Nuclear Instruments and Methods in Physics Research A

Size: px
Start display at page:

Download "Nuclear Instruments and Methods in Physics Research A"

Transcription

1 Nuclear Instruments and Methods in Physics Research A 621 (2010) Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: Preliminary experimental results of a quasi-monolithic detector with DOI capability for a small animal PET Yong Hyun Chung a,b,n, Cheol-Ha Baek a,b, Seung-Jae Lee a,b, Key Jo Hong c, Ji Hoon Kang d, Yong Choi c a Department of Radiological Science, College of Health Science, Yonsei University, 234 Meaji, Heungup, Wonju, Kangwon-Do , Republic of Korea b Institute of Health Science, Yonsei University, Wonju , Republic of Korea c Department of Electric Engineering, Sogang University, Seoul , Republic of Korea d Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul , Republic of Korea article info Article history: Received 16 January 2010 Received in revised form 5 April 2010 Accepted 12 April 2010 Available online 20 April 2010 Keywords: PET Maximum likelihood position estimation Monolithic crystal Depth of interaction abstract In our previous work, the new detector module with depth of interaction (DOI) based on a quasimonolithic crystal array and maximum-likelihood position-estimation (MLPE) algorithm was designed and the methodology for 3D event positioning in the detector was established by using Monte Carlo simulation. The aim of this study was to evaluate the performance of our detector module experimentally. The detector module consisted of a 1D array of eight LYSO crystals of 20.0(X) 2.0(Y) 10.0(Z) mm 3, optically coupled to a Hamamatsu H7546B position sensitive photomultiplier tube (PSPMT). The PSPMT was read out using a resistive charge divider, which multiplexes 64 anodes into 8(X)+8(Y) channels. Gaussian-based MLPE methods have been implemented using experimentally measured detector response functions (DRFs). The results demonstrated that the detector module could identify the position of the gamma ray interaction inside the crystal in all three directions. The new DOI detector for a small animal PET was developed and verified experimentally with a view towards achieving high sensitivity as well as high and uniform radial resolution. & 2010 Elsevier B.V. All rights reserved. 1. Introduction The demand for a high performance positron emission tomography (PET) is motivated by the need to perform molecular imaging of small animal models. However, thick scintillation crystals for high sensitivity in PET scanners cause radial elongation or parallax errors. To correct these problems and to improve the resolution uniformity across the field of view (FOV), there has been a growing interest in the design of small animal PET with the depth-of-interaction (DOI) determination capability. Most of developed DOI detector modules used matrices of small scintillating crystals, but the detection efficiency in such designs is reduced by the dead space between crystal elements [1 9]. As an alternative, monolithic crystal detectors with statistical positioning algorithms have been proposed. These detectors use several cubic centimeters of scintillating material coupled on one or more sides to photo sensor. These techniques require look-up tables (LUTs) containing information on the light distribution of scintillation events in the crystal as a function of source position, which are generated through computer simulation or machine training corresponding to a specific angle of incidence. Our group has introduced a new DOI detector concept based on an array of quasi-monolithic crystals and multi-channel photosensors used with a maximum-likelihood position-estimation (MLPE) algorithm [10 12]. The main advantage of the quasimonolithic detector is that the dependence of the light response on an event location can be measured experimentally, because the crystal elements are separated from each other in the axial direction and are monolithic in the trans-axial direction. In our previous work, we evaluated the performances through DETECT2000 simulation and the results showed that the detector could determine the position of the gamma events inside the crystal with a resolution of 2.0 mm in all three dimensions. The purpose of this study is to develop the proposed detector module and to evaluate the performances of it experimentally. 2. Materials and Methods 2.1. Quasi-monolithic detector n Corresponding author at: Department of Radiological Science, College of Health Science, Yonsei University, 234 Meaji, Heungup, Wonju, Kangwon-Do , Republic of Korea. Tel.: ; fax: address: ychung@yonsei.ac.kr (Y. Hyun Chung). Quasi-monolithic detector module employs a volumetric slab configuration, rather than a matrix of pixelated crystal elements. The module consists of a 1D array of eight LYSO crystals with /$ - see front matter & 2010 Elsevier B.V. All rights reserved. doi: /j.nima

