1 DoseLab for TomoTherapy Performing TG-148 QA with DoseLab Christine Gnaster, MS, DABR DoseLab Product Manager AAPM Task Group 148 Background TG-148 could be likened to the TG-142 for TomoTherapy machines Recommendations for TomoTherapyusers to verify the accuracy of their machines Have had users inquire for TG- 148 support With the help of an active TomoTherapyuser, a Quick Guide was developed 1
2 Available TG-148 Products DoseLab RITG148 Y-jaw divergence/beam centering Y-jaw/gantry rotation plane alignment Treatment field centering MLC alignment test Gantry angle consistency Couch speed uniformity Couch translation/gantry rotation Interrupted treatment Laser localization Cheese phantom image quality Y-jaw divergence/beam centering Y-jaw/gantry rotation plane alignment Treatment field centering MLC alignment test Gantry angle consistency Couch speed uniformity Couch translation/gantry rotation Interrupted treatment Laser localization Cheese phantom image quality Is DoseLab right for my clinic? DoseLab is able to provide the metrics required for tests: Some test are multi-step Require the use of a spreadsheet for final calculation User will utilize Custom Tests for data entry Best method for testing DoseLab workflow is with Demo license User Guides are easy-to-follow Provide us with feedback we truly investigate suggested workflow optimizations via user input 2
3 DoseLab stands out We are the only easily automated linear accelerator software (with AutoQA setup) that has this capability Sites with linear accelerators AND TomoTherapy units this opens doors that are otherwise closed TG-148 tests are for annual performance Workflows for TomoTherapymachine QA will easily be recuperated with our software automation for TG-142 QA of linear accelerators Customer-driven integrations Introduction to the Quick Guide 12 page User Guide Two sections Prior to guide usage, it is recommended the user is familiar with TG-148 Imaging setup and techniques not discussed Refer to TG Report 3
4 TG-148 Quick Guide Section First 2 pages of Guide Step by step instructions for tests that are able to be quickly summarized Reference to detailed instructions for 3 tests Involve more steps and extensive explanation TG-148 Detailed QA Section Remaining 10 pages Clear, detailed instructions for each test DoseLab interface graphics for ease of use General directions on utilizing DoseLab: Reference to example spreadsheets and example tolerances as applicable 4
5 Supplemental Materials Example Spreadsheet Example Tolerance Set Or free to create own Current users should add Tolerances manually to preserve current Sets Support can assist! Example Custom Tests created in Database: Machine B Example Images In-software Demonstration V.B.1.F. MLC Alignment V.B.3.A. Gantry Angle Consistency V.B.3.C. Couch Translation/Gantry Rotation 5
6 Why DoseLab? Unparalleled product support in the industry Clinicians on staff: 4 Medical Physicist, 2 Dosimetrists Feedback drives product development At least 1 new version per year Only marketed product that can perform fully automated TG-142 QA analysis for linear accelerators (with AutoQA setup) and have the ability to perform QA for TomoTherapy units No need for self-scripting workflows are pre-built into DoseLab! Try us out! If interested, please contact your Sales Representative or Distributor: Sales@mobiusmed.com All documentation will be provided: Quick Guide Example Spreadsheets for more complex analyses Example Tolerances sets Examples of Custom Test for Machine B sample machine Mobius Medical Systems Support is ready! 6
7 Questions? Monday Friday from 8am to 5pm Central Average response time: 1 hour Most tickets resolved: 24 hours 7
TG-148 overview. Introduction. System Overview. System Overview. QA for Helical Tomotherapy: Report of the AAPM Task Group 148. Conflict of Interest:
QA or Helical Tomotherapy: Report o the AAPM Tas Group 148 Members: Conlict o Interest: r. John Balog owns TomoTherapy stoc. Katja Langen (Co-chair) Nio Papaniolaou (Co-chair) Walter Grant Richard Crilly
More informationQuick Reference Datasheet For All RIT113 Packages
Quick Reference Datasheet For All RIT113 Packages For Rotational Therapies, IMRT & TG142 Highlights and selected product information only. A complete TG142 brochure is available. For more information on
More informationIMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays
IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays Sotiri Stathakis Outline Why IMRT/VMAT QA AAPM TG218 UPDATE Tolerance Limits and Methodologies for IMRT Verification QA Common sources
More informationTomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS
Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate
More informationTable of Contents. About AQUA
User Manual About AQUA Table of Contents 1 About AQUA... 6 1.1 Radiation Therapy... 6 1.1.1 Workflow manager... 6 1.1.2 Reports... 7 1.1.3 Tests... 7 1.1.4 Many Ways to Work... 7 1.2 Technology... 7 1.3
More informationComputational modelling and analysis: IOP Medical Physics Group
Computational modelling and analysis: IOP Medical Physics Group Using Bespoke Matlab Software for the Collection and Processing of Dynalog Files from the Varian Millennium 120 MLC system following delivery
More informationDAILY LINAC QA BEAM QA
BEAM QA DAILY LINAC QA The QA BeamChecker Plus allows for fast, reliable, and uncomplicated daily QA of Varian, Elekta, Siemens, and Accuray Treatment Machines. The QA BeamChecker Plus is specifically
More informationADVANCING CANCER TREATMENT
The RayPlan treatment planning system makes proven, innovative RayStation technology accessible to clinics that need a cost-effective and streamlined solution. Fast, efficient and straightforward to use,
More informationInitial Clinical Experience with 3D Surface Image Guidance
Initial Clinical Experience with 3D Surface Image Guidance Amanda Havnen-Smith, Ph.D. Minneapolis Radiation Oncology Ridges Radiation Therapy Center Burnsville, MN April 20 th, 2012 Non-funded research
More informationEvaluation of AutoQA Lite TM Image Quality Measurement Software
Evaluation of AutoQA Lite TM Image Quality Measurement Software Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4 2XU Phone: 0131 537 1161
More informationph fax Deming Way Middleton WI USA
www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 3120 Deming Way Middleton WI 53562-1461 USA 2007 Standard Imaging, Inc. 1239-21 (03.07) beam qa LINAC AND TOMOTHERAPY QA The NEW
More informationReal-time monitoring of linac performance using RT plans and logfiles
Real-time monitoring of linac performance using RT plans and logfiles NCCAAPM Wisconsin Dells 7 April 2017 Mark Wiesmeyer, PhD, DABR Technical Product Manager Technical talk with insights and results based
More informationMapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More informationMapCHECK 2 & 3DVH The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More informationTHE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!
THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! Confidence in complex treatments Modern radiation therapy uses complex plans with techniques such as IMRT, VMAT and Tomotherapy.
More informationExperiences of daily use of a database for linear accelerator quality control
Experiences of daily use of a database for linear accelerator quality control McGrath N, Stilwell E, Budgell G, Beck J, Aziz T Christie Medical Physics & Engineering Manchester Overview Background Daily
More informationRITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package.
So much is asked of medical physicists these days We Can Help Automated Radiation Therapy Phantom Analysis in Seconds QC for Therapy OBI s and CT simulators. Radia s Catphan/OBI module performs analysis
More information8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA
PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA Jennifer R. Clark, M.S., DABR Conflict of Interest Statement: I am an employee of Sun Nuclear Corporation Outline: Platform overview
More informationVersion 5.6. Quick Start Guide
WWW..COM Version 5.6 Quick Start Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 Jul / 2018 2018 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 DOC # 80714-05
More informationIMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE
QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac
More informationRaising the Bar in IMRT QA
MapCHECK 2TM Raising the Bar in IMRT QA The leader in quick and precise measurement of modulated radiotherapy beams Benefits Proven solution for film-less rotational delivery and IMRT QA - More than 1500
More informationHugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France. Sunday 17 July 2016
Hugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France Sunday 17 July 2016 AGENDA 1. Introduction 2. Material 3. Optimization process 4. Results 5. Comments 6. Conclusion
More informationURGENT IMPORTANT FIELD SAFETY NOTIFICATION
Subject: Incorrect Movement of the Treatment Table Product: MOSAIQ Scope: Sites affected will be those: 1. Running MOSAIQ and, 2. Treating on linear accelerators with the RATM license Notification Released:
More informationVolumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT
Volumetric Modulated Arc Therapy - Clinical Implementation Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA Acknowledgement David M. Shepard, Ph.D. Muhammad K. N. Afghan, Ph.D. Fan Chen, Ph.D.
