State of the art and challenges of time-of-flight PET

Size: px
Start display at page:

Download "State of the art and challenges of time-of-flight PET"

Transcription

1 Physica Medica (2009) 25, 1e11 available at journal homepage: REVIEW PAPER State of the art and challenges of time-of-flight PET Maurizio Conti Molecular Imaging, Siemens Healthcare, 810 Innovation Drive, Knoxville, TN 37932, USA Received 19 June 2008; received in revised form 1 October 2008; accepted 5 October 2008 Available online 19 December 2008 KEYWORDS Positron emission tomography; Time of flight; TOF reconstruction; TOF PET; Scintillation detectors; TOF gain; Signal-to-noise ratio Abstract After a brief review of the history of time-of-flight (TOF) positron emission tomography (PET) instrumentation from the 1980s to present, the principles of TOF PET are introduced,theconceptoftimeresolutionanditseffectontofgaininsignal-to-noiseratio (SNR) are discussed. The factors influencing the time resolution of a TOF PET scanner are presented, with focus on the intrinsic properties of scintillators of particular interest for TOF PET. Finally, some open issues, challenges and achievements of today s TOF PET reconstruction are reviewed: the structure of the data organization, the choice of analytical or iterative method, the recent experimental assessment of TOF image quality, and the most promising applications of TOF PET. ª 2008 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved. A brief history of TOF PET instrumentation The feasibility and advantages of time-of-flight (TOF) positron emission tomography (PET) were identified in the early 1980s [1e3], when the idea of using the time-of-flight information in positron emission tomography was implemented in the first generation of the TOF PET scanner. These scanners were mainly developed by three leading groups: at CEA-LETI in Grenoble with support and evaluation in Orsay, France [4,5]; at Washington University in St. Louis [6,7], Missouri, USA; and at the University of Texas, Houston, Texas, USA [8,9]. But other groups such as the National Institute for Radiological Sciences in Chiba, Japan [10,11] and the University of Washington in Seattle, address: maurizioconti@siemens.com Washington, USA [12] contributed with instrumentation research and development of reconstruction methods. This first generation of TOF PET scanners was based on CsF and BaF 2 scintillators [13e15]. Laval and colleagues [14] reported a best resolution for BaF 2 as low as 156 ps and 212 ps for CsF, for 511 kev. On the system level, however, time resolutions between 470 and 750 ps were reached [15e17]. Low density, low photoelectric fraction and low light output resulted in poor spatial resolution and sensitivity. Moreover, the ultraviolet emission of BaF 2 made light collection difficult and required the use of more expensive photomultipliers (PMT) with quartz windows. Thus, the first generation of TOF PET scanners was limited to the research environment. In a parallel development of non-tof PET systems during the late 1970s and early 1980s, bismuth germanate (Bi 4 Ge 3 O 12 or BGO) [18] began to be used for PET. This scintillator has high detection efficiency, acceptable light /$ - see front matter ª 2008 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved. doi: /j.ejmp

2 2 M. Conti output, and a light emission wavelength of around 480 nm. BGO became the standard in positron emission tomography. Its long decay time and low light output, however, made it unusable in TOF PET. With the success of BGO based PET systems and the low sensitivity of BaF 2, TOF PET development came to a halt. The discovery of new scintillators in the 1990s prepared the way to a new phase of TOF PET research and development. Cerium-doped lutetium orthosilicate or LSO (Lu 2 SiO 5 ) [19] has a relatively high light yield (about 30, 000 photons MeV 1 ), high effective Z (Z eff Z 66), and high density (7.4 g cm 3 ). These characteristics alone pushed LSO to become the standard PET detector for one of the major PET scanner manufacturers [20,21]. Another key parameter of LSO is its short decay time (40 ns). The short decay time was immediately used to reduce the coincidence window from the 12 ns typical of BGO scanners to 6 ns and it was later reduced to 4.5 ns with the development of faster electronics. The short time coincidence window reduced the random coincidences in the acquired data. It became clear that LSO could also be used as a detector for a new generation of TOF PET scanners. Time resolution down to 300 ps has been reported with two single LSO crystals in coincidence, making TOF PET a viable solution [22]. First attempts to perform TOF reconstruction on a commercial LSO PET scanner showed a measurable gain in signal-to-noise ratio due to the TOF reconstruction [23,24], even with a poor 1.2 ns time resolution. The recent discovery of cerium-doped lanthanum bromide (LaBr 3 ) [25] opened another avenue for TOF instrumentation. While lanthanum bromide showed lower stopping power than LSO, it had shorter decay time (16 ns), excellent energy resolution (about 3% at 662 kev), and twice as much light output, making it an attractive candidate for TOF PET [26]. The first LaBr 3 time-of-flight scanner was developed at the University of Pennsylvania [27]. It was based on mm 3 LaBr 3 crystals, assembled in 24 detector modules each of about cm 2, covering 25 cm axial FOV. A continuous light guide coupled each detector array to 24 PMTs. A 7.5% energy resolution allowed for a reduced scatter fraction. A 460 ps time resolution was measured. Recent improvements in triggering technique brought the time resolution to 420 ps, and there is evidence that time resolution could be brought down to 315e330 ps [28]. A first commercial time-of-flight PET scanner, the Gemini TF PET/CT, was recently introduced by Philips in 2006 [29], and uses a LYSO scintillator crystal. LYSO is a material with a structure very similar to LSO: a fraction of the lutetium atoms in the crystal are replaced with yttrium. LYSO properties are also very similar to those of LSO, the main difference being the lower density due to yttrium s lower weight. Gemini TF uses 28 flat modules made of mm 2 LYSO crystals, and covers an area of about cm 2 (18 cm axial coverage). A total of 420 PMTs are coupled to the 28 detector modules in an Anger logic architecture. Time resolution of the Gemini TF is 585 ps if measured with a low activity source. Recently, clinical results from a Siemens prototype TOF PET scanner have been presented [30]. The new TOF scanner has a time resolution of 550 ps. It comprises 192 independent detector blocks assembled in four rings around the patient on a cylindrical barrel configuration. The axial coverage is about 22 cm. Each block is made of LSO crystals ( mm 3 ), coupled to a set of four photomultiplier tubes. Additional valuable sources of information on TOF PET history and research can be found in some recent review papers available in the literature [17,31,32]. Introduction to the principles of TOF PET Principles of TOF PET Time-of-flight information is used in conventional (or non- TOF) positron emission tomography to determine if two detected photons are in time coincidence and therefore belong to the same positron annihilation event. Each detected photon is tagged with a detector position and detection time: if the detection time difference between two photons is smaller than a set coincidence window (traditionally 5e10 ns), the two events are considered physically correlated to the same annihilation event. Such measured difference in detection time is directly related to the actual photon time-of-flight difference, blurred by a measurement uncertainty named time resolution, which depends on several instrumental factors. Conventional PET reconstruction uses the time-of-flight information only to identify the line along which the annihilation occurred. It is unable, though, to determine which voxel along the line is the source of the two photons; therefore all the voxels along the line are given the same probability of emission. Analytical or iterative reconstruction algorithms are used to estimate the activity distribution most consistent with the measured projection data. TOF PET uses the time-of-flight difference to better locate the annihilation position of the emitted positron. The time-of-flight difference t is directly related to the distance x of the annihilation point from the center of the field of view (FOV) (x Z ct/2), along the line of response (LOR) identified by the two detectors in coincidence, as shown in Fig. 1. In Fig. 1, TOF A is proportional to the distance between the source and detector A, and TOF B is proportional to the distance between the source and detector B. The limitation in our ability to localize the annihilation point is mainly due to uncertainty in the measured time difference t, or time resolution Dt of the coincidence system. The time resolution is used in the reconstruction algorithm as a kernel for a localization probability function. The events are located along the LOR identified by the two detectors; its most probable position is set to the position corresponding to the measured TOF difference t. The FWHM of the probability function is the localization uncertainty Dx (FWHM) Z cdt/2. In Figs. 2e4, the difference between conventional and TOF reconstruction is illustrated with a simulation. A simulated uniform disk of activity with an internal hot spot is placed inside a ring of detectors (Fig. 2). A Maximum Likelihood Expectation Maximization (MLEM) iterative algorithm is used for image reconstruction for both TOF and non-tof. Iterative algorithms are based on repeated updates of an estimated image based on a comparison of