2 Y. Hyun Chung et al. / Nuclear Instruments and Methods in Physics Research A 621 (2010) (X) 2.0(Y) 10.0(Z) mm 3 size. The crystal elements are discrete in the Y direction and are monolithic in the X direction, as shown in Fig. 1. The mm 2 bottom surface of the crystal is polished and directly coupled to a position-sensitive PMT (PSPMT) or an array of silicon multipliers (SiPMs). The large sides are wrapped with Teflon reflector so that the crystals are separated from each other in the Y direction. To reduce the degradation of the resolution near the edge of the crystal, the short sides are painted with absorbent material [10]. All 64 channels of the PMT are acquired for each event. To compute the event positions in the Y direction, eight outputs on the same row are summed, and the eight summed signals along the Y direction are compared to each other. After determining the Y position of the event, the MLPE algorithm is employed to extract the X and Z positions using eight output signals in the selected monolithic crystal MLPE algorithm The MLPE method uses the likelihood between the detector response function (DRF) of each PMT channel and the event position. Because of the stochastic nature of signal formation in a Fig. 1. (a) Schematic diagram of a quasi-monolithic crystal module and (b) the simulated apparatus for training. detector, the probability of measuring a given set of PMT output signals M¼M 1, M 2, y, M 8 for a given event position x in the crystal is described by the probability distribution as Pr(M9x). Assuming that the noise process in the detector response follows a Gaussian distribution, the likelihood is given by Y N Pr M9x ¼! 1 pffiffiffiffiffiffi exp ðm i m i ðxþþ 2 s i ðxþ 2p 2s 2 i ðxþ where m i (x) is the mean and s i (x) is the standard deviation of the ith output of PMT as a function of event position. Performing the maximization with respect to log likelihood, the resulting approximation is X n lnpr M i 9x ¼ ðm i m i ðxþþ 2 2s 2 i ðxþ þ Xn lns i ðxþ The MLPE solution is achieved by minimizing the value between parenthesis in Eq. (2) Experimental setup for training DRFs of each crystal element were determined by experimental training. The single LYSO crystal of size 20.0(X) 2.0(Y) 10.0(Z) mm 3 was optically coupled to a Hamamatsu H7546B PSPMT. The PSPMT was read out using a resistive charge divider, which multiplexes 64 anodes into 8(X)+8(Y) channels. Fig. 2 is a diagram of the experimental apparatus employed. During the experimental training, only one crystal element was coupled to PSPMT to generate gamma events at any location. Fig. 3 shows the DRF training system consisting of a collimated source, motorized x z stage, LYSO crystal, PSPMT, signal processing electronics and data acquisition system. F-18 source was housed in a 5 cm thick lead block and collimated into a pencil beam 1 mm in diameter. To characterize DRFs, data were acquired on 4 5 evenly spaced points with 2.0 mm spacing in the X and Z directions using a computer controlled stepper motor stage and collimated F-18 source. Then DRFs were generated using Gaussian fitting and bicubic spline interpolation.! ð1þ ð2þ Fig. 2. Block diagram of the circuitry of a detector module. Fig. 3. Detector and training system consist of collimated source, motorized x z stage, LYSO crystal, PSPMT, signal processing electronics and data acquisition board.

3 592 Y. Hyun Chung et al. / Nuclear Instruments and Methods in Physics Research A 621 (2010) Fig. 4. The point source image and global energy spectrum for one training position. Fig. 5. Energy spectra of eight X signals were plotted and the mean and standard deviation were extracted to generate LUT by Gaussian fitting Performance measurements After generating DRFs for a crystal element, the point source imaging was performed. To measure the intrinsic spatial resolution of the detector in the X direction, the image of the collimated source was obtained and the full-width at half-maximum of the profile was calculated. The measured DOI (Z-axis) positions were compared to the true positions of the collimated source. The misclassification rates, defined as the number of mis-assigned events in Z direction divided by the total number of events, were calculated for each source position. 3. Results Fig. 4 illustrates the point source image and global energy spectrum for one training position. Energy spectra of eight X signals were plotted and the mean and standard deviation were extracted to generate LUT by Gaussian fitting as shown in Fig. 5.

