R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V4.30SP1

Size: px
Start display at page:

Download "R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V4.30SP1"

Transcription

1 R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V4.30SP1 These Release Notes are for following models of the R&S FSW Signal and Spectrum Analyzer: R&S FSW8, order no K08, R&S FSW13, order no K13, R&S FSW26, order no K26, R&S FSW43 order no K43, R&S FSW50, order no K50, R&S FSW67, order no K67, R&S FSW85, order no K85 R&S FSW8, order no K08, R&S FSW13, order no K13, R&S FSW26, order no K26, R&S FSW43 order no K43, R&S FSW50, order no K50, R&S FSW67, order no K67, R&S FSW85, order no K Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, Munich, Germany Phone: Fax: Internet: Subject to change. R&S is a registered trademark of Rohde & Schwarz GmbH & Co. KG. Trade names are trademarks of the owners. The following abbreviations are used throughout this document: R&S FSW is abbreviated as. Release Notes Test and Measurement PAD-T-M: /02.00/CI/1/EN/

2 Table of Contents New s Modified s Improvements Known Issues Windows Exchanging Solid State Disks Between Different Types of CPU Boards Missing smartcard Message Modifications to the Documentation Firmware Update Performing the Firmware Update on the Instrument Performing the Firmware Update from a Windows PC Operation with and without Administrator Rights Installing Firmware Options Firmware Options Included in the Basic Instrument Other Firmware Options within the FSWSetup.exe File Enabling Options by Entering Option Key Codes Customer Support Release Notes

3 1 1.1 New s The following table lists the new functions and indicates the version in which the new function was introduced: New functions of Firmware V4.30 Version V4.30 FSW-K97: Supports WLAN IEEE802.11ay measurements V4.30 Support for: FSW-K512RE: 512MHz Real-Time Measurement Application, POI > 15us FSW-K800RE: 800MHz Real-Time Measurement Application, POI > 15us V4.30 The behavior of the number block (on the front panel of the instrument) for text input in entry fields and dialogs can now be switched from text (default) to number V4.30 Support for Thermal Waveguide Power Sensor WR-15 NRP75TWG, order no Thermal Waveguide Power Sensor WR-12 NRP90TWG, order no Thermal Waveguide Power Sensor WR-10 NRP110TWG, order no V4.30 FSW-B10: Supports generator control for R&S SMB100B and R&S SMBV100B V4.30 FSW-K161R/B512R/B800R: Supports 4 vertical lines and multi-touch gestures. V4.30 FSW-K17: Offers generator control functionality interacting with signal generator SMW-K61 option. V4.30 FSW-K18: Generator control can now be activated or deactivated with an extra ON/OFFbutton V4.30 FSW-K18: The Read and Load Current Signal from Generator functionality now also supports the R&S SMW-K355 application V4.30 FSW-K19: Offers generator control functionality interacting with signal generator SMW-K811 to coordinate the notch settings between analyzer and generator V4.30 FSW-K30: Supports measurements with variable RBW and sweep time for frequencies below 10 MHz V4.30 FSW-K70: Supports equalizer data in Frequency Response Correction format.fres for K544 V4.30 FSW-K70M: Supports separate traces in the I/Q constellation plot (only pattern symbols/only data symbols/all symbols) V4.30 FSW-K70: Time marker coupling in MSRA V4.30 FSW-K144/K145: Signal demodulation and analysis in line with TS V Supports time alignment measurements in line with TS /2 V1.2.0 Supports transmit on/off power measurement in line with TS V1.2.0 Supports ACLR and SEM measurements for FR1/FR2 (in line with TS /2 V1.2.0 and TS /2 V15.0.0) Supports channel decoding via bitstream (PDSCH/PUSCH) Supports 3GPP test models in line with TS V1.2.0 Supports antenna port mapping and CDM group configuration Supports of segmented capture mode Supports channel estimation Pilot and Payload and Linear Interpolation Supports phase tracking Pilot + Payload Supports crest factor Supports SS/PBCH block offset optionally relative to reference point A Supports carrier axis optionally relative to center frequency Supports 15kHz subcarrier spacing for SS/PBCH blocks for channel bandwidths > 50 MHz Release Notes

4 New functions of Firmware V4.21 Version V4.21 When using the independent gate source in addition to the trigger source (double trigger), also independent trigger and gate offsets are now supported. New functions of Firmware V4.20 Version V4.20 Support for: FSW8, order no K08, FSW13, order no K13, FSW26, order no K26, FSW43, order no K43, FSW50, order no K50, FSW67, order no K67, FSW85, order no K85 (including 2 nd 1.85mm (V) connector up to 67 GHz) V4.20 Support for: FSW-B MHz Analysis Bandwidth for FSW26/43/50/67/85 with order no xy FSW-B MHz Analysis Bandwidth for FSW26/43/50/67/85 with order no xy V4.20 Support for: FSW-K161R 160 MHz Real-Time Measurement Application, POI 15 µs for FSW with order no xy FSW-B800R 800 MHz Real-Time Spectrum Analyzer, POI 15 µs for FSW26/43/50/67/85 with order no xy (including 2000 MHz analysis bandwidth) V4.20 FSW-B1200/B800R/B2001 can be combined with FSW-B5000 for FSW43/85 with order no xy V4.20 FSW-B21 supports: RPG FS-Z75, order no RPG FS-Z90, order no RPG FS-Z110, order no V4.20 SCPI recording supported for all FSW order numbers in following modes: Spectrum, I/Q Analyzer, MSRA Mode, K6, K7, K17, K18, K30, K40, K60, K70, K72/73, K91, K100, K106, K118/119, K144/145, K201 and Real-time Mode. V4.20 Support for multi-touch gestures (needs FSW with order no xy): Pinching and spreading with two fingers to adjust the traces in a diagram is now possible. V4.20 Support for dragging in x and y direction to adjust the traces in a diagram is now possible for all FSW order numbers: Traces can be comfortably shifted in a diagram in x- and y-axis by simply dragging to the new position. V4.20 Besides the graphical zoom, the new zoom mode measurement zoom is supported. In contrast to the graphical zoom, which is merely a visual tool, the measurement zoom adapts the measurement settings (e.g. reference level, level range, frequency span) such that the data is measured in the required detail. The measurement zoom is the default zoom mode for spectrum mode, I/Q analyzer and K7. The context menu of the measurement zoom toolbar icon can be opened with a long touch. Level-Lock (default on), X-Lock and Y-Lock can be used to prevent changing specific settings by the zoom operations. V4.20 New, Windows10-like, user interface design (industrial design theme) for all FSW order numbers Release Notes

5 New functions of Firmware V3.20 Version V3.20 Support of an independent gate source in addition to the trigger source. This allows double trigger measurement scenarios. In this case, the trigger source defines when a measurement is generally possible and the gate source determines which data is actually measured. V3.20 I/Q Analyzer & Options supporting I/Q data input: For the Input Source I/Q File a repetition number is supported which determines how often the data stream is copied in the I/Q data memory V3.20 FSW-B517 Dig IQ 40G streaming out interface is now supported by options FSW-K6, FSW- K7, FSW-K60 and FSW-K70 V3.20 FSW-K30: Support of calibration loss. Temperatures can be considered for input loss, output loss and calibration loss. V3.20 FSW-K40: Support for measuring up to 300 GHz frequency offset V3.20 FSW-K54: Support units dbµv/mhz, dbµa/mhz, dbmv/mhz and dbpw/mhz. With appropriate transducers also dbµv/mmhz and dbµa/mmhz is supported. V3.20 FSW-K91AX: Signal Content Detailed provides the L-LTF power result. :FETCh:SCDetailed:POWer:PPDU? LLTF V3.20 FSW-K91N/AC/AX: New setting to resolve the cyclic shift delay (CSD) ambiguity in case of SISO signals. :CONFigure:WLAN:MIMO:CSD <APPLy IGNore> V3.20 FSW-K100/101: Support for measuring additional operating band limits in Spectrum Emission Mask measurement V3.20 FSW-K106: New result RB power excluding E-UTRA (dbm) in result summary of downlink inband mode. This result is necessary for calculating the NB-IoT RB power dynamic range in 3GPP specification chapter V3.20 Support of FSW-K145 5G-NR Uplink Measurements Support of Transform Precoding V3.20 FSW-K144/K145: Signal demodulation and analysis in line with TS Support of ACLR and SEM measurements (in line with TS and TS ) Support of multi component carrier and multi frame analysis Support of channel decoder for PBCH (MIB) Support resource grid adjustment according to reference point A Support of phase compensation for upconversion to carrier frequency Support of EVM vs RB result display Support custom ID for DMRS pseudo-random sequence initialization (CORESET, PDSCH, PUCCH, PUSCH) Support DMRS power setting rel. to data (PBCH, CORESET, PDSCH, PUCCH, PUSCH) Support ignoring DC for analysis V3.20 FSW-K544: Support of indication for each file if it is included in the combined frequency response or not Release Notes

