QUICKStart Guide APT Versions 3.0 through 3.3

Size: px
Start display at page:

Download "QUICKStart Guide APT Versions 3.0 through 3.3"

Transcription

1 QUICKStart Guide APT Versions 3.0 through 3.3

2 PCS QUICKStart Guide Powering On & Off Seiko Printer Consumable Components Pipette Calibration Setup Pipette Calibration Completing Calibration Instrument Calibration Connecting to APT APT Inventory & Worklist Methods & Test Plans Calibration with APT Contact Information...28

3 Powering On & Off The color-coded bars within the text throughout this Guide indicate the sections where specific instructions are given. Vial block Power switch Powering On Connect system according to diagram below. Before performing this procedure, ensure that the Reagent Kit is at the same temperature as the Instrument. Switch the Instrument and the printer to ON. (The Instrument is on when the I on the rocker switch is depressed.) The Instrument will perform a system test. Be sure that the printer is ON LINE. Clean vials if necessary. Load CAL A and Blank vials into the vial block at the back of the Instrument at least 15 minutes prior to use. Powering Off Remove all vials from the Instrument. Switch the printer to OFF. Switch the Instrument to OFF. Store all reagents according to storage requirements. NOTE: If the STAR Dot Matrix Printer is being used, please refer to the STAR Printer User s Manual for printer connection and operation. 2

4 Always refer to the PCS Procedure Guide for further instruction 3 Powering On & Off

5 Seiko Printer Controls and Indicators: Power Switch: When the switch is in the ON position, power is supplied to the printer and the POWER and ON LINE indicators illuminate. When the printer is out of paper, the OFF LINE indicator flashes. Paper Feed Switch: Pressing this switch feeds the paper continuously. Paper feed will only work when the OFF LINE indicator illuminates or flashes. 4

6 Loading the Printer Paper: Turn the power switch to the ON position. Open the paper cover located on the top, right side of the printer. Cut the end of the paper roll cleanly and squarely. Insert the end of the paper into the inlet at the bottom of the paper holder until the auto-loader catches it and feeds the paper through. The outside of the roll is the printable surface for the thermal paper roll. Close the paper cover. Always refer to the PCS Procedure Guide for further instruction 5 Seiko Printer

7 Consumable Components Vial Cleaning Storage The Reagent Kit should be stored at room temperature (18-28 C). Avoid temperatures exceeding 32 C. Kits should be stored in the dark to the extent practical. Frozen reagents should be brought to room temperature and inverted several times prior to use. If Reagent Kit has been moved from another location, depending on the difference in temperature between the two locations, an hour may be required for equilibration. To ensure accurate results, all vials should be free of fingerprints and smudges prior to placing in the vial block for equilibration. All vials should be cleaned prior to use. Holding the vial by the cap, immerse the lower portion of the vial (about 2.5 cm) into clean isopropyl alcohol. Remove the vial from the alcohol and gently wipe the lower portion of the vial by rotating the vial against a lint-free wipe. Disposal All PCS reagents are non-hazardous and aqueous based. Small quantities may be disposed of in sewer. Large quantities should be disposed of in accordance with local, state or federal regulations. Removing the kit components from the box and letting them stand in free air will hasten equilibration. 6

8 Reagent Kit Blank Reagent Vials: Standard Reagent Kits contain 12 vials pre-filled with Blank Solution. Bulk Reagent Kits contain 50 vials pre-filled with Blank Solution. Sample Solutions: Range 1: For pipette volumes µl Range 2: For pipette volumes µl Range 3: For pipette volumes µl Range 4: For pipette volumes 2 9 µl Range 5*: For pipette volumes µl Range 6*: For pipette volumes µl *Range 5 and 6 are sold separately. Sample Aliquot Containers : Use aliquot containers to hold Sample Solutions used during calibration to avoid contamination of Sample Solution bottle. Transfer Pipettes: Plastic transfer pipettes are used to transfer Sample Solution from the reagent bottle to the aliquot container. Calibration Standard: CAL A vial is used to collect a zero reading for the Instrument. Kit Inserts: Declaration of Traceability Lot code card Always refer to the PCS Procedure Guide for further instruction 7 Consumable Components

9 Pipette Calibration Setup Figure 1 Figure 2 Figure 3 Figure 4 Main Menu Pipette Calibration? Enter Reagent Lot Code: Lot Code XXXXX Correct? (Y/N) Insert Cal A Vial Starting a Pipette Calibration At the Main Menu prompt (Figure 1), press YES to begin the pipette calibration procedure. The Instrument will perform a temperature check. The Instrument will prompt for lot code (Figure 2). The Instrument retains up to 6 previously entered lot codes. If the entered code matches one previously entered, press NO and re-enter the code. Press YES if entering a new lot code. Group numbers found on the correct side of the lot code card must be entered. Press ENTER after each group of digits is entered. A confirmation prompt will be displayed (Figure 3). Press YES. The Instrument will perform a brief self-calibration. Immediately following the self-calibration the Instrument will prompt to insert the CAL A vial (Figure 4). Open the vial holder cover and insert the CAL A vial pressing down until it meets the bottom of the vial holder. The Instrument will then perform a zero check and prompt to remove the CAL A when the test is complete. Return the CAL A vial to the Reagent Kit for future use. 8

10 Figure 5 Remove Cap and Insert Blank The Instrument will then prompt to insert a blue Blank vial into the vial holder (Figure 5). Remove the Blank vial from the vial block and remove the cap. Insert the open vial into the vial holder pressing down until it contacts the bottom of the vial holder. Close the Instrument cover. The Instrument will read the Blank vial. Important Notes: Use the CAL A vial from the Reagent Kit and not from the Instrument Calibrator Kit. Never remove the cover of the CAL A vial. The label for the CAL A vial should face the front of the Instrument with the vertical line facing forward. Each Blank vial contains a predetermined volume of solution. Any spillage will reduce accuracy. The lower portion of the Blank vial must be free of smudges. Touching this area should be avoided. Once the Blank vial has been inserted into the Instrument, it must not be turned or moved until the calibration procedure is complete for that vial. Always refer to the PCS Procedure Guide for further instruction 9