4 Y. Hyun Chung et al. / Nuclear Instruments and Methods in Physics Research A 621 (2010) Fig.6. The measured images of point sources located at nine different positions with 2.0 mm spacing. The numbers represent the positioning accuracies. Fig.6 represents the images of true source positions and the acquired images by MLPE algorithm. The positioning accuracies in the X and Z directions were calculated. The positioning accuracies for each source position were 91.7%, 97.8%, 99.8%, 95.5%, 98.6%, 99.7%, 96.8%, 91.3% and 99.9%, respectively, and the average is 96.8% in the X direction. The average misclassification rates in the DOI direction was 45% in the depth direction. 4. Discussion and conclusions A new DOI detector based on quasi-monolithic LYSO crystal and an MLPE algorithm was developed and tested. The source positions were accurately identified with 2.0 mm resolution in the X and Y directions, but the positioning accuracy in the depth direction was about 55%. Since the quasi-monolithic crystal elements were continuous in the X direction and separated in the Y direction, the spatial resolution in the Y direction was geometrically limited to 2.0 mm in this detector module. However, the spatial resolution can be improved by increasing the number of LUT sampling points and by reducing the physical crystal size in Y direction. The main advantage of the proposed detector module is that the LUTs needed for position estimates can be generated experimentally. For a monolithic crystal, generating a gamma event at a specific position is impossible, and for a pixelated crystal, the light distributions from events at different depths are indistinguishable. Also, the correction method for the nonuniform signals from PSPMT is not needed in this design because MLPE involves the use of LUTs to map from measured PSPMT signals to position estimates. Other advantages of this design include simple fabrication, higher packing fraction and high sensitivity. The positioning accuracy is affected by the light distribution in the crystal, which is largely varied by the surface conditions, especially for the DOI direction. In our previous simulation [10], the top surface of crystal was designed as diffuse reflecting material with a reflection coefficient (RC) of 0.5 and the position misclassification rate was calculated as 8% for the DOI direction. However, in the experiment, it is very hard to accurately wrap the crystal with specific RC value. Also, the wrapping material such as Teflon tape may introduce irregular dead spaces between crystals. Therefore, the optimization of the surface treatment of quasimonolithic crystal, a reflector material and an accurate surface treatment technique are required for further study to improve depth resolution. In this paper, PSPMT was considered as a photo-sensor in PET detector, but it can be replaced by an array of avalanche photodiodes (APDs) and SiPMs [8 9] in the future development due to their high quantum efficiency and positioning accuracy. In conclusion, using the quasi-monolithic detector, a small animal PET system with high sensitivity and uniform radial resolution might be realized. Acknowledgments This work was supported by a Korea Science and Engineering Foundation (KOSEF) grant funded by the Korea Government Ministry of Education, Science and Technology (MEST) ( ).

5 594 Y. Hyun Chung et al. / Nuclear Instruments and Methods in Physics Research A 621 (2010) References [1] A. Braem, et al., Nucl. Instr. and Meth. A 571 (2007) 419. [2] Y. Choi, et al., Nucl. Instr. and Meth. A 571 (2007) 669. [3] T. Hasegawa, et al., Phys. Med. Biol. 52 (2007) 213. [4] Y.H. Chung, et al., Phys. Med. Biol. 49 (2004) [5] A. Braem, et al., Nucl. Instr. and Meth. A 580 (2007) [6] Y. Shao, et al., Nucl. Instr. and Meth. A 580 (2007) 944. [7] Y. Yang, et al., Phys. Med. Biol. 51 (2006) [8] D. McElroy, et al., Nucl. Instr. and Meth. A 571 (2007) 106. [9] D.J. Herbert, et al., Nucl. Instr. and Meth. A 573 (2007) 84. [10] Y.H. Chung, et al., Nucl. Instr. and Meth. A 593 (2008) 588. [11] Y.H. Chung, et al., JKPS 54 (2009) 244. [12] S.-J. Lee, et al., JKPS 54 (2009) 2429.

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 602 (2009) 477 483 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997

218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997 218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997 Compton Scatter and X-ray Crosstalk and the Use of Very Thin Intercrystal Septa in High-Resolution PET Detectors Craig S. Levin, Member,

More information

An educational tool for demonstrating the TOF-PET technique

An educational tool for demonstrating the TOF-PET technique Nuclear Instruments and Methods in Physics Research A 471 (2001) 200 204 An educational tool for demonstrating the TOF-PET technique T.Bȧack a, *, J. Cederkȧall a, B. Cederwall a, A. Johnson a, A. Kerek

More information

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 3, JUNE 2003 405 Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging Edward J. Soares, Kevin W. Germino,

More information

A region segmentation based algorithm for building crystal position lookup table in scintillation detector

A region segmentation based algorithm for building crystal position lookup table in scintillation detector A region segmentation based algorithm for building crystal position lookup table in scintillation detector WANG Hai-Peng ( 王海鹏 ) 1,2,3 YUN Ming-Kai ( 贠明凯 ) 1,2 LIU Shuang-Quan ( 刘双全 ) 1,2 FAN Xin ( 樊馨

More information

Medical Imaging BMEN Spring 2016

Medical Imaging BMEN Spring 2016 Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,

More information

Geant4 Studies for the HPD-PET scintillators

Geant4 Studies for the HPD-PET scintillators F. Ciocia, a A. Braem, b E. Chesi, b, a1 C. Joram, b L. Lagamba, a E. Nappi, a J. Séguinot, b I. Vilardi a and P. Weilhammer b a Physics Dept. and INFN Section of Bari, Via Orabona 4, Bari, Italy b CERN

More information

Constructing System Matrices for SPECT Simulations and Reconstructions

Constructing System Matrices for SPECT Simulations and Reconstructions Constructing System Matrices for SPECT Simulations and Reconstructions Nirantha Balagopal April 28th, 2017 M.S. Report The University of Arizona College of Optical Sciences 1 Acknowledgement I would like

More information

SNIC Symposium, Stanford, California April The Hybrid Parallel Plates Gas Counter for Medical Imaging