6 New functions of Firmware V3.10 Version V3.10 Support of FSW-B517 DIG IQ 40G streaming out interface V3.10 Support of FSW-U85 improved sensitivity for frequencies > 43.5 GHz in YIG preselector bypass path (order number ) V3.10 Firmware supports Windows 10 operating system V3.10 RF Probe support with RT-ZA9 N-Type Adapter for RT-Zxx oscilloscope probes V3.10 FSW-B10: Support generator control for R&S SMA100B V3.10 FSW-B71: Support of RT-ZM Modular Probes V3.10 FSW-K7: Support of FSW-B2071 Oscilloscope Baseband Inputs V3.10 FSW-K7: New marker function AF Phase. This function allows relative signal delay measurements for the result displays AM/FM/PM spectrum. V3.10 FSW-K50: Support of carrier guard interval, carrier search range settings and display of spurs relative to the carrier V3.10 FSW-K70: Support for rotating PSK/QAM modulations in FSW-K70M V3.10 FSW-K70: Added DVB-S2(X) frame structure files and start of frame pattern V3.10 FSW-K70: A new softkey DVB-S2(X) Config has been added to the MEAS menu, this launches the DVB-S2(X) configuration tool. V3.10 FSW-K70: Support of new Bluetooth standards Bluetooth_5_LE_S2 and Bluetooth_5_LE_S8 V3.10 FSW-K70: New function Pattern Symbols are Part of PRBS added to FSW-K70P V3.10 FSW-K102/103: Support of carrier aggregation measurements for up to 5 CCs V3.10 FSW-K144: Support of Phase Tracking Reference Signal, Support of capture time < 10ms, Support of MSRA mode, Support Burst Set Periodicity of 5ms Analysis in line with TS V3.10 FSW-K544: Support of preview for selected file magnitude and phase, covered frequencies and combined frequency response for I/Q mode and spectrum mode V3.10 Remote Mode: To prevent a manually switch on of the screen with the softkey Display Update On/Off the remote command SYSTem:DISPlay:LOCK <on/off> can now be used to lock this soft key. Release Notes

7 1.2 Modified s The following table lists the modified functions and indicates the version in which the modification was carried out: Modified functions of Firmware V4.30 Version V4.30 Grid annotations displayed semi-transparent in front of traces to enhance the readability of the annotation values. V4.30 Online help now also supports result displays in FSW-K18, FSW-K30 and FSW-K40. V4.30 FSW-K18: Default Synchronization Confidence has been reduced to 80%. V4.30 FSW-K18: The data acquisition time now increases automatically when changing the File Repetition setting for file input. V4.30 FSW-K30: Maximum number of measurement points increased to V4.30 FSW-K144: SS/PBCH block results are obtained individually in allocation summary and bitstream. SCPI queries (TRACe<n>:DATA?) were adapted accordingly. V4.30 FSW-K144/K145: PDSCH/PUSCH DMRS rel. power is adapted according to TS V V4.30 FSW-K144: Update of PBCH channel decoder results for latest 3GPP specifications V4.30 FSW-K144: Improved auto detection of SS/PBCH configuration parameters V4.30 FSW-K145: Improved synchronization for signals containing PUCCH only. V4.30 FSW-K144/K145: If only one BWP is configured, the k_0 parameter can be set to -6, 0 or +6. Modified functions of Firmware V4.20 Version V4.20 Band breaks (e.g. 8 GHz) aligned with beginning of a measurement point. Modified functions of Firmware V3.20 Version V3.20 Hardware setup at 30.8 GHz optimized for satellite applications. V3.20 Max. frequency for center- / stop frequency can be entered via rotary knob. V3.20 FSW-K7: The Marker AF Phase Display result is now displayed with 5 decimal digits in unit rad, 2 digits in unit mrad and 3 digits in unit deg. V3.20 FSW-K18: The trace export dialog has been redesigned and improved. It is now also possible to export all traces of a selected window. V3.20 FSW-K18: The attenuator of the connected SMW is no longer changed by FSW-K18 during Parameter Sweep. This leads to more stable Parameter Sweep results when sweeping over Input Power. V3.10 FSW-K18: Regardless of the Evaluation Range settings, always the complete reference signal duration was evaluated when calculating the Input Power and the Gain. In future the evaluation range settings will be taken into account. V3.20 FSW-K40: It is now possible to edit the Nominal Frequency & Level during a tracked measurement V3.20 FSW-K70: The default directories for file manual loading (e.g. user modulation, user filter) have changed V3.20 FSW-K70: SCPI commands that involve file loading can now be defined relative to the default directory path, too. (Absolute paths still work as before). V3.20 FSW K144: Change of default settings: PDSCH mapping type A, SS/PBCH block 3 and 4 off V3.20 FSW K144: When loading I/Q data, the 5G NR signal configuration is now loaded from the I/Q file Release Notes

8 Version V3.20 FSW K144: CFO error estimation/compensation is now calculated sub frame wise V3.20 FSW K144: Modified trigger to frame accuracy V3.20 FSW K144: Total power is now measured over occupied slots (of minimum numerology) V3.20 FSW K144: RBs overlapping with SS/PBCH blocks are now viewed as not used for transmission as defined in TS V V3.20 FSW K144: New restrictions for PDSCH/PUSCH allocations according to TS Table / Modified functions of Firmware V3.10 Version V3.10 Legacy Pro: FSU emulation for electronic attenuation: Stop Frequency now limited to 3.6 GHz if electronic attenuation is switched on. V3.10 FSW-K7: The Modulation Frequency result is now display with 7 decimal digits. V3.10 FSW-K30: The frequency in trace export files now reflects either RF or IF. V3.10 FSW-K30: For ENR table import the ENR table values can now be separated by spaces or commas V3.10 FSW-K70: In K70M the ideal constellation points now takes boosting into account. V3.10 FSW-K106: Extended F_offset for SEM measurement to +/- 200 khz in case of standalone distribution according to table 5.6-3A of 3GPP of document revision 14 V2.90 B160R: Power vs Time Average Peak Detector algorithm modified. V3.10 FSW-K193: An information message is now displayed when preconditions for Synchronous Band Power measurements are not met Release Notes

9 1.3 Improvements The following tables list the improvements and indicate since which version the issue could be observed: Improvements of Firmware V4.30SP1 Since V2.80 Devices with the I/Q memory extension FSW-B106 and without a Real-Time option: I/Q measurements did not always finish if the extended memory was used. This issue is solved. Improvements of Firmware V4.30 Since V4.21 For a certain setup in spectrum emission mask measurement with active preamplifier, a level drop in the noise floor between 3 GHz and 4 GHz could be observed. This issue is solved. V4.21 When the virtual keyboard has been minimized and the keyboard hardkey was pressed, the virtual keyboard could not be brought to the foreground. This issue is solved. V2.60 FSW-B33 does not show up in the version and option dialog although the USB Mass Memory Write Protection is active. The functionality of the write protection was not affected. This issue is solved. V2.70 FSW-K7: Relative FM/PM results were not updated in the result table. Remote Control was not affected. This issue is solved. V2.40 FSW-K17: Phase difference trace showed unwanted phase-wrapping when measuring signals with strong phase distortion. This issue is solved. V4.21 FSW-K18: The indices returned for the TRACe<n>:DATA? command by the SCPI recorder did not match the selected result display. This issue is solved. V4.20 FSW-K91: Very short PPDUs (1 to 5 symbols) gave too optimistic EVM if payload channel estimation was used. In this case, a warning message was displayed. This issue is solved. V3.10 FSW-K91ax, improved channel estimation for subcarriers close to the DC subcarriers. This issue is solved. V3.10 FSW-K91ax HE-MU, improved power readings for leftmost or rightmost RU26. This issue is solved. V4.20 FSW-K161R/B512R/B800R: Setting a reference level other than 0 dbm prior to pressing the activate button in the edit frequency mask dialog applied an unintended level offset to the mask points. This issue is solved. V2.80 FSW-K161R/B512R/B800R: Persistence delta marker was misplaced for a reference level other than 0 dbm. This issue is solved. V2.10 FSW-K161R/B512R/B800R: "Trigger Armed" signal was not sent when the trigger was armed in Real-Time Spectrum or MSRT. This issue is solved. V3.20 FSW-K144: For DL configuration without SS/PBCH blocks at deploy frequency range <6GHz, synchronization did not work properly in certain cases. This issue is solved. V3.20 FSW-K144/K145: PDSCH/PUSCH DMRS power relative to PDSCH/PUSCH was not repeated per slot / not loaded from allocation file. This issue is solved. V3.20 FSW-K144/K145: Expected DMRS symbol positions were shifted for mapping type A and PDSCH/PUSCH configured with a symbol offset not equal to 0. This issue is solved. V3.20 FSW-K144/K145: In certain scenarios, the RF upconversion phase compensation was not adapted properly. This issue is solved. V4.20 FSW-K544: If files were loaded that did not have an overlapping frequency range, an interpolation was performed for the gaps in between. This issue is solved. V4.20 FSW-K544: Wraps in phase could cause drops in the frequency response correction. This issue is solved. Release Notes

