Monte-Carlo simulation for scatter correction compensation studies in SPECT imaging using GATE software package

Size: px
Start display at page:

Download "Monte-Carlo simulation for scatter correction compensation studies in SPECT imaging using GATE software package"

Transcription

1 Nuclear Instruments and Methods in Physics Research A 569 (2006) Monte-Carlo simulation for scatter correction compensation studies in SPECT imaging using GATE software package N.G. Sakellios a,, E. Karali a, D. Lazaro b, G.K. Loudos a, K.S. Nikita a a Nuclear Imaging Medical Group, Biomedical Simulations and Imaging Applications Laboratory, School of Electrical and Computer Engineering, National Technical University of Athens, 9 Iroon Polytechniou Street,GR15780 Athens, Greece b UMR 678 INSERM UPMC, CHU Pitié Salpêtière,91 Boulevard de l Hôpital, Paris Cedex 13, France Available online 23 October 2006 Abstract Simulations using Monte-Carlo packages are increasingly used in Nuclear Imaging, both for PET and SPECT applications. Either for modeling imaging systems or developing algorithms for analysis and improvement of image quantification. The aim of this work is the study of several scatter phenomena and the evaluation and improvement of two scatter correction compensation methods in pixelized scintillators. Firstly, the Dual Energy Window Subtraction Technique (DEWST) and then the Multi Energy Window and Filtering Technique (MEWFT) are investigated. Minimization Method has been used in order to achieve the best weighting co-efficient and the appropriate position and width of each low energy window. The position of them was found to be around kev. Following scatter rejection, both quality and quantitative accuracy of the images improved. r 2006 Elsevier B.V. All rights reserved. PACS: Pm Keywords: Image correction; Single photon emission computed tomography; Scatter; Minimization method 1. Introduction The aim of single photon emission computed tomography is to provide accurate images of the radionuclide distribution present in the object being scanned. The inclusion of scatter radiation is one of the main sources of error in SPECT imaging. Apparently, the detection of scattered events affects the contrast of the lesions and the resolution of the images. Hence, it will introduce extraneous activity in both cold and hot phantoms. During Compton scattering, photons lose part of their original energy and change their incident direction. So, energy loss can be used in order to identify scattered photons and eliminate them from detection. Several methods for the effects of scattered events have been proposed [1 8]. The difference between methods is the way of estimating the scatter contribution. For example, Corresponding author. Tel.: ; fax: address: snicol@biosim.ntua.gr (N.G. Sakellios). some methods involve events collection in one or more energy windows lower than the photopeak one, and these data can be subtracted from the true non-scattered data, before or after reconstruction. Other techniques try to estimate the shape of the scatter component within the photopeak window as a convolution of the photopeak projection with the scatter distribution function. Trying to evaluate and improve those methods, we use Monte-Carlo simulation code. The advantage of the simulation is the absolute control of any parameters involved and the wide range of analysis from the user while knowing the history of each photon created and the interactions in which it has participated. The aim of this work is to evaluate and improve two scatter correction compensation methods using simulation data obtained with GATE package. Dual Energy Window Subtraction Technique (DEWST), Multi Energy Window- Filtering Technique (MEWFT) and Convolution Subtraction Technique (CST) have been studied /$ - see front matter r 2006 Elsevier B.V. All rights reserved. doi: /j.nima

