014 Interface Powered RS-530 to V.35 Converter An ISO-9001 Certified Company Part #07M2014-C Doc. #070011UC Revised 4/22/98 SALES OFFICE (301) 975-1000 TECHNICAL SUPPORT (301) 975-1007

More information

EPS Power Supply

EPS Power Supply EPS - 600 Power Supply Installation and Operation Manual Version 1.0 *This instrument is intended for laboratory use only Index A. Important Notice ----------------------------------------------------------------

More information

NOTE: The F2 button and the 4 LED annunciators do not function on the LD-ACF-I. To use these features, ask about the LD-ACF-R4/R4A.

NOTE: The F2 button and the 4 LED annunciators do not function on the LD-ACF-I. To use these features, ask about the LD-ACF-R4/R4A. Large display AC frequency indicator The is a sophisticated microprocessorbased indicator designed specifically to monitor AC frequency. Intuitive scrolling text menus, (activated from the front panel),

More information

Industrial PCMCIA ATA CARD. HERMIT-A Series. Product Specification INDUSTRIAL PCMCIA ATA CARD. Version 01V1 Document No. 100-xAFCHA-HATS May 2016

Industrial PCMCIA ATA CARD. HERMIT-A Series. Product Specification INDUSTRIAL PCMCIA ATA CARD. Version 01V1 Document No. 100-xAFCHA-HATS May 2016 Industrial PCMCIA ATA CARD HERMIT-A Series Product Specification INDUSTRIAL PCMCIA ATA CARD Version 01V1 Document No. 100-xAFCHA-HATS May 2016 APRO CO., LTD. Phone: +88628226-1539 Fax: +88628226-1389 This

More information

User Guide. Moisture Meter. Model MO250

User Guide. Moisture Meter. Model MO250 User Guide Moisture Meter Model MO250 Introduction Congratulations on your purchase of the Extech MO250 Moisture Meter. The MO250 detects moisture in wood and other building materials such as brick, wall

More information

User's Guide. Model High Precision Quad Output DC Power Supply

User's Guide. Model High Precision Quad Output DC Power Supply User's Guide Model 382270 High Precision Quad Output DC Power Supply Introduction Congratulations on your purchase of the Extech 382270 DC Power Supply. The Model 382270 can be used for many applications

More information

MODEL 3150 SUSPENDED SOLIDS ANALYZER MODEL 35/35L OPTICAL SUSPENDED SOLIDS SENSOR

MODEL 3150 SUSPENDED SOLIDS ANALYZER MODEL 35/35L OPTICAL SUSPENDED SOLIDS SENSOR MODEL 3150 SUSPENDED SOLIDS ANALYZER MODEL 35/35L OPTICAL SUSPENDED SOLIDS SENSOR REVISION 22 August 2016 Insite Instrumentation Group, Inc 80 Whisperwood Blvd. Suite 107 Slidell, LA 70458 Phone: (985)

More information

ELECTROSTATIC FIELDMETER

ELECTROSTATIC FIELDMETER ELECTROSTATIC FIELDMETER Small hand-held Electro-Fieldmeter with digital display designed to measure electrostatic voltage potentials (with pre-selected distance) according the fieldmill influence principal.

More information

PMDX-170 Slotted Optical Sensor

PMDX-170 Slotted Optical Sensor PMDX-170 Slotted Optical Sensor User s Manual Date: 20 May 2009 PMDX Web: http://www.pmdx.com 9704-D Gunston Cove Rd Phone: +1 (703) 372-2975 Lorton, VA 22079-2366 USA FAX: +1 (703) 372-2977 PMDX-170_Manual_10.doc

More information

User's Guide. Temperature / Humidity Datalogger. Model 42270

User's Guide. Temperature / Humidity Datalogger. Model 42270 User's Guide Temperature / Humidity Datalogger Model 42270 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment

More information

ARIENSE SENSE YOUR WORLD. 3-Axis Force Sensor FSE103. User Guide. Version 2.1. July VARIENSE INC.

ARIENSE SENSE YOUR WORLD. 3-Axis Force Sensor FSE103. User Guide. Version 2.1. July VARIENSE INC. V ARIENSE SENSE YOUR WORLD 3-Axis Force Sensor FSE103 Version 2.1 July 2017 2016 VARIENSE INC. Contents 1 Product Overview 2 2 Safety 3 3 Setup 4 3.1 Mounting the sensor...........................................

