HAND-HELD THERMOCOUPLE THERMOMETER SERVICE MANUAL. CATALOG NUMBERS , and , and PROPRIETARY

Size: px
Start display at page:

Download "HAND-HELD THERMOCOUPLE THERMOMETER SERVICE MANUAL. CATALOG NUMBERS , and , and PROPRIETARY"

Transcription

1 HAND-HELD THERMOCOUPLE THERMOMETER SERVICE MANUAL CATALOG NUMBERS , and , and PROPRIETARY Information contained in this manual is proprietary to COLE-PARMER INSTRUMENT CO., INC. No reproduction for other than the intended use of maintaining the product described herein is permitted without the written permission of COLE-PARMER INSTRUMENT CO. Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, IL USA (847) (800) A Edition 03 9:07 AM 4/5/2004 Page 1 of 14 A

2 TABLE OF CONTENTS SECTION I SECTION II SECTION III SECTION IV FUNCTIONAL DESCRIPTION BLOCK DIAGRAM CIRCUIT DESCRIPTION SCHEMATIC DIAGRAM REPLACEMENT PARTS LIST TROUBLESHOOTING CALIBRATION PROCEDURE FUNCTIONAL CHECK 9:07 AM 4/5/2004 Page 2 of 14 A

3 FUNCTIONAL DESCRIPTION This Hand-Held Thermocouple Thermometer is a microprocessor-based, precision instrument, for use with any one of J, K or T type thermocouples. The unit features a large, 0.5-inch high liquid crystal display with annunciators, a four button silicon rubber touch keypad, a splashproof plastic case, and is intrinsically safe. Factory calibration is maintained in a non-volatile memory (EEPROM). Readings may be displayed in Fahrenheit or Celsius. A reading Hold function and field calibration for probe offset is also included. BLOCK DIAGRAM 9:07 AM 4/5/2004 Page 3 of 14 A

4 CIRCUIT DESCRIPTION Please refer to the schematic diagram (page 9) while reading this circuit description. Power Supply The power supply circuits perform multiple functions: convert the battery to a steady +3.3V, switch the meter on and off, and provide a reset to the microprocessor. Power to the meter is supplied by two 1.5-volt Alkaline AA batteries. R1 provides intrinsic safety current limiting. U1 and its associated circuitry comprise the power supply. This is configured as a step-up switching regulator with a +3.3 V output, and a 1.0 to 3.2 volt input. The +3.3 is applied to Q1, which acts as an electronic On/Off switch. With the shutdown ( SHDN ) pin connected to the output of this switch, whenever the switch is off, U1 enters a standby mode in which it consumes less than 1 microamp of current. To activate this circuit, the On/Off switch is pressed, which grounds one end of R6. This turns on Q1 which in turn brings the shutdown pin of U1 high, thus starting the regulator. With the +3.3 now up, and the On/Off switch still held down, the latch formed with U5B and U5C, latches Q2 on. Once this happens, you may then release the On/Off switch and the meter will stay on. The output of Q1 has a slow rise time, typically 100 milliseconds. This is too slow for the microprocessor to reset. Since the low battery input pin ( LBI ) of U1 is a precision comparator, it is used to sense when the power supply output has risen to greater than +2.5 volts. At this point the low battery output pin ( LBO ) goes high, which allows C5 to charge, and complete the microprocessor reset sequence. To switch the power supply off, the microprocessor first senses that you have again pressed the On/Off switch. It then blanks the display and waits for you to release the On/Off switch, at which time it clears port pin R40, which causes the U5 latch to toggle, thus turning off Q2, and in turn, Q1. Analog Circuitry The analog circuits consist of a bias and filter network for the thermocouple, an ambient sensor circuit with linearization, an eight channel multiplexer, a 1.25-volt voltage reference, and a discrete single slope analog to digital converter (A/D). Since all voltages within this meter must be positive, and since a thermocouple may produce a negative output voltage differential, a bias network 9:07 AM 4/5/2004 Page 4 of 14 A

5 is used to keep all thermocouple voltages positive. This is R20 and R23. Low frequency filtering is provided by C10, C11 and C12, while high frequencies are attenuated with L3 and L4. The two sides of the thermocouple are connected to two inputs of the multiplexer. Other inputs to the multiplexer are the reference voltage, the ambient sensor, a battery voltage sense and an offset voltage for the A/D converter process. Analog to Digital Converter (A/D) This A/D converter provides 16 bits of resolution ( 1 part in 65,535 ), with a conversion time of 50 milliseconds. It consists of a DC amplifier, window comparator, analog integrator, gated clock and the microprocessor. The output of the multiplexer is applied to a DC amplifier, U3A, which has a fixed gain of 14. This is connected to the upper side of a window detector, U4B. U3B forms an analog integrator. Its output is connected to the lower side of a window detector, U4A. Figure 1 shows a typical A/D conversion cycle. Vin is the output level at U3A, and has a normal range from 500 mv minimum to 2 V maximum. VL is the voltage present at U4A-2, typically 450 mv. TPA goes low to signify the start of the measuring period, while TPB goes low to signify the end of the measuring period. The microprocessor counts the clock pulses, at the output of U5D, while the gate is high, but only between T3 and T4. Beginning at T0, the emitter of Q3 is raised high. This allows the output of the integrator, U3B, to rise. At T2, when the integrator output crosses the input level, Vin, the microprocessor pulls Q3 emitter low. The delay from T2 until this occurs is the small overshoot. As the integrator again crosses the Vin level, at T3, the gate again opens and the microprocessor begins counting clock pulses. This continues until T4, the point where the integrator crosses VL. With the gate closed, the microprocessor reads the number of clock pulses and converts this to a voltage, which is then used for calculating the thermocouple or ambient temperature or battery voltage. One of these conversions occurs for each multiplexer input. The multiplexer is switched from one channel to the next at time T4. The total time from T0 to T4 is a maximum of 50 ms. There is a 50 ms delay between conversions (from T4 to the next T0). This delay allows for the DC amplifier to settle. 9:07 AM 4/5/2004 Page 5 of 14 A

6 Figure 1. A/D Timing Other Features Thermocouple voltages are measured by taking readings of each thermocouple lead, with respect to ground, and subtracting the two values with the microprocessor. Calibration information and thermocouple type are stored in the EEPROM. Thermocouple type is determined when the unit is calibrated on the test fixture, which makes connection through the calibration access port. Alternately, there is a secret key combination that allows this to be changed: Note: Begin with power off. Press and hold the C/ F and HOLD buttons and then press On/Off. Wait about 5 seconds until the display shows a single digit number, 0, 1 or 2. Release the C/ F and HOLD buttons. A 0 represents K, 1 represents J, and 2 represents T. Use the C/ F button to select 0, 1 or 2, as required. Press the HOLD button. The selected thermocouple type is now stored in EEPROM. If different, the front panel label should be changed to match this new type. 9:07 AM 4/5/2004 Page 6 of 14 A