2 N.G. Sakellios et al. / Nuclear Instruments and Methods in Physics Research A 569 (2006) Materials and methods 2.1. GATE simulation platform The GATE simulation platform is based on GEANT4 toolkit [9]. It comprises a large number of software components enabling the modeling of various aspects of nuclear medicine experiments, including the detector geometry and movement with source kinetics, the radiation detection, data output, etc. Using an extended version of GEANT4 scripting language the user can choose among these options without performing any C++ coding. New scanner geometries can be defined and simulations can be set up interactively according to the user requirements. The SPECT scanner that was simulated was a scintillation camera (3 mm thick CsI(Tl)crystal array) based on a PSPMT (Hamamatsu R2486) dedicated to small animal imaging and radiopharmaceuticals testing [10]. The prototype is equipped with a removable low-energy and highresolution collimator (2.75 cm thick, mm septal thickness, 1.12 mm diameter hexagonal holes) Methods of scatter compensation The methods that were studied are as follows DEWST The DEWST proposed by Jaszczak et al. [1] involves data collection in a lower energy window, which provides an estimate for the scatter component in the photopeak window. These data are appropriately weighted and subtracted from the photopeak data. Two images are used for scatter compensation; the photopeak image P(x) and the lower energy window image L(x). C(x) is obtained as CðxÞ ¼PðxÞ klðxþ (1) where x is the vector of image pixels and k is a weighting factor. Since this method can be applied to projection data before reconstruction of tomographic slices it can be used for scatter compensation of planar images as well. In the case of pixelized scintillators a lower energy window has to be determined for each crystal cell. In a previous work [11] we assumed that each crystal element can be treated as an independent detector. We tested several lower energy windows in order to find the window that carries the most significant scatter information in a number of different imaging cases. The central channel of the optimal window was at 65% of the photopeak channel of each crystal element and its width was 75% of the photopeak channel. The optimal value for factor k was found to be equal to 0.5. These parameters provided improved results as compared with the standard single window method and they have been used in this work MEWFT The multi energy window acquisition method has been proposed by Todd-Pokropek et al. [2]. It relies on the assumption that the scatter component can be calculated as a weighted mean of data, recorded in more than one low energy windows. Furthermore, filter functions are used for the convolution of the Compton projection in each low energy window, as a result of the variability of the system transfer functions with regard to energy Sðx; FÞ ¼ X K T i C i ðx; FÞh i ðxþ (2) i where S(x,F) is the scatter photopeak projection, C i (x,f) the Compton projection, K T i ðx; FÞ the weighting coefficient and h i (x) the filter function at the ith low energy window, respectively. The derivation of the filter functions h i (x) was achieved by using the Line Spread Function (LSF). It is supposed that LSFs are good approximations of the system transfer functions in each low window used LSF s ¼ LSF i h i (3) where LSF s and LSF i are the LSF of the Compton component at the photopeak and the ith low energy window. The last step for the calculation of the filter function is the use of the Fourier theorem, where F and F 1 indicates the Fourier and Inverse Fourier transform, respectively, FðLSF s Þ¼FðLSF i ÞFðh i Þ (4) and the filter functions are h i ¼ F 1 FðLSF s Þ W (5) FðLSF i Þ where W is a window for filter function stabilization [12]. The corrected images T(x,F) finally calculated by subtracting the scatter component S(x,F) of Eq. (2) from the photopeak projections P(x,F) Tðx; FÞ ¼Pðx; FÞ Sðx; FÞ. (6) The photopeak window was set at kev (20%) and the different low energy windows vary from 80 to 120 kev Minimization method The function that will be minimized is shown below. The free parameters are three. The weighting factor wf i for each one of the low energy windows I low. The position pc i of low windows and its width w i are calculated I min ðx; yþ ¼I ph ðx; yþ X wf i I sc ðx; yþ i I NULL ðx; yþ. ð7þ The I ph window is the photopeak energy window (20%). In I NULL window, GATE is storing all the photons that were detected without contributing energy in other interactions, such as Compton or Rayleigh scattering. The I min function is supposed to reach the minimum value, when the scatter correction method will finally create

3 406 ARTICLE IN PRESS N.G. Sakellios et al. / Nuclear Instruments and Methods in Physics Research A 569 (2006) an image close to the NULL events recorded in GATE. The subtraction of these two images should give a value close to zero; depending on the step we have chosen for the scatter parameters wf i,pc i and ww i. of the pot. The size and shape of each region is shown in Fig. 1. This phantom was imaged for 50 s and counts were detected. 3. Phantom experiments Flood source: The flood source is a cylinder, 4 cm inner diameter and 10 cm in height. The cylinder was filled with 99m Tc solution and was placed close to the head of the camera. The simulation was running for 50 s and counts were detected. The experiment is used in order to estimate the scatter contribution and calculate its parameters; the weighting factors, the position and the width of each one of the low energy windows. The results will be applied to the rest of the phantoms. Breast phantom: The breast phantom consists of two hot quantities of a 99m Tc solution, both 0.5 ml in volume, placed inside a pot, 8 cm in diameter and 10 cm in height, containing 99m Tc solution. The activity ratios of the spot to the background were 10:1 (S1) and 5:1 (S2), respectively. Furthermore, the initial activity for the two spots was and Bq, respectively. The detector was placed at a 2.5 cm distance from the top of the pot. The hot phantom was imaged for 50 s and counts were collected. myphantom: The phantom consists of a large cylinder with four cold regions. The inner diameter and the height of the cylinder are 4 and 8 cm, respectively. The depth of each cold region varies from 2, 4, 6 and 8 cm from the top Table 1 The results of the minimization method Compton windows Factor wf i Window width (kev) Win_1 Low 1of Win_2 Low 1of Low 2of Fig. 2. Images projections: the photopeak 20% window, the NULL window 20% and the correction image using two low energy windows. Fig. 1. myphantom geometry.

