CHAPTER 1. Introduction Introduction 3

Size: px
Start display at page:

Download "CHAPTER 1. Introduction Introduction 3"

Transcription

1 Technical & User Manual Weight Indicator: SMIT-3015 Manufactured By: Samyak Instrumentation Pvt.Ltd. F-4, Memnagar Complex, Opp.Petrol Pump, Memnagar Village, Ahmedabad, India Phone: /5600, Fax: , Mobile: , For Support Inquiry ID: For New Product Inquiry: Web: Skype ID: samyakin Gtalk ID: samyakahd

2 INDEX CHAPTER 1. Introduction Introduction 3 CHAPTER 2.Technical Specification Power Supply Requirements Performance Display Hardware Features Environmental Analog Input Outputs Analog Output Enclosure 4 CHAPTER 3.Installation & Mounting Details Installation Mounting 5 CHAPTER 4.Power Supply and Wiring Details Back Terminal Detail Power Supply Connection Load Cell Connection Analog Output Connections Serial Output Connections Digital Input Connections Control Relay Outputs Connections 8 CHAPTER 5.Operating Details Display and Keyboard 9 CHAPTER 6. Menu List Functions Display DISP Menu Operation OPER Menu Calibration CAL Menu Relay Operation 18 CHAPTER 7. Serial communication Details Modbus Registers 22 CHAPTER 8. Sales and Service Sales and Service 23

3 1. Introduction 1.1 Introduction This Manual provides general information on the installation, configuration, calibration and operation of Weight Indicator SMIT It is based on popular 8-bit micro controller. It displays analog signal (mv) input from the Load cell into engineering values on display. This unit is 5 Digit Red LED Display and 4 membrane key. All units include the following futures. Single channel load cell with 300 Ohms Resistor Supports 4 wire load cell connections RS232 or RS485 Serial Interface Outputs 2 Digital Inputs (Start/Stop and Zero) 2 Control Relay Outputs (Set points) Analog output 4mA to 20mA, 0-10VDC Output utilizing 16 bit DAC.

4 2. Technical Specifications 2.1 Power Supply Requirements Input Voltage Power Supply: 24VDC + 10% Power Consumption : <5.0W 2.2 Performance Sampling Rate : 15 Readings /Second Load Cell Signal Input Range: 0-15mV Filtering : Analog & Digital Decimal : 0, 0.0, 0.00, selectable Calibration Method : Via Front Panel Key Board or Computer Interface Using Serial Communication 2.3 Display Type Keyboard: 4 keys Membrane keypad 2.4 Hardware Features Non-volatile EEPROM 2.5 Environmental Operating Temperature: 0 to 55 C Storage Temperature : 0 to 70 C : 5 Digit 7 Segment 0.5 Red LED Display for displaying the weight. Humidity : 30 to 90%RH (Non condensing) Warm-up time : <10 Min 2.6 Analog Input Type : 24 bit A/D conversation Load cell : 4 wire single channel load cell input Digital Input: Maximum 2 Digital Inputs 2.7 Outputs Relay outputs : Maximum 2 Relay Control Outputs Digital outputs : 1 Digital Output Serial Output : RS 232 / RS485 serial port for remote display/printer/computer Standard Protocol: MODBUS Baud Rate : 9600, 19200, 38400, selectable 2.8 Analog Output Type : 16 bit D/A conversation Current: 4-20mA (Compliance up to 500 Ohms) Voltage: 0-10VDC 2.9 Enclosure Enclosure Mounting : General purpose made up Plastic : Panel Mounted Outer Dimension : 96 mm (W) x 48 mm (H) x 110 mm (D) Cutout Dimension: 92 mm (W) x 44 mm (H)

5 3. Installation & Mounting Details 3.1 Installation Instructions before Install an Instrument: The instrument is powered from an external 24VDC source. The instrument can be operated from a computer; therefore, a clean power source is required for reliable operation. The incoming power should come from a source that is isolated from other process equipment. Cables carrying primary and switched power should be routed away from load cell and other signal cables to avoid electrical interference. Relays, motor starters and other inductive devices connected to the equipment must have reliable and effective arc suppression. Plastic insulating tape should not be used on load cell connections. Heavy electrical equipment should not be installed close to the weighing equipment. Excessive vibration will affect the accuracy of the weigh system and depending on the severity can cause damage to electrical and electronic components. The atmosphere should be dust free and not contain any corrosive gasses or materials, which could adversely affect the equipment. Check proper earthing of the unit. Check proper voltage supply and frequency of the electric supply. Check proper connection of load cell. 3.2 Mounting Install the instrument in a location where it will not be subjected to excessive heat, humidity or vibration. The unit should be installed at eye level so as to allow viewing be installed at eye level so as to allow viewing the display and access to the front panel keys. Avoid direct sunlight on the front of the instrument. Mount the instrument in the panel cutout of 96mm * 48mm. Fix the instrument with the panel using two nos. brackets provided with the instrument.

6 4. Power Supply and Wiring Details Refer connections details for back panel connections. 4.1 Back Terminal Detail 4.2 Power Supply Connection Use Terminal No.1 for the Positive node (+) of power supply and Terminal No.2 for the Negative node (-) of power supply. The supply Voltage must be 24VDC. SMPS SMIT-3015 Terminal Details +V Terminal No.1 (+24V) -V Terminal No.2 (GND) Earth Terminal No.3 (EARTH)

7 4.3 Load Cell Connection From 4 wired load cell shielded cable connect to in unit Shielded wire to Terminal No.4, Green wire (SENSE+) to Terminal No.5, White wire (SENSE-) to Terminal No.6, Red wire (EXC+) to Terminal No.7, Black wire (EXC+) to Terminal No 8. Load Cell Wire SMIT-3015 Terminal Details (Four core shielded) SHIELD Terminal No.4 (SHIELD) SENSE+ (Green) Terminal No.5 (SENSE+) SENSE- (White) Terminal No.6 (SENSE-) EXC+ (Red) Terminal No.7 (EXC+) EXC- (Black) Terminal No.8 (EXC-) 4.4 Analog Output Connections Use Terminal No.9 for the Positive node (+) of Retransmission output (4-20mA / 0-10V) and Terminal No.10 for the Negative node (-) of Retransmission output (4-20mA / 0-10V). Analog Output OUT+ OUT- SMIT-3015 Terminal Details Terminal No.9 (OUT+) Terminal No.10 (OUT-) 4.5 Serial Output Connections This serial Port supports RS232 and RS485 communication For RS232 connection Use Terminal No.11 for TxD, Terminal No.12 for RxD and Terminal No.13 for GND For RS485 connection Use Terminal No.11 for D- and Terminal No. 12 D+ For Serial RS232 Connection: 9-PIN D TYPE CONNECTOR (COMPUTER) Pin No.2 (TxD) Pin No.3 (RxD) Pin No.5 (GND) SMIT-3015 Terminal Details Terminal No.12 (RxD) Terminal No.11 (TxD) Terminal No.13 (GND) For RS 485 Connection: INTERFACE DEVICE SMIT-3015 Terminal Details D- Terminal No.11 (D-) D+ Terminal No.12 (D+)

