Aqueous GPC Instructions - Detailed

Size: px
Start display at page:

Download "Aqueous GPC Instructions - Detailed"

Transcription

1 Aqueous GPC Instructions - Detailed Waste Bottle Columns PL Datastream Injection Port GPC Computer UV Detector GPC Pump Eluent Bottle The Aqueous GPC Set Up RI Detector Important The points below must be observed: Please check and fill in the GPC queue sheet before you use this instrument. Fill in the log book every time you use this instrument. The GPC pump must never be allowed to run dry, and must not be left running overnight. All sample solutions must be filtered before being injected into the GPC, using 0.2 µm nylon filters. Samples must form a stable solution in the analysis conditions used, to prevent solid deposits forming in the GPC columns and blocking them. The UV Detector must be turned off when not in use to preserve the life of the lamp. Make sure the eluent is properly degassed to prevent air bubbles forming in the filter and being pumped through the column. Either degas the eluent every few days or make up fresh eluent. Flush the piston and piston seal areas at the end of each day with at least 50 ml deionised water, using a syringe or wash bottle connected to the flush tube. Peter Iddon Aqueous GPC (Detailed).doc Page 1 of 8

2 Eluent and Sample Preparation Eluents/solvents must be degassed before use. Aqueous solutions should be stirred under vacuum for 1 hour. Co-solvents (e.g. methanol) are best degassed in bulk, added to the degassed aqueous solution, and then the mixture degassed for an additional 10 minutes. Samples are normally prepared at a concentration of 5 to 20 mg polymer per ml eluent. It may be necessary to adjust the ph of a sample solution to match that of the eluent. All samples must be filtered through a 0.2 µm nylon filter before injection into the GPC. Samples must form a stable solution in the conditions used. Turning on the Aqueous GPC Pump Display Pump Purge Valve The Pump Controls 1. Make sure the waste bottle is empty, and that there is enough eluent in the eluent bottle to fully cover the filter for the whole time the pump will be running. 2. Turn on the GPC pump (at the back on the left) and PL Datastream (at the back). The RI detector should already be switched on (it should not be switched off). 3. Turn on the GPC pump by pressing FLOW CTRL, then YES on the pump control panel. The current flow rate and pressure will be displayed on the pump display. 4. If necessary alter the flow rate by pressing FLOW, then entering in the new flow rate using the number keys, then pressing ENTER (pump control panel). 5. The GPC should be purged for 1 hour at the start of each day it is used. Once the pump has been started, press the PURGE button on the front of RI detector (a green light will show). Press this button again at the end of the hour to turn the purge off. 6. Turn on the computer, and the 7. Turn on the UV Detector if required. 8. Open the Datastream Monitor program by either clicking on the desktop shortcut icon or by going through the Start menu (Start > Programs > PL Cirrus > Datastream Monitor). 9. Click on the Channel menu and select Add Channel. 10. Select PL Datastream on COM2 and RI (and also UV if required). Click on OK. 11. A real-time display of the detector output will now be displayed on the monitor. Peter Iddon Aqueous GPC (Detailed).doc Page 2 of 8

3 Running Samples 1. Fill in the Aqueous GPC Log Book, entering in the details for the current run. 2. Open the GPC Online program by either clicking on the desktop shortcut icon or by going through the Start menu (Start > Programs > PL Cirrus > GPC Online). 3. Either select an existing workbook from the Recent list, or select Other and click on OK to browse the workbook folders. If you need to create a new workbook, follow the instructions in the Creating a New Workbook section. 4. Click on OK to verify that the PL Datastream is ready for data collection. 5. If using both the RI and UV detectors, you will have to use two separate workbooks (one for each detector). 6. Click on the Runlist icon on the left side of the screen. 7. Click on the Add Sample toolbar button or select it from the Runlist menu. Keep on adding samples until there are enough entries for all the samples to be run (more sample entries can be added later if necessary). 8. Enter information into the following columns (the important columns are highlighted in bold): Batch Name The program will use the date as a batch name if you leave this blank. Sample Name Necessary to remember which chromatogram is which! File Name The program will create its own if you leave this blank. Sample Type Choose Unknown for sample analysis or Narrow Std. for calibrations. Method Choose the method (containing the calibration) to use for data analysis. If you have chosen Narrow Std. as the sample type a new calibration will be added to the method you choose. Cal. Version Will be Use Latest for unknown samples, or Create for calibrations. K Mark-Howink-Sakurada equation coefficient. Alpha Mark-Howink-Sakurada equation coefficient. Once a sample had started running its details will be saved (and cannot be changed), but sample type, method, K and Alpha can be varied during reanalysis using GPC Offline. 9. Check the baseline is flat using the Datastream Monitor program. If is not flat, read the Common Problems section. 10. Follow the instructions in the Sample Injection section below. 11. When the data collection time exceeds the time specified in the method being used, or collection is manually stopped, GPC Online will automatically stop collecting data. It will then be ready for the next sample, if one exists. Chromatograms are saved to file automatically. 12. To analyse the chromatograms follow the instructions in the Data Analysis (GPC Online) section. 13. The data collection will begin again automatically when the next sample is injected. Sample Injection The Sample Injection Port in the Left and Right Positions Peter Iddon Aqueous GPC (Detailed).doc Page 3 of 8

