Applied Biosystems RapidFinder Software Version 1.0 and SDS Software Version Beta Release 10/05 Release Notes (PN Rev A)
|
|
- Arlene Nicholson
- 5 years ago
- Views:
Transcription
1 Applied Biosystems RapidFinder Software Version 1.0 and SDS Software Version Beta Release 10/05 Release Notes (PN Rev A) Software Release notes for: Applied Biosystems RapidFinder Software v1.0 (7300 Real-Time PCR System) (P/N ) RapidFinder Software v1.0 (7300 Real-Time PCR System with RQ Study) (P/N ) Applied Biosystems RapidFinder Software v1.0 (7500 Real-Time PCR System) (P/N ) Installation Refer to the RapidFinder Software Version 1.0 Install User Bulletin for information. Upgrading from an earlier version of SDS software: If you are installing this software on a computer with an earlier version of the SDS software installed, DO NOT uninstall the existing software. Uninstalling the previous version and installing this version will result in the inability to operate the connected instrument. The upgrade process for SDS Software from v to takes approximately 10 minutes. Upgrading from a version earlier than to v may require up to an hour, including loading new firmware to the instrument. Upgrades are supported only for a common instrument type. For example, 7300 SDS v will upgrade only from a previous version of 7300 SDS and 7500 SDS v will upgrade only from a previous version of 7500 SDS. Place the new version disk in the CD drive and allow the System Setup Wizard to launch automatically. If the System Setup Wizard does not launch automatically, double click the SystemSetupWizard.exe file on the CD to start the installation. Note: You must complete all steps of the Unpack and Set Up a New Instrument procedure from the System Setup Wizard at least once before you may perform the Calibrate the Instrument procedure, even if you are running the System Setup Wizard on a machine with an existing SDS installation. When the instrument is first connected, the software may detect that the instrument firmware is not the latest version. If the instrument firmware is not the most recent, the software will recommend that the most recent version be loaded to the instrument. The new firmware version should be loaded to the instrument for correct performance. The firmware loading requires approximately 40 minutes. Please refer to the 7300/7500/7500 Fast Installation and Maintenance guide (p/n ) for detailed instructions. Part Number Rev. A Page 1 of 8
2 Known Issues with RapidFinder Software v1.0 Beta Release General Issues DO NOT SAVE A RAPIDFINDER PLATE in the SDS Software. RapidFinder plates saved in the SDS software cannot be opened by the RapidFinder software or the SDS software, and RapidFinder results may be invalidated. If you attempt to launch the SDS software version while the RapidFinder software is running, the SDS software does not start. The SDS software cannot run at the same time as the RapidFinder software. Any comments entered in the Comment field (on the Enter Plate Information, Select Plate to Pipette, and Select Plate to Run pages) are not included in printed reports or exported results. Creating a Plate If you click Close when the Enter Sample Names page or the Select Pipetting Option pages are displayed (then click OK in the Save Changes dialog), the settings on these pages are not saved. Settings on previous pages are saved. Pipetting a Plate When pipetting a plate that does not have positive controls specified, the Close Tubes Before Pipetting Positive Controls page is displayed instead of the Close Tubes page. IMPORTANT! After you pipette the plate, make sure reagents are in the bottom of the wells. If available, use a centrifuge with a plate adapter to briefly centrifuge the plate (<1500xg). If an expected plate does not appear in the Select a Plate to Pipette list, make sure you have confirmed the plate configuration (on the Main page, select Open Plate in Progress, then click Next on all pages including the Confirm Plate Configuration page). If the plate still does not appear in the Select a Plate to Pipette list, click Close to return to the Main page, click View Results, click any plate in the list, click Close to return to the Main page, then click Pipette a Plate. The plate list should be reset and correctly list all plates. When pipetting plates, do not select Pipette Another Plate as your next step. Return to the Main page, then select Pipette a Plate to pipette another plate. Running a Plate Before starting a run on the Run page, change the Number of PCR Cycles from 50 to 45. On the Run page, the Cycle Counter (Cycle n of n) and Time Remaining clock do not display the correct number of cycles run or the time remaining in the run. However, the run will complete successfully. Do not enter a Run Operator name on the Select a Plate to Run page. If you do, the plate is not available to run. Viewing Results Part Number Rev. A Page 2 of 8
3 For all "Negative" Sample results, check the By Well results for the sample to ensure that there are no By Well Warnings associated with the sample. If there are, you may want to rerun your samples. If you enter a Run Operator name, then click Next, then click Close, the plate is not available in the Select a Plate to Run list the next time you view the page. To make the plate available, select Create/Edit a Plate from the Main page, select Open Plate in Progress, select the plate in question, then click Next on all pages including the Confirm Plate Configuration page. Creating Templates If you create a template using the name of an existing template, a message box is displayed asking if you want to overwrite the template. The existing template is overwritten, even if you select No in the message box. If you create a template from a plate and use the name of an existing template, the software does not detect whether the template name already exists. The existing template will be overwritten. New Features in the SDS Software v1.3.2 System Setup Wizard The System Setup Wizard guides a user through performing all steps of:? an initial instrument installation;? a software upgrade or installation; or? an