NEMESYS Hardware Manual and Reference

Size: px
Start display at page:

Download "NEMESYS Hardware Manual and Reference"

Transcription

1 NEMESYS Hardware Manual and Reference ORIGINAL INSTRUCTIONS 2.08 AUGUST 2018

2 CETONI GmbH Wiesenring Korbussen Germany T +49 (0) F +49 (0) E info@cetoni.de 2 nemesys Hardware Manual

3 The information and data contained in this document are subject to change without notice. CETONI GmbH is constantly striving to develop all its products. This means that there may be changes in form, equipment and technology. Claims cannot therefore be made on the basis of information, illustration or descriptions in these instructions. The description for the product specification in this manual does not constitute an integral part of the contract. If you control the products with a software from CETONI GmbH, you agree to the applicable license agreement, which can be read in the corresponding software manual. This and all other current product manuals can be found at The reproduction, distribution and utilization of this document as well as the communication of its contents to others without explicit authorization are prohibited. Offenders will be held liable for the payment of damages. We are always open to comments, corrections and requests. Please send them to info@cetoni.de. The general terms and conditions of CETONI GmbH shall apply. Alternative agreements must be in written form. Copyright CETONI GmbH Automation and Microsystems. All rights reserved. nemesys Hardware Manual 3

4 1 Overviews and Directories 1.1 Table of Contents 1 Overviews and Directories Table of Contents Revision History 7 2 Introduction Foreword Symbols and Key Words Used Norms and Guide Lines Application Purpose General Description of the Advice Intended Use Reasonably Foreseeable Faulty Application Safety Advice Measures for Safe Operation Safety Devices on the System Condition of the Devices Warranty and Liability Scope of Supply 13 3 Initial Start-Up Software Installation System Configuration Connecting the Base Module Connecting Additional Modules Module Configuration 18 4 Base Module BASE nemesys Hardware Manual

5 4.1 Technical Data Environment Mechanical Data Electrical Data Interfaces Transport and Storage Maintenance and Care Hardware Operation 21 5 Low Pressure Syringe Pump Technical Data Environment Mechanical Data Electrical Data Interfaces Dosing Performance Valve Transport and Storage Maintenance and Care Hardware Operation Mounting a Syringe Fluidic/Valve 29 6 Mid Pressure Syringe Pump Technical Data Environment Mechanical Data Electrical Data Interfaces Dosing Performance Transport and Storage Maintenance and Care 35 nemesys Hardware Manual 5

6 6.4 Hardware Operation Mounting a Syringe Fluidic/Valve Mounting the Blank Holder 41 7 High Pressure Syringe Pump Technical Data Environment Wetted Parts Mechanical Data Electrical Data Interfaces Dosing Performance Transport and Storage Maintenance and Care Hardware Operation Mounting the Safety Hood Fluidic Connections Mounting a Syringe Pressure Sensor 51 8 Accessory Port 53 9 RS232 Connection Pin Assignment of Module Interfaces OEM RS232 Cable Set RS232 Wiring Communication Settings Pin Assignment of the RS232 Cable Disposal 57 6 nemesys Hardware Manual

7 1.2 Revision History REV DATE CHANGE Manuals merged Adaption of the pin assignment table Correction of pin assignment table Removed cable colors for I/O interface Added safety advices about the Pressure Equipment Directive Correction of the pin assignment of the I/O-Interface for the High Pressure Syringe Pump New Corporate Design Correction of pin assignment table Pictures BASE module and USB cable updated Pressure sensor type in High Pressure Syringe Pump changed nemesys Hardware Manual 7

8 2 Introduction 2.1 Foreword Thank you for deciding to purchase a CETONI product. We would like to support you with this handbook as far as possible in your interaction with the nemesys syringe pump system. We are directly available for any questions or suggestions that you may have. The nemesys syringe pump system may only be taken in operation after carefully reading and understanding this manual. We wish you much success in your work with the nemesys syringe pump system. 2.2 Symbols and Key Words Used The following symbols are used in this manual and are designed to aid your navigation through this document: HINT. Describes practical tips and useful information to facilitate the handling of the software. IMPORTANT. Signifies important hints and other useful information that may not result in potentially dangerous or harmful situations. ATTENTION. Identifies a potentially harmful situation. Failure to avert this situation may result in damage to the product or anything in its proximity. CAUTION. Indicates a potentially dangerous situation. Failure to avert this situation may result in light or minor injuries or property damage. 8 nemesys Hardware Manual

9 2.3 Norms and Guide Lines CETONI GmbH declares under its sole responsibility, that the individual nemesys devices and the entire nemesys syringe pump system comply with the health and safety requirements of the relevant European directives. 2.4 Application Purpose General Description of the Advice The nemesys devices are syringe pumps. They allow emptying and filling syringes by the relative linear movement of a syringe- and a piston holder Intended Use The nemesys syringe pump system serves for precise and pulsation-free dosing of fluids in the range of nanolitres per second up to millilitres per second. Pressures of up to several hundred bar can be reached depending on the device. Application usually takes place in laboratory-like rooms Reasonably Foreseeable Faulty Application A use for applications distinct from the intended purpose can lead to dangerous situations and is to be omitted. CAUTION. The unit must not be used as a medical device or for medical purposes. nemesys Hardware Manual 9

10 2.4.4 Safety Advice The safety of the user and a failure-free operation of the devices are assured only if original parts are used. Only original accessories may be used. Warranty claims will not be accepted for damage due to the use of alien accessories or expendables. The devices have been developed and constructed in such a way as to largely rule out hazards due to its intended use. Nevertheless, you must observe the following security measures in order to exclude any remaining hazards: CETONI GmbH points out the responsibilities of the operator for the operation of the devices. The laws and regulations of the place of installation must be observed while operating the devices! To ensure a safe work routine, operators and users must assume responsibility for adhering to regulations. The devices must not be used as a medical device or for medical purposes. Before operating the unit, the user must at all times ensure the operational reliability and the adequate and orderly condition of the unit. The user must be familiar with the operation of the devices and the software. The devices and pipes must be checked for damage before operation. Damaged pipes and plug devices must be replaced immediately. Cables must be laid in a way that avoids any risk of stumbling. Any moving parts must not be touched whilst the devices are in operation. There is a risk of crushing! It is not allowed to use the devices in an explosive atmosphere or with potentially explosive substances. The device is designed and approved to work in fluidic systems, which fall within the scope of Article 4 Paragraph 3 of the Pressure Equipment Directive 2014/68/EU. This means that the system may not exceed a maximum volume of 1 liter. With the use of fluids from Group 1 according to Article 13 of the Pressure Equipment Directive 2014/68/EU, the maximum allowable system pressure is 200 bar. For fluids from Group 2 it is 1000 bar. If different, product-specific values for the maximum pressure are given in the section "Technical Data", these values must be complied with. Regarding the maximum operating temperature, the specification from the section "Technical Data" must be observed. CETONI GmbH is not liable for consequences that may arise if the user expands the system by peripheral devices, such that one of the values or both values are exceeded. 10 nemesys Hardware Manual

11 It is the user's responsibility to become familiar with the mentioned Pressure Equipment Directive and to comply with the prevailing requirements. Wear protective glasses if you are working with corrosive, hot or otherwise dangerous substances during assembly work on the device. Transportation, storage or operation of the devices below 0 C with water in the fluid passages may cause damage to the modules Measures for Safe Operation ELECTROMAGNETIC EMISSIONS The Qmix system is intended for use in any type of facility, connected directly to the public power supply network that supplies buildings used for domestic purposes ELECTROSTATIC DISCHARGE Floors should be made of wood, concrete, or ceramic tiles. If the flooring is made of a synthetic material; the relative humidity must be at least 30% ELECTRIC DISTURBANCES The quality of the supply voltage should be to the standard of a typical business or hospital environment MAGNETIC DISTURBANCES Do not place power connector cables, even of other appliances, in close proximity of the devices and their cables. Mobile communication devices may not be used in closer proximity of the devices or their cables than the recommended safety distance! Safety Devices on the System The system can be switched off at any time in an emergency using the mains switch on the Base Module (rocker switch on the side of the housing); this will cause no damage to the unit Condition of the Devices Irrespective of the faultless manufacture of the devices, damage can occur whilst the unit is in operation. With this in mind, always carry out a visual check of the components mentioned before use. Pay particular attention to crushed cables, damaged tubing, and deformed plugs. If you should notice any damage, please do not use the devices and inform CETONI GmbH without delay. CETONI will put nemesys Hardware Manual 11

12 your devices back to an operational condition at the earliest. Do not attempt to repair the devices yourself. 2.5 Warranty and Liability The devices left our company in perfect condition. Only the manufacturer is permitted to open the devices. All guarantee and liability entitlements, particularly damage entitlements due to personal injuries, are void if the devices are opened by an unauthorized person. The duration of the warranty is 1 year from the day of delivery. It is not extended or renewed due to work carried out under warranty. CETONI GmbH considers itself responsible for the devices with regard to safety, reliability and function only if assembly, new settings, changes, extensions and repairs are carried out by CETONI GmbH or an authorized centre, and if the devices have been used in accordance with the instruction manual. The nemesys syringe pump system conforms to the basic safety regulation standards. Industrial property rights are reserved on the circuits, methods, names, software programs, and units. 12 nemesys Hardware Manual

13 2.6 Scope of Supply The delivery should correspond to the order. The following items should be included in the scope of supply of the base module: COUNTRY SPECIFIC POWER CABLE FOR NON- HEATING APPARATUS USB CABLE 10 m ACTIVE BUS TERMINATING PLUG HARDWARE-MANUAL AND CD WITH SOFTWARE- MANUAL, SOFTWARE AND DRIVERS nemesys Hardware Manual 13