4 1. Make sure that the PURGE button on the RI detector is switched off (no green light). 2. Press the AUTO ZERO button on the RI detector. 3. Zero the UV detector if using it. 4. Make sure the sample injection port is turned to the left position (see diagram above). 5. For the first sample, rinse the sample injection port with about 1 ml eluent (making sure there are no air bubbles in the syringe). The eluent passes into the small bottle behind the sample injection port empty this if it becomes full. 6. For the first sample, rinse the 100 µl micro-syringe 3 times with distilled water. 7. Rinse and fill the micro-syringe with the sample solution, until there are no air bubbles visible. 8. Inject 100 µl of the sample solution into the sample injection port (being careful not to introduce air bubbles into the system). Do not remove the syringe. 9. Holding the syringe in the sample injection port, turn the sample injection port to the right position (see diagram above). Continue to hold the syringe in the port. After the sample has been flushed from the sample injection port, turn the sample injection port back to the left position, then remove the syringe (the time taken to flush the sample from the sample injection port will depend on the pump speed; allow about 12 seconds [0.20 minutes] at 1 mlmin -1 and about 24 seconds [0.40 minutes] at 0.5 mlmin -1 ). 10. When the sample injection port is moved to the right position, the data recording is started automatically. The Datastream Monitor output will change from blue to red, and will show the time since the sample was injected. 11. Rinse the 100 µl micro-syringe 3 times with distilled water or eluent, and leave with the barrel out. 12. Rinse the sample injection port with about 1 ml eluent (making sure there are no air bubbles in the syringe). Creating a New Calibration Calibrations are specific to individual workbooks, but identical calibrations can be created in other workbooks (using the same data) by importing the calibration chromatograms. Multiple calibrations can be created from the same chromatograms if necessary. 1. To create a calibration from narrow standards as you inject them, follow the instructions in the Running Samples section and then the Data Analysis (GPC Online) section. 2. Alternatively, calibrations can be created from previously run standards. Follow the instructions in the Analysis Using GPC Offline section to do this. Chromatograms run in other workbooks can be imported by clicking on Import Chromatograms in the File menu. Data Analysis (GPC Online) Samples can be analysed immediately after they are run using GPC Online, or analysed later using GPC Offline. 1. Once a chromatogram has been recorded, it can be analysed. 2. Click on the Analysis icon on the left side of the screen. 3. If there is no chromatogram waiting to be analysed, click on the Next Analysis toolbar button, or select it from the Processing menu. Depending on the options chosen in the method being used for analysis, the peaks may be assigned and results calculated automatically, or it may be done manually. 4. If analysing Unknown samples, peaks are assigned by clicking on the Detect Peak & Baseline toolbar button, or by selecting it from the Analysis menu. Click and hold the mouse pointer on one side of the peak to be analysed (on the baseline), then drag the baseline across to the other side of the peak. Peter Iddon Aqueous GPC (Detailed).doc Page 4 of 8

5 If analysing Narrow Standard samples for column calibrations: Peak maximums are assigned by clicking on the [Detect] Peak Max toolbar button or by selecting it from the Analysis menu, then clicking anywhere inside the peak. Once peak maximums are assigned, click on the Add to Calibration toolbar button or select it from the Processing menu. In the Assign Peak Molecular Weights window enter in the M p values for the standards used, next to the corresponding peak value (highest molecular weight standard will elute first). Then click on OK. 5. Peak parameters can be edited by clicking on the Move Markers [and Zoom] toolbar button, or selecting it from the Analysis menu. Move the mouse pointer and click under the baseline to move the baseline (pointer will change into a B) or move the mouse pointer and click above the baseline to move the peak limits (pointer will change into a P). Peaks can also be deleted by clicking on the Delete Peak toolbar button or by selecting it from the Analysis menu, then clicking anywhere inside the peak. 6. Once the peak assignments are satisfactory, click on the Calculate Results toolbar button, or select it from the Processing menu. 7. Results can be viewed by clicking on the Results icon on the left side of the screen. 8. Results files are not saved automatically. To save results, select Save Results from the Results menu. Chromatograms are saved automatically when the results are calculated, along with any peak/baseline assignments. 9. Results can be printed by selecting Print from the File menu (when in the Results view). 10. Once chromatograms are analysed, the analysis cannot be repeated you must reanalyse the file using GPC Offline (see the Analysis Using GPC Offline section below). Analysis Using GPC Offline Samples can be analysed at a later date using GPC Offline. To get exactly the same results as last time: If the sample peaks have been assigned previously, then the peak assignments will have been saved with the chromatograms, so the reanalysis of the peaks without altering any of the peak parameters (baseline and peak limits) will give the same results as the last time they were calculated (as long as the correct column calibration file for the date the samples were run is also used). 1. Open the GPC Offline program by either clicking on the desktop shortcut icon or by going through the Start menu (Start > Programs > PL Cirrus > GPC Offline). 2. Either select an existing workbook from the Recent list, or select Other and click on OK to browse the workbook folders. If you need to create a new workbook, follow the instructions in the Creating a New Workbook section. 3. Click on the Runlist icon on the left side of the screen. 4. Select the chromatogram files to be analysed from the middle file window. Double clicking on a batch folder (named from the Batch Name property) adds all the chromatograms with that batch name. 5. It is possible to import files from other workbooks. This option can be found in the File menu (File > Import Chromatograms > Cirrus ). These files are then stored in the workbook in the Imported batch folder. 6. Check the information into the following columns is correct and amend as necessary (the first three columns cannot be altered; the other important columns are highlighted in bold): Batch Name The default is the date the samples were run (unless changed by user). Sample Name The default is the date and time the samples were run (unless changed by user). File Name The default is the date and injection number (unless changed by user). Sample Type Choose Unknown for sample analysis or Narrow Std. for calibrations. Peter Iddon Aqueous GPC (Detailed).doc Page 5 of 8

