ACMP 25th Annual Meeting

Size: px
Start display at page:

Download "ACMP 25th Annual Meeting"

Transcription

1 Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation ACMP 25th Annual Meeting Seattle, WA May 3, 2008 Melissa C. Martin, M.S., FACR Therapy Physics, Inc. 879 West 190th St., Ste 419, Gardena, CA Office: Cell:

2 Learning Objective Become familiar with the recommendations and requirements of the ACR Stereotactic Breast Biopsy Accreditation Program (SBBAP) Quality Control Manual Information By product of conducting Quality Control tests for Stereotactic Breast Biopsy units is greater familiarity with the operation and performance of your SBB system, particularly image quality and the accuracy of needle placement during SBB.

3

4 LORAD Stereotactic Breast Biopsy System

5 LORAD Stereotactic Breast Biopsy System

6 LORAD Stereotactic Breast Biopsy System Control Pendant

7 Cartesian coordinate system Geometry is reset for each patient x x z Don Jacobson 2/07 y

8 Siemens (Fischer) MammoVision Biopsy System

9 Siemens (Fischer) MammoVision Biopsy System

10 Fixed relationship between Autoguide and breast support Spherical coordinate system Distance Vertical angle Horizontal angle Don Jacobson 2/07

11 Don Jacobson 2/07

12

13 Don Jacobson 2/07

14 Goals of QC for Stereotactic Breast Biopsy To ensure that image quality in Stereotactic Breast Biopsy equals or exceeds that of screening and diagnostic mammography To ensure that equipment designed specifically for Stereo Breast Biopsy performs properly To ensure that needle localizations are accurate

15 General Requirements for SBBAP Qualified TEAM: Physicians, Technologist, and Medical Physicist Equipment: Table or add-on ;; film or digital QA Program, Manual, and Committee Technologist s s QC Testing - daily, weekly, monthly, semi-annual - 6 tests Medical Physicist s s QC Testing - acceptance and annual - 11 tests

16 The Quality Assurance Team: Physician QC Technologist Medical Physicist

17 Medical Physicist s s Qualifications for SBBAP: Initial Certification by ABR, ABMP in Diagnostic Medical Physics OR Alternate Education, Training and Experience AND >6/1/97: Have Performed 1 SBB Survey Under the Guidance of a Medical Physicist Qualified to Perform SBB Surveys <6/1/97: Have Performed 3 SBB Surveys

18 Medical Physicist Qualifications for SBBAP: Re-Accreditation After 3 Years At Least 1 SBB Physics Survey Per Year At Least 3 Hours of Continuing Education in SBB Physics Documentation of Above

19 Quality Control: Medical Physicist s s Evaluation Acceptance Test Before Patient Use Annually Thereafter Report Required as Part of ACR Application

20 Quality Control: Medical Physicist s s Evaluation The 1999 ACR SBB Quality Control Manual has a section for the Medical Physicist It has suggested Test Procedures, Forms, and Summary Report Format Detailed instructions on 11 Required Physicist s s tests

21 ACR Quality Control Manuals Mammography QC Manual (1990, 1992, 1994, 1999) Barium Enema QC Manual (1998) Stereotactic Breast Biopsy QC Manual (1999) MRI QC Manual (2001) Sent free to all facilities in program To purchase, call ACR Pubs: (800) QC forms available to anyone on Web site

22 Rad Tech QC Tests Mammo QC Tests Also Apply if Screen-Film Used Localization Accuracy - daily Phantom Image - weekly Hardcopy Output Quality - monthly, if app Visual Checklist - monthly Compression Force - semi-annually Repeat Analysis - semi-annually Zero Alignment Test before ea patient, if app

23 Medical Physicist s s Quality Control Tests 1. Stereotactic Unit Assembly Evaluation 2. Collimation Assessment 3. Focal Spot Performance & Digital System Limiting Resolution 4. kvp Accuracy and Reproducibility 5. Beam Quality Assessment (HVL) 6. AEC or Manual Exposure Performance

24 Medical Physicist s s Quality Control Tests 7A. Uniformity of Screen Speed 7B. Digital Receptor Uniformity 8. Breast Entrance Exposure, Average Glandular Dose, and Exposure Reproducibility 9. Image Quality Evaluation 10. Artifact Evaluation 11. Localization Accuracy Test

25 Stereotactic Unit Assembly Evaluation

26 Collimation Assessment

27 Collimation Assessment

28 Collimation Assessment

29 Collimation Assessment SID = 84 cm

30 Digital Limiting Resolution/Focal Spot

31 Digital Limiting Resolution/Focal Spot

32 Digital Limiting Resolution/Focal Spot

33 kvp Accuracy and Reproducibility

34 kvp Accuracy and Reproducibility

35 Beam Quality Assessment

36 Beam Quality Specifications for SBB Units The minimum acceptable Half-Value Layer measurement on a digital or film/screen SBB unit is Action Limit: If measured HVL < (kvp/100) (in mm Al) or if measured HVL > (kvp/100) + C (in mm Al) where C = 0.12 for Mo/Mo, C = 0.19 for Mo/Rh, and C = 0.22 for Rh/Rh, then seek service correction.