8 4.6 Digital Input Connections Connect Positive of 24 VDC to Terminal No.15. Connect Tare PUSH button between Terminal 16 and Common of 24 VDC. Connect START/STOP PUSH button between Terminal no 17 and Common of 24 VDC. Digital Inputs SMIT-3015 Terminal Details 24V Terminal No. 15 TARE Terminal No. 16 START/STOP Terminal No Control Relay Outputs Connections Use Terminal No.18 for Relay 2 NO/NC2 and Terminal No. 19 for Relay 1 NO/NC1 and Terminal No. 20 for the common for the both outputs. Relays SMIT-3015 Terminal Details NO/NC2 Terminal No. 18 NO/NC1 Terminal No. 19 COMMON Terminal No. 20

9 5. Operating Details The following paragraphs give detailed description of how to operate the unit. For efficient use of the instrument one must study and understand this section. 5.1 Display and Keyboard Display: 5- digit 7-segment red Led display. Key Board: 4-key keypad organized as 1 x 3 lines. List of keys and their functions: 1. Index Key: If in Edit Mode Save New data Else it will select new Menu. 2. Select4Key: Enter into data entry/verification mode. The Edit Mode is identified with Blinking of Selected Display. When In Edit mode selects Next Digit. 3. Increment 5Key: Increment selected parameter value. 4. Esc Key: Cancel Selection and Terminate Edit mode Note: Select 4 Key also used for start stop function in non-edit mode in specific instruments. Increment 5Key also used as tare key in non-edit mode.

10 6. Menu List Functions Parameter in Weight Indicator SMIT 3015 is divided in three main Menus. Display DISP Menu Operation OPER Menu Calibration CAL Menu 6.1 Display DISP Menu Display Values DP (Decimal) 0, 0.0, 0.00, SVD (Step Width) 0, 0.2, 0.5, 1, 2, 5 ROUND (Rounding) 1, 2, 5

11 Decimal Position: DP Press Index key, to see DISP Menu after that press Right arrow key to see current selected DP decimal position. Press Right key to enter DP Menu after that use Up arrow' key to select desired decimal point position like for 0, 0.0, 0.00 and and select any one and press 'ENTER' key to store it in NVRAM. The decimal point will be displayed at newly selected position. Step width: Svd Use of Step width: Step width is the minimum change in weight required to display the weight. Display weight will only update when amount of change in weight is greater than or equal to value of Step width. Setting Svd = 0 will update Display weight directly. For example, if Svd = 2 is set and display weight is 15.0 kg now display weight will only update when consecutive weight is more than or equal to 15.2 because difference should be more than value set in Svd. Press Index key, you will see DISP Menu after pressing Right arrow key to display currently selected Svd. Press Right key for enter Svd Menu after Using 'UP arrow' key select desired Svd position like for 0,0.2,0.5,1,2 & 5 then select any one and press 'ENTER' key to store it in NVRAM. Rounding: round Least digit of Display weight will be rounded off by selected rounding value. For example, If round = 1, then weight will display 0, 1, 2, 3. If round = 2, then weight will display 0, 2, 4, 6. If round = 5, then weight will display 0, 5, 10, 15. Press Index key, you will see DISP Menu after pressing Right arrow key to display currently selected round.press Right key for enter round Menu after Using 'Up arrow' key select desired rounding value like for 1,2 and 5 and select any one and press 'ENTER' key to store it in NVRAM.

12 6.2 Operation OPER Menu Display Values OP-FS (Output Full Scale) NAV (Moving Average) PON (Power On Time) HI (High Alarm) CLR PEK (Clear Peak) PASSWORD 3210 UNIT (Unit No) HIHI (High-High Alarm) LO (Low Alarm) LOLO(Low-Low Alarm) CAP (Capacity) PLANT (Plant Name) Set Plant Name Up to 18 characters BAUD (Baud Rate) 9.6,19.2,38.4,57.6 and SPS (Sample per Second) 12.5,25,50,100 HYSTE (Hysteresis) REL 1 LO,LOLO,HI,HIHI REL 2 LO,LOLO,HI,HIHI

13

14 Full-scale OUTPUT: op-fs Right Key you will see op-fs and. Press again Right key for enter op-fs Menu, and using UP and RIGHT key set the value of Maximum KG of the parameter to the required value for getting Maximum current output from the unit will give 20mA. Or 10V and press 'ENTER' key to store it in NVRAM. Moving Average Time: nav Moving Average is use to removes scale vibration & other noise signal. Setting OFF provides no averaging; Setting 32 provide maximum averaging. Higher averaging number will give higher stability but will take longer time to settle weight display. Right Key you will see op-fs and. After Press Index key for select nav Menu, and now press RIGHT key entering nav menu and using Up arrow' key select desired valued like 1,2,3, up to 32 and OFF select any one and press 'ENTER' key to store it in NVRAM.

15 Power on delay Time: POn High Alarm: HI Clear Peak: CLr Pk Unit no: Unit Right Key you will see op-fs and. After Press Index key for select POn Menu, and now press RIGHT key entering POn menu and using Up arrow and RIGHT key for set Power on starting time and press 'ENTER' key to store it in NVRAM. Right Key you will see op-fs and. After Press Index key for select HiAlrn Menu, and now press RIGHT key entering HiAlrn menu and using Up arrow and RIGHT key for set point for High Alram and press 'ENTER' key to store it in NVRAM. Right Key you will see op-fs and. After Press index key for select CLr PEk Menu, and now press RIGHT key entering CLr PEk menu and using UP arrow and RIGHT key set password 3210 and press 'ENTER' key to Clear Peak Value. Right Key you will see op-fs and. After Press Index key for select Unit Menu, and now press RIGHT key entering Unit menu and using UP arrow and RIGHT key for set Values like 01,02, up to 32 and select any one and press 'ENTER' key to store it in NVRAM. High-High Alarm: HIHI Right Key you will see op-fs and. After Press Index key for select HIHI Menu, and now press RIGHT key entering HIHI menu and using Up arrow and RIGHT key for set High-High Alram and press 'ENTER' key to store it in NVRAM.

16 Low Alarm : Lo Right Key you will see op-fs and. After Press Index key for select Lo Menu, and now press RIGHT key entering Lo menu and using Up arrow and RIGHT key for set Low Alarm and press 'ENTER' key to store it in NVRAM. Low-Low Alarm: LoLo Right Key you will see op-fs and. After Press Index key for select LoLo Menu, and now press RIGHT key entering LoLo menu and using Up arrow and RIGHT key for set Low-Low Alarm and press 'ENTER' key to store it in NVRAM. Capacity weight For over message: CAP Plant Name PLAnt Baud Rate: baud Right Key you will see op-fs and. After Press Index key for select CAP Menu, and now press RIGHT key entering CAP menu and using UP arrow and RIGHT key for values of set Values of Kg for Maximum full capacity values of weight of showing over message and press 'ENTER' key to store it in NVRAM. Right Key you will see op-fs and. After Press Index key for select PLAnt Menu, and now press RIGHT key entering PLAnt menu and using UP arrow and RIGHT key entered through keyboard and will be displayed at power ON Time and press 'ENTER' key to store it in NVRAM. It is Maximum 18 characters can be entered. Right Key you will see op-fs and. After Press Index key for select baud Menu, and now press RIGHT key entering baud menu and using UP arrow and RIGHT key for set Values like 9600,19200,38400,57600 and and select any one and press 'ENTER' key to store it in NVRAM.