instrument recalibration. You must complete all steps of the Unpack and Set Up a New Instrument procedure from the System Setup Wizard at least once before you may perform the Calibrate the Instrument procedure, even if you are running the System Setup Wizard on a machine with an existing SDS installation. Firmware Version G2.09 This version of the SDS software includes version G2.09 of the instrument firmware. After installing this firmware, attempting to revert to a previous version of the firmware will result in a loss of all firmware calibration data. Usage Notes SDS Software Installation If you are installing this software on a machine with an existing installation of an earlier version of SDS, please DO NOT uninstall the existing software. Instead, upgrade the software by following the instructions in the Upgrading from an earlier version of SDS software section above. Uninstalling the previous version and installing this version will result in the inability to operate the connected instrument. Part Number Rev. A Page 3 of 8
4 Display Settings The display graphics require a video controller and system settings exercising at least the High (16 bit) color palette. Using color palette settings with fewer colors will result in the incorrect display of some detectors Instrument A maximum of five (5) dyes can be processed in each well on the 7500 system. A maximum of four (4) detectors can be applied per well in any one of the following combinations: 4 reporters, 1 quencher, 0 passive references 4 reporters, 0 quenchers, 1 passive reference 3 reporters, 1 quencher, 1 passive reference 2 reporters, 2 quenchers, 1 passive reference 3 reporters, 2 quenchers, 0 passive reference It is recommended to include a passive reference in all reactions. Allelic Discrimination Modifying the Use Pre-Read Data in Analysis selection causes results to be recalculated and causes allele calls to be re-evaluated. However, if that selection is modified and the Keep Manual Calls option is enabled, results will not be recalculated for wells that have manually assigned allele calls. It is recommended that the Keep Manual Calls option be disabled before the Use Pre-Read Data in Analysis selection is modified unless you want to keep the manual calls you have already made. RQ Study When performing an RQ study, all included RQ assays should have been run on the same instrument. Including RQ assays run on various instruments may produce unexpected results. The software will not allow you to include documents in an RQ study unless they share consistent plate type, instrument type, run mode, and thermal profile information; however, the software does not enforce that all included assays have been run on a single instrument. After omitting selected wells in an RQ Study and then resorting the grid, the displayed row positions that were selected remain selected, even if the row content has moved to a different display position due to the grid resorting. It is suggested that you always check the selected rows in the Well information grid after omitting wells and resorting the grid. A statistical p-value calculation is not supported in this version of the SDS application. Part Number Rev. A Page 4 of 8
5 Known Issues SDS Software System Setup Wizard When performing the Optical Calibration, the Preparing the Plate page contains a picture in which the plate is labeled Background plate. It should be labeled ROI plate. If the software has already been installed and the System Setup Wizard is launched and the user selects either the Unpack and Set Up a New Instrument option or the Install the SDS Software option, as part of the wizard series, the Software Installation page will be displayed and the Next > button will be disabled. You must select Install / Upgrade SDS Software in order to continue, although canceling the installer without performing any installation operations will still enable the Next > button if the current version of the software is already installed. Assay Setup If a thermal profile is set up using the auto-increment feature, the software typically rejects auto-increment parameter combinations that cannot be executed by the instrument (e.g. starting at 80 C and incrementing 2 C for each of 40 cycles would result in a temperature of 158 C). However, it is possible to enter some parameter combinations for which a run cannot be completed (e.g. by entering negative values or modifying values in an unusual order). If you set up such a thermal profile, the system will allow you to save the document and start the run, but the run will not be completed or may be completed improperly. Instrument Runs If an instrument run is stopped while the instrument is waiting for the cover to heat, the software will indicate that the run is terminated and no further data will be collected; however, the instrument will continue to thermally cycle. If the instrument front door is open when a run is requested to start, the block will be raised, but a door open error will be displayed and the run will not be performed. Pushing the Ok button on the dialog box should lower the block, allowing the plate to be ejected; however, if the Ok button on the dialog box is pushed while the instrument s filter wheel is turning, the block will not be lowered. To avoid that condition, if you are faced with the door open error situation, please wait until the filter wheel stops turning before pushing the Ok button on the dialog box. If the instrument lamp is not performing properly (defined as not drawing the correct current), the first attempt to perform a run after instrument power up will result in a Lamp Failure error. Any further attempts to perform runs on the instrument with the low lamp current condition (without first turning the instrument off and on again) will result in Instrument Communication Failure errors. For an instrument with the low lamp current condition, turning the instrument off and then on will reproduce the initial Lamp Failure error. Part Number Rev. A Page 5 of 8
