1 SCOPE This section is intended for the guidance of NDE personnel qualified in accordance with the requirements of M&NP Training and Qualification of Personnel. The inspection processes and procedures described in this document may be used for line pipe, piping components and pressure vessel plates. Ultrasonic Thickness testing may also be used to evaluate the thickness of the parent material after grinding etc. This procedure is not intended to be exhaustive in the treatment of ultrasonic technique as it is provided for general guidance only. SAFETY Prior to excavation for inspection, verification shall be made that the pipeline has been established at a safe working pressure in accordance with the requirements of O&MSM Section 12, Reference 17 - Safe Working Pressure. Defect excavation shall then be completed in accordance with the requirements of O&MSM Section 11, Reference 13 - Defect Excavation and Assessment. EQUIPMENT A. INITIAL SETUP 1. Introduction Figure 1 T-MIKE E For those settings which are seldom changed during operation of the T-Mike E the INITIAL SET UP procedure is designed to allow the operator access to those settings without unnecessary complication of the units normal operation. These settings include: Backlight on/off Selection of Metric or English units of measure Calibration Lock Page 1 of 10

2 The INITIAL SET UP procedure is accessed by holding the CAL key down and pressing the ON/OFF key. When this is done the T-Mike E will display the version code of the operating software contained in the unit until the CAL key is released. This information will be useful should the unit require service or updated software is desired. Upon release of the CAL key the SETUP procedure continues as follows: 2. Backlight On/Off When the CAL key is first released the T-Mike E will display the words ON or OFF to indicate the status of the backlight. The backlight setting can be selected by pressing the UP or DOWN arrow key. NOTE: It is recommended that the backlight be OFF when measurements are made in well-illuminated areas, as this will greatly increase battery life. 3. Metric/English Units Press the CAL key a second time. The T-Mike E will display the units in which it is set to measure thickness (IN or MM). Press the UP or DOWN arrow keys to select the desired unit of measurement. 4. Calibration Lock Press the CAL key a third time. The T-Mike E will display the status of the calibration lock function (CAL or LOC). When Cal is displayed the velocity calibration function can be performed as described in the 4 th step under section B, OPERATION. When LOC is displayed the calibration function is disabled. The velocity used in measurement cannot be changed without entering this SETUP procedure again to change settings to CAL. It is recommended that the lock feature be used when supervising personnel wish to prevent unauthorized change of the calibrated velocity. Pressing the Cal key a final time completes the SETUP procedure and returns the T-Mike E to the measurement mode. B. OPERATION Page 2 of 10

3 1. Lemo quick-disconnect transducer connectors. 2. Battery charger connector / data port. 3. Built-in calibration block. 4. Coupling indicator. 5. Backlit amber LCD indicates modes, measurements and icons. 6. Durable, lightweight, extruded aluminum case. 7. Environmentally sealed keypad. 8. Power key. 9. One-step material calibration key. 10. Arrow keys for calibration. 11. Operator selectable metric or English measurement mode. 1. ON/OFF Function (item 8 on Figure 2) Figure 2 T-Mike E - Front Panel Press this key to turn the T-Mike E on. The unit will display a reading of zero. The T-Mike E is now ready for measurement of calibration. In order to compensate for any change in the ultrasonic transducer, a PROBE ZERO function should be performed at this point. 2. Probe Zero Function The PROBE ZERO function compensates the T-Mike E for the fixed delay of the ultrasonic transducer. This function should be performed at the start of each day (to compensate for transducer wear) or whenever the transducer is changed. Apply a drop of couplant to the end of the transducer. Press the ON/OFF button on they keyboard to turn on the T-Mike E. Place the transducer in steady contact with the probe disk located on the top of the T-Mike E. DO NOT use another test piece since the velocity of the probe disk is carefully controlled and errors will result. The T-Mike E recognizes that the probe is in contact with the probe disk, and automatically performs the PROBE ZERO function. Upon completion of the function Prb 0 will appear on the display indicating that the appropriate transducer delay factor has been entered into memory. NOTE: During the PROBE function, the T-Mike E velocity is temporarily reset to the velocity of the probe disk ( in./msec.). Upon completion of the PROBE function the velocity will return to the last calibrated value. This allows the operator to continue measurements on the previously calibrated material. Page 3 of 10

4 3. Material Calibration to Known Thickness The following procedure is used to calibrate the T-Mike E sound velocity to a known sample of material. Some materials may exhibit sound velocity variations which can cause errors in measurement. Be sure to check a representative group of samples to determine if such variation is present. First follow the above PROBE ZERO procedure (step 2) to calibrate the delay of the ultrasonic transducer being used. Apply a drop of couplant to the representative piece of material of known thickness. This piece is called the sample. Place the probe in steady contact with the surface of the sample. The T-Mike E will display a thickness reading using the last calibrated velocity. Press the CAL key (item 9 in Figure 2) to lock in the reading. The transducer may now be removed from the sample. The display on the T-Mike E will momentarily display CAL. Small arrows will appear in the upper left portion of the display above the thickness readings. This indicates that the T-Mike E is in the calibration mode and enables the UP/DOWN arrow keys. U P Adjust the thickness shown on the T-Mike E display by using the UP and DOWN arrows (item 10 in Figure 2). Note that when these keys are held continuously, the speed at which the thickness changes will increase, making it easy to make the large adjustments for materials such as plastic and aluminum. When the correct thickness appears on the display, press the CAL key again to display the sound velocity of the calibrated material as calculated by the T-Mike E. The display of sound velocity will be indicated by the appearance of appropriate velocity units (IN/ms or M/S) above the displayed velocity value. Press CAL key one final time to accept this sound velocity and return to the measurement mode. Check the calibration by replacing the probe on the sample piece. The T-Mike E is in the CALIBRATION mode the unit will not make further measurement until the calibration procedure is completed. 4. Calibration from Known Velocity The following procedure is used to calibrate the T-Mike E to a given material velocity. Follow the above PROBE procedure to calibrate the delay of the ultrasonic transducer being used. Press the CAL key (item 9 in Figure 2). The arrow symbols will appear in the upper left portion of the display indicating the UP/DOWN arrow keys are active. Press the CAL key a second time. The display on the T-Mike E will display the current velocity and the appropriate units of measure. Page 4 of 10

5 IN /µ s Or if metric units are selected: m / s UP Use the UP and Down arrow keys to scroll to velocity to the known velocity of the material. Note that when these keys are held continuously, the speed that the velocity changes will increase, making it easy to make the large adjustments for materials such as plastic and aluminum. DOWN When the desired velocity appears on the display, press the CAL key again to enter the velocity for use in thickness measurements and to return to the measurement mode.. 5. Thickness Measurement Remove dirt, loose material, and couplant residue from the surface of material at points where measurements are desired. Verify that the PROBE and CAL functions have been performed. Place a drop of couplant on the material surface at the measurement point. Place the transducer in steady contact with the surface of the material at the measurement point. When a detectable echo is received from the back surface of the material, the echo indicator in the upper right of the display will indicate the detection of an echo by changing in appearance to represent contact between the pictured transducer and the same surface. NO ECHO ECHO DETECTED Page 5 of 10

