3D imaging with SPACE vs 2D TSE in MR guided prostate biopsy

Size: px
Start display at page:

Download "3D imaging with SPACE vs 2D TSE in MR guided prostate biopsy"

Transcription

1 3D imaging with SPACE vs 2D TSE in MR guided prostate biopsy Poster No.: C-2004 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper M. Garmer, S. Mateiescu, M. Busch, D. Groenemeyer; Bochum/ DE Pelvis, Oncology, Genital / Reproductive system male, MR, Image manipulation / Reconstruction, Biopsy, Imaging sequences, Neoplasia /ecr2011/C-2004 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 20

2 Purpose MR guided prostate biopsy is an up-and-coming method with good results and has been evaluated in our institute in more than 70 patients with good performance of standard 2D T2-weighted Turbospinecho sequences (2D T2w TSE) in combination with fast imaging. Localization and accuracy in biopsy of previously detected suspected lesions is limited especially at 1.5 T as the interventional procedure has to be performed without use of an endorectal coil. This reduces the image quality and results in a slightly different anatomic positioning of the prostate. A further limitation is given by the time consuming character of the procedure which excludes multiple long acquisitions with excellent image quality, time should be saved. Some lesions are small and difficult to reach, for example in an apical position (Fig:1-5: preceding diagnostic endorectal MRI of the prostate); high resolution and a good documentation are advisable. Optimized sequence strategies are essential for effective and fast performance of MR guided biopsy; 3D imaging with Sampling Perfection with Application-optimized Contrast using different flip-angle Evolutions (3D SPACE) is evaluated in this context as high resolution and the potential for reconstructions in other planes are promising. Images for this section: Page 2 of 20

3 Fig. 1: T2w TSE axial plane: small suspect lesion in the right peripheral zone Page 3 of 20

4 Fig. 2: T2w TSE koronal plane: apical position of the small suspect lesion Page 4 of 20

5 Fig. 3: ADC (apparent diffusion coefficient) map: restricted diffusion in the lesion Page 5 of 20

6 Fig. 4: T2w TSE, contrast enhanced imaging, colour coded map and conspicuous kinetic curve (red) of suspect lesion Page 6 of 20

7 Fig. 5: Chemical Shift Imaging with conspicuous spectrum in suspect lesion Page 7 of 20

8 Methods and Materials MR guided prostate biopsy was performed in a 1.5 Tesla wide bore scanner (Fig.1) using 2D and 3D imaging before and/or after puncture. All biopsy procedures started with 2D TSE (Fig.2) and/or 3D SPACE (Fig.3) T2-weighted sequences in order to localize the position of the suspected lesion(s) known from the preceding endorectal multifunctional MRI including dynamic contrast imaging, diffusionweighted sequences and 3D chemical shift imaging (spectroscopy). If the lesion was not well defined in T2-weighted images other landmarks (cysts, stones, capsula) were used. Positioning of the cylindric needle guide filled with gadolinium-chelat gel during the procedure was controlled with true fast Imaging with Steady-state Precession (truefisp) in sagittal and coronal plane (Fig.4). After biopsy the needle position was verified using 2D TSE (Fig.5) and/or 3D SPACE (Fig.6) in axial and/or coronal plane according to the location of the lesion. Reconstructions were performed in other planes as required. Fig. 7 shows a sagittal oblique reconstruction from 3D SPACE to indicate the position of the notch (green) in relation to the lesion. Pairs of 2D and 3D axial images were compared in prebiopsy and postbiopsy imaging, image quality and delineation of the suspected lesion / landmarks were rated in reconstructed images. Images for this section: Page 8 of 20

9 Fig. 1: Wide bore scanner with mounted biopsy device Page 9 of 20

10 Fig. 2: 2D T2w TSE axial plane with suspect lesion. TR/TE 4930/100, Slice 3mm TA 1:44 Page 10 of 20

11 Fig. 3: 3D SPACE axial plane with suspect lesion. TR/TE 1400/123 Slice 1.5mm, TA 3:21 Page 11 of 20

12 Fig. 4: Controlled targeting of the needle guide with sagittal and coronal truefisp (right) and axial SPACE (left) Page 12 of 20

13 Fig. 5: 2D TSE axial plane with needle tip in suspect lesion. Page 13 of 20

14 Fig. 6: 3D SPACE axial plane with needle tip in suspect lesion. Page 14 of 20

15 Fig. 7: Sagittal oblique reconstruction from axial 3D SPACE, notch (green)is located sufficiently inside the hypointense lesion; needle tip and base (red) are not capable of removing samples - histopathology: adenocarcinoma Page 15 of 20

16 Results In prebiopsy lesion detection pairs of axial 2D TSE and 3D SPACE could be compared in 27 patients. According to image quality and delineation of the suspected lesion/ landmarks 3D SPACE was found to be superior to 2D TSE in 22/27 patients with better contrast between lesion and normal prostate tissue, equal to 2D TSE in 3/27 patients and inferior to 2D TSE in 2/27 patients because of artifacts. In postbiopsy imaging pairs of axial 2D TSE and 3D SPACE could be compared in 40 specimens taken. 3D SPACE was rated superior to 2D TSE in 29/40 cores, equal to 2D TSE in 9/40 cores and inferior to 2D TSE in only 2/40 cores. 73 reconstruction sets from 3D SPACE sequences in other planes were performed; images were rated according to image quality and delineation of the suspected lesion/ landmarks and needle position; 48 were found to be excellent, 25 were rated sufficient and none was poor. Images for this section: Fig. 1: Prebiopsy lesion detection - comparison of 2D TSE and 3D SPACE Page 16 of 20

