2100 Digital Indicator Quick Start Manual For use with Software Versions 2.0 and above

Size: px
Start display at page:

Download "2100 Digital Indicator Quick Start Manual For use with Software Versions 2.0 and above"

Transcription

1 2100 Digital Indicator Quick Start Manual For use with Software Versions 2.0 and above

2 Rinstrum Digital Indicator Quick Start Manual Rev 2.50 Copyright All Rights Reserved. No part of this document may be copied, reproduced, republished, uploaded, posted, transmitted, distributed, stored in or introduced into a retrieval system in any form, or by any means (electronic, mechanical, photocopying, recording or otherwise) whatsoever without prior written permission of Rinstrum Pty Ltd. Disclaimer Rinstrum Pty Ltd reserves the right to make changes to the products contained in this manual in order to improve design, performance or reliability. The information in this manual is believed to be accurate in all respects at the time of publication, but is subject to change without notice. Rinstrum Pty Ltd assumes no responsibility for any errors or omissions and disclaims responsibility for any consequences resulting from the use of the information provided herein. SPECIAL NOTE Trade Use of the Rinstrum 2100 This manual may occasionally make reference to Trade Use settings of the Only properly marked Trade Certified versions of the 2100 can be used in Legal for Trade applications. Trade Certification is only available on 2100 instruments with software Versions 2.0 and above. Some individual settings may not be legal for trade use. Please check regulations with the appropriate Weights and Measures Authority. Everything should be made as simple as possible, but not simpler. - Albert Einstein - Page 2

3 Rinstrum Digital Indicator Quick Start Manual Rev 2.50 Table of Contents 1. INTRODUCTION Approvals Features rin-smart Software Options Manuals SPECIFICATIONS WARNINGS General DC Power Supply Load Cell Signals and Scale Build Configuration Issues INSTALLATION Special Function Key Electrical Safety CONNECTIONS Connecting Shields Cable Shield Connection and Earthing Unused Pins Load Cell Connection FULL SETUP SAFE SETUP OPERATOR MENU SETUP ENABLING SOFTWARE OPTIONS RS-485 COMMUNICATIONS ERROR MESSAGES BATTERY OPERATION DIAGNOSTIC ERRORS...23 Page 3

4 Rinstrum Digital Indicator Quick Start Manual Rev Introduction This manual contains information on the installation, calibration and setup of the Approvals NSC S403 approval (6000 divisions at 1µV/division). NMI TC6033 approval (6000 divisions at 1µV/division). C-tick approved and CE approved Features 27mm alphanumeric LCD display (LED back-lighting). Programmable special function key and three remote inputs. Full setup and calibration. Six wire load cell wiring. Checkweighing, kg/lb switching, totalising, intelligent batching, counting, live weight, hold/peak hold functions. Five point linearity correction. Real Time Clock and Calendar. RAM for storing Zero, Tare and Total settings, etc. Soft Power On/Off for inactivity auto power down rin-smart Software Options 0224 (SERIAL) Enable RS-232 serial communications 0225 (SETP) Enable setpoint output 1.4. Manuals For more information on the 2100 refer to the 2100 Reference Manual and 2100 Operator Manual (available free of charge from Page 4

5 Rinstrum Digital Indicator Quick Start Manual Rev 2.50 Performance Display Backlight Display Resolution Count-by Zero Cancellation Span Adjustment Stability/Drift Operating Environment Digital 2. Specifications Backlit alphanumeric LCD with six 27mm high digits LED backlight with adjustable brightness Up to 30,000 divisions, minimum of 0.25µV/division (Trade 6000 divisions at 1µV/division) 1, 2, 5, 10, 20, 50, 100 (Entered in Displayed Weight) + / 2.0mV/V 0.1mV/V to 3.0mV/V full scale Zero: < 0.1uV/ C, Span < 10ppm/ C, Linearity < 20ppm, Noise < 0.05µVp-p Temperature: 10 to +50 C ambient, Humidity: <90% non-condensing Setup and Calibration Full digital with visual prompting in plain messages Memory Retention Full non-volatile operation Digital Filter Averaging from 1 to 100 consecutive readings Zero Range Adjustable from +/ 2% to +/ 20% of full capacity A/D Converter Type 24bit Sigma Delta Resolution 8,388,608 internal counts A/D Sync Filter Selectable 25/30Hz, FIR filter > 80dB Load Cells Excitation 8 volts for up to 8 x 350 ohm load cells (6-wire + shield) Serial Comms (Software Option 0224) Serial output Single RS-232 as automatic transmit, network or printer drive Power Input Standard General 10 to15vdc (60mA to 400mA depending on load cells and backlight) ON/OFF key with override and Auto-Off software Variants AC AC Power: 110/240VAC 50/60Hz 10VA fitted in s/s housing DC DC Power: 12-24VDC 10VA fitted in s/s housing Battery Rechargeable: 12VDC Battery fitted in s/s housing Dimensions Body size 189mm (L) x 99mm (H) x 23mm (D) Panel cutout Flush mounted with cable holes drilled separately (template provided) Page 5

6 Rinstrum Digital Indicator Quick Start Manual Rev Warnings 3.1. General Instrument not to be subject to shock, excessive vibration or extremes of temperature (before or after installation). Inputs are protected against electrical interference, but excessive levels of electro-magnetic radiation and RFI may affect the accuracy and stability. For full EMC or for RFI immunity, termination of the load cell shield at the 2100 end is important. Instrument and load cell cable are sensitive to excessive electrical noise. Install well away from any power or switching circuits DC Power Supply DC supply need not be regulated provided it is free of excessive electrical noise and sudden transients. Instrument can be operated from high quality plug-pack provided there is sufficient capacity to drive both it and load cells. Use plug packs with a rating of 10VDC to 12VDC with current outputs of 0.5 to 1A Load Cell Signals and Scale Build Very low output scale bases can be used but may induce some instability in weight readings when used with higher resolutions (ie. higher output/lower number of divisions equals greater display stability/accuracy) Configuration Issues Configuration and calibration can be performed from front panel, using digital setup. When Full Setup is used, all menu items are accessible and care must be taken to ensure no accidental changes are made to calibration and trade settings. Enter a passcode to prevent unauthorised or accidental tampering. Page 6

7 Rinstrum Digital Indicator Quick Start Manual Rev Installation The following steps are required to set up the 2100 indicator. Inspect indicator to ensure good condition. Ensure mounting options and connectors are available. Use connection diagrams to wire up load cell, power and auxiliary cables as required. Connectors for all cables are supplied with indicator. Instrument has built in panel mounting screws. Use the "Panel Drilling Template" provided for hole locations. Connect Power to instrument and press <POWER> key (if not overridden) to start instrument. Follow instructions in Full Setup section page 12 to configure and calibrate instrument. Enter passcode to protect settings from tampering. Record passcode for future reference. To turn instrument OFF press and hold <POWER> key for three seconds (until display blanks) Special Function Key The Special Function Key on the 2100 ships as a blank key. If any of the special functions are to be used on the indicator it is important that the matching function key sticker (supplied) is applied to the keypad. Ensure keypad is clean and dry before affixing sticker. Cleaning Keypad: Wipe with a soft cloth slightly dampened with either methylated spirits or warm soapy water Electrical Safety For your protection all mains electrical hardware must be rated for environmental conditions of use. Pluggable equipment must be installed near an easily accessible power socket outlet. To avoid the possibility of electric shock or damage to the instrument, always switch off or isolate the instrument from the power supply before maintenance is carried out. Page 7