6 While an instrument run is in progress, the SDS software requires significant processor utilization to perform data collection activities. In order to allow the workstation the greatest processor usage, it is recommended to avoid the following activities during an instrument run: rapidly changing the well selection for any Results chart; surfing the Internet; writing to a CD or DVD; scanning a disk for viruses; defragmenting a disk; or any other activity that extensively uses the workstation s processor. Performing such an activity during an instrument run may result in data loss or run termination. Auto-baselining and Auto C T Amplification plots in which the baselines are very steep may not be baselined correctly in Auto Baseline mode. Plots with very steep baselines may appear to have an earlier than expected baseline end cycle. It is suggested that you manually set the start and end cycles on such plots. When Auto C T is enabled, amplification plots crossing the threshold in the baseline region will be assigned an undetermined C T. Results Report When handling AQ Run results, some wells may be filtered due to noisy Baselines. While these filtered wells do not appear in the analyzed data, they are not reported as Filtered in views, exports, and reports. For any well that is assigned more than one detector, there is a remote possibility that the second detector assigned to well will be reported as undetected (undet.) even if the signal shown for that detector on the Amplification Plot chart crosses the threshold. That may occur if: The combination of detectors is not assigned to any other well on the plate. The second detector is not assigned as the first detector to any other well on the plate. For each well on the plate to which the second detector is assigned, the first detector assigned to the well is not assigned to any well before that well, where the ordering of the wells is: A1,... A12, B1,... H12 Allelic Discrimination When the display of an AD scatter plot is refreshed, inconsistencies might arise between the plot selection and the table selection. Selecting another tab and then returning to the scatter plot can refresh the scatter plot view. One of the Apply, OK, or OK & Reanalyze buttons must be selected to change the Quality Value setting in the Analysis Settings dialog. Pushing the Enter key alone after inserting a new value does not commit the change. Part Number Rev. A Page 6 of 8
7 Preferences When setting preferences for the Image Viewer, user-entered parameters may not be validated. When setting preferences for the Amplification Plot, user-entered parameters may not be validated. Printing The Background Assay Report does not print in landscape mode, only in portrait mode. Instrument Communication In the event that the software loses communication with the instrument during a run (e.g. a cable is disconnected or there is a hardware malfunction), the best method for resetting the system and reconnecting the software is to: 1. quit the software application on the PC (SDS or System Setup Wizard), 2. cycle the power on the instrument, 3. re-launch the software application on the PC, and 4. re-open the document for connection. Part Number Rev. A Page 7 of 8
8 Notice to Purchaser: License Disclaimer Purchase of these software products alone does not imply any license under any process, instrument or other apparatus, system, composition, reagent or kit rights under patent claims owned or otherwise controlled by Applera Corporation, either expressly or by estoppel. Applied Biosystems is a registered trademark and AB (Design) and Applera and RapidFinder are trademarks of Applera Corporation or its subsidiaries in the US and certain other countries. Part Number Rev. A Page 8 of 8
About Data Collection Preferences. Before You Begin. Macintosh computer - Step One: Setting Folder Locations
To navigate within this document use the or keys ABI PRISM 310 Genetic Analyzer Data Collection Software Preferences Module Macintosh computer BEFORE PERFORMING ANY TROUBLESHOOTING WORK ON YOUR ABI PRISM
More informationDuring the calibration the following keys move the autosampler in the desired directions: Page Up. Hold SHIFT key together with the above key
To navigate within this document use the or keys ABI PRISM 310 Genetic Analyzer BEFORE PERFORMING ANY TROUBLESHOOTING WORK ON YOUR ABI PRISM 310 GENETIC ANALYZER, PLEASE READ THE INSTRUMENT USER S MANUAL
More informationSequencePro Data Analysis Application. User Guide
SequencePro Data Analysis Application User Guide SequencePro Data Analysis Application User Guide DRAFT October 31, 2001 12:52 pm, Title_page.fm Copyright 2001, Applied Biosystems. All rights reserved.
More informationUser Bulletin Applied Biosystems StepOne /StepOnePlus Real-Time PCR System
User Bulletin Applied Biosystems StepOne /StepOnePlus Real-Time PCR System July 30, 2007 SUBJECT: Network Port Calibration and StepOne Instrument Connection Procedures In This User Bulletin How to Check
More informationApplied Biosystems SDS Plate Utility v2.2 For the ABI PRISM 7900HT Sequence Detection System
Applied Biosystems SDS Plate Utility v2.2 For the ABI PRISM 7900HT Sequence Detection System User Guide Copyright 2004 Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic
More informationRelease Notes QuantStudio TM Design & Analysis Software v1.4.3
Release Notes QuantStudio TM Design & Analysis Software v1.4.3 CONTENTS OVERVIEW OVERVIEW FEATURES UPDATES FIXED DEFECTS LICENSE COMPATIBILITY ONLINE HELP KNOWN ISSUES SYSTEM REQUIREMENTS INSTALL SOFTWARE
More informationHID Real-Time PCR Analysis Software
USER GUIDE HID Real-Time PCR Analysis Software Version 1.3 for use with: 7500 Real-Time PCR Instruments QuantStudio 5 Real-Time PCR Instruments (with 0.2-mL 96-Well Block) Quantifiler DNA Quantification
More informationABI PRISM 7700 Sequence Detection System: Routine Maintenance
ABI PRISM 7700 Sequence Detection System: Routine Maintenance Overview The ABI PRISM 7700 Sequence Detection System (SDS) and the Macintosh computer that runs the SDS need to be properly and routinely
More informationUser Bulletin. GeneMapper Software Version 4.1. In This User Bulletin. Overview This user bulletin supplements the GeneMapper Software v4.
