4 Instrument Instructions for CE

Size: px
Start display at page:

Download "4 Instrument Instructions for CE"

Transcription

1 4 Instrument Instructions for CE Overview: the vial trays There are three instrumental steps in the experiment: a) Condition the capillary; b) Set up and, if necessary, optimize your method for a single sample; c) Set up and run a sequence, including all your samples and standards. You will need vials of liquid for each step. Although the steps are independent, you should prepare all the vials in advance and place them in the holders before starting the conditioning. The paragraphs that follow discuss each step, but before that we will have a quick look at the vial holders. The CE contains three vial trays. At the front left is the inlet vial tray; to the right is the outlet vial tray; and at the back left is the sample inlet tray. The front trays, which are identical, look like this: A B C D E F Inlet vial tray A B C D E F Outlet vial tray The sample inlet tray, at the rear, is similar, but has eight rows instead of six. Your samples, standards, spiked solutions and QC all go in the rear tray; remaining solutions go in the front trays. The inlet vial tray holds liquid used to condition or wash the column, and also buffer used during the separation. The outlet vial tray holds empty vials, into which waste is delivered, and more vials of buffer. So how do all these vials get used? When you condition the column, or do a rinse, liquid is extracted from a vial in the inlet vial tray, passed through the capillary, and dumped into an empty vial in the outlet vial tray. During a separation, all three trays are used. First the capillary is washed, again by extracting liquid from the inlet tray and expelling it to a vial in the outlet tray. Then a sample is selected from the rear tray and injected into the capillary. Finally, the left end of the capillary is placed in buffer in the left front tray, while the other end goes into buffer in the right front tray; the voltage is applied and the separation begins. The rinse solvents are gradually consumed and the separation buffer gets contaminated, so fresh solution is required periodically. However, you do not need to refill the vials. There is enough room on the trays for several vials of each solvent and of buffer, so instead of using one vial of each liquid you will use several. How many? Well, you will need fresh solvents and buffers for every four samples. Therefore, if you have, say, 17 samples (including standards, QC, ) you will need 5 sets of solvents and buffer. You will also need a set of the preconditioning solvents; see step 4 below.

2 So, let's get started. 1. Power on: Turn on the instrument by pressing the power button on the front. The CE takes several minutes to boot up, so start it 10 minutes before you need to use it. 2. CARE!: You can cause serious damage by opening the plastic cover whilst the instrument is moving the vial trays. Do not open the cover while the instrument is going through the startup. 3. Prepare the sample vials: Work carefully so that you do not muddle up your vials; your experiment will fail if the wrong vial is used for a separation. For each of your samples, standards etc. label a vial, fill it just to the shoulder and cap the vial. Count the total number (N) of standards, samples, spiked samples and QC sample vials you have. 4. Prepare the solvent and buffer vials: For every four of your N samples, you need (i) 3 vials of buffer, (ii) 1 vial of 0.1M NaOH, (iii) 1 vial of DI water, and (iv) 1 empty vial. In addition, for the preliminary conditioning you will need a further 3 vials of buffer, 1 vial of HPLC grade methanol, 1 vial of 0.1M NaOH, 1 vial of DI water, and an empty vial. Prepare all these vials and label clearly. 5. Start the software: Once the CE has completed its start up, turn on the computer and ask your TA to log on for you. Start the 32Karat 8.0 application on the desktop, then open the MDQ 2 module. When this new window opens, an Instrument Wizard window will open automatically. You don't need this yet, so click OK to close this wizard. 6. Initial control: MDQ2 establishes communication between the computer and the instrument. Under Control, select Direct Control, then View. This will create a window you can use for several instrument commands. Click the Load button. You will hear noise as the instrument brings the buffer trays forward for loading. Once the noise stops, open the cover. 7. Insert the vials for the conditioning: Lift the vial trays carefully out of the instrument. Place two buffer vials in positions A1 and A2 of the inlet tray; place the MeOH, NaOH and water vials in positions A3-A5 of the same tray. Place a second buffer vial in position A1 of the outlet vial tray and an empty vial (with lid) in position B1 of the outlet tray. Record the position of each vial. 8. Insert the remaining inlet solvent and buffer vials: Recall that for each four samples, you need a fresh set of wash solutions and buffers. From position A6 onwards on the inlet tray continue to add vials: two buffer solutions, an NaOH vial and a water vial for each set of four samples. The order of these vials does not matter, but you must make a note of where each vial is, as you must specify those positions later. You will find that a (neat and clear!) diagram will be very useful. 9. Insert the remaining outlet vials: Insert the remaining buffer and empty vials in the outlet tray. Do not use columns D, E or F on the outlet tray, or the trays may clash during operation. Again, record the position of each vial. 10. Insert the sample vials: Finally, place your standards, samples, QC, etc. in the sample inlet vial tray. As before, note the location of each vial. Return the trays to the instrument, making sure they are correctly seated, and close the cover. The vial trays will move back to the Home position. 11. Condition the Capillary: You will now prepare a fresh surface on the inside of the capillary, using the Direct Control window. First, flush the capillary with water. To do this, move your mouse over the Pressure reading in the Direct Control window. The cursor will change to a small dial. Click on this box, to open a dialog box. Enter a