SNIC Symposium, Stanford, California April The Hybrid Parallel Plates Gas Counter for Medical Imaging The Hybrid Parallel Plates Gas Counter for Medical Imaging F. Anulli, G. Bencivenni, C. D Ambrosio, D. Domenici, G. Felici, F. Murtas Laboratori Nazionali di Frascati - INFN, Via E. Fermi 40, I-00044 Frascati,

More information

Introduction to Positron Emission Tomography

Introduction to Positron Emission Tomography Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator

More information

Geant4 Studies of the scintillator crystals for the axial HPD-PET concept

Geant4 Studies of the scintillator crystals for the axial HPD-PET concept > FOR CONFERENCE-RELATED PAPERS, REPLACE THIS LINE WITH YOUR SESSION NUMBER, E.G., AB- (DOUBLE-CLICK HERE) < Geant4 Studies of the scintillator crystals for the axial HPD-PET concept F. Ciocia a, A. Braem

More information

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and

More information

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Artifact Mitigation in High Energy CT via Monte Carlo Simulation PIERS ONLINE, VOL. 7, NO. 8, 11 791 Artifact Mitigation in High Energy CT via Monte Carlo Simulation Xuemin Jin and Robert Y. Levine Spectral Sciences, Inc., USA Abstract The high energy (< 15 MeV) incident

More information

arxiv:physics/ v2 [physics.med-ph] 23 Nov 2005

arxiv:physics/ v2 [physics.med-ph] 23 Nov 2005 arxiv:physics/0511065v2 [physics.med-ph] 23 Nov 2005 Experimental study of a liquid Xenon PET prototype module M.-L. Gallin-Martel 1, P. Martin, F. Mayet, J. Ballon, G. Barbier, C. Barnoux, J. Berger,

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

CHAPTER 11 NUCLEAR MEDICINE IMAGING DEVICES

CHAPTER 11 NUCLEAR MEDICINE IMAGING DEVICES CHAPTER 11 M.A. LODGE, E.C. FREY Russell H. Morgan Department of Radiology and Radiological Sciences, Johns Hopkins University, Baltimore, Maryland, United States of America 11.1. INTRODUCTION Imaging

More information

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors MS116 Characterization of light production, propagation and collection for both organic and inorganic scintillators D10.2 R&D on new and existing scintillation materials: Report on the light production,

More information

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar Doctoral Thesis Time-of-Flight Time-of-Flight Positron Positron Emission Emission Tomography Tomography Using Using Cherenkov Cherenkov Radiation Radiation Rok Dolenec Supervisor: Prof. Dr. Samo Korpar

More information

Effects of system geometry and other physical factors on photon sensitivity of high-resolution

Effects of system geometry and other physical factors on photon sensitivity of high-resolution Home Search Collections Journals About Contact us My IOPscience Effects of system geometry and other physical factors on photon sensitivity of high-resolution positron emission tomography This article

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,500 108,000 1.7 M Open access books available International authors and editors Downloads Our

More information

Development of a high resolution animal PET with continuous crystals and SiPMs

Development of a high resolution animal PET with continuous crystals and SiPMs Development of a high resolution animal PET with continuous crystals and SiPMs 1, John Barrio1, Jorge Cabello1,2, Ane Etxebeste1, Carlos Lacasta1, Josep F. Oliver1, Magdalena Rafecas1, Carles Solaz1, Vera

More information

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial 8//0 AAPM0 Scientific Symposium: Emerging and New Generation PET: Instrumentation, Technology, Characteristics and Clinical Practice Aug Wednesday 0:4am :pm Solid State Digital Photon Counting PET/CT Instrumentation

More information

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology.

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology. Photon Counting Detectors and Their Applications in Medical Imaging Principles, Pitfalls and Progress in Photon-Counting-Detector Technology Taly Gilat Schmidt, PhD Associate Professor Department of Biomedical

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

SUBMITTED 15/08/2018 TO: IEEE TRPMS - SPEICIAL ISSUE PSMR2018-7TH CONFERENCE ON PET/MR AND SPECT/MR 1

SUBMITTED 15/08/2018 TO: IEEE TRPMS - SPEICIAL ISSUE PSMR2018-7TH CONFERENCE ON PET/MR AND SPECT/MR 1 SUBMITTED 15/08/2018 TO: IEEE TRPMS - SPEICIAL ISSUE PSMR2018-7TH CONFERENCE ON PET/MR AND SPECT/MR 1 A novel DOI Positioning Algorithm for Monolithic Scintillator Crystals in PET based on Gradient Tree

More information

In-vivo dose verification for particle therapy

In-vivo dose verification for particle therapy In-vivo dose verification for particle therapy D.R. Schaart, NCS Lustrum, 5-Oct-2012 HollandPTC 1 Protons vs. photons Photons Protons Dennis R. Schaart Delft University of Technology 2 Protons: the promise

More information

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa 18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa TOWARDS ESTABLISHMENT OF STANDARDIZED PRACTICE FOR ASSESSMENT OF SPATIAL RESOLUTION AND CONTRAST OF INTERNATIONAL