14 3 Initial Start-Up IMPORTANT. Please carefully read this manual and the associated software manual in their entirety before starting up your nemesys syringe pump system. To ensure failure-free initial start-up of your syringe pump system, the following sequence must be adhered to. The individual steps are described in more detail in the sections below. Software Installation Before initially connecting the devices to a PC System Configuration Adding modules to the system one by one Module Configuration Setting syringe size, pressure limits 3.1 Software Installation Before connecting the system, the supplied software and drivers must be installed. The procedure is described in the associated software manual, which you can find on the supplied CD or USB stick. IMPORTANT. Install the software and device drivers as described in the software manual, before connecting your device to a PC via USB. 3.2 System Configuration Place your nemesys modules on a firm and level surface in the desired sequence, without connecting them. You can arrange the system horizontally or vertically. 14 nemesys Hardware Manual

15 ATTENTION. In vertical arrangement the assembly will be less stable. Please take precautions to avoid tipping and place the devices at least 40 cm from the edge of the table Connecting the Base Module After installing the software and device driver, connect the USB connector of the base module (USB type B) with a free USB connector on your PC (USB type A). Plug the base module into a power outlet using the supplied power cable. The device can be connected to AC power sources ranging from 90 to 264 volts and 47 to 63 Hz. USB-socket (type B) Fuse Power socket Power switch CAUTION. Danger of injury due to damaged power lines and plugs! Check the device and power lines for damage! Never operate the device with damaged power lines or plugs! Only use supplied cables. CAUTION. Danger of stumbling due to connecting cables! Place cables in such way as to avoid any danger of stumbling! nemesys Hardware Manual 15

16 Flip the power switch to turn on the device. The on/off switch should light up while the device is turned on. If this is not the case, make sure that the power cord is firmly connected to the device and the power outlet Connecting Additional Modules The modules must be configured individually by the software before they can be used. The procedure is slightly different depending on whether you use the nemesys UserInterface or QmixElements. The exact process is described in the respective software manuals and will be touched on briefly here: In case of the nemesys UserInterface the modules are added to the base module one by one, before being configured. According to this sequence, each module is assigned a number internally. That is the reason devices have to be reset through Remove Dosing Unit before they can be used in a different system. This also resets the assigned number. In QmixElements every module is plugged into the base module, configured and unplugged individually. Once all modules are configured, the system can be assembled. Following, we will describe the mechanical connection of additional modules. To ensure correct function of the overall system, please read and observe the respective section in the associated software manual, before connecting additional modules. IMPORTANT. Please read and observe the respective section of the associated software manual before connecting additional devices. 16 nemesys Hardware Manual

17 Place the module you would like to connect next to your base module or your existing system in such way that the centering pins of the final system module face the centering holes of the module to be connected. Plug the new module into the system. The centering pins are inserted into the respective centering holes and the plug connectors are connected. To ensure clean contact between the modules, both modules must make physical contact across their entire surface. Avoid canting the modules. Connecting additional modules Insert the bus termination plug into the last connected module of your system. Make sure that this plug is always plugged in before the system is turned on. Otherwise, there may be disruptions in data communication. IMPORTANT. Always insert the bus termination plug into the socket of the final connected system. Otherwise there may be disruptions in data communication. nemesys Hardware Manual 17

18 Connecting Terminator Plug IMPORTANT. Deactivate the standby/sleep mode of your PC to avoid malfunctions. 3.3 Module Configuration Among other things, module configuration includes inserting syringes, using valves and connecting periphery devices. All necessary information with respect to technical data and hardware operation as well as configuration of the various models can be found in the following module-specific sections of this manual. If software actions become necessary, the manual will point you to the software manual at the appropriate time. There you will also find additional and helpful information. 18 nemesys Hardware Manual

19 4 Base Module BASE Technical Data Environment OPERATING TEMPERATURE 0 C to 50 C STORAGE TEMPERATURE -20 C to 75 C OPERATING AIR HUMIDI TY 20% to 90%, non-condensing STORAGE AIR HUMIDITY 20% to 90%, non-condensing Mechanical Data DIMENSIONS (L X W X H) WEIGHT 310 x 100 x 56 mm 1800 g nemesys Hardware Manual 19

20 4.1.3 Electrical Data SUPPLY VOLTAGE FREQUEN CY 90 to 260 VAC 47 to 63 Hz POWER OU TPUT 24 VDC, 5A, 120 W Interfaces USB 1.1 and Transport and Storage Please do not lift or transport the modules while they are plugged into each other. Transport in assembled state is only permissible when using the original packaging. Please refer to section for information about storage. ATTENTION. Danger of damaging the device! Do not transport modules while they are plugged into each other. 4.3 Maintenance and Care When used properly, the device is maintenance-free. In case of problems that you cannot fix yourself or that require opening the device, please contact CETONI GmbH to coordinate any further actions. The device may be opened only by CETONI GmbH or authorized service personnel. Failure to adhere to this rule will void the warranty. The software manual includes detailed information about malfunctions with respect to the operating software. Wipe the device with a moist (not wet) cloth in such way that no liquids get into the inside. In case of heavy soiling you may use some detergent or alcohol. 20 nemesys Hardware Manual

21 4.4 Hardware Operation The base module supplies all other modules with power and serves as an interface to your PC. The BASE 120 base module can continuously supply a power of 120 W. Under the technical data electrical data section you can find the power consumption of every module you may want to connect. Please make sure that the total power consumption of the assembly must not exceed the power supply capability of the base module. In case of overload the base module will deactivate itself and all connected devices and you will lose control of your application. Should the power supply capacity of the BASE 120 base module be insufficient, you may also use the BASE 600 base module or the BASE 600XT extension. ATTENTION. Danger of sudden deactivation! Make sure that the total power consumption of all connected devices does not exceed the power supply capacity of the base module. On one side of the device there is a power connector with a fuse box. A replacement fuse (Ø5x20 mm) can be found in the same box. The rated current of the fuse may vary depending on the country of use and can be seen on the type plate. The power switch is located to the right of the power socket and will light up red when the device is turned on. The type B USB-connector is used to connect the device to your PC. USB connector (type B) Fuse Power socket Power switch nemesys Hardware Manual 21

22 5 Low Pressure Syringe Pump 5.1 Technical Data Environment OPERATING TEMPERATU RE 0 C to 45 C STORAGE TEMPERATURE -20 C to 75 C OPERATION AIR HUMIDITY 20% to 90%, non-condensing STORAGE AIR HUMIDITY 20% to 90%, non-condensing Mechanical Data DIMENSIONS (L X W X H) WEIGHT 310 x 47 x 56 mm 1300 g 22 nemesys Hardware Manual

23 5.1.3 Electrical Data SUPPLY VOLTAGE CURRENT DRAW 24 VDC 0,3 A POWER CONSUMPTION 7 W Interfaces CAN 1 Mbit/s RS-232 section 9 ACCESSORY PORT section Dosing Performance The following table provides an overview of minimum and maximum dosing speeds of the various gear configurations as well as the resulting flow rates, using the example of a 1 ml syringe with a 60 mm stroke. Dosing precision slowly decreases below the speeds and flow rates referred to as pulsation-free. Gear w/o STANDARD Special Min. speed [µm/s] 0,065 0,065 0,065 Min. pulsation free speed [µm/s] 14,648 1,042 0,502 Max. speed [mm/s] 89 6,33 3,06 Min. flow [µl/min] 0,065 0,065 0,065 1 ml syringe with 60mm stroke Min. pulsation-free flow [µl/min] 14,648 1,042 0,502 Max. flow [ml/min] 89 6,33 3,05 nemesys Hardware Manual 23

24 5.1.6 Valve TECHNICAL DATA HOUSING MATERIAL SEALING MATERIAL PEEK FFKM (perfluoroelastomer) MEDIA TEMPERATURE 0 to +50 C MAX. VISCOSITY 21 mm²/s INTERNAL VOLUME approx. 45 µl MAX. PRESSURE 3 bar NOMINAL SIZE 0,6 mm FLUIDIC CONNECTIONS ¼ 28 UNF 5.2 Transport and Storage Please do not lift or transport the modules while they are plugged into each other. Transport in assembled state is only permissible when using the original packaging. Use the original packaging for shipping the modules. Please observe the information provided in with respect to storage. 24 nemesys Hardware Manual

25 ATTENTION. Danger of damaging the device! Never transport modules while they are plugged into each other. 5.3 Maintenance and Care When used properly, the device is maintenance-free. In case of problems that you cannot fix yourself or that require opening the device, please contact CETONI GmbH to coordinate any further actions. The device may be opened only by CETONI GmbH or authorized service personnel. Failure to adhere to this rule will void the warranty. The software manual includes detailed information about malfunctions with respect to the operating software. Wipe the device with a moist (not wet) cloth in such way that no liquids get into the inside. In case of heavy soiling you may use some detergent or alcohol. 5.4 Hardware Operation Connect the Low Pressure Syringe Pump to your base module / system as described in section and the software manual. If the module has not been configured, you will be required to perform a reference move during the configuration process. During the reference move the piston holder will move to its front position and be synchronized with the software display. To avoid damage the reference move may only be performed without a syringe. After deactivating the base module the piston holder can be moved by applying some force (e.g. through residual pressure in the system). Therefore, it is sensible to repeat the reference run from time to time. ATTENTION. The reference move must be performed without syringe. Otherwise the device or the syringe may be damaged. CAUTION. Do not touch any moving parts on the device during operation! There is a danger of crushing. nemesys Hardware Manual 25