4 N.G. Sakellios et al. / Nuclear Instruments and Methods in Physics Research A 569 (2006) Results Minimization with flood source: Using the flood source, and Eq. (7) we apply the minimization method, both for Fig. 3. myphantom Study: The Compton Low Energy Windows kev and kev Low Energy Windows. one (1) and two (2) low energy windows. For both cases, the scatter parameters were calculated. In addition, the Line Spread Functions, in air and in water environment have been used. The results of the minimization method are shown in Table 1. It is clearly seen in Fig. 2 that the resulting image of the minimization method has approximated the NULL photon image, which is supposed to be the ideal one. Scatter component: The scatter distribution within the photopeak window was calculated in both phantoms using one or two low energy windows, according to the results presented in Table 1. In the following figure, the scatter component is presented for the cold phantom study (myphantom). Fig. 3 shows that the scatter contribution is preserving the geometry of the phantom. This property was observed in all different low-energy windows. Image contrast: Image projections have been drawn. It can be seen in the plots of Fig. 4 that the scatter correction method provides an improvement in contrast, in both phantoms and in both cases, using one or two low energy windows. Signal/noise ratio: Signal-to-Noise Ratio (SNR) was calculated for both studies with and without the scatter correction methods. It was defined as the net signal divided by the noise level. Count density was calculated from ROIs, 3 3 pixels large. Background count density was calculated from ROIs, 2 5 pixels large, drawn around the hot and cold lesions, respectively. The values of the SNR Fig. 4. Breast phantom: (A) photopeak image 20%, (B) corrected image. myphantom: (C) photopeak image 20%, (D) corrected image.

5 408 ARTICLE IN PRESS N.G. Sakellios et al. / Nuclear Instruments and Methods in Physics Research A 569 (2006) Table 2 Signal/noise ratio Total 20% One low Two low Breast S1(10:1) S2(5:1) myphantom R1(2 cm) R2(4 cm) Discussion The use of one or more low-energy windows has improved image contrast. The increase in SNR and the reduction of noise followed by the rejection of scattered events can be clearly observed. Comparison between the total images and the corrected ones can be observed in Fig. 5. The scatter events subtraction in both studies provides better quality in the reconstructed images. MEWST seems to provide more robust results and this technique will be further investigated as well as the Dual Photopeak [13]. References Fig. 5. Breast Phantom: (A) photopeak image 20%, (B) corrected image. myphantom: (C) photopeak image 20%, (D) corrected image. are reported in Table 2, for the breast phantom study and for only two out of four cold regions of myphantom study. An increase in SNR was observed after the correction with both techniques. Using one or more Compton energy windows, the gain in the contrast of the image is greater than the increase of noise due to scatter rejection. [1] R.J. Jaszczak, K.L. Greer, C.E. Floyd, C.C. Harris, R.E. Coleman, J. Nucl. Med. 29 (1984) 893. [2] A.E. Todd-Pokropek, G. Clarke, R. Marsh, Preprocessing of SPECT data as a precursor for attenuation correction, in: F. Deckonic (Ed.), Information Processing in Medical Imaging, Martinus Nijhoff Publishers, Bruxelles, 1983, pp [3] P. Msaki, B. Axelsson, A. Israelsson, S.A. Larson, An improved scatter correction technique in SPECT using point spread scatter function, Presented at the International Conference on Medical Biology and Engineering and International Conference on Medical Physics, Espoo, Finland, [4] C.A. Lowry, D.N. Taylor, Br. J. Radiol. 59 (1986) 728. [5] C.A. Lowry, D.N. Taylor, R.S. Holt, M.J. Cooper, J.A.R. McIntosh, Single photon emission computed tomography: correction for Compton scatterd photons, in: Conference Abstracts, 43rd Annual Conference on HPA, Coventry, England, [6] C.E. Floyd Jr., R.J. Jaszczack, K.L. Greer, R.E. Coleman, J. Nucl. Med. 26 (1985) 403. [7] S.D. Egbert, R.S. May, IEEE Trans. Nucl. Sci. NS-27 (1980) 543. [8] D.J. Wagett, B.C. Wilson, Br. J. Radiol. 51 (1977) [9] G. Santin, et al., IEEE Trans. Nucl. Sci. NS-50 (2003) [10] G.K. Loudos, et al., Appl. Radiat. Isot. 58 (4) (2003) 501. [11] G. Loudos, et al., A modification of the dual energy window subtraction method for scatter compensation in pixelized scintillators for SPECT, in: IInd International Conference on Imaging Techniques in Biomedical Sciences ITBS2003, May 2003, Athens and Milos Island, Greece. [12] R. Marsh, Scatter correction for single photon emission computed tomography images, Thesis, University College London, London, UK, [13] M.A. King, G.J. Hademenos, S.J. Glick, J. Nucl. Med. 33 (4) (1992) 605.