User Bulletin GeneMapper Software Version 4.1 April 2009 SUBJECT: Autoanalysis In This User Bulletin This user bulletin covers: Autoanalysis Using GeneMapper Software v4.1.............. 2 Autoanalysis
More informationData Collection Software Release Notes. Real-Time PCR Analysis Software Release Notes. SNP Genotyping Analysis Software Release Notes
PN 101-6531 D1 RELEASE NOTES Biomark/EP1 Software To download the latest version of the software for Biomark HD, Biomark, and EP1, go to fluidigm.com/software. For more information about updating the software,
More informationInstalling the GeneMapper Software and Creating a Matrix on the ABI PRISM 310 Genetic Analyzer
User Bulletin GeneMapper Software Version 4.0 and Version 3.7 September 2005 SUBJECT: Installing the GeneMapper Software and Creating a Matrix on the ABI PRISM 310 Genetic Analyzer In This User Bulletin
More informationUse the AccuSEQ Software v2.0 Custom Experiment mode
QUICK REFERENCE Use the AccuSEQ Software v2.0 Custom Experiment mode Publication Part Number 4425585 Revision Date 25 January 2013 (Rev. B) Custom Experiments Note: For safety and biohazard guidelines,
More informationApplied Biosystems AutoCaller Software
User Guide Introduction Applied Biosystems AutoCaller Software SNP Genotyping Analysis Tool Preparation Tasks Manage Studies and Assay Information View and Edit Data Publish Data Copyright 2008, Applied
More informationABI PRISM GeneMapper Software Version 3.0 SNP Genotyping
ABI PRISM GeneMapper Software Version 3.0 SNP Genotyping Tutorial ABI PRISM GeneMapper Software Version 3.0 SNP Genotyping Tutorial September 25, 2002 1:20 pm, 7x9_Title.fm Copyright 2002, Applied Biosystems.
More informationUser Bulletin. Applied Biosystems StepOne Real-Time PCR System DRAFT. SUBJECT: StepOne System Updates. In This Bulletin Installation Updates...
Applied Biosystems StepOne Real-Time PCR System User Bulletin User Bulletin Applied Biosystems StepOne Real-Time PCR System January 2007 SUBJECT: StepOne System Updates In This Bulletin Installation Updates.................................................
More informationGeneMapper Software v3.5
GeneMapper Software v3.5 Installation Instructions This document includes: Before You Install v3.5................................... 2 Installing the Instrument Configuration...................... 6 Installing
More informationUser Bulletin. GeneMapper ID-X Software Version Update. GeneMapper ID-X Software Version Verification. In this user bulletin.
User Bulletin GeneMapper ID-X Software Version 1.1.1 Update October 2009 SUBJECT: GeneMapper ID-X Software Version 1.1.1 Verification In this user bulletin Overview This user bulletin covers: Overview......................................................
More informationABI PRISM 3100-Avant and 3100 Data Collection v2.0 Software Frequently Asked Questions (FAQ)
ABI PRISM 3100-Avant and 3100 Data Collection v2.0 Software Frequently Asked Questions (FAQ) Table of Contents General Information...2 Access Control/Auditing Features...2 Instrument/Plate Setup...3 Spectral
More informationVeriti Thermal Cycler
product brochure PCR Systems Veriti Thermal Cycler The Power of MORE Applied Biosystems, the most trusted name in PCR instruments, introduces a new line of innovative thermal cyclers The Veriti Thermal
More informationMaintenance and Troubleshooting Guide
Applied Biosystems 7900HT Fast Real-Time PCR System Maintenance and Troubleshooting Guide Performing Calibration and Verification Runs Maintaining the 7900HT Instrument Hardware Maintaining the Computer
More informationGuide to running KASP TM genotyping reactions on the Roche LC480-Series instruments
extraction sequencing genotyping extraction sequencing genotyping extraction sequencing genotyping extraction sequencing Guide to running KASP TM genotyping reactions on the Roche LC480-Series instruments
More informationNCSBI Forensic Biology Section DNA SOP Effective Date: January 30, 2007 Title: ABI PRISM 7000: DNA Quantitation Revision 02
1. Preparing the DNA Standards 1.1. Label eight microcentrifuge tubes to be used for dilution of standards (i.e. Std. 1, Std. 2, Std. 3, etc.) 1.2. Add required amount of TE/glycogen to each tube (See
More informationThermal Cycler Dice TM Real Time System Software. Operation Manual
Thermal Cycler Dice TM Real Time System Software Version 4.02 Code TP900/TP960 Operation Manual Attention: Thermal Cycler Dice TM Real Time System Software Operation Manual should be stored for reference
More informationGetting Started Guide. HID Real-Time PCR Analysis Software Version 1.0 For 7500 Real-Time PCR System
Getting Started Guide HID Real-Time PCR Analysis Software Version 1.0 For 7500 Real-Time PCR System Getting Started Guide Get Started HID Real-Time PCR Analysis Software Version 1.0 For 7500 Real-Time
More informationRELEASE NOTES TABLE OF CONTENTS SOFTWARE OVERVIEW
RELEASE NOTES GeneMapper TM ID-X Software v1.5.2 Life Technologies is a Thermo Fisher Scientific brand. 2016 Thermo Fisher Scientific Inc. All rights reserved October 2016 TABLE OF CONTENTS * Software
More informationRelease Notes Life Technologies Attune NxT Software v2.3
Release Notes Life Technologies Attune NxT Software v2.3 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance Software
More informationCFX Maestro Software for Mac
Real-Time PCR CFX Maestro Software for Mac User Guide Version 1.0 CFX Maestro Software for Mac User Guide Version 1.0 Bio-Rad Technical Support Department The Bio-Rad Technical Support department in the