17 Fig. 2: Postbiopsy imaging - comparison of 2D TSE and 3D SPACE Page 17 of 20

18 Fig. 3: Postbiopsy imaging - Rating of images reconstructed from 3D SPACE according to image quality and delineation of the suspected lesion/landmarks and needle position Page 18 of 20

19 Conclusion 3D imaging with SPACE offers high resolution with a superior image qualitiy and lesion detection compared to 2D TSE in MR guided prostate biopsy. Sufficient good quality of reconstructed images in other planes offers exact control of needle and notch position in an acceptable time. 3D SPACE offers effective imaging in MR guided prostate biopsy. References References [1] A.B. Rosenkrantz, J. Neil, X. Kong, J. Melamed, J.S. Babb, S.S. Taneja, B. Taouli. Prostate cancer: Comparison of 3D T2-weighted with conventional 2D T2-weighted imagin for image quality and tumor detection. Am J Roentgenol. 2010; 194(2): [2] M. Garmer, M. Busch, R. Brinkmann, S. Mateiescu, B. Wagener, K. Liebenrodt, D.H.W. Groenemeyer. Magnetic Resonance guided prostate biopsy in supine position using a standard wide bore 1.5 T scanner. Oral Presentation ESMRMB 2009 congress Antalya/TR. Personal Information Marietta Garmer, MD Head of Department of Diagnostic Radiology, Groenemeyer Institute for Microtherapy, Bochum, Germany. mail to: garmer@microtherapy.de Serban Mateiescu, graduate engineer Department of Diagnostic Radiology, Groenemeyer Institute for Microtherapy, Bochum, Germany. Page 19 of 20

20 Martin Busch, PhD Head of Medical Physics, Groenemeyer Institute for Microtherapy, Bochum, Germany. Dietrich Groenemeyer, MD Chair of Radiology and Microtherapy at the University of Witten/Herdecke; Groenemeyer Institute for Microtherapy, Bochum, Germany. Page 20 of 20

Dedifferentiated chondrosarcomas: A comprehensive review of imaging and pathologic features

Dedifferentiated chondrosarcomas: A comprehensive review of imaging and pathologic features Dedifferentiated chondrosarcomas: A comprehensive review of imaging and pathologic features Poster No.: C-2187 Congress: ECR 2010 Type: Educational Exhibit Topic: Musculoskeletal - Bone Authors: G. Bierry,

More information

Learning knee MRI the easy way: A multimedial online program for teaching multiplanar MRI anatomy of the knee

Learning knee MRI the easy way: A multimedial online program for teaching multiplanar MRI anatomy of the knee Learning knee MRI the easy way: A multimedial online program for teaching multiplanar MRI anatomy of the knee Poster No.: C-2225 Congress: ECR 2010 Type: Educational Exhibit Topic: Musculoskeletal Authors:

More information

3D temporal subtraction on chest MDCT images using nonlinear image registration technique

3D temporal subtraction on chest MDCT images using nonlinear image registration technique 3D temporal subtraction on chest MDCT images using nonlinear image registration technique Poster No.: C-1072 Congress: ECR 2010 Type: Topic: Authors: Keywords: DOI: Scientific Exhibit Computer Applications

More information

A client-server architecture for semi-automatic segmentation of peripheral vessels in CTA data

A client-server architecture for semi-automatic segmentation of peripheral vessels in CTA data A client-server architecture for semi-automatic segmentation of peripheral vessels in CTA data Poster No.: C-2174 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit A. Grünauer, E. Vuçini,

More information

Using a local Dose Index Registry system to determine Notification Values for the MITA XR 25 (Dose Check) Standard in CT

Using a local Dose Index Registry system to determine Notification Values for the MITA XR 25 (Dose Check) Standard in CT Using a local Dose Index Registry system to determine Notification Values for the MITA XR 25 (Dose Check) Standard in CT Poster No.: C-0913 Congress: ECR 2014 Type: Authors: Scientific Exhibit M. Shinozaki

More information

Digital breast tomosynthesis: comparison of different methods to calculate patient doses

Digital breast tomosynthesis: comparison of different methods to calculate patient doses Digital breast tomosynthesis: comparison of different methods to calculate patient doses Poster No.: C-2220 Congress: ECR 2011 Type: Scientific Paper Authors: A. Jacobs 1, L. Cockmartin 1, D. R. Dance

More information

A validation of the biexponential model in diffusion MRI signal attenuation using diffusion Monte Carlo simulator

A validation of the biexponential model in diffusion MRI signal attenuation using diffusion Monte Carlo simulator A validation of the biexponential model in diffusion MRI signal attenuation using diffusion Monte Carlo simulator Poster No.: C-0331 Congress: ECR 2014 Type: Scientific Exhibit Authors: D. Nishigake, S.