8 Rinstrum Digital Indicator Quick Start Manual Rev Connections 5.1. Connecting Shields To obtain full EMC or for RFI immunity with the 2100, the load cell shield MUST be connected electrically to the metal shell of the DB9 connector. Refer to diagrams below or to instructions supplied with the connector. Fold shield wires back over outside of cable insulation so cable clamp of backshell makes good electrical contact with shield when installed. Twist shield wires together and solder ends to DB9 casing. Page 8

9 Rinstrum Digital Indicator Quick Start Manual Rev Cable Shield Connection and Earthing Care should be taken when connecting shields to maximise RFI immunity and minimise earth loops and cross-talk (interference) between instruments. For RFI immunity, termination of the load cell shield at the 2100 end is important (ie. with connection to the 2100 case via the shield connection). The 2100 enclosure is directly connected to the shield connections on the cables. The 2100 should be connected to earth via a single reliable link to avoid earth loops. Where each instrument is separately earthed, interconnecting cable shields should be connected at one end only. Caution: Some load cells connect the cable shield directly to the load cell (and therefore the scale base). Connection of the shield in this situation may be site specific. The instrument complies with relevant EMC standards provided case ground connection is correctly made. Resistance measured between 2100 case and nearest earth point should be less than 2 ohms. If static problems are expected, options 0084/0085 may be required to protect the serial outputs Unused Pins Unused pins are NOT to be connected. Reason: The functions of the pins may not be compatible with equipment at the other end (eg. connecting output pins to a PC communications port may affect the operation of the PC). Page 9

10 Rinstrum Digital Indicator Quick Start Manual Rev Load Cell Connection Wire Connection Note: Sense lines MUST be connected Wire Connection *For more information on shielding refer to page 8. Page 10

11 Rinstrum Digital Indicator Quick Start Manual Rev Auxiliary Connection *For more information on shielding refer to page 8. Note: Do NOT connect unused pins. For more information refer to page 9. Remote Display 2100 Pin Remote Display Plug 3 (TXD) RXD / Receive 5 (GND) GND / Ground Printer 2100 Pin Printer Plug - DB25F 3 (TXD) RXD - Pin 3 5 (GND) GND - Pin 7 4 (DTR) DTR - Pin 20 Direct Computer Link 2100 Pin Computer DB-9F (DB-25F) 2 (RXD) TXD - Pin 3 (Pin 2) 3 (TXD) RXD - Pin 2 (Pin 3) 5 (GND) GND - Pin 5 (Pin 7) Page 11

12 Rinstrum Digital Indicator Quick Start Manual Rev IO Connections Power Warning For 2100 use only VDC Voltages outside this range may cause improper operation or damage. Note: Variants of the 2100 may require different power requirements. Refer to Variant Data Sheet. 6. Full Setup Full Setup provides access to configure and calibrate the instrument. Ensure instrument is On. Press the <ZERO> and <POWER> keys together for two seconds. To save settings, exit and return to normal weighing mode, press the <ZERO> and <POWER> keys together for two seconds or select - End - from the menus. Indicates setting is trade critical and calibration counter will be incremented if setting is changed. Indicates functions are only suitable for remote inputs. Page 12

13 Rinstrum Digital Indicator Quick Start Manual Rev 2.4 GRP ITM SEL EDT OK Underline = Defaults BUILD DP Decimal Point Position , , , , , CAP Maximum Capacity SEL changes position, EDT changes digit. RES Resolution (Count-By) 1, 2, 5, 10, 20, 50, 100 UNITS Units of Measure none, g, kg, lb, t, oz HI.RES Resolution x 10 Mode OFF, ON OPTION USE Industrial or Trade Use TRADE (Trade), INDUST (Industrial) (+ and weighing) FILTER Digital Filtering/Averaging 1, 2, 5, 10, 25, 50, 75, 100 (number of readings) MOTION Motion Detection Setting none, t, t, t (fine), t, t, t, t (coarse) Default = 0.5 graduations per 1.0 second (Time) AUTO.Z Auto Zero on Power Up OFF, ON Z.TRAC Zero Tracking Setting OFF, SLOW (0.5 div/sec), MED (2 div/sec), FAST (10 div/sec) Z.RANGE Zero Key Range 2+2, 1+3, (% of full scale) Z.BAND Zero 'Dead' Band SEL changes position, EDT changes digit. Page 13

14 Rinstrum Digital Indicator Quick Start Manual Rev 2.4 GRP ITM SEL EDT OK CAL ZERO Zero Calibration Routine (Current weight displays) SPAN Span Calibration Routine (Current weight displays) ED.LIN Edit Linearisation Points L Select Linearisation point 1 to 5 (L2, L3, L4, L5). (Approx. % of full scale) CLR.LIN Clear Linearisation Points L Select Linearisation point 1 to 5 (L2, L3, L4, L5). (Approx. % of full scale) FAC.CAL Factory Calibration Cont.N (No) Cont.Y (Yes) Underline = Defaults Remove all weight. OK starts routine (Z.in P displays). ITM key to exit, OK to repeat routine. Add test weight. Set correct weight. SEL changes position, EDT changes digit. OK starts routine (S.in P displays). ITM key to exit, OK to repeat routine. Set capacity of test weight. SEL changes position, EDT changes digit. OK starts routine (L.in P displays). ITM key to exit, OK to repeat routine. OK to clear point or ITM key to exit. Cont.N Warning: Choosing Cont.Y will restore default factory calibration in BUILD and CAL menus. Page 14

15 Rinstrum Digital Indicator Quick Start Manual Rev 2.4 SPEC OP.PC Security Passcode for (0000 no passcode). Operator Menu Setup Set 4 digit passcode (eg. 1234). SEL changes position, EDT changes digit. Activated only when FULL.PC is also set. FULL.PC Full Security Passcode for Digital Setup KEY.LOC Front Panel Key Locking Zero, Tare, Gross/Net, Print KEY.FN Key Functions Position 1: Keypad Function Key Positions 2, 3 & 4: Remote Inputs (0000 no passcode). Set 4 digit passcode (eg. 1234). SEL changes position, EDT changes digit. Z T G P Letter indicates key is unlocked. (-) Dash indicates key is locked. SEL changes position, EDT changes digit. ( no functions assigned). Keypad Function Key: U lb/kg switching, C Count, H Hold, E Peak Hold, L Live Weight, S Show Total, B Batch, (-) No function. Remote Inputs: Z Zero, T Tare, G Gross/Net, P Print, U lb/kg switching, C Count, H Hold, E Peak Hold, L Live Weight, S Show Total, B Batch, K Blank, (-) No function (enable IO pin as output). Front Panel Keys: Z Zero, T Tare, G Gross/Net, P Print B.LIGHT Backlight Operation ON, AUTO, OFF (Automatically turns instrument off after 10 seconds of inactivity) BRIGHT Backlight Brightness 01 to 10 (01=10%, 10=100%) Page 15