3 Time-of-flight PET 3 A second more general (common to iterative and analytical reconstruction) characteristic is that the additional information provided by TOF also results in a final image that is characterized by lower noise than the corresponding non-tof image, as is visible when comparing Figs. 3c and 4c. The TOF gain Figure 1 TOF PET scanner geometrical model. A photon pair is generated in the position indicated by the square pixel at x position from the center. Two detectors A and B are selected in the detector ring. TOF A is proportional to the distance between the source and detector A, TOF B is proportional to the distance between the source and detector B. the experimental projection data to the forward projection of the estimated image into projection space. In Fig. 3, the conventional non-tof approach is shown. A first estimate of the image is obtained by back projecting the events along the corresponding LOR. In Fig. 3a, one event detected by the central detectors at a 90 angle is evenly back projected into all pixels crossed by the LOR. In Fig 3b, all events detected at 0 and 90 are back projected. In Fig. 3c, the image obtained at the end of the first iteration of MLEM is shown. In Fig. 4, the corresponding TOF case is shown. Now detected events are back projected, assuming a probability distribution along the LOR that is not flat. A Gaussian probability kernel is used, centered at the position corresponding to the measured TOF, with a FWHM equal to a simulated time resolution. One can observe that the additional information provided by TOF allows for a much better localization of the original activity distribution even at the first estimate of the image (Fig. 4b). This is a first characteristic of TOF reconstruction: the iterative TOF reconstruction converges faster to the true image. Using the additional TOF information helps the algorithm to converge in a smaller number of iterations. Figure 2 A simulated uniform activity disk placed at the center of a PET scanner. A spot with higher activity is represented by a darker off-center pixel. Some photon pair tracks that reach the detector ring are traced. In fact, the TOF reconstruction reduces the noise propagation along the LOR during back projection of the data in the reconstruction. The reduction is related to the width of the kernel used, or to the time resolution of the system. This noise reduction was studied during the early years of TOF PET, mainly via modeling the noise propagation through analytical reconstruction processes in TOF and non- TOF reconstruction [2,7,10,33,34]. Conventional non-tof signal-to-noise ratio (SNR) has been modeled in a simple case by Strother using an analytical back-projection algorithm [35]. In a cylinder of diameter D, with uniform distribution of activity, one can estimate the SNR in an image element of size d starting from the data collected in the projection space: " # 1=2 SNRZconst,n 1=2 T 2 ð1þ ðt þ S þ RÞ where the T represents the total trues in the image, S scatter events, R the random coincidences, and n the number of volume elements influencing the noise in such image element. In the case of conventional non-tof PET, all volume elements that contain activity along the same LOR contribute to the same projection data. In reconstruction, each detected event is evenly back-projected in all image elements along the LOR, not only in the image element where it was originated. Therefore, all n elements contribute to the noise in each image element. In this case, n can be estimated as n conv Z D/d (Fig. 5a). If TOF is implemented, each event is TOF tagged. In reconstruction, each event is back-projected only in the position associated to such TOF information and into few volume elements adjacent to it, with a weight given by a TOF kernel or probability function of width Dx; Dx is the localization uncertainty, related to the time resolution Dt by the equation Dx Z cdt/2. In this case, n can be estimated as n TOF Z Dx/d (Fig. 5b). Using the appropriate values of n in Equation (1), one can obtain an estimate of the signal-to-noise gain introduced by TOF reconstruction: rffiffiffiffiffiffi D SNR TOF Z SNR non-tof ð2þ Dx Equation (2) is used as a measure of the TOF SNR gain in the literature. Other researchers estimated a slightly different value [10], or modified the formula for effect of the randoms [36], but Equation (2) is still a reasonable estimate of TOF gain in analytical reconstruction. Performing a similar analysis for iterative reconstruction is more complex because of the non-linearity of the reconstruction and the complication added by the arbitrary

4 4 M. Conti Figure 3 (a) One event detected by the central detectors at a 90 angle is evenly back projected into all pixels crossed by the LOR. (b) All events detected at 0 and 90 are back projected. (c) The image obtained at the end of the first MLEM iteration, using all projection data, is shown. choice of the iteration number. If iterative reconstruction is used, Equation (2) can only roughly estimate the SNR gain between a non-tof and a TOF image which reached similar contrast recovery but at a different iteration number, since TOF has a faster convergence. The quantity T 2 =ðt þ S þ RÞ in Equation (1) is commonly referred to as the noise equivalent count rate (NEC), which is a measure of the detected counts corrected for the noise contribution of scatter and random coincidences, or a measure of the effective sensitivity of the PET scanner. The NEC can be expressed as the square of the SNR; therefore the SNR gain can be seen as NEC gain or sensitivity gain. TOF reconstruction is equivalent to an amplification of the PET scanner sensitivity, as shown in Equation (3), derived from Equations (1) and (2): NEC TOF Z D Dx NEC non-tof It can be observed in Equation (3) that the NEC gain is directly proportional to the size of the patient and inversely proportional to the time resolution of the PET scanner. In other words, large patients will particularly benefit from TOF reconstruction. Equation (3) also indicates that improving the time resolution is the key for better performing TOF PET scanners. The factors influencing the time resolution will be discussed in the Section Time resolution below. In Table 1, the estimated TOF NEC gain is reported as a function of the time resolution, given a patient size ð3þ equivalent to a 40-cm diameter cylinder. It shows the tremendous potential of TOF reconstruction as a sensitivity amplifier. Time resolution The system time resolution is the key parameter that defines a TOF PET scanner. If only two detector elements A and B are considered, the time resolution can be measured by placing a point source between the detectors, and measuring the full width half maximum (FWHM) of the distribution of time of flight difference between detector A and B. The time resolution depends on several factors associated with different components of the detection process in each individual detector. In a more complex system using detector rings or panels, the system time resolution results from the average of the individual time resolution associated with each detector-to-detector pair, and might be also affected by variations in the detector quality and electronics, and by the time alignment procedure. In the present and past generations of TOF scanner, the detector is based on an inorganic crystal that converts the high energy photon into light photons, a light sensor and amplifier such as the photomultiplier tube (PMT), and the associated electronics for shaping, amplification, time and energy discrimination. We will focus on this present architecture and describe the factors that influence the time resolution. Figure 4 (a) One event detected by the central detectors at a 90 angle is back projected along the LOR according to a probability distribution with a width proportional to the time resolution. (b) All events detected at 0 and 90 are back projected. (c) The image obtained at the end of the first iteration of MLEM is shown.

5 Time-of-flight PET 5 smaller reflectivity decreases the light output of the crystal, with an indirect effect on time resolution. A long crystal implies a path length dispersion and consequently a transit time dispersion of light in the detector on the order of 50 ps/cm for a material with a refractive index of 1.7. Crystal arrays and light guides In most PET scanners, scintillator crystals are assembled in arrays and mounted on a light guide that is coupled with a group of PMTs that shares the light coming from the crystals. Reflective material is placed between the crystals to reduce crosstalk. Due to light sharing between several PMTs and the light guide design and optical properties, the time resolution of the crystal array is significantly worse than the time resolution of a single crystal [37,38]. Photomultipliers Figure 5 (a) In non-tof reconstruction, all volume elements n found in the object along the line of response contribute to the noise in each image element, and n conv Z D/d. (b) In TOF reconstruction, because of the better localization of each event along the LOR, only the volume elements n adjacent to the position identified by the measured TOF contribute to the local noise, and n TOF Z Dx/d. The time resolution Dt limits the number of elements contributing to the noise, since it determines the localization uncertainty Dx. Crystal material The intrinsic material characteristics are the first and often main limitations to the time resolution. In particular, it has been shown that the time resolution is controlled by rise time, decay time and absolute light output. If rise time is negligible compared to the decay time, only decay time and light output determine the intrinsic limits of the time resolution; in particular, faster decay time and higher light output reduce, i.e. improve, time resolution. Materials with high light output and fast decay time, for example LaBr 3 and LSO, are good candidates for TOF PET. The Section Scintillators for TOF presents a review and a discussion of these scintillators. Crystal shape and surface The crystal geometry and surface finish have a direct and indirect effect on the time resolution. A surface finish with Table 1 Time resolution, spatial uncertainty and estimated TOF NEC gain for a 40 cm diameter uniform cylinder. Time resolution (ns) Dx (cm) TOF NEC gain There are two groups of phenomena that originate in the PMTs: phenomena that affect the number of photoelectrons produced for each light photon, and phenomena that affect the dispersion of electron transit time in the PMT [39]. The number of photoelectrons produced for each light photon, or quantum efficiency (QE), is related to the photocathode sensitivity and the photoelectron collection efficiency of the first dynode. The dispersion of electron transit time is mainly related to the variation of electron transit time from the photocathode to the first dynode, which is due to the different positions of the emitted electrons in the photocathode (center or edge, for example) and to the different initial velocity (both direction and modulus) of the emitted electrons. Such dispersion can also be reduced by increasing the high voltage between the cathode and the first dynode. A TOF-grade PMT needs an optimization of the uniformity of the electric field and of the mechanical assembly in order to reduce transit path dispersions. Finally, a complete TOF PET system requires a large set of PMTs with consistent or uniform response in terms of time jitter and electron transit time. Electronics The signal from the PMT or similar light sensor is fed into the electronic chain for time triggering and stamping, which usually comprises a discriminator and a time to digital converter (TDC). Both constant fraction discriminators (CFD) and leading edge discriminators (LED) have been considered for TOF PET. Time jitter from electronics components, time slewing due to pick up of smaller amplitude signals and non optimal TDC design can contribute to degrade the time resolution of a PET scanner. While the electronics could be a major limitation for time resolution, it has been shown that proper design can reduce its contribution to a negligible fraction [28,38]. Scintillators for TOF If several factors contribute to the overall performance of a TOF PET scanner, the fundamental limits to the

