1 Instructions for Use US IVD Bacterial Test Standard Mass calibration standard containing a typical Escherichia coli DH5 alpha peptide and protein profile plus additional proteins. For quality control and validation of MALDI Biotyper CA Systems Language: US-en Revision C (July 2015)
2 1 Intended Use Product Description Inspection, Storage and Stability Inspection upon Receipt Storage on Arrival Accelerated and Shipping Stability Study of BTS Real-Time Stability Study of BTS Storage after Reconstitution and Aliquoting In-Use Stability Study of reconstituted US IVD BTS US IVD BTS Sample Preparation Procedure Preparation of US IVD BTS Aliquots Inoculation of US IVD BTS Sample on a US IVD 48 Spot Target Manufacturer Intended Use US IVD Bacterial Test Standard (BTS) is an in-vitro- diagnostic product used for quality control and validation of Bruker microflex LT/SH mass spectrometers when used for the identification of microorganisms isolated from human specimens. US IVD BTS must be used when processing test organisms for identification on the MALDI Biotyper CA System. Caution: Federal Law restricts this device to sale by or on the order of a licensed healthcare practitioner. 2 Product Description US IVD BTS contains a manufactured extract of Escherichia coli DH5 alpha that demonstrates a characteristic peptide and protein profile mass spectrum when tested on the MALDI Biotyper CA System. US IVD BTS is spiked with two additional proteins that extend the upper boundary of the mass range of US IVD BTS. The overall mass range covered by US IVD BTS is 3.6 to 17 kda. Each tube contains enough material for the inoculation of approximately 40 US IVD 48 Spot Target positions. US IVD BTS is an in-vitro diagnostic device designed to be used only with the Bruker MALDI Biotyper CA System. Page 2 of 8 US IVD Bacterial Test Standard Instructions for Use Revision C
3 3 Inspection, Storage and Stability 3.1 Inspection upon Receipt Check the US IVD BTS package upon receipt. If the tubes are damaged, do not use. 3.2 Storage on Arrival 18 C US IVD BTS is shipped at ambient temperature but must be stored at 18 C or below immediately on arrival. The expiry date on the package is valid for the enclosed US IVD BTS when stored at 18 C or below on arrival Accelerated and Shipping Stability Study of BTS During this study, the following three (3) lots of BTS material were used: Lot Number Manufacture Date Feb Feb Feb2012 Shipping was simulated by storing the reagents in a climate controlled chamber at 37±2 C for two (2) weeks while an accelerated stability study was conducted at the same temperature for one (1) week. All testing was conducted in duplicate. For each time interval tested (one (1), two (2) and three (3) weeks), BTS was removed from the climate chamber, allowed to acclimate to room temperature and reconstituted in accordance with the instructions for use. Two targets were then inoculated with the BTS solution on 16 target plate positions and three (3) cross-joint positions [BTS_0, BTS_1 and BTS_4]. Testing then proceeded in accordance with the instructions for use. Revision C US IVD Bacterial Test Standard Instructions for Use Page 3 of 8
4 Results All lots met all performance specifications and testing demonstrated marginal standard deviations. Table: Accelerated/Shipping Stability Results Test Condition Lot# , Lot# , Lot# , Lot# , Lot# , Lot# , No Aging ± ± ± ± ± ± week ± ± ± ± ± ± weeks ± ± ± ± ± ± weeks ± ± ± ± ± ± Real-Time Stability Study of BTS During this study, the following three (3) lots of BTS material were used: Lot Number Manufacture Date Jul Jun Jun2011 All three (3) lots of material were maintained at recommended storage conditions of 18 C. At designated time intervals [3, 6, 9, and 12 months], BTS was removed from the freezer, reconstituted, and tested utilizing the same testing process as described in the accelerated/shipping stability study Results Long-term stability testing is ongoing; however, data received to date confirms product stability for up to 12 months. Table: Real-Time Stability Results Test Condition Lot# , Lot# , Lot# , Lot# , Lot# , Lot# , No Aging ± ± ± ± ± ± months ± ± ± ± ± ± months ± ± ± ± ± ± months ± ± ± ± ± ± months ± ± ± ± ± ± Page 4 of 8 US IVD Bacterial Test Standard Instructions for Use Revision C
5 3.3 Storage after Reconstitution and Aliquoting US IVD BTS must be reconstituted before use. We strongly recommend that reconstituted US IVD BTS is aliquoted and frozen using 0.5 ml screw cap micro tubes with screw caps. 18 C Frozen, reconstituted US IVD BTS can be stored for up to 5 months at 18 C or below In-Use Stability Study of reconstituted US IVD BTS Testing for in-use (reconstituted) stability was done using four vials of a single lot of BTS reagent [ , manufacture date: 27 Sep 2011]. Four BTS vials were reconstituted in accordance with the instructions for use and two pools, each containing the contents of two vials, were created. Tests were carried out on eight (8) target plate positions and one (1) cross-joint position immediately after reconstitution in accordance with the instructions for use. The pooled BTS solutions were then placed into the freezer and tested at the following time intervals: 1, 2, 3, 4, 5 and 6 months. At each time point, both (pooled) BTS vials were brought to room temperature and tested as described above Results Reconstituted BTS was tested over a six (6) month period and testing confirmed sample stability at 18 C or below after five (5) freeze-thaw cycles. Because there was one BTS target spot at the six (6) month time point that yielded a <2.2, reconstituted stability is limited to five (5) months. Table: In-Use (Reconstituted) Stability Results Aging # of MBT ID, < 2.2 # of MBT ID, < 2.2 No Aging ± / ± /8 1 month ± / ± /8 2 months ± / ± /8 3 months ± / ± /8 4 months ± / ± /8 5 months ± / ± /8 Revision C US IVD Bacterial Test Standard Instructions for Use Page 5 of 8
