Digital breast tomosynthesis: comparison of different methods to calculate patient doses
|
|
- June Carson
- 5 years ago
- Views:
Transcription
1 Digital breast tomosynthesis: comparison of different methods to calculate patient doses Poster No.: C-2220 Congress: ECR 2011 Type: Scientific Paper Authors: A. Jacobs 1, L. Cockmartin 1, D. R. Dance 2, H. Bosmans 1 ; 1 Leuven/ BE, 2 Guildford/UK Keywords: Breast DOI: /ecr2011/C-2220 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 8
2 Purpose Digital breast tomosynthesis (DBT) has been proposed as a promising new development in the field of breast imaging. This pseudo three-dimensional (3D) technique provides images with reduced structure noise from overlaying tissues. While clinical studies are ongoing to evaluate the clinical performance of DBT systems, the associated radiation dose should also be examined. It is important to survey the dose delivered to the patient especially when it is aimed to use DBT systems in a breast screening program. Methods and Materials Data collection was performed on two systems at the University Hospital in Leuven, a Siemens MAMMOMAT Inspiration TOMO breast tomosynthesis system (Siemens, Erlangen, Germany) and a Hologic Selenia Dimensions tomosynthesis system (Hologic, Bedford, USA). Data from 100 clinical cases were collected for both medio-lateral-oblique (MLO) and cranio-caudal (CC) views on the Siemens tomosynthesis system. On the Hologic Dimensions, data of only 18 clinical cases were available. The average glandular dose (AGD) within the patient's breast was calculated using different approaches. The first method is described by Dance et al [1-3], and the following formula is used: AGD = Kgcs#t(#) (1) where K is the incident air kerma for a single projection, measured at zero degree projection angle and the remaining factors were obtained from Monte Carlo simulations. The g factor is the incident air kerma to average glandular dose conversion factor for 50% glandularity breasts, the c factor allows a correction for different glandularities and the s factor corrects for the use of different spectra. These factors are the same as used for 2D breast dose calculations. t(#) is the tomo-factor which is dependent on the projection angle # measured from the center of rotation. For each patient, we calculated a c- factor both for the glandularity as observed in the patient and for an age based glandularity estimate. The second method is described by Sechopoulos et al [4-5]. A different formula is used for the dose calculations: AGD = X CR D g N 0 #RGD(#) (2) Page 2 of 8
3 where AGD is the average glandular dose, X CR is the 'free in air' exposure of a single projection measured at a reference point on the breast support plate (expressed in röntgen) without the compression plate. This reference point is the point where the central x-ray meets the support plate. D g N 0 is the normalized glandular dose per unit exposure for the zero degree projection and RGD(#) is the relative glandular dose coefficient which is calculated for every tomosynthesis projection angle (#) measured from the support plate. The different methods were used to calculate the doses of the two x-ray systems. In a first step, the different methods were compared. Then, the calculated AGD values were compared with the doses indicated on the tomosynthesis systems. Finally, the doses from the tomosynthesis acquisition were compared with 2D doses from a large patient dosimetry study. Results 1. Comparison of different methods Doses were calculated once using the Dance et al method based on individual breast glandularity and once based on the Sechopoulos et al method. The ratio between those two AGD values was made. System 1 System 2 All CC MLO All CC MLO Mean ratio Standard deviation Maximum deviation Table 1: Mean ratio of AGD values calculated by the Dance et al method based on breast glandularity and Sechopoulos' method, for all views and CC and MLO view separately. Also the standard deviation and maximum deviation from the mean value are shown. Table 1 shows a difference of 23 % for the Siemens system (system 1) and 13% for the Hologic system (system 2), which is mainly due to differences in the breast model and the fact that Sechopoulos et al method takes the heel effect into account and Dance et al method does not. The mean ratio for the CC view is 1.15 for system 1 and 1.10 for Page 3 of 8