More information

Monte-Carlo simulation for scatter correction compensation studies in SPECT imaging using GATE software package

Monte-Carlo simulation for scatter correction compensation studies in SPECT imaging using GATE software package Nuclear Instruments and Methods in Physics Research A 569 (2006) 404 408 www.elsevier.com/locate/nima Monte-Carlo simulation for scatter correction compensation studies in SPECT imaging using GATE software

More information

Optical Modeling of Scintillation Detectors Using GATE

Optical Modeling of Scintillation Detectors Using GATE GATE Simulations of Preclinical and Clinical Scans in Emission Tomography, Transmission Tomography and Radiation Therapy Optical Modeling of Scintillation Detectors Using GATE Emilie Roncali Department

More information

Modeling and Incorporation of System Response Functions in 3D Whole Body PET

Modeling and Incorporation of System Response Functions in 3D Whole Body PET Modeling and Incorporation of System Response Functions in 3D Whole Body PET Adam M. Alessio, Member IEEE, Paul E. Kinahan, Senior Member IEEE, and Thomas K. Lewellen, Senior Member IEEE University of

More information

FOOT de/dx detector: simulation report

FOOT de/dx detector: simulation report FOOT de/dx detector: simulation report Matteo Bertazzoni, Maria Giuseppina Bisogni, Esther Ciarrocchi e Matteo Morrocchi February 16, 2017 1 Introduction The goals of the de/dx detector group are to determine

More information

A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations

A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations V. Rosso, G.Battistoni, N. Belcari, N. Camarlinghi, A. Ferrari, S. Ferretti, A. Kraan, A. Mairani, N. Marino, J. E.

More information

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Effects of the difference in tube voltage of the CT scanner on. dose calculation Effects of the difference in tube voltage of the CT scanner on dose calculation Dong Joo Rhee, Sung-woo Kim, Dong Hyeok Jeong Medical and Radiological Physics Laboratory, Dongnam Institute of Radiological

More information

Proton dose calculation algorithms and configuration data

Proton dose calculation algorithms and configuration data Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam

More information

THE EFFECT OF SURFACE ROUGHNESS OF CSI(TL) MICRO-COLUMNS ON THE RESOLUTION OF THE X-RAY IMAGE; OPTICAL SIMULATION STUDY

THE EFFECT OF SURFACE ROUGHNESS OF CSI(TL) MICRO-COLUMNS ON THE RESOLUTION OF THE X-RAY IMAGE; OPTICAL SIMULATION STUDY PAPER THE EFFECT OF SURFACE ROUGHNESS OF CSI(TL) MICRO-COLUMNS ON THE RESOLUTION OF THE X-RAY IMAGE; OPTICAL SIMULATION STUDY HYUNKI KIM, JUN HYUNG BAE, BO KYUNG CHA, HOSANG JEON, JONG YUL KIM, CHAN KYU

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Monte-Carlo Modeling of Scintillator Crystal Performance for Stratified PET Detectors with DETECT2000

Monte-Carlo Modeling of Scintillator Crystal Performance for Stratified PET Detectors with DETECT2000 Monte-Carlo Modeling of Scintillator Crystal Performance for Stratified PET Detectors with DETECT2 Francois Cayouette, Member, IEEE, Christian Moisan, Nan Zhang, Member, IEEE and Christopher J. Thompson,

More information

Continuation Format Page

Continuation Format Page C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar

More information

Large Plastic Scintillation Detectors for the Nuclear Materials Identification System

Large Plastic Scintillation Detectors for the Nuclear Materials Identification System Large Plastic Scintillation Detectors for the Nuclear Materials Identification System J.S. Neal, J.T. Mihalczo, M. T. Hiatt, J. D. Edwards Oak Ridge National Laboratory P. O. Box 2008, Oak Ridge, Tennessee

More information

Fast Timing and TOF in PET Medical Imaging

Fast Timing and TOF in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.

More information

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO Abstract C.S. Levin, Y-C Tai, E.J. Hoffman, M. Dahlbom, T.H. Farquhar UCLA School of Medicine Division

More information

Limitations in the PHOTON Monte Carlo gamma transport code

Limitations in the PHOTON Monte Carlo gamma transport code Nuclear Instruments and Methods in Physics Research A 480 (2002) 729 733 Limitations in the PHOTON Monte Carlo gamma transport code I. Orion a, L. Wielopolski b, * a St. Luke s/roosevelt Hospital, Columbia

More information

SINGLE-PHOTON emission computed tomography

SINGLE-PHOTON emission computed tomography 1458 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 4, AUGUST 2012 SPECT Imaging With Resolution Recovery Andrei V. Bronnikov Abstract Single-photon emission computed tomography (SPECT) is a method