37 Image Quality Evaluation (Phantom) Objective: Ensure Image Quality for SBB meets or exceeds that of mammography, and to detect temporal changes in image quality Procedure: Same as for Mammography, except ACR phantom must be imaged in 4 separate quadrants for digital because of small field of view

38 Two Types of Approved Phantoms Mini Stereotactic Breast Biopsy Accreditation Phantom Nuclear Associates Mammography Accreditation Phantom RMI 156 Nuclear Associates

39 Mini Stereotactic Breast Biopsy Accreditation Phantom Chest Wall Side

40 SBBAP Testing Criteria Dose and Phantom Dose - Must be less than 300 mrads (3 mgy) Phantom image quality Fibers Speck Groups Masses MAP Phantom F/S Digital Mini Phantom F/S Digital

41 Image Quality for SBB Units RMI 156 or NA MAP Phantom

42 on Jacobson 2/07 NA Mini- Phantom

43 RMI 156 Accreditation Phantom

44 Nuclear Associates Digital Mini Phantom

45 AEC or Manual Exposure Performance 2 cm Thickness

46 AEC or Manual Exposure Performance 4 cm Thickness - Also used for Uniformity and Artifacts

47 AEC or Manual Exposure Performance 6 cm Thickness 8 cm Thickness

48 AEC or Manual Exposure Performance

49 n AEC or Manual Exposure Control a l Performance Requirement e Action Limit (Digital): x c If the signal range e exceeds ±20% of signal for 4 cm phantom, e d s revise technique chart. s i g r a n g e ± 2 0 % Action Limit (Screen-Film): If the density o f range exceeds ±0.15 of mean, revise technique chart. s i g n a l f o r 4 c m

50 Digital Receptor Uniformity Requirements Digital Receptor Uniformity Requirements For Units with ROI statistics measurement For Units with ROI statistics measurement capability: capability: Action Limit: If SNR(I) / Action Limit: If SNR(I) / SNR(Center SNR(Center) is > 1.15 ) is > 1.15 or < 0.85, seek service correction. or < 0.85, seek service correction. I f m e a s u r e d H V L < ( k V p / ) ( i n m m A l ) I f m e a s u r e d H V L < ( k V p / ) ( i n m m A l )

51 Digital Receptor Uniformity Requirements Digital Receptor Uniformity Requirements For Units without ROI statistics For Units without ROI statistics measurement capability: measurement capability: I f m e a s u r e d H V L < ( k V p / ) ( i n m m A l ) I f m e a s u r e d H V L < ( k V p / ) ( i n m m A l ) If line w/o black dots > 1/4 length of image, seek service correction If non-uniform areas (w/o black dots) > 10% of image or Action Limits: If geometric pincushioning > 1 cm from edge of image or

52 Digital Receptor Uniformity

53 Digital Receptor Uniformity - Image Statistics

54 Breast Entrance Exposure, Average Glandular Dose, and Exposure Reproducibility Same Procedure as for Mammography Recommended Signal Level for Digital Digital Matrix Sizes Performance Criteria: a) Coefficient of Variation < 0.05 b) Av. Glandular Dose < 3 mgy for Screen/Film and for Digital Image Receptors

55 Entrance Exposure/Mid-Glandular Doses

56 Entrance Exposure/Mid-Glandular Doses

57 Entrance Exposure/Mid-Glandular Doses

58 Artifact Evaluation

59 11. Localization Accuracy: Gelatin Phantom Objective: To assure that the biopsy needle is accurately placed for sampling as directed from the stereotactic scout images Technologist to perform test Physicist to observe and analyze results End-to to-end test which supplements the daily in-air positioning accuracy test

60 Localization Accuracy: Gelatin Phantom Method 1. Position Needle: - Target Lesion Using Stereo Views - Position Core Needle to Proper X, Y, and Z Coordinates 2. Verify Needle Position: - Acquire Stereo Pre-fire Images - Needle Tip should be within Lesion 3. Fire Gun

61 Localization Accuracy: Method Gelatin Phantom 4. Verify Post-Fire Position - Acquire Post-Fire Stereo Images - Needle Tip should be beyond Center of Lesion 5. Verify Sampling of Lesion - Examine Contents of Core Sample

62 Gelatin Phantom Biopsy

63 Gelatin Biopsy Images

64 Localization Accuracy

65

66

67 SAMS Questions

68 0% 0% 0% The maximum allowable patient mid-glandular dose for a 4.5 cm breast equivalent phantom under MQSA is 1. 1 mgy 2. 2 mgy 3. 3 mgy 0% 0% 4. 5 mgy mgy 10

