Prospective Motion Correction of X-ray Images for Coronary Interventions

Size: px
Start display at page:

Download "Prospective Motion Correction of X-ray Images for Coronary Interventions"

Transcription

1 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 1 Prospective Motion Correction of X-ray Images for Coronary Interventions Guy Shechter, Barak Shechter, Jon R. Resar, Rafael Beyar Guy Shechter is with the Rappaport Faculty of Medicine, Technion - Israel Institute of Technology, Haifa, 31096, Israel. ( gshec[at]tx.technion.ac.il). Barak Shechter is with the Department of Anatomy and Neurobiology, University of Maryland School of Medicine, Baltimore, MD 21201, USA. Jon R. Resar is with the Division of Cardiology, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USA. Rafael Beyar is with the Rappaport Faculty of Medicine, Technion - Israel Institute of Technology, Haifa, 31096, Israel. c 2004 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE.

2 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 2 Abstract A method for prospective motion correction of x-ray imaging of the heart is presented. A 3D+t coronary model is reconstructed from a biplane coronary angiogram obtained during free breathing. The deformation field is parameterized by cardiac and respiratory phase, which enables the estimation of the state of the arteries at any phase of the cardiac respiratory cycle. The motion of the 3D coronary model is projected onto the image planes and used to compute a dewarping function for motion correcting the images. The use of a 3D coronary model facilitates motion correction of images acquired with the x-ray system at arbitrary orientations. The performance of the algorithm was measured by tracking the motion of selected left coronary landmarks using a template matching cross correlation. In three patients, we motion corrected the same images used to construct their 3D+t coronary model. In this best case scenario, the algorithm reduced the motion of the landmarks by 84-85%, from mean RMS displacements of pixels to pixels. Prospective motion correction was tested in five patients by building the coronary model from one dataset, and correcting a second dataset. The patient s cardiac and respiratory phase are monitored and used to calculate the appropriate correction parameters. The results showed a 48-63% reduction in the motion of the landmarks, from a mean RMS displacement of pixels to pixels. Index Terms Motion compensation, X-ray angiography, Chest imaging I. INTRODUCTION X-ray imaging is routinely used to guide intravascular therapy. Percutaneous transluminal coronary angioplasty and stent placement are two common cardiac procedures that are performed under x-ray fluoroscopy. First, a contrast-enhanced angiogram is acquired to diagnose and locate stenoses in the coronary artery tree. Then, a catheter carrying a balloon, or stent, is advanced to the site of the lesion and deployed under image guidance. Accurate placement of the therapeutic catheter is complicated by several factors. Contrast agent toxicity limits the number of times that the arteries can be opacified, such that guidance is performed in a fluoroscopic imaging mode, when the arteries are not visible. The interventional cardiologist must mentally register the diagnostic angiograms that depict the location of the stenosis to the fluoroscopic images that only show the catheter. Cardiac and respiratory motion of the heart can introduce large changes in image position of a given stenosis, requiring the

3 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 3 clinician to continuously estimate the target position. Final placement of the catheter is verified with small puffs of contrast agent. In this paper, we present a method for motion correcting x-ray images of the heart. The algorithm relies on motion information derived from a patient specific 3D+t (three dimensional + time) model of the coronary arteries. Methods for generating these models from diagnostic coronary angiograms acquired on biplane and rotational single plane systems have been described previously [1], [2]. We then perform a cardiac respiratory parametric decomposition of the 3D motion field, which allows us to predict the correction parameters for images acquired at any arbitrary cardiac respiratory phase [3]. Eck et. al. described an image based method for synthesizing angiographic images at any cardiac and respiratory phase [4]. Their warping function is view-angle dependent, and must be recomputed if the imaging C-arm is reoriented. In contrast, our method uses a 3D deforming model of the coronary tree, which makes it possible to prospectively correct images acquired at any imaging angle. A. X-ray projection imaging II. METHODS X-ray projection imaging is a 3D to 2D imaging process that can be described by a 3x4 projection matrix P : P = ( n u)(sid) n 0 u IS 2 ( n 0 v)(sid) n v IS R pa R sa 0... SOD which is the product of a 3x3 matrix representing a perspective projection, and a 3x4 matrix describing the orientation of the imaging system relative to a world coordinate system. n u and n v are the image dimensions in pixels. The 3x3 matrix R pa R sa represents the orientation of the imaging C-arm, as defined by the primary and secondary angles (Fig. 1). IS is the intensifier size, SOD is the source-to-isocenter distance, and SID is the source-to-intensifier distance (Fig. 2). Figure 3 is a diagram of the imaging coordinate system. The imaging process is modeled as a perspective projection along the -z axis. In this mathematical description, the patient is oriented within a fixed camera coordinate system. The primary angle rotation, R pa, is a straightforward (1)

4 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 4 rotation about the +y axis, with a positive angle rotation corresponding to an LAO view: cos(pa) 0 sin(pa) R pa = (2) sin(p A) 0 cos(pa) If PA=0, then the secondary angle rotation R sa is simply a rotation about the +x axis, with a positive angle rotation corresponding to a cranial view: R sa = 0 cos(sa) sin(sa). (3) 0 sin(sa) cos(sa) In most cases, PA 0, and a new axis of rotation must be computed for the secondary rotation. The new axis u is obtained by rotating the x axis by -(PA) around the +y axis. Using the transpose of R PA as the negative rotation, v = R T pa [ 1 0 0] T (4) and u = (u x,u y,u z ) T = v/ v. The secondary angle rotation matrix is then computed using ( ) 0 u z u y R sa = uu T + cos(sa) uut +sin(sa) u z 0 u x. (5) u y u x 0 which represents a 3D rotation about an arbitrary axis u [5]. In homogeneous coordinates, multiplying the 3D point q = (q x,q y,q z, 1) T by the projection matrix P, generates the 2D point r = (r u,r v,r w ) T : r = Pq. (6) The image coordinates of this point are obtained by dividing r by r w. B. Imaging protocol Patients undergoing diagnostic left heart catheterization were recruited to participate in an IRB approved protocol. All images were acquired on a Siemens biplane cardiovascular angiography system at a rate of 30 frames per second. The clinician was not constrained in positioning

5 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 5 the C-arms. The technicians were requested to not move the patient table during the image acquisition. Images of a calibration grid and phantom were acquired with the imaging system in the same configurations (primary and secondary angles, source-intensifier distance) used to acquire patient data. The grid of metal beads (1 cm horizontal and vertical spacing) was used to correct for the geometric distortion introduced by the image intensifier [6]. The plastic rectangular phantom ( cm) containing 18 metal beads was reconstructed and compared to its known geometry in order to optimize the imaging parameters and validate the accuracy of the 3D reconstruction [7]. C. Cardiac and Respiratory Phase A personal computer with digital acquisition system recorded the ECG and an image acquire signal from the Siemens digital acquisition subsystem (Fig. 4a). The cardiac phase (χ) of each image was then calculated so that χ represents the percentage of the cardiac cycle, with the QRS peak at χ = 0. Beat-to-beat variations in heart rate were normalized by calculating the systolic and diastolic intervals using the method of [8]. The systolic interval was then linearly rescaled between [0, 0.42), which corresponds to the systolic interval of a 60 beat/minute heart rate. The diastolic interval was rescaled between [0.42, 1). Thus, each dataset had a cardiac phase 0 χ < 1. Respiratory phase (ρ) was measured from the movement of the diaphragm as seen in the x-ray images. A profile perpendicular to the diaphragm lung interface was manually positioned on a first image (Fig. 4b), and all the images in a dataset were sampled at this location. The displacement of the diaphragm-lung interface over time generates a one dimensional curve (Fig. 4c) that was manually segmented, with the maximum and minimum values corresponding to endexpiration (EE, ρ = 0) and end-inspiration (EI, ρ = ±1) respectively. Intermediate values of ρ were computed as a linear distance between the respiratory extrema. The sign of ρ depends on whether it is an inspiratory (-) or expiratory (+) maneuver. D. Patient Specific Coronary Motion Models A patient specific model of coronary artery motion was used to compute the motion correction parameters. The model was reconstructed from biplane coronary angiograms using a stereo-vision

