FIDAMAT 5E-P, -AP, -G Gas analyzers for Measurement of Total Hydrocarbon Content 7MB Instruction Manual

Size: px
Start display at page:

Download "FIDAMAT 5E-P, -AP, -G Gas analyzers for Measurement of Total Hydrocarbon Content 7MB Instruction Manual"

Transcription

1 FIDAMAT 5E-P, -AP, -G Gas analyzers for Measurement of Total Hydrocarbon Content 7MB Instruction Manual Order no. C79000-B5276-C3-03

2 Contents Page Introduction 4. General 4.2 Staff Qualifications 4 2 Application 5 3 Design 6 4 Mode of Operation 7 5 Technical Data 8 6 Installation 6. Mounting 6.2 Gas Supply 6.3 Electric Connections 2 7 Startup, Calibration 5 7. Startup Calibration FIDAMAT 5E-G without zero and calibration gas connections, FIDAMAT 5E-AP FIDAMAT 5E-P without zero and calibration gas connections 6 8 Operation 7 8. Description of Control Panel Input Example Coding and Organization of Operating Function Groups Organization of Data Storage 20 9 Warnings and Error Messages 2 9. Warnings Faults 22 0 Spare Parts and Accessories FIDAMAT 5E-AP FIDAMAT 5E-P FIDAMAT 5E-G 29 Appendix A-. Gas Flow Diagrams A-.2 Layout Diagrams of Important Electronic Components A-4.3 Summary of Operating Functions A-5.4 Special Sheet for Storage of Current Coding A-23 2

3 Note Your attention is drawn to the fact that the contents of this Instruction Manual are not part of a previous or existing agreement, commitment or statutory right and do not change these. All commitments on the part of Siemens are contained in the respective sales contract which also contains the complete and solely applicable warranty conditions. These warranty conditions in the contract are neither extended nor limited by the contents of this Instruction Manual. Also note that for clarity reasons this Instruction Manual cannot describe every possible problem in conjunction with the use in systems. Should you require further information, or should particular problems occur which are not handled in sufficient depth in this Manual, help can be requested through your local Siemens office or representative. Terms with the following meanings are used in this Instruction Manual and in the warning information on the product: Danger in the sense of this Manual and the warning information on the product itself means that death, severe personal injury and/or substantial damage to property will occur if the appropriate safety precautions are not observed. Warning in the sense of this Manual and the warning information on the product itself means that death, severe personal injury and/or substantial damage to property can occur if the appropriate safety precautions are not observed. Caution in the sense of this Manual and the warning information on the product itself means that slight personal injury and/or damage to property can occur if the appropriate safety precautions are not observed. A note in the sense of this Manual is important information on the product or the respective part of the Manual to which particular attention should be paid. 3

4 Introduction. General Warning! Certain parts in this electrical device contain dangerous voltages during operation. Failure to observe the warnings could therefore result in severe personal injury and/or damage to property. Only suitably qualified staff should work on this device or in its vicinity. These should be thoroughly familiar with all warnings and maintenance measures according to this Manual. Correct and safe operation of this device is dependent on proper handling, installation, operation and maintenance. This device has left the factory in a perfect condition as regards safety. The notes and warnings in this Manual must be observed by the user if this state is to be maintained and hazard-free operation of the device assured..2 Staff Qualifications A qualified person in the sense of this Instruction Manual and the warning information is one who is familiar with the installation, construction and operation of the device and who has the appropriate qualifications, e.g.: is trained and authorized to energize, de-energize, ground and tag circuits and devices in accordance with established safety practices; is trained in the proper care and use of protective equipment in accordance with established safety practices; is trained in first aid. 4

5 2 Application The FIDAMAT 5E is an analyzer for measuring the concentration of total hydrocarbons. It operates according to the principle of flame ionization. The analyzer is suitable for the measurement of all gaseous hydrocarbons. The connection of a heated line is provided in order to prevent condensation and adsorption in the gas supply line. Caution! Explosive gas mixtures must not be measured using the FIDAMAT. The analyzer must not be used in potentially explosive atmospheres. Order No. 7MB MB MB Device type FIDAMAT 5E-P rack-mounted device for process measurements, combustion gas H2 FIDAMAT 5E-AP rack-mounted device for car exhaust measurements, combustion gas H2/He FIDAMAT 5E-G rack-mounted or benchtop devices for measurements of gases of highest-grade purity, combustion gas H2/He 5

6 3 Design The device is available for installation in a 9-inch rack or 9-inch cabinet, the FIDAMAT 5E-G also als benchtop unit. The FIDAMAT comprises two main sections, the analyzer section and the electronics with front panel controls. Analyzer section The analyzer section consists of the oven with detector chamber and various restrictors. The oven is accessible from the side to enable modifications or repairs to be carried out in the installed condition without loosening the electric connections or gas piping connections. The analyzer section also contains pressure regulators, solenoid valves, pressure sensors and a flow sensor. Electronics The electronics consist of the input board which is integrated into the front panel, the motherboard with preamplifier board and the control board. The motherboard contains the EPROMs for the software, the EEPROM for the parameter sets and the switch for the write-protected area on the EEPROM. 6

7 4 Mode of Operation Electrons are released when hydrocarbons are combusted in a hydrogen flame. These electrons are collected at an electrode by means of an electric field and measured using a highly sensitive amplifier. The current is proportional to the quantity of organically-bound C atoms in the sample gas. The sample gas is passed into the measuring chamber via an obstruction-resistant fused silica restrictor. The sample gas is mixed in the measuring chamber with hydrogen or hydrogen/helium (4:6) or hydrogen and a specific amount of air and routed via the nozzle into the combustion chamber. The hydrogen pressure is held constant by a pressure regulator. The balanced system of restrictors and pressure regulators ensures that the sample gas pressure is kept constant. The FIDAMAT operates largely automatically. If the parameters (pressures, temperatures) are set, the device starts up automatically when switched on and ignites when the setpoint temperature has been reached. The hydrogen and combustion air pressures are measured when switching on, and the control panel indicates if they are incorrectly set V DC Flame 2 Quartz nozzle 3 Combustion air 4 Sample gas 5 H2/He or H2 6 Measuring electrode 7 Amplifier Fig. 4 Principle of flame ionization detector (FID) 7

8 5 Technical Data Number of measuring ranges Switching ratio Autoranging Max. 4, freely parameterizable Optional Selectable Hysteresis: Switchover to larger range at 90 % of span of smaller range Switchover to smaller range at 80 % of span of smaller range Output signals At terminal strips X and X3. A cable with a plug connector (Order No. W79070-U60-M25) can be connected to these terminal strips Analog 0/2/4 to 20 ma, max. resolution 0. %, max. load 750 Ω Digital Programming facilities (for setting of device-specific parameters using control panel) RS232 serial interface as V.24 or TTY (20 ma) Measuring ranges, measured-value storage, time constant, 3-stage code for protection against unauthorized and unintentional use etc. Device output see Section 6.3 electrical connections Measured-value display Resolution of digital display Digital concentration display (5 digits with floating decimal point) 0. % of measured value Ripple of output signal <0.5 % of output signal Noise FIDAMAT 5E-P FIDAMAT 5E-AP FIDAMAT 5E-G Detection limit FIDAMAT 5E-P FIDAMAT 5E-AP FIDAMAT 5E-G Repeatability Characteristic Long-term drift Zero Sensitivity <0.5 % of 0 ppm C <0.5 % of 0 ppm C3 <5 vpb C 0. ppm C 0. ppm C vpb C < % of full-scale value Linearization error < % of f.s.v. <2 % of span/week, <50 vpb C/4 h with a full-scale value of vpm As for zero 8

9 Response time (T90 time) with time constant 0 s, set using function 3 Warm-up time Analyzer chamber temperature Permissible storage temperature Permissible ambient temperature (operation) Inlet pressure for hydrogen, hydrogen/helium and combustion air sample gas FIDAMAT 5E-AP FIDAMAT 5E-P, -G zero gas, calibration gas FIDAMAT 5E-AP (via sample gas inlet) FIDAMAT 5E-G FIDAMAT 5E-P Position of use < s Approx. 2 to 3 hours 9 C 30 to +70 C 5 to 45 C 30 psig 0 psig 5 psig 8 psig 5 psig 8 psig Front panel vertical Change in display through influencing variables Mains voltage variations (+0, -5 %) < % of measured value Ambient temperature variations < % of measured value/0 K Atmospheric pressure variations < % of measured value /50 mbar Frequency variations < % of measured value for the permissible frequency range Influence of position < % of measured value with inclination <5 Power supply Mains connection Power consumption AC 20 V, 48 to 63 Hz Approx. 350 VA in start-up phase, approx. 50 VA during operation Degree of protection IP 20 to DIN Dimensions Weight Material of parts in contact with sample gas See dimensional drawing 26 kg PTFE, quartz (nozzle, fused silica), graphite (gaskets), stainless steel.457 (36 Ti) Gas consumption data (sample gas at standard pressure) combustion air Approx. 350 ml/min (approx. 2 l/h) hydrogen/helium (5E-AP) Approx. 65 ml/min (approx. 4.0 l/h) hydrogen (5E-P, -G) Approx. 25 ml/min (approx..5 l/h) sample gas Approx l/min (approx. 20 l/h) Depending on the actual configuration of the analyzer and the sampling system the sample gas flow may vary up to 50 %. 9

10 FIDAMAT 5E rack-mounted unit Caution! Only mount FIDAMAT 5E on sliding rails or angled rails Sample gas connection: /4" screw connection Sample gas outlet: /8" PVDF screw connection Auxiliary gas connections: /4" screw connecction 4 All measures are in millimeters (mm) FIDAMAT 5E-G as bench top unit Fig. 5. Dimension drawings 0

11 6 Installation 6. Mounting The FIDAMAT 5E should be installed or mounted in a vibration-free location. Warning! The device must not be used to measure explosive gases. Operation in potentially explosive atmospheres is not permitted. The permissible ambient temperature (Section 5, Technical Data) must be observed during operation. The FIDAMAT 5E must only be mounted on sliding rails or angled rails. See Section 5 for dimensional drawings. 6.2 Gas Supply Warning! When measuring toxic gases, the exhaust gas must be routed such that there is no danger to the environment. Gases which may lead to explosive mixtures must not be measured using the FIDAMAT. A gas-tight hose or steel piping can be used for the sample gas inlet and outlet depending on the appliction, and must be connected leak-free to the gas connection. The sample gas must be supplied with a pressure of 0 psig (5E-AP) or 5 psig (5E-P, -G), the supply gases with a pressure of 30 psig. The exhaust line from the FIDAMAT 5E must always be routed with a downward gradient since water condenses in it. Mains 20 V X H2 - He comb. air X3 sample gas Fig. 6. Rear panel of FIDAMAT 5E-AP