6 The T-Mike E will display the thickness of the material and the appropriate units of measure (inches or millimeters). NOTE: If the detected echo is of marginal magnitude a question mark? Will appear to the left of the echo indicator. This will alert the operator that the measurement may be subject to error, and steps should be taken to improve the transducer coupling or position. ON OFF If no echo is detected for a period of 4 ½ minutes, the T-Mike E will automatically switch to its low power or dormant state as indicated by the display becoming blank. In this state probe and velocity calibration factors are retained in memory. Operations may be resumed immediately by pressing the power key. 6. Measurement Units The T-Mike E can display thickness in either English or Metric units. The unit system may be selected during the INITIAL SETUP of the T-Mike E as described previously. 7. Low Battery The T-Mike E automatically monitors the charge state of the battery to insure that sufficient charge is available to operate the unit properly. When the T-Mike E detects minimum allowable battery level the numbers on the display will blink on and off. This alerts the operator that the batteries should be replaced or recharged before further operation. When the battery level fails below the minimum allowable level the T-Mike E will turn off and not allow further operations until batteries are replaced or recharged. 8. Replacing or Recharging Batteries The T-Mike E comes standard with 4 AA alkaline batteries. The user may elect to use rechargeable NiCad AA batteries that can be inserted as a direct replacement of the alkaline cells. The NiCad batteries may be charged externally or in the T-Mike E as described below. To change the batteries, loosen the screw on the bottom end cap and remove the spent batteries. Replace batteries with fresh Alkaline AA cells or recharged NiCad AA cells. Care must be used to replace the batteries in the correct polarity as indicated on the battery holder. To recharge the NiCad batteries in the T-Mike E connect the optional battery charger to the combination DATA/CHARGER jack on the top end cap (item 2 of Figure 2) of the T-Mike E. The NiCad batteries will be fully charged in approximately 14 hours. NOTE: After changing batteries be sure to perform the probe zero procedure and recalibrate the T-Mike E for the material to be measured. Page 6 of 10

7 C. T-MIKE PROGRAMMABLE 1. Configuration and Calibration The T-Mike Programmable is preset to display in inches or millimeters; the back light enabled or disabled; and the sound velocity fixed. This can be done at the factory prior to shipping or this may be done by connecting the T-Mike programmable to an IBM compatible computer running special software, or from a T-Mike E. When using a T-Mike E, a special cable is required. 2. To Program From a T-Mike E Set the T-Mike E in the desired mode by holding down the CAL key while turning on the power. Now calibrate the sound velocity of the T-Mike E to the desired sound velocity. Turn off both gauges. Connect the special cable between the two data/charger ports. Turn on the T-Mike E; press the Cal key twice and the LCD will indicate the velocity. Now turn on the T-Mike Programmable. The T-Mike Programmable should display CAL, followed by the new velocity and then will go into the measurement mode. Be sure this new velocity matches with the sound velocity in the T-Mike E. 3. To Program from an IBM Compatible PC Remove rubber plug from the serial port of the T-Mike and insert the LEMO connector of the T-Mike serial cable (Part Number ). Connect the serial cable to COM1 or COM2 of the PC. Run SETMIKE from Start/Programs/Stresstel or from the appropriate installation directory. Select Measurement Unit: Inches or Millimeters. Enter sound velocity in inches/us or M/S. Press Enter to accept the new velocity. Check Backlight to illuminate T-Mike reading. Select Serial Port: COM1 or COM2. \\Stresstel SetMike T-Mike Settings Measurement Units Inches Millimeters Transmit Settings Serial Port COM1 COM2 X Velocity (M/S) Language English Use the Tab key to cycle through the SETMIKE Windows application controls. Backlight SEND CLOSE Press SEND to initiate update of the T-Mike. Turn on the T-Mike. If the T-Mike was already on, turn it off and turn on gain. \\Stresstel T-Mike Update X The T-Mike P/B may require several attempts to synchronize with SETMIKE. Repeat Off/ON cycles until the correct velocity appears in the display. Turn on instrument Now CLOSE Page 7 of 10

8 When you turn off the T-Mike Programmable, it will display the EPROM software version. When you turn on the T-Mike Programmable, it will momentarily display the calibrated sound velocity. Test the T- Mike Programmable on a material sample to verify that it reads the correct thickness. 4. Special Cables From a T-Mike E to T-Mike Programmable Part # From IBM Compatible Computer to T-Mike Programmable adaptor 9 pin to 25 pin Part # ; 9 pin Part # D. T-MIKE BASIC 1. Configuration and Calibration The T-Mike Basic has five operator selectable materials (steel, aluminum, stainless steel, cast iron, and plexiglass) with is preset sound velocities. One custom material ca be programmed by connecting the T- Mike B to an IBM compatible computer running special software, or from a T-Mike E. In either case, a special cable is required. 2. Material Selection First follow the above PROBE ZERO procedure to calibrate the delay of the ultrasonic transducer being used. MAT Press the MAT key to select the material being tested. Select the material by using the UP and DOWN arrows. Press the MAT key a second time to store the velocity for the selected material. Some materials may exhibit sound velocity variations which can cause errors in measurement. Be sure to check a representative group of samples to determine if such variation is present. 3. Initial Setup In order to change the settings of the T-Mike B, the instrument must be powered on in the Initial Setup mode. The settings, which may be changed, are the units of measure and the backlight. This procedure is defined in the INITIAL SETUP ( A page 1). Use the MAT key on the T-Mike B as CAL key referenced in the procedure. 4. To Program from a T-Mike E Set the T-Mike E in the desired mode by holding down the CAL key while turning on the power. Now calibrate the sound velocity of the T-Mike E to the desired sound velocity. Turn off both gauges. Connect the special cable between the two data/charger ports. Turn on the T-Mike E; press the CAL key twice and Page 8 of 10