17 Sample per second SPS Hysteresis : HyStE For Digital filter, Unit provided four options to set ADC output word. i.e. 12.5, 25, 50 & 100 Samples Per Seconds. For Fast Weight updation, higher value of SPS is set but will be more sensitive to external electro-mechanical disturbances. Right Key you will see op-fs and. After Press Index key for select SPS Menu, and now press RIGHT key entering SPS menu and using UP arrow and RIGHT key for set Values like 12.5, 25, 50 & 100 and select any one and press 'ENTER' key to store it in NVRAM. Right Key you will see op-fs and. After Press Index key for select HyStE Menu, and now press RIGHT key entering HyStE menu and using UP arrow and RIGHT key for set Value of Hysteresis and press 'ENTER' key to store it in NVRAM. Relay 1 Selection rel 1 Right Key you will see op-fs and. After Press Index key for select rel 1 Menu, and now press RIGHT key entering rel 1 menu and using UP arrow and RIGHT key for set Lo, LoLo, HI,HIHI and select any one and press 'ENTER' key to store it in NVRAM. Relay 2 Selection rel 2 Right Key you will see op-fs and. After Press Index key for select SPS Menu, and now press RIGHT key entering SPS menu and using UP arrow and RIGHT key for set Lo, LoLo, HI,HIHI and select any one and press 'ENTER' key to store it in NVRAM

18 6.3 Relay Operation Relay 1 and relay 2 can be configured as Low Alarm, Low Low Alarm, High Alarm and High High Alarm 1. Low Alarm: Low Alarm Relay will be ON When Current Weight value is less than Relay 1 (Low Alarm) and it will be OFF When Current Weight value is greater than (Low Alarm + hysteresis) 2. High Alarm: High Alarm Relay will be ON When Current Weight value is greater than Relay 2 (High Alarm) and it will be OFF When Current Weight value is Less than (High Alarm - hysteresis) 6.4 Calibration CAL Menu Display Value ZERO (Zero) SPAN (Span) OUTZ (Output Zero) OUTS (Output Span) COUNT (ADC Count)

19

20 Press INDEX key three time, you will see CAL Menu after pressing RIGHT Key you will see PASS and again press once time RIGHT Key for you will see Password 0000 and using UP and RIGHT Key set password 1234 then press Enter key so you will see below Menu. Calibration Zero: ZERO Calibration Span: SPAN Calibration Out Zero: OUTZ Calibration Out Span: OUTS COUNT Calibration Zero: ZERO After entering password you will see ZERO menu after pressing RIGHT key for entering ZERO Menu and using UP and RIGHT key set 0.0 kg values at empty load cell plate form and press 'ENTER' key to store it in NVRAM. Calibration Span: SPAN After entering password you will see SPAN menu after pressing once again INDEX key you will see SPAN Menu and using RIGHT key for entering SPAN Menu and now put any known weight like 5kg, 10kg, 50kg & 100kg on load cell plate form. And using UP and RIGHT key set known kg values and press 'ENTER' key to store it in NVRAM. Note: - Please take known weight maximum 10% value of your load cell capacity so you can take more accuracy. Calibration out Zero: OUTZ After entering password you will see ZERO menu after pressing twice time again INDEX key you will see OUTZ Menu and using RIGHT key for entering OUTZ Menu and now put Red probe of multimeter on back Terminal No. 9 OUT+ and also put Black probe of multimeter on back Terminal No.10 OUT-. and using UP and RIGHT key set values of Counts till getting Retransmission Zero Output (0V or 4mA) and press 'ENTER' key to store it in NVRAM. Calibration out Span: OUTS After entering password you will see ZERO menu after pressing three time again INDEX key you will see OUTS Menu and using RIGHT key for entering OUTS Menu and now put Red probe of multimeter on back Terminal No. 9 OUT+ and also put Black probe of multimeter on back Terminal No.10 OUT-. and using UP and RIGHT key set values of Counts till getting Retransmission SPAN Output (10V or 20mA) and press 'ENTER' key to store it in NVRAM. COUNT After entering password you will see ZERO menu after pressing four time again INDEX key you will see COUNT Menu and using RIGHT key for entering COUNT Menu and now you will see ADC counts of the unit.

21 7. Serial communication Details Refer back details for Serial Connection. This serial port supports RS232 communication. For RS232 connection: 9-PIN D TYPE CONNECTOR (COMPUTER) Pin No.2 (TxD) Pin No.3 (RxD) Pin No.5 (GND) SMIT-3015 Terminal Details Terminal No.12 (RxD) Terminal No.11 (TxD) Terminal No.13 (GND) How to Test: - To test serial interface, you may use MODSCAN32.EXE utility. Settings: 1. Set Device Address = 1 2. Set Datatype= Holding Register 3. Set Start Address=1 4. Set Length = BAUD RATE = Set Proper COM Port 7. Set Data Format= Decimal 8. Now click on Connect button.

22 7.1 Modbus Registers SR. NO. Register No Parameter Function Weight LSB of displayed weight (16-bits)R Weight MSB of displayed weight (16-bits)R Adc count LSB of displayed weight (16-bits)R Adc count MSB of displayed weight (16-bits)R Decimal point 0-3 R,W Svd 1,2,5 R,W Rounding value 1,2,5 R,W Full-scale lower byte LSB of displayed weight ( Full-scale higher MSB of displayed weight (16-10 byte Moving Average time 1-32 R,W Power on delay 0-30sec R,W LO ALarm lower byte LSB of displayed weight ( IO Alarm higher byte MSB of displayed weight ( LOLO Alarm lower byte LSB of displayed weight ( LOLO Alarm higher byte MSB of displayed weight ( Hi ALarm lower byte LSB of displayed weight ( Hi Alarm higher byte MSB of displayed weight ( HiHi Alarm lower byte LSB of displayed weight ( HiHi Alarm higher byte MSB of displayed weight ( Unit No R,W Capacity lower byte LSB of displayed weight ( Capacity higher byte MSB of displayed weight ( Tare, zero 1,2 W Zero Count Span Count Start/Stop 0 for stop, 1 for start SPS 15,30,60,120,240,480,960,1920

23 8. Sales and Service When the unit has occurring any problem you may be contact on below address and contact no. Samyak Instrumentation P Ltd. F-4, Memangar Complex, Opp. Petrol Pump, Memnagar Village, Ahmedabad, India Phone: /5600, Fax: , Mobile: , For Support Inquiry ID: -support@samyak.co.in For New Product Inquiry: -sales@samyak.com Web: - Skype ID: samyakin Gtalk ID: -samyakahd

OPERATING MANUAL FOR SMIT-301MH STREAM SELECTOR

OPERATING MANUAL FOR SMIT-301MH STREAM SELECTOR OPERATING MANUAL FOR SMIT-301MH STREAM SELECTOR Manufactured By: INSTRUMENTATION PVT. LTD. F-4 Memnagar complex, Opp. Petrol Pump Memnagar, Ahmedabad, India-380 052 Phone: +91-79_ 2749 5500/5600 FAX: +91-79-2741

More information

EXCELL PRECISION CO., LTD.

EXCELL PRECISION CO., LTD. Excell Precision Limited 24. All rights reserved Worldwide. The information contained herein is the property of Excell Precision Limited and is supplied without liability for errors or omissions. No part

More information

The LT1300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications.