6 6 M. Conti performance itself are rooted in the intrinsic properties of the scintillating material used as the photon detector. In this section we will overview some of the scintillators presently being investigated for their TOF capabilities. A good PET detector must have high density and high Z to increase the interaction probability of 511 kev photons and the probability of photoelectric interaction (over Compton and coherent scattering). These parameters have a direct effect on sensitivity and spatial resolution. In addition, a TOF detector must offer a good time resolution. If all the other contributors (Section Time resolution) to time resolution are minimized or optimized, the time resolution Dt decreases with shorter decay time t and with higher absolute light output N (number of light photons emitted per MeV of interacting high energy photon). More specifically, a theoretical model and experimental data are available that directly relate time resolution to the decay time and the inverse of the square root of the number of photoelectrons emitted by the PMT s photocathode [39e43]. This model assumes a rise time that is negligible compared to the decay time; it is an accurate model for materials such as LSO [39], and it is less than accurate if such approximation is not valid [44,45]. Even for materials in which the rise time is not negligible, this model provides at least a lower limit for the time resolution, but more complex modeling is needed for a more accurate description of the data. Most of the materials currently being studied and considered for TOF PET are Ce-doped. These have a fast decay time related to 5de4f transition of Ce 3þ ions [43,46]. Halidesor other compounds that contain heavy Lu or La are generally used, in order to reach high density and effective Z.Pr 3þ has similar characteristics as a dopant and has been recently considered for use instead of Ce 3þ. The emission light wavelength of all these materials ranges between 300 and 450 nm. A first group of scintillators is in the class of lutetium oxyorthosilicate, Ce-doped, such as Lu 2 SiO 5 (LSO) [19,47] and its modifications such as Lu x Y (2 x) SiO 5 (LYSO), where part of the lutetium atoms have been replaced by the lighter yttrium [48,49]. These scintillators are dense (up to 7.4 g/cm 3 ) and are characterized by a decay time of around 40 ns and by a fairly high light output (up to 40,000 photons/mev). A second group of scintillators is also in the class of lutetium compounds: lutetium aluminum perovskite LuAlO 3 (LuAP) and modifications such as Lu x Y (1 x) AlO 3 (LuYAP), which are all Ce-doped [50e53]. The growth of pure lutetium aluminum perovskite is a critical process that can terminate into formation of garnets. On the other end, a lutetium aluminum garnet, Lu 3 Al 5 O 12 (LuAG), Pr-doped, is by itself an interesting scintillator [54e58]. This class of materials can be as dense as 8.3 g/cm 3, with decay times around 20 ns, but light output is usually below 20,000 photons/mev. A third group is the lutetium or lanthanum halides, Cedoped, particularly LaCl 3 [59,60], LaBr 3 [44,61,62], and LuI 3 [63e66]. They are lighter materials (less than 6 g/cm 3 ), but have very high light output (up to 100,000 photons/mev) and decay times between 20 and 40 ns. They are hygroscopic. As historical reference, BaF 2 was one of the first scintillators to be used for TOF PET in the 1980s. It has an exceptionally short decay time (0.8 ns) but low density (4.9 g/cm 3 ) and very poor light output in the fast component (1800 photons/mev). The low light output worsens energy resolution and the ability to identify individual crystals in crystal arrays. The very short emission wavelength (220 nm) caused additional light collection problems. In Table 2 the main performance parameters of some materials considered for TOF PET are reported. A good candidate for TOF PET must have high density and photoelectric interaction probability for 511 kev photons [67], high light output and short decay time. It is worth mentioning that the decay time reported in Table 2 is the fast component of the decay time, if more than one component participates in the scintillation mechanism. The light yield is usually the total light yield integrated over a long enough time to include all components. Scintillators with one decay component like LSO release all light in the useful fast component. Two component scintillators such as LuAG or LuI 3 release only a fraction of the light in the fast component, for example about 50% in LuI 3 [65]. The fraction of the light emitted in the fast component is also reported in Table 2. A quantitative method to compare possible TOF-grade scintillators has been recently presented [58]: it defines a figure of merit that is computed using conventional efficiency parameters such as absorption coefficient and photopeak fraction, and TOF-relevant parameters such as decay time and absolute light yield. This and other figures of merit, based on quantitative intrinsic material characteristics, point in the right direction, but do not provide a unique and definitive answer to the question of which scintillator can offer the best performance in a TOF PET scanner. In general, as can be seen in Table 2, there is no such a thing as a best TOF PET scintillator. Sensible choices must be made to optimize the performance of the chosen material in terms of architecture, photomultiplier or other light sensor, and electronics. Also, other parameters not discussed here could determine the choice, such as fabrication complexity, hygroscopicity, cost and availability. For example, is the technology mature enough as to be able to deliver hundreds of thousands of crystals identical in terms of these key parameters, particularly light output? If not, large contribution to time resolution will simply come from components non-uniformity. Is the production cost competitive? Can other technical issues be properly addressed, hygroscopicity for example? In addition, one exceptionally good parameter cannot compensate for another extremely poor parameter: for example, BaF 2 has exceptionally fast light, but the light output is so poor that crystal positions cannot be properly separated if light is shared in a typical block structure. TOF reconstruction: questions and challenges TOF PET faces challenges not only in terms of instrumentation, but also in the field of image reconstruction algorithms. We will briefly discuss some of these challenges in this section. The first challenge is to make reconstruction time clinically viable, since TOF brings a non-negligible additional computation burden. The choice of algorithm (analytical or

7 Time-of-flight PET 7 Table 2 Properties of some TOF-grade scintillators: density, photoelectric absorption coefficient for 511 kev photons, decay time (fast component), total light output (all components), fraction of the light in the fast component Material Density (g/cm 3 ) Photoelectric absorption (cm 2 /g) Decay time (ns) Total light output (photons/mev) Fast component fraction (%) References LSO(Ce) [19,47] LYSO(Ce) (Lu 1.8,Y 0.2 ) [48,49] LuAP(Ce) [51] LuYAP(Ce) (Lu 0.7,Y 0.3 ) [52,53] LuAG(Pr) [57,58] LaBr 3 (Ce) [62,63] LaCl 3 (Ce) [59,60] LuI 3 (Ce) [66,67] BaF [13,17] The data were taken from the references listed in the last column. iterative) and the choice of data organization (sinogram or listmode) are only parts of this equation, together with other elements, such as computer architecture or software optimization. A second issue is the correctness of the reconstruction in terms of validating the individual components that must now properly include the TOF information; for example, the scatter correction. Finally, we must consider the clinical impact of TOF PET reconstruction in terms of better diagnosis or treatment of patients and where we expect to see the largest improvements. Even if it is too early to fully answer these questions, the first clinical studies show encouraging results and indicate some directions. Analytical or iterative reconstruction? Analytical reconstruction methods were the only reconstruction methods available at the beginning of TOF PET development. A variation of filtered back projection (FBP) with a time-of-flight one dimensional weight along the time-of-flight line was the natural extension of non TOF reconstruction [10]. This method, also known as confidence weighting (CW) time of flight reconstruction, performed better than non-tof and better than other analytical TOF methods, in terms of signal-to-noise [34]. The introduction of maximum likelihood iterative reconstruction [68] prompted an extension of this method to TOF PET [11,69]: a point spread function along the line of flight is applied in the projector, with a width corresponding to the system time resolution of the scanner. Iterative reconstruction methods are generally slower than analytical methods and in particular compared to FBP, but outperform FBP from the point of view of image noise [70]. In the early 1980s the authors admitted that computational speed constituted a problem, but they forecast that future generations of computers would close the gap and make iterative reconstruction viable [69]. In fact, today iterative reconstruction methods are the standard in clinical PET, and also appear to be the natural choice for TOF PET in both present and future clinical TOF scanners [71]. But there is still interest in developing analytical methods for TOF reconstruction, and a new adaptive CW kernel has been proposed for TOF analytical reconstruction, together with a method to improve the noise performance of FBP through a locally adaptive filter [72,73]. Fast analytical methods such as TOF-FBP might be viable solutions for some applications where speed and accurate quantification are preferred to low image noise. Sinogram or listmode reconstruction? Each pair of detector elements in coincidence in a PET scanner identifies a line-of-response (LOR), and a fully correct data organization should preserve the spatial information carried by each LOR. The first natural approach is a listmode reconstruction, in which data are stored as a list of detected events. Typically, an iterative reconstruction algorithm processes each event and correspondingly updates the image array. Reconstruction methods based on listmode organization of the data are usually slower, since back- and forwardprojection is independently executed for each event of the list and the reconstruction time depends on the length on the list, thus on the acquisition time. Recent publications [74e76] showed ways to efficiently incorporate new information such as TOF or local spatial resolution in listmode reconstruction. A second approach to PET data organization is to bin the acquired data into projection representation of the data or sinograms. Events are accumulated into a 3D sinogram array, where each element is identified by a radial position, a transaxial angle, and an axial position (with polar angle). The size of the sinogram array grows with the number of independent detector elements. The large axial coverage and fine spatial resolution of today s PET scanners requires 10 4 e10 5 detectors. Different methods, commonly called rebinning, can be used to reduce the size of the sinogram array. Rebinning techniques, by grouping LORs, reduce the number of transaxial angles (transaxial mashing) or polar angles (axial rebinning): some information is lost in this process and axial or transaxial blurring is possible and must be evaluated. TOF PET adds complexity to data organization and computation time to the reconstruction algorithm. If the reconstruction is sinogram based, TOF information adds a 4th dimension to the 3D sinogram representation,