3 pressure of 25 psi and a duration time of 2 minutes, then click on the Trays button to select the vial positions. On the inlet side, select the position in which you placed the vial of water. On the outlet side, select vial B1, where you placed the empty vial. This will serve as your waste vial for the conditioning. Check that forward direction and pressure type are selected, then click OK to close the Trays window. In the Rinse Window the tray positions are specified, with the BI: and BO: before the letter and numbers of the vial positions standing for the buffer inlet tray and buffer outlet tray, respectively. Click OK again to start the rinse. The vial trays will move as the instrument begins this operation. 12. Rinse in progress: You will see a little cloud in the Direct Control window, on the lefthand side by the thermometer, as pressure is applied and the capillary is rinsed. If you use the small LED torch that sits on top of the AA PC and shine it through the transparent window on the CE, you may be able to see drips coming out of the outlet end of the capillary as the rinse proceeds. 13. Continuing the conditioning: Check the countdown clock in the Direct Control Window. Wait for the rinse to complete (pressure will fall to zero) before starting the next rinse. Run the following additional rinses, in order, in the same manner. For the buffer rinse, choose the second buffer vial on the inlet tray that you put in place for this purpose (i.e. a rinse buffer vial). This should not be the vial in position A1. Solution Pressure (psi) Time (min) Methanol 25 1 Water 25 2 NaOH 20 3 Water 20 2 Buffer 20 2 Now the capillary should be like new and you can build a method. 14. Build a Method: A method combines several steps, including (each time) preconditioning of the column, sample injection, and applying voltage for the separation. To build a method, you need an offline instrument window. Leave the online window open, but go back to the 32 Karat main screen. Right click on the MDQ 2 module and select Open Offline to create an offline instrument window. Click OK to close the instrument wizard window that appears. Select File/Method/New from the menu bar, and then Method/Instrument Setup. A window with three tabs will open. Set the detection wavelength to 214 nm using the UV Detector Initial Conditions tab and the Wavelength option. Under the Electropherogram channel heading, select Acquisition enabled in order to collect and save data. The rest of the settings can be left as they are. 15. Using the Time tab: Select the Time Program tab. The Time Program window is arranged like a spreadsheet with one event per line, executed in order from top to bottom. To program a line, click the Event box. Click on the down arrow to open a menu of events. Select an event to open a dialog box for that event. In your first line, select Rinse pressure as the event. The Rinse dialog box will appear. The first rinse will be a pressure rinse in the forward direction, for two minutes, at 20 psi, with 0.1 M NaOH. The outlet for all rinses in this conditioning step will be the empty vial in the outlet tray (position B1). You can select the position of the trays graphically, by clicking on the Trays button. Don't worry about all the coloured vials that may be displayed; just click on the two positions required for this rinse and they will be highlighted in red. Select OK once

4 you have entered all this information. For the next 2 lines, select the following rinses: Solution Pressure (psi) Duration (min) Water 20 2 Buffer 20 3 After these rinses, the fourth line of your method will be for the injection of the sample. Click on the Event box and select Inject. In the Inject dialog box, enter a time of 5 seconds at a pressure of 0.5 psi. The pressure and time are small to ensure a small injection, which minimizes band broadening. Select the sample inlet tray position for the sample you want to inject. The outlet position should be the buffer vial you will use for the subsequent separation. You will be creating a sequence later; for now, select the option to Allow Override under Sequence Table. 16. Program the separation: On the next line, click on the Event box and select Separate from the drop-down menu. The Separate dialog controls the conditions for the separation process. You will be separating by voltage, so enter a voltage of 20 kv, and a duration time of 14 minutes. The tray positions must be specified so that both the inlet and the outlet of the capillary are in buffer. Remember, use fresh buffer every four runs. Enter a time value of 0 in the time column next to the Separate action. 17. Autozero: The next and last line of your method will be to autozero the detector. Under the Event drop-down menu, select Autozero. Enter a time of 0.5 minutes, and then click OK. 18. Save the method: Your method is now complete. Save it by selecting File/Method/Save As in the C:\CE data\chem361 directory. Create a folder in this directory (click on the icon of a file with a sparkle in the top of the window) with your own name for saving your methods, sequence, and data files. 19. Prepare to run the first method: Do not close the offline module you used to build the method, but return to the active MDQ 2 module. This is the initial window you opened when you were conditioning the capillary. Load the new method by clicking File/Method/Open. Select your method to load it. To make sure you are viewing it properly, select Method/Instrument Setup. Check that the uv light is on by looking to see that the relevant light is lit on the front of the instrument. To perform a single run, select Control/Single Run from the menu bar. Edit the data path so that your results are sent to the correct directory. You must provide a Sample ID and Data File. For the Data File, you may name your file after the Sample ID, Date, and/or several other options. You can view these by clicking on the arrow to the right of the Data File box. The remaining settings can be left unchanged. 20. Run the method: Click Start to begin the run. Don't open the cover after Start has been clicked! To begin with, nothing much will appear to happen, apart from a message "Running sample " below the display. However, once the washing steps have been completed, the electropherogram should start to appear in the window; if it does not, click on Window/UV to reveal the display. 21. View the results: After the sample has run, you can view it in the offline window. Go back to this window and select File/Data/Open. Find your data file and open it. You can switch to the UV (detector) window by choosing