9 the LCD will indicate the velocity. Now turn on the T-Mike B. The T-Mike B should display CAL, followed by the new velocity and then will go into the measurement mode. Be sure this new velocity matches with the sound velocity in the T-Mike E. 5. To Program from an IBM Compatible PC Remove rubber plug from the serial port of the T-Mike and insert the LEMO connector of the T-Mike serial cable (Part Number ). Connect the serial cable to COM1 or COM2 of the PC. Run SETMIKE from Start/Programs/Stresstel or from the appropriate installation directory. Select Measurement Unit: Inches or Millimeters. Enter sound velocity in inches/us or M/S. Press Enter to accept the new velocity. Check Backlight to illuminate T-Mike reading. Select Serial Port: COM1 or COM2. \\Stresstel SetMike T-Mike Settings Measurement Units Inches Millimeters Transmit Settings Serial Port COM1 COM2 X Velocity (M/S) Language English Use the Tab key to cycle through the SETMIKE Windows application controls. Backlight SEND CLOSE Press SEND to initiate update of the T-Mike. Turn on the T-Mike. If the T-Mike was already on, turn it off and turn on gain. \\Stresstel T-Mike Update X The T-Mike P/B may require several attempts to synchronize with SETMIKE. Repeat Off/ON cycles until the correct velocity appears in the display. Turn on instrument Now CLOSE When you turn off the T-Mike B, it will display the EPROM software version. When you turn on the T- Mike B, it will display the active material. Press MAT again to display the sound velocity. Test the T- Mike B on a material sample to verify that it reads the correct thickness. E. HELPFUL HINTS on OPERATION 1. Backlight The T-Mike E features a backlight LCD display. The display illuminates when tan echo is detected. Using the backlight will significantly reduce battery life. Therefore, you may wish to deactivate the backlight when using the T-Mike E in a well-lit environment. 2. Clean the Surface Clean the probe and the material to be material to be measured and apply only enough couplant to obtain a steady, valid reading. Usually one drop is enough. Page 9 of 10

10 3. Material Consistency Maximum measurement accuracy will be contained if the work piece and the calibration piece are the same temperature, shape, material, and size. NOTICE Since velocity of sound varies from test piece to work piece, accuracy of measurement depends on consistency of ultrasonic velocity. Ultrasonic velocity also varies with internal stress, so that heat treatment of the material will also affect accuracy. The following table expresses accuracy variations for some common materials. NOTE: These specifications are intended only as a general guide. Actual accuracy may be different. Aluminum +/- 2% Cast iron +/- 8% Steel +/- 0.5% Nylon +/- 10% Page 10 of 10

Quick Guide v. 3.0 Advanced model

Quick Guide v. 3.0 Advanced model Ultrasonic Thickness Gage Quick Guide v. 3.0 Advanced model Introduction The PosiTector UTG is a hand-held Ultrasonic Thickness Gage that uses the non-destructive ultrasonic pulse-echo principle to measure

More information

OPERATION MANUAL DAKOTA ULTRASONICS. Material & Coating Thickness Gauge

OPERATION MANUAL DAKOTA ULTRASONICS. Material & Coating Thickness Gauge OPERATION MANUAL DAKOTA ULTRASONICS CMX Material & Coating Thickness Gauge P/N P-170-0002 Rev 1.30, April 2008 CHAPTER ONE INTRODUCTION...1 CHAPTER TWO QUICK STARTUP GUIDE...2 CHAPTER THREE KEYBOARD,

More information

OPERATION MANUAL DAKOTA ULTRASONICS. Material & Coating Thickness Gauge

OPERATION MANUAL DAKOTA ULTRASONICS. Material & Coating Thickness Gauge OPERATION MANUAL DAKOTA ULTRASONICS CMX DL Material & Coating Thickness Gauge P/N P-171-0002 Rev 1.30, April 2008 CHAPTER ONE INTRODUCTION...1 CHAPTER TWO QUICK STARTUP GUIDE...2 CHAPTER THREE KEYBOARD,

More information

OPERATION MANUAL DAKOTA ULTRASONICS. Ultrasonic Thickness Gauge

OPERATION MANUAL DAKOTA ULTRASONICS. Ultrasonic Thickness Gauge OPERATION MANUAL DAKOTA ULTRASONICS ZX-5 Ultrasonic Thickness Gauge P/N P-303-0002 Rev 1.00, September 2018 CONTENTS CHAPTER ONE INTRODUCTION... 1 1.1 DISCLAIMER... 1 CHAPTER TWO KEYPAD, MENU, DISPLAY

More information

EHC-03 Ultrasonic Thickness Gage

EHC-03 Ultrasonic Thickness Gage QUICK START GUIDE EHC-03 Ultrasonic Thickness Gage Software version x.x Liability Ultrasonic testing is a function of using the proper equipment (electronics, transducer, cable and couplant combination)

More information

CL 5. Operating Manual. Ident-Nr Issue 01

CL 5. Operating Manual. Ident-Nr Issue 01 CL 5 Operating Manual Ident-Nr. 021-002-296 Issue 01 Issue 01, 04/2005 applies to software version XX.04.XX Contents 1 General Information... 1-1 1.1 Supplying Power to the CL 5... 1-2 1.2 Powering On

More information

OPERATION MANUAL DAKOTA ULTRASONICS. Data Logging Thickness Gauge

OPERATION MANUAL DAKOTA ULTRASONICS. Data Logging Thickness Gauge OPERATION MANUAL DAKOTA ULTRASONICS MMX-7 Data Logging Thickness Gauge P/N P-160-0002 Rev 1.10, January 2008 CHAPTER ONE INTRODUCTION...1 CHAPTER TWO QUICK STARTUP GUIDE...2 CHAPTER THREE KEYBOARD, MENU,

More information

Instruction Manual Ultrasonic Thickness gauge DC3020

Instruction Manual Ultrasonic Thickness gauge DC3020 Instruction Manual Ultrasonic Thickness gauge DC3020 V 7.11 CONTENTS 1.General Description... - 1-2.Technical Specifications... - 1-3. Standard Delivery... - 2-4. Overview the Display Unit... - 2-5. Keypad

More information

OPERATION MANUAL DAKOTA ULTRASONICS. Ultrasonic Multi-Echo Data Logging Thickness Gauge

OPERATION MANUAL DAKOTA ULTRASONICS. Ultrasonic Multi-Echo Data Logging Thickness Gauge OPERATION MANUAL DAKOTA ULTRASONICS ZX-6 DL Ultrasonic Multi-Echo Data Logging Thickness Gauge P/N P-306-0002 Rev 1.00, September 2018 CONTENTS CHAPTER ONE INTRODUCTION... 1 1.1 DISCLAIMER... 1 CHAPTER

More information

NOTES TABLE OF CONTENTS Introduction Quick Start Guide Keyboard, Menu and Connector Reference...12

NOTES TABLE OF CONTENTS Introduction Quick Start Guide Keyboard, Menu and Connector Reference...12 NOTES TABLE OF CONTENTS 01.0 Introduction......................................................02 02.0 Quick Start Guide...................................................3 03.0 Keyboard, Menu and Connector

More information

Advanced Test Equipment Rentals ATEC (2832) PANAMETRICS. 25DL Multi-Mode Ultrasonic Thickness Gage with Internal Datalogger

Advanced Test Equipment Rentals ATEC (2832) PANAMETRICS. 25DL Multi-Mode Ultrasonic Thickness Gage with Internal Datalogger Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) PANAMETRICS 25DL Multi-Mode Ultrasonic Thickness Gage with Internal Datalogger 25DL Easy to use Uses contact, delay