6 4 US IVD BTS Sample Preparation Procedure Note Poor sample preparation will impact product performance and will result in low resolution and poor reproducibility. The generation of ions through MALDI depends on the presence of an optimal ratio of matrix substance to analyte. For best results, use all materials as directed and use only chemicals recommended for use by Bruker (for example, CHROMASOLV LC-MS solvents). Chemicals and Materials Required Standard solvent (acetonitrile 50%, water 47.5% and trifluoroacetic acid 2.5%) from Sigma- Aldrich, which has been tested by Bruker and is recommended for reconstituting US IVD BTS. US IVD 48 Spot Target (# ). US IVD HCCA portioned (# ) Equipment and Tools Required but not Provided Centrifuge, pipettes, pipette tips Screw cap micro tubes (Sarstedt, # ) with screw caps (Sarstedt, # ). Page 6 of 8 US IVD Bacterial Test Standard Instructions for Use Revision C
7 4.1 Preparation of US IVD BTS Aliquots 1. Add 50 µl of standard solvent to the US IVD BTS pellet and dissolve by pipetting up and down at least 20 times at room temperature. Avoid foaming of the solution! 2. Incubate the US IVD BTS solution for 5 minutes at room temperature and then mix by pipetting up and down at least 20 times. 3. Centrifuge at 13,000 rpm for 2 minutes at room temperature. 4. Pipette 5 µl aliquots of the supernatant into screw cap micro tubes and close tubes tightly. 5. Store the aliquots at 18 C or below. 4.2 Inoculation of US IVD BTS Sample on a US IVD 48 Spot Target Note Frozen US IVD BTS aliquots should be thawed at room temperature and mixed by flicking the screw cap micro tube with a finger. Screw cap micro tubes should be closed and stored at 18 C or below immediately after preparation of US IVD BTS sample spots. Each 5 µl aliquot contains enough material for at least four sample spots. US IVD BTS must be inoculated onto each US IVD 48 Spot Target before processing. 1. Select two positions on a US IVD 48 Spot Target (cross-joint positions are recommended). Pipette 1 µl of US IVD BTS solution onto each position. 2. Allow the US IVD BTS sample spots to dry at room temperature. 3. After drying, overlay each US IVD BTS sample spot with 1 µl US IVD HCCA portioned solution. Note After the sample has dried, US IVD HCCA portioned must be added within 30 minutes. If US IVD HCCA portioned matrix solution is not applied within 30 minutes after the sample has dried, the MALDI target plate must be cleaned and the inoculation of US IVD BTS must be repeated. 4. Allow the US IVD BTS sample spots to dry at room temperature. Note: If US IVD BTS does not meet all required performance specifications, the test run will be invalid. If US IVD BTS is not inoculated onto a target before processing, the test run will be invalid. Revision C US IVD Bacterial Test Standard Instructions for Use Page 7 of 8
8 5 Manufacturer Bruker Daltonics Inc 40 Manning Road, Manning Park Billerica, MA USA Support Phone: +1 (877) access code 724 Fax: +1 (978) Internet: Sales Information Phone: +1 (978) Web: Hours: Monday - Friday: 5 am - 12 midnight EST Weekends/Holidays: 9 am - 9 pm EST For consumables orders, PO questions and Service Contract quotes, please call +1 (877) during business hours. Ordering Information Product Part Number US IVD Bacterial Test Standard # # Descriptions and specifications supersede all previous information and are subject to change without notice. Copyright 2015 Bruker Daltonics Inc. Page 8 of 8 US IVD Bacterial Test Standard Instructions for Use Revision C
MALDI Biotyper Troubleshooting Guide
Serial Number: MALDI Biotyper Troubleshooting Guide 605510 Revision B Calibration/Validation Issues If failed Validation, run Manual Calibration using flexcontrol software in RUO mode (shown in image below)
More informationMTP SmallAnchor Targets BC
Instructions for Use MTP SmallAnchor Targets BC Polished steel targets with hydrophilic sample and calibrant spots surrounded by a hydrophobic barrier for MALDI sample preparation CARE products are designed
More informationProtein Quantification Kit (BCA Assay)
Protein Quantification Kit (BCA Assay) Booklet Item NO. KTD3001 Product Name Protein Quantification Kit (BCA Assay) ATTENTION For laboratory research use only. Not for clinical or diagnostic use Version
More informationVITROS TDM Performance Verifier is prepared from bovine serum to which therapeutic drugs, inorganic salts, and preservatives have been added.