6 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 6 reconstruction and automatic motion tracking method [1]. The angiogram was acquired with a hand-injected bolus of contrast lasting 4-8 seconds while the patient was freely breathing. First, a static 3D model of the coronary arteries, Υ 0, was reconstructed from one image biplane image pair at end diastole. Using a semi interactive segmentation tool, a user segmented the arterial centerlines in one pair of projection images. This process was constrained by a potential field generated using multi-scale filters which detect rectilinear structures such as vessels [9]. The epipolar constraint was used to define point to point correspondences between the arteries two projections, and a tree of 3D cubic B spline curves was constructed using a dynamic programming algorithm [10]. Υ 0 was represented within a patient-specific coordinate system. The x-axis corresponds to the patient s right-to-left axis, with +x directed to the patient s left. The +y axis points toward the patient s feet, with the +z axis directed toward the posterior. A 0 LAO/RAO, 0 cranial/caudal x-ray view corresponds to a projection along the z axis of the patient s coordinate system. The arterial model Υ 0 served as the initialization for an automatic motion tracking algorithm that recovered the motion of the arteries from the cine-angiogram images. The algorithm computes the transformations M : R 3 R 3 which keep the arteries projection onto the image planes consistent with the temporally changing images [1]. If Υ 0 was reconstructed from image pair t 0, then the transformation M : R 3 R 3 maps any 3D point q 0 Υ 0 to the point s 3D position consistent with image t 0 + t: q t = M(q;t), (7) which we also write as Υ t = M[Υ 0,t] to represent the deformation of the whole arterial tree. Since every image t represents a cardiac and respiratory phase (χ t,ρ t ), M(q;t) M(q;χ t,ρ t ) (8) The deformation field M was recovered from free breathing angiograms that provided a discrete sampling of different cardiac and respiratory phases. In order to perform motion correction of x-ray images, we needed to represent the deformation field for any cardiac and respiratory phase combination. A cardiac respiratory parametric model, M, was fit to the observed data M [3]. The model uses approximating B spline basis functions to separate the motion into a cardiac contraction motion and a respiratory motion: M(q;χ,ρ) M(q;χ,ρ) = M χ (q;χ) + M ρ (q;ρ) (9)

7 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 7 where M χ is the motion due to the cardiac contraction, and M ρ is the respiratory motion of the heart. Fundamentally, splines are smooth piecewise polynomial functions that can be used to interpolate data points. They provide the mechanism that allows us to estimate the deformation field at any cardiac-respiratory phase based on the data recovered from the angiograms. E. Motion correction of X-ray images The acquisition of x-ray images of the heart at cardiac phase χ and respiratory phase ρ is described by the equation r = P M(q;χ,ρ) (10) where P represents the geometric configuration of the x-ray imaging system (Eq. 1), and M is the transformation of points of a coronary tree from some reference state, q Υ 0. r is the 2D image position of the 3D coronary point q. M was reconstructed from one free-breathing coronary angiogram for each patient. The coronary tree Υ 0 was generated from images acquired at (χ 0,ρ 0 ). We expressed the image projection of the coronary tree at the reference time by the set of points {r 0 }: r 0 = P M(q;χ 0,ρ 0 ). (11) At some later time, at cardiac respiratory phase (χ i,ρ i ), the coronary arteries could be found at {r i } in the image planes: r i = P M(q;χ i,ρ i ). (12) The 2D projection of the 3D motion of a point q from its reference position to its position at cardiac respiratory phase (χ i,ρ i ) can be expressed as the vector r0 r i. To motion correct an x-ray image, we first formed the point set {q} by sampling the 3D coronary tree Υ 0 at discrete points separated by 1 mm along the length of the arteries. The set of 2D displacements, { r0 r i }, was then computed for the set {q}. We formulated a 2D image warping function d : R 2 R 2 as a 2-dimensional B-spline function: d(r) = n i 1 i=0 n j 1 j=0 b i (r x )b j (r y )u ij (13) where {u ij = (u x ij,u y ij ) R2 } are the n i n j control points, and {b} are cubic B-spline basis functions defined over a uniform knot vector. The knot vector range spanned the dimensions of the image and a 10% boundary (i.e. [-51, 563] for a 512 pixel image).

8 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 8 The B-spline control points {u} were computed so that Eq. (13) approximated the set of observed displacements { r0 r i } of the projected coronary tree points. The solution was recovered using a linear minimum mean-square estimator with a second derivative smoothing constraint [11], [12]. A bounding box was defined in the image plane around the projection of the coronary tree model. Outside this region, we added zero displacement vectors at regularly spaced intervals of 20 pixels in the horizontal and vertical directions. This was done so that the border regions of the image would remain unaffected by the warping function. After fitting the 2D warping function to the vector field for an image I χ,ρ, we generated a motion compensated image Ĩχ,ρ. The 2D warping function from Eq. (13) was evaluated at integer values corresponding to image pixel coordinates of Ĩχ,ρ i.e. (0, 0), (0, 1), (0, 2),...,(512, 512). The function returned the 2D vector offset corresponding to the pixel s position in I χ,ρ, which was then sampled using bilinear interpolation. F. Performance Evaluation The performance of the motion correction algorithm was measured by tracking landmarks on the coronary tree. A set of landmarks was manually selected in the reference image near (1) the left main (LM) left anterior descending (LAD) bifurcation; (2) the origin of the first obtuse marginal (OM) branch from the left circumflex artery (LCx); and (3) the origin of the first or second diagonal (D) branch off the LAD [13]. After identifying a landmark in the reference image, its position was automatically tracked in the rest of the dataset s images. A pixel template T was generated around the landmark. A cross correlation metric was computed between the k-th landmark s template, T k, and image I χ,ρ : c(x,y) = N/2 N/2 i= N/2 j= N/2 T k (i,j) I χ,ρ (x + i,y + j). (14) The position of the k-th landmark in the image was determined by finding the peak of the cross correlation function. Since the coronary arteries are deforming during a cardiac cycle, this rigid template matching procedure was ineffective for isolated cardiac phases at some landmarks. Mistracked points were identified visually and corrected by manually adjusting the segmentation. For each landmark, we computed a baseline measure of motion by computing the 2D distance between the landmark s position in the reference image, and its position in the other images in

9 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 9 the original angiogram dataset. These values represent the amount of motion observed due to the cardiac contraction and respiratory motion of the heart. The root mean square (RMS) of these values provides a scalar measure of the landmark s variation throughout the image frames. The position of each landmark was similarly tracked in the motion corrected images, and its 2D distance from the landmark s reference position was computed. A perfect motion correction would produce a zero displacement measure for all the images in the dataset. III. RESULTS In five patients, we acquired two coronary angiograms and used one to generate a 3D+t coronary model. The average length of the reconstructed tree was 24±7 cm (range, cm). For each patient, the cardiac respiratory parametric model was fit using an average of 98±12 biplane image pairs. This corresponds to 3.3±0.4 seconds of free breathing data based on a frame rate of 30 frames per second. A. Retrospective Motion Correction The motion correction algorithm was first applied to the same images that were used to create the 3D+t coronary motion model. The algorithm requires that the user specify two parameters of the image-plane warping function: (1) the control point density (n i n j ) and (2) the smoothing parameter. We tested various combinations of these parameters on left coronary angiograms from three patients. Figure 5 shows the displacement plots for the three left coronary landmarks in a biplane dataset. The dashed lines represent the motion of the coronary landmarks in the original images. A periodic motion spanning three to four cardiac cycles is seen. The solid lines show the 2D displacement of each landmark after applying motion correction with a control point density of 6 6 and a smoothing factor of At five of the six points that are shown, there was an order of magnitude reduction in the 2D motion after motion correction. The behavior of the LAD-Diagonal bifurcation in the LAO projection is an artifact of the template matching algorithm. In this patient, contrast material flowed back into the aorta during the bolus injection. When the aortic valve was closed during diastole, the accumulation of contrast material opacified the aortic root, which overlapped the bifurcation and caused the

10 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 10 template matching to fail. Notice that the peaks in the motion corrected trace are out of phase with the peaks (systolic displacement) of the original trace. Figure 6 shows the image results from another patient. Image frame 132 (middle column) is shown next to the reference image (image 81). The reference image was acquired at middiastole, near end-expiration, while frame 132 was acquired at end-diastole, near end-inspiration. The third column of the image shows frame 132 after motion correction (compare to reference image in first column). Table I shows the average 2D RMS displacement for three left coronary landmarks on two x-ray projections (N=6). With control point grid size ranging from 6 6 to and smoothing factors ranging from to 1.0, the average RMS displacements remained between 2-3 pixels. In contrast, without motion correction, the average RMS displacements were 12.8±1.8, 14.3±1.3, and 14.6±1.8 pixels for the three patients. With a control point grid size of 10 10, and a smoothing factor of 0.01, the mean RMS displacements were 2.1±1.7, 2.2±1.4, and 2.2±2.1 pixels. This represents an 84% to 85% reduction in motion in these three cases. B. Prospective Motion Correction Prospective motion correction was studied by motion correcting a second dataset for each patient. That is, the 3D+t cardiac respiratory parametric motion model was generated from one dataset. It was then used to correct another biplane acquisition, acquired with different projection angles, at a later time. A control point grid size of and a smoothing parameter of 0.01 were used. Analysis was performed on 100±10 images per dataset. For each patient, the mean RMS displacement of the landmarks in the original images and in the motion corrected images is plotted in Figure 7. Each bar represents the motion of three coronary landmarks in two projection views (N=6). The mean RMS displacement in the original images was between 11.5 and 13.6 pixels, whereas the motion corrected displacements were between 4.4 and 7.1 pixels. The motion of the landmarks was reduced by 48-63% in these five datasets. The difference between the original and motion corrected measurements was statistically significant for each patient (two-tailed t-test for paired samples, significance level α=0.05).