7 The liquid crystal display is of the static drive type. This means that the drive voltage is a simple square wave, with levels of 0 and +3V. The result is a very wide viewing angle (as compared to a multiplexed LCD). The microprocessor runs at 2.09 MHZ; this frequency should only be measured at U8-2. 9:07 AM 4/5/2004 Page 7 of 14 A

8 9:07 AM 4/5/2004 Page 8 of 14 A

9 REPLACEMENT PARTS LIST Barnant P/N Cole-Parmer P/N Description E-2105 E-2105 PCB Assy E-2525 E-2525 Input Connector Assy B B Screw, #2-56 B B Lockwasher, #2 E-2073 E-2073 Case, Front B-3556 B-3553 Overlay, Window B-3564 B-3558 Overlay, J B-3566 B-3560 Overlay, K B-3565 B-3559 Overlay, T D-2828 D-2828 Lens D-2818 D-2818 LCD B-3574 B-3574 Zebra Strip E-2075 E-2075 Keypad B B Case, Back B B Wire Bail D D Battery Clips E-2076 E-2076 Battery Door B B Fiber Washer B B Screw, Battery Door A-4498 A-4498 Foot, Bottom A-4497 A-4497 Pad, Bottom B B Screw, Case 9:07 AM 4/5/2004 Page 9 of 14 A

10 SECTION III TROUBLESHOOTING The following is a list of typical problems that might occur and possible remedies. Does Not Turn On Dead batteries -- replace with new ones. Bad battery connections -- check the clips for spring tension to the PC board. Batteries installed backwards. On/Off switch not working -- clean switch contacts. Low Battery graphic is on. Dead batteries If the batteries are OK, then the EEPROM may have lost calibration information. Error Messages Err -- field cal outside of 0±10 C (32±18 F) Err1 -- EEPROM has lost factory cal info -- recalibrate. Err2 -- EEPROM has lost stored thermocouple type or bad field cal data -- recalibrate. Err3 -- EEPROM write verify error -- replace EEPROM. Err4 -- A/D not running correctly, signal at TPA is wrong. Note: Thermocouple channels display Open for this condition. Err5 -- A/D not running correctly, signal at TPB is wrong. Note: Thermocouple channels display Open for this condition. Err6 -- A/D overflow (overranged). Note: Thermocouple channels display instead of Err6 to indicate over probe limit -- probably wrong thermocouple type. Err7 -- Bad 1.2V reference value stored in EEPROM -- recalibrate. Note: This error is not displayed if Err1 also exists. 9:07 AM 4/5/2004 Page 10 of 14 A

11 SECTION IV CALIBRATION PROCEDURE Basic Factory Calibration, Without The Aid Of A Test Fixture. 1. Determine software version. Install batteries. Begin with power off. Press and hold the C/ F button and then press and release On/Off. At the end of the display test, a software version will be displayed in the form x.x. For 1.0 proceed to step 2. For 1.1 and higher, proceed to step Perform this first step only if Err7 is displayed or if D4 has been replaced. Otherwise, you may proceed directly to step 4. This step requires that the UUT have power but you must also have access to the electronics at the top of the UUT. One suggested method is to take a rear case and cut off the top. Then install batteries into the bottom and attach the rear case with screws. Press On/Off to activate UUT; ignore the display. Measure the voltage at pin 8 (referenced to pin 4) of D4, the 1.2-volt reference, and write it down. You need four significant digits right of the decimal point, as in Attach the rear case normally, with batteries installed. 4. Begin with power off. Press and hold the CAL and HOLD buttons and then press and release On/Off. Wait until the display blanks, then release only the HOLD button--continue pressing CAL. Wait for CAL to begin flashing. Press and hold C/ F and release CAL. In a few seconds, the display will show 2300 (assuming this unit has not been previously calibrated) or, from the above example in step 2, would show Release C/ F. Now use C/ F and HOLD to increment/decrement the display to match the 4 significant digits determined in step 2. Press CAL. Proceed to step Begin with power off. Press and hold the CAL and HOLD buttons and then press On/Off. Wait until the display blanks, then release only the HOLD button--continue pressing CAL. Wait for CAL to begin flashing. Release CAL. 6. Using only K type thermocouple connectors and wire, and with the calibration source set for K type, apply +950 C or F to the UUT. (This K selection is true even if this is a J or T type unit.) Wait for the reading to settle and press CAL. The display will show 1 for a few seconds, then the display returns. 7. Apply 0 C or +32 F to the UUT. Wait for the reading to settle and press CAL. The display will show 2 for a few seconds, then the display returns. 9:07 AM 4/5/2004 Page 11 of 14 A

12 8. Connect a copper short to the input. Wait for the reading to settle and press CAL. The display will show 3 for a few seconds, then the display returns. 9. Apply 0 C or +32 F to the UUT. Wait for the reading to settle and press CAL. The CAL indicator should go off, and the UUT should indicate 0 C/+32 F, if the unit is selected as a K type. Calibration of software version 1.0 is now complete Begin here for software version 1.1 and higher; go back to step 2 for 11. Perform this first step only if Err7 is displayed or if D4 has been replaced. Otherwise, you may proceed directly to step 14. This step requires that the UUT have power but you must also have access to the electronics at the top of the UUT. One suggested method is to take a rear case and cut off the top. Then install batteries into the bottom and attach the rear case with screws. Press On/Off to activate UUT; ignore the display. Measure the voltage at pin 8 (referenced to pin 4) of D4, the 1.2-volt reference, and write it down. You need four significant digits right of the decimal point, as in Attach the rear case normally, with batteries installed. 13. Begin with power off. Press and hold the CAL and HOLD buttons and then press and release On/Off. Wait until the display blanks, then release only the HOLD button--continue pressing CAL. Wait for CAL to begin flashing. Press and hold C/ F and release CAL. In a few seconds, the display will show 2300 (assuming this unit has not been previously calibrated) or, from the above example in step 11, would show Release C/ F. Now use C/ F and HOLD to increment/decrement the display to match the 4 significant digits determined in step 11. Press CAL. Proceed to step Begin with power off. Press and hold the CAL and HOLD buttons and then press On/Off. Wait until the display blanks, then release only the HOLD button--continue pressing CAL. Wait for CAL to begin flashing. Release CAL. 15. Using only K type thermocouple connectors and wire, and with the calibration source set for K type, apply +950 C or F to the UUT. (This K selection is true even if this is a J or T type unit.) Wait for the reading to settle and press CAL. The display will show 1 for a few seconds, then the display returns. 9:07 AM 4/5/2004 Page 12 of 14 A