More information

P/N: QSD 301 ENG Rev

P/N: QSD 301 ENG Rev I P/N: QSD 301 ENG Rev 3 2012 Chapter 1 Chapter 2 Chapter 3 First Steps 1.1 Know the QS5 1 1.1.1 Front panel 1 1.1.2 Connectors 2 1.2 Install or replace batteries 3 1.3 Connecting the transducer 5 1.4

More information

OPERATION MANUAL DAKOTA ULTRASONICS. Material & Coating Thickness Gauge

OPERATION MANUAL DAKOTA ULTRASONICS. Material & Coating Thickness Gauge OPERATION MANUAL DAKOTA ULTRASONICS CMX DL+ Material & Coating Thickness Gauge P/N P-172-0002 Rev 2.06, July 2010 CHAPTER ONE INTRODUCTION...1 CHAPTER TWO QUICK STARTUP GUIDE...2 CHAPTER THREE KEYBOARD,

More information

ISU-300D ULTRASONIC THICKNESS GAUGE OPERATION MANUAL

ISU-300D ULTRASONIC THICKNESS GAUGE OPERATION MANUAL MN-ISU-00D-E www.insize.com ISU-00D ULTRASONIC THICKNESS GAUGE OPERATION MANUAL Attention Specification Of Transducers In order to obtain good measurement precision, you need to clear away the rusty, sundry,

More information

QSV User Manual. Ultrasonic Velocity Gauge. P/N: QSD 101 ENG Rev

QSV User Manual. Ultrasonic Velocity Gauge. P/N: QSD 101 ENG Rev Ultrasonic Velocity Gauge P/N: QSD 101 ENG Rev 2 2012 Index Chapter 1 Chapter 2 Chapter 3 QSV User Manual First Steps 1.1 Know the QSV 1 1.1.1 Front panel 1 1.1.2 Connectors 2 1.2 Install or replace batteries

More information

27MG Ultrasonic Thickness Gage Getting Started Guide

27MG Ultrasonic Thickness Gage Getting Started Guide 27MG Ultrasonic Thickness Gage Getting Started Guide Intended Use The 27MG is designed to measure thicknesses of industrial and commercial materials. Do not use the 27MG for any purpose other than its

More information

ISU-250C ULTRASONIC THICKNESS GAUGE OPERATION MANUAL

ISU-250C ULTRASONIC THICKNESS GAUGE OPERATION MANUAL MN-ISU-0C-E www.insize.com ISU-0C ULTRASONIC THICKNESS GAUGE OPERATION MANUAL Attention Technical Specification In order to obtain good measurement precision, you need to clear away the rusty, sundry,

More information

Platinum RTD Thermometer MODEL NO

Platinum RTD Thermometer MODEL NO Platinum RTD Thermometer MODEL NO. 93410-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com

More information

Quick Guide v. 1.1 DeFelsko Corporation USA 2012

Quick Guide v. 1.1 DeFelsko Corporation USA 2012 Quick Guide v. 1.1 Introduction The PosiTector Surface Profile Gage (SPG) is a hand-held electronic instrument that measures the peak-to-valley height of the surface profile of abrasive blast cleaned surfaces.

More information

OPERATION MANUAL DAKOTA ULTRASONICS. Underwater Material & Coating Thickness Gauge

OPERATION MANUAL DAKOTA ULTRASONICS. Underwater Material & Coating Thickness Gauge OPERATION MANUAL DAKOTA ULTRASONICS UMX-2 Underwater Material & Coating Thickness Gauge P/N P-185-0004 Rev 1.0, May 2010 UMX-2 Underwater Material & Coating Thickness Gauge CHAPTER ONE INTRODUCTION -

More information

Portable Hardness Tester DHT-100

Portable Hardness Tester DHT-100 Portable Hardness Tester DHT-100 Large LCD display with back light Impact devices D DC DL C D+15 G and E are available for special applications RS 232 output Automatic conversions to: Brinell HB, Rockwell

More information

SigmaCheck. Fully Featured Eddy Current Conductivity Meter ET NDE

SigmaCheck. Fully Featured Eddy Current Conductivity Meter ET NDE Fully Featured Eddy Current Conductivity Meter ET NDE Introducing The SigmaCheck APPLICATIONS Material Verification / Metal Sorting. Heat Treatment Verification. Heat or Fire Damage Investigation. Non-conductive

More information

SIGMATEST Operating Instructions

SIGMATEST Operating Instructions SIGMATEST 2.069 Software Version 3.12 Operating Instructions Registered trade-mark Contents 1. Introduction...4 2. Applications and Features...5 3. Getting Started...6 4. Measuring System Specifications...8

More information

160 UTG-2841 UTG ISU 200C ISU 250C ISU-200D -P

160 UTG-2841 UTG ISU 200C ISU 250C ISU-200D -P + Quotation 92 Ultrasonic Thickness Gauge / D-meter Make Dipel Time Model B 2 TT100 Range mm 1. 5 to 99 0. 75-300 Least Count 0. 1~ 1.5 to 60 / 0. 2~ 60 to 99 mm 0. 1 mm With Probe - 1 Probe Rs. 14500

More information

ph/orp Meter SD Card real-time Datalogger

ph/orp Meter SD Card real-time Datalogger User's Guide ph/orp Meter SD Card real-time Datalogger Model SDL100 Introduction Congratulations on your purchase of the Extech SDL100 ph/orp Meter, an SD Logger Series meter. This meter displays and stores

More information

MONARCH INSTRUMENT. TACH-4A / TACH-4AR Programmable Optical Tachometer. Instruction Manual Rev 2.0

MONARCH INSTRUMENT. TACH-4A / TACH-4AR Programmable Optical Tachometer. Instruction Manual Rev 2.0 MONARCH INSTRUMENT Instruction Manual TACH-4A / TACH-4AR Programmable Optical Tachometer 1071-4850-001 Rev 2.0 15 Columbia Drive Amherst, NH 03031 E-Mail: Support@MonarchInstrument.com Phone: (603) 883-3390

More information

OPERATION MANUAL DAKOTA ULTRASONICS. Material & Coating Thickness Gauge. (Manual 1 of 2)

OPERATION MANUAL DAKOTA ULTRASONICS. Material & Coating Thickness Gauge. (Manual 1 of 2) OPERATION MANUAL DAKOTA ULTRASONICS DFX-7 Material & Coating Thickness Gauge (Manual 1 of 2) P/N P-220-0004 Rev 1.1, April 2014 CHAPTER ONE INTRODUCTION... 1 CHAPTER TWO QUICK STARTUP GUIDE... 2 CHAPTER

More information

Ultrasonic Thickness Gauge. Operation Manual

Ultrasonic Thickness Gauge. Operation Manual Ultrasonic Thickness Gauge Operation Manual SA40 Contents 1. Overview 4 2. Keypad 4 3. LCD display 4 4. Calibration 5 5. Change velocity 5 6. Measuring velocity 5 7. Data memory and read 6 8. Optional