69 Answer: C - 3 mgy Ref: Stereotactic Breast Biopsy Accreditation Manual American College of Radiology (1997)

70 Using the dedicated mini-phantom designed for SBB units, the minimum acceptable phantom image quality scores on a unit with digital imaging are: 0% 0% 0% 1. 2 fibers, 3 speck groups, 2.5 masses 2. 3 fibers, 3 speck groups, 2.5 masses 3. 4 fibers, 3 speck groups, 3.0 masses 0% 0% 4. 3 fibers, 2.5 speck groups, 2.5 masses 5. 3 fibers, 4 speck groups, 3.0 masses 10

71 Answer: b - 3 fibers, 3 speck groups, 2.5 masses Ref: Stereotactic Breast Biopsy Accreditation Manual American College of Radiology (1997)

72 The minimum acceptable Half Value Layer measurement on a digital SBB unit at 28 kvp is 0% 0% 0% 0% 0% mm Al mm Al mm Al mm Al mm Al 10

73 Answer: e mm Al Ref: Stereotactic Breast Biopsy Accreditation Manual American College of Radiology (1997) Min. HVL = (kvp/100) mm Al

74 The minimal acceptable uniformity level for a digital receptor on a SBB unit is 0% 0% 0% 0% 0%

75 Answer: d Ref: Stereotactic Breast Biopsy Accreditation Manual American College of Radiology (1997) Action Limit: If SNR(I) / SNR(Center) ) is > 1.15 or < 0.85, seek service correction.

76 The AEC or Manual Exposure Control Performance requirement on a SBB unit with digital imaging is 0% 0% 0% 0% 0% 1. +/ of the average signal for thickness of cm 2. +/ of the average signal for 4 cm phantom 3. +/ of the average signal for thicknesses of cm 4. +/ of the average signal for 4 cm phantom 5. +/ of the average signal for thicknesses of cm 10

77 Answer: d: +/- 20% of average signal for 4 cm phantom Ref: Stereotactic Breast Biopsy Accreditation Manual American College of Radiology (1997)

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc. Unique Features of the GE Senoclaire Tomosynthesis System Tyler Fisher, M.S., DABR Therapy Physics, Inc. Conflict of Interest Disclosure I have no conflicts to disclose. Learning Objectives Overview of

More information

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc.

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc. Lorad FFDM QC Procedures for Medical Physicist Tao Wu, Ph.D. Hologic, Inc. Lorad Selenia FFDM Tests Following Lorad QC Manual Collimation Assessment Artifact Evaluation System Resolution Phantom Image

More information

RECENTLY APPROVED GE FFDM DBT TESTING PROCEDURES.

RECENTLY APPROVED GE FFDM DBT TESTING PROCEDURES. RECENTLY APPROVED GE FFDM DBT TESTING PROCEDURES. S. G U R U P R A S A D, P H. D., D A B R, E M E R I T U S, M AT H E W H A L L M. S. N O RT H S H O R E U N I V E R S I T Y H E A LT H S Y S T E M A N D

More information

Certificate Extension II 8/1/2016. Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role

Certificate Extension II 8/1/2016. Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role Kish Chakrabarti, Ph.D., FAAPM Division of Mammography Quality Standards Center for Devices and Radiological

More information

8/2/2016. Hands-on GE SenoClaire DBT: technical characteristics & quality control. Disclosures and Acknowledgements

8/2/2016. Hands-on GE SenoClaire DBT: technical characteristics & quality control. Disclosures and Acknowledgements Hands-on GE SenoClaire DBT: technical characteristics & quality control Razvan Iordache, Ph.D. GE Healthcare Imagination at work Disclosures and Acknowledgements Razvan Iordache Employee of GE Healthcare

More information

DISC QA Mamchex Tool Use Instructions with CARESTREAM DIRECTVIEW CR Mammography Systems

DISC QA Mamchex Tool Use Instructions with CARESTREAM DIRECTVIEW CR Mammography Systems DISC QA Mamchex Tool Use Instructions with CARESTREAM DIRECTVIEW CR Mammography Systems Purpose This document describes and provides instructions for use of the QA Mamchex Tool in calibrating a mammography

More information

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance

More information

연구용유방초음파질관리 원광대학병원김혜원

연구용유방초음파질관리 원광대학병원김혜원 연구용유방초음파질관리 원광대학병원김혜원 Why US QC? Quality control (QC) testing of ultrasound scanners is important to verify the proper and consistent operation of these devices. main goal ; quality improvement Guidelines

More information

MAMMOMAT 1000 / 3000 Nova. Accessories. Accessories (optional) Compression plate with low edge 18 cm x 24 cm #

MAMMOMAT 1000 / 3000 Nova. Accessories. Accessories (optional) Compression plate with low edge 18 cm x 24 cm # MAMMOMAT 1000 / 3000 Nova Accessories Accessories (optional) Compression plate with low edge 18 cm x 24 cm # 44 94 378 Standard compression plate with 40 mm high front edge. Compression plate with high

More information

Discontinued Products

Discontinued Products Discontinued Products January 2019 CIRS was founded in 1983 to improve upon existing tissue simulation methodology and provide quantitative reference standards for Computed Tomography. Today CIRS is recognized