6 Method Choose the method (containing the calibration) to use for data analysis. If you have chosen Narrow Std. as the sample type a new calibration will be added to the method you choose. Cal. Version Will be Use Latest for unknown samples, or Create for calibrations. You can choose to analyse files using older calibrations if you wish. K Mark-Howink-Sakurada equation coefficient. Alpha Mark-Howink-Sakurada equation coefficient. Any altered information will be for the current analysis only, no original information in the chromatogram file will be changed. 7. Now follow the instructions in the Data Analysis (GPC Online) section above. Creating a New Workbook Use a new workbook when using different run conditions or a different polymer. It would be best if users had their own folder in the Cirrus Workbooks folder (shortcut icon on desktop) into which they could put their different workbooks. It is possible to move workbook folders after they are created if necessary. 1. Open the GPC Online program by either clicking on the desktop shortcut icon or by going through the Start menu (Start > Programs > PL Cirrus > GPC Online). 2. Select Create a New Workbook and click on OK. 3. Click on the Default workbook icon, and click on OK. 4. Enter in a name for the new workbook next to Name. The save location of the workbook can be altered by clicking on browse (i.e. it can be placed inside your own folder). Information can be entered under Experimental Title and Comments in this step or the next step. Click on OK. 5. The workbook options window will now be displayed, showing five tabs. Enter in the following details by clicking on each tab (the important information is highlighted in bold): Workbook Enter information under Experimental Title and Comments if not already done so. Conditions A warning will appear when you first select this tab just click on OK. It only warns that Channel 1 is selected by default. Click on Edit in the Detectors section and for the Channel Number select RI(2) (for the RI detector) or UV(1) (for the UV detector). Make sure PL Datastream Input Voltage is set to 1.28 V and PL Datastream Mode is set to Input Data. Enter in details in the Pump and Column Set sections if desired. Data Collection Change Run Length to that required for your experiments. Other details here can be ignored. Instruments Instrument details can be entered if desired. Options Options here can be ignored. When all the details have been entered correctly, click on OK. This information can be edited at a later date. 6. When asked, click on Yes to ready the PL Datastream for data collection (the PL Datastream can be made ready by pressing F8 or selecting the option from the Collection menu if you press no by mistake). 7. Click on OK to verify that the PL Datastream is ready for data collection. 8. Edit or create the methods by following the instructions in the Editing or Creating a New Method section below. Peter Iddon Aqueous GPC (Detailed).doc Page 6 of 8

7 Editing or Creating a New Method You can either use one of the default methods for your calibration, edit an existing method, or create an entirely new one. Editing the default method Manual Analysis is recommended. 1. Click on the Method Browser icon on the left of the screen. 2. Select Edit Method or New Method from the Method menu. 3. The workbook options window will now be displayed, showing six tabs. Enter in the following details by clicking on each tab (the important information is highlighted in bold): General Name the method. Do Data Processing should be ticked. Time Between Chromatograms is the time delay between the completion of one analysis and the start of another (most relevant when automatic processing is used). Column Calibration Generate Using Narrow Standard Calibrants should be selected. Choose the default order of the polynomial to use for calibration. Select whether the calibration will have automatic or manual processing (Manual Processing without Automatic Peak Detection is recommended). Sample Analysis Select whether the sample analysis will have automatic or manual processing (Manual Processing without Automatic Peak Detection is recommended). Perform GPC Calculations should be ticked. Choose the Calculation Method (See Help for more information). Peak Detection Can change the limits and threshold options for automatic peak detection, if it is being used. Reporting Options here can be ignored. User Programs Options here can be ignored. 4. When all the details have been entered correctly, click on OK. This information can be edited at a later date, although editing the default polynomial order doesn t change the calibrations that have already been done. Changing the Column(s) and/or Eluent When putting on two columns in series, the higher molecular weight range column goes first. Before removing a column, purge it thoroughly with degassed double distilled water. Make sure no air gets into the column or tubing when removing a column from the GPC. When putting a new column onto the GPC, make sure no air gets into the tubing or columns. Purge it thoroughly with the eluent (until a stable baseline is seen with the Datastream Monitor program). Make sure the flow rate and pressure are not too high for the columns used. When changing the eluent, make sure the columns are properly purged with the new eluent. Shutting Down 1. Make sure the micro-syringe has been rinsed out with distilled water, it is empty, and the barrel left fully out after use. 2. Make sure the 2 ml syringe used to rinse out the sample injection port has been rinsed with distilled water. 3. Press FLOW CTRL, and then YES. 4. Shut down the computer. Peter Iddon Aqueous GPC (Detailed).doc Page 7 of 8

8 5. Switch off the computer, GPC pump, PL DataStream, and the RI detector. 6. The UV detector must be turned off to preserve the life of the lamp. 7. Flush the piston and piston seal area with at least 50 ml deionised water. 8. If you have finished analysing samples, please tick the box on the GPC queue sheet. More help with Cirrus software is needed Workbook does not exist error occurs Small & sharp negative peaks appear occasionally Air bubbles have formed in the pump or tubing PRESSURE FAULT is showing on the pump display Large numbers of air bubbles are forming in the eluent filter The baseline is not flat Extra peaks appear too early or too late The wrong order polynomial has been used for the calibration No data appears to have been collected after sample injection Common Problems For more detail on using the Cirrus software, either refer to the instruction manual, or press F1 at any time while using GPC Online or GPC Offline to show help information relevant to the current task. All the information in the instruction manual can be accessed through the Help menu. If a program says that a workbook does not exist when you click on one from the Recent list, try browsing for the workbook by clicking on the File menu and selecting Open. This can happen if workbook folders are moved or reorganised. This is normal, and happens when the PL Datastream is checked by the software, such as when GPC Online is opened. If air bubbles have formed in the pump or tubing, the system can be purged by first opening the pump purge valve (shown on the diagram of the pump controls). Press PURGE, then ENTER on the pump control panel. The pump will now pump at a high rate, and the eluent will emerge from the outlet just under the valve. To end the purge, press ESC on the pump control panel, and close the pump purge valve. This is usually due to air bubbles forming or being trapped in the pump head. Follow the instructions in the previous paragraph, then press ESC on the pump control panel to cancel the error message. If bubbles are continuously forming in the eluent filter, either degas the eluent again or make up fresh eluent. If this does not work, clean the solvent filter in 6M nitric acid in a sonic bath for 10 minutes (rinse very thoroughly afterwards). Check that the PURGE button on the RI detector is switched off. Make sure the detectors are zeroed before every injection. If the GPC has been on for a long time, it may be necessary to purge the RI detector for about 5 minutes. If extra peaks or a poor baseline appears after the run has finished, or before any sample is expected to elute though the columns, there may be retained polymer, monomer, solvent or salt. Either allow extra time in the method for all the components to elute properly, or change the eluent conditions to reduce adsorption of components to the columns. The polynomial order has to be set before the calibration is created. Change the setting (as described in the section Editing or Creating a New Method), then create a new calibration using GPC Offline (using the same data). Check that the PL DataStream has been turned on. Check that the PL DataStream is ready for data collection (by pressing F8 or selecting the option from the Collection menu). Last Updated: 1 July 2005 Peter Iddon Aqueous GPC (Detailed).doc Page 8 of 8