4 system 2 with a maximum deviation from the mean of 5% for the first system and 10% for the second. 2. Comparison with dose indicated on the system The dose indicated on the system is compared with the calculated AGD values in order to know whether the dose indicated on the system is a good estimation of the doses calculated with the methods described by Dance and Sechopoulos. The results of the Dance et al method based on age was the closest to the system AGD (fig. 1). For both systems, the method of Dance et al based on age provided the best results with only a difference of 5 % between the indicated and the calculated value for the Siemens tomosynthesis system and a 12% difference for the Hologic Dimensions. The differences are larger when the other methods are used. 3. Comparison of AGD values for 2D mammography and breast tomosynthesis On the Siemens tomosynthesis system, a comparison was made between the calculated 3D AGD values and the 2D dose values from a large patient dosimetry study. 2D doses were calculated for about 1000 patients collected on the same system, during the same period of data collection of the DBT patient doses. The 2D dose calculation is based on the formula of Dance et al with an c factor for a screening population and therefore the same method is used for the 3D doses. Fig. 2 shows both the 2D and 3D dose values as a function of the compressed breast thickness. Table 2 presents an overview of the 2D and 3D AGD values for different ranges of compressed breast thicknesses. From this table, it can be seen that, on the Siemens system, doses in one view DBT are almost doubled compared to doses in one view 2D digital mammography. Compressed thickness (mm) breast 2D AGD (mgy) > D AGD (mgy) Table 2: Overview of 2D and 3D AGD values for different ranges of compressed breast thickness. Page 4 of 8
5 Images for this section: Fig. 1: Comparison between AGD indicated on the system and AGD calculated based on the age of the patient. Page 5 of 8
6 Fig. 2: Comparison between 2D and 3D doses in function of compressed breast thickness. The blue dots represent the 2D AGD values and the red squares are the 3D dose values based on the age of the patient. Also the 2D acceptable and achievable dose levels are shown. Page 6 of 8
7 Conclusion The methods proposed to calculate the AGD for breast tomosynthesis were applied and compared on the available dataset in our hospital. It is possible to calculate this AGD in a similar way as for 2D mammography using different methods that could be considered relatively easy and applicable on a larger dataset. There is a small but significant difference between the two methods which can be explained by the different breast models used in the simulations and by the fact that the heel effect was taken into account in Sechopoulos' method but not in the method of Dance et al. The calculated dose values were compared with the system AGD. A good agreement was found between the latter method based on the age of the patient and the indicated dose for both tomosynthesis systems with only a bias of 5% in case of Siemens and 12% in case of Hologic. Finally, we showed that, on the Siemens system, doses in one view DBT are almost double of the doses in one view 2D digital mammography. References [1] D R Dance, C L Skinner, K C Young, J R Beckett and C J Kotre, Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol, Phys. Med. Biol. 45 (2000) [2] D R Dance, K C Young and R E van Engen, Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA breast dosimetry protocols, Phys. Med. Biol. 54 (2009) [3] D R Dance, K C Young and R E van Engen, Estimation of mean glandular breast dose for digital tomosynthesis: factors for use with the UK, European and IAEA breast dosimetry protocols. Phys. Med. Biol. 56 (2011) [4] I Sechopoulos, S Suryanarayanan, S Vendantham, C J D'Orsi and A Karellas, Computation of the glandular radiation dose in digital tomosynthesis of the breast, Med. Phys. 34 (1), January 2007, [5] I Sechopoulos and C J D'Orsi, Glandular radiation dose in tomosynthesis of the breast using tungsten targets, Journal of Applied Clinical Medical Physics, Volume 9, Number 4, Fall 2008, Page 7 of 8
8 Personal Information This work was funded by the OPTIMAM project which is supported by CR-UK & EPSRC Cancer Imaging Programme, in association with the MRC and Department of Health (England). Page 8 of 8
Dedifferentiated chondrosarcomas: A comprehensive review of imaging and pathologic features
Dedifferentiated chondrosarcomas: A comprehensive review of imaging and pathologic features Poster No.: C-2187 Congress: ECR 2010 Type: Educational Exhibit Topic: Musculoskeletal - Bone Authors: G. Bierry,
More information3D temporal subtraction on chest MDCT images using nonlinear image registration technique
3D temporal subtraction on chest MDCT images using nonlinear image registration technique Poster No.: C-1072 Congress: ECR 2010 Type: Topic: Authors: Keywords: DOI: Scientific Exhibit Computer Applications