More information

What is New In MOC? Anne Roberts, M.D. UCSD Medical Center

What is New In MOC? Anne Roberts, M.D. UCSD Medical Center What is New In MOC? Anne Roberts, M.D. UCSD Medical Center American Board of Radiology Responsible for certifying candidates in Radiology Specialties Board Examinations MOC Quality of Radiologists ABR

More information

SUMMARY OF DENTAL MEASUREMENT PROCEDURES (Abridged Protocol)

SUMMARY OF DENTAL MEASUREMENT PROCEDURES (Abridged Protocol) SUMMARY OF DENTAL MEASUREMENT PROCEDURES (Abridged Protocol) INTRAORAL IMAGING PROCEDURE Entrance Skin Exposure / Air Kerma To measure the typical intraoral Entrance Skin Exposure (ESE) and Entrance Skin

More information

AAPM Standard of Practice: CT Protocol Review Physicist

AAPM Standard of Practice: CT Protocol Review Physicist AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose

More information

Acknowledgments and financial disclosure

Acknowledgments and financial disclosure AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories

More information

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives 1. Be familiar with recommendations

More information

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging Stephen J. Glick* J. Michael O Connor**, Clay Didier**, Mini Das*, * University of Massachusetts Medical

More information

LAMBDA DATA SHEET. ANALOGIC mammography unit. ANALOGIC mammography unit. Rev. n.05 of 15/01/13

LAMBDA DATA SHEET. ANALOGIC mammography unit. ANALOGIC mammography unit. Rev. n.05 of 15/01/13 NOTE The manufacturer reserves the right to make further improvements while keeping main features unchanged MAIN CHARACTERISTICS Line voltage 220/230/240 VAC 10% @ 50/60 Hz Power 6.6 kva (0.5 kva stand-by)

More information

Mammographic QC Kit Gammex 182B Routine Mammography QC Kit Gammex Processor QC Kit Gammex 185C... 13

Mammographic QC Kit Gammex 182B Routine Mammography QC Kit Gammex Processor QC Kit Gammex 185C... 13 INDEX PHANTOMS Mammographic Accreditation Phantom 156... 1 Digital Mammographic Phantom 156S... 2 Rachel Anthropomorphic Breast Phantom 169... 2 Ultrasound Biopsy Phantom 429... 3 Stereotactic Breast Biopsy

More information

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology.

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology. Fujifilm DR Solution FDR AcSelerate The new pinnacle in diagnostic imaging from Fujifilm CsI Scintillator ISS Technology Dynamic Visualization Welcome to the X-ray room of the future! A streamlined solution

More information

Product data. Mammographic System

Product data. Mammographic System Technologically advanced mammographic system, offering the best quality/price ratio. Lightweight and easy-to-use, facilitating the operator s job. Excellent diagnostic image quality, similar to the top-of

More information

Choosing a teleradiology provider

Choosing a teleradiology provider Choosing a teleradiology provider Tim Hunter, MD Professor Emeritus, Department of Medical Imaging University of Arizona Purpose of this talk Provide brief overview of teleradiology Provide a general guideline

More information

Kevin O Donnell Senior R&D Manager Toshiba Medical Research Institute Co-Chair, DICOM Standards Committee Past Chair, IHE Radiology Planning

Kevin O Donnell Senior R&D Manager Toshiba Medical Research Institute Co-Chair, DICOM Standards Committee Past Chair, IHE Radiology Planning 1 Kevin O Donnell Senior R&D Manager Toshiba Medical Research Institute Co-Chair, DICOM Standards Committee Past Chair, IHE Radiology Planning Committee Standards & Tools 2 Learning Objectives 1) DICOM

More information

Preparing for Part 2 of the ABR Diagnostic Physics Exam

Preparing for Part 2 of the ABR Diagnostic Physics Exam Preparing for Part 2 of the ABR Diagnostic Physics Exam Joseph Zambelli, Ph.D. Spectrum Health, Grand Rapids, MI Outline Exam Background Exam Preparation Suggested Resources Location, Time, and Format

More information

S. Guru Prasad, Ph.D., DABR

S. Guru Prasad, Ph.D., DABR PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information

More information

Chapter 1 General Information

Chapter 1 General Information Chapter 1 General Information System Description Chapter 1 General Information 1.0 System Description The Affirm attaches to the Selenia Dimensions. A biopsy device attaches to the Affirm. X- and Y-axes

More information

1 Overview All Breast Biopsy Recommendations & Incompletes All Breast Six Month Follow-Up... 3

1 Overview All Breast Biopsy Recommendations & Incompletes All Breast Six Month Follow-Up... 3 Planning Worksheet Exam Types for Mammography Tracking Date: Site Information Site Name: Site Number: Site Contacts Title Name Phone E-mail System Administrator MQSA Compliance Officer When finished, email

More information

James R Halama, PhD Loyola University Medical Center

James R Halama, PhD Loyola University Medical Center James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives Be familiar with the tests recommended