VITROS Chemistry Products TDM Performance Verifier I, II, and III TDM Performance Verifier I, II, and III 896 2540 179 2357 818 2172 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry
More informationQUICKStart Guide APT Versions 3.0 through 3.3
QUICKStart Guide APT Versions 3.0 through 3.3 PCS QUICKStart Guide Powering On & Off...2-3 Seiko Printer...4-5 Consumable Components...6-7 Pipette Calibration Setup...8-11 Pipette Calibration...12-13 Completing
More informationPROTOCOL: GCP MS Analysis (MSA)
PROTOCOL: GCP MS Analysis (MSA) Purpose Analyze and relatively quantify post translational modifications on core histone H3 with a fully targeted mass spectrometry assay. This protocol should be started
More informationVial block. Power switch
Powering On & Off The color-coded bars within the text throughout this Guide indicate the sections where specific instructions are given. Vial block Power switch Powering On Connect system according to
More informationab BCA protein assay kit reducing agent compatible (test tube)
ab207004 BCA protein assay kit reducing agent compatible (test tube) Instructions for use: For measuring total protein concentration of pure proteins, extracts or lysates in the presence of reducing agents.
More informationBCA protein assay kit for low concentrations. For measuring total protein concentration of pure proteins, extracts or lysates.
ab207002 BCA protein assay kit for low concentrations Instructions for use: For measuring total protein concentration of pure proteins, extracts or lysates. This product is for research use only and is
More informationInsulin Autoantibody Assays
Insulin Autoantibody Assays 24 May 2011 Page 1 of 13 Contents Initial dilution of Insulin Label.. 3 IAA Plating.....4 ciaa Plating... 5 Diluting IAA Label........6 Diluting ciaa Label...... 7 Washing PAS,
More informationBioRobot 9604 Operating Procedure
BioRobot 9604 Operating Procedure Note: An import table (directions given in number 20) must be made prior to a robot run. 1. Prior to operating the BioRobot 9604, perform all daily and weekly maintenance
More informationMTP TLC Adapter. Instructions for Use. Product description
CARE products are designed to support our worldwide customers with high-quality consumables, accessories and dedicated kits. The CARE product range is specifically optimized and certified for use with
More informationThermo Scientific Samco Transfer Pipettes. Precisely what you need
Thermo Scientific Samco Transfer Pipettes Precisely what you need Yesterday, today Thermo Scientific Samco transfer pipettes have been the industry leaders for over 35 years Design Grips easily with latex
More informationThermo Scientific Samco Transfer Pipettes. Precisely what you need
Thermo Scientific Samco Transfer Pipettes Precisely what you need Application Guide LIFE SCIENCE APPLICATIONS INDUSTRIAL, RESEARCH AND TESTING CLINICAL APPLICATIONS Top 9 Pipettes: 6, 3, 5, 20, 23, 24,
More informationPerkinElmer Life and Analytical Sciences, Inc. USING THE ALPHASCREEN OMNIBEADS CATALOG NUMBERS: D, M, R
PerkinElmer Life and Analytical Sciences, Inc. USING THE ALPHASCREEN OMNIBEADS CATALOG NUMBERS: 6760626D, 6760626M, 6760626R For Laboratory Use Only Research Chemicals for Research Purposes Only 2 Precautions
More informationBruker BioSpin FLOW NMR. Racks & Accessories Catalog. Version. NMR Spectroscopy. think forward
Bruker BioSpin FLOW NMR Racks & Accessories Catalog Version 004 think forward NMR Spectroscopy Copyright by Bruker BioSpin GmbH All rights reserved. No part of this publication may be reproduced, stored
More informationIssued by: LABORATORY MANAGER Original Date: March 14, 2001 Approved by: Laboratory Director Revision Date: June 6, 2003.