More informationPET-CT in Radiation Treatment Planning
PET-CT in Radiation Treatment Planning TA van de Water, Radiotherapeutic Institute Friesland, Leeuwarden JA van Dalen, Isala, Zwolle ACKNOWLEDGEMENTS A. van der Schaaf (University of Groningen, University
More informationQuality assurance of a helical tomotherapy machine
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 49 (2004) 2933 2953 PHYSICS IN MEDICINE AND BIOLOGY PII: S0031-9155(04)71892-8 Quality assurance of a helical tomotherapy machine J D Fenwick 1,2,WATomé
More informationFuture is Effortless
AQUATM Radiology is a machine Quality Assurance management solution that standardizes, streamlines and automates QA procedures throughout rediology department and across multiple sites. AQUATM Radiology
More informationINSTRUCTIONS RPD INFORMATION RPD PRODUCT INFORMATION. Address 5218 Barthel Industrial Drive Albertville, MN Website
Expect Service INSTRUCTIONS Radiation Products Design Inc RPD INFORMATION Address 5218 Barthel Industrial Drive Albertville, MN 55301 Website www.rpdinc.com Email sales@rpdinc.com Phone 763-497-2071 or
More informationURGENT IMPORTANT FIELD SAFETY NOTIFICATION
Subject: Product: Incorrect Rotation of the Collimator or Couch ERGO Scope: Sites affected will be those: 1. Running ERGO version 1.7.3 and higher and, 2. Using a Multileaf Collimator (MLC) device for
More informationADVANCING CANCER TREATMENT
3 ADVANCING CANCER TREATMENT SUPPORTING CLINICS WORLDWIDE RaySearch is advancing cancer treatment through pioneering software. We believe software has un limited potential, and that it is now the driving
More informationDelta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear
Delta 4 PT True Volumetric Pre-treatment Verification The difference is clear ScandiDos innovators in advanced volumetric dosimetry solutions ScandiDos ScandiDos is the innovative company that introduced
More informationCT Protocol Review: Practical Tips for the Imaging Physicist Physicist
CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance
More informationFAST, precise. qa software
qa software FAST, precise Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, rotational or brachytherapy plans no film, no phantoms, no linac
More informationGoals. Scope/exclusions. Outline: follow outline of MPPG (plus rationale & some implementation experiences)
AAPM MEDICAL PHYSICS PRACTICE GUIDELINE # : Commissioning and QA of Treatment Planning Dose Calculations: Megavoltage Photon and Electron Beams Jennifer Smilowitz (Chair), University of Wisconsin-Madison
More informationHealthInfoNet AUDITOR REFERENCE GUIDE. Revised: Page 1 of 17
HealthInfoNet AUDITOR REFERENCE GUIDE Revised: 9.2.2014 Page 1 of 17 HealthInfoNet Auditor Reference Guide INSIDE: Accessing HealthInfoNet (HIN) 3-4 Audit Activity 5-9 Resetting Your Password 10-13 Types
More informationIAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques
IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques Report of the Coordinated Research Project (CRP) on Development of Procedures
More information3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending
3DVH : On The Accuracy Of The Planned Dose Perturbation Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools *Patent Pending introduction State-of-the-art IMRT QA of static gantry
More informationAAPM Task Group IT considerations in the electronic chart - Luis Fong de los Santos Ph.D. Mark Parry
AAPM Task Group 262 - IT considerations in the electronic chart - Luis Fong de los Santos Ph.D. Mark Parry IT Infrastructure for Radiation Oncology What is IT? and Why is IT so important? E-Chart System/Environment
More informationS. Guru Prasad, Ph.D., DABR
PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information
More informationIndrin Chetty Henry Ford Hospital Detroit, MI. AAPM Annual Meeting Houston 7:30-8:25 Mon 08/07/28 1/30
Review of TG105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics,
More informationNew Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC
New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it
More informationMachine and Physics Data Guide
WWW..COM Machine and Physics Data Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 May / 2008 2008 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 www.standardimaging.com
More informationSpike for Sign & Graphics Industry. ikegps GROUP LIMITED l MEASURING UP FY15 RESULTS & COMMENTARY
Spike for Sign & Graphics Industry ikegps GROUP LIMITED l MEASURING UP FY15 RESULTS & COMMENTARY 1 Training Agenda About ikegps Spike Value Propositions How Sign Shops Use Spike Spike Product Overview
More informationREF User Manual DOC # STANDARD IMAGING, INC Deming Way Middleton, WI
WWW..COM REF 90501 User Manual STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 Jun / 2011 2011 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 www.standardimaging.com
More informationAutomated Quality Assurance for Image-Guided Radiation Therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Automated Quality Assurance for Image-Guided Radiation Therapy Eduard Schreibmann, a Eric Elder, Tim Fox Department of Radiation