16 Rinstrum Digital Indicator Quick Start Manual Rev 2.4 GRP ITM SEL EDT OK Underline = Defaults SPEC ctd. AUT.OFF Automatic Power Off NEVER, 20 S, 30 S, 60 S, 300 S, 600 S (Seconds of inactivity before power down) AUX.DSP Auxiliary Display (Time) OFF, TIME (Recommended for batching Operations) SYNC A/D Frequency 25Hz or 30Hz (This setting may affect calibration.) SERIAL TYPE Serial Output Type OFF, NET, AUTO, PRINT, MASTER ADDR Serial Address 00 to 31 (Default = 01) SEL changes position, EDT changes digit. BAUD Serial Baud Rate 1200, 2400, 4800, 9600, BITS Serial Format Options N (Default Serial Format Options) Position 1: Parity SEL changes position, EDT changes digit. Position 2: Data Bits Parity: N None, O Odd, E Even Position 3: Stop Bits Data Bits: 7 or 8 data bits Position 4: DTR Handshake Stop Bits: 1 or 2 stop bits DTR: (-) DTR disabled or d DTR enabled PRN.ID Printout ID OFF, ON (Adds ADDR ID to printouts) Page 16

17 Rinstrum Digital Indicator Quick Start Manual Rev 2.4 SERIAL PRN.ROW Row Padding on Printout 00.0 to ctd Number of line feeds to add before printout Number of line feeds to add after printout. If padding is set to 10 then a form feed character is sent. PRN.COL Column Padding on Printout 00 to 20 (Number of spaces to add at start of each line of printout.) SET.PTS OPTN A - G O H - - G O H - (Default Setpoint Settings) - none - None disables the setpoint when not in use Active A Active for level control and overload alarms Slow S Slow Fill used in conjunction with Fast Fill Gross G Gross (use gross weight) Net N Net (use net weight, Tares before operating relay) Reading R Reading (use current displayed weight) Held H Held (use Held reading) Over O Over (weight increasing) Under U Under (weight decreasing, negative weighing) High H High (Active High logic) Low L Low (Active Low logic) - none - None (no internal beep during relay operation) Single S Single (one internal beep during operation) Double D Double (double internal beeps during operation) Flash F Flash (Display flashes) TARG A Target A SEL changes position, EDT changes digit. (Primary target in batching) Page 17

18 Rinstrum Digital Indicator Quick Start Manual Rev 2.4 GRP ITM SEL EDT OK SET.PTS ctd. OPTN B - G O H - -none Gross Over High -none Active Net Under Low Single Fast Reading Double Held Flash TARG B Target B (Preliminary value in batching) OPTN C - G O H - -none Gross Over High -none Active Net Under Low Single Dump Reading Double Pass Held Flash Underline = Defaults - G O H - (Default) Same as OPTN A except: F Fast Fill used in conjunction with Slow Fill. SEL changes position, EDT changes digit. - G O H - (Default) Same as OPTN A except: D Dump signal for weight to return to zero or dump time (refer to D.TIME below). P Pass output for setpoint 3 only. (checkweigh application) TARG C Target C (Not used in batching) SEL changes position, EDT changes digit. FLIGHT In-Flight SEL changes position, EDT changes digit. HYS Hysteresis SEL changes position, EDT changes digit. D.TIME Dump Time 00.0 to 20.0 seconds 0 for dump to zero weight, otherwise operated for the time specified. Page 18

19 Rinstrum Digital Indicator Quick Start Manual Rev 2.4 CLOCK TIME Time Setting 24 hour clock Set time in the format HH.MM HH.MM.SS displays SEL changes position, EDT changes digit. DATE Date Setting Set day and month in the format DD.MM DD.MM.YY displays Set the year in the format YYYY QA.OPT Quality Assurance Reminder OFF, ON (Flashes QA Due on due date) QA.DATE Quality Assurance Due Date Set day and month in the format DD.MM Set the year in the format YYYY TEST SCALE Load Cell Test Displays load cell output in mv/v (Format=0.0000) Trade Mode=5 sec display FRC.OUT Force Outputs (Default display) EDT advances through each output (1--, -2-, --3) TST.INP Test Inputs (Default display) EDT advances through each input (1--, -2-, --3) FACTRY DEFLT Restore Factory Defaults Cont.N Cont.N (No) Warning: Choosing Cont.Y will clear all stored Cont.Y (Yes) data except calibration. CONFIG Configure Software Options Used to enter license codes for software options. - END - EXIT SETUP settings and return to normal weighing mode Page 19

20 Rinstrum Digital Indicator Quick Start Manual Rev Safe Setup Safe Setup provides access to configure the instrument without access to trade or calibration specific items. Ensure instrument is On. Press the <GROSS/NET> and <POWER> keys together for two seconds. To save settings, exit and return to normal weighing mode, press the <GROSS/NET> and <POWER> keys together for two seconds or select - End - from the menus. 8. Operator Menu Setup The Operator Menu provides access to setpoint target and flight settings. Ensure instrument is On. Press the <TARE> key for two seconds. To save settings, exit and return to normal weighing mode, press the <TARE> key for two seconds or select - End - from the menus. 9. Enabling Software Options To enable any of the rin-smart Software options a license code must be entered. The license codes are unique to each option and to each instrument. These options may be factory installed or installed in the field. Follow the steps below to discover which options are installed or to install another option: Ensure the instrument is On. Press the <POWER> and <FUNCTION> keys together for two seconds. Page 20

21 Rinstrum Digital Indicator Quick Start Manual Rev 2.4 The 2100 will display the installed options and prompt for a new license code. (eg. SERIAL is displayed if the RS-232 Serial Option is installed). The <GROSS/NET> key changes the position and the <PRINT/M+> key changes the digit. (A code of leaves the instrument unchanged.) Press the <FUNCTION> key when complete. The 2100 returns to normal operation. If successful the new rin-smart Option will be displayed, otherwise the 2100 will sound a beep. Page 21

22 Rinstrum Digital Indicator Quick Start Manual Rev RS-485 Communications The 2100 is compatible with the RI 0080 RS-232 to RS-485 serial converter. There is no need to provide external power to the RI 0080 module as this is provided directly by the Error Messages Error Description (U ) The weight reading is below the normal weighing range. (O ) The weight reading is above the maximum capacity of the equipment. (ZERO) The weight is outside the zero range (ERROR) tolerance setting. See Note below. (STABLE) The scale motion has prevented a zero, (ERROR) tare or print operation from occurring. See (QA) (DUE) Note below. Quality assurance testing is due. Press any key to clear this warning for one hour. Note: The ZERO and STABLE error messages are accompanied by a series of long beeps. The messages repeat until a key is pressed. 12. Battery Operation BAT is flashed on the auxiliary display if the battery voltage falls below 11V. If the battery voltage falls below 10.5V the instrument automatically powers down. Battery checking is only carried out if the Automatic Off (AUT.OFF) option is set to a value other than NEVER. Page 22

23 Rinstrum Digital Indicator Quick Start Manual Rev Diagnostic Errors Error Description Action E00001 Power supply voltage too low. Check supply E00002 Power supply voltage too high. Check scale / cables E00004 E00008 E00010 E00020 Load cell excitation voltage too low. (8 volts for up to 8 x 350 ohm load cells) Load cell excitation voltage too high. (8 volts for up to 8 x 350 ohm load cells) Temperature outside limits. ( 10 to +50 C ambient) Scale build incorrect. (100 to grads). Page 23 Check scale / supply Check scale / supply Check location Fix up scale build E00100 Digital setup information lost. Re-enter setup E00200 Calibration information lost. Re-calibrate E00300 All setup information lost. Enter setup and calibrate E00400 Factory information lost. (FATAL) Service E00800 EEPROM memory chip failed. Service (FATAL) E02000 ADC out of range. Possible load cell or cable damage. Check load cell cable E04000 Battery backed RAM data lost. Re-enter setup E08000 FLASH program memory incorrect. (FATAL) Service The E type error messages are additive. For example, E00005 ( ) would indicate that both Excitation and Power Supply Voltage are low. The numbers add in hexadecimal as follows: A - B - C - D - E - F (For example, = 6, or = C)