8 8 M. Conti changing data storage and dynamic memory requirements. If the reconstruction is listmode based, more computation time is added to the already slower method. However, TOF information can be used to actually improve present axial rebinning methods such as single slice rebinning (SSRB) and Fourier rebinning (FORE), and allows reduction of the transaxial angular sampling with minimal resolution degradation [77,78]. Emission data corrections for TOF PET In non-tof PET the coincidence emission data acquired in a PET scanner usually undergo a pre-reconstruction phase in which several corrections are applied in order to recover the true emission data. Random coincidences are estimated and subtracted; normalization is applied to equalize the response of single detectors or line-of-response; the scatter events contribution is estimated and subtracted; and an attenuation correction is applied to account for the photons self-absorption in the patient body or any other medium present in the scanner s FOV. The introduction of the TOF dimension requires a revision of the whole process and an accurate analysis in order to identify how the different elements change in the presence of TOF. It appears the randoms do not have a TOF structure and there is no reason to expect the attenuation to have TOF dependence. The current approach is therefore to apply conventional non-tof corrections to the new TOF data. Similarly, a non-tof normalization is usually applied to TOF data, a reasonable approximation that probably requires some further study. On the contrary, scatter correction is clearly identified as the component that definitely has a TOF structure and requires an appropriate TOF computation. Recent papers provided a TOF extension of conventional scatter estimate [79e81]. While work is necessary to refine and confirm the present form of TOF corrections, it appears that TOF reconstruction is much less sensitive to errors and improper approximations. The redundant information present in TOF data naturally corrects the data inconsistencies during the iterative reconstruction. It has been observed that TOF reconstruction reduces artifacts due to incorrect normalization, approximated scatter correction, truncated CT attenuation map, and metal and contrast agent artifacts in the PET image. Can we measure an image quality improvement due to TOF reconstruction? A TOF sensitivity gain was estimated and measured even in the early 1980s, using phantoms and analytical reconstruction. Making a quantitative and solid assessment on clinical patient images reconstructed with today s iterative methods is a more complex problem. First, clinical iterative reconstruction is usually stopped before convergence to a true solution. Second, the true solution might not be the same if different reconstruction methods are used, particularly in the presence of a high noise level. Third, this is complicated by a faster convergence of TOF reconstruction relative to non-tof, which makes it improper to compare TOF and non-tof images obtained with the same iteration number. Finally, the true solution in a patient image is in fact unknown. Nevertheless, attempts to quantify the TOF sensitivity gain in today s iterative reconstruction have been made and some results are available in the literature. The SNR improvement was measured in a simulation study of a LaBr 3 TOF scanner at the University of Pennsylvania [82]. The images were reconstructed using listmode iterative reconstruction and evaluated using a Non- PreWhitening Matched Filter (NPWMF) SNR observer. Time resolutions from 1000 ps to 300 ps were simulated. SNR in TOF images was consistently better than in non TOF images, and the improvement was larger if time resolution was better, i.e. smaller. Also, contrast recovery was observed to be better in TOF images, and the convergence of the iterative algorithm was faster if TOF reconstruction was used. A study published in 2005 [24] plotted image noise versus cold spot residuals for different counts statistics in a simulation study of a 40 cm diameter water phantom, for a PET scanner with 1.2 ns time resolution. The study showed that the TOF data lay on a curve that was consistently better (for any iteration number) than the non-tof data acquired with twice as many statistics. Also, a criterion was proposed for comparing TOF and non-tof images in iterative reconstruction: a TOF and a non-tof image can be selected with matched contrast recovery, even if produced with different iterations. This approach compensates for the faster convergence of TOF reconstruction. It was demonstrated that at matched contrast the SNR was higher in TOF reconstruction. A recent paper [83] presented the first quantitative analysis of TOF gain on oncology patient studies, for a 600 ps time resolution TOF scanner. In this paper, TOF and non-tof images of the same noise level were matched, and TOF contrast gain for small lesions (1e 2 cm size) was demonstrated to be correlated with the patient mass. Also, SNR versus noise level curves were presented for different patients and different small lesions, and TOF consistently outperformed non-tof. Even though the authors admitted that it was difficult to quantify a single sensitivity gain factor for TOF, the results showed a significant impact of TOF reconstruction on image quality. A study on a large set of oncology patients was recently presented [30], confirming that human observers prefer TOF images from the point of view of contrast recovery, noise level and clarity of detail in anatomical structure. In this phase, still more clinical studies are needed to determine how this improved image quality affects clinical diagnosis and subsequent patient therapy. Who benefits from TOF PET? In the literature, recent and past, it has been pointed out that TOF images, compared to equivalent non-tof images, in general show a lower level of noise and better resolution. But there are particular cases in which the improvement in image quality carried by TOF information is more significant. First, there is a correlation between patient size and TOF gain in signal-to-noise, and this correlation has been

9 Time-of-flight PET 9 demonstrated in recent papers [30,83]. The TOF gain in large patients, due to the better localization of coincidence events along the line-of-response as discussed in the sub- Section The TOF gain, may be even more significant when limited acquisition time and high attenuation reduce the total counts in the image. Second, another set of patients might find TOF PET beneficial; those whose PET scans have a low number of counts because of the chosen PET protocol. For example, in dynamic studies short frames are acquired with very low statistics. It has been observed that, using iterative reconstruction algorithms, scans with different levels of counts do not necessarily converge to the same image. This is due to a bias introduced by high noise level. TOF reconstruction, operating as a statistics booster, might reduce the initial image bias due to limited statistics and help the reconstruction to converge to a truer and less noisy answer, regardless of the patient size. Third, any PET application where there is a partial coverage of the patient can take advantage of TOF reconstruction, most commonly in limited transaxial angle coverage. Incomplete coverage of all angles creates image artifacts in conventional PET reconstruction. The additional TOF information can recover some of the missing information and reduce or eliminate the artifacts, as has been shown in two simulation studies: a dedicated breast TOF PET scanner and an inbeam PET system for hadron-therapy treatment facilities [84,85]. A fourth opportunity, related to the better performance of TOF in low statistics cases, is for high resolution studies. The limited acquisition time and noise level in the images favor a preference for a low resolution and low-passfiltered image, in particular for oncology imaging. The full spatial resolution of today s PET scanners is usually not sufficiently exploited. The lower noise and amplified sensitivity of TOF reconstruction could favor a better use of the full resolution potentiality of PET scanners, if computation speed allows it. Finally, the image quality might be kept at the same level as non-tof PET, but the TOF gain can be used to reduce acquisition time, improving the hospital workflow and patient comfort, and reducing costs. Acknowledgments I would like to thank Lars Eriksson, Mike Casey, Bernard Bendriem and David Townsend for providing experimental data, valuable references and deeper understanding through useful discussions. References [1] Budinger TF. Instrumentation trends in nuclear medicine. Semin Nucl Med 1977;7:285e97. [2] Budinger TF. Time-of-flight positron emission tomography: status relative to conventional PET. J Nucl Med 1983;24:73e8. [3] Mullani NA, Markham J, Ter-Pogossian MM. Feasibility of timeof-flight reconstruction in positron emission tomography. J Nucl Med 1980;21:1095e7. [4] Gariod R, Allemand R, Carmoreche E, et al. The LETI Positron tomograph architecture and time of flight improvements. Proceeding of the Workshop on Time-of-Flight tomography. IEEE Publication: Washington University; p [5] Bendriem B, Soussaline F, Campagnolo R, Verrey B, Wajnberg P, Syrota A. A technique for the correction of scattered radiation in a PET system using Time-of-Flight information. J. Comput Assist Tomogr 1986;10(2):287e95. [6] Ter-Pogossian MM, Ficke DC, Yamamoto M, et al. Super PETT I: a positron emission tomograph utilizing photon time-of-flight information. Trans Med Imaging 1982;1:179e86. [7] Yamamoto M, Ficke DC, Ter-Pogossian MM. Experimental assessment of the gain achieved by the utilization of time-offlight information in a positron emission tomograph (Super PETT I). IEEE Trans Med Imaging 1982;1:187e92. [8] Wong WH, Mullani NA, Philippe EA, Hartz RK, Gould KL. Image improvement and design optimization of the Time-of-Flight PET. J Nucl Med 1983;24:52e60. [9] Wong WH, Mullani NA, Philippe EA, Hartz RK, Bristow D, Yerian K, et al. Performance characteristics of the University of Texas TOFPET-I PET camera. J Nucl Med 1984;25:46e7. [10] Tomitani T. Image reconstruction and noise evaluation in photon time-of-flight assisted positron emission tomography. IEEE Trans Nucl Sci 1981;28:4582e8. [11] Yamamoto M, Nohara N, Tanaka E, Tomitani T, Murayama H, Sato N, et al. Time-of-flight positron imaging and the resolution improvement by an iterative method. IEEE Trans Nucl Sci 1989;36:998e1002. [12] Lewellen TK, Bice AN, Harrison RL, Pencke MD, Link JM. Performance measurements of the SP3000/UW time-of-flight positron emission tomograph. IEEE Trans Nucl Sci 1988;35(1):665e9. [13] Allemand R, Gresset C, Vacher J. Potential advantages of a cesium fluoride scintillator for a time of flight positron camera. J Nucl Med 1980;21:153e5. [14] Laval M, Moszynski M, Allemand R, Cormoreche E, Guinet P, Odru R, et al. Barium fluoridedinorganic scintillator for subnanosecond timing. Nucl Inst Methods 1983;206:169e76. [15] Ishii K, Orihara H, Matsuzawa T, Binkley DM, Nutt R. High resolution time-of-flight positron emission tomograph. Rev Sci Instr. 1990;61(12):3755e62. [16] Soussaline S, Comar D, Allemand R, Campagnolo R, Laval M, Vacher J, et al. New developments in positron emission tomography instrumentation using the time-of-flight information. In: Gretz T, editor. The metabolism of the human brain studied with positron emission tomography. New York: Raven Press; p. 1e11. [17] Lewellen TK. Time-of-flight PET. Semin Nucl Med 1998;28(3): 268e75. [18] Weber MJ, Monchamp RRlimits. Luminescence of Bi4Ge3O12dspectral and decay properties. J Appl Phys 1973; 44:5495e9. [19] Melcher CL, Schweitzer JS. Cerium-doped lutetium orthosilicate: a fast, efficient new scintillator. IEEE Trans Nucl Sci 1992;39:502e5. [20] Wienhard K, Schmand M, Casey ME, Baker K, Bao J, Eriksson L, et al. The ECAT HRRT: Performance and first clinical application of the new high resolution research tomograph. IEEE Trans Nucl Sci 2002;49(1):104e10. [21] Herzog H, Tellmann L, Hocke C, Pietrzyk U, Casey ME, Kuwert T. NEMA NU guided performance evaluation of four Siemens ECAT PET scanners. IEEE Trans Nucl Sci 2004; 51(5):2662e9. [22] Moses WW, Derenzo SE. Prospects for time-of-flight PET using LSO scintillator. IEEE Trans Nucl Sci 1999;46(3):783e8. [23] Conti M, Bendriem B, Casey M, Chen M, Kehren F, Michel C, et al. Implementation of time-of-flight on CPS HiRez PET scanner IEEE Nuclear Science Symposium Conference Record, Rome, Italy, October 16e22, 2004.