12 H2 Mains 20 V/ 230 V X X2 comb. air X3 zero gas sample gas cal. gas Fig. 6.2 Rear panel of FIDAMAT 5E-P sample gas H2 Mains 20 V/ 230 V X X2 comb. air X3 zero gas cal. gas Fig. 6.3 Rear panel of FIDAMAT 5E-G 6.3 Electric Connections Warning! The following must be observed for the electric connections: VDE 000 Regulations for the installation of power systems with mains voltages below 000 V. A mains disconnection device must be provided in the building installation (see rating plate for loading capacity). Check that the local power supply agrees with that specified on the rating plate of the device. Connect the required signal cables to the 25-pin trapezoidal plugs X and X3 (FID 5E-AP) or X, X2 and X3 (FID 5E-P, -G) according to the pin assignment diagrams. 2

13 Pin assignment diagram FIDAMAT 5E-P, -G Device side Pin Connection side MR X Device side Pin 4 Connection side Analog current MR 2 MR 3 MR 4 MR 4 Measuring range identification Contact closed = MR active Analog current Vacant Analog current Calibrate zero (funct. 5) Calibrate span (funct. 8) Measure Calibrate Status signal 2 Ready Not Ready Status signal 3 3 Device side Connection side Pin Valve for zero gas E2 3 Valve for calibration gas Valve for sample gas Valve for hydrogen E X2 Contact open = valve closed (when off) E closed triggers a single autocalibration E2 closed = E is activated Device side Pin Connection side Vacant Ground Warning No warning X3 2 3 Screen TXD RXD V24 7 GND 9 RXD RXD TXD+ TTY 25 TXD 3

14 Pin assignment diagram FIDAMAT 5E-AP X Device side Connection side Device side Connection side Pin Pin MR 4 Analog current MR 2 MR 3 MR 4 Measuring range identification Contact closed = MR active Analog current Vacant 4 MR 4 9 Calibrate zero 5 20 Calibrate span No warning Signal contact Calibrate (contact open = calibration being carried out Measure Calibrate Status signal 2 Ready Not Ready Status signal 3 3 X3 Screen 2 3 TXD RXD V24 7 GND 9 RXD RXD TXD+ TTY 25 TXD 4

15 7 Startup, Calibration 7. Startup You should be acquainted with the operation of the FIDAMAT 5! (see Section 8) A prerequisite for correct startup is the presence of hydrogen/helium or hydrogen and combustion air with the appropriate inlet pressure of 30 psig and the sample gas with an inlet pressure of 0 psig (5E-AP) or 5 psig (5E-P, -G). The settling time differs depending on the previous use of the device. It should be 2 to 3 hours for devices which have been used normally. The time is longer for the initial startup or with contaminated devices. The device indicates the current software version when the power supply is connected. Start then appears in the display. The device is in the status Not ready. The hydrogen/helium and sample gas/combustion air pressures are checked immediately at the beginning and when the oven temperatures have been reached. If the device has already been used correctly, the startup is usually carried out without readjustment of the pressure regulators (adjustable screws on front panel). If the message H2" appears in the display, the hydrogen pressure must be adjusted to 000 mbar (980 to 020 mbar) by means of function 35 and using the adjustable screw on the front panel. The device automatically returns to Start" at the end of function 35. If the message Air appears in the display, it must first be made sure, that combustion air (30 psig) and sample gas (0 psig with FID 5E-AP, 5 psig with FID 5E-P and -G) are supplied with the correct pressures. Then the sample gas pressure (combustion air pressure) inside the device must be adjusted to 340 mbar by means of function 36 and using the adjustable screw on the front panel. The device then automatically returns to Start. Once the temperatures have been reached, the device automatically ignites the flame. If there is an appropriate sample gas pressure (0 psig with FID 5E-AP, 5 psig with FID 5E-P and -G) the device automatically switches to the measure mode. If there is no pressure at all or insufficient pressure the device switches to the stand-by mode and waits until there is sufficient sample gas pressure. 7.2 Calibration The calibration should be carried out using gases which represent the gas to be measured as closely as possible. Zero will always be calibrated using function 5, span using function FIDAMAT 5E-G without zero and calibration gas connections, FIDAMAT 5E-AP The calibration is carried out via the sample gas line where the calibration gas pressure must be 8 psig (FID 5E-AP) or 5 psig (FID 5E-G). The setpoint for the sensitivity (function 7) must be entered prior to the calibration. This value can be checked using function 6 - one value must always be entered for function 7. Once the contacts have been closed for longer than sec. the calibration is started and finished after approx. 20 sec. The adjustment by floating contacts is only possible in the meas mode. 5

16 7.2.2 FIDAMAT 5E-P without zero and calibration gas connections The sample gas pressure must be lower than the pressures for zero and calibration gas. During calibration the excess zero and calibration gas flows back through the sample gas line. 6

17 8 Operation 8. Description of Control Panel mg/m 3 ppm C ppm C3 ppm C6 warning autocal lim not ready 0 C 0 Switch over measuring range 02 Display measuring range 05 Calibrate zero 06 Display sensitivity setpoint 07 Set sensitivity setpoint Calibrate sensitivity to setpoint 09 Display limit (lim) 35 Display and adjust H2 pressure Display and adjust sample gas pressure MEAS CAL Manual adjustment Range No. Relay No. Time Identifier Off/On ENTER Setpoint Limit No Fig. 8. Control panel of FIDAMAT 5E () Display of selected measuring range to 4 (2) Large display: concentration display, parameter display, fault display, status display (3) Small display: display of selected function codes; if the small display is dark, the large display outputs the measured value (4) Clear key to return to basic status or to delete faulty input (5) Display of operating states and concentration units: "ppm C etc. = display of concentration unit warning" = common status for warnings "autocal" = the LED lights up when autocal is activated, the LED flashes during the automatic calibration "lim" = limit value exceeded "not ready" = common status for error messages (6) Function table for the most important functions (7) Switchover from measurement to calibration mode and simultaneous signalling of these statuses via a contact 7

18 (8) Operator prompting: the corresponding LED flashes here if further inputs are required for the entered functions (9) Every input must be terminated by the ENTER key apart from the code number (0) Flame monitor: The LED lights up when the flame is alight and goes out when the flame is off. () LED Autoranging lights up when the corresponding function (20) is active. (2) Dot key with the following meaning: Decimal point Change of sign when entering values Start point of code entries 8.2 Input Example You wish to switch over the measuring range Every input - apart from the codes - must be terminated by ENTER. Press the key in the event of a faulty input. The analyzer is in measurement mode. This means: The green LED lights up for MEAS next to MEAS/CAL. Input:. Since the desired function is in function group, it is not necessary to enter any further codes. Input: press the MEAS/CAL key The green LED lights up next to CAL. The calibration status is thus signalled via pins 24/25 on plug X. Input: The appears in the small display and thus confirms the input. Input: ENTER The measured value disappears in the large display, and a line appears on the right. The green LED lights up in field 6 in front of Switch over measuring range. The green LED simultaneously flashes in front of Range No. in field 8 and thus requests the input of the new range number. Input: 3 Measuring range 3 is selected. The input is output in the large display. Input: ENTER The measured value appears in the large display. The LED 3 lights up in field and indicates that range 3 is active. Input: press the MEAS/CAL key The green LED lights up next to MEAS. If a control panel is present, the status Measure is signalled to it. 8

19 8.3 Coding and Organization of Operating Function Groups The functions are combined into groups according to their type: Function groups Types of function Display functions Calibration and adjustment functions Functions for input of application parameters Functions for input of device parameters To prevent important parameters from being deleted or modified by mistake, there are several function levels which can only be accessed using special codes. Note that the decimal point key must be pressed before entering a code. The factory settings for the codes of the three levels are: Level. Level Level The codes can be redefined using function 50. Function 50 also enables the device protection by codes to be switched off for a certain time or permanently. Code can also be set to zero so that it is omitted. The CAL key must be pressed to signal an intervention on the device to the control room if applicable. The device is automatically recoded by pressing the MEAS key. All operations permissible for a level can be carried out if you are in level, 2 or 3. Higher function levels also include the lower function levels. For example, from function level 3 you can execute all operations of levels 0 to 3, and from level 2 all operations of levels 0 to 2. A description of the operation functions is provided in the appendix. 9

20 8.4 Organization of Data Storage The EPROM and the EEPROM contain three parameter sets: Basic factory-set data User-specific data Current data. All parameters required to operate the FIDAMAT 5E are stored in the EPROM and are protected by hardware (basic factory-set data). These data can be loaded into the current parameter memory using function 53. The user-specific data set is overwritten in the process, however. The basic factory-set data can only be overwritten by a hardware intervention. A second parameter set (user-specific data) is present in an area of the EEPROM which can be overwritten. The user can store a standard parameter set for his application here (see functions 55 and 57). The third data set (current data) is the set which is automatically loaded when the analyzer is switched on. This data set is modified according to the inputs. Note therefore when switching off the analyzer and restarting that it is the current data set which was parameterized when the analyzer was switched off which is loaded and not the user-specific data set. EPROM RAM EEPROM Program Current data Function 57 User-specific data Function 55 Basic factoryset data Automatically when required Current data Restart Function 53 Fig. 8.2 Organization of data storage 20

21 9 Warnings and Error Messages 9. Warnings The symbol precedes each warning. The LED warning flashes on the control panel. The cause of the warning can be scanned using ENTER. Further measurements are possible in this state. The analyzer may fail if the cause of the warning is not eliminated. Number in display Cause of warning Pt00 temperature of physical section faulty or outside the limits Temperature of electronics section outside the limits (0 to 80 C) or sensor faulty Control air flow too low Temperature of physical section outside the limits (0 to 80 C) Hydrogen pressure outside the limits Sample gas pressure outside the limits Flame out Difference between actual and former zero adjustment >6 % of full-scale value Output current used for internal value Measured value exceeds full-scale value of largest measuring range Incorrect device configuration ( autocal only) Incorrect measured value Causes, notes The limits defined in function 37 have been violated. Hydrogen input pressure is too high or too low; this warning may also be produced if the limits are too close and there are extreme variations in atmosph. pressure. The limits defined in function 38 have been violated. Combustion (control) air pressure is too high or too low; sample gas inlet faulty (filter lines); this warning may also be produced if the limits are too close and there are extreme variations in atmospheric pressure. See function 62. Call function 96 to eliminate warning. define full-scale value (funct. 2). Inform service. Recalibrate device using function(s) 5 and/or 8. 2