26 5.4.1 Mounting a Syringe The syringe holder of Low Pressure Syringe Pump allows the use of syringes with outside diameters ranging from 6 to 30 mm and a piston stroke of up to 65 mm. IMPORTANT. Please use high-quality glass syringes with outside diameters ranging from 6 to 30 mm to ensure precise flow rates. Before mounting a syringe to the Low Pressure Syringe Pump, it must be configured and selected in the operating software. The respective process is described in the software manual. You need the volume (scale volume), the nominal stroke (scale length) and the total stroke (piston stroke), which may be different. Scale Length Scale Volume Piston Stroke 26 nemesys Hardware Manual

27 Use the following process to mount a syringe to the Low Pressure Syringe Pump: Loosen the knurled screw of the syringe receptacle. Now you can push the bracket off the pins and lift it up. Disassembly syringe holder Loosen the knurled screw of the piston holder and remove the adapter plate. Lift up the clamping piece. Disassembly piston holder In order to make use of the entire syringe volume, move the piston holder to the front position through the software. Place an empty syringe on the remaining base of the syringe holder in such way that the piston touches the piston holder. The syringe position can be varied somewhat by moving the piston holder. For this purpose loosen the screw with a 4 mm Allen wrench. nemesys Hardware Manual 27

28 ATTENTION. To avoid damage, please make sure that the remaining syringe stroke is always the same as or larger than that of the module. Screw Replace the bracket of the syringe holder, insert into the pins and lock the syringe by using the large knurled screw. Replace the clamping piece and insert the adapter plate suitable for the piston diameter in such way that the piston plate is located between piston holder and adapter plate. The piston is locked in place by slightly tightening the knurled screw on the back. Make sure that syringe and syringe piston are in alignment. 28 nemesys Hardware Manual

29 IMPORTANT. The syringes, particularly the seals, are wear parts. Check them on a regular basis and replace them, if necessary Fluidic/Valve The Low Pressure Syringe Pump can be fitted with an optional valve. The valve allows you to switch the syringe connection between your application (outlet) and a reservoir (intake), allowing an automatic refilling of the syringe. Through the software, you can set the valve in such way that it automatically switches to the intake during filling. The three connections are fitted with a ¼ -28 UNF thread and therefore allow the use of common HPLC fittings. Outlet (normally open) Intake (normally closed) Syringe connection A rocker inside the valve (green) connects the syringe port (1) alternatively with the outlet (2) or the intake (3). In the following image the syringe is connected with the outlet (blue), while the intake is closed (red). A FFKM membrane connected to the rocker (shaded in green) seals the fluidic system. This membrane limits the operating pressure of the valve to 3 bar. nemesys Hardware Manual 29

30 ATTENTION. To avoid damaging the valve, please observe the maximum operating pressure of 3 bar. ATTENTION. Before using the valve, please check the chemical resistance of the PEEK housing material and the FFKM sealing material (perfluoroelastomer) against the dosing liquid VALVE INSTALLATION AND REMOVAL The valve can be removed from the device in a few simple steps. This simplifies the fitting of tubes as well as cleaning. It also allows operation without a valve, in cases where it is not needed. To remove the valve press the rocking catch at the plug and remove the plug. When replacing it later please observe the orientation of the plug (the white surface must face the valve, the rocking catch must face away from the valve). 30 nemesys Hardware Manual

31 ATTENTION. Observe the orientation of the valve plug (the white surface must face the valve, the rocking catch must face away from the valve). Rocking Catch In the next step you can simply pull up the valve to remove it. nemesys Hardware Manual 31

32 6 Mid Pressure Syringe Pump 6.1 Technical Data Environment OPERATING TEMPERATUR E 0 C to 45 C STORAGE TEMPERATURE -20 C to 75 C OPERATING AIT HUMIDITY 20% to 90%, non-condensing STORAGE AIR HUMIDITY 20% to 90%, non-condensing Mechanical Data DIMENSIONS (L X W X H) WEIGHT 310 x 59 x 60 mm 2000 g 32 nemesys Hardware Manual

33 6.1.3 Electrical Data SUPPLY VOLTAGE CURRENT DRAIN 24 VDC 0,3 A POWER CONSUMPTION 15 W Interfaces CAN 1 Mbit/s RS-232 section 9 ACCESSORY PORT section 8 nemesys Hardware Manual 33

34 6.1.5 Dosing Performance The following table provides an overview of minimum and maximum dosing speeds. The resulting flow rates shown here are based on the example of a 1 ml glass syringe with a 60 mm stroke as well as CETONI stainless steel syringes. Dosing precision slowly decreases below the speeds and flow rates referred to as pulsation-free. The table also indicates the maximum pressure. In case of a glass syringe, the syringe itself is the limiting element. In case of stainless steel syringes it is the force of the module. You need the nominal and maximum stroke for the software configuration of syringes. Please read the relevant sections in the software and syringe manual. SPEEDS MIN [µm/s] MIN PULSATION-FREE [µm/s] MAX [MM/S] 0,065 1,005 6,00 SYRIN GE NOMINAL STROKE/ MAX. STROKE [MM]. MAX. PRESSURE MIN [µl/min] FLOW RATES MIN PULSATION-FREE [µl/min] MAX [ML/S] 1 ML-ILS GLASS SYRIN GE bar 530 psi 0,065 1,005 0,10 2,5 ML 49, ML 52, ML 49, ML 50, ML 51, bar 2900 psi 100 bar 1500 psi 50 bar 720 psi 20 bar 290 psi 10 bar 140 psi 0,196 3,015 0,30 0,370 5,703 0,57 0,784 12,083 1,20 1,923 29,631 2,95 3,768 58,064 5,78 34 nemesys Hardware Manual

35 6.2 Transport and Storage Please do not lift or transport the modules while they are plugged into each other. Transport in assembled state is only permissible when using the original packaging. Use the original packaging for shipping the modules. Please observe the information provided in with respect to storage. ATTENTION. Danger of damaging the device. Do not transport modules while they are plugged into each other! 6.3 Maintenance and Care When used properly, the device is maintenance-free. In case of problems that you cannot fix yourself or that require opening the device, please contact CETONI GmbH to coordinate any further actions. The device may be opened only by CETONI GmbH or authorized service personnel. Failure to adhere to this rule will void the warranty. The software manual includes detailed information about malfunctions with respect to the operating software. Wipe the device with a moist (not wet) cloth in such way that no liquids get into the inside. In case of heavy soiling you may use some detergent or alcohol. 6.4 Hardware Operation Connect the Mid Pressure Syringe Pump to your base module / system as described in section and the software manual. If the module has not been configured, you will be required to perform a reference move during the configuration process. During the reference move the piston holder will move to its front position and be synchronized with the software display. To avoid damage the reference move may only be performed without a syringe. After deactivating the base module the piston holder can be moved by applying some force (e.g. through residual pressure in the system). Therefore, it is sensible to repeat the reference run from time to time. nemesys Hardware Manual 35

36 ATTENTION. The reference move must be performed without a syringe. Otherwise the device or the syringe may be damaged. CAUTION. Do not touch moving parts during operation! There is a danger of crushing Mounting a Syringe The syringe holder of Mid Pressure Syringe Pump allows the use of syringes with a ¼ -28 UNF outside thread at the tip and an M3 thread in the piston. The maximum diameter is 50 mm, the maximum syringe stroke 65 mm. ATTENTION. On the Mid Pressure Syringe Pump please only use syringes offered by CETONI GmbH for this device. When using glass syringes, please make sure not to exceed their maximum operating pressure. Before mounting a syringe to the Mid Pressure Syringe Pump, it must be configured and selected in the operating software. The respective process is described in the software manual. You will need the volume (scale volume), the nominal stroke (scale length) and the maximum stroke (piston stroke), which may be different. The scale volume and scale length values are engraved on CETONI stainless steel syringes. Unless otherwise stated, the piston stroke is 60 mm. These values can also be found in the table in section For more information with respect to syringes, please refer to the syringe manual. 36 nemesys Hardware Manual

37 Scale Volume Scale Length Please follow these steps to mount a syringe on the Mid Pressure Syringe Pump: Loosen the screws on the syringe holder and push it all the way to the front, or completely remove it from the device. Loosen the knurled screws of the holding-down clamp, if installed. Knurled screws Fasting screw nemesys Hardware Manual 37

38 Screw the syringe into the syringe holder. Use one of the supplied PTFE disks to achieve a sealed connection. PTFE disk In order to make use of the entire syringe volume, move the piston holder to the front position through the software. Push back the syringe holder until the piston touches the piston holder. Retighten the fastening screws of the syringe holder. Lock the syringe piston with the knurled screw. To avoid bending the syringe in case of high pressure, lower the blank holder onto the syringe and fasten the two knurled screws. 38 nemesys Hardware Manual

39 Blank holder Knurled screw IMPORTANT. Syringes, and particularly the seals, are wear parts. Check them on a regular basis and replace them if necessary. ATTENTION. To avoid damage, please make sure that the remaining syringe stroke is always larger than or the same as that of the module. ATTENTION. Before use, please check the resistance of syringe seals against the dosing medium. Use a different material, if necessary Fluidic/Valve The syringe holder of the Mid Pressure Syringe Pump offers two ¼ -28 UNF threaded connections on the output side, allowing the use of common HPLC connectors. The Mid Pressure Syringe Pump can be fitted with an optional pair of non-return valves. They allow automatic switching of the syringe connector between your application (outlet) and a reservoir (intake) and thereby automatic refilling of the syringe. Please refer to the associated valve manual for more detailed information about valves. The valves are simply plugged onto the corresponding pins on the module. nemesys Hardware Manual 39

40 Connect tubes between the syringe holder and the valves as follows: Intake Outlet ATTENTION. Before using the valve please check its chemical resistance against the media you intend to pump. 40 nemesys Hardware Manual