24

IP69K Washdown Indicator Quick Start Manual (800)

IP69K Washdown Indicator Quick Start Manual (800) IP69K Washdown Indicator Quick Start Manual (800) 223-4227 www.cas-usa.com U21022050 Table of Contents 1. INTRODUCTION...2 1.1. Approvals...2 1.2. Manuals...2 2. SHIPPING CONTENTS...2 3. SPECIFICATIONS...3

More information

Digital Indicator Quick Start Manual. For use with Software Versions 4.6 and above

Digital Indicator Quick Start Manual. For use with Software Versions 4.6 and above 5000 Digital Indicator Quick Start Manual For use with Software Versions 4.6 and above 5000-601-442 Copyright All Rights Reserved. No part of this document may be copied, reproduced, republished, uploaded,

More information

DFI 250 DIGITAL INDICATOR QUICK START MANUAL

DFI 250 DIGITAL INDICATOR QUICK START MANUAL DFI 250 DIGITAL INDICATOR QUICK START MANUAL www.cooperinstruments.com PH: (540) 349-4746 FAX: (540) 347-4755 CONTENTS 1.0 INTRODUCTION... 1 1.1 Approvals (for trade versions only)...1 1.2 Manuals...1

More information

GSE Model 250. Digital Indicator. Technical Reference Manual Version 1.1

GSE Model 250. Digital Indicator. Technical Reference Manual Version 1.1 GSE Model 250 Digital Indicator Technical Reference Manual Version 1.1 Table of Contents 1. INTRODUCTION...3 1.1. Overview...3 1.2. Approvals (for trade versions only)...3 1.3. The Manuals Set...3 1.4.

More information

R310 & N310 R320 & N320 Digital Indicator

R310 & N310 R320 & N320 Digital Indicator T RINSTRUM R310 & N310 R320 & N320 Digital Indicator Trade 4,000 Divisions Compact ABS Housing Internal Batteries RS-232 Serial Output Special Functions 20 mm LCD Display Bright LED Backlighting NORDIC

More information

5200 (Totaliser/Checkweigher) Digital Indicator Reference Manual For use with Software Versions 1.0 and above

5200 (Totaliser/Checkweigher) Digital Indicator Reference Manual For use with Software Versions 1.0 and above 5200 (Totaliser/Checkweigher) Digital Indicator Reference Manual For use with Software Versions 1.0 and above 0025-603-110 Copyright All Rights Reserved. No part of this document may be copied, reproduced,

More information

Gateway 1400 Reference Manual

Gateway 1400 Reference Manual Profibus-DP Gateway 1400 Reference Manual Copyright All Rights Reserved. No part of this document may be copied, reproduced, republished, uploaded, posted, transmitted, distributed, stored in or introduced

More information

In-Motion Axle Scale (K422) Operator Manual

In-Motion Axle Scale (K422) Operator Manual In-Motion Axle Scale (K422) Operator Manual Copyright All Rights Reserved. No part of this document may be copied, reproduced, republished, uploaded, posted, transmitted, distributed, stored in or introduced

More information

5000 Digital Indicator Reference Manual. For use with Software Versions 4.6 and above

5000 Digital Indicator Reference Manual. For use with Software Versions 4.6 and above 5000 Digital Indicator Reference Manual For use with Software Versions 4.6 and above 5000-600-440 TABLE OF CONTENTS 1. INTRODUCTION... 4 1.1. THE MANUALS...4 1.2. THE 5000...4 2. SPECIFICATIONS... 5 3.

More information

1203 Weight Transmitter Reference Manual For use with Software Versions 1.0 and above

1203 Weight Transmitter Reference Manual For use with Software Versions 1.0 and above 1203 Weight Transmitter Reference Manual For use with Software Versions 1.0 and above 1203-600-180 Copyright All Rights Reserved. No part of this document may be copied, reproduced, republished, uploaded,

More information

N420 R420 R423. series

N420 R420 R423. series N420 R420 R423 series powerful flexible industrial weight controller The 400 series of indicators are designed with both the installer and operator in mind and cover a wide range of applications. They

More information

Weighing Indicator. Features. Description. Applications

Weighing Indicator. Features. Description. Applications Force Pressure Temperature Switch Weighing Indicator Description The compact Indicator of model E1932X is a precision and cost effective product for weighing applications. They are ideal for use in mobile

More information

TRACKER 240 SERIES. Load Cell and Weighing Indicators. A Precision Measurement Instrument with Outstanding Features

TRACKER 240 SERIES. Load Cell and Weighing Indicators. A Precision Measurement Instrument with Outstanding Features TRACKER 240 SERIES Load Cell and Weighing Indicators A Precision Measurement Instrument with Outstanding Features TRACKER 240 SERIES INDICATORS Ratiometric Measurement Tare and Auto Transducer Excitation

More information

400 Series (K404, K405, K422) Digital Indicator Reference Manual

400 Series (K404, K405, K422) Digital Indicator Reference Manual 400 Series (K404, K405, K422) Digital Indicator Reference Manual Copyright All Rights Reserved. No part of this document may be copied, reproduced, republished, uploaded, posted, transmitted, distributed,

More information

D840/D820. Remote Display Installation Manual

D840/D820. Remote Display Installation Manual D840/D820 Remote Display Installation Manual Copyright All Rights Reserved. No part of this document may be copied, reproduced, republished, uploaded, posted, transmitted, distributed, stored in or introduced

More information

The IQ300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications.

The IQ300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. IQ300 Wall Mount Load Cell Indicator Data sheet English 1.01 Introduction The IQ300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. The high

More information

C500 Series Data Sheet

C500 Series Data Sheet C500 Series Data Sheet - 16 x 350Ω load cells - AC and DC versions - Ethernet and USB host and slave standard - 10,000d @ 0.5 µv/d - 8 set points (option) - Analog module with 2 isolated outputs (option)

More information

PS-IN202 Weighing Indicator User Manual

PS-IN202 Weighing Indicator User Manual PS-IN202 Weighing Indicator User Manual Version 2015.05 Contents FRONT AND REAR VIEW OF THE INDICATOR CHAPTER 1 CONNECTIONS 1.1 WHAT'S IN THE BOX 1.2 CONNECTING TO THE WEIGH PLATFORM 1.3 CONNECTION TO

More information

DWP-98 Series. Counting scale. User s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT

DWP-98 Series. Counting scale. User s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT c DWP-98 Series Counting scale User s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT Specifications subject to change without prior notice CONTENTS 1. INSTALLATION

More information

! User Manual Rev. 1

! User Manual Rev. 1 User Manual Rev. 1 Warnings For safety operation of the weighing indicator, please follow the following warning/ safety instructions: Calibration inspection and maintenance of the indicator are prohibited

More information

The LT1300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications.