11 Enter Operator ID: The Instrument will prompt to enter an operator ID (Figure 6). Enter a number (not exceeding 20 digits) to identify the operator. For alpha character entry, refer to Appendix E of the PCS Procedure Guide. Figure 6 Enter the pipette ID (Figure 7). Type in a number not exceeding 20 digits and press ENTER. The printer will print the calibration header information. Enter Pipette ID: Enter the volume (in microliters) to be dispensed and press ENTER (Figure 8). The CLEAR key can be used to correct any typing errors. Figure 7 Enter Pipette Volume: µl Figure 8 10

12 Figure 9 XX Left this Vial Proceed? (Yes/No) Range X Pipette Sample #1 The number of samples that can be dispensed in this vial, at the selected volume, will be displayed (Figure 9). Press YES and the Instrument will display the appropriate Sample Solution to use for the pipette calibration. The Instrument will take a final reading of the Blank vial and then prompt for the first dispense of the Sample Solution with the pipette under test (Figure 10). Figure 10 Important Notes: The proper range of Sample Solution for the pipette volume to be tested must be used. Each Reagent Kit contains bottles of Sample Solution corresponding to various pipette volume ranges. If an improper Sample Solution has been dispensed into the Blank vial, the results obtained will not be valid and the vial is no longer usable. It must be discarded. Always refer to the PCS Procedure Guide for further instruction 11 Pipette Calibration Setup

13 Pipette Calibration Aspirating and Dispensing Sample Solution Using a plastic transfer pipette, transfer an aliquot of Sample Solution into an aliquot container. Both items are located in the Reagent Kit. Transfer only enough solution for half-hour increments. The volume gradations on the transfer pipette and the 1 ml mark on the aliquot container can be used as a guide for the volume of solution transferred. Aspirate the Sample Solution from the aliquot container with the pipette under test. Lift the vial holder cover and dispense the Sample Solution into the Blank vial using standard pipetting technique. Important Note: The Sample Solution should be dispensed against the inside wall of the vial with the pipette tip placed above the meniscus but below the shoulder of the Blank vial. The vortex mixing action will wash the solution from the inside wall of the vial. 12

14 Figure 11 Figure 12 Figure 13 Mixing Sample #1 Please Wait Reading Sample #1 Please Wait Result #1 XXX.X Pipette Sample #2 Reading the Sample Once the sample is dispensed, and the cover is closed, the Instrument will mix the sample and display a prompt (Figure 11). The Instrument will then read the absorbance of the sample and display a prompt (Figure 12). Finally, the result of the reading will be displayed on the screen and the Instrument will prompt for the dispense of the next sample (Figure 13). Follow the same steps listed on this page for the next sample and every subsequent sample. When performing a single volume calibration and the desired number of data points have been collected, press END OF RUN. When performing a multi-volume calibration, press ENTER (see shaded section below). The group statistics will be computed and the results printed. Rules for Adjustable Volume Pipettes: Calibration can be performed by starting at any volume. When the desired number of samples at the first volume have been dispensed, press ENTER. If a data point must be deleted, press YES at the reprint results prompt and follow the delete a data point instructions. Enter the next volume in the pipette s range. This process can be repeated. Press END OF RUN when complete. In some cases, more than one Sample Solution will be required. Special attention should be paid to the screen prompts to ensure usage of correct solution. Always refer to the PCS Procedure Guide for further instruction 13 Pipette Calibration

15 Completing Calibration Figure 14 Figure 15 Reprint Results? (Yes/No) Delete Data Points? (Yes/No) Enter Pt # to Delete If, at the end of the run, a second copy of the results is desired, press YES at the next prompt (Figure 14). If it is determined that a given data point may be eliminated from the statistical calculation, press YES at the next prompt (Figure 15). Enter the sample # of the data point to be deleted at the next prompt (Figure 16). At the next prompt to delete more data points, follow the same steps until all appropriate data points are deleted. The correct statistics will be printed. For suggested tolerance limits, refer to section 4.13 of the Procedure Guide Figure 16 14

16 Figure 17 Figure 18 Figure 19 Another Cal w/vial? (Yes/No) Use New Vial? (Yes/No) End of Vial The number of data points that can be collected in each Blank vial is dependent on the sample volume and will range from 1 point to 40. Each time a pipette calibration run is completed, the Instrument will display the remaining number of samples of the selected volume that may be dispensed into the vial. In any one Blank vial, calibrations may be performed at the same volume, at different volumes with the same Sample Solution or at different volumes with different Sample Solutions. To use the same vial after the completion of the first calibration, press YES at the first prompt (Figure 17). The calibration cycle will re-start. When the maximum capacity of the vial has been reached, alternating prompts will flash (Figures 18 & 19). If more data points are required, answer YES to the Use New Vial prompt. The Instrument will prompt to remove the old vial and insert a new one. This process may be repeated if needed. Always refer to the PCS Procedure Guide for further instruction 15 Completing Calibration

17 Instrument Calibration Figure 20 Main Menu Instr. Calibration? Instrument Cal. Menu Perform Instr. Cal.? Calibrating the PCS: To access the Instrument Calibration Menu, press MENU SCROLL until the Instrument Calibration prompt is displayed (Figure 20). Press YES at the prompt. The Instrument Calibration Menu will open. Press YES at the Perform Instrument Calibration prompt (Figure 21). Figure 21 Important Notes: Minimally, the Instrument Calibration should be performed every 30 days or prior to the next use, whichever is later. The Instrument is calibrated using the PCS Instrument Calibrator Kit, which contains 4 vials with calibrator values: CAL A, B, C and D. The contents of the kit are subject to degradation with time, temperature and exposure to light. Instrument calibration can now be initiated directly with Artel Pipette Tracker (APT) software (v.3.3). PCS Instrument calibration reports can be viewed, stored, and printed using Pipette Tracker. 16

18 Figure 22 Figure 23 Figure 24 Important Notes: CAL Lot Code XXXXX New Lot? (Y/N) Insert CAL A Vial Insert CAL B Vial If the calibration is successful, the date of the last Instrument calibration check will change to the current day. Never remove the caps of the calibrator vials. Printed results will include: - Reading data - Instrument accuracy data - Instrument repeatability data - Overall calibration result At the prompt, enter the Operator ID for the Instrument calibration. For alpha character entry, see appendix E of the Procedure Guide. The Instrument will prompt to enter the Calibrator Kit lot code. This code is a four or five digit number found on the inside label of the Instrument Calibrator Kit box lid. After entering the lot code, press ENTER. If this lot code matches one previously entered, the next prompt will be to insert the first Cal Vial. If the lot code is for a new kit, press YES when the Instrument identifies it as new (Figure 22). The new calibrator values are found on the box label and must be entered at the prompt. The leading zeros and the decimal points must be entered. Press YES to confirm when the lot code is completely entered. After the lot code is successfully entered, the Instrument will take a series of readings and then prompt for the first calibrator vial (Figure 23). Use the CAL A vial in the Instrument Calibrator Kit. When the readings are complete, the Instrument will then prompt for the next vial (Figure 24). The remainder of the calibrator vials must be inserted at the sequential prompts. Results will be printed at conclusion. The entire process takes approximately 12 minutes to complete. Always refer to the PCS Procedure Guide for further instruction 17 Instrument Calibration