10 10 M. Conti [24] Conti M, Bendriem B, Casey M, Chen M, Kehren F, Michel C, et al. First experimental results of Time-Of-Flight reconstruction on an LSO PET scanner. Phys Med Biol 2005;50:4507e26. [25] van Loef E, Dorenbos P, van Eijk CWE, Kramer KW, Gudel HU. High-energy-resolution scintillator: Ce3þ activated LaBr3. Appl Phys Lett 2005; e1573. [26] Surti S, Karp JS, Muehllehner G, Raby PS. Investigation of lanthanum scintillators for 3-D PET. IEEE Trans Nucl Sci 2003; 50(3):348e54. [27] Karp JS, Kuhn A, Perkins AE, Surti S, Werner ME, Daube- Witherspoon ME, et al. Characterization of a time-of-flight PET scanner based on lanthanum bromide IEEE Nuclear Science Symposium Conference Record, Puerto Rico, October 23e29, [28] Kyba CCM, Wiener RI, Newcomer FM, Van Berg R, Dressnandt N, Karp JS. Timing measurements from a TOF-PET scanner using local PMT triggering. IEEE 2007 Nuclear Science Symposium Conference Record, Honolulu, Hawaii, October 27e November 3, [29] Surti S, Kuhn A, Werner ME, Perkins AE, Kolthammer J, Karp JS. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med 2007;48(3):471e80. [30] Conti M, Townsend DW, Casey ME, Lois C, Jacoby BW, Long MJ, et al. Assessment of the clinical potential of a time-of-flight PET/CT scanner SNM Annual Meeting Proceedings, New Orleans, Louisiana, June 14e18, [31] Moses WW. Time of flight in PET revisited. IEEE Trans Nucl Sci 2003;50(5):1325e30. [32] Muehllehner G, Karp JS. Positron emission tomography. Phys Med Biol 2006;51:R117e37. [33] Lewellen TK, Harrison RL, Bice AN. An experimental evaluation of the effect of time-of-flight information in image reconstruction for the Scanditronix/PETT electronics SP-3000 positron emission tomograph - preliminary results. IEEE Trans Nucl Sci 1989;36:1095e9. [34] Snyder DL, Thomas LJ, Ter-Pogossian MM. A mathematical model for positron emission tomography systems having time-of-flight measurements. IEEE Trans Nucl Sci 1981;28(3):3575e83. [35] Strother SC, Casey ME, Hoffman EJ. Measuring PET scanner sensitivity: relating count rates to image signal-to-noise ratios using noise equivalent counts. IEEE Trans Nucl Sci 1990;37(2): 783e8. [36] Conti M. Effect of random reduction on signal-to-noise-ratio in TOF PET. IEEE Trans Nucl Sci 2006;53(3):1188e93. [37] Aykac M, Bauer F, Williams CW, Loope M, Schmand M. Timing performance of Hi-Rez detector for time-of-flight (TOF) PET. IEEE Trans Nucl Sci 2006;53(3):1184e9. [38] Moses WW, Ullisch M. Factors influencing timing resolution in a commercial LSO PET camera. IEEE Trans Nucl Sci 2006;53(1): 78e85. [39] Szczesniak T, Moszynski M, Swiderski L, Nassalski A, Lavoute P, Kapusta M. Fast photomultipliers for TOF PET. IEEE 2007 Nuclear Science Symposium Conference Record, Honolulu, Hawaii, October 27eNovember 3, [40] Hyman LG. Time resolution of photomultiplier systems. Rev Sci Instr 1965;36(3):193e6. [41] Moszynski M, Bengtson B. Status of timing with plastic scintillation detectors. Nucl Instr Methods 1979;158:1e31. [42] Szczesniak T, Moszynski M, Nassalski A, Lavoute P, Dehaine AG. A further study of timing with LSO on XP20D0 for TOF PET. IEEE Trans Nucl Sci 2007;54(5):1464e73. [43] Nassalski A, Kapusta M, Batsch T, Wolski D, Moeckel D, Enghardt W, et al. Comparative study of scintillators for PET/CT detectors. IEEE Trans Nucl Sci 2007;54(1):3e10. [44] Glodo J, Moses WW, Higgins WM, van Loef EV, Wong P, Derenzo SE, et al. Effects of Ce Concentration on scintillation properties of LaBr 3 :Ce. IEEE Trans Nucl Sci 2005;52(5):1805e8. [45] Shao Y. A new timing model for calculating the intrinsic timing resolution of a scintillator detector. Phys Med Biol 2007;52: 1103e17. [46] Dorenbos P. Light output and energy resolution of Ce 3þ -doped scintillators. Nucl Instr Methods A 2002;486:208e13. [47] Kapusta M, Szupryczynski P, Melcher CL, Moszynski M, Balcerzyk M, Carey AA, et al. Non-proportionality and thermoluminescence of LSO:Ce. IEEE Trans Nucl Sci 2005;52(4): 1098e104. [48] Prelude420 Data Sheet, Saint-Gobain Crystals; [49] Pidol L, Kahn-Harari A, Viana B, Virey E, Ferrand B, Dorenbos P, et al. High efficiency of lutetium silicate scintillators, Ce-doped LPS, and LYSO crystals. IEEE Trans Nucl Sci 2004;51(3):1084e7. [50] Lecoq P, Korzhik M. New inorganic scintillation materials development for medical imaging. IEEE Trans Nucl Sci 2002; 49(4):1651e4. [51] Moszynski M, Wolski D, Ludziejewski T, Kapusta M, Lempicki A, Brecher C, et al. Properties of the new LuAP:Ce scintillator. Nucl Instr Methods A 1997;385:123e31. [52] Weber S, Christ D, Kurzeja M, Engels R, Kemmerling G, Halling H. Comparison of LuYAP, LSO, BGO as scintillators for high resolution PET detectors. IEEE Trans Nucl Sci 2003;50(5):1370e2. [53] Korzhik M, Federov A, Annenkov A, Borissevitch A, Dossovitski A, Missevitch O, et al. Development of scintillator materials for PET scanners. Nucl Instr Methods A 2007;571: 122e5. [54] Nikl M, Ogino H, Krasnikov A, Beitleroval A, Yoshikawa A, Fukuda T. Photo- and radioluminescence of Pr-doped Lu3Al5O12 single crystal. Phys Stat Sol A 2005;202(1):R4e6. [55] Ogino H, Yoshikawa A, Nikl M, Pejchalm J, Fukuda T. Growth and luminescence properties of Pr-doped Lu3(Ga,Al)5O12 single crystals. Jpn J Appl Phys 2007;46:3514e7. [56] Kamada K, Tsutsumi K, Usuki Y, Yanagida T, Sato M, Ogino H, et al. Scintillation properties of 2-inch-diameter Pr:Lu3Al5O12 single crystal. IEEE 2007 Nuclear Science Symposium Conference Record, Hunolulu, Hawaii, October 27eNovember 3, [57] Pr:LuAG data sheet, Furukawa Co. Ltd., furukawakk.co.jp, [58] Conti M, Eriksson L, Rothfuss H, Melcher C. Comparing fast scintillators with TOF PET potentiality Symposium on Radiation Measurements and Applications, Berkeley, California, USA, June 2e5, 2008, to be published in IEEE Trans Nucl Sci, [59] van Loef EVD, Dorenbos P, van Eijk CWE, Krämer K, Güdel HU. Scintillation properties of LaCl3:Ce3þ crystals: fast, efficient, and high-energy resolution scintillators. IEEE Trans Nucl Sci 2001;48(3):341e5. [60] Brillance 350 Data Sheet, Saint-Gobain Crystals, detectors.saint-gobain.com/, [61] Shah KS, Glodo J, Klugerman M, Moses WW, Derenzo SE, Weber MJ. LaBr3:Ce scintillators for gamma-ray spectroscopy. IEEE Trans Nucl Sci 2003;50(6):2410e3. [62] Brillance 380 Data Sheet, Saint-Gobain Crystals, detectors.saint-gobain.com/, [63] Shah KS, Glodo J, Klugerman M, Higgins W, Gupta T, Wong P, et al. LuI3:Ceda new scintillator for gamma ray spectroscopy. IEEE Trans Nucl Sci 2004;51(5):2302e5. [64] Glodo J, Shah KS, Klugerman M, Wong P, Higgins W, Dorenbos P. Scintillation properties of LuI3:Ce. Nucl Instr Methods A 2005;537:279e81. [65] Birowosuto MD, Dorenbos P, van Eijk CWE, Krämer KW, Güdel HU. Scintillation properties of LuI3:Ce3þ-high light yield scintillators. IEEE Trans Nucl Sci 2005;52(4):1114e8. [66] Birowosuto MD, Dorenbos P, de Haas JTM, van Eijk CWE, Krämer KW, Güdel HU. Optical properties and luminescence quenching of LuI3:Ce3þ. J Luminesc 2006;118:308e16.