More information

DIAGNOSTIC IMAGING QA & PATIENT ALIGNMENT

DIAGNOSTIC IMAGING QA & PATIENT ALIGNMENT 2018 Product Solutions DIAGNOSTIC IMAGING QA & PATIENT ALIGNMENT CT Ultrasound Mammography DR/CR/Fluoro Radiation Oncology Patient Alignment PRODUCT SOLUTIONS // Introduction PROVIDING PROVEN SOLUTIONS

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

4/19/2016. Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT

4/19/2016. Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT 1 Three manufacturers approved for Tomo Hologic and GE, and Siemens Why 2D Digital Mammography 2D FFDM it appears to

More information

Photon beam dose distributions in 2D

Photon beam dose distributions in 2D Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation

More information

Planmed Sophie Classic Mobile. English. dedicated for mobile mammography

Planmed Sophie Classic Mobile. English. dedicated for mobile mammography Sophie Classic Mobile English at its best Sophie Classic Mobile proven mammography on wheels The Sophie Classic Mobile system takes transportable mammography to a new level. Generously equipped with integrated

More information

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners ImPACT Imaging Performance Assessment of CT Scanners A Medical Devices Agency Evaluation Group CT dosimetry and a data base for CTDI values EFOMP Workshop at ECR 2002 S. Edyvean, ImPACT St George s Hospital,

More information

ABR CERTIFICATION OF MEDICAL PHYSICISTS CERTIFICATION AMERICAN BOARD OF RADIOLOGY 7/22/2014

ABR CERTIFICATION OF MEDICAL PHYSICISTS CERTIFICATION AMERICAN BOARD OF RADIOLOGY 7/22/2014 ABR CERTIFICATION OF MEDICAL PHYSICISTS Geoffrey S. Ibbott, Ph.D. AAPM Annual Meeting July 22, 2014 CERTIFICATION Part of a life-long process From Initial Training to Retirement AMERICAN BOARD OF RADIOLOGY

More information

A Method for Producing Simulated Mammograms: Observer Study

A Method for Producing Simulated Mammograms: Observer Study A Method for Producing Simulated Mammograms: Observer Study Payam Seifi M.Sc. Michael R. Chinander Ph.D. Robert M. Nishikawa Ph.D., FAAPM Carl J. Vyborny Translational Laboratory for Breast Imaging Research

More information

RaySafe Solo. Specifications

RaySafe Solo. Specifications RaySafe Solo Specifications RAYSAFE SOLO GENERAL EMC TESTED According to EN 61000-6-1:2007 and EN 61000-6-3:2007 EXPOSURE NEEDED One RESET Automatic TEMP. 15 35 C (59 95 F) DETECTOR CABLE LENGTH 2 and

More information

Proposed Catalog Changes TECH-PREP ENHANCED SKILLS CERTIFICATE RADIOLOGIC TECHNOLOGY (Computed Tomography) These courses will prepare Registered Radiographer, Radiation Therapist, or Nuclear Medicine Technologist

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

OPERATION MANUAL 4th Edition

OPERATION MANUAL 4th Edition DRYPIX 5000/7000 Quality Control for Mammography OPERATION MANUAL 4th Edition October 2008 897N0370D 2 897N0370A 09.2005 FOREWORD The present document, DRYPIX 5000/7000 Quality Control for Mammography

More information

Initial Clinical Experience with 3D Surface Image Guidance

Initial Clinical Experience with 3D Surface Image Guidance Initial Clinical Experience with 3D Surface Image Guidance Amanda Havnen-Smith, Ph.D. Minneapolis Radiation Oncology Ridges Radiation Therapy Center Burnsville, MN April 20 th, 2012 Non-funded research

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique

Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique Elena Marimón *a,b, Hammadi Nait-Charif a, Asmar Khan b, Philip A. Marsden c, Oliver Diaz d a CDE

More information

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them Dose Distributions George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Purpose To familiarize the resident with dose distributions and the factors that affect them

More information

10/9/2018. Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT

10/9/2018. Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT 1 Mammography Five FFDM approved for Tomo Hologic, Ge Senoclair and GE Pristina, Siemens, and Fujifilm 2 Why 2D Digital

More information

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT. Progress report: Development of assessment and predictive metrics for quantitative imaging in chest CT Subaward No: HHSN6801000050C (4a) PI: Ehsan Samei Reporting Period: month 1-18 Deliverables: 1. Deployment

More information

CURRICULUM COMMITTEE MEETING Friday, March 18, :00 p.m. Student Life Center, Faculty Dining Room (Building 23, First Floor) AGENDA

CURRICULUM COMMITTEE MEETING Friday, March 18, :00 p.m. Student Life Center, Faculty Dining Room (Building 23, First Floor) AGENDA CURRICULUM COMMITTEE MEETING Friday, March 18, 2016-2:00 p.m. Student Life Center, Faculty Dining Room (Building 23, First Floor) I. Call to Order AGENDA II. Roll Call III. Minutes of meeting of January

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Optimisation of Toshiba Aquilion ONE Volume Imaging

Optimisation of Toshiba Aquilion ONE Volume Imaging Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was

More information

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60601-2-44 2001 AMENDMENT 1 2002-09 Amendment 1 Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment for computed tomography Amendement

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL Control-X Medical Z-MOTION Compact design, low ceiling height requirement Motorized and manual movement capability Wide motion / SID range Best-in-class image quality Flexible connectivity to PACS systems

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package.