More information

160 UTG-2841 UTG ISU 200C ISU 250C ISU-200D -P

160 UTG-2841 UTG ISU 200C ISU 250C ISU-200D -P + Quotation 92 Ultrasonic Thickness Gauge / D-meter Make Dipel Time Model B 2 TT100 Range mm 1. 5 to 99 0. 75-300 Least Count 0. 1~ 1.5 to 60 / 0. 2~ 60 to 99 mm 0. 1 mm With Probe - 1 Probe Rs. 14500

More information

FTE-6000 Hand Held Tuneable Laser Source

FTE-6000 Hand Held Tuneable Laser Source FTE-6000 Hand Held Tuneable Laser Source USER S MANUAL Avoid Exposure Laser Radiation Emitted From The Aperture This product conforms to CDRH standards for laser product Per 21 CFR 1040.10 & 1040.11 Table

More information

Calibration Procedure. DeFelsko Corporation. PosiTector UTG. Ultrasonic Thickness Gage. Table of Contents

Calibration Procedure. DeFelsko Corporation. PosiTector UTG. Ultrasonic Thickness Gage. Table of Contents Management Procedure 2585 Revision: C Date Issued: Feb 21, 2003 Date Revised: November 15, 2005 Calibration Procedure DeFelsko Corporation PosiTector UTG Ultrasonic Thickness Gage Table of Contents 1 Introduction

More information

1. Place your CD driver in your PC drive for software installation 2. Follow the instruction on the screen to finish the software installation

1. Place your CD driver in your PC drive for software installation 2. Follow the instruction on the screen to finish the software installation ENGLISH Hardware Installation 1. Install two rechargeable AA batteries into the mouse and two AAA batteries into the keyboard. 2. Plug in your USB receiver to a USB port of your PC. 3. Turn on your PC.

More information

BlueNote USER S GUIDE

BlueNote USER S GUIDE BlueNote USER S GUIDE SAFETY PRECAUTIONS SAFETY PRECAUTIONS To prevent injury and/or damage, take the following precautions: 1 Make sure you keep the antenna away from the mains or high voltage electricity

More information

Press the PEN HOME key to move the pen(s) to the outside of the chart. The pen(s) is automatically raised off the chart.

Press the PEN HOME key to move the pen(s) to the outside of the chart. The pen(s) is automatically raised off the chart. DICKSON GETTING STARTED 7-day chart rotation 0 to 100 o F KT8P0/2/3 & KT856 Remote Sensing Temperature Recorder QUICK START 1. Remove the protective pen cap(s). 2. Connect Probe(s): a. KT8P0/2/3: Make

More information

ProScale. Model 18W. Limited Edition. Model 18W

ProScale. Model 18W. Limited Edition. Model 18W ProScale Model 18W Limited Edition Model 18W WARRANTY Accurate Technology, Inc. warrants the ProScale Model 18W against defective parts and workmanship for 1 year commencing from the date of original purchase.

More information

User Manual. Heavy Duty Differential Pressure Manometer. Model SDL720. Additional User Manual Translations available at

User Manual. Heavy Duty Differential Pressure Manometer. Model SDL720. Additional User Manual Translations available at User Manual Heavy Duty Differential Pressure Manometer Model SDL720 Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the Extech SDL720 Pressure

More information

PV Remote Unit A 1378 Instruction manual Version 2.1, Code no

PV Remote Unit A 1378 Instruction manual Version 2.1, Code no PV Remote Unit A 1378 Instruction manual Version 2.1, Code no. 20 752 010 Distributor: Manufacturer: METREL d.d. Ljubljanska cesta 77 1354 Horjul Slovenia web site: http://www.metrel.si e-mail: metrel@metrel.si

More information

12-Channel Thermocouple Datalogger

12-Channel Thermocouple Datalogger User Manual 12-Channel Thermocouple Datalogger Model TM500 Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the Extech TM500 Thermometer,

More information

Platinum RTD Thermometer MODEL NO

Platinum RTD Thermometer MODEL NO Platinum RTD Thermometer MODEL NO. 93410-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com

More information

Sonic Meter SM1000. Operating Manual. 2-00, 1-11, Rev 1.1

Sonic Meter SM1000. Operating Manual. 2-00, 1-11, Rev 1.1 Sonic Meter SM1000 Operating Manual 2-00, 1-11, Rev 1.1 SyncroCraft, LLC 16225 CAMINO DEL Sol LOS GATOS, CA 95032 TELEPHONE 408-438-0597 syncrocraft@comcast.net Important: Do not attempt to operate this

More information

Quotation Ultrasonic Thickness Gauge - With 10% Discount Model Rs Page No

Quotation Ultrasonic Thickness Gauge - With 10% Discount Model Rs Page No Quotation 110 Ultrasonic Thickness Gauge / D-meter Make Dipel Time Time Model B 2 TT100 Extra Probe for TT 100 Range mm 1. 5 to 99 0. 75-300 - Least Count 0. 1~ 1.5 to 60 / 0. 2~ 60 to 99 mm 0. 1 mm -

More information

4-Channel Vibration Datalogger

4-Channel Vibration Datalogger User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com 4-Channel Vibration Datalogger Model VB500 Introduction Congratulations

More information

Introduction. Power-up / Power-down

Introduction. Power-up / Power-down Introduction The PowderSave XP is a hand-held, non-contact Coating Thickness Gauge that uses an airborne high frequency ultrasound to analyze coating powder applied to rigid substrates to calculate and

More information

KUHLMAN INSTRUMENT COMPANY 5000K WARRANTY

KUHLMAN INSTRUMENT COMPANY 5000K WARRANTY KUHLMAN INSTRUMENT COMPANY 5000K WARRANTY The 5000K is warranted for one year against defects in material or workmanship. This warranty does not cover water or physical damage, damage from over pressure

More information

Quotation. 110 Ultrasonic Thickness Gauge / D-meter Ultrasonic Thickness Gauge Transducers Insize Make

Quotation. 110 Ultrasonic Thickness Gauge / D-meter Ultrasonic Thickness Gauge Transducers Insize Make 110 Ultrasonic Thickness Gauge / D-meter Quotation Make Dipel Time Time Model B 2 TT100 Extra Probe for TT 100 Range mm 1. 5 to 99 0. 75-300 - Least Count 0. 1~ 1.5 to 60 / 0. 2~ 60 to 99 mm 0. 1 mm -

More information

Light Meter SD Card real-time datalogger

Light Meter SD Card real-time datalogger USER GUIDE Light Meter SD Card real-time datalogger Model SDL400 Introduction Congratulations on your purchase of the Extech SDL400 Light Meter. This meter displays and stores light meter readings in three

More information

---, PANAMETRICS. Advanced Test Equipment Rentals ATEC (2832) ULTRASONIC THICKNESS GAGES