More information

User's Guide. Extech AM A AC Analog Clamp Meter

User's Guide. Extech AM A AC Analog Clamp Meter User's Guide Extech AM300 300A AC Analog Clamp Meter Introduction Congratulations on your purchase of the Extech AM300 Analog Clamp Meter. This device measure AC Voltage and Current, DC Voltage, and Resistance.

More information

USER MANUAL. MODEL 2011 High Speed Asynchronous to Synchronous Converter

USER MANUAL. MODEL 2011 High Speed Asynchronous to Synchronous Converter USER MANUAL MODEL 011 High Speed Asynchronous to Synchronous Converter Part# 07M011-A Doc# 0601UA Revised 03/16/94 SALES ICE (301) 975-1000 TECHNICAL SUPPORT (301) 975-1007 http://www.patton.com 1.0 WARRANTY

More information

HI ma Simulator

HI ma Simulator Instruction Manual HI 931002 4-20 ma Simulator http://www.hannainst.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you for choosing a Hanna Instruments Product. Please

More information

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only.

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only. Concorde 600 Power Supply Operation Manual Cat.no. R10-1001011 *This instrument is intended for laboratory use only http://www.recenttec.com E-mail : support@recenttec.com Version 1.1 Packing List x 1

More information

User Guide. Model Temperature Datalogger Kit Model Temperature and Humidity Datalogger Kit Model SW276 Datalogging Software SW276

User Guide. Model Temperature Datalogger Kit Model Temperature and Humidity Datalogger Kit Model SW276 Datalogging Software SW276 User Guide Model 42265 Temperature Datalogger Kit Model 42275 Temperature and Humidity Datalogger Kit Model SW276 Datalogging Software SW276 Introduction Congratulations on your purchase of Extech Instrument

More information

User Guide. Control Box. RoscoLED TM.

User Guide. Control Box. RoscoLED TM. RoscoLED TM Control Box User Guide This guide applies to the following RoscoLED Control Box models: RoscoLED Control Box 300W/Static White (293 22250 0000) RoscoLED Control Box 400W/VariWhite (293 22260

More information

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810 Operation Manual Hand-held ph/mv/temperature Meter ph6810 CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS FUNCTIONS..... 4

More information

MODBUS MANUAL For the AccUView UV Transmission Monitor

MODBUS MANUAL For the AccUView UV Transmission Monitor MODBUS MANUAL For the AccUView UV Transmission Monitor HF scientific, inc. 3170 Metro Parkway Ft. Myers, FL 33916 Phone: 239-337-2116 Fax: 239-332-7643 E-Mail: info@hfscientific.com Website: www.hfscientific.com

More information

Integrating Sound Level Datalogger Model

Integrating Sound Level Datalogger Model User's Guide Integrating Sound Level Datalogger Model 407780 Introduction Congratulations on your purchase of the Extech 407780 Integrating Sound Level Meter. The 407780 with programmable integrating time

More information

PSM-2 POWER SUPPLY MODULE INSTRUCTIONS

PSM-2 POWER SUPPLY MODULE INSTRUCTIONS PSM2-09.doc 1 Rev: 8-28-01 PSM-2 POWER SUPPLY MODULE INSTRUCTIONS PSM Model 2 equipped with Field Programmable Output Voltages MODEL PRODUCT OFFERING / VARIATIONS: PSM-2 Metal case housed surface mountable

More information

BatteryCheck USER MANUAL BATTERY MANAGEMENT TECHNOLOGY THAT POWERS YOUR ADVENTURES.

BatteryCheck USER MANUAL BATTERY MANAGEMENT TECHNOLOGY THAT POWERS YOUR ADVENTURES. BatteryCheck USER MANUAL BM PRO - 19 Henderson Road, Knoxfield 3180, Victoria, Australia Phone +61 3 9763 0962 Fax +61 3 9763 8789 Email sales@teambmpro.com Web www.teambmpro.com BATTERY MANAGEMENT TECHNOLOGY

More information

PMDX-103. Parallel Port Isolator Board. User s Manual. Document Revision: 1.2 Date: 20 February 2007 PCB Revision: PCB-447B

PMDX-103. Parallel Port Isolator Board. User s Manual. Document Revision: 1.2 Date: 20 February 2007 PCB Revision: PCB-447B PMDX-103 Parallel Port Isolator Board User s Manual Date: 20 February 2007 PMDX Web: http://www.pmdx.com 9704-D Gunston Cove Rd Phone: +1 (703) 372-2975 Lorton, VA 22079-2366 USA FAX: +1 (703) 372-2977