19 Connecting to APT Connecting the PCS Instrument to Artel Pipette Tracker software With the PCS Instrument power switch turned OFF, use the included serial cable to connect the serial port on the PCS with the serial port on the PC on which the Artel Pipette Tracker software has been installed (Figure 25). Figure 25 - Back of PCS 18

20 Computer Link ready for command Turn the Instrument power switch ON, and use the MENU SCROLL key on the PCS to advance the menu to Link to Computer. Figure 26 Press YES to enter Link to Computer mode. This completes the link and the PCS instrument indicates that it is ready for use with Artel Pipette Tracker (Figure 26). To leave Link to Computer mode at any time, press RESET but note that any unused reagent in the PCS will need to be discarded. Always refer to the PCS Procedure Guide for further instruction 19 Connecting to APT

21 APT Inventory & Worklist Artel Pipette Tracker Inventory and Worklist The Inventory window shows a list of all devices* that are being tracked (Figure 27). The Inventory includes the size, model and number of channels for each pipette, based on the unique ID number assigned to it. It can also store optional information such as the pipette location, the assigned user and a description of how it is used. The Worklist window displays a list of the devices currently due for calibration. Each pipette may be listed up to three times, once for each Interval. The sort order of the list may be changed by clicking the heading of the column to sort on; for example, by ID number or by Location. *In Artel Pipette Tracker v.3.3, pipettes are referred to as Devices. 20

22 Figure 27 New in APT v.3.3 PCS Instrument calibration can now be initiated and tracked within Artel Pipette Tracker. Always refer to the Artel Pipette Tracker User Manual for further instruction 21 APT Inventory & Worklist

23 Methods & Test Plans Methods and Test Plans Methods and Test Plans are used to describe calibration procedures in the program. Each calibration requires the selection of both a Method and a Test Plan. A Method specifies the number of data points to be taken for the calibration and the criteria to be used to evaluate the calibration results (e.g. accuracy and/or precision (CV)). Pipettes and Intervals Calibration Intervals for each pipette include its calibration schedule, and associated Methods and Test Plans. Up to three Intervals may be associated with each pipette (for example, weekly performance checks with four data points, and monthly and quarterly calibrations with ten data points). A Test Plan specifies the set of volumes to be used for testing, and the tolerance limit values used to evaluate the test results at each volume. 22

24 To Set Up a Method: Choose Setup Method. Click ADD and give the new Method a name and description (Figure 28). Enter the number of samples per volume, the method of reporting, and the testing criteria. Include any other comments, and accept the changes by clicking OK on the Setup and List dialog boxes. New in APT v.3.3 More flexible entry for As Found and As Calibrated samples Figure 28 Always refer to the Artel Pipette Tracker User Manual for further instruction 23

25 To Set Up a Test Plan: Choose Setup Test Plan. Click ADD and give the new Test Plan a name and description (Figure 29). Choose a Limit Table from the list or create a new one by entering the volumes and tolerance limits. Include any other comments and accept the changes by clicking OK on the Setup and List dialog boxes. New in APT v.3.3 More flexible entry for As Found and As Calibrated samples New option for High to Low volume order Figure 29 24

26 To Set Up a Pipette: Choose Setup Device. Click ADD and give the new device a unique ID (Figure 30). Choose a Model for the device.* Choose the Interval 1 tab and select the desired Method and Test Plan. Select a Calibration Interval and then accept the changes by clicking OK on the Setup and List dialog boxes. Notice the new entry in the Inventory and Worklist. * In APT v.3.2, the Device model is entered in the Details Tab. Figure 30 Always refer to the Artel Pipette Tracker User Manual for further instruction 25 Methods & Test Plans

27 Calibration with APT Starting a Calibration Ensure PCS is set to Computer Link, Ready for Command. Launch Pipette Tracker and select the new device entry from the worklist. Click CALIBRATE SELECTED on the toolbar. The calibration dialog box opens (Figure 31). The program then communicates with the PCS Instrument, and reads the temperature and last date of Instrument calibration. If the PCS has not been calibrated in 30 days, a warning will appear. Perform the PCS Instrument Calibration* prior to continuing. The reagent lot code must match the lot code on the Blank vials and Sample Solution bottles in order for a calibration to be valid. - The calibration dialog box lists (up to) the six most recently used reagent lot codes. - If the correct lot code is not on the list, click NOT LISTED. - The PCS Instrument will prompt for any missing lot code information before beginning the calibration. There are optional fields to enter an operator name if not logged in through the security settings, any pipette tip information and a comment. Click START RUN in the calibration dialog box to initiate a pipette calibration using the PCS Instrument. * PCS Instrument Calibration can be initiated directly with Artel Pipette Tracker v3.3. Figure 31 New in APT v ➊ Choose from a user-created list of Measurement Standards at run time ➋ Create and store maintenance plans ➌ Graphically view data points in real time in relation to the device s tolerance limits (to access, click the Chart tab) ➋ ➊ ➌ ➌ 26

28 Performing a Calibration The Run Time tab will appear (Figure 32) and the program will initiate a pipette calibration with the PCS Instrument. Follow the prompts on the PCS Instrument to complete the calibration. As the data points are collected, the program calculates summary statistics for the calibration and displays them across the bottom of the calibration window. Problem data points may be redone immediately following a problem with pipetting. APT v Data points may be redone at any time during calibration. Outlier data points may be deleted at any time during the calibration to exclude them from the summary statistics. Once all data are collected, the result is reflected in the Status box, according to whether or not the device has performed within tolerance. In tolerance status is highlighted green, and out of tolerance status is highlighted red. Click END OF RUN to complete and save the calibration. To print additional reports or labels, click on ellipsis box to the right of PRINT RUN. Select desired report, click OK, then click PRINT RUN box. Once all desired reports have been printed, click CLOSE to close the window. Figure 32 New in APT v Appearance of this screen differs slightly from v.3.2. Always refer to the Artel Pipette Tracker User Manual for further instruction 27 Calibration with APT