More informationA novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems
A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems Amy S. Yu a) Tyler L. Fowler, and Piotr Dubrowski Department of Radiation
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More information7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT
CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance
More informationAcceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems. Bruce Curran University of Michigan Medical Center Ann Arbor, MI
Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems Bruce Curran University of Michigan Medical Center Ann Arbor, MI Disclosure Some of the work discussed in this talk was supported
More informationVirtual Phantoms for IGRT QA
TM Virtual Phantoms for IGRT QA Why ImSimQA? ImSimQA was developed to overcome the limitations of physical phantoms for testing modern medical imaging and radiation therapy software systems, when there
More informationIMS HEALTH INCORPORATED SALES FORCE EFFECTIVENESS SERVICE PLAN DETAILS
IMS HEALTH INCORPORATED SALES FORCE EFFECTIVENESS SERVICE PLAN DETAILS Subject to the terms set forth herein, IMS will use commercially reasonable efforts to provide to CLIENT the services and support
More information1 Getting Started. 1) Accessing Financial Compass. Gateway: Search for Financial Compass, click on link: Financial Compass. The home page will open:
1 Getting Started 1) Accessing Financial Compass Gateway: Search for Financial Compass, click on link: Financial Compass The home page will open: 2) Initial Steps Step #1) Become familiar with Home Page
More informationTG 132: Use of Image Registration and Fusion in RT
TG 132: Use of Image Registration and Fusion in RT Kristy K Brock, PhD, DABR, FAAPM Associate Professor Department of Radiation Oncology, University of Michigan Chair, AAPM TG 132: Image Registration and
More informationAn approach for measuring the spatial orientations of a computed-tomography simulation system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 2, 2014 An approach for measuring the spatial orientations of a computed-tomography simulation system Meng Chia Wu, a Ramani Ramaseshan Department
More informationTransitioning from pencil beam to Monte Carlo for electron dose calculations
Transitioning from pencil beam to Monte Carlo for electron dose calculations Jessie Huang-Vredevoogd (jyhuang4@wisc.edu) University of Wisconsin NCC AAPM October 12, 2019 1 Topics to cover Background RayStation
More informationThe team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.
Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge Dimitre Hristov Radiation Oncology Stanford University The team Renhui Gong 1 Magdalena Bazalova-Carter 1
More informationRadiation therapy plan checks in a paperless clinic
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Radiation therapy plan checks in a paperless clinic R. Alfredo Siochi, (a) Edward C. Pennington, Timothy J. Waldron, John E.
More informationIHE Radiation Oncology Technical Framework Supplement. Treatment Delivery Plan Content (TDPC) Rev. 1.1 Trial Implementation
Integrating the Healthcare Enterprise 5 IHE Radiation Oncology Technical Framework Supplement 10 Treatment Delivery Plan Content 15 Rev. 1.1 Trial Implementation 20 Date: November 16, 2016 Author: IHE
More informationIpanema Support Service Description. May 18, 2017
Ipanema Support Service Description May 18, 2017 Table of Contents Introduction...4 A. Technical Support Gold Service overview...5 A.1. Overview...5 A.2. Language...5 B. Accessing Technical Support...6
More information4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1
Measurement and Analysis.1 Overview and Underlying Principles.1.1 Introductory Remarks The physics and setup for film dosimetry have been described in the previous chapters. The measurement setup for IMRT
More informationMember Support Help Desk Solution
Member Support Help Desk Solution ComTrax, LLC. 18450 N. 46 th Drive Glendale, AZ. 85308 (800) 729-3083 1 P a g e The following instruction set will show how to use the ComTrax, LLC. Member Support Feature
More informationFeasibility study of performing IGRT system daily QA using a commercial QA device
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 3, summer 2011 Feasibility study of performing IGRT system daily QA using a commercial QA device Jean L.Peng, 1 Darren Kahler, 2 Jonathan
More informationRHM Client Support Plan
RHM Client Support Plan Issue 1 February 2017 Quick Contact Reference: We want to make contacting RHM as easy as possible for our customers therefore you may get your message across via a-few different
More informationCode of practice: Create a verification plan for OCTAVIUS Detector 729 in Philips Pinnacle³
Technical Note D655.208.03/00 Code of practice: Create a verification plan for OCTAVIUS Detector 729 in There are basically three ways to carry out verification of IMRT fluences. For all options, it is
More informationTOMOHD. Product Specifications
TOMOHD Treatment System Product Specifications 1. Complete Treatment System Dimensions Height Width Length Weight 8 ft., 3.25 in. (252.1 cm) 9 ft., 2.50 in. (280.6 cm) 15 ft., 2.25 in. (463.0 cm) 10,000