More information

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics PIECAL 322 Automated Thermocouple Calibrator Operating Instructions Product Description Easy to use With the PIECAL 322-1 you can check & calibrate all your thermocouple instruments and measure thermocouple

More information

USER MANUAL CCHP-6000-XX 6kJ/s Capacitor Charging Power Supply

USER MANUAL CCHP-6000-XX 6kJ/s Capacitor Charging Power Supply 26 Ward Hill Ave., Bradford, MA 01835 Phone: 978-241-8260 Fax: 978-241-8262 USER MANUAL CCHP-6000-XX 6kJ/s Capacitor Charging Power Supply 02001235 Rev 1 CCHP-6000-XX User Manual Page 1 of 14 Table of

More information

MS2601E High Resolution ER 4-20mA Transmitter Operator s Manual

MS2601E High Resolution ER 4-20mA Transmitter Operator s Manual MS2601E High Resolution ER 4-20mA Transmitter Operator s Manual Metal Samples Company A Division of Alabama Specialty Products, Inc. 152 Metal Samples Rd., Munford, AL 36268 Phone: (256) 358-4202 Fax:

More information

User s Guide. 600A AC Clamp Meter. Model 38387

User s Guide. 600A AC Clamp Meter. Model 38387 User s Guide 600A AC Clamp Meter Model 38387 Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.

More information

HI ma Simulator

HI ma Simulator Instruction Manual HI 931002 4-20 ma Simulator http://www.hannacan.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you for choosing a Hanna Instruments Product. Please

More information

INSTRUCTION MANUAL. Model 600C Radio Remote Control Handset

INSTRUCTION MANUAL. Model 600C Radio Remote Control Handset INSTRUCTION MANUAL Model 600C Radio Remote Control Handset The 600 Series remote control handset and controller is a wireless system comprised of a mobile, battery operated remote handset commanding numerous

More information

Rev. A. ANC Series RS-485/RS-422 Synchronous Clock Display. Antona Corporation (818) URL:

Rev. A. ANC Series RS-485/RS-422 Synchronous Clock Display. Antona Corporation (818) URL: Rev. A ANC - 7020 Series RS-485/RS-422 Synchronous Clock Display Antona Corporation, Los Angeles, CA Antona Corporation (818)783-4299 URL:http://www.antona.com 1 Antona Corporation Copyright Copyright

More information

SC-9

SC-9 OPERATING INSTRUCTIONS FOR SC-9 www.sportablescoreboards.com 1 Table of Contents INSTRUCTIONS FOR REPORTING SHIPPING DAMAGE... 3 GETTING STARTED... 6 CABLE SYSTEMS... 6 WIRELESS SYSTEMS... 6 SUPPLYING

More information

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389 User s Guide 600A True RMS AC/DC Clamp Meter Model 38389 Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further

More information

Model 8702 DP-CALC Micromanometer

Model 8702 DP-CALC Micromanometer Ventilation Testing/Balancing Model 8702 DP-CALC Micromanometer Operation and Service Manual 1980260, Revision F July 2006 Model 8702 DP-CALC Micromanometer Operation and Service Manual 1980260, Revision

More information

TC100 Precision Thermocouple Calibrator

TC100 Precision Thermocouple Calibrator TC100 Precision Thermocouple Calibrator Table of Contents 1. Introduction.....................1 2. Accessories....................5 3. Set-Up Basic/Advanced...........5 4. Operating Procedure.............6

More information

User Manual DOCUMENT #601119REF. EcoSense ph1000a. ph, mv (ORP), and Temperature Benchtop Meter USER MANUAL. English

User Manual DOCUMENT #601119REF. EcoSense ph1000a. ph, mv (ORP), and Temperature Benchtop Meter USER MANUAL. English User Manual DOCUMENT #601119REF EcoSense ph1000a ph, mv (ORP), and Temperature Benchtop Meter USER MANUAL English Item #601119REF Rev C, October 2017 For the latest version of this manual, visit ysi.com

More information

SATA III MLC CFast Card

SATA III MLC CFast Card SATA III MLC CFast Card HERMES-G Series Product Specification APRO MLC SATA III CFast Card Version 01V2 Document No. 100-xPCFA-JGTM April 2015 APRO CO., LTD. Phone: +88628226-1539 Fax: +88628226-1389 This

More information

Seven Channel Remote Control

Seven Channel Remote Control Seven Channel Remote Control Page 1 Versa7 Kit Contents The Versa7 is designed to remotely control up to seven devices, or functions. All seven functions can be controlled with the supplied four button

More information

SATA II HDD Canister KISS DA 435 Quick Reference Guide

SATA II HDD Canister KISS DA 435 Quick Reference Guide SATA II HDD Canister KISS DA 435 Quick Reference Guide If it s embedded, it s Kontron 1. Table of Contents SATA II HDD Canister KISS DA 435 1. Table of Contents 1. Table of Contents... 1 2. Important Information...