10 Improvements of Firmware V4.21 Since V3.20 FSW-K6: With segmented capture mode, the markers would not account for gaps between segments in the displayed stimulus. This issue is resolved. V3.20 For a DL configuration without SS/PBCH blocks and with a deploy frequency range under 6 GHz, synchronization did not work properly in certain cases. The issue is solved. Improvements of Firmware V3.20 Since V2.60 FSW-B512R: Persistence Diagram Style was not restored correctly when switching from Probability Mask to Free Run. This issue is solved. V3.10 RT-ZM probes: The setup of the probe modes differential mode (DM) and common mode (CM) was mixed up, resulting in the displayed DM level showing the CM level and vice versa. This issue is solved. V3.00 I/Q Analyzer: During I/Q Vector ASCII Trace Export, the I-values were sorted in ascending order. This issue is solved and these values are not sorted anymore. V3.10 FSW-K6: The order of table headers and table column data did not correctly match up in an exported CSV file. This issue is solved. V3.10 FSW-K7: With active AF Phase Marker in some situations, the sweep did not finish. This issue is solved. V1.10 FSW-K10: For GMSK modulated signals with very severe corruption of the TSC the estimator for the Modulation Accuracy table results amplitude droop, IQ Imbalance and IQ Offset might not converge. This issue is solved. V3.10 FSW-K18: The equalizer of FSW-K18 unintentionally also compensated for the Gain of the DUT. This issue is solved. V3.10 FSW-K18: Files created by the Trace Export functionality always contained the instrument settings, even when Include Instrument & Measurement Settings was disabled. This issue is solved. V3.10 FSW-K40: Reference Measurement did not work below 30 Hz, this issue is solved. V1.92 FSW-K60: Unit in Signal States Table changed to Hz/us for Signal Model Chirp V3.00 FSW-K70: Occasionally, mostly after remote operation, the blue marker that highlights the currently analyzed result range was not visible. This issue is solved. V3.00 FSW-K70: Improved measurement speed, if pattern search is active. V3.10 FSW K144: Improvements of SS/PBCH auto detection V3.10 FSW K144: In case 5ms synchronization burst set periodicity was selected, the application did not sync occasionally. The issue is solved. V3.10 FSW-K544: Gaps between files caused an interpolation. This issue is solved. V3.10 FSW-K544: Wraps in phase frequency response may have caused drops in magnitude diagrams for IQ data. This issue is solved. Release Notes

11 1.4 Known Issues The following table lists the known issues and indicates since which version the issue could be observed: since V4.20 The measurement zoom button in the toolbar is always available although some options are not intended to support this feature and the button should be unavailable. V4.30 License key installation over R&S License Manager webpage often shows a red error message in the last line. Nevertheless, the license key has been properly installed on the device, which can be seen after the reboot. V4.20 Touch gestures with a y-scaling in linear percent or linear with unit do not always work. V4.20 MSRA mode in I/Q Analyzer or K7: Dragging X-axis which are reflecting a time unit has no effect. V1.81 If the additional software R&S DigIConf should be installed directly on the FSW, please use the version V Build220 SP1 or newer (Beta available upon request) older version may have data transfer problems or show white screens. The R&S DigIConf software might prevent windows to shut down occasionally. If this happens, please select "Force shut down" and confirm. V2.80 FSW-B512R: Attenuation occurs further than 240MHz away from the center frequency in replay zoom. V2.40 FSW-B512R: For some resolution bandwidths, a single sweep causes artifacts in the spectrum that are caused by the instrument itself. V2.40 FSW-B160R/B512R: Kaiser window not supported. If selected a Blackmann Harris window is applied. V1.90 FSW-B160R: Auto Adjust functions are not available in MSRT mode. V1.80 FSW-B160R: No full support of trigger outputs. "Device Triggered" mode for "Frequency Mask Trigger" is available for trigger output 3 only. V2.20 FSW-B2000: In some situation after recalling a file with active B2000 configuration the connection to the R&S RTO is not established. Workaround: After the recall switching the B2000 state off and on again should solve the problem. V3.20 FSW-K6 with FSW-B512/B1200/B2001: First sweep with gauss filter at 200 MHz may stop. Change the BW and back to 200MHz will resume operation. V2.90 FSW-K6: When using B2000 or B5000 with segmented capture mode, the timestamps and PRI values may not be correctly shown V2.80 FSW-K6: Segmented capture with FSW-B512R can produce incorrect segment lengths. V2.70 FSW-K6 with FSW-B512/B1200/B2001: When using the external or IF power trigger then the segmented capture timing may become inaccurate for short segment lengths (under 400ns) or small amount of pre-trigger samples (under 100ns). V2.60 FSW-K6/K60: When using the IF power with bandwidths 1.2 GHz (B1200) or 2 GHz (B2001) then the segmented capture timing may become inaccurate for short segment lengths (under 400ns) or small amount of pre-trigger samples (under 100ns). V2.40 FSW-K6: The user can set a larger amount of segment data than is actually possible with certain RTO configurations (using FSW-B2000). In this case, the lower number of available segments will be analyzed and a warning will appear in the status bar. V2.20 FSW-K6: The view trace behaves like a clear write trace. V1.51 FSW-K6: After recalling a Pulse channel which was saved after a RUN SINGLE operation, the RUN SINGLE button is lit but no measurement is running. V1.41 FSW-K6: After aborting a measurement via SCPI (:ABORt) the status bar still shows "Measuring...". V1.41 FSW-K6: The following commands are not available: [SENSe:]FREQuency:CENTer:STEP [SENSe:]FREQuency:CENTer:STEP V1.60 FSW-K10: Cumulating of Modulation Spectrum limits of carriers of a multicarrier BTS not supported for frequency offsets < 1.8 MHz of the outmost carrier. Release Notes

12 since V4.30 FSW-K18: Due to an unintended correction mechanism, Baseband I Input Voltage and Baseband Q Input Voltage influence each other if there is a frequency error or a sample rate error. To avoid this, the external references of R&S SMW and need to be coupled. V4.30 FSW-K18: On instruments equipped with analog baseband inputs, the displayed capture time is sometimes not updated when reducing the capture length below a certain number. V3.20 FSW-K18: The crest factor of signals created by the Generate Own Reference Signal - functionality lies in a range between about 9.5 db and 10.5 db even if higher or lower values were selected in the dialog. Since version 4.30 it is no longer possible to select values lower than 9.5 db or higher than 10.5 db. V2.90 FSW-K18: Markers are not available for Parameter Sweep measurements. V2.70 FSW-K18: Pressing softkey "Update DPD Tables on Generator" directly after starting the application may lead to a lock up of the instrument. V2.60 FSW-K60: Adding a marker to the spectrum resets selected frame in the spectrogram. V2.70 FSW-K70: CALCulate<n>:MSRA:WINDow<n>:IVAL? and the displayed analysis range in the MSRA only works reliably for capture lengths <= Samples V2.70 FSW-K70: CALCulate<n>:MSRA:WINDow<n>:IVAL? and the displayed analysis range in the MSRA only works reliably for capture lengths <= Samples V2.60 FSW-K70: In remote mode, if the capture buffer length exceeds symbols and if the command SENSE:DDEMOD:SEARCH:MBURST:CALC is sent right after the finished measurement, occasionally the trace in the Mag(Capture Buffer) display is not visible. Workaround: Send refresh command (INIT:REFM) and resend SENSE:DDEMOD:SEARCH:MBURST:CALC command. V2.60 FSW-K91N: For specific n antenna 2, 3,.. signals with cyclic shift delay the EVM might drop by approx. 1 db. V2.60 FSW-K91AX HE Trigger-based PPDU analysis: In some cases the auto bandwidth detection might fail. A work around it to set the bandwidth of the signal to be measured manually. FSW-K91AX HE Trigger-based PPDU analysis: For 40MHz signals with short GI/HE-LTF size, the auto detection might fail. A work around it to set the GI/HE-LTF size of the signal to be measured manually. FSW-K91AX HE MU PPDU: Only manual PPDU Configuration is supported FSW-K91AX Ext Range SU PPDU: Only RU242 size is supported. No MIMO is supported. V2.50 FSW-K91/192/K193: In some cases (using FSW-B512) auto level might not produce optimal results. Manual adjustments of the amplitude settings may be required in these cases. V2.00 FSW-K91: Please use not more than three FSW-K91 channels - at maximum capture time - to ensure that the memory of the application is not exceeded. V1.90 FSW-K91: When measuring MIMO and recalling a save set containing the same IPaddresses of the connected devices the measurement will stop working. Please delete the shutdown file 'C:\R_S\Instr\results\Shutdown.dfl' and power cycle the FSW on the rear main switch. V1.42 FSW-K91, -K91N, -K91AC: In case the instrument is waiting for a trigger this message is not displayed in the status bar. V2.90 FSW-K95: When loading an I/Q file, an incorrect warning is displayed occasionally. The warning states that the sample rate of the file was different from the sample rate of the application even though the sample rate of the file is equal to the application sample rate. Changing the sample rate hence and forth makes the message disappear. Release Notes