Policy # MI/SER/06/v01 Page 1 of 5 Section: Subject Title: Epstein Barr Virus Serology Issued by: LABORATORY MANAGER Original Date: March 14, 2001 Approved by: Laboratory Director Revision Date: June 6,
More informationCustomer Responsibilities
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationMicro SAL. Small Animal Saliva Collection System. Catalog Number MRSAL-402
Micro SAL Small Animal Saliva Collection System Catalog Number MRSAL-402 Intended Use The Micro SAL Small Animal Saliva Collection Device is intended for the controlled, standardized collection and transportation
More informationCopyright Black Box Corporation. All rights reserved.
Copyright 1999. Black Box Corporation. All rights reserved. 1000 Park Drive Lawrence, PA 15055-1018 724-746-5500 Fax 724-746-0746 JUNE 1999 FT990A FT991 FT992 FT993 Fiber Optic Inspection Scope CUSTOMER
More informationSPS3 autosampler. Innovation with Integrity. For the Bruker 810/820-MS. Mass Spectrometry
SPS3 autosampler For the Bruker 810/820-MS Innovation with Integrity Mass Spectrometry SPS3 autosampler The SPS3 Sample Preparation System, Bruker s fastest-ever spectroscopy autosampler, meets the diverse
More informationA 2 O Advanced Automated Osmometer
A 2 O Advanced Automated Osmometer A fully automated, multi-sample osmometer that sets the new benchmark for analytical performance, ease of use, and true walkaway operation. www.aicompanies.com Discover
More informationReconstitution Application Guide
Reconstitution Application Guide For Research Use Only. Not for use in diagnostic procedures. This guide contains the following topics: About this guide on page 2 App description on page 3 Before you start
More informationFULLY AUTOMATED 2 MICROPLATE ANALYZER
FULLY AUTOMATED 2 MICROPLATE ANALYZER The new Personal Lab,(Plab) is an integration between our unique extensive expertise and information gathered from thousands of installations. The new Plab is innovative
More informationSkyla VB1 Quick Reference Guide
Skyla VB1 Quick Reference Guide 1 VB1 DM/JAC 2014 v1.0 Quick Reference Guide Contents External features of the analyser 3 VB1 Function Buttons 4 Testing a Sample 5 How to reconstitute a biochemistry control
More informationNuclease-free water (e.g. ThermoFisher, cat # AM9937) Freshly-prepared 70% ethanol in nucleasefree water (optional)
Before start checklist Materials Consumables Equipment Rapid Barcoding Sequencing Kit (SQK- RBK004) 1.5 ml Eppendorf DNA LoBind tubes Ice bucket with ice Flow Cell Priming Kit (EXP-FLP001) 0.2 ml thin-walled
More informationRIDASCREEN Aflatoxin M1
R-Biopharm AG Product information Reliable Extensively tested, excellent performance Sensitive Low limit of detection (5 ppt for milk) Easy All reagents ready to use Automatable Options available R-Biopharm
More informationThe Endosafe MCS Multi Cartridge System
The Endosafe MCS Multi Cartridge System USER S GUIDE 1023 Wappoo Road Suite 43-B, Charleston, SC 29407 Phone: 843.766.7575 Fax: 843.766.7576 www.criver.com endo-comments@crl.com Table of Contents Endosafe
More informationHow to Place an Order
How to Place an Order In this document, you will find information or step-by-step directions on the following topics: Add Items to Your Cart... 2 Add Programs, Publications, and Learning Items to Your
More informationA 2 O Advanced Automated Osmometer
A 2 O Advanced Automated Osmometer A fully automated, multi-sample osmometer that sets the new benchmark for analytical performance, ease of use, and true walkaway operation. www.aicompanies.com/a2o The
More informationPolicy # MI_SER_GAL Department of Microbiology. Page Quality Manual TABLE OF CONTENTS
Department of Microbiology Version: 1.0 CURRENT 1 of 23 Prepared by QA Committee Issued by: Laboratory Manager Revision Date: 3/26/2018 Approved by Laboratory Director: Annual Review Date: 9/1/2018 Microbiologist-in-Chief
More informationA 2 O Advanced Automated Osmometer
A 2 O Advanced Automated Osmometer A fully automated, multi-sample osmometer that sets the new benchmark for analytical performance, ease of use, and true walkaway operation. www.aicompanies.com/a2o Discover
More informationAgilent G1978B Multimode Source for 6100 Series Single Quad LC/MS
Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide Agilent Technologies Notices Agilent Technologies, Inc. 2008 No part of this manual may be reproduced in any form or by any
More informationCryoClear. Cryogenic Vials. Innovation. for the long-term storage of your precious samples.