More information

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners Poster No.: C-1748 Congress: ECR 2012 Type: Educational Exhibit Authors: S. Miyashita

More information

EVERNOTE: How radiologists can improve their productivity

EVERNOTE: How radiologists can improve their productivity EVERNOTE: How radiologists can improve their productivity Poster No.: C-0122 Congress: ECR 2013 Type: Educational Exhibit Authors: M. D. Monedero Picazo, M. R. Pastor Juan, M. Villar Garcia, 1 1 1 1 1

More information

Artefacts in body and breast MRI: lessons learned from second reading

Artefacts in body and breast MRI: lessons learned from second reading Artefacts in body and breast MRI: lessons learned from second reading Poster No.: C-0573 Congress: ECR 2013 Type: Scientific Exhibit Authors: M. Nadrljanski, P. Jovanovic, A. Jankovic, Z. C. Milosevic

More information

Artefacts in body and breast MRI: lessons learned from second reading

Artefacts in body and breast MRI: lessons learned from second reading Artefacts in body and breast MRI: lessons learned from second reading Poster No.: C-0573 Congress: ECR 2013 Type: Scientific Exhibit Authors: M. Nadrljanski, P. Jovanovic, A. Jankovic, Z. C. Milosevic

More information

Trabecular Bone Score (TBS) the new parameter of 2D texture analysis for the evaluation of 3D bone micro architecture status.

Trabecular Bone Score (TBS) the new parameter of 2D texture analysis for the evaluation of 3D bone micro architecture status. Trabecular Bone Score (TBS) the new parameter of 2D texture analysis for the evaluation of 3D bone micro architecture status. Poster No.: C-1961 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific

More information

Detector Noise evaluation by means of Continue Wavelet Transform. Comparison with Fourier Transform methods

Detector Noise evaluation by means of Continue Wavelet Transform. Comparison with Fourier Transform methods Detector Noise evaluation by means of Continue Wavelet Transform. Comparison with Fourier Transform methods Poster No.: C-0215 Congress: ECR 2014 Type: Scientific Exhibit Authors: N. Kalyvas 1, S. Angelakis

More information

Functional analysis with DTI and diffusion-neurography of cranial nerves

Functional analysis with DTI and diffusion-neurography of cranial nerves Functional analysis with DTI and diffusion-neurography of cranial nerves Poster No.: C-1942 Congress: ECR 2013 Type: Educational Exhibit Authors: J. P. Martínez Barbero, T. Martín Noguerol, A. Luna Alcalá;

More information

Evaluation of 3D shearwave(tm) elastography and its benefits for the characterization of breast lesions

Evaluation of 3D shearwave(tm) elastography and its benefits for the characterization of breast lesions Evaluation of 3D shearwave(tm) elastography and its benefits for the characterization of breast lesions Poster No.: C-2280 Congress: ECR 2011 Type: Scientific Paper Authors: D. AMY; Aix en Provence/FR

More information

AN EVALUATION OF IMSIMQA AS A TOOL FOR COMMISSIONING 4DCT

AN EVALUATION OF IMSIMQA AS A TOOL FOR COMMISSIONING 4DCT AN EVALUATION OF IMSIMQA AS A TOOL FOR COMMISSIONING 4DCT e-poster: 12474 Congress: ESTRO 2011 Type: eposter Topic: 11th Biennial / QA: of software Authors: N. Whilde; Any information contained in this

More information

Development and Evaluation of a New Method for Measuring of Signal-to-Noise Ratio in Magnetic Resonance Images

Development and Evaluation of a New Method for Measuring of Signal-to-Noise Ratio in Magnetic Resonance Images Development and Evaluation of a New Method for Measuring of Signal-to-Noise Ratio in Magnetic Resonance Images Poster No.: C-0707 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Fukuyama, K. Imai,

More information

Accelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT

Accelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT Accelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT Award: Poster No.: C-2161 Certificate of Merit Congress: ECR 2012 Type: Authors: Scientific Exhibit Q. Yang, M. Elter,

More information

Dynalog data tool for IMRT plan verification

Dynalog data tool for IMRT plan verification Dynalog data tool for IMRT plan verification Poster No.: R-0051 Congress: 2014 CSM Type: Scientific Exhibit Authors: V. Sashin; FOOTSCRAY/AU Keywords: Computer applications, Radiation physics, Experimental,

More information

Integrating information in Radiology: Teaching File System (TFS) implementation in a third-level hospital using MIRC

Integrating information in Radiology: Teaching File System (TFS) implementation in a third-level hospital using MIRC Integrating information in Radiology: Teaching File System (TFS) implementation in a third-level hospital using MIRC Poster No.: C-2964 Congress: ECR 2018 Type: Educational Exhibit Authors: S. Ibáñez Caturla,

More information

Building soft-tissue constraints through a mass spring system for liver surgical simulations

Building soft-tissue constraints through a mass spring system for liver surgical simulations Building soft-tissue constraints through a mass spring system for liver surgical simulations e-poster (content for oral presentation): 754 Congress: ESMRMB 2008 Type: Scientific Paper Topic: Magnetic Resonance