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide 2005 IEEE Nuclear Science Symposium Conference Record M04-8 Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide J. S. Karp, Senior Member, IEEE, A. Kuhn, Member, IEEE, A. E. Perkins,

More information

Recent advances and future advances in time-of-flight PET

Recent advances and future advances in time-of-flight PET ARTICLE IN PRESS Nuclear Instruments and Methods in Physics Research A 580 (2007) 919 924 www.elsevier.com/locate/nima Recent advances and future advances in time-of-flight PET William W. Moses Lawrence

More information

Fast Timing and TOF in PET Medical Imaging

Fast Timing and TOF in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

in PET Medical Imaging

in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

NIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2010 February 9.

NIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2010 February 9. NIH Public Access Author Manuscript Published in final edited form as: J Nucl Med. 2010 February ; 51(2): 237. doi:10.2967/jnumed.109.068098. An Assessment of the Impact of Incorporating Time-of-Flight

More information

Time-of-Flight Technology

Time-of-Flight Technology Medical Review Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems INTRODUCTION Improving the care for every patient while providing a high standard care to

More information

An educational tool for demonstrating the TOF-PET technique

An educational tool for demonstrating the TOF-PET technique Nuclear Instruments and Methods in Physics Research A 471 (2001) 200 204 An educational tool for demonstrating the TOF-PET technique T.Bȧack a, *, J. Cederkȧall a, B. Cederwall a, A. Johnson a, A. Kerek

More information

Introduction to Positron Emission Tomography

Introduction to Positron Emission Tomography Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator

More information

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 5 SPECT, PET Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar Doctoral Thesis Time-of-Flight Time-of-Flight Positron Positron Emission Emission Tomography Tomography Using Using Cherenkov Cherenkov Radiation Radiation Rok Dolenec Supervisor: Prof. Dr. Samo Korpar

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial 8//0 AAPM0 Scientific Symposium: Emerging and New Generation PET: Instrumentation, Technology, Characteristics and Clinical Practice Aug Wednesday 0:4am :pm Solid State Digital Photon Counting PET/CT Instrumentation

More information

Extremely Fast Detector for 511 kev Gamma

Extremely Fast Detector for 511 kev Gamma Extremely Fast Detector for 511 kev Gamma V. Sharyy, D. Yvon, G. Tauzin, E.Delagnes, Ph. Abbon, J P. Bard, M. Kebbiri, M. Alokhina, C. Canot IRFU, CEA D. Breton, J. Maalmi LAL,IN2P3 Journée 2015 du Labex

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.

More information

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO Abstract C.S. Levin, Y-C Tai, E.J. Hoffman, M. Dahlbom, T.H. Farquhar UCLA School of Medicine Division

More information

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and

More information

Improvement of contrast using reconstruction of 3D Image by PET /CT combination system

Improvement of contrast using reconstruction of 3D Image by PET /CT combination system Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(1):285-290 ISSN: 0976-8610 CODEN (USA): AASRFC Improvement of contrast using reconstruction of 3D Image

More information

Positron Emission Tomography

Positron Emission Tomography Physics 656 Seminar on Physical Fundamentals of Medical Imaging Positron Emission Tomography Ahmed Qamesh Outline What is PET? PET mechanism Radionuclide and its synthesis Detection concept and Development

More information

486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007

486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007 486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007 Count-Rate Dependent Component-Based Normalization for the HRRT Mario Rodriguez, Jeih-San Liow, Member, IEEE, Shanthalaxmi Thada, Merence

More information

Nuclear Instruments and Methods in Physics Research A

Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 641 (211) 121 127 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

Performance Evaluation of radionuclide imaging systems

Performance Evaluation of radionuclide imaging systems Performance Evaluation of radionuclide imaging systems Nicolas A. Karakatsanis STIR Users meeting IEEE Nuclear Science Symposium and Medical Imaging Conference 2009 Orlando, FL, USA Geant4 Application

More information

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation 26 IEEE Nuclear Science Symposium Conference Record M6-349 Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation Habib Zaidi, Senior Member, IEEE and Mohammad

More information

Effects of system geometry and other physical factors on photon sensitivity of high-resolution

Effects of system geometry and other physical factors on photon sensitivity of high-resolution Home Search Collections Journals About Contact us My IOPscience Effects of system geometry and other physical factors on photon sensitivity of high-resolution positron emission tomography This article

More information

A Further Study on Large Size LSO and LYSO Crystal Samples

A Further Study on Large Size LSO and LYSO Crystal Samples October 25, 2005 IEEE NSS, N12-6, Puerto-Rico 1 A Further Study on Large Size LSO and LYSO Crystal Samples Jianming Chen, Liyuan Zhang, Ren-yuan Zhu California Institute of Technology BGO, LSO & LYSO Samples

More information

Modeling and Incorporation of System Response Functions in 3D Whole Body PET

Modeling and Incorporation of System Response Functions in 3D Whole Body PET Modeling and Incorporation of System Response Functions in 3D Whole Body PET Adam M. Alessio, Member IEEE, Paul E. Kinahan, Senior Member IEEE, and Thomas K. Lewellen, Senior Member IEEE University of

More information

Medical Imaging BMEN Spring 2016

Medical Imaging BMEN Spring 2016 Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,

More information

Constructing System Matrices for SPECT Simulations and Reconstructions

Constructing System Matrices for SPECT Simulations and Reconstructions Constructing System Matrices for SPECT Simulations and Reconstructions Nirantha Balagopal April 28th, 2017 M.S. Report The University of Arizona College of Optical Sciences 1 Acknowledgement I would like

More information

2005 IEEE Nuclear Science Symposium Conference Record M Influence of Depth of Interaction on Spatial Resolution and Image Quality for the HRRT

2005 IEEE Nuclear Science Symposium Conference Record M Influence of Depth of Interaction on Spatial Resolution and Image Quality for the HRRT 2005 IEEE Nuclear Science Symposium Conference Record M03-271 Influence of Depth of Interaction on Spatial Resolution and Image Quality for the HRRT Stéphan Blinder, Marie-Laure Camborde, Ken R. Buckley,

More information

Future Prospects of Time-of-Flight PET: Scintillator, Detector and Depth-of- interaction (DOI) Technologies. TOF scanners. Clinical benefit of TOF PET

Future Prospects of Time-of-Flight PET: Scintillator, Detector and Depth-of- interaction (DOI) Technologies. TOF scanners. Clinical benefit of TOF PET Future Prospects of Time-of-Flight PET: Scintillator, Detector and Depth-of- interaction (DOI) Technologies Suleman Surti Physics & Instrumentation Group Department of Radiology University of Pennsylvania

More information

2007 IEEE Nuclear Science Symposium Conference Record M26-296

2007 IEEE Nuclear Science Symposium Conference Record M26-296 2007 IEEE Nuclear cience ymposium Conference Record M26-296 A Global and a egmented Plane catter Calibration: Improving the Quantitative Accuracy of Frames with High Random Fraction and/or Low Number of

More information

Introduction to Emission Tomography

Introduction to Emission Tomography Introduction to Emission Tomography Gamma Camera Planar Imaging Robert Miyaoka, PhD University of Washington Department of Radiology rmiyaoka@u.washington.edu Gamma Camera: - collimator - detector (crystal

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy

Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy Physics Department, Faculty of Applied Science,Umm Al-Qura

More information

218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997

218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997 218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997 Compton Scatter and X-ray Crosstalk and the Use of Very Thin Intercrystal Septa in High-Resolution PET Detectors Craig S. Levin, Member,

More information

HHS Public Access Author manuscript Phys Med Biol. Author manuscript; available in PMC 2016 February 19.