5 Window/UV 214nm. 22. Optimize the method: If the separation is inadequate, improve your method by changing the separation voltage (see step 17), and do another run to see the effect. Continue trying new voltages until you are happy with the result. In a good separation the peaks are well resolved, and the run time minimized. It may help in your choice of a voltage to know that the time at which a peak appears is roughly inversely related to the separation voltage. 23. Build the rest of the methods: Once you have a method you are happy with, you will need to build several methods to run all your samples. Each method will be slightly different since, for every four runs, you will be using different buffer, rinse buffer, NaOH rinse, water and empty waste vials. Be sure to record the information for each method so you can identify them later. Once you have prepared these methods, you can build a sequence. 24. Building a sequence: A sequence is a list of methods and data files that will be used to run a batch of samples. To create a new sequence, select File/Sequence/New from the Instrument window. Scroll over to the Method column. Click on the grey button in the box and select the method you want to use. This will automatically fill in most of the other boxes in the row. Because you allowed overrides when you built the method, you can change the Sample Inject Inlet and Outlet locations. Leave the run type as unknown for all runs. Fill in the Sample ID column with something to identify your sample. For example, standard 1 could be entered for standard #1. You can change the number of replicates for a sample by changing the number in the Reps column. Remember to use your different methods as you built them for the different buffer locations. 25. Save the sequence: Once you have finished the list for your injections, go to the Filename column. Double click in this box (NOT on the little green square in the bottom right of the square. If you click this it will open a window. Cancel to get out of this window). Enter a file name that includes your initials at the beginning and something to identify the sample. At the end of the name enter <D> to add the date and time to the filename. After you have entered all your runs, in the last line of the sequence, select the lamp off.met method from the C:\CE Data\Chem361 directory; this will turn off the uv lamp when all runs are completed. When you have finished editing the table, select File/Sequence/Save As from the menu bar. Give the sequence a suitable name and save it in the same location as your method and data files. 26. Check that the detector lamp is on: If you run your sequence with the uv lamp off, every method will fail in turn because of the lack of a detector signal, so check the indicator lamp at the front of the instrument to be sure the lamp is on. If it is not, go to the Instrument status window under Control and turn it on. 27. Start the sequence run: Before you leave the laboratory, start the sequence by clicking the green arrow icon, or selecting File/Sequence/Start. Bring up the uv signal window and check that all is well before you leave; the sequence will run overnight. A TA or SLI will print out the data the following day and leave it in the folder outside the laboratory for you to collect, BUT see point 28 below 28. Finish: You MUST remove your vials from the buffer trays the morning after your analysis. To do this, go back into the online instrument window (which should still be open from your runs yesterday). Go to Window/Direct Control. Click on the Load button to bring the vial trays forward. Do not open the cover until the trays have stopped

6 moving! Once movement has stopped, open the cover and remove your vials. Do not throw out these vials, but clean them, and once clean, return them to your TA. Problems and solutions Steps 3/4: The software doesn't communicate with the instrument: Call for help from a TA. The driver card in the PC is probably at fault. Go to the Windows control panel and click on Add new hardware; the PC should find a NI GPIB card, search for a suitable driver, download and install it. Communication should then work. Step 13: I get an autozero error: This usually indicates that the lamp is off: check the UV lamp indicator on the instrument. If it is off, click on the lamp icon in the Direct Control window if it is displayed, or use the Instrument status window, which you ll find under the Control menu, to turn it on. Steps 17/18: My results are poor, or totally missing: Check the current window and make sure it looks ok: there should be little noise, with a fairly steady current of around ma. If it is noisy, there is no buffer present for the injection, which suggests that the vial locations you have specified are wrong. If current is OK, then the injection has failed check the method to ensure that the sample vial has been correctly specified. I don't see three peaks on the electropherogram: First check that you are looking at the detector output (UV), not the display of the capillary current. If the detector output shows fewer than three peaks, try running the QC, which certainly should give three peaks. Still don't see three? You probably have chosen too short a separation time, or too low a voltage, so try increasing the separation time and/or increasing the separation voltage. If you see more than three peaks, the smaller ones are probably contamination.

Standard Operating Procedure for the Beckman PA/CE MDQ Electrophoresis System. Last Modified October 19, 2012

Standard Operating Procedure for the Beckman PA/CE MDQ Electrophoresis System. Last Modified October 19, 2012 Standard Operating Procedure for the Beckman PA/CE MDQ Electrophoresis System Last Modified October 19, 2012 Considerations Prior to Beginning Experiment All solutions must be filtered (using 0.45 m or

More information

Aqueous GPC Instructions - Detailed

Aqueous GPC Instructions - Detailed Aqueous GPC Instructions - Detailed Waste Bottle Columns PL Datastream Injection Port GPC Computer UV Detector GPC Pump Eluent Bottle The Aqueous GPC Set Up RI Detector Important The points below must

More information

Equipment Overview: Safety:

Equipment Overview: Safety: PerkinElmer Series 200 HPLC: Operating Procedure 12/2/11 R.C Equipment Overview: This document outlines the basic procedure for using the Sedex light scattering detector and the PE series 200 HPLC system

More information

Manual HPLC-4 Alliance-UV

Manual HPLC-4 Alliance-UV Manual HPLC-4 Alliance-UV MAKE SURE NEVER TO: 1. pump particles or air through the column. 2. inject when the injector is dry. 3. inject solutions containing particles. 4. apply large (sudden) pressure

More information

3 Starting the chiral HPLC and logging into LC Real Time Analysis.

3 Starting the chiral HPLC and logging into LC Real Time Analysis. Edition: 1.2 Date: 7-Oct-11 Page 1 of 5 1 Responsibilities. First person responsible: Mariëlle Delville. Second person responsible: Helene I.V. Amatdjais-Groenen. 2 Definitions. Read the information on

More information

PerkinElmer Series 200 HPLC: Operating Procedure

PerkinElmer Series 200 HPLC: Operating Procedure PerkinElmer Series 200 HPLC: Operating Procedure Chem Eng 5503 Version 1.01 Last Update 8/9/07 This document outlines the basic operating procedure for the PE series 200 HPLC, UV/Vis and SEDEX Light Scattering

More information

Operating Manual. High Performance Liquid Chromatograph. Scientific Equipment Center, Prince Of Songkla University

Operating Manual. High Performance Liquid Chromatograph. Scientific Equipment Center, Prince Of Songkla University Operating Manual High Performance Liquid Chromatograph Agilent 1100 ; VWD, DAD, FLD and RID Scientific Equipment Center, Prince Of Songkla University Operating Manual High Performance Liquid Chromatograph

More information

UniPoint System Software User s Guide

UniPoint System Software User s Guide UniPoint System Software User s Guide LT2137 1998 Gilson, Inc. All rights reserved Exercise 3-Creating operations list... 4-22 Create and name operations list... 4-22 Set up steps... 4-23 Identify sample

More information

P/ACE MDQ Basic Training Workbook

P/ACE MDQ Basic Training Workbook TM P/ACE MDQ Basic Training Workbook For 32 Karat Software Version 5.0 715091 AA August 2001 Beckman Coulter, Inc. 4300 N. Harbor Blvd., Fullerton, CA 92835 Copyright 2001 Beckman Coulter, Inc. PrintedinU.S.A.