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

VALIDATION OF GATE MONTE CARLO SIMULATIONS FOR INDIUM 111 IMAGING

VALIDATION OF GATE MONTE CARLO SIMULATIONS FOR INDIUM 111 IMAGING VALIDATION OF GATE MONTE CARLO SIMULATIONS FOR INDIUM IMAGING Karine Assié, Isabelle Gardin, Pierre Véra, and Irène Buvat Abstract-- Accurate quantification of indium ( In) Single Photon Emission Computed

More information

3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction

3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction 3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction Yuni K. Dewaraja, Member, IEEE, Michael Ljungberg, Member, IEEE, and Jeffrey A. Fessler, Member, IEEE Abstract- we

More information

Nuclear Instruments and Methods in Physics Research A

Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 621 (2010) 590 594 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

An educational tool for demonstrating the TOF-PET technique

An educational tool for demonstrating the TOF-PET technique Nuclear Instruments and Methods in Physics Research A 471 (2001) 200 204 An educational tool for demonstrating the TOF-PET technique T.Bȧack a, *, J. Cederkȧall a, B. Cederwall a, A. Johnson a, A. Kerek

More information

Comparison of 3D PET data bootstrap resampling methods for numerical observer studies

Comparison of 3D PET data bootstrap resampling methods for numerical observer studies 2005 IEEE Nuclear Science Symposium Conference Record M07-218 Comparison of 3D PET data bootstrap resampling methods for numerical observer studies C. Lartizien, I. Buvat Abstract Bootstrap methods have

More information

ATTENUATION CORRECTION IN SPECT DURING IMAGE RECONSTRUCTION USING INVERSE MONTE CARLO METHOD A SIMULATION STUDY *

ATTENUATION CORRECTION IN SPECT DURING IMAGE RECONSTRUCTION USING INVERSE MONTE CARLO METHOD A SIMULATION STUDY * Romanian Reports in Physics, Vol. 66, No. 1, P. 200 211, 2014 ATTENUATION CORRECTION IN SPECT DURING IMAGE RECONSTRUCTION USING INVERSE MONTE CARLO METHOD A SIMULATION STUDY * S. AHMADI 1, H. RAJABI 2,

More information

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 181 3-D Monte Carlo-Based Scatter Compensation in Quantitative I-131 SPECT Reconstruction Yuni K. Dewaraja, Member, IEEE, Michael Ljungberg,

More information

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce

More information

Monte-Carlo-Based Scatter Correction for Quantitative SPECT Reconstruction

Monte-Carlo-Based Scatter Correction for Quantitative SPECT Reconstruction Monte-Carlo-Based Scatter Correction for Quantitative SPECT Reconstruction Realization and Evaluation Rolf Bippus 1, Andreas Goedicke 1, Henrik Botterweck 2 1 Philips Research Laboratories, Aachen 2 Fachhochschule

More information

MEDICAL IMAGE ANALYSIS

MEDICAL IMAGE ANALYSIS SECOND EDITION MEDICAL IMAGE ANALYSIS ATAM P. DHAWAN g, A B IEEE Engineering in Medicine and Biology Society, Sponsor IEEE Press Series in Biomedical Engineering Metin Akay, Series Editor +IEEE IEEE PRESS

More information

Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy

Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy Physics Department, Faculty of Applied Science,Umm Al-Qura

More information

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and

More information

Master of Science Thesis. Activity quantification of planar gamma camera images

Master of Science Thesis. Activity quantification of planar gamma camera images Master of Science Thesis Activity quantification of planar gamma camera images Marie Sydoff Supervisor: Sigrid Leide-Svegborn The work has been performed at Department of Radiation Physics Malmö University

More information

Introduction to Positron Emission Tomography

Introduction to Positron Emission Tomography Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 5 SPECT, PET Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Beam Attenuation Grid Based Scatter Correction Algorithm for More Info at Open Access Database

More information

Planar and SPECT Monte Carlo acceleration using a variance reduction technique in I 131 imaging

Planar and SPECT Monte Carlo acceleration using a variance reduction technique in I 131 imaging Iran. J. Radiat. Res., 2007; 4 (4): 175-182 Planar and SPECT Monte Carlo acceleration using a variance reduction technique in I 131 imaging H.R. Khosravi 1, 2*, S. Sarkar 1, 3,A.Takavar 1, 4, M. Saghari

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives 1. Be familiar with recommendations

More information

Limitations in the PHOTON Monte Carlo gamma transport code

Limitations in the PHOTON Monte Carlo gamma transport code Nuclear Instruments and Methods in Physics Research A 480 (2002) 729 733 Limitations in the PHOTON Monte Carlo gamma transport code I. Orion a, L. Wielopolski b, * a St. Luke s/roosevelt Hospital, Columbia

More information

Small Angle Gamma Ray Scatter: What Is The Impact On Image Quality

Small Angle Gamma Ray Scatter: What Is The Impact On Image Quality ISPUB.COM The Internet Journal of Medical Technology Volume 4 Number 2 Small Angle Gamma Ray Scatter: What Is The Impact On Image Quality G CUrrie, J Wheat Citation G CUrrie, J Wheat.. The Internet Journal