Manual HPLC-4 Alliance-UV

Manual HPLC-4 Alliance-UV Manual HPLC-4 Alliance-UV MAKE SURE NEVER TO: 1. pump particles or air through the column. 2. inject when the injector is dry. 3. inject solutions containing particles. 4. apply large (sudden) pressure

More information

Operating Manual. High Performance Liquid Chromatograph. Scientific Equipment Center, Prince Of Songkla University

Operating Manual. High Performance Liquid Chromatograph. Scientific Equipment Center, Prince Of Songkla University Operating Manual High Performance Liquid Chromatograph Agilent 1100 ; VWD, DAD, FLD and RID Scientific Equipment Center, Prince Of Songkla University Operating Manual High Performance Liquid Chromatograph

More information

PerkinElmer Series 200 HPLC: Operating Procedure

PerkinElmer Series 200 HPLC: Operating Procedure PerkinElmer Series 200 HPLC: Operating Procedure Chem Eng 5503 Version 1.01 Last Update 8/9/07 This document outlines the basic operating procedure for the PE series 200 HPLC, UV/Vis and SEDEX Light Scattering

More information

Equipment Overview: Safety:

Equipment Overview: Safety: PerkinElmer Series 200 HPLC: Operating Procedure 12/2/11 R.C Equipment Overview: This document outlines the basic procedure for using the Sedex light scattering detector and the PE series 200 HPLC system

More information

3 Starting the chiral HPLC and logging into LC Real Time Analysis.

3 Starting the chiral HPLC and logging into LC Real Time Analysis. Edition: 1.2 Date: 7-Oct-11 Page 1 of 5 1 Responsibilities. First person responsible: Mariëlle Delville. Second person responsible: Helene I.V. Amatdjais-Groenen. 2 Definitions. Read the information on

More information

Quick Setup Instructions for the Knauer HPLC-System. Quick Setup Instructions for the Knauer HPLC-System For windows XP (using CHANCE vers. 2.

Quick Setup Instructions for the Knauer HPLC-System. Quick Setup Instructions for the Knauer HPLC-System For windows XP (using CHANCE vers. 2. For windows XP (using CHANCE vers. 2.0) Compiled by P. Madl Vers. 1.0 June 2011 1 / 9 1. Hardware Setup: Result 1.1. Attach RS 232 interface cable (designated as PC) to COM-port-1 of the laptop. 1.2. connect

More information

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017 Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017 Introduction This document is only for operation of the HPLC and ICP-MS instruments to obtain data. It contains no maintenance

More information

Chromeleon software orientation

Chromeleon software orientation Chromeleon software orientation Upon opening of Chromeleon shortcut, a blue screen should appear (called control panel). If this does not occur, the green circled shortcut will open this screen. To ensure

More information

FROG-4000TM Quick Start Guide

FROG-4000TM Quick Start Guide FROG-4000TM Quick Start Guide 1 Precautions DO NOT invert the instrument with the sparge bottle attached. DO NOT transport or store the instrument with liquid in the sparge bottle. DO NOT handle or carry

More information

4 Instrument Instructions for CE

4 Instrument Instructions for CE 4 Instrument Instructions for CE Overview: the vial trays There are three instrumental steps in the experiment: a) Condition the capillary; b) Set up and, if necessary, optimize your method for a single

More information

Agilent 1100 HPLC ChemStation 1) Turn on all the modules for HPLC system (CapPump G1376A, DAD G1315B) and allow for system to initialize.

Agilent 1100 HPLC ChemStation 1) Turn on all the modules for HPLC system (CapPump G1376A, DAD G1315B) and allow for system to initialize. Standard Operating Procedure Remcho Research Group P a g e 1 Agilent 1100 HPLC ChemStation 1) Turn on all the modules for HPLC system (CapPump G1376A, DAD G1315B) and allow for system to initialize. 2)

More information

CH142: Instructions for the Vernier LabPro

CH142: Instructions for the Vernier LabPro 1 CH142: Instructions for the Vernier LabPro INTRODUCTION: The Vernier LabPro is a versatile data collection interface that can be used in combination with more than 45 sensors, and with a computer, TI

More information

Tutorial EN

Tutorial EN Tutorial 8.102.8001EN Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com Tutorial 8.102.8001EN 07.2009 ek Teachware Metrohm AG CH-9101

More information

QAM-I-102 Operation and Calibration of the Autoanalyzer

QAM-I-102 Operation and Calibration of the Autoanalyzer 1. Applicability and Purpose a. This procedure applies to the operation and calibration of the LaChat Quikchem 8000 Automated Ion Analyzer. An equivalent system may be used with proper revision of this

More information

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.0.4 and up GETTING STARTED FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.0.4 and up Getting started

More information

Agilent GC/MSD Instructions

Agilent GC/MSD Instructions Agilent GC/MSD Instructions GC/MS SAMPLE PREPARATION: Sample Components to Avoid Completely: The following should never be injected: metals, strong acids or bases, salts, oligomeric and polymeric material.

More information

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Section 1: Starting an experiment 1.1 Opening a file 1. Open the File menu and select Open Activity. 2. In the Open dialog box, navigate to

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in the Thermo TraceFinder 3.0 analytical software. For detailed descriptions

More information

Getting low-resolution GC-MS data on the JEOL GCMate Introduction

Getting low-resolution GC-MS data on the JEOL GCMate Introduction Getting low-resolution GC-MS data on the JEOL GCMate Introduction GC-MS is a technique that combines gas phase separation and mass-based detection/characterization of analytes. It s a very sensitive method

More information

Sepacore X10 / X50 system Technical data sheet

Sepacore X10 / X50 system Technical data sheet Sepacore X10 / X50 system Technical data sheet Sepacore X10 and X50 flash chromatography systems address most requirements for the purification of organic compounds. Whether a crude synthesis mixture or

More information

Introduction to the HPLC ChemStation and Acquisition

Introduction to the HPLC ChemStation and Acquisition Introduction to the HPLC ChemStation and Acquisition In This Section, We Will Discuss: How to work in the Microsoft Windows Environment The structure of the ChemStation Software. How to set up an acquisition