11 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 11 IV. DISCUSSION In the first set of experiments, we tested the motion correction algorithm in a retrospective mode in which the 3D+t coronary model was generated from the angiogram images that were then corrected. This allowed us to study the motion correction algorithm, without immediately addressing the quality of patient-specific motion models for prospective correction of de novo images. The results showed that the performance of the algorithm remained flat over over a wide range of control point grid sizes and B-spline smoothing factors. The second set of experiments tested the practical prospective use of this algorithm. In the clinic, the patient-specific motion model would be constructed from one angiogram at the start of a procedure and then be used to motion correct angiographic and fluoroscopic images acquired subsequently. Although we performed our motion correction off-line in this study, the methods can be used in an on-line implementation. The only requirements are that the cardiac and respiratory phase be measured in real-time, and that the geometry of the imaging system be registered to the geometry used in building the patient specific coronary model. In the current set of experiments, we requested that the patient table not be moved during the acquisition. This translational motion could also be easily tracked and incorporated into the correction algorithm. The orientation of the imaging system (primary angle, secondary angle) was held constant during an acquisition, but was moved between acquisitions. Newer rotational angiography systems are designed to spin around the patient during a bolus injection. This should not pose a problem for the motion correction algorithm, as long as the orientation parameters can be obtained in real time. However, the rotation will likely interfere with measuring the displacement of the diaphragm. The changing view angle will capture different portions of the diaphragm over time, changing the profile of the diaphragmatic silhouette. The use of an external respiratory signal, such as a pneumatic bellow placed around the chest wall, could be an alternative measure of respiratory phase. The mean RMS displacements obtained with the prospective correction ( pixels) were larger than were measured in the first set of experiments ( pixels). This observation raises the possibility that the cardiac respiratory motion model has limitations. Consider that the cardiac respiratory motion model is generated from 3-4 seconds worth of data that spans a few heartbeats and approximately one respiratory cycle. Additional data could be used to improve the

12 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 12 reliability of the model. However, a typical bolus of contrast agent is 10 cc administered over 4-8 seconds. The bolus interferes with normal blood flow in the coronaries which precludes longer injections. Additional unessential injections are not tolerated due to concerns over radiation exposure and renal toxicity from total contrast dose. The model requires two inputs: cardiac and respiratory phase. Two parameters may be insufficient considering the variability that exists physiologically. The state of the heart during a cardiac contraction is dependent on multiple factors including preload, afterload, and contractility, and may not be adequately indexed by the ECG signal used to calculate cardiac phase. Respiratory phase is measured from diaphragmatic displacement, whereas breathing is also a function of chest wall motion. Finally, the cardiac-respiratory parametric motion model assumes that the cardiac motion and the respiratory motion are independent. In reality, cardiac and respiratory function are interdependent [14], [15]. The quality of the motion correction also depends on the completeness and accuracy of the 3D+t coronary tree model. Arteries that are not included in the 3D coronary model cannot influence the dewarping function. Inspection of the warped images showed that artifacts sometimes developed in areas not supported by coronary arteries (Fig. 8). The artifacts were more pronounced when the warping function was constructed using a large number of control points and small smoothing factor. Solutions to this problem include using a more constrained deformation field, or by reconstructing a more complete coronary tree model. Another source of error could be a 3D-2D misregistration between the model coordinate system and the imaging projection space. This would happen if the table or the patient moves between acquisitions. For the experiments we presented in this paper the arterial tree was visible in both sets of angiograms, the one used for generating the model, and the one subsequently used to test the prospective motion correction algorithm. To reduce the effect of this misregistration, we performed a rigid body 3D-to-2D registration of the reconstructed static coronary tree model to a pair of images from the second acquisition having nearly the same cardiac and respiratory phase as the model. This solution could not be used when motion correcting fluoroscopic images of a guiding catheter, when the arteries are not opacified and therefore not visible. For the general case, we propose the use of external radio-opaque fiducial markers to maintain the registration between the arterial model coordinate system and the imaging projection space.

13 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 13 A. Comparison to Image Based 2D Methods Motion stabilization of angiograms has been studied previously [16], [17], [18], [19]. However, those techniques are image-based methods in which the motion of the arteries is analyzed in the 2D images and subsequently corrected. The paradigm we present in this paper uses information from patient specific 3D+t anatomic models to effect motion correction. The power of the proposed method is in its ability to correct images in which the vessels are not visible. During a coronary intervention, the cardiologist inserts a guidewire catheter through the coronary tree under x-ray guidance. This is done blindly, in the sense that the arteries are not filled with contrast agent, and are therefore not visible. Our model based method can be used to stabilize the physiologic motion of the catheter in fluoroscopic images to assist the cardiologist in reaching a predetermined target stenosis location. An additional advantage of our method over 2D image based methods is that it is view angle independent. The anatomic models are initially generated from angiograms acquired with the x-ray C-arms in some orientation. Since the anatomic models are 3D objects, they can be rotated and translated to match a re orientation of the x-ray C-arms. Thus, the same anatomic model can be used to motion correct images acquired from different projection angles (Section III-B). B. Computation Time The computational complexity of the prospective motion correction approach we describe was studied in two parts. 1) Preparation Phase: First, there is the preparation phase during which a patient specific coronary motion model is generated. In this paper we propose generating this model from coronary angiograms acquired at the beginning of a cardiac catheterization procedure. This process should ideally be completed within minutes, corresponding to the amount of time between the initial diagnostic angiographic portion of the catheterization, and the subsequent intervention to deliver therapy. We relied on a semi-automatic technique for constructing an initial 3D model of the coronary arteries, which required about 20 minutes of user interaction [10]. Tracking the 3D motion of a coronary tree model through a biplane x-ray acquisition took 16±22 minutes per frame for the five datasets we used. The actual speed is proportional to the number of arteries being tracked, the spatial resolution of the model and deformation field, and the optimization termination criteria.

14 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 14 Then, fitting the cardiac respiratory parametric model to the motion field added an additional 74±21 minutes of processing time for each dataset. Thus, building a patient specific coronary motion model from a 4 second acquisition acquired at 30 frames per second, took more than 30 hours (20 min + 16 min/frame ( 4 30 frames) + 74 min). The tracking and modeling algorithms are currently implemented in MATLAB and little effort was made to optimize their performance. Reimplementation in a compiled language with better memory management should significantly speed up the algorithms. In addition, a simple gradient descent technique with stringent termination criteria was used to optimize the cost functions for motion tracking [1]. Conjugate gradient methods with variable step size should be studied for improving the performance of the algorithm. Finally, the paradigm we propose for prospective motion correction does not require that the 3D+t anatomic models be generated with the method we used. For example, Blondel et al. described a method for automatic reconstruction of coronary models from a rotational x-ray system which would eliminate the need to manually reconstruct the initial coronary tree [20]. They also described a 4D motion tracking method that may converge faster due to additional constraints on the temporal smoothness of the recovered motion field [2]. 2) Online Motion Correction: The second part of our prospective motion correction technique involves (1) estimating the patient s physiologic state, (2) predicting the current arterial state using the precomputed motion model, (3) computing the image warping function that effects the motion correction, and (4) interpolating a new motion corrected image. These series of events have to be completed in real time for each acquired x-ray image so that it may be used on-line during an intervention. The motion correction algorithms from Section II-E were implemented in MATLAB and were run on a Linux system with an Intel Xeon TM 2.8 GHz processor. Some code fragments were implemented in C MEX functions for faster performance. The following computation times were measured in one of the patient data sets. Projecting the 3D coronary tree displacement into the 2D imaging plane (Eqs. [11] and [12]), and fitting the B-spline image warping function (Eq. [13]) was accomplished in 52.7±0.5 milliseconds. Evaluating the warping function for every pixel in a image was done in 429.5±1.9 milliseconds. Finally, generating the motion corrected image through bilinear interpolation of the original image was accomplished in 327.2±1.9 milliseconds.