29 Trust Your Results 25 Bradley Drive Westbrook, Maine toll-free: tel: fax: ordering: technical support: A5721D

Vial block. Power switch

Vial block. Power switch Powering On & Off The color-coded bars within the text throughout this Guide indicate the sections where specific instructions are given. Vial block Power switch Powering On Connect system according to

More information

4 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (max.) Up to 7 plates / run Reaction / reading system

4 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (max.) Up to 7 plates / run Reaction / reading system Specifications & more > > System Specifications > > Consumables > > Accessories System Overview Cat. No. 16300 Analyzer type 4 plate fully automated ELISA analyzer Reagent system Open (Human tests predefined)

More information

scil Technical Support

scil Technical Support Vet ABC Hematology Analyzer Calibration ** Before Calibration is performed run a concentrated cleaning and make sure your Photometer is within range. After both are completed run a Startup by pressing

More information

GEMINI. Compact Microplate Processor. System Guide

GEMINI. Compact Microplate Processor. System Guide GEMINI Compact Microplate Processor System Guide 1 Gemini System Guide Table of Contents 1. Introduction 3 2. User Levels 4 3. Instrument Components & Accessories 6 4. Basic Operation - Quick Guide: 1-Lane

More information

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette Pipettes allow transfer of accurately known volume of solution from one container to another. Volumetric or transfer pipettes deliver a

More information

Chemical Oxygen Demand (COD Detector) COD-200. Operation Manual

Chemical Oxygen Demand (COD Detector) COD-200. Operation Manual Chemical Oxygen Demand (COD Detector) COD-200 Operation Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st. Israel 5887 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.0.6

More information

CHLORINE DIOXIDE (0 to 5.00 mg/l) Method 10126

CHLORINE DIOXIDE (0 to 5.00 mg/l) Method 10126 CHLORINE DIOXIDE (0 to 5.00 mg/l) Method 10126 DPD Method* Scope and Application: for measuring chlorine dioxide in clean water For water 530 nm 1. Enter the stored program for chlorine dioxide (ClO 2

More information

MVS Data Manager Integration

MVS Data Manager Integration MVS Data Manager Integration An Example: Executing an MVS Volume Verification of a Tecan EVO Using EVOware Software Alexis L. Rogers ARTEL INTRODUCTION: This paper describes how to run or call an Artel

More information

BioRobot 9604 Operating Procedure

BioRobot 9604 Operating Procedure BioRobot 9604 Operating Procedure Note: An import table (directions given in number 20) must be made prior to a robot run. 1. Prior to operating the BioRobot 9604, perform all daily and weekly maintenance

More information

Magnetic Immunoassay Reader Operator Manual

Magnetic Immunoassay Reader Operator Manual Page 1 of 8 Magnetic Immunoassay Reader Operator Manual Table of contents 1 Introduction...1 2 Magnetic Immunoassay Reader...1 2.1 Contact information and spare parts...1 3 Contents of delivery...1 4 Technical

More information

Picus & Picus NxT Electronic Pipette User Manual

Picus & Picus NxT Electronic Pipette User Manual Picus & Picus NxT Electronic Pipette User Manual Table of Contents 1. Introduction 3 1.1. Intended Use 4 1.2. Product Overview 4 1.2.1. Single and Multichannel Pipettes 4 1.2.2. Display 5 1.2.3. Optifit

More information

Reagent Transfer User Guide

Reagent Transfer User Guide Reagent Transfer User Guide For Research Use Only. Not for use in diagnostic procedures. This guide contains the following topics: About this guide on page 2 Utility description on page 3 Before you start

More information

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample

More information

Ion 540 Kit Chef QUICK REFERENCE. Dilute the libraries. Prepare the libraries and consumables. Create a Planned Run. Catalog Numbers A27759, A30011

Ion 540 Kit Chef QUICK REFERENCE. Dilute the libraries. Prepare the libraries and consumables. Create a Planned Run. Catalog Numbers A27759, A30011 QUICK REFERENCE Ion 540 Kit Chef Catalog Numbers A27759, A30011 Pub. No. MAN0014119 Rev. C.0 Note: For safety and biohazard guidelines, see the Safety appendix in the Ion 540 Kit Chef User Guide (Pub.

More information

Policy # MI_VTK Department of Microbiology. Page Quality Manual VITEK MANUAL TABLE OF CONTENTS

Policy # MI_VTK Department of Microbiology. Page Quality Manual VITEK MANUAL TABLE OF CONTENTS Department of Microbiology Version: 1.0 CURRENT 1 of 28 Prepared by QA Committee Issued by: Laboratory Manager Revision Date: 10/2/2017 Approved by Laboratory Director: Annual Review Date: 5/1/2018 Microbiologist-in-Chief

More information

MINISCAN INSTRUCTIONS

MINISCAN INSTRUCTIONS MINISCAN INSTRUCTIONS DR VIRGINIA GORDON Version 3, July 2011 General Hints on Running Mini-Scan Tests... 1 Total Glycerin Test (TG)... 3 GENERAL HINTS ON RUNNING MINI-SCAN TESTS PIPETTES To use the pipettes

More information

1 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (Max.) 1 plate /run Barcode reader

1 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (Max.) 1 plate /run Barcode reader Specifications & more Elisys Uno > > System Specifications > > Consumables > > Accessories System Overview Cat.. 17350 Analyzer type 1 plate fully automated ELISA analyzer Reagent system Open (Human tests

More information

US IVD Bacterial Test Standard

US IVD Bacterial Test Standard Instructions for Use US IVD Bacterial Test Standard Mass calibration standard containing a typical Escherichia coli DH5 alpha peptide and protein profile plus additional proteins. For quality control and

More information

VIAFLO 96 VIAFLO 384. Microplate Pipetting Easy and Affordable

VIAFLO 96 VIAFLO 384. Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable VITARIS AG Blegistrasse 9 CH-6340 Baar Tel.: 041 769 00 00 Fax: 041 769 00 01 E-mail: info@vitaris.com www.vitaris.com VIAFLO 96 VIAFLO 384

More information

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Section 1: Starting an experiment 1.1 Opening a file 1. Open the File menu and select Open Activity. 2. In the Open dialog box, navigate to