More information

Sciex QTrap Operational Steps for Trained Personnel

Sciex QTrap Operational Steps for Trained Personnel Sciex 6500+ QTrap Operational Steps for Trained Personnel Last Updated 09172017 1. If any of the following instructions does not make sense to you or was not covered during your hand-on training, stop

More information

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer Operating instructions - Cary 100 Bio UV-visible Spectrophotometer This document describes how to power up the spectrophotometer, set its measurement parameters, insert one or more of samples for analysis

More information

ChromQuest 5.0 Quick Reference Guide

ChromQuest 5.0 Quick Reference Guide ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User

More information

Data Acquisition with CP-2002/2003 Micro-GC Control

Data Acquisition with CP-2002/2003 Micro-GC Control Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598 Star Chromatography Workstation Version 6 Data Acquisition with CP-2002/2003 Micro-GC Control Operation Manual Varian, Inc. 2002

More information

Getting low-resolution GC-MS data on the JEOL GCMate Introduction

Getting low-resolution GC-MS data on the JEOL GCMate Introduction Getting low-resolution GC-MS data on the JEOL GCMate Introduction GC-MS is a technique that combines gas phase separation and mass-based detection/characterization of analytes. It s a very sensitive method

More information

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017 Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml If you have any problems with the

More information

HITACHI LACHROM ELITE

HITACHI LACHROM ELITE HITACHI LACHROM ELITE Clarity Control Module ENG Code/Rev.: M133/60D Date: 7/16/2018 Phone: +420 251 013 400 DataApex Ltd. Fax: +420 251 013 401 Petrzilkova 2583/13 clarity@dataapex.com 158 00 Prague 5

More information

Detailed procedures, operation instructions, maintenance, and emergency contact information list is attached.

Detailed procedures, operation instructions, maintenance, and emergency contact information list is attached. GC/FID Standard Operating Procedure Environmental health and Safety (EH&S) 2809 Daley Drive Ames, IA 50011-3660 Phone: (515)-294-5359 Fax: (515)-294-9357 www.ehs.iastate.edu Procedure: Department: Bioeconomy

More information

Standard Operating Procedure for the Horiba FluroMax-4

Standard Operating Procedure for the Horiba FluroMax-4 Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.

More information

Agilent Max-Light Cartridge Cell Information for G4212A/B DAD and G7117A/B DAD

Agilent Max-Light Cartridge Cell Information for G4212A/B DAD and G7117A/B DAD Agilent Max-Light Cartridge Cell Information for G4212A/B DAD and G7117A/B DAD Information about the Max-Light Cartridge Cells 2 Specifications 2 Description of the dispersion volume V(σ) 2 Recommendations

More information

Pictoral Guide to Genotyping on the Ceq8000

Pictoral Guide to Genotyping on the Ceq8000 Pictoral Guide to Genotyping on the Ceq8000 The Beckman Ceq8000. available in beige Index............ 1 Overview........... 2 1. Getting started.......... 3 Logging on to the computer....... 3 Turning

More information

Chromatography Software Training Materials. Contents

Chromatography Software Training Materials. Contents Chromatography Software Training Materials This document contains information on how to build a method, start the instrument to acquire data, and then process the data using the Galaxie Program. You will

More information

Fluorimeter User s Booklet

Fluorimeter User s Booklet Fluorimeter User s Booklet 1) Verify that the instrument power and the computer are both off. Then turn on the lamp. The red indicator light will glow when the lamp is on. Turn on lamp switch here. This

More information

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017 Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017 Introduction This document is only for operation of the HPLC and ICP-MS instruments to obtain data. It contains no maintenance

More information

USER GUIDE TO BECKMAN COULTER FC500

USER GUIDE TO BECKMAN COULTER FC500 USER GUIDE TO BECKMAN COULTER FC500 Create a New Protocol Purpose: This procedure describes how to define the parameters and create rough analysis plots for a new experimental set-up. Some protocol templates

More information

Procedure to set up an HPIB 82350B card on an M57. M58 or C20x PC. Table 1: Parts required. Part Number Description Qty

Procedure to set up an HPIB 82350B card on an M57. M58 or C20x PC. Table 1: Parts required. Part Number Description Qty 1 OF 21 Procedure to set up an HPIB 82350B card on an M57. M58 or C20x PC Parts required Table 1: Parts required Part Number Description Qty 289000764 Kit, HPIB Card, Inter, PCI 1 Procedure The procedure

More information

Joe von Fischer Lab Michaelis-Menton Enzyme Kinetics Procedure

Joe von Fischer Lab Michaelis-Menton Enzyme Kinetics Procedure 1 Joe von Fischer Lab Michaelis-Menton Enzyme Kinetics Procedure Revised 2/7/13 by Claire Griebenow Preparation and Incubation I. Vial Prep 1. Label a set of 60mL vials. Refer to Appendix A for the correct

More information

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample

More information

Seized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division

Seized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division SHIMADZU ULTRAVIOLET/VISIBLE SPECTROPHOTOMETER (UV/VIS) Instrument Shimadzu UV-2401 UV/VIS Spectrophotometer (Serial Number

More information

Depending on the computer you find yourself in front of, here s what you ll need to do to open SPSS.