More informationDetector Noise evaluation by means of Continue Wavelet Transform. Comparison with Fourier Transform methods
Detector Noise evaluation by means of Continue Wavelet Transform. Comparison with Fourier Transform methods Poster No.: C-0215 Congress: ECR 2014 Type: Scientific Exhibit Authors: N. Kalyvas 1, S. Angelakis
More informationUsing a local Dose Index Registry system to determine Notification Values for the MITA XR 25 (Dose Check) Standard in CT
Using a local Dose Index Registry system to determine Notification Values for the MITA XR 25 (Dose Check) Standard in CT Poster No.: C-0913 Congress: ECR 2014 Type: Authors: Scientific Exhibit M. Shinozaki
More informationLearning knee MRI the easy way: A multimedial online program for teaching multiplanar MRI anatomy of the knee
Learning knee MRI the easy way: A multimedial online program for teaching multiplanar MRI anatomy of the knee Poster No.: C-2225 Congress: ECR 2010 Type: Educational Exhibit Topic: Musculoskeletal Authors:
More informationA client-server architecture for semi-automatic segmentation of peripheral vessels in CTA data
A client-server architecture for semi-automatic segmentation of peripheral vessels in CTA data Poster No.: C-2174 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit A. Grünauer, E. Vuçini,
More informationClarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners
Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners Poster No.: C-1748 Congress: ECR 2012 Type: Educational Exhibit Authors: S. Miyashita
More informationA validation of the biexponential model in diffusion MRI signal attenuation using diffusion Monte Carlo simulator
A validation of the biexponential model in diffusion MRI signal attenuation using diffusion Monte Carlo simulator Poster No.: C-0331 Congress: ECR 2014 Type: Scientific Exhibit Authors: D. Nishigake, S.
More informationTrabecular Bone Score (TBS) the new parameter of 2D texture analysis for the evaluation of 3D bone micro architecture status.
Trabecular Bone Score (TBS) the new parameter of 2D texture analysis for the evaluation of 3D bone micro architecture status. Poster No.: C-1961 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific
More information3D imaging with SPACE vs 2D TSE in MR guided prostate biopsy
3D imaging with SPACE vs 2D TSE in MR guided prostate biopsy Poster No.: C-2004 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper M. Garmer, S. Mateiescu, M. Busch, D. Groenemeyer; Bochum/
More informationAN EVALUATION OF IMSIMQA AS A TOOL FOR COMMISSIONING 4DCT
AN EVALUATION OF IMSIMQA AS A TOOL FOR COMMISSIONING 4DCT e-poster: 12474 Congress: ESTRO 2011 Type: eposter Topic: 11th Biennial / QA: of software Authors: N. Whilde; Any information contained in this
More informationEVERNOTE: How radiologists can improve their productivity
EVERNOTE: How radiologists can improve their productivity Poster No.: C-0122 Congress: ECR 2013 Type: Educational Exhibit Authors: M. D. Monedero Picazo, M. R. Pastor Juan, M. Villar Garcia, 1 1 1 1 1
More informationDynalog data tool for IMRT plan verification
Dynalog data tool for IMRT plan verification Poster No.: R-0051 Congress: 2014 CSM Type: Scientific Exhibit Authors: V. Sashin; FOOTSCRAY/AU Keywords: Computer applications, Radiation physics, Experimental,
More informationDevelopment and Evaluation of a New Method for Measuring of Signal-to-Noise Ratio in Magnetic Resonance Images
Development and Evaluation of a New Method for Measuring of Signal-to-Noise Ratio in Magnetic Resonance Images Poster No.: C-0707 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Fukuyama, K. Imai,
More informationThe effect of automatic exposure control with inappropriate scout image on radiation dose: a chest phantom experiment with three different CT machines
The effect of automatic exposure control with inappropriate scout image on radiation dose: a chest phantom experiment with three different CT machines Poster No.: C-0891 Congress: ECR 2015 Type: Authors:
More informationMonte Carlo simulation of scatter field for calculation of contrast of discs in synthetic CDMAM images
Monte Carlo simulation of scatter field for calculation of contrast of discs in synthetic CDMAM images Oliver Díaz 1, Mary Yip 1, Jorge Cabello 1, David R. Dance 2,3, Kenneth C. Young 2,3, and Kevin Wells
More informationEvaluation of 3D shearwave(tm) elastography and its benefits for the characterization of breast lesions
Evaluation of 3D shearwave(tm) elastography and its benefits for the characterization of breast lesions Poster No.: C-2280 Congress: ECR 2011 Type: Scientific Paper Authors: D. AMY; Aix en Provence/FR