CryoClear Cryogenic Vials Innovation for the long-term storage of your precious samples. CryoClear Cryogenic Vials Innovation to protect your precious samples. Globe Scientific s CryoClear vials are designed
More informationNCSBI Forensic Biology Section DNA SOP Effective Date: January 30, 2007 Title: ABI PRISM 7000: DNA Quantitation Revision 02
1. Preparing the DNA Standards 1.1. Label eight microcentrifuge tubes to be used for dilution of standards (i.e. Std. 1, Std. 2, Std. 3, etc.) 1.2. Add required amount of TE/glycogen to each tube (See
More informationUser Guide: Illumina sequencing technologies Sequence-ready libraries
DECEMBER 17, 2018 MASSIVELY PARALLEL SEQUENCING SERVICES McGill University and Génome Québec Innovation Centre User Guide: Illumina sequencing technologies Sequence-ready libraries Version 6.0 Copyright
More informationURGENT: Hospira, Inc. Issues Voluntary Recall of Ketorolac Tromethamine Inj., USP
July 1, 2015 URGENT: Issues Voluntary Recall of Dear ASD Healthcare Customer: is voluntarily recalling multiple product lists and lots of. This recall action is being initiated due to the potential for
More informationHow to Place an Order
How to Place an Order In this document, you will find information or step-by-step directions on the following topics: Add Items to Your Cart... 2 Add Programs, Publications, and Learning Items to Your
More informationSymBiot I Sample Workstation
SymBiot I Sample Workstation Getting Started Guide DRAFT August 28, 2001 8:42 pm Symbfmgs.fm5 Copyright 1997, 2001, Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic
More informationTroubleshooting Guide
DS-11 FX Series / DS-C / QFX Technical Note 165 Troubleshooting Guide Rev January 2017 Table of Contents 1. Sample Result Concerns... 4 Microvolume Sample Concentration Accuracy... 4 Microvolume Sample
More informationDRI ph-detect APPLICATION BECKMAN COULTER AU640, AU680
Microgenics Corporation Part of Thermo Fisher Scientific DRI ph-detect APPLICATION BECKMAN COULTER AU640, AU680 Catalog No. 100054 Intended for the determination of urine ph. For In Vitro Diagnostic Use
More informationThermo. EASY-nLC User Guide for LC Devices
Thermo EASY-nLC 1200 User Guide for LC Devices 60053-97273 Revision A September 2015 2015 Thermo Fisher Scientific Inc. All rights reserved. AFC, EASY-nLC, EASY-Column, EASY-Spray, Foundation, LC TriPlus,
More informationMINISCAN INSTRUCTIONS
MINISCAN INSTRUCTIONS DR VIRGINIA GORDON Version 3, July 2011 General Hints on Running Mini-Scan Tests... 1 Total Glycerin Test (TG)... 3 GENERAL HINTS ON RUNNING MINI-SCAN TESTS PIPETTES To use the pipettes
More informationMagnetic Immunoassay Reader Operator Manual
Page 1 of 8 Magnetic Immunoassay Reader Operator Manual Table of contents 1 Introduction...1 2 Magnetic Immunoassay Reader...1 2.1 Contact information and spare parts...1 3 Contents of delivery...1 4 Technical
More informationMachine Starters Kit
Machine Starters Kit Spitfire 65/90 Extreme Dear Customer, Thank you for buying a Mutoh Spitfire 65/90 Extreme series Printer. This kit contains some tools to make your printer operative. Please refer
More information(11-nor-9-carboxy- 9-THC) PCP Phencyclidine (Phencyclidine)
Profile-V Reader System Quick Reference Instructions Before performing the test, refer to the PROFILE -V MEDTOXScan Drugs of Abuse Test System Insert and MEDTOXScan User Manual for complete operating instructions
More informationMaintenance and Troubleshooting Guide
Applied Biosystems 7900HT Fast Real-Time PCR System Maintenance and Troubleshooting Guide Performing Calibration and Verification Runs Maintaining the 7900HT Instrument Hardware Maintaining the Computer
More informationFoot switch. Installation, operation and maintenance manual
Installation, operation and maintenance manual ST Foot switch Read this manual before installing, operating or maintaining this actuator. Failure to follow safety precautions and instructions could cause
More informationCA846 THERMO-HYGROMETER ENGLISH. User Manual. 99 Washington Street Melrose, MA Phone Toll Free
THERMO-HYGROMETER CA846 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com ENGLISH User Manual Statement of Compliance Chauvin Arnoux,
More informationINSTRUCTION MANUAL HBPH DIFFERENTIAL 2-WIRE (4-20 ma) ph Sensor
INSTRUCTION MANUAL HBPH DIFFERENTIAL 2-WIRE (4-20 ma) ph Sensor TABLE OF CONTENTS 1. GENERAL INFORMATION 2 2. SPECIFICATIONS 2 3. INSTALLATION 3 3.1. General Instructions 3 3.2. HBPH 4-20 ma Sensor 3 4.