More information

Application of MCNP Code in Shielding Design for Radioactive Sources

Application of MCNP Code in Shielding Design for Radioactive Sources Application of MCNP Code in Shielding Design for Radioactive Sources Ibrahim A. Alrammah Abstract This paper presents three tasks: Task 1 explores: the detected number of as a function of polythene moderator

More information

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation 26 IEEE Nuclear Science Symposium Conference Record M6-349 Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation Habib Zaidi, Senior Member, IEEE and Mohammad

More information

A publicly accessible Monte Carlo database for validation purposes in emission tomography

A publicly accessible Monte Carlo database for validation purposes in emission tomography Short communication A publicly accessible Monte Carlo database for validation purposes in emission tomography I. Castiglioni 1, I. Buvat 2, G. Rizzo 1, M. C. Gilardi 1, J. Feuardent 2, F. Fazio 1 1 IBFM-CNR,

More information

Simulations in emission tomography using GATE

Simulations in emission tomography using GATE Simulations in emission tomography using GATE Irène Buvat buvat@imed.jussieu.fr Laboratory of Functional Imaging, U678 INSERM, Paris, France Outline Emission tomography and need for simulations GATE short

More information

Validation of the Small Animal Biospace Gamma Imager Model Using GATE Monte Carlo Simulations on the Grid

Validation of the Small Animal Biospace Gamma Imager Model Using GATE Monte Carlo Simulations on the Grid Validation of the Small Animal Biospace Gamma Imager Model Using GATE Monte Carlo Simulations on the Grid Joe Aoun, Vincent Breton, L. Desbat, B. Bzeznik, M. Leabadand, J. Dimastromatteo To cite this version:

More information

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam Home Search Collections Journals About Contact us My IOPscience Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam projections This content has been downloaded from IOPscience.

More information

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO Abstract C.S. Levin, Y-C Tai, E.J. Hoffman, M. Dahlbom, T.H. Farquhar UCLA School of Medicine Division

More information

Medical Imaging BMEN Spring 2016

Medical Imaging BMEN Spring 2016 Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,

More information

James R Halama, PhD Loyola University Medical Center

James R Halama, PhD Loyola University Medical Center James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives Be familiar with the tests recommended

More information

SPECT (single photon emission computed tomography)

SPECT (single photon emission computed tomography) 2628 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 5, OCTOBER 2006 Detector Blurring and Detector Sensitivity Compensation for a Spinning Slat Collimator Gengsheng L. Zeng, Senior Member, IEEE Abstract

More information

Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors

Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors Elena Marimon 1,2*, Hammadi Nait-Charif 1, Asmar Khan 2, Philip A. Marsden 3, Oliver Diaz 4 1 Centre for

More information

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques IMAGING A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques Lai K. Leong, Randall L. Kruger, and Michael K. O Connor Section of Nuclear Medicine, Department

More information

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 3, JUNE 2003 405 Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging Edward J. Soares, Kevin W. Germino,

More information

Iterative SPECT reconstruction with 3D detector response

Iterative SPECT reconstruction with 3D detector response Iterative SPECT reconstruction with 3D detector response Jeffrey A. Fessler and Anastasia Yendiki COMMUNICATIONS & SIGNAL PROCESSING LABORATORY Department of Electrical Engineering and Computer Science

More information

Continuation Format Page

Continuation Format Page C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar

More information

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide 2005 IEEE Nuclear Science Symposium Conference Record M04-8 Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide J. S. Karp, Senior Member, IEEE, A. Kuhn, Member, IEEE, A. E. Perkins,

More information

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar Doctoral Thesis Time-of-Flight Time-of-Flight Positron Positron Emission Emission Tomography Tomography Using Using Cherenkov Cherenkov Radiation Radiation Rok Dolenec Supervisor: Prof. Dr. Samo Korpar

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy.

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy. Modeling the Distance-Dependent Blurring in Transmission Imaging in the Ordered-Subset Transmission (OSTR) Algorithm by Using an Unmatched Projector/Backprojector Pair B. Feng, Member, IEEE, M. A. King,

More information

Scatter correction improvement based on the Convolution Subtraction Technique in SPECT imaging

Scatter correction improvement based on the Convolution Subtraction Technique in SPECT imaging Vol.2, Issue.3, May-June 2012 pp-1414-1418 ISSN: 2249-6645 Scatter correction improvement based on the Convolution Subtraction Technique in SPECT imaging Mohsen Zand 1 1 (Department of Computer, Islamic

More information

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Effects of the difference in tube voltage of the CT scanner on. dose calculation Effects of the difference in tube voltage of the CT scanner on dose calculation Dong Joo Rhee, Sung-woo Kim, Dong Hyeok Jeong Medical and Radiological Physics Laboratory, Dongnam Institute of Radiological

More information

Advanced Scatter Correction for Quantitative Cardiac SPECT

Advanced Scatter Correction for Quantitative Cardiac SPECT Advanced Scatter Correction for Quantitative Cardiac SPECT Frederik J. Beekman PhD Molecular Image Science Laboratory University Medical Center Utrecht The Netherlands Outline What is scatter? Scatter

More information

Emission Computed Tomography Notes

Emission Computed Tomography Notes Noll (24) ECT Notes: Page 1 Emission Computed Tomography Notes Introduction Emission computed tomography (ECT) is the CT applied to nuclear medicine. There are two varieties of ECT: 1. SPECT single-photon

More information

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.