15 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 15 Real time measurement of physiologic parameters was not addressed in our work. However, work with navigator echos for respiratory gated MR imaging has shown that it is possible to measure diaphragmatic displacement from a one-dimensional grayscale signal similar to that shown in Figure 4(b) in less than 22 milliseconds [21]. Taken together, the steps required for online motion correction currently take approximately one second to execute. When implemented using pipeline processing, this time can be regarded as the lag, or delay, introduced by the motion correction algorithm. The use of hardware processing, instead of the current software implementation, will reduce this computation time to tens or hundreds of milliseconds. A real-time processing delay of this order makes the motion correction algorithm clinically practical. V. CONCLUSION We developed a method for motion correcting x-ray images of the heart. Correction factors are computed using a patient specific 3D+t coronary motion model that is constructed from a biplane coronary angiogram acquired during free breathing. This 3D model allows the correction to be subsequently computed for any orientation of the imaging C-arm, and for images acquired at any cardiac and respiratory phase combination. Prospective motion correction reduced the in-plane motion of three landmarks on the left coronary tree by 48-63% in five patients. REFERENCES [1] G. Shechter, F. Devernay, E. Coste-Manière, A. Quyyumi, and E. McVeigh, Three dimensional motion tracking of coronary arteries in biplane cineangiograms, IEEE Trans. Med. Imaging, vol. 22, no. 4, pp , Apr [2] C. Blondel, G. Malandain, R. Vaillant, and N. Ayache, 4D deformation field of coronary arteries from monoplane rotational X ray angiography, in CARS Computer Assisted Radiology and Surgery. Proceedings of the 17th International Congress and Exhibition, June 2003, pp [3] G. Shechter, C. Ozturk, J. Resar, and E. McVeigh, Respiratory motion of the heart from free breathing coronary angiograms, IEEE Trans. Med. Imaging, vol. 23, no. 8, pp , Aug [4] K. Eck, I. Wachter, and J. Bredno, Synthesis of angiographic images using iterative approximation, in Proc. SPIE Med. Imaging, J. Fitzpatrick and M. Sonka, Eds., vol. 5370, 2004, pp [5] O. A. R. Board, M. Woo, J. Neider, T. Davis, and D. Shreiner, OpenGL Programming Guide, 3rd ed. Addison Wesley, 1999, pp [6] E. Gronenschild, The accuracy and reproducibility of a global method to correct for geometric image distortion in the x-ray imaging chain, Med. Phys., vol. 24, pp , [7] G. Shechter, F. Devernay, E. Coste-Manière, and E. McVeigh, Temporal tracking of 3D coronary arteries in projection angiograms, in Proc. SPIE Med. Imaging, M. Sonka and J. Fitzpatrick, Eds., vol. 4684, May 2002, pp

16 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 16 [8] A. Weissler, W. Harris, and C. Schoenfeld, Systolic time intervals in heart failure in man, Circulation, vol. 37, pp , [9] A. Frangi, W. Niessen, R. Hoogeveen, T. van Walsum, and M. Viergever, Model based quantitation of 3-D magnetic resonance angiographic images, IEEE Trans. Med. Imaging, vol. 18, no. 10, pp , Oct [10] F. Mourgues, F. Devernay, G. Malandain, and E. Coste-Manière, 3D+t modeling of coronary artery tree from standard non simultaneous angiograms, in Proc. MICCAI, vol. 2208, Oct. 2001, pp [11] J. Declerck, T. Denney, C. Ozturk, W. O Dell, and E. McVeigh, Left ventricular motion reconstruction from planar tagged MR images: a comparison, Phys. Med. Biol., vol. 45, pp , [12] T. Denney and J. Prince, Reconstruction of 3-D left ventricular motion from planar tagged cardiac MR images: An estimation theoretic approach, IEEE Trans. Med. Imaging, vol. 14, no. 5, pp , Dec [13] W. Austen, J. Edwards, R. Frye, G. Gensini, V. Gott, L. Griffith, D. McGoon, M. Murphy, and B. Roe, A reporting system on patients evaluated for coronary artery disease, Circ., vol. 51, no. 4 Suppl, pp. 5 40, Apr [14] W. Summer, S. Permutt, K. Sagawa, A. Shoukas, and B. Bromberger-Barnea, Effects of spontaneous respiration on canine left ventricular function, Circ. Res., vol. 45, pp , [15] S. Scharf, R. Brown, N. Saunders, and L. Green, Effects of normal and loaded spontaneous inspiration on cardiovascular function, J. Appl. Physiol., vol. 47, no. 3, pp , Sept [16] N. Eigler, M. Eckstein, K. Mahrer, and J. Whiting, Improving detection of coronary morphological features from digital angiograms. Effect of stenosis-stabilized display, Circ., vol. 89, no. 6, pp , June [17] S. Roehm, J. Smith, and R. Mangalik, Digital x-ray imaging system with automatic tracking, US Patent 5,293,574, Mar. 8, [18] M.-P. Dubuisson-Jolly, C. Liang, and A. Gupta, Optimal polyline tracking for artery motion compensation in coronary angiography, in Sixth International Conference on Computer Vision (ICCV), 1998, pp [19] E. Meijering, W. Niessen, and M. Viergever, Retrospective motion correction in digital subtraction angiography, IEEE Trans. Med. Imaging, vol. 19, no. 1, pp. 2 21, Jan [20] C. Blondel, R. Vaillant, F. Devernay, G. Malandain, and N. Ayache, Automatic trinocular 3D reconstruction of coronary artery centerlines from rotational X-ray angiography, in Computer Assisted Radiology and Surgery 2002 Proceedings. Paris: Springer Publishers, Heidelberg, June 2002, pp [21] Y. P. Du, M. Saranathan, and T. Foo, An accurate, robust, and computationally efficient navigator algorithm for measuring diaphragm positions, J. Cardiovasc. Magn. Reson., vol. 6, no. 2, pp , Apr

17 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 17 RAO LAO Cranial Caudal Primary Angle Secondary Angle Fig. 1. The primary angle (PA) and secondary angle (SA) define the geometric orientation of the imaging system with respect to the patient. Zero degree primary and secondary angles correspond to an anterior-posterior projection. The primary angle diagram is viewed from the patient s feet.

18 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 18 Image Intensifier IS Isocenter SOD SID X-ray source Fig. 2. Diagram of an x-ray imaging system. An x-ray source and image intensifier are arranged on a C-arm gantry. The isocenter is defined as the point of rotation of the imaging arm. The SOD, or source to object distance, is measured to the isocenter, where the imaged object is typically placed. SID is the source to image intensifier distance, and IS is the intensifier size. The figure is adapted from the Series 9800 Mobile C-Arm Operator s Guide (OEC Medical Systems Inc).

19 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 19 Y Z X Image Intensifier X-ray Source 90 LAO c Arai/Shechter Fig. 3. The coordinate system of the imaging C-arm. For the mathematical description of the imaging process, different projections of the patient are obtained by rotating the patient in the camera s fixed frame of reference. (Drawing courtesy of Yaniv Arai.)

20 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 20 (a) ECG X ray Trigger χ=0.5 χ=0 χ= Time (s) (b) (c) ρ = 0 End Expiration ρ=0 ρ =1 ρ = 1 End Inspiration ρ = 1 End Inspiration Image Number Seconds Fig. 4. (a) Cardiac phase (χ) was measured by digitizing an ECG in synchrony with the image acquisition. (b) Respiratory phase (ρ) was measured by tracking the displacement of the diaphragm along a profile in the angiogram images. (c) The displacement of the lung diaphragm interface is shown as a function of image number [3].

21 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) RAO, 0 Cranial 54 LAO, 2 Caudal 2D distance from first frame (pixels) Original Motion Corrected LM bifurcation 2D distance from first frame (pixels) Original Motion Corrected LM bifurcation Image Frame Number Image Frame Number 2D distance from first frame (pixels) Original Motion Corrected LCx OM bifurcation 2D distance from first frame (pixels) Original Motion Corrected LCx OM bifurcation Image Frame Number Image Frame Number 2D distance from first frame (pixels) Original Motion Corrected LAD Diag bifurcation 2D distance from first frame (pixels) Original Motion Corrected LAD Diag bifurcation Image Frame Number Image Frame Number Fig. 5. Two dimensional displacement of three left coronary tree landmarks in two projection angiograms. The dashed lines represent the displacement in the original images. The solid line is the displacement of the point after motion correcting the x-ray projection images. The behavior of the LAD-Diagonal bifurcation in the LAO projection is an artifact of the template matching algorithm used to track the landmarks. Contrast accumulation in the aortic root during diastole interfered with localizing the nearby bifurcation.