Depending on the computer you find yourself in front of, here s what you ll need to do to open SPSS. 1 SPSS 13.0 for Windows Introductory Assignment Material covered: Creating a new SPSS data file, variable labels, value labels, saving data files, opening an existing SPSS data file, generating frequency

More information

Operating Procedure for Horiba Raman Microscope

Operating Procedure for Horiba Raman Microscope Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch

More information

Wheatstone Corporation Technical Documentation

Wheatstone Corporation Technical Documentation Wheatstone Corporation Technical Documentation WheatNet-IP Scheduler Technical Manual Preliminary V 2.0.0 600 Industrial Drive New Bern, NC 28562 252.638.7000 www.wheatstone.com Rev 2.0 March, 2015 Dick

More information

New software features of the 7010 Particle Size. Analyzer Danielle Chamberlin May May 5, Page 1

New software features of the 7010 Particle Size. Analyzer Danielle Chamberlin May May 5, Page 1 New software features of the 7010 Particle Size May 5, 2009 Analyzer 2.0.0 Danielle Chamberlin May 5 2009 Page 1 Introduction to the new 7010 software Instrument operation New look and feel easy to operate

More information

cief Data Analysis Chapter Overview Chapter 12:

cief Data Analysis Chapter Overview Chapter 12: page 285 Chapter 12: cief Data Analysis Chapter Overview Analysis Screen Overview Opening Run Files How Run Data is Displayed Viewing Run Data Data Notifications and Warnings Checking Your Results Group

More information

IGSS 13 Configuration Workshop - Exercises

IGSS 13 Configuration Workshop - Exercises IGSS 13 Configuration Workshop - Exercises Contents IGSS 13 Configuration Workshop - Exercises... 1 Exercise 1: Working as an Operator in IGSS... 2 Exercise 2: Creating a new IGSS Project... 28 Exercise

More information

SCHEMATIC DESIGN IN QUARTUS

SCHEMATIC DESIGN IN QUARTUS SCHEMATIC DESIGN IN QUARTUS Consider the design of a three-bit prime number detector. Figure 1 shows the block diagram and truth table. The inputs are binary signals A, B, and C while the output is binary

More information

OCEAN OPTICS Chem 2000 Spectrophotometer

OCEAN OPTICS Chem 2000 Spectrophotometer OCEAN OPTICS Chem 2000 Spectrophotometer prepared for: Department of Chemistry and Biochemistry New Mexico State University Las Cruces, NM 88003 last updated May 2003 R.. Cowley TABLE OF CONTENTS I. Introduction...2

More information

Perkin-Elmer Lambda 3B Spectrophotometer

Perkin-Elmer Lambda 3B Spectrophotometer Written by Mark Csele Written by Jeff Brown Operation Turn on the Lambda 3B and the Computer Remove all samples from the sample chamber. Turn on the lamp for the range being investigated using the switches

More information

Computer. Test Fixture SPA. Key board

Computer. Test Fixture SPA. Key board Semiconductor Parameter Analyzer (Operating instructions) A PDF copy of this document should be available on the Windows Desktop of the SPA computer Read this document fully before attempting to operate

More information

Graphing on Excel. Open Excel (2013). The first screen you will see looks like this (it varies slightly, depending on the version):

Graphing on Excel. Open Excel (2013). The first screen you will see looks like this (it varies slightly, depending on the version): Graphing on Excel Open Excel (2013). The first screen you will see looks like this (it varies slightly, depending on the version): The first step is to organize your data in columns. Suppose you obtain

More information

Introduction to Windows

Introduction to Windows Introduction to Windows Naturally, if you have downloaded this document, you will already be to some extent anyway familiar with Windows. If so you can skip the first couple of pages and move on to the

More information

Dionex AS-AP Sample Conductivity and ph Accessory Setup and Operation Guide

Dionex AS-AP Sample Conductivity and ph Accessory Setup and Operation Guide Dionex AS-AP Sample Conductivity and ph Accessory Setup and Operation Guide Document No. 065470 Revision 02 February 2012 2012 by Thermo Fisher Scientific Inc. All rights reserved. Chromeleon is a registered

More information

Chromeleon-PA Analyzer User s Guide Dionex Corporation

Chromeleon-PA Analyzer User s Guide Dionex Corporation Chromeleon-PA Analyzer User s Guide 2003 Dionex Corporation Document No. 031964 Revision 01 October 2003 2003 by Dionex Corporation All rights reserved worldwide. Printed in the United States of America.

More information

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Section 1: Starting an experiment 1.1 Opening a file 1. Open the File menu and select Open Activity. 2. In the Open dialog box, navigate to

More information

SIMPLE TEXT LAYOUT FOR COREL DRAW. When you start Corel Draw, you will see the following welcome screen.

SIMPLE TEXT LAYOUT FOR COREL DRAW. When you start Corel Draw, you will see the following welcome screen. SIMPLE TEXT LAYOUT FOR COREL DRAW When you start Corel Draw, you will see the following welcome screen. A. Start a new job by left clicking New Graphic. B. Place your mouse cursor over the page width box.

More information

Easy user s guide <outline>

Easy user s guide <outline> Easy user s guide 1st edition(2008.07.01) Start computer Turn on computer and display power. Log in the windows. User Name:administrator Password: (nothing) Turn on sequencer power Turn on the

More information

R-09HR ReleaseNote. R-09HR Operating System Version 2.00 RO9HRRN200

R-09HR ReleaseNote. R-09HR Operating System Version 2.00 RO9HRRN200 R-09HR ReleaseNote R-09HR Operating System Version.00 009 Roland Corporation U.S. All rights reserved. No part of this publication may be reproduced in any form without the written permission of Roland

More information

Content: Step-by-Step instructions to duplicate MS Outlook files to a file. Purpose: To move content from one computer to another.