More information

SLA Battery Capacity Analyzer

SLA Battery Capacity Analyzer Model: 601B SLA Battery Capacity Analyzer USER MANUAL Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation,

More information

VELOCICALC Air Velocity Meter

VELOCICALC Air Velocity Meter ENERGY AND COMFORT Ventilation Testing VELOCICALC Air Velocity Meter Model 9525 Operation and Service Manual Copyright TSI Incorporated / May 2007 / All rights reserved. Address TSI Incorporated / 500

More information

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements www.adwainstruments.com Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting

More information

STYLISH LED DESK LAMP WITH DIFFERENT LIGHT MODES

STYLISH LED DESK LAMP WITH DIFFERENT LIGHT MODES STYLISH LED DESK LAMP WITH DIFFERENT LIGHT MODES User manual, Operating- and safety instructions, Warranty conditions LED 12 MON Item no.: MROS501 1 Getting started and operation 1. Setting up and connecting

More information

Heavy Duty Datalogger Module

Heavy Duty Datalogger Module User's Guide Heavy Duty Datalogger Module Model 380340 Introduction Congratulations on your purchase of Extech s 380340 Datalogger Module. The Datalogger connects to and records data from Extech Heavy

More information

User's Guide. Vane Thermo-Anemometer Datalogger. Model Introduction

User's Guide. Vane Thermo-Anemometer Datalogger. Model Introduction User's Guide Vane Thermo-Anemometer Datalogger Model 451126 Introduction Congratulations on your purchase of Extech's Thermo-Anemometer Datalogger. This Vane-type Anemometer can indicate Air Velocity in

More information

OPERATING AND SERVICE MANUAL. Universal Interface Device 47

OPERATING AND SERVICE MANUAL. Universal Interface Device 47 OPERATING AND SERVICE MANUAL Universal Interface Device 47 MAGNA-POWER ELECTRONICS, INC. 39 ROYAL ROAD, FLEMINGTON, NJ 08822 May 24, 2012 SAFETY NOTICE Universal Interface Device 47 (UID46) connects

More information

PHASETRONICS. SCR Power Control Specialists. EP1 Series Power Control Single Phase SCR Amps OPERATION & SERVICE MANUAL

PHASETRONICS. SCR Power Control Specialists. EP1 Series Power Control Single Phase SCR Amps OPERATION & SERVICE MANUAL PHASETRONICS Specialists EP1 Series Power Control Single Phase SCR 10-50 Amps OPERATION & SERVICE MANUAL Phasetronics, Inc. P.O. Box 5988 1600 Sunshine Drive Clearwater, FL 33765 (727)573-1900 FAX(727)573-1803

More information

Industrial Semi-Metal USB

Industrial Semi-Metal USB Industrial Semi-Metal USB Generation 3L HERMIT-C Series Product Specification INDUSTRIAL Semi-Metal USB Flash Disk Generation 3L Version 01V0 Document No. 100-xMUFD-HCT3L December, 2015 APRO CO., LTD.

More information

Product Description. S&C IntelliNode Interface Module. Table of Contents. Instruction Sheet

Product Description. S&C IntelliNode Interface Module. Table of Contents. Instruction Sheet S&C IntelliNode Interface Module Product Description Table of Contents Section Page Section Page Introduction Qualified Persons.......................... 2 Read this Instruction Sheet..................

More information

MP3 Speaker USER GUIDE

MP3 Speaker USER GUIDE MP3 Speaker USER GUIDE Jazwares, Inc. 2012 CONTENTS Please read the instructions along with the Speaker carefully before you use it, so that you can operate it conveniently. WELCOME, Warnings & Safety

More information

HCC-CW-201A. CWDM Analyzer. Operation Guide

HCC-CW-201A. CWDM Analyzer. Operation Guide CWDM Analyzer Operation Guide Table of Contents Description... 3 Features... 3 Specifications... 4 Safety Information... 5 Preparing for Operation... 6 Unpacking the instrument... 6 Front keys definition...