More information

Waters Alliance GPC 2000 Series System

Waters Alliance GPC 2000 Series System Waters Alliance GPC 2000 Series System Quick Start Guide 34 Maple Street Milford, MA 01757 71500021503, Revision A NOTICE The information in this document is subject to change without notice and should

More information

Dionex AS-AP Sample Conductivity and ph Accessory Setup and Operation Guide

Dionex AS-AP Sample Conductivity and ph Accessory Setup and Operation Guide Dionex AS-AP Sample Conductivity and ph Accessory Setup and Operation Guide Document No. 065470 Revision 02 February 2012 2012 by Thermo Fisher Scientific Inc. All rights reserved. Chromeleon is a registered

More information

Sciex QTrap Operational Steps for Trained Personnel

Sciex QTrap Operational Steps for Trained Personnel Sciex 6500+ QTrap Operational Steps for Trained Personnel Last Updated 09172017 1. If any of the following instructions does not make sense to you or was not covered during your hand-on training, stop

More information

Montgomery County Community College Document Number: DP DeKalb Pike Revision Number: 2. Page 1 of 9 SOP: BioLogic LP Chromatography System

Montgomery County Community College Document Number: DP DeKalb Pike Revision Number: 2. Page 1 of 9 SOP: BioLogic LP Chromatography System Page 1 of 9 Approvals Preparer: Jason McMillan Reviewer: Dr. Maggie Bryans Date: 18DEC14 Date: 18DEC14 1. Purpose: Operation of the BioLogic LP Chromatography System. 2. Scope: Applies to the BioLogic

More information

Agilent MassHunter Workstation Software 7200 Accurate-Mass Quadrupole Time of Flight GC/MS

Agilent MassHunter Workstation Software 7200 Accurate-Mass Quadrupole Time of Flight GC/MS Agilent MassHunter Workstation Software 7200 Accurate-Mass Quadrupole Time of Flight GC/MS Familiarization Guide Before you begin 3 Prepare your system 3 Prepare the samples required for data acquisition

More information

Chromatography Software Training Materials. Contents

Chromatography Software Training Materials. Contents Chromatography Software Training Materials This document contains information on how to build a method, start the instrument to acquire data, and then process the data using the Galaxie Program. You will

More information

Standard Operating Procedure for the Horiba FluroMax-4

Standard Operating Procedure for the Horiba FluroMax-4 Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.

More information

Quick Start Guide. Introduction. Importing Raw Data. Method 1: Use Open Data Panel

Quick Start Guide. Introduction. Importing Raw Data. Method 1: Use Open Data Panel ACD/Spectrus Processor: Chromatography Data Irina Oshchepkova and Laura Zepeda Advanced Chemistry Development, Inc. Toronto, ON, Canada www.acdlabs.com Introduction The following document outlines how

More information

The Olis Titrator Series

The Olis Titrator Series The Olis Titrator Series Perfect Titrations into any type of cuvette With considerable effort, one can do precise titration studies by hand. Or, one can let the OLIS titrators do the technician s work

More information

Steps for LCMS-8040 Triple Quadrupole

Steps for LCMS-8040 Triple Quadrupole Steps for LCMS-8040 Triple Quadrupole June 2017 Version 2 A. Data Acquisition 1. Check lights on MS. There should be a total of three lights lit by LED. One of them is blinking. Power Blinking Gas Status

More information

Procedures for Group Quantitation

Procedures for Group Quantitation Procedures for Group Quantitation Global Application Development Center Analytical & Measuring Instruments Division Shimadzu Corporation Group Quantitation In anionic surfactant analyses based on inspection

More information

Flexar FX-20 LC Pump: Ultimate Performance and Flexibility for UHPLC and HPLC

Flexar FX-20 LC Pump: Ultimate Performance and Flexibility for UHPLC and HPLC P R O D U C T N O T E Liquid Chromatography Key Features Improved productivity compatible with all UHPLC columns, the Flexar FX-20 Pump can operate at pressures as high as 18,000 psi, providing throughput

More information

ChromQuest 5.0 Quick Reference Guide

ChromQuest 5.0 Quick Reference Guide ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User

More information

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample

More information

Data Handling and Reports

Data Handling and Reports Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA Star Chromatography Workstation Version 6 Data Handling and Reports Tutorials Varian, Inc. 2002 03-914736-00:Rev. 4 Trademark

More information

Isolera. User Manual

Isolera. User Manual Isolera User Manual Safety and Document Conventions Isolera system should only be operated and maintained by trained individuals. Please read this manual carefully before working with the system. To guarantee

More information

Chromeleon-PA Analyzer User s Guide Dionex Corporation

Chromeleon-PA Analyzer User s Guide Dionex Corporation Chromeleon-PA Analyzer User s Guide 2003 Dionex Corporation Document No. 031964 Revision 01 October 2003 2003 by Dionex Corporation All rights reserved worldwide. Printed in the United States of America.

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in Thermo TraceFinder analytical software. For detailed descriptions

More information

MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL

MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL 470 350/REV A/10-03 MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1-2 Installation 2 Displays

More information

HITACHI LACHROM ELITE

HITACHI LACHROM ELITE HITACHI LACHROM ELITE Clarity Control Module ENG Code/Rev.: M133/60D Date: 7/16/2018 Phone: +420 251 013 400 DataApex Ltd. Fax: +420 251 013 401 Petrzilkova 2583/13 clarity@dataapex.com 158 00 Prague 5

More information

ibidi PumpControl Instruction Manual Version 1.5.0

ibidi PumpControl Instruction Manual Version 1.5.0 ibidi PumpControl Instruction Manual Version 1.5.0 II ibidi GmbH 2011, version 1.5.0, revision 2, 2011-10-26, Andreas Rainer Table of contents 1 General Information... 1 1.1 System Requirements... 1 1.2

More information

ChromGraph BAS-200B CONTROL

ChromGraph BAS-200B CONTROL BAS-200B CONTROL January, 1998 MF-9070 INSTRUCTION MANUAL ChromGraph BAS-200B CONTROL Bioanalytical Systems, Inc 2701 Kent Avenue West Lafayette Indiana 47906 MANUFACTURER S NOTE This instrument, either