More informationInstalling and Using VeritiLink Software Quick Reference Card
VeritiLink Remote Management Software Installing and Using VeritiLink Software Quick Reference Card For more information on installing VeritiLink Software, see the VeritiLink Remote Management Software
More informationAnalyzing a Project without a Reference
Applied Biosystems Variant Reporter Software v1.0 Analyzing a Project without a Reference Quick Reference Card This quick reference card summarizes the procedures for setting up a project in Applied Biosystems
More informationQuantStudio Design and Analysis Software
QuantStudio Design and Analysis Software USER GUIDE Command-Line Application for QuantStudio Design and Analysis desktop Software v1.2.x for use with: QuantStudio 3 and 5 Real-Time PCR Systems Publication
More informationUser Bulletin. 3730/3730xl DNA Analyzer Data Collection Software. New features. SUBJECT: Data Collection Software version 3.1. In this user bulletin
User Bulletin 3730/3730xl DNA Analyzer Data Collection Software May 2009 SUBJECT: Data Collection Software version 3.1 In this user bulletin This user bulletin covers: New features....................................................1
More informationDataAssist v2.0 Software User Instructions
Introduction DataAssist v2.0 Software User Instructions DataAssist Software is a simple, yet powerful data analysis tool for sample comparison when using the comparative C T (ΔΔC T ) method for calculating
More informationNext generation Confirmation (NGC) module
QUICK REFERENCE Next generation Confirmation (NGC) module Catalog Number A28221 Pub. No. MAN0015891 Rev. A.0 Product description The Applied Biosystems Next generation Confirmation (NGC) module analyzes
More informationGeneMapper ID-X Software Version 1.2
GeneMapper ID-X Software Version 1.2 USER BULLETIN DECEMBER 2010 SUBJECT: Verification Experiments and Installation Procedures Purpose of this user bulletin In this user bulletin This user bulletin: Describes
More informationSymBiot I Sample Workstation
SymBiot I Sample Workstation Getting Started Guide DRAFT August 28, 2001 8:42 pm Symbfmgs.fm5 Copyright 1997, 2001, Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic
More informationxponent Version 4.2 RELEASE NOTES
xponent Version 4.2 RELEASE NOTES Copyright Luminex Corporation, 1996-2013. All rights reserved. Part Number: 89-30000-00-462 Disclaimer Subject to change without notice. Although prepared to ensure accuracy,
More informationSetup and analysis QuantStudio 5 Real-Time PCR System
Setup and analysis QuantStudio 5 Real-Time PCR System Applied Biosystems Software 1 Open software Click on the QuantStudio Design and Analysis Software icon to start the software 2 Software 2 Home window
More informationSynthAssist Software. User Guide. Version 3.1. DRAFT September 7, :02 pm, SynthA_Title.fm
SynthAssist Software Version 3.1 User Guide September 7, 2004 2:02 pm, SynthA_Title.fm Copyright 2004, Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic procedures.
More informationMicroSeq ID Analysis Software Version 1.0
MicroSeq ID Analysis Software Version 1.0 Getting Started Guide Copyright 2003, Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. Information in this
More informationNetwork Upgrade Installation Instructions
Helper Instructions email: www.helper.com Copyright 2006 VantageMed Corporation. All rights reserved. The information in this document is subject to change without notice and should not be construed as
More informationApplied Biosystems TrueScience Aneuploidy STR Kits
Applied Biosystems TrueScience Aneuploidy STR Kits Software Setup and Data Analysis User Guide The results obtained from these or any other diagnostic kits should be used and interpreted only in the context
More informationUser Bulletin. GeneMapper ID Software v3.2.1 Patch. In This User Bulletin. Obtaining and Installing the GeneMapper ID Software v3.2.
User Bulletin GeneMapper ID Software v3.2.1 Patch 01/2007 SUBJECT: GeneMapper ID Software v3.2.1 Patch In This User Bulletin This user bulletin provides information about four minor changes provided in
More informationLiving Image. Release Notes Version 4.7
Living Image Release Notes Version 4.7 1. Purpose This document provides a brief overview of new features and improvements in Living Image 4.7. This release adds acquisition support for the new Lumina
More informationLTQ Velos SP3 Release Notes
LTQ Velos 2.6.0 SP3 Release Notes These release notes briefly list changes in the LTQ Velos 2.6.0 SP3 experiment setup, tuning, and calibration application; fixes to previously reported issues; and known
More informationFor additional information, see these documents on the Breeze 2 software DVD:
R E L E A S E N O T E S Waters Breeze 2 Read these release notes before you install and operate Breeze 2. They contain information necessary to ensure successful liquid chromatography results, including:
More informationRelease Notes QuantStudio 3D Software v 3.0
CONTENTS OVERVIEW FEATURES KNOWN ISSUES LICENSE DOWNLOAD SOFTWARE INSTALL SOFTWARE LEGAL STATEMENTS Release Notes QuantStudio 3D Software v 3.0 OVERVIEW QuantStudio 3D Software supports the Applied Biosystems
More informationHow to run a qpcr plate on QuantStudio 6
How to run a qpcr plate on QuantStudio 6 Important notes Check the compatibility of your mastermix with this instrument. In particular, mastermixes containing «high rox» concentrations should not be used.