More information

Design and simulation of continuous scintillator with pixellated photodetector

Design and simulation of continuous scintillator with pixellated photodetector University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 2001 Design and simulation of continuous scintillator with pixellated

More information

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Sensors & Transducers 2013 by IFSA http://www.sensorsportal.com Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Shinji Ohyama, Masato Mukouyama Graduate School

More information

arxiv:physics/ v1 [physics.ins-det] 18 Dec 1998

arxiv:physics/ v1 [physics.ins-det] 18 Dec 1998 Studies of 1 µm-thick silicon strip detector with analog VLSI readout arxiv:physics/981234v1 [physics.ins-det] 18 Dec 1998 T. Hotta a,1, M. Fujiwara a, T. Kinashi b, Y. Kuno c, M. Kuss a,2, T. Matsumura

More information

DETECT2000: An Improved Monte-Carlo Simulator for the Computer Aided Design of Photon Sensing Devices

DETECT2000: An Improved Monte-Carlo Simulator for the Computer Aided Design of Photon Sensing Devices DETECT2000: An Improved Monte-Carlo Simulator for the Computer Aided Design of Photon Sensing Devices François Cayouette 1,2, Denis Laurendeau 3 and Christian Moisan 3 1 Montreal Neurological Institute,

More information

Application of MCNP Code in Shielding Design for Radioactive Sources

Application of MCNP Code in Shielding Design for Radioactive Sources Application of MCNP Code in Shielding Design for Radioactive Sources Ibrahim A. Alrammah Abstract This paper presents three tasks: Task 1 explores: the detected number of as a function of polythene moderator

More information

Iterative SPECT reconstruction with 3D detector response

Iterative SPECT reconstruction with 3D detector response Iterative SPECT reconstruction with 3D detector response Jeffrey A. Fessler and Anastasia Yendiki COMMUNICATIONS & SIGNAL PROCESSING LABORATORY Department of Electrical Engineering and Computer Science

More information

Interpolating Silicon Photomultipliers

Interpolating Silicon Photomultipliers Interpolating Silicon Photomultipliers Peter Fischer, Heidelberg University, Germany (Presenter) Claudio Piemonte, FBK, Italy We present the novel Interpolating Silicon PhotoMultiplier (ISiPM) topology

More information

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY BACKSCATTER DATA Young S. Ham and C.F. Poranski Materials Chemistry Branch Naval Research Laboratory Washington DC 20375 E.C. Greenawald Geo-Centers,

More information

Forward Time-of-Flight Geometry for CLAS12

Forward Time-of-Flight Geometry for CLAS12 Forward Time-of-Flight Geometry for CLAS12 D.S. Carman, Jefferson Laboratory ftof geom.tex April 13, 2016 Abstract This document details the nominal geometry for the CLAS12 Forward Time-of- Flight System

More information

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 5 SPECT, PET Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce

More information

Supplementary Information

Supplementary Information Supplementary Information Interferometric scattering microscopy with polarization-selective dual detection scheme: Capturing the orientational information of anisotropic nanometric objects Il-Buem Lee

More information

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran. DXRaySMCS First User Friendly Interface Developed for Prediction of Diagnostic Radiology X-Ray Spectra Produced by Monte Carlo (MCNP-4C) Simulation in Iran M.T. Bahreyni Toosi a*, H. Moradi b, H. Zare

More information

A New Model for Optical Crosstalk in SinglePhoton Avalanche Diodes Arrays

A New Model for Optical Crosstalk in SinglePhoton Avalanche Diodes Arrays A New Model for Optical Crosstalk in SinglePhoton Avalanche Diodes Arrays I. Rech, A. Ingargiola, R. Spinelli, S. Marangoni, I. Labanca, M. Ghioni, S. Cova Dipartimento di Elettronica ed Informazione Politecnico

More information

CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona

CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona Introduction CBCT has become a routine procedure for

More information

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide 2005 IEEE Nuclear Science Symposium Conference Record M04-8 Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide J. S. Karp, Senior Member, IEEE, A. Kuhn, Member, IEEE, A. E. Perkins,

More information

Improved Modeling of System Response in List Mode EM Reconstruction of Compton Scatter Camera Images'

Improved Modeling of System Response in List Mode EM Reconstruction of Compton Scatter Camera Images' IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 48, NO. 1, FEBRUARY 2001 111 Improved Modeling of System Response in List Mode EM Reconstruction of Compton Scatter Camera Images' S.J. Wilderman2, J.A. Fe~sle?t~>~,

More information

Crystal Identification in Dual-Layer-Offset DOI-PET Detectors Using Stratified Peak Tracking Based on SVD and Mean-Shift Algorithm

Crystal Identification in Dual-Layer-Offset DOI-PET Detectors Using Stratified Peak Tracking Based on SVD and Mean-Shift Algorithm 2502 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 63, NO. 5, OCTOBER 2016 Crystal Identification in Dual-Layer-Offset DOI-PET Detectors Using Stratified Peak Tracking Based on SVD and Mean-Shift Algorithm