More information

Thermo Scientific Orion Star A223 and Star A323 Portable RDO/DO Meter. Instruction Sheet

Thermo Scientific Orion Star A223 and Star A323 Portable RDO/DO Meter. Instruction Sheet English RDO / DO Thermo Scientific Orion Star A223 and Star A323 Portable RDO/DO Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket

More information

STANDARD OPERATING PROCEDURES DIVISON OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

STANDARD OPERATING PROCEDURES DIVISON OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA STANDARD OPERATING PROCEDURES DIVISON OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1139.5 Date Issued: 12/03 Date Revised: 8/16 Page 1 of 5 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: Animal Care

More information

INSTRUCTION MANUAL. Instruction Manual. Analog Multi-Tube Vortexer Digital Multi-Tube Vortexer

INSTRUCTION MANUAL. Instruction Manual. Analog Multi-Tube Vortexer Digital Multi-Tube Vortexer INSTRUCTION MANUAL Instruction Manual Analog Multi-Tube Vortexer Digital Multi-Tube Vortexer Table of Contents Package Contents............ 1 Warranty............ 1 Installation............ 2 Maintenance

More information

EDP3-Plus. Advanced Ergonomic Electronic Pipette. 9 pipettes for volume ranges from 0.5 µl to 20 ml. 200 µl EDP3-Plus shown

EDP3-Plus. Advanced Ergonomic Electronic Pipette. 9 pipettes for volume ranges from 0.5 µl to 20 ml. 200 µl EDP3-Plus shown EDP3-Plus Advanced Ergonomic Electronic Pipette 9 pipettes for volume ranges from 0.5 µl to 20 ml 200 µl EDP3-Plus shown 1 TABLE OF CONTENTS Introduction...1 Unpacking...1 Set Up / Initial Charge...2 Sleep

More information

6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory

6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory 6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory Overvoltage Category II Pollution Degree 2 Equipment Class III Safety information Introduction Before using this accessory, you must read

More information

Peel & Present Option Rev.C

Peel & Present Option Rev.C Peel & Present Option 92-2479-01 Rev.C Contents of the Peel & Present Kit This kit contains the following items: Peel and Present Mechanism Assist Roller Bushing Follow the steps below to install these

More information

Optimizing Accuracy Performance on a Beckman Coulter Biomek Using the Artel MVS

Optimizing Accuracy Performance on a Beckman Coulter Biomek Using the Artel MVS Optimizing Accuracy Performance on a Beckman Coulter Biomek Using the Artel MVS Keith J. Albert Artel ABSTRACT: This application note discusses one approach for optimizing pipetting accuracy on a Beckman

More information

Accessory GP-EXT. EXTERNAL USB PRINTER User's Manual

Accessory GP-EXT. EXTERNAL USB PRINTER User's Manual Accessory GP-EXT EXTERNAL USB PRINTER User's Manual WaveSurfer is a trademark of LeCroy Corporation. Windows is a registered trademark or a trademark of Microsoft Corporation in the United States and other

More information

Thermo Scientific Matrix Hybrid Pipette User Manual

Thermo Scientific Matrix Hybrid Pipette User Manual Thermo Scientific Matrix Hybrid Pipette User Manual Trademarks and Copyright 2007 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and

More information

Thermo Scientific Orion Star A326 Portable ph/rdo/do Meter. Instruction Sheet

Thermo Scientific Orion Star A326 Portable ph/rdo/do Meter. Instruction Sheet English ph / RDO / DO Thermo Scientific Orion Star A326 Portable ph/rdo/do Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket plug

More information

Troubleshooting Guide

Troubleshooting Guide DS-11 FX Series / DS-C / QFX Technical Note 165 Troubleshooting Guide Rev January 2017 Table of Contents 1. Sample Result Concerns... 4 Microvolume Sample Concentration Accuracy... 4 Microvolume Sample

More information

Mini8 Plus Real-Time PCR System

Mini8 Plus Real-Time PCR System Mini8 Plus Real-Time PCR System Version:EN / 20160413 Coyote, highly specialized in research & development, offers integrate molecular diagnostic solutions by providing devices from sample pre to gene

More information

USER GUIDE Turbidity Meter Model TB400

USER GUIDE Turbidity Meter Model TB400 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 USER GUIDE Turbidity Meter Model TB400 Visit us at www.testequipmentdepot.com Introduction Congratulations on your purchase

More information

Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter

Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter English Cond Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket

More information

PTR. printing temperature recorder. User Manual. order code:

PTR. printing temperature recorder. User Manual. order code: PTR printing temperature recorder User Manual order code: 226-275 2 CONTENTS Introduction 4 probes 4 functions 4 display 4 PTR overview 5 PTR operation 6 switching the PTR on or off 6 changing input 6

More information

Loading paper and specialty media Learning about the printer. Stop all printer activity. 7 Back button Return to the previous screen.

Loading paper and specialty media Learning about the printer. Stop all printer activity. 7 Back button Return to the previous screen. Quick Reference Use the 6 Stop or Cancel button To Stop all printer activity. Loading paper and specialty media Learning about the printer Using the printer control panel 0 Use the 9 To Display View the

More information

Nuclease-free water (e.g. ThermoFisher, cat # AM9937) Freshly-prepared 70% ethanol in nucleasefree water (optional)

Nuclease-free water (e.g. ThermoFisher, cat # AM9937) Freshly-prepared 70% ethanol in nucleasefree water (optional) Before start checklist Materials Consumables Equipment Rapid Barcoding Sequencing Kit (SQK- RBK004) 1.5 ml Eppendorf DNA LoBind tubes Ice bucket with ice Flow Cell Priming Kit (EXP-FLP001) 0.2 ml thin-walled

More information

At 5% print coverage (A4 or Letter size). The actual number of printed pages will vary depending on the average type of print job and paper.

At 5% print coverage (A4 or Letter size). The actual number of printed pages will vary depending on the average type of print job and paper. Drum unit This printer can print up to 30,000 pages * with one drum unit. * At 5% print coverage (A4 or Letter size). The actual number of printed pages will vary depending on the average type of print

More information

ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE

ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE 1.2.80.1071 07MAN1181.A1 Effective Date: 30-Jan-11 ECO #: 3093 Contents What are Quick Set-Up Instructions?... 3 Starting up ClonePix 2...