More information3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools
3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground
More informationAAPM Standard of Practice: CT Protocol Review Physicist
AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose
More informationmykpc Notifications Proactive Notifications
With the goal in mind of delivering your project on-time, Kingery Printing has integrated a unique system of notifications into mykpc, our on-line client console. Through mykpc notifications, we can collaborate
More informationThank-You Members of TG147 TG 147: QA for nonradiographic
Thank-You Members of TG147 TG 147: QA for nonradiographic localization and positioning systems Twyla Willoughby, M.S. Medical Physicist Clinical AAPM Meeting March 2013 Department of Radiation Oncology
More informationStatement of Work IBM Support Services IBM Essential Care Service - Acquired from an IBM Business Partner -
IBM Switzerland Ltd Statement of Work IBM Support Services IBM Essential Care Service - Acquired from an IBM Business Partner - Edition: December 2015 1. Subject The subject of the Service is the provision
More informationUse of TrueBeam developer mode for imaging QA
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Use of TrueBeam developer mode for imaging QA Gilmer Valdes, 1,2a Olivier Morin, 1 Yanisley Valenciaga, 3 Niel Kirby, 1 Jean Pouliot,
More informationINTEGRATED DATA MIGRATION:
Technical Perspectives Integrated data migration INTEGRATED DATA MIGRATION: Patient record transfer with discrete RT data elements ARIA Database DATA MIGRATION Third-party Database The purpose of this
More informationICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588
ICARO Vienna April 27-29 2009 Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588 M. Saiful Huq, Ph.D., Professor and Director, Dept. of Radiation
More informationPhoton beam dose distributions in 2D
Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation
More informationVendor: Microsoft. Exam Code: MB Exam Name: Microsoft Dynamics CRM Online Deployment. Version: Demo
Vendor: Microsoft Exam Code: MB2-715 Exam Name: Microsoft Dynamics CRM Online Deployment Version: Demo Exam A QUESTION 1 You are a systems support specialist for your company. A sales manager is currently
More informationIntegrated proton-photon treatment planning
Pinnacle 3 Proton Planning Integrated proton-photon treatment planning Philips Pinnacle 3 Proton Planning specifications Pinnacle 3 Proton Planning is designed to simplify treatment planning for proton
More informationAdditional Tools and User Options
Overview Getting Started Site Navigation Reports Additional Tools and User Options Help and Contact Information OVERVIEW Welcome to U.S. Bank TrustNow. A secure Internet site that provides you with real-time
More informationTraxsolutions Mobile Scanner Training Guide
Traxsolutions Mobile Scanner Training Guide 6225 N. 24 th St., Suite GL100, Phoenix, AZ 85016 Phone 602.954.9557 Fax 602.954.9559 nfocus.com Austin, TX Chicago, IL Ft. Sill, OK Leavenworth, KS Louisville,
More informationMonte Carlo methods in proton beam radiation therapy. Harald Paganetti
Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization
More informationPrecedence Connector. Zedmed installation guide
Precedence Connector Zedmed installation guide Version 5.8.1 Precedence Health Care 2018 1 of 12 Contents 1.Requirements 3 2. Install Precedence Connector: Standalone Mode.. 4 3. Install Precedence Connector:
More informationOverview of Proposed TG-132 Recommendations
Overview of Proposed TG-132 Recommendations Kristy K Brock, Ph.D., DABR Associate Professor Department of Radiation Oncology, University of Michigan Chair, AAPM TG 132: Image Registration and Fusion Conflict
More informationCLAIMS FILING PROCESS USING OUR TICKET SYSTEM
2017-2018 CLAIMS FILING PROCESS USING OUR TICKET SYSTEM FOR ZERO DEDUCTIBLE STUDENT ATHLETIC / ACTIVITIES ACCIDENT MEDICAL SUPPLEMENTAL BENEFIT PLAN School Claim Filing Instruction Process 1. Download
More informationTPS COMMISSIONING. Laurence Court University of Texas MD Anderson Cancer Center 4/3/2017 1
TPS COMMISSIONING Laurence Court University of Texas MD Anderson Cancer Center 4/3/2017 1 Conflicts of interest Court receives funding from NIH, CPIRT, Varian and Elekta 4/3/2017 2 Resources your first
More informationA secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans
Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2013 A secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans Adam
More informationNHS e-referral Service Transition Planning WebEx May 2015
NHS e-referral Service Transition Planning WebEx May 2015 Issued 19.05.15 V3.0 Purpose of this webinar and key messages To reassure users that they do not need to worry about the change to NHS e-referral
More informationTRASER for Windows ~ QuickBooks. Link Reference
TRASER for Windows ~ Link Reference February 2009 TRA-SER for Windows ~ Link Reference Introduction Trade Service of Australia provides a powerful, easy-to-use electrical database known as TRASER for Windows.