---, PANAMETRICS. Advanced Test Equipment Rentals ATEC (2832) ULTRASONIC THICKNESS GAGES Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) INSTRUCTION MANUAL j ULTRASONIC THICKNESS GAGES MODELS 22DL, 22DLHR, &22DLHP Software V 1.06 Manual No. 910-02OG 7/13/94

More information

Hygro-Thermometer / Datalogger

Hygro-Thermometer / Datalogger USER GUIDE Hygro-Thermometer / Datalogger Model SDL500 Introduction Congratulations on your purchase of the Extech SDL500 Hygro-Thermometer, an SD Logger Series meter. This meter displays and stores Temperature

More information

Digital Measurement Metrology, Inc PRECISION IS OUR VISION

Digital Measurement Metrology, Inc PRECISION IS OUR VISION Contents Powering the Gage 2 Using the Gage 3 Optional Settings 8 Dimensions 10 BFG Specification Table 11 Introduction Thank you for choosing the Mecmesin Corporation Basic Force Gage (BFG) instrument.

More information

Moisture Content Meter Plus Dew Point, Wet Bulb, and Thermocouple Temperature functions

Moisture Content Meter Plus Dew Point, Wet Bulb, and Thermocouple Temperature functions User's Guide Moisture Content Meter Plus Dew Point, Wet Bulb, and Thermocouple Temperature functions SD Card real-time datalogger Model SDL550 %RH Moisture Content Meter SDL550 Introduction Congratulations

More information

THE TECHNOLOGY BEHIND THE MEASUREMENT

THE TECHNOLOGY BEHIND THE MEASUREMENT THE TECHNOLOGY BEHIND THE MEASUREMENT The USCX series non-invasive flowmeters work on the transit time ultrasonic principle. This involves sending and receiving ultrasonic pulses from a pair of sensors

More information

IDS-801 Series Compact Bench Counting Scale. User Manual

IDS-801 Series Compact Bench Counting Scale. User Manual IDS-801 Series Compact Bench Counting Scale User Manual 1. INTRODUCTION... 3 1.1. Product Description... 3 1.2. General Features... 3 1.3. Safety Precautions... 3 2. INSTALLATION... 4 2.1. Unpacking...

More information

Dual Type Coating Thickness Tester LZ-990. Operating Manual

Dual Type Coating Thickness Tester LZ-990. Operating Manual Dual Type Coating Thickness Tester LZ-990 Operating Manual Coating Thickness Tester Safety Precautions The Coating Thickness Tester may cause damage to objects being handled if the precautions are not

More information

DFS Series User s Manual (5N-1000N Capacity)

DFS Series User s Manual (5N-1000N Capacity) DFS Series User s Manual (5N-1000N Capacity) REV 3.8 Table of Contents Table of Contents 2 Introduction 3 Before Use 3 Operation Overview 4 Powering the DFS 5 Using the DFS. 6 Fitting Accessories 6 Mounting

More information

HI HI N HI HI

HI HI N HI HI Instruction Manual HI 935005 - HI 935005N HI 935002 - HI 935009 Portable K-Thermocouple Thermometers www.hannainst.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you

More information

Thickness Gauge Transducers & Accessories - Price List Date: 01/01/18

Thickness Gauge Transducers & Accessories - Price List Date: 01/01/18 Thickness Gauge Transducers & Accessories - Price List Date: 01/01/18 Dual Transducer Reference Chart Frequency & Disks Note: The frequency and crystal diameter are stamped and sorted according to color,

More information

PosiTector PC. Powder Checker Coating Thickness Gage. INSTRUCTION MANUAL v. 1.0

PosiTector PC. Powder Checker Coating Thickness Gage. INSTRUCTION MANUAL v. 1.0 PosiTector PC Powder Checker Coating Thickness Gage INSTRUCTION MANUAL v. 1.0 Introduction The PosiTector PC Powder Checker is a hand-held, non-contact Coating Thickness Gage that uses an airborne high

More information

Ultrasonic Load Monitoring

Ultrasonic Load Monitoring Ultrasonic Load Monitoring Because ensuring a leak-free solution is critical. It has to be Hydratight. For more than 100 years, Hydratight has provided world-class bolted joint solutions, assessing and

More information

CF3000 Dealer Diagnostic Tool Instruction Manual

CF3000 Dealer Diagnostic Tool Instruction Manual CF3000 Dealer Diagnostic Tool Instruction Manual Table of Contents: About the CF3000......3 Important Precautions......4 Components....5 Charging the CF3000......7 Licensing the CF3000.......8 Updating

More information

DMTFP Portable Ultrasonic Flow Meters

DMTFP Portable Ultrasonic Flow Meters DMTFP Portable Ultrasonic Flow Meters Series DMTFP Portable Transit Time Ultrasonic Flow Meter is a state-of-the-art universal transit-time flow meter using MultiPulse TM technology and low-voltage broadband

More information

1. Introduction Specifications Operation Guide Trouble-shooting Note... 13

1. Introduction Specifications Operation Guide Trouble-shooting Note... 13 Content 1. Introduction... 1 Notice... 1 Safety Guide... 1 2. Specifications... 2 Features... 2 Specifications... 3 Capacity and Resolution... 4 Schematic Diagram... 4 Dimension... 5 3. Operation Guide...

More information

Fully Featured Eddy Current Conductivity Meter ET NDE

Fully Featured Eddy Current Conductivity Meter ET NDE Fully Featured Eddy Current Conductivity Meter ET NDE Introducing The SigmaCheck APPLICATIONS Material Verification / Metal Sorting. Heat Treatment Verification. Heat or Fire Damage Investigation. Non-conductive

More information

Vibration Meter SD Card real-time datalogger

Vibration Meter SD Card real-time datalogger User's Guide Vibration Meter SD Card real-time datalogger Model SDL800 Introduction Congratulations on your purchase of the Extech SDL800 Vibration Meter. This meter displays and stores vibration readings

More information

EB300 Weighing Scale Operation Manual

EB300 Weighing Scale Operation Manual EB300 Weighing Scale Operation Manual EB300 User Manual Page 1 of 18 TABLE OF CONTENTS 1. Safety Warnings... 3 2. Features..... 3 3. Options.......... 4 4. Product packaging.. 4 5. Display and keyboard.....

More information

200J Series Coating Thickness Testers LE-200J/LH-200J/LZ-200J. Operating Manual

200J Series Coating Thickness Testers LE-200J/LH-200J/LZ-200J. Operating Manual 200J Series Coating Thickness Testers LE-200J/LH-200J/LZ-200J Operating Manual Coating Thickness Tester Safety Precautions Be sure to carefully follow all safety precautions. Carefully read operating manual.