The LT1300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. LT1300 Wall Mount Load Cell Indicator Datasheet English 1.00 Introduction The LT1300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. The high

More information

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD AD-8923-BCD Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD4002137 2010 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated

More information

QC-3265 Checkweigher User s Manual

QC-3265 Checkweigher User s Manual QC-3265 Checkweigher User s Manual CAUTION Risk of electrical shock. Do not remove cover. No user serviceable parts inside. Refer servicing to qualified service personnel. Weigh-Tronix reserves the right

More information

APEX Junior Cased Weighing Indicator/Controller

APEX Junior Cased Weighing Indicator/Controller APEX Junior Cased Weighing Indicator/Controller Non-Trade User Manual www.pcm-uk.com APPROVED DISTRIBUTORS OF: Electronic Weighing Services 1 Index Section 1: Setting Up (Terminations and time/date routines)

More information

Digital Bench Scale. Revision 1.2 September 14, 2000 Contents subject to change without notice.

Digital Bench Scale. Revision 1.2 September 14, 2000 Contents subject to change without notice. Digital Bench Scale Revision 1.2 September 14, 2000 Contents subject to change without notice. Salter Brecknell Weighing Products 1000 Armstrong Drive Fairmont, MN 56031 Tel (800) 637-0529 Tel (507) 238-8702

More information

The IQ240 panel mount load cell indicator is a precision digital indicator for load cell and strain gauge applications.

The IQ240 panel mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. IQ240 Panel Mount Load Cell Indicator Data sheet English 1.01 Introduction The IQ240 panel mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. The high

More information

USER INSTRUCTION MANUAL FOR LOADCELL TRANSMITTER MODEL TDC/I/0550 (SOFTWARE: VER2A) INDEX

USER INSTRUCTION MANUAL FOR LOADCELL TRANSMITTER MODEL TDC/I/0550 (SOFTWARE: VER2A) INDEX USER INSTRUCTION MANUAL FOR LOADCELL TRANSMITTER MODEL TDC/I/0550 (SOFTWARE: VER2A) INDEX DOCUMENT NO: TDC 0550 MANUAL - 2 1.0) INTRODUCTION. PAGE 2 1.1) ABOUT THIS MANUAL. PAGE 2 1.2) INTRODUCTION. PAGE

More information

Calibration & Connectivity

Calibration & Connectivity Calibration & Connectivity TS-700 Series Digital Indicators This document supplements the User s Guide TS-700 MS TS-700 SS TS-700 WB Full Function/Advanced Function Digital Indicator www.trinerscale.com

More information

EXCELL PRECISION CO., LTD.

EXCELL PRECISION CO., LTD. Excell Precision Limited 24. All rights reserved Worldwide. The information contained herein is the property of Excell Precision Limited and is supplied without liability for errors or omissions. No part

More information

Digital Weight Indicator Setup / Operation Manual

Digital Weight Indicator Setup / Operation Manual Digital Weight Indicator Setup / Operation Manual Load Cell Central follows a policy of continuous improvement and reserves the right to change specifications without notice. 2013 Load Cell Central Toll

More information

The PM1000 series is a universal 4 digit LED plug-on display for transmitters with 4-20mA 2 wire output and fitted with DIN43650 connector.

The PM1000 series is a universal 4 digit LED plug-on display for transmitters with 4-20mA 2 wire output and fitted with DIN43650 connector. PM1000 SERIES PLUG-ON DISPLAY BRIGHT LED DISPLAY INDICATION RANGE -999 TO +9999 FITS TO DIN 43650 CONNECTOR PLUG-ON TO ANY TRANSMITTER WITH 4-20MA OUTPUT EASY TO SCALE ON SITE ROBUST DESIGN SET POINT OPTION

More information

COMPLETE MANUAL. WEIGHING INDICATOR TI-1200 and TI-1200-S TI-1200 with rechargeable battery.

COMPLETE MANUAL. WEIGHING INDICATOR TI-1200 and TI-1200-S TI-1200 with rechargeable battery. COMPLETE MANUAL WEIGHING INDICATOR TI-1200 and TI-1200-S TI-1200 with rechargeable battery. TI-1200-S without rechargeable battery and with a smaller cabinet. Note: When TI-1200 is shipped as a part of

More information

DRAFT. Customer site gravity factor. Filtering parameters. Five points for linearity correction to ensure maximum accuracy

DRAFT. Customer site gravity factor. Filtering parameters. Five points for linearity correction to ensure maximum accuracy Avery Weigh-Tronix Technical Specification Description General This specification describes the indicator, capable of stand-alone operation, or of integration into a larger system via serial interface.

More information

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION UWE-1707 SERIES CONTENTS ELECTRONIC WEIGHING INDICATOR OPERATION MANUAL 1. SPECIFICATIONS PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE SCALE 2. INSTALLATION 3. ROUTINE OPERATION

More information

D-500s Indicator D LED. User Instructions. English Issue September SangR. Page 2. User s manual of Sang D-500s LED

D-500s Indicator D LED. User Instructions. English Issue September SangR. Page 2. User s manual of Sang D-500s LED D-500s Indicator D - 500 LED User Instructions English 8921-004210 Issue September 2010 1 2-2 0 1 0 D 6 0 0 4 6 7 5 Page 2 D-500s Indicator Page 2 Table of Content Section 1: Introduction. 1.1 Introduction

More information

1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882

1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 This manual is intended for use in support of installation, commissioning and configuration of the 1/32-DIN Temperature

More information

TI-500 Plus DIGITAL WEIGHT INDICATOR Setup / Operation Manual

TI-500 Plus DIGITAL WEIGHT INDICATOR Setup / Operation Manual TI-500 Plus DIGITAL WEIGHT INDICATOR Setup / Operation Manual Revision 1.0 975 Deerfield Parkway Buffalo Grove, IL 60089 April 3, 2008 Tel (847) 419-9180 Fax (847) 419-1515 http://www.transcell.net i TABLE

More information

LCI User Manual mantracourt.com

LCI User Manual mantracourt.com LCI User Manual mantracourt.com LCI Load Cell Junction Box with Fault Monitor Contents Chapter 1 Introduction to the LCI... 2 Chapter 2 Installing the LCI... 3 Chapter 3 Setting up the LCI... 4 Sequence

More information

Operating manual. Digital Weighing Indicator WE2110. A en

Operating manual. Digital Weighing Indicator WE2110. A en Operating manual Digital Weighing Indicator WE2110 WE2110 Contents...................................................... 3 Page Safety instructions................................................ 4 1

More information

7561-PSD Manual Portable Battery Powered Indicator

7561-PSD Manual Portable Battery Powered Indicator 7561-PSD Manual Portable Battery Powered Indicator Lebow Products Inc. 1728 Maplelawn Drive P.O. Box 1089 Troy, Michigan 48084-1089 (800) 803-1164 Phone: (248) 643-0220 FAX: (248) 643-0259 Visit our web

More information

400 Series (K410, K411, K412) Batching Indicator Reference Manual

400 Series (K410, K411, K412) Batching Indicator Reference Manual 400 Series (K410, K411, K412) Batching Indicator Reference Manual RI00-613-120 Copyright All Rights Reserved. No part of this document may be copied, reproduced, republished, uploaded, posted, transmitted,

More information

FG-7000 Digital Force Gauge Operation Manual

FG-7000 Digital Force Gauge Operation Manual FG-7000 Digital Force Gauge Operation Manual Operators should wear protection such as a mask and gloves in case pieces or components break away from the unit under test. Whether the unit is ON or OFF,