22 9.2 Faults The symbol precedes each fault. The LED not ready flashes on the control panel. One of the following error messages appears when you press ENTER. Number in display Error message Comment Slave faulty Hardware fault Heating has switched off EEPROM jumper must be set Parameter storage test not carried out EPROM faulty Sensitivity too low Oven temperature outside limits or Pt00 faulty Flame monitor faulty Sample gas pressure too high or too low or pressure sensor faulty Hydrogen pressure too high or too low or pressure sensor faulty High-voltage outside limits Operating voltages outside setpoints Oven temperature setpoint exceeded or fallen below Combustion air/sample gas pressure too low Flame does not ignite (30 min) Pull mains plug immediately, inform service Inform service RESET Inform service See u 8, next page Order replacement from manufacturer Incorrect calibration gas, incorrect measuring range, see functions 2 and 6 Check contacts Check contacts Check cylinder pressure, readjust, check pressure sensor Check cylinder pressure, readjust, check pressure sensor Check whether high-voltage line is connected Check power supply and connections on transformer Combustion air line interrupted Check gas flows 22

23 Number in display Cause of error Comment 8 Parameter storage test not carried out A: Analyzer does not react to ENTER B: One of the following error numbers is displayed when you press ENTER. A further number may be displayed when the error has been eliminated. Press dot key for 30 sec until function 50 appears in the small display; recode the analyzer. Then reload basic data into RAM using function 53 or 55. Correct the value in the corrresponding function using the following table Code Code 2 Code 3 Cause Assignment of current output Number of measuring ranges Measuring range Sensitivity adjustment Sensitivity setpoint Start-of-scale value Full-scale value Limits Zero Sensitivity Limits of time constant T90 Dynamic noise rejection Baud rate Transmission procedure Start, end, don t care characters Calibration of analog current Adjustment of preamplifier Check/ input using function No. 62, identifier / / 7 / Contact service Contact service 23

24 0 Spare Parts and Accessories 0. FIDAMAT 5E-AP Order No. Flame ionization detector, complete FID cover complete with thermocouple line Quartz nozzle for FIDAMAT 5E-AP Clamping screw in the FID Nut for M5 connection Outer ring for M5 connection Graphite gasket for M5 connection Clamping ring for M5 connection Nut for M7 x 0.75 connection, tube.5 to 3.0 Outer ring for M7 x 0.75 connection, tube.5 to 3.0 Graphite gasket for M7 x 0.75 connection, pipe.5 Clamping ring for M7 x 0.75 connection, pipe.5 Graphite gasket for M7 x 0.75 connection, pipe 3.0 Clamping ring for M7 x 0.75 connection, pipe 3.0 Coarse filter Sample gas restrictor (fused silica), complete, 5E-AP Crimp restrictor, hydrogen, 5E-AP Crimp restrictor 2, combustion air, 5E-AP Crimp restrictor 3, auxiliary air, 5E-AP Crimp restrictor 4, exhaust, 5E-AP C7945-A3405-B578 C7945-A3405-B50 C79402-Z282-C2 C792-A3003-C5 C7945-A3040-D26 C7945-A3040-D2 C7945-A3040-D02 C7945-A3040-D2 C7945-A3040-D27 C7945-A3040-D22 C7945-A3040-D03 C7945-A3040-D3 C7945-A3040-D05 C7945-A3040-D5 C7945-A3405-C69 C7945-A3405-B90 C7945-A3405-C63 C7945-A3405-C226 C7945-A3405-C224 C7945-A3405-C227 Crimp restrictor 5, sample gas, 5E-AP C7945-A3405-C92 Temperature sensor, complete, FID oven and physical section Flat heater element, FID oven, top Flat heater element, FID oven, bottom Cable connector for input signal Cable connector for FID voltage C7945-A3405-B03 C7945-A3405-B02 C7945-A3405-B0 C7945-A3405-B09 C7945-A3405-B0 24

25 Order No. Cable connector for pressure sensor for combustion air Cable connector for pressure sensor for hydrogen Solenoid valve for combustion air, complete Solenoid valve for H2, sample gas, combustion air complete Solenoid valve for control air, combustion air, H2 Gasket, diameter 4 mm Gasket, diameter 6 mm Gasket, diameter 0 mm O-ring, diameter 0 mm Pressure controller, hydrogen, combustion air/sample gas PTFE gasket for M7 x 0.75 connection, tube.5 Elbow compression gland C7945-A3405-B05 C7945-A3405-B06 C7945-A3040-B20 C7945-A3040-B206 C79402-Z87-A5 C7945-A3040-C33 C7945-A3040-C354 C7945-A3040-C353 F79402-E87-A C7945-A3405-B20 C7945-A3040-D0 C75304-Z209-C2 Union nut for pipe R/4" Clamping ring, front, pipe R/4" Clamping ring, rear, pipe R/4" available only as set of parts, order no. C7945-A3405-D62 Analyzer fuse, T 6.3 A/20 V Motherboard Preamplifier board Control board ROM, motherboard position D3, 5E-AP, master ROM, motherboard position D4, 5E-AP, slave Display board (front panel) Touch pad keyboard with labels, German Touch pad keyboard with labels, English Touch pad keyboard with labels, French Touch pad keyboard with labels, Spanish Touch pad keyboard with labels, Italian W79054-L0-T630 C7945-A3405-B507 C7945-A3405-B508 C7945-A3405-B509 S7960-G52-A90 S7960-G52-A902 C7945-A320-A502 C7945-A3405-D52 C7945-A3405-D53 C7945-A3405-D54 C7945-A3405-D55 C7945-A3405-D56 25

26 0.2 FIDAMAT 5E-P Order No. Flame ionization detector, complete FID cover complete with thermocouple line Quartz nozzle for FIDAMAT 5E-P Clamping screw in the FID Nut for M5 connection Outer ring for M5 connection Graphite gasket for M5 connection Clamping ring for M5 connection Nut for M7 x 0.75 connection, tube.5 to 3.0 Outer ring for M7 x 0.75 connection, tube.5 to 3.0 Graphite gasket for M7 x 0.75 connection, pipe.5 Clamping ring for M7 x 0.75 connection, pipe.5 Sample gas filter, complete Gasket, sample gas filter Gasket, sample gas filter Filter cylinder 3 µm Spring washer for sample gas filter Sample gas restrictor (fused silica), complete, 5E-P Crimp restrictor, hydrogen, 5E-P Crimp restrictor 2, combustion air, 5E-P Crimp restrictor 3, auxiliary air, 5E-P Crimp restrictor 4, exhaust, 5E-P Crimp restrictor 5, zero gas, calibration gas(es), 5E-P Temperature sensor, complete, FID oven and physical section Flat heater element, FID oven, bottom Flat heater element, FID oven, top C7945-A3405-B50 C7945-A3405-B50 C79402-Z282-C C792-A3003-C5 C7945-A3040-D26 C7945-A3040-D2 C7945-A3040-D02 C7945-A3040-D2 C7945-A3040-D27 C7945-A3040-D22 C7945-A3040-D03 C7945-A3040-D3 C7945-A3405-B2 C7945-A3405-C43 C7945-A3405-C44 C7927-Z970-C C7945-A3405-C47 C7945-A3405-B73 C7945-A3405-C52 C7945-A3405-C53 C7945-A3405-C54 C7945-A3405-C55 C7945-A3405-C223 C7945-A3405-B03 C7945-A3405-B0 C7945-A3405-B02 26

27 Order No. Cable connector for pressure sensor for combustion air Cable connector for pressure sensor for hydrogen Cable connector for input signal Cable connector for FID voltage Flow monitor Solenoid valve for control air, complete Solenoid valve for H2, combustion air, complete Solenoid valve for zero gas/calibration gas, complete Solenoid valve for control air, H2, combustion air Solenoid valve for for zero gas, calibration gas Gasket, diameter 4 mm Gasket, diameter 6 mm Gasket, diameter 0 mm O-ring, diameter 0 mm Pressure controller, hydrogen, combustion air/sample gas C7945-A3405-B05 C7945-A3405-B06 C7945-A3405-B09 C7945-A3405-B0 C7945-A3407-B53 C7945-A3405-B20 C7945-A3040-B206 C7945-A3405-B37 C79402-Z87-A5 C79402-Z87-A7 C7945-A3040-C33 C7945-A3040-C354 C7945-A3040-C353 F79402-E87-A C7945-A3405-B20 Union nut for pipe 6 Clamping ring, front, pipe 6 Clamping ring, rear, pipe 6 available only as set of parts, order no. C7945-A3405-D60 Union nut for pipe 8 Clamping ring, front, pipe 8 Clamping ring, rear, pipe 8 available only as set of parts, order no. C7945-A3405-D6 Union nut for pipe R/4" Clamping ring, front, pipe R/4" Clamping ring, rear, pipe R/4" available only as set of parts, order no. C7945-A3405-D62 PTFE gasket for M7 x 0.75 connection, tube.5 Analyzer fuse, T 4.0 A/250 V Analyzer fuse, T 6.3 A/250 V C7945-A3040-D0 W79054-L0-T400 W79054-L0-T630 27

28 Order No. Motherboard Preamplifier board Control board ROM, motherboard D3, 5E-P ROM, motherboard D4, 5E-P Display board (front panel) Touch pad keyboard with labels, German Touch pad keyboard with labels, English Touch pad keyboard with labels, French Touch pad keyboard with labels, Spanish Touch pad keyboard with labels, Italian C7945-A3405-B507 C7945-A3405-B508 C7945-A3405-B509 S7960-G42-A90 S7960-G42-A902 C7945-A320-A502 C7945-A3405-D52 C7945-A3405-D53 C7945-A3405-D54 C7945-A3405-D55 C7945-A3405-D56 28

29 0.3 FIDAMAT 5E-G Order No. Flame ionization detector, complete FID cover complete with thermocouple line Quartz nozzle for FIDAMAT 5E-G Clamping screw in the FID Nut for M5 connection Outer ring for M5 connection Graphite gasket for M5 connection Clamping ring for M5 connection Nut for M7 x 0.75 connection, tube.5 to 3.0 Outer ring for M7 x 0.75 connection, tube.5 to 3.0 Graphite gasket for M7 x 0.75 connection, pipe.5 Clamping ring for M7 x 0.75 connection, pipe.5 Graphite gasket for M7 x 0.75 connection, pipe 3.0 Clamping ring for M7 x 0.75 connection, pipe 3.0 Coarse filter Sample gas restrictor (fused silica), complete, 5E-G Crimp restrictor, hydrogen, 5E-G Crimp restrictor 2, combustion air, 5E-G Crimp restrictor 3, auxiliary air, 5E-G Crimp restrictor 4, exhaust, 5E-G Crimp restrictor 5, sample gas, 5E-G Temperature sensor, complete, FID oven and physical section Flat heater element, FID oven, bottom Flat heater element, FID oven, top Cable connector for input signal Cable connector for FID voltage Cable connector for pressure sensor for combustion air C7945-A3405-B50 C7945-A3405-B50 C79402-Z282-C C792-A3003-C5 C7945-A3040-D26 C7945-A3040-D2 C7945-A3040-D02 C7945-A3040-D2 C7945-A3040-D27 C7945-A3040-D22 C7945-A3040-D03 C7945-A3040-D3 C7945-A3040-D05 C7945-A3040-D5 C7945-A3405-C69 C7945-A3405-B74 C7945-A3405-C52 C7945-A3405-C53 C7945-A3405-C222 C7945-A3405-C55 C7945-A3405-C223 C7945-A3405-B03 C7945-A3405-B0 C7945-A3405-B02 C7945-A3405-B09 C7945-A3405-B0 C7945-A3405-B05 29