More information

FIELD MICROOPTODE METER USER MANUAL

FIELD MICROOPTODE METER USER MANUAL FIELD MICROOPTODE METER USER MANUAL Field MicroOptode Meter User Manual Copyright 2017 Unisense A/S Version May 2017 FIELD MICROOPTODE METER USER MANUAL UNISENSE A/S TABLE OF CONTENTS FIELD MICROOPTODE

More information

MLC. Semi Metal USB. Generation 5 / 5L HERCULES-PB Series. Product Specification MLC. Semi Metal USB Flash Disk Generation 5/5L

MLC. Semi Metal USB. Generation 5 / 5L HERCULES-PB Series. Product Specification MLC. Semi Metal USB Flash Disk Generation 5/5L MLC Semi Metal USB Generation 5 / 5L HERCULES-PB Series Product Specification MLC Semi Metal USB Flash Disk Generation 5/5L Version 01V0 Document No. 100-xMUFD-MPBTMB5 MAY 2016 APRO CO., LTD. Phone: +88628226-1539

More information

MLC. Semi-Metal USB. Generation 3L HERMIT-C Series. Product Specification MLC. Semi-Metal USB Flash Disk Generation 3L

MLC. Semi-Metal USB. Generation 3L HERMIT-C Series. Product Specification MLC. Semi-Metal USB Flash Disk Generation 3L MLC Semi-Metal USB Generation 3L HERMIT-C Series Product Specification MLC Semi-Metal USB Flash Disk Generation 3L Version 01V0 Document No. 100-xMUFD-HCTMB3L JUNE, 2016 APRO CO., LTD. Phone: +88628226-1539

More information

FTC-200 X-ray Tube Controller Users Manual

FTC-200 X-ray Tube Controller Users Manual Tubes FTC-200 X-ray Tube Controller TUB-MAN-2101 Rev. A 09/09/2009TTUB Table of Contents List of Tables...3 List of Figures...3 Introduction...3 Manual Set...3 About this Manual...4 X-Ray Safety Information...4

More information

4-20mA Display. Installation and Operation Manual. Rev 4/3/2019 Part #

4-20mA Display. Installation and Operation Manual. Rev 4/3/2019 Part # 4-20mA Display Installation and Operation Manual Rev 4/3/2019 Part # 12050769 Table of Contents DOCUMENTATION CONVENTIONS... 2 Section 1: System Description... 3 Function and Theory... 3 Section 2: System

More information

PC 150. Digital Bench Scale. Operation Manual

PC 150. Digital Bench Scale. Operation Manual PC 150 Digital Bench Scale Operation Manual Revision 1.2 September 14, 2000 1996-2000 Transcell Technology, Inc. Contents subject to change without notice. Transcell Technology, Inc. 35 Waltz Drive Wheeling,

More information

PIECAL 520B & 521B Thermocouple Source Operating Instructions

PIECAL 520B & 521B Thermocouple Source Operating Instructions PIECAL 520B & 521B Thermocouple Source Operating Instructions (Shown without optional boot) Product Description (Shown with optional boot) Easy to use With the PIECAL 520B/521B you can check & calibrate

More information

OPERATING AND SERVICE MANUAL. Universal Interface Device 47

OPERATING AND SERVICE MANUAL. Universal Interface Device 47 OPERATING AND SERVICE MANUAL Universal Interface Device 47 MAGNA-POWER ELECTRONICS, INC. 39 ROYAL ROAD, FLEMINGTON, NJ 08822 May 24, 202 SAFETY NOTICE Universal Interface Device 47 (UID47) connects two

More information

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide Model P4017 Single Channel USB Oscilloscope Quick Start Guide General Warranty BNC warrants that the product will be free from defects in materials and workmanship for 3 years from the date of purchase

More information

AquaController Apex Probe Module 2

AquaController Apex Probe Module 2 AquaController Apex Probe Module 2 Setup Guide Table of Contents PROBE MODULE 2 INTRODUCTION... 1 FEATURES... 1 PHYSICAL INSTALLATION... 1 Initial Connections... 1 Startup... 2 VERIFY THE INSTALLATION...

More information

User s Guide. RP7000S Series Single-Ended Active Probe. Nov RIGOL Technologies, Inc.