13 16. Apply 0 C or +32 F to the UUT. Wait for the reading to settle and press CAL. The display will show 2 for a few seconds, then the display returns. 17. Apply 0 C or +32 F to the UUT. Wait for the reading to settle and press CAL. The CAL indicator should go off, and the UUT should indicate 0 C/+32 F, if the unit is selected as a K type. Calibration of software version 1.1 and higher is now complete. FUNCTIONAL CHECK 1. Press and hold ON/OFF. Verify all LCD segments and characters as being present. This display should remain on as long as ON/OFF is being held down. 2. Release ON/OFF. If any Err # is displayed, tag the unit with the Err # and set aside. NOTE: The tolerances given in these steps do not allow for any error(s) originating from the calibration source. 3. Apply +0.0 C. Use C/ F key to select C. Verify the reading to be -0.5 to +0.5 C. 4. Use C/ F key to select F. Verify the reading to be to F. 5. Press HOLD. The HOLD designator on the LCD should be on. 6. Apply the following input value: J type C K type C T type +400 C 9:07 AM 4/5/2004 Page 13 of 14 A

14 The reading shall not change from above. 7. Press HOLD. The HOLD designator on the LCD should go off. 8. Use C/ F to select C. Verify the reading to be: J type +997 to C K type to C T type +398 to +402 C 9. Remove the thermocouple probe from the top of the UUT. The readings should begin decreasing (reading more negative), until the word OPEN is displayed. 10. Press and hold CAL for about 5 seconds until CAL blinks on the display. Release CAL. 11. Press ON/OFF to turn the UUT off. Test complete. 9:07 AM 4/5/2004 Page 14 of 14 A

********SERVICE MANUAL******** MODELS:

********SERVICE MANUAL******** MODELS: ********SERVICE MANUAL******** MODELS: 75211-10 Variable Speed Drive, 5000 RPM, 115 V 75211-15 Variable Speed Drive, 5000 RPM, 230 V, CE Mark 75211-60 Variable Speed Drive w/pump, 9000 RPM, 115 V 75211-62

More information

ph/orp 300 (Cat. No ) MICROCOMPUTER BASED OPERATION MANUAL ph /ORP/ Temperature CONTROLLER Phone: , Fax:

ph/orp 300 (Cat. No ) MICROCOMPUTER BASED OPERATION MANUAL ph /ORP/ Temperature CONTROLLER Phone: , Fax: OPERATION MANUAL /ORP 300 (Cat. No. 0580212) MICROCOMPUTER BASED /ORP/ Temperature CONTROLLER 625 East Bunker CT, Vernon Hills, IL 60061-1844 Phone: 800-32-4340, Fax: 847-247-2929 E-mail: info@coleparmer.com,

More information

Specifications

Specifications Specifications 18200-10 Cole-Parmer Instrument Company 625 East Bunker Court Vernon Hills, Illinois 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com e-mail: techinfo@coleparmer.com

More information

Cole-Parmer Instrument Company

Cole-Parmer Instrument Company Operating Instructions Cole-Parmer ph/orp 400 ph/orp Controller Cole-Parmer Instrument Company 625 East Bunker Court, Vernon Hills, Illinois 60061 Tel: 1-847-549-7600 or Toll-free: 1-800-323-4340 Fax:

More information

Platinum RTD Thermometer MODEL NO

Platinum RTD Thermometer MODEL NO Platinum RTD Thermometer MODEL NO. 93410-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com

More information

Platinum RTD Thermometer MODEL NO

Platinum RTD Thermometer MODEL NO Platinum RTD Thermometer MODEL NO. 93410-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com

More information

DIGITAL METERS Universal Analog Input Process Meter With Rate/Totalizer/Batch Control Features Instruction Manual

DIGITAL METERS Universal Analog Input Process Meter With Rate/Totalizer/Batch Control Features Instruction Manual DIGITAL METERS Universal Analog Input Process Meter With Rate/Totalizer/Batch Control Features 4-20 ma, 1-5 V, 0-5 V, or 0-10 V Field Selectable Inputs Full Six Digit Display for Total, 4½ Digit + Extra

More information

Cole-Parmer Instrument Company

Cole-Parmer Instrument Company Operating Instructions Cole-Parmer ph/orp 350 ph/orp Transmitter Cole-Parmer Instrument Company 625 East Bunker Court, Vernon Hills, Illinois 60061 Tel: 1-847-549-7600 or Toll-free: 1-800-323-4340 Fax:

More information

********SERVICE MANUAL********* MODELS: Gear Pump Drive, w/4-20 ma, 3600 RPM, 115 V Gear Pump Drive, w/4-20 ma, 3600 RPM, 230 V CE

********SERVICE MANUAL********* MODELS: Gear Pump Drive, w/4-20 ma, 3600 RPM, 115 V Gear Pump Drive, w/4-20 ma, 3600 RPM, 230 V CE ********SERVICE MANUAL********* MODELS: 75211-50 Gear Pump Drive, w/4-20 ma, 3600 RPM, 115 V 75211-55 Gear Pump Drive, w/4-20 ma, 3600 RPM, 230 V CE PROPRIETARY Information contained in this manual is

More information

Specifications

Specifications Specifications 18200-40 Cole-Parmer Instrument Company 625 East Bunker Court Vernon Hills, Illinois 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com e-mail: techinfo@coleparmer.com

More information

IMPROVED HARDWARE FEATURES

IMPROVED HARDWARE FEATURES IMPROVED HARDWARE FEATURES 1. FOURTEEN SEGMENT DISPLAYS Improvement from the seven-segment displays on the DynaTrol 600. All letters can be represented on these displays; some messages will change from

More information

CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM

CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM-6650.02 1 Instruction Manual DM-6650.02 TABLE OF CONTENTS I.0 INTRODUCTION 3 1.1 General Description... 3 1.2 Specifications... 3 1.3

More information

350 SMART MANOMETER OPERATING INSTRUCTIONS

350 SMART MANOMETER OPERATING INSTRUCTIONS 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com 350 SMART MANOMETER OPERATING INSTRUCTIONS Meriam Instrument s 350 Smart Manometer

More information

Professional Radio GM Series. Controlhead Service Information

Professional Radio GM Series. Controlhead Service Information Professional Radio GM Series Controlhead Service Information Issue: September 2000 ii Computer Software Copyrights The Motorola products described in this manual may include copyrighted Motorola computer

More information

INSTRUCTION MANUAL STATION CONTROLLER SC1000 MOTOR PROTECTION ELECTRONICS, INC.