More information

Flexar FX-20 LC Pump: Ultimate Performance and Flexibility for UHPLC and HPLC

Flexar FX-20 LC Pump: Ultimate Performance and Flexibility for UHPLC and HPLC P R O D U C T N O T E Liquid Chromatography Key Features Improved productivity compatible with all UHPLC columns, the Flexar FX-20 Pump can operate at pressures as high as 18,000 psi, providing throughput

More information

Waters 600E Multisolvent Delivery System Quick Start Guide

Waters 600E Multisolvent Delivery System Quick Start Guide Waters 600E Multisolvent Delivery System Quick Start Guide 34 Maple Street Milford, MA 01757 WAT174-04TP, Revision 3 NOTICE The information in this document is subject to change without notice and should

More information

MICROFLUIDIC EOF KIT 9015

MICROFLUIDIC EOF KIT 9015 MICROFLUIDIC EOF KIT 9015 User Manual - Version 1.1 Visiting address Colosseum 15 7521 PV Enschede Postal address P.O. Box 214 7500 AE Enschede The Netherlands Phone +31 53 850 6 850 Fax +31 53 850 6 851

More information

Making your first runs

Making your first runs Making your first runs Begin here with ÄKTA FPLC 18-1140-80 Important user information All users must read this entire manual to fully understand the safe use of ÄKTAFPLC. WARNING! The Warning sign highlights

More information

GE Healthcare. Getting Started with

GE Healthcare. Getting Started with GE Healthcare Getting Started with ÄKTAavant and UNICORN 6 Table of Contents Table of Contents 1 About this guide... 1.1 Get started... 1.2 Regulatory information... 2 Safety instructions... 2.1 Safety

More information

Agilent 6400 Series Triple Quadrupole LC/MS System

Agilent 6400 Series Triple Quadrupole LC/MS System Agilent 6400 Series Triple Quadrupole LC/MS System Quick Start Guide Where to find information 4 Getting Started 6 Step 1. Start the Data Acquisition software 7 Step 2. Prepare the LC modules 13 Step 3.

More information

Thermo Xcalibur Getting Started (Quantitative Analysis)

Thermo Xcalibur Getting Started (Quantitative Analysis) Thermo Xcalibur Getting Started (Quantitative Analysis) XCALI-97207 Revision B September 2010 2010 Thermo Fisher Scientific Inc. All rights reserved. Xcalibur, Surveyor, and Accela are registered trademarks

More information

UniPoint System Software User s Guide

UniPoint System Software User s Guide UniPoint System Software User s Guide LT2137 1998 Gilson, Inc. All rights reserved Exercise 3-Creating operations list... 4-22 Create and name operations list... 4-22 Set up steps... 4-23 Identify sample

More information

Gallios TM Quick Reference

Gallios TM Quick Reference Gallios TM Quick Reference Purpose: The purpose of this Quick Reference is to provide a simple step by step outline of the information needed to perform various tasks on the system. We begin with basic

More information

ACQUITY UPLC Systems with 2D Technology Capabilities Guide

ACQUITY UPLC Systems with 2D Technology Capabilities Guide ACQUITY UPLC Systems with 2D Technology Capabilities Guide Revision A Copyright Waters Corporation 2012 All rights reserved Copyright notice 2012 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA

More information

2. create the workbook file

2. create the workbook file 2. create the workbook file Excel documents are called workbook files. A workbook can include multiple sheets of information. Excel supports two kinds of sheets for working with data: Worksheets, which

More information

Standard Operating Procedure for the Beckman PA/CE MDQ Electrophoresis System. Last Modified October 19, 2012

Standard Operating Procedure for the Beckman PA/CE MDQ Electrophoresis System. Last Modified October 19, 2012 Standard Operating Procedure for the Beckman PA/CE MDQ Electrophoresis System Last Modified October 19, 2012 Considerations Prior to Beginning Experiment All solutions must be filtered (using 0.45 m or

More information

For additional information, see these documents on the Breeze 2 software DVD:

For additional information, see these documents on the Breeze 2 software DVD: R E L E A S E N O T E S Waters Breeze 2 Read these release notes before you install and operate Breeze 2. They contain information necessary to ensure successful liquid chromatography results, including:

More information

New software features of the 7010 Particle Size. Analyzer Danielle Chamberlin May May 5, Page 1

New software features of the 7010 Particle Size. Analyzer Danielle Chamberlin May May 5, Page 1 New software features of the 7010 Particle Size May 5, 2009 Analyzer 2.0.0 Danielle Chamberlin May 5 2009 Page 1 Introduction to the new 7010 software Instrument operation New look and feel easy to operate

More information

Guidebook for Installation and Familiarization. HP 1100 Series Combinatorial Chemistry Analysis System

Guidebook for Installation and Familiarization. HP 1100 Series Combinatorial Chemistry Analysis System Guidebook for Installation and Familiarization HP 1100 Series Combinatorial Chemistry Analysis System Copyright Hewlett- Packard Company 1997 All rights reserved. Reproduction, adaption, or translation

More information

Using SIMDIS Expert 9 (Separation Systems) with Clarity

Using SIMDIS Expert 9 (Separation Systems) with Clarity Using SIMDIS Expert 9 (Separation Systems) with Clarity The SIMDIS Expert 9 may be used with Clarity or Clarity Lite software to process the acquired chromatograms according to Simulated Distillation.