30 Order No. Cable connector for pressure sensor for hydrogen Flow monitor Solenoid valve for H2, combustion air, complete Solenoid valve for zero gas, calibration gas, sample gas, compl. Solenoid valve for control air, H2, Solenoid valve for for zero gas, calibration gas, sample gas Gasket, diameter 4 mm Gasket, diameter 6 mm Gasket, diameter 0 mm O-ring, diameter 0 mm Pressure controller, hydrogen, combustion air, sample gas PTFE gasket for M7 x 0.75 connection, tube.5 Elbow compression gland, 5E-G C7945-A3405-B06 C7945-A3407-B53 C7945-A3040-B206 C7945-A3405-B38 C79402-Z87-A5 C79402-Z87-A7 C7945-A3040-C33 C7945-A3040-C354 C7945-A3040-C353 F79402-E87-A C7945-A3405-B20 C7945-A3040-D0 C79304-Z209-C2 Union nut for pipe 6 Clamping ring, front, pipe 6 Clamping ring, rear, pipe 6 available only as set of parts, order no. C7945-A3405-D60 Union nut for pipe R/4" Clamping ring, front, pipe R/4" Clamping ring, rear, pipe R/4" available only as set of parts, order no. C7945-A3405-D62 Analyzer fuse, T 4.0 A/230 V Analyzer fuse, T 6.3 A/230 V Motherboard Preamplifier board Control board ROM, motherboard D3, 5E-G ROM, motherboard D4, 5E-G Display board (front panel) W79054-L0-T400 W79054-L0-T630 C7945-A3405-B507 C7945-A3405-B508 C7945-A3405-B509 S7960-G68-A90 S7960-G68-A902 C7945-A320-A502 30

31 Order No. Touch pad keyboard with labels, German Touch pad keyboard with labels, English Touch pad keyboard with labels, French Touch pad keyboard with labels, Spanish Touch pad keyboard with labels, Italian C7945-A3405-D52 C7945-A3405-D53 C7945-A3405-D54 C7945-A3405-D55 C7945-A3405-D56 3

32 32

33 Appendix. Gas Flow Diagram Pressure regulastor Solenoid valve Combustion gas (H2/He), inlet pressure 30 psig Measuring chamber Pressure sensor 2 outlet (atmospheric pressure) Restrtictors Pressure sensor Solenoid valve Pressure regulator Solenoid valve combustion air, inlet pressure 30 psig Restrictor Fused silica Flow monitor external solenoid valve, controllable via pins 3 and 4 of plug X2 Coarse filter sample gas, inlet pressure 5 psig zero gas, inlet pressure 8 psig calibration gas, inlet pressure 8 psig optional Fig.. FIDAMAT 5E-P for process measurements A-

34 Measuring chamber Pressure sensor Pressure regulastor Solenoid valve Combustion gas inlet pressure 30 psig 2 outlet (atmospheric pressure) Pressure sensor Pressure regulator Solenoid valve combustion air, inlet pressure 30 psig Restrictor Fused silica Restrtictors 3 Solenoid valve sample gas, inlet pressure 0 psig 9 C Fig..2 FIDAMAT 5E-AP for car exhaust measurements A-2

35 Measuring chamber Pressure sensor Pressure regulastor Solenoid valve 2 Combustion gas (H2/He), inlet pressure 30 psig outlet (atmospheric pressure) Pressure sensor Pressure regulator Solenoid valve combustion air, inlet pressure 30 psig Restrictor Fused silica Restrtictors Flow monitor Coarse filter sample gas, inlet pressure 5 psig zero gas, inlet pressure 5 psig calibration gas, inlet pressure 5 psig Fig..3 FIDAMAT 5E-G for measurements of gases of highest-grade purity A-3

36 .2 Layout Diagrams of Important Electronic Components Slave Master Motherboard Cooling rail EEPROM protection switch Terminal strips Control board Heat sink Potentiometer for adjustment of flow sensor Fig..4 Motherboard (top) and control board A-4

37 .3 Summary of Operating Functions Symbols in the sequence diagrams Display of set value and possible adjustmnet for a new value. When entering values make sure to enter the sign first! ENTER key (importing of numbers) Display output in large display (the measured value is displayed again after approx. 3 s) Input * ** *** Function protected by code (function level ) Function protected by codes and 2 (function level 2) Function protected by codes, 2 and 3 (function level 3) Symbols in the display Enter identifier CAL ILL The MEAS/CAL key must be pressed Illegal code, illegal identifier, non-existent function number entered cod Enter code according to function level Error message Warning Remarks The input can be interrupted at any time using the key; the old values are then retained and the system returns to the start of the function or partial function. A-5

38 Survey of operating functions A Display and switch over measuring ranges, switch over display * Switch over measuring range 2 Display measuring range ** 20 Autoranging on/off * 96 Switch over between mg/m 3 and ppm B Calibration of measuring ranges * 5 Calibrate zero 6 Display sensitivity setpoint * 7 Adjust sensiitivity setpoint * 8 Calibrate sensitivity to setpoint D Temperatures, pressures, flow, limits ** 34 Display, set max. temperature deviation * 35 Display, set H2 or H2/He pressure * 36 Display, set sample gas pressure * 37 Enter tolerance for hydrogen or hydrogen/helium pressure * 38 Enter tolerance for sample gas pressure A-6

39 E Time constants ** 3 Set time constant T90 ** 4 Set dynamic time constant TD (dynamic noise rejection) F Measured-value memory, current output ** 2 Measured-value memory on/off ** 24 Set current output G Redefine measuring ranges ** Define start-of-scale values of measuring ranges ** 2 Define full-scale values of measuring ranges ** 5 Select number of valid measuring ranges I Further display functions 5 Display compensated background current (zero calibration) 60 Display internal values 6 Display measured variables K Adjustment and calibration functions *** 7 Compensation of temperature influence on zero *** 73 Compensation of temperature influence on measured value A-7

40 L Test functions, RESET, switch-off * 29 Measure during warm-up phase ** 49 Switch valves individually *** 53 Load basic factory-set data *** 55 Load user-specific data *** 57 Store user-specific data *** 62 Reproduce internal values on analog signal, channel *** 67 Ignite flame *** 68 RESET 69 LED check, software no. *** 77 Suppression of brief noise signals *** 78 Suppression of negative measured values * 92 Display warnings and brief errors * 94 Switch over between mg/m 3 and ppm *** 99 Switch analyzer on/off M Reprogram, activate, deactivate codes *** 50 Reprogram, activate/deactivate codes A-8

41 * Switch over measuring range 2 Display measuring range * 5 Calibrate zero A A B 6 Display sensitivity setpoint B * 7 Adjust sensitivity setpoint B * 8 Calibrate sensitivity to setpoint B ** Define start-of-scale values of measuring ranges G ** 2 Define full-scale values of measuring ranges G ** 3 Set time constant T90 E ** 4 Set dynamic time constant TD (dynamic noise rejection) E 5 Display compensated background current (zero calibration) I ** 20 Autoranging on/off A ** 2 Measured-value memory on/off F ** 24 Set current output F * 29 Measure during warm-up phase L ** 34 Display, set max. temperature deviation * 35 Display, set H2 or H2/He pressure * 36 Display, set sample gas pressure * 37 Enter tolerance for hydrogen or hydrogen/helium pressure * 38 Enter tolerance for sample gas pressure ** 49 Switch valves individually *** 50 Reprogram, activate/deactivate codes ** 5 Select number of valid measuring ranges D D D D D L M G A-9

42 *** 53 Load basic factory-set data *** 55 Load user-specific data *** 57 Store user-specific data 60 Display internal values 6 Display measured variables *** 62 Reproduce internal values on analog signal, channel *** 67 Ignite flame *** 68 RESET 69 LED check, software no. *** 7 Compensation of temperature influence on zero *** 73 Compensation of temperature influence on measured value *** 77 Suppression of brief noise signals *** 78 Suppression of negative measured values * 92 Display warnings and brief errors L L L I I L L L L K K L L L * 94 Switch over between mg/m 3 and ppm A 96 Display zero adjustment *** 99 Switch analyzer on/off L L A-0

43 A Display and switch over measuring ranges A * Switch over measuring range Measuring range No. There are four measuring ranges. The measuring range which is switched on is indicated to the right of the display. The number of active measuring ranges can be defined using function 5. The start-of-scale and full-scale values of the measuring range can be freely selected. Factory setting: 2 Display measuring range The start-of-scale and full-scale values of the measuring range are output in succession in the display. 2 Display Measuring range No. ** 20 Autoranging on/off 2 0 Autoranging 0 Manual switchover This function only applies to the recorder output (see plug connector X). The current measured value is always output on the display even if it exceeds the current measuring range. The displayed value flashes when the measuring range(s) is exceeded. Factory setting: 0 * 94 Switch over between mg/m 3 and ppm The LED on the right of the display indicates the concentration unit of the display mg C/m 3 ppm C 2 ppm C3 3 ppm C6 Factory setting: 0 A-

44 B Calibration of measuring ranges B * 5 Calibrate zero (FIDAMAT 5E-AP only) 5 Prior to the activation of this function zero gas must be applied via the sample gas input. 20 s. after activation of function 5 the calibration will be carried out automatically. * 5 Calibrate zero (FIDAMAT 5E-P only) 5 Display Zero gas valve opens after pressing 5 and EN- TER. If there is no such valve, zero gas must be feeded via the sample gas line. After having pressed the ENTER key once more the calibration will be carried out and the zero gas valve (if present) will be closed. 6 Display sensitivity setpoint The setpoint is displayed which was used for the last calibration. The measuring range in which the setpoint is present is output to the right of the display. 6 Display Measuring range No. * 7 Adjust sensitivity setpoint The setpoint must be reset when the calibration gas is changed. All four ranges are calibrated using the same calibration gas. 7 n Identifier Setpoint Identifier: = calibration gas for range (to 4) * 8 Calibrate sensitivity to setpoint The sequence is analogous to function 5. A calibration gas is used to calibrate measuring range. The other measuring ranges are also calibrated. A-2