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package. So much is asked of medical physicists these days We Can Help Automated Radiation Therapy Phantom Analysis in Seconds QC for Therapy OBI s and CT simulators. Radia s Catphan/OBI module performs analysis

More information

Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors

Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors Elena Marimon 1,2*, Hammadi Nait-Charif 1, Asmar Khan 2, Philip A. Marsden 3, Oliver Diaz 4 1 Centre for

More information

Nicholas Marshall Department of Radiology, University Hospitals Leuven, Herestraat 49, 3000 Leuven, Belgium

Nicholas Marshall Department of Radiology, University Hospitals Leuven, Herestraat 49, 3000 Leuven, Belgium Practical Methods for Assessing Image Quality in Diagnostic Radiology Nicholas Marshall Department of Radiology, University Hospitals Leuven, Herestraat 49, 3000 Leuven, Belgium general x-ray conebeam

More information

AUDIT PROGRAM. Revision 6 Dated September 29, Management Systems Analysis, Inc. P.O. Box 136, Royersford, PA

AUDIT PROGRAM. Revision 6 Dated September 29, Management Systems Analysis, Inc. P.O. Box 136, Royersford, PA AUDIT PROGRAM Revision 6 Dated September 29, 2010 Management Systems Analysis, Inc. P.O. Box 136, Royersford, PA 19468 610-409-0168 jhighl@aol.com Approved: 1.0 Objective 2.0 Scope 3.0 General To describe

More information

Future Topics. Projection Imaging Dose Reporting, XA 3D Volume Objects. for DICOM WG-02. Presented by Heinz Blendinger, Siemens Medical Solutions

Future Topics. Projection Imaging Dose Reporting, XA 3D Volume Objects. for DICOM WG-02. Presented by Heinz Blendinger, Siemens Medical Solutions Future Topics for Projection Imaging Dose Reporting, XA 3D Volume Objects DICOM WG-02 Presented by Heinz Blendinger, Siemens Medical Solutions 1 Presentation outline Dose Reporting Why Dose Reporting?

More information

Monte Carlo simulation of scatter field for calculation of contrast of discs in synthetic CDMAM images

Monte Carlo simulation of scatter field for calculation of contrast of discs in synthetic CDMAM images Monte Carlo simulation of scatter field for calculation of contrast of discs in synthetic CDMAM images Oliver Díaz 1, Mary Yip 1, Jorge Cabello 1, David R. Dance 2,3, Kenneth C. Young 2,3, and Kevin Wells

More information

3D printing technologies

3D printing technologies AFTERNOON SESSION / 5 / 3:30 4:00 / 9 Sep, 2016 method for secondary James Jung Department of Radiation Oncology, University of Florida Courtesy of ForUsDocs/consultant360 webpage

More information

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell

More information

ABR Update: MOC & PQI. Lane F. Donnelly MD Nemours Children s Hospital Orlando, FL

ABR Update: MOC & PQI. Lane F. Donnelly MD Nemours Children s Hospital Orlando, FL ABR Update: MOC & PQI Lane F. Donnelly MD Nemours Children s Hospital Orlando, FL Expiration / Disclosure The following slides are true through May 2013 ABR - Pediatric Radiology Trustees Donald P. Frush

More information

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4

More information

Scattered radiation in DBT geometries with flexible breast compression paddles: A Monte Carlo simulation study

Scattered radiation in DBT geometries with flexible breast compression paddles: A Monte Carlo simulation study Scattered radiation in DBT geometries with flexible breast compression paddles: A Monte Carlo simulation study Oliver Diaz, Eloy García, Arnau Oliver, Joan Martí, Robert Martí a a VICOROB research institute,

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

ASSESSMENT CRITERIA FOR RADIOGRAPHER PROGRAM. To determine whether the identified outcomes are being met, the Program will assess its effectiveness.

ASSESSMENT CRITERIA FOR RADIOGRAPHER PROGRAM. To determine whether the identified outcomes are being met, the Program will assess its effectiveness. ASSESSMENT CRITERIA FOR RADIOGRAPHER PROGRAM To determine whether the identified outcomes are being met, the will assess its effectiveness. Responsibility: Director, Instructors, Clinical Staff. Standard:

More information

Thank-You Members of TG147 TG 147: QA for nonradiographic

Thank-You Members of TG147 TG 147: QA for nonradiographic Thank-You Members of TG147 TG 147: QA for nonradiographic localization and positioning systems Twyla Willoughby, M.S. Medical Physicist Clinical AAPM Meeting March 2013 Department of Radiation Oncology

More information

Certificate in Clinician Performed Ultrasound (CCPU)