Content: Step-by-Step instructions to duplicate MS Outlook  files to a file. Purpose: To move  content from one computer to another. Content: Step-by-Step instructions to duplicate MS Outlook e-mail files to a file. Purpose: To move e-mail content from one computer to another. Definitions: Old Computer == the computer you want to copy

More information

Quick Start Guide. Introduction. Importing Raw Data. Method 1: Use Open Data Panel

Quick Start Guide. Introduction. Importing Raw Data. Method 1: Use Open Data Panel ACD/Spectrus Processor: Chromatography Data Irina Oshchepkova and Laura Zepeda Advanced Chemistry Development, Inc. Toronto, ON, Canada www.acdlabs.com Introduction The following document outlines how

More information

TopPoint User's Guide

TopPoint User's Guide TopPoint User's Guide LT2152/ 2000 Gilson, Inc. All rights reserved August 2000 Table of Contents 1 Introduction 2 Operation 3 Troubleshooting Steps to Using TopPoint... 1-2 Accessing Online Help... 1-3...

More information

QAM-I-102 Operation and Calibration of the Autoanalyzer

QAM-I-102 Operation and Calibration of the Autoanalyzer 1. Applicability and Purpose a. This procedure applies to the operation and calibration of the LaChat Quikchem 8000 Automated Ion Analyzer. An equivalent system may be used with proper revision of this

More information

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.0.4 and up GETTING STARTED FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.0.4 and up Getting started

More information

e-bridge Color Profile Tool Quick Start Guide

e-bridge Color Profile Tool Quick Start Guide e-bridge Color Profile Tool Quick Start Guide 1 Contents 1. Installation... 3 1.1. Installing the e-bridge Color Profile Tool Software... 3 1.1. Removing the e-bridge Color Profile Tool... 4 1.2. Installing

More information

CH142: Instructions for the Vernier LabPro

CH142: Instructions for the Vernier LabPro 1 CH142: Instructions for the Vernier LabPro INTRODUCTION: The Vernier LabPro is a versatile data collection interface that can be used in combination with more than 45 sensors, and with a computer, TI

More information

SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS

SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS0051-04 Introduction This document details the installation and known issues with OMNISEC v10.30. This release is a major release to fix a

More information

Waters 600E Multisolvent Delivery System Quick Start Guide

Waters 600E Multisolvent Delivery System Quick Start Guide Waters 600E Multisolvent Delivery System Quick Start Guide 34 Maple Street Milford, MA 01757 WAT174-04TP, Revision 3 NOTICE The information in this document is subject to change without notice and should

More information

Agilent GC/MSD Instructions

Agilent GC/MSD Instructions Agilent GC/MSD Instructions GC/MS SAMPLE PREPARATION: Sample Components to Avoid Completely: The following should never be injected: metals, strong acids or bases, salts, oligomeric and polymeric material.

More information

Steps for LCMS-8040 Triple Quadrupole

Steps for LCMS-8040 Triple Quadrupole Steps for LCMS-8040 Triple Quadrupole June 2017 Version 2 A. Data Acquisition 1. Check lights on MS. There should be a total of three lights lit by LED. One of them is blinking. Power Blinking Gas Status

More information

Galaxie Photodiode Array Software

Galaxie Photodiode Array Software Varian, Inc. 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA Galaxie Photodiode Array Software User s Guide Varian, Inc. 2002-2006 03-914950-00:6 Table of Contents Introduction... 3 General... 3 PDA

More information

Burning CDs in Windows XP

Burning CDs in Windows XP B 770 / 1 Make CD Burning a Breeze with Windows XP's Built-in Tools If your PC is equipped with a rewritable CD drive you ve almost certainly got some specialised software for copying files to CDs. If

More information

Fiery proserver User Guide

Fiery proserver User Guide Fiery proserver User Guide Congratulations on your purchase of the EFI Fiery proserver, a powerful production solution for your printer. Before you begin Make sure that you have: About this document Installed

More information

The Newsletter will contain a Title for the newsletter, a regular border, columns, Page numbers, Header and Footer and two images.

The Newsletter will contain a Title for the newsletter, a regular border, columns, Page numbers, Header and Footer and two images. Creating the Newsletter Overview: You will be creating a cover page and a newsletter. The Cover page will include Your Name, Your Teacher's Name, the Title of the Newsletter, the Date, Period Number, an

More information

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation The GVSU chemistry department owns 2 NMR spectrometers. A JEOL Eclipse 300 MHz and a Varian Inova 400 MHz. Due to their strong, constant magnetic

More information

BUSINESS LINE COMMPORTAL GUIDE

BUSINESS LINE COMMPORTAL GUIDE wowforbusiness.com BUSINESS LINE COMMPORTAL GUIDE WOW! Business BCSE.U.1505.O Business Line CommPortal Guide Table of Contents Getting Started...3 Dashboard...4 Messages and Calls...5 Messages...5 Missed

More information

Series 6 Technical Admin Guide Page 1

Series 6 Technical Admin Guide Page 1 Series 6 Technical Admin Guide Page 1 TABLE OF CONTENTS PRODUCT REGISTRATION... 6 Free Trial Registration and the Wizard...7 How the Trial Works...8 Register Invu Document Management...8 Privileges for

More information

Thermo Scientific Spectronic 200 Visible Spectrophotometer. The perfect tool. for routine measurements

Thermo Scientific Spectronic 200 Visible Spectrophotometer. The perfect tool. for routine measurements Thermo Scientific Spectronic 200 Visible Spectrophotometer The perfect tool for routine measurements The Standard for Routine Measurements Thermo Scientific SPECTRONIC spectrophotometers have served as