41 6.4.3 Mounting the Blank Holder The blank holder prevents the syringe from bending upward, which is an effect that occurs in case of high pressure. As a consequence syringe wear is decreased as well. To mount the blank holder simply place it into the two gaps on the device and fasten it on both sides with the provided screws (2 mm Allen head). nemesys Hardware Manual 41

42 7 High Pressure Syringe Pump 7.1 Technical Data Environment OPERATING TEMPERATURE 0 C to 45 C STORAGE TEMPERATURE -20 C to 75 C OPERATING AIR HUMIDI TY 20% to 90%, non-condensing STORAGE AIR HUMIDITY 20% to 90%, non-condensing Wetted Parts SYRIN GE HOLDER AND FITTING PRESSU RE SENSOR Stainless steel (316L) Aluminium oxide (Al2O3) PRESSU RE SENSOR SEAL Perfluoroelastomer FFKM 42 nemesys Hardware Manual

43 7.1.3 Mechanical Data DIMENSIONS (L X W X H) WEIGHT 310 x 110 x 136 mm 4500 g Electrical Data SUPPLY VOLTAGE CURRENT DRAIN 24 VDC 1,9 A POWER CONSUMPTION 45 W Interfaces CAN 1 Mbit/s RS-232 section 9 ACCESSORY PORT section 8 nemesys Hardware Manual 43

44 7.1.6 Dosing Performance The following table provides an overview of minimum and maximum dosing speeds. The resulting flow rates shown here are based on CETONI stainless steel syringes. Dosing precision slowly decreases below the speeds and flow rates referred to as pulsation-free. The table also indicates the maximum pressure that can be achieved with the High Pressure Syringe Pump in combination with the respective syringe. You need the nominal stroke and the maximum stroke for the software configuration of syringes. Please read the relevant sections in the software and syringe manual. SPEED MIN [µm/s] MIN PULSATION-FREE [µm/s] MAX [MM/S] 0,061 0,636 7,00 SYRIN GE NOMINAL / MAX STROKE [MM]. MAX. PRESSURE MIN [µl/min] FLOW RATES MIN PULSATION-FREE [µl/min] MAX [ML/S] 2,5 ML 49, ML 52, ML 49, ML 50, ML 51, ML 50, bar 7400 psi 270 bar 3900 psi 120 bar 1740 psi 50 bar 360 psi 25 bar 360 psi 12 bar 175 psi 0,182 1,908 0,35 0,344 3,609 0,66 0,728 7,646 1,40 1,785 18,751 3,44 3,499 36,743 6,74 7,137 74,960 13,76 44 nemesys Hardware Manual

45 7.2 Transport and Storage Please do not lift or transport the modules while they are plugged into each other. Transport in assembled state is only permissible when using the original packaging. Use the original packaging for shipping the modules. Please observe the information provided in with respect to storage. ATTENTION. Danger of damaging the device. Do not transport modules while they are plugged into each other. 7.3 Maintenance and Care When used properly, the device is maintenance-free. In case of problems that you cannot fix yourself or that require opening the device, please contact CETONI GmbH to coordinate any further actions. The device may be opened only by CETONI GmbH or authorized service personnel. Failure to adhere to this rule will void the warranty. The software manual includes detailed information about malfunctions with respect to the operating software. Wipe the device with a moist (not wet) cloth in such way that no liquids get into the inside. In case of heavy soiling you may use some detergent or alcohol. 7.4 Hardware Operation Connect the High Pressure Syringe Pump to your base module / system as described in section and the software manual. The High Pressure Syringe Pump only works with an installed safety hood and may only be operated with it in place. Please refer to section for more information about using the safety hood. If the module has not been configured, you will be required to perform a reference move during the configuration process. During the reference move the piston holder will move to its front position and be synchronized with the software display. To avoid damage the reference move may only be performed without a syringe. nemesys Hardware Manual 45

46 After deactivating the base module the piston holder can be moved by applying some force (e.g. through residual pressure in the system). Therefore, it is sensible to repeat the reference run from time to time. ATTENTION. The reference move must be performed without a syringe. Otherwise the device or the syringe may be damaged Mounting the Safety Hood Place the safety hood on the device in such way that the pins slide into the gaps provided for them (blue in image). Push down on the opposite end of the hood until the spring-loaded balls (also shown in blue) settle into the holes provided for them in the hood. To remove the hood, simply reverse this process. Pins Balls CAUTION. Only use the High Pressure Syringe Pump with the mounted safety hood! Do not touch any moving parts on the device during operation! IMPORTANT. If you remove the safety hood during operation, the High Pressure Syringe Pump will stop automatically. 46 nemesys Hardware Manual

47 7.4.2 Fluidic Connections A Swagelok tube fitting is used for fluidic connection. It is suitable for using capillaries made from metal (e.g. stainless steel, titanium) and plastic (e.g. PTFE, PEEK). Please refer to information provided by the respective manufacturer with respect to maximum pressure. The High Pressure Syringe Pump is supplied with two tube fittings for capillaries with an external diameter of 1/16 and 1/18. Following is a description of using the tube fittings: FIRST-TIME INSTALLATION (1) Fully insert the tube / hose into the fitting and against the shoulder; rotate the nut finger-tight. (2) Mark the nut at the 6 o clock position. (3) Tighten the nut three-quarters turn to the 3 o clock position with an open-end wrench DISMANTLING ATTENTION. Release pressure from the system before loosening the fittings. (1) Before dismantling, draw a marker line across the nut and the fitting body. In this way you create a reference for retightening the cap nut to exactly the same position it was in before. (2) Pull out the capillaries. The nut and the ferrules remain on the capillary REASSEMBLY (1) To reassemble, insert the capillary with preassembled ferrules into the fitting body until the front ferrule seats against the fitting body. (2) Rotate the nut with open-end wrench to the previously pulled-up position as indicated by the marks you made before; at this point you will feel a significant increase in resistance. (3) Retighten the nut slightly. Done! You can purchase additional connecting material from Swagelok, such as replacement clamping rings (ordering number SS-100-SET). nemesys Hardware Manual 47

48 IMPORTANT. Only use capillaries specified for the anticipated pressure levels. The supplied 1/16 PEEK tubes are suitable for pressures up to 200 bar. ATTENTION. After connecting, check the tightness of all fluidic connections on a regular basis REPLACING THE CONDUIT FITTING The tube fitting is screwed into the syringe holder and can be loosened or tightened with a 9/16 open end wrench. For this purpose, place the wrench onto the large hexagon nut (not the small one of the cap nut)! There is a metal bushing behind the conduit fitting for sealing purposes. This bushing can remain in the borehole while the conduit fitting is replaced. Firmly tighten the newly mounted conduit fitting to achieve a sealed connection. 48 nemesys Hardware Manual

49 7.4.3 Mounting a Syringe Before mounting a syringe to the High Pressure Syringe Pump, it must be configured and selected in the operating software. The respective process is described in the software manual. You will need the volume (scale volume), the nominal stroke (scale length) and the maximum stroke (piston stroke), which may be different. The scale volume and scale length values are engraved on CETONI stainless steel syringes. Unless otherwise stated, the piston stroke is 60 mm. These values can also be found in the table in section Additional information can be found in the syringe manual. Scale Volume Scale Length Follow these steps to mount a syringe on the High Pressure Syringe Pump: To be able to screw the syringe into the mounting bore while establishing a sealed connection, please insert one of the provided O-rings (3 mm inside diameter and 1.5 mm cord thickness) into the indentation at the syringe outlet (blue in image). nemesys Hardware Manual 49

50 O-ring Position the piston holder in such way as to make sufficient space to screw in the syringe. Remove the safety hood as described in section Screw the syringe into the syringe holder as far as it will go. Put the safety hood back on the device and advance the piston holder until it touches the piston. Take off the safety hood and attach the piston to the piston holder using the knurled screw. 50 nemesys Hardware Manual

51 After putting the safety hood back on, you are ready for dosing. ATTENTION. Before operation, check the resistance of the syringe seal against the dosing medium. If necessary, replace the seal with a seal made from a different material. IMPORTANT. Syringes, and particularly seals, are wear parts. Check them on a regular basis and replace them if necessary Pressure Sensor There is an integrated pressure sensor in the syringe holder of the High Pressure Syringe Pump. This allows the device to stop automatically upon reaching the preset maximum pressure. ATTENTION. Before using the device, check the chemical resistance of wetted materials against the fluid to be metered. nemesys Hardware Manual 51

52 The pressure sensor must be configured in the software before use. The associated process is described in the software manual. For the configuration you will need the pressure range of the sensor (e.g bar) and the range of the output signal (e.g V). These values can be found on the nameplate on the side of the device under sensor: ATTENTION. Configure the pressure sensor before use, to avoid damaging the device or your application. 52 nemesys Hardware Manual

53 8 Accessory Port The Low, Mid and High Pressure Syringe Pumps, are equipped with an accessory port or can be equipped with it as an optional extra. The additional port allows the use of a pressure sensor, for example. The pin assignment of the connector at the module and the wire colors of the connecting cable, which can be purchased from CETONI, can be found in the table on the next page. Of course, you can also purchase ready-made periphery devices from CETONI GmbH. A matching connector plug is also available from Hirose (order number HR10A-10P-12P(73)). The configuration of pressure sensors is described in the software manual. Read and observe the relevant section before connecting a pressure sensor. IMPORTANT. Pin 1 is not assigned at the accessory port of the High Pressure Syringe Pump, because analogue input AI1 is already used for the built-in pressure sensor. nemesys Hardware Manual 53