INSTRUCTION MANUAL STATION CONTROLLER SC1000 MOTOR PROTECTION ELECTRONICS, INC. INSTRUCTION MANUAL STATION CONTROLLER SC1000 MOTOR PROTECTION ELECTRONICS, INC. 2464 Vulcan Road, Apopka, Florida 32703 Phone: (407) 299-3825 Fax: (407) 294-9435 Revision Date: 9-11-08 Applications: Simplex,

More information

Barnant Company. Humidity Data Logger MODEL NO YR MIN HOLD MO MAX DAY STORE LOG RECALL PRINT C/ F PRINT SETUP.

Barnant Company. Humidity Data Logger MODEL NO YR MIN HOLD MO MAX DAY STORE LOG RECALL PRINT C/ F PRINT SETUP. Barnant Company Humidity Data Logger MODEL NO. 600-2050 %RH CAL F DP YR MIN HOLD MO MAX DAY STORE LOG RECALL PRINT STORE C/ F HOLD RECALL LOG MAX CLEAR PRINT MIN ON /OFF SETUP CAL Barnant Company 28W092

More information

4-Channel Thermometer

4-Channel Thermometer 4-Channel Thermometer 800023 4-Channel Thermometer - 800023 Copyright 2015 by Sper Scientific ALL RIGHTS RESERVED. Printed in the USA The contents of this manual may not be reproduced or transmitted in

More information

HI HI N HI HI

HI HI N HI HI Instruction Manual HI 935005 - HI 935005N HI 935002 - HI 935009 Portable K-Thermocouple Thermometers www.hannainst.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you

More information

Copyright [W.L. Walker Co., Inc.]. All Rights Reserved

Copyright [W.L. Walker Co., Inc.]. All Rights Reserved Copyright [W.L. Walker Co., Inc.]. All Rights Reserved Operating the MKIIs For units originally equipped with bonding straps: To reduce the risk of fire or explosion, this device must be bonded according

More information

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics PIECAL 322 Automated Thermocouple Calibrator Operating Instructions Product Description Easy to use With the PIECAL 322-1 you can check & calibrate all your thermocouple instruments and measure thermocouple

More information

ACTIONI/Q Q498. DC Powered DC Input Field Configurable Isolator with Math Functions

ACTIONI/Q Q498. DC Powered DC Input Field Configurable Isolator with Math Functions ACTIONI/Q Q498 DC Powered DC Input Field Configurable Isolator with Math Functions Q498-0000 Provides 2 Isolated Analog Inputs, 1 Analog Output & 1 Frequency Output, as well as a Discrete Input & Output

More information

Multifunction Calibrator

Multifunction Calibrator Multifunction Calibrator Model CL TC - - - C V _ TECPEL CO., LTD. Features:. - (KΩ load, V Loop Supply). -.,-.V, -.V. K, J, E, T Thermocouple ( C and ). Frequency -.. Basic Accuracy. Easy Key-pad Operation.

More information

Autoranging True RMS Multimeter User Manual

Autoranging True RMS Multimeter User Manual Autoranging True RMS Multimeter User Manual Please read this manual before switching the unit on. Important safety information inside. Contents Page 1. Safety Information... 4 2. Safety Symbols... 5 3.

More information

PRO 360. Owner's Manual

PRO 360. Owner's Manual One-Year Limited Warranty If, within one year from the date of original purchase, the Pro 360 Digital Protractor fails to function because of defects in materials or workmanship, the manufacturer will,

More information

OPERATION MANUAL. ph 500 (Cat. No ) MICROCOMPUTER BASED ph / Temperature TRANSMITTER & CONTROLLER

OPERATION MANUAL. ph 500 (Cat. No ) MICROCOMPUTER BASED ph / Temperature TRANSMITTER & CONTROLLER OPERATI MANUAL 500 (Cat. No. 05802-16) MICROCOMPUTER BASED / Temperature TRANSMITTER & CTROLLER 625 East Bunker CT Vernon Hills, IL 60061-1844 Phone: 800-32-4340 Fax: 847-247-2929 E-mail: info@coleparmer.com

More information

DSTCAL Software User Manual 09/16/09 WD1037 Rev E Revised 09/09/14

DSTCAL Software User Manual 09/16/09 WD1037 Rev E Revised 09/09/14 DSTCAL Software User Manual 09/16/09 WD1037 Rev E Revised 09/09/14 Wahl Instruments, Inc. 234 Old Weaverville Road Asheville, NC 28804 Toll Free 800-421-2853 Phone 828-658-3131 Fax 828-658-0728 www.palmerwahl.com

More information

Instruction Manual Temp JKT Temperature Meter Series

Instruction Manual Temp JKT Temperature Meter Series Instruction Manual Temp JKT Temperature Meter Series 68X243615 Technology Made Easy... Rev. 2 11/03 Preface This manual serves to explain the use of the Temp JKT temperature meter. This manual functions

More information

Model ST-610B. Instruction Manual. Digital Thermocouple Thermometer. reedinstruments. www. com

Model ST-610B. Instruction Manual. Digital Thermocouple Thermometer. reedinstruments. www. com Model ST-610B Digital Thermocouple Thermometer Instruction Manual reedinstruments com Table of Contents Safety... 2 Features... 3 Specifications...3-4 Operating Instructions... 5 Additional Functions...

More information

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter S-14 Compact Digital Multimeter S-14 Compact Digital Multimeter SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation at this meter

More information

SINGLE-CHANNEL PROCESS CONTROLLERS & TEMP. DISPLAYS D3820/D3830 Series

SINGLE-CHANNEL PROCESS CONTROLLERS & TEMP. DISPLAYS D3820/D3830 Series The Digitec Series D3800 Panel Meters offer many features and performance capabilities to suit a wide range of industrial applications. The D3820 and D3830 are available in the ranges of 4-20mA, 0-10 VDC,

More information

5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL Fruitland Ave Los Angeles, CA UM Channel Monitor.