The LT1300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. LT1300 Wall Mount Load Cell Indicator Datasheet English 1.00 Introduction The LT1300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. The high

More information

The IQ240 panel mount load cell indicator is a precision digital indicator for load cell and strain gauge applications.

The IQ240 panel mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. IQ240 Panel Mount Load Cell Indicator Data sheet English 1.01 Introduction The IQ240 panel mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. The high

More information

OPERATING MANUAL FOR STREAM SELECTOR

OPERATING MANUAL FOR STREAM SELECTOR OPERATING MANUAL FOR STREAM SELECTOR Manufactured By: INSTRUMENTATION PVT. LTD. F-4 Memnagar complex, Opp. Petrol Pump Memnagar, Ahmedabad, India-380 052 Phone: +91-79-2749 5500/5600 FAX: +91-79-27417997

More information

The IQ300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications.

The IQ300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. IQ300 Wall Mount Load Cell Indicator Data sheet English 1.01 Introduction The IQ300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. The high

More information

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION UWE-1707 SERIES CONTENTS ELECTRONIC WEIGHING INDICATOR OPERATION MANUAL 1. SPECIFICATIONS PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE SCALE 2. INSTALLATION 3. ROUTINE OPERATION

More information

! User Manual Rev. 1

! User Manual Rev. 1 User Manual Rev. 1 Warnings For safety operation of the weighing indicator, please follow the following warning/ safety instructions: Calibration inspection and maintenance of the indicator are prohibited

More information

The PM1000 series is a universal 4 digit LED plug-on display for transmitters with 4-20mA 2 wire output and fitted with DIN43650 connector.

The PM1000 series is a universal 4 digit LED plug-on display for transmitters with 4-20mA 2 wire output and fitted with DIN43650 connector. PM1000 SERIES PLUG-ON DISPLAY BRIGHT LED DISPLAY INDICATION RANGE -999 TO +9999 FITS TO DIN 43650 CONNECTOR PLUG-ON TO ANY TRANSMITTER WITH 4-20MA OUTPUT EASY TO SCALE ON SITE ROBUST DESIGN SET POINT OPTION

More information

M260 Load Cell Amplifiers

M260 Load Cell Amplifiers E L 4 - + 2-3 1-2 M260 Load Cell Amplifiers Description. This fast start guide shows basic installation, setup, and calibration for most applications. For more specific information contact Minsterport.

More information

Masibus Automation And Instrumentation Pvt. Ltd.

Masibus Automation And Instrumentation Pvt. Ltd. Operator s Manual DIGITAL CONTROLLER 5006H Masibus Automation And Instrumentation Pvt. Ltd. B/30, GIDC Electronics Estate, Sector-25, Gandhinagar-382044, Gujarat, India Ph: 91 79 23287275-79 Fax: 91 79

More information

DMW. Digital Weighing Meter. Model Number Legend SALIENT FEATURES. Sensing

DMW. Digital Weighing Meter. Model Number Legend SALIENT FEATURES. Sensing Digital Meter Model Number Legend 1 2 3 1. Series 2. Output Type N: Indication Only R: 2 Relay Output M: 2 Relay Output + RS485 Modbus SALIENT FEATURES External Tare 96 x 48 Housing Dual Alarm Outputs

More information

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD AD-8923-BCD Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD4002137 2010 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated

More information

Trade Approved Version NMI: S520

Trade Approved Version NMI: S520 @Weigh BW/BWS Series Trade Approved Version NMI: S520 Weighing Indicator User s guide UGBW-E1.09 Table of Contents SECTION 1 INTRODUCTION. 1 SECTION 2 SPECIFICATIONS... 2 SECTION 3 INSTALLATION.3 3. 1

More information

AD-4532B. Digital Indicator INSTRUCTION MANUAL 1WMPD A

AD-4532B. Digital Indicator INSTRUCTION MANUAL 1WMPD A AD-4532B Digital Indicator INSTRUCTI MANUAL 1WMPD4001683A WARNING DEFINITIS The warnings described in this manual have the following meanings: A potentially hazardous situation which, if not avoided, could

More information

USER INSTRUCTION MANUAL FOR LOADCELL TRANSMITTER MODEL TDC/I/0550 (SOFTWARE: VER2A) INDEX

USER INSTRUCTION MANUAL FOR LOADCELL TRANSMITTER MODEL TDC/I/0550 (SOFTWARE: VER2A) INDEX USER INSTRUCTION MANUAL FOR LOADCELL TRANSMITTER MODEL TDC/I/0550 (SOFTWARE: VER2A) INDEX DOCUMENT NO: TDC 0550 MANUAL - 2 1.0) INTRODUCTION. PAGE 2 1.1) ABOUT THIS MANUAL. PAGE 2 1.2) INTRODUCTION. PAGE

More information

KW serial. User s guide. Bench Scales UGKW-E0302

KW serial. User s guide. Bench Scales UGKW-E0302 KW serial Bench Scales User s guide UGKW-E0302 KW serial bench scale user s manual Table of Contents SECTION 1 INTRODUCTION 1 SECTION 2 SPECIFICATIONS.. 2 SECTION 3 INSTALLATION.3 SECTION 4 KEY DESCRIPTIONS

More information

D-500s Indicator D LED. User Instructions. English Issue September SangR. Page 2. User s manual of Sang D-500s LED

D-500s Indicator D LED. User Instructions. English Issue September SangR. Page 2. User s manual of Sang D-500s LED D-500s Indicator D - 500 LED User Instructions English 8921-004210 Issue September 2010 1 2-2 0 1 0 D 6 0 0 4 6 7 5 Page 2 D-500s Indicator Page 2 Table of Content Section 1: Introduction. 1.1 Introduction

More information

+GF+ SIGNET ph/orp Transmitter Instructions

+GF+ SIGNET ph/orp Transmitter Instructions GF SIGNET 8750 ph/orp Transmitter Instructions ENGLISH 8750.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.

More information

DCS Series. Digital Counting Scale. Setup Manual

DCS Series. Digital Counting Scale. Setup Manual DCS Series Digital Counting Scale Setup Manual Revision 1.1 February 12, 1997 1996-1997 Transcell Technology, Inc. Contents subject to change without notice. Transcell Technology, Inc. 975 Deerfield Parkway

More information

CDD Carbon Dioxide Transmitter

CDD Carbon Dioxide Transmitter Introduction The OSA CO2 transmitter uses Infrared Technology to monitor CO2 levels within a range of 0 2000 ppm and outputs a linear 4-20 ma or 0-5/0-10 Vdc signal. The enclosure is designed to operate

More information

Gateway 1400 Reference Manual

Gateway 1400 Reference Manual Profibus-DP Gateway 1400 Reference Manual Copyright All Rights Reserved. No part of this document may be copied, reproduced, republished, uploaded, posted, transmitted, distributed, stored in or introduced

More information

SY021 Portable Load/Force Meter User instructions

SY021 Portable Load/Force Meter User instructions SY021 Portable Load/Force Meter User instructions Relates to firmware version 5.2 INTRODUCTION The SY021 is a portable load meter, which can indicate the load present on any attached cell. A dual channel

More information

UniPak UP448 Bridge Input Isolating Signal Conditioner

UniPak UP448 Bridge Input Isolating Signal Conditioner 6-wire Bridge Connection Excitation for to 350 Ω Load Cells 500 Volt (3-way) Isolation Excellent Accuracy (0.0%) DIP Switch Configuration Digital Input Tare Calibration RS85 Modbus RTU Superior Flexible

More information

Analogue Signal Transmitter AST 3 From prog.name A001A140. Operating instructions, Quick installation

Analogue Signal Transmitter AST 3 From prog.name A001A140. Operating instructions, Quick installation GB Analogue Signal Transmitter AST 3 From prog.name A001A140 Operating instructions, Quick installation Operating instructions, Quick operation Contents Introduction... 1 Operating instruction General...