54 PIN SIGNAL DESCRIPTION 1 Analog input AI1 0-5 V (to Pin 12) 2 Analog input AI2 0-5 V (to Pin 12) 3 Digital input 1 <0,8 V Low >2 V High 24 V max. 4 Digital input 2 <0,8 V Low >2 V High 24 V max. 5 Digital input 3 <1,7 V Low >4,2 V High 24 V max. 6 Digital output 1 Valve voltage NPN Max. 1 A Active: 0 V (GND) Inactive: open 7 Digital output 2 Switch valve NPN Max. 1 A Active: 0 V (GND) Inactive: open 8 Digital output 3 NPN Max. 1 A Active: 0 V (GND) Inactive: open 9 Digital ground V Out +24 VDC / <1 A V Out +5 VDC / <150 ma 12 Analog ground 54 nemesys Hardware Manual

55 9 RS232 Connection 9.1 Pin Assignment of Module Interfaces PLUG SOCKET A1 1 2 A2 A2 2 1 A PIN 1 Not connected RS232 RX 2 Not connected RS232 TX 3 CAN High CAN High 4 CAN Low CAN Low 5 Signal GND Signal GND A1 +24 V +24 V A2 GND GND 9.2 OEM RS232 Cable Set RS232 Wiring Insert the mixed D-Sub plug of the cable into the socket of the final module. The system should be deactivated when you do this. Tighten both screws on the plug manually. You do not need a bus termination plug, since the plug of the RS232 cable already contains a bus termination resistor. Now, plug the 9-pin D-Sub socket of the cable into an RS232 connection on your PC or other controller. For greater distances to the socket please use a 1:1 cable with a 9-pin D-Sub plug. nemesys Hardware Manual 55

56 Now, you can reactivate your system and send or receive data through RS232. Since every module contains a gateway from RS232 to the system s internal CAN bus, you can now address each module of your system with only one RS232 cable Communication Settings For a functioning communication with the nemesys modules you have to make the following communication settings for the serial interface on your PC or other controller: Baud rate: Data bit rate: 8 Parity: none Stop bits: 1 Flow control: none Pin Assignment of the RS232 Cable The OEM RS232 cable adapts the nemesys device interface to a standard 9-pin D-Sub plug. The following table shows the pin assignment of the nemesys interface and the 9-pin D-Sub: nemesys Interface (socket) 9-pin D-Sub socket A2 2 1 A Pin Cable Pin 1 RS232 RX Orange 3 TXD Transmit Data 2 RS232 TX Brown 2 RXD Receive Data 5 Signal GND Black 5 GND Signal GND D-Sub housing = shielding Shielding 56 nemesys Hardware Manual

57 10 Disposal Please send your old devices back to CETONI GmbH. We will take care of proper disposal according to electric devices regulations. If necessary, please decontaminate the device before sending it back and attach a completed decontamination declaration with your shipment. nemesys Hardware Manual 57

NEMESYS ULTRA HIGH PRESSURE Hardware Manual

NEMESYS ULTRA HIGH PRESSURE Hardware Manual NEMESYS ULTRA HIGH PRESSURE Hardware Manual ORIGINAL MANUAL 2.02 APRIL 2016 CETONI GmbH Wiesenring 6 07554 Korbussen Germany T +49 (0) 36602 338-0 F +49 (0) 36602 338-11 E info@cetoni.de www.cetoni.de

More information

December OEM Module. Hardware Reference

December OEM Module. Hardware Reference December 2008 OEM Module Hardware Reference The information and data contained in this documentation can be amended without notice. No part of this documentation can be duplicated or transferred for any

More information

NEMESYS OEM Hardware Manual

NEMESYS OEM Hardware Manual NEMESYS OEM Hardware Manual ORIGINAL INSTRUCTIONS 2.04 MAY 2018 CETONI GmbH Wiesenring 6 07554 Korbussen Germany T +49 (0) 36602 338-0 F +49 (0) 36602 338-11 E info@cetoni.de www.cetoni.de 2 nemesys OEM

More information

SKIVING MACHINE MAINTENANCE & INSTRUCTIONS MANUAL ASSK2. Allswage UK. Roebuck Street, West Bromwich, B70 6RB

SKIVING MACHINE MAINTENANCE & INSTRUCTIONS MANUAL ASSK2. Allswage UK. Roebuck Street, West Bromwich, B70 6RB SKIVING MACHINE ASSK2 MAINTENANCE & INSTRUCTIONS MANUAL A. WARRANTY AND RESPONSIBILITY Warranty: It's the supplier's responsibility to guarantee the conformity of the product, assuring that it's manufactured

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS TS1000C TS700C INSTALLATION INSTRUCTIONS CONTENT: 1. Important safety instructions. 2. Specifications and main dimensions. 3. Parts included. 4. Installation. 5. RF remote system. 6. Feature functions

More information

icore Kiosk system Installation Guide

icore Kiosk system Installation Guide icore Kiosk system Installation Guide The reproduction, transmission or use of this document or its contents is not permitted without express authority. Offenders will be liable for damages. All rights,

More information

PKP Prozessmesstechnik GmbH. Borsigstrasse 24. D Wiesbaden-Nordenstadt. Tel: / Fax: / Operating manual PSA06

PKP Prozessmesstechnik GmbH. Borsigstrasse 24. D Wiesbaden-Nordenstadt. Tel: / Fax: / Operating manual PSA06 PKP Prozessmesstechnik GmbH Borsigstrasse 24 D-65205 Wiesbaden-Nordenstadt Tel: 06122 / 7055-0 Fax: 06122 / 7055 50 Operating manual PSA06 Electronical pressure switch Content Page 19-34 1. Foreword 19

More information

RST ROOM TEMPERATURE TRANSMITTER. Mounting and operating instructions

RST ROOM TEMPERATURE TRANSMITTER. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 4 OPERATIONAL DIAGRAM 5 WIRING AND

More information

Operating instructions. Switching amplifier DN0210 DN / / 2015

Operating instructions. Switching amplifier DN0210 DN / / 2015 Operating instructions Switching amplifier DN0210 DN0220 UK 80011079 / 00 01 / 2015 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5

More information

To connect the AC adapter:

To connect the AC adapter: Replacing the AC Adapter Replacing the AC Adapter 3 Plug the power cord into a wall outlet. The power indicator turns on. To connect the AC adapter: Connect the power cord to the AC adapter. Power indicator

More information

Expansion module EM-BAC-MOD-01. Installation manual

Expansion module EM-BAC-MOD-01. Installation manual Installation manual GB/en Expansion module EM-BAC-MOD-01 BACnet MS/TP or Modbus RTU interface card for EASYLAB controller TCU3 and adapter module TAM Firmware from 4.0 TROX GmbH Heinrich-Trox-Platz 47504

More information

5. Operating instructions: Octagon 200CL

5. Operating instructions: Octagon 200CL 1. Technical specifications 5. Operating instructions: Octagon 200CL Copyright 2015 by Endecotts Ltd. 33 1. Setting up Technical specifications SIEVE SHAKER MODEL: Octagon 200CL General Information The

More information

TRC-190 User s Manual

TRC-190 User s Manual First Edition, November 2008 www.moxa.com/product 2008 Moxa Inc. All rights reserved. Reproduction without permission is prohibited. The software described in this manual is furnished under a license agreement

More information

The power behind competitiveness. Delta Infrasuite Power Management. Power Distribution Unit. User Manual.

The power behind competitiveness. Delta Infrasuite Power Management. Power Distribution Unit. User Manual. The power behind competitiveness Delta Infrasuite Power Management Power Distribution Unit User Manual www.deltapowersolutions.com Save This Manual This manual contains important instructions and warnings

More information

C9900-P223 and C9900-P224

C9900-P223 and C9900-P224 Installation and Operating instructions for C9900-P223 and C9900-P224 Power Supply Units Version: 1.2 Date: 2012-02-22 Table of contents Table of contents 1. Foreword 2 Notes on the Documentation 2 Liability

More information

E2KQ-X. Fluororesin-coated Capacitive Sensor with Sensitivity Adjuster. Chemical-resistant Proximity Sensor. Ordering Information

E2KQ-X. Fluororesin-coated Capacitive Sensor with Sensitivity Adjuster. Chemical-resistant Proximity Sensor. Ordering Information Chemical-resistant Proximity Sensor E2KQ-X CSM_E2KQ-X_DS_E_4_3 Fluororesin-coated Capacitive Sensor with Sensitivity Adjuster Excellent resistance against chemicals and oil with fluororesincoated case.

More information

Removal and Installation8

Removal and Installation8 8 Screw Types 8-4 Top Cover Assembly 8-5 Left Hand Cover 8-6 Right Hand Cover 8-10 Front Panel Assembly 8-14 Left Rear Cover 8-15 Right Rear Cover 8-16 Extension Cover (60" Model only) 8-17 Media Lever

More information

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26 E39.0.31.6C-02 Data logger for humidity temperature as of version V1.0 Operating Manual T-Logg 160 GREISINGER electronic GmbH D - 93128 Regenstauf, Hans-Sachs-Straße 26 +49 (0) 9402 / 9383-0 +49 (0) 9402

More information

Installation manual plugs and connectors with screw connection (16/32 A)

Installation manual plugs and connectors with screw connection (16/32 A) EN Installation manual plugs and connectors with screw connection (16/32 60003213 Issue 04.2016 2016-04-01 Table of contents 1 About this manual 3 1.1 Structure of the warnings 3 1.2 Symbols used 4 1.3

More information

PDS100 Programmable Dispensing System SAFETY INSTRUCTIONS

PDS100 Programmable Dispensing System SAFETY INSTRUCTIONS PDS100 Programmable Dispensing System INSTRUCTION MANUAL SAFETY INSTRUCTIONS Before using any Fluid Metering, Inc. product read the following safety instructions as well as specific product specifications

More information

Operating instructions. Speed monitor D / / 2014

Operating instructions. Speed monitor D / / 2014 Operating instructions Speed monitor D200 80005257 / 00 05 / 2014 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5 2.2 Target group...5

More information

Installation Guide V290 (Color) This guide provides basic information for Unitronics LCD color touchscreen models V C30B and V T40B.