More information

Stock Weigh 300. Operators Manual HELLO. Ft. Atkinson, Wisconsin USA. Panningen, The Netherlands

Stock Weigh 300. Operators Manual HELLO. Ft. Atkinson, Wisconsin USA. Panningen, The Netherlands Stock Weigh 300 Operators Manual HELLO Ft. Atkinson, Wisconsin USA Panningen, The Netherlands www.digi-star.com D3734-US REV E October 31 st, 2012 SW300 User s Manual D3734-US REV E TABLE OF CONTENTS Table

More information

High Precision Quad Output DC Power Supply

High Precision Quad Output DC Power Supply User's Guide High Precision Quad Output DC Power Supply Model 382270 Introduction Thank you for selecting the Extech Model 382270. This device is shipped fully tested and calibrated and, with proper use,

More information

User Manual. Waterproof Weighing Scale. ELW Plus / ESW Plus ELW Plus Wipower

User Manual. Waterproof Weighing Scale. ELW Plus / ESW Plus ELW Plus Wipower User Manual Waterproof Weighing Scale ELW Plus / ESW Plus ELW Plus Wipower Excell Precision Limited 2016. All rights reserved Worldwide. The information contained herein is the property of Excell Precision

More information

DCS Series. Digital Counting Scale. Setup Manual

DCS Series. Digital Counting Scale. Setup Manual DCS Series Digital Counting Scale Setup Manual Revision 1.1 February 12, 1997 1996-1997 Transcell Technology, Inc. Contents subject to change without notice. Transcell Technology, Inc. 975 Deerfield Parkway

More information

KW serial. User s guide. Bench Scales UGKW-E0302

KW serial. User s guide. Bench Scales UGKW-E0302 KW serial Bench Scales User s guide UGKW-E0302 KW serial bench scale user s manual Table of Contents SECTION 1 INTRODUCTION 1 SECTION 2 SPECIFICATIONS.. 2 SECTION 3 INSTALLATION.3 SECTION 4 KEY DESCRIPTIONS

More information

Model 815 User s Manual

Model 815 User s Manual Model 815 User s Manual CAUTION Risk of electrical shock. Do not remove cover. No user serviceable parts inside. Refer servicing to qualified service personnel. Weigh-Tronix reserves the right to change

More information

D730. Remote Display Installation Manual

D730. Remote Display Installation Manual D730 Remote Display Installation Manual Copyright All Rights Reserved. No part of this document may be copied, reproduced, republished, uploaded, posted, transmitted, distributed, stored in or introduced

More information

CHAPTER 1. Introduction Introduction 3

CHAPTER 1. Introduction Introduction 3 Technical & User Manual Weight Indicator: SMIT-3015 Manufactured By: Samyak Instrumentation Pvt.Ltd. F-4, Memnagar Complex, Opp.Petrol Pump, Memnagar Village, Ahmedabad, India-380052. Phone: +91-79-27495500/5600,

More information

Emerson Network Power provides customers with technical support. Users may contact the nearest Emerson local sales office or service center.

Emerson Network Power provides customers with technical support. Users may contact the nearest Emerson local sales office or service center. Liebert PSA iton User Manual Version: V2.8 Revision date: November 14, 2005 Emerson Network Power provides customers with technical support. Users may contact the nearest Emerson local sales office or

More information

CALOG CALOG - LC DISPLAY. Loadcell Display. English Users Manual. Rev. 1.10

CALOG CALOG - LC DISPLAY. Loadcell Display. English Users Manual. Rev. 1.10 CALOG C A L I B R AT O R S CALOG - LC DISPLAY Loadcell Display English Users Manual Rev. 1.10 Table of Contents! Contents! 1.0 Introduction...2! 2.0 Getting Started...2! 2.1 Keypad...2! 2.2 Display...2!

More information

DWP-PC SERIES PRICE COMPUTING SCALE User s Guide

DWP-PC SERIES PRICE COMPUTING SCALE User s Guide DWP-PC SERIES PRICE COMPUTING SCALE User s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT Specifications subject to change without prior notice CONTENTS 1. INSTALLATION

More information

MRC 7000 Proven Recording Reliability that Became the Industry Standard!

MRC 7000 Proven Recording Reliability that Became the Industry Standard! Data Sheet Recorder//Profiler DESCRIPTION The MRC7000 Recorder/ is a microprocessor based circular chart instrument capable of measuring, displaying, recording and controlling up to two process variables

More information

GMC-206/215/230 Digital Desktop Weighing Scale Operation Manual

GMC-206/215/230 Digital Desktop Weighing Scale Operation Manual GMC-206/215/230 Digital Desktop Weighing Scale Operation Manual Version 2.1 Gamma Scale Tel: (905) 455-8333 Fax: (905) 455-0658 E-mail: info@gammascale.com Website: www.gammascale.com 1. Technical Parameters

More information

Digital Bench Scale. Revision 8.93 August 12, 1993

Digital Bench Scale. Revision 8.93 August 12, 1993 Digital Bench Scale Revision 8.93 August 12, 1993 Salter Brecknell Weighing Products 1000 Armstrong Drive Fairmont, MN 56031 Tel (800) 637-0529 Tel (507) 238-8702 Fax (507) 238-8271 E-mail: sales@salterbrecknell.com

More information

FG-7000L Digital Force Gauge Operation Manual

FG-7000L Digital Force Gauge Operation Manual FG-7000L Digital Force Gauge Operation Manual Operators should wear protection such as a mask and gloves in case pieces or components break away from the unit under test. Whether the unit is ON or OFF,

More information

Table of Contents. 3-1 Specifications & Features Front Panel Display Keyboard. 3-3 Rear Panel. 3-4 Power Supply. 4.

Table of Contents. 3-1 Specifications & Features Front Panel Display Keyboard. 3-3 Rear Panel. 3-4 Power Supply. 4. Table of Contents 1. Introduction 1 2. Precaution... 1 3. Product Introduction 3-1 Specifications & Features... 3-2 Front Panel 3-2-1 Display 3-2-2 Keyboard. 3-3 Rear Panel. 3-4 Power Supply. 4. Installation

More information

Weighing Indicator Service Manual V1.24X

Weighing Indicator Service Manual V1.24X Weighing Indicator Service Manual V1.24X CONTENTS 1. PRECAUTIONS.3 2. SPECIFICATIONS....4 3. INTRODUCTION....5 4. INSTALLATION..6 Unpacking 6 Installation 6 Load cell connections......7 Connect Adaptor

More information

M260 Load Cell Amplifiers

M260 Load Cell Amplifiers E L 4 - + 2-3 1-2 M260 Load Cell Amplifiers Description. This fast start guide shows basic installation, setup, and calibration for most applications. For more specific information contact Minsterport.

More information

Userʼs Guide. 1200g High Precision Scale. 1200g High Precision Scale. Legal for Trade. Includes RS-232 interface port. UG Version 1.