22 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 22 Fig. 6. Motion correction of an x-ray angiogram acquired during free breathing. The images from the middle column were corrected, with the images from the first column serving as the reference state. Compare the results in the third column, to the reference images from the first column.

23 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 23 TABLE I THE EFFECT OF CONTROL POINT DENSITY AND B-SPLINE SMOOTHING FACTOR ON THE QUALITY OF MOTION CORRECTION OF THE LEFT CORONARY TREE. THE 2D DISPLACEMENT OF A CORONARY LANDMARK WAS MEASURED FROM A REFERENCE IMAGE. A SCALAR MEASURE OF THE LANDMARK S MOTION IN TIME THROUGH ONE DATASET WAS COMPUTED AS THE RMS OF THE DISPLACEMENTS. THE REPORTED VALUES (IN PIXELS) ARE THE MEAN AND STANDARD DEVIATION OF THE RMS DISTANCES FOR THREE LANDMARKS IN TWO PROJECTIONS (N=6). Patient I: Without motion correction, the RMS displacement was 14.6±1.8 pixels. Number of Smoothing Factor Control Points ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± 1.7 Patient II: Without motion correction, the RMS displacement was 12.8±1.8 pixels. Number of Smoothing Factor Control Points ± ± ± ± ± ± ± ± ± ± ± ± 1.5 Patient III: Without motion correction, the RMS displacement was 14.3±1.3 pixels. Number of Smoothing Factor Control Points ± ± ± ± ± ± ± ± ± ± ± ± 1.6

24 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) RMS 2D Displacement Original Images Motion Corrected Images * P<0.004 * P< * P< * P< * P< I II III IV V Patient Number Fig. 7. Prospective motion correction of left coronary angiograms in five patients. The 3D+t coronary model was reconstructed from a first angiogram, and used to motion correct another angiogram acquired at a later time. Mean RMS 2D displacement was measured in the original images and in the motion corrected images, at three coronary landmarks in two projection views (N=6). In three of the five patients, the displacement in the motion corrected images was significantly less than in the original images (paired t-test, α = 0.05).

25 IEEE TRANSACTION ON MEDICAL IMAGING (PREPRINT) 25 Fig. 8. Artifacts can appear in the motion corrected images when using a dense 2D B-spline control point grid (n i = n j = 14), and a weak smoothing parameter (0.001).

Model-Based Respiratory Motion Compensation for Image-Guided Cardiac Interventions

Model-Based Respiratory Motion Compensation for Image-Guided Cardiac Interventions Model-Based Respiratory Motion Compensation for Image-Guided Cardiac Interventions February 8 Matthias Schneider Pattern Recognition Lab Friedrich-Alexander-University Erlangen-Nuremberg Imaging and Visualization

More information

Respiratory Motion Estimation using a 3D Diaphragm Model

Respiratory Motion Estimation using a 3D Diaphragm Model Respiratory Motion Estimation using a 3D Diaphragm Model Marco Bögel 1,2, Christian Riess 1,2, Andreas Maier 1, Joachim Hornegger 1, Rebecca Fahrig 2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg 2

More information

Intensity-Based 3D-Reconstruction of Non-Rigid Moving Stenosis From Many Angiographies erschienen in

Intensity-Based 3D-Reconstruction of Non-Rigid Moving Stenosis From Many Angiographies erschienen in Intensit-Based 3D-Reconstruction of Non-Rigid Moving Stenosis From Man Angiographies erschienen in @inproceedings{bouattour24i3o, pages = {45-49}, crossref = {proceedings24bfd}, title = {Intensit-Based

More information

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy Chenyang Xu 1, Siemens Corporate Research, Inc., Princeton, NJ, USA Xiaolei Huang,

More information

4D Motion Modeling of the Coronary Arteries from CT Images for Robotic Assisted Minimally Invasive Surgery

4D Motion Modeling of the Coronary Arteries from CT Images for Robotic Assisted Minimally Invasive Surgery 4D Motion Modeling of the Coronary Arteries from CT Images for Robotic Assisted Minimally Invasive Surgery Dong Ping Zhang 1, Eddie Edwards 1,2, Lin Mei 1,2, Daniel Rueckert 1 1 Department of Computing,

More information

Towards Projector-based Visualization for Computer-assisted CABG at the Open Heart

Towards Projector-based Visualization for Computer-assisted CABG at the Open Heart Towards Projector-based Visualization for Computer-assisted CABG at the Open Heart Christine Hartung 1, Claudia Gnahm 1, Stefan Sailer 1, Marcel Schenderlein 1, Reinhard Friedl 2, Martin Hoffmann 3, Klaus

More information

Simulation environment of X-ray rotational angiography using 3D+t coronary tree model

Simulation environment of X-ray rotational angiography using 3D+t coronary tree model Simulation environment of X-ray rotational angiography using 3D+t coronary tree model Guanyu Yang, Yining Hu, Xi Huang, Huazhong Shu, Christine Toumoulin To cite this version: Guanyu Yang, Yining Hu, Xi

More information

3D Guide Wire Navigation from Single Plane Fluoroscopic Images in Abdominal Catheterizations

3D Guide Wire Navigation from Single Plane Fluoroscopic Images in Abdominal Catheterizations 3D Guide Wire Navigation from Single Plane Fluoroscopic Images in Abdominal Catheterizations Martin Groher 2, Frederik Bender 1, Ali Khamene 3, Wolfgang Wein 3, Tim Hauke Heibel 2, Nassir Navab 2 1 Siemens

More information

Motion Compensation from Short-Scan Data in Cardiac CT

Motion Compensation from Short-Scan Data in Cardiac CT Motion Compensation from Short-Scan Data in Cardiac CT Juliane Hahn 1,2, Thomas Allmendinger 1, Herbert Bruder 1, and Marc Kachelrieß 2 1 Siemens Healthcare GmbH, Forchheim, Germany 2 German Cancer Research

More information

Coronary Artery Tracking from Dynamic Cardiac CT Sequences

Coronary Artery Tracking from Dynamic Cardiac CT Sequences Coronary Artery Tracking from Dynamic Cardiac CT Sequences Dong Ping Zhang a, Ose Pedro a, Kensaku Mori b, Philip Edwards a,c and Daniel Rueckert a a Department of Computing, Imperial College London, UK

More information

Automatic Extraction of quasi-synchronous Views from Rotational Angiographic Sequence without ECG-Data

Automatic Extraction of quasi-synchronous Views from Rotational Angiographic Sequence without ECG-Data Automatic Extraction of quasi-synchronous Views from Rotational Angiographic Sequence without ECG-Data Sahla Bouattour and Dietrich Paulus Institute of Computational Visualistics, University of Koblenz-Landau,

More information

Respiratory Motion Compensation for C-arm CT Liver Imaging

Respiratory Motion Compensation for C-arm CT Liver Imaging Respiratory Motion Compensation for C-arm CT Liver Imaging Aline Sindel 1, Marco Bögel 1,2, Andreas Maier 1,2, Rebecca Fahrig 3, Joachim Hornegger 1,2, Arnd Dörfler 4 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg

More information

Estimating 3D Respiratory Motion from Orbiting Views

Estimating 3D Respiratory Motion from Orbiting Views Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

A Comprehensive Method for Geometrically Correct 3-D Reconstruction of Coronary Arteries by Fusion of Intravascular Ultrasound and Biplane Angiography

A Comprehensive Method for Geometrically Correct 3-D Reconstruction of Coronary Arteries by Fusion of Intravascular Ultrasound and Biplane Angiography Computer-Aided Diagnosis in Medical Imaging, September 20-23, 1998, Chicago IL Elsevier ICS 1182 A Comprehensive Method for Geometrically Correct 3-D Reconstruction of Coronary Arteries by Fusion of Intravascular

More information

Guide wire tracking in interventional radiology

Guide wire tracking in interventional radiology Guide wire tracking in interventional radiology S.A.M. Baert,W.J. Niessen, E.H.W. Meijering, A.F. Frangi, M.A. Viergever Image Sciences Institute, University Medical Center Utrecht, rm E 01.334, P.O.Box

More information

4-D Tomographic Representation of Coronary Arteries from One Rotational X-Ray Sequence

4-D Tomographic Representation of Coronary Arteries from One Rotational X-Ray Sequence 4-D Tomographic Representation of Coronary Arteries from One Rotational X-Ray Sequence Christophe Blondel, Grégoire Malandain, Régis Vaillant, Frédéric Devernay, Ève Coste-Manière, Nicholas Ayache To cite

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION

FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION Philips J. Res. 51 (1998) 197-201 FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION This special issue of Philips Journalof Research includes a number of papers presented at a Philips