5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL Fruitland Ave Los Angeles, CA UM Channel Monitor. 5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL 33309 3211 Fruitland Ave Los Angeles, CA 90058 UM-600 6-Channel Monitor Version 2 Installation and Operation Manual Rev. G P/N145F-12990 PCO 00007462 (c)

More information

MAINTENANCE MANUAL KEYPAD/DISPLAY PANEL ASSEMBLIES 19D902913G3, G4, G10 (CONVENTIONAL) 19D902913G5, G6 (EDACS)

MAINTENANCE MANUAL KEYPAD/DISPLAY PANEL ASSEMBLIES 19D902913G3, G4, G10 (CONVENTIONAL) 19D902913G5, G6 (EDACS) A MAINTENANCE MANUAL KEYPAD/DISPLAY PANEL ASSEMBLIES 19D902913G3, G4, G10 (CONVENTIONAL) 19D902913G5, G6 (EDACS) TABLE OF CONTENTS Page SPECIFICATIONS............................................... Front

More information

MANUAL PCE-123

MANUAL PCE-123 www.pce-industrial-needs.com Tursdale Technical Services Ltd Unit N12B Tursdale Business Park Co. Durham DH6 5PG United Kingdom Phone: +44 ( 0 ) 191 377 3398 Fax: +44 ( 0 ) 191 377 3357 info@tursdaletechnicalservices.co.uk

More information

FH125/FT121. Temperature / Humidity LCD Graphing Logger FH125 SET CURSOR. 1 Hour/Div 1 1 : 12 : 23 01/03/02 (ESC)

FH125/FT121. Temperature / Humidity LCD Graphing Logger FH125 SET CURSOR. 1 Hour/Div 1 1 : 12 : 23 01/03/02 (ESC) 72. 1 F- 27.8 H-- 1 Hour/Div 1 1 : 12 : 23 01/03/02 DICKSON FH125/FT121 Temperature / Humidity LCD Graphing Logger 100 80 60 40 20 0 SET CURSOR (ESC) FH125 FH121 Specifications FH121 Temperature Scales:

More information

Model R2400. Instruction Manual. Digital Thermocouple Thermometer. reedinstruments

Model R2400. Instruction Manual. Digital Thermocouple Thermometer. reedinstruments Model R2400 Digital Thermocouple Thermometer Instruction Manual reedinstruments com 1-877-849-2127 info@reedinstruments.com.reedinstruments.com Table of Contents Safety... 2 Features... 3 Specifications...3-4

More information

TD-700 FLUOROMETER SERVICE MANUAL

TD-700 FLUOROMETER SERVICE MANUAL TD-700 FLUOROMETER SERVICE MANUAL July 1996 CONTENTS Page Section 1 INTRODUCTION 2 Section 2 PRELIMINARY CHECKS 3 Section 3 TROUBLESHOOTING GUIDE 5 A. Lamp (Fluorescent) 5 B. Lamp Heater 7 C. Fan 8 D.

More information

Model User Manual Revision E 04/29/99. OEM Gaging System. Part Number R01

Model User Manual Revision E 04/29/99. OEM Gaging System. Part Number R01 04/29/99 Model 3800 OEM Gaging System User Manual Revision E Part Number 028585-R01 Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted

More information

7561-PSD Manual Portable Battery Powered Indicator

7561-PSD Manual Portable Battery Powered Indicator 7561-PSD Manual Portable Battery Powered Indicator Lebow Products Inc. 1728 Maplelawn Drive P.O. Box 1089 Troy, Michigan 48084-1089 (800) 803-1164 Phone: (248) 643-0220 FAX: (248) 643-0259 Visit our web

More information

SI3300. user and installation manual. 4-20mA/DC-Digital Display

SI3300. user and installation manual. 4-20mA/DC-Digital Display SI3300 4-20mA/DC-Digital Display The SI3300 is a member of the SI3000 Readout Family. All members of the family are marked SI3000 on the front panel. This manual is specifically for the SI3300 Model with

More information

user and installation manual

user and installation manual SI3500 SI3500 ORBIT Digital Display The SI3500 is a member of the SI3000 Readout Family. All members of the family are marked SI3000 on the front panel. This manual is specifically for the SI3500 Model

More information

PRO 360. Owner's Manual

PRO 360. Owner's Manual PRO 360 Owner's Manual Introduction The Pro 360 Digital Protractor is a revolutionary measuring tool that provides an immediate, digital reading of all angles in a 360 circle. The machined aluminum frame

More information

SC2000 MOTOR PROTECTION ELECTRONICS, INC. INSTRUCTION MANUAL. (407) Phone: Website:

SC2000 MOTOR PROTECTION ELECTRONICS, INC. INSTRUCTION MANUAL. (407) Phone: Website: SC2000 INSTRUCTION MANUAL MOTOR PROTECTION ELECTRONICS, INC. 2464 Vulcan Road Apopka, Florida 32703 Phone: Website: (407) 299-3825 www.mpelectronics.com Operating Program Revision: 12 Revision Date: 8-27-14

More information

4-Channel Datalogging Thermometer

4-Channel Datalogging Thermometer 4-Channel Datalogging Thermometer 800024 4-Channel Datalogging Thermometer - 800024 Copyright 2015 by Sper Scientific ALL RIGHTS RESERVED. Printed in the USA The contents of this manual may not be reproduced

More information

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: This meter is

More information

E3B USER'S GUIDE. for QT310 / QT320 QProx IC evaluation and development Overview. Fast Start

E3B USER'S GUIDE. for QT310 / QT320 QProx IC evaluation and development Overview. Fast Start E3B USER'S GUIDE for QT310 / QT320 QProx IC evaluation and development Overview The E3B board works with Quantum s QT310 and QT320 QProx ICs. Either device can be inserted into the 8-pin DIP socket. QT310

More information

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE

More information

DIGITAL PANEL THERMOMETER

DIGITAL PANEL THERMOMETER Love Controls DIGITAL PANEL THERMOMETER LCI108-1x LCI108J-1x INSTRUCTION MANUAL EDITION: January 2001 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This manual does not constitute a formal agreement.

More information

Victoreen 190I. Operators Manual. Survey Meter

Victoreen 190I. Operators Manual. Survey Meter Victoreen 190I Survey Meter Operators Manual March 2005 Manual No. 191001 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective

More information

D115 The Fast Optimal Servo Amplifier For Brush, Brushless, Voice Coil Servo Motors

D115 The Fast Optimal Servo Amplifier For Brush, Brushless, Voice Coil Servo Motors D115 The Fast Optimal Servo Amplifier For Brush, Brushless, Voice Coil Servo Motors Ron Boe 5/15/2014 This user guide details the servo drives capabilities and physical interfaces. Users will be able to

More information

DTM-321 REL HOLD T2 K-TYPE TEMP: 20 C 60 C 140 F C F HUMIDITY: %RH POWER-UP OPTIONS

DTM-321 REL HOLD T2 K-TYPE TEMP: 20 C 60 C 140 F C F HUMIDITY: %RH POWER-UP OPTIONS DTM-321 T1 T2 K-TYPE REL HOLD C F TEMP: ( ) HUMIDITY: 20 C 4 F 200 328 0 C F POWER-UP OPTIONS 60 C 140 F 1370 2498 100 %RH C F CONTENTS TITLE PAGE I. Safety Information... 1 II. Introduction...... 1 III.