Userʼs Guide. 1200g High Precision Scale. 1200g High Precision Scale. Legal for Trade. Includes RS-232 interface port. UG Version 1. Userʼs Guide 1200g High Precision Scale Legal for Trade 1200g High Precision Scale Includes RS-232 interface port UG Version 1.2_12-30-13 !!!! CALIBRATION WARNING!!!! Calibration AND inspection of calibration

More information

Installation / Operator / Service Manual. Bench Scales Revision 6 07/ by Fairbanks Scales Inc. All rights reserved

Installation / Operator / Service Manual. Bench Scales Revision 6 07/ by Fairbanks Scales Inc. All rights reserved Installation / Operator / Service Manual Bench Scales 2005-2010 by Fairbanks Scales Inc. All rights reserved 50778 Revision 6 07/10 Amendment Record 50778 Manufactured by Fairbanks Scales Inc. 821 Locust

More information

BW/BWS Weighing Indicator

BW/BWS Weighing Indicator BW/BWS Weighing Indicator - 1 - CONTENTS 1. SPECIFICATION... 2 2. INTRODUCTION... 3 3. INSTALLATION. 4 Precautions 4 Parts 4 Installation.. 5 4. KEYS DESCRIPTION... 6 Key Board... 6 Secondary functions

More information

1.0 Description. 2.0 Unpacking. 3.0 Installation

1.0 Description. 2.0 Unpacking. 3.0 Installation ES-H, ES-HA Series Precision Balance Thank you for purchasing the Model ES-H and ES-HA precision balance. Please read all operating instructions carefully before using and note the following items to ensure

More information

PC 150. Digital Bench Scale. Operation Manual

PC 150. Digital Bench Scale. Operation Manual PC 150 Digital Bench Scale Operation Manual Revision 1.2 September 14, 2000 1996-2000 Transcell Technology, Inc. Contents subject to change without notice. Transcell Technology, Inc. 35 Waltz Drive Wheeling,

More information

IGB Series. Operation Manual. Platform Scale. Kg/Lb Specification IMPORTANT

IGB Series. Operation Manual. Platform Scale. Kg/Lb Specification IMPORTANT Platform Scale IGX/ Kg/Lb Specification Operation Manual IMPORTANT Do not carry out installation, operation, service, or maintenance until thoroughly understanding the contents of this manual. Keep this

More information

User's Guide. Model High Precision Quad Output DC Power Supply

User's Guide. Model High Precision Quad Output DC Power Supply User's Guide Model 382270 High Precision Quad Output DC Power Supply Introduction Congratulations on your purchase of the Extech 382270 DC Power Supply. The Model 382270 can be used for many applications

More information

QuickSilver Series Bench Scale

QuickSilver Series Bench Scale Operating Manual QuickSilver Series Bench Scale 2004 by Fairbanks Scales Inc. All rights reserved 50738 Issue #1 1/04 Amendment Record QuickSilver Series Bench Scale 50738 Manufactured by Fairbanks Scales

More information

Multi-Function Smart Indicator User Manual

Multi-Function Smart Indicator User Manual Multi-Function Smart Indicator User Manual EC30006 Content 1 Notice... 2 2 Specification... 2 3 Main parameter... 3 4 How to fix... 3 5 Port... 4 6 On/Off... 4 7 Charging... 4 8 Display... 5 9 Screen keypad...

More information

1. INTRODUCTION This manual contains installation, operation and maintenance instructions. Please read the manual completely before using the balance.

1. INTRODUCTION This manual contains installation, operation and maintenance instructions. Please read the manual completely before using the balance. 1. INTRODUCTION This manual contains installation, operation and maintenance instructions. Please read the manual completely before using the balance. 1.1 Safety Precautions Please follow these safety

More information

Division Services. Model: DPAS-300. Division Portable Bulk Load Accumulator Scale INDUSTRIAL PH (901) FAX (901)

Division Services. Model: DPAS-300. Division Portable Bulk Load Accumulator Scale INDUSTRIAL PH (901) FAX (901) Division Services 5680 E. Shelby Drive Memphis, TN 38141 PH (901) 366-4220 FAX (901) 365-3934 Model: DPAS-300 VERSION 5 Division Portable Bulk Load Accumulator Scale INDUSTRIAL Division Model DPAS-300

More information

INSTALLATION MANUAL. LC 200 Electronic Overload Guard. Software versione PW0501 R 0.3

INSTALLATION MANUAL. LC 200 Electronic Overload Guard. Software versione PW0501 R 0.3 INSTALLATION MANUAL LC 200 Electronic Overload Guard Software versione PW0501 R 0.3 CONTENTS MAIN FEATURES LC 200 TECHNICAL FEATURES Page 2 SYMBOLS Page 3 WARNINGS Page 3 IDENTIFICATION DATA PLATE Page

More information

LCI Load Cell Junction Box with Fault Monitor

LCI Load Cell Junction Box with Fault Monitor LCI Load Cell Junction Box with Fault Monitor User Manual www.mantracourt.co.uk The LCI Load Cell Failure Alarm Manual Chapter 1 Introduction to the LCI...2 Chapter 2 Installing the LCI...3 Figure 2.1

More information

1.6. Counters, Panel Meters, Tachometers and Timers. Contents Description Fusion Integrated Machine Control Standards and Certifications...

1.6. Counters, Panel Meters, Tachometers and Timers. Contents Description Fusion Integrated Machine Control Standards and Certifications... .6 Contents Standards and Certifications............... Product Selection....................... Technical Data and Specifications........... Dimensions............................ Learn Online Page V3-T-04

More information

Trade Approved Version NMI: S520

Trade Approved Version NMI: S520 @Weigh BW/BWS Series Trade Approved Version NMI: S520 Weighing Indicator User s guide UGBW-E1.09 Table of Contents SECTION 1 INTRODUCTION. 1 SECTION 2 SPECIFICATIONS... 2 SECTION 3 INSTALLATION.3 3. 1

More information

1. Introduction Page 2. Specifications Page Installation Page Key Functions Page Operations Page 6. Check Weighing Page 6

1. Introduction Page 2. Specifications Page Installation Page Key Functions Page Operations Page 6. Check Weighing Page 6 Contents: 1. Introduction Page 2 Specifications Page 2 2. Installation Page 3 3. Key Functions Page 4 4. Operations Page 6 Check Weighing Page 6 Accumulation Page 7 Parts Counting Page 7 Print (RS232C)

More information

SY021 Portable Load/Force Meter User instructions

SY021 Portable Load/Force Meter User instructions SY021 Portable Load/Force Meter User instructions Relates to firmware version 5.2 INTRODUCTION The SY021 is a portable load meter, which can indicate the load present on any attached cell. A dual channel

More information

805HP. Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website:

805HP. Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website: 805HP Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website: www.anyload.com Email: info@anyload.com TABLE OF CONTENTS 1. Introduction and Product Features 3 2.

More information

Digital Bench Scale. Revision 1.2 August 24, Contents subject to change without notice.

Digital Bench Scale. Revision 1.2 August 24, Contents subject to change without notice. Digital Bench Scale Revision 1. August 4, 000 Contents subject to change without notice. Salter Brecknell Weighing Products 1000 Armstrong Drive Fairmont, MN 56031 Tel (800) 637-059 Tel (507) 38-870 Fax

More information

EB300 Weighing Scale Operation Manual

EB300 Weighing Scale Operation Manual EB300 Weighing Scale Operation Manual EB300 User Manual Page 1 of 18 TABLE OF CONTENTS 1. Safety Warnings... 3 2. Features..... 3 3. Options.......... 4 4. Product packaging.. 4 5. Display and keyboard.....