More information

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction In this lab exercise, you will investigate the linearity of the DeskCAT scanner by making measurements

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Introduction to Emission Tomography

Introduction to Emission Tomography Introduction to Emission Tomography Gamma Camera Planar Imaging Robert Miyaoka, PhD University of Washington Department of Radiology rmiyaoka@u.washington.edu Gamma Camera: - collimator - detector (crystal

More information

Extremely Fast Detector for 511 kev Gamma

Extremely Fast Detector for 511 kev Gamma Extremely Fast Detector for 511 kev Gamma V. Sharyy, D. Yvon, G. Tauzin, E.Delagnes, Ph. Abbon, J P. Bard, M. Kebbiri, M. Alokhina, C. Canot IRFU, CEA D. Breton, J. Maalmi LAL,IN2P3 Journée 2015 du Labex

More information

Customizable and Advanced Software for Tomographic Reconstruction

Customizable and Advanced Software for Tomographic Reconstruction Customizable and Advanced Software for Tomographic Reconstruction 1 What is CASToR? Open source toolkit for 4D emission (PET/SPECT) and transmission (CT) tomographic reconstruction Focus on generic, modular

More information

Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector

Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector K. Karafasoulis, K. Zachariadou, S. Seferlis, I. Papadakis, D. Loukas, C. Lambropoulos, C. Potiriadis Abstract Spectroscopic

More information

Improvement of contrast using reconstruction of 3D Image by PET /CT combination system

Improvement of contrast using reconstruction of 3D Image by PET /CT combination system Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(1):285-290 ISSN: 0976-8610 CODEN (USA): AASRFC Improvement of contrast using reconstruction of 3D Image

More information

Validation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation camera dedicated to small-animal imaging

Validation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation camera dedicated to small-animal imaging INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 49 (24) 271 285 PHYSICS IN MEDICINE AND BIOLOGY PII: S31-9155(4)68367-9 Validation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation

More information

3-D PET Scatter Correction

3-D PET Scatter Correction Investigation of Accelerated Monte Carlo Techniques for PET Simulation and 3-D PET Scatter Correction C.H. Holdsworth, Student Member, IEEE, C.S. Levin", Member, IEEE, T.H. Farquhar, Student Member, IEEE,

More information

Design and Fabrication of Kidney Phantoms for Internal Radiation Dosimetry Using 3D Printing Technology

Design and Fabrication of Kidney Phantoms for Internal Radiation Dosimetry Using 3D Printing Technology Design and Fabrication of Kidney Phantoms for Internal Radiation Dosimetry Using 3D Printing Technology Johannes Tran-Gia, PhD NPL MEMPHYS Workshop Applications of 3D Printing for Medical Phantoms Klinik

More information

SINGLE-PHOTON emission computed tomography

SINGLE-PHOTON emission computed tomography 1458 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 4, AUGUST 2012 SPECT Imaging With Resolution Recovery Andrei V. Bronnikov Abstract Single-photon emission computed tomography (SPECT) is a method

More information

Mathematical methods and simulations tools useful in medical radiation physics

Mathematical methods and simulations tools useful in medical radiation physics Mathematical methods and simulations tools useful in medical radiation physics Michael Ljungberg, professor Department of Medical Radiation Physics Lund University SE-221 85 Lund, Sweden Major topic 1:

More information

Addition of Noise by Scatter Correction Methods in PVI

Addition of Noise by Scatter Correction Methods in PVI IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 41, NO. 4, AUGUST 1994 1551 Addition of Noise by Scatter Correction Methods in PVI.J. Scott Barneyt, Ronald Harropl and M. Stella Atkinsi t Div. of Nuclear Medicine,

More information

Reconstruction from Projections

Reconstruction from Projections Reconstruction from Projections M.C. Villa Uriol Computational Imaging Lab email: cruz.villa@upf.edu web: http://www.cilab.upf.edu Based on SPECT reconstruction Martin Šámal Charles University Prague,

More information

Positron. MillenniumVG. Emission Tomography Imaging with the. GE Medical Systems

Positron. MillenniumVG. Emission Tomography Imaging with the. GE Medical Systems Positron Emission Tomography Imaging with the MillenniumVG GE Medical Systems Table of Contents Introduction... 3 PET Imaging With Gamma Cameras PET Overview... 4 Coincidence Detection on Gamma Cameras...