More information

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet English ph / ISE / COND / RDO / DO Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the

More information

EDP 3-Plus. Advanced Ergonomic Electronic Pipette. 10 pipettes for volume ranges from 0.5 µl to 20 ml

EDP 3-Plus. Advanced Ergonomic Electronic Pipette. 10 pipettes for volume ranges from 0.5 µl to 20 ml EDP 3-Plus Advanced Ergonomic Electronic Pipette 10 pipettes for volume ranges from 0.5 µl to 20 ml TABLE OF CONTENTS Introduction...1 Unpacking...1 Set Up / Initial Charge...2 Sleep Mode...2 Rapid Charge

More information

* IMPORTANT * REGISTERING YOUR MACHINE

* IMPORTANT * REGISTERING YOUR MACHINE * IMPORTANT * REGISTERING YOUR MACHINE Thank you for your purchase of the Keyline 994 Laser. Before continuing with machine setup and use, please complete the following; COMPLETE PRODUCT REGISTRATION FORM

More information

Thermo Scientific Orion Star A214 Benchtop ph/ise Meter. Instruction Sheet

Thermo Scientific Orion Star A214 Benchtop ph/ise Meter. Instruction Sheet English ph / ISE Thermo Scientific Orion Star A214 Benchtop ph/ise Meter Instruction Sheet Preparation Power Source 1. Power adapter (included with meter) a. Select the appropriate wall socket plug plate.

More information

Introduction General description Technology - Mechanical Technology - Viability measurement

Introduction General description Technology - Mechanical Technology - Viability measurement Instruction Manual Table of Contents Introduction General description Technology - Mechanical Technology - Viability measurement 4-6 4 5 6 Product description Packing list Front view of the ADAM-rWBC

More information

DS2 Automated ELISA System

DS2 Automated ELISA System DS2 Automated ELISA System Operator s Manual For Use with DS-Matrix version 1.24 and above IMPORTANT Please read carefully before using the DS2 Revision History Manual Version Revision Date: Revision F

More information

VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable

VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Handheld microplate processing Life science labs are confronted with a growing need to process more samples in parallel

More information

Thermo Scientific Orion Star A324 Portable ph/ise Meter. Instruction Sheet

Thermo Scientific Orion Star A324 Portable ph/ise Meter. Instruction Sheet English ph / ISE Thermo Scientific Orion Star A324 Portable ph/ise Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket plug plate.

More information

Thermo Scientific Orion Star A325 Portable ph/conductivity Meter. Instruction Sheet

Thermo Scientific Orion Star A325 Portable ph/conductivity Meter. Instruction Sheet English ph / COND Thermo Scientific Orion Star A325 Portable ph/conductivity Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket plug

More information

Issued by: LABORATORY MANAGER Original Date: March 14, 2001 Approved by: Laboratory Director Revision Date: June 6, 2003.

Issued by: LABORATORY MANAGER Original Date: March 14, 2001 Approved by: Laboratory Director Revision Date: June 6, 2003. Policy # MI/SER/06/v01 Page 1 of 5 Section: Subject Title: Epstein Barr Virus Serology Issued by: LABORATORY MANAGER Original Date: March 14, 2001 Approved by: Laboratory Director Revision Date: June 6,

More information

Reference Guide for the KODAK Picture Kiosk GS Compact. Software Version 2.0

Reference Guide for the KODAK Picture Kiosk GS Compact. Software Version 2.0 Reference Guide for the KODAK Picture Kiosk GS Compact Software Version 2.0 Eastman Kodak Company, 2008 Contents Product Overview... 4 Console... 4 605 Photo Printer... 5 Product Choices... 6 Startup and

More information

Data Acquisition with CP-2002/2003 Micro-GC Control

Data Acquisition with CP-2002/2003 Micro-GC Control Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598 Star Chromatography Workstation Version 6 Data Acquisition with CP-2002/2003 Micro-GC Control Operation Manual Varian, Inc. 2002

More information

Visit our video library at epson.com/support/f2100 Maintenance Schedule SureColor F2100 White Edition

Visit our video library at epson.com/support/f2100 Maintenance Schedule SureColor F2100 White Edition Visit our video library at epson.com/support/f2100 Maintenance Schedule SureColor F2100 White Edition Shake White Ink Cartridges DAILY (when in operation) WEEKLY MONTHLY AS NEEDED LONG-TERM STORAGE (more

More information

2 1.1 Safety using methods Definition of warning symbols Standard accessories Installation...

2 1.1 Safety using methods Definition of warning symbols Standard accessories Installation... Table of Contents 1 Satety precautions... 2 1.1 Safety using methods...2 1.2 Definition of warning symbols...2 2 Standard accessories......6 3 Installation... 7 3.2 Leg frame... 7 3.3 Installation and

More information

MVPplus Quick Reference Guide

MVPplus Quick Reference Guide MVPplus Quick Reference Guide Use this guide to operate your printer on a daily basis. For more detailed information, refer to the User Guide. Contents External View...........................................................

More information

Applications Overview...

Applications Overview... Applications Overview........................................ 3 Overview Introduction This chapter explains how to set up for a new board and define parameters for tooling, vision, squeegees, stencil cleaning

More information

User Manual. Page 0 of 57

User Manual. Page 0 of 57 User Manual Page 0 of 57 Table of Contents General Information... 4 Quidel Contact Information... 4 Main Menu Structure v.2.0 or higher... 5 Supervisor Menu Structure v.2.0 or higher... 6 Main Menu Structure

More information

BA-88A Semi-auto Chemistry Analyzer

BA-88A Semi-auto Chemistry Analyzer BA-88A Semi-auto Chemistry Analyzer Main Contents Features Working principles Improvements from BA-88 Competitors Operation Price FAQ Features Test mode: flowcell or cuvette 7.0 TFT touch screen & popup

More information

Installing a Printhead Cartridge

Installing a Printhead Cartridge Installing a Printhead Cartridge Summary: Procedure for installing a new printhead cartridge. Applicable Products: ijetpress Tools/Parts Required: A new printhead, deionized or distilled water, lint free

More information

User's Guide QL-700. ENG ver.a.