45 D Temperatures, pressures, flow, limits D ** 34 Display, set max. temperature deviation This function defines how far the oven temperatures may deviate upwards or downwards from the setpoint temperature. A warning is output if the temperature deviates by more than this value for one of the four heaters. 3 4 temperature deviation in C 2 to 0 Factory setting: 5 * 35 Display, set H2/He or H2 pressure 3 5 Display The hydrogen/helium pressure is output on the display in mbar and can be adjusted using the top screw on the front panel. The technical data refer to an overpressure of 000 mbar. * 36 Display, set sample gas pressure 3 6 Display The sample gas pressure is output on the display in mbar and can be adjusted using the bottom screw on the front panel. The technical data refer to an overpressure of 340 mbar (FID 5E-AP) or 500 mbar (FID 5E-P and 5E-G). The pressures will be less if there is no sample gas or combustion air. * 37 Enter tolerance for hydrogen pressure 3 7 Pressure deviation in mbar (0-255) The entered pressure deviation, usually 50 to 00 mbar, specifies the maximum permissible deviation from the setpoint which has been set using function 35. A warning is output if the hydrogen pressure exceeds or falls below this value. Factory setting: 50 * 38 Enter tolerance for sample gas pressure 3 8 Pressure deviation in mbar (0-255) The entered pressure deviation, usually 30 to 00 mbar, specifies the maximum permissible deviation from the setpoint which has been set using function 36. A warning is output if the sample gas pressure exceeds or falls below this value. Factory setting: 30 A-3

46 E Time constants E ** 3 Set time constant T90 The time constant T90 (0. to 00 s) is used to reduce the noise. The signal delay increases with increasing T90. Factory setting: T to 00 seconds ** 4 Set dynamic time constant TD (dynamic noise rejection) 4 Identifier (% of f.s.v.)0.0 to Identifier TD (0.0 to 00 s) The time constant TD applies to small changes in signal without a defined range (in % of respective f.s.v.). The time T90 defined in function 3 applies to larger changes in signal. TD must be larger than T90. In this manner, noise is optimally suppressed in the case of small changes; in the case of larger changes (> ), the modified signal is recorded faster by automatic switching over to the time constant T90. Factory setting: 3.0/.0 F Measured-value memory, current output F ** 2 Measured-value memory on/off 2 On 0 Off If the measured-value memory is switched on, the last measured value is stored when the CAL key is pressed. It is then present at the ma output until the analyzer is reset to the measurement status by pressing the MEAS key. The measured value is also stored in the display in the event of an error. Factory setting: 0 ** 24 Set current output 2 4 Status Status 0 0 to 20 ma 2 to 20 ma 2 4 to 20 ma Factory setting: 2 A-4

47 G Redefine measuring ranges G ** Define start-of-scale values of measuring ranges Range Start-of-scale value The start-of-scale values can be set as desired. The only condition is that the full-scale value set using function 2 is greater than the start-of-scale value set here. The unit displayed when the setting is made applies as the unit to the value entered here. Factory setting: 0 ** 2 Define full-scale values of measuring ranges 2 Range Full-scale value The full-scale values can be set as desired. The only condition is that the start-of-scale value set using function is greater than the full-scale value set here. The unit displayed when the setting is made applies as the unit to the value entered here. Factory setting: 0/50/200/000 ** 5 Select number of valid measuring ranges The number of valid measuring ranges is selected. Autoranging (function 20) only takes place within the selected ranges. 5 n Identifier Identifier: = measuring range 2 = measuring ranges and 2 3 = measuring ranges, 2 and 3 4 = measuring ranges, 2, 3 and 4 Factory setting: 4 A-5

48 I Further display functions I 5 Display compensated background current (zero calibration) The current compensated in the zero is displayed here in pa. 5 Display * 60 Display internal values Identifier: = Oven temperature (00 to 200 C) 2 = Physical section temperature (0 to 60 C) 6 0 Identifier Display 7 = Flame temperature (230 to 300 C) 9 = Temperature of board (0 to 70 C) = Pressure of sample gas/combustion air (mbar abs.) 2 = Hydrogen/helium pressure (~000 mbar absolute) 4 = FID voltage (200 V) 5 = +5 V 6 = 5 V 7 = +24 V 6 Display measured variables 6 n Identifier n not equal to 0 n Identifier Identifier: n = 0 = Measured value = Raw measured value (pa) 2 = Analog current 3 = Current following adjustment of preamplifier 4 = Current following zero adjustment Terminating this function via key sets the identifier automatically to 0. A-6

49 K Adjustment and calibration functions K *** 7 Compensation of temperature influence on zero 7 n Identifier value Identifier Service staff only! Setpoint Temperature of analyzer section with average ambient temperature (in C; see function 60, identifier 2) 2 Change in display with increase in temperature (function 6, identifier 4) 3 Change in display with reduction in temperature (function 6, identifier 4) Procedure. Calibrate zero of basic temperature (function 5) 2. Switch on measuring range 3. Select function 7 and enter a value of zero for identifiers 2 and 3 4. Measure (see example) 5. Enter identifier and setpoints under function 7 Identifier Value Caution! Function 7 might influence zero! Example: Value with Temp. < Temp. 0 Value 0 with Temp. 0 Basic temperature (Temperature 0) (Value 2 - Value 0) 2 * 000 (Temp. 2 Temp. 0) (Value 0 - Value ) 3 * 000 (Temp. 0 Temp. ) Value 2 with Temp. 2 > Temp. 0 Zero at average ambient temperature (e. g. 25 C) Value 0,, 2 is the measured value (as indicated under function 6, identifier 4) A-7

Operating and installation instructions Differential pressure monitor

Operating and installation instructions Differential pressure monitor Operating and installation instructions Differential pressure monitor RM-DPC Micro SensorLine Table of contents 1 Safety instructions...3 2 Equipment specification...3 3 Assembly...4 4 Step-by-step installation...4

More information

USER MANUAL VIBRATION CONTROL RMA-POWER-BOX 107/230

USER MANUAL VIBRATION CONTROL RMA-POWER-BOX 107/230 USER MANUAL VIBRATION CONTROL Version 2.2 1 IMPORTANT NOTES Electrical danger within the meaning of this documentation or the warning labels on the product itself respectively means that death, serious

More information

Operating instructions Edition 07/2006. FIDAMAT 6-G Gas analyzer for measuring total hydrocarbon concentrations. gas g analysis y

Operating instructions Edition 07/2006. FIDAMAT 6-G Gas analyzer for measuring total hydrocarbon concentrations. gas g analysis y Operating instructions Edition 07/2006 Gas analyzer for measuring total hydrocarbon concentrations gas g analysis y General Information 1 Description 2 Installation and connection 3 Gas analyzer for determining

More information

2-Axis Counter BDD 622

2-Axis Counter BDD 622 2-Axis Counter BDD 622 Technical Description, User's Guide english No. 622-221 D/E. Edition 0701; Subject to modification. www.balluff.com Balluff GmbH Schurwaldstrasse 9 73765 Neuhausen a.d.f. Germany

More information

OPERATING INSTRUCTIONS PA AMPLIFIER P-1812

OPERATING INSTRUCTIONS PA AMPLIFIER P-1812 OPERATING INSTRUCTIONS PA AMPLIFIER P-1812 Please follow the instructions in this manual to obtain the optimum results from this unit. We also recommend that you keep this manual handy for future reference.

More information

Continuous gas analysis

Continuous gas analysis Oxygen analyzer for standard applications 7MB2001 Operating instructions Continuous gas analysis General information 1 Description 2 Continuous gas analysis Oxygen analyzer for standard applications Operating

More information

Electronic Temperature Controller. Instruction Manual Version

Electronic Temperature Controller. Instruction Manual Version Electronic Temperature Controller 701 Instruction Manual Version 1.00.01 Dear Customer, we have made up this operating manual in such a way that all necessary information about the product can be found

More information

Operating Instructions Differential Pressure Transducer PS27

Operating Instructions Differential Pressure Transducer PS27 Operating Instructions Differential Pressure Transducer PS27 halstrup-walcher GmbH Stegener Straße 10 79199 Kirchzarten/Germany Phone: +49 (0) 76 61/39 63 0 Fax: +49 (0) 76 61/39 63 99 E-mail: info@halstrup-walcher.com

More information

CDD4 Duct Carbon Dioxide Transmitter

CDD4 Duct Carbon Dioxide Transmitter Drill or punch a 1-1/8 or 1-1/4 hole in the duct at the preferred location and insert the probe into the hole to mark the enclosure mounting holes. Remove the unit and drill the four mounting holes. Clean

More information

SINAMICS G130. Terminal Module 150 (TM150) Operating Instructions 03/2013 SINAMICS

SINAMICS G130. Terminal Module 150 (TM150) Operating Instructions 03/2013 SINAMICS SINAMICS G130 Operating Instructions 03/2013 SINAMICS s Safety information 1 General information 2 SINAMICS SINAMICS G130 Mechanical installation 3 Electrical installation 4 Technical specifications 5

More information

Indicating Digital Transmitter for Differential Pressure and Flow. Media 6 with LCD Media 6 with LED

Indicating Digital Transmitter for Differential Pressure and Flow. Media 6 with LCD Media 6 with LED Indicating Digital Transmitter for Differential Pressure and Flow Media 6 with LCD Media 6 with LED Application Microprocessor-controlled transmitter for measuring and indicating the differential pressure

More information

SIMADYN D Digital Control System. Fiber-Optic Rack Coupling CS12. User Manual. Edition DK No

SIMADYN D Digital Control System. Fiber-Optic Rack Coupling CS12. User Manual. Edition DK No SIMADYN D Digital Control System User Manual Fiber-Optic Rack Coupling CS12 Edition 05.95 DK No. 237741 User Manual, Fiber-Optic Rack Coupling CS12 Edition Status 1 Fiber-Optic Rack Coupling CS12 05.95

More information

PROCESS AUTOMATION MANUAL. Universal Temperature Converter KCD2-UT2-(Ex)1 ISO9001

PROCESS AUTOMATION MANUAL. Universal Temperature Converter KCD2-UT2-(Ex)1 ISO9001 PROCESS AUTOMATION MANUAL Universal Temperature Converter KCD2-UT2-(Ex)1 ISO9001 With regard to the supply of products, the current issue of the following document is applicable: The general terms of delivery

More information

Intrinsically safe batch controller Batching Master 110i

Intrinsically safe batch controller Batching Master 110i Intrinsically safe batch controller Batching Master 110i Installation Guide BVS 04 AT E 172 Revision 12.2 IBS BatchControl GmbH Im Sträßchen 2-4 Tel.: ++49 2441 9199 801 53925 Kall Fax.: ++49 2441 9199

More information

Installation and operation manual ReciFlow Gas

Installation and operation manual ReciFlow Gas Installation and operation manual ReciFlow Gas 1 1. Measurement principle... 3 2. Installation... 5 3. Operation... 7 4. Electrical interfaces... 11 5. Communication protocol... 14 6. Software update and

More information

Model VDL 115 Vacuum drying chambers for flammable solvents

Model VDL 115 Vacuum drying chambers for flammable solvents Model VDL 115 Vacuum drying chambers for flammable solvents A BINDER safety vacuum drying oven of the VDL series ensures maximum safety when drying organic solvents standard with TÜV/GS. The inner chamber

More information

PROCESS ANALYSERS. SERVOTOUGH FluegasExact Gas Analyser. QuickStart Manual. Part Number: D Revision: 1 Language: UK English

PROCESS ANALYSERS. SERVOTOUGH FluegasExact Gas Analyser. QuickStart Manual. Part Number: D Revision: 1 Language: UK English PROCESS ANALYSERS SERVOTOUGH FluegasExact Gas Analyser QuickStart Manual Part Number: 02700003D Revision: 1 Language: UK English This page intentionally blank TABLE OF CONTENTS 1. INTRODUCTION............................................