More information

218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997

218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997 218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997 Compton Scatter and X-ray Crosstalk and the Use of Very Thin Intercrystal Septa in High-Resolution PET Detectors Craig S. Levin, Member,

More information

S rect distortions in single photon emission computed

S rect distortions in single photon emission computed IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 41, NO. 6, DECEMBER 1994 2807 A Rotating and Warping Projector/Backprojector for Fan-Beam and Cone-Beam Iterative Algorithm G. L. Zeng, Y.-L. Hsieh, and G. T.

More information

SPECT reconstruction

SPECT reconstruction Regional Training Workshop Advanced Image Processing of SPECT Studies Tygerberg Hospital, 19-23 April 2004 SPECT reconstruction Martin Šámal Charles University Prague, Czech Republic samal@cesnet.cz Tomography

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel Proc. Of SPIE Vol 0626, Application of Optical Instrumentation in Medicine XIV and Picture Archiving and Communication Systems(PACS IV) for Medical Applications IV, ed. S J Dwyer/R H Schneider (Jan 1986)

More information

ACOMPTON-SCATTERING gamma camera records two

ACOMPTON-SCATTERING gamma camera records two IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 5, OCTOBER 2006 2787 Filtered Back-Projection in 4 Compton Imaging With a Single 3D Position Sensitive CdZnTe Detector Dan Xu, Student Member, IEEE, and

More information

PET Quantification using STIR

PET Quantification using STIR PET Quantification using STIR STIR User s Meeting Charalampos Tsoumpas, PhD King s College London Hammersmith Imanet 1 PET Quantification Image elements should correspond to concentration of the injected

More information

Performance Evaluation of radionuclide imaging systems

Performance Evaluation of radionuclide imaging systems Performance Evaluation of radionuclide imaging systems Nicolas A. Karakatsanis STIR Users meeting IEEE Nuclear Science Symposium and Medical Imaging Conference 2009 Orlando, FL, USA Geant4 Application

More information

Recognition and Measurement of Small Defects in ICT Testing

Recognition and Measurement of Small Defects in ICT Testing 19 th World Conference on Non-Destructive Testing 2016 Recognition and Measurement of Small Defects in ICT Testing Guo ZHIMIN, Ni PEIJUN, Zhang WEIGUO, Qi ZICHENG Inner Mongolia Metallic Materials Research

More information

arxiv: v2 [cond-mat.mtrl-sci] 5 Jan 2010

arxiv: v2 [cond-mat.mtrl-sci] 5 Jan 2010 Gamma scattering scanning of concrete block for detection of voids. Shivaramu 1, Arijit Bose 2 and M. Margret 1 1 Radiological Safety Division, Safety Group, IGCAR, Kalpakaam - 63 12 (India) 2 Chennai

More information

The use of the modulation transfer function for comparison and image quality assessment of commercially available hybrid PET-CT scanners

The use of the modulation transfer function for comparison and image quality assessment of commercially available hybrid PET-CT scanners The use of the modulation transfer function for comparison and image quality assessment of commercially available hybrid PET-CT scanners Poster No.: C-2943 Congress: ECR 2010 Type: Topic: Scientific Exhibit

More information

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Sensors & Transducers 2013 by IFSA http://www.sensorsportal.com Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Shinji Ohyama, Masato Mukouyama Graduate School

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Effects of system geometry and other physical factors on photon sensitivity of high-resolution

Effects of system geometry and other physical factors on photon sensitivity of high-resolution Home Search Collections Journals About Contact us My IOPscience Effects of system geometry and other physical factors on photon sensitivity of high-resolution positron emission tomography This article

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Improved activity estimation with MC-JOSEM versus TEW-JOSEM in 111 In SPECT

Improved activity estimation with MC-JOSEM versus TEW-JOSEM in 111 In SPECT Improved activity estimation with MC-JOSEM versus TEW-JOSEM in 111 In SPECT Jinsong Ouyang, a Georges El Fakhri, and Stephen C. Moore Department of Radiology, Harvard Medical School and Brigham and Women

More information

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial 8//0 AAPM0 Scientific Symposium: Emerging and New Generation PET: Instrumentation, Technology, Characteristics and Clinical Practice Aug Wednesday 0:4am :pm Solid State Digital Photon Counting PET/CT Instrumentation

More information

Constructing System Matrices for SPECT Simulations and Reconstructions

Constructing System Matrices for SPECT Simulations and Reconstructions Constructing System Matrices for SPECT Simulations and Reconstructions Nirantha Balagopal April 28th, 2017 M.S. Report The University of Arizona College of Optical Sciences 1 Acknowledgement I would like