More informationUpgrading Luminex IS 2.3 to Bio-Plex Manager 6.1 Software. For technical support, call your local Bio-Rad office, or in the US, call
Upgrading Luminex IS 2.3 to Bio-Plex Manager 6.1 Software For technical support, call your local Bio-Rad office, or in the US, call 1-800-424-6723. Bio-Rad Laboratories, Inc., 2000 Alfred Nobel Drive,
More informationUser Bulletin. ABI PRISM GeneScan Analysis Software for the Windows NT Operating System. Introduction DRAFT
User Bulletin ABI PRISM GeneScan Analysis Software for the Windows NT Operating System SUBJECT: June 2002 Overview of the Analysis Parameters and Size Caller Introduction In This User Bulletin Purpose
More informationOlink NPX Manager. User Guide
Olink NPX Manager User Guide Table of contents INTRODUCTION 3 Process 3 Limitations 3 Requirements for analysis 3 List of abbreviations 3 TO CREATE A HEAT MAP CT CSV FILE 4 OLINK NPX MANAGER STEP BY STEP
More informationformat The new Gradient Palm-CyclerTM
format The new Gradient Palm-CyclerTM combines robust engineering, advanced performance, and a palmtop PC interface for supreme user-friendly flexibility. Palmtop PC Interface palmtop PC interface provides
More informationEcoStudy Software User Guide
EcoStudy Software User Guide FOR RESEARCH USE ONLY What is EcoStudy? 3 Setting Up a Study 4 Specifying Analysis Settings for your Study 6 Reviewing the Data in your Study 8 Exporting Study Data to a Report
More informationRotor-Gene AssayManager v1.0 UDT Basic Plug-in User Manual
February 2018 Rotor-Gene AssayManager v1.0 UDT Basic Plug-in User Manual R3 QIAGEN GmbH, QIAGEN Strasse 1, 40724 Hilden, Germany Sample to Insight Contents 1 UDT Basic Plug-in User Manual...1-1 1.1 Safety
More informationRoche LightCycler 480
Roche LightCycler 480 Instrument Setup Instructions for RT 2 Profiler PCR Array, TaqMan Version A. Preparation Before the Experiment (Presetting the Run Template will save time for your run): 1. Make sure
More informationTA Instruments New Features in ITCRun TM Software
TA Instruments New Features in ITCRun TM Software Notice The material contained in this manual, and in the online help for the software used to support TA Instruments products, is believed adequate for
More informationFirmware Flash Upgrade Wizard. User Manual
Firmware Flash Upgrade Wizard User Manual November 2018 1696956 Rev A ETS-Lindgren Inc. reserves the right to make changes to any products herein to improve functioning or design. Although the information
More informationTA Instruments New Features in DSCRun TM Software
TA Instruments New Features in DSCRun TM Software Notice The material contained in this manual, and in the online help for the software used to support TA Instruments products, is believed adequate for
More informationOperating Instructions METTLER TOLEDO Calibry Software for Pipette Calibration Calibry
Operating Instructions METTLER TOLEDO Calibry Software for Pipette Calibration Calibry Version 3.1 Contents Contents 1 Introduction... 5 1.1 Short description of Calibry... 5 1.2 Important information
More informationRegister your instrument! PCR-Assistant. Software manual
N) alanual Register your instrument! www.eppendorf.com/myeppendorf Software manual Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the
More informationMic qpcr. Fast. Accurate. Compact. Ultimate Performance for Real-Time PCR. Speed Accuracy Size Connectivity
Mic qpcr Ultimate Performance for Real-Time PCR Fast. Accurate. Compact. Speed Accuracy Size Connectivity Hardware Style meets substance - ultimate performance, beautiful design. Fast results without compromise
More informationDell SonicWALL SonicOS 5.9 Upgrade Guide
Dell SonicWALL Upgrade Guide April, 2015 This provides instructions for upgrading your Dell SonicWALL network security appliance to from a previous release. This guide also provides information about importing
More informationLabChip EZ Reader V3.0 SP2 Software Release Notes
LabChip EZ Reader V3.0 SP2 Software Release Notes Installation Notes... 2 Starting the Installation... 2 Upgrading From V2.0, V2.1, or V2.2 to V3.0 SP2... 2 Uninstalling the Software... 3 Documentation...