Certificate in Clinician Performed Ultrasound (CCPU) Certificate in Clinician Performed Ultrasound (CCPU) Syllabus Physics Tutorial Physics Tutorial Purpose: Training: Assessments: This unit is designed to cover the theoretical and practical curriculum for

More information

2011 Clinical. Breast Imaging. Superior Image Quality. FREE SYLLABUS with purchase of entire set 16 AMA PRA Category 1 Credit(s) tm

2011 Clinical. Breast Imaging. Superior Image Quality. FREE SYLLABUS with purchase of entire set 16 AMA PRA Category 1 Credit(s) tm A DVD Teaching Program 2011 Clinical Breast Imaging Digital Mammography 1.5 AMA PRA Category 1 Credit(s) TM Breast Ultrasound 1.0 AMA PRA Category 1 Credit(s) TM Breast MR 1.75 AMA PRA Category 1 Credit(s)

More information

Advanced Systems, Inc. Course. Registration Form. (Please complete all information)

Advanced Systems, Inc. Course. Registration Form. (Please complete all information) MARCH 9 TO 11, 2015 Registration Form (Please complete all information) To register on-line for the Environmental Laboratory Assessments Basic Assessor Training. Course offered in Olathe, Kansas on March

More information

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods Stereoscopic Breast Imaging Andrew D. A. Maidment, Ph.D. Chief, Physics Section Department of Radiology University of Pennsylvania Limitations of Projection Radiography Mammography is a projection imaging

More information

Technical aspects of SPECT and SPECT-CT. John Buscombe

Technical aspects of SPECT and SPECT-CT. John Buscombe Technical aspects of SPECT and SPECT-CT John Buscombe What does the clinician need to know? For SPECT What factors affect SPECT How those factors should be sought Looking for artefacts For SPECT-CT Issues

More information

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan

More information

MOC Update for Medical Physicists ABR MOC Medical Physics 3/14/2017 Disclosures NOTE Maintenance of Certification Why do we have MOC?

MOC Update for Medical Physicists ABR MOC Medical Physics 3/14/2017 Disclosures NOTE Maintenance of Certification Why do we have MOC? Disclosures MOC Update for Medical Physicists AAPM Spring Clinical Meeting I am a part time paid employee of the American Board of Radiology G. Donald Frey Associate Executive Director American Board of

More information

Philips SPECT/CT Systems

Philips SPECT/CT Systems Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET

More information

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce

More information

QUALITY CONTROL IN MEDICAL IMAGING FULL MEDICAL IMAGING QA SOLUTIONS: BEAM QA IMAGE QA DISPLAY QA PATIENT DOSE QA

QUALITY CONTROL IN MEDICAL IMAGING FULL MEDICAL IMAGING QA SOLUTIONS: BEAM QA IMAGE QA DISPLAY QA PATIENT DOSE QA iba-dosimetry.com 1 QUALITY CONTROL IN MEDICAL IMAGING FULL MEDICAL IMAGING QA SOLUTIONS: BEAM QA IMAGE QA DISPLAY QA PATIENT DOSE QA TABLE OF CONTENTS Quality Control for Diagnostic Imaging Modalities

More information

Micro-CT Methodology Hasan Alsaid, PhD

Micro-CT Methodology Hasan Alsaid, PhD Micro-CT Methodology Hasan Alsaid, PhD Preclinical & Translational Imaging LAS, PTS, GlaxoSmithKline 20 April 2015 Provide basic understanding of technical aspects of the micro-ct Statement: All procedures

More information

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND

More information

Rad Tech Study Guide

Rad Tech Study Guide Rad Tech Study Guide If you are looking for a ebook Rad tech study guide in pdf form, then you have come on to correct website. We present complete variation of this ebook in epub, doc, txt, DjVu, PDF

More information

NDTMA 2015 Annual Conference

NDTMA 2015 Annual Conference NDTMA 2015 Annual Conference Golden Nugget Hotel, Las Vegas, NV February 10-13, 2015 What was removed from the NAS 410 standard What was added to the NAS 410 standard What needs to be added to the employer

More information

Philips MicroDose L30

Philips MicroDose L30 English DICOM Conformance Statement Philips MicroDose L30 Software 8.3 P1 DICOM Conformance Statement - Philips MicroDose L30, Software 8.3 P1 Issued by: Philips Medical Systems Nederland BV, a Philips

More information

Monte Carlo Reference Data Sets for Imaging Research

Monte Carlo Reference Data Sets for Imaging Research AAPM REPORT NO. 195 Monte Carlo Reference Data Sets for Imaging Research The Report of AAPM Task Group 195 September 2015 DISCLAIMER: This publication is based on sources and information believed to be

More information

Accessories for RaySafe Diagnostic X-ray Equipment

Accessories for RaySafe Diagnostic X-ray Equipment Accessories for RaySafe Diagnostic X-ray Equipment RaySafe Solo RaySafe Xi RaySafe X2 Solo RaySafe X2 CASES 1922014/1922016* 4559871/4559880* RAYSAFE Xi STORM CASE Heavy duty, waterproof case with customized

More information

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE / Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis

More information

COMPUTED RADIOGRAPHY. Mammography. and more. DX-M - Computed Radiography from Agfa HealthCare. For Digital Mammography and General Radiography

COMPUTED RADIOGRAPHY. Mammography. and more. DX-M - Computed Radiography from Agfa HealthCare. For Digital Mammography and General Radiography COMPUTED RADIOGRAPHY Mammography and more DX-M - Computed Radiography from Agfa HealthCare For Digital Mammography and General Radiography Mammo Needle-based technology delivers quality images Agfa HealthCare

More information

Effective detective quantum efficiency (edqe) and effective noise equivalent quanta (eneq) for system optimization purposes in digital mammography

Effective detective quantum efficiency (edqe) and effective noise equivalent quanta (eneq) for system optimization purposes in digital mammography Effective detective quantum efficiency (edqe) and effective noise equivalent quanta (eneq) for system optimization purposes in digital mammography Elena Salvagnini a,b, Hilde Bosmans a, Lara Struelens

More information

The smaller, simpler, safer alternative to radioisotope irradiators. Precision s MultiRad160 is ideal for irradiating larger cell cultures and/or

The smaller, simpler, safer alternative to radioisotope irradiators. Precision s MultiRad160 is ideal for irradiating larger cell cultures and/or The smaller, simpler, safer alternative to radioisotope irradiators. Precision s MultiRad160 is ideal for irradiating larger cell cultures and/or superficial small animal irradiation. STOP Off On The MultiRad160

More information

Breast Tomosynthesis: Impact on IT Systems. dondennison.com

Breast Tomosynthesis: Impact on IT Systems. dondennison.com Breast Tomosynthesis: Impact on IT Systems MIIT 2016, HAMILTON Don K Dennison Industry consultant with 14+ years experience in imaging IT Society for Imaging Informatics in Medicine (SIIM) Board of Directors

More information

PCXMC 2.0 SUPPLEMENTARY PROGRAMS USER'S GUIDE

PCXMC 2.0 SUPPLEMENTARY PROGRAMS USER'S GUIDE FEBRUARY 2012 PCXMC 2.0 SUPPLEMENTARY PROGRAMS USER'S GUIDE PCXMC20Rotation.exe Autocalc-sheet.xls AutocalcRotation-sheet.xls Markku Tapiovaara STUK - Radiation and Nuclear Safety Authority P.O. Box 14

More information

NIAC Membership Application Checklists

NIAC Membership Application Checklists NIAC Membership Application Checklists Thank you for your interest in joining NIAC. To ensure the NIAC membership requirements are met, please review your processes for qualifying lead auditors and performing

More information

QIBA Profile: CT Tumor Volume Change for Advanced Disease (CTV-AD)

QIBA Profile: CT Tumor Volume Change for Advanced Disease (CTV-AD) 5 QIBA Profile: CT Tumor Volume Change for Advanced Disease (CTV-AD) Stage 3: Technically Confirmed June 22, 2018 When referencing this document, please use the following format: QIBA CT Volumetry Technical

More information

UNCOMPROMISING QUALITY

UNCOMPROMISING QUALITY ION CHAMBERS UNCOMPROMISING QUALITY Designed with over 30 years of scientific integrity for a broad range of dosimetry measurements in diverse radiation beams Farmer-type Chambers For absolute dosimetry

More information

QIBA Profile: CT Tumor Volume Change for Advanced Disease (CTV-AD)

QIBA Profile: CT Tumor Volume Change for Advanced Disease (CTV-AD) 5 QIBA Profile: CT Tumor Volume Change for Advanced Disease (CTV-AD) Stage: Consensus+ Draft Tech Confirmed When referencing this document, please use the following format: QIBA CT Volumetry Technical

More information

QIBA Profile: CT Tumor Volume Change for Advanced Disease (CTV-AD)

QIBA Profile: CT Tumor Volume Change for Advanced Disease (CTV-AD) 5 QIBA Profile: CT Tumor Volume Change for Advanced Disease (CTV-AD) Stage: Consensus When referencing this document, please use the following format: QIBA CT Volumetry Technical Committee. CT Tumor Volume

More information

An Employer Guide to AMT Certification

An Employer Guide to AMT Certification An Employer Guide to AMT Certification Why Certification Is Important AS AN EMPLOYER, you want to hire qualified personnel to ensure that your patients get the best care possible. AMT shares in this goal

More information

Design Considerations in Optimizing a Breast Tomosynthesis System

Design Considerations in Optimizing a Breast Tomosynthesis System Design Considerations in Optimizing a Breast Tomosynthesis System Andrew Smith, Ph.D., Vice President - Imaging Science, Hologic Introduction Breast tomosynthesis, also referred to as three-dimensional

More information

ENGLISH. Digital Breast Tomosynthesis

ENGLISH. Digital Breast Tomosynthesis ENGLISH Digital Breast Tomosynthesis Medical imaging excellence Planmed is strongly committed to improving the early detection of breast cancer. Our mission is to provide top quality imaging systems for

More information