More information

Serial PROFIBUS Interface

Serial PROFIBUS Interface Installation Manual Serial PROFIBUS Interface Version: EN-062016-2.3 Copyright 2016 Softing Industrial Automation GmbH Disclaimer of liability The information contained in these instructions corresponds

More information

Manual Control Unit GFC 32

Manual Control Unit GFC 32 Manual Control Unit 1400004_EN/05.2017 Index 1. Main features 3 2. Technical features 3 3. Installation guidelines 4 4. Preliminary checks 5 5. Electrical connections 5 6. Settings 6 7. Remote control

More information

Pressure transmitter for highest pressure applications Up to 217,500 psi (15,000 bar) Model HP-2

Pressure transmitter for highest pressure applications Up to 217,500 psi (15,000 bar) Model HP-2 Electronic pressure measurement Pressure transmitter for highest pressure applications Up to 217,500 psi (15,000 bar) Model HP-2 WIKA data sheet PE 81.53 Applications Test bench construction Water jet

More information

Temperature and humidity regulator H3020 H3021 H3023 H3027. Instruction Manual

Temperature and humidity regulator H3020 H3021 H3023 H3027. Instruction Manual Temperature and humidity regulator H3020 H3021 H3023 H3027 Instruction Manual Table of content 1. GENERAL SAFETY RULES... 3 2. GENERAL DESCRIPTION... 4 3. REGULATOR MOUNTING AND CONNECTION... 4 4. DESCRIPTION

More information

H Series PLC. ! : Indicates Compulsion. EH-150 Analog input module EH-AXH8M Instruction manual. Safety precautions DANGER CAUTION COMPULSION

H Series PLC. ! : Indicates Compulsion. EH-150 Analog input module EH-AXH8M Instruction manual. Safety precautions DANGER CAUTION COMPULSION H Series PLC EH-150 Analog input module EH-AXH8M Instruction manual Thank you for purchasing a Hitachi Programmable Logic Controller. To operate it safely, please read this instruction manual and all the

More information

DMT PERFORMING UNDER PRESSURE DMT. Applications. Special features. Description 1/9

DMT PERFORMING UNDER PRESSURE DMT. Applications. Special features. Description 1/9 Applications Industry (laboratory, workshop and production) Transmitter and pressure gauge manufacturers Calibration service companies and service industry Research and development laboratories National

More information

PROCESS AUTOMATION MANUAL. Temperature Trip Amplifier. KFD2-GU-(Ex)1 ISO9001

PROCESS AUTOMATION MANUAL. Temperature Trip Amplifier. KFD2-GU-(Ex)1 ISO9001 PROCESS AUTOMATION MANUAL Temperature Trip Amplifier KFD2-GU-(Ex)1 ISO9001 With regard to the supply of products, the current issue of the following document is applicable: The General Terms of Delivery

More information

NDIR Infrared Gas Analyzer

NDIR Infrared Gas Analyzer QUICKSTART GUIDE MODEL 7600 Teledyne Analytical Instruments Quickstart Guide Teledyne Analytical Instruments Model 7600 Quickstart Guide GETTING STARTED This Quickstart Guide is designed to get you set

More information

MAINTENANCE MANUAL. EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS

MAINTENANCE MANUAL. EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS MAINTENANCE MANUAL EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS SPECIFICATIONS*... 2 INTRODUCTION... 3 DESCRIPTION...

More information

Model VD 23 Vacuum drying chambers for non-flammable solvents

Model VD 23 Vacuum drying chambers for non-flammable solvents Model VD 23 Vacuum drying chambers for non-flammable solvents A BINDER vacuum drying oven VD series gently dries materials with homogeneous temperature distribution. The patented expansion shelf technology

More information

HITACHI. EH-150 series PLC EH-RTD8 Resistance Temperature Detective input module Instruction manual. Safety precautions

HITACHI. EH-150 series PLC EH-RTD8 Resistance Temperature Detective input module Instruction manual. Safety precautions HITACHI EH-150 series PLC Resistance Temperature Detective input module Instruction manual Thank you for purchasing a Hitachi Programmable Logic Controller. To operate it safely, please read this instruction

More information

REDUNDANCY MODULE TSP-REM360 AND TSP-REM600

REDUNDANCY MODULE TSP-REM360 AND TSP-REM600 REDUNDANCY MODULE TSP-REM360 AND TSP-REM600 Operating Instructions Seite 1 Dimensions drawings: TSP-REM360 Weight: 0.882lb Gewicht: 0.40kg Seite 2 Dimensions drawings: TSP-REM600 Bottom view Top view Side

More information

SD1. Differential Controller. Operating and Installation Instructions F2 72.3

SD1. Differential Controller. Operating and Installation Instructions F2 72.3 SD1 Differential Controller Operating and Installation Instructions F2 72.3 i Please follow the safety information and read these Instructions carefully before putting the system into operation. Safety

More information

Electronic SD1, AS 8, ASR 14, ASR 20

Electronic SD1, AS 8, ASR 14, ASR 20 Electronic SD1, AS 8, ASR 14, ASR 20 2 Electronic SD1, AS 8, ASR 14, ASR 20 Contents Page List of contents..................................................... 3 Function and characteristics.........................

More information

Plug-in module SIRAX TV 808, 2 channels Isolating Amplifier unipolar/bipolar

Plug-in module SIRAX TV 808, 2 channels Isolating Amplifier unipolar/bipolar Plug-in module SIRAX TV 808, channels For electrically insulating, amplifying and converting DC signals Application The purpose of the isolating amplifi er SIRAX TV 808 (Fig. ) is to electrically insulate

More information

Description. Monitoring functions Evaluation of brush length monitoring, digital, non-floating A00

Description. Monitoring functions Evaluation of brush length monitoring, digital, non-floating A00 Monitoring functions Evaluation of brush length monitoring, digital, non-floating Evaluation of brush length monitoring, digital, floating A00 A06 Non-floating brush length monitoring with use of KM01

More information

Manual Simatek Control Unit GFC 32

Manual Simatek Control Unit GFC 32 Manual Simatek Control Unit GFC 32 Original instructions 1400004_GB Ver. 2014.01.09 Index 1. Main features 3 2. Technical features 3 3. Installation guidelines 4 4. Preliminary checks 5 5. Electrical connections

More information

Electronic pressure switch with display Model PSD-4

Electronic pressure switch with display Model PSD-4 Electronic pressure measurement Electronic pressure switch with display Model PSD-4 WIKA data sheet PE 81.86 Applications Machine tools Hydraulics and pneumatics Pumps and compressors Special-purpose machine

More information

Industriefunkuhren. Technical Manual. Signal Converter. for DIN Rail Mounting Series 4800xx-yy ENGLISH

Industriefunkuhren. Technical Manual. Signal Converter. for DIN Rail Mounting Series 4800xx-yy ENGLISH Industriefunkuhren Technical Manual Signal Converter for DIN Rail Mounting Series 4800xx-yy ENGLISH Version: 01.01-19.07.2007 2 / 23 Signal Converter 4800 - V01.01 INPORTANT NOTES Downloading Technical

More information

NIR Moisture Analyzer KB-30

NIR Moisture Analyzer KB-30 NIR Moisture Analyzer KB-30 Operation Manual Safety Precautions FPL950701 Improper use of the NIR moisture analyzer in violation of the following safety notes may result in death, injury or damage to property

More information

IV-30 Operating Manual for Pulse Distributor Cassette with potential separation

IV-30 Operating Manual for Pulse Distributor Cassette with potential separation IV-30 Operating Manual for Pulse Distributor Cassette with potential separation Edition-/Rev.-Date: 09/08/2006 Document-/Rev.-No.: TR - EAK - BA - GB - 0093-02 Software version: - File name: TR-EAK-BA-GB-0093-02.DOC

More information

TSP-BCM48, TSP-BCM48A. Operating Instruction Manual. Page 1 of 13

TSP-BCM48, TSP-BCM48A. Operating Instruction Manual.  Page 1 of 13 Battery Controller Module TSP-BCM24, TSP-BCM24A TSP-BCM48, TSP-BCM48A Operating Instruction Manual http://www.tracopower.com Page 1 of 13 Dimensions drawings: TSP-BCM24 & TSP-BCM48 Weight: 0.816lb Gewicht:

More information

Data Sheet T 6493 EN. TROVIS 6400 Automation System TROVIS 6493 Compact Controller. For panel mounting (front frame 48 x 96 mm/1.89 x 3.