More information

Waters 2707 Autosampler ICS v1.20 Installer

Waters 2707 Autosampler ICS v1.20 Installer R E L E A S E N O T E S Waters 2707 Autosampler ICS v1.20 Installer Read these release notes before you install and operate the Waters 2707 autosampler instrument control software (ICS). They contain information

More information

CALIBRATION of a STREAM

CALIBRATION of a STREAM CALIBRATION of a STREAM PRELIMINARY SET UP: STANDARD GAS CYLINDER @ 5 PSIG; FLOW METER: 20/30 CC/MIN. DIP TUBE LIQUIDS AND WELKER PISTON CYLINDERS ACCORDINGLY. ALL OPERATIONS ARE INITIATED FROM THE MAIN

More information

Quick Guide to the Star Bar

Quick Guide to the Star Bar Saturn View Saturn Writer Quick Guide to the Star Bar Last active Chromatogram in SaturnView Last active method in the Method Editor Click with the right mouse button on the Star Bar to get this menu Sample,

More information

Canto II Operation Protocol

Canto II Operation Protocol Canto II Operation Protocol Make sure the following actions have been taken before running your samples. Make sure the waste tank is empty and refilled with 500 ml clean (see page 15 for details) Make

More information

Jasco 4700 FTIR. Laser Spectroscopy Labs. Jasco 4700 FTIR. Operation instructions. Laser Spectroscopy Labs, UCI

Jasco 4700 FTIR. Laser Spectroscopy Labs. Jasco 4700 FTIR. Operation instructions. Laser Spectroscopy Labs, UCI Laser Spectroscopy Labs Jasco 4700 FTIR Operation instructions Turn the system on 1. Turn on the computer in case it has been switched off. Normally computer stays on. 2. The instrument should be always

More information

Hitachi U-2001 UV/Vis Spectrometer Updated November 14, 2017

Hitachi U-2001 UV/Vis Spectrometer Updated November 14, 2017 Hitachi U-2001 UV/Vis Spectrometer Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml If you have any problems with the

More information

DataPro Quick Start Guide

DataPro Quick Start Guide DataPro Quick Start Guide Introduction The DataPro application provides the user with the ability to download and analyze data acquired using the ULTRA-LITE PRO range of Auto Meter products. Please see

More information

PART 7. Getting Started with Excel

PART 7. Getting Started with Excel PART 7 Getting ed with Excel When you start the application, Excel displays a blank workbook. A workbook is a file in which you store your data, similar to a three-ring binder. Within a workbook are worksheets,

More information

MS Word MS Outlook Mailbox Maintenance

MS Word MS Outlook Mailbox Maintenance MS Word 2007 MS Outlook 2013 Mailbox Maintenance INTRODUCTION... 1 Understanding the MS Outlook Mailbox... 1 BASIC MAILBOX MAINTENANCE... 1 Mailbox Cleanup... 1 Check Your Mailbox Size... 1 The Quota

More information

Soprane II Documentation

Soprane II Documentation Soprane II Documentation p. 1 / 198 Table of contents 1. INTRODUCTION 6 2. INSTALLATION 6 3. ANALYZER CONFIGURATION 8 3.1. Creation 9 3.1.1. Analyzer creation with network connection 11 3.1.2. Analyzer

More information

Importing data in a database with levels

Importing data in a database with levels BioNumerics Tutorial: Importing data in a database with levels 1 Aim In this tutorial you will learn how to import data in a BioNumerics database with levels and how to replicate and summarize level-specific

More information

TopPoint User's Guide

TopPoint User's Guide TopPoint User's Guide LT2152/ 2000 Gilson, Inc. All rights reserved August 2000 Table of Contents 1 Introduction 2 Operation 3 Troubleshooting Steps to Using TopPoint... 1-2 Accessing Online Help... 1-3...

More information

Client Care Desktop V4

Client Care Desktop V4 Client Care Desktop V4 V4.1 Quay Document Manager V4.1 Contents 1. LOCATIONS...3 1.1. Client... 3 1.2. Holdings... 4 1.3. Providers... 4 1.4. Contacts/Introducers... 5 1.5. Adviser... 5 2. ADDING EXISTING

More information

Introduction to Microsoft Office 2016: Word

Introduction to Microsoft Office 2016: Word Introduction to Microsoft Office 2016: Word Last Updated: September 2018 Cost: $2.00 Microsoft Word is a word processing software. You can use it to type letters, reports, and other documents. This class

More information

HEL Crystal Eyes Manual

HEL Crystal Eyes Manual HEL Crystal Eyes Manual Contents 1. Introduction... 3 2.Temperature Probe... 3 2.1. Temperature Calibration... 3 3. Turbidity Probe... 3 3.1. Set up of the reflection probe... 4 3.2. Turbidity calibration...

More information

Using Microsoft Excel

Using Microsoft Excel About Excel Using Microsoft Excel What is a Spreadsheet? Microsoft Excel is a program that s used for creating spreadsheets. So what is a spreadsheet? Before personal computers were common, spreadsheet

More information

ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE

ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE 1.2.80.1071 07MAN1181.A1 Effective Date: 30-Jan-11 ECO #: 3093 Contents What are Quick Set-Up Instructions?... 3 Starting up ClonePix 2...

More information

Unpacking and Installing the Flora 2512 UV Printer. Steps 1: Unscrew the 10mm bolts holding the top. Then remove the top and put in a safe place.

Unpacking and Installing the Flora 2512 UV Printer. Steps 1: Unscrew the 10mm bolts holding the top. Then remove the top and put in a safe place. Unpacking and Installing the Flora 2512 UV Printer Steps 1: Unscrew the 10mm bolts holding the top. Then remove the top and put in a safe place. Step 2: Unscrew 10mm bolts holding the end panels. On the

More information

Biolight Luminescence System GmbH ( SOFTWARE MANUAL. BioLight Studio Software. Version

Biolight Luminescence System GmbH (  SOFTWARE MANUAL. BioLight Studio Software. Version Biolight Luminescence System GmbH (www.biolight.com) SOFTWARE MANUAL BioLight Studio Software Version 1.04.00 BL Studio Software Manual Notice This document is being furnished by BioLight Luminescence

More information

The first time you open Word

The first time you open Word Microsoft Word 2010 The first time you open Word When you open Word, you see two things, or main parts: The ribbon, which sits above the document, and includes a set of buttons and commands that you use

More information

Tube Washing Upgrade Guide. for Epson SureColor F2000

Tube Washing Upgrade Guide. for Epson SureColor F2000 Tube Washing Upgrade Guide for Epson SureColor F2000 2016 guides.equipmentzone.com Page 1 of 14 Part 1 -How to Upgrade Epson F2000 Firmware Open your internet browser Select address bar located up top

More information

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette Pipettes allow transfer of accurately known volume of solution from one container to another. Volumetric or transfer pipettes deliver a

More information

CyAn ADP Guide. Starting Up

CyAn ADP Guide. Starting Up CyAn ADP Guide Starting Up 1. Check the sheath and waste fluids, if needed, fill and empty: 1. To fill the Sheath Tank: a. Unscrew the top of the tank labeled Sheath Tank, and set it on top of the tank.