More informationeprep Sample Preparation Workstation
eprep Sample Preparation Workstation Robotic Sample Preparation for the Chromatography Laboratory Easy, precise and affordable workflow automation Area (mau*min) Highly Flexible and Adaptable The eprep
More informationDRI ETHYL ALCOHOL ASSAY APPLICATION Beckman Coulter AU400, AU480, AU640, AU680, AU2700, AU5400, AU5800
DRI ETHYL ALCOHOL ASSAY APPLICATION Beckman Coulter AU400, AU480, AU640, AU680, AU2700, AU5400, AU5800 Catalog No. 0037, 0038 Intended for the quantitative determination of alcohol in human urine, serum
More informationURGENT: Cubist Pharmaceuticals issues Voluntary Nationwide Recall of Certain Lots of Cubicin (daptomycin for injection) 500 mg
3101 Gaylord Parkway Frisco, Texas 75034 P: 1.800.746.6273 www.asdhealthcare.com August 7, 2014 URGENT: Cubist Pharmaceuticals issues Voluntary Nationwide Recall of Certain Lots of Cubicin (daptomycin
More informationSOP:11:VL:023:02:NIBT PAGE 1 of 9
SOP:11:VL:023:02:NIBT PAGE 1 of 9 Northern Ireland Blood Transfusion Service STANDARD OPERATING PROCEDURE (Operational Copy) Document Details Document Number: SOP:11:VL:023:02:NIBT Supersedes Number: SOP:11:VL:023:01:NIBT
More informationDeFelsko PosiTector IRT Infrared Thermometer Probes
Management Procedure 2572 Revision: A Date Issued: April 12, 2018 Date Revised: Calibration Procedure DeFelsko PosiTector IRT Infrared Thermometer Probes Table of Contents 1 Introduction and UUC Performance
More informationTDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT. Pool Professional s Meter
TDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT Pool Professional s Meter TRACER EC/TDS/SAL POCKETESTER TM CODE 1749-KIT TABLE OF CONTENTS Introduction... 4 Specifications... 4 Contents... 5 Parts &
More informationTG Autoantibody Assays
TG Autoantibody Assays 24 May 2011 Page 1 of 14 Contents Making Label.....3 Plating.....6 Diluting Label........7 Washing and adding PAS/IgA-Agarose...9 Washing Plates and Transferring 10 Getting Counts
More informationAnaPrep 12. Instruction Manual. BioChain Institute Inc. Automated Nucleic Acid Preparation System
BioChain Institute Inc. Instruction Manual Automated Nucleic Acid Preparation System AnaPrep 12 Catalog No.: Z1321001 Rev. Date: April 30, 2015 Manual No. Z1001 Version: Z1321001UA (MV1.14) For General
More informationRadiometer AQT90 FLEX Troponin I ERROR CODES
Radiometer AQT90 FLEX Troponin I ERROR CODES The meter automatically performs electronic self-tests, i.e. environmental check ensures that the analyser is ready for testing and that consumables have not
More informationGPS Explorer Software For Protein Identification Using the Applied Biosystems 4700 Proteomics Analyzer
GPS Explorer Software For Protein Identification Using the Applied Biosystems 4700 Proteomics Analyzer Getting Started Guide GPS Explorer Software For Protein Identification Using the Applied Biosystems
More informationINVITATION TO BID. (This is not an order.)
Page 1 of 9 SCHEDULE OF EVENTS Event Description 1 / 1 Roche Diagnostics POC Chemstrip Urine Test Strips 2.00000 / Event Date #: BC417109; For Urinalysis; Product Type: Chemstrip 9; Detectable Analytes:
More informationExpansion Port SWITCH USER MANUAL
Expansion Port SWITCH USER MANUAL Used Symbols Danger Information regarding user safety or potential damage to the router. Attention Problems that can arise in specific situations. Information, notice
More informationOperation Manual MODEL 3173/3173R. MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R
Operation Manual MODEL 3173/3173R MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R INITIAL INSPECTION..... 2 GENERAL INTRODUCTION.. 2 USING THE METER. 3 THE DISPLAY....
More informationSTANDARD OPERATING PROCEDURES DIVISON OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA
STANDARD OPERATING PROCEDURES DIVISON OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1139.5 Date Issued: 12/03 Date Revised: 8/16 Page 1 of 5 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: Animal Care
More informationNMR Users Guide Organic Chemistry Laboratory
NMR Users Guide Organic Chemistry Laboratory Introduction The chemistry department is fortunate to have a high field (400 MHz) Nuclear Magnetic Resonance (NMR) spectrometer. You will be using this instrument
More information1/4 Specifications Hidex 300 SL automatic TDCR liquid scintillation counter
1/4 Specification Hidex 300 SL 100-240V AC and 24 V DC. Features Model (425-201) Automatic TDCR Liquid Scintillation Counter General Description Hidex 300SL automatic TDCR liquid scintillation counter
More informationFast Facts. for Operating the scil Vet abc Plus. Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education
Fast Facts for Operating the scil Vet abc Plus Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education SCIL Vet ABC Plus Fast Facts Table of Contents Start Up Cycle pg. 2 Reagents Status
More informationDPP Micro Reader. The DPP Micro Reader is maintenance-free, not configurable by the user and is operated with a single, multi-function button.