More information

2.2. Facilities Requirements

2.2. Facilities Requirements 2.2. Facilities Requirements Facilities requirements for the alpha-se system are listed in Table 2-1 and the system dimensions are given in Figure 2-1. As shown in Figure 2-2, the preferred clear work

More information

ADR View Software for Three Phase Mains Supply Analysers

ADR View Software for Three Phase Mains Supply Analysers ADR View Software for Three Phase Mains Supply Analysers Software Systems ADR-VIEW PRESENTATION AND CONTENTS Three phase mains power supply analysis software from the VEMER ADR range. The instruments are

More information

Microsoft Access Database How to Import/Link Data

Microsoft Access Database How to Import/Link Data Microsoft Access Database How to Import/Link Data Firstly, I would like to thank you for your interest in this Access database ebook guide; a useful reference guide on how to import/link data into an Access

More information

Agilent MassHunter Workstation Software 7200 Accurate-Mass Quadrupole Time of Flight GC/MS

Agilent MassHunter Workstation Software 7200 Accurate-Mass Quadrupole Time of Flight GC/MS Agilent MassHunter Workstation Software 7200 Accurate-Mass Quadrupole Time of Flight GC/MS Familiarization Guide Before you begin 3 Prepare your system 3 Prepare the samples required for data acquisition

More information

Using Mail Merge in Microsoft Word XP/2002

Using Mail Merge in Microsoft Word XP/2002 Using Mail Merge in Microsoft Word XP/2002 Open Microsoft Word XP/2002 Beginning the Merge You may start Mail Merge with a blank document or one that you have already created. We will use a blank document

More information

Installing the Printer Software

Installing the Printer Software 4 Printing Your First Page Attaching the Paper Support 7 1 Checking the Printer 6 2 Plugging in the Printer 3 Installing the Ink Cartridges 5 Installing the Printer Software Connecting the Printer 4012581-00

More information

ECSE-323 Digital System Design. Lab #1 Using the Altera Quartus II Software Fall 2008

ECSE-323 Digital System Design. Lab #1 Using the Altera Quartus II Software Fall 2008 1 ECSE-323 Digital System Design Lab #1 Using the Altera Quartus II Software Fall 2008 2 Introduction. In this lab you will learn the basics of the Altera Quartus II FPGA design software through following

More information

Quick Setup Instructions for the Knauer HPLC-System. Quick Setup Instructions for the Knauer HPLC-System For windows XP (using CHANCE vers. 2.

Quick Setup Instructions for the Knauer HPLC-System. Quick Setup Instructions for the Knauer HPLC-System For windows XP (using CHANCE vers. 2. For windows XP (using CHANCE vers. 2.0) Compiled by P. Madl Vers. 1.0 June 2011 1 / 9 1. Hardware Setup: Result 1.1. Attach RS 232 interface cable (designated as PC) to COM-port-1 of the laptop. 1.2. connect

More information

Help Volume Agilent Technologies. All rights reserved. Instrument: Agilent Technologies 16550A Logic Analyzer

Help Volume Agilent Technologies. All rights reserved. Instrument: Agilent Technologies 16550A Logic Analyzer Help Volume 1992-2002 Agilent Technologies. All rights reserved. Instrument: Agilent Technologies 16550A Logic Analyzer Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer The Agilent

More information

Upchurch Scientific Nano Flow Sensor User s Guide

Upchurch Scientific Nano Flow Sensor User s Guide Upchurch Scientific Nano Flow Sensor User s Guide Oak Harbor, WA 98277 Tel: 800.426.0191 360.679.2528 Fax: 800.359.3460 360.679.3830 Email: customerservice.upchurch@idexcorp.com Internet: www.upchurch.com

More information

Taskbar: Working with Several Windows at Once

Taskbar: Working with Several Windows at Once Taskbar: Working with Several Windows at Once Your Best Friend at the Bottom of the Screen How to Make the Most of Your Taskbar The taskbar is the wide bar that stretches across the bottom of your screen,

More information

Quick Start Guide - Contents. Opening Word Locating Big Lottery Fund Templates The Word 2013 Screen... 3

Quick Start Guide - Contents. Opening Word Locating Big Lottery Fund Templates The Word 2013 Screen... 3 Quick Start Guide - Contents Opening Word... 1 Locating Big Lottery Fund Templates... 2 The Word 2013 Screen... 3 Things You Might Be Looking For... 4 What s New On The Ribbon... 5 The Quick Access Toolbar...

More information

DCN Next Generation Synoptic Microphone Control. en Software User Manual LBB 4171/00

DCN Next Generation Synoptic Microphone Control. en Software User Manual LBB 4171/00 DCN Next Generation Synoptic Microphone Control en Software User Manual LBB 4171/00 About this manual This user manual is divided into five chapters. Chapters 1 and 2 provide background information; chapter

More information

This tutorial walks you through the process of using CC BATCH to simulate a batch distillation column.

This tutorial walks you through the process of using CC BATCH to simulate a batch distillation column. CC-BATCH Tutorial This tutorial walks you through the process of using CC BATCH to simulate a batch distillation column. Description of the Problem The simulation you will create is a five step batch distillation

More information

Agilent 1100 HPLC ChemStation 1) Turn on all the modules for HPLC system (CapPump G1376A, DAD G1315B) and allow for system to initialize.