More information

DFI 9000 PORTABLE DIGITAL INDICATOR USER S GUIDE

DFI 9000 PORTABLE DIGITAL INDICATOR USER S GUIDE DFI 9000 PORTABLE DIGITAL INDICATOR USER S GUIDE www.cooperinstruments.com PH: 540-349-4746 FAX: 540-347-4755 CONTENTS 1.0 INTRODUCTION...1 2.0 SETUP AND USE...1 2.1 Your DFI 9000...1 2.2 Additional Tools...1

More information

BC-7000 Field Calibration Procedure 55 AMP

BC-7000 Field Calibration Procedure 55 AMP BC-7000 Field Calibration Procedure 55 AMP CF1_FIELDCAL_BC7000 Revised 9/8/2012 INTRODUCTION: The BC-7000 Battery Capacity Analyzer will perform capacity testing of 12 and 24 volt lead-acid batteries.

More information

Installation & Operation

Installation & Operation LED Readout Installation & Operation WARRANTY Accurate Technology, Inc. warrants the ProScale Systems against defective parts and workmanship for 1 year commencing from the date of original purchase. Upon

More information

6-Channel Monitor. Installation and Operation Manual

6-Channel Monitor. Installation and Operation Manual 3211 Fruitland Ave Los Angeles, CA 90058 Catalyst Monitor 6-Channel Monitor Version 2 Installation and Operation Manual Rev. H P/N145F-12964 PCO - 00009743 (c) Copyright 2015, Barksdale, Inc. All Rights

More information

TC100 Precision Thermocouple Calibrator

TC100 Precision Thermocouple Calibrator TC100 Precision Thermocouple Calibrator Table of Contents 1. Introduction.....................1 2. Accessories....................5 3. Set-Up Basic/Advanced...........5 4. Operating Procedure.............6

More information

LAUREL ELECTRONICS, INC.

LAUREL ELECTRONICS, INC. LAUREL ELECTRONICS, INC. Laureate RTD Temperature Panel Meter / Controller Features Factory calibrated for 100Ω platinum, 10Ω copper & 120Ω nickel RTDs 2, 3 or 4-wire connection with lead resistance compensation

More information

HGM9580 Bus Tie Bus Parallel Unit USER MANUAL. Smartgen Technology

HGM9580 Bus Tie Bus Parallel Unit USER MANUAL. Smartgen Technology HGM9580 Bus Tie Bus Parallel Unit USER MANUAL Smartgen Technology This manual is suitable for HGM9580 bus tie bus parallel unit only. Clarification of notation used within this publication. SIGN INSTRUCTION

More information

User Guide. TP3016M Series Switching Mode DC Power Supply

User Guide. TP3016M Series Switching Mode DC Power Supply User Guide TP3016M Series Switching Mode DC Power Supply 1 Introduction I. Quick Start 1.1 Front Panel Description 1.2 Pre-Checking 1.3 Quick Start 1.4 Output Checking II. Specifications 2.1 Major Specification

More information

Section 30. In-Circuit Serial Programming (ICSP )

Section 30. In-Circuit Serial Programming (ICSP ) Section 30. In-Circuit Serial Programming (ICSP ) HIGHLIGHTS This section of the manual contains the following major topics: 30. Introduction... 30-2 30.2 Entering In-Circuit Serial Programming Mode...

More information

HI HI N HI HI

HI HI N HI HI Instruction Manual HI 935005 - HI 935005N HI 935002 - HI 935009 Portable K-Thermocouple Thermometers w w w. h a n n a i n s t. c o m Dear Customer, Thank you for choosing a Hanna product. Please read this

More information

1. Introduction 2. Specifications 3. Symbol Definition and Button Location 4. Operation Instructions 5. Temperature Measurement 6.

1. Introduction 2. Specifications 3. Symbol Definition and Button Location 4. Operation Instructions 5. Temperature Measurement 6. CONTENTS TITLE PAGE 1. Introduction.. 1 2. Specifications... 1 3. Symbol Definition and Button Location... 2 4. Operation Instructions.... 3 4.1 Power-Up...3 4.2 Connection of the Temperature Probe...

More information

This Datasheet for the IC670ALG310. Analog Output Voltage, 8 Channel, +/-10V, 0 to 10V.

This Datasheet for the IC670ALG310. Analog Output Voltage, 8 Channel, +/-10V, 0 to 10V. This Datasheet for the IC670ALG310 Analog Output Voltage, 8 Channel, +/-10V, 0 to 10V http://www.qualitrol.com/shop/p-14493-ic670alg310.aspx Provides the wiring diagrams and installation guidelines for

More information

2-Axis Counter BDD 622

2-Axis Counter BDD 622 2-Axis Counter BDD 622 Technical Description, User's Guide english No. 622-221 D/E. Edition 0701; Subject to modification. www.balluff.com Balluff GmbH Schurwaldstrasse 9 73765 Neuhausen a.d.f. Germany

More information

Transducers for Pt-100/1000, Resistors

Transducers for Pt-100/1000, Resistors Transducers for Pt-100/1000, Resistors Microprocessor-based technology General Description Isolating transducers with digital programming of ranges, for DIN-rails or for printed circuit boards. Modules

More information

GV3000/SE Operator Interface Module (OIM) User Guide Version 2.0 M/N 2RK3000

GV3000/SE Operator Interface Module (OIM) User Guide Version 2.0 M/N 2RK3000 GV3000/SE Operator Interface Module (OIM) User Guide Version 2.0 M/N 2RK3000 Instruction Manual D2-3342-2 The information in this manual is subject to change without notice. Throughout this manual, the

More information

Type K/J Thermometers

Type K/J Thermometers Type K/J Thermometers 800005 CONTENTS INTRODUCTION....2 FRONT PANEL DESCRIPTION...3 LCD DISPLAY....4 OPERATION INSTRUCTION...5 Power Up....5 Connecting the Thermocouples....5 Selecting Temperature Scale....5

More information

DICKSON TC200/TH300 DICKSON. Handheld Temperature and Temperature / RH Indicators. Useful Features. Applications & Product. Specifications.