More information

Hubbell Power Systems

Hubbell Power Systems Hubbell Power Systems Polymer Insulator Tester Operating Instructions Manual PSC403-3679 Rev.3 Hubbell has a policy of continuous product improvement. Please visit hubbellpowersystems.com to confirm current

More information

Integrating Sound Level Datalogger Model

Integrating Sound Level Datalogger Model User's Guide Integrating Sound Level Datalogger Model 407780 Introduction Congratulations on your purchase of the Extech 407780 Integrating Sound Level Meter. The 407780 with programmable integrating time

More information

GI-4000 User Manual. Kobetron, Inc. P.O. Box 5489 Navarre, FL 32566

GI-4000 User Manual. Kobetron, Inc. P.O. Box 5489 Navarre, FL 32566 GI-4000 User Manual Kobetron, Inc. P.O. Box 5489 Navarre, FL 32566 TEL: (850) 939-5222 FAX: (850) 939-0490 EMAIL: sales@kobetron.com WEB: www.kobetron.com "Testing Tomorrow's Technology Today" TABLE OF

More information

OPERATING MANUAL H A W K E Y E Portable Digital Slit Lamp 1 INDEX Turning HAWK EYE ON and OFF... p. 3 Automatic shut down... p. 3 Adjusting the illumination intensity... p. 3 Taking pictures... p. 4 Playing

More information

USER INSTRUCTIONS - TP7C & TP9

USER INSTRUCTIONS - TP7C & TP9 USER INSTRUCTIONS - TP7C & TP9 INTRODUCTION This manual describes the basic function, use and safety instructions for a model TP7C or TP9 portable digital thermometer instrument. These ThermoProbe instruments

More information

Cygnus 3 Data Logger. Operating & Accessories Manual. Doc No. M4-CYG3-M-ENG_Iss2.doc 16 April 2013 (Mk4 Cygnus 3 Gauges)

Cygnus 3 Data Logger. Operating & Accessories Manual. Doc No. M4-CYG3-M-ENG_Iss2.doc 16 April 2013 (Mk4 Cygnus 3 Gauges) Cygnus 3 Data Logger Operating & Accessories Manual Doc No. M4-CYG3-M-ENG_Iss2.doc 16 April 2013 (Mk4 Cygnus 3 Gauges) Cygnus 3 Operating Manual M4-CYG3-M-ENG_Iss2.doc QUALITY POLICY STATEMENT Cygnus Instruments

More information

ATG II ULTRASONIC THICKNESS GAUGE INTRODUCTION MANUAL

ATG II ULTRASONIC THICKNESS GAUGE INTRODUCTION MANUAL ATG II ULTRASONIC THICKNESS GAUGE INTRODUCTION MANUAL 1. GENERAL INTRODUCTION ATGIIA-scan ultrasonic thickness gauge, Launch the ultrasonic through one side of the object, measure the digitized thickness

More information

Paint Test Equipment. Eban 4000 Coating Thickness Meter Data Sheet

Paint Test Equipment. Eban 4000 Coating Thickness Meter Data Sheet Paint Test Equipment Eban 4000 Coating Thickness Meter Data Sheet Coating Thickness Gloss Porosity Adhesion Surface Roughness Surface Cleanliness Climatic Conditions Electrostatic Inspectors Accessories

More information

373 Series Coating Thickness Testers LE-373/LH-373/LZ-373. Operating Manual

373 Series Coating Thickness Testers LE-373/LH-373/LZ-373. Operating Manual 373 Series Coating Thickness Testers LE-373/LH-373/LZ-373 Operating Manual Coating Thickness Tester Safety Precautions Be sure to carefully follow all safety precautions. Carefully read operating manual.

More information

Elcometer DL. Ultrasonic Thickness Gauge. Operating Instructions

Elcometer DL. Ultrasonic Thickness Gauge. Operating Instructions Elcometer 207 207DL English Ultrasonic Thickness Gauge Operating Instructions English The Elcometer 207 and 207DL Ultrasonic Thickness Gauges have been tested in accordance with EU regulations governing

More information

Sturtevant Richmont. Operating Instructions 3000-Series Exacta 2 Digital Torque Wrenches (All Models)

Sturtevant Richmont. Operating Instructions 3000-Series Exacta 2 Digital Torque Wrenches (All Models) Sturtevant Richmont Global Reach. Local Support. Division of Ryeson Corporation 3203 N. Wolf Road Franklin Park, IL 60131 Phone: 847/455-8677 800/877-1347 Fax: 847/455-0347 Website: www.srtorque.com E-mail:

More information

HAND-HELD THERMOCOUPLE THERMOMETER SERVICE MANUAL. CATALOG NUMBERS , and , and PROPRIETARY

HAND-HELD THERMOCOUPLE THERMOMETER SERVICE MANUAL. CATALOG NUMBERS , and , and PROPRIETARY HAND-HELD THERMOCOUPLE THERMOMETER SERVICE MANUAL CATALOG NUMBERS 600-1000, 600-1010 and 600-1020 91100-00, 91100-10 and 91100-20 PROPRIETARY Information contained in this manual is proprietary to COLE-PARMER

More information

TransPort Model PT878

TransPort Model PT878 GE Sensing & Inspection Technologies Flow TransPort Model PT878 Portable Liquid Flowmeter User s Manual 910-219 Rev. F November 2009 GE Sensing & Inspection Technologies TransPort Model PT878 Portable

More information

Instruction Manual TU_US

Instruction Manual TU_US DIGITAL ULTRASONIC THICKNESS GAUGE 6. Servicing 7. Transport and storage Appendix A Sound velocities Appendix B Application notes 8. Declaration of conformity 1. Overview Model: TU-US Table of contents

More information

TC41-Thermocouple Thermometers User Manual

TC41-Thermocouple Thermometers User Manual TC41-Thermocouple Thermometers User Manual Product Registration cs@perfect-prime.com Product Introduction This thermometer can be used to measure up to 4 different temperatures simultaneously, and with

More information

INTRODUCTION CT87E FEATURES AND CONTROLS

INTRODUCTION CT87E FEATURES AND CONTROLS INTRODUCTION The CT87E is a precision instrument used to monitor and record the presence or absence of voltage, light, or sound level such as what would be produced by an operating electric motor or compressor.