More informationDetection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis
Detection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis Andria Hadjipanteli a, Premkumar Elangovan b, Padraig T Looney a, Alistair Mackenzie a,
More informationAccelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT
Accelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT Award: Poster No.: C-2161 Certificate of Merit Congress: ECR 2012 Type: Authors: Scientific Exhibit Q. Yang, M. Elter,
More informationIntegrating information in Radiology: Teaching File System (TFS) implementation in a third-level hospital using MIRC
Integrating information in Radiology: Teaching File System (TFS) implementation in a third-level hospital using MIRC Poster No.: C-2964 Congress: ECR 2018 Type: Educational Exhibit Authors: S. Ibáñez Caturla,
More informationTeleradiology in Italy: results of an online survey
Teleradiology in Italy: results of an online survey Poster No.: B-1044 Congress: ECR 2015 Type: Scientific Paper Authors: F. Coppola, E. Neri, D. Regge ; Bologna/IT, Pisa/IT, Turin/IT Keywords: Computer
More informationArtefacts in body and breast MRI: lessons learned from second reading
Artefacts in body and breast MRI: lessons learned from second reading Poster No.: C-0573 Congress: ECR 2013 Type: Scientific Exhibit Authors: M. Nadrljanski, P. Jovanovic, A. Jankovic, Z. C. Milosevic
More informationArtefacts in body and breast MRI: lessons learned from second reading
Artefacts in body and breast MRI: lessons learned from second reading Poster No.: C-0573 Congress: ECR 2013 Type: Scientific Exhibit Authors: M. Nadrljanski, P. Jovanovic, A. Jankovic, Z. C. Milosevic
More informationPower Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis
Power Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis Lesley Cockmartin 1,*, Predrag R. Bakic 2, Hilde Bosmans 1, Andrew
More informationPurpose. Methods and Materials
A systematic approach for the objective evaluation of low-contrast performance in MDCT: combination of a fullreference image fidelity metric and a software phantom Poster No.: C-1496 Congress: ECR 2012
More informationThe use of the modulation transfer function for comparison and image quality assessment of commercially available hybrid PET-CT scanners
The use of the modulation transfer function for comparison and image quality assessment of commercially available hybrid PET-CT scanners Poster No.: C-2943 Congress: ECR 2010 Type: Topic: Scientific Exhibit
More information4/19/2016. Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT
Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT 1 Three manufacturers approved for Tomo Hologic and GE, and Siemens Why 2D Digital Mammography 2D FFDM it appears to
More informationB-230 Skeletal maturity assessment web application
B-230 Skeletal maturity assessment web application Scientific Paper B-230 Skeletal maturity assessment web application S. D. Bolboaca (Cluj Napoca/RO) Topic: Computer Applications 1 Purpose In pediatrics,
More informationA comparison of different VNA philosophies, their associated configurations and respective properties in deployment
A comparison of different VNA philosophies, their associated configurations and respective properties in deployment Poster No.: C-2178 Congress: ECR 2013 Type: Authors: Keywords: DOI: Educational Exhibit
More informationA fast scatter field estimator for Digital Breast Tomosynthesis
A fast scatter field estimator for Digital Breast Tomosynthesis Oliver Díaz a, David R. Dance b,c, Kenneth C. Young b,c, Premkumar Elangovan a, Predrag R. Bakic d and Kevin Wells a a Centre for Vision,
More informationFunctional analysis with DTI and diffusion-neurography of cranial nerves
Functional analysis with DTI and diffusion-neurography of cranial nerves Poster No.: C-1942 Congress: ECR 2013 Type: Educational Exhibit Authors: J. P. Martínez Barbero, T. Martín Noguerol, A. Luna Alcalá;
More informationCertificate Extension II 8/1/2016. Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role
Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role Kish Chakrabarti, Ph.D., FAAPM Division of Mammography Quality Standards Center for Devices and Radiological
More informationBuilding soft-tissue constraints through a mass spring system for liver surgical simulations
Building soft-tissue constraints through a mass spring system for liver surgical simulations e-poster (content for oral presentation): 754 Congress: ESMRMB 2008 Type: Scientific Paper Topic: Magnetic Resonance
More informationAcknowledgments and financial disclosure
AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories
More informationBackground 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.
Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging Stephen J. Glick* J. Michael O Connor**, Clay Didier**, Mini Das*, * University of Massachusetts Medical
More informationScattered radiation in DBT geometries with flexible breast compression paddles: A Monte Carlo simulation study
Scattered radiation in DBT geometries with flexible breast compression paddles: A Monte Carlo simulation study Oliver Diaz, Eloy García, Arnau Oliver, Joan Martí, Robert Martí a a VICOROB research institute,
More informationEffective detective quantum efficiency (edqe) and effective noise equivalent quanta (eneq) for system optimization purposes in digital mammography
Effective detective quantum efficiency (edqe) and effective noise equivalent quanta (eneq) for system optimization purposes in digital mammography Elena Salvagnini a,b, Hilde Bosmans a, Lara Struelens
More informationLesion Ground Truth Estimation for a Physical Breast Phantom
Lesion Ground Truth Estimation for a Physical Breast Phantom Suneeza Hanif 1, Frank Schebesch 1, Anna Jerebko 2, Ludwig Ritschl 2, Thomas Mertelmeier 2, Andreas Maier 1 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg,
More informationINVESTIGATION OF PHYSICAL PROCESSES IN DIGITAL X-RAY TOMOSYNTHESIS IMAGING OF THE BREAST. A Thesis Presented to The Academic Faculty
INVESTIGATION OF PHYSICAL PROCESSES IN DIGITAL X-RAY TOMOSYNTHESIS IMAGING OF THE BREAST A Thesis Presented to The Academic Faculty By Ioannis Sechopoulos In Partial Fulfillment Of the Requirements for
More informationDetailed analysis of scatter contribution from different simulated geometries of X-ray detectors
Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors Elena Marimon 1,2*, Hammadi Nait-Charif 1, Asmar Khan 2, Philip A. Marsden 3, Oliver Diaz 4 1 Centre for
More informationSetting up a teleradiology network in a tertiary care neuroradiology unit: Problems, pitfalls, and solutions
Setting up a teleradiology network in a tertiary care neuroradiology unit: Problems, pitfalls, and solutions Poster No.: C-0690 Congress: ECR 2013 Type: Educational Exhibit Authors: C. Ozdoba, A. Gennert,
More informationUnique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.
Unique Features of the GE Senoclaire Tomosynthesis System Tyler Fisher, M.S., DABR Therapy Physics, Inc. Conflict of Interest Disclosure I have no conflicts to disclose. Learning Objectives Overview of
More informationMonte Carlo Reference Data Sets for Imaging Research
AAPM REPORT NO. 195 Monte Carlo Reference Data Sets for Imaging Research The Report of AAPM Task Group 195 September 2015 DISCLAIMER: This publication is based on sources and information believed to be
More informationValidation of a digital mammography image simulation chain with automated scoring of CDMAM images
Validation of a digital mammography image simulation chain with automated scoring of CDMAM images Mary Yip 1, Abdulaziz Alsager 2, Emma Lewis 1, Kevin Wells 1, and Kenneth C. Young 3 1 Centre for Vision,
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationAverage glandular dose coefficients for pendant-geometry breast CT using realistic breast phantoms
Average glandular dose coefficients for pendant-geometry breast CT using realistic breast phantoms Andrew M. Hernandez Department of Radiology, Biomedical Engineering Graduate Group, University of California
More informationDesign Considerations in Optimizing a Breast Tomosynthesis System
Design Considerations in Optimizing a Breast Tomosynthesis System Andrew Smith, Ph.D., Vice President - Imaging Science, Hologic Introduction Breast tomosynthesis, also referred to as three-dimensional
More informationSynthesized 2D Mammographic Imaging
Synthesized 2D Mammographic Imaging Theory and Clinical Performance Synthesized 2D Mammographic Imaging Theory and Clinical Performance Andrew Smith, Ph.D., Vice President, Image Research Hologic, Inc.,
More information8/2/2016. Hands-on GE SenoClaire DBT: technical characteristics & quality control. Disclosures and Acknowledgements
Hands-on GE SenoClaire DBT: technical characteristics & quality control Razvan Iordache, Ph.D. GE Healthcare Imagination at work Disclosures and Acknowledgements Razvan Iordache Employee of GE Healthcare
More informationHidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.