HHS Public Access Author manuscript Phys Med Biol. Author manuscript; available in PMC 2016 February 19. Monte Carlo calculations of PET coincidence timing: single and double-ended readout Stephen E Derenzo, Woon-Seng Choong, and William W Moses Life Sciences Division, Lawrence Berkeley National Laboratory,

More information

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce

More information

Performance Evaluation of the Philips Gemini PET/CT System

Performance Evaluation of the Philips Gemini PET/CT System Performance Evaluation of the Philips Gemini PET/CT System Rebecca Gregory, Mike Partridge, Maggie A. Flower Joint Department of Physics, Institute of Cancer Research, Royal Marsden HS Foundation Trust,

More information

3D-OSEM Transition Matrix for High Resolution PET Imaging with Modeling of the Gamma-Event Detection

3D-OSEM Transition Matrix for High Resolution PET Imaging with Modeling of the Gamma-Event Detection 3D-OSEM Transition Matrix for High Resolution PET Imaging with Modeling of the Gamma-Event Detection Juan E. Ortuño, George Kontaxakis, Member, IEEE, Pedro Guerra, Student Member, IEEE, and Andrés Santos,

More information

The great interest shown toward PET instrumentation is

The great interest shown toward PET instrumentation is CONTINUING EDUCATION PET Instrumentation and Reconstruction Algorithms in Whole-Body Applications* Gabriele Tarantola, BE; Felicia Zito, MSc; and Paolo Gerundini, MD Department of Nuclear Medicine, Ospedale

More information

Detector simulations for in-beam PET with FLUKA. Francesco Pennazio Università di Torino and INFN, TORINO

Detector simulations for in-beam PET with FLUKA. Francesco Pennazio Università di Torino and INFN, TORINO Detector simulations for in-beam PET with FLUKA Francesco Pennazio Università di Torino and INFN, TORINO francesco.pennazio@unito.it Outline Why MC simulations in HadronTherapy monitoring? The role of

More information

NIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2009 December 1.

NIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2009 December 1. NIH Public Access Author Manuscript Published in final edited form as: J Nucl Med. 2009 August ; 50(8): 1315 1323. doi:10.2967/jnumed.109.063016. Impact of Time-of-Flight on PET Tumor Detection Dan J.

More information

The PET components in the most recent generation

The PET components in the most recent generation Impact of Time-of-Flight on PET Tumor Detection Dan J. Kadrmas 1, Michael E. Casey 2, Maurizio Conti 2, Bjoern W. Jakoby 2 4, Cristina Lois 5, and David W. Townsend 3 1 Department of Radiology, University

More information

Continuation Format Page

Continuation Format Page C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar

More information

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 3, JUNE 2003 405 Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging Edward J. Soares, Kevin W. Germino,

More information

THE general task in emission computed tomography (CT)

THE general task in emission computed tomography (CT) 228 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 17, NO. 2, APRIL 1998 List-Mode Likelihood: EM Algorithm and Image Quality Estimation Demonstrated on 2-D PET Lucas Parra* and Harrison H. Barrett, Member,

More information

Evaluation of Centrally Located Sources in. Coincidence Timing Calibration for Time-of-Flight PET

Evaluation of Centrally Located Sources in. Coincidence Timing Calibration for Time-of-Flight PET Evaluation of Centrally Located Sources in Coincidence Timing Calibration for Time-of-Flight PET by Richard Ryan Wargo Graduate Program in Medical Physics Duke University Date: Approved: Timothy G. Turkington,

More information

Simultaneous measurement of noise and spatial resolution in PET phantom images

Simultaneous measurement of noise and spatial resolution in PET phantom images IOP PUBLISHING Phys. Med. Biol. 55 (2010) 1069 1081 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/55/4/011 Simultaneous measurement of noise and spatial resolution in PET phantom images Martin

More information

Large Plastic Scintillation Detectors for the Nuclear Materials Identification System

Large Plastic Scintillation Detectors for the Nuclear Materials Identification System Large Plastic Scintillation Detectors for the Nuclear Materials Identification System J.S. Neal, J.T. Mihalczo, M. T. Hiatt, J. D. Edwards Oak Ridge National Laboratory P. O. Box 2008, Oak Ridge, Tennessee

More information

Improving Positron Emission Tomography Imaging with Machine Learning David Fan-Chung Hsu CS 229 Fall

Improving Positron Emission Tomography Imaging with Machine Learning David Fan-Chung Hsu CS 229 Fall Improving Positron Emission Tomography Imaging with Machine Learning David Fan-Chung Hsu (fcdh@stanford.edu), CS 229 Fall 2014-15 1. Introduction and Motivation High- resolution Positron Emission Tomography

More information

Pragmatic Fully-3D Image Reconstruction for the MiCES Mouse Imaging PET Scanner

Pragmatic Fully-3D Image Reconstruction for the MiCES Mouse Imaging PET Scanner Pragmatic Fully-3D Image Reconstruction for the MiCES Mouse Imaging PET Scanner Kisung Lee 1, Paul E. Kinahan 1, Jeffrey A. Fessler 2, Robert S. Miyaoka 1, Marie Janes 1, and Tom K. Lewellen 1 1 Department

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Nuclear Instruments and Methods in Physics Research A

Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 621 (2010) 590 594 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,

More information

Attenuation map reconstruction from TOF PET data

Attenuation map reconstruction from TOF PET data Attenuation map reconstruction from TOF PET data Qingsong Yang, Wenxiang Cong, Ge Wang* Department of Biomedical Engineering, Rensselaer Polytechnic Institute, Troy, NY 80, USA *Ge Wang (ge-wang@ieee.org)

More information

Effect of normalization method on image uniformity and binding potential estimates on micropet

Effect of normalization method on image uniformity and binding potential estimates on micropet Effect of normalization method on image uniformity and binding potential estimates on micropet Marie-Laure Camborde, Arman Rhamim, Danny F. Newport, Stefan Siegel, Ken R. Buckley, Eric Vandervoort, Thomas

More information

If it matters to you, it matters to us

If it matters to you, it matters to us If it matters to you, it matters to us Philips clinical innovations in nuclear medicine Innovation with insight We understand that clinical innovations are only as valuable as the day-to-day difference

More information

CHAPTER 11 NUCLEAR MEDICINE IMAGING DEVICES

CHAPTER 11 NUCLEAR MEDICINE IMAGING DEVICES CHAPTER 11 M.A. LODGE, E.C. FREY Russell H. Morgan Department of Radiology and Radiological Sciences, Johns Hopkins University, Baltimore, Maryland, United States of America 11.1. INTRODUCTION Imaging

More information

AX-PET : A novel PET concept with G-APD readout

AX-PET : A novel PET concept with G-APD readout AX-PET : A novel PET concept with G-APD readout Matthieu Heller CERN - PH/DT Marie Curie network MC-PAD Matthieu.heller@cern.ch On behalf of the AX-PET collaboration https://twiki.cern.ch/twiki/bin/view/axialpet

More information

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Sensors & Transducers 2013 by IFSA http://www.sensorsportal.com Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Shinji Ohyama, Masato Mukouyama Graduate School

More information

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors MS116 Characterization of light production, propagation and collection for both organic and inorganic scintillators D10.2 R&D on new and existing scintillation materials: Report on the light production,

More information

SNIC Symposium, Stanford, California April The Hybrid Parallel Plates Gas Counter for Medical Imaging

SNIC Symposium, Stanford, California April The Hybrid Parallel Plates Gas Counter for Medical Imaging The Hybrid Parallel Plates Gas Counter for Medical Imaging F. Anulli, G. Bencivenni, C. D Ambrosio, D. Domenici, G. Felici, F. Murtas Laboratori Nazionali di Frascati - INFN, Via E. Fermi 40, I-00044 Frascati,

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

Identification of Shielding Material Configurations Using NMIS Imaging

Identification of Shielding Material Configurations Using NMIS Imaging Identification of Shielding Material Configurations Using NMIS Imaging B. R. Grogan, J. T. Mihalczo, S. M. McConchie, and J. A. Mullens Oak Ridge National Laboratory, P.O. Box 2008, MS-6010, Oak Ridge,

More information

Addition of Noise by Scatter Correction Methods in PVI

Addition of Noise by Scatter Correction Methods in PVI IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 41, NO. 4, AUGUST 1994 1551 Addition of Noise by Scatter Correction Methods in PVI.J. Scott Barneyt, Ronald Harropl and M. Stella Atkinsi t Div. of Nuclear Medicine,

More information

A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations

A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations V. Rosso, G.Battistoni, N. Belcari, N. Camarlinghi, A. Ferrari, S. Ferretti, A. Kraan, A. Mairani, N. Marino, J. E.