Read this User Guide completely before using the product. Storage conditions: Store between -20 to 80 C (-4 to 176 F) DPP Micro Reader For use under Emergency Use Authorization (EUA) only. For In Vitro
More informationTotal WBC count and differential WBC count HemoCue WBC DIFF system
TLM_PRO_029 and Version No. 001 Page 1 of 6 Department: Clinical Sciences Unit: Tropical Laboratory Medicine Total WBC count and differential WBC count HemoCue WBC DIFF system Project/study: Not applicable
More informationDRI CREATININE-DETECT APPLICATION BECKMAN COULTER AU640, AU680
Microgenics Corporation Part of Thermo Fisher Scientific DRI CREATININE-DETECT APPLICATION BECKMAN COULTER AU640, AU680 Catalog No. 1797 Intended for the quantitative determination of creatinine in human
More informationDri-Block Heaters DB-2A DB-2D & DB-2DH DB-2D
Dri-Block Heaters Techne's Dri-block heaters provide a safe, dry, constant temperature source in the laboratory. The units are particularly suitable for microbiology and clinical laboratories for incubation,
More informationIFA AUTOMATION BEYOND HUMAN BOUNDARIES HELMED IFA PROCESSOR
IFA AUTOMATION BEYOND HUMAN BOUNDARIES HELMED IFA PROCESSOR WHAT IS THE HELMED? The HELMED is a fully automated and open immunofluorescence processor capable of performing all IFA steps. It is suitable
More informationOrbital Shaker OS-10. Operating Manual Specifications Certificate
Orbital Shaker OS-10 Operating Manual Specifications Certificate Contents 1. Safety P recautions 2. General Information 3. Getting Started 4. Operation of the Shaker 5. Specifications 6. Technical Maintenance
More informationReagent Transfer User Guide
Reagent Transfer User Guide For Research Use Only. Not for use in diagnostic procedures. This guide contains the following topics: About this guide on page 2 Utility description on page 3 Before you start
More informationToxicology Gas Chromatography-Mass Spectrometry (GC-MS)
Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent Gas Chromatograph interfaced to the Agilent
More informationDNA Shearing with E220 Focused-ultrasonicator.
Quick Guide: DNA Shearing with Focused-ultrasonicator. This Quick Guide provides DNA Shearing protocols when using -130, -50, -15, -500, or minitube and a Covaris Focused-ultrasonicator. Revision History
More informationVIAFLO 96 VIAFLO 384. Microplate Pipetting Easy and Affordable
VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable VITARIS AG Blegistrasse 9 CH-6340 Baar Tel.: 041 769 00 00 Fax: 041 769 00 01 E-mail: info@vitaris.com www.vitaris.com VIAFLO 96 VIAFLO 384
More informationAgilent G1978B Multimode Source for 6410 Triple Quad LC/MS
Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS Set-Up Guide Agilent Technologies Notices Agilent Technologies, Inc. 2008 No part of this manual may be reproduced in any form or by any means
More informationCOBAS INTEGRA 400 plus
COBAS INTEGRA 400 plus COBAS INTEGRA 400 plus is designed to consolidate testing and increase efficiency while reducing total running costs of the laboratory. The COBAS INTEGRA systems extensive assay
More informationInstructions for How to Setup the Robot to Print
Instructions for How to Setup the Robot to Print Introduction This document provides guidelines for setting up a Pat O. Brown arrayer * to print antigens onto glass slides. Pat Brown s website, http://cmgm.stanford.edu/pbrown/mguide/index.html,
More informationConductivity and TDS Meter Pen Style Water Quality Meter
User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Conductivity and TDS Meter Pen Style Water Quality Meter Model EC150
More informationDDS-12DW Benchtop Conductivity Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD
DDS-12DW Benchtop Conductivity Meter Instruction Manual BANTE INSTRUMENTS CO., LTD DDS-12DW Benchtop Conductivity Meter 1 Introduction Thank you for selecting the DDS-12DW benchtop conductivity meter.