2-D D Dose-CT Mapping in Geant4 Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. Table of Contents Background & Purpose Materials Methods
More informationSimulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images
Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),
More informationMegan A. Wood, M.S. Under the direction of Larry DeWerd, Ph.D. University of Wisconsin Medical Radiation Research Center (UWMRRC)
Megan A. Wood, M.S. Under the direction of Larry DeWerd, Ph.D. University of Wisconsin Medical Radiation Research Center (UWMRRC) NCCAAPM Spring Meeting May 3, 2013 Introduction FFF background Project
More informationGeneralized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction
Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction Klaus Erhard a, Michael Grass a, Sebastian Hitziger b, Armin Iske b and Tim Nielsen a a Philips Research Europe Hamburg,
More informationMonte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters
Monte Carlo Simulation for Neptun 1 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters M.T. Bahreyni Toossi a, M. Momen Nezhad b, S.M. Hashemi a a Medical Physics Research Center,
More information10/9/2018. Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT
Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT 1 Mammography Five FFDM approved for Tomo Hologic, Ge Senoclair and GE Pristina, Siemens, and Fujifilm 2 Why 2D Digital
More informationPCXMC 2.0 SUPPLEMENTARY PROGRAMS USER'S GUIDE
FEBRUARY 2012 PCXMC 2.0 SUPPLEMENTARY PROGRAMS USER'S GUIDE PCXMC20Rotation.exe Autocalc-sheet.xls AutocalcRotation-sheet.xls Markku Tapiovaara STUK - Radiation and Nuclear Safety Authority P.O. Box 14
More informationScatter reduction for grid-less mammography using the convolutionbased image post-processing technique
Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique Elena Marimón *a,b, Hammadi Nait-Charif a, Asmar Khan b, Philip A. Marsden c, Oliver Diaz d a CDE
More informationRECENTLY APPROVED GE FFDM DBT TESTING PROCEDURES.
RECENTLY APPROVED GE FFDM DBT TESTING PROCEDURES. S. G U R U P R A S A D, P H. D., D A B R, E M E R I T U S, M AT H E W H A L L M. S. N O RT H S H O R E U N I V E R S I T Y H E A LT H S Y S T E M A N D
More informationDisclosure 7/24/2014. Validation of Monte Carlo Simulations For Medical Imaging Experimental validation and the AAPM Task Group 195 Report
Validation of Monte Carlo Simulations For Medical Imaging Experimental validation and the AAPM Task Group 195 Report Ioannis Sechopoulos, Ph.D., DABR Diagnostic Imaging Physics Lab Department of Radiology
More informationCBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona
CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona Introduction CBCT has become a routine procedure for
More informationBreast Tomosynthesis: Impact on IT Systems. dondennison.com
Breast Tomosynthesis: Impact on IT Systems MIIT 2016, HAMILTON Don K Dennison Industry consultant with 14+ years experience in imaging IT Society for Imaging Informatics in Medicine (SIIM) Board of Directors
More informationContrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis
Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision
More informationABSTRACT 1. INTRODUCTION
Estimating breast thickness for dual-energy subtraction in contrastenhanced digital mammography using calibration phantoms Kristen C. Lau 1, Young Joon Kwon, Moez Karim Aziz, Raymond J. Acciavatti, Andrew
More informationComparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms.
Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. D.Leone, A.Häußler Intitute for Nuclear Waste Disposal, Karlsruhe Institute for Technology,
More informationNoise power spectrum and modulation transfer function analysis of breast tomosynthesis imaging
Noise power spectrum and modulation transfer function analysis of breast tomosynthesis imaging Weihua Zhou a, Linlin Cong b, Xin Qian c, Yueh Z. Lee d, Jianping Lu c,e, Otto Zhou c,e, *Ying Chen a,b a
More informationENGLISH. Digital Breast Tomosynthesis
ENGLISH Digital Breast Tomosynthesis Medical imaging excellence Planmed is strongly committed to improving the early detection of breast cancer. Our mission is to provide top quality imaging systems for
More informationMedical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.