More informationCFX96 and CFX384 Real-Time PCR Detection Systems Instruction Manual
CFX96 and CFX384 Real-Time PCR Detection Systems Instruction Manual Catalog # 184-5384 # 185-5384 # 184-5096 # 185-5096 Copyright 2008 Bio-Rad Laboratories, Inc. Reproduction in any form, either print
More informationGuide to running KASP genotyping on the BIO-RAD CFX-series instruments
extraction sequencing genotyping extraction sequencing genotyping extraction sequencing genotyping extraction sequencing genotypin Guide to running KASP genotyping on the BIO-RAD CFX-series instruments
More informationLabVIEW Datalogging and Supervisory Control Module
RELEASE AND UPGRADE NOTES LabVIEW Datalogging and Supervisory Control Module Version 8.6 Contents This document describes the system requirements and the process of installing the LabVIEW 8.6 Datalogging
More informationUser Guide 701P Wide Format Solution Wide Format Scan Service
User Guide 701P44865 6204 Wide Format Solution Wide Format Scan Service Xerox Corporation Global Knowledge & Language Services 800 Phillips Road Bldg. 845-17S Webster, NY 14580 Copyright 2006 Xerox Corporation.
More informationDataView Release Notes Version Released
DataView Release Notes Version 3.53.0013 Released 9-25-2018 CATEGORY (v 3.53.0013): Simple Logger II Version 1.01.0059 Open button is now active when selecting recordings from the session s list. Corrected
More informationOne Identity Active Roles 7.2
One Identity December 2017 This document provides information about the Active Roles Add_on Manager7.2. About Active Roles Add_on Manager New features Known issues System requirements Getting started with
More informationInstallation of Hot Disk v 7.3 and later on a PC with Windows 7, 8 or 10
Installation Instruction for Hot Disk TPS 7 2.1 2017-05-09 1(44) Installation of Hot Disk v 7.3 and later on a PC with Windows 7, 8 or 10 Installation Instruction for Hot Disk TPS 7 2.1 2017-05-09 2(44)
More informationXpert Calibration GXCAL-CE-5. InVitro Diagnostic Medical Device
Xpert Calibration GXCAL-CE-5 InVitro Diagnostic Medical Device 301-1186, Rev. A September 2012 This product is sold under license from Molecular Probes, Inc. Copyright 2012 Cepheid. Trademark Information
More informationE-bot PRO Firmware Upgrade Manual for v1.3.0
E-bot PRO Firmware Upgrade Manual for v1.3.0 This manual introduces Firmware upgrade and TTS date upgrade. The firmware upgrade is necessary. The TTS data upgrade is not necessary if the version of E-bot
More informationTorch Melting Curve Analysis with FilmArray 2.0 Software
Torch Melting Curve Analysis with FilmArray 2.0 Software 1. Introduction FilmArray PCR reactions contain the DNA binding dye LCGreen Plus. LCGreen Plus is incorporated into the copies of DNA as they are
More informationIRF90 - Rotating Focuser
IRF90 - Rotating Focuser Part # 600180 REV092111 Page 1 Contents Introduction and Overview... 3 Limitations... 3 Packing List... 4 Installation... 5 Remove Existing Focuser... 5 Installing the Rotating
More informationDMS Local. User Manual. For Projector Management V 1.01
DMS Local User Manual For Projector Management V 1.01 2 Copyright Copyright Copyright 2018 BenQ Corporation. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed,
More informationRelease Notes Life Technologies Attune NxT Software v2.2
Release Notes Life Technologies Attune NxT Software v2.2 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance Software
More informationFor Research Use Only. Not for use in diagnostic procedures.
Release Notes Agilent Feature Extraction 12.1 (12.1.0.3) Product Numbers # G2566AA - Feature Extraction provided with the Agilent Scanner # G4460AA - G4462AA - Feature Extraction Commercial License # G4463AA
More informationThe following documents are included with your Sony VAIO computer.