More information

Image registration for motion estimation in cardiac CT

Image registration for motion estimation in cardiac CT Image registration for motion estimation in cardiac CT Bibo Shi a, Gene Katsevich b, Be-Shan Chiang c, Alexander Katsevich d, and Alexander Zamyatin c a School of Elec. Engi. and Comp. Sci, Ohio University,

More information

3D+t Modeling of Coronary Artery Tree from Standard Non Simultaneous Angiograms

3D+t Modeling of Coronary Artery Tree from Standard Non Simultaneous Angiograms 3D+t Modeling of Coronary Artery Tree from Standard Non Simultaneous Angiograms Fabien Mourgues, Frédéric Devernay, Grégoire Malandain, Ève Coste-Manière To cite this version: Fabien Mourgues, Frédéric

More information

A Non-Linear Image Registration Scheme for Real-Time Liver Ultrasound Tracking using Normalized Gradient Fields

A Non-Linear Image Registration Scheme for Real-Time Liver Ultrasound Tracking using Normalized Gradient Fields A Non-Linear Image Registration Scheme for Real-Time Liver Ultrasound Tracking using Normalized Gradient Fields Lars König, Till Kipshagen and Jan Rühaak Fraunhofer MEVIS Project Group Image Registration,

More information

Basic fmri Design and Analysis. Preprocessing

Basic fmri Design and Analysis. Preprocessing Basic fmri Design and Analysis Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial filtering

More information

Nonrigid Motion Compensation of Free Breathing Acquired Myocardial Perfusion Data

Nonrigid Motion Compensation of Free Breathing Acquired Myocardial Perfusion Data Nonrigid Motion Compensation of Free Breathing Acquired Myocardial Perfusion Data Gert Wollny 1, Peter Kellman 2, Andrés Santos 1,3, María-Jesus Ledesma 1,3 1 Biomedical Imaging Technologies, Department

More information

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

Functional MRI in Clinical Research and Practice Preprocessing

Functional MRI in Clinical Research and Practice Preprocessing Functional MRI in Clinical Research and Practice Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization

More information

SIGMI Meeting ~Image Fusion~ Computer Graphics and Visualization Lab Image System Lab

SIGMI Meeting ~Image Fusion~ Computer Graphics and Visualization Lab Image System Lab SIGMI Meeting ~Image Fusion~ Computer Graphics and Visualization Lab Image System Lab Introduction Medical Imaging and Application CGV 3D Organ Modeling Model-based Simulation Model-based Quantification

More information

Volumetric Analysis of the Heart from Tagged-MRI. Introduction & Background

Volumetric Analysis of the Heart from Tagged-MRI. Introduction & Background Volumetric Analysis of the Heart from Tagged-MRI Dimitris Metaxas Center for Computational Biomedicine, Imaging and Modeling (CBIM) Rutgers University, New Brunswick, NJ Collaboration with Dr. Leon Axel,

More information

Image Guidance and Beam Level Imaging in Digital Linacs

Image Guidance and Beam Level Imaging in Digital Linacs Image Guidance and Beam Level Imaging in Digital Linacs Ruijiang Li, Ph.D. Department of Radiation Oncology Stanford University School of Medicine 2014 AAPM Therapy Educational Course Disclosure Research

More information

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most

More information

A preliminary study on adaptive field-of-view tracking in peripheral digital subtraction angiography

A preliminary study on adaptive field-of-view tracking in peripheral digital subtraction angiography Journal of X-Ray Science and Technology 11 (2003) 149 159 149 IOS Press A preliminary study on adaptive field-of-view tracking in peripheral digital subtraction angiography James R. Bennett a,b,er-weibai

More information

Constrained Reconstruction of Sparse Cardiac MR DTI Data

Constrained Reconstruction of Sparse Cardiac MR DTI Data Constrained Reconstruction of Sparse Cardiac MR DTI Data Ganesh Adluru 1,3, Edward Hsu, and Edward V.R. DiBella,3 1 Electrical and Computer Engineering department, 50 S. Central Campus Dr., MEB, University

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 116: 3D X-Ray Storage SOP Class

Digital Imaging and Communications in Medicine (DICOM) Supplement 116: 3D X-Ray Storage SOP Class 2 4 6 Digital Imaging and Communications in Medicine (DICOM) 8 Supplement 116: 3D X-Ray Storage SOP Class 10 12 14 16 18 20 22 Prepared by: 24 DICOM Standards Committee, Working Groups 02, Projection Imaging

More information

3D/2D Model-to-Image Registration Applied to TIPS Surgery

3D/2D Model-to-Image Registration Applied to TIPS Surgery 3D/2D Model-to-Image Registration Applied to TIPS Surgery Julien Jomier 1, Elizabeth Bullitt 2, Mark Van Horn 2, Chetna Pathak 2, and Stephen R. Aylward 1 1 Kitware Inc. 28 Corporate Drive Clifton Park,

More information

Computer-Aided Diagnosis in Abdominal and Cardiac Radiology Using Neural Networks

Computer-Aided Diagnosis in Abdominal and Cardiac Radiology Using Neural Networks Computer-Aided Diagnosis in Abdominal and Cardiac Radiology Using Neural Networks Du-Yih Tsai, Masaru Sekiya and Yongbum Lee Department of Radiological Technology, School of Health Sciences, Faculty of

More information

A POINT-EVENT DETECTION ALGORITHM FOR THE ANALYSIS OF CONTRAST BOLUS IN FLUOROSCOPIC IMAGES OF THE CORONARY ARTERIES

A POINT-EVENT DETECTION ALGORITHM FOR THE ANALYSIS OF CONTRAST BOLUS IN FLUOROSCOPIC IMAGES OF THE CORONARY ARTERIES A POINT-EVENT DETECTION ALGORITHM FOR THE ANALYSIS OF CONTRAST BOLUS IN FLUOROSCOPIC IMAGES OF THE CORONARY ARTERIES Alexandru Condurache and Alfred Mertins Institute for Signal Processing, University

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing Functional Connectivity Preprocessing Geometric distortion Head motion Geometric distortion Head motion EPI Data Are Acquired Serially EPI Data Are Acquired Serially descending 1 EPI Data Are Acquired

More information

Brilliance CT Big Bore.

Brilliance CT Big Bore. 1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.

More information

SPM8 for Basic and Clinical Investigators. Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

Automatic 3D Reconstruction of Coronary Artery Centerlines from Monoplane X-ray Angiogram Images

Automatic 3D Reconstruction of Coronary Artery Centerlines from Monoplane X-ray Angiogram Images Automatic 3D Reconstruction of Coronary Artery Centerlines from Monoplane X-ray Angiogram Images Ali Zifan 1, Panos Liatsis 1, Panagiotis Kantartzis 1, Manolis Gavaises 1, Nicos Karcanias 1, Demosthenes

More information

Intraoperative Prostate Tracking with Slice-to-Volume Registration in MR

Intraoperative Prostate Tracking with Slice-to-Volume Registration in MR Intraoperative Prostate Tracking with Slice-to-Volume Registration in MR Sean Gill a, Purang Abolmaesumi a,b, Siddharth Vikal a, Parvin Mousavi a and Gabor Fichtinger a,b,* (a) School of Computing, Queen

More information

Three-Dimensional Motion Tracking of Coronary Arteries in Biplane Cineangiogram

Three-Dimensional Motion Tracking of Coronary Arteries in Biplane Cineangiogram Three-Dimensional Motion Tracking of Coronary Arteries in Biplane Cineangiogram Guy Shechter, Frédéric Devernay, Ève Coste-Manière, Arshed Quyyumi, Elliott Mcveigh To cite this version: Guy Shechter, Frédéric

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 167: X-Ray 3D Angiographic IOD Informative Annex

Digital Imaging and Communications in Medicine (DICOM) Supplement 167: X-Ray 3D Angiographic IOD Informative Annex 5 Digital Imaging and Communications in Medicine (DICOM) Supplement 167: X-Ray 3D Angiographic IOD Informative Annex 10 15 Prepared by: DICOM Standards Committee, Working Group 2, Projection Radiography

More information

Overview of Proposed TG-132 Recommendations

Overview of Proposed TG-132 Recommendations Overview of Proposed TG-132 Recommendations Kristy K Brock, Ph.D., DABR Associate Professor Department of Radiation Oncology, University of Michigan Chair, AAPM TG 132: Image Registration and Fusion Conflict

More information

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ORAL DEFENSE 8 th of September 2017 Charlotte MARTIN Supervisor: Pr. MP REVEL M2 Bio Medical

More information

Image-Based Device Tracking for the Co-registration of Angiography and Intravascular Ultrasound Images