More information

Teleradiology in Italy: results of an online survey

Teleradiology in Italy: results of an online survey Teleradiology in Italy: results of an online survey Poster No.: B-1044 Congress: ECR 2015 Type: Scientific Paper Authors: F. Coppola, E. Neri, D. Regge ; Bologna/IT, Pisa/IT, Turin/IT Keywords: Computer

More information

The use of the modulation transfer function for comparison and image quality assessment of commercially available hybrid PET-CT scanners

The use of the modulation transfer function for comparison and image quality assessment of commercially available hybrid PET-CT scanners The use of the modulation transfer function for comparison and image quality assessment of commercially available hybrid PET-CT scanners Poster No.: C-2943 Congress: ECR 2010 Type: Topic: Scientific Exhibit

More information

Purpose. Methods and Materials

Purpose. Methods and Materials A systematic approach for the objective evaluation of low-contrast performance in MDCT: combination of a fullreference image fidelity metric and a software phantom Poster No.: C-1496 Congress: ECR 2012

More information

A comparison of different VNA philosophies, their associated configurations and respective properties in deployment

A comparison of different VNA philosophies, their associated configurations and respective properties in deployment A comparison of different VNA philosophies, their associated configurations and respective properties in deployment Poster No.: C-2178 Congress: ECR 2013 Type: Authors: Keywords: DOI: Educational Exhibit

More information

The effect of automatic exposure control with inappropriate scout image on radiation dose: a chest phantom experiment with three different CT machines

The effect of automatic exposure control with inappropriate scout image on radiation dose: a chest phantom experiment with three different CT machines The effect of automatic exposure control with inappropriate scout image on radiation dose: a chest phantom experiment with three different CT machines Poster No.: C-0891 Congress: ECR 2015 Type: Authors:

More information

B-230 Skeletal maturity assessment web application

B-230 Skeletal maturity assessment web application B-230 Skeletal maturity assessment web application Scientific Paper B-230 Skeletal maturity assessment web application S. D. Bolboaca (Cluj Napoca/RO) Topic: Computer Applications 1 Purpose In pediatrics,

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

New Technology Allows Multiple Image Contrasts in a Single Scan

New Technology Allows Multiple Image Contrasts in a Single Scan These images were acquired with an investigational device. PD T2 T2 FLAIR T1 MAP T1 FLAIR PSIR T1 New Technology Allows Multiple Image Contrasts in a Single Scan MR exams can be time consuming. A typical

More information

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street MRI is located in the sub basement of CC wing. From Queen or Victoria, follow the baby blue arrows and ride the CC south

More information

Setting up a teleradiology network in a tertiary care neuroradiology unit: Problems, pitfalls, and solutions

Setting up a teleradiology network in a tertiary care neuroradiology unit: Problems, pitfalls, and solutions Setting up a teleradiology network in a tertiary care neuroradiology unit: Problems, pitfalls, and solutions Poster No.: C-0690 Congress: ECR 2013 Type: Educational Exhibit Authors: C. Ozdoba, A. Gennert,

More information

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Outline High Spectral and Spatial Resolution MR Imaging (HiSS) What it is How to do it Ways to use it HiSS for Radiation

More information

Computational Medical Imaging Analysis Chapter 4: Image Visualization

Computational Medical Imaging Analysis Chapter 4: Image Visualization Computational Medical Imaging Analysis Chapter 4: Image Visualization Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky Lexington,

More information

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Doug Dean Final Project Presentation ENGN 2500: Medical Image Analysis May 16, 2011 Outline Review of the

More information

Storage optimization of multi-slice computed tomographic data on PACS systems

Storage optimization of multi-slice computed tomographic data on PACS systems Storage optimization of multi-slice computed tomographic data on PACS systems Poster No.: C-3033 Congress: ECR 2010 Type: Scientific Exhibit Topic: Radiographers Authors: B. Begenau, C. Beigelman, O. Lucidarme,

More information

Human Brain Stereology From Vertical Sections

Human Brain Stereology From Vertical Sections Human Brain Stereology From Vertical Sections Poster No.: C-2276 Congress: ECR 2014 Type: Authors: Scientific Exhibit N. Roberts 1, J. Gelšvartas 1, L. M. Cruz-Orive 2 ; 1 Edinburgh/UK, 2 Santander/ES

More information

e-anatomy User Guide IMAIOS - e-anatomy

e-anatomy User Guide  IMAIOS - e-anatomy e-anatomy User Guide IMAIOS - e-anatomy 3.1.5 Content 1 3 Chapter 1 : The IMAIOS site 4 How the site works 4 The color coding system 4 Princing system 5 Chapter 2 : e-anatomy 5 Presentation 6 The modules

More information

Automatic Quantification of DTI Parameters along Fiber Bundles

Automatic Quantification of DTI Parameters along Fiber Bundles Automatic Quantification of DTI Parameters along Fiber Bundles Jan Klein 1, Simon Hermann 1, Olaf Konrad 1, Horst K. Hahn 1, and Heinz-Otto Peitgen 1 1 MeVis Research, 28359 Bremen Email: klein@mevis.de