13 1.5 Windows 10 FSW running operation system Windows 10 are using Windows 10 IoT Enterprise LTSB. (IoT stands for Internet of Things and is the embedded version of Windows 10. LTSB stands for Long Term Servicing Branch which means a long support period of Windows patches.) 1.6 Exchanging Solid State Disks Between Different Types of CPU Boards For the FSW spectrum analyzer, different CPU boards are in use: IPC10 with order no and IPC11 with order no If the solid state disk is exchanged between FSWs with different types of CPU boards, the screen resolution is only 800x600 and a "Drive Access Error" is displayed by the firmware. Switch off the device and change back to the solid state drive which fits to the CPU board. The device will boot as usual. 1.7 Missing smartcard Message For FSW spectrum analyzers produced July 2017 or later, the message Missing smartcard or smartcard not initialized may appear after starting the device. This only happens if a firmware below V2.60 is used. Either because a downgrade to a version below V2.60 had been performed, or the solid state drive was exchanged with a version below V2.60. Solution: Please install firmware V2.60 or higher. The device will boot as usual. 1.8 Modifications to the Documentation The latest documentation is available for download from the Rohde & Schwarz website at: Release Notes

14 Firmware Update 2 Firmware Update The firmware update file for the is one file including the main firmware version number, e.g. FSWSetup_V1.60.exe. It will be referred to as FSWSetup.exe later in the text. The file can be found on the Rohde & Schwarz web page at Performing the Firmware Update on the Instrument There are three ways to make the setup FSWSetup.exe visible to the device: Using a memory stick: 1. Copy the file to a directory of the memory stick and insert the memory stick into one of the USB sockets of the. Using the remote desktop and copying the installation files to a directory of the instrument: 1. Connect the to your LAN. 2. Start the remote desktop on your PC (C:\winnt\system32\mstsc.exe). 3. Enter the TCP/IP address of the instrument you want to update. Ensure that the local resources > drives option is selected and press the Connect button. (To get the TCP/IP address of the, press the Setup key and then Network + Remote. The IP address consists of 4 numbers between 0 and 255.) 4. Log in to the instrument (user name: instrument and default password ). 5. Copy the FSWSetup.exe from your PC to a new folder, e.g. C:\FWUpdate. You can now access this directory with the FSWSetup.exe from the analyzer firmware. Using a network drive: 1. Connect your to your LAN, and establish a connection to one of your servers. (Ask the local IT administrator for support.) 2. Copy the FSWSetup.exe from your PC to a directory on this server. You can now access the directory with the FSWSetup.exe from the analyzer firmware. Performing the update on the instrument: The firmware update process is performed by the following steps: 1. Switch the instrument on and wait until the analyzer has resumed operation. 2. Press the "SETUP" key, then System Config, and select the Firmware Update tab. A file browser is displayed to select the proper FSW*.exe setup file. Change the path to the drive and directory which you prepared in the step before (USB stick Release Notes

15 Firmware Update directory, remote PC directory or directory on a server) and close the dialog with the Install button. 3. Press the "Next" button to come to the selection of the firmware packages. By default, all applications should be installed. Ensure that the required applications are selected. 4. Press the "Install" button. The firmware is stopped and the installation starts. After a few minutes the system restarts automatically. After the restart the firmware installation is complete. After the firmware update the "UNCAL" flag appears. A self alignment is necessary. 5. Press the "SETUP" key, select "Alignment" and then "Start Self Alignment" to invoke the alignment procedure. 2.2 Performing the Firmware Update from a Windows PC If the firmware version 1.10 or newer is installed on the instrument the new firmware can also be uploaded without using a memory stick or a network drive. Just a LAN connection from the instrument and a Windows PC is necessary. 1. Run FSWSetup.exe on your PC. 2. Select Remote Installation and then Next. 3. Select the packages to install and then Next. HINT FOR FIRE WALL USERS: The FSWSetup.exe communicates with the instruments via LAN. Therefore it is necessary that the FSWSetup.exe may pass the fire wall. After adding it to the fire wall rules, restart the scan by clicking on Rescan. 4. After scanning your LAN subnet all found instruments are listed. Select the instruments you want to update. It is possible to select up to 5 instruments for updating in parallel. Be careful and check twice if you have selected the correct instruments. Depending on your company s network structure, also instruments of other departments will show up! 5. Select Help to display additional help. Select Options for further options. Release Notes

16 Firmware Update 6. Select Install to start the installation. 7. Confirm that you want to reboot the instrument in order to activate the firmware update (the instrument then restarts automatically) 2.3 Operation with and without Administrator Rights With firmware version V1.41 or later, you can operate the analyzer with or without administrator rights. Some administrative tasks (e.g. LXI functions or network configuration) do require administrator rights. Since firmware V1.60, a firmware update is also possible without administrator rights. In the default configuration, auto-login is enabled, and the Instrument account with administrator rights is active. This means that no password is required, and the full functionality of the analyzer is available. An additional user account (user name NormalUser with default password ) is pre-defined. Use standard Windows functionality if you wish to deactivate the auto-login mechanism and activate the NormalUser account. Also refer to the Getting Started Manual. Release Notes

17 Firmware Update 2.4 Installing Firmware Options Firmware Options Included in the Basic Instrument The -K7, -K19, -K33, -K54, - K512RE, FSW-K800RE and -K544 application software packages are included in the basic instrument firmware. Therefore, they do not have a separate item in the installer to be selected Other Firmware Options within the FSWSetup.exe File The -K6, -K10, -K15, -K17, -K18, -K30, -K40, -K50, -K60, -K70, R&S FSW-K72/73, -K76/77, -K82/83, -K84/85, - K91, -K95, -K97, -K100/101/102/104/105, - K106, -K118/K119, -K144/K145, -K161R, - K192/193 and -K201 application software packages have their own installation item and are therefore added to the selection list during the firmware update. Ensure that the checkbox is checked if the installation is requested. Note: The functionality of FSW-K6S is integrated within FSW-K6 and is activated by its own key code. The functionality of FSW-K18D is integrated within FSW-K18 and is activated by its own key code. The functionality of FSW-K70P is integrated within FSW-K70 and is activated by its own key code. The functionality of the FSW-K91p, FSW-K91n, FSW-K91ac and FSW-K91ax are integrated within FSW-K91 and are activated by their own key code. The functionality of the FSW-K60C and FSW-K60H are integrated within FSW-K60 and are activated by their own key code Enabling Options by Entering Option Key Codes This section can be skipped if the option key was entered once. To activate application software packages, you must enter a license key for validation. If a XML-file with an option key was sent to you see the install description below. The license key is in the device certificate or delivered as a part of the software package. The process is performed in the following steps: 1. Press the "SETUP" key. 2. Go to the Versions + Options tab. 3. Press the "Install Option" button. A dialog box is displayed. 4. Enter the option key number using the keypad. Release Notes

18 Firmware Update 5. Press "ENTER". After a successful validation, the "Option Key valid" message is displayed. If the validation failed, the option software is not installed. 6. Reboot the device. Installation of options via XML-file 1. Press the "SETUP" key. 2. Go to the Versions + Options tab. 3. Press the "Install Option by XML" button. A file browser is displayed. 4. Select the path to the XML file (e.g. network drive or USB stick) 5. Press "Select". After a successful validation, the "Option Key valid" message is displayed. If the validation failed, the option software is not installed. 6. Reboot the device. Release Notes