Installation Guide V290 (Color) This guide provides basic information for Unitronics LCD color touchscreen models V C30B and V T40B. Vision OPLC Installation Guide V290 (Color) This guide provides basic information for Unitronics LCD color touchscreen models V290-19-C30B and V290-19-T40B. General Description Vision OPLCs are programmable

More information

Documentation KM2042. Sixteen channel digital output module with D-Sub Connector. Version: Date:

Documentation KM2042. Sixteen channel digital output module with D-Sub Connector. Version: Date: Documentation Sixteen channel digital output module with D-Sub Connector Version: Date: 2.0.0 2017-11-20 Table of contents Table of contents 1 Foreword... 5 1.1 Notes on the documentation... 5 1.2 Safety

More information

Installation and operation manual ReciFlow Gas

Installation and operation manual ReciFlow Gas Installation and operation manual ReciFlow Gas 1 1. Measurement principle... 3 2. Installation... 5 3. Operation... 7 4. Electrical interfaces... 11 5. Communication protocol... 14 6. Software update and

More information

ITB THROTTLE WITH INTEGRATED STEPPER MOTOR INSTALLATION INSTRUCTION

ITB THROTTLE WITH INTEGRATED STEPPER MOTOR INSTALLATION INSTRUCTION ITB THROTTLE WITH INTEGRATED STEPPER MOTOR INSTALLATION INSTRUCTION MOTORTECH Gas Engine Accessories P/N 01.50.014 Rev. 12/2014 Copyright Copyright 2014 MOTORTECH GmbH. All rights reserved. Distribution

More information

C9900-P208 and C9900-P209

C9900-P208 and C9900-P209 Installation and Operating instructions for C9900-P208 and C9900-P209 Power Supply Units Version: 1.5 Date: 2012-02-22 Table of contents Table of contents 1. Foreword 2 Notes on the Documentation 2 Liability

More information

RXTP ROOM TEMPERATURE

RXTP ROOM TEMPERATURE ROOM TEMPERATURE CONTROLLER WITH PI CONTROL Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4

More information

DXTH DUCT SENSOR / SWITCH FOR TEMPERATURE AND HUMIDITY. Mounting and operating instructions

DXTH DUCT SENSOR / SWITCH FOR TEMPERATURE AND HUMIDITY. Mounting and operating instructions DUAL DUCT SENSOR / SWITCH FOR TEMPERATURE AND HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL

More information

Operating and Installation Manual. EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage

Operating and Installation Manual. EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage Operating and Installation Manual EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage Type EM-TRF-USV for 230 V AC mains voltage; provides uninterruptible power supply (UPS) The art of handling

More information

3-Phase, Dual-Input 6-Slot Power Supply System STARTUP GUIDE

3-Phase, Dual-Input 6-Slot Power Supply System STARTUP GUIDE 3-Phase, Dual-Input 6-Slot Power Supply System STARTUP GUIDE -ST-01 Page 1 of 10 November 2016 2016 Copyright Lite-On Technology Corporation ALL RIGHTS RESERVED. Lite-On is a trademark of Lite-On Technology

More information

Sliding shutter control Aluminium shutters

Sliding shutter control Aluminium shutters Sliding shutter control Aluminium shutters Operating and maintenance instructions FOR THE USER Valid from 09.2011 E 916.3 Notes These operating instructions describe how to operate the EHRET sliding shutter

More information

Operation Manual WARNING. Be sure to read this Operation Manual before use. Universal Space Amusement Equipment Ltd.

Operation Manual WARNING. Be sure to read this Operation Manual before use. Universal Space Amusement Equipment Ltd. WARNING Be sure to read this Operation Manual before use. Universal Space Amusement Equipment Ltd. CONTENTS 1. The company..2 2. Specifications.. 3 3. Package Contents..5 4. Installation, Fix and Transport..6

More information

C-pan arm USERS INSTRUCTIONS

C-pan arm USERS INSTRUCTIONS C-pan arm USERS INSTRUCTIONS Designed 1 of 12 and made in Denmark Thanks for purchasing a 9.Solutions product. With great passion, we design our products to be as versatile as possible. We hope that our

More information

Nanoject III. Programmable Nanoliter Injector INSTRUCTION MANUAL

Nanoject III. Programmable Nanoliter Injector INSTRUCTION MANUAL Nanoject III Programmable Nanoliter Injector INSTRUCTION MANUAL 3-000-207 MADE LABORATORY EQUIPMENT E21997 IN USA Rules for Safe Operation For indoor use only. Never operate unit in an explosive atmosphere.

More information

ZUA 2000 Universal Applicator. Instruction manual

ZUA 2000 Universal Applicator. Instruction manual ZUA 2000 Universal Applicator Instruction manual Index Exclusion of liability... 4 1 Description of device... 5 2 Safety information... 5 2.1 Symbols used... 5 2.2 Safety notes and hints... 5 3 Delivery

More information

MDM 011-Z1 Regen Resistor

MDM 011-Z1 Regen Resistor MDM 011-Z1 Regen Resistor Date of creation: 10.04.2017 Version date: 10.04.2017 Article number: 09-402-011-Z1-E Publisher: SIGMATEK GmbH & Co KG A-5112 Lamprechtshausen Tel.: 06274/4321 Fax: 06274/4321-18

More information

Flow Sensor SONOFLOW CO.55

Flow Sensor SONOFLOW CO.55 Flow Sensor SONOFLOW CO.55 Operating Manual Device: Model: Type: Flow Sensor SONOFLOW CO.55 Manufacturer: SONOTEC Ultraschallsensorik Halle GmbH Nauendorfer Str. 2 06112 Halle (Saale), Germany Phone: +49

More information

PTB 01 ATEX Example / Beispiel / Exemple. Type Operating Instructions

PTB 01 ATEX Example / Beispiel / Exemple. Type Operating Instructions Device with II 2G EX i approval Geräte mit II 2G EX i Zulassung Appareils avec mode de protection II 2G EX i Example / Beispiel / Exemple Type 6106 Operating Instructions Bedienungsanleitung Manuel d utilisation

More information

BK MIKRO. Air Barrier Adapter. Assembly Instructions Issue 1.00 dated

BK MIKRO. Air Barrier Adapter. Assembly Instructions Issue 1.00 dated BK MIKRO Air Barrier Adapter Assembly Instructions Issue 1.00 dated 5.6.2008 MSC Tuttlingen GmbH Rudolf-Diesel-Straße 17 78532 Tuttlingen Germany Tel. +49 7461 925-200 Fax +49 7461 925-268 E-Mail sales@msc-tuttlingen.de

More information

H Series PLC. ! : Indicates Compulsion. EH-150 Analog input module EH-AXH8M Instruction manual. Safety precautions DANGER CAUTION COMPULSION

H Series PLC. ! : Indicates Compulsion. EH-150 Analog input module EH-AXH8M Instruction manual. Safety precautions DANGER CAUTION COMPULSION H Series PLC EH-150 Analog input module EH-AXH8M Instruction manual Thank you for purchasing a Hitachi Programmable Logic Controller. To operate it safely, please read this instruction manual and all the

More information

1 Channel Strobe Controller ORDERCODE 40226

1 Channel Strobe Controller ORDERCODE 40226 1 Channel Strobe Controller ORDERCODE 40226 Congratulations! You have bought a great, innovative product from Showtec. The Showtec Strobe Controller brings excitement to any venue. Whether you want simple

More information

SIWAREX JB Aluminum Housing. Instruction Manual Edition 03/2006

SIWAREX JB Aluminum Housing. Instruction Manual Edition 03/2006 s SIWAREX JB Aluminum Housing Instruction Manual Edition 03/2006 Table of Contents Table of Contents... 2 Warning and Safety Terms... 3 General... 4 1 Technical Description... 5 1.1 Area of Application...

More information

JUMO PINOS L01. Calorimetric flow sensor. Brief description. Special features. Customer benefits. Data Sheet

JUMO PINOS L01. Calorimetric flow sensor. Brief description. Special features. Customer benefits. Data Sheet sales@jumo.co.uk info.us@jumo.net Page 1/9 JUMO PINOS L01 Calorimetric flow sensor Brief description The JUMO PINOS L01 flow sensor serves the measuring and monitoring of flow velocities of water and aqueous

More information

This guide provides basic information for Unitronics Models 230/260/280/290 (Non-color Screens).

This guide provides basic information for Unitronics Models 230/260/280/290 (Non-color Screens). Vision OPLC Installation Guide Models 230/260/280/290 (Non-color Screens) This guide provides basic information for Unitronics Models 230/260/280/290 (Non-color Screens). General Description Vision OPLCs

More information

ASI Systems ML5740A Series

ASI Systems ML5740A Series ML5740A Series Electric Linear Valve Actuators Non-Spring Return APPLICATION The ML5740A series actuators operate standard ASI Systems valves in heating, ventilating and air conditioning (HVAC) applications.

More information

RTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions

RTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA STANDARDS OPERATIONAL DIAGRAMS WIRING AND CONNECTIONS

More information

Table of Contents Quick Install Guide page Introduction Safety Rack System Precautions ESD Precautions...