DXRaySMCS First User Friendly Interface Developed for Prediction of Diagnostic Radiology X-Ray Spectra Produced by Monte Carlo (MCNP-4C) Simulation in Iran M.T. Bahreyni Toosi a*, H. Moradi b, H. Zare
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 60601-2-44 2001 AMENDMENT 1 2002-09 Amendment 1 Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment for computed tomography Amendement
More informationI. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital
International Journal of Computational Engineering Research Vol, 04 Issue, 4 Simulation of Photon and Electron dose distributions 5 code for the treatment area using the linear electron accelerator (LINAC)
More informationImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners
ImPACT Imaging Performance Assessment of CT Scanners A Medical Devices Agency Evaluation Group CT dosimetry and a data base for CTDI values EFOMP Workshop at ECR 2002 S. Edyvean, ImPACT St George s Hospital,
More informationInformatics I - The digital breast image interoperability and management
Informatics I - The digital breast image interoperability and management Alain Gauvin Montebello, 2017-02-01 DICOM HL7 VARIABLE Orders, demography, (HL7 ORM, ADT) Images + metadata (dicom storage) Modality
More informationUniversità degli Studi di Napoli Federico II
Università degli Studi di Napoli Federico II DOTTORATO DI RICERCA IN FISICA Ciclo XXIX Coordinatore: Prof. Salvatore Capozziello Dose and image quality in X-ray phase contrast breast imaging Settore Scientifico
More informationBreast Tissue Classification in Digital Tomosynthesis Images Based on Global Gradient Minimization and Texture Features
Breast Tissue Classification in Digital Tomosynthesis Images Based on Global Gradient Minimization and Texture Features Xulei Qin, Emory University Guolan Lu, Emory University Ioannis Sechopoulos, Emory
More informationDIGITALIZATION OF A MAMMOGRAPHIC PHANTOM VIEW THROUGH A MONTE CARLO SIMULATION
DIGITALIZATION OF A MAMMOGRAPHIC PHANTOM VIEW THROUGH A MONTE CARLO SIMULATION M. Ramos 1, S. Ferrer 1, J.I. Villaescusa 2, G. Verdú 1, J.M. Campayo 3 1 Departamento de Ingeniería Química y Nuclear. Universidad
More informationMAMMOMAT 1000 / 3000 Nova. Accessories. Accessories (optional) Compression plate with low edge 18 cm x 24 cm #
MAMMOMAT 1000 / 3000 Nova Accessories Accessories (optional) Compression plate with low edge 18 cm x 24 cm # 44 94 378 Standard compression plate with 40 mm high front edge. Compression plate with high
More informationDesign and development of a phantom for tomosynthesis with potential for automated analysis via the cloud
Received: 21 August 2017 Revised: 25 October 2017 Accepted: 7 January 2018 DOI: 10.1002/acm2.12297 MEDICAL IMAGING Design and development of a phantom for tomosynthesis with potential for automated analysis
More informationIterative CT Reconstruction Using Curvelet-Based Regularization
Iterative CT Reconstruction Using Curvelet-Based Regularization Haibo Wu 1,2, Andreas Maier 1, Joachim Hornegger 1,2 1 Pattern Recognition Lab (LME), Department of Computer Science, 2 Graduate School in
More informationScaling Calibration in the ATRACT Algorithm
Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,
More informationInvestigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers
Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Krista Chytyk MSc student Supervisor: Dr. Boyd McCurdy Introduction The objective of cancer radiotherapy
More informationHuman Brain Stereology From Vertical Sections
Human Brain Stereology From Vertical Sections Poster No.: C-2276 Congress: ECR 2014 Type: Authors: Scientific Exhibit N. Roberts 1, J. Gelšvartas 1, L. M. Cruz-Orive 2 ; 1 Edinburgh/UK, 2 Santander/ES
More informationAssessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT
Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan
More informationApplication of MCNP Code in Shielding Design for Radioactive Sources
Application of MCNP Code in Shielding Design for Radioactive Sources Ibrahim A. Alrammah Abstract This paper presents three tasks: Task 1 explores: the detected number of as a function of polythene moderator
More information8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit
Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant
More informationTHE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 THE SIMULATION OF
More informationTerms of Use. Changes. General Use.