More information

Imaging protocols for navigated procedures

Imaging protocols for navigated procedures 9732379 G02 Rev. 1 2015-11 Imaging protocols for navigated procedures How to use this document This document contains imaging protocols for navigated cranial, DBS and stereotactic, ENT, and spine procedures

More information

Development and Optimization of Fourdimensional Magnetic Resonance Imaging (4D-MRI) for Radiation Therapy

Development and Optimization of Fourdimensional Magnetic Resonance Imaging (4D-MRI) for Radiation Therapy Development and Optimization of Fourdimensional Magnetic Resonance Imaging (4D-MRI) for Radiation Therapy by Yilin Liu Medical Physics Graduate Program Duke University Date: Approved: Jing Cai, co-advisor

More information

A 3D Speech Visualization Tool for Turkish based on MRI

A 3D Speech Visualization Tool for Turkish based on MRI A 3D Speech Visualization Tool for Turkish based on MRI Maviş Emel Kulak Kayıkcı, PhD, Hacettepe University, Turkey Can Ölçek, MSc, Sobee Studios, Turkey Erdal Yılmaz, PhD, Sobee Studios, Turkey Burce

More information

Magnetic Resonance Elastography (MRE) of Liver Disease

Magnetic Resonance Elastography (MRE) of Liver Disease Magnetic Resonance Elastography (MRE) of Liver Disease Authored by: Jennifer Dolan Fox, PhD VirtualScopics Inc. jennifer_fox@virtualscopics.com 1-585-249-6231 1. Overview of MRE Imaging MRE is a magnetic

More information

Fiber Selection from Diffusion Tensor Data based on Boolean Operators

Fiber Selection from Diffusion Tensor Data based on Boolean Operators Fiber Selection from Diffusion Tensor Data based on Boolean Operators D. Merhof 1, G. Greiner 2, M. Buchfelder 3, C. Nimsky 4 1 Visual Computing, University of Konstanz, Konstanz, Germany 2 Computer Graphics

More information

Automatic Quantification of DTI Parameters along Fiber Bundles

Automatic Quantification of DTI Parameters along Fiber Bundles Automatic Quantification of DTI Parameters along Fiber Bundles Jan Klein, Simon Hermann, Olaf Konrad, Horst K. Hahn, Heinz-Otto Peitgen MeVis Research, 28359 Bremen Email: klein@mevis.de Abstract. We introduce

More information

Automatic Determination of Arterial Input Function for Dynamic Contrast Enhanced MRI in Tumor Assessment

Automatic Determination of Arterial Input Function for Dynamic Contrast Enhanced MRI in Tumor Assessment Automatic Determination of Arterial Input Function for Dynamic Contrast Enhanced MRI in Tumor Assessment Jeremy Chen, Jianhua Yao, and David Thomasson Diagnostic Radiology Department, Clinical Center,

More information

Intraoperative Prostate Tracking with Slice-to-Volume Registration in MR

Intraoperative Prostate Tracking with Slice-to-Volume Registration in MR Intraoperative Prostate Tracking with Slice-to-Volume Registration in MR Sean Gill a, Purang Abolmaesumi a,b, Siddharth Vikal a, Parvin Mousavi a and Gabor Fichtinger a,b,* (a) School of Computing, Queen

More information

Geometric distortion evaluation using a multi-orientated water-phantom at 0.2 T MRI

Geometric distortion evaluation using a multi-orientated water-phantom at 0.2 T MRI Bio-Medical Materials and Engineering 26 (2015) S1431 S1438 OI 10.3233/BME-151441 IOS Press S1431 Geometric distortion evaluation using a multi-orientated water-phantom at 0.2 T MRI Jochi Jao a, Mingye

More information

New 4-D Imaging for Real-Time Intraoperative MRI: Adaptive 4-D Scan

New 4-D Imaging for Real-Time Intraoperative MRI: Adaptive 4-D Scan New 4-D Imaging for Real-Time Intraoperative MRI: Adaptive 4-D Scan Junichi Tokuda 1, Shigehiro Morikawa 2, Hasnine A. Haque 3, Tetsuji Tsukamoto 3, Kiyoshi Matsumiya 1, Hongen Liao 4, Ken Masamune 1,

More information

Software Strategy for Robotic Transperineal Prostate Therapy in Closed-Bore MRI

Software Strategy for Robotic Transperineal Prostate Therapy in Closed-Bore MRI Software Strategy for Robotic Transperineal Prostate Therapy in Closed-Bore MRI Junichi Tokuda 1,GregoryS.Fischer 2,CsabaCsoma 2, Simon P. DiMaio 3, David G. Gobbi 4, Gabor Fichtinger 4,ClareM.Tempany

More information

A high accuracy multi-image registration method for tracking MRIguided

A high accuracy multi-image registration method for tracking MRIguided A high accuracy multi-image registration method for tracking MRIguided robots Weijian Shang* a, Gregory S. Fischer* a a Worcester Polytechnic Institute, 100 Institute Road, Worcester, MA, USA 01609; ABSTRACT

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

By choosing to view this document, you agree to all provisions of the copyright laws protecting it.