More information

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Krista Chytyk MSc student Supervisor: Dr. Boyd McCurdy Introduction The objective of cancer radiotherapy

More information

AX-PET : A novel PET concept with G-APD readout

AX-PET : A novel PET concept with G-APD readout AX-PET : A novel PET concept with G-APD readout Matthieu Heller CERN - PH/DT Marie Curie network MC-PAD Matthieu.heller@cern.ch On behalf of the AX-PET collaboration https://twiki.cern.ch/twiki/bin/view/axialpet

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

A Radiometry Tolerant Method for Direct 3D/2D Registration of Computed Tomography Data to X-ray Images

A Radiometry Tolerant Method for Direct 3D/2D Registration of Computed Tomography Data to X-ray Images A Radiometry Tolerant Method for Direct 3D/2D Registration of Computed Tomography Data to X-ray Images Transfer Function Independent Registration Boris Peter Selby 1, Georgios Sakas 2, Stefan Walter 1,

More information

Performance Evaluation of 3-Axis Scanner Automated For Industrial Gamma- Ray Computed Tomography

Performance Evaluation of 3-Axis Scanner Automated For Industrial Gamma- Ray Computed Tomography National Seminar & Exhibition on Non-Destructive Evaluation, NDE 2014, Pune, December 4-6, 2014 (NDE-India 2014) Vol.20 No.6 (June 2015) - The e-journal of Nondestructive Testing - ISSN 1435-4934 www.ndt.net/?id=17822

More information

Project GAMMA: Adaptive Methods for Medical Applications

Project GAMMA: Adaptive Methods for Medical Applications Project GAMMA: Adaptive Methods for Medical Applications V. Solovov, A. Morozov, V. Chepel, V. Domingos, R.Martins, L. Pereira, F. Neves, A. Lindote, F. Fraga, I.Lopes, F. Alves LIP-Coimbra and Coimbra

More information

Estimating 3D Respiratory Motion from Orbiting Views

Estimating 3D Respiratory Motion from Orbiting Views Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing

More information

Reconstruction from Projections

Reconstruction from Projections Reconstruction from Projections M.C. Villa Uriol Computational Imaging Lab email: cruz.villa@upf.edu web: http://www.cilab.upf.edu Based on SPECT reconstruction Martin Šámal Charles University Prague,

More information

Department of Game Mobile Contents, Keimyung University, Daemyung3-Dong Nam-Gu, Daegu , Korea

Department of Game Mobile Contents, Keimyung University, Daemyung3-Dong Nam-Gu, Daegu , Korea Image quality enhancement of computational integral imaging reconstruction for partially occluded objects using binary weighting mask on occlusion areas Joon-Jae Lee, 1 Byung-Gook Lee, 2 and Hoon Yoo 3,

More information

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation 26 IEEE Nuclear Science Symposium Conference Record M6-349 Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation Habib Zaidi, Senior Member, IEEE and Mohammad

More information

Rectification of distorted elemental image array using four markers in three-dimensional integral imaging

Rectification of distorted elemental image array using four markers in three-dimensional integral imaging Rectification of distorted elemental image array using four markers in three-dimensional integral imaging Hyeonah Jeong 1 and Hoon Yoo 2 * 1 Department of Computer Science, SangMyung University, Korea.

More information

SPECT reconstruction

SPECT reconstruction Regional Training Workshop Advanced Image Processing of SPECT Studies Tygerberg Hospital, 19-23 April 2004 SPECT reconstruction Martin Šámal Charles University Prague, Czech Republic samal@cesnet.cz Tomography

More information

The optimum design of a moving PM-type linear motor for resonance operating refrigerant compressor

The optimum design of a moving PM-type linear motor for resonance operating refrigerant compressor International Journal of Applied Electromagnetics and Mechanics 33 (2010) 673 680 673 DOI 10.3233/JAE-2010-1172 IOS Press The optimum design of a moving PM-type linear motor for resonance operating refrigerant

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

Technical Basis for optical experimentation Part #4

Technical Basis for optical experimentation Part #4 AerE 545 class notes #11 Technical Basis for optical experimentation Part #4 Hui Hu Department of Aerospace Engineering, Iowa State University Ames, Iowa 50011, U.S.A Light sensing and recording Lenses

More information

Minimizing Thin Glass Deflection in Flexible Display Manufacturing via Pin Map Optimization

Minimizing Thin Glass Deflection in Flexible Display Manufacturing via Pin Map Optimization Minimizing Thin Glass Deflection in Flexible Display Manufacturing via Pin Map Optimization Nam Hoon Jo BS. Scholar, Department of Mechanical Engineering, Graduate School Kongju National University (KNU),

More information

Performance Evaluation of the Philips Gemini PET/CT System

Performance Evaluation of the Philips Gemini PET/CT System Performance Evaluation of the Philips Gemini PET/CT System Rebecca Gregory, Mike Partridge, Maggie A. Flower Joint Department of Physics, Institute of Cancer Research, Royal Marsden HS Foundation Trust,

More information

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION 2009 IEEE Nuclear Science Symposium Conference Record M03-7 FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION Wonseok Huh, Jeffrey A. Fessler, Adam M. Alessio, and Paul E. Kinahan Department

More information

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy.