More information

MODEL 3150 SUSPENDED SOLIDS ANALYZER MODEL 35/35L OPTICAL SUSPENDED SOLIDS SENSOR

MODEL 3150 SUSPENDED SOLIDS ANALYZER MODEL 35/35L OPTICAL SUSPENDED SOLIDS SENSOR MODEL 3150 SUSPENDED SOLIDS ANALYZER MODEL 35/35L OPTICAL SUSPENDED SOLIDS SENSOR REVISION 22 August 2016 Insite Instrumentation Group, Inc 80 Whisperwood Blvd. Suite 107 Slidell, LA 70458 Phone: (985)

More information

Instruction Manual Temp JKT Temperature Meter Series

Instruction Manual Temp JKT Temperature Meter Series Instruction Manual Temp JKT Temperature Meter Series 68X243615 Technology Made Easy... Rev. 2 11/03 Preface This manual serves to explain the use of the Temp JKT temperature meter. This manual functions

More information

350 SMART MANOMETER OPERATING INSTRUCTIONS

350 SMART MANOMETER OPERATING INSTRUCTIONS 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com 350 SMART MANOMETER OPERATING INSTRUCTIONS Meriam Instrument s 350 Smart Manometer

More information

PLATINUM PRO6000 PROFESSIONAL SERIES DIGITAL COUNTERTOP / PORTABLE BALANCE

PLATINUM PRO6000 PROFESSIONAL SERIES DIGITAL COUNTERTOP / PORTABLE BALANCE PLATINUM PRO6000 PROFESSIONAL SERIES DIGITAL COUNTERTOP / PORTABLE BALANCE ongratulations on your purchase of the CProfessional Series model Platinum PRO6000 precision countertop/portable balance by GemOro

More information

One (1) acoustic and three (3) seismic channels. Three (3) component velocity output. All units are interchangeable with accuracy within two percent.

One (1) acoustic and three (3) seismic channels. Three (3) component velocity output. All units are interchangeable with accuracy within two percent. Chapter 2 Specifications Data Recorded Frequency Response Transducer Microphone Memory Clock Timer Mode Display Keyboard Power On Log Printer Chart Width Chart Length Battery Operating Time External Battery

More information

HOBO Waterproof Shuttle (U-DTW-1) Manual

HOBO Waterproof Shuttle (U-DTW-1) Manual HOBO Waterproof Shuttle (U-DTW-1) Manual The HOBO Waterproof Shuttle performs several major functions: Reads out all logger information (serial number, deployment number, data, etc.) from loggers in the

More information

Switch on / off. For switching off please press button for approx. 5 sec.

Switch on / off. For switching off please press button for approx. 5 sec. DUC-MP Quickstart A User Interface: Switch on / off. For switching off please press button for approx. 5 sec. Activates the background lighting To navigate through the different menus please use the keys

More information

Model ST-FT1 DIGITAL FORCE / TORQUE INDICATOR. User s Guide

Model ST-FT1 DIGITAL FORCE / TORQUE INDICATOR. User s Guide Model ST-FT1 DIGITAL FORCE / TORQUE INDICATOR Thank you Thank you for purchasing a Mesa Labs Model ST-FT1 digital force / torque indicator, designed for use with a remote torque sensor. With proper usage,

More information

Portable Conductivity Meter M1410. User Manual 1.002

Portable Conductivity Meter M1410. User Manual 1.002 Portable Conductivity Meter M1410 User Manual 1.002 General information This manual describes the operation of the Portable Conductivity Meter M1410. Please read this entire manual before operating this

More information

SL 200 INTEGRATING-AVERAGING SOUND LEVEL METER

SL 200 INTEGRATING-AVERAGING SOUND LEVEL METER SL 200 INTEGRATING-AVERAGING SOUND LEVEL METER 1. Introduction SL-200 is a high performance, integrating-averaging sound level meter with an easy-to-use interface for quick and simple measurements. This

More information

SP-400 Portable Water Analyzer Operation Manual. Rev. A Firmware version 1.02

SP-400 Portable Water Analyzer Operation Manual. Rev. A Firmware version 1.02 SP-400 Portable Water Analyzer Operation Manual Rev. A Firmware version 1.02 Pyxis Lab, Inc 120 Jeffrey Ave Holliston, MA 01746 USA www.pyxis-lab.com service@pyxis-lab.com 1 Contents 1 General Description...

More information

HI HI N HI HI

HI HI N HI HI Instruction Manual HI 935005 - HI 935005N HI 935002 - HI 935009 Portable K-Thermocouple Thermometers w w w. h a n n a i n s t. c o m Dear Customer, Thank you for choosing a Hanna product. Please read this

More information

MIYAWAKI ULTRASONIC TESTER PM11 OPERATOR S MANUAL. Be sure to read this manual for safe and proper operation. Store this manual carefully after use.

MIYAWAKI ULTRASONIC TESTER PM11 OPERATOR S MANUAL. Be sure to read this manual for safe and proper operation. Store this manual carefully after use. GB MIYAWAKI ULTRASONIC TESTER PM11 OPERATOR S MANUAL Be sure to read this manual for safe and proper operation. Store this manual carefully after use. TABLE OF CONTENTS TO THE OWNER GB-0-2 1. SYMBOL INDICATION

More information

24/7 Sprinkler Monitor. The Ultimate Rain/Freeze Sensor

24/7 Sprinkler Monitor. The Ultimate Rain/Freeze Sensor 24/7 Sprinkler Monitor The Ultimate Rain/Freeze Sensor User s Manual PIONEER SALES, LTD. 5529 Redfield St. Dallas, TX 75235 Phone: (214) 276-0306 Fax: (214) 631-4218 Toll Free: 1-(866) 501-7745 1 Table

More information

Copyright [W.L. Walker Co., Inc.]. All Rights Reserved

Copyright [W.L. Walker Co., Inc.]. All Rights Reserved Copyright [W.L. Walker Co., Inc.]. All Rights Reserved Operating the MKIIs For units originally equipped with bonding straps: To reduce the risk of fire or explosion, this device must be bonded according

More information

INSTALLATION & USER MANUAL

INSTALLATION & USER MANUAL INSTALLATION & USER MANUAL MODELS EVMS & GSVMS with Rain Sensor 00--0 Fax: 0-- csm@wascoproducts.com www.wascoskylights.com Eastern Facility: Spencer Drive, Unit A, Wells, ME 000 Western Facility: Echo

More information

Manometers 2PSI, 5PSI, 15PSI, 30PSI, 100PSI , , ,

Manometers 2PSI, 5PSI, 15PSI, 30PSI, 100PSI , , , Manometers 2PSI, 5PSI, 15PSI, 30PSI, 100PSI 840080, 840081, 840082 840083, 840085 Manometers 840080, 840081, 840082, 840083, 840085 Copyright 2015 by Sper Scientific ALL RIGHTS RESERVED Printed in the

More information

Installation / Operator / Service Manual. Bench Scales Revision 6 07/ by Fairbanks Scales Inc. All rights reserved

Installation / Operator / Service Manual. Bench Scales Revision 6 07/ by Fairbanks Scales Inc. All rights reserved Installation / Operator / Service Manual Bench Scales 2005-2010 by Fairbanks Scales Inc. All rights reserved 50778 Revision 6 07/10 Amendment Record 50778 Manufactured by Fairbanks Scales Inc. 821 Locust

More information

User's Guide Video Borescope Model BR200

User's Guide Video Borescope Model BR200 User's Guide Video Borescope Model BR200 Introduction Congratulations on your purchase of this Extech BR200 Video Borescope. This instrument was designed for use as a remote inspection device. It can be

More information