By choosing to view this document, you agree to all provisions of the copyright laws protecting it. Copyright 2009 IEEE. Reprinted from 31 st Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2009. EMBC 2009. Sept. 2009. This material is posted here with permission

More information

A Study of Medical Image Analysis System

A Study of Medical Image Analysis System Indian Journal of Science and Technology, Vol 8(25), DOI: 10.17485/ijst/2015/v8i25/80492, October 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 A Study of Medical Image Analysis System Kim Tae-Eun

More information

DEEP LEARNING WITH ORTHOGONAL VOLUMETRIC HED SEGMENTATION AND 3D SURFACE RECONSTRUCTION MODEL OF PROSTATE MRI

DEEP LEARNING WITH ORTHOGONAL VOLUMETRIC HED SEGMENTATION AND 3D SURFACE RECONSTRUCTION MODEL OF PROSTATE MRI DEEP LEARNING WITH ORTHOGONAL VOLUMETRIC HED SEGMENTATION AND 3D SURFACE RECONSTRUCTION MODEL OF PROSTATE MRI Ruida Cheng a, Nathan Lay b, Francesca Mertan c, Baris Turkbey c, Holger R. Roth b, Le Lu b,

More information

Use of MRI in Radiotherapy: Technical Consideration

Use of MRI in Radiotherapy: Technical Consideration Use of MRI in Radiotherapy: Technical Consideration Yanle Hu, PhD Department of Radiation Oncology, Mayo Clinic Arizona 04/07/2018 2015 MFMER slide-1 Conflict of Interest: None 2015 MFMER slide-2 Objectives

More information

Elastic Registration of Prostate MR Images Based on State Estimation of Dynamical Systems

Elastic Registration of Prostate MR Images Based on State Estimation of Dynamical Systems Elastic Registration of Prostate MR Images Based on State Estimation of Dynamical Systems Bahram Marami a,b,c, Suha Ghoul a, Shahin Sirouspour c, David W. Capson d, Sean R. H. Davidson e, John Trachtenberg

More information

Breast MRI Accreditation Program Clinical Image Quality Guide

Breast MRI Accreditation Program Clinical Image Quality Guide Breast MRI Accreditation Program Clinical Image Quality Guide Introduction This document provides guidance on breast MRI clinical image quality and describes the criteria used by the ACR Breast MRI Accreditation

More information

Orthopedic MRI Protocols. Philips Panorama HFO

Orthopedic MRI Protocols. Philips Panorama HFO Orthopedic MRI Protocols Philips Panorama HFO 1 2 Prepared in collaboration with Dr. John F. Feller, Medical Director of Desert Medical Imaging, Palm Springs, CA. Desert Medical Imaging will provide the

More information

Biomedical Image Processing for Human Elbow

Biomedical Image Processing for Human Elbow Biomedical Image Processing for Human Elbow Akshay Vishnoi, Sharad Mehta, Arpan Gupta Department of Mechanical Engineering Graphic Era University Dehradun, India akshaygeu001@gmail.com, sharadm158@gmail.com

More information

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals.

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals. GE Healthcare CLINICAL GALLERY Discovery * MR750w 3.0T This brochure is intended for European healthcare professionals. NEURO PROPELLER delivers high resolution, motion insensitive imaging in all planes.

More information

ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T

ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T Revision date: 12/13/2010 Overview Susceptibility Weighted Imaging (SWI) is a relatively new data acquisition and processing

More information

Technician s role in the postprocessing of CT and MRI images.

Technician s role in the postprocessing of CT and MRI images. Technician s role in the postprocessing of CT and MRI images. Poster No.: C-1628 Congress: ECR 2013 Type: Educational Exhibit Authors: C. Fraga Piñeiro, M. Centeno, M. González Vázquez, D. Castellón Plaza,

More information

EPILOG PREOP. Viewer Interpretation Guideline

EPILOG PREOP. Viewer Interpretation Guideline EPILOG PREOP Viewer Interpretation Guideline Starting the viewer Launch the viewer for a patient by clicking this icon. A new browser tab or window will open. Data loading To ensure smooth switching between

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

MRI Physics II: Gradients, Imaging

MRI Physics II: Gradients, Imaging MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes

More information

Fiber Selection from Diffusion Tensor Data based on Boolean Operators

Fiber Selection from Diffusion Tensor Data based on Boolean Operators Fiber Selection from Diffusion Tensor Data based on Boolean Operators D. Merhofl, G. Greiner 2, M. Buchfelder 3, C. Nimsky4 1 Visual Computing, University of Konstanz, Konstanz, Germany 2 Computer Graphics

More information

M R I Physics Course

M R I Physics Course M R I Physics Course Multichannel Technology & Parallel Imaging Nathan Yanasak, Ph.D. Jerry Allison Ph.D. Tom Lavin, B.S. Department of Radiology Medical College of Georgia References: 1) The Physics of

More information

TOF-MRA Using Multi-Oblique-Stack Acquisition (MOSA)

TOF-MRA Using Multi-Oblique-Stack Acquisition (MOSA) JOURNAL OF MAGNETIC RESONANCE IMAGING 26:432 436 (2007) Technical Note TOF-MRA Using Multi-Oblique-Stack Acquisition (MOSA) Ed X. Wu, PhD, 1,2 * Edward S. Hui, BEng, 1,2 and Jerry S. Cheung, BEng 1,2 Purpose:

More information

ThE ultimate, INTuITIVE Mr INTErFAcE

ThE ultimate, INTuITIVE Mr INTErFAcE ThE ultimate, INTuITIVE Mr INTErFAcE Empowering you to do more The revolutionary Toshiba M-power user interface takes Mr performance and flexibility to levels higher than ever before. M-power is able to

More information

Linear Intensity-Based Image Registration

Linear Intensity-Based Image Registration Linear Intensity-Based Image Registration Yasmin Mumtaz Ahmad 1, Shahnorbanun Sahran 2, Afzan Adam 3 Center for Artificial Intelligence Technology, Faculty of Information Sciences and Technology, Universiti

More information

The protocols used to scan the musculoskeletal system are tailored to each patient and

The protocols used to scan the musculoskeletal system are tailored to each patient and Chapter 22. Musculoskeletal Protocols The protocols used to scan the musculoskeletal system are tailored to each patient and region being examined. The clinical indication for the examination will also

More information

USAGE PROFILES FOR SYSTEM REQUIREMENTS

USAGE PROFILES FOR SYSTEM REQUIREMENTS USAGE PROFILES FOR SYSTEM REQUIREMENTS Understanding how the customer uses the system, and how its behavior deviates from the expected (and designed) behavior, is the main question that Philips MR wanted

More information

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ORAL DEFENSE 8 th of September 2017 Charlotte MARTIN Supervisor: Pr. MP REVEL M2 Bio Medical

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Page 1 of 9. Protocol: adult_other_adni3basichumanprotocol25x_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION

Page 1 of 9. Protocol: adult_other_adni3basichumanprotocol25x_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION 3 Localizer FOV 26.0 Slice Thickness 5.0 Slice Spacing 0.0 Freq 256 Phase 128 3-PLANE 3 Localizer Unswap Phase Correction Gradient Echo Imaging Options Seq, Fast Recon All Images 3 Localizer Pause / SCIC

More information

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous,

More information

ECE1778 Final Report MRI Visualizer

ECE1778 Final Report MRI Visualizer ECE1778 Final Report MRI Visualizer David Qixiang Chen Alex Rodionov Word Count: 2408 Introduction We aim to develop a mobile phone/tablet based neurosurgical MRI visualization application with the goal

More information

Alternate Biplanar MR Navigation for Microwave Ablation of Liver Tumors

Alternate Biplanar MR Navigation for Microwave Ablation of Liver Tumors Magnetic Resonance in Medical Sciences, Vol. 4, No. 2, p. 89 94, 2005 TECHNICAL NOTE Alternate Biplanar MR Navigation for Microwave Ablation of Liver Tumors Koichiro SATO 1 *, Shigehiro MORIKAWA 2,ToshiroINUBUSHI

More information

Greater TEXAS Federal Credit Union 6411 North Lamar Blvd. Austin, TX (512) Toll Free: (800) Fax: (512)

Greater TEXAS Federal Credit Union 6411 North Lamar Blvd. Austin, TX (512) Toll Free: (800) Fax: (512) Terms and Conditions for Remote Deposit This Remote Deposit Capture Agreement contains the terms and conditions that govern the use of Remote Deposit Capture provided to you by Greater Texas Federal Credit

More information

Role of Parallel Imaging in High Field Functional MRI

Role of Parallel Imaging in High Field Functional MRI Role of Parallel Imaging in High Field Functional MRI Douglas C. Noll & Bradley P. Sutton Department of Biomedical Engineering, University of Michigan Supported by NIH Grant DA15410 & The Whitaker Foundation

More information

Slice-to-Volume Registration and its Potential Application to Interventional MRI-Guided Radio-Frequency Thermal Ablation of Prostate Cancer

Slice-to-Volume Registration and its Potential Application to Interventional MRI-Guided Radio-Frequency Thermal Ablation of Prostate Cancer IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 22, NO. 4, APRIL 2003 515 Slice-to-Volume Registration and its Potential Application to Interventional MRI-Guided Radio-Frequency Thermal Ablation of Prostate

More information

Multi-slice-to-volume registration for MRI-guided transperineal prostate biopsy

Multi-slice-to-volume registration for MRI-guided transperineal prostate biopsy DOI 10.1007/s11548-014-1108-7 ORIGINAL ARTICLE Multi-slice-to-volume registration for MRI-guided transperineal prostate biopsy Helen Xu Andras Lasso Andriy Fedorov Kemal Tuncali Clare Tempany Gabor Fichtinger

More information

Outline: Contrast-enhanced MRA

Outline: Contrast-enhanced MRA Outline: Contrast-enhanced MRA Background Technique Clinical Indications Future Directions Disclosures: GE Health Care: Research support Consultant: Bracco, Bayer The Basics During rapid IV infusion, Gadolinium

More information

6 credits. BMSC-GA Practical Magnetic Resonance Imaging II

6 credits. BMSC-GA Practical Magnetic Resonance Imaging II BMSC-GA 4428 - Practical Magnetic Resonance Imaging II 6 credits Course director: Ricardo Otazo, PhD Course description: This course is a practical introduction to image reconstruction, image analysis