More information

Transmitter WST 3 From prog. name W001A100. Technical Manual

Transmitter WST 3 From prog. name W001A100. Technical Manual GB Transmitter WST 3 From prog. name W001A100 Technical Manual Transmitter WST 3 Contents 1. Introduction Functions... 1-2 Technical data... 1-4 2. Installation General... 2-1 Electrical installation...

More information

Process displays For current and voltage

Process displays For current and voltage Features Voltage input TRMS-AC/DC up to 600 V Current input AC/DC 1 A, 5 A and shunt (precision resistor) 60 or 100 mv With 2 or 4 limits Display range can be linearised Min, Max, Hold functions 4 20 ma

More information

RTD-500 Precision RTD Simulator. Operations Manual.

RTD-500 Precision RTD Simulator. Operations Manual. RTD-500 Precision RTD Simulator Operations Manual. Page 1 of 16 Table of Content. Table of Content....1 1 Scope...3 2 Package Content....3 3 Technical Data....3 4 Preparation for use....4 4.1 Switching

More information

DAT 500 Series Digital / Analog Transmitters. Installation & Operating Manual (version 1.3) 9 Richmond St Picton, Ontario, Canada K0K 2T0

DAT 500 Series Digital / Analog Transmitters. Installation & Operating Manual (version 1.3) 9 Richmond St Picton, Ontario, Canada K0K 2T0 Precise Precise DAT 500 Series Digital / Analog Transmitters Installation & perating Manual version 1.3) Industrial Weighing Systems 9 Richmond St Picton, ntario, Canada K0K 2T0 Tel: 613-786-0016 Web:

More information

Dual channel temperature logger with two voltage inputs 0-5V Instruction Manual

Dual channel temperature logger with two voltage inputs 0-5V Instruction Manual LOGGER S0541 Dual channel temperature logger with two voltage inputs 0-5V Instruction Manual Instruction Manual for use of S0541 logger Instrument is designed for measurement and record of temperature

More information

ioselect Z-NET Z-SG Bridge Input Isolating I/O Module

ioselect Z-NET Z-SG Bridge Input Isolating I/O Module -wire Bridge Connection Excitation for to 30 Ω Load Cells 00 Volt (3-way) Isolation Excellent Accuracy (0.0%) DIP Switch Configuration Digital Input Tare Calibration RS8 Modbus RTU Superior Flexible Power:

More information

10 DIGITAL MICRO-PROCESS TOTALIZER METER with ALARMS / ANALOG OUTPUT / PULSE OUTPUT / RS-485. T Total Alarm x 1 + FRONT PANEL & KEY FUNCTIONS

10 DIGITAL MICRO-PROCESS TOTALIZER METER with ALARMS / ANALOG OUTPUT / PULSE OUTPUT / RS-485. T Total Alarm x 1 + FRONT PANEL & KEY FUNCTIONS 0 DIGITAL MICRO-PROCESS TOTALIZER METER with ALARMS / ANALOG OUTPUT / PULSE OUTPUT / RS-85 + FEATURES! Accuracy: + - 0.00% F.S. High brightness 0." LED; Rate of display range: 0~99999 Rate / Total decimal

More information

Installation and Maintenance Instructions for GC55 Differential Pressure Transducer Version 1.0 5/11

Installation and Maintenance Instructions for GC55 Differential Pressure Transducer Version 1.0 5/11 Installation and Maintenance Instructions for GC55 Differential Pressure Transducer Version 1.0 5/11 2011 Ashcroft Inc. 250 East Main Street, Stratford, CT 06614 USA Tel: 203-378-8281, Fax: 203-385-0402

More information

1. Introduction Page 2. Specifications Page Installation Page Key Functions Page Operations Page 6. Check Weighing Page 6

1. Introduction Page 2. Specifications Page Installation Page Key Functions Page Operations Page 6. Check Weighing Page 6 Contents: 1. Introduction Page 2 Specifications Page 2 2. Installation Page 3 3. Key Functions Page 4 4. Operations Page 6 Check Weighing Page 6 Accumulation Page 7 Parts Counting Page 7 Print (RS232C)

More information

575F-2051, 575F F-2051, 1000F F-2051, 1800F-2251

575F-2051, 575F F-2051, 1000F F-2051, 1800F-2251 Nokeval Manual Scaleable frequency display 7F-0, 7F- 000F-0, 000F-0 800F-0, 800F- No 008/09 v. Manufacturer: Nokeval Oy Yrittäjäkatu 700 NOKIA Finland Tel. +8 800 Fax. +8 0 email mail@nokeval.com http://

More information

USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER TPLR-96 Series

USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER TPLR-96 Series USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER TPLR-96 Series TEMPSEN DEVICES Plot No : 2&3, Balaji Nagar, 4 th Street, Mettukuppam, Thoraipakkam, Chennai-600097 Tele fax : +91-44-24581758,Mobil

More information

LAUREL. Laureate Digital Panel Meter for Load Cell & Microvolt Input ELECTRONICS, INC. Features. Description

LAUREL. Laureate Digital Panel Meter for Load Cell & Microvolt Input ELECTRONICS, INC. Features. Description Description LAUREL ELECTRONICS, INC. Features Laureate Digital Panel Meter for Load Cell & Microvolt Input 20, 50, 100, 250 & 500 mv ranges Span adjust from 0 to ±99,999, zero adjust from -99,999 to +99,999

More information

dual-line 6-dIGIt process MeteR

dual-line 6-dIGIt process MeteR dual-line 6-dIGIt process MeteR ProVu Se r i e S provu Model pd6000 0-20 ma, 4-20 ma, 0-5 V, 1-5 V, and ±10 V Inputs NEMA 4X, IP65 Front Universal 85-265 VAC or 12/24 VDC Input Power Large Dual-Line 6-Digit

More information

USER MANUAL V1.1. Setpoint unit 2022Setp

USER MANUAL V1.1. Setpoint unit 2022Setp USER MANUAL 0.4.005 V. Setpoint unit 0Setp INTRODUCTION 0Setp is a special edition of Nokeval 0 multi-input indicator. It is available in wall-mounted 800 enclosure or in a /8 DIN panel enclosure. Normally

More information

CDD4 Duct Carbon Dioxide Transmitter

CDD4 Duct Carbon Dioxide Transmitter Drill or punch a 1-1/8 or 1-1/4 hole in the duct at the preferred location and insert the probe into the hole to mark the enclosure mounting holes. Remove the unit and drill the four mounting holes. Clean

More information

CONFIGURATION HANDBOOK

CONFIGURATION HANDBOOK MINIATURE DIGITAL INDICATOR / CHASSIS THERMOSTAT CONFIGURATION HANDBOOK 94000 94000L LOREME 12, rue des Potiers d'etain Actipole BORNY - B.P. 35014-57071 METZ CEDEX 3 Phone 03.87.76.32.51 Contact : Commercial@Loreme.fr