Image-Based Device Tracking for the Co-registration of Angiography and Intravascular Ultrasound Images Image-Based Device Tracking for the Co-registration of Angiography and Intravascular Ultrasound Images Peng Wang 1, Terrence Chen 1, Olivier Ecabert 2, Simone Prummer 2, Martin Ostermeier 2, and Dorin

More information

Isogeometric Analysis of Fluid-Structure Interaction

Isogeometric Analysis of Fluid-Structure Interaction Isogeometric Analysis of Fluid-Structure Interaction Y. Bazilevs, V.M. Calo, T.J.R. Hughes Institute for Computational Engineering and Sciences, The University of Texas at Austin, USA e-mail: {bazily,victor,hughes}@ices.utexas.edu

More information

QAngio XA 7.3. Quick Start Manual. July 31, v6.0

QAngio XA 7.3. Quick Start Manual. July 31, v6.0 QAngio XA 7.3 Quick Start Manual July 31, 2018 9.04.250.73.6 v6.0 Medis medical imaging systems bv Schuttersveld 9, 2316 XG Leiden, the Netherlands http://www.medis.nl Medis medical imaging systems bv

More information

Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data

Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data Christopher M Rank 1, Thorsten Heußer 1, Andreas Wetscherek 1, and Marc Kachelrieß 1 1 German Cancer

More information

Automatic Extraction of 3D Dynamic Left Ventricle Model from 2D Rotational Angiocardiogram

Automatic Extraction of 3D Dynamic Left Ventricle Model from 2D Rotational Angiocardiogram Automatic Extraction of 3D Dynamic Left Ventricle Model from 2D Rotational Angiocardiogram Mingqing Chen 1, Yefeng Zheng 1, Kerstin Mueller 2,3, Christopher Rohkohl 2, Guenter Lauritsch 2,JanBoese 2, Gareth

More information

Finite Element Simulation of Moving Targets in Radio Therapy

Finite Element Simulation of Moving Targets in Radio Therapy Finite Element Simulation of Moving Targets in Radio Therapy Pan Li, Gregor Remmert, Jürgen Biederer, Rolf Bendl Medical Physics, German Cancer Research Center, 69120 Heidelberg Email: pan.li@dkfz.de Abstract.

More information

Depth-Layer-Based Patient Motion Compensation for the Overlay of 3D Volumes onto X-Ray Sequences

Depth-Layer-Based Patient Motion Compensation for the Overlay of 3D Volumes onto X-Ray Sequences Depth-Layer-Based Patient Motion Compensation for the Overlay of 3D Volumes onto X-Ray Sequences Jian Wang 1,2, Anja Borsdorf 2, Joachim Hornegger 1,3 1 Pattern Recognition Lab, Friedrich-Alexander-Universität

More information

This Time. fmri Data analysis

This Time. fmri Data analysis This Time Reslice example Spatial Normalization Noise in fmri Methods for estimating and correcting for physiologic noise SPM Example Spatial Normalization: Remind ourselves what a typical functional image

More information

Gradient-Based Differential Approach for Patient Motion Compensation in 2D/3D Overlay

Gradient-Based Differential Approach for Patient Motion Compensation in 2D/3D Overlay Gradient-Based Differential Approach for Patient Motion Compensation in 2D/3D Overlay Jian Wang, Anja Borsdorf, Benno Heigl, Thomas Köhler, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-University

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

Family of Diagnostic and Therapeutic Catheters

Family of Diagnostic and Therapeutic Catheters Family of Diagnostic and Therapeutic Catheters Full line of Percutaneous Transluminal Angioplasty (PTA), Percutaneous Transluminal Valvuloplasty (PTV) and Aortic Valvuloplastly (BAV) Catheters PTS and

More information

Lecture 6: Medical imaging and image-guided interventions

Lecture 6: Medical imaging and image-guided interventions ME 328: Medical Robotics Winter 2019 Lecture 6: Medical imaging and image-guided interventions Allison Okamura Stanford University Updates Assignment 3 Due this Thursday, Jan. 31 Note that this assignment

More information

Reconstruction of Coronary Arteries from a Single Rotational X-Ray Projection Sequence

Reconstruction of Coronary Arteries from a Single Rotational X-Ray Projection Sequence Reconstruction of Coronary Arteries from a Single Rotational X-Ray Projection Sequence Christophe Blondel, Grégoire Malandain, Régis Vaillant and Nicholas Ayache Abstract Cardiovascular diseases remain

More information

Reconstruction of Time-Varying 3-D Left-Ventricular Shape From Multiview X-Ray Cineangiocardiograms

Reconstruction of Time-Varying 3-D Left-Ventricular Shape From Multiview X-Ray Cineangiocardiograms IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 21, NO. 7, JULY 2002 773 Reconstruction of Time-Varying 3-D Left-Ventricular Shape From Multiview X-Ray Cineangiocardiograms Masamitsu Moriyama, Yoshinobu Sato*,

More information

MR Advance Techniques. Vascular Imaging. Class III

MR Advance Techniques. Vascular Imaging. Class III MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 31, NO. 8, AUGUST David Rivest-Hénault*, Hari Sundar, and Mohamed Cheriet

IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 31, NO. 8, AUGUST David Rivest-Hénault*, Hari Sundar, and Mohamed Cheriet IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 31, NO. 8, AUGUST 2012 1557 Nonrigid 2D/3D Registration of Coronary Artery Models With Live Fluoroscopy for Guidance of Cardiac Interventions David Rivest-Hénault*,

More information

Compressed Sensing for Rapid MR Imaging

Compressed Sensing for Rapid MR Imaging Compressed Sensing for Rapid Imaging Michael Lustig1, Juan Santos1, David Donoho2 and John Pauly1 1 Electrical Engineering Department, Stanford University 2 Statistics Department, Stanford University rapid

More information

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging 1 CS 9 Final Project Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging Feiyu Chen Department of Electrical Engineering ABSTRACT Subject motion is a significant

More information

Texture Mapping using Surface Flattening via Multi-Dimensional Scaling

Texture Mapping using Surface Flattening via Multi-Dimensional Scaling Texture Mapping using Surface Flattening via Multi-Dimensional Scaling Gil Zigelman Ron Kimmel Department of Computer Science, Technion, Haifa 32000, Israel and Nahum Kiryati Department of Electrical Engineering

More information

Assessing Accuracy Factors in Deformable 2D/3D Medical Image Registration Using a Statistical Pelvis Model

Assessing Accuracy Factors in Deformable 2D/3D Medical Image Registration Using a Statistical Pelvis Model Assessing Accuracy Factors in Deformable 2D/3D Medical Image Registration Using a Statistical Pelvis Model Jianhua Yao National Institute of Health Bethesda, MD USA jyao@cc.nih.gov Russell Taylor The Johns

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

GENERAL AUTOMATED FLAW DETECTION SCHEME FOR NDE X-RAY IMAGES

GENERAL AUTOMATED FLAW DETECTION SCHEME FOR NDE X-RAY IMAGES GENERAL AUTOMATED FLAW DETECTION SCHEME FOR NDE X-RAY IMAGES Karl W. Ulmer and John P. Basart Center for Nondestructive Evaluation Department of Electrical and Computer Engineering Iowa State University

More information

Total Variation Regularization Method for 3D Rotational Coronary Angiography

Total Variation Regularization Method for 3D Rotational Coronary Angiography Total Variation Regularization Method for 3D Rotational Coronary Angiography Haibo Wu 1,2, Christopher Rohkohl 1,3, Joachim Hornegger 1,2 1 Pattern Recognition Lab (LME), Department of Computer Science,

More information

2D Vessel Segmentation Using Local Adaptive Contrast Enhancement

2D Vessel Segmentation Using Local Adaptive Contrast Enhancement 2D Vessel Segmentation Using Local Adaptive Contrast Enhancement Dominik Schuldhaus 1,2, Martin Spiegel 1,2,3,4, Thomas Redel 3, Maria Polyanskaya 1,3, Tobias Struffert 2, Joachim Hornegger 1,4, Arnd Doerfler

More information

University College London, Gower Street, London, UK, WC1E 6BT {g.gao, s.tarte, p.chinchapatnam, d.hill,

University College London, Gower Street, London, UK, WC1E 6BT {g.gao, s.tarte, p.chinchapatnam, d.hill, Validation of the use of photogrammetry to register preprocedure MR images to intra-procedure patient position for image-guided cardiac catheterization procedures Gang Gao 1, Segolene Tarte 1, Andy King

More information

Automatic Lung Surface Registration Using Selective Distance Measure in Temporal CT Scans