More informationPAGE Electrophoresis and Blotting Eco-Line. Electrophoresis and Blotting
PAGE Electrophoresis and Blotting Eco-Line Electrophoresis and Blotting 2 Eco-Line The Eco-Line offers tank-style systems for vertical gel electrophoresis with polyacrylamide gels. Two sizes are available:
More informationDRI OXYCODONE APPLICATION BECKMAN COULTER AU480, AU680, AU5800
Microgenics Corporation Part of Thermo Fisher Scientific DRI OXYCODONE APPLICATION BECKMAN COULTER AU480, AU680, AU5800 Catalog No. 100248, 100249, 10015632 The DRI Oxycodone Enzyme Immunoassay is intended
More informationCEDIA METHADONE APPLICATION Beckman Coulter AU480 & AU680
Microgenics Corporation Part of Thermo Fisher Scientific CEDIA METHADONE APPLICATION Beckman Coulter AU480 & AU680 Catalog No. 100088, 100097, 1730916 Intended for the Qualitative and Semiquantitative
More informationBluelab ph Meter TM Instruction Manual
TM Instruction Manual www.getbluelab.com Table of Contents 1.0 Introduction to Bluelab ph Meter 2 2.0 Preparing the Bluelab ph Meter for use 3 3.0 Calibrating Meter 4 4.0 Measure ph Value 5 5.0 Cleaning
More informationOperation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M
Operation Manual Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS
More informationSupramolecular Approach to Enzyme Sensing on Paper Discs Using Lanthanide Photoluminescence
Supporting Information for Supramolecular Approach to Enzyme Sensing on Paper Discs Using Lanthanide Photoluminescence Tumpa Gorai and Uday Maitra* Department of Organic Chemistry, Indian Institute of
More informationNew Brunswick Products. Operating Manual M S Revision B+
New Brunswick Products Operating Manual Revision B+ The enclosed Operating Manual was originally produced by the New Brunswick Scientific Company, Inc. About Eppendorf Eppendorf is a biotech company which
More informationDiscovery. The Professional s Choice For Analytical Applications! Semi-Micro and Analytical Balances
Discovery Semi-Micro and Analytical Balances The Professional s Choice For Analytical Applications! The OHAUS Discovery series of semi-micro and analytical balances combine unmatched weighing performance
More informationUser guide. The utterly reliable analytical laboratory
User guide The utterly reliable analytical laboratory How to contact us For all enquiries (quotes for analysis, technical advice, container requirements or collection of samples), email us at ask@envlab.co.uk
More informationMil-Spec Night Vision Goggles Kit Gen 3, SHD-3 & XD-4 SKU# TY-N16
Mil-Spec Night Vision Goggles Kit Gen 3, SHD-3 & XD-4 SKU# TY-N16 Directions to use 1 Mil-Spec Night Vision Goggles Kit Gen 3, SHD-3 & XD-4 SKU# TY-N16 1. On-Off, I/R Switch 2. Mounting socket 3. Eyepiece
More informationPI31xx Product Family Reflow Soldering Guidelines
APPLICATION NOTE AN:302 PI31xx Product Family Reflow Soldering Guidelines Chris Swartz Principal Applications Engineer Contents Page Introduction 1 Storage 1 Solder Paste 1 Leaded Applications 1 PCB Considerations
More informationModel T28 Transmitter
Model T28 Transmitter Product Training Presented by: Joe Bradley October 2009 ELECTRO-CHEMICAL DEVICES Model T28 Transmitter Two Wire Transmitter Loop Powered 24 VDC @ 500 Ω Measurement Choices ph ORP
More informationLIGHT SOURCE Quartz-iodine lamp 12V-20W.
SPECIFICATIONS THROUGHPUT Up to 180 tests per hour; Up to 300 tests per hour with ISE unit. REAGENT SYSTEM Rotor with 24 positions for 25 ml bottles and 8 positions for 5 ml bottles. All positions can
More informationDisposable Glassware CO R N I N G A N D P Y R E X D I S P O S A B L E G L A S SWA R E D I G I TA L ST I R R I N G H OT P L AT E S...
Disposable Glassware and Equipment CO R N I N G A N D P Y R E X D I S P O S A B L E WA R E........................... 128 D I G I TA L ST I R R I N G H OT P L AT E S............. 135 Digital Stirring Hot
More informationExpansion Port RS232 USER MANUAL
Expansion Port RS232 USER MANUAL Used Symbols Danger Information regarding user safety or potential damage to the router. Attention Problems that can arise in specific situations. Information, notice Useful
More informationStandard Operating Procedure for the Kaye Validator AVS System
Document #: Revision: A Kaye Validator AVS Standard Operating Version 1.0 This is a version controlled document that is released with Amphenol internal approvals. The Approvals section is provided for
More informationDRI METHADONE METABOLITE ASSAY APPLICATION BECKMAN COULTER AU400, AU480, AU640, AU680, AU2700, AU5400, AU5800
Microgenics Corporation Part of Thermo Fisher Scientific DRI METHADONE METABOLITE ASSAY APPLICATION BECKMAN COULTER AU400, AU480, AU640, AU680, AU2700, AU5400, AU5800 Catalog Nos. 100115 and 100116 Intended
More informationAgilent 2100 Bioanalyzer System & Applications Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More information