More information

COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER

COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER By RAMI RIMON ABU-AITA A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY

More information

In-vivo dose verification for particle therapy

In-vivo dose verification for particle therapy In-vivo dose verification for particle therapy D.R. Schaart, NCS Lustrum, 5-Oct-2012 HollandPTC 1 Protons vs. photons Photons Protons Dennis R. Schaart Delft University of Technology 2 Protons: the promise

More information

In the short span since the introduction of the first

In the short span since the introduction of the first Performance Characteristics of a Newly Developed PET/CT Scanner Using NEMA Standards in 2D and 3D Modes Osama Mawlawi, PhD 1 ; Donald A. Podoloff, MD 2 ; Steve Kohlmyer, MS 3 ; John J. Williams 3 ; Charles

More information

SPECT reconstruction

SPECT reconstruction Regional Training Workshop Advanced Image Processing of SPECT Studies Tygerberg Hospital, 19-23 April 2004 SPECT reconstruction Martin Šámal Charles University Prague, Czech Republic samal@cesnet.cz Tomography

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004 Spiral ASSR Std p = 1.0 Spiral EPBP Std p = 1.0 Kachelrieß et al., Med. Phys. 31(6): 1623-1641, 2004 256 slices (0/300) Advantages of Cone-Beam Spiral CT Image quality nearly independent of pitch Increase

More information

Positron. MillenniumVG. Emission Tomography Imaging with the. GE Medical Systems

Positron. MillenniumVG. Emission Tomography Imaging with the. GE Medical Systems Positron Emission Tomography Imaging with the MillenniumVG GE Medical Systems Table of Contents Introduction... 3 PET Imaging With Gamma Cameras PET Overview... 4 Coincidence Detection on Gamma Cameras...

More information

Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015

Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015 Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI Floris Jansen, GE Healthcare July, 2015 PET/MR 101 : challenges Thermal Workflow & Apps RF interactions?!!

More information

A dedicated tool for PET scanner simulations using FLUKA

A dedicated tool for PET scanner simulations using FLUKA A dedicated tool for PET scanner simulations using FLUKA P. G. Ortega FLUKA meeting June 2013 1 Need for in-vivo treatment monitoring Particles: The good thing is that they stop... Tumour Normal tissue/organ

More information

Pragmatic fully 3D image reconstruction for the MiCES mouse imaging PET scanner

Pragmatic fully 3D image reconstruction for the MiCES mouse imaging PET scanner INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 49 (2004) 4563 4578 PHYSICS IN MEDICINE AND BIOLOGY PII: S0031-9155(04)79577-9 Pragmatic fully 3D image reconstruction for the MiCES mouse imaging PET scanner

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

3-D PET Scatter Correction

3-D PET Scatter Correction Investigation of Accelerated Monte Carlo Techniques for PET Simulation and 3-D PET Scatter Correction C.H. Holdsworth, Student Member, IEEE, C.S. Levin", Member, IEEE, T.H. Farquhar, Student Member, IEEE,

More information

The Design and Implementation of COSEM, an Iterative Algorithm for Fully 3-D Listmode Data

The Design and Implementation of COSEM, an Iterative Algorithm for Fully 3-D Listmode Data IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 20, NO. 7, JULY 2001 633 The Design and Implementation of COSEM, an Iterative Algorithm for Fully 3-D Listmode Data Ron Levkovitz, Dmitry Falikman*, Michael Zibulevsky,

More information

3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction

3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction 3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction Yuni K. Dewaraja, Member, IEEE, Michael Ljungberg, Member, IEEE, and Jeffrey A. Fessler, Member, IEEE Abstract- we

More information

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

PURE. ViSION Edition PET/CT. Patient Comfort Put First. PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion

More information

High-resolution 3D Bayesian image reconstruction using the micropet small-animal scanner

High-resolution 3D Bayesian image reconstruction using the micropet small-animal scanner Phys. Med. Biol. 43 (1998) 1001 1013. Printed in the UK PII: S0031-9155(98)90627-3 High-resolution 3D Bayesian image reconstruction using the micropet small-animal scanner Jinyi Qi, Richard M Leahy, Simon

More information

Improving Reconstructed Image Quality in a Limited-Angle Positron Emission

Improving Reconstructed Image Quality in a Limited-Angle Positron Emission Improving Reconstructed Image Quality in a Limited-Angle Positron Emission Tomography System David Fan-Chung Hsu Department of Electrical Engineering, Stanford University 350 Serra Mall, Stanford CA 94305

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques IMAGING A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques Lai K. Leong, Randall L. Kruger, and Michael K. O Connor Section of Nuclear Medicine, Department

More information

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube T. Donath* a, F. Pfeiffer a,b, O. Bunk a, W. Groot a, M. Bednarzik a, C. Grünzweig a, E. Hempel c, S. Popescu c, M. Hoheisel

More information

Investigation on reconstruction methods applied to 3D terahertz computed Tomography

Investigation on reconstruction methods applied to 3D terahertz computed Tomography Investigation on reconstruction methods applied to 3D terahertz computed Tomography B. Recur, 3 A. Younus, 1, P. Mounaix 1, S. Salort, 2 B. Chassagne, 2 P. Desbarats, 3 J-P. Caumes, 2 and E. Abraham 1

More information

SUV Analysis of F-18 FDG PET Imaging in the Vicinity of the Bladder. Colleen Marie Allen. Graduate Program in Medical Physics Duke University

SUV Analysis of F-18 FDG PET Imaging in the Vicinity of the Bladder. Colleen Marie Allen. Graduate Program in Medical Physics Duke University SUV Analysis of F-18 FDG PET Imaging in the Vicinity of the Bladder by Colleen Marie Allen Graduate Program in Medical Physics Duke University Date: Approved: Timothy Turkington, Supervisor Terence Wong

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

Reconstruction from Projections

Reconstruction from Projections Reconstruction from Projections M.C. Villa Uriol Computational Imaging Lab email: cruz.villa@upf.edu web: http://www.cilab.upf.edu Based on SPECT reconstruction Martin Šámal Charles University Prague,

More information

Emission Computed Tomography Notes

Emission Computed Tomography Notes Noll (24) ECT Notes: Page 1 Emission Computed Tomography Notes Introduction Emission computed tomography (ECT) is the CT applied to nuclear medicine. There are two varieties of ECT: 1. SPECT single-photon

More information

FRONT-END DATA PROCESSING OF NEW POSITRON EMIS- SION TOMOGRAPHY DEMONSTRATOR

FRONT-END DATA PROCESSING OF NEW POSITRON EMIS- SION TOMOGRAPHY DEMONSTRATOR SOUDABEH MORADI FRONT-END DATA PROCESSING OF NEW POSITRON EMIS- SION TOMOGRAPHY DEMONSTRATOR Master of Science Thesis Examiners: Prof. Ulla Ruotsalainen MSc Defne Us Examiners and topic approved by the

More information

Simulations in emission tomography using GATE

Simulations in emission tomography using GATE Simulations in emission tomography using GATE Irène Buvat buvat@imed.jussieu.fr Laboratory of Functional Imaging, U678 INSERM, Paris, France Outline Emission tomography and need for simulations GATE short

More information

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Enhancement Image Quality of CT Using Single Slice Spiral Technique Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,

More information

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology.

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology. Photon Counting Detectors and Their Applications in Medical Imaging Principles, Pitfalls and Progress in Photon-Counting-Detector Technology Taly Gilat Schmidt, PhD Associate Professor Department of Biomedical

More information

An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste

An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste S.C. Kane 1, S. Croft 1, P. McClay 1, W.F. Mueller

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

FUZZY CLUSTERING ALGORITHM FOR GAMMA RAY TRACKING IN SEGMENTED DETECTORS

FUZZY CLUSTERING ALGORITHM FOR GAMMA RAY TRACKING IN SEGMENTED DETECTORS Romanian Reports in Physics, Vol. 62, Nos. 1, P. 27 36, 2010 FUZZY CLUSTERING ALGORITHM FOR GAMMA RAY TRACKING IN SEGMENTED DETECTORS G. SULIMAN, D. BUCURESCU Horia Hulubei National Institute for Physics

More information