Automatic Lung Surface Registration Using Selective Distance Measure in Temporal CT Scans Automatic Lung Surface Registration Using Selective Distance Measure in Temporal CT Scans Helen Hong 1, Jeongjin Lee 2, Kyung Won Lee 3, and Yeong Gil Shin 2 1 School of Electrical Engineering and Computer

More information

Scaling Calibration in the ATRACT Algorithm

Scaling Calibration in the ATRACT Algorithm Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,

More information

Make the Most of Time

Make the Most of Time Make the Most of Time Temporal Extension of the itv Algorithm for 4D Cardiac C-Arm CT Viktor Haase 1,5, Oliver Taubmann 2,3, Yixing Huang 2, Gregor Krings 4, Günter Lauritsch 5, Andreas Maier 2,3, Alfred

More information

Figure 11.: Detail of final frames from image sequences shown in Figure C 11.D

Figure 11.: Detail of final frames from image sequences shown in Figure C 11.D Figure 11.: Detail of final frames from image sequences shown in Figure 10. Figures 11.a, 11.b and 11.c are larger versions of the highlighted images shown in figure 10.a, 10.b and 10.c respectively. 11.A

More information

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance

More information

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization

More information

4D Magnetic Resonance Analysis. MR 4D Flow. Visualization and Quantification of Aortic Blood Flow

4D Magnetic Resonance Analysis. MR 4D Flow. Visualization and Quantification of Aortic Blood Flow 4D Magnetic Resonance Analysis MR 4D Flow Visualization and Quantification of Aortic Blood Flow 4D Magnetic Resonance Analysis Complete assesment of your MR 4D Flow data Time-efficient and intuitive analysis

More information

ONE of the promising techniques for noninvasive study of

ONE of the promising techniques for noninvasive study of IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL 20, NO 6, JUNE 2001 499 Fast LV Motion Estimation Using Subspace Approximation Techniques Yu-Ping Wang, Member, IEEE, Yasheng Chen, Amir A Amini*, Senior Member,

More information

Compression of Light Field Images using Projective 2-D Warping method and Block matching

Compression of Light Field Images using Projective 2-D Warping method and Block matching Compression of Light Field Images using Projective 2-D Warping method and Block matching A project Report for EE 398A Anand Kamat Tarcar Electrical Engineering Stanford University, CA (anandkt@stanford.edu)

More information

2005 IEEE Nuclear Science Symposium Conference Record M10-2

2005 IEEE Nuclear Science Symposium Conference Record M10-2 25 IEEE Nuclear Science Symposium Conference Record M1-2 Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng 1, Jeffrey A. Fessler 1, and James M. Balter 2 rzeng@eecs.umich.edu, fessler@eecs.umich.edu,

More information

Improvement and Evaluation of a Time-of-Flight-based Patient Positioning System

Improvement and Evaluation of a Time-of-Flight-based Patient Positioning System Improvement and Evaluation of a Time-of-Flight-based Patient Positioning System Simon Placht, Christian Schaller, Michael Balda, André Adelt, Christian Ulrich, Joachim Hornegger Pattern Recognition Lab,

More information

Accurate and Dense Wide-Baseline Stereo Matching Using SW-POC

Accurate and Dense Wide-Baseline Stereo Matching Using SW-POC Accurate and Dense Wide-Baseline Stereo Matching Using SW-POC Shuji Sakai, Koichi Ito, Takafumi Aoki Graduate School of Information Sciences, Tohoku University, Sendai, 980 8579, Japan Email: sakai@aoki.ecei.tohoku.ac.jp

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Prostate Brachytherapy Seed Reconstruction Using C-Arm Rotation Measurement and Motion Compensation

Prostate Brachytherapy Seed Reconstruction Using C-Arm Rotation Measurement and Motion Compensation Prostate Brachytherapy Seed Reconstruction Using C-Arm Rotation Measurement and Motion Compensation Ehsan Dehghan 1, Junghoon Lee 2, Mehdi Moradi 3,XuWen 3, Gabor Fichtinger 1, and Septimiu E. Salcudean

More information

An Improved Tracking Technique for Assessment of High Resolution Dynamic Radiography Kinematics

An Improved Tracking Technique for Assessment of High Resolution Dynamic Radiography Kinematics Copyright c 2008 ICCES ICCES, vol.8, no.2, pp.41-46 An Improved Tracking Technique for Assessment of High Resolution Dynamic Radiography Kinematics G. Papaioannou 1, C. Mitrogiannis 1 and G. Nianios 1

More information

Interpol, Routines for Interpolation

Interpol, Routines for Interpolation Interpol, Routines for Interpolation Alan Louis Scheinine, Senior Researcher, CRS4 CRS4 Centro di Ricerca, Sviluppo e Studi Superiori in Sardegna Sesta Strada, Ovest Zona Industriale Macchiareddu 09010

More information

BLUT : Fast and Low Memory B-spline Image Interpolation

BLUT : Fast and Low Memory B-spline Image Interpolation BLUT : Fast and Low Memory B-spline Image Interpolation David Sarrut a,b,c, Jef Vandemeulebroucke a,b,c a Université de Lyon, F-69622 Lyon, France. b Creatis, CNRS UMR 5220, F-69622, Villeurbanne, France.

More information

Image Thickness Correction for Navigation with 3D Intra-cardiac Ultrasound Catheter

Image Thickness Correction for Navigation with 3D Intra-cardiac Ultrasound Catheter Image Thickness Correction for Navigation with 3D Intra-cardiac Ultrasound Catheter Hua Zhong 1, Takeo Kanade 1,andDavidSchwartzman 2 1 Computer Science Department, Carnegie Mellon University, USA 2 University

More information

Image Registration. Prof. Dr. Lucas Ferrari de Oliveira UFPR Informatics Department

Image Registration. Prof. Dr. Lucas Ferrari de Oliveira UFPR Informatics Department Image Registration Prof. Dr. Lucas Ferrari de Oliveira UFPR Informatics Department Introduction Visualize objects inside the human body Advances in CS methods to diagnosis, treatment planning and medical

More information

Fmri Spatial Processing

Fmri Spatial Processing Educational Course: Fmri Spatial Processing Ray Razlighi Jun. 8, 2014 Spatial Processing Spatial Re-alignment Geometric distortion correction Spatial Normalization Smoothing Why, When, How, Which Why is

More information

Cardiac Dual Energy CT: Technique

Cardiac Dual Energy CT: Technique RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de

More information

MEDICAL IMAGE ANALYSIS

MEDICAL IMAGE ANALYSIS SECOND EDITION MEDICAL IMAGE ANALYSIS ATAM P. DHAWAN g, A B IEEE Engineering in Medicine and Biology Society, Sponsor IEEE Press Series in Biomedical Engineering Metin Akay, Series Editor +IEEE IEEE PRESS

More information

NIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13.

NIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13. NIH Public Access Author Manuscript Published in final edited form as: Med Phys. 2008 February ; 35(2): 660 663. Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic

More information

Human Heart Coronary Arteries Segmentation

Human Heart Coronary Arteries Segmentation Human Heart Coronary Arteries Segmentation Qian Huang Wright State University, Computer Science Department Abstract The volume information extracted from computed tomography angiogram (CTA) datasets makes

More information

Reduced Oder Modeling techniques to predict the risk of fatigue fracture of peripheral stents. Michel Rochette

Reduced Oder Modeling techniques to predict the risk of fatigue fracture of peripheral stents. Michel Rochette Reduced Oder Modeling techniques to predict the risk of fatigue fracture of peripheral stents Michel Rochette 1 RT3S: Real Time Simulation for Safer vascular Stenting 2 The vascular disease in peripheral

More information

GPU Based Region Growth and Vessel Tracking. Supratik Moulik M.D. Jason Walsh

GPU Based Region Growth and Vessel Tracking. Supratik Moulik M.D. Jason Walsh GPU Based Region Growth and Vessel Tracking Supratik Moulik M.D. (supratik@moulik.com) Jason Walsh Conflict of Interest Dr. Supratik Moulik does not have a significant financial stake in any company, nor

More information

Computational Medical Imaging Analysis Chapter 4: Image Visualization

Computational Medical Imaging Analysis Chapter 4: Image Visualization Computational Medical Imaging Analysis Chapter 4: Image Visualization Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky Lexington,

More information

Initial Clinical Experience with 3D Surface Image Guidance

Initial Clinical Experience with 3D Surface Image Guidance Initial Clinical Experience with 3D Surface Image Guidance Amanda Havnen-Smith, Ph.D. Minneapolis Radiation Oncology Ridges Radiation Therapy Center Burnsville, MN April 20 th, 2012 Non-funded research

More information