More information

masibus : UT-94 Page 1 of 6 INTRODUCTION SPECIFICATIONS Dimensions Document Ref No. m98a/om/101 Issue No: 08

masibus : UT-94 Page 1 of 6 INTRODUCTION SPECIFICATIONS Dimensions Document Ref No. m98a/om/101 Issue No: 08 INTRODUCTION This is a micro-controller based Universal Transmitter which retransmits the Universal Inputs given, into the required Voltage or current. This instrument provides isolation at 4 levels: (a)

More information

PCE Americas Inc. PCE Instruments UK Ltd. 711 Commerce Way Units 12/13, Southpoint Business Park Suite 8 Ensign Way Jupiter, FL 33458 Hampshire / Southampton USA United Kingdom, SO31 4RF Phone: 561-320-9162

More information

Manual DTM Digital Pressure Transmitter

Manual DTM Digital Pressure Transmitter TetraTec Instruments GmbH Gewerbestr. 8 71144 Steinenbronn Deutschland E-Mail: info@tetratec.de Tel.: 07157/5387-0 Fax: 07157/5387-10 Manual Digital Pressure Transmitter *** VERSION 1.1 *** Update: 22.11.2006

More information

USER MANUAL FOR FIOA-0402-U-16

USER MANUAL FOR FIOA-0402-U-16 USER MANUAL FOR FIOA-0402-U-16 COPYRIGHT NOTICE This manual is a publication of Renu Electronics Pvt. Ltd. and is provided for use by its customers only. The contents of the manual are copyrighted by Renu

More information

Characteristics and functioning

Characteristics and functioning Characteristics and functioning 1/1 enod4-t Characteristics and functioning 216702-C_NU-eNod4T-ETH-E-0314 2/2 enod4-t Characteristics and functioning 216702-C_NU-eNod4T-ETH-E-0314 1 ENOD4 PRODUCT RANGE...

More information

High Voltage DC Meter

High Voltage DC Meter High Voltage DC Meter Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications

More information

Chennai Weighing System

Chennai Weighing System Chennai Weighing System No:27/1, New Boag Road, T.Nagar,Chennai- 600017 Cell:98400 29989, Ph:044-2432 0078 E-mail:chennai_weighing@yahoo.co.in Contact Person: D.Dayabaran CWS CONTENTS : Product Information

More information

Digital Indicator Solutions

Digital Indicator Solutions Digital Indicator Solutions Loop-Powered Attachable Compact Attachable Loop-Powered Smart Compact System Loop-Powered Compact Smart System Intelligent Smart System "Intelligent" Options Dual & Display

More information

DTM Distributed Transmitter Monitor

DTM Distributed Transmitter Monitor Distributed Transmitter Monitor 96 Communication Module The 96 functions as an RS232 to RS485 convertor and / or enables (32) modules to be networked together via Modbus. Combined with the -CFG configuration

More information

USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER MODEL : ARC2020

USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER MODEL : ARC2020 USER MANUAL MULTI COLOR MODEL : ARC2020 ACCSYS ELECTRONICS 140/6B, GOLDEN INDUSTRIAL ESTATE, JAWAHARLAL NEHRU ROAD, GERUGAMBAKKAM, CHENNAI - 600122 Tel: 044 60505599 / 60505511 E-mail : sales@accsyselectronics.com

More information

Model 620A Maintenance and Calibration

Model 620A Maintenance and Calibration Model 0A Maintenance and Calibration CAUTION! See Ch Pg, A & B below prior to connecting inputs to TB of the A PCB. Determine Rev. of the A board on the Model 0/0A for correct input connections. Failure

More information

SW-6KS / SW-15KS SW-15KM / SW-30KM / SW-60KM / SW-150KM SW-60KL / SW-150KL

SW-6KS / SW-15KS SW-15KM / SW-30KM / SW-60KM / SW-150KM SW-60KL / SW-150KL SW Series Super Washdown Scales SW-6KS / SW-15KS SW-15KM / SW-30KM / SW-60KM / SW-150KM SW-60KL / SW-150KL 1WMPD4002080A Warning Definition The warning described in this manual has the following meaning:

More information

Trend-1 Vibration Data Logger. User Manual

Trend-1 Vibration Data Logger. User Manual Trend-1 Vibration Data Logger User Manual Copyright College of Engineering-University of Basrah 2013 1 1. General Specifications 1.1 Features Thank you for being interested in purchasing Trend-1 Vibration

More information

North American Service Manual

North American Service Manual North American Service Manual MODEL PC SERIES Version 2 March 25, 1997 Revision B1.1 April 4, 2000 Prepared by TRANSCELL TECHNOLOGY, INC. TABLE OF CONTENTS Page Chapter 1: Introduction... 1-1 Chapter 2:

More information

HIGH-VOLTAGE DC METER

HIGH-VOLTAGE DC METER HIGH-VOLTAGE DC METER Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications

More information

The TRSUN is the complete solution for the remote monitoring of photovoltaic plants.

The TRSUN is the complete solution for the remote monitoring of photovoltaic plants. The TRSUN is the complete solution for the remote monitoring of photovoltaic plants. Monitoring of inverters and string combiner boxes (up to 64 devices); Complete configuration of the monitoring system

More information

Soft Starter Remote Operator. Section 1.0 Introduction 1.1 Important user information General Manual description...2.

Soft Starter Remote Operator. Section 1.0 Introduction 1.1 Important user information General Manual description...2. Section 1.0 Introduction 1.1 Important user information... 2 1.2 General... 2 1.3 Manual description...2 Contents Section 2.0 Specification 2.1 General technical data...3 2.2 Dimensions...3 Section 3.0

More information

AD-4401 WEIGHING INDICATOR A-IE

AD-4401 WEIGHING INDICATOR A-IE AD-440 WEIGHING INDICATOR 386-2A-IE 998 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any

More information

MIC-2 MKII, Multi-instrument DATA SHEET

MIC-2 MKII, Multi-instrument DATA SHEET MIC-2 MKII, Multi-instrument DATA SHEET Measurements All 3-phase AC measurements True RMS 4-Quadrant energy Power Quality Analysis Replaces analogue meters RS-485 Modbus RTU protocol TCP/IP Modbus (optional)

More information

Characteristics and functioning

Characteristics and functioning Characteristics and functioning 1/27 enod4-c Characteristics and functioning NU-eNod4C-ETH-E-1014_216706-A 1 ENOD4 PRODUCT RANGE... 4 1.1 General presentation... 4 1.2 Versions and options... 4 1.2.1 Versions...