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy. Modeling the Distance-Dependent Blurring in Transmission Imaging in the Ordered-Subset Transmission (OSTR) Algorithm by Using an Unmatched Projector/Backprojector Pair B. Feng, Member, IEEE, M. A. King,

More information

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 5, OCTOBER

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 5, OCTOBER IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 5, OCTOBER 2006 2549 Positioning Annihilation Photon Interactions in a Thin LSO Crystal Sheet With a Position-Sensitive Avalanche Photodiode Angela M.

More information

Optimization of scanner parameters for dual energy micro-ct

Optimization of scanner parameters for dual energy micro-ct Optimization of scanner parameters for dual energy micro-ct E. PAUWELS* 1, J. DHAENE 1, A. DE MUYNCK 1 E., M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics and Astronomy, Ghent University, Proeftuinstraat

More information

664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005

664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005 664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005 Attenuation Correction for the NIH ATLAS Small Animal PET Scanner Rutao Yao, Member, IEEE, Jürgen Seidel, Jeih-San Liow, Member, IEEE,

More information

Luminous. Optoelectronic Device Simulator 4/15/05

Luminous. Optoelectronic Device Simulator 4/15/05 Optoelectronic Device Simulator 4/15/05 Contents Overview Key Benefits Applications Charge Coupled Devices (CCDs) Separate Absorption Multiplication (SAM) reach through avalanche photo detectors High speed

More information

UNIVERSITY OF SOUTHAMPTON

UNIVERSITY OF SOUTHAMPTON UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions

More information

Introduction to Emission Tomography

Introduction to Emission Tomography Introduction to Emission Tomography Gamma Camera Planar Imaging Robert Miyaoka, PhD University of Washington Department of Radiology rmiyaoka@u.washington.edu Gamma Camera: - collimator - detector (crystal

More information

Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector

Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector K. Karafasoulis, K. Zachariadou, S. Seferlis, I. Papadakis, D. Loukas, C. Lambropoulos, C. Potiriadis Abstract Spectroscopic

More information

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry Sasha A. Philips, Frazier Bronson, Ram Venkataraman, Brian M. Young Abstract--Activity measurements require knowledge of the

More information

Single Polarizer Liquid Crystal Display Mode with Fast Response

Single Polarizer Liquid Crystal Display Mode with Fast Response Mol. Cryst. Liq. Cryst., Vol. 543: pp. 101=[867] 106=[872], 2011 Copyright # Taylor & Francis Group, LLC ISSN: 1542-1406 print=1563-5287 online DOI: 10.1080/15421406.2011.568342 Single Polarizer Liquid

More information

Validation of the Small Animal Biospace Gamma Imager Model Using GATE Monte Carlo Simulations on the Grid

Validation of the Small Animal Biospace Gamma Imager Model Using GATE Monte Carlo Simulations on the Grid Validation of the Small Animal Biospace Gamma Imager Model Using GATE Monte Carlo Simulations on the Grid Joe Aoun, Vincent Breton, L. Desbat, B. Bzeznik, M. Leabadand, J. Dimastromatteo To cite this version:

More information

486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007

486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007 486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007 Count-Rate Dependent Component-Based Normalization for the HRRT Mario Rodriguez, Jeih-San Liow, Member, IEEE, Shanthalaxmi Thada, Merence

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER

COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER By RAMI RIMON ABU-AITA A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig

More information

3D-OSEM Transition Matrix for High Resolution PET Imaging with Modeling of the Gamma-Event Detection

3D-OSEM Transition Matrix for High Resolution PET Imaging with Modeling of the Gamma-Event Detection 3D-OSEM Transition Matrix for High Resolution PET Imaging with Modeling of the Gamma-Event Detection Juan E. Ortuño, George Kontaxakis, Member, IEEE, Pedro Guerra, Student Member, IEEE, and Andrés Santos,

More information

in PET Medical Imaging

in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives 1. Be familiar with recommendations

More information

New Edge-Enhanced Error Diffusion Algorithm Based on the Error Sum Criterion

New Edge-Enhanced Error Diffusion Algorithm Based on the Error Sum Criterion New Edge-Enhanced Error Diffusion Algorithm Based on the Error Sum Criterion Jae Ho Kim* Tae Il Chung Hyung Soon Kim* Kyung Sik Son* Pusan National University Image and Communication Laboratory San 3,

More information