Terms of Use THESE TERMS AND CONDITIONS (THE TERMS ) ARE A LEGAL CONTRACT BETWEEN YOU AND SPIN TRANSFER TECHNOLOGIES ( SPIN TRANSFER TECHNOLOGIES, STT, WE OR US ). THE TERMS EXPLAIN HOW YOU ARE PERMITTED
More information(12) Patent Application Publication (10) Pub. No.: US 2007/ A1
(19) United States (12) Patent Application Publication (10) Pub. No.: US 2007/0268999 A1 Ullberg et al. US 20070268.999A1 (43) Pub. Date: (54) (75) (73) (21) (22) (30) APPARATUS AND METHOD FOR CREATING
More informationShielding factors for traditional safety glasses
Shielding factors for traditional safety glasses Malcolm McEwen, Hong Shen and Ernesto Mainegra-Hing Ionizing Radiation Standards, National Research Council Canada Alan DuSautoy, Radiation and Health Sciences
More information3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH
3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk
More informationFuture Topics. Projection Imaging Dose Reporting, XA 3D Volume Objects. for DICOM WG-02. Presented by Heinz Blendinger, Siemens Medical Solutions
Future Topics for Projection Imaging Dose Reporting, XA 3D Volume Objects DICOM WG-02 Presented by Heinz Blendinger, Siemens Medical Solutions 1 Presentation outline Dose Reporting Why Dose Reporting?
More informationDUAL energy X-ray radiography [1] can be used to separate
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy
More informationLorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc.
Lorad FFDM QC Procedures for Medical Physicist Tao Wu, Ph.D. Hologic, Inc. Lorad Selenia FFDM Tests Following Lorad QC Manual Collimation Assessment Artifact Evaluation System Resolution Phantom Image
More informationLinköping University Post Print A MONTE CARLO-BASED MODEL FOR SIMULATION OF DIGITAL CHEST TOMOSYNTHESIS
Linköping University Post Print A MONTE CARLO-BASED MODEL FOR SIMULATION OF DIGITAL CHEST TOMOSYNTHESIS Gustaf Ullman, David R. Dance, Michael Sandborg, Gudrun Alm Carlsson, Angelica Svalkvist and Magnus
More informationAir Kerma Strength Measurements and Monte Carlo based Dosimetric Characterization of a directional Pd 103 planar source array.
Air Kerma Strength Measurements and Monte Carlo based Dosimetric Characterization of a directional Pd 103 planar source array. MANIK AIMA UWMRRC, MADISON 2014 NCAAPM Autumn Meeting October 24 th, 2014
More informationAnalysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code
American Journal of Biomedical Engineering 216, 6(4): 124-131 DOI: 1.5923/j.ajbe.21664.3 Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code Ankit Kajaria 1,*, Neeraj Sharma
More informationOptimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago
Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic
More informationACMP 25th Annual Meeting
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation ACMP 25th Annual Meeting Seattle, WA May 3, 2008 Melissa C. Martin, M.S., FACR Therapy Physics, Inc. 879 West 190th St., Ste 419,
More informationDesign and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon
Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization
More informationDosimetric optimization of a conical breast brachytherapy applicator for improved skin dose sparing
Dosimetric optimization of a conical breast brachytherapy applicator for improved dose sparing 5 Yun Yang Biomedical Engineering and Biotechnology, University of Massachusetts Lowell, Massachusetts 01854
More informationPowerLook AMP DICOM Conformance Statement
icad, Inc. PowerLook AMP DICOM Conformance Statement 0071-4003-D Form 1 Proprietary Information of icad, Inc. Page 1 of 113 2015 icad, Inc. All rights reserved. Reproduction of this document without prior
More informationDigital Breast Tomosynthesis: The Fundamentals
Digital Breast Tomosynthesis: The Fundamentals 8:00 am Registration and Coffee 8:30 am Introduction What is DBT? Why do we need DBT? How does DBT work? Clinical issues Clinical advantages 10:10 am Informal
More information