19 Customer Support 3 Customer Support Technical support where and when you need it For quick, expert help with any Rohde & Schwarz equipment, contact one of our Customer Support Centers. A team of highly qualified engineers provides telephone support and will work with you to find a solution to your query on any aspect of the operation, programming or applications of Rohde & Schwarz equipment. Up-to-date information and upgrades To keep your instrument up-to-date and to be informed about new application notes related to your instrument, please send an to the Customer Support Center stating your instrument and your wish. We will take care that you will get the right information. Europe, Africa, Middle East Phone customersupport@rohde-schwarz.com North America Phone TEST-RSA ( ) customer.support@rsa.rohde-schwarz.com Latin America Phone customersupport.la@rohde-schwarz.com Asia/Pacific Phone customersupport.asia@rohde-schwarz.com China Phone / customersupport.china@rohde-schwarz.com Release Notes

R&S ESW Test Receiver Release Notes Firmware Version V1.40 SP1

R&S ESW Test Receiver Release Notes Firmware Version V1.40 SP1 R&S ESW Test Receiver Release Notes Firmware Version V1.40 SP1 These Release Notes are for following models of the R&S ESW EMI Test Receiver: R&S ESW8, order no. 1328.4100.08 R&S ESW26, order no. 1328.4100.26

More information

R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V4.21

R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V4.21 R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V4.21 These Release Notes are for following models of the R&S FSW Signal and Spectrum Analyzer: R&S FSW8, order no. 1331.5003K08, R&S

More information

R&S ESR EMI Test Receiver Release Notes Firmware Version V3.36 SP2

R&S ESR EMI Test Receiver Release Notes Firmware Version V3.36 SP2 R&S ESR EMI Test Receiver Release Notes Firmware Version V3.36 SP2 These Release Notes are for following models of the R&S EMI Test Receiver: R&S ESR3, order no. 1316.3003K03 R&S ESR7, order no. 1316.3003K07

More information

R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V3.10

R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V3.10 R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V3.10 These Release Notes are for following models of the R&S FSW Signal and Spectrum Analyzer: R&S FSW8, order no. 1312.8000K08, R&S

More information

R&S ESU EMI Test Receiver. Release Notes. Firmware Version V5.74 SP1. Only an ESU with Windows 7 Operating System is supported

R&S ESU EMI Test Receiver. Release Notes. Firmware Version V5.74 SP1. Only an ESU with Windows 7 Operating System is supported Release Notes Test and Measurement PAD-T-M: 3574.3288.02/02.00/CI/1/EN R&S ESU EMI Test Receiver Release Notes Firmware Version V5.74 SP1 These Release Notes are for following models of the R&S EMI Test

More information

R&S FSW Signal and Spectrum Analyzer

R&S FSW Signal and Spectrum Analyzer Release Notes Test and Measurement PAD-T-M: 3574.3288.02/02.00/CI/1/EN/ R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V2.00 These Release Notes are for following models of the R&S

More information

R&S RTO and R&S RTE Oscilloscopes Release Notes Firmware Version

R&S RTO and R&S RTE Oscilloscopes Release Notes Firmware Version R&S RTO and R&S RTE Oscilloscopes Release Notes Firmware Version 3.70.1.0 These Release Notes are for following models of the R&S RTO and R&S RTE: R&S RTO 2002, order no. 1329.7002K02, R&S RTO 2004, order

More information

R&S Spectrum Rider FPH Release Notes Firmware Version V1.50

R&S Spectrum Rider FPH Release Notes Firmware Version V1.50 R&S Spectrum Rider FPH Release Notes Firmware Version V1.50 2017 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S RTA4000 Digital Oscilloscope Release Notes Firmware Version

R&S RTA4000 Digital Oscilloscope Release Notes Firmware Version Digital Oscilloscope Release Notes Firmware Version 01.100 These Release Notes describe the following models and options of the R&S Digital Oscilloscope: R&S RTA4004, Digital Oscilloscope order no. 1335.7700K04

More information

R&S SMA100B Release Notes Firmware Version

R&S SMA100B Release Notes Firmware Version R&S SMA100B Release Notes Firmware Version 4.30.040.33 The Release Note is for the following model of the R&S SMA100B RF and Microwave Signal Generator: R&S SMA100B, order no. 1419.8888K02 2018 Rohde &

More information

R&S Spectrum Rider FPH Release Notes Firmware Version V1.40

R&S Spectrum Rider FPH Release Notes Firmware Version V1.40 R&S Spectrum Rider FPH Release Notes Firmware Version V1.40 2017 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

Release Notes. R&S FSC Spectrum Analyzer

Release Notes. R&S FSC Spectrum Analyzer Release Notes Revision: 02 R&S FSC Spectrum Analyzer Firmware Release 2.20 These Release Notes describe the following models and options of the R&S FSC Spectrum Analyzer: R&S Spectrum Analyzer FSC3, order

More information

R&S RTC1002 Digital Oscilloscope Release Notes Firmware Version

R&S RTC1002 Digital Oscilloscope Release Notes Firmware Version R&S RTC1002 Digital Oscilloscope Release Notes Firmware Version 06.100 These Release Notes describe the following models and options of the R&S Digital Oscilloscope: R&S RTC1002 Digital Oscilloscope, order

More information

R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V2.80

R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V2.80 R&S FSW Signal and Spectrum Analyzer Release Notes Firmware Version V2.80 These Release Notes are for following models of the R&S FSW Signal and Spectrum Analyzer: R&S FSW8, order no. 1312.8000K08, R&S

More information

R&S RTB2000 Oscilloscope Release Notes Firmware Version

R&S RTB2000 Oscilloscope Release Notes Firmware Version Oscilloscope Release Notes Firmware Version 02.202 These Release Notes describe the following models and options: R&S RTB2004 Oscilloscope, order no. 1333.1005K04 R&S RTB2002 Oscilloscope, order no. 1333.1005K02

More information

R&S FSVA and R&S FSV Signal and Spectrum Analyzer Release Notes Firmware Version V3.40

R&S FSVA and R&S FSV Signal and Spectrum Analyzer Release Notes Firmware Version V3.40 R&S FSVA and R&S FSV Signal and Spectrum Analyzer Release Notes Firmware Version V3.40 These Release Notes are for following models of the R&S FSVA Signal and Spectrum Analyzer: Running operating system

More information

R&S Pulse Sequencer DFS Software Release Notes Software Version 1.7

R&S Pulse Sequencer DFS Software Release Notes Software Version 1.7 R&S Pulse Sequencer DFS Software Release Notes Software Version 1.7 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S Pulse Sequencer DFS Software Release Notes Software Version 1.6

R&S Pulse Sequencer DFS Software Release Notes Software Version 1.6 R&S Pulse Sequencer DFS Software Release Notes Software Version 1.6 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S DSA DOCSIS Signal Analyzer Release Notes Firmware Version 2.2.1

R&S DSA DOCSIS Signal Analyzer Release Notes Firmware Version 2.2.1 R&S DSA DOCSIS Signal Analyzer Release Notes Firmware Version 2.2.1 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12 164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S VSE Vector Signal Explorer Release Notes Firmware Version V1.40

R&S VSE Vector Signal Explorer Release Notes Firmware Version V1.40 R&S VSE Vector Signal Explorer Release Notes Firmware Version V1.40 These Release Notes are for following models of the R&S VSE Vector Signal Explorer: R&S VSE, order no. 1320.7500.06 2017 Rohde & Schwarz

More information

R&S Broadcast Drive Test Release Notes Software Version 03.13

R&S Broadcast Drive Test Release Notes Software Version 03.13 R&S Broadcast Drive Test Release Notes Software Version 03.13 2015 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12 164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S RECAL + Release Notes Software Version 4.01

R&S RECAL + Release Notes Software Version 4.01 R&S RECAL + Release Notes Software Version 4.01 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S CMW Bluetooth Release Notes Software Version V3.2.70

R&S CMW Bluetooth Release Notes Software Version V3.2.70 R&S CMW Bluetooth Release Notes Software 2014 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S SGT100A RF Source Base Unit Release Notes FW Version

R&S SGT100A RF Source Base Unit Release Notes FW Version R&S SGT100A RF Source Base Unit Release Notes FW Version 3.50.124.71 2017 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12 164 E-mail: mailto:info@rohde-schwarz.com

More information

Release Notes Revision: 01

Release Notes Revision: 01 Release Notes Revision: 01 R&S FSH4/FSH8/FSH13/FSH20 Handheld Spectrum Analyzer Firmware Release 2.81 These Release Notes describe the following models and options of the R&S Handheld Spectrum Analyzer:

More information

R&S SMB100B Release Notes Firmware Version

R&S SMB100B Release Notes Firmware Version R&S SMB100B Release Notes Firmware Version 4.50.074.45 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S ELEKTRA Release Notes Software Version V2.10

R&S ELEKTRA Release Notes Software Version V2.10 R&S ELEKTRA Release Notes Software V2.10 ELEMI-E 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S GNSS Test Automation Release Notes Software Version 1.3.2

R&S GNSS Test Automation Release Notes Software Version 1.3.2 R&S GNSS Test Automation Release Notes Software Version 1.3.2 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S NRP-Z51/-Z52/-Z55/-Z56/- Z57/-Z58 Thermal Power Sensors Release Notes

R&S NRP-Z51/-Z52/-Z55/-Z56/- Z57/-Z58 Thermal Power Sensors Release Notes R&S NRP-Z51/-Z52/-Z55/-Z56/- Z57/-Z58 Thermal Power Sensors Release Notes 2016 Rohde & Schwarz GmbH & Co. KG 81671 Munich, Germany Printed in Germany Subject to change Data without tolerance limits is

More information

R&S TSMA Release Notes Firmware Version

R&S TSMA Release Notes Firmware Version R&S TSMA Release Notes Firmware Version 1.32.04.00 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S WinIQSIM2 Simulation Software Release Notes Firmware Version

R&S WinIQSIM2 Simulation Software Release Notes Firmware Version R&S WinIQSIM2 Simulation Software Release Notes Firmware Version 4.00.048.23 2017 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12 164 E-mail:

More information

Firmware Release V1.83 SP3

Firmware Release V1.83 SP3 Test and Measurement Division Release Notes for R&S ESL EMI Test Receivers with order number: 1300.5001.xx Release Note Revision: 7 Printed in the Federal Republic of Germany Contents History... 3 General

More information

R&S TSMAx Release Notes Firmware Version

R&S TSMAx Release Notes Firmware Version R&S TSMAx Release Notes Firmware Version 03.00.11.00 2019 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

How to deal with the effects of CVE

How to deal with the effects of CVE How to deal with the effects of CVE-2018-0886 Restore Remote Access to R&S instruments V 2.10 Document date: 26.06.2018 This document describes possible ways to access R&S instruments after the security

More information

Release Notes. Revision: 01. R&S FSH4/FSH8/FSH13/FSH20 Handheld Spectrum Analyzer. Firmware Release 2.40

Release Notes. Revision: 01. R&S FSH4/FSH8/FSH13/FSH20 Handheld Spectrum Analyzer. Firmware Release 2.40 R&S FSH4/FSH8/FSH13/FSH20 Handheld Spectrum Analyzer Release Notes Revision: 01 R&S FSH4/FSH8/FSH13/FSH20 Handheld Spectrum Analyzer Firmware Release 2.40 These Release Notes describe the following models

More information

R&S AVG Pattern Import Release Notes Firmware Version 02.25

R&S AVG Pattern Import Release Notes Firmware Version 02.25 R&S AVG Pattern Import Release Notes Firmware Version 02.25 2010-2015 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

How to Apply MS17-10 to prevent WannaCrypt Attacks

How to Apply MS17-10 to prevent WannaCrypt Attacks How to Apply MS17-10 to prevent WannaCrypt Attacks Document date: 8.8.2017 This document describes the procedure how to apply the Microsoft patches to R&S devices running Windows XP or Win7 operating systems

More information

R&S FPS Release Notes Firmware Version V1.40 SP2

R&S FPS Release Notes Firmware Version V1.40 SP2 R&S FPS Release Notes Firmware V1.40 SP2 These Release Notes are for following models of the R&S FPS Signal and Spectrum Analyzer: R&S FPS4, order no. 1319.2008K04 R&S FPS7, order no. 1319.2008K07 R&S

More information

How to Apply MS17-10 to prevent WannaCrypt Attacks

How to Apply MS17-10 to prevent WannaCrypt Attacks How to Apply MS17-10 to prevent WannaCrypt Attacks For Video Tester R&S VTC, R&S VTE, R&S VTS and Video Generator R&S DVSG. Document date: 2017-08-03 This document describes the procedure how to apply

More information

R&S VISA Release Notes Software Version 5.8.6

R&S VISA Release Notes Software Version 5.8.6 R&S VISA Release Notes Software Version 5.8.6 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S InstrumentView Spectrum Analyzer Software Manual

R&S InstrumentView Spectrum Analyzer Software Manual R&S InstrumentView Spectrum Analyzer Software Manual (=E:X2) Software Manual 1321.1040.02 04.00 The software manual describes the following: R&S InstrumentView The R&S InstrumentView supports the following

More information

Firmware Release V2.40

Firmware Release V2.40 Test and Measurement Division Release Notes Firmware Release V2.40 for R&S FSL Spectrum Analyzers with order number: 1300.2502.xx New Features of V2.20: Support for Power Sensors R&S NRP-Z211 and R&S NRP-Z221

More information

R&S CMW500 CDMA2000 1XRTT and 1xEV-DO Release Notes Software Version

R&S CMW500 CDMA2000 1XRTT and 1xEV-DO Release Notes Software Version R&S CMW500 CDMA2000 1XRTT and 1xEV-DO Release Notes Software Version 3.2.81 2014 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail:

More information

Release Note for LTE Measurement

Release Note for LTE Measurement Release Note 3.2.80 for LTE Measurement New Features compared to version 3.2.70: PUCCH / PUSCH / SRS time mask and SRS blanking time mask also for TDD Peak values of the EVM vs subcarrier trace are available

More information

R&S QuickStep Test Executive Software Flexibility and excellent performance

R&S QuickStep Test Executive Software Flexibility and excellent performance Product Brochure Version 05.00 R&S QuickStep Test Executive Software Flexibility and excellent performance QuickStep_bro_en_3607-2249-12_v0500.indd 1 21.12.2017 16:45:09 R&S QuickStep Test Executive Software

More information

R&S InstrumentView Software Manual

R&S InstrumentView Software Manual R&S InstrumentView Software Manual (=E:X2) Version 06.00 1321104002 The software manual describes the following: R&S InstrumentView The R&S InstrumentView supports the following product series: R&S FPH

More information

R&S ETL TV Analyzer Release Notes Firmware Version 3.40

R&S ETL TV Analyzer Release Notes Firmware Version 3.40 R&S ETL TV Analyzer Release Notes Firmware Version 3.40 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12 164 E-mail: mailto:info@rohde-schwarz.com

More information

TD-SCDMA Mobile Station Test Application Firmware R&S FS-K77

TD-SCDMA Mobile Station Test Application Firmware R&S FS-K77 Test and Measurement Division Release Notes TD-SCDMA Mobile Station Test Application R&S FS-K77 Release 4.10 for R&S FSP, FSU, FSQ, FMU, FSUP Analyzer 4.1x New Features: High Dynamic Power vs. Time Measurement

More information

Copyright ANRITSU. All rights reserved.

Copyright ANRITSU. All rights reserved. General Caution --------------- Please use a USB Memory Stick for firmware updates. If there is a firmware update question, please contact Anritsu service support at https://techsupport.anritsu.com, and

More information

R&S InstrumentView Software Manual

R&S InstrumentView Software Manual R&S InstrumentView Software Manual (=E:X2) Version 07.00 1321104002 The software manual describes the following: The supports the following product series: R&S FPH R&S ZPH R&S FPC R&S FSH4 R&S ZVH R&S

More information

R&S MSD Modular System Device Flexible antenna switching and rotator control

R&S MSD Modular System Device Flexible antenna switching and rotator control MSD_bro_en_3606-7082-12_v0300.indd 1 Product Brochure 03.00 Radiomonitoring & Radiolocation Modular System Device Flexible antenna switching and rotator control 28.10.2015 12:46:19 Modular System Device

More information

R&S FSL Spectrum Analyzer Resolving Security Issues When Working in Secure Areas

R&S FSL Spectrum Analyzer Resolving Security Issues When Working in Secure Areas Spectrum Analyzer Resolving Security Issues When Working in Secure Areas Based upon the user s security requirements, this document describes the Rohde&Schwarz options available to address the user s spectrum

More information

R&S NRPV Virtual Power Meter Release Notes Application Version 3.2

R&S NRPV Virtual Power Meter Release Notes Application Version 3.2 R&S NRPV Virtual Power Meter Release Notes Application Version 3.2 2011-2017 Rohde & Schwarz GmbH & Co. KG 81671 Munich, Germany Subject to change Data without tolerance limits is not binding. R&S is a