User's Guide QL-700. ENG ver.a. User's Guide QL-700 Be sure to read and understand this guide before using the machine. We recommend that you keep this guide nearby for future reference. www.brother.com ENG ver.a Thank you for purchasing

More information

Reconstitution Application Guide

Reconstitution Application Guide Reconstitution Application Guide For Research Use Only. Not for use in diagnostic procedures. This guide contains the following topics: About this guide on page 2 App description on page 3 Before you start

More information

A 2 O Advanced Automated Osmometer

A 2 O Advanced Automated Osmometer A 2 O Advanced Automated Osmometer A fully automated, multi-sample osmometer that sets the new benchmark for analytical performance, ease of use, and true walkaway operation. www.aicompanies.com Discover

More information

ClonePix TM FL Quick Set-Up Instructions

ClonePix TM FL Quick Set-Up Instructions ClonePix TM FL Quick Set-Up Instructions Software Release: 1.2.13.971 07MAN1055.A2 Effective Date: 01-Jun-10 QUICK INSTRUCTIONS > CONTENTS Contents What are Quick Set Up Instructions?...3 Starting up ClonePix

More information

DSX Automated ELISA System

DSX Automated ELISA System DSX Automated ELISA System Operator s Manual For Revelation 6.0 and above IMPORTANT Please read carefully before using the DSX ii Revision History Revision Date: September 19, 2005 12-16-2005 Part No.

More information

VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable

VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Handheld microplate processing Have you reached the limits of productivity with traditional handheld pipettes? Would you

More information

Quick Start Guide. Introduction. P4T/RP4T Overview

Quick Start Guide. Introduction. P4T/RP4T Overview 7 6 Thank you for choosing one of our Zebra P4T Series Mobile s. The P4T Series consists of two models. The P4T can print on direct thermal or thermal transfer media. The RP4T can print on direct or thermal

More information

Zebra XiII-Series Printer Quick Reference Guide

Zebra XiII-Series Printer Quick Reference Guide Zebra XiII-Series Printer Quick Reference Guide Contents Media and Ribbon Loading...67 Media Loading...67 Ribbon Loading...70 Operator Controls...72 Front Panel Keys...72 Front Panel Lights...72 Calibration...74

More information

Skyla VB1 Quick Reference Guide

Skyla VB1 Quick Reference Guide Skyla VB1 Quick Reference Guide 1 VB1 DM/JAC 2014 v1.0 Quick Reference Guide Contents External features of the analyser 3 VB1 Function Buttons 4 Testing a Sample 5 How to reconstitute a biochemistry control

More information

VIAFLO 96 VIAFLO 384. Microplate Pipetting Easy and Affordable

VIAFLO 96 VIAFLO 384. Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Handheld microplate processing Have you reached the limits of productivity with traditional handheld pipettes? Would you

More information

Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.

Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2. Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.6 Contents 1. Description 1.1 Background information 1.2 Reagent and

More information

EPSON ActionLaser Read This First. eepa POLLUTION PREVENTER

EPSON ActionLaser Read This First. eepa POLLUTION PREVENTER EPSON ActionLaser 1400 Read This First eepa POLLUTION PREVENTER This booklet is to help you set up your printer and begin printing quickly. It also gives you instructions for routine maintenance. If you

More information

REFILLABLE CARTRIDGE SYSTEM FOR THE EPSON STYLUS PRO 7800 & 9800

REFILLABLE CARTRIDGE SYSTEM FOR THE EPSON STYLUS PRO 7800 & 9800 REFILLABLE CARTRIDGE SYSTEM FOR THE EPSON STYLUS PRO 7800 & 9800 Thank you for purchasing an Ink2image bulk ink feed system. You have just taken the first step to enjoying the cost saving and superior

More information

Fast Facts. for Operating the scil Vet abc Plus. Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education

Fast Facts. for Operating the scil Vet abc Plus. Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education Fast Facts for Operating the scil Vet abc Plus Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education SCIL Vet ABC Plus Fast Facts Table of Contents Start Up Cycle pg. 2 Reagents Status

More information

Protein Quantification Kit (BCA Assay)

Protein Quantification Kit (BCA Assay) Protein Quantification Kit (BCA Assay) Booklet Item NO. KTD3001 Product Name Protein Quantification Kit (BCA Assay) ATTENTION For laboratory research use only. Not for clinical or diagnostic use Version

More information

MAKE THE CHANGE FOR BETTER NEONATAL SCREENING. Introducing the new AutoDELFIA immunoassay system for newborn screening

MAKE THE CHANGE FOR BETTER NEONATAL SCREENING. Introducing the new AutoDELFIA immunoassay system for newborn screening Introducing the new AutoDELFIA immunoassay system for newborn screening MAKE THE CHANGE FOR BETTER NEONATAL SCREENING THE WORLD S LEADING SCREENING PLATFORM PerkinElmer s AutoDELFIA immunoassay system

More information

VIAFLO Fixed and Adjustable Spacing Electronic Pipettes

VIAFLO Fixed and Adjustable Spacing Electronic Pipettes VIAFLO Fixed and Adjustable Spacing Electronic Pipettes VIAFLO Electronic Pipettes All VIAFLO pipettes function as repeaters, mixers, serial dilutors, manual pipettes, handheld sample processors and more!

More information

DS2 Automated ELISA System

DS2 Automated ELISA System DS2 Automated ELISA System Operator s Manual Part No. 91000200 For Use with DS-Matrix version 1.24 and above IMPORTANT Please read carefully before using the DS2 Revision History Manual Version Revision

More information

Getting low-resolution GC-MS data on the JEOL GCMate Introduction

Getting low-resolution GC-MS data on the JEOL GCMate Introduction Getting low-resolution GC-MS data on the JEOL GCMate Introduction GC-MS is a technique that combines gas phase separation and mass-based detection/characterization of analytes. It s a very sensitive method

More information

LED Dot Matrix Digital Clock. Instructions. Contents. v (5x7-1x6 Module)

LED Dot Matrix Digital Clock. Instructions. Contents. v (5x7-1x6 Module) LED Dot Matrix Digital Clock v1.0.2 (5x7-1x6 Module) Instructions Designer: Yan Zeyuan. China Website: E-mail: yanzeyuan@163.com Contents Attention... 2 Functional Features... 2 Technical Specifications..

More information

Data acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2.