More informationThis table connects the content provided by Education Perfect to the NSW Syllabus..
Education Perfect Maths is a comprehensive online learning and assessment resource. Designed by teachers and written by our in-house team of maths experts, our content aligns to the NSW Syllabus and covers
More informationSecondary 3D Dose QA Fully Automated using MOSAIQ's IQ Engine. MOSAIQ User Meeting May Antwerp
Secondary 3D Dose QA Fully Automated using MOSAIQ's IQ Engine MOSAIQ User Meeting May 31 2013 - Antwerp Contents Project goal and collaboration Secondary 3D Dose QA project justification Secondary 3D Dose
More informationUnique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.
Unique Features of the GE Senoclaire Tomosynthesis System Tyler Fisher, M.S., DABR Therapy Physics, Inc. Conflict of Interest Disclosure I have no conflicts to disclose. Learning Objectives Overview of
More information1. Create a customer account.
The web team request process has been update to provide a faster and fairer service to the entire university. This document will help you through the new process. Please bookmark our new portal. https://westernsydneyu.atlassian.net/servicedesk/customer/portal/2/
More informationAdvanced Parcel Editing. Amy Andis Tim Hodson
Advanced Parcel Editing Amy Andis Tim Hodson Overview What to expect in this technical workshop Review of the Parcel Fabric Data Model Advanced Tips and tricks for Parcel entry Assessing Quality of Parcel
More informationImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing
ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND
More informationSPS Online Store User Tutorial
SPS Online Store User Tutorial New Improvements 2018 The following improvements were made to the SPS Online Store based on criteria extracted from your feedback and user data. Our goal is to continue to
More informationSage Construction Anywhere Connector
Sage Construction Anywhere Connector Release Notes Sage 300 Construction and Real Estate Integration Builds 2.0 and 2.1 10/29/2013 Notice This is a publication of Sage Software, Inc. 2013 Sage Software,
More informationAIDR 3D Iterative Reconstruction:
Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of
More informationWith FARO PointSense for Revit. Jörg Braunes Platform Owner Software Chris Palmer Sales Engineer 1
From Point Clouds to BIM Models With FARO PointSense for Revit Jörg Braunes Platform Owner Software Chris Palmer Sales Engineer 1 Agenda Scan to BIM with Revit PointClouds in Revit Advanced Workflow with
More informationVA EWSR Nuance Support for Dragon Medical 360 Network Edition (DMNE)
For all support requests such as problems, issues, product support, training, webcasts, etc. contact Nuance directly (now open to Clinical End Users, not just Local Dragon Administrators and ITOPS) at
More informationPROVEN TECHNOLOGY IMPROVED
PHANTOMS PROVEN TECHNOLOGY IMPROVED The DoseView 3D is a new 3-axis water phantom for performing comprehensive linear accelerator beam commissioning and quality assurance measurements quickly, accurately
More informationRadiotherapy Plan Competition TomoTherapy Planning System. Dmytro Synchuk. Ukrainian Center of TomoTherapy, Kirovograd, Ukraine
Radiotherapy Plan Competition 2016 TomoTherapy Planning System Dmytro Synchuk Ukrainian Center of TomoTherapy, Kirovograd, Ukraine Beam Geometry 6MV fan beam 3 jaw options 1.0, 2.5 and 5 cm 64 leaves binary
More informationTHE SURVEYING AND ALIGNMENT FOR THE PROSCAN PROJECT. Jean-Luc Pochon Paul Scherrer Institute, 5232 Villigen PSI
THE SURVEYING AND ALIGNMENT FOR THE PROSCAN PROJECT Jean-Luc Pochon Paul Scherrer Institute, 5232 Villigen PSI October 2002 Content 1 The PROSCAN Project...2 2 The Survey Task...3 2.1 The Network...3 2.1.1
More information