More information

Validation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation camera dedicated to small animal imaging

Validation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation camera dedicated to small animal imaging Validation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation camera dedicated to small animal imaging D Lazaro 1, I Buvat 2, G Loudos 3,4, D Strul 5, G Santin 6, N Giokaris

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste

An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste S.C. Kane 1, S. Croft 1, P. McClay 1, W.F. Mueller

More information

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Artifact Mitigation in High Energy CT via Monte Carlo Simulation PIERS ONLINE, VOL. 7, NO. 8, 11 791 Artifact Mitigation in High Energy CT via Monte Carlo Simulation Xuemin Jin and Robert Y. Levine Spectral Sciences, Inc., USA Abstract The high energy (< 15 MeV) incident

More information

Quantitative imaging for clinical dosimetry

Quantitative imaging for clinical dosimetry Quantitative imaging for clinical dosimetry Irène Buvat Laboratoire d Imagerie Fonctionnelle U678 INSERM - UPMC CHU Pitié-Salpêtrière, Paris buvat@imed.jussieu.fr http://www.guillemet.org/irene Methodology

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

3D X-ray Laminography with CMOS Image Sensor Using a Projection Method for Reconstruction of Arbitrary Cross-sectional Images

3D X-ray Laminography with CMOS Image Sensor Using a Projection Method for Reconstruction of Arbitrary Cross-sectional Images Ke Engineering Materials Vols. 270-273 (2004) pp. 192-197 online at http://www.scientific.net (2004) Trans Tech Publications, Switzerland Online available since 2004/08/15 Citation & Copright (to be inserted

More information

Image Reconstruction Methods for Dedicated Nuclear Medicine Detectors for Prostate Imaging

Image Reconstruction Methods for Dedicated Nuclear Medicine Detectors for Prostate Imaging Image Reconstruction Methods for Dedicated Nuclear Medicine Detectors for Prostate Imaging Mark F. Smith, Ph.D. Detector and Imaging Group, Physics Division Thomas Jefferson National Accelerator Facility

More information

Modeling and Incorporation of System Response Functions in 3D Whole Body PET

Modeling and Incorporation of System Response Functions in 3D Whole Body PET Modeling and Incorporation of System Response Functions in 3D Whole Body PET Adam M. Alessio, Member IEEE, Paul E. Kinahan, Senior Member IEEE, and Thomas K. Lewellen, Senior Member IEEE University of

More information

Scatter Correction Methods in Dimensional CT

Scatter Correction Methods in Dimensional CT Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University

More information

By choosing to view this document, you agree to all provisions of the copyright laws protecting it.

By choosing to view this document, you agree to all provisions of the copyright laws protecting it. Copyright 2007 IEEE. Reprinted from 2007 IEEE Nuclear Science Symposium Conference Record, Paper M13-5, p. 2838-41. This material is posted here with permission of the IEEE. Internal or personal use of

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

Slab-by-Slab Blurring Model for Geometric Point Response Correction and Attenuation Correction Using Iterative Reconstruction Algorithms

Slab-by-Slab Blurring Model for Geometric Point Response Correction and Attenuation Correction Using Iterative Reconstruction Algorithms 2168 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL 45, NO. 4, AUGUST 1998 Slab-by-Slab Blurring Model for Geometric Point Response Correction and Attenuation Correction Using Iterative Reconstruction Algorithms

More information

Fits you like no other

Fits you like no other Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room

More information

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct J. DHAENE* 1, E. PAUWELS 1, T. DE SCHRYVER 1, A. DE MUYNCK 1, M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics

More information

ISO ISO ISO OHSAS ISO

ISO ISO ISO OHSAS ISO ISO 9001 ISO 13485 ISO 14001 OHSAS 18001 ISO 27001 Pro-NM Performance 08-101 - standard version 08-103 - version with the PET Lid Phantom for NM and PET systems performance evaluation (collimator, artifacts,

More information

Evaluation of a Combined Array-Planar Crystal for Gamma-ray Scintillation Imagers

Evaluation of a Combined Array-Planar Crystal for Gamma-ray Scintillation Imagers Evaluation of a Combined Array-Planar Crystal for Gamma-ray Scintillation Imagers R.Pani 1, R.Scafè 2, R.Pellegrini 1, A.Soluri 3, N.Burgio 2, C.Ciavola 2, M.N.Cinti 1, L.Indovina 1, G.De Vincentis 1,

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

A new approach to scatter correction in SPECT images based on Klein_Nishina equation

A new approach to scatter correction in SPECT images based on Klein_Nishina equation A new approach to scatter correction in SPECT images based on Klein_Nishina equation Mohsen Hajizadeh Saffar 1, Shabnam Oloomi 2, Peter Knoll 3, Hadi Taleshi 4 Original Article 1 Medical Physics Research

More information