Data acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2. Data acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2.8 Contents 1. Description 1.1 Background information 1.2 Reagent and instrument

More information

Preprocessing Methods for the Maxprep Liquid Handler Technical Manual

Preprocessing Methods for the Maxprep Liquid Handler Technical Manual Preprocessing Methods for the Maxprep Liquid Handler Technical Manual Instructions for use of AS9100, AS9101, AS9200 and AS9201 TM529 11/17 Table of Contents 1 Description...1 2...2 2.1. Maxprep Liquid

More information

TT230SM THERMAL TRANSFER PRINTER USER S MANUAL

TT230SM THERMAL TRANSFER PRINTER USER S MANUAL TT230SM THERMAL TRANSFER PRINTER USER S MANUAL Operations Overview Unpacking and Inspection This printer has been specially packaged to withstand damage during shipping. Please carefully inspect the packaging

More information

Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc.

Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc. Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc. Version 1.06 Metertech Inc. provides this publication as is without warranty of any kind, either express or implied, including, but

More information

Inventory Field Reference Guide

Inventory Field Reference Guide DATA PLUS HOSPITALITY SOLUTIONS Inventory Field Reference Guide Version 8.0 Table of Contents Overview...4 Inventory...5 File Maintenance...6 File Maintenance/Lists...6 Parameters... 7 Issue/Adjustment

More information

Sartorius Biohit Picus Electronic Pipettes

Sartorius Biohit Picus Electronic Pipettes Sartorius Biohit Picus Electronic Pipettes Enjoy pipetting Achieve better results Learn to use it in a minute Description Picus electronic pipette is intended to be used in liquid handling applications

More information

HC100 Troubleshooting Worksheet

HC100 Troubleshooting Worksheet HC100 Troubleshooting Worksheet Printer Lights Bottom* Cartridge Side Problem Solution Media Out Replace with new cartridge No wristband under mouth of cartridge No wristband on roll Wristband broken off

More information

QC-PRO Gage Management

QC-PRO Gage Management QC-PRO Gage Management User Guide www.pister.com Version 9.1 User Guide Revision 1.5 Table of Contents Introduction 1-1 Overview 1-1 Set Up Information 1-1 Skip Holidays 1-1 Calibration Reports 1-2 Measurement

More information

LASER PRINTER. Operation Manual (for copier) Be sure to become thoroughly familiar with this manual to gain the maximum benefit from the product.

LASER PRINTER. Operation Manual (for copier) Be sure to become thoroughly familiar with this manual to gain the maximum benefit from the product. LASER PRINTER Operation Manual (for copier) Be sure to become thoroughly familiar with this manual to gain the maximum benefit from the product. Before installing this product, be sure to read the installation

More information

Automated Analytical Syringe User Manual

Automated Analytical Syringe User Manual XR Every One an Expert Automated Analytical Syringe User Manual WARNING Only trained laboratory personnel that have read this manual and the warnings and cautions contained in it should operate the evol.

More information

Quick Start Guide Axxin T8-ISO. Instrument. In association with: E: W: D Version 1.5

Quick Start Guide Axxin T8-ISO. Instrument. In association with: E: W:   D Version 1.5 Quick Start Guide Axxin T8-ISO Instrument In association with: E: support@axxin.com W: D006093 Version 1.5 Office DOCUMENT CONTROL DETAILS TITLE DOC No. CUSTOMER CONTACT QSG, Quick Start Guide, Axxin T8-ISO

More information

MAKE THE CHANGE FOR BETTER NEONATAL SCREENING. Introducing the new AutoDELFIA immunoassay system for newborn screening

MAKE THE CHANGE FOR BETTER NEONATAL SCREENING. Introducing the new AutoDELFIA immunoassay system for newborn screening MAKE THE CHANGE FOR BETTER NEONATAL SCREENING Introducing the new AutoDELFIA immunoassay system for newborn screening THE WORLD S LEADING SCREENING PLATFORM PerkinElmer s AutoDELFIA immunoassay system

More information

KYOCERA Net Viewer User Guide

KYOCERA Net Viewer User Guide KYOCERA Net Viewer User Guide Legal Notes Unauthorized reproduction of all or part of this guide is prohibited. The information in this guide is subject to change without notice. We cannot be held liable

More information

DC100 Digital Dispenser Operating Manual

DC100 Digital Dispenser Operating Manual DC100 Digital Dispenser Operating Manual 2015 Fisnar Phone: (262)253-8600 E-mail: info@fisnar.com Table of Contents Product Safety Statement 4 Specifications 6 Accessories 6 External Controls 7 Machine

More information

VWR Protocole Builder Software User manual Summary

VWR Protocole Builder Software User manual Summary VWR Protocole Builder Software User manual Summary 1. Introduction... 2 1.1. What is VWR Protocole Builder Software?... 2 1.2. Passwords and User information... 2 1.3. Forgotten password... 2 2. Home screen...

More information

Chromatography Software Training Materials. Contents

Chromatography Software Training Materials. Contents Chromatography Software Training Materials This document contains information on how to build a method, start the instrument to acquire data, and then process the data using the Galaxie Program. You will

More information

MONARCH 9416 XL QUICK REFERENCE

MONARCH 9416 XL QUICK REFERENCE MONARCH 9416 XL QUICK REFERENCE This Quick Reference contains ribbon loading, supply loading, and general care, maintenance, and troubleshooting procedures for the 9416 XL Thermal Direct and 9416 XL Thermal

More information

Tecan crystallisation robot user guide

Tecan crystallisation robot user guide Tecan crystallisation robot user guide The TECAN crystallisation robot in RRB S9 provides an automated facility for setting up crystallisation trays. This user guide is meant as online support for trained

More information

OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide

OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide 2004-2009, AFL Telecommunications, all rights reserved. COM4-00-1001 Revision E, 2009-06-16 Specifications

More information

CSS1 Contractor Series Light Sources and CSM1 Contractor Series Optical Power Meters User s Guide

CSS1 Contractor Series Light Sources and CSM1 Contractor Series Optical Power Meters User s Guide CSS1 Contractor Series Light Sources and CSM1 Contractor Series Optical Power Meters User s Guide www.aflglobal.com or (800) 321-5298, (603) 528-7780 Limited Warranty All NOYES test equipment products

More information

Ascribe Pharmacy System User Guide

Ascribe Pharmacy System User Guide Ascribe Pharmacy System User Guide i Ascribe Pharmacy System User Guide Contents Introduction... 1 Terminology... 1 Navigation... 1 Hot keys... 2 Required fields... 2 Resetting a forgotten password...

More information

SymBiot I Sample Workstation

SymBiot I Sample Workstation SymBiot I Sample Workstation Getting Started Guide DRAFT August 28, 2001 8:42 pm Symbfmgs.fm5 Copyright 1997, 2001, Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic

More information