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

Visualisation : Lecture 1. So what is visualisation? Visualisation

Visualisation : Lecture 1. So what is visualisation? Visualisation So what is visualisation? UG4 / M.Sc. Course 2006 toby.breckon@ed.ac.uk Computer Vision Lab. Institute for Perception, Action & Behaviour Introducing 1 Application of interactive 3D computer graphics to

More information

MR-Guided Mixed Reality for Breast Conserving Surgical Planning

MR-Guided Mixed Reality for Breast Conserving Surgical Planning MR-Guided Mixed Reality for Breast Conserving Surgical Planning Suba Srinivasan (subashini7@gmail.com) March 30 th 2017 Mentors: Prof. Brian A. Hargreaves, Prof. Bruce L. Daniel MEDICINE MRI Guided Mixed

More information

5. USER INTERFACE. We use a commercially available six-degree-of-freedom tracking system called the FASTRAK, manufactured by

5. USER INTERFACE. We use a commercially available six-degree-of-freedom tracking system called the FASTRAK, manufactured by We are now working to expand the capabilities of the system in several ways. We intend to integrate localization equipment in the operating room in order to use the system as a navigation aid during surgery.

More information

Deep Learning for Fast and Spatially- Constrained Tissue Quantification from Highly-Undersampled Data in Magnetic Resonance Fingerprinting (MRF)

Deep Learning for Fast and Spatially- Constrained Tissue Quantification from Highly-Undersampled Data in Magnetic Resonance Fingerprinting (MRF) Deep Learning for Fast and Spatially- Constrained Tissue Quantification from Highly-Undersampled Data in Magnetic Resonance Fingerprinting (MRF) Zhenghan Fang 1, Yong Chen 1, Mingxia Liu 1, Yiqiang Zhan

More information

Introducing Computer-Assisted Surgery into combined PET/CT image based Biopsy

Introducing Computer-Assisted Surgery into combined PET/CT image based Biopsy Introducing Computer-Assisted Surgery into combined PET/CT image based Biopsy Santos TO(1), Weitzel T(2), Klaeser B(2), Reyes M(1), Weber S(1) 1 - Artorg Center, University of Bern, Bern, Switzerland 2

More information

NIH Public Access Author Manuscript Proc SPIE. Author manuscript; available in PMC 2010 December 1.

NIH Public Access Author Manuscript Proc SPIE. Author manuscript; available in PMC 2010 December 1. NIH Public Access Author Manuscript Published in final edited form as: Proc SPIE. 2010 February 23; 7625(8): 76251A. Design of a predictive targeting error simulator for MRI-guided prostate biopsy Shachar

More information

Technical Publications

Technical Publications GE Medical Systems Technical Publications Direction 2188003-100 Revision 0 Tissue Volume Analysis DICOM for DICOM V3.0 Copyright 1997 By General Electric Co. Do not duplicate REVISION HISTORY REV DATE

More information

3-D Gradient Shimming

3-D Gradient Shimming 3-D Gradient Shimming 3-D Gradient Shimming on imaging and micro-imaging systems Technical Overview Introduction Advantage statement The GE3DSHIM method is a 3-D gradient shimming procedure developed for

More information

Evaluation of AutoQA Lite TM Image Quality Measurement Software

Evaluation of AutoQA Lite TM Image Quality Measurement Software Evaluation of AutoQA Lite TM Image Quality Measurement Software Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4 2XU Phone: 0131 537 1161

More information

MR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner

MR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner MR-guided radiotherapy: Vision, status and research at the UMC Utrecht Dipl. Ing. Dr. Markus Glitzner About myself Training Medizintechnik TU Graz PhD UMC Utrecht Clinical work Software implementation

More information

Parallel Imaging. Marcin.

Parallel Imaging. Marcin. Parallel Imaging Marcin m.jankiewicz@gmail.com Parallel Imaging initial thoughts Over the last 15 years, great progress in the development of pmri methods has taken place, thereby producing a multitude

More information

Lecture 6: Medical imaging and image-guided interventions

Lecture 6: Medical imaging and image-guided interventions ME 328: Medical Robotics Winter 2019 Lecture 6: Medical imaging and image-guided interventions Allison Okamura Stanford University Updates Assignment 3 Due this Thursday, Jan. 31 Note that this assignment

More information

Data. MAGNETOM Symphony, A Tim System Matrix Coils

Data. MAGNETOM Symphony, A Tim System Matrix Coils Data MAGNETOM Symphony, A Tim System Matrix Coils Head Matrix Coil The Head Matrix coil is part of the standard system configuration. 12-element design with 12 integrated preamplifiers, two rings of 6

More information

An Automatic Algorithm for Quality Assurance of MRI Scanners using a DWI Phantom

An Automatic Algorithm for Quality Assurance of MRI Scanners using a DWI Phantom An Automatic Algorithm for Quality Assurance of MRI Scanners using a DWI Phantom Aalborg University Biomedical Engineering and informatics School of Medicine and Health Master s Thesis By Andreas Ormstrup

More information