DICKSON TC200/TH300 DICKSON. Handheld Temperature and Temperature / RH Indicators. Useful Features. Applications & Product. Specifications. / Handheld Temperature and Temperature / RH Indicators Contents: Applications and Accessories Warranty / This versatile temperature indicator accepts J, K and T thermocouple probes with mini-connectors.

More information

IC693PWR331 High Capacity Power Supply, 24 VDC Input

IC693PWR331 High Capacity Power Supply, 24 VDC Input IC693PWR331 High Capacity Power Supply, 24 VDC Input The Series 90-30 DC input High Capacity power supply (IC693PWR33) is a 30 watt wide range supply designed for 24 VDC nominal inputs. For applications

More information

DVC-350C CALIBRATORS, INC HANDHELD BATTERY-POWERED VOLTAGE CALIBRATOR FEATURES INTRODUCTION GENERAL DESCRIPTION

DVC-350C CALIBRATORS, INC HANDHELD BATTERY-POWERED VOLTAGE CALIBRATOR FEATURES INTRODUCTION GENERAL DESCRIPTION CALIBRATORS, INC HANDHELD BATTERY-POWERED VOLTAGE CALIBRATOR FEATURES HAND-HELD, LIGHT WEIGHT 11 OUNCES (342 GRAMS) LABORATORY ACCURACY OF +/- 0.015% OF FSR AT 25c, TRACEABLE TO N.I.S.T. 4 ½ DIGIT LCD

More information

I. PANEL DESCRIPTION... 1

I. PANEL DESCRIPTION... 1 Table of Contents I. PANEL DESCRIPTION... 1 II. OPERATING INSTRUCTION... 7 1. MA OUTPUT... 7 1A. GENERAL OPERATION 4-20MA... 7 1B. SELECT 0-20MA OR 0-24MA... 8 1C. ENTER A VALUE LESS THAN 1... 9 2. % (PERCENTAGE)

More information

GFC114. Low Power Flow Computer FEATURES

GFC114. Low Power Flow Computer FEATURES Low Power Flow Computer FEATURES l UNIVERSAL INPUT, 5 khz, CAN POWER THE l SIMPLIFIED MENUS AND PROGRAMMING l NO NEED OF AMPLIFIERS OR LINEARIZERS l SEVEN DIGIT RATE WITH PROGRAMMABLE DECIMAL PLACES l

More information

OPERATING INSTRUCTION

OPERATING INSTRUCTION OPERATING INSTRUCTION AUTORANGING MULTIMETER MAX Ω F C 10A MAX every 15 min. COM V SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation

More information

Model INSTRUCTION MANUAL DIGITAL MULTIMETER

Model INSTRUCTION MANUAL DIGITAL MULTIMETER Model 57040 INSTRUCTION MANUAL DIGITAL MULTIMETER SAFETY INFORMATION This multimeter has been designed according to IEC 1010 concerning electronic measuring instruments with an overvoltage category (CAT

More information

SC2000 MOTOR PROTECTION ELECTRONICS, INC. INSTRUCTION MANUAL. Phone: (407) Fax: (407) Vulcan Road Apopka, Florida 32703

SC2000 MOTOR PROTECTION ELECTRONICS, INC. INSTRUCTION MANUAL. Phone: (407) Fax: (407) Vulcan Road Apopka, Florida 32703 SC2000 INSTRUCTION MANUAL MOTOR PROTECTION ELECTRONICS, INC. 2464 Vulcan Road Apopka, Florida 32703 Phone: (407) 299-3825 Fax: (407) 294-9435 Operating Program Revision: 10 Revision Date: 1-9-12 STATION

More information

BTH-1208LS Wireless Multifunction DAQ Device

BTH-1208LS Wireless Multifunction DAQ Device Wireless Multifunction DAQ Device Features Eight 11-bit single-ended (SE) or four 12-bit differential (DIFF) analog input channels Acquires data over Bluetooth or USB connection Maximum sampling rate of

More information

Trend-1 Vibration Data Logger. User Manual

Trend-1 Vibration Data Logger. User Manual Trend-1 Vibration Data Logger User Manual Copyright College of Engineering-University of Basrah 2013 1 1. General Specifications 1.1 Features Thank you for being interested in purchasing Trend-1 Vibration

More information

RVDT Setup and Polarity Checking Tool for EG Governors. Product Manual (Revision B) Original Instructions T87444

RVDT Setup and Polarity Checking Tool for EG Governors. Product Manual (Revision B) Original Instructions T87444 Product Manual 01504 (Revision B) Original Instructions RVDT Setup and Polarity Checking Tool for EG Governors T87444 Operation and Calibration Manual General Precautions Read this entire manual and all

More information

AS Channels Capacitive Touch Sensor IC From Santa Clara, United States of America

AS Channels Capacitive Touch Sensor IC From Santa Clara, United States of America ASI Competitor Equivalent A Competitor Equivalent B Volts Leading Performance: ESD HBM >8k Volts (Directly Applied to All IC Pins) Operating Temperature up to >+95 0 C Features Overview Analog and Digital

More information

Chapter 1 Introducing the OM-USB Functional block diagram... 5

Chapter 1 Introducing the OM-USB Functional block diagram... 5 Table of Contents Preface About this User's Guide... 4 What you will learn from this user's guide... 4 Conventions in this user's guide... 4 Where to find more information... 4 Chapter 1 Introducing the

More information

WalkLAB ph mv/orptemperature

WalkLAB ph mv/orptemperature 16,1,16,1,2,15,2,15,14,3,14,3,4,13,4,13,12,5,12,5,6,11,6,11,10,7,10,7,8,9,8,9 WalkLAB ph mv/orptemperature meter HP9000 Operations Manual TABLE OF CONTENTS PAGE 1. INTRODUCTION 1 1.1 Features 1 1.2 Technical

More information

Preamplifiers for Amptek X-Ray Detectors. OEM Preamplifiers for Amptek X-Ray Detectors

Preamplifiers for Amptek X-Ray Detectors. OEM Preamplifiers for Amptek X-Ray Detectors Preamplifiers for Amptek X-Ray Detectors OEM Preamplifiers for Amptek X-Ray Detectors PA-210 PA-230 NOTE: If you are using the Amptek DP5/PC5 with the PA-210/PA-230 you do not need any of the information

More information

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller Operation Manual TX10 1/8 DIN Microcomputer Based /ORP Controller 11751 Markon Drive Garden Grove, CA. 92841 U.S.A. pg. 1 CONTENTS GENERAL INTRODUCTION.....3 INITIAL INSPECTION......3 USING THE Sensorex

More information

MONARCH INSTRUMENT. TACH-4A / TACH-4AR Programmable Optical Tachometer. Instruction Manual Rev 2.0

MONARCH INSTRUMENT. TACH-4A / TACH-4AR Programmable Optical Tachometer. Instruction Manual Rev 2.0 MONARCH INSTRUMENT Instruction Manual TACH-4A / TACH-4AR Programmable Optical Tachometer 1071-4850-001 Rev 2.0 15 Columbia Drive Amherst, NH 03031 E-Mail: Support@MonarchInstrument.com Phone: (603) 883-3390

More information

OVEN INDUSTRIES, INC.