More information

Pennsylvania Scale Company Model 400B Battery Powered Digital Indicator Operation & Calibration Manual

Pennsylvania Scale Company Model 400B Battery Powered Digital Indicator Operation & Calibration Manual Pennsylvania Scale Company Model 400B Battery Powered Digital Indicator Operation & Calibration Manual Pennsylvania Scale Company 1042 New Holland Avenue Lancaster PA 17601 For online interactive tech

More information

BB 8100 SS Operator s Manual & Troubleshooting Guide

BB 8100 SS Operator s Manual & Troubleshooting Guide FCC NOTICE BB 8100 SS Operator s Manual & Troubleshooting Guide This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Subpart J of Part 15 of the

More information

SHIMPO INSTRUMENTS. FG-7000T Digital Torque Gauge Operation Manual

SHIMPO INSTRUMENTS. FG-7000T Digital Torque Gauge Operation Manual FG-7000T Digital Torque Gauge Operation Manual SHIMPO INSTRUMENTS Operators should wear protection such as a mask and gloves in case pieces or components break away from the unit under test. Whether the

More information

User's Guide. Programmable DC Power Supply 200 Watt (40 Volts / 5 Amps) Model Introduction

User's Guide. Programmable DC Power Supply 200 Watt (40 Volts / 5 Amps) Model Introduction User's Guide Programmable DC Power Supply 200 Watt (40 Volts / 5 Amps) Model 382280 382280 Introduction Congratulations on your purchase of the Extech 382280 Programmable DC Power Supply. This 200 watt

More information

Multi-Function Indicator. Operation Manual (Full Version) PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE OPERATING THIS INSTRUMENT

Multi-Function Indicator. Operation Manual (Full Version) PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE OPERATING THIS INSTRUMENT 19eA & 19eS Multi-Function Indicator Operation Manual (Full Version) PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE OPERATING THIS INSTRUMENT Specifications subject to change without prior notice V110 Rev

More information

GS100P. Weighing Controller. Contents. Order Codes

GS100P. Weighing Controller. Contents. Order Codes GS100P Weighing Controller 1 This intelligent weighing controller comes precalibrated to suit either a 4-wire or a 6-wire strain gauge input, and has a number of advanced functions designed specifically

More information

TR150 Digital Readout Unit SL755. Impact Test Equipment Ltd & User Guide. User Guide

TR150 Digital Readout Unit SL755. Impact Test Equipment Ltd   &   User Guide. User Guide TR150 Digital Readout Unit SL755 Impact Test Equipment Ltd www.impact-test.co.uk & www.impact-test.com User Guide User Guide Impact Test Equipment Ltd. Building 21 Stevenston Ind. Est. Stevenston Ayrshire

More information

Precision Balance W3200 Series Operating Manual

Precision Balance W3200 Series Operating Manual Precision Balance W3200 Series Operating Manual Version 9-19-16 1 2 CONTENTS 1. INTRODUCTION. 4 2. INSTALLATION... 4 2-1. Unpacking.... 5 2-2. Specifications.... 5 2-3. Environmental Requirements... 5

More information

Copyright Western Scale Co. Limited. All rights reserved.

Copyright Western Scale Co. Limited. All rights reserved. MAX DIGITAL WEIGHT INDICATOR OPERATION MANUAL SOFTWARE RELEASE 5.03, 04/2011 Copyright 2007-2011 Western Scale Co. Limited. All rights reserved. Published by: Western Scale Co. Limited. Information in

More information

Multi-Point Gas Detection and Control System

Multi-Point Gas Detection and Control System Multi-Point Gas Detection and Control System Specifications subject to change without notice. USA 40804 Page of 7 DESCRIPTION Wall mounted, microprocessor-based, multi-point, analog electronic control

More information

LTS DR2100A/AN Indicator WARRANTY POLICY

LTS DR2100A/AN Indicator WARRANTY POLICY TABLE OF CONTENTS WARRANTY POLICY... 1 CONTACT INFORMATION... 2 GENERAL INFORMATION... 2 STANDARD FEATURES... 3 INDICATOR OPTIONS... 3 DISPLAY DESCRIPTION... 4 DR2100 KEYBOARD LAYOUT... 4 KEYBOARD CONTROLS...

More information

A0-10 ANALOG OUTPUT. Instruction Manual April 2001 NALOG OUTPUT

A0-10 ANALOG OUTPUT. Instruction Manual April 2001 NALOG OUTPUT A0-10 ANALOG OUTPUT Instruction Manual April 2001 NALOG OUTPUT Safety Guidelines Warning notices must be observed to ensure personal safety as well as that of others, and to protect the product and the

More information

PCE Americas Inc. PCE Instruments UK Ltd. 711 Commerce Way Units 12/13, Southpoint Business Park Suite 8 Ensign Way Jupiter, FL 33458 Hampshire / Southampton USA United Kingdom, SO31 4RF Phone: 561-320-9162

More information

Industrial SBU Product User s Manual

Industrial SBU Product User s Manual Industrial SBU Product User s Manual EUROPEAN COUNTRIES WARNING This is a Class A product. In a domestic environment this product may cause radio interference in which the user may be required to take

More information

Model S200 Series. Revision 1.0 July, Contents subject to change without notice.

Model S200 Series. Revision 1.0 July, Contents subject to change without notice. Model S200 Series Revision 1.0 July, 2003 Contents subject to change without notice. Salter Brecknell Weighing Products 1000 Armstrong Drive Fairmont, MN 56031 Tel (800) 637-0529 Tel (507) 238-8702 Fax

More information

Power TracKer VII. User s Manual 2006 AVID. Multi Mode Reader. Manufactured under one or more of the following Patents:

Power TracKer VII. User s Manual 2006 AVID. Multi Mode Reader. Manufactured under one or more of the following Patents: Power TracKer VII User s Manual 2006 AVID Multi Mode Reader Manufactured under one or more of the following Patents: 5,235,326-5,266,926-5,559,507 6,172,609 FCC ID: IOL-134-AV1028 The device complies with

More information

Digital Indicator Solutions

Digital Indicator Solutions Digital Indicator Solutions Loop-Powered Attachable Compact Attachable Loop-Powered Smart Compact System Loop-Powered Compact Smart System Intelligent Smart System "Intelligent" Options Dual & Display

More information

DC-10. Dual Counting Indicator. Operation Manual (Full Version) PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE OPERATING THIS INSTRUMENT

DC-10. Dual Counting Indicator. Operation Manual (Full Version) PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE OPERATING THIS INSTRUMENT DC-10 Dual Counting Indicator Operation Manual (Full Version) PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE OPERATING THIS INSTRUMENT Specifications subject to change without prior notice V225 December

More information

SAC / SHC / SHC-CH SERIES

SAC / SHC / SHC-CH SERIES This document hosted by: www.oldwillknottscales.com SAC / SHC / SHC-CH SERIES HIGH PRECISION COUNTING SCALE OPERATION MANUAL PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE SCALE July

More information

Totalcomp TLI Indicator. Operation and Maintenance Manual Manual

Totalcomp TLI Indicator. Operation and Maintenance Manual Manual Totalcomp TLI Indicator Operation and Maintenance Manual Manual V1.08-0.06E DT: 04/29/2013 - - CONTENTS Precautions... 2 1. SPECIFICATIONS... 3 2. INTRODUCTION. 4 3. INSTALLATION... 5 Unpacking 5 Parts

More information