More information

Lesson 2. Using the Macro Recorder

Lesson 2. Using the Macro Recorder Lesson 2. Using the Macro Recorder When the recorder is activated, everything that you do will be recorded as a Macro. When the Macro is run, everything that you recorded will be played back exactly as

More information

Client Care Desktop v4.3. Document Manager V4.3

Client Care Desktop v4.3. Document Manager V4.3 Client Care Desktop v4.3 Document Manager V4.3 Contents 1. LOCATIONS... 3 1.1. Client... 3 1.2. Enquiries... 4 1.3. Holdings... 4 1.4. Providers... 5 1.5. Contacts/Introducers... 6 1.6. Adviser... 6 2.

More information

Excel Select a template category in the Office.com Templates section. 5. Click the Download button.

Excel Select a template category in the Office.com Templates section. 5. Click the Download button. Microsoft QUICK Excel 2010 Source Getting Started The Excel Window u v w z Creating a New Blank Workbook 2. Select New in the left pane. 3. Select the Blank workbook template in the Available Templates

More information

EnCal 3000 Gas Chromatograph Software Manual d

EnCal 3000 Gas Chromatograph Software Manual d EnCal 3000 Gas Chromatograph Software Manual 73022345 d 27.08.2015 Contact Information... 2 Gas quality measurement system EnCal 3000... 3 1 Introduction... 5 1.1 System Requirements... 5 1.2 Installation

More information

HI FORMOL NUMBER MINITITRATOR for Fruit Juices and Wines analysis

HI FORMOL NUMBER MINITITRATOR for Fruit Juices and Wines analysis Instruction Manual HI 84533 FORMOL NUMBER MINITITRATOR for Fruit Juices and Wines analysis www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction

More information

LSRFortessa Guide. Cell Evaluation and Therapy Core HO324

LSRFortessa Guide. Cell Evaluation and Therapy Core HO324 LSRFortessa Guide Cell Evaluation and Therapy Core HO324 Before Use: Prime the instrument twice with no tube and arm to the side Ensure that sheath is full and waste is empty To Begin: Log into ilabs tracker

More information

Agilent Max-Light Cartridge Cell Information for G4212A/B DAD and G7117A/B DAD

Agilent Max-Light Cartridge Cell Information for G4212A/B DAD and G7117A/B DAD Agilent Max-Light Cartridge Cell Information for G4212A/B DAD and G7117A/B DAD Information about the Max-Light Cartridge Cells 2 Specifications 2 Description of the dispersion volume V(σ) 2 Recommendations

More information

Section 3 Formatting

Section 3 Formatting Section 3 Formatting ECDL 5.0 Section 3 Formatting By the end of this Section you should be able to: Apply Formatting, Text Effects and Bullets Use Undo and Redo Change Alignment and Spacing Use Cut, Copy

More information

Agilent 6300 Ion Trap LC/MS Systems Quick Start Guide

Agilent 6300 Ion Trap LC/MS Systems Quick Start Guide Agilent 6300 Ion Trap LC/MS Systems Quick Start Guide A guide to help you get started with the Agilent 6300 Series Ion Trap LC/MS system What is the Agilent 6300 Series Ion Trap LC/MS System? 2 Where to

More information

870 KF Titrino plus. Installation and tutorial EN

870 KF Titrino plus. Installation and tutorial EN 870 KF Titrino plus Installation and tutorial 8.870.8004EN Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 870 KF Titrino plus Installation

More information

Data Acquisition with CP-2002/2003 Micro-GC Control

Data Acquisition with CP-2002/2003 Micro-GC Control Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598 Star Chromatography Workstation Version 6 Data Acquisition with CP-2002/2003 Micro-GC Control Operation Manual Varian, Inc. 2002

More information

Detailed procedures, operation instructions, maintenance, and emergency contact information list is attached.

Detailed procedures, operation instructions, maintenance, and emergency contact information list is attached. GC/FID Standard Operating Procedure Environmental health and Safety (EH&S) 2809 Daley Drive Ames, IA 50011-3660 Phone: (515)-294-5359 Fax: (515)-294-9357 www.ehs.iastate.edu Procedure: Department: Bioeconomy

More information

Q Exactive HF-X QuickStart Guide

Q Exactive HF-X QuickStart Guide Q Exactive HF-X QuickStart Guide About this Guide Q Exactive HF-X Tune This Q Exactive HF-X QuickStart Guide gives an introduction on setting up and using the Thermo Scientific Q Exactive HF-X mass spectrometer.

More information

Introduction to Microsoft Excel 2007

Introduction to Microsoft Excel 2007 Introduction to Microsoft Excel 2007 Microsoft Excel is a very powerful tool for you to use for numeric computations and analysis. Excel can also function as a simple database but that is another class.

More information

CPM-200 User Guide For Lighthouse for MAX

CPM-200 User Guide For Lighthouse for MAX CPM-200 User Guide For Lighthouse for MAX Contents Page Number Opening the software 2 Altering the page size & Orientation 3-4 Inserting Text 5 Editing Text 6 Inserting Graphics 7-8 Changing the Colour

More information

UNC Flow Cytometry Core Facility LSR II Operation 2018

UNC Flow Cytometry Core Facility LSR II Operation 2018 Reference: Startup BD FACSDiva 6 Online Course BD Training & elearning 1. Check Sheath and Waste containers: Empty Waste container in the sink Add bleach to the black mark ( ~10% final concentration).

More information

SPA III Quick Reference Guide for Custom Panels

SPA III Quick Reference Guide for Custom Panels SPA III Quick Reference Guide for Custom Panels This guide contains instructions for using the BD FACS Sample Prep Assistant III (SPA III) with BD FACS SPA software version.0 and later. Initial Setup Custom

More information

Agilent G2070BA GC ChemStation

Agilent G2070BA GC ChemStation Agilent G2070BA GC ChemStation Familiarization Guide Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in any form or by any means (including electronic

More information