More information

R&S AVBrun Sequencer Software. Release Notes. Firmware Version Release Notes. Broadcasting

R&S AVBrun Sequencer Software. Release Notes. Firmware Version Release Notes. Broadcasting Release Notes Broadcasting PAD-T-M: 3574.3288.02/02.00/CI/1/EN/ R&S AVBrun Sequencer Software Release Notes Firmware Version 01.93 2012-2016 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich,

More information

Replacing R&S CMU200 with R&S CMW500 in 2G and 3G Speech Test Applications Application Note

Replacing R&S CMU200 with R&S CMW500 in 2G and 3G Speech Test Applications Application Note Replacing R&S CMU200 with R&S CMW500 in 2G and 3G Speech Test Applications Thomas Lechner, 1GPM 17-Jul-14- Replacing R&S CMU200 with R&S CMW500 in 2G and 3G Speech Test Applications Application Note Products:

More information

R&S NRP-Z27/-Z37 Power Sensor Module Release Notes Firmware Version 04.18a

R&S NRP-Z27/-Z37 Power Sensor Module Release Notes Firmware Version 04.18a R&S NRP-Z27/-Z37 Power Sensor Module Release Notes Firmware Version 04.18a 2010 Rohde & Schwarz GmbH & Co. KG 81671 Munich, Germany Printed in Germany Subject to change Data without tolerance limits is

More information

R&S FPS Release Notes Firmware Version V1.50

R&S FPS Release Notes Firmware Version V1.50 R&S FPS Release Notes Firmware V1.50 These Release Notes are for following models of the R&S FPS Signal and Spectrum Analyzer: R&S FPS4, order no. 1319.2008K04 R&S FPS7, order no. 1319.2008K07 R&S FPS13,

More information

R&S FSH4/8/13/20 Spectrum Analyzer Quick Start Guide

R&S FSH4/8/13/20 Spectrum Analyzer Quick Start Guide R&S FSH4/8/13/20 Spectrum Analyzer Quick Start Guide 1309.6269.12 07 Test & Measurement Quick Start Guide The Operating Manual describes the following R&S FSH models and options: R&S FSH4 (1309.6000.04)

More information

Firmware Version Build 170 SP1

Firmware Version Build 170 SP1 Test and Measurement Release Notes R&S DigIConf Software Release Notes Firmware Version 2.20.360.86 Build 170 SP1 2012 Rohde & Schwarz GmbH & Co. KG 81671 Munich, Germany Printed in Germany Subject to

More information

TSML-W Radio Network Analyzer

TSML-W Radio Network Analyzer Version 02.00 TSML-W Radio Network Analyzer February 2007 WCDMA PN scanner WCDMA PN scanning (bands I to IX) with BCH (SIB) demodulation Low power consumption Attractive pricing Handy, portable, and compact

More information

R&S DigIConf Software Release Notes Firmware Version Build 170 SP1

R&S DigIConf Software Release Notes Firmware Version Build 170 SP1 R&S DigIConf Software Release Notes Firmware Version 2.20.360.86 Build 170 SP1 2012 Rohde & Schwarz GmbH & Co. KG 81671 Munich, Germany Printed in Germany Subject to change Data without tolerance limits

More information

3G FDD BTS/3GPP HSDPA BTS

3G FDD BTS/3GPP HSDPA BTS Test and Measurement Division Release Notes 3G FDD BTS/3GPP HSDPA BTS Application R&S FS-K72/FS-K74 Release 4.17 for R&S FSUP Analyzer 4.17 Release Note Revision: 1 Printed in the Federal Republic of Germany

More information

R&S DVMS DTV Monitoring System Release Notes Firmware Version

R&S DVMS DTV Monitoring System Release Notes Firmware Version R&S DVMS DTV Monitoring System Release Notes Firmware Version 02.72.0 2017 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12 164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S SMBV100B Release Notes Firmware Version

R&S SMBV100B Release Notes Firmware Version R&S SMBV100B Release Notes Firmware Version 4.50.074.45 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information

R&S LXI Class C Support V1.21 (XP)

R&S LXI Class C Support V1.21 (XP) Test and Measurement Division Release Notes for R&S LXI Class C Support V1.21 (XP) R&S FSP Spectrum Analyzers R&S FSU Spectrum Analyzers R&S FSG Signal Analyzers R&S FSQ Signal Analyzers Printed in the

More information

1xEV-DO Mobile Station Test

1xEV-DO Mobile Station Test Test and Measurement Division Release Notes 1xEV-DO Mobile Station Test Application R&S FS-K85 Release 4.10 for R&S FSP, FSU, FSQ, FMU, FSUP Analyzer 4.1x Release Note Revision: 2 Printed in the Federal

More information

PC Software R&S FS300-K1

PC Software R&S FS300-K1 Copyright 0 Copyright Copyright Licence Agreement Rohde & Schwarz grants you the right to install the R&S FS300-K1 software package on one or more PCs of your choice. The licence included in the software

More information

R&S ETL TV Analyzer Release Notes Firmware Version 3.51

R&S ETL TV Analyzer Release Notes Firmware Version 3.51 R&S ETL TV Analyzer Release Notes Firmware Version 3.51 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12 164 E-mail: mailto:info@rohde-schwarz.com

More information

cdma2000/1xev-dv Mobile Station Test

cdma2000/1xev-dv Mobile Station Test Test and Measurement Division Release Notes cdma2000/1xev-dv Mobile Station Test Application R&S FS-K83 Release 4.10 for R&S FSP, FSU, FSQ, FMU, FSUP Analyzer 4.1x Release Note Revision: 2 Printed in the

More information

R&S Power Meter NRP Release Notes Firmware Version

R&S Power Meter NRP Release Notes Firmware Version R&S Power Meter NRP Release Notes Firmware Version 06.03.020 2011 Rohde & Schwarz GmbH & Co. KG 81671 Munich, Germany Printed in Germany Subject to change Data without tolerance limits is not binding.

More information

R&S Browser in the Box Release Notes

R&S Browser in the Box Release Notes R&S Browser in the Box 2017-07-07 2017 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com Internet:

More information

Application Firmware R&S FS-K5 Release for R&S FSP, FSU, FSQ, FMU, FSUP Analyzer Firmware 4.1x

Application Firmware R&S FS-K5 Release for R&S FSP, FSU, FSQ, FMU, FSUP Analyzer Firmware 4.1x Test and Measurement Division Release Notes GSM/EDGE Application Firmware R&S FS-K5 Release 4.10 for R&S FSP, FSU, FSQ, FMU, FSUP Analyzer Firmware 4.1x New Features: Extended Configuration Mode for mixed

More information

Agilent 53210A 350 MHz RF Counter Agilent 53220A 350 MHz Universal Frequency Counter/Timer Agilent 53230A 350 MHz Universal Frequency Counter/Timer

Agilent 53210A 350 MHz RF Counter Agilent 53220A 350 MHz Universal Frequency Counter/Timer Agilent 53230A 350 MHz Universal Frequency Counter/Timer Agilent 53210A 350 MHz RF Counter Agilent 53220A 350 MHz Universal Frequency Counter/Timer Agilent 53230A 350 MHz Universal Frequency Counter/Timer Firmware Update 2.01 Release Notes 53210A/53220A/53230A

More information

R&S SFC / SFC-U / VT-B600 Compact Modulator / USB Compact Modulator / Broadcast TX Modulator Release Notes Firmware Version 02.96

R&S SFC / SFC-U / VT-B600 Compact Modulator / USB Compact Modulator / Broadcast TX Modulator Release Notes Firmware Version 02.96 R&S SFC / SFC-U / VT-B600 Compact Modulator / USB Compact Modulator / Broadcast TX Modulator Release Notes Firmware Version 02.96 2012-2017 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich,

More information

Resolving Security Issues When Working with the R&S ESIB in Secure Areas

Resolving Security Issues When Working with the R&S ESIB in Secure Areas Products: R&S EMI Test Receiver ESIB Resolving Security Issues When Working with the R&S ESIB in Secure Areas Based upon the user s requirements, this document describes the Rohde&Schwarz options available

More information

BitLocker White Paper Windows 10

BitLocker White Paper Windows 10 BitLocker White Paper Windows 10 This white paper gives detailed instructions how to enable BitLocker on a measurement device. BitLocker White Paper Windows 10 7.2018 1178.8859.02-01 Table of Contents

More information

AKELA Vector Network Analyzer (VNA) Quick Start Guide

AKELA Vector Network Analyzer (VNA) Quick Start Guide AKELA Vector Network Analyzer (VNA) Quick Start Guide Copyright AKELA, Inc. 2012, all rights reserved http:\\akelainc.com LabVIEW and LabWindows are registered trademarks of National Instruments Incorporated

More information

Data Sheet. ProLock. 3G test set for service. Boosting wireless efficiency

Data Sheet. ProLock. 3G test set for service. Boosting wireless efficiency Data Sheet 2201 ProLock 3G test set for service Boosting wireless efficiency Willtek 2201 ProLock Service Testing for 2G and 3G Mobile Communications Service centres for mobile phones perform repair on

More information

This document contains the firmware release notes for two models. The SDS1202X-E is a dual channel 200 MHz scope.