More information

DIGITAL PROCESS CONTROLLER

DIGITAL PROCESS CONTROLLER 1 DPC 965 M R1 R2 R3 R4 TX RX PROG [OPERATOR S MANUAL] 2 Contents Section 1 specification.page 4 Section 2 system Programming Page 4 to 11 Section 3 Precautions..Page 11 Section 4 Connection Diagram.Page

More information

INSTALLATION MANUAL. LC 200 Electronic Overload Guard. Software versione PW0501 R 0.3

INSTALLATION MANUAL. LC 200 Electronic Overload Guard. Software versione PW0501 R 0.3 INSTALLATION MANUAL LC 200 Electronic Overload Guard Software versione PW0501 R 0.3 CONTENTS MAIN FEATURES LC 200 TECHNICAL FEATURES Page 2 SYMBOLS Page 3 WARNINGS Page 3 IDENTIFICATION DATA PLATE Page

More information

TRACKER 240 SERIES. Load Cell and Weighing Indicators. A Precision Measurement Instrument with Outstanding Features

TRACKER 240 SERIES. Load Cell and Weighing Indicators. A Precision Measurement Instrument with Outstanding Features TRACKER 240 SERIES Load Cell and Weighing Indicators A Precision Measurement Instrument with Outstanding Features TRACKER 240 SERIES INDICATORS Ratiometric Measurement Tare and Auto Transducer Excitation

More information

Note: the wiring of the dead-man buttons has been changed since October 2001.

Note: the wiring of the dead-man buttons has been changed since October 2001. UniOP epalm10 The epalm10 is a state-of-the-art handheld HMI device with a graphic display and a keypad. The rugged polyamide enclosure offers a high level of shock and environmental resistance making

More information

Analog Current Input - 16 Channel IC693ALG223. Current Ranges. Power Requirements and LEDs. Location in System. Analog Input Modules

Analog Current Input - 16 Channel IC693ALG223. Current Ranges. Power Requirements and LEDs. Location in System. Analog Input Modules Analog Input Modules Analog Current Input - 16 Channel IC693ALG223 The 16-Channel Analog Current Input module provides up to 16 single-ended input channels, each capable of converting an analog input signal

More information

Manual of Electrical Installation for Pressure Probes KS and KS-SIL2 series

Manual of Electrical Installation for Pressure Probes KS and KS-SIL2 series Manual of Electrical Installation for Pressure Probes KS and KS-SIL2 series GEFRAN SAFETY GUARANTEE Code 8594A Edition 2/204 Contents. General precautions pag 2 2. Transmitters with amplified analog output

More information

Scanner 2000 Steam Mass Flow Transmitter

Scanner 2000 Steam Mass Flow Transmitter 3352051/2 IM-P335-24 MI Issue 2 Scanner 2000 Steam Mass Flow Transmitter Installation and Maintenance Instructions 1. Safety information 2. Mechanical installation 3. Configuring software 4. Wiring procedures

More information

FLS M9.06. ph/orp MONITOR

FLS M9.06. ph/orp MONITOR FLS M9.06 /ORP MONITOR Safety Instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments

More information

MICRO SERIES PROCESS DISPLAYS LARGE DIGIT MODELS

MICRO SERIES PROCESS DISPLAYS LARGE DIGIT MODELS The MICRO SERIES PROCESS DISPLAYS LARGE DIGIT MODELS Mighty-5 DPM MODELS Micro-P & Mighty-1 Mighty-1 Micro-P ELECTRO-NUMERICS, INC. Introduction The Electro-Numerics family of Digital Panel Meters and

More information

BW/BWS Weighing Indicator

BW/BWS Weighing Indicator BW/BWS Weighing Indicator - 1 - CONTENTS 1. SPECIFICATION... 2 2. INTRODUCTION... 3 3. INSTALLATION. 4 Precautions 4 Parts 4 Installation.. 5 4. KEYS DESCRIPTION... 6 Key Board... 6 Secondary functions

More information

Multi-Function Smart Indicator User Manual

Multi-Function Smart Indicator User Manual Multi-Function Smart Indicator User Manual EC30006 Content 1 Notice... 2 2 Specification... 2 3 Main parameter... 3 4 How to fix... 3 5 Port... 4 6 On/Off... 4 7 Charging... 4 8 Display... 5 9 Screen keypad...

More information

Analog Amplifier FAA-27. Technical Manual. Flintec GmbH Bemannsbruch Meckesheim GERMANY.

Analog Amplifier FAA-27. Technical Manual. Flintec GmbH Bemannsbruch Meckesheim GERMANY. Analog Amplifier FAA-27 Technical Manual Flintec GmbH Bemannsbruch 9 74909 Meckesheim GERMANY www.flintec.com Table of Contents 1. Safety Instructions... 2 2. Declaration Of Conformity... 3 3. Front View,

More information

Dieter Wirth, Mess & Regeltechnik Neidlingen

Dieter Wirth, Mess & Regeltechnik Neidlingen Dieter Wirth, Mess & Regeltechnik Neidlingen Temperature controller Mod. BI50, BT31/32 and BI500, BT310/320/321 Made in Germany contents BT series 1. General Description... 2 2. Operating modes... 2 3.

More information

COMPACT FLEXIBLE ALL IN ONE INDUSTRIAL PUMP CONTROL ELECTRONIC MONITORING SYSTEM CONTROLLER

COMPACT FLEXIBLE ALL IN ONE INDUSTRIAL PUMP CONTROL ELECTRONIC MONITORING SYSTEM CONTROLLER COMPACT FLEXIBLE ALL IN ONE INDUSTRIAL PUMP CONTROL ELECTRONIC MONITORING SYSTEM CONTROLLER ONE EASY-TO-USE CONTROL. MANY APPLICATIONS. The EMS PRO is a rugged, all-in-one industrial controller designed

More information

MI101 / MI102 / MI102S INDICATOR

MI101 / MI102 / MI102S INDICATOR MI101 / MI102 / MI102S INDICATOR USER GUIDE Preface Thank you for purchasing the MI series indicator from Millennium Mechatronics Limited, New Zealand. The MI101/MI102 Indicator is an accurate, fast and

More information

600 g x 0.02 g DIGITAL BALANCE

600 g x 0.02 g DIGITAL BALANCE 600 g x 0.02 g DIGITAL BALANCE Model : GM-600P Your purchase of this DIGITAL BALANCE marks a step forward for you into the field of precision measurement. Although this DIGITAL BALANCE is a complex and

More information

DWYER INSTRUMENTS, INC. Series AVUL Air Velocity Transmitter. Specifications - Installation and Operating Instructions.

DWYER INSTRUMENTS, INC. Series AVUL Air Velocity Transmitter. Specifications - Installation and Operating Instructions. Series AVUL Air Velocity Transmitter Specifications - Installation and Operating Instructions Bulletin P-AVUL 3-49/64 [95.71] 2-43/64 [67.92] 1/2 NPS 3-3/16 [80.81] 1-19/32 [40.59] 31/32 24.58 3-33/64

More information

Nokeval. No Manual. Models 575F, 1000F, 1100F, Scalable counter for pulse sensors

Nokeval. No Manual. Models 575F, 1000F, 1100F, Scalable counter for pulse sensors No 90 Nokeval Manual Models 7F, 000F, 00F, 800-0 Scalable counter for pulse sensors Manufacturer: Nokeval Oy Yrittäjäkatu 700 NOKIA Finland Tel. +8 800 Fax. +8 0 Counter / batch controller XX-0 for pulse

More information

KCD-HP. KCD-HP200x, 300X. [Figures] Top : 1% Sensor probe Bottom left : 10% / 20% Sensor probe Bottom right : Holding bracket(optional) Measurement

KCD-HP. KCD-HP200x, 300X. [Figures] Top : 1% Sensor probe Bottom left : 10% / 20% Sensor probe Bottom right : Holding bracket(optional) Measurement Our CO2 gas sensors get a small deviation unlike NDIR Single type. So they keep long term stability. KCD-HP100x Excellent stability and accuracy - through testing and calibration with sophisticated process