Table of Contents Quick Install Guide page Introduction Safety Rack System Precautions ESD Precautions... Table of Contents Quick Install Guide page 1 EN English Table of Contents 1. Introduction... 2 1.1 Safety... 2 1.2 Rack System Precautions... 2-3 1.3 ESD Precautions... 3... 3 1... 3 2 Fitting PSU s...

More information

TRC-190 User s Manual

TRC-190 User s Manual User s Manual Edition 3.2, May 2017 www.moxa.com/product 2017 Moxa Inc. All rights reserved. User s Manual The software described in this manual is furnished under a license agreement and may be used only

More information

Manual Version: V1.00. Video Decoder Quick Guide

Manual Version: V1.00. Video Decoder Quick Guide Manual Version: V1.00 Video Decoder Quick Guide Thank you for purchasing our product. If there are any questions, or requests, please do not hesitate to contact the dealer. Copyright Copyright 2016 Zhejiang

More information

BS 287 DUAL CHANNEL POWER SUPPLY. User Manual. January 2017 V1.0

BS 287 DUAL CHANNEL POWER SUPPLY. User Manual. January 2017 V1.0 BS 287 DUAL CHANNEL POWER SUPPLY User Manual January 2017 V1.0 Table of contents 1.0 SAFETY INSTRUCTIONS... 3 2.0 GENERAL DESCRIPTION PS 289... 4 3.0 MECHANICAL INSTALLATION... 5 4.0 MAINS POWER & SAFETY

More information

801 Stirrer. Manual EN

801 Stirrer. Manual EN 801 Stirrer Manual 8.801.8001EN Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 801 Stirrer Manual 8.801.8001EN 06.2010 ek Teachware

More information

ARGO LAB SKO-D XL Orbital and Linear digital shaker. User manual

ARGO LAB SKO-D XL Orbital and Linear digital shaker. User manual ARGO LAB SKO-D XL Orbital and Linear digital shaker User manual 1 Contents 1. Warranty... 2 2. Safety instruction... 3 3. Proper use... 3 4. Receiving Inspection... 3 4.1 Listing of Items... 3 5. Trial

More information

B63/ NS MS. EtherNet/IP LINK

B63/ NS MS. EtherNet/IP LINK 3 609 929 B63/ IMenip 2008-09 NS MS EtherNet/IP LINK 3 609 929 B63/2008-09 IMenip Bosch Rexroth AG 15/76 Table of Contents About this document................. 16 General safety instructions............

More information

Installation- and Operating instructions for CU Ethernet Controller with USB Input. Version: 1.4 Date:

Installation- and Operating instructions for CU Ethernet Controller with USB Input. Version: 1.4 Date: Installation- and Operating instructions for CU8880-0010 Ethernet Controller with USB Input Version: 1.4 Date: 2018-04-12 Table of contents Table of contents 1. 2. 3. 4. 5. General instructions 2 Notes

More information

Version-E Manual PR1a. Photocell. Manual. Page 1

Version-E Manual PR1a. Photocell. Manual. Page 1 Version-E16-06-02 Photocell Manual Page 1 Important Information General Before using your ALGE-TIMING device read the complete manual carefully. It is part of the device and contains important information

More information

Type Quickstart (from device Version 2)

Type Quickstart (from device Version 2) Type 8022 Flow Transmitter / Pulse divider Durchflusstransmitter / Impulsteiler Transmetteur de débit / Diviseur d impulsions We reserve the right to make technical changes without notice. Technische Änderungen

More information

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide Model P4017 Single Channel USB Oscilloscope Quick Start Guide General Warranty BNC warrants that the product will be free from defects in materials and workmanship for 3 years from the date of purchase

More information

9 Maintenance and Repair

9 Maintenance and Repair Agilent 1200 Infinity II Series RID User Manual 9 Maintenance and Repair Introduction to Maintenance 120 Warnings and Cautions 121 Overview of Maintenance 123 Cleaning the Module 124 Storage of the Detector

More information

ISDN PABX System. Installation Manual. Profi Line: AS 100 IT

ISDN PABX System. Installation Manual. Profi Line: AS 100 IT ISDN PABX System Installation Manual Profi Line: AS 100 IT Safety Instructions - Equipment setup and maintenance should only be carried out by specially trained personnel. - Caution! The PABX needs to

More information

Led Illuminator and Light Stimulator. Model # LSD-1. Ver

Led Illuminator and Light Stimulator. Model # LSD-1. Ver Led Illuminator and Light Stimulator Model # LSD-1 Ver. 2.0.0 5.16 ALA Scientific Instruments Inc. 60 Marine Street, Farmingdale, NY 11735 Voice: +1 631.393.6401 fax: +1 631.393.6407 www.alascience.com

More information

AI-102. IR Illuminator. Installation Guide. Rev IP Surveillance

AI-102. IR Illuminator. Installation Guide. Rev IP Surveillance AI-102 IR Illuminator Installation Guide Rev. 1.1 IP Surveillance Revision History: Rev. 1.0: Initial release Rev. 1.1: Changed effective IR range and power consumption. AI-102 Tube Type Infrared Illuminator

More information

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 4 OPERATIONAL DIAGRAMS 5 WIRING AND

More information

Operating Instruction for Turbine-wheel Flow Meter. Model: DRS

Operating Instruction for Turbine-wheel Flow Meter. Model: DRS Operating Instruction for Turbine-wheel Flow Meter Model: DRS . Contents. Contents... 2 2. Note... 3 3. Instrument Inspection... 3 4. Regulation Use... 3 5. Operating Principle... 4 6. Mechanical Connection...

More information

Operating Instruction for Rotating Vane Flow Meter. Model: DRG-...

Operating Instruction for Rotating Vane Flow Meter. Model: DRG-... Operating Instruction for Rotating Vane Flow Meter Model: DRG-... 1. Contents 1. Contents... 2 2. Note... 3 3. Regulation Use... 3 4. Operating Principle... 3 5. Instrument Inspection... 4 6. Mechanical

More information

Mounting and Operating Instructions EB EN. Electric Actuators Type 5824 without fail-safe action Type 5825 with fail-safe action

Mounting and Operating Instructions EB EN. Electric Actuators Type 5824 without fail-safe action Type 5825 with fail-safe action Electric Actuators Type 5824 without fail-safe action Type 5825 with fail-safe action Three-step version Translation of original instructions Mounting and Operating Instructions EB 5824-1 EN Edition January

More information

CU USB-Extender-Rx (USB and DVI Extender) Version: 0.3 Date:

CU USB-Extender-Rx (USB and DVI Extender) Version: 0.3 Date: CU8860-0000 USB-Extender-Rx (USB and DVI Extender) Version: 0.3 Date: 2006-02-20 Table of Contents Table of Contents 1 Foreword 1 1.1 Notes on the documentation 1 1.1.1 Liability Conditions 1 1.1.2 Conditions

More information

GV204 Impulse Splitter for Incremental Encoders

GV204 Impulse Splitter for Incremental Encoders Operating Manual www.genesisautomationonline.com GV204 Impulse Splitter for Incremental Encoders Product features: Programmable encoder input A, /A, B, /B, Z, /Z, for use with RS422 or TTL differential

More information

Model No. ET-JPF200BE

Model No. ET-JPF200BE Operating Instructions Floor Stand Kit Commercial Use Model No. ET-JPF200BE ET-JPF200WE ENGLISH FRANÇAIS ESPAÑOL DEUTSCH ITALIANO * The above illustration is of this product mounted to an optional projector.

More information

BPL SERIES INSTALLATION INSTRUCTIONS THIS SHEET CONTAINS IMPORTANT SAFETY INSTRUCTIONS. SAVE THESE INSTRUCTIONS.

BPL SERIES INSTALLATION INSTRUCTIONS THIS SHEET CONTAINS IMPORTANT SAFETY INSTRUCTIONS. SAVE THESE INSTRUCTIONS. BPL SERIES INSTALLATION INSTRUCTIONS Important Warning THIS SHEET CONTAINS IMPORTANT SAFETY INSTRUCTIONS. SAVE THESE INSTRUCTIONS. This product must be installed in accordance with National Electrical

More information

Operating instructions. Standstill monitor A / / 2011

Operating instructions. Standstill monitor A / / 2011 Operating instructions Standstill monitor A300 UK 1 2 3 4 5 6 7 8 7390337 / 01 02 / 2011 1 2 3 4 5 6 7 8 switchpoint min max pulse/min power Made in Germany ifm electronic gmbh D 45127 Essen func. I II

More information

M60 SMS Reference Manual V5

M60 SMS Reference Manual V5 Last revision 02/2017 M60 SMS Reference Manual V5 An M60 fitted with our SMS Module fitted works exactly the same as a standard M60, but has the added facility to send SMS TEXT Messages. Each channel/input

More information

Fog machine PT Users Manual

Fog machine PT Users Manual Fog machine PT-1500 Users Manual Table of content 1. Safety instructions... 3 1.1. For safe and efficient operation... 3 2. Intended use of the device... 4 2.1. Overhead installation... 5 2.2. Electrical

More information

Operating manual. Diagnostic adapter VAS Operating manual Diagnostic adapter VAS Client documentation Revision: 00 Version: 05/2018

Operating manual. Diagnostic adapter VAS Operating manual Diagnostic adapter VAS Client documentation Revision: 00 Version: 05/2018 Operating manual Diagnostic adapter VAS 611 009 CAR-connect GmbH info@car-connect.cc www.car-connect.cc Page 1 of 11 Imprint Title: Operating manual of diagnostic adapter VAS 611 009 Manufacturer: CAR-connect

More information

ASI Systems ML5720A Series

ASI Systems ML5720A Series ML5720A Series Electric Linear Valve Actuators Non-Spring Return APPLICATION The ML5720A series actuators operate standard ASI Systems valves in heating, ventilating and air conditioning (HVAC) applications.