OVEN INDUSTRIES, INC. OVEN INDUSTRIES, INC. OPERATING MANUAL MODEL 5C7-195 THERMOELECTRIC MODULE TEMPERATURE CONTROLLER TABLE OF CONTENTS Page Features...1 Description...2 Block Diagram...2 Mechanical Package Drawing...3 RS232

More information

NFC110. Flow Computer For Low Pulses FEATURES

NFC110. Flow Computer For Low Pulses FEATURES Flow Computer For Low Pulses FEATURES l STABLE READING WITH PULSES DOWN TO ONE PULSE EVERY 5 MINUTES l PULSE FLOW INPUT UP TO 100 Hz, CAN POWER THE SENSOR l WORKS WITH OPTO-COUPLER FLOW METERS l SEVEN

More information

Specifications PMD-1208FS

Specifications PMD-1208FS Document Revision 1.1, May, 2005 Copyright 2005, Measurement Computing Corporation Typical for 25 C unless otherwise specified. Specifications in italic text are guaranteed by design. Analog input section

More information

Programmable Meter for Analog Sensors/Transducers

Programmable Meter for Analog Sensors/Transducers Programmable Meter for Analog Sensors/Transducers Supply Voltage: 24 V ± 20 % Temperature Range: 0 to 70 C Interface: 0 to 10 V or 4 to 20 ma with over-voltage protection Resolution: 16-bit analog to digital

More information

PF2100 MODBUS LOGGER CARD SYSTEM SPECIFICATION. v1.0 DRAFT Revised Dec 4, 2014 Last Revised by Alex Messner

PF2100 MODBUS LOGGER CARD SYSTEM SPECIFICATION. v1.0 DRAFT Revised Dec 4, 2014 Last Revised by Alex Messner PF2100 MODBUS LOGGER CARD SYSTEM SPECIFICATION Revised Last Revised by Alex Messner This page was intentionally left blank. Table of Contents 1 Overview... 2 2 User Interface... 3 2.1 LEDs... 3 2.2 Buttons...

More information

SERIES CMT CARBON MONOXIDE GAS TRANSMITTER

SERIES CMT CARBON MONOXIDE GAS TRANSMITTER SERIES CMT CARBON MONOXIDE GAS TRANSMITTER INSTALLATION OPERATION AND MAINTENANCE MANUAL DWYER INTRUMENTS, INC. PO BOX 373, MICHIGAN CITY, IN. 46360 USA PHONE: 800-872-9141 FAX: 219-872-9057 Web: www.dwyer-inst.com

More information

Digital Thermometer MEM. PC-Link READ

Digital Thermometer MEM. PC-Link READ Digital Thermometer MEM PC-Link READ CONTENTS Digital Thermometer 1. Safety Information...1 2. Product Overview...3 3. Recognize Meter..... 4 3.1 Component...4 3.2 Display...5 3.3 Keys Description...6

More information

MAX197 Evaluation Kit. Evaluates: MAX197/MAX199. Features

MAX197 Evaluation Kit. Evaluates: MAX197/MAX199. Features 9-0398; Rev ; 7/95 MAX97 Evaluation Kit General Description The MAX97 evaluation system (EV system) is a complete, low-cost, 8-channel data-acquisition system consisting of a MAX97 evaluation kit (EV kit)

More information

Operating instructions: STABILA Electronic-IP65

Operating instructions: STABILA Electronic-IP65 TECH Manual Operating instructions: STABILA 196-2 Electronic-IP65 ON/OFF Illumination MODE Hold Key lock sound calibration REF TECH 196E Manual Revision IP65 Code: International Protection Rating. 6: Dust

More information

Atmos Engineering, Inc. External Specification PN PC104 Air Data Atmodule PN Revision 6.0

Atmos Engineering, Inc. External Specification PN PC104 Air Data Atmodule PN Revision 6.0 A T M O S E N G I N E E R I N G I N C. Atmos Engineering, Inc External Specification PN 410042 PC104 Air Data Atmodule PN 840025 Revision 6.0 Atmos Engineering, Inc 443 Dearborn Park, Rd Pescadero, CA

More information

! User Manual Rev. 1

! User Manual Rev. 1 User Manual Rev. 1 Warnings For safety operation of the weighing indicator, please follow the following warning/ safety instructions: Calibration inspection and maintenance of the indicator are prohibited

More information

C600 2-wire Conductivity Transmitter

C600 2-wire Conductivity Transmitter C600 2-wire Conductivity Transmitter Installation and Operating Instructions 900596-C - Sensolution C600 Isolated 2-wire transmitter instruction manual.doc 1 of 24 1. Contents 1. Contents... 3 2. Introduction...

More information

PRO Owner's Manual

PRO Owner's Manual PRO 3600 Owner's Manual Feature Overview The Pro 3600 operates normally in a standard reference mode where level is displayed as 0.00. However, a new reference point for 0.00 can easily be established

More information

Chapter 1 Introducing the OM-USB-1608FS-Plus... 6 Functional block diagram... 6

Chapter 1 Introducing the OM-USB-1608FS-Plus... 6 Functional block diagram... 6 Table of Contents Preface About this User's Guide... 5 What you will learn from this user's guide... 5 Conventions in this user's guide... 5 Where to find more information... 5 Chapter 1 Introducing the

More information

QC-3265 Checkweigher User s Manual

QC-3265 Checkweigher User s Manual QC-3265 Checkweigher User s Manual CAUTION Risk of electrical shock. Do not remove cover. No user serviceable parts inside. Refer servicing to qualified service personnel. Weigh-Tronix reserves the right

More information

Digital Keypad Introduction

Digital Keypad  Introduction K2 Digital Keypad Introduction The K02 uses the latest microprocessor technology to operate door strikes and security systems that require a momentary (timed) or latching dry contact closure. All programming

More information

Energy Management System. Operation and Installation Manual

Energy Management System. Operation and Installation Manual Energy Management System Operation and Installation Manual AA Portable Power Corp 825 S 19 TH Street, Richmond, CA 94804 www.batteryspace.com Table of Contents 1 Introduction 3 2. Packing List 5 3. Specifications

More information