Data Sheet T 6493 EN. TROVIS 6400 Automation System TROVIS 6493 Compact Controller. For panel mounting (front frame 48 x 96 mm/1.89 x 3. Data Sheet T 6493 EN TROVIS 6400 Automation System TROVIS 6493 Compact Controller For panel mounting (front frame 48 x 96 mm/1.89 x 3.78 inch) Application Digital controller to automate industrial and

More information

DINALOG A 144 x 36 A1400 Light-Strip Indicator

DINALOG A 144 x 36 A1400 Light-Strip Indicator DINALOG A x 6 A00 Light-Strip Indicator -9-0-0 /.00 Front panel dimensions: x 6 mm Light-strip indicator with 7 high-contrast LEDs Red LED display color Digital display range for portrait format: 999 to

More information

English. Operating manual. Universal transmitter UT125. Save for later reference. Companies / brands of GHM

English. Operating manual. Universal transmitter UT125.  Save for later reference. Companies / brands of GHM English Operating manual Universal transmitter UT125 Companies / brands of GHM www.ghm-messtechnik.de Save for later reference. Table of contents page 1. Intended use (areas of application)... 3 1.1 Safety

More information

Tempco Instruction Manual

Tempco Instruction Manual Tempco Instruction Manual 1/16 DIN Solid State Temperature Controller Relay Output Solid State Output For Heating Model Numbers: TEC-901, TEC-902, TEC-905 Temperature controls in this series are designed

More information

CDD Carbon Dioxide Transmitter

CDD Carbon Dioxide Transmitter Introduction The OSA CO2 transmitter uses Infrared Technology to monitor CO2 levels within a range of 0 2000 ppm and outputs a linear 4-20 ma or 0-5/0-10 Vdc signal. The enclosure is designed to operate

More information

SITOP power supply. Selectivity modules. Overview. Safety notes. Description, device design, dimension drawing. Mounting/removal 3

SITOP power supply. Selectivity modules. Overview. Safety notes. Description, device design, dimension drawing. Mounting/removal 3 Overview Safety notes 1 SITOP power supply Manual Description, device design, dimension drawing 2 Mounting/removal 3 Mounting position, mounting clearances 4 Installation 5 Technical data 6 Safety, approvals,

More information

Operating instructions. Speed monitor D / / 2014

Operating instructions. Speed monitor D / / 2014 Operating instructions Speed monitor D200 80005257 / 00 05 / 2014 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5 2.2 Target group...5

More information

Monitoring technique. VARIMETER Voltage relay MK 9064N, MH 9064

Monitoring technique. VARIMETER Voltage relay MK 9064N, MH 9064 Monitoring technique VARIMETER Voltage relay MK 9064N, MH 9064 0269462 Your Advantages Preventive maintenance For better productivity Quicker fault locating Precise and reliable Min-, Max. value or window

More information

MINI MCR-SL-PT100-UI-200

MINI MCR-SL-PT100-UI-200 Configurable temperature transducer for PT00 INTERFACE Data sheet 078_en_0 PHOENIX CONTACT 00-0-5 Description The MINI MCR-SL-PT00-UI-00... is a configurable -way isolated temperature transducer. It is

More information

Flow VA 500/

Flow VA 500/ VA 500/ 520 Verbrauch 79 What are the advantages of the flow measuring technology of CS Instruments? 1) 500 is mounted by means of a standard 1/2 ball valve. During mounting and dismounting the safety

More information

DS 96/48 PK. Digital indicator. Manual 42/33-90 EN Rev. 00 DIM

DS 96/48 PK. Digital indicator. Manual 42/33-90 EN Rev. 00 DIM DS 96/48 PK Digital indicator Manual 42/33-90 EN Rev. 00 A1 A2 DIM E DS 96/48 PK Digital Indicator Manual Document No. 42/33-90 EN Edition 06.01 Revision 00 Manufacturer: ABB Automation Products GmbH Höseler

More information

Switched Mode Power Supply Operating manual PAP800

Switched Mode Power Supply Operating manual PAP800 Features: Switched mode power supply Wide output 0 144Vdc Analog control by an external 0 5Vdc Power failure alarm output Master-slave connection Powerfinn PAP series is a high power, lightweight, advanced

More information

MINI MCR-SL-PT100-UI(-SP)(-NC)

MINI MCR-SL-PT100-UI(-SP)(-NC) Configurable Temperature Transducer for PT100 Data Sheet 04/2005 Functions MINI MCR-SL-PT100-UI(-SP)(-NC) is a configurable 3-way isolated temperature measuring transducer. It is suitable for the connection

More information

Carbon Monoxide Sensor - ModBus

Carbon Monoxide Sensor - ModBus Introduction The CO Sensor uses an electrochemical sensor to monitor CO level in a range of 0 to 500 ppm and communicates via an RS-485 network configured for ModBus protocol. Before Installation Read

More information

Non-communicating room controllers

Non-communicating room controllers 3 882 DESIGO RXA Non-communicating room controllers RXA29.1 For fan-coil systems The RXA29.1 room controller are used for temperature control in individual rooms. For 2-pipe or 4-pipe fan-coil systems,

More information

DIGITAL PANEL THERMOMETER

DIGITAL PANEL THERMOMETER Love Controls DIGITAL PANEL THERMOMETER LCI108-1x LCI108J-1x INSTRUCTION MANUAL EDITION: January 2001 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This manual does not constitute a formal agreement.

More information

NMID. Remote control. SAPIM I/O number. SAPIM I/O number NBRN (RS485) ZM100. Relay, 24 VDC - + UO07 UI11 UI10 UI09 UI08 UI07 UI06 UI05 UI04 UI03 UI02

NMID. Remote control. SAPIM I/O number. SAPIM I/O number NBRN (RS485) ZM100. Relay, 24 VDC - + UO07 UI11 UI10 UI09 UI08 UI07 UI06 UI05 UI04 UI03 UI02 1/8 NRK16/A, NRK16-B/A, (NRK16-T /A, NRK14-T /A, NRK16-WEB/A) Control and interlock devices with pre-programmed system-specific application modules The controllers may be used as stand-alone control and

More information

1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882

1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 This manual is intended for use in support of installation, commissioning and configuration of the 1/32-DIN Temperature

More information

Pressure sensor For highest pressure applications to 15,000 bar Model HP-2

Pressure sensor For highest pressure applications to 15,000 bar Model HP-2 Pressure Pressure sensor For highest pressure applications to 15,000 bar Model HP-2 WIKA data sheet PE 81.53 for further approvals see page 6 Applications Test bench construction Water jet cutting High-pressure

More information

Operation Manual Series W-500

Operation Manual Series W-500 Operation Manual Series W-500 Table of Content 1 Safety information... 2 1.1 Place of application of the unit... 2 1.2 Instructions for installation... 3 2 Start-up and adjustment of controller 4 3 General

More information

OPERATOR S MANUAL 9000U ISOLATOR

OPERATOR S MANUAL 9000U ISOLATOR OPERATOR S MANUAL 9000U ISOLATOR Masibus Automation And Instrumentation Pvt. Ltd. B/30, GIDC Electronics Estate, Sector-25, Gandhinagar-382044, Gujarat, India Ph: 91 79 23287275-79 Fax: 91 79 23287281,

More information

Operating instructions AS-i SmartLine module AC3200 AC /00 06/2016

Operating instructions AS-i SmartLine module AC3200 AC /00 06/2016 Operating instructions AS-i SmartLine module AC3200 AC3201 80237876/00 06/2016 Contents 1 Preliminary note...3 1.1 Symbols used...3 1.2 Warnings used...3 2 Safety instructions...3 2.1 General...3 2.2 Target

More information

Mechanical, analog system comprising sensor and processor for accurately recording the position of gripper jaws.

Mechanical, analog system comprising sensor and processor for accurately recording the position of gripper jaws. APS-M1 Accessories Sensor System Analog Position Sensor Analog Position Sensor System Mechanical, analog system comprising sensor and processor for accurately recording the position of gripper jaws. Function

More information

Thermal Mass Flowmeter Sensyflow FMT200-D

Thermal Mass Flowmeter Sensyflow FMT200-D Data Sheet Thermal Mass Flowmeter Sensyflow FMT200-D Direct measurement of mass flow or standard volume flow of compressed air, nitrogen and biogas independent of media pressure and temperature Wide measuring

More information

SINAMICS G130. Voltage Sensing Module 10 (VSM10) Operating instructions 03/2011 SINAMICS

SINAMICS G130. Voltage Sensing Module 10 (VSM10) Operating instructions 03/2011 SINAMICS SINAMICS G130 Operating instructions 03/2011 SINAMICS s Safety information 1 General 2 SINAMICS SINAMICS G130 Voltage Sensing Module 10 (VSM10) Mechanical installation 3 Electrical installation 4 Technical

More information

JUMO PINOS L01. Calorimetric flow sensor. Brief description. Special features. Customer benefits. Data Sheet

JUMO PINOS L01. Calorimetric flow sensor. Brief description. Special features. Customer benefits. Data Sheet sales@jumo.co.uk info.us@jumo.net Page 1/9 JUMO PINOS L01 Calorimetric flow sensor Brief description The JUMO PINOS L01 flow sensor serves the measuring and monitoring of flow velocities of water and aqueous

More information

System 6000 Electropneumatic Converters (proportional valves) Electronic Controllers Signal Converters

System 6000 Electropneumatic Converters (proportional valves) Electronic Controllers Signal Converters System 6000 Electropneumatic Converters (proportional valves) Electronic Controllers Signal Converters Edition November 2003 Information Sheet T 6000 EN Proportional valves e/p Converters Type 6 66 626

More information

SINAMICS G130. Voltage Sensing Module 10 (VSM10) Operating Instructions 05/2010 SINAMICS

SINAMICS G130. Voltage Sensing Module 10 (VSM10) Operating Instructions 05/2010 SINAMICS SINAMICS G130 Operating Instructions 05/2010 SINAMICS s Safety information 1 General 2 SINAMICS SINAMICS G130 Voltage Sensing Module 10 (VSM10) Mechanical installation 3 Electrical installation 4 Technical

More information

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies ADC7520 SERIES 1600W Battery Chargers and Power Supplies Wide output adjustment range 0 72VDC Analog control by external 0-5VDC voltage Temp.comp charging, sense as on option Power fail relay alarm Master-Slave

More information

B5W-LD0101-1/2 B5W- L D Air Quality Sensor. Highly Sensitive and Compact Air Quality Sensor. Model Number Structure. Ordering Information

B5W-LD0101-1/2 B5W- L D Air Quality Sensor. Highly Sensitive and Compact Air Quality Sensor. Model Number Structure. Ordering Information Air Quality Sensor Highly Sensitive and Compact Air Quality Sensor Detection of particles down to 0.5 m in diameter is possible while using an LED light source Features a unique flow path structure that

More information

DIGITAL PANEL THERMOMETER

DIGITAL PANEL THERMOMETER Love Controls DIGITAL PANEL THERMOMETER LCI108-2x LCI108J-2x INSTRUCTIONS MANUAL EDITION: February 2001 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This manual does not constitute a formal agreement.

More information

Operating instructions Evaluation system for flow sensors VS / / 2013

Operating instructions Evaluation system for flow sensors VS / / 2013 Operating instructions Evaluation system for flow sensors VS3000 7097 / 0 07 / 203 Contents Preliminary note...2 2 Safety instructions...3 3 Function and features... Mounting.... Mounting of the sensors...

More information

WSK INTELLIGENT TEMPERATURE AND HUMIDITY CONTROLLER USER MANUAL

WSK INTELLIGENT TEMPERATURE AND HUMIDITY CONTROLLER USER MANUAL WSK INTELLIGENT TEMPERATURE AND HUMIDITY CONTROLLER USER MANUAL WSK INTELLIGENT TEMPERATURE AND HUMIDITY CONTROLLER USER MANUAL v 1.0 Thanks for choosing the Intelligent temperature and humidity controller

More information

Data sheet MK 20 GR. Customized solution or off-the-shelf: The choice is yours!