Agilent 1100 HPLC ChemStation 1) Turn on all the modules for HPLC system (CapPump G1376A, DAD G1315B) and allow for system to initialize. Standard Operating Procedure Remcho Research Group P a g e 1 Agilent 1100 HPLC ChemStation 1) Turn on all the modules for HPLC system (CapPump G1376A, DAD G1315B) and allow for system to initialize. 2)

More information

Fiery proserver User Guide

Fiery proserver User Guide Fiery proserver User Guide Congratulations on your purchase of the Fiery proserver, a powerful production solution for your printer. Before you begin Make sure that you have: About this document Installed

More information

A GET YOU GOING GUIDE

A GET YOU GOING GUIDE A GET YOU GOING GUIDE To Your copy here TextHELP Read and Write 11 June 2013 Advanced 10.31.34 1 Learning Support Introduction to TextHELP Read & Write 11 Advanced TextHELP Read and Write is a tool to

More information

Word 2007 Mail Merge

Word 2007 Mail Merge Word 2007 Mail Merge Mail Merge is the production of multiple documents from a single template form and a structured data source. This helps to create personalized letters and pre-addressed envelopes or

More information

Nuance Field Certification Program 2-09

Nuance Field Certification Program 2-09 08/09/2013 PageCounter Professional (PCP) Equitrac Professional 3.5 (EP3.5) Nuance Field Certification Program 2-09 Nuance Document Imaging Paul Bahry NUANCE DOCUMENT IMAGING Paul Bahry PageCounter Professional

More information

IGSS Configuration Workshop - Exercises

IGSS Configuration Workshop - Exercises IGSS Configuration Workshop - Contents Exercise 1: Working as an Operator in IGSS... 2 Exercise 2: Creating a New IGSS Project... 18 Exercise 3: Create Areas and Diagrams in Definition... 23 Exercise 4:

More information

Appendix C. Vernier Tutorial

Appendix C. Vernier Tutorial C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in the Thermo TraceFinder 3.0 analytical software. For detailed descriptions

More information

Lesson 4: Introduction to the Excel Spreadsheet 121

Lesson 4: Introduction to the Excel Spreadsheet 121 Lesson 4: Introduction to the Excel Spreadsheet 121 In the Window options section, put a check mark in the box next to Formulas, and click OK This will display all the formulas in your spreadsheet. Excel

More information

Adobe Flash CS3 Reference Flash CS3 Application Window

Adobe Flash CS3 Reference Flash CS3 Application Window Adobe Flash CS3 Reference Flash CS3 Application Window When you load up Flash CS3 and choose to create a new Flash document, the application window should look something like the screenshot below. Layers

More information

DS400 / DS404. Software Installation and Operations Manual. DS400 Docking Station. (800) (734)

DS400 / DS404. Software Installation and Operations Manual. DS400 Docking Station. (800) (734) DS400 / DS404 33 Docking Stations Software Installation and Operations Manual DS400 Docking Station DS404 Multi-inlet Docking Station 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734)

More information

Calibration. Calibration Overview CHAPTER

Calibration. Calibration Overview CHAPTER CHAPTER 21 Calibration Calibration Overview Sensitivity Determination The primary role of sound level meter calibration is to establish a numerical relationship between the sound level at the diaphragm

More information

Part 2 Uploading and Working with WebCT's File Manager and Student Management INDEX

Part 2 Uploading and Working with WebCT's File Manager and Student Management INDEX Part 2 Uploading and Working with WebCT's File Manager and Student Management INDEX Uploading to and working with WebCT's File Manager... Page - 1 uploading files... Page - 3 My-Files... Page - 4 Unzipping

More information

Steps for using MagicConnect

Steps for using MagicConnect Getting Started Getting Started MagicConnect is a remote access service that allows you safe remote access to a target device from your remote device. What is a remote device? It is a Windows device, iphone/ipad,

More information

Instructions for Numbering and/or Reordering a Bibliography Roberta E. Sonnino, M.D., Judy Roberts, M.A.

Instructions for Numbering and/or Reordering a Bibliography Roberta E. Sonnino, M.D., Judy Roberts, M.A. Instructions for Numbering and/or Reordering a Bibliography Roberta E. Sonnino, M.D., Judy Roberts, M.A. STEP 1 CLEAN UP PARAGRAPHS IN YOUR CURRENT DOCUMENT Open the Word file that contains your bibliography

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in Thermo TraceFinder analytical software. For detailed descriptions

More information

Direct Image Viewer User Tutorial

Direct Image Viewer User Tutorial Direct Image Viewer User Tutorial A guide to understanding the functions and features within the Direct Image Viewer to best present the data and to get the most understanding from the Direct Image logs.

More information

Chapter 11 Dealing With Data SPSS Tutorial

Chapter 11 Dealing With Data SPSS Tutorial Chapter 11 Dealing With Data SPSS Tutorial 1. Visit the student website at for this textbook at www.clowjames.net/students. 2. Download the following files: Chapter 11 Dealing with Data (SPSS data file)

More information

XPS1 Automated Multi-Sample Run Procedure

XPS1 Automated Multi-Sample Run Procedure XPS1 Automated Multi-Sample Run Procedure Follow the XPS Operating Procedure to load samples into the SAC chamber. Once the samples are in the SAC chamber, the following procedure can be used to automate

More information

NSCC SUMMER LEARNING SESSIONS MICROSOFT OFFICE SESSION

NSCC SUMMER LEARNING SESSIONS MICROSOFT OFFICE SESSION NSCC SUMMER LEARNING SESSIONS MICROSOFT OFFICE SESSION Module 1 Using Windows Welcome! Microsoft Windows is an important part of everyday student life. Whether you are logging onto an NSCC computer or

More information

RECORDING WITH CAMTASIA 9

RECORDING WITH CAMTASIA 9 RECORDING WITH CAMTASIA 9 Griffith College GRIFFITH COLLEGE South Circular Road Table of Contents 1.0 Preparation... 2 1.1 Surroundings... 2 1.2 Lighting... 2 2.0 Pre-recording... 3 2.1 Audio and Equipment

More information