More information

RS 485 Mini Modbus 1AO

RS 485 Mini Modbus 1AO RS 485 Mini Modbus 1AO Version 1.0 14/08/2014 Manufactured for Thank you for choosing our product. This manual will help you with proper support and proper operation of the device. The information contained

More information

OPERATIONAL MANUAL FOR 85XX

OPERATIONAL MANUAL FOR 85XX TECHNICAL DETAILS OPERATIONAL MANUAL FOR 85XX Ref. No. : m83/om/402 This is a versatile microcontroller based Scanner with a very high performance to price ratio. The features offered in this product are

More information

LAUREL. Laureate Digital Panel Meter for Process and Ratiometric Signals ELECTRONICS, INC. Features. Description

LAUREL. Laureate Digital Panel Meter for Process and Ratiometric Signals ELECTRONICS, INC. Features. Description LAUREL ELECTRONICS, INC. Laureate Digital Panel Meter for Process and Ratiometric Signals Features Reads process signals from ±200 mv to ±600V or ±2 ma to ±5A full scale Ratiometric mode for bridges and

More information

PROGRAMMABLE NUMERIC CONVERTER CPL 23 CONFIGURATION HANDBOOK

PROGRAMMABLE NUMERIC CONVERTER CPL 23 CONFIGURATION HANDBOOK PROGRAMMALE NUMERIC CONVERTER CPL 23 CONFIGURATION HANDOOK LOREME 12, rue des Potiers d'etain Actipole ORNY.P. 35014 57071 METZ CEDEX 3 Telephone 03.87.76.32.51 Fax 03.87.76.32.52 Contact: Commercial@Loreme.fr

More information

Calibration & Connectivity

Calibration & Connectivity Calibration & Connectivity TS-700 Series Digital Indicators This document supplements the User s Guide TS-700 MS TS-700 SS TS-700 WB Full Function/Advanced Function Digital Indicator www.trinerscale.com

More information

Batch Controller with numerical keypad, remote control inputs and 3 control / alarm / pulse outputs

Batch Controller with numerical keypad, remote control inputs and 3 control / alarm / pulse outputs Datasheet N410 User-friendly batch controller Reliable Batch Controller with numerical keypad, remote control inputs and 3 control / alarm / pulse outputs Advantages Save time and cost with the easy to

More information

IPM650 Intelligent Panel-Mount Display

IPM650 Intelligent Panel-Mount Display Quick Start Guide IPM650 Intelligent Panel-Mount Display Sensor Solutions Source Load Torque Pressure Multi Component Calibration Instruments Software www.futek.com Getting Help TECHNICAL SUPPORT For more

More information

EB300 Weighing Scale Operation Manual

EB300 Weighing Scale Operation Manual EB300 Weighing Scale Operation Manual EB300 User Manual Page 1 of 18 TABLE OF CONTENTS 1. Safety Warnings... 3 2. Features..... 3 3. Options.......... 4 4. Product packaging.. 4 5. Display and keyboard.....

More information

MODEL: M5DY SEN TRONIC AG CURRENT LOOP SUPPLY. [2] POWER INPUT AC Power M: V AC (Operational voltage range V, Hz)

MODEL: M5DY SEN TRONIC AG CURRENT LOOP SUPPLY. [2] POWER INPUT AC Power M: V AC (Operational voltage range V, Hz) Super-mini Terminal Block Signal Conditioners M5-UNIT CURRENT LOOP SUPPLY Functions & Features Powering a 0 ma DC current loop Applicable to smart transmitters Isolation between the input and output Fast

More information

TR150 Digital Readout Unit SL755. Impact Test Equipment Ltd & User Guide. User Guide

TR150 Digital Readout Unit SL755. Impact Test Equipment Ltd   &   User Guide. User Guide TR150 Digital Readout Unit SL755 Impact Test Equipment Ltd www.impact-test.co.uk & www.impact-test.com User Guide User Guide Impact Test Equipment Ltd. Building 21 Stevenston Ind. Est. Stevenston Ayrshire

More information

Installation and User Manual. version 1.00 MODWF 2014/53/UE EN : /30/UE EN55022:2010 EN :2005 EN :2007

Installation and User Manual. version 1.00 MODWF 2014/53/UE EN : /30/UE EN55022:2010 EN :2005 EN :2007 Installation and User Manual version 1.00 MODWF 2014/53/UE EN300 328:2016 2014/30/UE EN55022:2010 EN61000-6-2:2005 EN61000-6-4:2007 SYSTEM IDENTIFICATION Load Cell Systems reserves the right to change

More information

USB-1208LS Specifications

USB-1208LS Specifications Specifications Document Revision 1.1, February, 2010 Copyright 2010, Measurement Computing Corporation Typical for 25 C unless otherwise specified. Specifications in italic text are guaranteed by design.

More information

Zenith DS9000 Dispensing System

Zenith DS9000 Dispensing System Page Date: 04/2009 Zenith DS9000 Dispensing System Installation & Operation Manual Page 2 Table of Content Introduction... 3 Control Specification... 4 Wiring Diagram... 5 Wiring Instructions... 6 Field

More information

Trident and Trident X2 Digital Process and Temperature Panel Meter

Trident and Trident X2 Digital Process and Temperature Panel Meter Sign In New User ISO 9001:2008 Certified Quality System Home Products Online Tools Videos Downloads About Us Store Contact Policies Trident and Trident X2 Digital Process and Temperature Panel Meter Products

More information

GS100P. Weighing Controller. Contents. Order Codes

GS100P. Weighing Controller. Contents. Order Codes GS100P Weighing Controller 1 This intelligent weighing controller comes precalibrated to suit either a 4-wire or a 6-wire strain gauge input, and has a number of advanced functions designed specifically

More information

Model 332 Differential Pressure Transmitter

Model 332 Differential Pressure Transmitter Model Differential Pressure Transmitter USERS GUIDE Please read carefully before using. Manufactured by: FURNESS CONTROLS LIMITED Beeching Road, Bexhill, East Sussex, TN9 LJ, England. Telephone (044) 706

More information

H3 Series. Nanmac Corporation. Dual-Line 6-Digit Temperature Meter.

H3 Series. Nanmac Corporation. Dual-Line 6-Digit Temperature Meter. H3 Series Dual-Line 6-Digit Temperature Meter Temperature J, K, T, E, R, S, B, N, C Thermocouples 100 or 1000 Ω Platinum, 10 Ω Copper, 120 Ω Nickel RTDs 1 or 0.1 Resolution Averages up to 10 RTD Sensors

More information

UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS

UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS EN0I-6041 12/99 PRODUCT SPECIFICATION SHEET OVERVIEW The UDC 1000 and UDC 1500 are microprocessor-based 1/16 DIN and 1/8 DIN controllers

More information

Vibration-Resistant Weighing Indicator

Vibration-Resistant Weighing Indicator Vibration-Resistant Weighing Indicator Powerful vibration-cancelling function (High Performance Digital Filter) High speed sampling (100 times/second)/high accuracy Circuits equipped with powerful noise

More information

FLS M9.06 PH/ORP MONITOR. safety instructions. packing List

FLS M9.06 PH/ORP MONITOR. safety instructions. packing List FLS M9.06 PH/ORP MONITOR safety instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments

More information