Data sheet MK 20 GR. Customized solution or off-the-shelf: The choice is yours! Data sheet MK 20 GR Customized solution or off-the-shelf: The choice is yours! MK20-GR data sheet Looking for a quick and reliable solution? Then rely on our standard products: Our universal controls provide

More information

Surface-mounting Thermostats AM Series with 1 or 2 single-pole snap-action switches IP40 protection

Surface-mounting Thermostats AM Series with 1 or 2 single-pole snap-action switches IP40 protection Page 1/6 Surface-mounting Thermostats AM Series with 1 or 2 single-pole snap-action switches IP40 protection Brief description Surface-mounting thermostats control and monitor thermal processes. Thermostats

More information

RXL24.1. Room controller RXL. Communicating controller for chilled ceiling and radiator applications CC-02

RXL24.1. Room controller RXL. Communicating controller for chilled ceiling and radiator applications CC-02 s 3 878 RXL Room controller RXL24.1 Communicating controller for chilled ceiling and radiator applications CC-02 The RXL24.1 room controller is used for temperature control in individual rooms. For chilled

More information

Manual Control Unit GFCD 16

Manual Control Unit GFCD 16 Manual Control Unit 1400002_EN/04.2017 Index 1. Main features 3 2. Technical features 3 3. Installation guidelines 4 4. Preliminary checks 5 5. Electrical connections 5 6. Filter taps 5 7. Settings 6 8.

More information

14 Pressure Measurement. Measuring Devices Description Pressure range Connection Device Page

14 Pressure Measurement. Measuring Devices Description Pressure range Connection Device Page Measuring Devices Pressure range Connection Device Page Digital display mounting, für niedrige Drücke 0 2. mbar / 2. bar 4 mm tube MPV, MPA.02 portable, hand-operated 0 1 mbar / 10 bar 4 mm tube MHA.03

More information

MINI MCR-SL-PT100-UI. Active resistance thermometer measuring transducer. Data sheet. 1 Description

MINI MCR-SL-PT100-UI. Active resistance thermometer measuring transducer. Data sheet. 1 Description Active resistance thermometer measuring transducer Data sheet 099_en_06 PHOENIX CONTACT 0-0- Description The 6. mm wide MINI MCR-SL-PT00-UI-... is a configurable, -way isolated temperature measuring transducer.

More information

Operating Instructions for Digital Thermometers. Model: DTM

Operating Instructions for Digital Thermometers. Model: DTM Operating Instructions for Digital Thermometers Model: 1. Contents 1. Contents... 2 2. Note... 3 3. Instrument Inspection... 3 4. Regulation Use... 3 5. Operating Principle... 4 6. Mechanical Connection...

More information

MODEL: M5DY SEN TRONIC AG CURRENT LOOP SUPPLY. [2] POWER INPUT AC Power M: V AC (Operational voltage range V, Hz)

MODEL: M5DY SEN TRONIC AG CURRENT LOOP SUPPLY. [2] POWER INPUT AC Power M: V AC (Operational voltage range V, Hz) Super-mini Terminal Block Signal Conditioners M5-UNIT CURRENT LOOP SUPPLY Functions & Features Powering a 0 ma DC current loop Applicable to smart transmitters Isolation between the input and output Fast

More information

Type 6150 AS-Interface Module. Type 3372 Electropneumatic Actuator with AS-interface module. Mounting and Operating Instructions EB 6150 EN

Type 6150 AS-Interface Module. Type 3372 Electropneumatic Actuator with AS-interface module. Mounting and Operating Instructions EB 6150 EN Type 6150 AS-Interface Module Type 3372 Electropneumatic Actuator with AS-interface module Mounting and Operating Instructions EB 6150 EN Edition July 2013 Definition of the signal words used in these

More information

RXB39.1. Room controller RXB. Communicating room controller for fan-coil applications FC-13

RXB39.1. Room controller RXB. Communicating room controller for fan-coil applications FC-13 s 3 875 RXB Room controller Communicating room controller for fan-coil applications FC-13 RXB39.1 The RXB39.1 room controller is used for temperature control in individual rooms. For 2-pipe or 4-pipe fan-coil

More information

NT PRESSURE TRANSDUCER MODELS 4100, 4210, 4130, 4230

NT PRESSURE TRANSDUCER MODELS 4100, 4210, 4130, 4230 P/N 0-027289 (Rev. A 04/4) NT PRESSURE TRANSDUCER MODELS 400, 420, 430, 4230 User Guide Table of Contents Safety Alert Symbol... 2 Introduction... 3 Installation... 3 Mechanical Installation... 3 Electrical

More information

Operating Manual LF1-LF4 i

Operating Manual LF1-LF4 i Operating Manual Oct. 2005 Fillerkampsweg 1 5 31832 Springe OT Eldagsen Tel.: 05044/887-0 (Fax: -99) E-Mail: info@iotronic.de Internet: http://www.iotronic.de Page 3 Contents 1.0 General Guidelines for

More information

Model HM-535 Power Supply Installation and Service Instructions

Model HM-535 Power Supply Installation and Service Instructions Model HM-535 Power Supply Installation and Service Instructions 430-535 0104 2004 Heritage MedCall, Inc SENTRY INSTALLATION & SERVICE INSTRUCTIONS POWER SUPPLY UNIT Model HM-535 IMPORTANT SAFETY INSTRUCTIONS

More information

EX-RC1 Remote I/O Adapter

EX-RC1 Remote I/O Adapter EX-RC1 Remote I/O Adapter The EX-RC1 interfaces between Unitronics Vision OPLCs and remote I/O Expansion Modules distributed throughout your system. The adapter is connected to a PLC via CANbus. Each adapter

More information

SIMOREG 6RA70 DC MASTER SIMOREG CM

SIMOREG 6RA70 DC MASTER SIMOREG CM SIMOREG RA70 DC MASTER /2 Application /2 Design /3 Technical Data /3 Standards /4 Block diagram / Options Siemens DA 21.1 200 /1 SIMOREG RA70 DC MASTER Application An important application for the converter

More information

Product data sheet accessories control and regulation Accessories AIRWIN

Product data sheet accessories control and regulation Accessories AIRWIN Product data sheet accessories control and regulation BO-CT1 Continuous controller Item no. 6100101 Control of humidification performance BO-CT1 Option board 2 binary inputs Item no. 002212 Setpoint switching,

More information

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 4 OPERATIONAL DIAGRAMS 5 WIRING AND

More information

VL BPC MINI. A configurable industrial computer platform. Data sheet 2930_en_F. 1 Description. 2 Features

VL BPC MINI. A configurable industrial computer platform. Data sheet 2930_en_F. 1 Description. 2 Features A configurable industrial computer platform Data sheet 90_en_F Description PHOENIX CONTACT 0-08- Features The VL BPC MINI is an embedded box PC and is part of the Valueline family of industrial computers.

More information

IO-AO6X I/O Expansion Module 6 Isolated Analog Outputs

IO-AO6X I/O Expansion Module 6 Isolated Analog Outputs IO-AO6X I/O Expansion Module 6 Isolated Analog Outputs The IO-AO6X is an I/O Expansion Module that can be used in conjunction with specific Unitronics OPLC controllers. The module offers 6 12-bit isolated

More information

Sartorius Research. R 160 P Electronic Semi-Microbalance

Sartorius Research. R 160 P Electronic Semi-Microbalance Sartorius Research. R 160 P Electronic Semi-Microbalance Installation and Operating Instructions 1 Power receptacle, fuse, 8 Level indicator voltage selector 9 ON/OFF key 2 Menu access switch 10 Print

More information

Communicative controller for VAV systems

Communicative controller for VAV systems FCR015 Communicative controller for VAV systems Summary FCR015 is a communicative controller for heating and cooling panels and a VAV (variable air volume) damper. It measures temperature and CO 2 concentration

More information

VL BPC MINI. A configurable industrial computer platform Intel Atom Z510PT CMAT IPC Module Option [I28] AUTOMATION Data Sheet 2930_en_A.

VL BPC MINI. A configurable industrial computer platform Intel Atom Z510PT CMAT IPC Module Option [I28] AUTOMATION Data Sheet 2930_en_A. A configurable industrial computer platform Intel Atom Z0PT CMAT IPC Module Option [I8] AUTOMATION Data Sheet 90_en_A Description PHOENIX CONTACT 0-0-0 Features The VL BPC MINI is an embedded box PC and

More information

EE360. High-End Moisture in Oil Transmitter. Typical applications. Features EE360 EE360. (+34)

EE360. High-End Moisture in Oil Transmitter. Typical applications. Features EE360 EE360. (+34) is dedicated for reliable monitoring of lubrication, hydraulic and insulation oils as well as diesel fuel. In addition to highly accurate measurement of water activity (a w ) and temperature (T), calculates

More information

UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS

UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS EN0I-6041 12/99 PRODUCT SPECIFICATION SHEET OVERVIEW The UDC 1000 and UDC 1500 are microprocessor-based 1/16 DIN and 1/8 DIN controllers

More information

TEMPERATURE AND HUMIDITY TRANSDUCER P18 TYPE USER S MANUAL

TEMPERATURE AND HUMIDITY TRANSDUCER P18 TYPE USER S MANUAL TEMPERATURE AND HUMIDITY TRANSDUCER P18 TYPE USER S MANUAL 1 2 CONTENTS 1. APPLICATION... 5 2. BASIC REQUIREMENTS, OPERATIONAL SAFETY... 5 3. INSTALLATION... 5 3.1. Assembly...5 3.2. Electrical Connections...7

More information

Operating instructions. Switching amplifier DN0210 DN / / 2015

Operating instructions. Switching amplifier DN0210 DN / / 2015 Operating instructions Switching amplifier DN0210 DN0220 UK 80011079 / 00 01 / 2015 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5

More information

Instruction Manual. Instrument Amplifier HE 017

Instruction Manual. Instrument Amplifier HE 017 Instruction Manual Instrument Amplifier HE 017 List of contents INSTRUCTION MANUAL: HE 017 Date: 20.4.99 Art.-No.: 320999 S General Safety Instructions.......................................... S 1 General

More information

IO-PT4. Component identification. User safety and equipment protection guidelines. Unitronics Industrial Automation Systems 1

IO-PT4. Component identification. User safety and equipment protection guidelines. Unitronics Industrial Automation Systems 1 IO-PT4 I/O Expansion Module 4 PT100 Inputs (-50 to 460 C) The IO-PT4 is an I/O expansion module that can be used in conjunction with specific Unitronics OPLC controllers. The module offers 4 PT100 inputs

More information