Documentation The following documents are included with your Sony VAIO computer. Printed Documentation Quick Start Guide Describes the process from unpacking to starting up your VAIO. Troubleshooting and
More informationAchievement Series Working with Scan Station
Achievement Series Working with Scan Station Information in this document is subject to change without notice and does not represent a commitment on the part of Scantron. The software described in this
More informationUser Manual (Vers. 1.03)
PLUSSoftUsb Power Meter Mode User Manual (Vers. 1.03) Laserpoint srl - Via Burona, 51-20090 Vimodrone (Milano) - Italy Phone +39 02 27 400 236 - Telefax +39 02 25 029 161 www.laserpoint.it 1 Contents CONTENTS
More informationmiscript mirna PCR Array Data Analysis v1.1 revision date November 2014
miscript mirna PCR Array Data Analysis v1.1 revision date November 2014 Overview The miscript mirna PCR Array Data Analysis Quick Reference Card contains instructions for analyzing the data returned from
More informationUpgrade Instructions. Note
Upgrade Instructions Prior to upgrading, please review the following: Before running an upgrade, make sure to back up your existing installation using the Backup option on the Tools menu. If you are upgrading
More informationqpcr Hand Calculations - Multiple Reference Genes
qpcr Hand Calculations - Multiple Reference Genes This word document will take you through the hand calculations needed to determine the Normalized Gene Expression, using the Pfaffl method for efficiency
More informationManaging Modular Infrastructure by using OpenManage Essentials (OME)
Managing Modular Infrastructure by using OpenManage Essentials (OME) This technical white paper describes how to manage the modular infrastructure by using Dell EMC OME. Dell Engineering June 2017 A Dell
More informationSOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS
SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS0051-05 Introduction This document details the installation and known issues with OMNISEC v10.31. This release is a maintenance release to
More informationOperating Instructions. Windows. Video Wall Adjustment & Calibration Software. Video Wall Manager Ver. 1.8
Operating Instructions Video Wall Adjustment & Calibration Software Windows Video Wall Manager Ver. 1.8 Thank you for purchasing this Panasonic product. Before using this software, please read the instructions
More informationJuly SonicWall SonicOS 6.2 Upgrade Guide
SonicWall SonicOS 6.2 July 2017 This provides instructions for upgrading your SonicWall network security appliance from SonicOS 6.1 firmware or a previous version of SonicOS 6.2 to the latest version of
More information(1) DirectCD. Software Operating Instructions MVC-CD200/CD Sony Corporation
3-067-952-12(1) DirectCD Software Operating Instructions MVC-CD200/CD300 2001 Sony Corporation Notice for users Program Copyright 1999 Adaptec, Inc. All rights reserved./ Documentation 2001 Sony Corporation
More informationUser Bulletin. ABI PRISM 310 Genetic Analyzer. DRAFT February 6, :09 pm, A.fm. G5v2 Module for Use with Dye Set 33 (DS-33)
User Bulletin ABI PRISM 310 Genetic Analyzer Issue Date 2/06/03 SUBJECT: G5v2 Module for Use with Dye Set 33 (DS-33) In This User Bulletin Overview This user bulletin describes how to install the for the
More information1. What s New in Bio-Plex Manager 6.1
Bio-Rad Laboratories, Inc. Release Notes Bio-Plex Manager Software, Version 6.1 September 26, 2011 This document has the following sections: 1. What s New in Bio-Plex Manager 6.1 2. Resources for New Users
More informationC (1) Remote Controller. Setup software RM-IP Setup Tool guide Software Version Sony Corporation
C-499-100-12 (1) Remote Controller Setup software RM-IP Setup Tool guide Software Version 1.1 2017 Sony Corporation Table of Contents Introduction... 3 Using This Manual...3 Downloading and Starting the
More informationAgilent 6400 Series Triple Quadrupole LC/MS System
Agilent 6400 Series Triple Quadrupole LC/MS System Quick Start Guide Where to find information 4 Getting Started 6 Step 1. Start the Data Acquisition software 7 Step 2. Prepare the LC modules 13 Step 3.
More informationCopyright 2017 Xi'an NovaStar Tech Co., Ltd. All Rights Reserved. No part of this document may be copied, reproduced, extracted or transmitted in any
Product Version: Document Number: SmartLCT Screen Configuration Software V3.2.0 NS110100363 Copyright 2017 Xi'an NovaStar Tech Co., Ltd. All Rights Reserved. No part of this document may be copied, reproduced,
More informationApplied Biosystems PCR System 9700, Firmware v3.12 Release Notes Contents:
Applied Biosystems PCR System 9700, Firmware v3.12 Release Notes Contents: 1. Overview 2. Features 3. Confirm Firmware Version 4. Downloading Firmware 5. Upgrading Firmware 6. Known Issues 7. Legal Section
More informationLegal Notes. Regarding Trademarks KYOCERA MITA Corporation
Legal Notes Unauthorized reproduction of all or part of this guide is prohibited. The information in this guide is subject to change without notice. We cannot be held liable for any problems arising from
More informationMedivators Data Management Software Installation / User Manual
Medivators Data Management Software Installation / User Manual For DSD and SSD Endoscope Reprocessors Medivators is a registered trademark of Minntech Corporation. Windows and Windows Explorer, Excel,
More informationFUTEK USB Software Version User s Manual
FUTEK USB Software Version 2.0.0.0 User s Manual 10 Thomas, Irvine, CA 92618, USA Toll Free: (800) 23-FUTEK Telephone: (949) 465-0900 Fax: (949) 465-0905 futek@futek.com www.futek.com 2 Table of Contents
More informationED40. Development Kit. Quick Start Guide
ED40 Development Kit Quick Start Guide Disclaimer Honeywell International Inc. ( HII ) reserves the right to make changes in specifications and other information contained in this document without prior
More informationChromQuest 4.2 Chromatography Data System
ChromQuest 4.2 Chromatography Data System Installation Guide CHROM-97200 Revision A April 2005 2006 Thermo Electron Corporation. All rights reserved. Surveyor is a registered trademark and ChromQuest is
More informationWaters 2707 Autosampler ICS v1.20 Installer
R E L E A S E N O T E S Waters 2707 Autosampler ICS v1.20 Installer Read these release notes before you install and operate the Waters 2707 autosampler instrument control software (ICS). They contain information
More informationLudlum Lumic Calibration Software Manual
Ludlum Lumic Calibration Software Manual March 2016 Version 1.1.0.4 Calibration Wizard Version 1.0.5 or Higher Plateau Wizard Version 1.0.3 or Higher Table of Contents Software License Agreement... i
More information