User s Guide. RP7000S Series Single-Ended Active Probe. Nov RIGOL Technologies, Inc. User s Guide RP7000S Series Single-Ended Active Probe Nov. 2013 RIGOL Technologies, Inc. Guaranty and Declaration Copyright 2013 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL

More information

Kinetix 300 Memory Module Programmer

Kinetix 300 Memory Module Programmer Kinetix 300 Memory Module Programmer Catalog Number 2097-PGMR Topic About the Memory Module Programmer 1 Parts List 3 Batteries Operation 4 Using Memory Module Programmer 6 Switch On/Off Memory Module

More information

PV Remote Unit A 1378 Instruction manual Version 2.1, Code no

PV Remote Unit A 1378 Instruction manual Version 2.1, Code no PV Remote Unit A 1378 Instruction manual Version 2.1, Code no. 20 752 010 Distributor: Manufacturer: METREL d.d. Ljubljanska cesta 77 1354 Horjul Slovenia web site: http://www.metrel.si e-mail: metrel@metrel.si

More information

WS115 BREAKOUT ADAPTER USER MANUAL

WS115 BREAKOUT ADAPTER USER MANUAL STATIC WATER LEVEL MEASUREMENT WS115 BREAKOUT ADAPTER USER MANUAL www.enoscientific.com Page 1 Eno Scientific 1606 Faucette Mill Rd Hillsborough, NC 27278 USA www.enoscientific.com 910-778-2660 Copyright

More information

Global Water Instrumentation, Inc.

Global Water Instrumentation, Inc. Instrumentation, Inc. 11390 Amalgam Way Gold River, CA 95670 T: 800-876-1172 Int l: (916) 638-3429, F: (916) 638-3270 Display: EZ100 11/12/04-1 - Congratulations on your purchase of the EZ100 Display.

More information

MODEL ATV 4/6 BATTERY POWERED DIGITAL VACUUM INSTRUMENT

MODEL ATV 4/6 BATTERY POWERED DIGITAL VACUUM INSTRUMENT MODEL ATV 4/6 BATTERY POWERED DIGITAL VACUUM INSTRUMENT Contains Operating and Programming Information Model ATV 4/6 Battery Powered Digital Vacuum Instrument ã 1996, Teledyne Electronic Technologies,

More information

Industrial Semi Metal USB

Industrial Semi Metal USB Industrial Semi Metal USB Generation 5 HERCULES-PB Series Product Specification INDUSTRIAL Semi Metal USB Flash Disk Generation 5 Version 01V0 Document No. 100-xMUFD-MPBT5 July 2016 APRO CO., LTD. Phone:

More information

CL600 SIMPLE LOGGER RMS CLAMP-ON MODULE E N G L I S H. User Manual

CL600 SIMPLE LOGGER RMS CLAMP-ON MODULE E N G L I S H. User Manual SIMPLE LOGGER RMS CLAMP-ON MODULE CL600 E N G L I S H User Manual Owner s Record The serial number for the Simple Logger Model CL600 is located on the back of the instrument. Please record this number

More information

Industrial Rugged Metal USB

Industrial Rugged Metal USB Industrial Rugged Metal USB Generation 4SB HERCULES-PB Series Product Specification INDUSTRIAL Rugged Metal USB Flash Disk Generation 4SB Version 01V0 Document No. 100-xRUFD-MPB4SB August 2017 APRO CO.,

More information

SATA III MLC CFast Card

SATA III MLC CFast Card SATA III MLC CFast Card PHANES-C Series Product Specification APRO MLC SATA III CFast Card Version 01V0 Document No. 100-xPCFA-JGTM April 2015 APRO CO., LTD. Phone: +88628226-1539 Fax: +88628226-1389 This

More information

E2KQ-X. Fluororesin-coated Capacitive Sensor with Sensitivity Adjuster. Chemical-resistant Proximity Sensor. Ordering Information

E2KQ-X. Fluororesin-coated Capacitive Sensor with Sensitivity Adjuster. Chemical-resistant Proximity Sensor. Ordering Information Chemical-resistant Proximity Sensor E2KQ-X CSM_E2KQ-X_DS_E_4_3 Fluororesin-coated Capacitive Sensor with Sensitivity Adjuster Excellent resistance against chemicals and oil with fluororesincoated case.

More information

3000 Series Instructions Read Instructions Carefully!

3000 Series Instructions Read Instructions Carefully! Series Instructions Read Instructions Carefully! With buttons Without buttons CURTIS INSTRUMENTS, INC. Kisco Avenue, Mt. Kisco, NY 9 Tel. (9) 9 www.curtisinst.com. Technical Specifications. Electrical

More information

Heavy Duty Datalogger Module

Heavy Duty Datalogger Module User's Guide Heavy Duty Datalogger Module Model 380340 Introduction Congratulations on your purchase of Extech s 380340 Datalogger Module. The Datalogger connects to and records data from Extech Heavy

More information