This document contains the firmware release notes for two models. The SDS1202X-E is a dual channel 200 MHz scope. This document contains the firmware release notes for two models of SIGLENT SDS1000X-E oscilloscopes. The SDS1202X-E is a dual channel 200 MHz scope. The SDS1004X-E is a four channel scope available in

More information

PSA-Manager Version 3 User Manual

PSA-Manager Version 3 User Manual PSA-Manager Version 3 User Manual CONTENTS 1 Introduction...2 2 File Types and Locations...3 3 PSA-Manager Windows...4 4 Viewing Trace Files...7 4.1.1 The Trace Memo sub-window...8 4.1.2 Printing Trace

More information

Release Notes. R&S ES-SCAN EMI Measurement Software

Release Notes. R&S ES-SCAN EMI Measurement Software Release Notes Revision: 09 R&S ES-SCAN EMI Measurement Software Software Release V2.70 These Release Notes describe the following types and options of the R&S ES-SCAN EMI Measurement Software. R&S ES-SCAN,

More information

CPRI RF Analyzer and BBU Emulator

CPRI RF Analyzer and BBU Emulator Measurement Guide CPRI RF Analyzer and BBU Emulator for Anritsu RF and Microwave Handheld Instruments CPRI Common Public Radio Interface BBU Base Band Unit Anritsu Company 490 Jarvis Drive Morgan Hill,

More information

R&S NGE100 Power Supply Series Reduced to the max

R&S NGE100 Power Supply Series Reduced to the max R&S NGE100 Power Supply Series Reduced to the max year Product Brochure Version 01.01 R&S NGE100 Power Supply Series At a glance The R&S NGE100 power supply series consists of robust, high-performance,

More information

N4010A Wireless Connectivity Test Set Firmware Revision History

N4010A Wireless Connectivity Test Set Firmware Revision History N4010A Wireless Connectivity Test Set Firmware Revision History November 17 th 2009 Revision A.05.03.14 November 12 th 2009 - Maintenance release The latest Firmware Upgrade Utility (version 2.0.16) is

More information

Using the Spectrum Management Tools

Using the Spectrum Management Tools CHAPTER 4 Revised: June 28, 2007, Using the Trace Window CBT 3.3 interfaces with the Cisco CMTS to provide return path spectrum analysis, which is provided in the following tasks: Trace Window To monitor

More information

Using the Spectrum Management Tools

Using the Spectrum Management Tools CHAPTER 4 CBT 3.4 interfaces with the Cisco CMTS to provide return path spectrum analysis. The following tasks allow you to use the spectrum management tool effectively: Trace Window Monitors power and

More information

Voice Quality Measurements with R&S ROMES. Application Note

Voice Quality Measurements with R&S ROMES. Application Note Voice Quality Measurements with R&S ROMES Application Note 1SP52 Products: R&S ROMES This document introduces Voice Quality Measurements with R&S ROMES and describes the configuration of the drive test

More information

1. Press "Speed Test" to find out your actual uplink and downlink speed.

1. Press Speed Test to find out your actual uplink and downlink speed. ASRock G10 Gaming Router 6.4 QoS Settings QoS Add Gaming Boost web page. Gaming Boost: Enable or disable the Gaming Boost. Bandwidth The router supports Gaming Boost natively, which identifies and intelligently

More information

SpectrumShot. Software for the GSP-9330 Spectrum Analyzer QUICK START GUIDE REVISION MAY 2016 ISO-9001 CERTIFIED MANUFACTURER

SpectrumShot. Software for the GSP-9330 Spectrum Analyzer QUICK START GUIDE REVISION MAY 2016 ISO-9001 CERTIFIED MANUFACTURER SpectrumShot Software for the GSP-9330 Spectrum Analyzer QUICK START GUIDE REVISION 2.0.0.1 MAY 2016 ISO-9001 CERTIFIED MANUFACTURER This manual contains proprietary information, which is protected by

More information

R&S SMW200A Creating Tutorials and Interactive Examples Application Sheet

R&S SMW200A Creating Tutorials and Interactive Examples Application Sheet Creating Tutorials and Interactive Examples Application Sheet (;ÜT;2) 1178.3611.02 04 Test & Measurement Application Sheet This manual describes firmware version FW 3.20.459.xx and later of the R&S SMW200A.

More information

Using the Spectrum Management Tools

Using the Spectrum Management Tools CHAPTER 4 CBT 3.5 interfaces with the Cisco CMTS to provide return path spectrum analysis. The following tasks allow you to use the spectrum management tool effectively: Trace Window Monitors power and

More information

ThinkRF R5500. Real-Time Spectrum Analyzer 9 khz to 8 GHz / 18 GHz / 27 GHz. Technical Data Sheet Preliminary

ThinkRF R5500. Real-Time Spectrum Analyzer 9 khz to 8 GHz / 18 GHz / 27 GHz. Technical Data Sheet Preliminary Technical Data Sheet Preliminary ThinkRF R5500 Real-Time Spectrum Analyzer 9 khz to 8 GHz / 18 GHz / 27 GHz Featuring Real-Time Bandwidth (RTBW) up to 100 MHz Spurious Free Dynamic Range (SFDR) up to 100

More information

R&S ZN-Z15x Calibration Units User Manual

R&S ZN-Z15x Calibration Units User Manual R&S ZN-Z15x Calibration Units User Manual (=UÉÍ2) Version 02 User Manual 1337596302 This User Manual describes the following R&S ZN-Z15x models: R&S ZN-Z150, N (f), 2 ports, 5 khz to 6 GHz (1335.6710.72)

More information

Fluke 190-Series II Firmware Upgrade V11.46

Fluke 190-Series II Firmware Upgrade V11.46 Fluke 190-Series II Firmware Upgrade V11.46 Requirements 1. Fluke 190-II Series ScopeMeter with firmware prior to V11.46 2. Supported models are: 190-102, 190-104, 190-062, 190-202, 190-204, 190-502, 190-504

More information

TBS2000 Oscilloscope Demo Guide 28 JULY 2016

TBS2000 Oscilloscope Demo Guide 28 JULY 2016 TBS2000 Oscilloscope Demo Guide 28 JULY 2016 Table of Contents TBS2000 Series Oscilloscopes About This Guide & Required Equipment Setting up the Equipment Front Panel Tour Understanding the Display 1.

More information

SensorConnect Changelog. Updated:

SensorConnect Changelog. Updated: SensorConnect Changelog Updated: 2018-09-24 v9.12.11 (2018-09-24) Sensor Output Mode: The latest firmware for the G-Link-200 allows configuring the device to output Tilt channels. Last Known State: The

More information

S2 Family Reversing 2-port VNAs of Planar, S, and Cobalt Series

S2 Family Reversing 2-port VNAs of Planar, S, and Cobalt Series Version 18.3.1 Fixed a bug with the Vector Mixer Calibration menu. Now the *.s2p file of the mixer automatically de-embeds upon completing the calibration process Version 18.2.4 Added T4311 and Z5411 to

More information

Cisco Spectrum Expert Software Local Settings

Cisco Spectrum Expert Software Local Settings CHAPTER 13 Local Settings Overview There are a number of settings you can change to determine how Cisco Spectrum Expert Software operates and presents data. The available settings are for: Sensors and

More information

New Features in Firmware V2.02 (Compared to V2.01) New Features in Firmware V2.01 (Compared to V2.00)

New Features in Firmware V2.02 (Compared to V2.01) New Features in Firmware V2.01 (Compared to V2.00) What's New in this Revision This help describes version V2.10 of the R&S ZVA firmware. Compared to the previous version V2.02, the following True differential mode (option R&S ZVA-K6) Automatic calibration

More information

R&S QuickStep Release Notes Software Version 04.60

R&S QuickStep Release Notes Software Version 04.60 R&S QuickStep Release Notes Software Version 04.60 2018 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29-0 Fax: +49 89 41 29 12-164 E-mail: mailto:info@rohde-schwarz.com

More information