More information

Service: SLR Laboratory Stirrer glass made of ideas. Service Instructions for the SLR Laboratory Stirrer

Service: SLR Laboratory Stirrer glass made of ideas. Service Instructions for the SLR Laboratory Stirrer Service Instructions for the SLR Laboratory Stirrer SLR Service eng.doc 17 Version: 03.04.2001 Contents Page Contents... 18 General way of proceeding... 19 1 Safety information... 20 2 Introduction...

More information

English. Operating manual. Universal transmitter UT125. Save for later reference. Companies / brands of GHM

English. Operating manual. Universal transmitter UT125.  Save for later reference. Companies / brands of GHM English Operating manual Universal transmitter UT125 Companies / brands of GHM www.ghm-messtechnik.de Save for later reference. Table of contents page 1. Intended use (areas of application)... 3 1.1 Safety

More information

CubePro. Main PCB Replacement Guide. Prosumer 3D Printer. Original Instructions

CubePro. Main PCB Replacement Guide. Prosumer 3D Printer. Original Instructions CubePro Prosumer 3D Printer Main PCB Replacement Guide Original Instructions 1 INTRODUCTION COPYRIGHT 2014 by All rights reserved. This document is subject to change without notice. This document is copyrighted

More information

E2EY. A Proximity Sensor for Aluminum, Brass and Other Non-ferrous Metals. Iron Is Not Detected. Aluminum-detecting Proximity Sensor

E2EY. A Proximity Sensor for Aluminum, Brass and Other Non-ferrous Metals. Iron Is Not Detected. Aluminum-detecting Proximity Sensor Aluminum-detecting Proximity Sensor EEY CSM_EEY_DS_E_5_1 A Proximity Sensor for Aluminum, Brass and Other Non-ferrous Metals. Iron Is Not Detected. Non-ferrous metals, such as aluminum and brass, are detected.

More information

Serial PROFIBUS Interface

Serial PROFIBUS Interface Installation Manual Serial PROFIBUS Interface Version: EN-062016-2.3 Copyright 2016 Softing Industrial Automation GmbH Disclaimer of liability The information contained in these instructions corresponds

More information

Operating Instructions (Translation) 3. Safety Information. 1. Description. 2. Explosion Protection. Supply module Type 17-21BB-170x

Operating Instructions (Translation) 3. Safety Information. 1. Description. 2. Explosion Protection. Supply module Type 17-21BB-170x 1. Description The supply module was developed specially for direct mounting in hazardous areas in Zone 1 and 21 and is ATEX-certified. The supply module is a permanently installed piece of electrical

More information

3M AEROSPACE & COMMERCIAL TRANSPORTATION DIVISION St. Paul, MN USA or or Made in USA

3M AEROSPACE & COMMERCIAL TRANSPORTATION DIVISION St. Paul, MN USA  or or Made in USA 3M AEROSPACE & COMMERCIAL TRANSPORTATION DIVISION St. Paul, MN 55144-1000 USA www.3m.com/aerospace or 1-800-364-3577 or 1-800-235-2376 Made in USA with US and Globally Sourced Materials 3M 2015. All rights

More information

Documentation. KM2604 and KM2614. Four channel relay module. Version: Date:

Documentation. KM2604 and KM2614. Four channel relay module. Version: Date: Documentation KM2604 and KM2614 Four channel relay module Version: Date: 2.1.0 2017-12-01 Tabel of contents Tabel of contents 1 Foreword... 5 1.1 Notes on the documentation... 5 1.2 Safety instructions...

More information

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: This meter is

More information

RТTH DUAL ROOM SWITCH

RТTH DUAL ROOM SWITCH DUAL ROOM SWITCH FOR TEMPERATURE AND RELATIVE HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA

More information

LW217 / LW218 Optical fiber modules for transmission of SSI encoders and measuring systems

LW217 / LW218 Optical fiber modules for transmission of SSI encoders and measuring systems Operating Manual LW217 / LW218 Optical fiber modules for transmission of SSI encoders and measuring systems Product features: The optical fiber modules for data transmission of SSI signals by means of

More information

PTZ Control VMC Joystick

PTZ Control VMC Joystick Installation and Configuration English PTZ Control VMC Joystick Rev. 1.0.0 / 2010-07-20 Information about Copyright, Trademarks, Design Patents 2010 Dallmeier electronic The reproduction, distribution

More information

TOP - 1. Instruction Manual. Version 1.0 Produced in Jan. 2004

TOP - 1. Instruction Manual. Version 1.0 Produced in Jan. 2004 Version 1.0 Produced in Jan. 2004 Instruction Manual LCD monitor IV-08MP Thank you for purchasing the SHARP IV-08MP LCD monitor. Read this introductory instruction manual carefully to thoroughly familiarize

More information

HST -TZ1 Guard-locking mechanism (Translation of Original Manual)

HST -TZ1 Guard-locking mechanism (Translation of Original Manual) Installation and Operating Manual for Components HST -TZ1 Guard-locking mechanism (Translation of Original Manual) HST-TZ1 Ident.-No.: 10234 HST-TZ1 Ident.-No.: 10236 HST-TZ1 Ident.-No.: 10235 HST-TZ1

More information

Documentation EM2042. Sixteen Channel Digital Output Module with D-Sub Connector. Version: Date:

Documentation EM2042. Sixteen Channel Digital Output Module with D-Sub Connector. Version: Date: Documentation Sixteen Channel Digital Output Module with D-Sub Connector Version: Date: 2.0 2016-08-03 Table of contents Table of contents 1 Foreword... 4 1.1 Notes on the documentation... 4 1.2 Safety

More information

User Manual. 1U LCD Keyboard Drawer. KwikDraw - A Series. Manual. IT and Instrumentation for industry. - With KVM options - 15", 17, 19 screen size

User Manual. 1U LCD Keyboard Drawer. KwikDraw - A Series. Manual. IT and Instrumentation for industry. - With KVM options - 15, 17, 19 screen size User Manual 1U LCD Keyboard Drawer KwikDraw - A Series - With KVM options - 15", 17, 19 screen size 1.1 Important Safeguards Please read all of these instructions carefully before you use the device. Save

More information

Universal Tristop Cylinder for Cam Brakes

Universal Tristop Cylinder for Cam Brakes Tristop for Cam Brakes 3rd Edition This publication is not subject to any update service. New versions are available in INFORM at www.wabco-auto.com Copyright WABCO 2006 Vehicle Control Systems An American

More information

Assembly and Setup Manual

Assembly and Setup Manual M-12 Series Copyboard / C-12 Series Captureboard Assembly and Setup Manual This is the installation and assembly manual for the M-12 series Copyboard and C-12 series Captureboard. (The copyboard and/or

More information

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com Model 2460-KIT Screw Terminal Connector Kit Description The Model 2460-KIT Screw Terminal Connector Kit

More information

RXTH DUAL ROOM SENSOR / SWITCH

RXTH DUAL ROOM SENSOR / SWITCH DUAL ROOM SENSOR / SWITCH FOR TEMPERATURE AND RELATIVE HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE

More information

BIC 1B0-ITA50-Q40KFU-SM4A4A BIC 1B0-IT005-Q40KFU-SM4A4A BIC 2B0-ITA50-Q40KFU-SM4A5A. User s Guide

BIC 1B0-ITA50-Q40KFU-SM4A4A BIC 1B0-IT005-Q40KFU-SM4A4A BIC 2B0-ITA50-Q40KFU-SM4A5A. User s Guide BIC 1B0-ITA50-Q40KFU-SM4A4A BIC 1B0-IT005-Q40KFU-SM4A4A BIC 2B0-ITA50-Q40KFU-SM4A5A User s Guide Table of contents 1 Safety 2 1.1. Installation and startup 2 1.2. General safety instructions 2 Hazardous

More information

Operating manual LRD 325. Conductivity measuring cell to be installed in pipelines

Operating manual LRD 325. Conductivity measuring cell to be installed in pipelines Operating manual Conductivity measuring cell to be installed in pipelines ba55321de02 10/2010 Copyright Weilheim 2010, WTW GmbH Reprinting - even as excerpts - is only allowed with the explicit written

More information

LCD LCDS LCDRI CONTROLLER OPERATING MANUAL

LCD LCDS LCDRI CONTROLLER OPERATING MANUAL This manual contains important safety information about installation and use of this equipment. Ignoring this information could result in injuries or damages. It is strictly forbidden to use this equipment

More information

PRODUCT MARKING AND BARCODE IDENTIFICATION. up to Serial No Barcode/Label printer MACH4. Service Manual. Edition 11/06

PRODUCT MARKING AND BARCODE IDENTIFICATION. up to Serial No Barcode/Label printer MACH4. Service Manual. Edition 11/06 PRODUCT MARKING AND BARCODE IDENTIFICATION up to Serial No. 9999 Barcode/Label printer MACH4 Service Manual Edition 11/06 copyright by cab / 9008553 / Q49 / 1 All specifications about delivery, design,

More information

Installation Job Aid for VSP 4850GTS

Installation Job Aid for VSP 4850GTS Installation Job Aid for VSP 4850GTS Notices Release 6.1.0.0 NN46251-308 Issue 02.01 November 2017 Notice paragraphs alert you about issues that require your attention. The following paragraphs describe

More information

Operation Manual SRZ Helical Flow Meters

Operation Manual SRZ Helical Flow Meters Operation Manual SRZ Helical Flow Meters Küppers Elektromechanik Contents Liebigstr. 2 85757 Karlsfeld Tel. +49 81 31/59 39 10 Fax +49 81 31/9 26 04... Fehler! Textmarke nicht definiert. 1 General Information...

More information